SOP 3-Marken-Kolben Dispensette
SOP 3-Marken-Kolben Dispensette
Reinigung:
Den Messkolben vollständig mit 3%iger Mucasol®-Lösung (ca 2,5 ml Mucasol® pro 100 ml Wasser) füllen und 1 Tag einwirken
lassen. Dann zehnmal mit Leitungswasser und fünfmal mit E-Wasser kräftig ausspülen.
Beim letzten Ausspülen mit E-Wasser den Messkolben in der senkrechten Position halten und kontrollieren, ob das Wasser
tropfenfrei abläuft.
Nur wenn das ablaufende Wasser auf der Innenoberfläche des Halses keine Tropfen hinterlässt, ist der Messkolben sauber.
Bleiben einzelne Tropfen haften, muss die Reinigung wiederholt werden.
Trocknen:
Den Messkolben in ein Trockengestell kopfüber einhängen und über Nacht trocknen lassen.
Cleaning:
Fill the volumetric flask completely with a 3% Mucasol® solution (ca. 2.5 ml Mucasol® per 100 ml water) and let it soak for 1 day
before rinsing it vigorously 10 times with tap water and 5 times with deionized water.
On the last rinse with deionized water, hold the volumetric flask vertically and check whether the water is draining off without
leaving drops at the wall.
The volumetric flask is clean when the water leaves no drops on the inside surface of the flask neck when it runs out.
The cleaning procedure must be repeated if any drops of water remain on the wall.
Drying:
Place the volumetric flask upside-down in a drying rack and let it dry overnight.
In case of error:
Repeat the experiment if the meniscus is located outside of the error limits. If, after the second attempt, the meniscus still is
located outside of the error limits it could be either caused by a faulty dispenser or by improper operation. We recommend look-
ing for the cause of the error by following the troubleshooting procedure described in the operating manual and, if necessary,
cleaning the bottle-top dispenser carefully. Please subsequently perform a gravimetric volume test as described in the SOP.
If this gravimetric volume test also returns results outside of the error limits, please contact the manufacturer.
The dispenser may be faulty and may need to be sent for repair.