Kaye CTR 40 Users Guide Rev 371501
Kaye CTR 40 Users Guide Rev 371501
World Headquarters GE Kaye Instruments, Inc. 101 Billerica Avenue, Building #7 North Billerica, MA 01862 USA tel 978-262-0005, 800-964-5293 (US & Canada) fax 978-439-8181 email [email protected] European Headquaters GE Kaye Europe Sinsheimer Strasse 6 D-75179 Pforzheim Germany tel +49 (0) 7231 14335 0 fax +49 (0) 7231 14335 29 email [email protected] www.gekaye.com
Subject to change without notice. Copyright 2002 Printed in USA
Rev. 371501
Table of Contents
1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 3
Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 4.2 4.3 4.4 Unpacking. . . . . . . . Set Up . . . . . . . . . . Power . . . . . . . . . . Setting the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 . 13 . 14 . 14
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Bath Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 5.2 Dry-out Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 5.3 Bath Preparation and Filling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.3.1 Drain Valve Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 5.3.2 Filling With Fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.4 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
General Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.1 Heat Transfer Fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.1.1 Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.2 Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.3 Specific Heat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.4 Thermal Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.5 Thermal Expansion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.6 Electrical Resistivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.7 Fluid Lifetime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.8 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.9 Cost . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.10 Commonly Used Fluids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.10.1 Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.10.2 Ethylene Glycol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.10.3 Mineral Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.10.4 Silicone Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.11 Fluid Characteristics Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.11.1 Limitations and Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.11.2 About the Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 23 24 24 24 25 25 25 26 26 26 26 27 27 27 27 28
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. 30 . 30 . 31 . 31 . 32
Controller Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.1 Bath Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 8.2 Reset Cutout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 8.3 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.3.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 8.3.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 8.3.3 Set-point Vernier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.4 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
8.4.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 8.4.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
ii
Secondary Menu . . . . Heater Power . . . . . . Proportional Band . . . Cutout . . . . . . . . . . Controller Configuration Probe Parameters. . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. 43 . 43 . 43 . 46 . 47 . 47
10 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.1 10.2 10.3 10.4 Calibration Points . . . . . . . Measuring the Set-point Error Computing R0 and ALPHA . . Calibration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 . 63 . 63 . 65
11 Charging Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.1 Leak Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
iii
12 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 13 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
13.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 13.2 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
13.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 13.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
iv
Figures
Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Drain Valve Installation IMPORTANT: Do Not Over Tighten. . . . . . . . . . . . . . . . . . 18 Tank Baffle Showing Minimum and Maximum Fill Levels . . . . . . . . . . . . . . . . . . . . 19 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Chart of Various Bath Fluids and Their Properties . . . . . . . . . . . . . . . . . . . . . . . . 29 Controller Operation Flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Bath Temperature Fluctuation At Various Proportional Band Settings . . . . . . . . . . . . . . 44 Serial Communications Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Calibration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Wiring Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Tables
Table 1 Table 2 Table 3 Table 4 Table 5 Table 5 Table 5 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Table of Various Bath Fluids and Their Properties . . . . . . . . . . . . . . . . . . . . . . . . 28 Program Mode Setting Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Typical Proportional Band Settings for Various Fluids . . . . . . . . . . . . . . . . . . . . . . 45 Interface Command Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Interface Command Summary Continued. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Interface Command Summary Continued. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
vi
1
1.1
Symbol
Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances. C-TIC Australian EMC mark
CAT
1.2
Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information below. The following definitions apply to the terms Warning and Caution. Warning identifies conditions and actions that may pose hazards to the user. Caution identifies conditions and actions that may damage the instrument being used.
1.2.1
Warnings
To avoid personal injury, follow these guidelines. GENERAL DO NOT use the instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the unit may cause unknown hazards to the user. DO NOT use the unit in environments other than those listed in the users guide. DO NOT overfill the bath. Overflowing extremely cold or hot fluid may be harmful to the operator. See Section 5.3, Bath Preparation and Filling, for specific instructions. Follow all safety guidelines listed in the users manual. Calibration Equipment should only be used by Trained Personnel.
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Before initial use, or after transport, or after storage in humid or semi-humid environments, or anytime the instrument has not been energized for more than 10 days, the instrument needs to be energized for a "dry-out" period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1. If the product is wet or has been in a wet environment, take necessary measures to remove moisture prior to applying power such as storage in a low humidity temperature chamber operating at 50 degree centigrade for 4 hours or more. DO NOT operate high temperature baths (500C) near flammable materials. Extreme temperatures could ignite the flammable material. Overhead clearance is required. Do not place the instrument under a cabinet or other structure. Always leave enough clearance to allow for safe and easy insertion and removal of probes. The instrument is intended for indoor use only. BURN HAZARD Extremely cold temperatures may be present in this equipment. Freezer burns and frostbite may result if personnel fail to observe safety precautions. High temperatures may be present in this equipment. Fires and severe burns may result if personnel fail to observe safety precautions. ELECTRICAL HAZARD These guidelines must be followed to ensure that the safety mechanisms in this instrument will operate properly. This instrument must be plugged into a 115 VAC, 60Hz (230 VAC, 50Hz optional), AC only electric outlet. The power cord of the instrument is equipped with a three-pronged grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded three-prong receptacle. The receptacle must be installed in accordance with local codes and ordinances. Consult a qualified electrician. DO NOT use an extension cord or adapter plug. DO use a ground fault interrupt device. This unit contains a liquid. A ground fault device is advised in case liquid is present in the electrical system and could cause an electrical shock. Always replace the power cord with an approved cord of the correct rating and type. If you have questions, contact Kaye Customer Service.
3
High voltage is used in the operation of this equipment. Severe injury or death may result if personnel fail to observe the safety precautions. Before working inside the equipment, turn off the power and disconnect the power cord. BATH FLUIDS Fluids used in this unit may produce noxious or toxic fumes under certain circumstances. Consult the fluid manufacturers MSDS (Material Safety Data Sheet). Proper ventilation and safety precautions must be observed. The instrument is equipped with a soft cutout (user settable firmware) and a hard cutout (set at the factory). Check the flash point, boiling point, or other fluid characteristic applicable to the circumstances of the unit operation. Ensure that the soft cutout is adjusted to the fluid characteristics of the application. As a guideline, the soft cutout should be set 10C to 15C below the flash point of the bath fluid. See Section 7.1, Heat Transfer Fluid, for specific information on bath fluids and Section 8.10, Cutout.
1.2.2
Cautions
THE DRAIN VALVE MUST BE INSTALLED ON THE BACK OF THE BATH BEFORE ATTEMPTING TO FILL THE TANK WITH FLUID. See Section 5.3, page 17 for drain installation instructions. Always operate this instrument at room temperature between 41F and 122F (5C to 50C). Allow sufficient air circulation by leaving at least 6 inches (15 cm) of clearance around the instrument. DO NOT overfill the bath. Overflowing liquid may damage the electrical system. Be sure to allow for thermal expansion of the fluid as the bath temperature increases. See Section 5.3, Bath Preparation and Filling, for specific instructions. DO NOT change the values of the bath calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the unit. The refrigeration may be damaged or the lifetime shortened if the set-point temperature is set above 60C for more than one hour with the refrigeration manually on. Ensure that the refrigeration is off when the unit is used above 60C. The Factory Reset Sequence should be performed only by authorized personnel if no other action is successful in correcting a malfunction. You must have a copy of the most recent Report of Test to restore the test parameters.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environment. The unit is a precision instrument. Although it has been designed for optimum durability and trouble free operation, it must be handled with care. Position the unit before the tank is filled with fluid. Use the handles provided to move the unit. Due to the weight of the compressor, it may require two people to safely move the bath. If two people are used, place one person in the front and one person in the back of the unit, carefully slide hands under the unit and lift in unison. The area containing the compressor will be heavier than the rest of the unit. Do not move a unit filled with fluid. Most probes have handle temperature limits. Be sure that the probe handle temperature limit is not exceeded in the air above the instrument. The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care. Do not allow them to be dropped, struck, stressed, or overheated. COLD BATHS Refrigerated baths require that the condensing coil be cleaned periodically. Accumulation of dust and dirt on the condenser will result in premature failure of the compressor. This bath has been equipped with a brownout and over voltage protection device as a safety feature to protect the system components. Mode of Operation: This bath needs to be plugged into the line voltage for at least 2 minutes before operation. This is only necessary for the first time that the bath is energized or when it is moved from one location to another. Turning the bath ON or OFF does not trigger the delay. If a High/Low voltage condition exists for longer than 5 seconds, the bath de-energizes. An amber indicator on the back panel lights when this condition exists. Re-energization is automatic upon correction of the fault condition and after a delay cycle of about 2 minutes. If a fault condition exists upon application of power, the bath will not energize. Under and Over Voltage Protection at 115 VAC
Voltage Cutout: 12.5% (101 - 129 VAC) Voltage Cut In: 7.5% (106 - 124 VAC)
1.3
2 Introduction
Introduction
The Kaye Cold Temperature Reference is a bench-top constant temperature bath useful in temperature calibration and other applications requiring stable temperatures. An innovative state of the art solid-state temperature controller has been incorporated which maintains the bath temperature with extreme stability. The temperature controller uses a micro-controller to execute the many operating functions. User interface is provided by the 8-digit LED display and four key-switches. Digital remote communications is standard with an RS-232 interface. The CTR 40 bath was designed to be compact and low cost without compromising performance. The CTR 40 bath operates over a wide temperature range from 40C to 150C. The refrigeration permits sub-ambient temperature control. The CTR 40 features: Rapid heating and cooling RS-232 (standard) Temperature scan rate control Ramp and soak functions Compact size Eight set-point memory Adjustable readout in C or F Automatic refrigeration control
3
3.1
Uniformity
Heating Time Cooling Time Stabilization Time Temperature Setting Set-point Resolution
Display Temperature Resolution 0.01 Digital Setting Accuracy Digital Setting Repeatability Heater Access Opening Depth Wetted Parts Power
0.5C 0.01C 700 Watts @ 115V (230V) Nominal 6.8 x 3.7 (172mm x 94mm) 9.25 (234 mm) 304 Stainless Steel 115 VAC (10%), 60 Hz, 15 amps [230 VAC (10%), 50 Hz, 8 amps optional] Note: If the voltage is outside 10%, the compressor may be damaged. Check the back panel label for the correct voltage and frequency prior to energizing the unit. 2.4 gal. (9.2 liters) 78 lb. (35.4 kg) 12 x 24.5 x 23 (305 mm x 622 mm x 584 mm) OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC 1010-1 R-507 single stage RS-232 included
3.2
Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual. The instrument operates safely under the following conditions: temperature range: 540C (41104F) ambient relative humidity: maximum 80% for temperatures < 31C decreasing linearly to 50% at 40C pressure: 75kPa - 106kPa mains voltage within 10% of nominal vibrations in the calibration environment should be minimized altitudes less than 2000 meters indoor use only
3.3
3.3.1
3.3.2
In-Warranty Repairs
Customers are requested to discuss their problem with a Kaye Service Representative to insure a prompt and accurate assessment of their needs. Frequently, a problem can be resolved via phone or FAX with minimal inconvenience or delay. If necessary, the Customer Service Representative will send replacement parts or authorize the return of the instrument to the factory for repair. Instruments
10
serviced in this manner will be repaired, completely tested, and calibrated prior to shipment. When an instrument is returned to the factory, the customer must prepay the freight charges. Kaye will prepay freight charges for the instrument's return via a comparable shipment method. If Field Service is required under the warranty, the customer is responsible for travel and living expenses incurred by the Field Service Representative.
