Technical Information Proline Promass 84F


Coriolis flowmeter

Premium accuracy, robustness and transmitter for custody transfer


Application Measuring principle operates independently of physical fluid properties such as viscosity or density Highest measurement performance for liquids and gases under varying, demanding process conditions Device properties Mass flow: measured error 0,05% (PremiumCal) Rated secondary containment up to 40 bar (580 psi) Nominal diameter: DN 8 to 250 ( to 10") 4-line backlit display with touch control Device in compact or remote version Phase-shifted pulse, HART, Modbus RS485 Your benefits Highest process safety immune to fluctuating and harsh environments Fewer process measuring points multivariable measurement (flow, density, temp) Space-saving installation no in/outlet run needs Quality designed for custody transfer; featuring worldwide recognized metrological approvals Flexible data transfer options numerous communication types Automatic recovery of data for servicing

Proline Promass 84F

Table of contents
Function and system design . . . . . . . . . . . . . . . . . . . . . .3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . 20

Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . Medium density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Medium pressure range (nominal pressure) . . . . . . . . . . . . . . Pressure-temperature ratings . . . . . . . . . . . . . . . . . . . . . . . . . . Rupture disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Thermal insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Heating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 20 20 22 27 27 27 27 28 28

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring range in non-custody transfer mode . . . . . . . . . . . Measuring range in custody transfer mode . . . . . . . . . . . . . . . Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 5 6 6 6

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 7 7 7 7 7

Custody transfer measurement. . . . . . . . . . . . . . . . . . 29


Custody transfer variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Suitability for custody transfer, metrological control, obligation to subsequent verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Verification (Example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Stamp points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Electrical connection remote version . . . . . . . . . . . . . . . . . . . . . 9 Switching on the power supply in custody transfer mode . . . 10 Potential equalisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Remote version cable specifications . . . . . . . . . . . . . . . . . . . . . 10

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 31
Design/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 66 66 67

Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Performance characteristics . . . . . . . . . . . . . . . . . . . . 10
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Influence of medium temperature . . . . . . . . . . . . . . . . . . . . . . . Influence of medium pressure . . . . . . . . . . . . . . . . . . . . . . . . . . Design fundamentals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 10 12 12 12 13 13

Certificates and approvals . . . . . . . . . . . . . . . . . . . . . . 68


CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Modbus RS485 certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . Measuring Instruments Directive . . . . . . . . . . . . . . . . . . . . . . . Approval for custody transfer . . . . . . . . . . . . . . . . . . . . . . . . . . Suitability for custody transfer measurement . . . . . . . . . . . . . Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . 68 68 68 68 68 68 68 69 69 70

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . Special mounting instructions . . . . . . . . . . . . . . . . . . . . . . . . . . 15 16 17 17 17 18

Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . 71 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71


Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Communication-specific accessories . . . . . . . . . . . . . . . . . . . . . Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 71 72 72

Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ambient class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CIP cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SIP cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 20 20 20 20 20 20 20

Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . . 73

Endress+Hauser

Proline Promass 84F

Function and system design


Measuring principle The measuring principle is based on the controlled generation of Coriolis forces. These forces are always present when both translational and rotational movements are superimposed. FC = 2 m (v ) FC = Coriolis force m = moving mass = rotational velocity v = velocity of the moving mass in a rotating or oscillating system The amplitude of the Coriolis force depends on the moving mass m, its velocity v in the system, and thus on the mass flow. Instead of a constant angular velocity , the Promass sensor uses oscillation. The measuring tubes through which the measured material flows are brought into oscillation. The Coriolis forces produced at the measuring tubes cause a phase shift in the tube oscillations (see illustration): At zero flow, in other words when the fluid is at a standstill, the two tubes oscillate in phase (1). Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).

A B 1 2

A B 3

A B

a0003383

The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube oscillations at the inlet and outlet. System balance is ensured by the antiphase oscillation of the two measuring tubes. The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow profile. Density measurement The measuring tubes are continuously excited at their resonance frequency. A change in the mass and thus the density of the oscillating system (comprising measuring tubes and fluid) results in a corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The micro-processor utilises this relationship to obtain a density signal. Temperature measurement The temperature of the measuring tubes is determined in order to calculate the compensation factor due to temperature effects. This signal corresponds to the process temperature and is also available as an output. The temperature measurement cannot be used to generate data for invoicing in applications subject to legal metrology controls.

Endress+Hauser

Proline Promass 84F

Measuring system

The measuring system consists of a transmitter and a sensor. Two versions are available: Compact version: transmitter and sensor form a single mechanical unit. Remote version: transmitter and sensor are installed separately. Transmitter
Promass 84
Esc + E

Four-line liquid-crystal display Operation with "Touch control" Application-specific Quick Setup Mass flow, volume flow, density and temperature measurement as well as calculated variables (e.g. corrected volume flow)

a0003672

Sensor
F (Standard version) Universal sensor for fluid temperatures up to +200 C (+392 F). Nominal diameters DN 8 to 250 (" to 10"). Material: Stainless Steel EN 1.4539/ASTM 904L, EN 1.4404/ASTM 316L, Alloy C22 DIN 2.4602 Material: Sensor: Stainless Steel: 1.4301/1.4307 (304L), optional 1.4404 (316/316L) Measuring tube: Stainless Steel: 1.4539 (904L), 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022) Process connections: Stainless Steel: 1.4301 (F304), 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022)

a0003673

F (High-temperature version) Universal high-temperature sensor for fluid temperatures up to +350 C (+662 F). Nominal diameters DN 25, 50, 80 (1", 2", 3") Material: Process connections: Rostfreier Stahl 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022)
a0003675

Endress+Hauser

Proline Promass 84F

Input
Measured variable Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to register a phase shift in the oscillation) Fluid density (proportional to resonance frequency of the measuring tube) Fluid temperature (measured with temperature sensors) Measuring ranges for liquids
DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10 Range for full scale values (liquids) min(F) to max(F) [kg/h] 0 to 2000 0 to 6500 0 to 18000 0 to 45000 0 to 70000 0 to 180000 0 to 350000 0 to 800000 0 to 2200000 [lb/min] 0 to 73.50 0 to 238.9 0 to 661.5 0 to 1654 0 to 2573 0 to 6615 0 to 12860 0 to 29400 0 to 80850

Measuring range in noncustody transfer mode

Measuring ranges for gases The full scale values depend on the density of the gas. Use the formula below to calculate the full scale values: max(G) = max(F) (G) x [kg/m] max(G) = Max. full scale value for gas [kg/h] max(F) = Max. full scale value for liquid [kg/h] (G) = gas density in [kg/m] under process conditions
DN [mm] 8 15 25 40 50 [in] 1 1 2 X 60 80 90 90 90 DN [mm] 80 100 150 250 [in] 3 4 6 10 X 110 130 200 200

Here, max(G) can never be greater than max(F) Calculation example for gas: Sensor type: Promass F, DN 50 Gas: air with a density of 60.3 kg/m (at 20 C and 50 bar) Measuring range (liquids): 70000 kg/h x = 90 (for Promass F DN 50)

Max. possible full scale value: max(G) = max(F) (G) x [kg/m] = 70000 kg/h 60.3 kg/m 90 kg/m = 46900 kg/h Recommended full scale values See information in chapter "Limiting flow" 27

Endress+Hauser

Proline Promass 84F

Measuring range in custody transfer mode

The following are example data for German PTB approval (liquids other than water). Measuring ranges for liquids in mass flow
DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10 Mass flow (liquids) Qmin to Qmax [kg/min] 1.5 to 30 5 to 100 15 to 300 35 to 700 50 to 1000 150 to 3000 200 to 4500 350 to 12000 1500 to 35000 [lbs/min] 3.3075 to 66.15 11.025 to 220.5 33.075 to 661.5 77.175 to 1543.5 110.25 to 2205.0 330.75 to 6615.0 441.00 to 9922.5 771.75 to 26460 3307.5 to 77175 Smallest measured quantity [kg] 0.5 2 5 20 50 100 200 500 1000 [lbs] 1.10 4.41 11.0 44.1 110.25 220.50 441.00 1102.5 2205.0

Measuring ranges for liquids in volume flow (also LPG)


DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10 Volume flow (liquids) Qmin to Qmax [l/min] 1.5 to 30 5 to 100 15 to 300 35 to 700 50 to 1000 150 to 3000 200 to 4500 350 to 12000 1500 to 35000 [gal/hr] 23.76 to 475.20 79.20 to 1584.0 237.6 to 4752.0 554.4 to 11088 792.0 to 15840 2376 to 47520 3168 to 71280 5544 to 190 080 23760 to 554 400 Smallest measured quantity [l] 0.5 2.0 5.0 20 50 100 200 500 1000 [gal] 0.132 0.528 1.320 5.280 13.20 26.40 52.80 132.0 264.0

!
Operable flow range Input signal

Note! For information about the other approvals see corresponding certicate. Over 20 : 1 for verified device Status input (auxiliary input), HART: U = 3 to 30 V DC, Ri = 5 k, galvanically isolated. Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start. Status input (auxiliary input), Modbus RS485: U = 3 to 30 V DC, Ri = 3 k, galvanically isolated, switch level: 3 to 30 V DC, independent of polarity. Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start.

Endress+Hauser

Proline Promass 84F

Output
Output signal Current output, HART Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value selectable, temperature coefficient: typically 0.005% o.r./C, resolution: 0.5 A Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 ) Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri 150 Pulse / frequency output, HART For custody transfer measurement, two pulse outputs can be operated. Passive, galvanically isolated, open collector, 30 V DC, 250 mA Frequency output: Full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max. 2 s. In "Phase-shifted pulse outputs" operating mode, the end frequency is limited to a maximum of 5000 Hz. Pulse output: Pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms) Pulse / frequency output, Modbus Active/passive selectable, galvanically isolated Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Passive: Open Collector, 30 V DC, 250 mA Frequency output: Full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max. 2 s. Pulse output: Pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms) Modbus interface Modbus device type: slave Address range: 1 to 247 Functions codes supported: 03, 04, 06, 08, 16, 23 Broadcast: supported with the function codes 06, 16, 23 Physical interface: RS485 in accordance with standard EIA/TIA-485 Baud rate supported: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud Transmission mode: RTU or ASCII Response time: Direct data access = typically 25 to 50 ms Auto-scan buffer (data area) = typically 3 to 5 ms Possible output combinations 8 Signal on alarm Current output: Failsafe mode selectable (e.g. in accordance with NAMUR Recommendation NE 43). Pulse/frequency output: Failsafe mode selectable. Relay output: De-energised by fault or power supply failure. Modbus RS485: If an error occurs, the value NaN (not a number) is output for the process variables.

Load Low flow cutoff Galvanic isolation Switching output

See "Output signal" Switch points for low flow cutoff are selectable. All circuits for inputs, outputs, and power supply are galvanically isolated from each other. Relay output max. 30 V / 0.5 A AC; 60 V / 0.1 A DC galvanically isolated Normally closed (NC or break) or normally open (NO or make) contacts available (factory setting: relay 1 = NO, relay 2 = NC)

Endress+Hauser

Proline Promass 84F

Power supply
Terminal assignment The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable (flexible), depending on the version ordered (see table). Replacements for modules which are defective or which have to be replaced can be ordered as accessories.
Order characteristic Terminal No. (inputs/outputs) for "inputs/outsputs" 20 (+) / 21 () 22 (+) / 23 () Fixed communication boards (permanent assignment) S T Pulse/freq. output Ex i, passive Pulse/freq. output Ex i, passive Current output Ex i active, HART Current output Ex i passive, HART

24 (+) / 25 ()

26 (+) / 27 ()

Flexible communication boards D M N Q 1 2 7 Status input Status input Current output Relay output Relay output Relay output 2 Relay output Pulse/frequency output Current output HART Current output HART Modbus RS485 Modbus RS485 Current output HART Current output 1 HART Modbus RS485

Pulse/frequency out- Pulse/frequency output 2 put 1 Pulse/frequency outStatus input put Status input

Pulse/frequency out- Pulse/frequency output 2 put 1 Current output 2 Relay output 1 Pulse/frequency output Status input

Supply voltage

85 to 260 V AC, 45 to 65 Hz 20 to 55 V AC, 45 to 65 Hz 16 to 62 V DC AC: <15 VA (including sensor) DC: <15 W (including sensor) Switch-on current Max. 13.5 A (<50 ms) at 24 V DC Max. 3 A (<5 ms) at 260 V AC

Power consumption

Power supply failure

Lasting min. 1 power cycle: EEPROM or T-DAT save measuring system data if the power supply fails. HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial number, calibration factor, zero point, etc.) See also "Switching on the power supply in custody transfer mode" 10.

