Series M22/M23/M24 Instruction Manual Table of Contents

VorTek Series M22, M23, M24, and M24R


Pro-V™ Vortex Volumetric and Mass Flow Meters

Instruction Manual
Document Number M-000-00010
Rev 04/2019

8475 W I-25 Frontage Rd


Suite 300
Longmont, CO 80504
(303) 682-9999 (888) 386-7835 Fax (303) 682-4368
http://www.vortekinst.com

M-000-00010 0-1
Table of Contents Series M22/M23/M24 Instruction Manual

Customer Notice for Oxygen Service


Unless you have specifically ordered VorTek’s optional O 2 cleaning, this flow meter
may not be fit for oxygen service. Some models can only be properly cleaned during
the manufacturing process. VorTek Instruments, LLC., is not liable for any damage
or personal injury, whatsoever, resulting from the use of VorTek Instruments stand-
ard mass flow meters for oxygen gas.

Specific Conditions of Use


(ATEX/IECEx)
Contact Manufacturer regarding Flame path information.

Clean with a damp cloth to avoid any build-up of electrostatic charge.

The Model M22, M23, M24 and M24R Pro-VTM Multivariable Mass Vortex Flow-
meters standard temperature option (ST) process temperature range is -40°C to
260°C. The high temperature option (HT) process temperature range is -40°C up
to +400°C.

Pro-VTM Multivariable Mass Vortex Flowmeters


Tmax Temperature Class Value (Gas)
(Process) ST Version HT Version
80°C T6 85°C
95°C T5 100°C
130°C T4 135°C
195°C T3 200°C
260°C T2 300°C
400°C N/A 405°C

Notice to Users
Warning based on Electric Appliances Safety Control Act of Korea. This product
should be handled as electromagnetic radiation emitting equipment, and is in-
tended for use by industrial dealers and end-users. It is not for residential use.

NIST Standard Reference Database 23, NIST Reference Fluid Thermodynamic and
Transport Properties: REFPROP Version 9.x. Standard Reference Databases are copy-
righted by the U.S. Secretary of Commerce on behalf of the United States of America. All
rights reserved. No part of the database may be reproduced, stored in a retrieval system
or transmitted, in any form or by any means, electronic, mechanical, photocopying, record-
ing or otherwise without prior permission.

0-2 M-000-00010
Series M22/M23/M24 Instruction Manual Table of Contents

© COPYRIGHT VORTEK INSTRUMENTS 2019


No part of this publication may be copied or distributed, transmitted, transcribed, stored in
a retrieval system, or translated into any human or computer language, in any form or by
any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties with-
out the express written permission of VorTek Instruments. The information contained in this
manual is subject to change without notice.

TRADEMARKS
Pro-V™ is a trademark of VorTek Instruments, LLC. Other product and company names
listed in this manual are trademarks or trade names of their respective manufacturers.

M-000-00010 0-3
Table of Contents Series M22/M23/M24 Instruction Manual

Table of Contents
Chapter 1 Introduction
Pro-V™ Vortex Mass Flow Meters ..................................................... 1-1
Using this Manual ......................................................................... 1-1
Note and Safety Information ......................................................... 1-2
Receipt of System Components .................................................... 1-2
Technical Assistance ..................................................................... 1-2
How the Pro-V Vortex Meter Operates ............................................... 1-3
Velocity Measurement .................................................................. 1-3
Vortex Shedding Frequency .......................................................... 1-4
Vortex Frequency Sensing ............................................................ 1-4
Flow Velocity Range ..................................................................... 1-5
Pressure Drop ................................................................................ 1-6
Minimum Back Pressure ............................................................... 1-6
Temperature Measurement ............................................................ 1-7
Pressure Measurement................................................................... 1-7
Flow Meter Configurations .................................................................. 1-8
Multivariable Options ................................................................... 1-8
Line Size / Process Conditions / Materials .................................... 1-9
Flow Meter Electronics ................................................................. 1-9

Chapter 2 Installation
Installation Overview ........................................................................... 2-1
Flow Meter Installation Requirements .......................................... 2-1
Unobstructed Flow Requirements ................................................. 2-2
Recommended Meter Locations .................................................... 2-3
Series M22 and M24 In-Line Flow Meter Installation ........................ 2-4
Wafer-Style Flow Meter Installation ............................................. 2-5
Flange-Style Flow Meter Installation ............................................ 2-6
Series M23 Insertion Flow Meter Installation ..................................... 2-7
Cold Tap Guidelines...................................................................... 2-8
Hot Tap Guidelines ....................................................................... 2-9
Flow Meter Insertion ......................................................................... 2-11
Installing Meters with a Compression Connection ..................... 2-12
Installing Meters with a Packing Gland Connection ................... 2-14
Installing Meters (Packing Gland), No Insertion Tool ................ 2-17
Adjusting Meter Orientation .............................................................. 2-19
Display/Keypad Adjustment ....................................................... 2-19
Enclosure Adjustment ................................................................. 2-20
Loop Power Flow Meter Wiring Connections ................................... 2-21
Input Power Connections ............................................................ 2-21
4-20 mA Output Connections ..................................................... 2-22
Pulse Output Connections ........................................................... 2-23
Frequency Output Connections ................................................... 2-24
Optional Backlight Connections ................................................. 2-24
Remote Electronics Wiring ......................................................... 2-25
High Power Flow Meter Wiring Connections ................................... 2-27
Input Power Connections ............................................................ 2-27

0-4 M-000-00010
Series M22/M23/M24 Instruction Manual Table of Contents

4-20 mA Output Connections ..................................................... 2-29


Frequency Output Connections ................................................... 2-30
Pulse Output Connections ........................................................... 2-31
Alarm Output Connections.......................................................... 2-33
Remote Electronics Wiring ......................................................... 2-34
Optional Input Electronics Wiring .............................................. 2-35
Optional Energy EMS RTD Input Wiring .................................. 2-35
Optional External 4-20 mA Input Wiring ................................... 2-36
Optional Contact Closure Input Wiring ...................................... 2-37

Chapter 3 Operating Instructions


Flow Meter Display/Keypad ................................................................ 3-1
Start Up ................................................................................................ 3-2
Run Mode Screens ............................................................................... 3-3
Using the Setup Menus ........................................................................ 3-4
Programming the Flow Meter ....................................................... 3-5
Output Menu ................................................................................. 3-6
Display Menu ................................................................................ 3-8
Alarms Menu ................................................................................. 3-9
Totalizer #1 Menu ....................................................................... 3-10
Totalizer #2 Menu ....................................................................... 3-11
Energy Menu ....................................................................... 3-12
Fluid Menu .................................................................................. 3-13
Units Menu .................................................................................. 3-14
Time and Date Menu ................................................................... 3-15
Diagnostics Menu........................................................................ 3-16
Calibration Menu......................................................................... 3-17
Password Menu ........................................................................... 3-18

Chapter 4 Serial Communications


HART Communications ...................................................................... 4-1
Wiring ........................................................................................... 4-1
HART Commands with the DD Menu .......................................... 4-3
HART Commands with Generic DD Menu .................................. 4-8
MODBUS Communications ................................................................ 4-9
Wiring ........................................................................................... 4-9
Menu Items.................................................................................. 4-10
Register Definitions..................................................................... 4-12
BACnet MSTP Communications....................................................... 4-19
BACnet MSTP Description......................................................... 4-19
Baud Rates on the MS/TP Bus .................................................... 4-19
Supported BACnet Objects ......................................................... 4-20
ANNEX - BACnet Protocol Implementation Conformance
Statement ..................................................................................... 4-26
Acronyms and Definitions .......................................................... 4-32

Chapter 5 Troubleshooting and Repair


Hidden Diagnostics Menus .................................................................. 5-1
Level One Hidden Diagnostics Values ......................................... 5-3

M-000-00010 0-5
Table of Contents Series M22/M23/M24 Instruction Manual

Level Two Hidden Diagnostics Values ......................................... 5-4


Analog Output Calibration .................................................................. 5-7
Display Contrast Adjustment .............................................................. 5-7
Troubleshooting the Flow Meter ......................................................... 5-8
First Check Items ................................................................................. 5-8
Record Values ...................................................................................... 5-8
Determine the Fault ............................................................................. 5-9
Symptom: Output at no Flow ........................................................ 5-9
Symptom: Erratic Output .............................................................. 5-9
Symptom: No Output .................................................................. 5-11
Symptom: Meter Displays Temperature Fault ............................ 5-12
Symptom: Meter Displays Pressure Fault ................................... 5-13
Electronics Assembly Replacement ................................................... 5-14
Pressure Sensor Replacement (Series M22 and M24 Only) ............. 5-15
Returning Equipment to the Factory .................................................. 5-15

Appendix A Product Specifications


Appendix B Approvals
Appendix C Flow Meter Calculations
Appendix D Glossary

Figures
1-1. In-Line Vortex Multi-Parameter Mass Flow Meter ................ 1-3
1-2. Measurement Principle of Vortex Flow Meters...................... 1-4
1-3. Reynolds Number Range of the Pro-V ................................... 1-6

2-1. Recommended Pipe Length Required for Installation ............ 2-2


2-2. Flange Bolt Torquing Sequence ............................................. 2-3
2-3. Wafer-Style Flow Meter Installation ...................................... 2-4
2-4. Flange-Style Flow Meter Installation ..................................... 2-5
2-5. Hot Tap Sequence ................................................................... 2-9
2-6. Insertion Calculation (Compression Type) ........................... 2-11
2-7. Flow Meter with Compression Type Fitting......................... 2-12
2-8. Insertion Calculation (Meters with Insertion Tool) .............. 2-13
2-9. Flow Meter with Permanent Insertion Tool .......................... 2-14
2-10. Flow Meter with Removable Insertion Tool......................... 2-15
2-11. Insertion Calculation (Meters without Insertion Tool) ......... 2-16
2-12. Display/Keypad Viewing Adjustment .................................. 2-18
2-13. Enclosure Viewing Adjustment ............................................ 2-19
2-14. Loop Power Wiring Terminals ............................................. 2-20
2-15. DC Power Connections......................................................... 2-20
2-16. Load Resistance Versus Input Voltage ................................. 2-21
2-17. Isolated Pulse Output Using External Power Supply ........... 2-22
2-18. Non-Isolated Pulse Output Using External Power Supply ... 2-22

0-6 M-000-00010
Series M22/M23/M24 Instruction Manual Table of Contents

2-19. Isolated Frequency Output Using External Power Supply ... 2-23
2-20. Non-Isolated Freq. Out. Using External Power Supply........ 2-23
2-21. Backlight Using External Power Supply .............................. 2-23
2-22. Loop Power Volumetric Flowmeter Junction Box ............... 2-24
2-23. Loop Power Volumetric Flowmeter Junction Box Prior ...... 2-25
2-24. Loop Power Mass Flowmeter Junction Box ......................... 2-25
2-25. AC Wiring Terminals ........................................................... 2-26
2-26. AC Power Connections......................................................... 2-26
2-27. DC Wiring Terminals ........................................................... 2-27
2-28. DC Power Connections......................................................... 2-27
2-29. Load Resistance Versus Input Voltage ................................. 2-28
2-30. Isolated 4-20 Output Using External Power Supply ............. 2-28
2-31. Non-Isolated 4-20 Output Using Input Power Supply.......... 2-29
2-32. Isolated 4-20 Output Using Meter Power Supply (AC only) 2-29
2-33. Isolated Frequency Output Using External Power Supply ... 2-30
2-34. Non-Isolated Frequency Output Using Input Power Supply 2-30
2-35. Isolated Freq. Out. Using Meter Power Sup. (AC only) ....... 2-30
2-36. Isolated Pulse Output Using External Power Supply ........... 2-31
2-37. Non-Isolated Pulse Output Using Input Power Supply ........ 2-31
2-38. Isolated Pulse Output Using Meter Power Sup. (AC only) .. 2-31
2-39. Isolated Alarm Output Using External Power Supply .......... 2-32
2-40 Non-Isolated Alarm Output Using Meter Power Supply ...... 2-32
2-41 Isolated Alarm Output Using Meter Power Sup. (AC only) 2-33
2-42 High Power Flow Meter Junction Box ................................ 2-33
2-43. Optional Energy EMS RTD Input Wiring ............................ 2-34
2-44. External 4-20 mA Input Wiring – External Power Supply ... 2-35
2-45. External 4-20 mA Input Wiring – DC Powered Meter ......... 2-35
2-46. External 4-20 mA Input Wiring – AC Powered Meter ......... 2-36
2-47. Optional External Contact Closure Input Wiring ................. 2-36

3-1. Flow Meter Display/Keypad................................................... 3-1

4-1. Loop Powered Meter Wiring (HART).................................... 4-1


4-2. DC Powered Meter Wiring (HART) ...................................... 4-2
4-3. AC Powered Meter Wiring (HART) ...................................... 4-2
4-4. RS-485 Wiring (MODBUS) ................................................... 4-9

5-1. Electronics Stack Sensor Connections .................................. 5-10


5-2. Remote Feed Through Board Sensor Connections ............... 5-11
5-3. Vortex Sensor Connector ...................................................... 5-11
5-4. Temperature Sensor Connector ............................................ 5-12
5-5. Pressure Sensor Connector ................................................... 5-13

Tables
4-1. Byte Order (MODBUS) ........................................................ 4-11
4-2. Register Definitions (MODBUS) ......................................... 4-13

M-000-00010 0-7
Table of Contents Series M22/M23/M24 Instruction Manual

Warnings and Cautions

Warning!
Consult the flow meter nameplate for specific flow meter approvals before any hazardous
location installation.

Hot tapping must be performed by a trained professional. U.S. regulations often require a
hot tap permit. The manufacturer of the hot tap equipment and/or the contractor perform-
ing the hot tap is responsible for providing proof of such a permit.

All flow meter connections, isolation valves and fittings for cold/hot tapping must have the
same or higher pressure rating as the main pipeline.

For Series M23 insertion flow meter installations, an insertion tool must be used for any
installation where a flow meter is inserted under pressure greater than 50 psig.

To avoid serious injury, DO NOT loosen a compression fitting under pressure.

To avoid potential electric shock, follow National Electric Code or your local code when
wiring this unit to a power source. Failure to do so could result in injury or death. All AC
power connections must be in accordance with published CE directives. All wiring proce-
dures must be performed with the power Off.

Before attempting any flow meter repair, verify that the line is not pressurized. Always re-
move main power before disassembling any part of the mass flow meter.

Caution!
Calibration must be performed by qualified personnel. VorTek Instruments, Inc., strongly rec-
ommends that you return your flow meter to the factory for calibration.

In order to achieve accurate and repeatable performance, the flow meter must be in-
stalled with the specified minimum length of straight pipe upstream and downstream of
the flow meter’s sensor head.

When using toxic or corrosive gases, purge the line with inert gas for a minimum of four
hours at full gas flow before installing the flow meter.

For Series M23 insertion flow meter installations, the sensor alignment pointer must point
downstream in the direction of flow.

The AC wire insulation temperature rating must meet or exceed 85°C (185°F)

0-8 M-000-00010
Series M22/M23/M24 Instruction Manual Table of Contents

Revision History
Rev. 2/2019 to 4/2019
Page Change Description
4-3 Updated HART Online Menu
4-4 Updated HART Analog Output Menu
4-7 Updated HART Sensor Cal Menu
B-1 Updated Appendix B

M-000-00010 0-9
Table of Contents Series M22/M23/M24 Instruction Manual

0-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 1 Introduction

Chapter 1 Introduction
Pro-V™ Multi-Parameter Vortex Mass Flow Meters
The VorTek Instruments’ Series M22/M24 In-Line and the Series M23
Insertion Pro-V™ Vortex Flow Meters provide a reliable solution for
process flow measurement. From a single entry point in the pipeline,
Pro-V meters offer precise measurements of mass or volumetric flow
rates.

Multi-Parameter Mass Flow Meters


Mass flow meters utilize three primary sensing elements: a vortex shed-
ding velocity sensor, an RTD temperature sensor, and a solid state pres-
sure sensor to measure the mass flow rate of gases, liquids, and steam.
Meters are available as loop powered devices or with up to three 4-20
mA analog output signals for monitoring your choice of the five process
variables (mass flow, volumetric flow, temperature, pressure and fluid
density). The Energy Monitoring option permits real-time calculation of
energy consumption for a facility or process.

Volumetric Flow Meters


The primary sensing element of a volumetric flow meter is a vortex
shedding velocity sensor. Meters are loop powered. The analog 4-20
mA output signal offers your choice of volumetric or mass flow rate.
Mass flow rate is based on a constant value for fluid density stored in the
instrument’s memory.

Both the mass and volumetric flow meters can be ordered with a local
keypad/display which provides instantaneous flow rate, total, and process
parameters in engineering units. A pulse output signal for remote totali-
zation and MODBUS, HART, or BACnet communications are also
available. Pro-V digital electronics allows for easy reconfiguration for
most gases, liquids and steam. The VorTek Series M22, M23, and M24
Pro-V Meters’ simple installation combines with an easy-to-use interface
that provides quick set up, long term reliability and accurate mass flow
measurement over a wide range of flows, pressures and temperatures.

Using This Manual


This manual provides information needed to install and operate both the
Series M22/M24 In-Line and Series M23 Insertion Pro-V Flow Meters.
• Chapter 1 includes the introduction and product description
• Chapter 2 provides information needed for installation
• Chapter 3 describes system operation and programming
• Chapter 4 information on HART, MODBUS, and BACnet protocols
• Chapter 5 covers troubleshooting and repair
Appendix A - Product Specifications, Appendix B – Approvals,
Appendix C – Flow Meter Calculations, Appendix D – Glossary of
Terms

M-000-00010 1-1
Chapter 1 Introduction Series M22/M23 Instruction Manual

Note and Safety Information


We use note, caution and warning statements throughout this book to
draw your attention to important information.

Warning! Caution! Note


This statement appears with This statement appears with This statement appears with
information that is important to information that is important a short message to alert you
protect people and equipment for protecting your equipment to an important detail.
from damage. Pay very close and performance. Read and
attention to all warnings that follow all cautions that apply
apply to your application. to your application.

Receipt of System Components


When receiving a VorTek mass flow meter, carefully check the outside
packing carton for damage incurred in shipment. If the carton is dam-
aged, notify the local carrier and submit a report to the factory or distrib-
utor. Remove the packing slip and check that all ordered components are
present. Make sure any spare parts or accessories are not discarded with
the packing material. Do not return any equipment to the factory without
first contacting VorTek Customer Service.

Technical Assistance
If you encounter a problem with your flow meter, review the configura-
tion information for each step of the installation, operation and set up pro-
cedures. Verify that your settings and adjustments are consistent with fac-
tory recommendations. Refer to Chapter 5, Troubleshooting, for specific
information and recommendations.

If the problem persists after following the troubleshooting procedures


outlined in Chapter 5, contact VorTek Instruments, Technical Support
Via Email at [email protected] or by phone at (888) 386-7835 or
(303) 682-9999 between 8:00 a.m. and 5:00 p.m. MST. When calling
Technical Support, have the following information on hand:
• the serial number and VorTek order number (all marked on
the meter nameplate)
• the problem you are encountering and any corrective action
taken
• application information (fluid, pressure, temperature and
piping configuration)

1-2 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 1 Introduction

How the Pro-V Vortex Mass Flow Meter Operates

Figure 1-1. In-Line Vortex Multi-Parameter Mass Flow Meter

VorTek Series M22, M23, and M24 Pro-V™ Multi-Parameter Vortex


Mass Flow Meters use a unique sensor head to monitor mass flow rate by
directly measuring three variables–fluid velocity, temperature and pres-
sure. The built-in flow computer calculates the mass flow rate and volu-
metric flow rate based on these three direct measurements. The velocity,
temperature and pressure sensing head is built into the vortex meter’s
flow body. To measure fluid velocity, the flow meter incorporates a bluff
body (shedder bar) in the flow stream and measures the frequency of vor-
tices created by the shedder bar. Temperature is measured using a plati-
num resistance temperature detector (PRTD). Pressure measurement is
achieved using a solid-state pressure transducer. All three elements are
combined into an integrated sensor head assembly located downstream
of the shedder bar within the flow body.

Velocity Measurement
The Pro-V vortex velocity sensor is a patented mechanical design that
minimizes the effects of pipeline vibration and pump noise, both of which
are common error sources in flow measurement with vortex flow meters.
The velocity measurement is based on the well-known Von Karman vor-
tex shedding phenomenon. Vortices are shed from a shedder bar, and the
vortex velocity sensor located downstream of the shedder bar senses the
passage of these vortices. This method of velocity measurement has many
advantages including inherent linearity, high turndown, reliability and
simplicity.

M-000-00010 1-3
Chapter 1 Introduction Series M22/M23 Instruction Manual

Vortex Shedding Frequency


Von Karman vortices form downstream of a shedder bar into two distinct
wakes. The vortices of one wake rotate clockwise while those of the oth-
er wake rotate counterclockwise. Vortices generate one at a time, alter-
nating from the left side to the right side of the shedder bar. Vortices in-
teract with their surrounding space by over-powering every other nearby
swirl on the verge of development. Close to the shedder bar, the distance
(or wave length) between vortices is always constant and measurable.
Therefore, the volume encompassed by each vortex remains constant, as
shown below. By sensing the number of vortices passing by the velocity
sensor, the Pro-V™ Flow Meter computes the total fluid volume.

Figure 1-2. Measurement Principle of Vortex Flow Meters

Vortex Frequency Sensing


The velocity sensor incorporates a piezoelectric element that senses the
vortex frequency. This element detects the alternating lift forces pro-
duced by the Von Karman vortices flowing downstream of the vortex
shedder bar. The alternating electric charge generated by the piezoelec-
tric element is processed by the transmitter’s electronic circuit to obtain
the vortex shedding frequency. The piezoelectric element is highly sensi-
tive and operates over a wide range of flows, pressures and temperatures.

1-4 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 1 Introduction

Flow Velocity Range


To ensure trouble-free operation, vortex flow meters must be correctly
sized so that the flow velocity range through the meter lies within the
measurable velocity range (with acceptable pressure drop) and the linear
range.

The measurable range is defined by the minimum and maximum velocity


using the following table.

Gas Liquid

Vmin 25 ft/s 1 ft/s English  (lb/ft3)



Vmax
0.5 in 175 ft/s 30 ft/s
0.75 in 250 ft/s 30 ft/s
1 in 250 ft/s 30 ft/s
1.5 in and 300 ft/s 30 ft/s
greater

Vmin 3 7 m/s 0.3 m/s Metric  (kg/m3)



Vmax
DN15 53 m/s 9.1 m/s
DN20 76 m/s 9.1 m/s
DN25 76 m/s 9.1 m/s
DN40 and 91 m/s 9.1 m/s
greater

The linear range is defined by the Reynolds number. The Reynolds number is the
ratio of the inertial forces to the viscous forces in a flowing fluid and is defined
as:
 V D
Re =

Where
Re = Reynolds Number
 = mass density of the fluid being measured
V = velocity of the fluid being measured
D = internal diameter of the flow channel
 = viscosity of the fluid being measured

M-000-00010 1-5
Chapter 1 Introduction Series M22/M23 Instruction Manual

The Strouhal number is the other dimensionless number that quantifies the vortex
phenomenon. The Strouhal number is defined as:

St = f d
V

Where
St = Strouhal Number
f = frequency of vortex shedding
d = shedder bar width
V = fluid velocity

As shown in Figure 1-3, Pro-V™ meters exhibit a constant Strouhal number


across a large range of Reynolds numbers, indicating a consistent linear output
over a wide range of flows and fluid types. Below this linear range, the intelligent
electronics in Pro-V automatically corrects for the variation in the Strouhal num-
ber with the Reynolds number. The meter’s smart electronics corrects for this
non-linearity via its simultaneous measurements of the process fluid temperature
and pressure. This data is then used to calculate the Reynolds number in real
time. Pro-V meters automatically correct down to a Reynolds number of 5,000.
Strouhal Number, St

Corrected range
0.3
Linear range
0.2

0.1

0.0
10
3 4
10 10 10 10 10
5 6 7 8

5000 Rey nolds Number, Re

Figure 1-3. Reynolds Number Range for the Pro-V

Pressure Drop
The pressure drop for series M23 insertion meters is negligible. The pressure drop
for series M22/M24 in-line meters is defined as:
P = .00024  V2 English units (P in psi,  in lb/ft3, V in ft/sec)
P = .000011  V2 Metric units (P in bar,  in kg/m3, V in m/sec)

Minimum Back Pressure


A minimum back pressure is required to prevent cavitation in the vortex
flow meter when installed in liquid applications. Cavitation is a phenom-
enon wherein a liquid passing through a low pressure zone no longer can
remain in the liquid phase and therefore partially flashes (or “boils”) into

1-6 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 1 Introduction

its vapor phase. The resulting two-phase flow degrades the liquid calibra-
tion accuracy. For some applications, a valve downstream of the flow me-
ter may be required to increase the pressure in the meter, thereby avoiding
cavitation. The following equation defines the minimum back pressure to
prevent cavitation.

P = 2.9 P + 1.3 PV

Where:
P = Minimum line pressure five pipe diameters downstream of the flow
meter required to avoid cavitation (psia or bara).

P = Permanent pressure loss across the flow meter (psia or bara).

PV = Liquid vapor pressure at actual flowing conditions (psia or bara).

Temperature Measurement
Pro-V Flow Meters use a 1000 ohm platinum resistance temperature detec-
tor (PRTD) to measure fluid temperature.

Pressure Measurement
Pro-V Flow Meters incorporate a solid-state pressure transducer isolated
by a 316 stainless steel diaphragm. The transducer itself is micro-
machined silicon, fabricated using integrated circuit processing technol-
ogy. A nine-point pressure/temperature calibration is performed on every
sensor. Digital compensation allows these transducers to operate within a
0.3% of full scale accuracy band within the entire ambient temperature
range of -40°F to 140°F (-40 to 60°C). Thermal isolation of the pressure
transducer ensures the same accuracy across the allowable process fluid
temperature range of -330°F to 750°F (-200 to 400°C).

M-000-00010 1-7
Chapter 1 Introduction Series M22/M23 Instruction Manual

Flow Meter Configurations


Pro-V™ Vortex Mass Flow Meters are available in two model
configurations:
• Series M22/M24 in-line flow meter (replaces a section of the pipeline)
• Series M23 insertion flow meter (requires a “cold” tap or a “hot”
tap into an existing pipeline)

Both the in-line and insertion configurations are similar in that they both
use identical electronics and have similar sensor heads. Besides installa-
tion differences, the main difference between an in-line flow meter and
an insertion flow meter is their method of measurement.

