Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Designation: D6377 − 16

Standard Test Method for


Determination of Vapor Pressure of Crude Oil:
VPCRx (Expansion Method)1
This standard is issued under the fixed designation D6377; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope* D4057 Practice for Manual Sampling of Petroleum and


1.1 This test method covers the use of automated vapor Petroleum Products
pressure instruments to determine the vapor pressure exerted in D4177 Practice for Automatic Sampling of Petroleum and
vacuum of crude oils. This test method is suitable for testing Petroleum Products
samples that exert a vapor pressure between 25 kPa and D5191 Test Method for Vapor Pressure of Petroleum Prod-
180 kPa at 37.8 °C at vapor-liquid ratios from 4:1 to 0.02:1 (X ucts (Mini Method)
= 4 to 0.02). D5853 Test Method for Pour Point of Crude Oils
NOTE 1—This test method is suitable for the determination of the vapor D6708 Practice for Statistical Assessment and Improvement
pressure of crude oils at temperatures from 0 °C to 100 °C and pressures of Expected Agreement Between Two Test Methods that
up to 500 kPa, but the precision and bias statements (see Section 14) may Purport to Measure the Same Property of a Material
not be applicable.
1.2 This test method also allows the determination of vapor 3. Terminology
pressure for crude oil samples having pour points above 15 °C 3.1 Definitions:
provided the proper sample handling, transfer, and analysis
3.1.1 platinum resistance thermometer, n—temperature
procedures are followed.
measuring device constructed with a length of platinum wire,
1.3 The values stated in SI units are to be regarded as whose electrical resistance changes in relation to temperature.
standard. No other units of measurement are included in this 3.1.2 vapor-liquid ratio (V/L), n—the ratio of the vapor
standard. volume to the liquid volume of specimen, in equilibrium, under
1.4 This standard does not purport to address all of the specified conditions.
safety concerns, if any, associated with its use. It is the
3.2 Definitions of Terms Specific to This Standard:
responsibility of the user of this standard to establish appro-
3.2.1 dead crude oil, n—crude oil with sufficiently low
priate safety and health practices and determine the applica-
vapor pressure that, when exposed to normal atmospheric
bility of regulatory limitations prior to use. For specific
pressure at room temperature, does not result in boiling of the
warning statements, see 7.2.1 – 7.3.2.
sample.
2. Referenced Documents 3.2.1.1 Discussion—Sampling and handling of dead crude
2.1 ASTM Standards:2 oils can usually be done without loss of sample integrity or
D323 Test Method for Vapor Pressure of Petroleum Products other problems by using normal, non-pressurized sample
(Reid Method) containers such as cans.
D2892 Test Method for Distillation of Crude Petroleum 3.2.2 live crude oil, n—crude oil with sufficiently high vapor
(15-Theoretical Plate Column) pressure that it would boil if exposed to normal atmospheric
D3700 Practice for Obtaining LPG Samples Using a Float- pressure at room temperature.
ing Piston Cylinder 3.2.2.1 Discussion—Sampling and handling live crude oils
requires a pressurized sample system and pressurized sample
1
This test method is under the jurisdiction of ASTM Committee D02 on containers to ensure sample integrity and prevent loss of
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---

Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of volatile components.
Subcommittee D02.08 on Volatility.
Current edition approved April 1, 2016. Published April 2016. Originally
3.2.3 Reid vapor pressure equivalent (RVPE), n—a value
approved in 1999. Last previous edition approved in 2015 as D6377 – 15. DOI: calculated by a defined correlation equation (see Eq X1.1) from
10.1520/D6377-16. VPCR4 at 37.8 °C that is expected to be equivalent to the vapor
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or pressure value obtained by Test Method D323.
contact ASTM Customer Service at [email protected]. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on 3.2.3.1 Discussion—As noted in 14.3 and Appendix X1,
the ASTM website. subsection X1.2, since Eq X1.1 for estimating RVPE is not