Endress+Hauser

Proline Promass 84F

Electrical connection

HART b/c /b a/a Modbus RS485

A b/c /b a/a

a/a b/b /c HART


27 + 26 25 + 24 23 + 22 21 + 20 N (L-) 2 L1 (L+) 1

Modbus RS485
A (RxD/TxD-N) 27 B (RxD/TxD-P) 26 25 + 24 f 23 + 22 21 + 20 N (L-) 2 L1 (L+) 1

b e

c e b d a
a0006816

d a

Connecting the transmitter, cable cross-section: max. 2.5 mm A B C a b c d e f View A (field housing) View B (stainless steel field housing) View C (wall-mount housing) Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC Terminal No. 1: L1 for AC, L+ for DC Terminal No 2: N for AC, L- for DC Signal cable: Terminal assignment 8 Fieldbus cable: Terminal assignment 8 Ground terminal for protective earth Ground terminal for Signal cable/RS485 cable Service connector for connecting service interface FXA 193 with Proline adapter cable (Fieldcheck, FieldCare)

Electrical connection remote version

S1 S1 S2 S2 GND TM TM TT TT + + + + 4 5 6 7 8 9 10 11 12 41 42

d d d e c 4 5 6 7 8 9 10 11 12 + + + + S1 S1 S2 S2 GND TM TM TT TT
Connection of remote version a b c d e Wall-mount housing: non-hazardous area and ATEX II3G / zone 2 see separate "Ex documentation" Wall-mount housing: ATEX II2G / Zone 1 /FM/CSA see separate "Ex documentation" Connection housing sensor Cover for connection compartment or connection housing Connecting cable

41 42
a0003681

Terminal No.: 4/5 = grey; 6/7 = green; 8 = yellow; 9/10 = pink; 1 = white; 41/42 = brown

Endress+Hauser

Proline Promass 84F

Switching on the power supply in custody transfer mode

If the device is started in custody transfer mode, for example also after a power outage, system error No. 271 "POWER BRK. DOWN" flashes on the local display. The fault message can be acknowledged or reset using the "Enter" key or by means of the status input configured accordingly. Note! For correct measuring operation, it is not mandatory to reset the fault message. No special measures for potential equalization are required. For instruments for use in hazardous areas, observe the corresponding guidelines in the specific Ex documentation. Power-supply and signal cables (inputs/outputs) Cable entry M20 1.5 (8 to 12 mm / 0.31" to 0.47") Thread for cable entries, " NPT, G " Connecting cable for remote version Cable entry M20 1.5 (8 to 12 mm / 0.31" to 0.47") Thread for cable entries, " NPT, G "

Potential equalisation

Cable entries

Remote version cable specifications

6 0.38 mm PVC cable with common shield and individually shielded cores Conductor resistance: 50 /km (0.015 /ft) Capacitance: core/shield: 420 pF/m (128 pF/ft) Cable length: max. 20 m (65 ft) Operating temperature: max. +105 C (+221 F)

Operation in zones of severe electrical interference: The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC requirements of ICE/EN 61326, and NAMUR recommendation NE 21/43.

Performance characteristics
Reference operating conditions Error limits following ISO/DIN 11631 Water with 15 to 45 C (59 to 113 F); 2 to 6 bar (29 to 87 psi) Data according to calibration protocol Accuracy based on accredited calibration rigs according to ISO 17025

To obtain measured errors, use the Applicator sizing tool Applicator: 72. Maximum measured error o.r. = of reading; 1 g/cm3= 1 kg/l; T = fluid temperature Base accuracy Mass flow and volume flow (liquids) 0.05% o.r. (PremiumCal, for mass flow) 0.10% o.r Mass flow (gases) 0.35% o.r. Density (liquids) Reference conditions: 0.0005 g/cm3 Field density calibration: 0.0005 g/cm3 (valid after field density calibration under process conditions) Standard density calibrations: 0.01 g/cm3 (valid over the entire measuring range of the sensor 20) Temperature 0.5 C 0.005 T C (1 F 0.003 (T 32) F)

10

Endress+Hauser

Proline Promass 84F

Zero point stability Promass F (standard version)


DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10 [kg/h] 0.030 0.200 0.540 2.25 3.50 9.00 14.00 32.00 88.00 Zero point stability [lb/min] 0.001 0.007 0.019 0.083 0.129 0.330 0.514 1.17 3.23

Promass F (high-temperature version)


DN [mm] 25 50 80 [in] 1 2 3 [kg/h] 1.80 7.00 18.0 Zero point stability [lb/min] 0.0661 0.2572 0.6610

Flow values Flow values as turndown parameter depending on nominal diameter. SI units
DN [mm] 8 15 25 40 50 80 100 150 250 1:1 [kg/h] 2000 6500 18000 45000 70000 180000 350000 800000 2200000 1:10 [kg/h] 200.0 650.0 1800 4500 7000 18000 35000 80000 220000 1:20 [kg/h] 100.0 625.0 900.0 2250 3500 9000 17500 40000 110000 1:50 [kg/h] 40.00 130.0 360.0 900.0 1400 3600 7000 16000 44000 1:100 [kg/h] 20.00 65.00 180.0 450.0 700.0 1800 3500 8000 22000 1:500 [kg/h] 4.000 13.00 36.00 90.00 140.0 360.0 700.0 1600 4400

Endress+Hauser

11

Proline Promass 84F

US units
DN [in] 1 1 2 3 4 6 10 1:1 [lb/min] 73.50 238.9 661.5 1654 2573 6615 12860 29400 80850 1:10 [lb/min] 7.350 23.89 66.15 165.4 257.3 661.5 1286 2940 8085 1:20 [lb/min] 3.675 11.95 33.08 82.70 128.7 330.8 643.0 1470 4043 1:50 [lb/min] 1.470 4.778 13.23 33.08 51.46 132.3 257.2 588.0 1617 1:100 [lb/min] 0.735 2.389 6.615 16.54 25.73 66.15 128.6 294.0 808.5 1:500 [lb/min] 0.147 0.478 1.323 3.308 5.146 13.23 25.72 58.80 161.7

Accuracy of outputs o.r. = of reading; o.f.s. = of full scale value The output accuracy must be factored into the measured error if analog outputs are used, but can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP). Current output Accuracy: Max. 0,05 % o.f.s. or 5 A Pulse/frequency output Genauigkeit: Max. 50 % ppm o.r. Repeatability o.r. = of reading; 1 g/cm3= 1 kg/l; T = fluid temperature Design fundamentals 13. Base repeatability Mass flow and volume flow (liquids) 0.025% o.r. (PremiumCal, for mass flow) 0.05% o.r. Mass flow (gases): 0.25% o.r. Density (liquids) 0.00025 g/cm3 Temperature 0.25 C 0.0025 T C (0.45 F 0.0015 (T 32) F) Response time The response time depends on the configuration (damping). Response time in the event of erratic changes in the measured variable (only mass flow): after 100 ms95 % of the full scale value. When there is a difference between the temperature for zero point adjustment and the process temperature, the typical measured error of the Promass sensor is 0.0002% of the full scale value / C (0.0001% of the full scale value / F).

Influence of medium temperature

12

Endress+Hauser

Proline Promass 84F

Influence of medium pressure

The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure and process pressure.
DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10 Promass F (standard version) [% o.r./bar] no influence no influence no influence 0.003 0.008 0.009 0.007 0.009 0.009 Promass F (high-temperature version) [% o.r./bar] no influence 0.008 0.009

Design fundamentals

o.r. = of reading BaseAccu = base accuracy in % o.r. BaseRepeat = base repeatability in % o.r. MeasValue = measured value (in flow units consistent with the zero point stability value 11) ZeroPoint = zero point stability Calculation of the maximum measured error depending on flowrate
Flowrate (in flow units consistent with the zero point stability value 11) Maximum measured error in % o.r.

ZeroPoint BaseAccu

100
A0021332

BaseAccu
A0021339

ZeroPoint < BaseAccu 100


A002133

ZeroPoint MeasValue 100

A0021334

Calculation of the repeatability depending on flowrate


Flowrate (in flow units consistent with the zero point stability value 11) Repeatability in % o.r.

ZeroPoint BaseRepeat ZeroPoint BaseRepeat

100
A0021335 A0021340

<

100
A0021336

ZeroPoint MeasValue 100


A0021337

Endress+Hauser

13

Proline Promass 84F

Example for maximum measured error

E [%] 2.5

2.0 1.5 1.0 0.5 0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 Q [%]


A0016708

E = Error: Maximum measured error as % o.r. (example) Q = Flow rate as %

14

Endress+Hauser

Proline Promass 84F

Installation
Mounting location Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors Avoid the following locations: Highest point of a pipeline. Risk of air accumulating. Directly upstream of a free pipe outlet in a vertical pipeline.

a0003605

Mounting location

The proposed configuration in the following diagram, however, permits installation in a vertical pipeline. Pipe restrictors or the use of an orifice plate with a smaller cross-section than the nominal diameter prevent the sensor from running empty during measurement.

3 4

5
a0003597

Installation in a down pipe (e.g. for batching applications) 1 2 3 4 5 Supply tank Sensor Orifice plate, pipe restriction (see Table following page) Valve Batching tank

Endress+Hauser

15

Proline Promass 84F

DN [mm] 8 15 25 40 50 80 100 150 250 [in] 1 1 2 3 4 6 10

Orifice plate, pipe restriction [mm] 6 10 14 22 28 50 65 90 150 [in] 0.24 0.40 0.55 0.87 1.10 2.00 2.60 3.54 5.91

Orientation

Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow (direction in which the fluid flows through the pipe). Vertical (Fig. V) Recommended orientation with upward direction of flow (Fig. V). When fluid is not flowing, entrained solids will sink down and gases will rise away from the measuring tube. The measuring tubes can be completely drained and protected against solids build-up. Horizontal (Fig. H1, H2) The measuring tubes must be horizontal and beside each other. When installation is correct the transmitter housing is above or below the pipe (Fig. H1, H2). Always avoid having the transmitter housing in the same horizontal plane as the pipe.
Orientation Vertical Horizontal, Transmitter head up Horizontal, Transmitter head down

a0004572

a0004576

a0004580

Fig. V

Fig. H1

Fig. H2

Standard version, Compact version Standard version, Remote version High-temperature, Compact version High-temperature, Remote version

1 TM > 200 C (> 392F) 1 TM > 200 C (> 392F)

2 2 2 2

= Recommended orientation; = Orientation recommended in certain situations; = Impermissible orientation

In order to ensure that the maximum permissible ambient temperature for the transmitter is not exceeded, we recommend the following orientations:
1

= For fluids with low temperatures, we recommend the horizontal orientation with the transmitter head pointing upwards (Fig. H1) or the vertical orientation (Fig. V).

2 = For fluids with high-temperatures, >200 C (>392 F), we recommend the horizontal orientation with the transmitter head pointing downwards (Fig. H2) or the vertical orientation (Fig. V).

16

Endress+Hauser

Proline Promass 84F

"

Caution! The two measuring tubes are slightly curved. The position of the sensor, therefore, has to be matched to the fluid properties when the sensor is installed horizontally .

a0004581

Installed horizontally 1 2 Not suitable for fluids with entrained solids. Risk of solids accumulating. Not suitable for outgassing fluids. Risk of air accumulating.