3.3.3
After-Warranty Repairs
Customers are requested to discuss their problem with a Kaye Service Representative to insure a prompt assessment of their needs. Frequently, a customer installed exchange part will solve the problem with minimal inconvenience and expense. Factory repairs can frequently be completed on a fixed price basis. A base service fee plus labor and materials will be charged in lieu of the fixed repair price upon customer request or if extensive repairs are required. Customers are requested to obtain a return authorization number prior to returning any instrument for service. All instruments serviced at the factory will be repaired, updated, calibrated, and completely tested prior to shipment.
3.3.4
3.3.5
Customer Support
Within 90 days of shipment from the factory, installation and initial configuration assistance will be provided by a Customer Service Representative via phone or FAX at no charge. After 90 days from shipment, technical assistance or consultation will be limited to identification and resolution of instrument failures, unless a Customer Support Agreement has been purchased.
11
3.3.6
3.3.7
12
4 Quick Start
Quick Start
This chapter gives a brief summary of the steps required to set up and operate the bath. This should be used as a general overview and reference and not as a substitute for the remainder of the manual. Please read Section 5, Installation, through Section 7, General Operation, carefully before operating the bath.
4.1
Unpacking
Unpack the bath carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that all components are present: CTR 40 Bath Access Hole Cover Manual RS-232 Cable Report of Test Drain Valve
4.2
Set Up
Caution: The drain valve must be installed on the back of the bath before attempting to fill the tank with fluid. See Section 5.3, page 17 for drain installation instructions.
Warning: The instrument is equipped with a soft cutout (user settable firmware) and a hard cutout (set at the factory). Check the flash point, boiling point, or other fluid characteristic applicable to the circumstances of the unit operation. Ensure that the soft cutout is adjusted to the fluid characteristics of the application. As a guideline, the soft cutout should be set 10C to 15C below the flash point of the bath fluid. See Section 7.1, Heat Transfer Fluid, for specific information on bath fluids and Section 8.10, Cutout. Set up of the bath requires careful unpacking and placement of the bath, installing the drain valve, filling the bath with fluid, and connecting power. Consult Section 5, Installation, for detailed instructions for proper installation of the bath. Install the drain valve onto the pipe fitting at the back of the bath and
13
4 Quick Start
make sure the valve is closed. See Section 5.3, Bath Preparation and Filling. Be sure to place the bath in a safe, clean and level location. Fill the bath tank with an appropriate liquid. Be sure to select the correct fluid for the temperature range of the calibration. Bath fluids should be selected to operate safely with adequate thermal properties to meet the application requirements. For operation at moderate bath temperatures, clean distilled water works well. Carefully pour the fluid into the bath tank through the large rectangular access hole above the tank avoiding spilling any fluid. The fluid must not exceed a height of 12.720.3 mm (0.50.8 inches) below the bottom of the lid (NOT the access cover). Note: As the temperature of the bath increases the fluid level will increase, see Section 7.1.5 on page 24.
4.3
Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency, and current capability. Refer to Section 3.1, Specifications, for power details. Refer to and read the CAUTION at the front of the manual concerning brownout and over voltage protection. Check the back panel label for the correct voltage and frequency prior to energizing the unit. Turn the bath on using the front panel POWER switch. The bath will turn on and begin to heat or cool to reach the previously programmed temperature set-point. The front panel LED display will indicate the actual bath temperature. Set the cooling switch to ON for below or near ambient temperatures. Note: The actual temperature where cooling is required depends on the bath fluid and whether a cover is used over the access well.
4.4
24.68 C
When SET is pressed the display shows the set-point memory that is currently being used and its value. Eight set-point memories are available.
S
14
4 Quick Start
1. 25.0
Press SET to select this memory and access the set-point value.
S U S E
C 25.00
Press UP or DOWN to change the set-point value. Increment display New set-point value
C 30.00
Press SET to accept the new value and display the vernier value. The bath begins heating or cooling to the new set-point. Store new set-point, access vernier Current vernier value
0.00000
Press EXIT and the bath temperature will be displayed again. Return to the temperature display Bath temperature display
24.73 C
The bath heats or cools until it reaches the new set-point temperature. Turn off the cooling to reach and control at higher temperatures. When setting the set-point temperature be careful not to exceed the temperature limit of the bath fluid. The over-temperature cutout should be correctly set for added safety. See Section 8.10, Cutout. To obtain optimum control stability adjust the proportional band as discussed in Section 8.9, Proportional Band.
15
5 Installation
5
5.1
Installation
Bath Environment
The CTR 40 Bath is a precision instrument, which should be located in an appropriate environment. The location should be free of drafts, extreme temperatures and temperature changes, dirt, etc. The surface where the bath is placed must be level. Allow plenty of space around the bath for air circulation. The top surface of the bath may become hot at high temperatures. Beware of the danger of accidental fluid spills. A fume hood should be used to remove any vapors given off by hot bath fluid.
5.2
Dry-out Period
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Before initial use, or after transport, or after storage in humid or semi-humid environments, or anytime the bath has not been energized for more than 10 days, the instrument needs to be energized for a "dry-out" period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1. If the product is wet or has been in a wet environment, take necessary measures to remove moisture prior to applying power such as storage in a low humidity temperature chamber operating at 50 degree centigrade for 4 hour or more.
5.3
5.3.1
2.
The valve should be secured in place at this point. If not, repeat step number 2 being careful not to over tighten the assembly.
17
5 Installation
Ensure the valve handle is in the closed position before attempting to add fluid to the tank.
Valve body
Figure 1 Drain Valve Installation IMPORTANT: Do Not Over Tighten. Follow the installation instructions above.
5.3.2
18
5 Installation
bath fluid level as the bath temperature rises to prevent overflow or splashing. Remove excess hot fluid if necessary with caution.
MAX FILL
MIN FILL
5.4
Power
With the bath power switch off, plug the bath into an AC mains outlet of the appropriate voltage, frequency, and current capacity. Refer to Section 3.1, Specifications, for power details. Refer to and read the CAUTION at the front of the manual concerning brownout and over voltage protection. Check the back panel label for the correct voltage and frequency prior to energizing the unit.
19
6
6.1
CTR 40
1 4
Heater Mode
72.84 C
Set Down Up Exit
Cooling
Power
I
O
I
O
1. The digital display is an important part of the temperature controller. It displays the set-point temperature and bath temperature as well as the various other bath functions, settings, and constants. The display shows temperatures according to the selected scale units C or F. 2. The control buttons (SET, DOWN, UP, and EXIT) are used to set the bath temperature set-point, access and set other operating parameters, and access and set bath calibration parameters. A brief description of the functions of the buttons follows: SET - Used to display the next parameter in a menu and to set parameters to the displayed value. DOWN - Used to decrement the displayed value of parameters. UP - Used to increment the displayed value. EXIT - Used to exit from a menu. When EXIT is pressed any changes made to the displayed value will be ignored. 3. The on/off switch controls power to the entire bath including the stirring motor. 4. The control indicator is a two color light emitting diode (LED). This indicator lets the user visually see the ratio of heating to cooling. When the indicator
21
is red the heater is on, and when it is green the heater is off and the bath is cooling. 5. The cooling switch turns on the refrigeration for control below 50C and rapid cool down. The cooling shuts off automatically above 60C. Manually turn off the refrigeration at temperatures where it has been determined to be unnecessary. See Sections 7.6, Refrigeration, 8.13.2, Cooling Mode, and 8.13.3, Hot Gas Bypass Mode.
6.2
6.3
Back Panel
On the back of the bath are the system fuses, removable power cord, high/low voltage indicator, drain, and serial port.
22
7 General Operation
7
7.1
General Operation
Heat Transfer Fluid
Many fluids will work with CTR 40 bath. Choosing a fluid requires consideration of many important characteristics of the fluid. Among these are temperature range, viscosity, specific heat, thermal conductivity, thermal expansion, electrical resistivity, fluid lifetime, safety, and cost.
7.1.1
Temperature Range
One of the most important characteristics to consider is the temperature range of the fluid. Few fluids work well throughout the entire temperature range of the bath. The temperature at which the bath is operated must always be within the safe and useful temperature range of the fluid used. The lower temperature range of the fluid is determined either by the freeze point of the fluid or the temperature at which the viscosity becomes too great. The upper temperature is usually limited by vaporization, flammability, or chemical breakdown of the fluid. Vaporization of the fluid at higher temperatures may adversely affect temperature stability because of cool condensed fluid dripping into the bath from the lid. The bath temperature should be limited by setting the safety cutout (see Section 8.10, Cutout) or the high limit (see Section 8.15, Calibration Parameters) so that the bath temperature cannot exceed the safe operating temperature limit of the fluid.