For an in-line vortex flow meter, the shedder bar is located across the en-
tire diameter of the flow body. Thus, the entire pipeline flow is included
in the vortex formation and measurement. The sensing head, which di-
rectly measures velocity, temperature and pressure is located just down-
stream of the shedder bar.

Insertion vortex flow meters have a shedder bar located across the di-
ameter of a short tube. The velocity, temperature and pressure sensor are
located within this tube just downstream of a built-in shedder bar. This
entire assembly is called the insertion sensing head. It fits through any
entry port with a 1.875 inch minimum internal diameter.

The sensing head of an insertion vortex flow meter directly monitors the
velocity at a point in the cross-sectional area of a pipe, duct, or stack (re-
ferred to as “channels”). The velocity at a point in the pipe varies as a func-
tion of the Reynolds number. The insertion vortex flow meter computes the
Reynolds number and then computes the total flow rate in the channel. The
output signal of insertion meters is the total flow rate in the channel. The
accuracy of the total flow rate computation depends on adherence to the
piping installation requirements given in Chapter 2. If adherence to those
guidelines cannot be met, contact the factory for specific installation ad-
vice.

Multivariable Options
The M22, M23, or M24 models are available with the following options:
V, volumetric flowmeter; VT, velocity and temperature sensors; VTP,
velocity, temperature, and pressure sensors; VT-EM energy output op-
tions; VTP-EM, energy options with pressure; VT-EP, external pressure
transmitter input; VETEP, external RTD temperature input, external 4-20
mA input.

1-8 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 1 Introduction

Line Size / Process Connections / Materials


The M22 and M24 Non-reducing In-line models are built for line sizes ½
through 4-inch wafer or ½ through 12-inch flanged design using ANSI
150, 300, 600, or DIN PN16, 40, or 63 class flanges. JIS flanges are
available upon request. These can be built with A105 carbon steel (1 ½
through 12 inch), 316/316L stainless steel, or Hastelloy C-276.

The M24 Non-reducing and M24R Reducing In-line models have face to
face lengths that are different from M22. The M24R In-line flanged or
wafer model reduces by one pipe size to increase the velocity through the
meter and is available for ANSI 150, 300, or 600 class flanges and
316/316L stainless steel.

The M23 Insertion model can be used in line sizes 2 inch and greater and
is built with a compression fitting or packing gland design using 2-inch
NPT, or 2-inch flanged connections (ANSI 150, 300, 600, DIN PN16,
40, or 63 class flanges). The packing gland design can be ordered with a
permanent or removable retractor. The M23 Insertion model can be built
with 316/316L stainless steel or Hastelloy C-276.

Flow Meter Electronics


Pro-V Flow Meter electronics are available mounted directly to the flow
body, or remotely mounted. The electronics housing may be used indoors
or outdoors, including wet environments. Available input power options
are: DC loop powered (2-wire), DC powered, or AC powered. Three an-
alog output signals are available for your choice of three of the five pro-
cess variables: mass flow rate, volumetric flow rate, temperature, pres-
sure or fluid density. A pulse output signal for remote totalization and
MODBUS, HART, and BACnet communications are also available.

Pro-V Flow Meters include a local 2 x 16 character LCD display housed


within the enclosure. Local operation and reconfiguration is accom-
plished using six pushbuttons operated via finger touch. For hazardous
locations, the six buttons can be operated with the electronics enclosure
sealed using a hand-held magnet, thereby not compromising the integrity
of the hazardous location certification.

The electronics include nonvolatile memory that stores all configuration


information. The nonvolatile memory allows the flow meter to function
immediately upon power up, or after an interruption in power. All
flowmeters are calibrated and configured for the customer’s flow
application.

M-000-00010 1-9
Chapter 1 Introduction Series M22/M23/M24 Instruction Manual

1-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Chapter 2 Installation
Installation Overview
VorTek’s Pro-V Vortex Flow Meter installations are simple and straight-
forward. Both the Series M22/M24 In-Line and Series M23 Insertion type
flow meter installations are covered in this chapter. After reviewing the
installation requirements given below, see page 2-3 for Series M22/M24
installation instructions. See page 2-6 for Series M23 installation instruc-
tions. Wiring instructions begin on page 2-20.

Flow Meter Installation Requirements


Before installing the flow meter, verify the installation site allows for
these considerations:
Warning!
Consult the flow meter nameplate for 1. Line pressure and temperature will not exceed the flow meter
specific flow meter approvals before any rating.
hazardous location installation.

2. The location meets the required minimum number of pipe di-


ameters upstream and downstream of the sensor head as illus-
trated in Figure 2-1.

3. Safe and convenient access with adequate overhead clear-


ance for maintenance purposes.

4. Verify that the cable entry into the instrument meets the
specific standard required for hazardous area installations.
The cable entry device shall be of a certified flameproof
type, suitable for the conditions of use and correctly in-
stalled. The degree of protection of at least IP66 to EN
60529 is only achieved if certified cable entries are used
that are suitable for the application and correctly installed.
Unused apertures shall be closed with suitable blanking el-
ements.

5. For remote installations, verify the supplied cable length is


sufficient to connect the flow meter sensor to the remote
electronics.

Also, before installation check your flow system for anomalies such as:

• leaks
• valves or restrictions in the flow path that could create disturb-
ances in the flow profile that might cause unexpected flow rate in-
dications

M-000-00010 2-1
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Unobstructed Flow Requirements


Select an installation site that will minimize possible distortion in the flow
profile. Valves, elbows, control valves and other piping components may
cause flow disturbances. Check your specific piping condition against the
examples shown below. In order to achieve accurate and repeatable per-
formance install the flow meter using the recommended number of
straight run pipe diameters upstream and downstream of the sensor.

Note: For liquid applications in vertical pipes, avoid installing with flow
in the downward direction because the pipe may not be full at all points.
Choose to install the meter with flow in the upward direction if possible.

A B A B

Flow meter
C' C Flow meter

Flow conditioner
Example 1. Example 4. (if used)
One 90° elbow before meter Reduction before meter

A B
A B
C' C Flow meter
C' C Flow meter
Flow conditioner
(if used)
Example 2. Example 5. Flow conditioner
Two 90° elbows before meter in one plane Expansion before meter (if used)

A B
A B
C' C Flow meter
C' C Flow meter

Flow conditioner
(if used)
Flow conditioner (if used)
Example 3. Example 6.
Two 90° elbows before meter out of plane (if three Regulator or valve partially closed before meter
90° bends present, double recommended length) (If valve is always wide open, base length
requirements on fitting directly preceding it)

Minimum Required Minimum Required


Upstream Diameters Downstream Diameters
No Flow No Flow With Flow
Conditioner With Flow Conditioner Conditioner Conditioner
Example A A C C´ B B
1 10 D N/A N/A N/A 5D 5D
2 15 D 10 D 5D 5D 5D 5D
3 30 D 10 D 5D 5D 10 D 5D
4 10 D 10 D 5D 5D 5D 5D
5 20 D 10 D 5D 5D 5D 5D
6 30 D 10 D 5D 5D 10 D 5D
D = Internal diameter of channel. N/A = Not applicable
Figure 2-1. Recommended Pipe Length Requirements for Installation, Series M22/M23/M24

2-2 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation
Recommended Meter Locations
Liquid Horizontal

Liquid Vertical

Gas or Steam Horizontal

Gas or Steam Vertical

M-000-00010 2-3
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Series M22/M24 In-Line Flow Meter Installation


Install the Series M22/M24 In-Line Flow Meter between two convention-
al pipe flanges as shown in Figures 2-3 and 2-4.

The meter inside diameter is equal to the same size nominal pipe ID in
schedule 80. For example, a 2” meter has an ID of 1.939” (2” schedule
80). Do not install the meter in a pipe with an inside diameter smaller
than the inside diameter of the meter. For schedule 160 and higher
pipe, a special meter is required. Consult the factory before purchasing
the meter.

Series M22/M24 Meters require customer-supplied gaskets. When select-


ing gasket material make sure that it is compatible with the process fluid
and pressure ratings of the specific installation. Verify that the inside di-
ameter of the gasket is larger than the inside diameter of the flow meter
and adjacent piping. If the gasket material extends into the flow stream, it
will disturb the flow and cause inaccurate measurements.

Flange Bolt Specifications


Stud bolt lengths may be calculated using the following equation:

L = Meter face to face length + 2 (mounting flange thickness + flange


raised face) + 2 (gasket thickness) + 4 (mounting nut thickness)

Refer to the mounting flange specification to select the correct stud


bolt diameter.

The required bolt load for sealing the gasket joint is affected by several
application-dependent factors; therefore the required torque for each ap-
plication may be different. Refer to the ASME Pressure Vessel Code
guidelines for bolt tightening standards.

1
1 1 12 5
8 5 8 9

4 3 4 3 4 3

10 7
6 7
2 2 6 11
2

4-bolt 8-bolt 12-bolt

Figure 2-2. Flange Bolt Torquing Sequence

2-4 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Wafer-Style Flow Meter Installation


Install the wafer-style meter between two conventional pipe flanges of the
same nominal size as the flow meter. If the process fluid is a liquid, make
sure the meter is located where the pipe is always full. This may require
locating the meter at a low point in the piping system. Note: Vortex flow
meters are not suitable for two-phase flows (i.e., liquid and gas mixtures).
For horizontal pipelines having a process temperature above 300° F,
mount the meter at a 45 or 90-degree angle to avoid overheating the elec-
tronics enclosure. To adjust the viewing angle of the enclosure or dis-
play/keypad, see page 2-18 and 2-19.

Caution!
When using toxic or corrosive Figure 2-3. Wafer-Style Flow Meter Installation
gases, purge the line with inert
gas for a minimum of four hours
When installing the meter make sure the section marked with a flow arrow is
at full gas flow before installing positioned upstream of the outlet, with the arrow head pointing in the direc-
the flow meter. tion of flow. (The mark is on the wafer adjacent to the enclosure mounting
neck.) This ensures that the sensor head is positioned downstream of the vor-
tex shedder bar and is correctly aligned to the flow. Installing the meter oppo-
site this direction will result in completely inaccurate flow measurement. To
install the meter:
1. Turn off the flow of process gas, liquid or steam. Verify that the line
is not pressurized. Confirm that the installation site meets the required
minimum upstream and downstream pipe diameters.
2. Insert the studs for the bottom side of the meter body between the pipe
flanges. Place the wafer-style meter body between the flanges with the
end stamped with a flow arrow on the upstream side, with the arrow
head pointing in the direction of flow. Center the meter body inside the
diameter with respect to the inside diameter of the adjoining piping.
3. Position the gasket material between the mating surfaces. Make sure
both gaskets are smooth and even with no gasket material extending in-
to the flow profile. Obstructions in the pipeline will disturb the flow and
cause inaccurate measurements.

M-000-00010 2-5
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

4. Place the remaining studs between the pipe flanges. Tighten the nuts in
the sequence shown in Figure 2-2. Check for leaks after tightening the
flange bolts

Flange-Style Flow Meter Installation


Install the flange-style meter between two conventional pipe flanges of
the same nominal size as the flow meter. If the process fluid is a liquid,
make sure the meter is located where the pipe is always full. This may re-
quire locating the meter at a low point in the piping system. Note: Vortex
flow meters are not suitable for two-phase flows (i.e., liquid and gas mix-
tures). For horizontal pipelines having a process temperature above 300°
F, mount the meter at a 45 or 90-degree angle to avoid overheating the
electronics enclosure. To adjust the viewing angle of the enclosure or dis-
play/keypad, see page 2-18 and 2-19.

Caution! Figure 2-4. Flange-Style Flow Meter Installation


When using toxic or corrosive
gases, purge the line with inert
gas for a minimum of four hours When installing the meter make sure the flange marked with a flow arrow is
at full gas flow before installing positioned upstream of the outlet flange, with the arrow head pointing in the
the flow meter. direction of flow. (The mark is on the flange adjacent to the enclosure
mounting neck.) This ensures that the sensor head is positioned downstream
of the vortex shedder bar and is correctly aligned to the flow. Installing the
meter opposite this direction will result in completely inaccurate flow meas-
urement. To install the meter:

1. Turn off the flow of process gas, liquid or steam. Verify that the line
is not pressurized. Confirm that the installation site meets the required
minimum upstream and downstream pipe diameters.

2-6 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

2. Seat the meter level and square on the mating connections with the flange
stamped with a flow arrow on the upstream side, with the arrow head
pointing in the direction of flow. Position a gasket in place for each side.
Make sure both gaskets are smooth and even with no gasket material ex-
tending into the flow profile. Obstructions in the pipeline will disturb the
flow and cause inaccurate measurements.

3. Install bolts in both process connections. Tighten the nuts in the se-
quence shown in Figure 2-2. Check for leaks after tightening the flange
bolts.

Series M23 Insertion Flow Meter Installation


Prepare the pipeline for installation using either a cold tap or hot tap
method described on the following pages. Refer to a standard code for all
pipe tapping operations. The following tapping instructions are general in
nature and intended for guideline purposes only. Before installing the me-
ter, review the mounting position and isolation value requirements given
below.

Mounting Position

Allow clearance between the electronics enclosure top and any other ob-
struction when the meter is fully retracted.

Isolation Valve Selection


An isolation valve is available as an option with Series M23 meters. If
you supply the isolation valve, it must meet the following requirements:

1. A minimum valve
bore diameter of 1.875-inch min.
v alv e bore
1.875 inches is re-
2-inch min.
quired, and the
valve’s body size
should be two inches.
Normally, gate 2-inch
valves are used. v alv e size

2. Verify that the


valve’s body and Isolation Valve Requirements
flange rating are
within the flow meter’s maximum operating pressure and tempera-
ture.

3. Choose an isolation valve with at least two inches existing between


the flange face and the gate portion of the valve. This ensures that the
flow meter’s sensor head will not interfere with the operation of the
isolation valve.

M-000-00010 2-7
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Cold Tap Guidelines


Refer to a standard code for all pipe tapping operations. The following
tapping instructions are general in nature and intended for guideline pur-
poses only.

1. Turn off the flow of process gas, liquid or steam. Verify that the line
Caution! is not pressurized.
When using toxic or
corrosive gases, purge 2. Confirm that the installation site meets the minimum upstream and
the line with inert gas downstream pipe diameter requirements. See Figure 2-1.
for a minimum of four
hours at full gas flow
before installing the 3. Use a cutting torch or sharp cutting tool to tap into the pipe. The pipe
flow meter. opening must be at least 1.875 inches in diameter. (Do not attempt to
insert the sensor probe through a smaller hole.)

4. Remove all burrs from the tap. Rough edges may cause flow profile
distortions that could affect flow meter accuracy. Also, obstructions
could damage the sensor assembly when inserting into the pipe.

5. After cutting, measure the thickness of the cut-out and record this
number for calculating the insertion depth.

Warning! 6. Weld the flow meter pipe connection


All flow meter connec- on the pipe. Make sure this connection
tions, isolation valves is within ± 5° perpendicular to the pipe
and fittings for cold tap- centerline.
ping must have the same
or higher pressure rating
as the main pipeline. 7. Install the isolation valve (if used).

8. When welding is complete and all fittings are installed, close the
isolation valve or cap the line. Run a static pressure check on the
welds. If pressure loss or leaks are detected, repair the joint and re-
test.

9. Connect the meter to the pipe process connection.

10. Calculate the sensor probe insertion depth and insert the sensor probe
into the pipe as described on the following pages.

2-8 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Hot Tap Guidelines


Refer to a standard code for all pipe tapping operations. The following
tapping instructions are general in nature and intended for guideline pur-
poses only.
Warning!
Hot tapping must be 1. Confirm that the installation site meets the minimum upstream and
performed by a trained downstream pipe diameter requirements.
professional. US. regulations
often require a hot tap permit.
The manufacturer of the hot 2. Weld a two inch mounting adapter on the pipe. Make sure the mount-
tap equipment and/or the ing adapter is within ± 5° perpendicular to the pipe centerline (see
contractor performing the hot previous page). The pipe opening must be at least 1.875 inches in di-
tap is responsible for provid-
ameter.
ing proof of such a permit.

3. Connect a two inch process connection on the mounting adapter.

4. Connect an isolation valve on the process connection. The valve’s full


open bore must be at least 1.875 inches in diameter.

5. Run a static pressure check on the welds. If pressure loss or leaks are
Warning!
All flow meter connections,
detected, repair the joint and re-test.
isolation valves, and fittings
for hot tapping must have the 6. Connect the hot tapping equipment to the isolation valve, open the
same or higher pressure isolation valve and drill at least a 1.875 inch diameter hole.
rating as the main pipeline.
7. Retract the drill, close the isolation valve, and remove the hot tapping
equipment.

8. Connect the flow meter to the isolation valve and open the isolation
valve.

9. Calculate the sensor probe insertion depth and insert the sensor probe
into the pipe as described on the following pages.

M-000-00010 2-9
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Connect isolation
valve and test for
leaks

xxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxx
Purge pipe
xxxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxxx

Figure 2-5. Hot Tap Sequence

2-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Flow Meter Insertion


The sensor head must be properly positioned in the pipe. For this reason,
it is important that insertion length calculations are carefully followed. A
sensor probe inserted at the wrong depth in the pipe will result in inaccu-
rate readings.

Insertion flow meters are applicable to pipes 2 inch and larger. For pipe
sizes ten inches and smaller, the centerline of the meter’s sensing head is
located at the pipe’s centerline. For pipe sizes larger than ten inches, the
centerline of the sensing head is located in the pipe’s cross section five
inches from the inner wall of the pipe; i.e., its “wetted” depth from the
wall to the centerline of the sensing head is five inches.

Insertion flow meters are available in three probe lengths:

Standard Probe configuration is used with most flow meter process


connections. The length, S, of the stem is 29.47 inches.

Compact Probe configuration is used with compression fitting process


connections. The length, S, of the stem is 13.1 inches.

12-Inch Extended Probe configuration is used with exceptionally lengthy


flow meter process connections. The length, S, of the stem is 41.47 inch-
es.

Use the Correct Insertion Formula


Depending on your flow meter’s process connection, use the applicable
insertion length formula and installation procedure as follows:

• Flow meters with a compression type connection (NPT or flanged)


follow the instructions beginning on page 2-11.

Warning! • Flow meters with a packing gland type connection (NPT or flanged)
An insertion tool must be
used for any installation configured with an insertion tool, follow the instructions beginning on
where a flow meter is page 2-13.
inserted under pressure
greater than 50 psig. • Flow meters with a packing gland type connection (NPT or flanged)
without an insertion tool, follow the instructions beginning on page
2-16.

M-000-00010 2-11
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Installing Flow Meters with a Compression Connection*


Use the following formula to determine insertion length for flow meters
(NPT and flanged) with a compression process connection. The installa-
tion procedure is given on the next page.

Insertion Length Formula


I=S–F–R–t
Where:
I = Insertion length.
S = Stem length – the distance from the center of the sensor head to the base
of the enclosure adapter (S = 29.47 inches for standard probes; S = 13.1
inches for compact; S = 41.47 inches for 12-inch extended).
F = Distance from the raised face of the flange or top of NPT stem housing to
the outside of the pipe wall.
R = Pipe inside diameter  2 for pipes ten inches and smaller.
R = Five inches for pipe diameters larger than ten inches.
t = Thickness of the pipe wall. (Measure the disk cut-out from the tapping
procedure or check a piping handbook for thickness.)

Figure 2-6. Insertion Calculation (Compression Type)


Example:
To install a Series M23 meter with a standard probe (S = 29.47 inches) into a 14
inch schedule 40 pipe, the following measurements are taken:
F=3 inches
R=5 inches
t=0.438 inches
The insertion length for this example is 21.03 inches. Insert the stem through
the fitting until an insertion length of 21.03 inches is measured with a ruler.

*All dimensions are in inches

2-12 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation
Insertion Procedure for Meters with a Compression Connection

Figure 2-7. Flow Meter with Compression Type Fitting

1. Calculate the required sensor probe insertion length.


Caution! 2. Fully retract the stem until the sensor head is touching the bottom of
The sensor alignment
pointer must point
the stem housing. Slightly tighten the compression nut to prevent
downstream, in the slippage.
direction of flow.
3. Bolt or screw the flow meter assembly into the process connection.
Use Teflon tape or pipe sealant to improve the seal and prevent seiz-
ing on NPT styles.
Warning! 4. Hold the meter securely while loosening the compression fitting. In-
To avoid serious injury,
DO NOT loosen the sert the sensor into the pipe until the calculated insertion length, I, is
compression fitting measured between the base of the enclosure adapter and the top of
under pressure. the stem housing, or to the raised face of the flanged version. Do
not force the stem into the pipe.

5. Align the sensor head using the sensor alignment pointer. Adjust the
alignment pointer parallel to the pipe and pointing downstream.

6. Tighten the compression fitting to lock the stem in position. When


the compression fitting is tightened, the position is permanent.

M-000-00010 2-13
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Installing Flow Meters with a Packing Gland Connection*


Use the formula below to determine the insertion depth for flow meters
(NPT and flanged) equipped with an insertion tool. To install, see the next
page for instructions for meters with a permanent insertion tool. For me-
ters with a removable insertion tool, see page 2-15.

Insertion Length Formula


I = F + R + t – 1.35

Where:

I = Insertion length.

F = Distance from the raised face of the flange or top of


the process connection for NPT style meters to the top
outside of the process pipe.

R = Pipe inside diameter  2 for pipes ten inches & small-


er.
R = Five inches for pipe diameters larger than ten inches.

t = Thickness of the pipe wall. (Measure the disk cut-


out from the tapping procedure or check a piping hand-
book for thickness.)

Figure 2-8. Insertion Calculation (Meters with Insertion Tool)


Example 1: Flange Style Meters:
To install a Series M23 Flow Meter into a 14 inch schedule 40 pipe, the fol-
lowing measurements are taken:
F = 12 inches
R = 5 inches
t = 0.438 inches
The example insertion length is 16.09 inches.
Example 2: NPT Style Meters:
The length of thread engagement on the NPT style meters is also subtracted in
the equation. The length of the threaded portion of the NPT meter is 1.18
inches. Measure the thread portion still showing after the installation and sub-
tract that amount from 1.18 inches. This gives you the thread engagement
length. If this cannot be measured use .55 inch for this amount.
F = 12 inches
R = 5 inches
t = 0.438 inches
The example insertion length is 15.54 inches.

*All dimensions are in inches.

2-14 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Insertion Procedure for Flow Meters with Permanent Insertion Tool

Figure 2-9. Flow Meter with Permanent Insertion Tool

1. Calculate the required sensor probe insertion length (see previous


page). Measure from the depth marker arrow down the stanchion and
scribe a mark at the calculated insertion depth.
Caution!
The sensor alignment
pointer must point 2. Fully retract the flow meter until the sensor head is touching the bot-
downstream, in the tom of the stem housing. Attach the meter assembly to the two inch
direction of flow. full-port isolation valve, if used. Use Teflon tape or pipe sealant to
improve seal and prevent seizing on NPT style.

3. Loosen the two packing gland nuts on the stem housing of the meter.
Loosen the stem lock bolt adjacent to the sensor alignment pointer.
Align the sensor head using the sensor alignment pointer. Adjust the
Note alignment pointer parallel to the pipe and pointing downstream. Tighten
If line pressure is above the stem lock bolt to secure the sensor position.
500 psig, it could require
up to 25 ft lb of torque to 4. Slowly open the isolation valve to the full open position. If necessary,
insert the flow meter.
Do not confuse this with
slightly tighten the two packing gland nuts to reduce the leakage
possible interference around the stem.
in the pipe.
5. Turn the insertion tool handle clockwise to insert the sensor head into
the pipe. Continue until the top of the upper retractor bracket aligns
with the insertion length position scribed on the stanchion. Do not
force the stem into the pipe.
6. Tighten the packing gland nuts to stop leakage around the stem. Do
not torque over 20 ft-lb.

M-000-00010 2-15
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Insertion Procedure for Flow Meters with Removable Insertion Tool

Figure 2-10. Flow Meter with Removable Insertion Tool

1. Calculate the required sensor probe insertion length. Measure from


the depth marker arrow down the stanchion and scribe a mark at the
calculated insertion depth.
Caution!
The sensor alignment
pointer must point
2. Fully retract the flow meter until the sensor head is touching the bot-
downstream, in the tom of the stem housing. Attach the meter assembly to the two inch
direction of flow. full-port isolation valve, if used. Use Teflon tape or pipe sealant to
improve seal and prevent seizing on NPT style.

3. Remove the two top stem clamp nuts and loosen two stem clamp
bolts. Slide the stem clamp away to expose the packing gland nuts.

4. Loosen the two packing gland nuts. Loosen the stem lock bolt adja-
cent to the sensor alignment pointer. Align the sensor head using the
Note sensor alignment pointer. Adjust the alignment pointer parallel to the
If line pressure is above pipe and pointing downstream. Tighten the stem lock bolt to secure
500 psig, it could require the sensor position.
up to 25 ft lb of torque to
insert the flow meter.
Do not confuse this with 5. Slowly open the isolation valve to the full open position. If necessary,
possible interference slightly tighten the two packing gland nuts to reduce the leakage
in the pipe. around the stem.

6. Turn the insertion tool handle clockwise to insert the stem into the
pipe. Continue until the top of the upper retractor bracket lines up
with the insertion length mark scribed on the stanchion. Do not force
the stem into the pipe.

2-16 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

7. Tighten the packing gland nuts to stop leakage around the stem. Do not
torque over 20 ft-lbs.

8. Slide the stem clamp back into position. Torque stem clamp bolts to 15
ft-lbs. Replace the stem clamp nuts and torque to 10-15 ft-lbs.

9. To separate the insertion tool from the flow meter, remove four socket
head cap bolts securing the upper and lower retractor brackets. Remove
the insertion tool.
Installation of Meters with Packing Gland Connection (No Insertion Tool)*
Use the following formula to determine insertion depth for meters with a
packing gland connection (NPT and flanged) without an insertion tool.

Insertion Length Formula


I=S–F–R–t
Where:

I = Insertion length.

S = Stem length – the distance from the


center of the sensor head to the base of the
enclosure adapter (S = 29.47 inches for
standard probes; S = 41.47 inches for 12 inch
extended probes).

F = Distance from the raised face of the


flange or top of NPT stem housing to the
outside of the pipe wall.

R = Pipe inside diameter  2 for pipes ten


inches & smaller.
R = Five inches for pipe diameters larger than ten
inches.

t = Thickness of the pipe wall. (Measure the disk


cut-out from the tapping procedure or check a piping
handbook for thickness.)