Installation instructions

Note the following points: No special measures such as supports are necessary. External forces are absorbed by the construction of the instrument, for example the secondary containment. The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by pipe vibrations. No special precautions need to be taken for fittings which create turbulence (valves, elbows, Tpieces, etc.), as long as no cavitation occurs. For mechanical reasons and in order to protect the pipe, it is advisable to support heavy sensors. Please refer to the verification ordinances for the installation conditions of the approval for custody transfer in question. The necessary steps for creating a measuring system and obtaining approval from the Standards Authorities must be clarified with the authority for legal metrology controls responsible.

Inlet and outlet runs Length of connecting cable

There are no installation requirements regarding inlet and outlet runs. max. 20 m (65 ft), remote version

Endress+Hauser

17

Proline Promass 84F

Special mounting instructions

Rupture disk Make sure that the function and operation of the rupture disk is not impeded through the installation of the device. The position of the rupture disk is indicated on a sticker beside it. For additional information that is relevant to the process ( 27). The existing connecting nozzles are not designed for a rinse or pressure monitoring function.

45

1
RUPTURE DISK

3 B 2
A0008361

DN 8 to DN 150(" to 6") 1 = Rupture disk label, 2 = Rupture disk with 1/2" NPT internal thread with 1" width across flat, 3 = Transport protection

Dimensions
DN [mm] 8 15 25 40 50 80 100 150 [in] 1 1 2 3 4 6 [mm] ca. 42 ca. 42 ca. 42 ca. 42 ca. 42 ca. 42 ca. 42 ca. 42 A [in] ca. 1.65 ca. 1.65 ca. 1.65 ca. 1.65 ca. 1.65 ca. 1.65 ca. 1.65 ca. 1.65 B [in] SW 1 SW 1 SW 1 SW 1 SW 1 SW 1 SW 1 SW 1 C [in] NPT NPT NPT NPT NPT NPT NPT NPT [mm] 62.0 62.0 62.0 67.0 79.0 101.0 120.0 141.0 D [in] 2.44 2.44 2.44 2.64 3.11 3.98 4.72 5.55 [mm] 216 220 260 310 452 560 684 880 E [in] 8.50 8.66 10.26 12.20 17.78 22.0 27.0 34.6

18

Endress+Hauser

Proline Promass 84F

D
i

A C

1
RUPTURE DISK

2
DN 250 (10")

B
A0009733

1 = Rupture disk label, 2 = Rupture disk with 1/2" NPT internal thread with 1" width across flat, 3 = Transport protection

Dimensions
DN [mm] 250 [in] 10 [mm] ca. 42 A [in] ca. 1.65 B [in] SW 1 C [in] NPT [mm] 182 D [in] 7.17 [mm] 380 E [in] 14.96

Zero point adjustment All measuring devices are calibrated to state-of-the-art technology. Calibration takes place under reference operating conditions 10. Consequently, the zero point adjustment is generally not necessary. Experience shows that the zero point adjustment is advisable only in special cases: To achieve highest measuring accuracy also with small flow rates Under extreme process or operating conditions (e.g. very high process temperatures or very highviscosity fluids).

Endress+Hauser

19

Proline Promass 84F

Environment
Ambient temperature range Sensor, transmitter Standard: 20 to +60 C (4 to +140 F) Optional: 40 to +60 C (40 to +140 F) Note! Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions. At ambient temperatures below 20 C (4 F) the readability of the display may be impaired. 40 to +80 C (40 to +175 F), preferably +20 C (+68 F) B, C, I Standard: IP 67 (NEMA 4X) for transmitter and sensor According to IEC 68-2-31 Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6 yes yes To ICE/EN 61326 and NAMUR recommendation NE 21

!
Storage temperature Ambient class Degree of protection Shock resistance Vibration resistance CIP cleaning SIP cleaning Electromagnetic compatibility (EMC)

Process
Medium temperature range Sensor 50 to +200 C (58 to +392 F) High-temperature version: 50 to +350 C (58 to +660 F) Medium density Medium pressure range (nominal pressure) 0 to 5000 kg/m(0 to 312 lb/ft) Flanges Standard version: according to DIN PN 16 to 100 according to ASME B16.5 Cl 150, Cl 300, Cl 600 JIS 10K, 20K, 40K, 63K High-temperature version: according to DIN PN 40, 64, 100 according to ASME B16.5 Cl 150, Cl 300, Cl 600 JIS 10K, 20K, 63K

20

Endress+Hauser

Proline Promass 84F

Secondary containment pressure rating The sensor housing is filled with dry nitrogen and protects the electronics and mechanics inside.
Secondary containment rating (designed with a safety factor 4) [in] 1 1 2 3 4 6 10 [bar] 40 40 40 40 40 25 16 16 10 [psi] 580 580 580 580 580 362 232 232 145 Burst pressure of secondary containment [bar] 255 200 280 180 195 105 85 80 57 [psi] 3695 2900 4060 3610 2825 1520 1230 1160 825

DN [mm] 8 15 25 40 50 80 100 150 250

Note! In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary containment in the event of tube failure can be bled off. This is especially important in high pressure gas applications. These connections can also be used for gas circulation and/or gas detection (Dimensions 31). Do not open the purge connections unless the containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum pressure: 5 bar (72.5 psi). If a device equipped with purge connections is connected to the purge system, the maximum pressure rating is defined by the purge system itself or the device, whichever is lower. If the device is equipped with a rupture disk, the maximum pressure rating is defined by the rupture disk properties ( 65).

Endress+Hauser

21

Proline Promass 84F

Pressure-temperature ratings

Warning! The following material load curves refer to the entire sensor and not just the process connection. Flange connection according to EN 1092-1 (DIN 2501) Flange material: 1.4404 (F316/F316L), Alloy C22

[psi] [bar] 100 1400 90 1300 80 1200 800 600 400 200 0 70 60 50 40 30 20 10 0

PN 100

PN 63

PN 40

PN 16 -50 0 0 80 50 160 100 240 150 320 200 400 250 480 300 560 350 [C] 640 [F]
A0022596-EN

-80

The values for the temperature range from +200 C to +350 C (+392 F to +662 F) are exclusively valid for the high-temperature version.

Flange connection according to ASME B16.5 Flange material: 1.4404 (F316/F316L), standard version

[psi] [bar] 1400 1200 1000 800 600 400 200 0 100 90 80 70 60 50 40 30 20 10 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0021017-EN

Class 600

Class 300

Class 150

22

Endress+Hauser

Proline Promass 84F

Flange material: 1.4404 (F316/F316L), high-temperature version

[psi] [bar] 1400 1300 1200 800 600 400 200 0 100 90 80 70 60 50 40 30 20 10 0 -50 -80 0 0 80 50 160 100 240 150 320 200 400 250 480 300 560 350 [C] 640 [F]
A0022598-EN

Class 600

Class 300

Class 150

The values for the temperature range are exclusively valid for high-temperature version.

Flange material: Alloy C22, high-temperature version

[psi] [bar] 1400 1300 1200 800 600 400 200 0 100 90 80 70 60 50 40 30 20 10 0 -50 -80 0 0 80 50 160 100 240 150 320 200 400 250 480 300 560 350 [C] 640 [F]
A0022597-EN

Class 600

Class 300

Class 150

The values for the temperature range are exclusively valid for high-temperature version.

Endress+Hauser

23

Proline Promass 84F

Flange connection to JIS B2220 Flange material: 1.4404 (F316/F316L), Alloy C22

[psi] [bar] 70 1200 60 800 50 600 40 400 200 0 30 20 10 0 -80

63K

40K 20K 10K -50 0 0 80 50 160 100 240 150 320 200 400 250 480 300 560 350 [C] 640 [F]
A0022599-EN

The values for the temperature range from +200 C to +350 C (+392 F to +662 F) are exclusively valid for high-temperature version.

Lap joint flange according to EN 1092-1 (DIN 2501) Flange material: 1.4301 (F304); wetted parts: Alloy C22

[psi] [bar] 600 400 200 0 40 30 20 10 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0021313-EN

PN 40

Lap joint flange according to ASME B16.5 Flange material: 1.4301 (F304); wetted parts: Alloy C22

[psi] [bar] 1400 1200 1000 800 600 400 200 0 100 90 80 70 60 50 40 30 20 10 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0021345-EN

Class 600

Class 300

Class 150

24

Endress+Hauser

Proline Promass 84F

Lap joint flange JIS B2220 Flange material: 1.4301 (F304); wetted parts: Alloy C22

[psi] [bar] 30 400 200 0 20 10 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0021346-EN

20K

Process connection to DIN 11851 Connection material: 1.4404 (316/316L)

[psi] [bar] 50 600 40 400 200 0 30 20 10 0

DN 840 DN 50100

-50 -50 0

50

100

150

200 [C]

50 100 150 200 250 300 350 400 [F]


A0021027-EN

DIN 11851 allows for applications up to +140 C (+284 F) if suitable sealing materials are used. Please take this into account when selecting seals and counterparts as these components can limit the pressure and temperature range.

Process connection to SMS 1145 Connection material: 1.4404 (316/316L)

[psi] [bar] 8 100 6 50 2 0 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0013056

SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please take this into account when selecting seals and counterparts as these components can limit the pressure and temperature range.

Tri-Clamp process connection The Clamp connections are suited up to a maximum pressure of 16 bar (232 psi). Please observe the operating limits of the clamp and seal used as they could be under 16 bar (232 psi). The clamp and the seal are not included in the scope of supply.

Endress+Hauser

25

Proline Promass 84F

Aseptic threaded connection to DIN 11864-1 Form A Connection material: 1.4404 (316/316L)

[psi] [bar] 50 600 40 400 200 0 30 20 10 0

DN 8...40 DN 50...100

-50 -50 0

50

100

150

200 [C]

50 100 150 200 250 300 350 400 [F]


a0021028-EN

Flange connection to DIN 11864-2 Form A (aseptic flat flange with groove) Flange material: 1.4404 (316/316L)

[psi] [bar] 40 400 200 0

30 20 10 0

DN 8...40 DN 50...100 -50 -50 0 0 50 50 100 150 200 [C]

100 150 200 250 300 350 400 [F]


A0004659

Threaded hygienic connection to ISO 2853 Connection material: 1.4404 (316/316L)

[psi] [bar] 30 400 200 0 20 10 0 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0020964-EN

26

Endress+Hauser

Proline Promass 84F

VCO process connection Connection material: 1.4404 (316/316L)

[psi] [bar] 1400 1200 1000 100 90 80 70 -50 -50 0 0 50 100 150 200 [C] 50 100 150 200 250 300 350 400 [F]
A0020963-EN

Rupture disk

To increase the level of safety, a device version with a rupture disk with a triggering pressure of 10 to 15 bar (145 to 217.5 psi) can be used. Special mounting instructions: ( 18). Rupture disks cannot be combined with the separately available heating jacket ( 71). See "Measuring range" section 5 Select nominal diameter by optimising between required flow range and permissible pressure loss. An overview of max. possible full scale values can be found in the "Measuring range" Section. The minimum recommended full scale value is approx. 1/20 of the max. full scale value. In most applications, 20 to 50% of the maximum full scale value can be considered ideal. Select a lower full scale value for abrasive substances such as fluids with entrained solids (flow velocity <1 m/s (<3 ft/s)). For gas measurement the following rules apply: Flow velocity in the measuring tubes should not be more than half the sonic velocity (0.5 Mach). The maximum mass flow depends on the density of the gas: formula 5

Limiting flow

Pressure loss System pressure

To calculate the pressure loss, use the Applicator sizing tool ( 71). It is important to ensure that cavitation does not occur, because it would influence the oscillation of the measuring tube. No special measures need to be taken for fluids which have properties similar to water under normal conditions. In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is important to ensure that pressure does not drop below the vapour pressure and that the liquid does not start to boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects can be prevented when system pressure is sufficiently high. Consequently, it is generally best to install the sensor: downstream from pumps (no danger of vacuum), at the lowest point in a vertical pipe.