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

Copyright ASTM International


Provided by IHS under license with ASTM
1Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
universally applicable to all crude oils, it is recommended to shall contain a movable piston with a minimum dead volume of
report the VPCR4 (37.8 °C) result for a crude oil sample. less than 1 % of the total volume at the lowest position to allow
3.2.4 vapor pressure of crude oil (VPCR x), n—the pressure sample introduction into the measuring chamber and expansion
exerted in an evacuated chamber at a vapor-liquid ratio of X:1 to the desired V/L. A static pressure transducer shall be
by conditioned or unconditioned crude oil, which may contain incorporated in the piston. The measuring chamber shall
gas, air or water, or a combination thereof, where X may vary contain an inlet/outlet valve combination for sample introduc-
from 4 to 0.02. tion and expulsion. The piston and the valve combination shall
be at the same temperature as the measuring chamber to avoid
3.3 Abbreviations: any condensation or excessive evaporation.
3.3.1 ARV, n—accepted reference value 6.1.1 The measuring chamber shall be designed to have a
3.3.2 RVPE, n—Reid vapor pressure equivalent total volume of 5 mL to 15 mL and shall be capable of
3.3.3 V/L, n—vapor liquid ratio maintaining a V/L of 4:1 to 0.02:1. The accuracy of the
3.3.4 VPCRx, n—vapor pressure of crude oil at x vapor adjusted V/L shall be within 0.01.
liquid ratio NOTE 3—The measuring chambers employed by the instruments used in
generating the precision and bias statements were constructed of nickel
4. Summary of Test Method plated aluminium, stainless steel and brass with a total volume of 5 mL.
Measuring chambers exceeding a 5 mL capacity and different design can
4.1 Employing a measuring chamber with a built-in piston, be used, but the precision and bias statement (see Section 14) may not be
a sample of known volume is drawn from the sample container applicable.
into the temperature controlled chamber at 20 °C or higher. 6.1.2 The pressure transducer shall have a minimum opera-
After sealing the chamber, the volume is expanded by moving tional range from 0 kPa to 500 kPa with a minimum resolution
the piston until the final volume produces the desired V/L of 0.1 kPa and a minimum accuracy of 60.5 kPa. The pressure
value. The temperature of the measuring chamber is then measurement system shall include associated electronics and
regulated to the measuring temperature. readout devices to display the resulting pressure reading.
4.2 After temperature and pressure equilibrium, the mea- 6.1.3 Electronic temperature control shall be used to main-
sured pressure is recorded as the VPCRX of the sample. The tain the measuring chamber at the prescribed temperature
test specimen shall be mixed during the measuring procedure within 60.1 °C for the duration of the test.
by shaking the measuring chamber to achieve pressure equi- 6.1.4 A platinum resistance thermometer shall be used for
librium in a reasonable time of 5 min to 30 min. measuring the temperature of the measuring chamber. The
minimum temperature range of the measuring device shall be
4.3 For results related to Test Method D323, the final
from 0 °C to 100 °C with a resolution of 0.1 °C and an
volume of the measuring chamber shall be five times the test
accuracy of 60.1 °C.
specimen volume and the measuring temperature shall be
6.1.5 The vapor pressure apparatus shall have provisions for
37.8 °C.
rinsing the measuring chamber with the next sample to be
5. Significance and Use tested or with a solvent of low vapor pressure.
6.1.6 The vapor pressure apparatus shall have provisions for
5.1 Vapor pressure of crude oil at various V/Ls is an shaking the sample during the measuring procedure with a
important physical property for shipping and storage. minimum frequency of 1.5 cycles per second.
NOTE 2—A vapor-liquid ratio of 0.02:1 (X = 0.02) mimics closely the
situation of an oil tanker. 6.2 Vacuum Pump for Calibration, capable of reducing the
5.2 Vapor pressure of crude oil is important to crude oil pressure in the measuring chamber to less than 0.01 kPa
producers and refiners for general handling and initial refinery absolute.
treatment. 6.3 McLeod Vacuum Gage or Calibrated Electronic Vacuum
5.3 The vapor pressure determined by this test method at a Measuring Device for Calibration, to cover at least the range
vapor-liquid ratio of 4:1 (VPCR4) of crude oil at 37.8 °C can be of 0.01 kPa to 0.67 kPa. The calibration of the electronic
related to the vapor pressure value determined on the same measuring device shall be regularly verified in accordance with
material when tested by Test Method D323 (see Appendix X1). Annex A of Test Method D2892.
5.4 Air saturation of crude oil shall not be done to avoid 6.4 Pressure Measuring Device for Calibration, capable of
potential vapor loss. However air saturation of the chilled measuring local station pressure with an accuracy and a
verification fluid is mandatory (see 7.2 and Section 11). resolution of 0.1 kPa or better, at the same elevation relative to
sea level as the apparatus in the laboratory.
5.5 This test method can also be applied in online applica-
tions. NOTE 4—This standard does not give full details of instruments suitable
for carrying out this test. Details on the installation, operation and
maintenance of each instrument may be found in the manufacturer’s
6. Apparatus manual.
6.1 The apparatus suitable for this test method employs a
small volume, cylindrically shaped measuring chamber with 7. Reagents and Materials
associated equipment to control the chamber temperature 7.1 Purity of Reagents—Use chemicals of at least 99 %
within the range from 0 °C to 100 °C. The measuring chamber
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---
purity for verification of instrument performance (see Section