Figure 2-11. Insertion Calculation (Meters without Insertion Tool)

Example:
To install a Series M23 Flow Meter with a standard probe (S = 29.47) into
a 14 inch schedule 40 pipe, the following measurements are taken:

F = 3 inches
R = 5 inches
t = 0.438 inches
The example insertion length is 21.03 inches.

*All dimensions are in inches.

M-000-00010 2-17
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Insertion Procedure for Flow Meters with No Insertion Tool


(Packing Gland Connection)

1. Calculate the required sensor probe insertion length.


Warning!
The line pressure 2. Fully retract the stem until the sensor head is touching the bottom of the
must be less than stem housing. Remove the two top stem clamp nuts and loosen two
50 psig for installation. stem clamp bolts. Slide the stem clamp away to expose the packing
gland nuts. Loosen the two packing gland nuts.

3. Align the sensor head using the sensor alignment pointer. Adjust the
alignment pointer parallel to the pipe and pointing downstream.
Caution!
The sensor alignment 4. Insert the sensor head into the pipe until insertion length, I, is
pointer must point achieved. Do not force the stem into the pipe.
downstream, in the
direction of flow.
5. Tighten the packing gland nuts to stop leakage around the stem. Do
not torque over 20 ft-lbs.

6. Slide the stem clamp back into position. Torque stem clamp bolts to
15 ft-lbs. Replace the stem clamp nuts and torque to 10-15 ft-lbs.

2-18 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Adjusting Meter Orientation


Depending on installation requirements, you may need to adjust the meter
orientation. There are two adjustments available. The first rotates the posi-
tion of the LCD display/keypad and is available on both in-line and inser-
tion meters. The second is to rotate the enclosure position. This adjust-
ment is only allowed on Series M22/M24 In-Line meters.

Display/Keypad Adjustment (All Meters)

Figure 2-12. Display/Keypad Viewing Adjustment

The electronics boards are electrostatically sensitive. Wear a grounding


wrist strap and make sure to observe proper handling precautions required
for static-sensitive components. To adjust the display:

1. Disconnect power to the flow meter.

2. Loosen the small set screw which secures the electronics enclosure
cover. Unscrew and remove the cover.

3. Loosen the 4 captive screws.

4. Carefully pull the display/microprocessor board away from the meter


standoffs. Make sure not to damage the connected ribbon cable.

5. Rotate the display/microprocessor board to the desired position. Max-


imum turn, two positions left or two positions right (180-degrees).

6. Align the board with the captive screws. Check that the ribbon cable
is folded neatly behind the board with no twists or crimps.

7. Tighten the screws. Replace the cover and set screw. Restore power to
the meter.

M-000-00010 2-19
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Enclosure Adjustment (Series M22/M24 Only)

Figure 2-13. Enclosure Viewing Adjustment

To avoid damage to the sensor wires, do not rotate the enclosure beyond
180-degrees from the original position. To adjust the enclosure:

1. Remove power to the flow meter.

2. Loosen the three set screws shown above. Rotate the display to the
desired position (maximum 180-degrees).

3. Tighten the three set screws. Restore power to the meter.

2-20 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Loop Power Flow Meter Wiring Connections


The NEMA 4X enclosure contains an integral wiring compartment with
one dual strip terminal block (located in the smaller end of the enclosure).
Two 3/4-inch female NPT conduit entries are available for separate power
Warning!
To avoid potential electric and signal wiring. For all hazardous area installations, only suitable certi-
shock, follow National Electric fied cable glands, blanking plugs or thread adapters may be used. The ca-
Code safety practices or your ble entry device shall be of a certified flameproof type, suitable for the
local code when wiring this conditions of use and correctly installed. The degree of protection of at
unit to a power source and to
peripheral devices. Failure to
least IP66 to EN 60529 is only achieved if certified cable entries are used
do so could result in injury or that are suitable for the application and correctly installed. Unused aper-
death. All wiring procedures tures shall be closed with suitable blanking elements. If conduit seals are
must be performed with the used, they must be installed within 18 inches (457 mm) of the enclosure.
power off.

Use a Class I or Class II


power supply.

A power switch is not provid-


ed with this meter, an ap-
proved switch meeting the
power requirements listed in OPTIONAL
Appendix A must be provided LOOP PULSE FREQ BACKLIGHT
POWER OUT OUT POWER OPTION 2
by the user. It must be easily + - + - + - + - 1 2 3 4 5
accessible and marked as the
disconnect for the flow meter.

Only the connectors supplied


with the meter are to be used
for connecting wiring.

If the equipment is used in a


manner not specified the pro-
tection provided by the
equipment may be impaired.

Figure 2-14. Loop Power Wiring Terminals


Input Power Connections
To access the wiring terminal blocks, locate and loosen the small set
screw which locks the small enclosure cover in place. Unscrew the cover
to expose the terminal block.

Caution! DC Power Wiring


The DC wire insulation tem- Connect 4-20 mA loop power (12 to 36 VDC at 25 mA, 1W max.) to the
perature rating must meet or +Loop Power and –Loop Power terminals on the terminal block. Torque
exceed 85°C (185°F), maxi-
mum operating voltage 300 all connections to 4.43 to 5.31 in-lbs (0.5 to 0.6 Nm). The DC power wire
VRMS. size must be 20 to 12 AWG with the wire stripped 1/4 inch (7 mm).

12 to 36 VDC
25 mA max.

Figure 2-15. DC Power Connections


M-000-00010 2-21
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

4-20 mA Output Connections


The Pro-V meter has a single 4-20 mA loop. The 4-20 mA loop current is
controlled by the meter electronics. The electronics must be wired in se-
ries with the sense resistor or current meter. The current control electron-
ics require 12 volts at the input terminals to operate correctly.

The maximum loop resistance (load) for the current loop output is de-
pendent upon the supply voltage and is given in Figure 2-16. The 4-20
mA loop is optically isolated from the flow meter electronics.

Rload is the total resistance in the loop, including the wiring resistance
(Rload = Rwire + Rsense ). To calculate Rmax, the maximum Rload for the loop,
subtract the minimum terminal voltage from the supply voltage and divide
by the maximum loop current, 20 mA. Thus:

The maximum resistance Rload = Rmax = (Vsupply – 12V) / 0.020 A

Figure 2-16. Load Resistance Versus Input Voltage

The current loop range is 3.8 to 20.5 mA.

2-22 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Pulse Output Connections


The pulse output is used for a remote counter. When the preset volume or
mass (defined in the totalizer settings, see page 3-10) has passed the me-
ter, the output provides a 50 millisecond square pulse.

The pulse output requires a separate 5 to 36 VDC power supply. The


pulse output optical relay is a normally-open single-pole relay. The relay
can conduct a current up to 40 mA. It is isolated from the meter electron-
ics and power supply.

Figure 2-17. Isolated Pulse Output Using External Power Supply

Figure 2-18. Non-Isolated Pulse Output Using External Power Supply

M-000-00010 2-23
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Frequency Output Connections


The frequency output is used for a remote counter. It can be scaled to out-
put a 1 to 10 kHz signal proportional to mass or volume flow, tempera-
ture, pressure or density.

The frequency output requires a separate 5 to 36 VDC power supply. The


frequency output optical relay is a normally-open single-pole relay. The
output can conduct a current up to 40 mA. It is isolated from the meter
electronics and power supply.

Freq. Out + Freq. Out voltage = +V


Select resistor so that current
Freq. Out - through Freq . Out <= 40 mA

Figure 2-19. Isolated Frequency Output Using External Power Supply

Freq. Out +
Freq. Out -

Freq. Out voltage = +V


Select resistor so that current through Freq. Out <= 40 mA

Figure 2-20. Non-Isolated Frequency Output Using External Power Supply

Optional Backlight Connection


The loop power meter has an optional backlight connection provided. It is
intended to be powered by a separate 12 to 36 VDC at 35 mA max. power
supply or by the pulse power input. Both options are shown below.

12 to 36 VDC
35 mA max.

Figure 2-21.Backlight Using External Power Supply

2-24 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Remote Electronics Wiring


The remote electronics enclosure should be mounted in a convenient, easy
to reach location. For hazardous location installations, make sure to ob-
serve agency requirements for installation. Allow some slack in the inter-
face cable between the junction box and the remote electronics enclosure.
To prevent damage to the wiring connections, do not put stress on the
terminations at any time.

The meter is shipped with temporary strain relief glands at each end of the
cable. Disconnect the cable from the meter’s terminal block inside the
junction box–not at the remote electronics enclosure. Remove both glands
and install appropriate conduit entry glands and conduit. The cable entry
device shall be of a certified flameproof type, suitable for the conditions
of use and correctly installed. The degree of protection of at least IP66 to
EN 60529 is only achieved if certified cable entries are used that are suit-
able for the application and correctly installed. Unused apertures shall be
closed with suitable blanking elements. When installation is complete, re-
connect each labeled wire to the corresponding terminal position on the
junction box terminal block. Make sure to connect each wire pair’s shield.
Note: incorrect connection will cause the meter to malfunction.

Note: Numeric code on junction box label matches wire labels.

NOT USED

FLOW
PWR BLK 1
GND RED 1
SENSOR V1 BLK 2
SENSOR V2 RED 2
SHIELD SHLD 1&2
NOT USED

Figure 2-22. Loop Power Volumetric Flowmeter Junction Box Sensor Connections
(Wires enter the flow connector from the left side of the connector shown
above.)

M-000-00010 2-25
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

VORTEX
PWR BLK 1
GND RED 1
SENSOR V1 BLK 2
SENSOR V2 RED 2
SHIELD SHLD 1&2

Figure 2-23. Loop Power Volumetric Flowmeter Junction Box Sensor Con-
nections Supplied Prior to Jan. 1, 2014 (Wires enter the flow connector
from the right side of the connector shown above.)

TEMPERATURE
SHIELD SHLD 5&6
E1 T1 RED 6
S1 T2 BLK 6
S2 T3 BLK 5
VORTEX E2 T4 RED 5
PWR BLK 1
GND RED 1
SENSOR V1 BLK 2
SENSOR V2 RED 2
SHIELD SHLD 1&2
SHIELD SHLD 3&4
P1 BLK 3
P2 BLK 4
P3 RED 4
P4 RED 3
PRESSURE

Figure 2-24. Loop Power Mass Flowmeter Junction Box Sensor Connections

2-26 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

High Power Meter Wiring Connections


The NEMA 4X enclosure contains an integral wiring compartment with
one dual strip terminal block (located in the smaller end of the enclosure).
Two 3/4-inch female NPT conduit entries are available for separate power
and signal wiring. For all hazardous area installations, only suitable certi-
fied cable glands, blanking plugs or thread adapters may be used. The ca-
Warning! ble entry device shall be of a certified flameproof type, suitable for the
To avoid potential electric shock, follow conditions of use and correctly installed. The degree of protection of at
National Electric Code safety practices or least IP66 to EN 60529 is only achieved if certified cable entries are used
your local code when wiring this unit to a that are suitable for the application and correctly installed. Unused aper-
power source and to peripheral devices.
Failure to do so could result in injury or tures shall be closed with suitable blanking elements. If conduit seals are
death. All AC power connections must be used, they must be installed within 18 inches (457 mm) of the enclosure.
in accordance with published CE direc-
tives. All wiring procedures must be per-
formed with the power off. OPTION 1 OPTION 2
1 2 3 4 5 1 2 3 4 5

A power switch is not provided with this


meter, an approved switch meeting the NEUT / L2
HOT / L1

FREQ PULSE ALARM ALARM ALARM


OUT OUT 1 2 3
power requirements listed in Appendix A + - + - + - + - + -
must be provided by the user. It must be
easily accessible and marked as the dis-
connect for the flow meter.
AC
PWR
Only the connectors supplied with the me- IN
+ - + - + - + - +
ter are to be used for connecting wiring.
RS485 GND

-
24 4-20 4-20 4-20

RS485
RS485
VDC mA 1 mA 2 mA 3
OUT
If the equipment is used in a manner not
specified the protection provided by the
equipment may be impaired.

Figure 2-25. AC Wiring Terminals


Input Power Connections
To access the wiring terminal blocks, locate and loosen the small set
screw which locks the small enclosure cover in place. Unscrew the cover
to expose the terminal block.
AC Power Wiring
The AC power wire size must be 20 to 10 AWG with the wire stripped
1/4 inch (7 mm). The wire insulation temperature must meet or exceed
Caution! 90°C (194°F). Connect 100 to 240 VAC (5 W maximum) to the Hot and
The AC wire insulation temperature rating Neutral terminals on the terminal block. Connect the ground wire to the
must meet or exceed 90°C (194°F), max- safety ground lug. Torque all connections to 4.43 to 5.31 in-lbs (0.5 to 0.6
imum operating voltage 600 VRMS.
Nm). Use a separate conduit entry for signal lines to reduce the possibility
of AC noise interference.

100 to 240 VAC, 50/60 Hz, 5 Watts Max.

Chassis screw safety


ground must be used
Hot / L1
for proper operation.
Neutral / L2

Figure 2-26. AC Power Connections


M-000-00010 2-27
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

OPTION 1 OPTION 2
1 2 3 4 5 1 2 3 4 5

FREQ PULSE ALARM ALARM ALARM


OUT OUT 1 2 3
+ - + - + - + - + -

+ - + - + - + - +

RS485 GND

-
DC 4-20 4-20 4-20

RS485
RS485
PWR mA 1 mA 2 mA 3

Figure 2-27. DC Wiring Terminals

DC Power Wiring
The DC power wire size must be 20 to 12 AWG with the wire
Warning! stripped 1/4 inch (7 mm). Connect 18 to 36 VDC (300 mA, 9 W max-
Use a Class I or Class II power supply. imum) to the +DC Pwr and –DC Pwr terminals on the terminal block.
A power switch is not provided with this me-
Torque all connections to 4.43 to 5.31 in-lbs (0.5 to 0.6 Nm).
ter, an approved switch meeting the power
requirements listed in Appendix A must be
provided by the user. It must be easily ac-
18 to 36 VDC @ 300 mA Max.
cessible and marked as the disconnect for
the flow meter.

Only the connectors supplied with the meter


are to be used for connecting wiring.

If the equipment is used in a manner not


specified the protection provided by the Figure 2-28. DC Power Connections
equipment may be impaired

Caution!
The DC wire insulation temperature rating
must meet or exceed 85°C (185°F), maxi-
mum operating voltage 300 VRMS.

2-28 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

4-20 mA Output Connections


The standard Pro-V Flow Meter has a single 4-20 mA loop. Two addi-
tional loops are available on the optional communication board. The 4-20
mA loop current is controlled by the meter electronics. The electronics
must be wired in series with the sense resistor or current meter. The cur-
rent control electronics require 12 volts at the input terminals to operate
correctly.

The maximum loop resistance (load) for the current loop output is depend-
ent upon the supply voltage and is given in Figure 2-29. The 4-20 mA loop
is optically isolated from the flow meter electronics.

Rload is the total resistance in the loop, including the wiring resistance
(Rload = Rwire + Rsense ). To calculate Rmax, the maximum Rload for the loop,
subtract the minimum terminal voltage from the supply voltage and divide
by the maximum loop current, 20 mA. Thus:

The maximum resistance Rload = Rmax = (Vsupply – 12V) / 0.020 A

Figure 2-29. Load Resistance Versus Input Voltage

The current loop range is 3.8 to 20.5 mA.

mA
Meter For Hart Communications
signal loop must have a
minimum of 250 ohms load
RL > 250 resistance R L

Figure 2-30. Isolated 4–20 mA Output Using External Power Supply

M-000-00010 2-29
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

mA Meter For HART


communications
the signal loop
must have a
DC Power minimum of 250
ohms load
DC Common resistance.

RL > 250 Ohm

DC powered meters only

Figure 2-31. Non-Isolated 4–20 mA Output Using Meter Input Power Supply

mA Meter For HART


communications
the signal loop
24 VDC must have a
minimum of 250
24 VDC ohms load
resistance.
RL > 250 Ohm
AC units only
Meter Provided DC
Power

Figure 2-32. Isolated 4–20 mA Output Using Meter Provided Power Supply

Frequency Output Connections


The frequency output is used for a remote counter. It can be scaled to out-
put a 1 to 10 kHz signal proportional to mass or volume flow, tempera-
ture, pressure or density.

The frequency output requires a separate 5 to 36 VDC power supply. The


frequency output optical relay is a normally-open single-pole relay. The
output can conduct a current up to 40 mA. It is isolated from the meter
electronics and power supply.

There are three connection options for the frequency output–the first with a
separate power supply (Figure 2-33), the second using the flow meter power
supply (Figure 2-34)(DC powered units only), and the third using the inter-
nal 24 VDC power supply (Figure 2-35)(AC powered units only). Use the
first option with a separate power supply (5 to 36 VDC) if a specific voltage
is needed for the frequency output. Use the second configuration if the volt-
age at the flow meter power supply is an acceptable driver voltage for the
load connected. (Take into account that the current used by the frequency
load comes from the meter’s power supply). Use the third configuration if
you have an AC powered unit only. In any case, the voltage of the frequen-
cy output is the same as the voltage supplied to the circuit.

2-30 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Freq. Out + Freq. Out voltage = +V


Select resistor so that current
Freq. Out - through Freq . Out <= 40 mA

Figure 2-33. Isolated Frequency Output Using External Power Supply

PWR Freq. Out +


PWR Freq. Out -
DC Powered meters only
Freq. Out voltage =

Figure 2-34. Non-Isolated Frequency Output Using Input Power Supply

+ 24 VDC Out Freq. Out +


- 24 VDC Out Freq. Out -
AC units only
Meter provided DC Power
Freq. Out voltage = +24 VDC

Figure 2-35. Isolated Frequency Output Using Meter Provided Power Supply

Pulse Output Connections


The pulse output is used for a remote counter. When the preset volume or
mass (defined in the totalizer settings, see page 3-10) has passed the me-
ter, the output provides a 50 millisecond square pulse.

The pulse output requires a separate 5 to 36 VDC power supply. The


pulse output optical relay is a normally-open single-pole relay. The relay
can conduct a current up to 40 mA. It is isolated from the meter electron-
ics and power supply.

There are three connection options for the pulse output–the first with a sepa-
rate power supply (Figure 2-36), the second using the flow meter power
supply (Figure 2-37)(DC powered units only), and the third using the inter-
nal 24 VDC power supply (Figure 2-38)(AC powered units only). Use the
first option with a separate power supply (5 to 36 VDC) if a specific voltage
is needed for the pulse output. Use the second configuration if the voltage at
the flow meter power supply is an acceptable driver voltage for the load
connected. (Take into account that the current used by the pulse load comes
from the meter’s power supply). Use the third configuration if you have an

M-000-00010 2-31
Chapter 2 Installation Series M22/M23/M24 Instruction Manual
AC powered unit only. In any case, the voltage of the pulse output is the
same as the voltage supplied to the circuit.

Figure 2-36. Isolated Pulse Output Using External Power Supply

PWR
PWR

DC Powered meters only

Figure 2-37. Non-Isolated Pulse Output Using Input Power Supply

+ 24 VDC Out
- 24 VDC Out
AC units only
Meter provided DC Power
+24 VDC

Figure 2-38. Isolated Pulse Output Using Meter Provided Power Supply

2-32 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Alarm Output Connections


One alarm output (Alarm 1) is included on the standard Pro-V™ Flow
Meter. Two or more alarms (Alarm 2 and Alarm 3) are included on the
optional communication board. The alarm output requires a separate 5 to
36 VDC power supply. The alarm output optical relay is a normally-open
single-pole relay. The relay can conduct a current up to 40 mA. It is iso-
lated from the meter electronics and power supply. When the alarm relay
is closed, the current draw will be constant. Make sure to size Rload appro-
priately.

There are three connection options for the alarm output–the first with a sep-
arate power supply (Figure 2-39), the second using the flow meter power
supply (Figure 2-40)(DC powered units only) and the third with the meter
provided power supply (Figure 2-41)(AC powered units only). Use the first
option with a separate power supply (5 to 36 VDC) if a specific voltage is
needed for the alarm output. Use the second configuration if the voltage at
the flow meter power supply is an acceptable driver voltage for the load
connected. (Take into account that the current used by the alarm load comes
from the meter’s power supply). Use the third if you have an AC powered
unit only. In any case, the voltage of the alarm output is the same as the
voltage supplied to the circuit.

The alarm output is used for transmitting high or low process conditions
as defined in the alarm settings (see page 3-9).

AC or DC powered meters

ALARM

ALARM

Figure 2-39. Isolated Alarm Output Using External Power Supply

DC units only

Figure 2-40. Non-Isolated Alarm Output Using Internal Power Supply

M-000-00010 2-33
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

+ 24 VDC Out
- 24 VDC Out
AC units only
Meter provided
DC Power +24 VDC

Figure 2-41. Isolated Alarm Output Using Meter Provided Power Supply
Remote Electronics Wiring
The remote electronics enclosure should be mounted in a convenient, easy
to reach location. For hazardous location installations, make sure to ob-
serve agency requirements for installation. Allow some slack in the inter-
face cable between the junction box and the remote electronics enclosure.
To prevent damage to the wiring connections, do not put stress on the
terminations at any time.

The meter is shipped with temporary strain relief glands at each end of the
cable. Disconnect the cable from the meter’s terminal block inside the
junction box–not at the remote electronics enclosure. Remove both glands
and install appropriate conduit entry glands and conduit. The cable entry
device shall be of a certified flameproof type, suitable for the conditions
of use and correctly installed. The degree of protection of at least IP66 to
EN 60529 is only achieved if certified cable entries are used that are suit-
able for the application and correctly installed. Unused apertures shall be
closed with suitable blanking elements. When installation is complete, re-
connect each labeled wire to the corresponding terminal position on the
junction box terminal block. Make sure to connect each wire pair’s shield.
Note: incorrect connection will cause the meter to malfunction.

TEMPERATURE
SHIELD SHLD 5&6
E1 T1 RED 6
S1 T2 BLK 6
S2 T3 BLK 5
VORTEX E2 T4 RED 5
PWR BLK 1
GND RED 1
SENSOR V1 BLK 2
SENSOR V2 RED 2
SHIELD SHLD 1&2
SHIELD SHLD 3&4
P1 BLK 3
P2 BLK 4
P3 RED 4
P4 RED 3
PRESSURE

Figure 2-42. High Power Flow Meter Junction Box Sensor Connections

Note: Numeric code on junction box label matches wire labels.


2-34 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Optional Input Electronics Wiring


The meter has two optional input wiring terminals, maximum wire size is
16 AWG. These can be used to input a Remote or Second RTD input in
the case of an Energy Monitoring meter, for the input of a Remote Pres-
sure Transducer, to pass a Contact Closure or for a Remote Density meas-
urement to name a few. In any case, the wiring diagram will be included
with the meter if any of the options are specified. Otherwise, the optional
terminal blocks will be left blank and nonfunctional.

Option1 Option2
1 2 3 4 5 1 2 3 4 5

Optional Energy EMS RTD Input Wiring

Loop Power
OPTIONAL
LOOP PULSE FREQ BACKLIGHT
POWER OUT OUT POWER OPTION 2
+ - + - + - + - 1 2 3 4 5

Red Red Black Black


R = 1000 ohm
* Other input options available.

High Power

Option1 Option2
1 2 3 4 5 1 2 3 4 5

Red Red Black Black


R = 1000 ohm

Figure 2-43.Optional Energy EMS RTD Input Wiring

The recommended customer supplied second RTD is a Class A 1000 ohm


4-wire platinum RTD. If a second RTD is not being used, then the factory
supplied 1000 ohm resistor needs to be installed in its place.

M-000-00010 2-35
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

Optional External 4-20 mA Input Wiring


The meter is set to have Option 1 used for the external input. Program-
ming menus that pertain to the optional 4-20 mA input are located in the
Hidden Diagnostics Menu in Chapter 5.

Option1 Option2
1 2 3 4 5 1 2 3 4 5

+
External
Power
- Ext. 4-20 mA
Input Device

Figure 2-44. External 4-20 mA Input Wiring - External Power Supply

Follow the above diagram to wire the external 4-20 mA input into the
flow meter using an external power supply.

Option1 Option2
1 2 3 4 5 1 2 3 4 5

DC powered meter only.

DC PWR + DC PWR
DC COMMON - DC PWR
Ext. 4-20 mA
Input Device

Figure 2-45. External 4-20 mA Input Wiring - DC Powered Meter

Follow the above diagram to wire the external 4-20 mA input into the
flow meter using power supplied to the input of a DC powered meter.

2-36 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Option1 Option2
1 2 3 4 5 1 2 3 4 5

AC units only. Meter


provided DC power.

+ 24 VDC OUT
- 24 VDC OUT
Ext. 4-20 mA
Input Device

Figure 2-46. External 4-20 mA Input Wiring - AC Powered Meter

Follow the above diagram to wire the external 4-20 mA input into the
flow meter using power from the 24 VDC output of an AC powered me-
ter.

Optional External 4-20 mA Input and RTD Wiring

Loop Power
OPTIONAL
LOOP PULSE FREQ BACKLIGHT
OUT OUT POWER OPTION 2
POWER
+ - + - + - + - 1 2 3 4 5

Red Black
R = 1000 ohm

+
External
Power
- Ext. 4-20 mA
Input Device

Figure 2-47. External 4-20 mA Input and RTD Wiring – Loop Power

M-000-00010 2-37
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

High Power

Option1 Option 2
1 2 3 4 5 1 2 3 4 5

Red Red Black Black


External
+ R = 1000 ohm
Power
- Ext. 4-20 mA
Input Device

Figure 2-48. External 4-20 mA Input and RTD Wiring – High Power

Optional Energy EMS External 4-20 mA Input and RTD Wiring


Option1 Option 2
1 2 3 4 5 1 2 3 4 5

Red Black Red Red Black Black


R = 1000 ohm R = 1000 ohm

+
External
Power
- Ext. 4-20 mA
Input Device

Figure 2-49. Energy EMS External 4-20 mA Input and RTD Wiring – High Power

2-38 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 2 Installation

Optional External Contact Closure Input Wiring

Option1 Option2
1 2 3 4 5 1 2 3 4 5

Figure 2-50. Optional External Contact Closure Input Wiring

Follow the above diagram to wire an external switch input into the flow
meter. The meter is configured to have Option 1 used for the external in-
put. If the above switch is used to remotely reset the totalizer a pushbut-
ton switch with a momentary contact closure is recommended.

