Download as pdf or txt
Download as pdf or txt
You are on page 1of 42

Portable Ultrasound System

VISION 4000.

Service Manual

P/N:4720-0012-01A

INTELLIGENT VISION Company limited


VISION 4000 Portable Ultrasound System Service Manual

Table of contents

1 GENERAL INFORMATION.................................................................................................2

1.1 THE CHARACTERISTIC OF THE PRODUCTS ...............................................................................2


1.2 APPLICATIONS ..........................................................................................................................2
1.3 OPERATION MODES..................................................................................................................2
1.4 MEASUREMENTS AND CALCULATIONS ....................................................................................2

2 SYSTEM CONFIGURATION AND FUNCTION ...............................................................3

2.1 PROBE SELECTION BOARD.......................................................................................................3


2.2 PREAMPLIFIER BOARD (DRV BOARD).....................................................................................3
2.3 FRONT-END BOARD (BFM BOARD).........................................................................................4
2.4 DSC BOARD ............................................................................................................................4
2.5 KEYBOARD ..............................................................................................................................4
2.6 MONITOR BOARD ....................................................................................................................4
2.7 REAR PANEL ............................................................................................................................5
2.8 POWER SUPPLY ........................................................................................................................5

3 SYSTEM OPERATION THEORY .......................................................................................6

3.1 OPERATION THEORY ................................................................................................................6

4 COMPOSITION PRINCIPLE OF THE SYSTEM ...........................................................18

4.1 GENERAL INFORMATION OF THE CIRCUIT BOARDS .................................................................18


4.2 DRAWING OF CONNECTION BETWEEN CIRCUIT BOARDS.........................................................29

5 ASSEMBLY /DISASSEMBLY OF THE SYSTEM ............................................................31

6 MALFUNCTION ANALYSIS .............................................................................................35

6.1 COMMON MALFUNCTION ANALYSIS OF FRONT END SUBSYSTEM .........................................35


6.2 MALFUNCTION ANALYSIS OF DSC BOARD ............................................................................38
6.3 MALFUNCTION ANALYSIS OF POWER .....................................................................................39

7 MALFUNCTION ANALYSIS OF MONITOR BOARD ...................................................40

P/N: 4720-0012-01A
1
-2- -2- VISION 4000 Portable
Ultrasound System Service Manual

1 General Information

1.1 The Characteristic of the Products


The VISION 4000 portable ultrasound diagnostic system, an electronic
linear/convex scanner, can provide fine image which owns high spatial
resolution and detail tissue information.
The equipment is suitable for operating in dark ultrasound room,
because of backlit key on its control panel.
VISION 4000 owns a perfect measurement and calculation package
including FGC(fetal grow curve)and detail report.
The equipment has two probe sockets, which can be connected with
more than ten types of probe.
The system can create and play Cine loop, store image without losing
after powering off, transfer image to PC through a standard RS232
interface for father processing, connect many types of video printer,
monitor or VCR through standard video port.
1.2 Applications
With a varied selection of probes, VISION 4000 can be used for
examination of abdominal, OB/GYN (including vagina), cardiac and
thyroid.
WARNING
It is not applicable for ophthalmology or any examination that ultrasound
beam would penetrate through eye.
1.3 Operation Modes
— B Mode
— M Mode
— B/M Mode
— B/B Mode
1.4 Measurements and Calculations

— General measurements: distance, perimeter, area, volume, angle,

heart rate, time, slope

— OB :BPD, CRL, FL, AC,HC,GS,UT,OFD,CI,AFI,FGC,VOL,ANG

— Cardiac: MV,LV,AO,ET,EDV,ESV,EF,CO,SV,FS Others: RUV,PV

P/N: 4720-0012-01A
2
VISION 4000 Portable Ultrasound System Service Manual

2 System configuration and function

This system consists of a main unit and probes. The main unit is composed of a probe
select board, a preamplifier board, a front-end board, a DSC board, a keyboard panel, a
monitor board, a CRT monitor and power supply. It is shown in Fig. 1.

CRT
Probe Probe Preampli Front- Monitor
A Select fier End
DSC Monitor
Board Board Board
Board Board

① ② ③
Probe ④ ⑥
B

Keyboard
Power Supply⑧ Rear Panel⑦
Main Unit Panel⑤

Video Video RS-232 Footswitch


Out I Out II Port

Fig. 1

2.1 Probe Selection Board

By connecting two 156-pin probe sockets to two probes at one time, the
probe selecting board can select one as current operating probe .It can
receive 24-channel high voltage drive signal and sent them to operating
probe element, at the same time it can also sent echo signal form operating
probe to preamplifier board.

2.2 Preamplifier Board (DRV Board)

This board consists of a 24-channel high-voltage transmitting drive signals generating


circuit and a 24-channel preamplifier, which will generate high-voltage transmitting drive
signals and amplify the echo signals sent from the probe select board.

P/N: 4720-0012-01A
3
VISION 4000 Portable Ultrasound System Service Manual

2.3 Front-End Board (BFM Board)

To receive echo signals from the preamplifier board, to process for focusing and
aperture control, and to generate receiving beam signals. Processed through dynamic
filtering, TGC (Time Gain Compensation) control, logarithm compression, detection and
edge enhancement, the receiving beam signals become video echo signals and send
them to the DSC board.
Transmitting excitation pulse signals are generated by the front-end board and sent to
the preamplifier board.

2.4 DSC Board

The DSC board is composed of three circuits: DSC (Digital Scan Converter), Main

Controller and Auxiliary Controller.


The DSC completes the conversion from ultrasonic scanning timing sequence to TV and
progressive scanning display timing sequence, achieves ultrasonic image digital
processing such as frame correlation, interpolation and γ correction, and generates
composite video signal I and II.
The main controller receives operation commands from the keyboard and sends

corresponding control information according to the system current state. It also

generates various graphs and character data displayed on the screen and achieves all

the measurement and calculation functions.


The auxiliary controller is for image & data asynchronous communication via RS-232C
port, image storage and recalling.

2.5 Keyboard

There are 62 keys, 1 control knob (overall gain), 7 TGC control knobs and 1 trackball
mounted on the keyboard circuit board, which is the user’s console for selecting imaging
mode, imaging parameters, measurement and calculation functions, inputting
characters and adjusting TGC.

2.6 Monitor Board

To amplify the composite video signals sent from the DSC board and generate signals
that drive line and field scanning, and to generate the voltage required by the normal
operation of CRT electrodes.

P/N: 4720-0012-01A
4
VISION 4000 Portable Ultrasound System Service Manual

2.7 Rear Panel

In addition to an AC power socket and a fuse, video signal output ports I and II, an

RS-232C port and a footswitch socket are installed on the rear panel.

