RAPIDPoint® 500 Blood Gas System

Quick Reference Guide

www.siemens.com/diagnostics

Step-by-Step
1 2 3

Make sure correct sample type is highlighted


Remove air, cap sample, and mix thoroughly. Scan or enter password, if prompted.
or touch screen to change selection.

4 5 6

Insert the sample device as shown on the Enter the demographics if prompted,
Scan sample ID if available.
screen and touch “start.” and touch the arrow button.

7 8 9

When prompted, remove the sample and View the results. A record of results will be Touch the arrow button to return to the
touch the arrow button. sent to the LIS and/or printed. ready screen to analyze another sample.

Troubleshooting

1. If there is an obstruction (possible clot in the sample), you will be required to change
the sample port as shown on the screen.
2. To view a video of the procedure, press the black camera icon next to point 1.
3. When you have attached the new sample port, touch the arrow button.

Answers for life.

RAPIDPoint and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc.
All other trademarks and brands are the property of their respective owners.
Order No. A91DX-110748-GC1-4A00 | 10-2011 | All rights reserved | © 2011 Siemens Healthcare Diagnostics Inc.