M-000-00010 2-39
Chapter 2 Installation Series M22/M23/M24 Instruction Manual

2-40 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Chapter 3 Operating Instructions


After installing the Pro-V Vortex Flow Meter, you are ready to
begin operation. The sections in this chapter explain the dis-
play/keypad commands, meter start-up and programming. The
meter is ready to operate at start up without any special pro-
Warning! gramming. To enter parameters and system settings unique to
Do not open covers when an ex-
your operation, see the following pages for instructions on using
plosive atmosphere may be pre-
sent. the setup menus.

Flow Meter Display/Keypad


The flow meter’s digital electronics allow you to set, adjust and
monitor system parameters and performance. A full range of
commands are available through the display/keypad. The LCD
Pro-VTM
display gives 2 x 16 characters for flow monitoring and pro-
VorTek
gramming. The six push-buttons can be operated with the enclo-
sure cover removed. Or, the explosion-proof cover can remain
Instruments,
LLC

in place and the keypad operated with a hand-held magnet posi-


tioned at the side of the enclosure as shown in the illustration at
the left. To secure the enclosure cover use a 1/16” hex key
wrench to tighten the 6-32 x 1/4 cup point socket set locking
screw.

Display/Keypad From the Run Mode, the ENTER key allows


access to the Setup Menus (through a
Commands
password screen). Within the Setup Menus,
pressing ENTER activates the current field.
To set new parameters, press the ENTER key

EXIT
 ENTER
until an underline cursor appears. Use the
 keys to select new parameters.
Pro-V™ Press ENTER to continue. (If change is not
allowed, ENTER has no effect.) All outputs
are disabled when using the Setup Menus.
VorTek
 
INSTRUMENTS
The EXIT key is active within the Setup Menus.
 When using a Setup Menu, EXIT returns you to the
Run Mode. If you are changing a parameter and
make a mistake, EXIT allows you to start over.

The  keys advance through each screen


of the current menu. When changing a system
parameter, all  keys are available to enter
new parameters.

Figure 3-1. Flow Meter Display/Keypad

M-000-00010 3-1
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Display Contrast Adjustment


The flow meter display contrast is set at the factory but if the dis-
play characters appear too dark or too light proceed as follows:

1) Hold down the “Exit” button on the front panel for 5 to 10 sec-
onds. “Setting Contrast” will appear.

2) Push the “Up” arrow to darken the display or the “Down” arrow
to lighten it.

3 Push the “Enter” button to save the contrast setting.

Start-Up
To begin flow meter operation:

1. Verify the flow meter is installed and wired as described in Chapter 2.


2. Apply power to the meter. At start up, the unit runs a series of self-
Note tests that check the RAM, ROM, EPROM and all flow sensing com-
Starting the flow meter ponents. After completing the self-test sequence, the Run Mode
or pressing EXIT will screens appear.
always display the
Run Mode screens. 3. The Run Mode displays flow information as determined by system
settings. Some screens depicted on the next page may not be dis-
played based on these settings. Press the  arrow keys to view
the Run Mode screens.
4. Press the ENTER key from any Run Mode screen to access the Setup
Menus. Use the Setup Menus to configure the meter’s multi-
parameter features to fit your application.

3-2 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Run Mode
Screens
ENTER
Mass Flow
Rate Password

Volume ENTER
Flow Rate
Setup Press Exit to return
Temperature
Menus to Run Mode

Pressure

* Energy * Energy EMS


Meters Only

Density

Total
Use
keys to access
Alarm 1 each item
Status

Alarm 2
Status

Alarm 3
Status

Fluid

Date & Time

M-000-00010 3-3
3-4
Run Mode
Screens
ENTER
Chapter 3 Operation

Mass Flow Password


Rate

ENTER
Setup Menus
Volume
Flow Rate
Output Display Alarms Totalizer #1
*Totalizer #2 * Energy Fluid Units Time & Date Diagnostics Calibration Password
Menu Menu Menu Menu Menu Menu Menu Menu Menu Menu Menu Menu
Temperature
Using the Setup Menus

Pressure 4-20 mA Cycle Time Relay Loc in Sent Flowing Mass Flow Sim Vor Meter Size Set
Totaling Totaling Flow Set Time
Output 1 (sec) Alarm 1 Fluid Unit Freq or Pipe ID Password
Yes / No
* Energy 4-20 mA Number of Relay Unit per Heating System STD Temp Volume
Reset Total Set Date Sim Temp 1, 2 Base Kf
Output 2 Digits Alarm 2 Pulse Yes / No (F) Flow Unit

Density 4-20 mA Display TC Relay STD Press Temperature Base


Reset Total % Returned Sim Pressure
Output 3 (sec) Alarm 3 (PSIA) Unit Reynolds

Total Scaled Mf Vf Te NORM * Energy Highest Head


Alarm Log
Frequency Pr De T Temp (C) Unit Velocity Diameter
Alarm 1
Modbus A1 A2 A3 Clear Alarm NORM Density Highest Abs
Status
Units Fl Dt *E Log Press (KPA) Unit Temp 1, 2 Roughness
Alarm 2
Status Modbus Pressure Highest Vortex Coef
Order Unit Pressure Ck
Alarm 3
Comm * Energy EMS Gauge Highest Low Flow
Status
Protocol Meters Only Pressure Int Temp Cutoff
(Abs/Gauge)
Fluid Lowest Serial
Baud Rate Int Temp Number
Atm. Press
14.6959
Date & Time
System Log
Address

Clear Sys
** BACnet Log
Device ID

** BACnet
MaxMaster ** BACnet Meters Only
Series M22/M23/M24 Instruction Manual

M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Programming the Flow Meter


1. Enter the Setup Menu by pressing the ENTER key until prompted for a password. (All
outputs are disabled while using the Setup Menus.)

2. Use the  keys to select the password characters (1234 is the factory-set
password). When the password is correctly displayed, press ENTER to continue.

3. Use the Setup Menus described on the following pages to customize the multi-
parameter features of your Pro-V Flow Meter. (The entire lower display line is
available for entering parameters.) Some items depicted in the graphic on the
preceding page may not be displayed based on flow meter configuration settings

4. To activate a parameter, press ENTER. Use the  keys to make selections. Press
ENTER to continue. Press EXIT to save or discard changes and return to Run Mode.

5. Program the UNITS menu first because later menus will be based on the units
selected.

M-000-00010 3-5
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Output Menu
ENTER
Run Mode Password

ENTER

Output Use
Menu keys to access menus

< Measure >


None
Mass
4-20 mA Output 1 Volume < 4 mA = xxxx > < 20mA = xxxx > < Time Const (Sec)
More > *** Energy xxxx xxxx xxxx
***
see Temp 1,2
below
Press
Density

< Measure >


None
4-20 mA Output 2 Mass
* see below
More > Volume < 4 mA = xxxx > < 20mA = xxxx > < Time Const (Sec)
***
see *** Energy
below xxxx xxxx xxxx
Temp 1,2
Press
Density

4-20 mA Output 3 < Measure >


* see below None
More >
Mass
Volume < 4 mA = xxxx > < 20mA = xxxx > < Time Const (Sec)
***
see *** Energy xxxx xxxx xxxx
below
Temp 1,2
Press
Density

Scaled Frequency < Max. Frequency > < Measure >


More > xxxxx None
***** Mass
Volume
*** Energy
Modbus Units Temp 1,2
** see below (Internal/Display) Press
Density

Modbus Order
** see below 0-1:2-3 < 0 Hz = (units) > < Max Hz = (units) > < Time Const (Sec)
3-2:1-0 xxxx xxxx xxxx
2-3:0-1
1-0:3-2

Comm Protocol
Modbus RTU
** see below (None1, None2,
Odd, Even)

Baud Rate
** see below 19200

Address
** see below 1 * - Physical Layer not available on Two
Wire Mass – Accessible via HART
** - Modbus not available on Two
Wire Mass
BACnet Device ID *** - Energy available on EMS meters only
****see below xxx **** - BACnet meters only
***** - 10000 Hz maximum

BACnet MaxMaster
****see below xxx

3-6 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Example for Setting an Output


The following shows how to set Output 1 to measure mass flow with 4 mA = 0 lb/hr and 20 mA = 100 lb/hr with
a time constant of 5 seconds. (All outputs are disabled while using the Setup Menus.)

First, set the desired units of measurement:


1. Use  keys to move to the Units Menu (see page 3-12).
2. Press  key until Mass Flow Unit appears. Press ENTER.
3. Press  key until lb appears in the numerator. Press  key to move the underline cursor to the
denominator. Press the  key until hr appears in the denominator. Press ENTER to select.
4. Press  key until Units Menu appears.

Second, set the analog output:


1. Use  keys to move to the Output Menu.
2. Press the  key until 4-20 mA Output 1 appears.
3. Press  key to access Measure selections. Press ENTER and press the  key to select Mass. Press ENTER.
4. Press  key to set the 4 mA point in the units you have selected for mass of lb/hr. Press ENTER and
use  keys to set 0 or 0.0. Press ENTER.
5. Press  key to set the 20 mA point. Press ENTER and use  keys to set 100 or 100.0. Press ENTER.
6. Press  key to select the Time Constant. Press ENTER and use  keys to select 5. Press ENTER.
7. Press the EXIT key and answer YES to permanently save your changes.

Example for Calculating Output Current


Assume Output 1 is scaled as above for mass flow with 4 mA = 0 lb/hr and 20 mA = 100 lb/hr.
Output 1 current = ((Flow rate / 100) x 16 mA) + 4 mA

M-000-00010 3-7
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Display Menu
ENTER
Run Mode Password

ENTER

Display Use
Menu keys to access menus

Cycle Time (sec)


If Cycle Time is set to zero, manual advance is required
0

Number of Digits
Used to set the number of digits displayed after the decimal point
2

Display TC (sec)
TC = Display Time constant, used to smooth display
1

MF Vf Te Pr De T MF = Mass Flow
Y or N Vf = Volume Flow
Te = Temperature
Pr = Pressure
De = Density
T = Total For each parameter:
Select Yes to view parameter in Run Mode
Select No to hide parameter in Run Mode
A1 A2 A3 Fl Dt E A1 = Alarm 1 Status
Y or N A2 = Alarm 2 Status
A3 = Alarm 3 Status
Fl = Fluid
Dt = Date and Time
* E = Energy * Energy EMS Meters Only

Use the Display Menu to set the cycle time for automatic screen sequencing used in the Run
Mode, change the precision of displayed values, smooth the values or enable or disable each item
displayed in the Run Mode screens.

Example for Changing a Run Mode Display Item


The following shows how to remove the temperature screen from the Run Mode screens. Note: all outputs
are disabled while using the Setup Menus.
1. Use  keys to move to the Display Menu.
2. Press  key until Mf Vf Pr Te De T appears.
3. Press ENTER to select. Press  key until the cursor is positioned below Te.
4. Press  key until N appears. Press ENTER to select.
5. Press EXIT and then ENTER to save changes and return to the Run Mode.

3-8 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Alarms Menu
ENTER
Run Mode Password

ENTER

* see below
Alarms Use
Menu keys to access menus

<Measure>
None
Mass <Mode>
Volume
Relay Alarm 1 ** Energy
None <Measure> units
More > Temp 1,2 HIGH Alarm (>) xxxx
Press LOW Alarm (<)
Density

<Measure>
None
Relay Alarm 2 Mass <Mode>
More > Volume None <Measure> units
** Energy
HIGH Alarm (>) xxxx
Temp 1,2
Press
LOW Alarm (<)
Density

Relay Alarm 3 <Measure>


More > None <Mode>
Mass
Volume None <Measure> units
** Energy HIGH Alarm (>) xxxx
Alarm LOG Temp 1,2 LOW Alarm (<)
xx Files (ENTER) Press
Density

Clear Alarm LOG?


Alarm File
YES or NO

Time
Date

(Press EXIT to return


* Physical Layer does not exist on Two Wire
to Alarm LOG) ** Energy EMS Meters Only
Mass - Accessible via HART

Example for Setting an Alarm


The following shows how to set Relay Alarm 1 to activate if the mass flow rate is greater than 100 lb/hr. You can
check the alarm configuration in the Run Mode by pressing the  keys until Alarm [1] appears. The lower line
displays the mass flow rate at which the alarm activates. Note: all outputs are disabled while using the Setup Menus.

First, set the desired units of measurement:


1. Use  keys to move to the Units Menu (see to page 3-12).
2. Press  key until Mass Flow Unit appears. Press ENTER.
3. Press  key until lb appears in the numerator. Press  key to move the underline cursor to the
denominator. Press the  key until hr appears in the denominator. Press ENTER to select.
4. Press  key until Units Menu appears.

Second, set the alarm:


1. Use  keys to move to the Alarms Menu.
2. Press the  key until Relay Alarm 1 appears.
3. Press  key to access Measure selections. Press ENTER and use the  key to select Mass. Press ENTER.
4. Press  key to select the alarm Mode. Press ENTER and use  key to select HIGH Alarm. Press ENTER.
5. Press  key to select the value that must be exceeded before the alarm activates. Press ENTER and use
 keys to set 100 or 100.0. Press ENTER.
6. Press the EXIT key to save your changes. (Alarm changes are always permanently saved.)
(Up to three relay alarm outputs are available depending on meter configuration.)

M-000-00010 3-9
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Totalizer #1 Menu
ENTER
Run Mode Password

ENTER

Totalizer Use
Menu keys to access menus

Totaling
Inactive
Mass
Example:
Volume
Energy Maximum flow rate = 600 gallons per minute
(600 gallons per minute = 10 gallons per second)
(unit) / Pulse
xxxx If unit per pulse is set to 600 gallons per pulse,
the totalizer will pulse once every minute.
Reset Total ? If unit per pulse is set to 10 gallons per pulse,
YES or NO
the totalizer will pulse once every second.

Use the Totalizer Menu to configure and monitor the totalizer. The totalizer maximum count is
999,999,999 at which point it will roll over to 0. The totalizer output is a 50 millisecond (.05 second)
positive pulse (relay closed for 50 milliseconds). The totalizer cannot operate faster than one pulse
every 100 millisecond (.1 second). A good rule to follow is to set the unit per pulse value equal to the
maximum flow in the same units per second. This will limit the pulse to no faster than one pulse
every second.

Example for Setting the Totalizer


The following shows how to set the totalizer to track mass flow in kg/sec. (All outputs are disabled while using
the Setup Menus.)

First, set the desired units of measurement:


1. Use  keys to move to the Units Menu (see to page 3-12).
2. Press  key until Mass Flow Unit appears. Press ENTER.
3. Press  key until kg appears in the numerator. Press  key to move the underline cursor to the
denominator. Press the  key until sec appears in the denominator. Press ENTER to select.
4. Press  key until Units Menu appears.

Second, set the pulse output:


1. Use  keys to move to the Totalizer Menu.
2. Press the  key until Totaling appears.
3. Press ENTER and press the  key to select Mass. Press ENTER.
4. Press  key to set the pulse output in the units you have selected for mass flow of kg/sec. Press
ENTER and use  keys to set the pulse value equal to the maximum flow in the same units per
second. Press ENTER.
5. To reset the totalizer, press  key until Reset Total? appears. Press ENTER and the  key to reset the
totalizer if desired. Press ENTER.
6. Press the EXIT key and answer YES to permanently save your changes.

3-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Totalizer #2 Menu
ENTER
Run Mode Password

ENTER

Totalizer Use
Menu keys to access menus

Totaling
Inactive
Mass
Volume
Energy

Reset Total ?
YES or NO

Use the Totalizer #2 to Monitor Flow or Energy. The totalizer maximum count is 999,999,999 at
which point it will roll over to 0. Note that Totalizer #2 does not operate a relay, it is for monitoring
only.

M-000-00010 3-11
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Energy Menu – For EMS Energy Meters Only


ENTER
Run Mode Password

ENTER

Energy Use
Menu keys to access menus

Loc in Sent
Flow
Yes or No

Heating
System
Yes or No

% Returned
xxx

Configuration:

There are several possibilities regarding the measurement of water or steam energy given the location
of the meter and the use of a second RTD. The table below summarizes the possibilities:

Fluid Meter Location Second RTD Measurement


Water “Sent” Flow Line “Return Flow Line Change in Energy
Water “Return” Flow Line “Sent” Flow Line Change in Energy
Water “Sent” Flow Line None Outgoing Energy
Steam “Sent” Flow Line “Return” Flow Line Change in Energy
(condensate)
Steam “Sent” Flow Line None Outgoing Energy

As above, you must properly configure the meter in the Energy Menu.
1. Loc in Sent Flow? Select Yes or No based on where the meter is located. Refer to the above
table
2. Heating System? Select Yes for a hot water system used for heating. Select No for a chilled
water system used for cooling. Always select Yes for a steam system.
3. % Returned. Select a number between 0% and 100%. Estimate the amount of water that returns.
It is usually 100%, or can be less than 100% if historical data shows the amount of makeup water
used. If a second RTD is not used, set to 0%. When 0% is selected, the energy calculation
represents the outgoing energy only (no return energy is subtracted). NOTE: the meter ships
from the factory assuming 0% return and has a 1000 ohm resistor installed in the RTD #2
wiring location. This needs to be removed if the meter is to be used in a manner other than
with 0% return and with the customer supplied RTD in its place.

3-12 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Fluid Menu
ENTER
Use
Run Mode Password
keys to access menus

ENTER < Liquid


Water
Fluid Ammonia
Menu Chlorine

< Density > < AL > < BL


xxxx xxxx xxxx

Flowing Fluid < Mole Weight > < CRIT PRESS > < CRIT TEMP > < CRIT Z > < AL > < BL
Liquids > xxxx xxxx xxxx xxxx xxxx xxxx
Other Liquids >
Goyal-Dorais >
< Density @ 60F > < K0 > < K1 > < AL > < BL
API 2540 >
xxxx xxxx xxxx xxxx xxxx
Nat Gas AGA8 >
Real Gas >
Other Gas > < Specific Gravity > < MoleFract N2 > <MoleFract CO2> <Ref Temp(F)> < Ref Press(PSIA)
Liquified Gas > xxxx xxxx xxxx xxxx xxxx
Thermal Oil >
Ethylene Glycol >
< Real Gas
Steam T & P Comp Select “Steam T & P Comp” for VT and
Air VTP models. The VT model will display
STD Temp (F) Argon “Sat Steam T Comp” for the fluid type
xxxx Ammonia in the run mode screens.
CO
CO2 For a V model in any fluid, enter nominal operating temperature and
STD Press (PSIA) Helium pressure as simulated values in the diagnostics menu.
xxxx Hydrogen
Methane
Nitrogen
NORM Temp (C) Oxygen
xxxx Propane

< Specific Gravity > < Compress (Z) > < Viscosity
NORM Press (KPA) xxxx xxxx xxxx
xxxx

< Liquified Gas


Carbon Dioxide
Nitrogen
Hydrogen
Oxygen
Argon

< Thermal Oil


Paratherm
Therminol VP-1
Dowtherm A

< % Volume @ 68 Deg. F


xx

Use the Fluid Menu to configure the flow meter for use with common gases, liquids and steam. Your
flow meter is pre-programmed at the factory for your application’s process fluid.

Reference Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition, 1996),
page 2-75 for definition and use of the Goyal-Doraiswamy equation and page 2-76 for the
definition and use of the API 2540 equation. Also, see Appendix C for Fluid Calculation
equations.

The units of measurement used in the Fluid Menu are preset and are as follows:
Mole Weight = lbm/(lbm·mol), CRIT PRESS = psia, CRIT TEMP = °R, Density = lbm/ft3 and
Viscosity = cP (centipoise).

M-000-00010 3-13
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Units Menu
ENTER
Run Mode Password

ENTER

Units Use
Menu keys to access menus

Mass Flow Unit


Ston = Short Ton = 2000 lb
lb, Ston, Lton, gram Lton = Long Ton = 2240 lb
/ sec, min, hr, day
kg, Mton, scf, nm3 Mton = Metric Ton = 1000 kg

Volume Flow Unit MilG = Million gallons


ImpG = Imperial gallon = 1.20095 US gallons
gal, MilG, ImpG, bbl bbl = barrels = 42 US gallons
/ sec, min, hr, day
lit, MilL, m3, ft3 MilL = Million liters

Temperature Unit
Deg F, Deg C, Kelvin, Rankine

Energy Unit
BTU, MBTU, MMBT, MWHr sec, min, hr, day MBTU = Thousand BTU
/
kWHr, HPHr, MCal,MJ MMBTU = Million BTU

Density Unit
lbm/ft3, kg/m3, gm/cc, lbm/gal,
gm/mlit, kg/lit, gm/lit, lbm/in3

Pressure Unit
psi, inH2O, ftH2O, mmH2O, in HG, 4inH2O and 4mmH2O are based on
mmHG, ATM, Bar, mBar, gm/cm2, kg/cm2, water at 4 degrees C.
Pascal, KiloPa, MegaPa, Torr, 4inH20, 4mmH2O

Gauge Pressure Unit Transducer reads in Absolute – if Gauge is desired


Absolute / Gauge then atmospheric pressure at meter is needed

Atm. Pressure Menu is only activated if Gage Pressure is chosen


14.6959 Enter the value in PSIA

Use the Units Menu to configure the flow meter with the desired units of measurement. (These
are global settings and determine what appears on all screens.

3-14 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

Time & Date Menu


ENTER
Run Mode Password

ENTER

Time & Date Use


Menu keys to access menus

Set Time
xx:xx:xx

Set Date
xx/xx/xx

Use the Time and Date Menu to enter the correct time and date into the flow meter’s
memory. The parameters are used in the Run Mode and the alarm and system log
files.

Note: Time is displayed in AM/PM format, but military format is used to set the time.
For example, 1:00 PM is entered as 13:00:00 in the Set Time menu.

Example for Setting the Time


How to set the time to 12:00:00. You can check the time in the Run Mode by pressing the  keys
until the Time & Date screen appears. Note: all outputs are disabled while using the Setup Menus.

1. Use  keys to move to the Time and Date Menu.


2. Press  key until Set Time appears. Press ENTER.
3. Press  key until 1 appears. Press  key to move the underline cursor to the next digit.
Press the  key until 2 appears. Continue sequence until all desired parameters are en-
tered. Press ENTER to return to the Time and Date Menu.
4. Press EXIT to return to the Run Mode.

M-000-00010 3-15
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Diagnostics Menu
ENTER
Run Mode Password

ENTER

Diagnositcs Use
Menu keys to access menus

Simulate Vortex Sim Vor Freq


Frequency (Hz) xxx

* Simulate Temperature Sim Temp 1, 2 * The unit of measure of the displayed value is the
xxx same as the unit configured for the flow meter.

* Simulate Pressure Sim Pressure For a V model in any fluid, enter nominal operating temperature and
xxx pressure as simulated values in the diagnostics menu.
For a VT model in any fluid except steam, enter nominal operating
pressure as a simulated value in the diagnostics menu.
Highest Recorded Highest Velocity
Velocity xxx

Highest Recorded Highest Temp 1, 2


Temperature xxx

Highest Recorded Highest Pressure


Pressure xxx

Highest Electronics Highest Int Temp


Temperature xx

Lowest Electronics Lowest Int Temp


Temperature xx
Momentarily displayed
Use Left and Right
System LOG
SysLog File #xx System Log File arrows to access all
xx Files (ENTER)
system log files

Clear Sys LOG? Time


YES or NO Date

Press EXIT to return


to System LOG

Use the Diagnostics Menu to simulate operation and review the system files. The system log files
contain time/date stamped messages including: power on, power off, programming time outs,
parameter faults, incorrect password entry and other various information relative to system
operation and programming.
The simulated inputs are for testing the meter to verify that the programming is correct. They are
also used to enter nominal operating temperature and pressure for the V only model. Simulated
vortex frequency allows you to enter any value for the sensor input in Hz. The meter will
calculate a flow rate based on the corresponding value and update all analog outputs (the
totalizer display and output is not affected by a simulated frequency). The simulated pressure
and temperature settings work the same way. The meter will output these new values and will use
them to calculate a new density for mass flow measurement. Note: when your diagnostic work is
complete, make sure to return the values to zero to allow the electronics to use the actual
transducer values. For the V only model keep the temperature and pressure at nominal operating
conditions.
3-16 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 3 Operation

If the meter display indicates a temperature or pressure fault, a substitute value can be entered to
allow flow calculations to continue at a fixed value until the source of the fault is identified and
corrected. The units of measure of the displayed values are the same as the units configured for
the flow meter.

Calibration Menu
ENTER
Run Mode Password

ENTER

Calibration Use
Menu keys to access menus

Meter Size Inline style meter – meter size


or Pipe ID Insertion style meter – pipe internal diameter (inches)

Meter calibration constant


Base Kf Inline style meter – pulses / ft 3
xx.xxx Insertion style meter – pulses / ft

Base Reynolds Meter calibration constant


xxxx.xx Typical Surface Roughness (inches)
Type of Pipe Roughness
Steel, drawn, new 0.0024
Steel, welded, new 0.003
Head Diameter in. ID of insertion head Steel, sheet, smooth 0.0028
x.xxx
Steel, galvanized, new 0.006
Steel, lightly corroded 0.01
Steel, severely corroded 0.068
Abs Roughness in. Pipe roughness
Steel, light scaling 0.05
x.xxxx Default value = 0.003
Steel, heavy scaling 0.11
Cast Iron, new 0.025
Cast Iron, corroded 0.06
Vortex Coef Ck Adaptive filter setting
xx Cast Iron, with scaling 0.1
Drawn Tubing, plastic 0.00006

Low Flow Cutoff < Vol (xxx/xxx) > < Mass (xxx/xxx)
xx xxx xxx
Low Flow Cutoff Low Flow Cutoff
setting displayed setting displayed
Serial Number in volumetric flow in mass flow
xxxxxxxxx units (view only) units (view only)

The Calibration Menu contains the calibration coefficients for the flow meter. These values
should by changed only by properly trained personnel. The Vortex Coef Ck and Low Flow Cutoff
are set at the factory. Consult the factory for help with these settings if the meter is showing
erratic flow rate.

M-000-00010 3-17
Chapter 3 Operation Series M22/M23/M24 Instruction Manual

Password Menu
ENTER
Run Mode Password

ENTER

Password Use
Menu keys to access menus

Set Password
1234

Use the Password Menu to set or change the system password. The factory-set password
is 1234.

3-18 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Chapter 4 Serial Communications

HART Communications
The HART Communications Protocol (Highway Addressable Remote Trans-
ducer Protocol) is a bidirectional digital serial communications protocol. The
HART signal is based on the Bell 202 standard and is superimposed on 4-20 mA
Output 1. Peer-to-peer (analog / digital) and multi-drop (digital only) modes are
supported.

Wiring

The diagrams below detail the proper connections required for HART
communications:

Warning!
Place controls in manual
Loop Powered Meter Wiring
mode when making
configuration changes to
the vortex meter.

