Determination of PT in Biological Fluids With ICP-MS: Evaluation of Analytical Uncertainty
Determination of PT in Biological Fluids With ICP-MS: Evaluation of Analytical Uncertainty
104
Vol. 25(3), May/June 2004
105
Linearity
The standard addition method
was used for plasma, ultrafiltrate,
and urine matrices.
The calibration curves were
established by plotting the current
intensity (ions/sec) ratio between
195Pt and the internal standard
106
Vol. 25(3), May/June 2004
TABLE III
Accuracy and Precision in the Determination of Pt in Human Plasma Samples,
Evaluated on Three Different Days
Nominal Concentration (µg L ) QC1 (2 µg L–1)
–1
QC2 (41 µg L–1) QC3 (152 µg L–1) LOQ (1 µg L–1)
1st day
Found Concentrations (µg L–1)
Intra-day Mean ± S.D. 2.06 ± 0.02 41.14 ± 0.29 152.16 ± 0.55 1.032 ± 0.005
Intra-day C.V. (%) 0.87 0.70 0.36 0.46
Intra-day Accuracy (%) 103.2 100.3 100.1 103.2
2nd day
Intra-day Mean ± S.D. 2.01 ± 0.04 40.88 ± 0.06 152.14 ±0.32 1.032 ± 0.028
Intra-day C.V. (%) 2.01 0.14 0.21 2.69
Intra-day Accuracy (%) 100.5 99.7 100.1 103.2
3rd day
Intra-day Mean ± S.D. 1.99 ± 0.03 41.17 ± 0.70 151.93 ±0.50 1.037 ± 0.017
Intra-day C.V. (%) 1.57 1.70 0.33 1.64
Intra-day Accuracy (%) 99.7 100.4 99.9 103.7
Inter-day Accuracy and Precision
Mean (µg L–1) 2.02 ± 0.04 41.06 ± 0.40 152.08 ± 0.42 1.034 ± 0.017
C.V. (%) 2.05 0.98 0.28 1.61
Accuracy (%) 101.1 100.2 100.1 104.6
TABLE IV
Accuracy and Precision in the Determination of Pt in Human Plasma Ultrafiltrate Samples,
Evaluated on Three Different Days
Nominal Concentration (µg L ) QC1( 0.02 µg L–1)
–1
QC2 ( 6 µg L–1) QC3 (18 µg L–1) LOQ (0.001 µg L–1)
1st day
Found Concentrations (µg L–1)
Intra-day Mean ± S.D. 0.019 ± 0.001 6.024 ± 0.142 17.78 ± 0.01 0.0098 ± 0.0008
Intra-day C.V. (%) 3.36 2.36 0.03 8.10
Intra-day Accuracy (%) 97.50 100.40 98.78 98.0
2nd day
Intra-day Mean ± S.D. 0.022 ± 0.001 5.975 ± 0.070 18.02 ± 0.14 0.0108 ± 0.0007
Intra-day C.V. (%) 5.22 1.18 0.77 6.52
Intra-day Accuracy (%) 108.00 99.59 100.13 107.7
3rd day
Intra-day Mean ± S.D. 0.0201 ± 0.0003 5.981 ± 0.061 18.06 ± 0.24 0.0108 ± 0.0006
Intra-day C.V. (%) 1.55 1.02 1.33 5.56
Intra-day Accuracy (%) 103.83 99.69 100.31 108.0
Inter-day Accuracy and Precision
Mean ± S.D. 0.021 ± 0.001 5.993 ± 0.088 17.95 ± 0.19 0.0105 ± 0.0008
C.V. (%) 5.51 1.47 1.06 7.49
Accuracy (%) 103.11 99.89 99.74 104.6
107
TABLE V
Accuracy and Precision in the Determination of Pt in Human Urine Samples,
Evaluated on Three Different Days
–1
Nominal Concentration (µg L ) QC1( 5 µg L–1) QC2 (40 µg L–1) QC3 (160 µg L–1) LOQ (2 µg L–1)
1st day
Found Concentrations (µg L-1)
Intra-day Mean ± S.D. 5.037 ± 0.010 40.07 ± 0.09 160.12 ± 0.42 2.091 ± 0.012
Intra-day precision (C.V.%) 0.19 0.23 0.26 0.58
Intra-day Accuracy (mean found/added %) 100.7 100.2 100.1 104.6
2nd day
Intra-day Mean ± S.D. 5.009 ± 0.004 40.05 ± 0.05 159.96 ± 0.94 2.068 ± 0.017
Intra-day precision (C.V.%) 0.08 0.11 0.59 0.81