2.8 Power Supply

The power supply of this unit is switch power supply. It generates all the DC voltages

required for normal operation of all the boards.

P/N: 4720-0012-01A
5
VISION 4000 Portable Ultrasound System Service Manual

3 System Operation Theory

3.1 Operation Theory

The operation principle diagram of the VISION 4000 is as shown in Fig. 1. The system

is chiefly composed of a probe select board, a preamplifier board, a front-end board, a

DSC board, a control panel and a monitor. On the probe switch board there are two

probe connectors (156-pin) for connecting two probes at the same time. There is an

RS232C port on the DSC board, which can be connected to a PC of 486 or above via a

socket on the rear panel. Thus an image displayed on the monitor can be sent to the PC

for display, storage or further processing.

The keys for operation modes on the panel can be set for scan modes (B, B/B, B/M or

M mode) and other operation conditions through the main controller. The main controller

will test whether the probe connected to the system exists, and it will set the parameters

such as scan depth, focus position, the relation between aperture size and depth,

repetition frequency of transmitting ultrasound pulse according to the code of the probe

connected.

Based on the probe code and the parameters from the main controller on the DSC

board (see Fig. 8) such as the transmitting trigger pulse, the transmitting focus position,

the transmitting excitation pulse, focus position and aperture, and scan line number sent

from FPGA2, FPGA1 on the front-end board (see Fig. 4) will determine the specific

element number (23 or 24 elements altogether) for ultrasonic reception and

transmission, and the transmitting pulse timing relation of elements (subject to the

parameters such as transmitting focus position). The transmitting circuit and receiving

pre-amplification will transform the transmitting trigger pulses from FPGA1 to

high-voltage transmitting signals, excite the elements to transmit ultrasound, and amplify

P/N: 4720-0012-01A
6
VISION 4000 Portable Ultrasound System Service Manual

the echo signals reflected from the imaging target (e.g. human organ tissue) about

20dB.

The pre-amplified 23-channel or 24-channel echo signals go through the receiving focus
and aperture control circuit, compound into beam signals. Going through the main
amplifier for dynamic filtering, TGC control and logarithm compression, the signals will
be detected for its amplitude and preprocessed for its video (edge enhancement, etc.).

The video echo signals detected are sent to the first level A/D converter on the
DSC board, and converted to 8-bit digital signals. The 8-bit data stream acquired from
ultrasound scanning sequence converts to a 2D ultrasound image with line field
scanning that can be displayed on a monitor. It is the task of digital scan converter
(DSC).

On the main channel of DSC signals, the output of A/D converter is processed by frame

correlation and saved to the frame memory. In the meantime, the data read from the

frame memory in line field scanning timing sequence are processed by interpolation

(achieved in FPGA3), and then outputs a pixel signal at a rate of 12.5 MHz in TV line

field scanning sequence. After this signal is transformed (such as γ correction) with the

grayscale signal generated by FPGA3, in the signal composite circuit they synthesize

with characters generated by CPLD1, graph DOT signals (signal of characters and

graphs), composite synchronous signals and composite blanking signal into digital TV

signals and progressive scanning composite video signals. By way of the D/A converter,

they are converted to 1.2V-some analog TV signals and analog progressive scanning

signals, and sent to the rear panel and the progressive scanning monitor.

The digitized signals (after A/D conversion) of ultrasonic scanning can be displayed as a
real-time image on the monitor via the DSC, or stored in the CINE memory. After the
real-time image is frozen, the stored images (64 frames altogether) can be played back
in single frames or frame by frame continuously. The read-write address signals of the
frame memory are generated by FPGA2. CPLD1 generates timing sequence signals
related to TV scanning and progressive scanning. Based on the related signals of TV
scanning, FPGA2 in combination with some matched EPROM and Flash Memory, will
generate read-write addresses and read-write signals of the frame memory, addresses
and read-write signals of CINE memory, and provide timing sequence signals related to
P/N: 4720-0012-01A
7
VISION 4000 Portable Ultrasound System Service Manual

ultrasonic scanning for FPGA1.


The digital TV signals outputted by the signal composite circuit can be displayed directly
on the monitor via D/A conversion. In the meantime, they are sent to CPLD2 and stored
in the transmitting image memory under the control of the auxiliary controller. As long as
the RS-232C interface of this system is connected to a PC (with related software), an
image (including characters) saved in the transmitting image memory can be sent to the
PC at its request. Images in the transmitting image memory can also be saved in the
image memory by pressing “STORE” key.

You can freeze a real-time ultrasonic image on the monitor by pressing FREEZE-KEY
on the keyboard panel or the footswitch. Various graphics measurement and edit of
characters on the screen can be operated at this time. The Cineloop can be done in
frozen state.

Keyboard composition and functions


The keyboard of VISION 4000 is as shown in Fig. 2. Here is the
introduction to the keyboard function.

● B work mode select keys


a) Press key, and it will display a single B mode image.

b)Press key. It is in dual B mode, and the live image will be

displayed on the left of the screen.

Press key, and the live image will be displayed on the right of
the screen, while the left image is frozen.

z M work mode select keys

Press key, and it will display both M mode and B mode


images.On the left is a B mode image. The sampling line position will
decide the sampling position of the M mode image displayed on the
right of the screen. The sampling line position can be changed by
moving the trackball.

Press key to enter full M mode.

P/N: 4720-0012-01A
8
VISION 4000 Portable Ultrasound System Service Manual

z M mode scan speed select key

Press repeatedly to select one of the four scan speeds (8

sec/frame, 4 sec/frame, 2 sec/frame, 1 sec/frame ) in turn.

z Image zoom rate(ZOOM)select keys

(a) Press key to zoom in.

Press key to zoom out.

(b)Press key repeatedly, and the zoom rate will be: ×0.8→×1.0→

×1.2→ ×1.5→×1.7 → ×2.0

z Image vertical shift keys


In B mode and when the image zoom rate is over 1.0, press the keys
to shift the image vertically. Press once to shift the image at an
increment of about 2mm.

Fig. Keyboard
P/N: 4720-0012-01A
9
VISION 4000 Portable Ultrasound System Service Manual

z Focus position select keys


To select the focus position in depth direction, and the selected focus
will be shown on the top right of the screen.

key: to select near-field focusing, i.e. the focus is at the


surface.

key: to select middle-field focusing, i.e. the focus is at a


medium distance.

key: to select far-field focusing, i.e. the focus is deeper.

key: to select a further-field focusing, i.e. the focus is much


deeper.

● Image processing function key

Press key to enter image processing function menu

(1)CORRM -to change correlation coefficient.