Copyright ASTM International


Provided by IHS under license with ASTM
2Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
TABLE 1 Accepted Reference Value (ARV) and Acceptable Testing Range for Air Saturated Reference FluidsA
Recommended Instrument Acceptable Testing Range
ARV [Ptot] ± Uncertainty,
Reference Fluid Manufacturer Tolerance, for Reference Fuel [Ptot],
(kPa)
(kPa) (kPa)
Pentane 112.8 ± 0.2 ± 1.0 112.8 ± 1.2 (111.6 to 114.0)
2,2 Dimethylbutane 74.1 ± 0.2 ± 1.0 74.1 ± 1.2 (72.9 to 75.3)
2,3 Dimethylbutane 57.1 ± 0.2 ± 1.0 57.1 ± 1.2 (55.9 to 58.3)
A
Supporting data have been filed at ASTM International Headquarters and may be obtained by requesting Research Report RR:D02-1619.

11). Unless otherwise indicated, it is intended that all reagents with Practice D3700. Sampling in accordance with Practice
conform to the specifications of the Committee on Analytical D4057 shall only be used for dead crude oil and if Practice
Reagents of the American Chemical Society where such D3700 is impractical.
specifications are available.3 Other grades may be used, pro-
NOTE 5—Sampling in accordance with Practice D4177 may also be
vided it is first ascertained that the reagent is of sufficient purity used instead of Practice D4057.
to permit its use without lessening the accuracy of the
determination. 8.1.2 For sampling in accordance with Practice D3700, a
7.1.1 The chemicals in 7.2.1, 7.2.2, and 7.2.3 are suggested floating piston cylinder with a minimum sample volume of
for verification of instrument performance (see Section 11), 200 mL shall be used if the overall volume of the test specimen
based on the air saturated reference fuels analyzed in the Test required for the vapor pressure determination, including the
Method D5191 2003 interlaboratory study (ILS)4 (see Table 1). rinsing procedure is not larger than 20 mL. Larger floating
Ptot used in Test Method D5191 is equivalent to VPCR4 used in piston cylinders can be used. The minimum piston back-
this test method). Such reference fuels are not to be used for pressure shall be higher than the sample vapor pressure at the
instrument calibration. Table 1 identifies the accepted reference introduction temperature of the measuring chamber plus
value (ARV) and uncertainty limits, as well as the acceptable 100 kPa for the shifting of the piston. The maximum back-
testing range for each of the reference fuels listed. pressure shall not exceed the maximum measurement pressure
7.1.2 The chemicals in 7.3.1 and 7.3.2 are suggested for use of the apparatus pressure transducer. Compressed air, or any
as rinsing solvents capable of cleaning the measuring chamber, other, non-flammable compressed gas, can be used as the
the valves and the inlet and outlet tubes. back-pressuring agent. The floating piston cylinder shall have
provisions for mechanical stirring of the sample and a second
7.2 Verification Fluids: valve at the inlet for rinsing.
7.2.1 2,2-Dimethylbutane, (Warning—2,2-dimethylbutane 8.1.3 Do not unnecessarily expose the samples to tempera-
is flammable and a health hazard.) tures exceeding 30 °C during sampling and storage. For
7.2.2 2,3-Dimethylbutane, (Warning—2,3-dimethylbutane prolonged storage, store the samples in an appropriate room or
is flammable and a health hazard.) refrigerator.
7.2.3 Pentane, (Warning—Pentane is flammable and a 8.1.4 Perform the vapor pressure determination on the first
health hazard.) test specimen withdrawn from the cylinder after the rinsing
7.3 Solvents: step in 9.3. Do not use the remaining sample in the floating
7.3.1 Toluene, (Warning—Toluene is flammable and a piston cylinder for more than three repeat vapor pressure
health hazard.) determinations.
7.3.2 Acetone, (Warning—Acetone is flammable and a 8.1.5 In the case of quality control checks with pure
health hazard.) compounds (see 11.1), smaller sample containers without
applied pressure can be used.
