LOGIQ™S6 ST-CW Option Kit Installation

Operating Documentation

Copyright© 2006 by General Electric Co.


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

Page 1-2 -
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

Important Precautions
Language

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
AVERTISSEMENT • NE PAS TENTER D’INTERVENTION SUR LES é QUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS é Té CONSULTé ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAîNER CHEZ LE
TECHNICIEN, L’OPé RATEUR OU LE PATIENT DES BLESSURES DUES à DES
DANGERS é LECTRIQUES, Mé CANIQUES OU AUTRES.

• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENö TIGT, IST
ES AUFGABE DES KUNDEN Fü R EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES
WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

- A-1
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

• ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLé S.


• SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLé S, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIó N.
• NO SE DEBERá DAR SERVICIO Té CNICO AL EQUIPO, SIN HABER
AVISO CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELé CTRICAS, MECá NICAS O DE OTRA
NATURALEZA.

• ESTE MANUAL DE ASSISTê NCIA Té CNICA Só SE ENCONTRA DISPONíVEL EM


INGLê S.
• SE QUALQUER OUTRO SERVIç O DE ASSISTê NCIA Té CNICA, QUE Nã O A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, é DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIç OS DE TRADUç ã O.
ATENÇÃO • Nã O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTê NCIA Té CNICA.
• O Nã O CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANç A
DO Té CNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELé TRICOS,
MECâ NICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE è DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE è TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVVERTENZA AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

A-2 -
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

- A-3
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (262) 827-3468 or 8*285-3468 immediately after
damage is found. At this time be ready to supply name of carrier, delivery date, consignee name, freight
or express bill number, item damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy And Procedures
Bulletins.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE Medical Systems
personnel. In performing all electrical work on these products, GE will use its own specially trained field
engineers. All of GE’s electrical work on these products will comply with the requirements of the
applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.

OMISSIONS & ERRORS


If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Medical Systems Global Documentation Group with specific information listing the system type,
manual title, part number, revision number, page number and suggestion details. Mail the information
to : Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226.
GE Medical Systems employees should use the iTrak System to report all documentation errors or
omissions.

A-4 -
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

Revision History

Revision Date Reason for change


1 April 1, 2006 Initial Release

List of Effected Pages

PAGES REVISION PAGES REVISION PAGES REVISION


Title Page 1 1 to 13 1 Back Cover N/A
Warning 1
Revision History 1
List of Effective
1
Pages (A)

- A-5
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

A-6 -
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

LOGIQ™S6 ST-CW Option Kit Installation

Section 1-1
Introduction
This Service instructions describe the installation of the ST-CW options for LOGIQ™ S6 System.

Section 1-2
Tools Required
• Standard Field Service Tool kit
• Loopback Connector

Section 1-3
Time Required
• Approximately 1 hours plus travel time for ST-CW Option

Section 1-4
Parts Required
1-4-1 ST-CW Option (Cat#: H41912LL, P/N:5173158)
ITEM PART NAME QTY PART# REMARKS
1. STCW ASSY 1 2277244-3
2. BTXCW ASSY 2 5111346 Stud assembled
3. Spacer 8 U0055AG
4. Screw 16 N9306HR For Installing BTXCW Assy
5 Option Label 1 P9524PZ
6 Installation Instructions 1 5173050

Section 1-5
Prerequisite
- None.

- 1-1
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

Section 1-6
ST-CW Option Kit Installation Steering Guide

ST-CW OPTION

1-7-2 INSTALLING STCW AND


BTXCW ASSY

1-7-3 BOARD CHECK

1-7-5 FINAL PROCEDURE

End

1-2 Section 1-6 - ST-CW Option Kit Installation Steering Guide


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

Section 1-7
Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
Perform each step in sequence and check it off when completed. If a problem occurs after completion,
go back through the procedures and check for implementation errors before contacting your Ultrasound
Region Engineer.

CAUTION Possible operational damage. Failure to strictly follow ESD (Electric Discharge) precautions
during this upgrade may cause constant or intermittent operational abnormalities. Strictly
follow all precausions.

Check each stem as it is completed.

1-7-1 Functional Check-out


Perform functional checks to verify that the system is functioning properly. Refer to the Functional
Checks in Chapter 4, LOGIQ™ S6 Service Manual.

- 1-3
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

1-7-2 Installing STCW and BTXCW Assy


1.) Turn the system power OFF and unplug the unit.
2.) Remove the three screws, then gently remove the right side cover rearward.

Screws

1.) Loose the seven screws and remove the nest rear cover.
2.) Remove the four screws and nest front cable cover.
3.) Remove the five screws and loosen the three screws, then remove the nest cover.

Nest Front Nest cover


Nest rear cover Cable Cover

1-4 Section 1-7 - Procedure


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
4.) Write down the number to the four Tx cables as shown before removing the cables.
5.) Disconnect the four Tx cables from the BTRAP5AHV boards.