(2)ENHCM -to change edge enhancement

(3)GREYM -to change grayscale curve (i.e. γ correction

curve)

(4)P/N M - positive/negative inverse

(5)L/R M - left/right inverse

(6)U/D M -up/down inverse

z Frequency select key

Press key to achieve change of probe work frequency. There are


4 work frequencies for each probe. The current work frequency is
displayed on the left of the screen.

P/N: 4720-0012-01A
10
VISION 4000 Portable Ultrasound System Service Manual

z Probe select key

Press key to select the required operating probe. The model


and frequency of the selected probe are shown on the left of the
screen. If the selected socket is not connected to any probe, the
original image will be frozen and “NO PROBE” is displayed.
Note: When the VISION 4000 is turned on, it will select probe A as the
current operating probe automatically, except that only probe B is
connected to the main unit.

z Freeze key
Press this key to freeze or unfreeze the image. When the image is frozen,
all the characters and graphs displayed on the screen will be retained.
When it is unfrozen, the ultrasound image will be refreshed, and all the
characters and graphs in the image area will be cleared.

z Cineloop control key

In image frozen state, press to enter cineloop in single

frames. In single B mode, press again to enter cineloop


continuously from single frames, and vice versa. However, cineloop
continuously cannot be played in M mode.

If one of the keys , , , keys is active (in

high brightness), it is invalid to press key.

z Body mark select key

Press key to enter the function menu related to body marks,

P/N: 4720-0012-01A
11
VISION 4000 Portable Ultrasound System Service Manual

which is as below:
BODYS : To enter the body mark select list
MOROP :To move and rotate the probe mark
OFFBY :To turn off the body mark

z Distance measurement key

In image frozen state, press this key to start distance measurement.

z Measurement function operating keys

Key: different measurement options are available in B,

B/B and M, B/M modes.

key: In image frozen state, press this key to display the

measurement function menu.

In B, B/M or B/B work state, measurement function menu options 1-9


are displayed:
BPD* Biparietal Diameter
GS Gestational Sac
CRL Crown-up Length
FL* Femur Length
AC* Abdominal Circumference
HC* Head Circumference

VOL-B Volume (Bi-plane method)

ANGLE Angle
OBRPT OB report
Note: Items with “*” are required measurements for displaying an OB
report.

● key
Press this key, and the patient’s PID can be inputted directly on the

P/N: 4720-0012-01A
12
VISION 4000 Portable Ultrasound System Service Manual

screen. After input, press key again to confirm and exit.

● key
Press this key, and the cursor “_” will be shown at the screen center.
Move the trackball to shift the cursor to where annotations to
be added on the image display area and enter characters.

After input, press key again to exit.

z key
To enter the system preset menu for setup. There are 5 setups for
selection.

z Biopsy guide line


For the probes listed in Section 1.3, they are equipped with biopsy
guidelines. The procedures for adding a biopsy line are as follows: in

live state, press key (it is lighted). Press key and a


biopsy line will be displayed. The biopsy guide line for 6U1 can be

moved. Press once and the position of biopsy line will be

shifted downward for 2mm; the biopsy line can be shifted within the
range of 1 to 5 cm.
z Numeral/character keys
The date and annotations can be inputted with the group of
numeral/character keys on the keyboard. Some special instructions for
some keys are:

key: to clear annotations and measurement graphs on the screen.


If it is pressed during adding annotation, only the cursor will be

P/N: 4720-0012-01A
13
VISION 4000 Portable Ultrasound System Service Manual

cleared, which is equivalent to pressing for the second


time.

key: for shift control of upper and lower cases of numeral/character


keys. When this key is lighted, it is in the upper case state.

z key
Press this key to start the image recorder connected with the main
unit for printing the image. It is a command to remote control the
printer, which equals to the Print button on the printer panel.

z key (used when an image storage part is


configured)

z key (used when it is configured with an image


storage part)
Press this key to display an image previously saved in the image
memory.

z Overall gain control knob GAIN : Turn clockwise to

increase image overall gain.


z TGC Control
There are 7 slide potentiometers for 7 sections of TGC control. The
slide potentiometers close to the monitor are for controlling the near
gain, while the further slide potentiometers are for controlling the far
gain. Toggle the potentiometer rightward to increase the gain of the
corresponding section, and leftward to decrease the gain.

P/N: 4720-0012-01A
14
VISION 4000 Portable Ultrasound System Service Manual

1.3 Performance specifications:


1.3 .1 Probes
Angle/ Scan Horizontal Vertical Application
Model Frequency(MHz)
Width depth resolution resolution range
C343 4mm 2mm Abdominal,
2.5/3.0/3.5/5.0 80° 200mm
convex (depth<80mm) (depth<80mm) OB/GYN
L751 1mm 1mm
5.0/6.0/7.5/10.0 50mm 90mm Small parts
linear (depth<60mm) (depth<60mm)

5V1 3mm 1mm


3.5/4.0/5.0/6.5 130° 120mm Gynecology
micro-convex (depth<60mm) (depth<60mm)

C511 3mm 1mm


3.5/4.0/5.0/6.5 130° 120mm Pediatrics
micro-convex (depth<60mm) (depth<60mm)
C361 4mm 2mm Abdominal,
2.5/3.0/3.5/5.0 70° 200mm
convex (depth<80mm) (depth<80mm) OB/GYN
C321 4mm 2mm
2.5/3.0/3.5/5.0 110° 170mm Cardiology
convex (depth<80mm) (depth<80mm)

C521 3mm 1mm


3.5/4.0/5.0/6.5 110° 120mm Pediactrics
micro-convex (depth<60mm) (depth<60mm)
C541 3mm 1mm Pediatrics,
3.5/4.0/5.0/6.5 80° 140mm
convex (depth<60mm) (depth<60mm) abdominal

6U1
4.0/5.0/6.5/8.0 60mm 110mm 1mm 1mm Transrectal
linear

1.3.2 Main performance specifications:


Scanning modes: electronic convex, linear scanning
Scanning method: electronic convex scanning, linear scanning

z Display Mode: B, B/B, B/M, M

z Gray shades: 256

z Frequency of probe: 2.5MHz to 10.0MHz

z Image display multiple: ×0.8,×1.0,×1.2,×1.5,×1.7,×2.0; shift approx. 2mm a

step

z Focusing method: Variable aperture, 1-4 focal zone electronic focusing

(1) Display range (Max):

Convex: Depth 200mm angle 80°

Linear: Depth 140mm width 65mm

P/N: 4720-0012-01A
15
VISION 4000 Portable Ultrasound System Service Manual

z Image control:
Overall gain control knob
8 TGC control slides
Grayscale curve: 4 types
Frame correlation: 4 steps
Edge enhancement: 4 steps
z Sweep Speed in M Mode: 1, 2, 4, 8 s/frame

z Image Display: Left/ Right, Positive/ Negative, Up/Down

z Cineloop: Up to 64 frames of continuous/ single frame replay in B mode, up to 256

seconds in M mode

z Image storage: 16 images and related characters as well as graphs. (Option)

z DSC Memory capacity: 512 x 512 x 8 bit

z Monitor: 10-inch B/W progressive scanning monitor

z Built-in backlit keyboard: all the keys on the control keyboard have backlight. For

the backlight of some function keys, there are two brightness types. The function of

the functional key is valid if it is in high brightness.