7.1.2
Viscosity
Viscosity is a measure of the thickness of a fluid or how easily it can be poured and mixed. Viscosity affects the temperature uniformity and stability of the bath. With lower viscosity fluid mixing is better. This creates a more uniform temperature throughout the bath. This improves the bath response time allowing it to maintain a more constant temperature. For good control the viscosity should be less than 10 centistokes. 50 centistokes is the practical upper limit of allowable viscosity. Viscosity greater than this causes very poor control stability because of poor stirring and may also overheat or damage the stirring motor. Viscosity may vary greatly with temperature, especially with oils. When using fluids with higher viscosities the controller proportional band (see Section 8.9, Proportional Band) may need to be increased to compensate for the reduced response time. Otherwise the temperature may begin to oscillate.
23
7 General Operation
7.1.3
Specific Heat
Specific heat is the measure of the heat storage ability of the fluid. Specific heat, to a small degree, affects the control stability and the heating and cooling rates. Generally, a lower specific heat means quicker heating and cooling. The proportional band may require some adjustment depending on the specific heat of the fluid.
7.1.4
Thermal Conductivity
Thermal conductivity measures how easily heat flows through the fluid. Thermal conductivity of the fluid affects the control stability, temperature uniformity, and temperature settling time. Fluids with higher conductivity distribute heat more quickly and evenly improving bath performance.
7.1.5
Thermal Expansion
Thermal expansion describes how much the volume of the fluid changes with temperature. Thermal expansion of the fluid must be considered since the increase in fluid volume as the bath temperature increases may cause overflow. It may be dangerous to permit the fluid to overflow the tank. It may also cause loss of valuable bath fluid. Excessive thermal expansion may also be undesirable in applications where constant liquid level is important. Thermal expansion coefficients of several fluids are shown in Table 2 on page 28. Fluid manufacturers can also provide this information. The thermal expansion coefficients are shown in units of cm/cm/C. However, the values are the same for any unit of length. Divide the value by 1.8 for F coefficients. The following equation may be used to find the desired depth: DE = DS [K(TETS)+1] Or DS = DE / [K(TETS)+1] where DE The Maximum Fill Depth Where: K=Expansion coefficient TE=Ending temperature TS=Starting temperature DE=Ending depth DS=Starting depth The maximum fill depth is typically 0.5 to 0.8 inches below the level of the gasket at the top of the bath tank (not the top of the bath lid). Judgement must be made with different stirring arrangements to prevent splashing on the gasket or lid of the bath.
24
7 General Operation
Example: The final depth of Dow Corning 710 silicone oil in the bath tank is to be 9.2 inches when heated from 25 to 300C. What should the starting depth be? Expansion coefficient for 710 oil on Table 2, K= 0.00077 inch/inch/C Ending temperature, TE = 300C Starting temperature, TS = 25C Ending depth, DE = 9.2 inches DS = 9.2/[0.00077 (300-25) + 1] = 7.59 inches
7.1.6
Electrical Resistivity
Electrical resistivity describes how well the fluid insulates against the flow of electric current. In some applications, such as measuring the resistance of bare temperature sensors, it may be important that little or no electrical leakage occur through the fluid. In such conditions choose a fluid with very high electrical resistivity.
7.1.7
Fluid Lifetime
Many fluids degrade over time because of vaporization, water absorption, gelling, or chemical break-down. Often the degradation becomes significant near the upper temperature limit of the fluid, substantially reducing the fluids lifetime.
7.1.8
Safety
When choosing a fluid always consider the safety issues associated. Obviously where there are extreme temperatures there can be danger to personnel and equipment. Fluids may also be hazardous for other reasons. Some fluids may be considered toxic. Contact with eyes, skin, or inhalation of vapors may cause injury. A proper fume hood must be used if hazardous or bothersome vapors are produced.
Warning: Fluids at high temperatures may pose danger from BURNS, FIRE, and TOXIC FUMES. Use appropriate caution and safety equipment. Fluids may be flammable and require special fire safety equipment and procedures. An important characteristic of the fluid to consider is the flash point. The flash point is the temperature at which there is sufficient vapor given off so that when there is adequate oxygen present and an ignition source is applied the va25
7 General Operation
por will ignite. This does not necessarily mean that fire will be sustained at the flash point. The flash point may be either of the open cup or closed cup type. Either condition may occur in a bath situation. The open cup flash point is measured under the condition of vapors escaping the tank. The closed cup flash point is measured with the vapors being contained within the tank. Since oxygen and an ignition source is less available inside the tank the closed cup flash point will be lower than the open cup flash point. Environmentally hazardous fluids require special disposal according to applicable federal or local laws after use.
7.1.9
Cost
Cost of bath fluids may vary greatly, from cents per gallon for water to hundreds of dollars per gallon for synthetic oils. Cost may be an important consideration when choosing a fluid.
7.1.10
7.1.10.1
Water
Water is often used because of its very low cost, its availability, and its excellent temperature control characteristics. Water has very low viscosity and good thermal conductivity and heat capacity, which make it among the best fluids for good control stability at lower temperatures. Temperature stability is much poorer at higher temperatures because water condenses on the lid, cools and drips into the bath. Water is safe and relatively inert. The electrical conductivity of water may prevent its use in some applications. Water has a limited temperature range, from a few degrees above 0C to a few degrees below 100C. At higher temperatures evaporation becomes significant. Water used in the bath should be distilled or deionized to prevent mineral deposits. Consider using an algicide chemical in the water to prevent contamination. Note: Water used at temperatures greater than 75C (167F) may have stability problems.
7.1.10.2
Ethylene Glycol
The temperature range of water may be extended by using a solution of one part water and one part ethylene glycol (antifreeze). The characteristics of the ethylene glycol-water solution are similar to water but with higher viscosity. Use caution with ethylene glycol since this fluid is very toxic. Ethylene glycol must be disposed of properly.
26
7 General Operation
7.1.10.3
Mineral Oil
Mineral oil or paraffin oil is often used at moderate temperatures above the range of water. Mineral oil is relatively inexpensive. At lower temperatures mineral oil is quite viscous and control may be poor. At higher temperatures vapor emission becomes significant. The vapors may be dangerous and a fume hood should be used. As with most oils, mineral oil expands as temperature increases. Be careful not to fill the bath too full to avoid overflows when heated. The viscosity and thermal characteristics of mineral oil is poorer than water so temperature stability will not be as good. Mineral oil has very low electrical conductivity. Use caution with mineral oil since it is flammable and may also cause serious injury if inhaled or ingested.
7.1.10.4
Silicone Oil
Silicone oils are available which offer a much wider operating temperature range than mineral oil. Like most oils, silicone oils have temperature control characteristics, which are somewhat poorer than water. The viscosity changes significantly with temperature and thermal expansion also occurs. These oils have very high electrical resistivity. Silicone oils are fairly safe and non-toxic, but can be fairly expensive.
7.1.11
7.1.11.1
27
7 General Operation
100C (v)**
70C (e)
NONE
5.7 @ 50C 0.8 @ 40C 0.5 @ 70C 1.3 @ 35C 0.66 @ 0C 0.45 @ 20C 1 @ 25C 0.4 @ 75C 7 @ 0C 2 @ 50C 0.7 @ 100C 15 @ 75C 5 @ 125C 5 @ 25C 10 @ 25C 3 @ 135C 20 @ 25C
1.71 @ 40C
0.2
0.0004
0.0011
96C (fr)
10C (fl,cc)
12C
0.6
0.0005 @ 20C
0.0014 @ 25C
Water Ethylene Glycol50% #5020 Mineral Oil No.7 #5011 Silicone Oil Type 200.05 #5010 Silicone Oil Type 200.10 #5012 Silicone Oil Type 200.20 #5013 Silicone Oil Type 200.50 #5014
NONE NONE
1.00 0.8 @ 0C
0.0014 0.001
0.0002 @ 25C
10C (v)
166C (fl)
168C
0.87 @ 25C 0.84 @ 75C 0.81 @ 125C 0.92 @ 25C 0.934 @ 25C
0.48 @ 25C 0.53 @ 75C 0.57 @ 125C 0.4 0.43 @ 40C 0.45 @ 100C 0.482 @ 200C 0.370 @ 40C 0.393 @ 100C 0.420 @ 200C 0.4 0.358 @ 40C 0.386 @ 100C 0.433 @ 200C 0.363 @ 40C 0.454 @ 100C 0.505 @ 200C 0.34 @ 100C 0.33
5 @ 25C
133C 211C
1000 @ 25C 10 @ 150C 1000 @ 25C 50 @ 150C 1000 @ 25C 50 @ 150C 1000 @ 25C 50 @ 150C 100 @ 25C 1 @ 150C 100 @ 25C 1 @ 150C 100 @ 25C 1 @ 150C 1.7 /cm3
10C (v)
232C
0.949 @ 25C
30C (v)
278C (fl, cc) 230C (fl, cc) 300C (fl, oc) 300C (fl, oc)
280C 232C
50 @ 25C 50 @ 70C 10 @ 104C 50 @ 80C 7 @ 204C 50 @ 66C 14 @ 204C 34 @ 150C 6.5 @ 300C 2.4 @ 500C
Silicone Oil Type 550 70C (v) #5016 Silicone Oil Type 710 80C (v) #5017 Silicone Oil Type 210-H Heat Transfer Salt #5001 66C (v) 180C (fr)
302C
1.11 @ 25C
315C NONE
0.0003 0.0014
0.00095 0.00041
*Limiting Factors b - boiling point e - high evaporation fl - flash point fr - freeze point **Very low water solubility, ice will form as a slush from condensation below freezing.
flash points, toxicity, and such issues must be considered. You are responsible for reading the MSDS (material safety data sheets) and acting accordingly.
7.1.11.2
28
7 General Operation
100C
0C
100C
200C
300C
400C
500C
600C
Silicone Oil 5017 Silicone Oil 5014 Silicone Oil 5013 Silicone Oil 5012 Silicone Oil 5010 HFE 7500 5023 Halocarbon 5019 Mineral Oil 5011 Water PP PP
PP
10 CS
FL 302C
10 CS FL 232C
FL 280C
10 CS
PP
10 CS
FL 133C
FL 211C
PP
10 CS
10 CS EP
EP 100C
10 CS
10 CS FR BP
FL 177C
Methanol
FR (Pure)
FL 12C
BP
Ethanol Ethylene Glycol (50/50 with H2O) 5020 Dynalene HF/LO 5022 Bath Salt 5001
10 CS FR
FL 16C
10 CS 10 CS FL 60C
BP
FR
Legend
Shaded area represents usable range of fluid starting at 50 centistokes. Lighter shading represents decreasing viscosity, while vaporization and decomposition increase. Black area represents liquid range with excessive viscosity. Range over which a fume hood is recommended.