Vortex Meter

+
LOOP PULSE FREQ BACKLIGHT
OPTIONAL

DC
POWER OUT OUT POWER

R load,
Current Power
+ - + - + - + -

250 ohm
Meter Supply
minimum
_

Field Connection Remote Connection


for Communicator for Communicator

Figure 4-1.Loop Powered Meter Wiring (HART)

M-000-00010 4-1
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

DC Powered Meter Wiring

Vortex Meter

- RS485

ALARM ALARM ALARM

-
RS485
+

+
+
RS485 GND

-
DC

2
OPTION 2

mA 3
4-20
R load,
4

+
Current Power

+
3
250 ohm
Meter

-
Supply
2

minimum

mA 2
4-20
+
1

+
PULSE

-
OUT
5

mA 1
DC PWR 4-20
+
OPTION 1
4

+
3

FREQ

-
OUT

-
2

+
1

Field Connection Remote Connection


for Communicator for Communicator

Figure 4-2.DC Powered Meter Wiring (HART)

AC Powered Meter Wiring

Vortex Meter

- RS485
PULSE ALARM ALARM ALARM

RS485
3

+
+

RS485 GND
5

-
2

-
OPTION 2

mA 3
4-20
4

R load,
+

Current
3

250 ohm
-
2

Meter
1

mA 2
4-20

minimum
1

+
-
OUT
5

mA 1
4-20
+
OPTION 1
4

+
FREQ
3

-
OUT

VDC
OUT
2

24
+

+
1

NEUT / L2
PWR
AC

IN

HOT / L1

Field Connection Remote Connection


for Communicator for Communicator

Figure 4-3.AC Powered Meter Wiring (HART)


4-2 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

HART Commands with the DD Menu

Online Menu

1 Process Variables 1 Mass Flo


2 Vol
3 Temp
4 Temp 2
5 Delta Temp.
6 Energy flo
7 Pres
8 Dens
9 Total
Total 2

1 Enter Password To Enter Password Menu


2 Device Setup

2 Display Unit 1 Mass flo unit


2 Vol unit
3 Temp unit
4 Energy flo unit
5 Line pr es unit 1 Norm Temp
6 Dens unit 2 Norm Press
7 Totalizer units 3 Std Temp
8 Std & Norm Cond 4 Std Press

3 Analog Output To Analog Output Menu

4 External Loop 1 External Input Inactive


2 Ext. Loop Zero Value Temp1
3 Ext. Loop Full Scale Temp 2
4 Ext. Loop Zero Count Pressure
5 Ext. Loop Span Count
6 Set Ext. 4 mA
7 Set Ext. 20 mA

1 Disp Cycle
5 Meter Display
2 Disp Digits
3 Disp Damping
4 Init Displ.
5 Disp Show/Hide
1 Alarm 1 var
2 Alarm 1 typ
3 Alarm 1 set pt
1 Alarm Status
6 Alarm Setup
2 Alarm 1 Setup
3 Alarm 2 Setup 1 Alarm 2 var
4 Alarm 3 Setup 2 Alarm 2 typ
5 Records in Log 3 Alarm 2 set pt
6 Read Alar m Log
7 Alarm Log Clear
1 Alarm 3 var
2 Alarm 3 typ
3 Alarm 3 set pt
7 Totalizer 1 Total
2 Totalize
3 Amount/Pulse * If insertion vor tex
4 Total 2 ** If inline vortex
5 Totalize 2
6 Clear Totalizer 1 Date
2 Change Password
3 Line size
8 Fluid Menu To Fluid Menu 4 Change Line size*
4 Meter size**
1 Meter Location 5 Meter size*
9 Energy Setup
2 Heating or Cooling 5 Change Meter size**
3 % Return 6 Dev id
7 Tag
8 Descr iptor
Device Menu
9 Message
To Sensor Cal Menu Final asmbly num
Sensor Cal
Poll addr
To Review Menu Num req preams
Review
Config Code
Compile Date
3 Maintenance To Maintenance Menu Compile Time
Signal Board Ver sion
4 Diagnostics To Diagnostics Menu Har dware rev
Software rev
Master reset

M-000-00010 4-3
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

HART Commands with the DD Menu Continued

Analog Output Menu

1 Fix Analog Output


2 Trim Analog Output
From Online Menu
3 Configure AO1 1 PV is
4 PV is 2 PV Loop cur rent
5 PV AO 1 Out 3 PV
6 PV % range 4 PV Unit
7 Configure AO2 5 PV % range
8 SV is 6 Apply values
9 SV AO 2 Out 7 PV LRV
SV % range 8 PV URV
Configure AO3 9 PV AO 1 Lo end pt
TV is PV AO1 Hi end pt
TV AO3 Out PV AO1 Added damp
TV % range
Configure AO4
QV is
QV AO4 Out 1 SV is
QV % range 2 SV AO 2 Out
3 SV
4 SV Unit
5 SV % range
6 Apply values
7 SV LRV
8 SV URV
9 SV AO 2 Lo end pt
SV AO2 Hi end pt
SV AO2 Added damp

1 TV is
2 TV AO3 Out
3 TV
4 TV Unit
5 TV % range
6 Apply values
7 TV LRV
8 TV URV
9 TV AO3 Lo end pt
TV AO3 Hi end pt
TV AO3 Added damp

1 Q V is
2 Q V AO4 Out
3 4V
4 4V Unit
5 Q V % r ange
6 Q V LRV
7 Q V URV
8.Q V AO4 Lo end pt
9 Q V AO4 Hi end pt
QV AO4 Added damp

4-4 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

HART Commands with the DD Menu Continued

Fluid Menu

From Online Menu 1 Fluid Liquid Water


2 Fluid Type Other Liquid Ammonia
Goyal-Dor ais Chlor ine
API-2540
Nat Gas AGA8
Real Gas Other Liquid Density
Other Gas Viscosity Coef AL
Liquified Gas Viscosity Coef BL
Thermal Oil
Ethylene Glycol
Mol Weight
Crit Press
Crit Temp
Crit Z
AL
BL

Density @ 60F
API K0
API K1
Maintenance Menu API AL
API BL

From Online Menu 1 Analog output 1 Fix analog output AGA Ref Temp
2 External Loop 2 Trim analog output AGA Ref Press
3 Vor tex Cal Specific Gravity
Mole Fract N2
1 External Input Mole Fract CO2
2 Ext. Loop Zero Value
3 Ext. Loop Full Scale Value
4 Ext. Loop Zero Count Steam
5 Ext. Loop Span Count Air
6 Set Ext. 4 mA Argon
7 Set Ext. 20 mA Ammonia
CO
CO2
Helium
1 Ck Value Hydrogen
2 Low flow cutoff Methane
3 Auto Set Ck Nitrogen
4 Auto Set LFC Oxygen
Pr opane

Specific gravity
Compr ess
Viscosity

Car bon Dioxide


Nitrogen
Hydrogen
Oxygen
Ar gon

M-000-00010 4-5
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

HART Commands with the DD Menu Continued

Diagnostics Menu

From Online Menu 1 Vel 1 Vtx Freq


2 Pres 2 Sim Vtx Freq
3 Temp 3 Vtx AtoD
4 Temp 2 4 Filter Set
5 Internal Temp 5 Change Filter Set
6 Vor tex Diag. 6 High Pass Filter
7 Press Diag. 7 G ain Set
8 Temp Diag. 8 Change Gain Set
9 Int Temp Diag 9 Re
Log Vel
Status Max Vel
AD1
AD2
AD3
AD4
Ck
Level

1 Press
2 Sim Press
3 Excite
4 Excite AtoD
5 Sense
6 Sense AtoD
7 Max Press

Review Menu 1 Temp


2 Sim Temp
3 RTD1
4 RTD1 AtoD
From Online Menu 1 Model 5 Max temp
2 Distributor 6 Temp 2
3 W rite protect 7 Sim Temp 2
4 Manufacturer 8 RTD2
5 Dev id 9 RTD2 AtoD
6 Tag Max temp 2
7 Descriptor
8 Message SPI not communicating
9 Date Freq. Input Overrange
1 Internal Temp (F)
Final asmbly num FRAM CRC error
2 Inter nal Temp (C)
Universal rev Signal Board Power ...
3 Int. Max temp
Fld dev r ev RTD1 Fault
4 Int. Min temp
Software rev RTD2 Fault
Poll addr Press. Transducer Fault
Num req preams
1 Records in Log
2 Read System Log
Totalizer Relay Overrange
3 System Log Clear

1 Status group 0 Alar m 1 Set


2 Status group 1 Alarm 2 Set
3 Status group 2 Alar m 3 Set

4-6 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

HART Commands with the DD Menu Continued

Sensor Cal Menu


* If insertion vortex

1 K Factor
1 Calibr ation Review 2 Base Reynolds *
From Online Menu 3 SR Coefficient *
2 Vor tex Sensor 1 Unit 1 Vtx Freq
3 Vor tex Cal 2 USL 2 Sim Vtx Freq 4 Ck Value
4 Press Sensor 3 LSL 3 Vtx AtoD 5 Low flow cutoff
5 Press Cal 4 Min Span 4 Filter Set 6 RTD1 R0
6 Temp Sensor 5 Snsr s/n 5 Change Filter Set 7 RTD1 alpha
7 Temp 1 & 2 Cal 6 Sim Vtx freq 6 High Pass Filter 8 RTD1 beta
8 Temp 2 Sensor 7 Max Vel 7 G ain Set 9 RTD2 R0
9 Cal. Cor rection 8 Vor tex Diag 8 Change Gain Set RTD2 alpha
9 Re RTD2 beta
Vel Pcal B00, Pcal B01
Max Vel Pcal B02, Pcal B10
1 K Factor AD1 Pcal B11, Pcal B12
2 Base Reynolds * AD2 Pcal B20, Pcal B21
3 SR Coefficient * AD3 Pcal B22
1 Flow 1 4 Ck Value AD4 Ref. Resistance
2 Deviation 1 5 Low flow cutoff Ck Internal Temp. Cal
3 Flow 2 6 Auto Set Ck Level Cal current
4 Deviation 2 7 Auto Set LFC Flow 1
5 Flow 3 Deviation 1
6 Deviation 3 Flow 2
7 Flow 4 1 Unit Deviation 2
8 Deviation 4 2 USL 1 Press Flow 3
9 Flow 5 3 LSL 2 Sim Press Deviation 3
Deviation 5 4 Min span 3 Excite Flow 4
Flow 6 5 Snsr s/n 4 Excite AtoD Deviation 4
Deviation 6 6 Sim Press 5 Sense Flow 5
Flow 7 7 Max Press 6 Sense AtoD Deviation 5
Deviation 7 8 Press Diag 7 Max Press Flow 6
Flow 8 Deviation 6
Deviation 8 Flow 7
Flow 9 1 PCal B00 Deviation 7
Deviation 9 2 PCal B01 Flow 8
Flow 10 3 PCal B02 Deviation 8
Deviation 10 4 PCal B10 Flow 9
5 PCal B11 Deviation 9
6 PCal B12 Flow 10
7 PCal B20 Deviation 10
8 PCal B21 Ext. Loop Zero Count
9 PCal B22 Ext. Loop Span Count
Ref. Resistance
Internal Temp. Cal
Cal Current

1 Temp
1 Unit 2 Sim Temp
2 USL 3 RTD1
3 LSL 4 RTD1 AtoD
4 Min span 5 Max Temp
5 Snsr s/n 6 Temp 2
6 Sim Temp 7 Sim Temp 2
7 Max Temp 8 RTD2
8 Temp Diag 9 RTD2 AtoD
Max temp 2

1 RTD1 Ro
2 RTD1 alpha
3 RTD1 beta
4 RTD2 Ro
5 RTD2 alpha
6 RTD2 beta

1 Temp
1 Unit 2 Sim Temp
2 USL 3 RTD1
3 LSL 4 RTD1 AtoD
4 Min span 5 Max Temp
5 Sim Temp 2 6 Temp 2
6 Max Temp 2 7 Sim Temp 2
7 Temp Diag 8 RTD2
9 RTD2 AtoD
Max temp 2

M-000-00010 4-7
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

HART Commands with Generic DD Menu


Online Menu

1 Device setup 1 Process Variables 1 PV From Device setup / Review 1 Model


2 PV 2 PV % Rnge 1 4 mA 2 Distributor
3 PV Loop current 3 PV Loop current 2 20 mA 3 PV Unit
3 Other 4 PV USL
4 End 1 4 mA 5 PV LSL
1 Test Device 2 20 mA 6 PV Min span
2 Diag/Service 2 Loop Test 3 Exit 7 PV Damp
3 Calibration 1 Apply Values 8 PV % rnge
4 D/A Trim 2 Enter Values 9 PV Xfer fnctn
1 PV URV PV Rnge unit
2 PV LRV PV URV
3 PV USL PV LRV
1 Tag 1 PV URV 4 PV LSL PV Loop current
2 PV unit 2 PV LRV
3 Range Values PV Alrm typ
3 PV Unit
3 Basic Setup 4 Device Information PV Snsr s/n
4 PV LSL
5 PV Xfer fnctn Write protect
5 PV USL
6 PV Damp Manufacturer
Dev id
Tag
1 Distributor Descriptor
2 Model Message
3 Dev id Date
4 Tag Universal rev
5 Date Fld dev rev
6 Write Protect Software rev
7 Descriptor Burst mode
8 Message Burst option
9 PV snsr s/n Poll addr
Final assy # 1 Universal Rev Num req preams
Revision #'s 2 Fld dev Rev
3 Software Rev

1 Sensors 1 PV
2 PV Sensor Unit 1 PV LSL
3 Sensor information 2 PV USL
3 PV Min span

1 PV Damp
2 Signal Condition 2 PV URV 1 PV URV
3 PV LRV 2 PV LRV
4 PV Xfer fnctn
5 PV % rnge

1 Loop current 1 4 mA
4 Detailed Setup 1 Analog Output 2 AO alarm typ 2 20 mA
3 Output Condition
5 Review 2 HART Output 3 Loop test 3 Other
4 D/A trim 4 End

4 PV LRV 1 PV URV
5 PV URV 2 PV LRV 1 Distributor
2 Model 1 Poll addr
3 Dev id 2 Num req. preams
4 Tag 3 Burst mode
5 Date 4 Burst option
4 Device Information 6 Write Protect
7 Descriptor
8 Message
9 PV snsr s/n
Final assy # 1 Universal Rev
Revision #'s 2 Fld dev Rev
3 Software Rev

4-8 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Modbus Communications
Applicable Flow Meter Models
Warning!
Place controls in manual VorTek Pro-V® Mass Flow Meters, Models M22, M23, and M24 with Modbus
mode when making communication protocol and firmware version 4.00.58 and above.
configuration changes to
the vortex meter.
Overview

This document describes the preliminary implementation of the Modbus


communication protocol for use in monitoring common process variables in the
VorTek Pro-V® Vortex flow meter. The physical layer utilizes the half-duplex
RS-485 port, and the Modbus protocol.

Reference Documents

The following documents are available online from www.modbus.org.

Modbus Application Protocol Specification V1.1


Modbus Over Serial Line Specification & Implementation Guide V1.0
Modicon Modbus Protocol Reference Guide PI–MBUS–300 Rev. J

Wiring
An RS485 daisy chained network configuration as depicted below is
recommended. Do not use a star, ring, or cluster arrangement.

RS-485 Master

RS-485 -

RS-485 +

RS-485 GND
RS-485 GND

RS-485 GND

RS-485 GND
RS-485 +

RS-485 +

RS-485 +
RS-485 -

RS-485 -

RS-485 -

Other Device 1 Vortex Meter Other Device 2, etc.

Figure 4-4.RS-485 Wiring (MODBUS)

Pin Labeling (among devices)


“RS-485 –” = “A” = “TxD-/RxD-” = “Inverting pin”
“RS-485 +” = “B” = “TxD+/RxD+” = “Non-Inverting pin”
“RS-485 GND” = “GND” = “G” = “SC” = “Reference”
M-000-00010 4-9
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

Menu Items

The following menu items are in the Output Menu and allow selection and
control of the Modbus communication protocol.

Address

When the Modbus protocol is selected, the Modbus address is equal to the
user programmable device address if it is in the range 1…247, in
accordance with the Modbus specification. If the device address is zero or
is greater than 247, then the Modbus address is internally set to 1.

Comm Protocol
The Comm Protocol menu allows selection of “Modbus RTU Even,” “Modbus
RTU Odd,” or “Modbus RTU None2,” or “Modbus RTU None1,” (non-standard
Modbus) with Even, Odd and None referring to the parity selection. When even
or odd parity is selected, the unit is configured for 8 data bits, 1 parity bit and 1
stop bit; with no parity, the number of stop bits is 1 (non-standard) or 2. When
changing the protocol, the change is made as soon as the Enter key is pressed.

Modbus Units

The Modbus Units menu is to control what units, where applicable, the meter’s
variables will be displayed in. Internal – these are the base units of the meter, °F,
3 3
psia, lbm/sec , ft /sec, Btu/sec , lbm/ft Display – variables are displayed in user
selected display unit.

Modbus Order

The byte order within registers and the order in which multiple registers
containing floating point or long integer data are transmitted may be changed
with this menu item. According to the Modbus specification, the most
significant byte of a register is transmitted first, followed by the least significant
byte. The Modbus specification does not prescribe the order in which registers
are transmitted when multiple registers represent values longer than 16 bits.
Using this menu item, the order in which registers representing floating point or
long integer data and/or the byte order within the registers may be reversed for
compatibility with some PLCs and PC software.

4-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

The following four selections are available in this menu; when selecting an
item, the protocol is changed immediately without having to press the
Enter key.

0-1:2-3 Most significant register first, most significant byte first (default)
2-3:0-1 Least significant register first, most significant byte first
1-0:3-2 Most significant register first, least significant byte first
3-2:1-0 Least significant register first, least significant byte first
Table 4-1.Byte Order

Note that all of the registers are affected by the byte order, including strings and
registers representing 16-bit integers; the register order only affects the order of
those registers representing 32-bit floating point and long integer data, but does
not affect single 16-bit integers or strings.

Modbus Protocol

The Modbus RTU protocol is supported in this implementation. Supported baud


rates are 1200, 2400, 4800, 9600, 19200, 38400, 57600, and 115200. The default
baud rate is 19200 baud. Depending upon the Modbus protocol selected, data are
transmitted in 8-bit data frames with even or odd parity and 1 stop bit, or no
parity and 2 or 1 (non-standard) stop bits.

The current Modbus protocol specification does not define register usage, but
there is an informal register numbering convention derived from the original
(now obsolete) Modicon Modbus protocol specification, and used by many
vendors of Modbus capable products.

Registers Usage Valid Function Codes


00001–09999 Read/write bits ("coils") 01 (read coils) 05 (write single coil)
15 (write multiple coils)

10001–19999 Read-only bits ("discrete inputs") 02 (read discrete inputs)


30001–39999 Read-only 16 bit registers ("input 03 (read holding registers) 04 (read
registers"), IEEE 754 floating point input registers)
register pairs, arbitrary length strings
encoded as two ASCII characters per 16-
bit register
40001–49999 Read/write 16-bit registers ("holding 03 (read holding registers) 06 (write
registers"), IEEE 754 floating point single register) 16 (write multiple
register pairs, arbitrary length strings registers)
encoded as two ASCII characters per 16-
bit register

M-000-00010 4-11
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

Each range of register numbers maps to a unique range of addresses that are
determined by the function code and the register number. The address is equal to
the least significant four digits of the register number minus one, as shown in the
following table.

Registers Function Codes Data Type and Address Range


00001-09999 01, 05, 15 Read/write bits 0000-9998
10001-19999 02 Read-only bits 0000-9999
30001-39999 03, 04 Read-only 16-bit registers 0000-9998
40001-49999 03, 06, 16 Read/write 16-bit registers 0000-9998

Register Definitions

The meter serial number and those variables that are commonly monitored (mass,
volume and energy flow rates, total, pressure, temperature, density, viscosity,
Reynolds number, and diagnostic variables such as frequency, velocity, gain,
amplitude and filter setting) are accessible via the Modbus protocol. Long integer
and floating point numbers are accessed as pairs of 16-bit registers in the register
order selected in the Modbus Order menu. Floating point numbers are formatted
as single precision IEEE 754 floating point values.

The flow rate, temperature, pressure, and density variables may be accessed as
either the flow meter internal base units or in the user-programmed display
units, which is determined by the programming Output Menu’s “Modbus
Units” item. The display units strings may be examined by accessing their
associated registers. Each of these units string registers contain 2 characters of
the string, and the strings may be 2 to 12 characters in length with unused
characters set to zero. Note that the byte order affects the order in which the
strings are transmitted. If the Modbus Order menu (see page 2) is set to 0-1:2-3
or 2-3:0-1, then the characters are transmitted in the correct order; if set to 1-
0:3-2 or 3-2:1-0, then each pair of characters will be transmitted in reverse
order.

4-12 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Registers Variable Data type Units Function Addresses


code
65100-65101 Serial number unsigned long — 03, 04
30525-30526 Totalizer unsigned long display units* 03, 04 524-525
32037-32042 Totalizer units string — 03, 04 2036-2041
30009-30010 Mass flow float display units* 03, 04 8-9
30007-30008 Volume flow float display units* 03, 04 6-7
30005-30006 Pressure float display units* 03, 04 4-5
30001-30002 Temperature float display units* 03, 04 0-1
30029-30030 Velocity float ft/sec 03, 04 28-29
30015-30016 Density float display units* 03, 04 14-15
30013-30014 Viscosity float cP 03, 04 12-13
30031-30032 Reynolds number float — 03, 04 30-31
30025-30026 Vortex frequency float Hz 03, 04 24-25
34532 Gain char — 03, 04 4531
30085-30086 Vortex amplitude float Vrms 03, 04 84-85
30027-30028 Filter setting float Hz 03, 04 26-27

The following registers are available with the energy meter firmware:

Registers Variable Data type Units Function Addresses


code
30527-30528 Totalizer #2 unsigned long display units* 03, 04 526-527
32043-32048 Totalizer #2 units string — 03, 04 2042-2047
30003-30004 Temperature #2 float display units* 03, 04 2-3
30011-30012 Energy flow float display units* 03, 04 10-11

The following registers contain the display units strings:

Registers Variable Data type Units Function code Addresses


32007-32012 Volume flow units string — 03, 04 2006-2011
32001-32006 Mass flow units string — 03, 04 2000-2005
32025-32030 Temperature units string — 03, 04 2024-2029
32019-32024 Pressure units string — 03, 04 2018-2023
32031-32036 Density units string — 03, 04 2030-2035
32013-32017 Energy flow units string — 03, 04 2012-2017
Table 4-2.Register Definitions

Function codes 03 (read holding registers) and 04 (read input registers) are the
only codes supported for reading these registers, and function codes for writing
holding registers are not implemented. We recommend that the floating point and
long integer registers be read in a single operation with the number of registers
being a multiple of two. If these data are read in two separate operations, each
reading a single 16-bit register, then the value will likely be invalid.

M-000-00010 4-13
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

The floating point registers with values in display units are scaled to the same
units as are displayed, but are instantaneous values that are not smoothed. If
display smoothing is enabled (non-zero value entered in the Display TC item
in the Display Menu), then the register values will not agree exactly with the
displayed values.

Exception Status Definitions

The Read Exception Status command (function code 07) returns the exception
status byte, which is defined as follows. This byte may be cleared by setting
“coil” register #00008 (function code 5, address 7, data = 0xff00).

Bit(s) Definition
0-1 Byte order (see Modbus Order on page 2)
0 = 3-2:1-0 1 = 2-3:0-1
2 = 1-0:3-2 3 = 0-1:2-3
2 Not used
3 Not used
4 Not used
5 Not used
6 Not used
7 Configuration changed

Discrete Input Definitions

The status of the three alarms may be monitored via the Modbus Read Discrete
Input command (function code 02). The value returned indicates the state of the
alarm, and will be 1 only if the alarm is enabled and active. A zero value is
transmitted for alarms that are either disabled or inactive,

Registers Variable Function Code Address


10001 Alarm #1 state 02 0
10002 Alarm #2 state 02 1
10003 Alarm #3 state 02 2

4-14 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Control Register Definitions

The only writeable registers in this implementation are the Reset Exception
Status, Reset Meter and Reset Totalizer functions, which are implemented as
”coils” which may be written with the Write Single Coil command (function
code 05) to address 7 through 9, respectively, (register #00008 through #00010).
The value sent with this command must be either 0x0000 or 0xff00, or the meter
will respond with an error message; the totalizer will be reset or exception status
cleared only with a value of 0xff00.

Error Responses

If an error is detected in the message received by the unit, the function code
in the response is the received function code with the most significant bit set,
and the data field will contain the exception code byte, as follows:

Exception
Code Description
01 Invalid function code — function code not supported by device
02 Invalid data address — address defined by the start address and number of registers
is out of range
03 Invalid data value — number of registers = 0 or >125 or incorrect data with the Write
Single Coil command

If the first byte of a message is not equal to the unit’s Modbus address, if the
unit detects a parity error in any character in the received message (with even or
odd parity enabled), or if the message CRC is incorrect, the unit will not
respond.

Command Message Format

The start address is equal to the desired first register number minus one. The
addresses derived from the start address and the number of registers must all be
mapped to valid defined registers, or an invalid data address exception will occur.

Normal Response Message Format

M-000-00010 4-15
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

Exception Response Message Format

Examples
Read the exception status byte from the device with address 1:
01 07 41 E2

01 Device address
07 Function code, 07 = read exception status
41 E2 CRC

A typical response from the device is as follows:


01 07 03 62 31

01 Device address
07 Function code
03 Exception status byte
62 31 CRC

Request the first 12 registers from device with address 1:

01 04 00 00 00 0C F0 0F

01 Device address
04 Function code, 04 = read input register
00 00 Starting address
00 0C Number of registers = 12
F0 0F CRC

A typical response from the device is as follows: *note these are the older
register definitions
01 04 18 00 00 03 E8 00 00 7A 02 6C 62 00 00 41 BA 87 F2 3E BF
FC 6F 42 12 EC 8B 4D D1

01 Device address
04 Function code
18 Number of data bytes = 24
00 00 03 E8 Serial number = 1000 (unsigned long)
00 00 7A 02 Totalizer = 31234 lbm (unsigned long)
6C 62 00 00 Totalizer units = “lb” (string, unused characters
are 0)
41 BA 87 F2 Mass flow rate = 23.3164 lbm/sec (float)
3
3E BF FC 6F Volume flow rate = 0.3750 ft /sec (float)
42 12 EC 8B Pressure = 36.731 psia (float)
4D D1 CRC

4-16 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

An attempt to read register(s) that don’t exist


01 04 00 00 00 50 F1 D2

01 Device address
04 Function code 4 = read input register
00 00 Starting address
00 50 Number of registers = 80
F0 36 CRC

results in an error response as follows:


01 84 02 C2 C1

01 Device address
84 Function code with most significant bit set indicates
error response
02 Exception code 2 = invalid data address
C2 C1 CRC

Request the state all three alarms:

01 02 00 00 00 03 38 0B

01 Device address
02 Function code 2 = read discrete inputs
00 00 Starting address
00 03 Number of inputs = 3
38 0B CRC

and the unit responds with:


01 02 01 02 20 49

01 Device address
02 Function code
01 Number of data bytes = 1
02 Alarm #2 on, alarms #1 and #3 off
20 49 CRC

To reset the totalizer:

01 05 00 00 FF 00 8C 3A

01 Device address
05 Function code 5 = write single coil
00 09 Coil address = 9
FF 00 Data to reset totalizer
5C 38 CRC

M-000-00010 4-17
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

The unit responds with an identical message to that transmitted, and the totalizer
is reset. If the “coil” is turned off as in the following message, the response is
also identical to the transmitted message, but the totalizer is not affected.