Intra-day Accuracy (mean found/added %) 100.2 100.1 99.9 103.4
3rd day
Intra-day Mean ± S.D. 5.021 ± 0.083 39.98 ± 0.08 159.94 ± 0.33 2.119 ± 0.012
Intra-day precision (C.V.%) 1.66 0.21 0.21 0.57
Intra-day Accuracy (mean found/added %) 100.4 99.9 100.0 105.9
Inter-day Accuracy and Precision
Mean ± S.D. 5.023 ± 0.044 40.032 ± 0.078 160.01 ± 0.55 2.093 ± 0.025
Precision (C.V.%) 0.87 0.20 0.34 1.20
Accuracy (mean found/added %) 100.46 100.08 100.00 104.6
Urine 0.59 (with only one value higher Uncertainty Evaluation From
Matrix-matched calibration was than 1%), and intra-day accuracy, Validation Data
achieved by adding dilute Pt stan- calculated as the percentage recov- The validation study of the pre-
dard solution to pooled urine sam- ery of the spiked samples, ranged sent assay was implemented by
ples, previously centrifuged to on average between 99.9 and assessing the uncertainty evaluation
remove the sediment, and kept at 100.7%. Precision lower than 1% for Pt in the different matrices at
–20oC until use. and accuracy very close to 100% the four levels of concentration.
were achieved during the inter-day According to the EURACHEM/
The mean slope and intercept evaluation. Accuracy and precision CITAC Guide (23), the combined
values of the calibration curves for urine samples at the LOQ con- uncertainty (ucombined) has been cal-
(0–200 µg/L) obtained over a three- centration are reported in Table V. culated by estimating the compo-
day validation period were: slope = nents associated with the
0.128 ± 0.04 (CV= 2.97%) and inter- The results obtained in the pre-
sent study meet to the FDA criteria repeatability (urepetability), the instru-
cept=0.033 ± 0.041 with a correla- mental calibration (ucalibration), and
tion coefficient better than 0.9999. for the validation of analytical meth-
ods (3). For each concentration the sampling (usampling). The uncer-
The detection limit (LOD) was level, the mean value is within tainty component due to the recov-
0.002 µg L–1 with a minimum quan- ±15% of the nominal value (at the ery (urecovery) was not considered
tifiable platinum concentration LOQ within ±20%), and the coeffi- because the calibration curve was
(LOQ) set at 2 µg L–1. cient of variation (CV) around the performed by standard addition
mean value is lower than ±15% (at method (in the same matrix as the
Intra-day and inter-day accuracy sample), determining the analytes
the LOQ ±20% for the CV is
and precision of the method for the according to the procedure
accepted).
determination of platinum in described in the Experimental sec-
human urine samples are reported These results prove that, in tion.
in Table V. keeping with international
standards, the method is adequate Consequently, the equation that
Intra-day precision, expressed as best represents the mathematical
for the assay of platinum in human
CV%, ranged between 0.08 and model upon which the relative
plasma and urine samples.
108
Vol. 25(3), May/June 2004
.
u(Comb.) =
. . .