Endress+Hauser

27

Proline Promass 84F

Thermal insulation

Some fluids require suitable measures to avoid loss of heat at the sensor. A wide range of materials can be used to provide the required thermal insulation.

max. 60 (2.4)

max. 60 (2.4)

a0004614

High-temperature version: maximum insulation thickness of 60 mm (2.4") in the area of the electronics/neck

If the Promass F high-temperature version is installed horizontally (with transmitter head pointing upwards), an insulation thickness of min. 10 mm (0.4") is recommended to reduce convection. The maximum insulation thickness of 60 mm (2.4") must be observed. Heating Some fluids require suitable measures to avoid loss of heat at the sensor. Heating can be electric, e.g. with heated elements, or by means of hot water or steam pipes made of copper or heating jackets. Caution! Risk of electronics overheating! Make sure that the maximum permissible ambient temperature for the transmitter is not exceeded. Consequently, make sure that the adapter between sensor and transmitter and the connection housing of the remote version always remain free of insulating material. Note that a certain orientation might be required, depending on the fluid temperature 20. With a fluid temperature between +200 C to +350 C (+392 F to +662 F) the remote version of the high-temperature version is preferable. When using electrical heat tracing whose heat is regulated using phase control or by pulse packs, it cannot be ruled out that the measured values are influenced by magnetic fields which may occur, (i.e. at values greater than those permitted by the EC standard (Sinus 30 A/m)). In such cases, the sensor must be magnetically shielded. The secondary containment can be shielded with tin plates or electric sheets without privileged direction (e.g. V330-35A) with the following properties: Relative magnetic permeability r 300 Plate thickness d 0.35 mm (d 0.014") Information on permissible temperature ranges 20 Special heating jackets which can be ordered as accessories from Endress+Hauser are available for the sensors.

"

28

Endress+Hauser

Proline Promass 84F

Custody transfer measurement


Promass 84 is a flowmeter suitable for custody transfer measurement for liquids (other than water) and gases. Custody transfer variables Mass flow Volume flow Density Promass 84 flowmeters are usually verified on site using reference measurements. Only once it has been verified on site by the Verification Authority for legal metrology controls may the measuring device be regarded as verified and used for applications subject to legal metrology controls. The associated seal (stamp) on the measuring device ensures this status. Caution! Only flowmeters verified by the Verification Authorities may be used for invoicing in applications subject to legal metrology controls. For all verification processes, both the corresponding approvals and the country-specific requirements resp. regulations (e.g. such as the German Verification Act) must be observed. The owner / user of the instrument is obliged to subsequent verification. Approval for custody transfer The requirements of the following legal metrology authorities are taken into consideration: PTB, Germany; (www.eichamt.de) NMi, Netherlands; (www.nmi.nl) METAS, Switzerland; (www.metas.ch) BEV, Austria; (www.bev.gv.at) NTEP, USA; (www.ncwm.net) MC, Canada; (www.ic.gc.ca) Switching on the power supply in custody transfer mode If the device is started in custody transfer mode, for example also after a power outage, system error No. 271 "POWER BRK. DOWN" flashes on the local display. The fault message can be acknowledged or reset using the "Enter" key or by means of the status input configured accordingly. Note! For correct measuring operation, it is not mandatory to reset the fault message. Type-approved measuring systems for liquids other than water are always verified at their place of deployment. For this purpose, the facility's owner-operator must make everything available when the Verification Authorities come to inspect and verify the system. This includes: Scales or container with a reading unit with a load or volumetric capacity that corresponds to the operation of the system at Qmax for one minute. The resolution of the scales display or the reading unit must be at least 0.1 % of the minimum measured quantity. Unit for removing the medium being measured after the totalizer to fill the scales or the container. Making a sufficient quantity of the medium being measured available. The quantity is derived from the operation of the system. The following rule of thumb applies - quantity at: 3 1 minute at Qmin, plus 3 1 minute at Qmax, plus 3 1 minute at Qmax, plus adequate quantity in reserve. Approval certificates Note! All issues should be clarified in advance with the authority responsible to ensure the successful verification of the measuring system. Setting up custody transfer mode A detailed description of the "setting up custody transfer mode" process is provided in the Operating Instructions supplied with the device.

Suitability for custody transfer, metrological control, obligation to subsequent verification

"

!
Verification (Example)

Endress+Hauser

29

Proline Promass 84F

Stamp points

t unter Spa Nich ffnen nnung


cover tight Keep w circuits are alive hile

over tigh Keep cits tw hl circu are aliv e ie

C
Examples of how to seal the various device versions.

Disabling custody transfer mode A detailed description of the "disabling custody transfer mode" process is provided in the Operating Instructions supplied with the device.

30

as ouvrir lapp Ne psous tensio n are

il

D
A0001778

Endress+Hauser

Proline Promass 84F

Mechanical construction
Design/dimensions
Abmessungen Field housing compact version, powder-coated die-cast aluminum Transmitter compact version, stainless steel Transmitter remote version, connection housing (II2G/Zone 1) Transmitter remote version, wall-mount housing (non Ex-zone and II3G/Zone 2) Sensor remote version, connection housing Sensor remote version, connection housing with an extended neck High-temperature version (compact) High-temperature version (remote) Process connections in SI units Flange connections EN (DIN) Flange connections ASME B16.5 Flange connections JIS B2220 Lap joint flange EN (DIN) Lap joint flange ASME B16.6 Lap joint flange JIS Tri-Clamp DIN 11851 (threaded hygienic connection) DIN 11864-1 Form A (aseptic threaded connection) DIN 11864-2 Form A (aseptic flat flange with groove) ISO 2853 (threaded hygienic connection) SMS 1145 (threaded hygienic conncetion) VCO connections Process connection in US units Flange connections ASME B16.5 Tri-Clamp SMS 1145 (Threaded hygienic conncetion) VCO connections Purge connections / secondary containment monitoring 57 62 63 64 65 41 43 45 48 48 49 50 51 52 53 54 55 56 32 34 35 36 37 38 39 40

Endress+Hauser

31

Proline Promass 84F

Field housing compact version, powder-coated die-cast aluminum

A A*

B C

D F

Esc

di

L
a0006996

Dimensions in SI units
DN 8 15 25 40 50 80 100 150 250 A 227 227 227 227 227 227 227 227 227 A* 207 207 207 207 207 207 207 207 207 B 187 187 187 187 187 187 187 187 187 C 168 168 168 168 168 168 168 168 168 D 160 160 160 160 160 160 160 160 160 E 75 75 75 105 141 200 254 378 548 F 266 266 266 271 283 305 324 362 390 G 341 341 341 376 424 505 578 740 938 L
1) 1) 1) 1) 1) 1) 1) 1) 1)

G
di
1) 1) 1) 1) 1) 1) 1) 1) 1)

* Blind version (without local display) 1) dependent on respective process connection All dimensions in [mm]

Dimensions in US units
DN " " 1" 1" 2" 3" 4" 6" 10" A 8.94 8.94 8.94 8.94 8.94 8.94 8.94 8.94 8.94 A* 8.15 8.15 8.15 8.15 8.15 8.15 8.15 8.15 8.15 B 7.68 7.68 7.68 7.68 7.68 7.68 7.68 7.68 7.68 C 6.61 6.61 6.61 6.61 6.61 6.61 6.61 6.61 6.61 D 6.30 6.30 6.30 6.30 6.30 6.30 6.30 6.30 6.30 E 2.95 2.95 2.95 4.13 5.55 7.87 10.0 14.9 21.6 F 10.5 10.5 10.5 10.7 11.1 12.0 12.8 14.3 15.4 G 13.4 13.4 13.4 14.8 16.7 19.9 22.8 29.1 36.9 L
1) 1) 1) 1) 1) 1) 1) 1) 1)

di
1) 1) 1) 1) 1) 1) 1) 1) 1)

*Blind version (without local display) 1) dependent on respective process connection All dimensions in [in]

!
32

Note! Dimensions for transmitters II2G/Zone 1 35.

Endress+Hauser

Proline Promass 84F

Field housing compact version (II2G/Zone 1), powder-coated die-cast aluminum

A A*
Nic
Non-intrinsically safe circuits Ip40 protected Boucles de courant sans scurit intrinsque protges par Ip40

B C
e r S p a nnu ht unt ng ffne n

o ve r tig ht wh ep c K e irc uits a re a liv ile c e

Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschtzt

D F
Ne

di
Dimension unit in mm (in)

L
A0021933

Dimensions in SI units
DN 8 15 25 40 50 80 100 150 250 A 240 240 240 240 240 240 240 240 240 A* 217 217 217 217 217 217 217 217 217 B 206 206 206 206 206 206 206 206 206 C 186 186 186 186 186 186 186 186 186 D 178 178 178 178 178 178 178 178 178 E 75 75 75 105 141 200 254 378 548 F 284 284 284 289 301 323 342 380 408 G 359 359 359 394 442 523 596 758 956 L
1) 1) 1) 1) 1) 1) 1) 1) 1)

G
di
1) 1) 1) 1) 1) 1) 1) 1) 1)

* Blind version (without local display) 1) dependent on respective process connection All dimensions in [mm]

Dimensions in US units
DN " " 1" 1" 2" 3" 4" 6" 10" A 9.45 9.45 9.45 9.45 9.45 9.45 9.45 9.45 9.45 A* 8.54 8.54 8.54 8.54 8.54 8.54 8.54 8.54 8.54 B 8.11 8.11 8.11 8.11 8.11 8.11 8.11 8.11 8.11 C 7.32 7.32 7.32 7.32 7.32 7.32 7.32 7.32 7.32 D 7.01 7.01 7.01 7.01 7.01 7.01 7.01 7.01 7.01 E 2.95 2.95 2.95 4.13 5.55 7.87 10.0 14.9 21.6 F 11.2 11.2 11.2 11.4 11.8 12.7 13.5 15.0 16.1 G 14.1 14.1 14.1 15.5 17.4 20.6 23.5 29.8 37.6 L
1) 1) 1) 1) 1) 1) 1) 1) 1)

*Blind version (without local display) 1) dependent on respective process connection All dimensions in [in]

!
Endress+Hauser

Note! Dimensions for transmitters II2G/Zone 1 35.

uv rir la p a s o te p p a re i l s o u s ns io n

di
1) 1) 1) 1) 1) 1) 1) 1) 1)

33

Proline Promass 84F

Transmitter compact version, stainless steel

a0002245

Dimensions in SI and US units


A [mm] 225 [in] 8.86 [mm] 153 B [in] 6.02 [mm] 168 C [in] 6.61

Transmitter compact version (II2G/Zone 1), stainless steel

A A*
Nic

B C
e r S p a nnu ht unt ng ffne n

o ve r tig ht wh ep c K e irc uits a re a liv ile c e

Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschtzt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans scurit intrinsque protges par Ip40

p c o v er tig ht w K e e irc uits a re ali hile ve c

Esc

Dimensions in SI and US units


A [mm] 227 [in] 8.94 [mm] 207 A* [in] 8.15 [mm] 187 B [in] 7.68 [mm] 168 C [in] 6.61 [mm] 160 D [in] 6.30

34

Ne

uv rir la p a s o te p p a re i l s o u s ns io n

A0022059

Endress+Hauser

Proline Promass 84F

Transmitter remote version, connection housing (II2G/Zone 1)

A A* B B*
Nic

C D
e r S p a nnu ht unt ng ffne n

o ve r tig ht wh ep c K e irc uits a re a liv ile c e

Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschtzt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans scurit intrinsque protges par Ip40

p c o v er tig ht w K e e irc uits a re ali hile ve c

Esc + E

E K J L M

G H
a0002128

Dimensions in SI units
A 265 A* 242 B 240 B* 217 C 206 D 186 E 178 F 8.6 (M8) G 100 H 130 J 100 K 144 L 170 M 348

* Blind version (without display) All dimensions in [mm]

Dimensions in US units
A 10.4 A* 9.53 B 9.45 B* 8.54 C 8.11 D 7.32 E 7.01 F 0.34 (M8) G 3.94 H 5.12 J 3.94 K 5.67 L 6.69 M 13.7

* Blind version (without display) All dimensions in [in]