6.) Remove the EBUS3T board and EBUS3B board while taking care not to damage the pins.

EBUS3T

EBUS3B

7.) Install STCW Assy into slot 2 (second slot from the left).
When it is hard to remove the board, you can use the board extracting tools specially.(Part Number:
2315492)

Table 1-1 Boards into the NEST Assy

Slot No. Board Cables disconnected

2 STCW (Option) -

3 BTRAPHV (with BTxCW) Tx Cables

4 BTRAPHV (with BTxCW) Tx Cables

- 1-5
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
8.) Install the BTXCW Assy on the BTRAPHV board below:
a.) Remove the two BTRAPHV Assy from slots 3 and 4.
b.) Place BTXCW on a static-free place as shown.

c.) Connect the BTXCW Assy onto the BTRAPHV Assy.


d.) Visually verify that it is connected to the BTRAPHV..

1-6 Section 1-7 - Procedure


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
e.) Hold it using the four screws (N9306HR) firmly from the backside of the BTRAPHV board.
BTXCW

Screw (N9306HR)

Stud (U0055AG)

f.) Install the two BTRAPHV+BTXCW Assy into the slots 3 and 4
g.) Install the EBUS boards (EBUS3T and EBUS3B). Note that both EBUS for TOP and for
BOTTOM are not interchangable.

- 1-7
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
h.) Connect the Tx cables to BTRAPHV+BTXCW as shown.
EBUS3T for TOP

Both EBUS for TOP


and for BOTTOM are
NOT interchangable.

EBUS3B for BOTTOM

.Tx Cable Connection Notice

QCON BTRAPHV+BTXCW

Tx Cable
Connection
#4

#3

#2

#1

Connect to BTxCW
slot with CW Option

1-8 Section 1-7 - Procedure


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
i.) Connect the #3 and #4 Tx cables to the BTRAPHV connectors, so that the triangle mark of the
connector faces RIGHT. .

marks
j.) Connect the #1 and #2 Tx cables to the Piggyback board (BTxCW) connector on the
BTRAPHV, so that the triangle mark of the connector faces left..

mark

k.) Reinstall the NEST cover and NEST bracket in the reverse order of removal.
l.) Go to 1-7-3 Board Check on page 1 - 10.

- 1-9
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

1-7-3 Board Check


Check if the STCW and BTRAPHV board has been installed properly using the FRU diagnostics. To
complete each test, it will take:

FRU test Time (to be tested) Tools required

STCW-FULL 3 minutes None

BTRAP-FULL 20 minutes Loopback Connector

1.) Power OFF the scanner and disconnect any probes from the scanner.
2.) Power ON the scanner.
Make sure that the LOGIQ7 software is activated. Press the Scan button to access the Scan Mode.
This Procedure is not necessary if the Scan Mode have been displayed already.
3.) Click on Utility > Service. It will take about ten (10) seconds for activating.

4.) Make sure that CAPS is OFF (should be dimmed) for password entry performed later.

1-10 Section 1-7 - Procedure


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
5.) The Service Login window for Service Platform will be shown on the monitor display.

OFF ON

CAUTION If the Login window for Service Platform is not displayed on the monitor, the installation of the
Service Platform has failed. Reload the application software or Base System Software (OS) +
application software. Contact a Technical Support for details.

6.) Select GE Service at the “Select User Level” field.


7.) Enter the password for the Service Platform.
8.) Click on Okay.

- 1-11
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION
9.) Verify that the following screen (Service Platform) is displayed on the monitor.

10.)Select Diagnostics > LOGIQS6 Diagnostics > FRU Test> STCW>FULL.


Perform STCW FULL test.

11.)Verify that all of the STCW tests complete with no error.

12.) :Select Diagnostics BTRAP > FULL.


Perform BTRAPHV FULL test and verify that all of the BTRAP tests complete with no error.

Click on x located at the upper right corner of the service platform screen to close the Service
Platform and return to the scan panel.

1-7-4 Functional Check


Power OFF the system, then ON. Press the CW key to verify that CW can be used with no error.

1-12 Section 1-7 - Procedure


GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

1-7-5 Final Procedure


1.) Unscrew four screws and remove the Right Side Cover.
2.) Attach the Option Label onto the Nest Cover.

System Version Label


Software Label

Nest Cover
Option Labels

3.) Re-install the Right Side Cover.


4.) Fill out the proper customer information on the Product Locator Card. Mail the installation card
“Product Locator” to the address corresponding to your region.

Example

5.) Properly dispose of excess material.

- 1-13
GE MEDICAL SYSTEMS
DIRECTION 5173050, REVISION 1 LOGIQ™S6 ST-CW OPTION KIT INSTALLATION

1-14 Section 1-7 - Procedure