z Character display

z Patient's ID

z Hospital Name

z Comment

z Automatically Display Items: Date & Time, probe frequency, gain and other

operating parameters, and various measured values.

z Body Marks: 25 types

z Measuring functions:

z Basic measurement: Distance, Circumference, Area, Volume, Angle, HR, Time

z Obstetrics measurement: BPD, CRL, FL, AC, HC, GS, VOL, ANG

z Other measurement: MV, LV, AO and slope

z I/O Port

P/N: 4720-0012-01A
16
VISION 4000 Portable Ultrasound System Service Manual

z RS-232C port for transmitting images to PC (Option)

z Two ports for convex or linear array probe

z Video output system:

Video I: Progressive scanning, 50 frames/ second

Video II: PAL format, 625 lines/ frame, 50 fields/ second

z Dimensions: Approx. 417×310×260(mm)

z Net weight: Approx. 12kg

z Power Consumption: ~220V 50Hz 120VA

z Environmental Requirements:

Operation Storage and transportation

Environment temperature: 5~40℃ -40℃~55℃

Relative humidity: ≤ 80% ≤ 95%

Atmospheric Pressure: 700~1600hPa 700~1600hPa

P/N: 4720-0012-01A
17
VISION 4000 Portable Ultrasound System Service Manual

4 Composition Principle of the System

4.1 General information of the circuit boards

Beam Former (BFM)


The beam former board is as shown in Fig. 3. it receives 24-channel echo signals

(CH1-CH24) from the preamplifier board. Since there is symmetry in the delay inequality

of the 24-channel echo signals, the signals can be incorporated to 12-channel signals

(XX0-XX11) after being processed in the folding circuits (U12, U22 and U23). Through

buffering, focusing (U24~45) and aperture control, the signals are synthesized to radio

frequency receiving beam signals RF. The RF signals are processed by TGC

compensation and dynamic filter, and then compressed in logarithm (U53, U54).

Finally video echo signals DET are formed after detection (U56) and edge enhancement

control (U57) and sent to DSC board. On the beam former board, there is a FPGA1 chip

with 100,000-gate as a core control component of the beam former board. FPGA1 will

start achieving signal control (such as timing signals) based on probe codes, as well as

scan line numbers, focus positions, scan lines from the DSC board. Its major functions

are:

(1) To generate various control signals required by the front-end, such as edge

enhancement control signal ENHCON1-2, dynamic filter control signal FLT1-3;

(2) To generate 24-channel acoustic wave transmit drive signal CHX[24:1] after focus

processing;

(3) To generate control data of high-voltage switch.

The layout of the major components on the beam former board is as shown in Fig. 4.

P/N: 4720-0012-01A
18
VISION 4000 Portable Ultrasound System Service Manual

U21-U23] CH[24:1]
CH[24:1] Folding XX[11:0] U24-U45 delay line set and aperture TGC level and
circuit buffer delay control control
RF
dynamic filter

U53,U54 U57
U56
logarithm edge video echo DET
detection enhancement output level
compression

Fig3

P/N: 4720-0012-01A
19
VISION 4000 Portable Ultrasound System Service Manual

FPGA1

Fig. 4

Probe Board

P/N: 4720-0012-01A
20
VISION 4000 Portable Ultrasound System Service Manual

The probe board includes two PCBs that are housed in one shield box. One of the
PCBs has two probe sockets and 48 DPDT relays. Two probes can be connected at the
same time. This PCB is called Probe Board 1 or Probe Select Board (PBCON). The
other PCB is mounted mainly with a transmit pulse drive circuit, a preamplifier and a
high-voltage switch. This PCB is called Probe Board 2 or Preamplifier Board (DRV
board). The block diagram of Probe Board is as shown in Fig. 5. On DRV board, the
transmit pulse signals CHX[24:1] sent by FPGA1 (U20) on BFM board go through
MOS Drive (U1-U12) and form transmit pulse drive signals CHXD[24:1]. This group
of signals are sent to high voltage transmit pulse generating circuit to generate high
voltage transmit pulse signals CX[24:1] with sufficient amplitude. The echo signals
received from the same working element are amplified by 24 preamplifiers (marked as
CH1-CH24) and then sent to the BFM board. The high voltage transmit pulse
generating circuit of the same channel and the preamplifier are installed on a
small circuit board, which is marked as U41-U64 (shown in Fig. 7).
On PR1 board, CX [24:1] are distributed to the current working element by the high
voltage switch. The maximum elements of all the probes are 96. Among the 96
elements (or 80 elements), only 24 elements are effective, i.e. they have high voltage
transmit pulse drive for transmitting acoustic waves. The 96 signals SW1-SW96
outputted by the high voltage switch are switched by 48 lock-mode DPDT relays and
then sent to two 156-pin probe sockets.
The component layout of Probe Board 1 and Probe Board 2 are as shown in Fig. 6 and
Fig.7.

P/N: 4720-0012-01A
21
- 22 - - 22 - VISION 4000 Portable
Ultrasound System Service Manual

TX2SA
U41-U64
U1-U12
12 Pieces of ECN3290 Probe
CHXD[24:1] 24-channel
MOS socket A
receive/transmit 24-channel high
48 relays
drive circuit
From BFM board voltage switch

Probe
socket B

CE[24:1]

PID[7:0]
TO BFM board
Probe
code
PS
selection

Fig. 5

P/N: 4720-0012-01A
22
VISION 4000 Portable Ultrasound System Service Manual

Fig. 6

P/N: 4720-0012-01A
23
VISION 4000 Portable Ultrasound System Service Manual

Fig. 7

P/N: 4720-0012-01A
24
VISION 4000 Portable Ultrasound System Service Manual

DSC Board
The component layout on the DSC board is as shown in Fig. 8.

Fig. 8

P/N: 4720-0012-01A
25
VISION 4000 Portable Ultrasound System Service Manual

The video echo signals DET sent from the BFM board are converted to 8-bit digital
signals ADC[7:0] by the A/D converter (D3) on the DSC board. ADC[7:0] are sent to
the frame correlation circuit for processing. The digital signal processing functions
including frame correlation calculation are done by a part FPGA3 (D45). Some major
functions that are done by FPGA2 and FPGA3 are: (1) frame correlation calculation; (2)
interpolation calculation; (3) γ correction; (4) generating read/write address for the frame
memory and the cineloop memory; (5) generating sampling pulse signal for variable
sampling frequency in accordance with the coordinate transform requirement; (6)
generating read/write signals and other control signals for frame memory; (7) providing
related time-sequence signals for the front-end circuit; (8) providing data exchange
channels with CPLD2; (9) displaying signal synthesis such as images, grayscale signals,
character and graph signals as well as composite synchronous signals, composite blank
signals.

