Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

4 PROGRAM M E CONTR OL L E D FA N SP EED LA RG E SURFACE A LL-RO UN D H EAT ING

Universal Oven UN and UF with SingleDISPLAY


Universal Oven UNplus and UFplus with Twin DISPLAY
Natural convection or forced ventilation
AtmoCONTROL software

Model sizes:
30 / 55 / 75 / 110 / 160 / 260 / 450 / 750
+30° C up to +300° C

UNIVERSAL OVENS U The all-round genius among the


heating ovens cover a multitude of applications, ideally at
temperatures above +50 °C. Without compromises! Thanks to
two model variants and eight sizes, optionally with natural or
forced convection, industry, science and research institutes will
find a heating and drying oven which combines top precision and
safety with optimal operating comfort.
AGEING BURNING- IN DRY ING T EMPERAT U R E S TO R AG E 5

Defined and programme-controlled fan speed


Air exchange rates and air flap position can be controlled electronically at the ControlCOCKPIT.
More inlet and outlet openings lead to a higher air exchange and reduced drying times. Various
applications recommend or even require controlled ventilation. When drying powder, sand or
corn, reducing the ventilation prevents undesired swirls. Other applications like testing of wires
or cables demand for defined air exchange rates. UFplus appliances feature easy programming
of temperature and air exchange rates with the AtmoCONTROL software.

Fresh air is preheated


Temperature deviations caused by fresh air can influence sample characteristics or prolong drying.
In Memmert unviersal ovens, the fresh air is therefore fed through a pre-heating chamber and
introduced into the working chamber.

100

90

80
Pre-warmed air
entering working
70 chamber
60

50

40

30

20

Temperature [°C] Pre-heating chamber below


the working chamber

Air supply from outside

Intended purpose as a medical device:

Heating ovens UF and UFplus are applied for heating of non-sterile fabrics
and covers.
6 TECHNICAL DATA OVE RVI E W UN IVERSA L OVEN S U

UNIVERSAL OVENS U
according to 12 880: 2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010

D F 56

(93) A (93) C
Standard equipment
Interior: Stainless steel, material 1.4301 (ASTM 304),
with all-round deep-drawn ribs to integrate
the large-area heating with ceramic-metal sheath

256

292
Internals: Stainless steel grids (sizes 30 and 55: 39
1 grid, sizes 75 – 750: 2 grids)
Housing: Textured stainless steel, rear zinc-plated steel,
intuitively operated SingleDISPLAY or exhaust air
TwinDISPLAY with Multi-Touchscreen inner Ø 57 mm

E
(from size 450 two leaves)

B
Fresh air: Admixture of pre-heated fresh air by
electronically adjustable air flap
Connection: Mains cable with plug (German Type)
CEE plug for 400 V

33
Installation: 4 feet; sizes 450 and 750
mounted on lockable castors
fresh air
Interfaces: Ethernet USB (only TwinDISPLAY)

LAN

Model sizes/Description 30 55 75 110 160 260 450 750

Stainless steel interior Volume approx. l 32 53 74 108 161 256 449 749
Width (A) mm 400 400 400 560 560 640 1040 1040
Height (B) mm 320 400 560 480 720 800 720 1200
Depth (less 39 mm for fan) (C) mm 250 330 330 400 400 500 600 600
Stainless steel grids (standard equipment) number 1 1 2 2 2 2 2 2
Max. number of grids number 3 4 6 5 8 9 8 14
Max. loading per grid kg 30
Max. loading of chamber kg 60 80 120 175 210 300 300 300

Textured stainless steel Width (D) mm 585 585 585 745 745 824 1224 1224
exterior
Height (size 450, 750 with castors) (E) mm 707 787 947 867 1107 1186 1247 1726
Depth (without door handle),
mm 434 514 514 584 584 684 784 784
door handle + 56 mm (F)
Further data Electrical load at 230 V, 50/60 Hz approx. W 1600 2000 2500 2800 3200 3400 –
Electrical load at 115 V, 50/60 Hz approx. W 1600 2000 2400 2400 2400 2400 –
Electrical load at 400 V and 3 x 230 V
approx. W – 5800 7000
w/o neutral, 50/60 Hz
Working-temperature range °C at least 5 K (UN/UNplus) at least 10 K (UF/UFplus) above ambient temperature to +300
Setting temperature range °C +20 to +300
Setting accuracy K up to 99.9 °C: 0.1 / from 100 °C: 0.5