01 05 00 09 00 00 1D C8

01 Device address
05 Function code 5 = write single coil
00 09 Coil address = 9
00 00 Data to “turn off coil” does not reset totalizer
1D C8 CRC

4-18 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

BACnet MS/TP Communications

1. BACnet MS/TP Description


The BACnet Master-Slave/Token-Passing (MSTP) driver
implements a data link protocol that uses the services of the RS-
485 physical layer. The MS/TP bus is based on BACnet standard
protocol SSPC-135, Clause 9. BACnet MS/TP protocol is a peer-
to-peer, multiple master protocols based on token passing. Only
master devices can receive the token, and only the device holding
the token is allowed to originate a message on the bus. The token is
passed from master device to master device using a small message.
The token is passed in consecutive order starting with the lowest
address. Slave devices on the bus only communicate on the bus
when responding to a data request from a master device.

2. Baud Rates on the MS/TP Bus


An MS/TP bus can be configured to communicate at one of four
different baud rates. It is very important that all of the devices on
an MS/TP bus communicate at the same baud rate. The baud rate
setting determines the rate at which devices communicate data
over the bus. The baud rate settings available on Pro-V™ Vortex
Mass Flow Meters are 9600, 19200 and 38400.

2.1. Baud Rate and MAC address configuration


1. Power on the IUT
2. Press Enter to go to the configuration menu
3. Enter the factory password 16363 (Use Up and Down arrows to enter
the digits)
4. Navigate with the left arrow to Diagnostics menu
5. Press Enter and press right button immediately
6. Navigate to Config Code screen by continuous pressing down button
7. After reaching config code screen, press right to navigate to Comm.
Type screen
8. Change the comm. Type to “BACnet” and press Enter
Note: BACnet will enable Baud Rate and MAC address screens
9. Press Exit twice to go back to the Diagnostics menu
10. Navigate to Output Menu by using right or left arrow buttons
11. Press Down button and reach Baud Rate and MAC address screens

M-000-00010 4-19
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

12. Change the required settings and press Exit & Enter buttons to save
the configuration
13. Reboot the device by power off and on for the changes to take effect

Note:
a. IUT support 9600, 19200, 38400 baud rates
b. MAC address range is 0-127

3. Supported BACnet Objects


A BACnet object represents physical or virtual equipment
information, as a digital input or parameters. The Pro-V™ Vortex
Mass Flow Meters presents the following object types:

a. Device Object
b. Analog Input
c. Binary Input
d. Binary Value

Each object type defines a data structure composed by properties


that allow the access to the object information. The below table
shows the implemented properties for each Vortex Mass Flow
Meters object type.

Properties Object Types

Device Analog Input Binary Input Binary


Value
Object_Identifier    
Object_Name    
Object_Type    
System_Status 
Vendor_Name 
Vendor_Identifier 
Model_Name 
Firmware_Revision 
Application-Software-Version 
Protocol_Version 
Protocol_Revision 
Protocol_Services_Supported 
Protocol_Object_Types_Supported 

4-20 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Object_List 
Max_ADPU_Length_Accepted 
Segmentation_Supported 
ADPU_Timeout 
Number_Of_ADPU_Retries 
Max_Masters 
Max_Info_Frames 
Device_Address_Binding 
Database_Revision 
Status_Flags
Event_State   
Reliability
Out_Of_Service  (W)  (W)  (W)
Units 
Polarity  (W)
Priority_Array
Relinquish_Default
Status_Flag   
Present_Value  (W)  (W)  (W)
Inactive_Text
Active_Text
(W) – Writable Property.

3.1. Device Object:


The Device object default property values are as follows –

Property Name Default Values


object-identifier 7
object-name Device,1
object-type Device
system-status operational
vendor-name VorTek Instruments
vendor-identifier 558
model-name Multivariable Flowmete
firmware-revision N/A
application-software- 1.07
version
protocol-version 1
protocol-revision 4
protocol-services- {F,F,F,F,F,F,F,F,F,F,F,F,T,F,T,T,T,T,F,F,F,F,F,F,F,F,F,F,F,F,T,T,F,F,F,F,F}
supported
protocol-object-types- {T,F,F,T,F,T,F,F,T,F,F,F,F,F,F,F,F,F,F,F,F,F,F,F,F}
M-000-00010 4-21
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

supported
object-list {(analog-input,1),(analog-input,2),(analog-input,3),(analog-input,4),
(analog-input,5), (analog-input,6),(analog-input,7),(analog-input,8)
(analog-input,9),(analog-input,10), (analog-input,11), (analog-
input,12), (analog-input,13),(analog-input,14), (analog-
input,15),(analog-input,16),(analog-input,17), (analog-
input,18),(analog-input,19),(binary-input,1),(binary-input,2),(binary-
input,3),(binary-input,4), (binary-value,1), (device,7) }
max-apdu-length- 300
accepted
segmentation-supported no-segmentation
apdu-timeout 3000
number-of-APDU-retries 1
max-master 127
max-info-frames 1
device-address-binding ()
database-revision 0

Note - Device Communication Control: Password – “vortek”

3.2. Analog Input Object:


Vortex Mass Flow Meters Analog Input type objects are described
in the below Table –

Object Instance Object Name Unit Description


1 Volume Flow cubic-feet-per-second, This AI object is used to
cubic-feet-per-minute, measure volume flow.
us-gallons-per-minute,
imperial-gallons-per-
minute,
liters-per-minute,
liters-per-second,
liters-per-hour,
cubic-meters-per-second,
cubic-meters-per-minute,
cubic-meters-per-hour
2 Mass Flow pounds-mass-per-second, This AI object is used to
grams-per-second, measure mass flow.
kilograms-per-second ,
kilograms-per-minute ,
kilograms-per-hour,
pounds-mass-per-minute ,
pounds-mass-per-hour,

4-22 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

tons-per-hour,
grams-per-second ,
grams-per-minute

3 Temperature 1 degrees-Celsius, This AI object measures


degrees-Kelvin, Temperature in one of
degrees-Fahrenheit the given Unit.
4 Temperature 2 degrees-Celsius, This AI object measures
degrees-Kelvin, Temperature in one of
degrees-Fahrenheit the given Unit.
5 Pressure pounds-force-per-square- TBD
inch,
inches-of-water,
inches-of-mercury,
millimeters-of-mercury,
bars,
millibars,
pascals,
kilopascals
6 Density kilograms-per-cubic-meter TBD
7 Energy Flow Kilowatts, TBD
Horsepower,
btus-per-hour,,
kilo-btus-per-hour,
megawatts
8 Totalizer 1 If Totalizer selection for An electronic counter
& Mass measure – which records the total
Totalizer 2 pounds-mass-per-second, accumulated flow over
grams-per-second, a certain range of time.
kilograms-per-second ,
kilograms-per-minute ,
kilograms-per-hour,
pounds-mass-per-minute ,
pounds-mass-per-hour,
tons-per-hour,
grams-per-second ,
grams-per-minute
If Totalizer selection for
Volume measure –
cubic-feet-per-second,
cubic-feet-per-minute,
us-gallons-per-minute,
imperial-gallons-per-
minute,

M-000-00010 4-23
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

liters-per-minute,
liters-per-second,
liters-per-hour,
cubic-meters-per-second,
cubic-meters-per-minute,
cubic-meters-per-hour
If Totalizer selection for
Energy measure –
Kilowatts,
Horsepower,
btus-per-hour,,
kilo-btus-per-hour,
megawatts
10 StatusRegister NO UNITS TBD
11 Channel 1 milliamperes TBD
(4-20mA)
12 Channel 2 milliamperes TBD
(4-20mA)
13 Channel 3 milliamperes TBD
(4-20mA)
14 Scaled Freq hertz TBD
15 Flow Velocity feet-per-second TBD
16 Viscosity centipoises TBD
17 Frequency hertz TBD
18 VorTex Amp millivolts TBD
19 FilterSetting hertz TBD

4-24 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

3.3. Binary Input Object:


Vortex Mass Flow Meters Binary Input type objects are described
in the below Table.

Object Object Description


Instance Name
1 Alarm1 The status of the three alarms may be monitored
2 Alarm2 via the Modbus command. The value returned
3 Alarm3 indicates the state of the alarm, and will be 1 only
if the alarm is enabled and active. A zero value is
transmitted for alarms that are either disabled or
inactive
4 External TBD

Note - Binary Input 4, Present value always read zero, because no


information available from client, so the polarity property doesn’t
impact on Present value property when the Out of service
property is false.

3.4. Binary Value Object:


Vortex Mass Flow Meters Binary Value type objects are described
in the below Table.

Object Instance Object Name Description


1 Reset Reset’s Totalizer

M-000-00010 4-25
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

4. ANNEX - BACnet PROTOCOL IMPLEMENTATION CONFORMANCE


STATEMENT

Date: 19-April-2012

Vendor Name: VorTek Instruments

Product Name: Pro-V M22/M23/M24 multivariable flow-meter

Product Model Number: M22/M23/M24 VT/VTP

Applications Software Version: 1.07

Firmware Revision: N/A

BACnet Protocol Revision: 4

Product Description: VorTek multivariable flow-meter

BACnet Standardized Device Profile (Annex L):


 BACnet Operator Workstation (B-OWS)
 BACnet Advanced Operator Workstation (B-AWS)
 BACnet Operator Display (B-OD)
 BACnet Building Controller (B-BC)
 BACnet Advanced Application Controller (B-AAC)
 BACnet Application Specific Controller (B-ASC)
 BACnet Smart Sensor (B-SS)
 BACnet Smart Actuator (B-SA)

List all BACnet Interoperability Building Blocks Supported


(Annex K):

BIBBs
DS-RP-B
DS-WP-B
DM-DDB-B
DM-DOB-B
DM-DCC-B
DS-RPM-B
DS-WPM-B

4-26 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Services Supported
Read Property Execute
Write Property Execute
Read Property Multiple Execute
Write Property Multiple Execute
Who-Is Execute
I-Am Initiate
Who-Has Execute
I-Have Initiate
Device Communication Control Execute

Segmentation Capability:
 Able to transmit segmented messages Window Size
Able to receive segmented messages Window Size

M-000-00010 4-27
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

Standard Object Types Supported:

Standard Object Types Supported


Additional
Dynamically Dynamically Range
Object Type Writable
Creatable Delete-able Restrictions
Properties
Analog No No None None
Input (AI)
Binary Input No No None None
(BV)
Binary No No None None
Value
Device No No None None

Standard Object Types


Supported Writable Properties
Object Type Properties
Analog Input (AI) Present Value Out-Of-Service
Binary Input (BV) Present Value Out-Of-Service Polarity
Binary Value Present Value Out-Of-Service
Device

4-28 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

Object List:

Properties of Analog Input/Value Objects Type


Present Status Event Out of
ID Name Units
Value Flags State Service
AI1 Volume Flow ? F,F,F,F Normal False ?
AI2 Mass Flow ? F,F,F,F Normal False ?
AI3 Temperature 1 ? F,F,F,F Normal False ?
AI4 Temperature 2 ? F,F,F,F Normal False ?
AI5 Pressure ? F,F,F,F Normal False ?
AI6 Density ? F,F,F,F Normal False ?
AI7 Energy Flow ? F,F,F,F Normal False ?
AI8 Totalizer 1 ? F,F,F,F Normal False ?
AI9 Totalizer 2 ? F,F,F,F Normal False ?
AI10 StatusRegister ? F,F,F,F Normal False ?
Channel 1 (4-
AI11 ? F,F,F,F Normal False ?
20mA)
Channel 2 (4-
AI12 ? F,F,F,F Normal False ?
20mA)
Channel 3 (4-
AI13 ? F,F,F,F Normal False ?
20mA)
AI14 Scaled Freq ? F,F,F,F Normal False ?
AI15 Flow Velocity ? F,F,F,F Normal False ?
AI16 Viscosity ? F,F,F,F Normal False ?
AI17 Frequency ? F,F,F,F Normal False ?
AI18 VorTex Amp ? F,F,F,F Normal False ?
AI19 FilterSetting ? F,F,F,F Normal False ?

Properties of Analog Input/Value Objects Type


Present Status Event Out of
ID Name Polarity
Value Flags State Service
BI1 Alarm1 ? F,F,F,F Normal False Normal
BI2 Alarm2 ? F,F,F,F Normal False Normal
BI3 Alarm3 ? F,F,F,F Normal False Normal
BI4 External ? F,F,F,F Normal False Normal

M-000-00010 4-29
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

Properties of Analog Input/Value Objects Type


Present Status Event Out of
ID Name out-of-service
Value Flags State Service
BV1 Reset ? F,F,F,F Normal False False

Data Link Layer Options:

BACnet IP, (Annex J)


BACnet IP, (Annex J), Foreign Device
ISO 8802-3, Ethernet (Clause 7)
ANSI/ATA 878.1, 2.5 Mb. ARCNET (Clause 8)
ANSI/ATA 878.1, EIA-485 ARCNET (Clause 8), baud rate(s)
MS/TP master (Clause 9), baud rate(s): 9600, 19200, 38400
MS/TP slave (Clause 9), baud rate(s):
Point-To-Point, EIA 232 (Clause 10), baud rate(s):
Point-To-Point, modem, (Clause 10), baud rate(s):
LonTalk, (Clause 11), medium:
Other:

Device Address Binding:

Is static device binding supported? (This is currently necessary for two-


way communication with MS/TP slaves and certain other devices.) :
Yes No

Networking Options:

 Router, Clause 6 - List all routing configurations, e.g., ARCNET-


Ethernet, Ethernet-MS/TP, etc.
 Annex H, BACnet Tunneling Router over IP
 BACnet/IP Broadcast Management Device (BBMD)
Does the BBMD support registrations by Foreign Devices?
 Yes  No
Does the BBMD support network address translation?
 Yes  No

Network Security Options:


 Non-secure Device - is capable of operating without BACnet Network
Security
 Secure Device - is capable of using BACnet Network Security (NS-SD
BVBB)

4-30 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 4 Serial Communications

 Multiple Application-Specific Keys:


 Supports encryption (NS-ED BVBB)
 Key Server (NS-KS BVBB)

Character Sets Supported:


Indicating support for multiple character sets does not imply that they can
all be supported simultaneously.

 ANSI X3.4 IBM™/Microsoft™DBCS


ISO 8859-1 ISO 10646 (UCS-2)
ISO 10646 (UCS-4) JIS C 6226

If this product is a communication gateway, describe the types of non-


BACnet equipment/networks(s) that the gateway supports:

• N/A

M-000-00010 4-31
Chapter 4 Serial Communications Series M22/M23/M24 Instruction Manual

5. Acronyms and Definitions

Item Description
APDU Application Protocol Data Unit
BACnet Building Automation and Control Network- Data
communication protocol
MS/TP Master-Slave Token passing(a twisted pair RS485 network
created by BACnet)
BIBB BACnet Interoperability Building Block (Specific
individual function blocks for data exchange between
interoperable devices).
BV Binary Value
BI Binary Input
AI Analog Input
RP Read Property
WP Write Property
RPM Read Property Multiple
WPM Write Property Multiple.
DDB Dynamic Device Binding
DOB Dynamic Object Binding
DCC Device communication Control

4-32 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Chapter 5 Troubleshooting and Repair


Warning!
Before attempting any flow
meter repair, verify that the
Hidden Diagnostics Menus
line is not pressurized. The menus shown on the following page can be accessed using the password
Always remove main power 16363, then moving to the display that reads “Diagnostics Menu” and pressing
before disassembling any ENTER (rather than one of the arrow keys).
part of the mass flow meter.

Use the right arrow key to move to the second level. Press EXIT to move from
the second level back to the first, press EXIT while in the first level to return to
the setup menus.

Caution: password 16363 will allow full access to the configuration and should
be used carefully to avoid changes that can adversely alter the function of the
meter.

Each of the menus on the following page will first be defined followed by
specific troubleshooting steps.

M-000-00010 5-1
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

Hidden Diagnostics Menus

{ ------------ Level One Values ------------ } { ------------------------------ Level Two Values ------------------------------ }

f fi Ck 4-20(1), Zero Alarm(1) Test Output Type


G A Lvl xxxx Low None

A1 A2 Adj. Filter 4-20(1), Fscale Alarm(2) Test Calibration Mode


A3 A4 xx dB xxxx Low

* Kc
Kb
It Iso. Power Volts
x.x vdc
4-20(2), Zero
xxxx
Alarm(3) Test
Low
A2D Ref. Resistor
2700

V O 4-20(2), Fscale Pres Cal Current


Reynolds Corr.
Re I xxxx 0.0003

Profile Factor Pulse Out Queue 4-20(3), Zero Press 9 C’s


Gain Control
+Inf xxxxxxxxxx xxxx More >

Obscuration TOF 4-20(3), Fscale RTD 1


x.xxxxxx Filter Control
G f xxxx More >

Rtd1 = x.x Sig. Rev. Ext. 4mA Cal. High Pass Filt. RTD 2
Rtd2 = x.x Micro. Rev. x 0.33 More >

Ext. 20mA Cal. Correction Pairs


Pe(v) = 0.0 AD R T Factory Defaults
x
Pv(v) = 0.0 F PT V

External Input Roughness


Std = 1.000 Spi Err Inactive Meter Type
xe-xx
Nrml = 1.000 Rcv Sent Temp 1
Temp 2
Config Code Force Recal?
Press
Visc = xxxx Version Retries 1BFE
Cp x
Ext. Full Scale
Enthalpy1 BTU/hr Internal Temp
x Test Pulse Out ** Min Delta H
1
x.xx xx.xx Deg F
Ext. Zero Scale
Test Scaled Freq Init Displ. (sec)
x
Enthalpy2 BTU/hr x xxx
Dens. Calc Time
x.xx
x.xxx Sec

x Cnts Rel. Roughness


Ext x.xxx mA x.xxxxxx * Not Present on M22 Models
** Energy EMS Meters Only

5-2 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Level One Hidden Diagnostics Values


• f = Vortex shedding frequency (Hz).
• fi = Adaptive filter – should be approximately 25% higher than
the vortex shedding frequency, this is a low-pass filter. If the
meter is using the Filter Control (see below) in the manual
mode, fi will be displayed as fm.
• G = Gain (applied to vortex signal amplitude). Gain defaults
to 1.0 and can be changed using the Gain Control (see below).
• A = Amplitude of vortex signal in Volts rms.
• A1, A2, A3, A4 = A/D counts representing the vortex signal
amplitude. Each stage (A1-A4) cannot exceed 512. Beginning
with stage A1, the A/D counts increase as the flow increases.
When stage A1 reaches 512, it will shift to stage A2. This will
continue as the flow rate increases until all 4 stages read 512 at
high flow rates. Higher flow rates (stronger signal strength)
will result in more stages reading 512.
• Kc, It, Kb = Profile equation (factory use only). Model M23
only
• V = Calculated average pipe velocity (ft/sec).
• Re = Calculated Reynolds number.
• Profile Factor = Factory use only.
• Obscuration = Factory use only.
• RTD1 = Resistance value of integral RTD in ohms.
• RTD2 = Optional RTD resistance value in ohms.
• Pe(v) = Pressure transducer excitation voltage
• Pv(v) = Pressure transducer sense voltage.
• Stnd = Density of fluid at standard conditions.
• Nrml = Density of fluid at normal conditions.
• Viscosity = Calculated viscosity of flowing fluid.
• Enthalpy1 BTU/hr = Factory use only.
• Enthalpy2 BTU/hr = Factory use only.
• x Cnts = A/D counts from the external 4-20 mA input.
• Ext x.xxx mA = Calculated external 4-20 mA input from the
digital counts.
• Ck = Calculated Ck at current operating conditions. Ck is a
variable in the equation that relates signal strength, density, and
velocity for a given application. It is used for noise rejection
purposes. Ck directly controls the fi value (see above). If the
Ck is set too low (in the calibration menu), then the fi value
will be too low and the vortex signal will be rejected resulting
in zero flow rate being displayed. The calculated Ck value in

M-000-00010 5-3
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

this menu can be compared to the actual Ck setting in the


calibration menu to help determine if the Ck setting is correct.
• Lvl = Threshold level. If the Low Flow Cutoff in the
calibration menu is set above this value, the meter will read
zero flow. The Lvl level can be checked at no flow. At no
flow, the Lvl must be below the Low Flow Cutoff setting or the
meter will have an output at no flow.
• Adj. Flilter = Adjustable filter. Displays the filtering in
decibels. Normally reads zero. If this value is consistently -5
or -10, for example, the Ck or density setting may be wrong.
• Iso. Power Volts = Nominally 2.7 VDC, if less than this check
the flow meter input power.
• O,I = Factory use only.
• Pulse Out Queue = Pulse output queue. This value will
accumulate if the totalizer is accumulating faster than the pulse
output hardware can function. The queue will allow the pulses
to “catch up” later if the flow rate decreases. A better practice
is to slow down the totalizer pulse by increasing the value in
the (unit)/pulse setting in the totalizer menu.
• TOF, G, f = Factory use only.
• Sig. Rev = Signal board hardware and firmware revision.
• Miro Rev = Microprocessor board hardware and firmware
revision.
• AD, R, T, F, PT, V = Factory use only.
• SPI Err, Rcv, Sent = Factory use only.
• Version Retries = Factory use only.
• Internal Temperature = Electronics temperature.
• Dens. Calc Time = Factory use only.
• Rel. Roughness = Factory use only.

Level Two Hidden Diagnostics Values

• 4-20(1) Zero = Analog counts to calibrate zero on analog


output 1.
• 4-20(1) FScale = Analog counts to cal. full scale on analog
output 1.
• 4-20(2) Zero = Analog counts to calibrate zero on analog
output 2.
• 4-20(2) FScale = Analog counts to cal. full scale on analog
output 2.

5-4 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

• 4-20(3) Zero = Analog counts to calibrate zero on analog


output 3.
• 4-20(3) FScale = Analog counts to cal. full scale on analog
output 3.
• Ext. 4 mA Cal. = Enter 0 for auto calibration or enter factory
supplied A/D counts. Note: You must connect a known 4.00
mA input if you are going to calibrate the unit.
• Ext. 20 mA Cal. = Enter 0 for auto calibration or enter factory
supplied A/D counts. Note: You must connect a known 20.00
mA input if you are going to calibrate the unit.
• External Input = Enter what the external 4-20 mA input
represents, i.e. Temperature 1, Temperature 2, or Pressure. The
meter will use this for its internal calculations.
• Ext. Full Scale = Enter the full scale units that correlate to the
20 mA point. Note: It must be in the units for the selected input
type such as Deg F, Deg C, PSIA, Bar A, etc.
• Ext. Zero Scale = Same as above but for the 4 mA point.
• Alarm (1) Test = Used as a test to verify that the alarm circuit
is functioning. When low is selected the alarm will initiate a
low alarm on the output. When High is selected it will give a
high alarm on the output.
• Alarm (2) Test = Used as a test to verify that the alarm circuit
is functioning. When low is selected the alarm will initiate a
low alarm on the output. When High is selected it will give a
high alarm on the output.
• Alarm (3) Test = Used as a test to verify that the alarm circuit
is functioning. When low is selected the alarm will initiate a
low alarm on the output. When High is selected it will give a
high alarm on the output.
• Reynolds Corr. = Reynolds number correction for the flow
profile. Set to Enable for M23 insertion and set to Disable for
M22/M24 inline.
• Gain Control = Manual gain control (factory use only). Leave
set at 1.
• Filter control = Manual filter control. This value can be
changed to any number to force the fi value to a constant. A
value of zero activates the automatic filter control which sets fi
at a level that floats above the f value.
• High Pass Filter = Filter setting – Factory use only
• Factory Defaults = Reset factory defaults. If you change this
to Yes and press Enter, all the factory configuration is lost and
you must reconfigure the entire program. Consult the factory

M-000-00010 5-5
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

before performing this process, it is required only in very rare


cases.
• Meter Type = Insertion (M23) or Inline (M22/M24) meter.
• Config Code = Factory use only.
• Test Pulse Out = Force totalizer pulse. Set to Yes and press
enter to send one pulse. Very useful to test totalizer counting
equipment.
• Test Scaled Freq = Enter a frequency value in order to test the
scaled frequency output. Return to 0 to stop the test.
• Output Type = Factory use only.
• Calibration Mode = Factory use only.
• A2D Ref. Resistor = Factory use only.
• Pressure Cal Current = Calibration value for the electronics
and pressure transducer combination. Consult Factory for
value.
• Pressure 9Cs = Nine pressure coefficients unique to the
pressure transducer. Use the RIGHT ARROW to access all
nine coefficients.
o Press. Max psi = Based on installed sensor.
o Press. Min psi = 0 psia
• RTD1. Press the RIGHT ARROW to access:
o Ro = RTD resistance at 0°C (1000 ohms).
o A = RTD coefficient A (.0039083).
o B = RTD coefficient B (-5.775e-07).
o RTD1 Max Deg. F = 500
o RTD1 Min Deg. F = -330
• RTD2 = Second RTD configuration, for special applications
only.
• Correction Pairs
o ft3/sec (1 through 10)
o %Dev. (1 through 10)
• Roughness = Factory use only.
• Force Recal? = Factory use only.
• Min. Delta H – Energy EMS meters only. Sets the deadband
for totalization to begin. Must be greater than this number (1
default) to initiate the totalizer.
• Init Displ. (sec) = Enter a value in seconds to initialize the
display every xxx seconds. Enter a value of 0 to disable
initializing the display.