The echo digital signals processed by frame correlation in FPGA3 are saved in the frame
memory (D42) and then cineloop memory (D37). In cineloop state, data are read from the
cineloop memory and then displayed. In live and freezing state, data are read from the
frame memory directly and displayed. In RECALL state, the displayed image information
is sent from the image memory (D7) via CPLD2 and then reach FPGA3 for display.

The echo digitals are read from the frame memory according to the line field scan display
timing sequence requirement. Then after interpolation calculation, the transform from
ultrasound scanning sequence to line field scan display format sequence will be
completed.

Another important circuit on the DSC board is the controller and the character-graph
generating circuit. The controller is composed of the main controller and the auxiliary
controller. The core part of the main controller is P87C52X2BA(D28), which consists
of P87C52X2BA, the program memory (D32), the data memory (D32) and the relative

circuits (D29,D9,D10,D19,D20,D34, etc.). The main controller mostly serves

for: (1) scanning the keyboard and generating corresponding control commands
according to the operating requirement and the current system state to ensure proper
functioning of the system as per operational requirements; (2) completing the calculation
required by all the calculation and measurement functions; (3) generating character and
graph data for the user interface; (4) managing TGC and overall gain knob, etc.

The auxiliary controller is composed of P87C52X2BA SCM (single chip microcomputer)


D39, CPLD (D27), image memory (D7), the memory for storing and transmitting image
data (D40) and the RS-232C level changeover circuit (D31). It provides an RS-232 port,
and the image storage function. It will transmit the data such as images, characters and
graphs displayed on the screen to the PC via the RS-232C port. In STORE state, the
displayed information such as images will be saved in the image memory.

P/N: 4720-0012-01A
26
VISION 4000 Portable Ultrasound System Service Manual

A circuit consisting of CPLD1 (34) and two static RAMs (D34 and D39) will generate
various time-sequence signals required for line field scan (interlaced and progressive)
on one hand, and generate dot-matrix signals DOT for characters and graphs with the
assistance of the main controller on the other hand.

The digital composite video signals of interlaced and progressive scanning


sent from FPGA3 go through D/A converter (D44 and D12), then are outputted
to the monitor and the rear panel.

Monitor (MON) Board


The SSI-600 is configured with a 10-inch build-in monitor, which is composed of a
main circuit board, a deflection coil, a CRT type display tube and a small circuit board at
the end of the display tube. The power is supplied by the switch power board of the main
unit.
The fundamental principle is the same as that of a common B/W monitor. It main consists
of a video amplification channel (which is for amplifying the video signals from the main
unit to a certain range and adding to the kinescope to display a corresponding image),
the sync separating circuit (whose major function is to separate the inputted video signals
into line synchronous and field synchronous signals. The line and field scan circuits
receive line and field synchronous signals, and generate sawtooth current with a certain
frequency, sufficient amplitude and good linearity. The line and field deflection coils
receive sawtooth wave current from line and field sweep circuit respectively, control the
horizontal and vertical movement directions of the electronic beams in the display tube
and sweep the corresponding grating on the screen. The dynamic focusing circuit is for
improving the focus performance of image scan. The grating amplitude
automatic-correction circuit is for adjusting the grating amplitude change induced by
brightness and contrast change. The high-voltage coil is for generating voltage required
by video amplification circuit and line output circuit, and medium voltage and high voltage
required by display tube operation.

Explanation of ports:
CH1: video in
CH3: interface of the main circuit board and the small circuit board at the end of the
display tube
CH6: power supply in
CH7: output port for line/field sweep signals
CH8: interface for the front panel brightness and contrast potentiometers
The layout of components is as shown in Fig. 9.