√(u(Repeat))2+(u(Calib.))2+(u(Samp.))2
where:
.
u(Comb) = combined relative uncer-
tainty
.
u(Repeat.) = relative uncertainty of
repeatability
.
u(Calib.)= relative uncertainty of cali-
bration
. Fig. 4. Relative Uncertainty Values (%) obtained at different concentration levels
for Pt in human plasma.
u(samp.) = relative uncertainty of sam-
pling
In order to investigate the contri-
butions of all components to the
combined relative uncertainty as a
function of Pt concentration (in
plasma, ultrafiltrate, and urine),
these uncertainties were evaluated
and plotted in Figures 4, 5, and 6,
respectively.
Whit respect to Pt in plasma and
urine, the graphs show that the rela-
tive uncertainty ranges from 4–7% at
1–2 µg L–1 and go down to 2–3% for
the higher concentrations (40–150
µg L–1).
While the uncertainty component
associated with the repeatability is
constant (0.5–2%), overall the differ-
ent concentration levels and the dif-
ferent days, the component due to
the instrumental calibration
increases significantly (7–8%) rela-
tive to the lowest concentrations of
analyte.
With respect to Pt in the plasma Fig. 5. Relative Uncertainty Values (%) obtained at different concentration levels
ultrafiltrate, Figure 5 shows that the for Pt in human plasma ultrafiltrate.
relative uncertainty ranges from
4–5% at 8–16 µg L–1 and increases up
109
to 20–30% for the lower concentra-
tions (at 0.01–0.02 µg L–1).
This occurrence means that the
quantitative determination of low
concentrations of Pt in the ultrafil-
trate is greatly affected by the
uncertainty assignable to the instru-
mental calibration as done in the
present study (0–20 µg L–1). Further
tests performed show that by
reducing the calibration range but
keeping the same number of con-
centration levels (n=6) gives better
values of combined uncertainty as a
consequence of a significant
improvement of the uncertainty
component due to calibration. The
mathematical function (9) used for
the calculation of uncertainty of
(xpred) due to calibration is the fol-
lowing: _
s2y/q 1 1 (xpred – xCal)2
u(xpred)= + + _
b √ m n Σ(xi – xCal)2
Fig. 6. Relative Uncertainty Values (%) obtained at different concentration levels
for Pt in human urine.
where:
s2y/q is the residual variance of our study, with a calibration range regression line or make more than
the regression model of 0–20 µg L–1 and a centroid corre- one measurement and use the mean
b is the slope of the regression .
sponding to (5.18; 1.84), the
u(Calib.) for a predicted mean value
value in the calculation of the
predicted value.
curve
of 0.36 µg L–1 Pt in the ultrafiltrate
m is the number of repetitions (n=3 replicates) was about 21.9%; Analysis of Real Samples
from which is derived the the same results became lower The present method was
predicted value (0.22%) when the calibration range employed to determine the pharma-
was set at 0–2 µg L–1 with the same cokinetic profiles of total Pt and
n is the number of calibration number of calibration points and
points on the regression line ultrafiltrable Pt in patients who
with a centroid corresponding to received JM216 (an oral bis
x(pred) is the predicted value (0.527, 0.163). (acetate) amine dichloro cyclohexy-
These results show the impor- lammine platinum (IV) analogue)
x(Cal) is the mean value of the tance of calibration in evaluating brought into clinical development
concentration levels used in the the measurement uncertainty of for phase II evaluation. Forty-six
calibration curve low Pt concentrations in biological patients were treated at doses rang-
Inspection of the previously fluids. Keeping the repeatability of ing from 10 mg/m2/d to 50
reported equation confirms that measurements constant as, in the mg/m2/d and 39 patients were eval-
as xpred approaches xCal, the third case of a wider calibration range, uated for hematologic toxicity over
term under the square root we should accept larger uncertainty 74 cycles. The pharmacokinetics of
u(xpred)approaches zero and thus of calibration which will reflect sig- total and ultrafilterable Pt were
approaches a minimum value. In a nificantly on the combined uncer- studied on days 1 and 14 of the first
practical analysis, therefore, a cali- tainty. cycle, and the results were recently
bration of this type will give the published by Sessa (3). Figure 7
If we wished to improve (i.e., shows, as example, the pharmacoki-
most precise results when the mea- narrow) the confidence limits of
sured instrument signal corresponds netics profiles of four patients who
calibration, we could reduce the received 45 mg/m2/d of agent.