8. Sampling and Sample Introduction 8.2 Sampling Procedures:
8.1 General Requirements: 8.2.1 If the sample is contained in a pressurized source like
8.1.1 The extreme sensitivity of vapor pressure measure- a pipeline, use a floating piston cylinder and obtain the sample
ments to losses through evaporation and the resulting changes directly from the source under pressure. Rinse the cylinder by
in composition requires the utmost precaution and the most opening the rinsing valve until the crude oil emerges at the
meticulous care in the drawing and handling of samples. second inlet. Close the rinsing valve and let the piston move
Sampling of live crude oil shall be performed in accordance slowly until at least 200 mL of sample has entered the cylinder.
Close the inlet valve and apply the back pressure immediately.
Check the filling of the cylinder to be at least 200 mL.
3
Reagent Chemicals, American Chemical Society Specifications , American
NOTE 6—The current precision statements were derived from the 2005
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
ILS using samples in 250 mL floating piston cylinders (see 14.3).
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia 8.2.2 If the sample is taken from a non-pressurized source
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
like a storage tank, oil tanker, drum or other container, obtain
MD.
4
Supporting data have been filed at ASTM International Headquarters and may a sample and test specimen in accordance with Practice D4057.
be obtained by requesting Research Report RR:D02-1619. Use either a 250 mL or 1 L sized container filled between 70 %
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
3Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
and 80 % with sample. For best testing precision concerning the calibration of electronic vacuum measuring devices and
(reproducibility), it is recommended that a 1 L sized container proper maintenance of McLeod gages.
be used. 10.1.3 Open the measuring chamber of the apparatus to
NOTE 7—The current precision statements were derived from the 2005 atmospheric pressure, and observe the corresponding pressure
ILS using samples in 1 L amber glass containers (see 14.3). However, value of the transducer. Ensure that the apparatus is set to
samples in containers of other sizes, as prescribed in Practice D4057, may
be used with the same ullage requirement if it is recognized that the display the total pressure and not a calculated or corrected
precision can be affected. value. Compare this pressure value with the pressure obtained
8.3 Sample Transfer: from a pressure measuring device (see 6.4), as the pressure
8.3.1 Transfer the sample at room temperature but at least reference standard. The pressure measuring device shall mea-
5 °C above the pour point (as determined by Test Method sure the local station pressure at the same elevation as the
D5853) of the sample from the container into the measuring apparatus in the laboratory at the time of pressure comparison.
chamber. When the instrument is used over the full pressure range, a
8.3.2 If the sample is contained in a pressurized floating calibration with a dead weight balance shall be carried out.
piston cylinder, apply a back-pressure which is higher than the (Warning—Many aneroid barometers, such as those used at
vapor pressure of the sample at the introduction temperature weather stations and airports, are pre-corrected to give sea
plus a minimum of 100 kPa for the piston movement. The level readings; these shall not be used for calibration of the
applied back-pressure shall not exceed the maximum limit of apparatus.)
the pressure transducer used in the vapor pressure apparatus. 10.1.4 Repeat 10.1.2 and 10.1.3 until the zero and baromet-
8.3.3 If the sample is contained in a nonpressurized ric pressures read correctly without further adjustments.
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---