Endress+Hauser

Ne

uv rir la p a s o te p p a re i l s o u s ns io n

35

Proline Promass 84F

Transmitter remote version, wall-mount housing (non Ex-zone and II3G/Zone 2)

Esc

D C B

F A H J E K

S R T P Q

L
a0001150

Dimensions (SI units)


A 215 L 95
1)

B 250 M 53

C 90.5 N 102

D 159.5 O 81.5

E 135 P 11.5

F 90 Q 192

G 45 R 8 M5

H > 50 S 20

J 81 T 1) 2 6.5

M
K 53 G 1.77 R 8 M5 H > 1.97 S 0.79 J 3.18 T
1)

Securing screw for wall mounting: M6 (screw head max. 10.5 mm) All dimensions in [mm]

Dimensions (US units)


A 8.46 L 3.74
1)

B 9.84 M 2.08

C 3.56 N 4.01

D 6.27 O 3.20

E 5.31 P 0.45

F 3.54 Q 7.55

N
K 2.08

2 0.26

Securing screw for wall mounting: M6 (screw head max. 0.41") All dimensions in [in]

36

Endress+Hauser

Proline Promass 84F

Sensor remote version, connection housing

a0002516

Dimensions in SI units
DN 8 15 25 40 50 80 100 150 250 All dimensions in [mm] A 118.5 118.5 118.5 118.5 118.5 118.5 118.5 118.5 118.5 B 137.5 137.5 137.5 137.5 137.5 137.5 137.5 137.5 137.5 C 113 113 113 118 130 152 171 209 237

Dimensions in US units
DN " " 1" 1" 2" 3" 4" 6" 10" All dimensions in [in] A 4.67 4.67 4.67 4.67 4.67 4.67 4.67 4.67 4.67 B 5.41 5.41 5.41 5.41 5.41 5.41 5.41 5.41 5.41 C 4.52 4.52 4.52 4.72 5.20 6.08 6.84 8.36 9.48

Endress+Hauser

37

Proline Promass 84F

Sensor remote version, connection housing with an extended neck Note! Use this version in case of insulation or application of heating jacket.

D B C E

a0002517

Dimensions in SI and US units


A [mm] 129 [in] 5.08 [mm] 80 B [in] 3.15 [mm] 110 C [in] 4.33 [mm] 102 D [in] 4.02 [mm] 292 E [in] 11.5

38

Endress+Hauser

Proline Promass 84F

High-temperature version (compact)

A B

a0002518

Dimensions in SI units
DN 25 50 80 All dimensions in [mm] A 187 187 187 B 168 168 168 C 100 141 200 D 350 365 385 E 450 506 585

Dimensions in US units
DN 1" 2" 3" All dimensions in [in] A 7.36 7.36 7.36 B 6.61 6.61 6.61 C 3.94 5.55 7.87 D 13.78 14.37 15.16 E 17.72 19.92 23.03

Endress+Hauser

39

Proline Promass 84F

High-temperature version (remote)


A

a0002519

Dimensions in SI units
DN 25 50 80 All dimensions in [mm] A 129 129 129 B 292 307 327 C 105 141 200 D 397 448 527

Dimensions in US units
DN 1" 2" 3" All diemensions in [in] A 5.08 5.08 5.08 B 11.50 12.09 12.87 C 4.13 5.55 7.87 D 15.63 17.64 20.75

40

Endress+Hauser

Proline Promass 84F

Process connections in SI units Flange connections EN (DIN), ASME B16.5, JIS

N
di L 2.0 (0.08)*
+1.5 (+0.06)* *DN 150250: 3.5 (6" 10": 0.14)
A0002501

Dimension unit in mm (in)

Flange connections EN (DIN)


Flange according to EN 1092-1 (DIN 2501 / DIN 2512N 1)) / PN 16: 1.4404 (F316/F316L) Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 m DN 100 150 2503)
1) 2)

G 220 285 405

L 1128/14002) 1330/17002) 1780

N 8 18 8 22 12 26

S 20 22 26

LK 180 240 355

U LK G
S
U 107.1 159.3 260.4 di 51.20 68.90 102.26 Flange with groove according to EN 1092-1 Form D (DIN 2512N) available Installation length in accordance with NAMUR recommendation NE 132 optionally available (order code for "Process connection", option D1N or D5N (with groove)) 3) Not available in Alloy All dimensions in [mm] Flange according to EN 1092-1 (DIN 2501 / DIN 2512N 1)) / PN 40: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 m DN 8
2)

G 95 95 115 150 165 200 235 300 450

L 370/510 404/510
3) 3)

N 4 14 4 14 4 14 4 18
3)

S 16 16 18 18 20 24 24 28 38

LK 65 65 85 110 125 160 190 250 385

U 17.3 17.3 28.5 43.1 54.5 82.5 107.1 159.3 258.8

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90 102.26

15 25 40 50 80 100 150 2504)


1) 2)

440/6003) 550 715/715

4 18 8 18 8 22 8 26 12 33

840/9153) 1128 1370 1850

Flange with groove according to EN 1092-1 Form D (DIN 2512N) available DN 8 standard with DN 15 flange 3) Installation length in accordance with NAMUR recommendation NE 132 optionally available (order code for "Process connection", option D2N or D6N (with groove)) 4) Not available in Alloy All dimensions in [mm]

Endress+Hauser

41

Proline Promass 84F

Flange according to EN 1092-1 (DIN 2501) / PN 40 (with DN 25-Flanges): 1.4404 (F316/F316L) Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 m DN 8 15 G 115 115 L 440 440 N 4 14 4 14 S 18 18 LK 85 85 U 28.5 28.5 di 5.35 8.30

All dimensions in [mm]

Flange according to EN 1092-1 (DIN 2501 / DIN 2512N ) extension-reduction / PN 16: 1.4404 (F316/ F316L) Only for nominal diameter DN 250 (on request) Surface roughness (flange): Ra 0.8 to 3.2 m DN 150 200 300 G 285 340 460 L 1980 1940 1940 N 8 22 12 22 12 26 S 22 24 28 LK 240 295 410 U 159.3 207.3 309.7 di 102.26 102.26 102.26

All dimensions in [mm]

Flange according to EN 1092-1 (DIN 2501 / DIN 2512N ) extension-reduction / PN 40: 1.4404 (F316/ F316L) Only for nominal diameter DN 250 (on request) Surface roughness (flange): Ra 0.8 to 3.2 m DN 150 200 300 G 300 375 515 L 1980 1940 1940 N 8 26 12 30 16 33 S 28 34 42 LK 250 320 450 U 159.3 206.5 307.9 di 102.26 102.26 102.26

All dimensions in [mm]

Flange according to EN 1092-1 (DIN 2501 / DIN 2512N ) / PN 63: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 m DN 50 80 100 150 250
1) 2) 2)

G 180 215 250 345 470

L 724 875 1128 1410 1890

N 4 22 8 22 8 26 8 33 12 36

S 26 28 30 36 46

LK 135 170 200 280 400

U 54.5 81.7 106.3 157.1 255.4

di 26.00 40.50 51.20 68.90 102.26

Flange with groove according to EN 1092-1 Form D (DIN 2512N) available Not available in Alloy All dimensions in [mm]

42

Endress+Hauser

Proline Promass 84F

Flange EN 1092-1 (DIN 2501 / DIN 2512N ) / PN 100: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 m DN 8
2)

G 105 105 140 170 195 230 265 355

L 400 420 470 590 740 885 1128 1450

N 4 14 4 14 4 18 4 22 4 26 8 26 8 30 12 33

S 20 20 24 26 28 32 36 44

LK 75 75 100 125 145 180 210 290

U 17.3 17.3 28.5 42.5 53.9 80.9 104.3 154.0

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90

15 25 40 50 80 100 150
2)

Flange with groove according to EN 1092-1 Form D (DIN 2512N) available DN 8 standard with DN 15 flange All dimensions in [mm]

Flange connections ASME B16.5


Flange according to ASME B16.5 / Cl 150: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN 8
2)

G 88.9 88.9 108.0 127.0 152.4 190.5 228.6 279.4 406.4

L 370.0 404.0 440.0 550.0 715.0 840.0 1128.0 1398.0 1836.8

N 4 15.7 4 15.7 4 15.7 4 15.7 4 19.1 4 19.1 8 19.1 8 22.4 12 25.4

S 11.2 11.2 14.2 17.5 19.1 23.9 23.9 25.4 30.2

LK 60.5 60.5 79.2 98.6 120.7 152.4 190.5 241.3 362.0

U 15.7 15.7 26.7 40.9 52.6 78.0 102.4 154.2 254.5

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90 102.26

15 25 40 50 80 100 150 2501)


1) 2)

Not available in Alloy DN 8 standard with DN 15 flange All dimensions in [mm]

Endress+Hauser

43

Proline Promass 84F

Flange according to ASME B16.5 / Cl 300: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN 8
2)

G 95.2 95.2 123.9 155.4 165.1 209.5 254.0 317.5 444.5

L 370.0 404.0 440.0 550.0 715.0 840.0 1128.0 1417.0 1868.2

N 4 15.7 4 15.7 4 19.0 4 22.3 8 19.0 8 22.3 8 22.3 12 22.3 16 28.4

S 14.2 14.2 17.5 20.6 22.3 28.4 31.7 36.5 47.4

LK 66.5 66.5 88.9 114.3 127.0 168.1 200.1 269.7 387.3

U 15.7 15.7 26.7 40.9 52.6 78.0 102.4 154.2 254.5

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90 102.26

15 25 40 50 80 100 150 2501)


1) 2)

Not available in Alloy DN 8 standard with DN 15 flange All dimensions in [mm]

Flange according to ASME B16.5 / Cl 600: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN 82) 15 25 40 50 80 100 150 250
1) 2) 1)

G 95.3 95.3 124.0 155.4 165.1 209.6 273.1 355.6 508.0

L 400.0 420.0 490.0 600.0 742.0 900.0 1158.0 1467.0 1951.2

N 4 15.7 4 15.7 4 19.1 4 22.4 8 19.1 8 22.4 8 25.4 12 28.4 16 35.1

S 20.6 20.6 23.9 28.7 31.8 38.2 48.4 47.8 69.9

LK 66.5 66.5 88.9 114.3 127.0 168.1 215.9 292.1 431.8

U 13.9 13.9 24.3 38.1 49.2 73.7 97.3 154.2 254.5

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90 102.26

Not available in Alloy DN 8 standard with DN 15 flange All dimensions in [mm]

Flange according to ASME B16.5 extension-reduction / Cl 150: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (auf Anfrage) Surface roughness (flange): Ra 3.2 to 6.3 m DN 150 200 300 G 279.4 342.9 482.6 L 1980 1940 1940 N 8 22.4 8 22.4 12 25.4 S 25.4 28.4 31.8 LK 241.3 298.5 431.8 U 154.2 202.7 304.80 di 102.26 102.26 102.26

All dimensions in [mm]

44

Endress+Hauser

Proline Promass 84F

Flange according to ASME B16.5 extension-reduction / Cl 300: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (auf Anfrage) Surface roughness (flange): Ra 3.2 to 6.3 m DN 150 200 300 G 317.5 381.0 520.7 L 1980 1940 1940 N 12 22.4 12 25.4 16 31.7 S 36.5 41.1 50.8 LK 269.7 330.2 450.8 U 154.2 202.7 304.80 di 102.26 102.26 102.26

All dimensions in [mm]

Flange according to ASME B16.5 extension-reduction / Cl 600: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (auf Anfrage) Surface roughness (flange): Ra 3.2 to 6.3 m DN 150 200 G 355.6 419.1 L 1980 1940 N 12 28.4 12 31.8 S 54.2 62.0 LK 292.1 349.3 U 154.2 202.7 di 102.26 102.26

All dimensions in [mm]

Flange connections JIS B2220


Flange JIS B2220 / 10K: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN 50 80 100 150 2501) G 155 185 210 280 400 L 715 832 1128 1354 1780 N 4 19 8 19 8 19 8 23 12 25 S 16 18 18 22 24 LK 120 150 175 240 355 U 50 80 100 150 250 di 26.00 40.50 51.20 68.90 102.26