Decomposition Starts BP - Boiling Point CS - Centistokes EP - Evaporation Point (fluid loss due to evaporation) FL - Flash Point FR - Freeze Point PP - Pour Point
Temperature Range: The temperature scale is shown in degrees Celsius. The shaded bands indicate the fluids general range of application. Qualities including pour point, freeze point, important viscosity points, flash point, boiling point and others may be shown. Freezing Point: The freezing point of a fluid is an obvious limitation to stirring. As the freezing point is approached high viscosity may also limit performance.
29
7 General Operation
Pour Point: This represents a handling limit for the fluid. Viscosity: Points shown are at 50 and 10 centistokes viscosity. When viscosity is greater than 50 centistokes stirring is very poor and the fluid is unsatisfactory for bath applications. Optimum stirring generally occurs at 10 centistokes and below. Fume Point: Indicates the point at which a fume hood should be used. This point is very subjective in nature and is impacted by individual tolerance to different fumes and smells, how well the bath is covered, the surface area of the fluid in the bath, the size and ventilation of the facility where the bath is located and other conditions. This is also subject to company policy. In the graph we assume the bath is well covered at this point. Flash Point: The point at which ignition may occur. The point shown may be either the open or closed cup flash point. Refer to the flash point discussion in Section 7.1.8, Safety. Boiling Point: At or near the boiling point of the fluid the temperature stability is difficult to maintain. Fuming or evaporation is excessive. Large amounts of heater power may be required to maintain the temperature because of the heat of vaporization. Decomposition: The temperature may reach a point at which decomposition of the fluid begins. Further increasing the temperature may accelerate decomposition to the point of danger or impracticality.
7.2
Stirring
Stirring the bath fluid is very important for stable temperature control. The fluid must be mixed well for good temperature uniformity and fast controller response. The stirrer is precisely adjusted for optimum performance.
7.3
Power
Power to the bath is provided by an AC mains supply. Refer to Section 3.1, Specifications, for power details. Refer to and read the CAUTION at the front of the manual concerning brownout and over voltage protection. Check the back panel label for the correct voltage and frequency prior to energizing the unit. Power to the bath passes through a filter to prevent switching spikes from being transmitted to other equipment. To turn on the bath press the control panel power switch to the ON position. The stirring motor will turn on, the LED display will begin to show the bath temperature, and the heater will turn on or off until the bath temperature reaches the programmed set-point.
30
7 General Operation
When powered on the control panel display will briefly show a four digit number. This number indicates the number of times power has been applied to the bath. Also briefly displayed is data, which indicates the controller hardware configuration. This data is used in some circumstances for diagnostic purposes.
7.4
Heater
The temperature controller precisely controls the bath heater to maintain a constant bath temperature. Power is controlled by periodically switching the heater on for a certain amount of time using a solid-state relay. The front panel red/green control indicator shows the state of the heater. The control indicator glows red when the heater is on and green when the heater is off. The indicator pulses at a constant rate when the bath is maintaining a stable temperature.
7.5
Temperature Controller
A unique hybrid digital/analog temperature controller controls the bath temperature. The controller offers the tight control stability of an analog temperature controller as well as the flexibility and programmability of a digital controller. The bath temperature is monitored using a platinum resistance sensor as the control probe. The signal is electronically compared with the programmable reference signal, amplified, and then passed to a pulse-width modulator circuit that controls the amount of power applied to the bath heater. The bath is operable within the temperature range given in the specifications. For protection against a solid-state relay failure or other circuit failure, the micro-controller automatically turns off the heater with a second mechanical relay anytime the bath temperature is more than a certain amount above the set-point temperature. In addition to this protection, the controller is also equipped with a separate thermocouple temperature monitoring circuit that shuts off the heater if the temperature exceeds the cutout set-point. The controller allows the operator to set the bath temperature with high resolution, set the cutout temperature, adjust the proportional band, monitor the heater output power, and program the controller configuration and calibration parameters. The controller may be operated in temperature units of degrees Celsius or Fahrenheit. The controller is operated and programmed from the front control panel using the four key switches and digital LED display. Remote digital operation with the controller is possible via the standard RS-232 serial port. Operation of the controller using the front control panel is discussed following in Section 8, Controller Operaton. Operation using the digital interface is discussed in Section 9, Digital Communication Interface. When the controller is set to a new set-point, the bath heats or cools to the new temperature. Once the new temperature is reached, the bath usually takes 1520
31
7 General Operation
minutes for the temperature to settle and stabilize. There may be a small overshoot or undershoot of about 0.5C during this process.
7.6
Refrigeration
Cooling is provided by a compact refrigeration system utilizing the ozone safe R-507 HFC refrigerant. Refrigeration is often not needed when the bath is above 45 to 60C. The automated system automatically turns the system off when above 60C. The refrigeration system automatically changes modes depending upon bath temperature and operation needs as described below. The refrigeration system is activated when the switch on the front panel is turned on. If the bath is below the 60C cutoff point and is set to a temperature above 60C, the refrigeration shuts off to protect it from overheating and creating excessive internal pressures. If the bath is above 60C and is set to a new temperature below the cutoff temperature, the refrigeration turns on again at 59C. When the bath is controlling at temperatures between the 60C and 0C, the bath operates in a low cooling capacity mode with the hot gas bypass valve on (HGb on). The HGb system reduces the cooling capacity and helps to improve the bath stability within that temperature range. Since more cooling capacity is generally required below 0C, the hot gas bypass is turned off (HGb off) providing more cooling capacity. When the bath is cooled from temperatures below the cutoff point to 0C and if the new set-point is at least 2C below the current bath temperature, the cooling turns to maximum capacity (HGb off) until the bath is within about 0.5C of the new set-point. This provides the fastest cooling possible to achieve the new set-point temperature. As indicated, the cooling is always maximum below 0C. When heating the bath from temperatures of 40C and the new set-point is at least 10C above the current bath temperature, the refrigeration turns off until the bath is about 1C below the new set-point. This permits the maximum heating rate to reach the higher temperatures as quickly as possible. The cooling remains off at temperatures above 60C. The automatic cooling mode may not be ideal for all circumstances. The automatic mode may be turned off permitting manual control. These controls are available through the front panel as described in Section 8.13, Operating Parameters, or through the digital interface as described in Sections 9.2, Interface Commands and 9.3, Cooling Control. The cooling mode control may be set to automatic, on, or off. With these selections the refrigeration is either in automatic as described above, always on, or always off. The hot gas bypass modes are similarly selectable between automatic, on, or off. This system is either in automatic as described above, always on (low cooling capacity) or always off (high cooling capacity).
32
7 General Operation
The following situations benefit by changing cooling modes. If the scan mode has been selected and the desired cooling scan requires maximum cooling capacity in a range where the hot gas bypass would normally be on. The application requires maximum cooling capacity when the hot gas bypass would normally be on. When using the bath at temperatures below 0C in an ambient of less than 23C, a small benefit to stability may be realized by using the hot gas bypass. The refrigeration may be beneficially used for short times (less than one hour) above 60C but less than 100C. This list is not intended to be complete, but only suggests some of the situations when automatic modes may not be best. Most of the time, the automatic functions are adequate and should be used.
33
8 Controller Operation
Controller Operation
This chapter discusses in detail how to operate the bath temperature controller using the front control panel. Using the front panel key switches and LED display the user may monitor the bath temperature, set the temperature set-point in degrees C or F, monitor the heater output power, adjust the controller proportional band, set the cutout set-point, and program the probe calibration parameters, operating parameters, serial interface configuration, and controller calibration parameters. Operation is summarized in Figure 5 on page 36.
8.1
Bath Temperature
The digital LED display on the front panel allows direct viewing of the actual bath temperature. This temperature value is what is normally shown on the display. The units, C or F, of the temperature value are displayed at the right. For example,
25.00 C
The temperature display function may be accessed from any other function by pressing the EXIT button.
8.2
Reset Cutout
If the over-temperature cutout has been triggered then the temperature display will alternately flash,
Cut-out
The message continues to flash between the actual temperature and Cut-out until the temperature is reduced and the cutout is reset. The cutout has two modes automatic reset and manual reset. The mode determines how the cutout is reset which allows the bath to heat up again. When in automatic mode, the cutout will reset itself as soon as the temperature is lowered below the cutout set-point. With manual reset mode the cutout must be reset by the operator after the temperature falls below the set-point. When the cutout is active and the cutout mode is set to manual (reset) then the display will flash cutout until the user resets the cutout. To access the reset cutout function press the SET button.
35
8 Controller Operation
Display Temperature
SET EXIT Cutout Active Reset Cutout SET
SET
UP EXIT Program Menu SET Number of Set-points SET Adjust Set-point SET Select Set-point EXIT
EXIT
SECONDARY FUNCTIONS
SET + EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
UP
EXIT
EXIT
UP DOWN
EXIT
UP DOWN
Probe Menu
SET
DOWN
Calibration Menu
SET
X5
R0
SET EXIT
BAUD Rate
EXIT SET
CTO
EXIT
Adj. R0
SET/EXIT
Sample Period
C0
EXIT
SET
EXIT
ALPHA
SET EXIT
Cooling Mode
SET EXIT SET
Duplex Mode
EXIT
CG
SET EXIT
Adj. ALPHA
SET/EXIT
Cool: Auto/on/off
SET EXIT
Linefeed
SET/EXIT
H
SET EXIT
L
SET EXIT
HGb Auto/on/off
SET/EXIT
HGbt
SET EXIT
Adj. bt
SET/EXIT
36
8 Controller Operation
rESEt ?
Reset cutout
This action switches the display to the set temperature function. To return to displaying the temperature display press the EXIT button. If the cutout is still in the over-temperature fault condition, the display continues to flash cutout. The bath temperature must drop a few degrees below the cutout set-point before the cutout can be reset.
8.3
Temperature Set-point
The bath temperature can be set to any value within the range and with resolution as given in the specifications. The operator must know the temperature range of the particular fluid used in the bath and the bath should only be operated well below the upper temperature limit of the liquid. In addition, the cutout temperature should also be set below the upper limit of the fluid. Setting the bath temperature involves three steps: 1) selecting the set-point memory, 2) adjusting the set-point value, and 3) adjusting the vernier, if desired.