5-6 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Analog Output Calibration


To check the 4–20 mA circuit, connect a DVM in series with the
output loop. Select zero or full scale (from the second level of the
hidden diagnostics) and then actuate the enter key twice. This
action will cause the meter to output its 4 mA or 20 mA condition.
If the DVM indicates a current greater than ± 0.006 mA from 4 or
20, adjust the setting up or down until the output is calibrated.

Note: these settings are not for adjusting the output zero and span
to match a flow range, that function is located in the Output Menu.

Display Contrast Adjustment


The flow meter display contrast is set at the factory but if the
display characters appear too dark or too light proceed as follows:

1) Hold down the “Exit” button on the front panel for 5 to 10


seconds. “Setting Contrast” will appear.

2) Push the “Up” arrow to darken the display or the “Down” arrow
to lighten it.

3) Push the “Enter” button to save the contrast setting.

M-000-00010 5-7
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

Troubleshooting the Flow Meter

Warning!
Before attempting any flow meter repair, verify that the line is not pressurized. Always remove main
power before disassembling any part of the mass flow meter. Use hazardous area precautions if
applicable. Static sensitive electronics - use electro-static discharge precautions.

First Check Items:


 Installation Direction Correct
 Installation Depth Correct (Insertion style meter)
 Power and Wiring Correct
 Application Fluid Correct
Please record what the fluid is __________________________

 Meter Range Correct for the Application


 Meter Configuration Correct
 Describe Installation Geometry i.e. upstream diameters, valve
position, downstream diameters, etc.

Record Values:
Record the following values from the Run Menu with the
meter installed in order to determine the operating state of
the flow meter:
With No Flow
With Flow
(if possible)
Flow =
Temperature =
Pressure =
Density =
Error Messages? =
Record the following values from the Hidden Diagnostics
Menu with the meter installed: (Use password 16363 to
access.)
With No Flow
With Flow
(if possible)
f=
fi =
A=
A1 =
A2 =
A3 =
A4 =
V=

5-8 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Record values - Hidden Diagnostics Menu continued:

With No Flow
With Flow
(if possible)
Ck =
Lvl =
Adj. Filter =
Iso. Power Volts =

Record the following values from the Calibration Menu.

Meter Size / Pipe ID =


Meter Factor =
Vortex Coef Ck =
Low Flow Cutoff =
Serial Number =

Determine the Fault


Symptom: Output at no Flow
1. The low flow cutoff is set too low. At no flow, go to the first
level of the hidden diagnostics menu and record the Lvl value.
The low flow cutoff must be set above this value.
2. Example: at no flow , Lvl = 25. Set the low flow cutoff in the
Calibration Menu to approximately 28 and the meter will no
longer read a flow rate at no flow.

Symptom: Erratic Output


1. The flow rate may be too low, just at the cutoff of the meter
range, and the flow cycles above and below the cutoff making
an erratic output. Consult the factory if necessary to confirm
the meter range based on current operating conditions. It may
be possible to lower the low flow cutoff to increase the meter
range. See the example above for output at no flow, only this
time the low flow cutoff is set too high. You can lower this
value to increase the meter range as long as you do not create
the output at no flow condition previously described.

2. Mechanical installation may be incorrect. Verify the straight


run is adequate as described in Chapter 2. For in-line meters,
make sure the meter is not installed backwards and there are no
gaskets protruding into the flow stream. For insertion meters,
verify the insertion depth and flow direction.

3. The meter may be reacting to actual changes in the flow


stream. The output can be smoothed using a time constant.

M-000-00010 5-9
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

The displayed values can be smoothed using the time constant


in the Display Menu. The analog outputs can be smoothed
using the time constant in the Output Menu. A time constant
of 1 will result in the change in value reaching 63% of its final
value in one second. A time constant of 4 is 22%, 10 is 9.5%
and 50 is 1.9% of the final value in one second. The time
constant equation is shown below (TC = Time Constant).

% change to final value


in one second = 100 (1 – e(-1/TC))

4. The vortex coefficient Ck may be incorrectly set. The Ck is a


value in the equation used to determine if a frequency
represents a valid vortex signal given the fluid density and
signal amplitude. In practice, the Ck value controls the
adaptive filter, fi, setting. During flow, view the f and fi values
in the first level of the hidden diagnostics. The fi value should
be approximately 10-20 % higher than the f value. If you raise
the Ck setting in the Calibration Menu, then the fi value will
increase. The fi is a low pass filter, so by increasing it or
lowering it, you can alter the range of frequencies that the
meter will accept. If the vortex signal is strong, the fi value
will increase to a large number – this is correct.

PRESSURE
VORTEX TEMPERATURE

Figure 5-1.Electronics Stack Sensor Connections

5-10 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

VORTEX

PRESSURE TEMPERATURE

Figure 5-2.Remote Feed Through Board Sensor Connections

Symptom: No Output
1. For remote mounted electronics, carefully check all the wiring
connections in the remote mount junction box. There are 18
connections that must be correct, verify each color (black and
red), shield, and wire number.

2. Turn on the pressure and temperature display in the Display


Menu and verify that the pressure and temperature are correct.

3. Using ESD precautions and hazardous area precautions,


remove the electronics enclosure window cover. Disconnect
the vortex sensor from the electronics stack or remote feed
through board. Refer to Figure 5-1 or 5-2. Measure the
resistance from each outside pin to the meter ground - each
should be open. Measure the resistance from the center pin to
the meter ground – this should be grounded to the meter.

OUTSIDE

INSIDE

OUTSIDE

Figure 5-3.Vortex Sensor Connector

With the sensor still disconnected, go to the first level of the


hidden diagnostics and display the vortex shedding frequency,

M-000-00010 5-11
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

f. Hold a finger on the three exposed pins on the analog board.


The meter should read electrical noise, 60 Hz for example. If
all readings are correct, re-install vortex sensor wires.

4. Verify all meter configuration and troubleshooting steps


previously described. There are many possible causes of this
problem, consult factory if necessary.

Symptom: Meter Displays Temperature Fault


1. For remote mounted electronics, carefully check all the wiring
connections in the remote mount junction box. There are 15
connections that must be correct, verify each color (black and
red), shield, and wire number.

2. Go to the first level of the hidden diagnostics and check the


resistance of the rtd1. It should be about 1080 ohms at room
temperature.

3. Using ESD precautions and hazardous area precautions,


remove the electronics enclosure window cover. Disconnect
the temperature sensor from the electronics stack or the remote
feed through board. Refer to Figures 5-1 ,5-2 and 5-4.
Measure the resistance across the outside pins of the
temperature sensor connector. It should read approximately
1080 ohms at room temperature (higher resistance at higher
temperatures). Measure the resistance across the inside pins,
they should read the same.

4. Measure the resistance from one of the outside pins to case


ground then from one of the inside pins to case ground. They
should read open.

OUTSIDE

INSIDE

OUTSIDE

Figure 5-4.Temperature Sensor Connector

5. Consult the factory with the findings

5-12 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Symptom: Meter Displays Pressure Fault


1. For remote mounted electronics, carefully check all the wiring
connections in the remote mount junction box. There are 15
connections that must be correct, verify each color (black and
red), shield, and wire number.

2. Using ESD precautions and hazardous area precautions,


remove the electronics enclosure window cover. Disconnect
the pressure sensor from the electronics stack or the remote
feed through board. Measure the resistance across the outside
pins of the pressure sensor connector, then across the inside
pins. Both readings should be approximately 4400 ohms.

OUTSIDE

INSIDE

OUTSIDE

Figure 5-5.Pressure Sensor Connector

3. Measure the resistance from one of the outside pins to case


ground then from one of the inside pins to case ground. They
should read open.

4. Go to the first level of the hidden diagnostics and record the


Pe(V) and Pv(V) values and consult the factory with findings.

M-000-00010 5-13
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

Electronics Assembly Replacement (All Meters)


The electronics boards are electrostatically sensitive. Wear a
grounding wrist strap and make sure to observe proper handling
precautions required for static-sensitive components.

1. Turn off power to the unit.

Warning! 2. Locate and loosen the small set screw which locks the larger
Before attempting any flow
meter repair, verify that the
enclosure cover in place. Unscrew the cover to expose the
line is not pressurized. electronics stack.
Always remove main power
before disassembling any
part of the mass flow meter.
3. Locate the sensor harnesses which come up from the neck of
the flow meter and attaches to the circuit boards. Make note of
the location of each sensor connection. Refer to figures 5-1
and 5-2. The vortex sensor connection is on the left, the
temperature sensor connection (if present) is second form the
left, and the pressure sensor connection (if present) is the right
most connector. Use small pliers to pull the sensor wiring
connectors off of the circuit boards.

4. Locate and loosen the small set screw which locks the smaller
enclosure cover in place. Unscrew the cover to expose the field
wiring strip. Tag and remove the field wires.

5. Remove the screws that hold the black wiring label in place,
remove the label.

6. Locate the 4 Phillips head screws which are spaced at 90-


degrees around the terminal board. These screws hold the
electronics stack in the enclosure. Loosen these screws (Note:
that these are captive screws, they will stay inside the
enclosure).

7. Carefully remove the electronics stack from the opposite side


of the enclosure. If the electronics stack will not come out,
gently tap the terminal strip with the screw driver handle. This
will loosen the rubber sealing gasket on the other side of the
enclosure wall. Be careful that the stack does not hang up on
the loose sensor harnesses.

8. Repeat steps 1 through 6 in reverse order to install the new


electronics stack.

5-14 M-000-00010
Series M22/M23/M24 Instruction Manual Chapter 5 Troubleshooting & Repair

Pressure Sensor Replacement (Series M22/M24 Only)


1. For local mounted electronics, remove the electronics stack as
previously described. For remote mount electronics, remove
all wires and sensor connectors from the remote feed through
board in the junction box at the meter.
2. Loosen the three set screws at the center of the adapter between
the meter and the enclosure.
3. Remove the top half of the adapter to expose the pressure
transducer.
4. Remove the transducer and replace it with the new one using
appropriate thread sealant.
5. Reassemble in reverse order.

Returning Equipment to the Factory


Before returning any Pro-V flow meter to the factory, you
must request a Return Material Authorization (RMA) number.
To obtain an RMA number and the correct shipping address,
contact Customer Service at:

888-386-7835 or 303-682-9999 in the USA,

When contacting Customer Service, be sure to have the meter


serial number and model code.

Please see the Meter Troubleshooting Checklist for additional


items which may help with problem isolation. When requesting
further troubleshooting guidance, please record the values on the
checklist at no flow and during flow if possible.

M-000-00010 5-15
Chapter 5 Troubleshooting & Repair Series M22/M23/M24 Instruction Manual

5-16 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications

Appendix A Product Specifications


Accuracy

Process M22/M24 Series In-Line Meters M23 Series Insertion Meters(1)


Variables
Liquids Gas & Steam Liquids Gas & Steam
Mass Flow ±1% of rate ±1.5% of rate(2) ±1.5% of ±2% of rate(2)
Rate over a 30:1 over a 30:1 rate over a over a 30:1
range(3) range(3) 30:1 range(3)
range(3)
Volumetric ±0.7% of ±1% of rate over ±1.2% of ±1.5% of rate
Flow Rate rate over a a 30:1 range(3) rate over a over a 30:1
30:1 30:1 range(3)
range(3) range(3)
Temperature ± 2° F ± 2° F ± 2° F ± 2° F
(± 1° C) (± 1° C) (± 1° C) (± 1° C)
Pressure 0.3% of 0.3% of 0.3% of 0.3% of
transducer transducer full transducer transducer full
full scale scale full scale scale
Density 0.3% of 0.5% of 0.3% of 0.5% of
reading reading(2) reading reading(2)
Notes: (1) Accuracies stated are for the total mass flow through the pipe.
(2) Over 50 to 100% of the pressure transducer’s full scale.
(3) Nominal rangeability is stated. Precise rangeability depends on fluid and pipe size.

Repeatability Mass Flow Rate: 0.2% of rate.


Volumetric Flow Rate: 0.1% of rate.
Temperature: ± 0.2° F (± 0.1° C).
Pressure: 0.05% of full scale.
Density: 0.1% of reading.

Stability Over 12 Months Mass Flow Rate: 0.2% of rate maximum.


Volumetric Flow Rate: Negligible error.
Temperature: ± 0.1° F (± 0.5° C) maximum.
Pressure: 0.1% of full scale maximum.
Density: 0.1% of reading maximum.

Response Time Adjustable from 1 to 100 seconds.

Material Capability Series M22/M24 In-Line Flow Meter:


Any gas, liquid or steam compatible with 316L stainless steel, C276
hastelloy or A105 carbon steel. Not recommended for multi-phase flu-
ids.

Series M23 Insertion Flow Meter:


Any gas, liquid or steam compatible with 316L stainless steel. Not
recommended for multi-phase fluids.

Flow Rates Typical mass flow ranges are given in the following table. Precise
flow depends on the fluid and pipe size. M23 insertion meters are
applicable to pipe sizes from 2 inch and above. Consult factory for
sizing program.

Water Minimum and Maximum Flow Rates


½-inch ¾-inch 1-inch 1.5-inch 2-inch 3-inch 4-inch 6-inch 8-inch
15 mm 20 mm 25 mm 40 mm 50 mm 80 mm 100 mm 150 mm 200 mm
gpm 1 1.3 2.2 5.5 9.2 21 36 81 142
22 40 67 166 276 618 1076 2437 4270
m3/hr .23 .3 0.5 1.3 2.1 4.7 8.1 18 32
5 9.1 15 38 63 140 244 554 970

M-000-00010 A-1
Appendix A Specifications Series M22/M23/M24 Instruction Manual

Typical Air Minimum and Maximum Flow Rates (SCFM)


Air at 70oF
Nominal Pipe Size (in)
Pressure 0.5 0.75 1 1.5 2 3 4 6 8
1.8 3 5 13 22 50 87 198 347
0 psig
18 41 90 221 369 826 1437 3258 5708
5 9 15 38 63 141 245 555 972
100 psig
138 325 704 1730 2890 6466 11254 25515 44698
7 13 21 52 86 193 335 761 1332
200 psig
258 609 1322 3248 5427 12140 21131 47911 83931
8 15 25 63 104 234 407 922 1615
300 psig
380 896 1944 4775 7978 17847 31064 70431 123375
10 18 29 72 120 269 467 1060 1857
400 psig
502 1183 2568 6309 10542 23580 41043 93057 163000
11 20 33 80 134 300 521 1182 2071
500 psig
624 1472 3195 7849 13115 28034 51063 115775 203000

Typical Air Minimum and Maximum Flow Rates (nm 3/hr)


Air at 20oC
Nominal Pipe Size (mm)
Pressure 15 20 25 40 50 80 100 150 200
3 5 9 21 36 79 138 313 549
0 barg
28 66 142 350 584 1307 2275 5157 9034
7 13 21 52 87 194 337 764 1339
5 barg
165 390 847 2080 3476 7775 13533 30682 53749
9 17 29 70 117 262 457 1035 1814
10 barg
304 716 1554 3819 6381 14273 24844 56329 98676
11 21 34 85 142 317 551 1250 2190
15 barg
442 1044 2265 5565 9299 20801 36205 82087 143801
13 24 40 97 162 363 632 1434 2511
20 barg
582 1373 2979 7318 12229 27354 47612 107949 189105
16 29 48 118 198 442 770 1745 3057
30 barg
862 2034 4414 10843 18119 40529 70544 159942 280187

Linear Range Smart electronics corrects for lower flow down to a Reynolds number of 5,000. The Reynolds
number is calculated using the fluid’s actual temperature and pres-
sure monitored by the meter. Rangeability depends on the fluid, pro-
cess connections and pipe size. Consult factory for your application.
Typical velocity range ability in standard applications is as follows:

Liquids 30:1 1 foot per second velocity minimum


30 feet per second velocity maximum
Gases 30:1 10 feet per second velocity minimum
300 feet per second velocity maximum

A-2 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications

Typical Saturated Steam Minimum and Maximum Flow Rates (lb/hr)


Nominal Pipe Size (in)
Pressure 0.5 0.75 1 1.5 2 3 4 6 8
6.5 12 20 49 82 183 318 722 1264
5 psig
52 122 265 650 1087 2431 4231 9594 16806
15 27 46 112 187 419 728 1652 2893
100 psig
271 639 1386 3405 5690 12729 22156 50233 87998
20 37 62 151 253 565 983 2229 3905
200 psig
493 1163 2525 6203 10365 23184 40354 91494 160279
24 45 74 182 304 680 1184 2685 4704
300 psig
716 1688 3664 9000 15040 33642 58556 132763 232575
28 51 85 209 349 780 1358 3079 5393
400 psig
941 2220 4816 11831 19770 44222 76971 174516 305717
31 57 95 233 389 870 1514 3433 6014
500 psig
1170 2760 5988 14711 24582 54987 95710 217001 380148

Typical Saturated Steam Minimum and Maximum Flow Rates (kg/hr)


Nominal Pipe Size (mm)
Pressure 15 20 25 40 50 80 100 150 200
3 5 8 19 32 72 126 286 500
0 barg
18 42 91 224 375 838 1459 3309 5797
6 11 18 45 75 167 290 658 1153
5 barg
95 224 485 1192 1992 4455 7754 17581 30799
8 15 24 59 99 222 387 877 1537
10 barg
168 397 862 2118 3539 7915 13777 31237 54720
9 17 29 71 119 266 463 1050 1840
15 barg
241 569 1236 3036 5073 11347 19750 44779 78444
11 20 33 81 136 304 529 1199 2100
20 barg
314 742 1610 3956 6611 14787 25738 58355 102226
13 24 40 99 165 369 642 1455 2548
30 barg
463 1092 2370 5822 9729 21763 37880 85884 150451

Linear Range Smart electronics corrects for lower flow down to a Reynolds number of 5,000. The Reynolds
number is calculated using the fluid’s actual temperature and pres-
sure monitored by the meter. Rangeability depends on the fluid, pro-
cess connections and pipe size. Consult factory for your application.
Typical velocity range ability in standard applications is as follows:

Liquids 30:1 1 foot per second velocity minimum


30 feet per second velocity maximum
Gases 30:1 10 feet per second velocity minimum
300 feet per second velocity maximum
Process Fluid Pressure

M22/M24 Pressure Ratings


Process Material Rating
Connection
Flanged 316L SS, A105 Carbon Steel, C276 Hastelloy 150, 300, 600 lb, PN16,
PN40, PN63
Wafer 316L SS, A105 Carbon Steel, C276 Hastelloy 600 lb, PN63

M-000-00010 A-3
Appendix A Specifications Series M22/M23/M24 Instruction Manual

M23 Pressure Ratings


Probe Seal Process Material Rating Ordering
Connection Code
Compression
Fitting 2-inch MNPT 316L SS ANSI 600 lb CNPT
2-inch 150 lb flange, 316L SS ANSI 150 C150,
DN50 PN16 lb, PN16 C16
2-inch 300 lb flange, 316L SS ANSI 300 C300,
DN50 PN40 lb, PN40 C40
2-inch 600 lb flange, 316L SS ANSI 600 C600,
DN50 PN63 lb, PN63 C64
Packing Gland 2-inch MNPT 316L SS 50 psig PNPT
2-inch 150 lb flange, 316L SS 50 psig P150,
DN50 PN16 P16
2-inch 300 lb flange, 316L SS 50 psig P300,
DN50 PN40 P40
Packing Gland with
Removable Retractor 2-inch MNPT 316L SS ANSI 300 lb PM, RR
2-inch 150 lb flange, 316L SS ANSI 150 lb P150,
DN50, PN16 P16,RR
2-inch 300 lb flange 316L SS ANSI 300 lb P300,
P40, RR
Packing Gland with
Permanent Retractor 2-inch MNPT 316L SS ANSI 600 lb PNPTR
2-inch 150 lb flange, 316L SS ANSI 150 lb P150R,
DN50 PN16 P16R
2-inch 300 lb flange, 316L SS ANSI 300 lb P300R,
DN50, PN40 P40R
2-inch 600 lb flange, 316L SS ANSI 600 lb P600R,
DN50 PN63 P64R

Pressure Transducer Ranges


Pressure Sensor Ranges(1), psia (bara)
Full Scale Operating Pres- Maximum Over-Range Pressure
sure
psia (bara) psia (bara)
30 2 60 4
100 7 200 14
300 20 600 40
500 35 1000 70
1500 100 2500 175

Note: (1) To maximize accuracy, specify the lowest full scale operat-
ing pressure range for the application. To avoid damage, the
flow meter must never be subjected to pressure above the
over-range pressure shown above.

Power Requirements 12 to 36 VDC, 25 mA, 1 W max., Loop Powered Volumetric or Mass


12 to 36 VDC, 300 mA, 9 W max. Multiparameter Mass options
100 to 240 VAC, 50/60 Hz, 5 W max. Multiparameter Mass
options

Use a Class I (Grounded Type) or Class 2 (double insulated) power


supply.

Installation (Over-voltage) Category II for transient over-voltages

AC & DC Mains supply voltage fluctuations are not to exceed


+/-10% of the rated supply voltage range.

A-4 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications
User is responsible for the provision of an external disconnect
means, disconnect line 1 and line 2 when 220 / 240 VAC power is
used, also provide over-current protection for the equipment (both
AC and DC models).

Display Alphanumeric 2 x 16 LCD digital display.


Six push-button switches (up, down, right, left, enter, exit) operable
through explosion-proof window using hand-held magnet. Viewing
at 90-degree mounting intervals.

Process Fluid and Process Fluid:


Ambient Temperature Standard temperature sensor: –330 to 500° F (–200 to 260°C)
High temperature sensor: to 750° F (400° C)

Ambient:
Operating temperature range: –40 to 140° F (–40 to 60° C)
Storage temperature range: –40 to 185° F (–40 to 85° C)
Maximum relative humidity: 0-98%, non-condensing conditions
Maximum altitude: -2000 to 14,000 feet (-610 to 4268 meters)

Pollution Degree 2 for the ambient environment

Output Signals (1)


Analog: Volumetric Meter: field rangeable linear 4-20 mA output
signal (1200 Ohms maximum loop resistance) selected by user for
mass flow rate or volumetric flow rate.

Communications: HART, MODBUS, RS485, BACnet

Multiparameter Meter: up to three field rangeable linear 4-20 mA


output signals (1200 Ohms maximum loop resistance) selected
from the five parameters–mass flow rate, volumetric flow rate, tem-
perature, pressure and density.

Pulse:Pulse output for totalization is a 50-millisecond duration pulse


operating a solid-state relay capable of switching 40 VDC, 40 mA
maximum.
Note: (1) All outputs are optically isolated and require external power
for operation.

Alarms Up to three programmable solid-state relays for high, low or window


alarms capable of switching 40 VDC, 40 mA maximum.

Totalizer Based on user-determined flow units, six significant figures in scien-


tific notation. Total stored in non-volatile memory.

Wetted Materials Series M22/M24 In-Line Flow Meter:


316L stainless steel standard.
C276 hastelloy or A105 carbon steel optional.

Series M23 Insertion Flow Meter:


316L stainless steel standard.
Teflon® packing gland below 500° F (260° C).
Graphite packing gland above 500° F (260° C).

Enclosure Protection
Classification NEMA 4X and IP66 cast enclosure.

Electrical Ports Two 3/4-inch female NPT ports.

Mounting Connections Series M22/M24: Wafer, 150, 300, 600 lb ANSI flange, PN16, PN40,
PN63 flange.
Series M23 Permanent installation: 2-inch MNPT; 150, 300, 600 lb
ANSI flange, PN16, PN40, PN63 flange with compression fitting
probe seal.
Series M23 Hot Tap(1) Installation: 2-inch MNPT; 150, 300, 600 lb
ANSI flange, PN16, PN40, PN63 flange and optional retractor with
packing gland probe seal.
Note: (1) Removable under line pressure.

M-000-00010 A-5
Appendix A Specifications Series M22/M23/M24 Instruction Manual
Mounting Position Series M22/M24 In-Line Flow Meter: No effect.
Series M23 Insertion Flow Meter: Meter must be perpendicular with-
in ± 5° of the pipe centerline.

Certifications Material Certificate – US Mill certs on all wetted parts


Pressure Test Certificate
Certificate of Conformance
NACE Certification (MR0175)
Oxygen Cleaning (CGA G-4.

A-6 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications

TM
Model Number Information - Pro-V Model M22 Inline Mass Vortex Flowmeter

* See follow ing page.


M22 * * *
Parent Num ber Code
M22 In-line Multivariable Mass
Vortex Flow Meter
Feature 1 : Multivariable Options
V Volumetric Flow Meter for liquid, gas and steam
VT Velocity and Temperature Sensors
VTP Velocity, Temperature and Pressure Sensors
VTEP Veloctiy, Temperature and External 4-20mA Input
(T or P)
VETEP Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)
VT-EM Energy output options
VTP-EM Energy options w ith Pressure Sensor
VTEP-EM Veloctiy, Temperature and External 4-20mA Input
(T or P)
VETEP-EM Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)

Feature 2: Flow Body


04 1/2-inch Nominal Bore (15mm)
06 3/4-inch Nominal Bore (20mm)
08 1-inch Nominal Bore (25mm)
12 1.5-inch Nominal Bore (40mm)
16 2-inch Nominal Bore (50mm)
24 3-inch Nominal Bore (80mm)
32 4-inch Nominal Bore (100mm)
48 6-inch Nominal Bore (150mm)
64 8-inch Nominal Bore (200mm)
80 10-inch Nominal Bore (250mm)
96 12-inch Nominal Bore (300mm)
Feature 3: Meter Body Material
C Carbon Steel **(Only for 1.5-inch and up)**
S 316 Stainless Steel
H Hastelloy
Feature 4: Process Connection
150 ANSI 150# Flange 16 PN 16
300 ANSI 300# Flange 40 PN 40
600 ANSI 600# Flange 63 PN 63
W Wafer ANSI 600# 100 PN 100
* JIS Flanges available upon request
Feature 5: Electronics Enclosure
L NEMA 4X, IP66 Enclosure
R( ) Remote Electronics NEMA 4X, IP66, Specify cable length in parentheses, Standard 50 ft
Feature 6: Display Option
DD Digital Display and Programming Buttons
Feature 7: Input Pow er
DCL 12-36 VDC, 25mA, 1W max. required on loop pow ered meters, 1AHL only
DCH 12-36 VDC, 300mA, 9W max. - use w ith 1AH, 1AM, 3AH, 3AM
DCHPOE 12-28 VDC or Pow er over Ethernet, 5W maximum, required on 1AMIP, 1ABIP, 3AMIP, 3ABIP.
AC 100-240 VAC, 50/60 Hz line pow er, 5W max. - use w ith 1AH, 1AM, 3AH, 3AM

M-000-00010 A-7
Appendix A Specifications Series M22/M23/M24 Instruction Manual

Model Number Information - Pro-V TM Model M22 Inline Mass Vortex Flowmeter

* See previous page.