container, transfer the sample into the measuring chamber 10.2 Temperature Sensor—Verify the calibration of the
using a transfer tube or a syringe. platinum resistance thermometer used to monitor the measur-
ing chamber temperature at least every six months against a
9. Preparation of Apparatus
thermometer which is traceable to National Institute of Stan-
9.1 Prepare the instrument for operation in accordance with dards and Technology (NIST) or to national authorities in the
the manufacturer’s instructions. country the equipment is used.
9.2 Rinse the measuring chamber, if necessary with a
solvent. Toluene has a low vapor pressure and can be used 11. Quality Control Checks
successfully. Rinsing is performed by drawing the solvent into 11.1 Use a verification fluid (see 7.2) of known volatility as
the chamber by the piston and expelling the solvent into the an independent check against the instrument calibration each
waste container. day the instrument is in use. For pure compounds, multiple test
9.3 To avoid contamination of the test specimen with the specimens may be taken from the same container over time.
previous sample or the solvent, rinse the measuring chamber a Air saturate the verification fluid at temperatures between 0 °C
minimum of three times with the sample to be tested. Fill the to 1 °C as described in Test Method D5191 (see 8.4). Transfer
measuring chamber with sample to at least half the total the verification fluid into the measuring chamber using a
volume of the chamber for each rinse. This rinsing procedure transfer tube or a syringe. The temperature of the verification
shall always be carried out immediately before the measuring fluid shall be at 0 °C to 3 °C during the sample introduction,
procedure (see Section 12). and the measuring procedure shall be in accordance with
Section 12 with a V/L ratio of 4:1 and a measuring temperature
10. Calibration of 37.8 °C.
10.1 Pressure Transducer: 11.2 Table 1 provides the accepted reference value (ARV)
10.1.1 Perform a calibration check of the transducer as [Ptot] and uncertainty limits (at least 95 % confidence interval)
indicated from the quality control checks (see Section 11). The of reference fluids tested in the 2003 D5191 ILS.4 As stated in
calibration of the transducer is checked using two reference 7.1.1, the “Ptot” reference used in Test Method D5191 (and
points, zero pressure (<0.1 kPa) and the ambient barometric Table 1 of this test method) is equivalent to VPCR4 (37.8 °C)
pressure. in this test method. This information, combined with the
NOTE 8—Calibration frequency of the pressure transducer may vary tolerance value recommended by instrument manufacturers,
with instrument type and frequency of use. A calibration check of the was used to establish the acceptable testing range for the
pressure transducer at least once every six months is recommended. reference fuels to verify instrument performance.
10.1.2 Connect a McLeod gage or a calibrated electronic 11.3 Values obtained within the acceptable testing range
vacuum measuring device to the vacuum source in line with the intervals in Table 1 indicate that the instrument is performing
measuring chamber (see Note 7). Apply vacuum to the mea- at the level deemed acceptable by this standard. If values
suring chamber (see Note 9). When the vacuum measuring outside the acceptable testing range intervals are obtained,
device registers a pressure less than 0.1 kPa, adjust the verify the quality of the pure compound(s) and re-check the
transducer control to zero or to the actual reading on the calibration of the instrument (see Section 10). (Warning—The
vacuum measuring device as dictated by the instrument design use of single component verification materials such as those
or manufacturer’s instructions. listed in Table 1 will only prove the calibration of the
NOTE 9—Refer to Annex A6.3 of Test Method D2892 for further details equipment. It will not check the accuracy of the entire test

Copyright ASTM International


Provided by IHS under license with ASTM
4Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
method, including sample handling, because losses due to 12.8 Wait for temperature equilibrium between measuring
evaporation will not decrease the sample vapor pressure as chamber and specimen, and monitor the total vapor pressure
happens with losses of light ends in multi component mixtures. every 30 s 6 5 s. When three successive readings agree within
The vapor pressure measurement process (including operator 0.3 kPa, record this resulting vapor pressure as VP-
technique) can be checked periodically by performing this test CRX (Tm°C).
method on previously prepared samples from one batch of
product, as per procedure described in 8.1.2. Samples should 13. Report
be stored in an environment suitable for long term storage 13.1 Report the results to the nearest 0.1 kPa and specify the
without sample degradation. Analysis of result(s) from these test temperature and vapor-liquid ratio.
quality control samples can be carried out using control chart
techniques.5) VPCRX ~ Tm°C ! 5 ##.# kPa (1)
NOTE 10—It is recommended that at least one type of verification fluid where:
used in 11.1 have a vapor pressure representative of the crude(s) regularly
tested by the equipment. X = vapor-liquid ratio, and
Tm = measuring temperature.
12. Procedure
12.1 Set the sample introduction temperature of the measur- 14. Precision and Bias
ing chamber. 14.1 Precision—The precision of this test method as deter-
12.1.1 For samples contained in a pressurized floating mined by the statistical examination of interlaboratory test
piston cylinder, set the sample introduction temperature of the results is as follows:
measuring chamber between 20 °C and the measuring tempera- 14.1.1 Repeatability—The difference between successive
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---