to a point close to the centroid of calibration range, increase the num-
the regression line (xCal, yCal, ). In ber of calibration points on the
110
Vol. 25(3), May/June 2004
Fig. 7. Pharmacokinetics profiles of four patients treated with 45 mg/m2/d of agent. Plasma levels of total platinum (Pt) and
plasma ultrafiltrate platinum (UPt).
JM216 was rapidly absorbed, The present study was implemented REFERENCES
with Pt and UPt detectable in by assessing the uncertainty evalua- 1. J.A. Gottlieb and B. Drewinko, Cancer
plasma as early as 30 min after the tion for Pt determination in the dif- Treat. Rep. 59, 621 (1975).
administration. On day 1, Pt and ferent matrices according to the 2. A.B. Khan, B.J. D’Souza, M.D.
UPt peak levels were achieved in EURACHEM/CITAC Guide. The Wharam,L.A.A. Champion, L.F. Sinks,S.Y.
combined uncertainty, calculated Woo, D.C. McCullough, and B.C. Leven-
most cases within 3 hours with thal, Cancer Treat. Rep. 66, 2013 (1982).
median Tmax values of 2.5 and 2, by estimating the components asso- 3. C. Sessa, C. Minoia A. Ronchi, M. Zuc-
respectively. Elimination was vari- ciated with the repeatability, the chetti, J. Bauer, M. Corner, M., J. de Joung,
able and slow with Pt and UPt con- instrumental calibration and the O. Pagani, J. Renard, C. Weil, C., and M.
centrations of, respectively, 100 µg sampling, was always lower than DÅfincalci, Annals of Oncology 9, 1315
(1988).
L–1 and 20 µg L–1 and still 10% for Pt concentrations higher
4. U.S. Pharmacopeia 23, 1982-84, United
detectable two weeks after the last than 1 µg/L. States Pharmacopeial Convention, Inc.,
dosing. (1994).
While the uncertainty compo-
5. International Conference on Harmoniza-
nent associated to the repeatability tion, Draft Guidance on Validation of Ana-
CONCLUSION
was constant overall, the different lytical Procedures: Definitions and
An ICP-MS method for the deter- concentration levels and the sam- Terminology, Federal Register, Volume
60, pp. 1260 (March 1, 1995)
mination of Pt in biological fluids pling uncertainty was negligible,
6. FDA – Guidance for Industries, Bioanalyti-
(plasma, ultrafiltrate, and urine) of the component due to the instru- cal Method Validation, U.S. Department of
patients treated with antitumor mental calibration increased signifi- Health and Human Services Food and
agents has been developed and vali- cantly in relation to the lowest Drug Administration (May 2001).
dated. The limits of quantification concentrations of analyte. 7. ISO, Guide to the Expression of Uncer-
tainty in Measurement, International Stan-
(LOQ) in the three matrices were dards Organization, Geneva, Switzerland
1.0, 0.1, and 2.0 µg L–1, (1993).
respectively. Received December 23, 2003. 8. EURACHEM/CITAC Guide (2000), Quanti-
fying Uncertainty in Analytical Measure-
Intra-day and inter-day precision ment, 2nd Edition.
and accuracy was in good agree- 9. S.J. Bannister, Y. Chang, L.A. Sterson, and
ment with the FDA criteria for the A.J. Repta, Clin. Chem 24, 877 (1978).
validation of analytical methods. 10. E.H. Evans and J.J. Giglio, J. Anal. At.
Spectrom. 8, 1 (1993).
11. V.J. Berwick and S.L.R. Ellison, Accred.
Qual. Assur. 5, 47 (2000).
111