M22 * * * * * * *

Feature 8: Output
1AHL Loop pow ered option - one analog output (4-20 mA), one scaled frequency, one pulse, HART, DCL input pow er only
1AH One analog output (4-20 mA), one alarm, one pulse, HART Communication Protocol, DCH or AC option only *
1AM One analog output (4-20 mA), one alarm, one pulse, MODBUS RTU Communication Protocol, DCH or AC option only *
1AMIP One analog output (4-20 mA), one alarm, one pulse, MODBUS TCP/IP Communication Protocol, DCHPOE ONLY*
1AB One analog output (4-20 mA), one alarm, one pulse, BACnet MS/TP Communication Protocol, DCH or AC option only *
1ABIP One analog output (4-20 mA), one alarm, one pulse, BACnet/IP Communication Protocol, DCHPOE ONLY *
3AH Three analog outputs (4-20 mA), three alarms, one pulse, HART (VT,VTP only), DCH or AC option only *
3AM Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS RTU (VT,VTP only), DCH or AC option only *
3AMIP Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS TCP/IP (VT,VTP only), DCHPOE ONLY*
3AB Three analog outputs (4-20 mA), three alarms, one pulse, BACnet MS/TP (VT,VTP only), DCH or AC option only *
3ABIP Three analog outputs (4-20 mA), three alarms, one pulse, BACnet/IP (VT,VTP only), DCHPOE ONLY *
* Includes scaled frequency output.
Feature 9: Tem perature Options
ST Standard Temperature
Process temperature -330° to 500°F (-200° to 260°C)
HT High Temperature
Process temperature 750°F (400°C)

Feature 10: Pressure Options


P0 No Pressure Sensor
P1 Maximum 30 psia (2 bara), Proof 60 psia (4 bara)
P2 Maximum 100 psia (7 bara), Proof 200 psia (14 bara)
P3 Maximum 300 psia (20 bara), Proof 600 psia (41 bara)
P4 Maximum 500 psia (34 bara), Proof 1000 psia (64 bara)
P5 Maximum 1500 psia (100 bara), Proof 2500 psia (175 bara)

A-8 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications
TM
Model Number Information - Pro-V Model M23 Insertion Mass Vortex Flowmeter

* See follow ing page.


M23 * *
Parent Num ber Code
M23 Insertion Multivariable Mass
Vortex Flow Meter

Feature 1 : Multivariable Options


V Volumetric Flow Meter for liquid, gas and steam
VT Velocity and Temperature Sensors
VTP Velocity, Temperature and Pressure Sensors
VTEP Veloctiy, Temperature and External 4-20mA Input (T
or P)
VETEP Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)
VT-EM Energy output options
VTP-EM Energy options w ith Pressure Sensor
VTEP-EM Veloctiy, Temperature and External 4-20mA Input (T
or P)
VETEP-EM Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)

Feature 2: Probe Length


SL Standard Length
CL Compact Length
EL Extended Length

Feature 3: Electronics Enclosure


L NEMA 4X, IP66 Enclosure
R( ) Remote Electronics NEMA 4X, IP66, Specify cable length in parentheses, Standard 50 ft
Feature 4: Dipsplay Option
DD Digital Display and Programming Buttons

Feature 5: Input Pow er


DCL 12-36 VDC, 25mA, 1W max. required on loop pow ered meters, 1AHL only
DCH 12-36 VDC, 300mA, 9W max. - use w ith 1AH, 1AM, 3AH, 3AM
DCHPOE 12-28 VDC or Pow er over Ethernet, 5W maximum, required on 1AMIP, 1ABIP, 3AMIP,3ABIP
AC 100-240 VAC, 50/60 Hz line pow er, 5W max. - use w ith 1AH, 1AM, 3AH, 3AM

Feature 6: Output Signal


1AHL Loop pow ered option - one analog output (4-20 mA), one scaled frequency, one pulse, HART, DCL input pow er only
1AH One analog output (4-20 mA), one alarm, one pulse, HART Communication Protocol, DCH or AC option only *
1AM One analog output (4-20 mA), one alarm, one pulse, MODBUS RTU Communication Protocol, DCH or AC option only*
1AMIP One analog output (4-20 mA), one alarm, one pulse, MODBUS TCP/IP Communication Protocol, DCHPOE ONLY*
1AB One analog output (4-20 mA), one alarm, one pulse, BACnet MS/TP Communication Protocol, DCH or AC option only *
1ABIP One analog output (4-20 mA), one alarm, one pulse, BACnet/IP Communication Protocol, DCHPOE ONLY *
3AH Three analog outputs (4-20 mA), three alarms, one pulse, HART (VT,VTP only), DCH or AC option only *
3AM Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS RTU (VT,VTP only), DCH or AC option only *
3AMIP Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS TCP/IP (VT,VTP only), DCHPOE ONLY *
3AB Three analog outputs (4-20 mA), three alarms, one pulse, BACnet MS/TP (VT,VTP only), DCH or AC option only *
3ABIP Three analog outputs (4-20 mA), three alarms, one pulse, BACnet/IP (VT,VTP only), DCHPOE ONLY *

Feature 7: Tem perature Options * Includes scaled frequency output.


ST Standard Temperature
Process temperature -330° to 500°F (-200° to 260°C)
HT High Temperature
Process temperature to 750°F (400°C)

M-000-00010 A-9
Appendix A Specifications Series M22/M23/M24 Instruction Manual

Model Number Information - Pro-V TM Model M23 Insertion Mass Vortex Flowmeter

* See previous page.


M23 * * * * * * *

Feature 8: Pressure Options


P0 No Pressure Sensor
P1 Maximum 30 psia (2 bara), Proof 60 psia (4 bara)
P2 Maximum 100 psia (7 bara), Proof 200 psia (14 bara)
P3 Maximum 300 psia (20 bara), Proof 600 psia (41 bara)
P4 Maximum 500 psia (34 bara), Proof 1000 psia (64 bara)
P5 Maximum 1500 psia (100 bara), Proof 2500 psia (175 bara)

Feature 9: Process Connections


CNPT Compression, 2 inch NPT P300 Packing Gland*, 2 inch 300# Flange
C150 Compression, 2 inch 150# Flange P40 Packing Gland*, DN50 PN40 Flange
C16 Compression, DN50 PN16 Flange PNPTR Packing Gland, 2 inch NPT, Retractor
C300 Compression, 2 inch 300# Flange P150R Packing Gland, 2 inch 150# Flange, Retractor
C40 Compression, DN50 PN40 Flange P16R Packing Gland, DN50 PN16 Flange, Retractor
C600 Compression, 2 inch 600# Flange P300R Packing Gland, 2 inch 300# Flange, Retractor
C63 Compression, DN50 PN63 Flange P40R Packing Gland, DN50 PN40 Flange, Retractor
PNPT Packing Gland*, 2 inch NPT P600R Packing Gland, 2 inch 600# Flange, Retractor
P150 Packing Gland*, 2 inch 150# Flange P63R Packing Gland, DN50 PN63 Flange, Retractor
P16 Packing Gland*, DN50 PN16 Flange * JIS flanges available upon request

A-10 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix A Specifications
TM
Model Number Information - Pro-V Model M24 Inline Mass Vortex Flowmeter
* See follow ing page.
M24 ____ * * *
Parent Num ber Code
M24 In-line Multivariable Mass Vortex Flow Meter
M24R In-line Reducing Multiv. Mass Vortex Flow Meter
Feature 1 : Multivariable Options
V Volumetric Flow Meter for liquid, gas and steam
VT Velocity and Temperature Sensors
VTP Velocity, Temperature and Pressure Sensors
VTEP Veloctiy, Temperature and External 4-20mA Input
(T or P)
VETEP Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)
VT-EM Energy output options
VTP-EM Energy options w ith Pressure Sensor
VTEP-EM Veloctiy, Temperature and External 4-20mA Input
(T or P)
VETEP-EM Veloctiy, External RTD Temperature Input,
External 4-20mA Input (T or P)

Feature 2: Flow Body (M24) Feature 2: Reducing Flow Body (M24R)


04 1/2-inch Nominal Bore (15mm)
06 3/4-inch Nominal Bore (20mm) 3/4-inch by 1/2-inch Reducing Meter (15mm)
08 1-inch Nominal Bore (25mm) 1-inch by 1/2-inch Reducing Meter (15mm)
12 1.5-inch Nominal Bore (40mm) 1.5-inch by 1-inch Reducing Meter (25mm)
16 2-inch Nominal Bore (50mm) 2-inch by 1.5-inch Reducing Meter (40mm)
24 3-inch Nominal Bore (80mm) 3-inch by 2-inch Reducing Meter (50mm)
32 4-inch Nominal Bore (100mm) 4-inch by 3-inch Reducing Meter (80mm)
48 6-inch Nominal Bore (150mm) 6-inch by 4-inch Reducing Meter (100mm)
64 8-inch Nominal Bore (200mm) 8-inch by 6-inch Reducing Meter (150mm)
80 10-inch Nominal Bore (250mm) 10-inch by 8-inch Reducing Meter (200mm)
96 12-inch Nominal Bore (300mm) 12-inch by 10-inch Reducing meter (250mm)
Feature 3: Meter Body Material
C Carbon Steel *(Only for 1.5-inch and up, not available for reducing m eters)*
S 316 Stainless Steel
H Hastelloy **(Not available for reducing m eters)**
Feature 4: Process Connection
150 ANSI 150# Flange 16 PN 16
300 ANSI 300# Flange 40 PN 40
600 ANSI 600# Flange 63 PN 63
W Wafer ANSI 600# 100 PN 100
* JIS Flanges available upon request
Feature 5: Electronics Enclosure
L NEMA 4X, IP66 Enclosure
R( ) Remote Electronics NEMA 4X, IP66, Specify cable length in parentheses, Standard 50 ft
Feature 6: Display Option
DD Digital Display and Programming Buttons
Feature 7: Input Pow er
DCL 12-36 VDC, 25mA, 1W max. required on loop pow ered meters, 1AHL only
DCH 12-36 VDC, 300mA, 9W max. - use w ith 1AH, 1AM, 3AH, 3AM
DCHPOE 12-28 VDC or Pow er over Ethernet, 5W maximum, required on 1AMIP, 1ABIP, 3AMIP, 3ABIP.
AC 100-240 VAC, 50/60 Hz line pow er, 5W max. - use w ith 1AH, 1AM, 3AH, 3AM

M-000-00010 A-11
Appendix A Specifications Series M22/M23/M24 Instruction Manual

Model Number Information - Pro-V TM Model M24 Inline Mass Vortex Flowmeter

* See previous page.


M24 * * * * * * *

Feature 8: Output
1AHL Loop pow ered option - one analog output (4-20 mA), one scaled frequency, one pulse, HART, DCL input pow er only
1AH One analog output (4-20 mA), one alarm, one pulse, HART Communication Protocol, DCH or AC option only *
1AM One analog output (4-20 mA), one alarm, one pulse, MODBUS RTU Communication Protocol, DCH or AC option only *
1AMIP One analog output (4-20 mA), one alarm, one pulse, MODBUS TCP/IP Communication Protocol, DCHPOE ONLY*
1AB One analog output (4-20 mA), one alarm, one pulse, BACnet MS/TP Communication Protocol, DCH or AC option only *
1ABIP One analog output (4-20 mA), one alarm, one pulse, BACnet/IP Communication Protocol, DCHPOE ONLY *
3AH Three analog outputs (4-20 mA), three alarms, one pulse, HART (VT,VTP only), DCH or AC option only *
3AM Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS RTU (VT,VTP only), DCH or AC option only *
3AMIP Three analog outputs (4-20 mA), three alarms, one pulse, MODBUS TCP/IP (VT,VTP only), DCHPOE ONLY*
3AB Three analog outputs (4-20 mA), three alarms, one pulse, BACnet MS/TP (VT,VTP only), DCH or AC option only *
3ABIP Three analog outputs (4-20 mA), three alarms, one pulse, BACnet/IP (VT,VTP only), DCHPOE ONLY *
* Includes scaled frequency output.
Feature 9: Tem perature Options
ST Standard Temperature
Process temperature -330° to 500°F (-200° to 260°C)
HT High Temperature
Process temperature 750°F (400°C)

Feature 10: Pressure Options


P0 No Pressure Sensor
P1 Maximum 30 psia (2 bara), Proof 60 psia (4 bara)
P2 Maximum 100 psia (7 bara), Proof 200 psia (14 bara)
P3 Maximum 300 psia (20 bara), Proof 600 psia (41 bara)
P4 Maximum 500 psia (34 bara), Proof 1000 psia (64 bara)
P5 Maximum 1500 psia (100 bara), Proof 2500 psia (175 bara)

A-12 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix B Approvals

Appendix B Approvals
ATEX-IECEx Specifications / Approval
EN 60079-0 (2012+A11:2013)
IEC 60079-0 (2011)
Electrical Apparatus for explosive gas atmospheres
General Requirements

EN IEC 60079-1 (2014)


Electrical Apparatus for explosive gas atmospheres
Flameproof enclosures “d”

EN 60079-31 (2014)
Explosive atmospheres. Equipment dust ignition protection by enclosure "t"

Directive 2014/34/EU
Equipment Intended for use in Potentially Explosive Atmospheres (ATEX)

M-000-00010 B-1
Appendix B Approvals Series M22/M23/M24 Instruction Manual

Declaration of Conformity
VorTek Instruments, LLC
8475 W I-25 Frontage Rd
Longmont, CO 80504

Declares in sole responsibility that the product:

Vortex Flow Meter


Type M22 / M23 / M24 / Pro-T

Conforms to the regulations of the European Directives:

EMC Directive 2014/30/EU


Low Voltage Directive 2014/35/EU
Pressure Equipment Directive 2014/68/EU
Explosion Protection Directive 2014/34/EU

Applied harmonized standards or normative documents:

EMC Directive 2014/30/EU


EN 61000-6-2: 2005 (immunity industrial environment)
EN 61000-6-3: 2007 +A1:2011 (emission residential, commercial)
EN 55011: 2009 +A1:2010 Group 1, Class A (ISM radio-frequency equipment)

Low Voltage Directive 2014/35/EU


EN IEC 61010-1: 2010 Safety requirements for electrical measuring, control and laboratory devices

Pressure Equipment Directive 2014/68/EU


Category I , Annex III, Module H Full Quality Assurance
Article 4, Paragraph 3 Sound Engineering Practice

RoHS Directive 2011/65/EU


EN IEC 63000:2018 Technical assessment of electrical and electronic products with respect to RoHS

Explosion Protection Directive 2014/34/EU


EN IEC 60079-0: 2011 Electrical Apparatus for explosive gas atmospheres, General Requirements
EN IEC 60079-1: 2014-06 Electrical Apparatus for explosive gas atmospheres, flameproof enclosures “d”
EN IEC 60079-31: 2013 Explosive atmospheres. Equipment dust ignition protection by enclosure "t"

EC – Type Examination Certificate: FM 17ATEX0018X

Name and Address of the Notified Body

FM Approvals Europe Ltd TÜV – Industrie Service GmbH


Element 78 TÜV SÜD Gruppe
1 Georges Quay Plaza DudenstraBe 28
Dublin D02 E440, Ireland D-68167 Mannheim

2809 0036

Eric Sanford Steve Donjon

__________________________ ________________________
Managing Director Quality Manager

B-2 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix B Approvals

Registrant: VorTek Instruments, LLC (Azbil VorTek, LLC)


Equipment Name: Vortex Flowmeter
Registration No’s: MSIP-REM-A1V-A001
MSIP-REM-A1V-A002
MSIP-REM-A1V-A003
Date of Registration: 2014-02-14

Technical assistance may be obtained by contacting Customer Service at:


(888) 386-7835 or (303) 682-9999 in the USA

M-000-00010 B-3
Appendix B Approvals Series M22/M23/M24 Instruction Manual

B-4 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix C Flow Meter Calculations

Appendix C Flow Meter Calculations


In-Line Flow Meter Calculations
Volume Flow Rate
f
QV =
K

Flowing Velocity
QV
Vf =
A

Mass Flow Rate


QM = QV 

Where:
A = Cross sectional area of the pipe (ft^2)
f = Vortex shedding frequency (pulses / sec)
K = Meter factor corrected for thermal expansion (pulses / ft^3)
QM = Mass flow rate (lbm / sec)
Qv = Volume flow rate (ft^3 / sec)
Vf = Flowing velocity (ft / sec)
 = Density (lbm / ft^3)

M-000-00010 C-1
Appendix C Flow Meter Calculations Series M22/M23/M24 Instruction Manual

Insertion Flow Meter Calculations


Flowing Velocity
f
Vf =
Kc

Volume Flow Rate


QV = V f A

Mass Flow Rate


QM = V f A

Where:
A = Cross sectional area of the pipe (ft^2)
f = Vortex shedding frequency (pulses / sec)
Kc = Meter factor corrected for Reynolds Number (pulses / ft)
Qv = Volume flow rate (ft^3 / sec)
QM = Mass flow rate (lbm / sec)
Vf = Flowing velocity (ft / sec)
 = Density (lbm / ft^3)

C-2 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix C Flow Meter Calculations

Energy Flow Calculations


Energy is calculated for a steam supply/condensate return or hot/chilled water
system. For steam/water, the meter must be located in the supply line; otherwise,
the meter may be located in either the supply line or in the return line.

Steam supply, water return, meter steam supply

Energy = mdot * (h0 - pctRet * h1)

Where:
mdot = mass flow at the meter, lbm/sec
pctRet = estimated percent of mass flow returned
h0 = Steam Enthalpy (t0, p)
h1 = Water Enthalpy (t1)
t0 = steam temperature
t1 = return water temperature
p = steam pressure

Water supply and return, meter supply

Energy = mdot * (h0 - pctRet * h1)

Where:
h0 = Water Enthalpy (t0)
h1 = Water Enthalpy (t1)
t0 = supply water temperature
t1 = return water temperature

Water supply and return, meter return

Energy = mdot * (h1/pctRet - h0)

Where:
h0 = Water Enthalpy (t0)
h1 = Water Enthalpy (t1)
t0 = return water temperature
t1 = supply water temperature

The energy flow is positive if less energy is returned than is supplied. This im-
plies that chilled water systems will indicate negative energy flow. Positive and
negative energy flows are accumulated in separate totalizers.

M-000-00010 C-3
Appendix C Flow Meter Calculations Series M22/M23/M24 Instruction Manual

Fluid Calculations
Calculations for Steam T & P
When “Steam T & P” is selected in the “Real Gas” selection of the Fluid
Menu, the calculations are based on the equations below.

Density
The density of steam is calculated from the formula given by Keenan and
Keys. The given equation is for the volume of the steam.

4.555.04 • T
v= +B
p
2 4 13
B = B0 + B0 g1 ( )  p + B0 g 2 ( ) 3  p 3 − B0 g 3 ( ) 12  p12
2
B0 = 1.89 − 2641.62  1080870
g1 ( ) = 82.546  −1.6246 105  2
g2 ( ) = 0.21828 −1.2697 105  2
g3 ( ) = 3.63510−4 − 6.7681064  24

Where tau is 1/ temperature in Kelvin.

The density can be found from 1/(v/ standard density of water).

Viscosity
The viscosity is based on an equation given by Keenan and Keys.
1.50110 −5 T
 ( poise) =
1 + 446.8 / T

Where T is the temperature in Kelvin

C-4 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix C Flow Meter Calculations

Calculations for Gas (“Real Gas” and “Other Gas”)


Use this formula to determine the settings for “Real Gas; Gas” selections
and “Other Gas” selections entered in the Fluid Menu. The calculations
for gas were taken from Richard W. Miller, Flow Measurement Engi-
neering Handbook (Third Edition, 1996).

Density
The density for real gases is calculated from the equation:

GM w, Air p f
=
Z f R0T f

Where G is the specific gravity, Mw is the molecular weight of


air, pf is the flowing pressure, Z is flowing compressibility, Ro is
the universal gas constant, and T is the flowing temperature.

The specific gravity, and Ro are known and are stored in a table used by
the Vortex meter.

The hard coefficient to find is the compressibility, Z. Z is found using the


Redlich-Kwong Equation (Miller page 2-18).

The Redlich-Kwong Equation uses the reduced temperature and pressure


to calculate the compressibility factor. The equations are non linear and
an iterative solution is used. The Vortex program uses Newton’s Method
on the Redlich-Kwong equations to iteratively find the compressibility
factor. The critical temperature and pressure used in the Redlich-Kwong
equation are stored in the fluid data table with the other coefficients.

Viscosity
The viscosity for real gases is calculated using the exponential equation
for two known viscosities. The equation is:

 cP = aTKn

Where a and n are found from two known viscosities at


two temperatures.

ln[( cP ) 2 /( cP )1 ]
n=
ln(TK 2 / TK 1 )
and
(  cP )1
a=
TKn1

M-000-00010 C-5
Appendix C Flow Meter Calculations Series M22/M23/M24 Instruction Manual

Calculations for Liquid


Use this formula to determine the settings for “Goyal-Dorais” selections
and “Other Liquid” selections entered in the Fluid Menu. The liquid cal-
culations were taken from Richard W. Miller, Flow Measurement Engi-
neering Handbook (Third Edition, 1996).

Density
The liquid density is found using the Goyal-Doraiswamy Equation. Goy-
al-Doraiswamy uses the critical compressibility, critical pressure and
critical temperature, along with the molecular weight to find the density.
The equation for specific gravity is:

pc Mw  0.008 Tf 
GF =  − 0.01102 
Tc  Z c0.773 TC 

The specific gravity can then be converted into density.

Viscosity
The liquid viscosity is found by Andrade's equation. This uses two vis-
cosities at different temperatures to extrapolate the viscosity.

Andrade's equation:

BL
 = AL exp
Tdeg R

To find A and B

Tdeg R1Tdeg R 2 ln(1 /  2 )


BL =
Tdeg R 2 − Tdeg R1

1
AL =
exp( BL / Tdeg R1 )

The temperatures are all in degrees Rankin. Do not believe the subscript
R means they are reduced temperatures.

C-6 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix D Glossary

Appendix D Glossary
A B C D
A Cross sectional area.

ACFM Actual Cubic Feet Per Minute (volumetric flow rate).

ASME American Society of Mechanical Engineers.

Bluff Body A non-streamlined body placed into a flow stream to


create vortices. Also called a Shedder Bar.

BTU British Thermal Unit, an energy measurement.

Cenelec European Electrical Code.

Compressibility A factor used to correct for the non-ideal changes in


Factor a fluid’s density due to changes in temperature
and/or pressure.

CSA Canadian Standards Association.

d Width of a bluff body or shedder bar.

D Diameter of a flow channel.

E F G H
f Frequency of vortices generated in a vortex flow
meter, usually in Hz.

Flow Channel A pipe, duct, stack, or channel containing flowing fluid.

Flow Profile A map of the fluid velocity vector (usually non-


uniform) in a cross-sectional plane of a flow channel
(usually along a diameter).

FM Factory Mutual.

Ft Foot, 12 inches, a measure of length.

Ft^2 Square feet, measure of area.

Ft^3 Cubic feet, measure of volume.

GPM Gallons Per Minute.

Hz Hertz, cycles per second.

M-000-00010 D-1
Appendix D Glossary Series M22/M23/M24 Instruction Manual

I J K L
In-Line Flow Meter A flow meter which includes a short section of piping
which is put in-line with the user’s piping.

Insertion Flow Meter A flow meter which is inserted into a hole in the us-
er’s pipeline.

Joule A unit of energy equal to one watt for one second. Al-
so equal to a Newton-meter.

LCD Liquid crystal display.

M N O P
m Mass flow rate.

mA Milli-amp, one thousandth of an ampere of current.

 Viscosity, a measure of a fluid’s resistance to shear stress.


Honey has high viscosity, alcohol has low viscosity.

nm3/hr Normal cubic meters per hour (flow rate converted to


normal conditions, as shipped 101 kPa and 0° C).
User definable.

P Permanent pressure loss.

P Line pressure (psia or bar absolute).

 act The density of a fluid at the actual temperature and


pressure operating conditions.

 std The density of a fluid at standard conditions (usually


14.7 psia and 20° C).

Permanent Unrecoverable drop in pressure.


Pressure Loss

Piezoelectric Crystal A material which generates an electrical charge


when the material is put under stress.

PRTD An resistance temperature detector (RTD) with plati-


num as its element. Used because of high stability.

psia Pounds per square inch absolute


(equals psig + atmospheric pressure). Atmospheric
pressure is typically 14.696 psi at sea level.

psig Pounds per square inch gauge.

D-2 M-000-00010
Series M22/M23/M24 Instruction Manual Appendix D Glossary

PV Liquid vapor pressure at flowing conditions (psia or


bar absolute).

Q R S T
Q Flow rate, usually volumetric.

Rangeability Highest measurable flow rate divided by the lowest


measurable flow rate.

Reynolds Number A dimensionless number equal to the density of a fluid


or Re times the velocity of the fluid times the diameter of the
fluid channel, divided by the fluid viscosity (i.e., Re =
VD/). The Reynolds number is an important num-
ber for vortex flow meters because it is used to deter-
mine the minimum measurable flow rate. It is the ratio
of the inertial forces to the viscous forces in a flowing
fluid.

RTD Resistance temperature detector, a sensor whose


resistance increases as the temperature rises.

scfm Standard cubic feet per minute (flow rate converted


to standard conditions, as shipped 14.696 psia and
59° F). User definable.

Shedder Bar A non-streamlined body placed into a flow stream to


create vortices. Also called a Bluff Body.

Strouhal Number A dimensionless number equal to the frequency


or St of vortices created by a bluff body times the width of
the bluff body divided by the velocity of the flowing
fluid (i.e., St = fd/V). This is an important number for
vortex flow meters because it relates the vortex fre-
quency to the fluid velocity.

Totalizer An electronic counter which records the total accu-


mulated flow over a certain range of time.

Traverse The act of moving a measuring point across the


width of a flow channel.

U V W X Y Z
Uncertainty The closeness of agreement between the result of a
measurement and the true value of the measurement.

V Velocity or voltage.

VAC Volts, alternating current.

VDC Volts, direct current.

VORTEX An eddy of fluid.

M-000-00010 D-3
Appendix D Glossary Series M22/M23/M24 Instruction Manual

D-4 M-000-00010