8.3.1
Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be quickly recalled to conveniently set the bath to a previously programmed temperature set-point. To set the bath temperature one must first select the set-point memory. This function is accessed from the temperature display function by pressing SET. The number of the set-point memory currently being used is shown at the left on the display followed by the current set-point value.
25.00 C
S
1.
25.0
U
4.
Increment memory
40.0
37
8 Controller Operation
Press SET to accept the new selection and access the set-point value.
S
8.3.2
Set-point Value
After selecting the set-point memory and pressing SET, the set-point value may be adjusted in increments of 0.01 (C or F). The set-point value is displayed with the units, C or F, at the left.
40.00
Set-point 4 value in C
If the set-point value does not need to be changed, press EXIT to resume displaying the bath temperature. Press UP or DOWN to adjust the set-point value.
U
C
Increment display
42.50
When the desired set-point value is reached, press SET to accept the new value and access the set-point vernier. If EXIT is pressed, any changes made to the set-point are ignored.
S
8.3.3
Set-point Vernier
The user may want to adjust the set-point slightly to achieve a more precise bath temperature. The set-point vernier allows one to adjust the temperature below or above the set-point by a small amount with very high resolution. Each of the 8 stored set-points has an associated vernier setting. The set-point vernier can be set in increments of 0.00018C. The vernier is accessed from the set-point by pressing SET. The vernier setting is displayed as a 6 digit number with five digits after the decimal point. This is a temperature offset in degrees of the selected units, C or F.
0.00000
To adjust the vernier, press UP or DOWN. Unlike most functions the vernier setting has immediate effect as the vernier is adjusted. SET need not be pressed. This allows the bath temperature to be continually adjusted as it is displayed.
U
38
Increment display
8 Controller Operation
0.00090
Next press EXIT to return to the temperature display or SET to access the temperature scale units selection.
S
8.4 Scan
The scan rate can be set and enabled so that when the set-point is changed the instrument heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the instrument heats or cools at the maximum possible rate.
8.4.1
Scan Control
The scan is controlled with the scan on/off function that appears in the main menu after the set-point function.
ScAn=OFF
ScAn=On
Scan function on
S
8.4.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set from 0.001 to 5.0 C/min. The maximum scan rate however is actually limited by the natural heating or cooling rate of the instrument. This is less than 2.5C/min, especially when cooling. The scan rate function appears in the main menu after the scan control function. The scan rate units are in degrees per minute, degrees C or F depending on the selected units.
Sr=0.010
Sr=2.0
39
8 Controller Operation
8.5
Un= C
U S
8.6
Un= F
Press SET to accept the new selection and resume displaying the bath temperature. Set the new units and resume temperature display
| 100.00 C
S+U
ProG
S
8.6.1
40
8 Controller Operation
set-points are independent from the programmable set-points described in Section 8.3.1, Programmable Set-points.
Pn=8
Pn=3
Press SET to continue. Press EXIT to ignore any changes and to continue.
S
8.6.2 Set-points
1 50.0
First set-point
3 30.0
Third set-point
C 30.00
Set-point value
C 40.00
Press SET to save the new set-point value. The other set-points can also be set in the same manner. Once the set-points are programmed as desired press EXIT to continue.
E
8.6.3
Pt=15
41
8 Controller Operation
Pt=5
S
8.6.4
Pf=1
Program mode
Pf=4
New mode
S
8.6.5
Program Control
The final parameter in the program menu is the control parameter. Three options are available for controlling the ramp and soak program. The options are to start the program from the beginning, (GO), continue the program from where it was when it was stopped (Cont), or stop the program (OFF).
Pr=OFF
8 Controller Operation
Pr=Cont
Press SET to activate the new program control command and return to the temperature display.
S
8.7
Secondary Menu
Functions, which are used less often, are accessed within the secondary menu. The secondary menu is accessed by pressing SET and EXIT simultaneously and then releasing. The first function in the secondary menu is the heater power display.
8.8
Heater Power
The temperature controller controls the temperature of the bath by pulsing the heater on and off. The total power being applied to the heater is determined by the duty cycle or the ratio of heater on time to the pulse cycle time. This value may be estimated by watching the red/green control indicator light or read directly from the digital display. By knowing the amount of heating the user can tell if the bath is heating up to the set-point, cooling down, or controlling at a constant temperature. Monitoring the percent heater power lets the user know the stability of the bath temperature. With good control stability the percent heating power should not fluctuate more than 1% within one minute. The heater power display is accessed in the secondary menu. Press SET and EXIT simultaneously and release. The heater power is displayed as a percentage of full power.
S+E
12 Pct
To exit out of the secondary menu press EXIT. To continue on to the proportional band setting function press SET.
E
8.9
Proportional Band
In a proportional controller such as this the heater output power is proportional to the bath temperature over a limited range of temperatures around the set-point. This range of temperature is called the proportional band. At the bottom of the proportional band the heater output is 100%. At the top of the pro43
8 Controller Operation
portional band the heater output is 0. Thus as the bath temperature rises the heater power is reduced, which consequently tends to lower the temperature back down. In this way the temperature is maintained at a fairly constant temperature. The temperature stability of the bath depends on the width of the proportional band (see Figure 6). If the band is too wide, the bath temperature deviates excessively from the set-point due to varying external conditions. This is because the power output changes very little with temperature and the controller cannot respond very well to changing conditions or noise in the system. If the proportional band is too narrow, the bath temperature may swing back and forth because the controller overreacts to temperature variations. For best control stability the proportional band must be set for the optimum width.
The optimum proportional band width depends on several factors among which are fluid volume, fluid characteristics (viscosity, specific heat, thermal conductivity), heater power setting, operating temperature, and stirring. Thus the proportional band width may require adjustment for best bath stability when any of these conditions change. Of these, the most significant factors affecting the optimum proportional band width are heater power setting and fluid viscosity. The proportional band should be wider when the higher power setting is used so that the change in output power per change in temperature remains the same. The proportional band should also be wider when the fluid viscosity is higher because of the increased response time. The proportional band width is easily adjusted from the bath front panel. The width may be set to discrete values in degrees C or F depending on the selected units. The optimum proportional band width setting may be determined by monitoring the stability with a high resolution thermometer or with the control44
8 Controller Operation
ler percent output power display. Narrow the proportional band width to the point at which the bath temperature begins to oscillate and then increase the band width from this point to 3 or 4 times wider. Table 4 lists typical proportional band settings for optimum performance with a variety of fluids at selected temperatures.
Table 4 Typical Proportional Band Settings for Various Fluids Fluid Water Water Eth-Gly 50% Eth-Gly 50% Eth-Gly 50% Oil 200, 10cs Oil 200, 10cs Oil 200, 10cs Oil 710 Temperature 30C 60C 35C 60C 100C 35C 60C 100C 200C Proportional Band 0.31C 0.31C 0.31C 0.31C 0.4C 0.6C 0.6C 0.6C 0.4C Stability 0.003C 0.003C 0.005C 0.005C 0.010C 0.004C 0.004C 0.004C 0.008C
The proportional band adjustment may be accessed within the secondary menu. Press SET and EXIT to enter the secondary menu and show the heater power. Then press SET to access the proportional band.
S+E
12 Pct
S D
Pb=0.101C
To change the proportional band press UP or DOWN. Decrement display New proportional band setting
Pb=0.060C
To accept the new setting and access the cutout set-point press SET. Pressing EXIT will exit the secondary menu ignoring any changes just made to the proportional band value.
45
8 Controller Operation
S
8.10 Cutout
As a protection against software or hardware fault, shorted heater triac, or user error, the bath is equipped with an adjustable heater cutout device that shuts off power to the heater if the bath temperature exceeds a set value. This protects the heater and bath materials from excessive temperatures and, most importantly, protects the bath fluids from being heated beyond the safe operating temperature preventing hazardous vaporization, breakdown, or ignition of the liquid. The cutout temperature is programmable by the operator from the front panel of the controller. It must always be set below the upper temperature limit of the fluid and no more than 10 degrees above the upper temperature limit of the bath. If the cutout is activated because of excessive bath temperature, power to the heater is shut off and the bath cools. The bath cools until it reaches a few degrees below the cutout set-point temperature. At this point the action of the cutout is determined by the setting of the cutout mode parameter. The cutout has two selectable modes automatic reset or manual reset. If the mode is set to automatic, the cutout automatically resets itself when the bath temperature falls below the reset temperature allowing the bath to heat up again. If the mode is set to manual, the heater remains disabled until the user manually resets the cutout. The cutout set-point may be accessed within the secondary menu. Press SET and EXIT to enter the secondary menu and show the heater power. Then press SET twice to access the cutout set-point.
S+ E
12 Pct
S S D
46
Pb=0.101C
CO= 110C
8 Controller Operation
CO= 75C
S
8.11
The next function is the configuration menu. Press EXIT to resume displaying the bath temperature.
Controller Configuration
The controller has a number of configuration and operating options and calibration parameters, which are programmable via the front panel. These are accessed from the secondary menu after the cutout set-point function by pressing SET. There are 5 sets of configuration parameters - probe parameters, operating parameters, serial interface parameters, and controller calibration parameters. The menus are selected using the UP and DOWN keys and then pressing SET. Pressing EXIT in any secondary menu exits and returns to displaying the temperature (see Figure 5 on page 36, Controller Operation Flowchart).
8.12
Probe Parameters
The probe menu is indicated by,
PrObE
Press SET to enter the menu. The probe parameters menu contains the parameters, R0 and ALPHA, which characterize the resistance-temperature relationship of the platinum control probe. These parameters may be adjusted to improve the accuracy of the bath. This procedure is explained in detail in Section 10, Calibration Procedure. The probe parameters are accessed by pressing SET after the name of the parameter is displayed. The value of the parameter may be changed using the UP and DOWN buttons. After the desired value is reached press SET to set the parameter to the new value. Press EXIT to skip the parameter ignoring any changes that have been made.
8.12.1
R0
This probe parameter refers to the resistance of the control probe at 0C. Normally this is set for 100.000 ohms.
47
8 Controller Operation
8.12.2
ALPHA
This probe parameter refers to the average sensitivity of the probe between 0 and 100C. Normally this is set for 0.00385C1.