1) Not available in Alloy All dimensions in [mm]

Endress+Hauser

45

Proline Promass 84F

Flange JIS B2220 / 20K: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 1.6 to 3.2 m DN 8
2)

G 95 95 125 140 155 200 225 305 430

L 370 404 440 550 715 832 1128 1386 1850

N 4 15 4 15 4 19 4 19 8 19 8 23 8 23 12 25 12 27

S 14 14 16 18 18 22 24 28 34

LK 70 70 90 105 120 160 185 260 380

U 15 15 25 40 50 80 100 150 250

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90 102.26

15 25 40 50 80 100 150 2501)


1) 2)

Not available in Alloy DN 8 standard with DN 15 flange All dimensions in [mm]

Flange JIS B2220 / 40K: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 1.6 to 3.2 m DN 81) 15 25 40 50 80 100 150
1)

G 115 115 130 160 165 210 250 355

L 400 425 485 600 760 890 1168 1498

N 4 19 4 19 4 19 4 23 8 19 8 23 8 25 12 33

S 20 20 22 24 26 32 36 44

LK 80 80 95 120 130 170 205 295

U 15 15 25 38 50 75 100 150

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90

All dimensions in [mm] DN 8 standard with DN 15 flange

Flange JIS B2220 / 63K: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 1.6 to 3.2 m DN 81) 15 25 40 50 80 100 150
1)

G 120 120 140 175 185 230 270 365

L 420 440 494 620 775 915 1168 1528

N 4 19 4 19 4 23 4 25 8 23 8 25 8 27 12 33

S 23 23 27 32 34 40 44 54

LK 85 85 100 130 145 185 220 305

U 12 12 22 35 48 73 98 146

di 5.35 8.30 12.00 17.60 26.00 40.50 51.20 68.90

All dimensions in [mm] DN 8 standard with DN 15 flange

46

Endress+Hauser

Proline Promass 84F

Flange JIS B2220 extension-reduction / 10K: 1.4404 (F316/F316L) Only for nominal diameter DN 250 (on request) Surface roughness (flange): Ra 1.6 to 3.2 m DN 150 200 300 G 280 330 445 L 1980 1940 1940 N 8 23 12 23 16 25 S 22 22 24 LK 240 290 400 U 150 200 300 di 102.26 102.26 102.26

All dimensions in [mm]

Flange JIS B2220 extension-reduction / 20K: 1.4404 (F316/F316L) Only for nominal diameter DN 250 (on request) Surface roughness (flange): Ra 1.6 to 3.2 m DN 150 200 300 G 305 350 480 L 1980 1940 1940 N 12 25 12 25 16 27 S 28 30 36 LK 260 305 430 U 150 200 300 di 102.26 102.26 102.26

All dimensions in [mm]

Endress+Hauser

47

Proline Promass 84F

Lap joint flange EN (DIN), ASME B16.5, JIS

B E F

C
D
+1,5 (+0.06)

L -2,0 (-0.08)
A0022221

Measuring unit: mm (in)

Lap joint flange EN (DIN)


Lap joint flange according to EN 1092-1 (DIN 2501/DIN 2512N) PN 40: 1.4301 (F304) (order code "Process connection", option DAC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN 8 2) 15 25 40 50 80 100
1) 2)

A 95 95 115 150 165 200 235

B 65 65 85 110 125 160 190

C 4 14 4 14 4 14 4 18 4 18 8 18 8 22

D 14.5 14.5 16.5 21.0 23.0 29.0 34.0

E 45 45 68 88 102 138 162

F 17.3 17.3 28.5 41.3 54.5 82.5 107.1

L 370.0 404.0 444.0 560.0 719.0 848.0 1132

Ldiff 1) 0 0 +4 +10 +4 +8 +4

Difference to the installation length of the welding neck flange (order code "Process connection", option D2C) DN 8 standard with DN 15 flange All dimensions in [mm]

Lap joint flange ASME B16.6


Lap joint flange according to ASME B16.5 Cl 150: 1.4301 (F304) (order code "Process connection", option ADC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN 82) 15 25 40 50 80 100
1) 2)

A 88.9 88.9 108.0 127.0 152.4 190.5 228.6

B 60.5 60.5 79.2 98.6 120.7 152.4 190.5

C 4 15.7 4 15.7 4 15.7 4 15.7 4 19.1 8 19.1 8 19.1

D 15.0 15.0 16.0 15.9 19.0 22.3 26.0

E 35.1 35.1 50.8 73.2 91.9 127.0 157.2

F 15.7 15.7 26.7 40.9 52.6 78.0 102.4

L 370.0 404.0 444.0 550.0 715.0 840.0 1128

Ldiff 1) 0 0 0 0 0 0 0

Difference to the installation length of the welding neck flange (order code "Process connection", option AAC) DN 8 standard with DN 15 flange All dimensions in [mm]

48

Endress+Hauser

Proline Promass 84F

Lap joint flange according to ASME B16.5 Cl 300: 1.4301 (F304) (order code "Process connection", option AEC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN 82) 15 25 40 50 80 100
1) 2)

A 95.2 95.2 123.9 155.4 165.1 209.5 254.0

B 66.5 66.5 88.9 114.3 127.0 168.1 200.1

C 4 15.7 4 15.7 4 19.1 4 22.3 8 19.1 8 22.3 8 22.3

D 16.5 16.5 21.0 23.0 25.5 31.0 32.0

E 35.1 35.1 50.8 73.2 91.9 127.0 157.2

F 15.7 15.7 26.7 40.9 52.6 78.0 102.4

L 376.0 406.0 450.0 564.0 717.0 852.6 1140

Ldiff 1) +6 +2 +10 +14 +2 +12.6 +12

Difference to the installation length of the welding neck flange (order code "Process connection". option ABC) DN 8 standard with DN 15 flange All dimensions in [mm]

Lap joint flange according to ASME B16.5 Cl 600: 1.4301 (F304) (order code "Process connection", option AFC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN 8 2) 15 25 40 50 80 100
1) 2)

A 95.2 95.2 123.9 155.4 165.1 209.6 273.1

B 66.5 66.5 88.9 114.3 127.0 168.1 215.9

C 4 15.7 4 15.7 4 19.1 4 22.3 4 19.1 8 22.3 8 25.4

D 17.0 17.0 21.5 25.0 28.0 35.0 44.0

E 35.1 35.1 50.8 73.2 91.9 127.0 157.2

F 13.9 13.9 24.3 38.1 49.2 73.7 97.3

L 400 420 490 600 742 900 1168

Ldiff 1) 0 0 0 0 0 0 +10

Difference to the installation length of the welding neck flange (order code "Process connection", option ACC) DN 8 standard with DN 15 flange All dimensions in [mm]

Lap joint flange JIS


Lap joint flange JIS B2220 20K: 1.4301 (F304) (order code "Process connection", option NIC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN 8 2) 15 25 40 50 80 100
1) 2)

A 95 95 125 140 155 200 225

B 70 70 90 105 120 160 185

C 4 15 4 15 4 19 4 19 8 19 8 23 8 23

D 14.0 14.0 18.5 18.5 23.0 29.0 29.0

E 51 51 67 81 96 132 160

F 15 15 25 40 50 80 100

L 370 404 440 550 715 844 1128

Ldiff 1) 0 0 0 0 0 +12 +0

Difference to the installation length of the welding neck flange (order code "Process connection", option NEC) DN 8 standard with DN 15 flange All dimensions in [mm]

Endress+Hauser

49

Proline Promass 84F

Tri-Clamp

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022593

Dimension unit in mm (in)

Tri-Clamp: 1.4404 (316/316L) DN 8 15 25 40 50 80 100 Clamp 1" 1" 1" 1" 2" 3" 4" G 50.4 50.4 50.4 50.4 63.9 90.9 118.9 L 367 398 434 560 720 900 1128 U 22.1 22.1 22.1 34.8 47.5 72.9 97.4 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [mm]

"-Tri-Clamp: 1.4404 (316/316L) DN 8 15 Clamp " " G 25.0 25.0 L 367 398 U 9.5 9.5 di 5.35 8.30

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [mm]

50

Endress+Hauser

Proline Promass 84F

DIN 11851 (threaded hygienic connection)

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022592

Dimension unit in mm (in)

Threaded hygienic connection DIN 11851: 1.4404 (316/316L) DN 8 15 25 40 50 80 100 G Rd 34 " Rd 34 " Rd 52 1/6" Rd 65 1/6" Rd 78 1/6" Rd 110 1/4" Rd 130 1/4" L 367 398 434 560 720 900 1128 U 16 16 26 38 50 81 100 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

3A version also available (Ra 0.8 m/150 grit) All dimensions in [mm]

Endress+Hauser

51

Proline Promass 84F

DIN 11864-1 Form A (aseptic threaded connection)

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022591

Dimension unit in mm (in)

Aseptic threaded connection DIN 11864-1 Form A: 1.4404 (316/316L) DN 8 15 25 40 50 80 100 G Rd 28 " Rd 34 " Rd 52 1/6" Rd 65 1/6" Rd 78 1/6" Rd 110 1/4" Rd 130 1/4" L 367 398 434 560 720 900 1128 U 10 16 26 38 50 81 100 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

3A version also available (Ra 0.8 m/150 grit) All dimensions in [mm]

52

Endress+Hauser

Proline Promass 84F

DIN 11864-2 Form A (aseptic flat flange with groove)

N
A A di L 2.0 (0.08)
+1.5 (+0.06)

U LK G

A0022590

Detail A: The flange has the smaller groove for the O-ring on the sensor side. When mounting the sensor the corresponding flange must have accordingly a larger groove. Dimension unit in mm (in)

DIN 11864-2 Form A (aseptic flat flange with groove): 1.4404 (316/316L) DN 8 15 25 40 50 80 100 G 54 59 70 82 94 133 159 L 387 418 454 560 720 900 1128 N 4 9 4 9 4 9 4 9 4 9 8 11 8 11 S 10 10 10 10 10 12 14 LK 37 42 53 65 77 112 137 U 10 16 26 38 50 81 100 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [mm]

Endress+Hauser

53

Proline Promass 84F

ISO 2853 (threaded hygienic connection)

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022589

Dimension unit in mm (in)

Threaded hygienic connection ISO 2853: 1.4404 (316/316L) DN 8 15 25 40 50 80 100 G 37.13 37.13 37.13 52.68 64.16 91.19 118.21 L 367 398 434 560 720 900 1128 U 22.6 22.6 22.6 35.6 48.6 72.9 97.6 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

Max. thread diameter to ISO 2853 Annex A 3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit)) All dimensions in [mm]

54

Endress+Hauser

Proline Promass 84F

SMS 1145 (threaded hygienic conncetion)

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022588

Dimension unit in mm (in)

Threaded hygienic conncetions SMS 1145: 1.4404 (316/316L) DN 8 15 25 40 50 80 100 G Rd 40 x 1/6" Rd 40 x 1/6" Rd 40 x 1/6" Rd 60 x 1/6" Rd 70 x 1/6" Rd 98 x 1/6" Rd 132 x 1/6" L 367 398 434 560 720 900 1128 U 22.6 22.6 22.6 35.6 48.6 72.9 97.6 di 5.35 8.30 12.00 17.60 26.00 40.50 51.20

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [mm]

Endress+Hauser

55

Proline Promass 84F

VCO connections

-Esc + E

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A002257

Dimension unit in mm (in)

8-VCO-4 ("): 1.4404 (316/316L) DN 8 All dimensions in [mm] G AF 1" L 390 U 10.2 di 5.35

12-VCO-4 ("): 1.4404 (316/316L) DN 15 All dimensions in [mm] G AF 1" L 430 U 15.7 di 8.30

56

Endress+Hauser

Proline Promass 84F

Process connections in US units Flange connections ASME B16.5

N
di L 2.0 (0.08)*
+1.5 (+0.06)* *DN 150250: 3.5 (6" 10": 0.14)
A0002501

Dimension unit in mm (in)

Flange according to ASME B16.5 / Cl 150: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN " 2) " 1" 1" 2" 3" 4" 6" 10"
1) 2) 1)