P/N: 4720-0012-01A
27
VISION 4000 Portable Ultrasound System Service Manual

TMP-JOLX-A2
1C5

7C15 104
4R31 7R11
4D9
5V1 750

470u/16V
223

50/60Hz-SW
4148

47u

1C1
4V9

2619
202
1R2

2619
5C3 202 4R32

CH1
4148 5D4 5R19 102
683

+
684 4R30
IL-4P-S3EN2-1

512 202
+
5R3 1R4
5R17 5R18 1V1 1R3
334 103
1 CH4 4

181

60Hz V.SIZE

5VR5
104 8C11

500K
V.SIZE

5VR2

5R21 304
4R45 8IC1 8R13 1R5 223

50K
5D1

823 5R5B
102 470 331 2619

(LM1881N)
4148

LM1881M
200K

5VR4
5IC2 8R14 7R10
5R5A
5C12

1C2
822 5R14 330

7C3

47u
103 8C12

1u
124

V.HOLD

5VR1
100K
7C12
+
+ 0.1u 8R12
1000u/25V

0.15u/100V
5C10

+
684 105

HD74LS123P
22u

7C4

10u
5C4

8C13
5C15 8R16

103
101 +
154 5R4 3R3
8R15 102

7C13
5R22

7C5
682 5R12

103
5R15

564

10u
8R11

2619

8V1
104 5C11

M52721SP
220
+
5R10
+

7IC1
+
5C17

7C14
1/0.5W

103
5C14

7C6
47u

470u/16V

47u
TDA1175P
1.2K/0.5W

512 5R9 0.1u

CH05W3.5V
8D1
8C14
+
5R11

7R6
100u/50V

222
5D3 7C9A 104
224 5R13
7C9B 104

5C5
151 5C9 CH05W5.1V 7C7 103

1R19 750
103

202
7R9
IL-2P-S3EN2-1

6R1 7C11

7C8
1.2K

GND
5D2

Iu
5IC1 5R20

7R8
123
+
CH2

102 5C8
1 2

220/0.5W

TP1
513 5R8 1N4007
7C10 104

10uH
1L2

B 3DA3417

SUB CONT
105 5R7
1000u/25V

363 5R6 1R13 331 153

7VR2
5K
5C13

1C14 101 7R7


6L1

6L2

IL-4P-S3EN2-1

2619
C 1V3
181

1V2
2.2K/5W
1R16
1 CH3
50uH

50uH

1R18
4

1R17 150 1R14 1C6


5VR3
V.LIN

1C11
50K

102 103
87633-4

1D2 331

1N4007

E
1
CH6

CH05W7.5V 1L1 47uH

1D3
5R16
4
6L3

103
6L4

5C7 1R15
5C6 1C12
50uH

50uH

6C2 15
1N4007

0.22u/63V
1

+
1D4

223

IL-6P-S3EN2-1
4C26

1
0.1u/100V

4.7u
33u/250V
1C13

4R47 123 4R46


470u/25V
6C1

104 4R41

CH8
154 4R42
103

564
6C4
DCY5.757.250

2619

2619
4R48

4V7
182
TJC2-04A

4V8
+
4L2

6 4R44 332 4R43


10u/50
4C17
CH7

470u/50V

LSR-10 682
4R39 472
472 4R40
6C3

4R24 4R37
104
4148

4R29
4D1

4VR6
102

5K
WIDTH

4L1 4R23

3.3u
222 4D8
H.LIN

473 4R28 4148


+
470 0.1u/100V 124 4C25
472 4R38
4C23

4R22 4C14 4C16


2619
1u/160V

4V6
C
B

CH05W15V
E

4700/100V
4V3
DCY7.820.904A MONIT S1

10K/1W

3DG2271E
1K/2W
4R49

4D7
4R36

103
4R26 4R25 4C27 3906
103
4V5

47u/50V
4R27 4C20 101
4C15

4C24
390K/1W
5.6M/0.5W
4C18
220u/50V

6.8u/250V 4C19 103 330 +


3300/160V
4C21

4R35 4R34
0.01u/1200V 4148
4R21 5D5
FOCUS 1M/0.5W

5R23
50uH
4L3

10K/0.5W
W208-2A-500K 203
4V4

4R20
820K/0.5W

SUB BRT

5R24
4VR3

4VR4
4R16

500K

100K/0.5W
4R51

472 9R1 9R3


4R18 2SB834
B C E 222 123
100K/0.5W B C E
5V2

3DG2383
A3

9C1 9R14 9R2 4R33


4R17 4D4 101 153 223 100
V09E 9IC1 9C7
333
9R7
H.PHASE 220 4R11

103
S IRF840A G

820K/0.5W 4R19
9VR1

5K

22/0.5W
220K/0.5W 5600 9C8
4R14
10u/250V

4V2

4C11
4C12

0.01u/1600V 123 9R8


A1

9C2 102
A2

LA7851

4D3
H.HOLD

9R4 183
9VR2

V11L
5K

4D2 9C3 331


47u/100V

9C11 0.047u
V11L
4C22

+
BSH10-N8

102 9R9
9R5
183

4R13
9C12
4B1

330u

9C9

390/0.5W 0.015u
4R15 +
9C5
1u

9C4 6800
22/0.5W 4D5 183
+
4C10
4C9

4C28

9C6

822 9R10
V19G(600V)

V09E
1u

15
9R12 9R6 153
4C13
4D6

9R13
10u/250V

510/0.5W 3DG8050
0.01u/630V

0.01u/630V

104
6800/630V

4V1
9C13

4R10
C
B
E

Fig. Component Layout on MON board

P/N: 4720-0012-01A
28
VISION 4000 Portable Ultrasound System Service Manual

4.2 Drawing of connection between circuit boards

Cables and wires connecting with DSC boards

(1) DSC PCB BFM PCB

XS4 P2
50pin

(2) DSC PCB Keyboard PCB

XS7 XS1
40pin
1
DET

+5V
BFM PCB
(3) +12V P4

-12V

DSC PCB XS2 POWER PCB


8pin

Monitor

DSC PCB
(4)
VIDEOⅠ
VIDEOⅡ
XS6 PRINT
10pin Rear Panel

RS232C(9pin)
(5) DSC PCB Serial Port
Rear Panel

Connectors connected with FE Board


Besides P2 and P4 connected to DSC, the other two connectors are:

P/N: 4720-0012-01A
29
VISION 4000 Portable Ultrasound System Service Manual

-12V

+12V POWER PCB

+5V

BFM PCB

P1(8pin) P2

BFM PCB DRV PCB

P3 P1
96pin

Wires and cables connected to DRV Board


Besides P1 connected to FE Board, the other two connectors are:

-100V

+100V POWER PCB

+80V

DRV PCB
P3 P2

DRV PCB PBRON PCB

P2 P3
96pin

Connectors connected to PBCON Board


Besides P3 connected to DRV Board, the other two connectors are 156PIN
probe connectors P1 and P2.

P/N: 4720-0012-01A
30
VISION 4000 Portable Ultrasound System Service Manual

5 Assembly /Disassembly of the system

The corresponding PCB of the system can be disassembled according


to the following steps. (See fig2.3A).

1. To disassembly the trackball PCB(7):

a) Remove Rotating Shaft Top on two sides and torque the six bolts(2)
from crossing frame, so that the keyboard subassembly can be
removed.
b) Unscrew two cylindrical head bolts (3), then disassemble the keyboard

lower enclosure(4) and upper enclosure(5)

c) torque Self Tapping Screws(6).


d) Remove the trackball PCB.

2. To disassemble the system Keyboard PCB(11)


a) According to above step 1, remove the keyboard upper enclosure.
b) Take off the seven potentiometer caps(8)
c) Pull out knob (9) forcibly.
d) Unscrew the 15 Self Tapping Screws (10)
e) Remove keyboard PCB(11).

3. To remove the Digital PCB(17)

a) Unscrew 4 pan head screws, 4 round head screws and 2 screws(14)


from foot pad of the system rear cover.
b) Remove rear cover (15)
c) Unscrew 4 screws (16)
d) Remove the digital PCB..

4. To remove the Front-end PCB(19)


According to Step 3, remove the system rear board and unscrew the four
screws (18), so that front end PCB(19) can be removed.

P/N: 4720-0012-01A
31
VISION 4000 Portable Ultrasound System Service Manual

5. disassembly of the probe selection PCB(25)and Pre-amplifier PCB


a) According to step 4, remove front end PCB(19)
b) Pull out connection cable from the pre-amplifier PCB
c) Torque eight screws (20) to remove probe casing subsystem (21)
d) After unscrewing the four screws (22), Pull out Probe selecting PCB
and pre-amplifier PCB backwards.

Caution:
Operator must remove Front End Board (19) and its connection cables
firstly, or the FEB PCB can not be disassembled or its connection port can
be damaged.

6. disassembly of the Monitor PCB(30)

a) Remove the system rear cover (15)


b) Pull out cable connecting Digital PCB and front end PCB
c) Unscrew four screws (26)located in the left and right side
d) Take off the Monitor Pressure Plate(27)
e) Rotate the fixed frame of monitor PCB 90 degree and unscrew the four
screws (29)
f) De-install monitor PCB (30).