ture. test results obtained by the same operator with the same
12.1.2 For samples contained in a non-pressurized apparatus under constant operating conditions on identical test
container, set the sample introduction temperature of the material would, in the long run, in the normal and correct
measuring chamber to 20 °C. operation of the test method, exceed the values calculated as
12.1.3 For samples with a pour point higher than 15 °C, set per the following equations only in one case in twenty:
the sample introduction temperature of the measuring chamber 14.1.1.1 For samples in pressurized Floating Piston Cylin-
at least 5 °C above the pour point temperature of the sample ders:
but no higher than the measuring temperature. V/L 5 4 and Tm 5 37.8°C:repeatability 5 2.48 kPa ~ 37.8°C ! (2)
12.1.4 For measurements with V/L ratios < 1, the sample
may not be exposed to the atmosphere and shall be contained V/L 5 0.02 and Tm 5 37.8°C:repeatability 5 5.61 kPa ~ 37.8°C ! (3)
in a floating piston cylinder. The sample introduction tempera- 14.1.1.2 For samples in nonpressurized sample containers:
ture of the measuring chamber shall be equal to the measuring
V/L 5 4 and Tm 5 37.8°C:repeatability 5 2.29 kPa ~ 37.8°C ! (4)
temperature to avoid any influence due to sample expansion.
12.2 Set the V/L to the desired value X:1 (for test results 14.1.2 Reproducibility—The difference between two single
related to Test Method D323, set the V/L to 4:1). and independent results obtained by different operators work-
ing in different laboratories on identical material would, in the
12.3 If the sample is contained in a pressurized floating long run, in the normal and correct operation of the test
piston cylinder, mix it vigorously with the mechanical stirrer, method, exceed the following values in only one case in
otherwise shake the container with the non-pressurized sample, twenty:
to ensure a homogenous sample. 14.1.2.1 For samples in pressurized Floating Piston Cylin-
12.4 Follow the manufacturer’s instructions for introducing ders:
the test specimen into the measuring chamber. The volume of V/L 5 4 and Tm 5 37.8°C:reproducibility 5 4.26 kPa ~ 37.8°C ! (5)
the specimen shall be such that after the expansion to the final
V/L 5 0.02 and Tm 5 37.8°C:reproducibility 5 20.3 kPa ~ 37.8°C !
volume the programmed V/L is achieved.
(6)
12.5 After closing the inlet valve, expand the volume of the
measuring chamber to the final volume. 14.1.2.2 For samples in nonpressurized sample containers:
12.6 Switch-on the shaker, and leave it on during the entire V/L 5 4 and Tm 5 37.8°C:reproducibility 5 5.26 kPa ~ 37.8°C ! (7)
measuring procedure. 14.2 Bias—Since there is no accepted reference material
12.7 Adjust the temperature control to the measuring tem- suitable for determining the bias for the procedures in this test
perature (for results related to Test Method D323, adjust to a method, a bias cannot be determined.
temperature of 37.8 °C). The measuring temperature shall not 14.3 Relative Bias to Test Method D323—A statistically
be lower than at least 10 °C above the pour point temperature significant relative bias was observed for crude oil in a limited
of the sample. interlaboratory test program the 2005 ILS6 between

5
MNL 7, Manual on Presentation of Data Control Chart Analysis, “Section 3:
Control Charts for Individuals,” 6th edition, ASTM International, W. Conshohocken, 6
Supporting data have been filed at ASTM International Headquarters and may
PA. be obtained by requesting Research Report RR:D02-1643.