G 3.50 3.50 4.25 5.00 6.00 7.50 9.00 11.0 16.0

L 14.6 15.9 17.3 21.7 28.1 33.1 44.4 55.0 72.3

N 4 0.62 4 0.62 4 0.62 4 0.62 4 0.75 4 0.75 8 0.75 8 0.88 12 1.0

S 0.44 0.44 0.56 0.69 0.75 0.94 0.94 0.99 1.19

LK 2.38 2.38 3.12 3.88 4.75 6.00 7.50 9.50 14.25

U 0.62 0.62 1.05 1.61 2.07 3.07 4.03 6.07 10.0

U LK G
S
di 0.21 0.33 0.47 0.69 1.02 1.59 2.01 2.71 4.03 Not available in Alloy DN " standard with DN " ange All dimensions in [in]

Endress+Hauser

57

Proline Promass 84F

Flange according to ASME B16.5 / Cl 300: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN "
2)

G 3.75 3.75 4.88 6.12 6.50 8.25 10.00 12.50 17.50

L 14.57 15.91 17.32 21.65 28.15 33.07 44.41 55.79 73.55

N 4 0.62 4 0.62 4 0.75 4 0.88 8 0.75 8 0.88 8 0.88 12 0.88 16 1.12

S 0.56 0.56 0.69 0.81 0.88 1.12 1.25 1.44 1.87

LK 2.62 2.62 3.50 4.50 5.00 6.62 7.88 10.62 15.25

U 0.62 0.62 1.05 1.61 2.07 3.07 4.03 6.07 10.02

di 0.21 0.33 0.47 0.69 1.02 1.59 2.02 2.71 4.03

" 1" 1" 2" 3" 4" 6" 10"1)


1) 2)

Not available in Alloy DN " standard with DN " ange All dimensions in [in]

Flange according to ASME B16.5 / Cl 600: 1.4404 (F316/F316L), Alloy C22 Surface roughness (flange): Ra 3.2 to 6.3 m DN " 2) " 1" 1" 2" 3" 4" 6" 10"
1) 2) 1)

G 3.75 3.75 4.88 6.12 6.50 8.25 10.75 14.00 20.00

L 15.75 16.54 19.29 23.62 29.21 35.43 45.59 57.76 76.82

N 4 0.62 4 0.62 4 0.75 4 0.88 8 0.75 8 0.88 8 1.00 12 1.12 16 1.38

S 0.81 0.81 0.94 1.13 1.25 1.50 1.91 1.88 2.75

LK 2.62 2.62 3.50 4.50 5.00 6.62 8.50 11.50 17.00

U 0.55 0.55 0.96 1.50 1.94 2.90 3.83 6.07 10.02

di 0.21 0.33 0.47 0.69 1.02 1.59 2.02 2.71 4.03

Not available in Alloy DN " standard with DN " ange All dimensions in [in]

Flange according to ASME B16.5 extension-reduction / Cl 150: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (on request) Surface roughness (flange): Ra 3.2 to 6.3 m DN 6" 8" 12" G 11.00 13.50 19.00 L 77.95 76.38 76.38 N 8 22.4 8 22.4 12 25.4 S 1.00 1.12 1.25 LK 9.50 11.75 17.00 U 6.07 7.98 12.00 di 4.03 4.03 4.03

All dimensions in [in]

58

Endress+Hauser

Proline Promass 84F

Flange according to ASME B16.5 extension-reduction / Cl 300: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (on request) Surface roughness (flange): Ra 3.2 to 6.3 m DN 6" 8" 12" G 12.5 15.0 20.5 L 78.0 76.4 76.4 N 12 0.88 12 1.00 16 1.25 S 1.44 1.62 2.00 LK 10.6 13.0 17.7 U 6.07 7.98 12.0 di 4.03 4.03 4.03

All dimensions in [in]

Flange according to ASME B16.5 extension-reduction / Cl 600: 1.4404 (F316/F316L) Only for nominal diameter DN 250 /10" (on request) Surface roughness (flange): Ra 3.2 to 6.3 m DN 6" 8" G 14.0 16.5 L 78.0 76.4 N 12 1.12 12 1.25 S 2.13 2.44 LK 11.5 13.7 U 6.07 7.98 di 4.03 4.03

All dimensions in [in]

Endress+Hauser

59

Proline Promass 84F

Lap joint flange ASME B16.5

B E F

C
D
+1,5 (+0.06)

L -2,0 (-0.08)
A0022221

Measuring unit: mm (in)

Lap joint flange according to ASME B16.5 Cl 150: 1.4301 (F304) (order code "Process connection", option ADC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN " 2) " 1" 1" 2" 3" 4"
1) 2)

A 3.50 3.50 4.25 5.00 6.00 7.50 9.00

B 2.38 2.38 3.12 3.88 4.75 6.00 7.50

C 4 0.62 4 0.62 4 0.62 4 0.62 4 0.75 8 0.75 8 0.75

D 0.59 0.59 0.63 0.63 0.75 0.88 1.02

E 1.38 1.38 2.00 2.88 3.62 5.00 6.19

F 0.62 0.62 1.05 1.61 2.07 3.07 4.03

L 14.57 15.91 17.32 21.65 28.15 33.07 44.41

Ldiff 1) 0 0 0 0 0 0 0

Difference to the installation length of the welding neck flange (order code "Process connection", option AAC) DN " standard with DN " ange All dimensions in [in]

Lap joint flange according to ASME B16.5 Cl 300: 1.4301 (F304) (order code "Process connection", option AEC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN " 2) " 1" 1" 2" 3" 4"
1) 2)

A 3.75 3.75 4.88 6.12 6.50 8.25 10.0

B 2.62 2.62 3.50 4.50 5.00 6.62 7.88

C 4 0.62 4 0.62 4 0.75 4 0.88 4 0.75 8 0.88 8 0.88

D 0.65 0.65 0.83 0.91 1.00 1.22 1.26

E 1.38 1.38 2.00 2.88 3.62 5.00 6.19

F 0.62 0.62 1.05 1.61 2.07 3.07 4.03

L 14.80 15.98 17.72 22.22 28.23 33.57 44.88

Ldiff 1) +0.23 +0.07 +0.40 +0.55 +0.08 +0.50 +0.47

Difference to the installation length of the welding neck flange (order code "Process connection", option ABC) DN " standard with DN " ange All dimensions in [in]

60

Endress+Hauser

Proline Promass 84F

Lap joint flange according to ASME B16.5 Cl 600: 1.4301 (F304) (order code "Process connection", option AFC); wetted parts: Alloy C22 Surface roughness (flange): Ra 3.2 to 12.5 m DN " 2) " 1" 1" 2" 3" 4"
1) 2)

A 3.75 3.75 4.88 6.12 6.50 8.25 10.75

B 2.62 2.62 3.50 4.50 5.00 6.62 8.50

C 4 0.62 4 0.62 4 0.75 4 0.88 4 0.75 8 0.88 8 1

D 0.67 0.67 0.85 0.98 1.10 1.38 1.73

E 1.38 1.38 2.00 2.88 3.62 5.00 6.19

F 0.55 0.55 0.96 1.50 1.94 2.90 3.83

L 15.75 16.54 19.29 23.62 29.21 35.43 45.98

Ldiff 1) 0 0 0 0 0 0 +0.39

Difference to the installation length of the welding neck flange (order code "Process connection", option ACC) DN " standard with DN " ange All dimensions in [in]

Endress+Hauser

61

Proline Promass 84F

Tri-Clamp

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022593

Dimension unit in mm (in)

Tri-Clamp: 1.4404 (316/316L) DN " " 1" 1" 2" 3" 4" Clamp 1" 1" 1" 1" 2" 3" 4" G 1.98 1.98 1.98 1.98 2.52 3.58 4.68 L 14.4 15.7 17.1 22.0 28.3 35.4 44.4 U 0.87 0.87 0.87 1.37 1.87 2.87 3.83 di 0.21 0.33 0.47 0.69 1.02 1.59 2.01

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [in]

"-Tri-Clamp: 1.4404 (316/316L) DN " " Clamp " " G 0.98 0.98 L 14.4 15.7 U 0.37 0.37 di 0.21 0.33

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [in]

62

Endress+Hauser

Proline Promass 84F

SMS 1145 (Threaded hygienic conncetion)

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022588

Dimension unit in mm (in)

Threaded hygienic connection SMS 1145: 1.4404 (316/316L) DN " " 1" 1" 2" 3" 4" G Rd 40 1/6" Rd 40 1/6" Rd 40 1/6" Rd 60 1/6" Rd 70 1/6" Rd 98 1/6" Rd 132 1/6" L 14.68 15.92 17.36 22.40 28.80 36.00 45.12 U 0.904 0.904 0.904 1.424 1.944 2.916 3.904 di 0.214 0.332 0.480 0.704 0.104 1.620 2.048

3A version also available (Ra 0.8 m/150 grit, option: Ra 0.4 m/240 grit) All dimensions in [in]

Endress+Hauser

63

Proline Promass 84F

VCO connections

-Esc + E

di

L 2.0 (0.08)

+1.5 (+0.06)

U G

A0022587

Dimension unit in mm (in)

8-VCO-4 ("): 1.4404 (316/316L) DN " All dimensions in [in] G AF 1" L 15.35 U 0.40 di 0.21

12-VCO-4 ("): 1.4404 (316/316L) DN " All dimensions in [in] G AF 1" L 16.93 U 0.62 di 0.33

64

Endress+Hauser

Proline Promass 84F

"

Purge connections / secondary containment monitoring Caution! Purge connections or secondary containment monitoring can not be combined with separately available heating jacket. Dimensions (not available for the Promass F high-temperature version)

45

a0002537

DN 8 bis DN 150 (" to 6")

DN [mm] 8 15 25 40 50 80 100 150 [in] 1 1 2 3 4 6

G [mm] "-NPT "-NPT "-NPT "-NPT "-NPT "-NPT "-NPT "-NPT 62 62 62 67 79 101 120 141

H [in] 2.44 2.44 2.44 2.64 3.11 3.98 4.72 5.55 [mm] 216 220 260 310 452 560 684 880

L [in] 8.50 8.66 10.24 12.20 17.78 22.0 27.0 34.6

H L
a0009734

DN 250 (10")

DN [mm] 250 [in] 10

G [mm] "-NPT 182

H [in] 7.17 [mm] 380

L [in] 14.96

Endress+Hauser

65

Proline Promass 84F

Weight

Compact version: see table below Remote version Sensor: see table below Wall-mount housing: 5 kg (11 lbs) Weight information in SI units
DN [mm] Compact version High-temperature compact version Compact version Ex d Remote version High-temperature remote version
1)

8 11 20 9

15 12 21 10

25 14 14.7 23 12 13.5

40 19 28 17

50 30 30.7 39 28 29.5

80 55 55.7 64 53 54.5

100 96 105 94

150 154 163 152

250 1) 400 409 398

With 10" according to ASME B16.5 Cl 300 flanges All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg].

Weight information in US
DN [in] Compact version High-temperature compact version Compact version Ex d Remote version High-temperature remote version 24 44 20 26 46 22 1 31 32 51 26 29 1 42 62 37 2 66 68 86 62 65 3 121 123 141 117 120 4 212 232 207 6 339 359 335 10 1) 882 902 877

1) With 10" according to ASME B16.5 Cl 300 flanges All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [lbs].