7. To disassemble the power PCB(39):


a) Put the system on flat and clean plane and rotate it 90 degree
according to fig .2.3B
b) unscrew the six screws(36)
c) Pull out power housing board
d) Unscrew the four screws (38)
e) Remove power PCB (39).

P/N: 4720-0012-01A
32
- 33 -- 33 - VISION 4000 Portable Ultrasound System Service Manual

FIG2.3A

P/N: 4720-0012-01A
33
- 34 - - 34 - VISION 4000 Portable
Ultrasound System Service Manual

38

39

37

36

FIG2.3B

P/N: 4720-0012-01A
34
VISION 4000 Portable Ultrasound System Service Manual

6 Malfunction Analysis

6.1 Common Malfunction Analysis of Front End Subsystem

The front end is composed of three boards: front end board (BFM Board), probe selection
board (PBCON), pre-amplifier board (DRV). The function of the BFM board is to receive
the control signals from the DSC, and to control transmitting and receiving of the system,
to generate transmitting drive pulse, and to incorporate, amplify and detect the echo
signal. Then send detected echo signals to DSC board. The function of PBCON board is
to connect the probes, switch the probe sockets and select the corresponding probe
array based on the command from DSC. The function of the pre-amplifier is to generate
high voltage required for transmitting, and to pre-amplify the echo signals.

The common failures of Front end board are as follows:

To cause power supply protection


The BFM board provides power through socket P1. The power supplies used on this
board are –12V, +12V, +5V. What’s more, the BFM board can transform –12V to
–6V, from +12V to +6V. At the same time, BFM board transmits –12V, +12V, +5V
to PBCON and DRV through XS4.
In addition, the system supplies the high voltage to the PBCON and DRV through P3
(6-hole socket) on DRV. The voltages are +100V, -100V, +80V.

When the power is in protecting situation (the LEDs on the keyboard are not on and the
ventilation fans are off), a diagnosis can be made following the steps below:
a) Turn off the power supply of the main unit and open the body case
b) Pull out P3 (6-pin socket) on the pre-amplification board and turn on the unit
c) If the system power resumes to work after the above steps, it means the probe
connector board or the pre-amplification board has some failures. Since these two
boards are housed in one metal box, dismantle the whole metal box and replace a new
one.
d) If the system still does not work orderly, power off the unit and pull out P2, then power
on again.
e) If the system power still does not work orderly, it means the cause of the failure lies in
other parts; if the power works again, then the cause of the failure may be in the
following parts: BFM, the PBCON board, the DRV board. The following steps can be
taken to further figure out the failure:
1) Turn off the unit
2) Disconnect the connector P3 (96-PIN) on BFM with DRV board.
3) Connect P1 and power on.
f) If the power recovers to work orderly, it means that the failure is caused by the DRV
P/N: 4720-0012-01A
35
VISION 4000 Portable Ultrasound System Service Manual

board or the PBCON board. Then dismantle the metal board and replace it. If the unit
still cannot work orderly, it means that the power failure is caused by the BFM board.

Power supply is sent via XS2 on the DSC board. There are +12V, -12V and +5V. +5V
is also sent from the board to the keyboard board. It is very likely that current
protection is due to +5V current shortcut, which can be determined by measurement
with a multimeter.

Bad quality of imaging


Because the front end is directly related to quality of ultrasound images, if image quality
is not good, it may be caused by one of the three front-end boards. The bad quality of
images are illustrated as below:

a) Image dots become coarse


The possible cause of dots becoming coarse is the worsening performance of the
amplifier on the BFM board, or the resonance frequency shifts. Therefore, it may be
caused by the BFM board failure.

b) Image depth decreases


The possible causes to make the image depth decreasing are:
A certain level of amplification circuit is inefficient, imaging channels are decreasing, and
transmitting pulses are in disorder. If the image depth decreases, a little metal bar (e.g. a
small screwdriver) can be glided along the probe surface. Under the normal
circumstances, a tower shape bright bar with left-right symmetry will appear on the
screen and shift. If the tower shape is complete, only the display depth is a little smaller
than usual, one that means a certain level of amplification circuit on the BFM board is
damaged, or the transmitting pulses are abnormal (less than the normal quantity).
Therefore, the BFM board should be replaced. If one fixed part of the tower shape is lost,
it means that a certain part from the BFM board is damaged and it needs replacing. If the
tower shape is not complete, and the lost part changes according to the movement of the
tower shape, then it is the DRV or the PBCON board is damaged. At this moment, the
box where these two boards are housed should be taken out and replaced.

Abnormal conditions of imaging


When there is a problem with ultrasound system images, usually it is shown on the image.
The abnormality of images is as follows:
a) There are regular dark strips on the image
There are many reasons that cause the dark strips on the image. The main reasons
are the front end system (including BFM board, PBCON board, DRV board) is
damaged; or the probe itself is defective. There are some symptoms of the dark strips:
a dark strip on a fix position of the image; several dark strips repeatedly appear on the
image.

P/N: 4720-0012-01A
36
VISION 4000 Portable Ultrasound System Service Manual

Because the probe is a kind of precise equipment, improper use may easily
lead to damage. Therefore, when a dark strip appears on the image, firstly
check the probe to see whether it is damaged. The method is very simple.
Connect another good probe (it can be a different model) to the system, if the
similar failure is still on the image, it means it is the system itself with some
trouble, otherwise the problem comes to the probe. If the system is in trouble,
it may be caused by the front end system. The DRV board of the front end
system and the relays on the PBCON board, the high voltage switch and the
pre-amplifier may cause such trouble. Therefore, the metal box housed with
these two boards should be replaced firstly.
b) Noise on the image
There are many reasons that cause noise on the image, e.g. electromagnetic
interference, power interference and interference from the system itself, etc. When
there is any interference appears on the image, remove the external interference
sources firstly. Only when there is no external interference can a further confirmation of
the system failure be made.
External electromagnetic interference can cause the whole screen (not limited to the
ultrasound imaging area) appears being disturbed. If the failure of the front end system
causes the interference, the interference will only appear on the ultrasound imaging
area. Therefore, whether the interference source is from external electromagnetic
interference can be distinguished.
External power interference may also cause noise on the ultrasound imaging areas. If
the outward power interference is detected, the system ground terminal should be
checked to see whether it is well grounded. In the meantime, keep away from the
power interference source.
If the noise is caused by system failure, the location of the noise can be identified roughly
according to the noise feature.