8.13
Operating Parameters
The operating parameters menu is accessed by pressing UP when the probe menu is displayed. The operating parameters menu is indicated by,
PAr
Press UP to enter the menu. The operating parameters menu contains the cutout reset mode parameter, cooling mode, and hot gas bypass mode.
8.13.1
CtorSt
Press SET to access the parameter setting. Normally the cutout is set for automatic mode.
Cto=Auto
The cooling mode determines whether refrigeration is in Auto mode, On, or Off. Normally the cooling mode is set to Auto mode. In the Auto mode, the refrigeration is On below approximately 60C. Note: If the fluid is cooling from above 60C, the refrigeration turns on at approximately 59C. If the fluid is heating from below 60C, the refrigeration shuts off at 60C. There may be times when Auto mode is undesirable. In that case, the refrigeration may be set to On or Off. When the refrigeration is set to On or Off, the refrigeration is on or off for all temperatures.
48
8 Controller Operation
Caution: The refrigeration may be damaged or the lifetime shortened if used above 60C for more than one hour. The parameter is indicated by,
CooL
C=Auto
To change to On or Off mode, press DOWN until the desired mode appears and then press SET.
C=On C=Off
8.13.3
Hgb
Hgb=Auto
To change to On or Off mode, press DOWN until the desired mode appears and then press SET.
49
8 Controller Operation
Hgb=On Hgb=Off
8.14
SErIAL
The serial interface parameters menu contains parameters which determine the operation of the serial interface. The parameters in the menu arebaud rate, sample period, duplex mode, and linefeed.
8.14.1
Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting determines the serial communications transmission rate. The baud rate parameter is indicated by,
bAUd
Press SET to choose to set the baud rate. The current baud rate value will then be displayed.
9600 b
The baud rate of the bath serial communications may be programmed to 300, 600, 1200, 2400, or 9600 baud. Use UP or DOWN to change the baud rate value.
2400 b
Press SET to set the baud rate to the new value or EXIT to abort the operation and skip to the next parameter in the menu.
8.14.2
Sample Period
The sample period is the next parameter in the serial interface parameter menu. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the bath transmits the current measurement over the serial interface approximately
50
8 Controller Operation
every five seconds. The automatic sampling is disabled with a sample period of 0. The sample period is indicated by,
SAnPLE
Press SET to choose to set the sample period. The current sample period value will be displayed.
SA= 1
Adjust the value with UP or DOWN and then use SET to set the sample rate to the displayed value.
The next parameter is the duplex mode. The duplex mode may be set to full duplex or half duplex. With full duplex any commands received by the bath via the serial interface are immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed. The duplex mode parameter is indicated by,
dUPL
dUP=FULL
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return. The linefeed parameter is indicated by,
LF
LF= On
8 Controller Operation
LF= OFF
8.15
Calibration Parameters
The operator of the bath controller has access to a number of the bath calibration constants namely CTO, C0, CG, H, L, and HGbt. These values are set at the factory and must not be altered. The correct values are important to the accuracy and proper and safe operation of the bath. These parameters should not be adjusted. In the event the controllers memory fails, the user may restore these values to the factory settings. A list of these constants and their settings are supplied to the user on the Report of Test with the manual.
Caution: DO NOT change the values of the bath calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the bath. The calibration parameters menu is indicated by,
CAL
8.15.1
CTO
Parameter CTO sets the calibration of the over-temperature cutout. This is not adjustable by software but is adjusted with an internal potentiometer. This parameter is set at the factory.
8.15.2
CO and CG
These parameters calibrate the accuracy of the bath set-point. These are programmed at the factory when the bath is calibrated. Do not alter the value of these parameters. If the user desires to calibrate the bath for improved accuracy, calibrate R0 and ALPHA according to the procedure given in Section 10, Calibration Procedure.
8.15.3
H and L
These parameters set the upper and lower set-point limits of the bath. Note: A high limit (H) setting below the flash point of the fluid in the bath is highly recommended.
52
8 Controller Operation
8.15.4
HGbt
This parameter is the temperature where the hot gas bypass activates. This parameter is factory set. To insure the bath's best performance without damaging its compressor, DO NOT alter the value of this parameter.
53
9.1
Serial Communications
The RS-232 serial interface allows serial digital communications over fairly long distances (15.24 meters). With the serial interface the user may access any of the functions, parameters and settings discussed in Section 8, Controller Operation with the exception of the baud rate setting. The serial interface operates with eight data bits, one stop bit, and no parity.
55
9.1.1
Wiring
The serial communications cable attaches to the bath through the DB-9 connector on the back of the instrument. Figure 7 shows the pin-out of this connector and the suggested cable wiring. To eliminate noise, the serial cable should be shielded with low resistance between the connector (DB-9) and the shield.
9.1.2
Setup
Before operation the serial interface of the bath must first be set up by programming the baud rate and other configuration parameters. These parameters are programmed within the serial interface menu.
Computer (DTE) Connector (DB 25-Pin) 2 TxD 3 RxD 4 RTS 5 CTS 6 DSR 7 GND 8 DCD 20 DTR
To enter the serial parameter programming menu, press EXIT while holding down SET, then release both buttons to enter the Figure 7 Serial Communications Cable Wiring secondary menu. Press SET repeatedly until the display reads ProbE. This is the menu selection. Press UP repeatedly until the serial interface menu is indicated with SErIAL. Finally press SET to enter the serial parameter menu. In the serial interface parameters menu are the baud rate, the sample rate, the duplex mode, and the linefeed parameter.
9.1.2.1
Baud Rate
The baud rate is the first parameter in the menu. The display prompts with the baud rate parameter by showing BAUd. Press SET to choose to set the baud rate. The current baud rate value is displayed. The baud rate of the bath may be programmed to 300, 600, 1200, 2400, or 9600 baud. The baud rate is pre-programmed to 9600 baud. Use UP or DOWN to change the baud rate value. Press SET to set the baud to the new value or EXIT to abort the operation and skip to the next parameter in the menu.
56
9.1.2.2
Sample Period
The sample period is the next parameter in the menu and prompted with SAnPLE. The sample period is the time period, in seconds, between temperature measurements transmitted from the serial interface. If the sample period is set to 5, the bath transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press SET to choose to set the sample period. Adjust the period with UP or DOWN. Press SET to set the sample period to the new value or EXIT to abort and skip to the next parameter.
9.1.2.3
Duplex Mode
The next parameter is the duplex mode indicated with dUPL. The duplex mode may be set to half duplex (HALF) or full duplex (FULL). With full duplex any commands received by the bath via the serial interface are executed and immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed. The default setting is full duplex. The mode may be changed using UP or DOWN. Press SET to save the new setting or EXIT to abort and skip to the next parameter.
9.1.2.4
Linefeed
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (On) or disables (OFF) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return. The default setting is with linefeed on. The mode may be changed using UP or DOWN. Press SET to set the sample period to the new value or EXIT to abort and skip to the next parameter.
9.1.3
Serial Operation
Once the cable has been attached and the interface set up properly the controller will immediately begin transmitting temperature readings at the programmed rate. The set-point and other commands may be sent to the bath via the serial interface to set the bath and view or program the various parameters. The interface commands are discussed in Section 9.2, Interface Commands.
9.2
Interface Commands
The various commands for accessing the bath controller functions via the digital interfaces are listed in this section (see Table 5 starting on page 59). These commands are used with the RS-232 serial interface. The commands are terminated with a carriage-return character. The interface makes no distinction between upper and lower case letters, hence either may be used. Commands may be abbreviated to the minimum number of letters, which determines a unique
57
command. A command may be used to either set a parameter or display a parameter depending on whether or not a value is sent with the command following a = character. For example an s <cr> returns the current set-point and an s=50.00" <cr> sets the set-point to 50.00 degrees. In the list of commands, characters or data within brackets, [ and ], are optional for the command. A slash, /, denotes alternate characters or data. Numeric data, denoted by n, may be entered in decimal or exponential notation. Characters are shown in lower case although upper case may be used. Spaces may be added within command strings and will simply be ignored. Backspace (BS, ASCII 8) may be used to erase the previous character. A terminating CR is implied with all commands.
9.3
Cooling Control
The CTR 40 bath has a fully automated refrigeration control system when the cooling power switch on the front panel is activated. Under normal conditions, the refrigeration is on at any temperature below approximately 60C (see Section 7.6, Refrigeration). When the bath is controlling at temperatures between 0C and 60C, the refrigeration is in the hot gas bypass or reduced cooling mode. Below 0C the refrigeration is in the high cooling mode, the hot gas bypass is turned off automatically. When the bath is cooling from one temperature to another below 60C, cooling is at maximum until the bath is within 1 degree C of the set-point. These conditions are the default conditions but may be altered in the following ways. The cooling (refrigeration) may be set to operate in the auto, on, or off modes. See Section 8.13.2, Cooling Mode. The hot gas bypass or reduced cooling mode may also be set to on, off, or auto. The auto modes allow the bath to operate automatically in the manner described above. The cooling on function eliminates this auto feature and the refrigeration is on at all times unless the front panel switch is turned off or until some other menu selection is made. The hot gas bypass mode selection works the same way. The auto function is the default and active until off or on is selected. On means it is always on with no automatic selection and off is always off (or cooling always on maximum). It may desirable for different reasons to eliminate the automatic functioning; therefore, these selections are made available either through the front panel or through the digital communications interface.