Material

Transmitter housing Compact version Compact version: powder coated die-cast aluminium Stainless steel housing: stainless steel 1.4301/ASTM 304 Window material: glass or polycarbonate Remote version Remote field housing: powder coated die-cast aluminium Wall-mount housing: powder coated die-cast aluminium Window material: glass Sensor housing / containment Acid and alkali-resistant outer surface Stainless steel 1.4307 (304L) Connection housing, sensor (remote version) Stainless steel 1.4301 (304) (standard) Powder coated die-cast aluminium (high-temperature version and version for heating)

66

Endress+Hauser

Proline Promass 84F

Process connections Stainless steel 1.4404 (F316/F316L) Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220 Stainless steel 1.4404 (316/316L) DIN 11864-2 Form A (aseptic flat flange with groove) Threaded hygienic connection: DIN 11851 SMS 1145 ISO 2853 DIN 11864-1 Form A Tri-Clamp (OD-Tubes) VCO connections Stainless Steel 1.4301 (F304), wetted parts: Alloy C22 Lap joint flange according to EN 1092-1 (DIN 2501) /according to ASME B16.5 / JIS B2220 Alloy C22, 2.4602 (UNS N06022) Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220 High-temperature version Stainless steel 1.4404 (316/316L) Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220 Alloy C22, 2.4602 (UNS N06022) Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220 Measuring tubes DN 8 to 100 (" to 4"): Stainless Steel 1.4539 (904L); manifold: 1.4404 (316/316L) DN 150 (6"): Stainless Steel 1.4404 (316/316L); manifold: 1.4404 (316/316L) DN 250 (10"): Stainless Steel 1.4404 (316/316L); manifold: CF3M/316L DN 8 to 150 (" to 6"): Alloy C22, 2.4602 (UNS N06022); manifold: Alloy C22, 2.4602 (UNS N06022)

High-temperature version DN 25, 50, 80: Alloy C22, 2.4602 (UNS N06022) Process connections Flanges according to EN 1092-1 (DIN 2501), according to ASME B16.5, JIS B2220, VCO connections Sanitary connections: Tri-Clamp, threaded hygienic connections (DIN 11851, SMS 1145, ISO 2853, DIN 11864-1 Form A), flange to DIN 11864-2 Form A (aseptic flat flange with groove)

Operability
Local operation Display elements Liquid crystal display: illuminated, four lines with 16 characters per line Selectable display of different measured values and status variables At ambient temperatures below 20 C (4 F) the readability of the display may be impaired. Operating elements Local operation with three optical sensors () Application specific Quick Setup menus for straightforward commissioning

Endress+Hauser

67

Proline Promass 84F

Language groups

Language groups available for operation in different countries: Western Europe and America (WEA): English, German, Spanish, Italian, French, Dutch and Portuguese Eastern Europe and Scandinavia (EES): English, Russian, Polish, Norwegian, Finnish, Swedish and Czech South and east Asia (SEA): English, Japanese, Indonesian China (CIN): English, Chinese You can change the language group via the operating program "FieldCare".

Remote operation

Operation via HART, Modbus RS485

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. The measuring system meets the EMC requirements of the Australian Communication and Media Authority (ACMA). Information about currently available Ex versions (ATEX, FM, CSA, IECEx, NEPSI) can be supplied by your Endress+Hauser Sales Centre on request. All explosion protection data are given in a separate documentation which is available upon request. 3A approval EHEDG tested The measuring device meets all the requirements of the Modbus/TCP conformity and integration test and has the "Modbus/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed all the test procedures carried out and is certified by the "Modbus/TCP Conformance Test Laboratory" of the University of Michigan. The measuring devices can be ordered with or without PED (Pressure Equipment Directive). If a device with PED is required, this must be ordered explicitly. For devices with nominal diameters less than or equal to DN 25 (1"), this is neither possible nor necessary. With the identification PED/G1/III on the sensor nameplate, Endress+Hauser confirms conformity with the "Basic safety requirements" of Appendix I of the Pressure Equipment Directive 97/23/EC. Devices with this identification (with PED) are suitable for the following types of fluid: Fluids of Group 1 and 2 with a steam pressure greater than, or smaller and equal to 0.5 bar (7.3 psi) Unstable gases Devices without this identification (without PED) are designed and manufactured according to good engineering practice. They correspond to the requirements of Art. 3, Section 3 of the Pressure Equipment Directive 97/23/EC. Their application is illustrated in Diagrams 6 to 9 in Appendix II of the Pressure Equipment Directive 97/23/EC. Measuring Instruments Directive 2004/22/EG (MID) Annex MI-002 (gas meter) The measuring device is approved as gas meter for use under legal control (in commercial transactions) acc. the European Measuring Instruments Directive, Annex MI-002 (DE-08-MI002-PTB007).

C-tick mark

Ex approval

Sanitary compatibility

Modbus RS485 certification

Pressure measuring device approval

Measuring Instruments Directive

68

Endress+Hauser

Proline Promass 84F

Annex MI-005 (liquids other than water) This flowmeter which is a suitable component in measuring systems subject to legal metrology controls in accordance with Annex MI-005 of the European Measuring Instruments Directive 2004/ 22/EC (MID) Note: According to the Measuring Instruments Directive, however, only the complete measuring system is licensable, covered by an EC type-examination certificate and bears conformity marking. This flowmeter is qualified to OIML R117-1 and has an MID Evaluation Certificate (1) which confirms compliance with the essential requirements of the Measuring Instruments Directive. The Evaluation Certificate results from the WELMEC (cooperation between the legal metrology services of the member states of the European Union and EFTA) towards voluntary modular approval for measuring systems in accordance with Annex MI-005 (measuring systems for the continuous and dynamic measurement of quantities of liquids other than water) of the Measuring Instruments Directive 2004/22/EC. Approval for custody transfer Promass 84 is a flowmeter suitable for custody transfer measurement for liquids (other than water) and for fuel gases under high pressure (<100 bar (<1450 psi)). The requirements of the following test centres are taken into consideration: PTB, Germany NMi, The Netherlands METAS, Switzerland BEV, Austria NTEP, USA MC, Canada Information on custody transfer measurement 29 Suitability for custody transfer measurement PTB/NMi/METAS/BEV approval PTB, METAS and BEV approval for determining the mass and volume of liquids, other than water, and of fuel gases.
Sensor Promass F [mm] 8 to 250 DN [in] to 10 Mass YES PTB /NMi/METAS /BEV approval For liquids other than water Volume YES Density YES High pressure gas (CNG) Mass NO

MID approval, Annex MI-002 (gas meter) The device is qualified to OIML R137/D11.
Sensor Promass F [mm] 8 to 250 DN [in] to 10 MID Type Examination Certificate MI-002 (Europe) Fuel gases up to 100 bar (1450 psi) Mass YES Volume YES* Density NO

* at pure gases only (invariable gas density)

MID approval, Annex MI-005 (liquids other than water) The device is qualified to OIML R117-1.
Sensor Promass F [mm] 8 to 250 DN [in] to 10 OIML R117-1/MID Evaluation Certificate (Europe) Liquids other than water Mass YES Volume YES Density YES

Endress+Hauser

69

Proline Promass 84F

NTEP approval The measuring instrument is qualified in accordance with the National Type Evaluation Program (NTEP) Handbook 44 ("Specifications and Tolerances and other Technical Requirements for Weighing and Measuring Devices").
Sensor Promass F [mm] 15 to 150 DN [in] to 6 Mass YES NTEP approval Liquids other than water Volume YES High-pressure gas (CNG) Mass NO

MC approval The measuring instrument is qualified in accordance with "The Draft Ministerial Specifications Mass Flow Meters" (1993-09-21).
Sensor Promass F [mm] 8 to 150 DN [in] to 6 Mass YES MC approval Liquids other than water Volume YES

Other standards and guidelines

EN 60529 Degrees of protection by housing (IP code) IEC/EN 60068-2-6 Environmental influences: Test procedure - Test Fc: vibrate (sinusoidal) IEC/EN 60068-2-31 Environmental influences: Test procedure - Test Ec: shocks due to rough handling, primarily for devices EN 61010-1 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures IEC/EN 61326 "Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements) NAMUR NE 21 Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output signal NAMUR NE 53 Software of field devices and signal-processing devices with digital electronics NACE MR 103 Materials resistant to sulfide stress cracking in corrosive petroleum refining environments NACE MR 0175/ISO 15156-1 Materials for use in H2S-containing Environments in Oil and Gas Production

70

Endress+Hauser

Proline Promass 84F

Ordering Information
Detailed ordering information is available from the following sources: In the Product Configurator on the Endress+Hauser website: www.endress.com Select country Instruments Select device Product page function: Congure this product From your Endress+Hauser Sales Center: www.endress.com/worldwide Note! Product Configurator - the tool for individual product configuration Up-to-the-minute configuration data Depending on the device: Direct input of measuring point-specific information such as measuring range or operating language Automatic verification of exclusion criteria Automatic creation of the order code and its breakdown in PDF or Excel output format Ability to order directly in the Endress+Hauser Online Shop

Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com. Device-specific accessories For the Transmitter
Accessories Mounting set for transmitter Description Mounting set for wall-mount housing (remote version). Suitable for: Wall mounting Pipe mounting Installation in control panel Mounting set for aluminum field housing: Suitable for pipe mounting (3/4" to 3")

For the Sensor


Accessories Heating jacket Description Is used to stabilize the temperature of the fluids in the sensor. Water, water vapor and other non-corrosive liquids are permitted for use as fluids. If using oil as a heating medium, please consult with Endress+Hauser. Heating jackets cannot be used with sensors fitted with a rupture disk. For details, see Operating Instructions BA132D

Communication-specific accessories

Accessories HART Communicator Field Xpert handheld terminal Commubox FXA195 HART

Description Handheld terminal for remote parameterization and for obtaining measured values via the current output HART (4 to 20 mA). Contact your Endress +Hauser representative for more information. The Commubox FXA195 connects intrinsically safe smart transmitters with the HART protocol with the USB port of a personal computer. This enables remote operation of the transmitter with operating software (e.g. FieldCare). Power is supplied to the Commubox via the USB port.

Endress+Hauser

71

Proline Promass 84F

Service-specific accessories
Accessories Applicator Description Software for selecting and sizing Endress+Hauser measuring devices: Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, accuracy or process connections Graphic illustration of the calculation results Administration, documentation and access to all project-related data and parameters over the entire life cycle of a project. Applicator is available: Via the Internet: https://wapps.endress.com/applicator On CD-ROM for local PC installation W@M Life cycle management for your plant. W@M supports you with a wide range of software applications over the entire process: from planning and procurement, to the installation, commissioning and operation of the measuring devices. All the relevant device information, such as the device status, spare parts and device-specific documentation, is available for every device over the entire life cycle. The application already contains the data of your Endress+Hauser device. Endress+Hauser also takes care of maintaining and updating the data records. W@M is available: Via the Internet: www.endress.com/lifecyclemanagement On CD-ROM for local PC installation Fieldcheck Tester/simulator for testing flowmeters in the field. When used in conjunction with the "FieldCare" software package, test results can be imported into a database, printed and used for official certification. Contact your Endress+Hauser representative for more information. FieldCare is Endress+Hausers FDT-based plant asset management tool and allows the configuration and diagnosis of intelligent field devices. By using status information, you also have a simple but effective tool for monitoring devices. The Proline flowmeters are accessed via a service interface or via the service interface FXA193. Service interface from the measuring device to the PC for operation via FieldCare.

FieldCare

FXA193

System components
Accessories Memograph M graphic display recorder Description The Memograph M graphic display recorder provides information on all the relevant process variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on a DSD card or USB stick. Memograph M boasts a modular design, intuitive operation and a comprehensive security concept. The ReadWin 2000 PC software is part of the standard package and is used for configuring, visualizing and archiving the data captured. The mathematics channels which are optionally available enable continuous monitoring of specific power consumption, boiler efficiency and other parameters which are important for efficient energy management.

Documentation
Flow measuring technology (FA00005D) Technical Information Promass 84A (TI00068D) Promass 84M (TI00104D) Operating Instructions/Description of Device Functions Promass 84 (BA00109D/BA00110D) Promass 84 Modbus (BA00129D/BA00130D) Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI

72

Endress+Hauser

Proline Promass 84F

Registered trademarks
KALREZ and VITON Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA SWAGELOK Registered trademark of Swagelok & Co., Solon, USA HART Registered trademark of HART Communication Foundation, Austin, USA Modbus Registered trademark of the SCHNEIDER AUTOMATION, INC. Applicator, FieldCare, Fieldcheck, HistoROM, F-CHIP, S-DAT, T-DAT Registered or registration-pending trademarks of the Endress+Hauser Group