1) Noise appears in tower shape, and there are one or more tower shapes on the image.
It is usually the failure of the pre-amplifier of the DRV board, or high voltage switch or
relay on the PBCON board. The whole metal box should be replaced.
2) Noise appears irregular, and appears in the whole ultrasound imaging area. The BFM
board usually is damaged, and the board should be replaced.
3) Noise appears in irregular geometry figures, such as dark strips, light lines, black
blocks or bright blocks, etc. probably it is not caused by the front end part. Careful
inspection should be done to check whether the DSC board is damaged.
Besides the above listed failure symptoms, there may be other problems due to the
complicated features of the system. If any failure occurs, it is important to note down the
operating environment, operating status and the failure phenomenon. Then contact a
professional service person as soon as possible to troubleshoot the problem.

P/N: 4720-0012-01A
37
VISION 4000 Portable Ultrasound System Service Manual

6.2 Malfunction Analysis of DSC Board

a) Failure: After power on, the icon and character display is normal, but the imaging area
is completely dark.
Causes:
(1) FPGA2 part on the DSC board has some trouble. The DSC board should
be replaced for repair.
(2) D21 part for saving interpolation data on the DSC board is damaged. The
DSC board should be replaced for repair.
b) Failure: Completely dark on the screen when turning on the unit.
Causes: 1) CPLD1 (D34) on the DSC board is damaged. Data may be damaged or
completely lost ; 2) It could be the problem with video out D/A converter (D12) circuit.
The DSC board should be replaced for service.
c) Failure: After turning on the unit, there are some blocks and big dots on the screen.
The keyboard does not work.
Cause: The main controller circuit on the DSC board is damaged. Some parts of the
bus are damaged.
d) Failure: The unit cannot transmit images through RS-232 to computer.
Causes: (1) CPLD2 (D27) on the DSC board is damaged. Its data may be completely
damaged or lost.
(2) The auxiliary controller (D39) on the DSC board is damaged. Its built-in
program data is damaged.
e) Failure: The keyboard does not work, while the cable connection is completely correct.

Cause: The bus buffers D19 and D20 on the DSC board are damaged.
f) Failure: Fail to read the correct probe code. Or “NO PROBE” is displayed on the
screen after the probe is connected.
Cause: D15 on the DSC board is damaged. The DSC board should be replaced for
service. It also could be a problem with the BFM board, which fails to send the probe
code to the DSC board.
g) Failure: the image display is normal. The keyboard can work, but there is no character

display.
Cause: Some circuits of CPLD1 (D34) on the DSC board are damaged.
h) Failure: The unit cannot be powered on. The unit is in protection state when turning on

the power.
Cause: If the unit can be powered on after dismantling the DSC board, it means that
the power supply of the DSC board is short circuit. The DSC board should be replaced
for service.

i) Failure: After turning on the unit, the display screen cannot display the characters after
system initialization in full. What shall be cleared cannot be cleared.
Cause: A problem with the main controller bus of the DSC board, which leads to

P/N: 4720-0012-01A
38
VISION 4000 Portable Ultrasound System Service Manual

improper execution of initialization program. The DSC board should be replaced.

6.3 Malfunction Analysis of power

The power applies to SSI-600 is PA8101 with 110V/220V AC input. There are several
output voltages. They are: +5V, +12V, -12V, +100V, -100V, +80V, +12M, +12F.
When inspecting power failure, it should be inspected to see whether there is any
obvious damage or default parts firstly. Then inspect the fuses, the connection of
every group of power cords. The output cords of each group vary from colors.

Annex: PA8101 output adjusting table

Voltage/Signal Color And Diameter


+5V Red 22#1007
+12V Yellow 22#1007
+12VM Yellow 22#1007
+12VF Yellow 22#1007
-12V Blue 22#1007
+100V Brown 22#1007
-100V White 22#1007
+80V Orange 22#1007
GND(Signal) Green 22#1007
GND(AC) Olivine 18#1015
L(AC) Brown 18#1015
N(AC) Red 18#1015

Measure the output of each group to see if it is within the rating or not. If the output is
abnormal, then disconnect all other circuit. Connect the power board to the relevant load
circuit board. Then to measure the output if it is normal or not. If the output is abnormal, it
means that the power has some failure. Please record the measurement result of each
group of circuit and contact our company for further instruction.

P/N: 4720-0012-01A
39
VISION 4000 Portable Ultrasound System Service Manual

7 Malfunction Analysis of Monitor Board

When the monitor display is abnormal, refer to other chapters of this manual to inspect
the whole unit. As for inspection of video output signals, if possible, connect the video
output of the main unit to an outward monitor to inspect whether the video output signals
are normal or not.
As for the monitor part failure, the diagnosis can be carried out based on the
following aspects (Note: Inspection should be done by technicians, because the
monitor parts contain some high voltage circuit.)
a) Observe the circuit board and other parts to see whether there is any abnormality,
burning, peculiar smell, etc.; and to see whether the down-leads or plugs are shed.
b) Power on and inspect the power input terminal P501, use multimeter to measure if the
input voltage of +12V is normal. If the voltage is abnormal, then pull out the terminal
and measure the terminal voltage without load. If the output still abnormal, it can be
basically judged as the power supply board failure.
c) When the power input and video output signals of the main unit are all normal, it can
be diagnosed primarily as a monitor board failure.

Caution: because hazardous voltages are present inside the monitor, all of
service should be perform by professional technicians.
Some common failures and relevant repair of monitor are illustrated below:
a) No display: adjust the brightness and contrast potentiometers on the front panel to see
if there is any change or if there is a scanning grating. If there is any grating, the
problem may usually lie in the video amplification channel; then observe if the filament
of the display tube is on or not. Furthermore to check if the power supply is normal or
not.
b) If the lines do not display synchronously, try to adjust potentiometer VR201 and VR202.
If they are still not synchronous, it means the circuit has some failure.
c) If the fields do not display synchronously, try to adjust potentiometer VR301. If they are
still not synchronous, it means the circuit has some failure.
d) If the brightness is abnormal, adjust the brightness potentiometer on the front panel, or
adjust the auxiliary brightness potentiometer VR601 on the main circuit board. If it still
has no change, the circuit is damaged.
e) If the contrast is abnormal, adjust the contrast knob on the front panel, or adjust the
auxiliary contrast potentiometer VR601 on the main circuit board. If it still has no
change, the circuit is damaged.
f) If the size of the displayed grating is abnormal, adjust the potentiometer VR303 to
change the image size in vertical direction. Adjust the magnetic core of coil L102, and
the image size in horizontal direction can be changed.
g) If the focusing is not clear, adjust VR104, and see whether there is any change in
image focusing.
If there is any abnormal on brightness, contract, focus, scanning grating, fine adjustment
can be done under the instruction of specialist. After the failure of the circuit board is
P/N: 4720-0012-01A
40
VISION 4000 Portable Ultrasound System Service Manual

confirmed, if possible, repair can be done with reference to the instruction of the circuit
principle and diagrams. Otherwise, the unit should be sent to our company for service.

P/N: 4720-0012-01A
41

You might also like