58
Table 5 Interface Command Summary Command Format s[etpoint] s[etpoint]=n v[ernier] v[ernier]=n sc[an] Command Example s s=150 v v=.00001 sc scan: {ON or OFF} scan: ON ON or OFF sc[an]=on sc[an]=of[f] sr[ate] sr[ate]=n t[emperature] u[nits] u[nits]=c/f u[nits]=c u[nits]=f u=c u=f sc=on sc=of sr sr=5 t u t: 9999.99 {C or t: 55.69 C F} u: x u: c C or F srat: 9.999 {C or srat: 0.010 C/min F}/min 0.001 to 5.000C/min 0.002 to 9.000F/min v: 9.99999 v: 0.00000 Depends on Configuration Returned Example
Command Description Display Temperature Read current set-point Set current set-point to n Read vernier Set vernier to n Read scan function Set scan function: Turn scan function on Turn scan function off Read scan rate Set scan rate to n degrees per minute Read temperature Read temperature units Set temperature units: Set temperature units to Celsius Set temperature units to Fahrenheit Ramp and Soak Menu Read number of programmable set-points Set number of programmable set-points to n Read programmable set-point number n
Returned
Acceptable Values
pn pn=n psn
pn: 9
pn: 2 2 to 8
psn: 9999.99 {C ps1: 50.00 C or F} 1 to 8, Instrument Range ti: 999 ti: 5 0 to 500 prog: {OFF or ON} prog: OFF GO or STOP or CONT
Set programmable set-point num- psn=n ber n to n Read program set-point soak time pt Set program set-point soak time to pt=n n minutes Read program control mode Set program control mode: Start program pc pc=g[o]/s[top]/c[ont] pc=g[o]
pc=g
59
9 Digital Communication Interface Interface Command Summary Continued Command Format pc=s[top] pc=c[ont] pf pf=n pr[op-band] pr[op-band]=n c[utout] c[utout]=n/r[eset] c[utout]=n c[utout]=r[eset] po[wer] c=160 c=r po po: 9999 po: 1 Temperature Range Command Example pc=s pc=c pf pf=2 pr pr=0.326 c cu: 9999 {x},{xxx} cu: 160 C, in pr: 999.999 pr: 0.326 Depends on Configuration pf: 9 pf: 3 1 to 4 Returned Example
Command Description Stop program Continue program Read program function Set program function to n Secondary Menu Read proportional band setting Set proportional band to n Read cutout setting Set cutout setting: Set cutout to n degrees Reset cutout now Read heater power (duty cycle) Configuration Menu Probe Menu Read R0 calibration parameter Set R0 calibration parameter to n
Returned
Acceptable Values
r[0] r[0]=n
r r=100.324 al al=0.0038433
Read ALPHA calibration parameter al[pha] Set ALPHA calibration parameter to al[pha]=n n Operating Parameters Menu Read cutout mode Set cutout mode: Set cutout to be reset manuallySet cutout to be reset automatically Read serial sample setting Set serial sampling setting to n seconds Set serial duplex mode: Set serial duplex mode to full Set serial duplex mode to half Set serial linefeed mode: Set serial linefeed mode to on Set serial linefeed mode to off cm[ode] cm[ode]=r[eset]/a[uto] cm[ode]=r[eset] cm[ode]=a[uto]
cm cm=r cm=a
cm: {xxxx}
sa sa=0
sa: 9999
60
9 Digital Communication Interface Interface Command Summary Continued Command Format *c0 *c0=n *cg *cg=n co[ol] co[ol]=of[f]/on/au[to] co[ol]=of[f] co[ol]=on co[ol]=au[to] hg[b] hg[b]=of[f]/on/au[to] hg[b]=of[f] hg[b]=on hg[b]=au[to] *tl[ow] *tl[ow]=n *th[igh] *th[igh]=n *ver[sion] h[elp] [] Optional Command data {} Returns either information n Numeric data supplied by user 9 Numeric data returned to user x Character data returned to user Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column. hg=of hg=on hg=au *tl *tl=40 *th *th=150 *ver h ver.9999,9.99 list of commands ver.7340,1.00 th: 999 th: 150 150 to 30 tl: 999 tl: 40 60 to 20 co=of co=on co=au hg hgb:xxx hgb: Auto On, Off, Auto Command Example *c0 *c0=0 *cg *cg=406.25 co co:xxx co: Auto On, Off, Auto cg: 999.999 cg: 406.25 Unlimited Returned Example c0: 0.0002 Unlimited
Command Description Calibration Menu Read C0 calibration parameter Set C0 calibration parameter to n Read CG calibration parameter Set CG calibration parameter to n Read Cool mode Set Cool mode: Set Cool mode to Off Set Cool mode to On Set Cool mode to Auto Read HGb mode Set HGb mode: Set HGb mode to Off Set HGb mode to On Set HGb mode to Auto Read low set-point limit value Set low set-point limit to n Read high set-point limit value Set high set-point limit to n Miscellaneous (not on menus) Read firmware version number Read structure of all commands Legend:
Acceptable Values
61
10 Calibration Procedure
10
Calibration Procedure
In some instances the user may want to calibrate the bath to improve the temperature set-point accuracy. Calibration is done by adjusting the controller probe calibration constants R0 and ALPHA so that the temperature of the bath as measured with a standard thermometer agrees more closely with the bath set-point. The thermometer used must be able to measure the bath fluid temperature with higher accuracy than the desired accuracy of the bath. By using a good thermometer and carefully following procedure the bath can be calibrated to an accuracy of better than 0.02C over a range of 100 degrees.
10.1
Calibration Points
In calibrating the bath R0 and ALPHA are adjusted to minimize the set-point error at each of two different bath temperatures. Any two reasonably separated bath temperatures may be used for the calibration however best results will be obtained when using bath temperatures which are just within the most useful operating range of the bath. The further apart the calibration temperatures, the greater the calibrated temperature range and the calibration error. If, for instance, 0C and 100C are chosen as the calibration temperatures, the bath may achieve an accuracy of 0.03C over the range 10 to 110C. Choosing 30C and 70C may allow the bath to have a better accuracy of 0.01C over the range 25 to 75C, but, outside that range the accuracy may be only 0.05C.
10.2
10.3
10 Calibration Procedure
Compute errors, errL = 29.843 - 30.00C = -0.157C errH = 79.914 - 80.00C = -0.086C
Compute R0,
( 0.086 ) 30.0 ( 0157) 80.0 . 0.00385 + 1 R0 = 80.0 30.0 100.000 = 100.077
Compute ALPHA,
(1 + 0.00385 80.0 )( 0157) (1 + 0.00385 30.0 )( 0.086 ) . ALPHA = + 1 0.00385 = 0.0038416 80.0 30.0
stored in the future. The new values R0 and ALPHA are computed by entering the old values for R0 and ALPHA, the calibration temperature set-points tL and tH, and the temperature errors errL and errH into the following equations,
err t errL tH ALPHA + 1R0 R0 = H L tH tL
64
10 Calibration Procedure
If R0 and ALPHA were previously set for 100.000 and 0.0038500 respectively and the data for tL, tH, errL, and errH were as given above then the new values R0 and ALPHA would be computed as 110.116 and 0.0038302 respectively. Program the new values R0 and ALPHA into the controller. Check the calibration by setting the temperature to tL and tH and measuring the errors again. If desired, the calibration procedure may be repeated again to further improve the accuracy.
10.4
Calibration Example
The bath is to be used between 25C and 75C and it is desired to calibrate the bath as accurately as possible for operation within this range. The current values for R0 and ALPHA are 100.000 and 0.0038500 respectively. The calibration points are chosen to be 30.00 and 80.00C. The measured bath temperatures are 29.843 and 79.914C respectively. Refer to Figure 8 for applying equations to the example data and computing the new probe constants.
65
11 Charging Instructions
11
Charging Instructions
The CTR 40 uses R-507 refrigerant with a polyolester oil. Care must be taken to avoid contamination from other types of refrigerants and oils.
11.1
Leak Testing
Leak testing should be done with equipment designed for use with R-507. Bubble or ultra-sonic leak testing may be viable in some instances.
11.2
Evacuation
DO NOT leave the system open for more than 15 minutes. Polyolester oils are very hygroscopic. Evacuate the system to a minimum of 200 microns. Evacuate from both high and low sides of the system. Schrader valves provide access to the system.
11.3
Charging
After evacuation, charge the system with approximately 280 grams of R-507. Check the head pressure and verify that the pressure is less than or equal to 260 psi at a room temperature of 23C (73.4F) with the bath operating at 25C. There should be very few bubbles, if any, in the sight glass under these conditions.
67
12 Maintenance
12
Maintenance
The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in dirty or dusty environments. If the outside of the bath becomes soiled, it may be wiped clean with a damp cloth and mild detergent. Do not use harsh chemicals on the surface, which may damage the paint. Periodically check the fluid level in the bath to ensure that the level has not dropped. A drop in the fluid level affects the stability of the bath. Changes in fluid level are dependent upon several factors specific to the conditions in which the equipment is used. A schedule cannot be outlined to meet each set of conditions. Therefore, the bath should be checked weekly and adjustments made as required. Heat transfer medium lifetime is dependent upon the type of medium and the conditions of use. The fluid should be checked at least every month for the first year and regularly thereafter. This fluid check provides a baseline for knowledge of bath operation with clean, usable fluid. Once some fluids have become compromised, the break down can occur rapidly. Particular attention should be paid to the viscosity of the fluid. A significant change in the viscosity can indicate that the fluid is contaminated, being used outside of its temperature limits, contains ice particles, or is close to a chemical breakdown. Once data has been gathered, a specific maintenance schedule can be outlined for the instrument. Refer to Section 7, General Operation, for more information about the different types of fluids used in calibration baths. Depending on the cleanliness of the environment, the internal parts (parts behind the front cover only) of the cold bath should be cleaned and/or checked at least every six months for dust and dirt. Particular attention should be paid to the condensing coil fins. The fins should be vacuumed or brushed free of dust and dirt on a regular basis. Dust and dirt inhibit the operation of the condensing coil and thus compromise the performance and lifetime of the cooling system. To clean or check the internal parts, remove the four screws on the fornt panel display. Remove the two screws under the front panel on the left and right sides. Pull the front panel up and out to remove. If a hazardous material is spilt on or inside the equipment, the user is responsible for taking the appropriate decontamination steps as outlined by the national safety council with respect to the material. MSDS sheets ap69
12 Maintenance
plicable to all fluids used in the baths should be kept in close proximity to the instrument. If the mains supply cord becomes damaged, replace it with a cord of the appropriate gauge wire for the current of the bath. If there are any questions, call Kaye Customer Service for more information. Before using any cleaning or decontamination method except those recommended by Hart, users should check with Kaye Customer Service to be sure that the proposed method will not damage the equipment. If the instrument is used in a manner not in accordance with the equipment design, the operation of the bath may be impaired or safety hazards may arise. The over-temperature cutout should be checked every 6 months to see that it is working properly. In order to check the user selected cutout, follow the controller directions (Section 8.2, Reset Cutout) for setting the cutout. Both the manual and the auto reset option of the cutout should be checked. Set the bath temperature higher than the cutout. Check to see if the display flashes cutout and if the temperature is decreasing. Note: When checking the over-temperature cutout, be sure that the temperature limits of the bath fluid are not exceeded. Exceeding the temperature limits of the bath fluid could cause harm to the operator, lab, and instrument.