Copyright ASTM International


Provided by IHS under license with ASTM
5Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
VPCR4 (37.8 °C) obtained using this test method and the vapor 15. Keywords
pressure obtained using Test Method D323 (see Appendix X1). 15.1 crude oil; expansion method; Reid vapor pressure;
14.4 The precision statements were derived from a 2005 vapor pressure
interlaboratory cooperative test program involving 7 laborato-
ries and 6 crude oil samples covering a vapor pressure range
between 34 kPa and 117 kPa.6 The sample introduction tem-
perature for this test program was allowed to be set between
20 °C and 37.8 °C.

APPENDIXES

(Nonmandatory Information)

X1. RELATIVE BIAS TO TEST METHOD D323

X1.1 The degree of expected agreement between in Eq X1.1 is not universally valid for all crude oils and it is
VPCR4 (37.8 °C) results by this test method for samples in recommended to report the VPCR4 (37.8 °C) result for all
either pressurized floating piston cylinders and nonpressurized crude oils. If Eq X1.1 is used to estimate RVPE, confirm its

--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---
1 L containers, and RVP results by Test Method D323 using validity for the crude oil samples under test by analyzing
nonpressurized 1 L containers has been assessed in accordance samples by both Test Methods D323 and D6377. In any case,
with Practice D6708. Sample specific biases were observed. Eq X1.1 only applies to dead crude oil samples (see 3.2.1 and
The reproducibility of Test Method D323 for the 2005 ILS 3.2.1.1) with VPCR4 (37.8 °C) results between 34 kPa and
crude oil sample set was R=9 kPa.6 Due to the low number of 117 kPa (see 14.4), since live crude oil samples (see 3.2.2 and
the crude oil samples in the ILS, a direct comparison between 3.2.2.1) if tested by Test Method D323 in non-pressurized
the results of this test method and Test Method D323 is containers (for example, sample bottles or cans) will result in
suggestive. significant loss of vapors during the air saturation procedure in
Test Method D323 prior to analysis.
X1.2 As stated in 14.3, a statistically significant relative bias
was observed for crude oil in a limited 2005 interlaboratory test RVPE 5 A 3 VPCR4 ~ 37.8°C ! (X1.1)
program involving VPCR4 (37.8 °C) results obtained by this
where:
test method and results obtained using Test Method D323. For
the six crude oil samples evaluated in the limited interlabora- A = 0.834 (for samples in pressurized floating piston
tory test program, RVP results determined by Test Method cylinders), and
A = 0.915 (for samples in nonpressurized 1-L sample
D323, based on VPCR4 (37.8 °C) results obtained by this test
containers)
method, can be predicted by Eq X1.1. However, the correlation

X2. VAPOR PRESSURE OF CRUDE OIL VERSUS TEMPERATURE

X2.1 For several purposes, for example, transport and X2.1.2 Increase the measuring temperature in steps to the
storage, it may be useful to measure the vapor pressure highest temperature of interest but not higher than 100 °C.
behavior of crude oil versus temperature. To obtain such Each step is measured according with 12.7 and 12.8.
information proceed as follows. X2.1.3 The result is either recorded as a series of values, or
X2.1.1 Follow the instructions outlined in 12.1 – 12.8. The plotted as log (VPCRX) versus 1/T, where T is the absolute
initial measuring temperature is programmed to the lowest measuring temperature in °K (°C + 273.2).
temperature of interest but not lower than the temperature NOTE X2.1—No precision data have been developed for temperatures
stated in 12.7. different from 37.8 °C.

Copyright ASTM International


Provided by IHS under license with ASTM
6Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST
D6377 − 16
SUMMARY OF CHANGES

Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue
(D6377 – 15) that may impact the use of this standard. (Approved April 1, 2016.)

(1) Revised 1.2, 8.3.1, 8.3.3, 11.1, 3.2.3.1, and 14.4. (2) Updated 12.1.

Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue
(D6377 – 14) that may impact the use of this standard. (Approved Oct. 1, 2015.)

(1) Updated 5.4 and 5.5.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
--``,``,`,``,,,``,,`,,,``,``,``,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
7Licensee=YPF/5915794100, User=Sarmiento, David Daniel
No reproduction or networking permitted without license from IHS Not for Resale, 02/20/2017 08:59:27 MST

You might also like