Packing data Net weight approx. kg 44 55 64 72 80 96 160 192


Gross weight (packed in carton) approx. kg 55 67 76 86 96 114 185 242
Width approx. cm 69 70 70 83 83 93 134 134
Height approx. cm 86 94 111 104 127 134 141 189
Depth approx. cm 66 73 73 79 79 89 99 99
Order No. Universal Ovens UN30 UN55 UN75 UN110 UN160 UN260 UN450 UN750
U = Universal Oven
N = Natural convection UN30plus UN55plus UN75plus UN110plus UN160plus UN260plus UN450plus UN750plus
F = Forced air circulation
plus = Model with TwinDISPLAY UF30 UF55 UF75 UF110 UF160 UF260 UF450 UF750
UF30plus UF55plus UF75plus UF110plus UF160plus UF260plus UF450plus UF750plus
T EC HNIC AL DATA OV ERV IEW UNIV ER S A L OV E N S U 7

Options 30 55 75 110 160 260 450 750

Full-sight glass door (4 insulating glass) B0


Chamber modification for the application of reinforced perforated stainless steel
shelves or stainless steel grids (bearing rails mounted in the working chamber) – K1
– includes replacement of 2 standard grids by 2 reinforced grids
Fresh-air filter (filtration efficiency 80 %) mounted at the bottom (for UF/UFplus) R8
Interior lighting
R0
(up to size 260: 15 W, sizes 450/750: 2 x 15 W)
Interior socket (can only be ordered with limited temperature-range – max. +70 °C)
current carrying ampacity 230 V, 2.2 A can be switched off with the On/Off switch, R3
cannot be switched individually
Interior nearly gastight K2
Ditto, with possibility for gas inlet/outlet through 2 tubes with ball valves K3
Entry port, 23 mm clear diameter, for introducing connections
at the side, can be closed by flap, standard positions left centre/centre F0
left centre top F1
right centre/centre F2
right centre top F3
Entry port, 23 mm clear diameter for introducing connections at the side,
can be closed by flap in special positions (please, state location) left F4
right F5
rear F6
Entry port, 14 mm clear diameter, can be closed by flap,
D6
in special positions at the back (please, state location)
Entry port, 38 mm clear diameter, can be closed by flap,
F7
in special positions at the back (please, state location)
Entry port, 57 mm clear diameter, can be closed by flap,
F8
in special positions at the back (please, state location)
Entry port, 100 mm clear diameter, can be closed by flap,
F9
in special positions at the back (please, state location)
Entry port, 120 mm clear diameter, can be closed by flap,
D7
in special positions at the back (please, state location)
4 – 20 mA current loop interface (0 to +310 °C 4 – 20 mA)
Temperature controller actual value V3
Temperature of a Pt100 sensor positioned flexibly in chamber V6
(max. 1 SingleDISPLAY, max. 3 TwinDISPLAY)
Fan speed monitoring – optional for UFplus only V4
Works calibration certificate for 3 temperatures: +100 °C, +160 °C, +220 °C D00128

Accessories 30 55 75 110 160 260 450 750

Stainless steel grids (standard equipment) E28884 E20164 E20165 E28891 E20182
Reinforced stainless steel grid, max. loading 60 kg
– E29767 E29766 E20185
(from size 450 only in connection with option K1)
Perforated stainless steel shelf B29727 B03916 B00325 B29725 B00328
Reinforced perforated stainless steel shelves, max. loading 60 kg
– B29777 B29724 B00844
(from size 450 only in connection with option K1)
Stainles steel tray (non-perforated) 15 mm rim
E02070 E02072 E02073 E29726 E02075
(may affect the temperature distribution)
Bottom drip tray (may affect the temperature distribution) B04356 B04358 B04359 B29722 B04362
Wall bracket (tubular frame for wall mounting) B29755 B29756 B29757 B29758 B29759 –
Guarantee extension by 1 year GA1Q5 GA2Q5