70
13 Troubleshooting
13
Troubleshooting
This section contains information on troubleshooting, CE Comments, and a wiring diagram. This information pertains to a number of bath models and certain specifics may not pertain to your model.
13.1
Troubleshooting
In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions. If a problem arises, please read this section carefully and attempt to understand and solve the problem. If the probe seems faulty or the problem cannot otherwise be solved, contact Kaye Customer Service for assistance. Be sure to have the instrument model number, serial number, voltage, and problem description available.
Problem The heater indicator LED stays red but the temperature does not increase Causes and Solutions If the display does not show Cut-out nor displays an incorrect bath temperature, and the controller otherwise appears to operate normally, the problem may be insufficient heating, no heating at all, or too much cooling. The heater power setting being too low, especially at higher operating temperature One or more burned out heaters or blown heater fuses may also cause this problem. If the heaters seem to be burned out, contact Kaye Customer Service for assistance.
71
13 Troubleshooting
Problem
The controller display flashes The display flashs Cut-out alternately with the process Cut-out and the heater does temperature. not operate If the process temperature displayed seems grossly in error, consult the following problem: The display flashes Cut-out and an incorrect process temperature. Normally, the cutout disconnects power to the heater when the bath temperature exceeds the cutout set-point causing the temperature to drop back down to a safe value. If the cutout mode is set to AUTO, the heater switches back on when the temperature drops. If the mode is set to RESET, the heater only comes on again when the temperature is reduced and the cutout is manually reset by the operator, see Section 8.10, Cutout. Check that the cutout set-point is adjusted to 10 or 20C above the maximum bath operating temperature and that the cutout mode is set as desired. If the cutout activates when the bath temperature is well below the cutout set-point or the cutout does not reset when the bath temperature drops and it is manually reset, then the cutout circuitry or the cutout thermocouple sensor may be faulty or disconnected. Contact Kaye Customer Service for assistance. The display flashes cutout and an The problem may be that the controllers voltmeter circuit is not functionincorrect process temperature ing properly. A problem could exist with the memory back-up battery. If the battery voltage is insufficient to maintain the memory, data may become scrambled causing problems. A nearby large static discharge may also affect data in memory. Verify that the parameters on the Report of Test. are accurate. Cycle the power off, disconnect the bath from AC, and then restart the bath. If the problem reoccurs, the battery should be replaced. Contact Kaye Customer Service for assistance. If initializing the memory does not remedy the problem, there may be a failed electronic component. Contact Kaye Customer Service for assistance. The controller may need to be reset. Perform the following Factory Reset Sequence. Factory Reset Sequence. Hold the SET and EXIT buttons down at the same time while powering up the instrument. The instrument display shows '-init-', the model number, and the firmware version. Each of the controller parameters and calibration constants must be reprogrammed. The values can be found on the Report of Test that was shipped with the instrument.
72
13 Troubleshooting
Problem
The displayed process temperature Possible causes may be either a faulty control probe or erroneous data in is in error and the controller rememory. mains in the cooling or the heating The probe may be disconnected, burned out, or shorted. The probe is lostate at any set-point value cated inside the stirrer motor cover. Check that the probe is connected properly. The probe may be checked with an ohmmeter to see if it is open or shorted. The probe is a platinum 4-wire Din 43760 type. The resistance should read 0.2 to 2.0 ohms between pins 1 and 2 on the probe connector and 0.2 to 2.0 ohms between pins 3 and 4. It should read 100 to 300 ohms between pins 1 and 4 depending on the temperature. If the probe appears to be defective, contact Kaye Customer Service for assistance. If the problem is not the probe, erroneous data in memory may be the cause. Re-initialize the memory as discussed in the problem The display flashes cutout and an incorrect process temperature. If the problem remains, the cause may be a defective electronic component, contact Kaye Customer Service for assistance. The controller controls or attempts to control at an inaccurate temperature The controller operates normally except when controlling at a specified set-point. At this set-point, the temperature displayed does not agree with the temperature measured by the users reference thermometer to within the specified accuracy. This problem may be caused by an actual difference in temperature between the points where the control probe and thermometer probe measure temperature, by erroneous bath calibration parameters, or by a damaged control probe. Check that the bath has an adequate amount of fluid in the tank and that the stirrer is operating properly. Check that the thermometer probe and control probe are both fully inserted into the bath to minimize temperature gradient errors. Check that the calibration parameters are all correct according to the Report of Test. If not, re-program the constants. The memory backup battery may be weak causing errors in data as described in the problem: The display flashes cutout and an incorrect process temperature. Check that the control probe has not been struck, bent, or damaged. If the cause of the problem remains unknown, contact Kaye Customer Service for assistance. The controller shows that the output power is steady but the process temperature is unstable Possible causes are an improper proportional band setting or the fluid being used. If the bath temperature does not achieve the expected degree of stability when measured using a thermometer, try adjusting the proportional band to a narrower width as discussed in Section 8.9, Proportional Band. Check to ensure the fluid has not deteriorated or is not too thick.
73
13 Troubleshooting
Causes and Solutions The bath is not stable and the duty cycle is not constant. The proportional band being too narrow typically causes this oscillation. Increase the width of the proportional band until the temperature stabilizes as discussed in Section 8.9, Proportional Band.
The controller erratically heats then If both the bath temperature and output power do not vary periodically cools, control is unstable but in a very erratic manner, the problem may be excess noise in the system. Noise due to the control sensor should be less than 0.001C. However, if the probe has been damaged or has developed an intermittent short, erratic behavior may exist. The probe is located inside the stirrer motor cover. Check for a damaged probe or poor connection between the probe and bath. Intermittent shorts in the heater or controller electronic circuitry may also be a possible cause. Contact Kaye Customer Service for assistance. The bath does not achieve low temperatures Too much heating or not enough cooling can cause this problem. Check that the control indicator glows green showing that the controller is attempting to cool. The heaters may be disabled as a test by temporarily removing the heater fuses. Insufficient cooling may be caused by lack of refrigerant due to a leak in the system. Refer to the Charging Instructions, Section 11. Check the sight glass to verify the presence of liquid refrigerant. It may be difficult to tell if the glass is completely full or completely empty. Verify by watching the glass while the compressor is being turned on.
13.2
13.2.1
Comments
EMC Directive
This equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested.
13.2.2
74
4 3 2 1
REVISIONS
Cooling Power INITIAL RELEASE ST 5/20/02 8/7/2002 ST JG JG Included 4022, corrected drawing # to HSG265 A
V1
13.3
VAC INPUT
GRN/YEL 18
DATE
Blue 18
BRN 14
BLU 14
WHT 14
WHT/BLK 14
GRN/YEL 18
WHT/BLU 18
WHT/BLU/BLK 18
DISPLAY
WHT/ORG/BLK 16
6
MT2 MT1
ORG 18 GRY 16 0 WHT/ORG 16 GRY 18
DIGITAL ANALOG
WHT/BLK 18
K1
WHT/YEL WHT
W43 GRY 18
BLU 14
BLU 18
BRN 18
6
CUT-OUT RELAY
BRN 14
F3 2 F4 2
COOLING SWITCH
FAN 3
GRY 16
S2
1A 2A 2 1
BRN 18
WHT/RED/BLK 16
STIR MOTOR
WHT/VIO 14 WHT/VIO/BLK 14
H G N
GROUNDING STUD
Wht/Yel/Blk 18 Blue 18
D
13
YELLOW
15 J1
TB1
1 2 3 4
5
14
Pin 9 COM Pin 7 NO
ICM 490
Pin 2
TERMINAL BLOCK
WHT/YEL/BLK 18 WHT/YEL 18 GRN/YEL 18 H N G
Pin 6 NC
AC
NOTES: 9. ALL WIRES ARE STRADED COPPER WITH PVC INSULATION RATED @ 300V, 105C AS PER CSA C22.2 (8/9/2002) OR EQUIVALENT. 10. WIRE GAUGE SIZES AS INDICATED MAY NOT BE SUBSTITUTED WITH LARGER GAUGE SIZES WITHOUT PRIOR APPROVAL. 11. CONNECT TO APPLICABLE VOLTAGE. 12. SET THE TIME DELAY TO 2 MINUTES. 13. ADJUST THE INPUT VOLTAGE TO 115/230 V. 14. SET "CHA SEL" TO F2 "CHB SEL" TO F3
Pin 4 115V
P1 J1
Brown 18
WHT/ORG/BLK 18 VIO 18 WHT/ORG 18 MT2 G MT1
12
TRIAC
F1
F2
TRIAC 2
P6 J6
Wiring Diagram
S1
POWER SWITCH
1A
RELAY
P8 J8 P2A P7
2A
WHT/BLK 18
J2
PROBE 12345 J7
TC + -
1 6 B A
CMP P100
7 3
J101 J100
GRN/YEL 16
P101
SUBASSY 6 FUSES
B
LID
J200 K3
B
GRN/YEL 18
4
P200
7
N H
FAN2
WHT/VIO 18
STIR MOTOR
G
W27 WHT/BLU 18 WHT BLK W28 WHT/BLU/BLK 18 BLU RED INSULATE BRN GRN/YEL W29 GRN/YEL 18
WHT BLK BLU INSULATE RED INSULATE BRN W29 GRN/YEL 18 GRN/YEL
WHT/VIO/BLK 18
HEATER
GND CONNECT TO TB1-1
FAN1
WHT/VIO/BLK 18
8
H
W30 WHT/BLU/BLK 18
BLK
W30 WHT/BLU/BLK 18
BLK
WHT/VIO 18
FAN 3
N W31 WHT/BLU 18
A
MT1 4
1- H 2- N
1- H 2- GND 3- N
Gate
5
1- H 2- N 3- N/C
DRAWN BY -
A
1 4 2 3 3
1- RED 2- RED 3- WHT 4- WHT
4 2
1 1- RED
2- RED 3- WHT 4- WHT
D. DAVIS
09/18/00
MT2
3- GND
Q4025P
TRIAC
MOLEX PLUG CONNECTOR P100 MOLEX RECEPTACLE CONNECTOR J100
REV
C
SCALE
64841
NONE
HSG265.TCW
SHEET
1 OF 1
13 Troubleshooting
75