N2 Generator ENG Manual
N2 Generator ENG Manual
N2 Generator ENG Manual
NITROGEN GENERATOR
USER MANUAL
High Purity Nitrogen Generator
Contents
i
High Purity Nitrogen Generator
ii
High Purity Nitrogen Generator
Chapter 1.Introduction
This document provides the necessary information for the installation and operation for the
NITROGEN GENERATOR and his options; it also describes the simple maintenance operations as
well as its troubleshooting.
The operating manual consider and describe an instrument generally equipped with the most
complex configuration ; should the explanation concerning the most complex instrument be too
different from the simpler instrument, both cases will be described separately.
1
High Purity Nitrogen Generator
Should this instrument produce some significant interference to the radio and television, this case
can be identified by plugging and unplugging the instrument. We recommend, in order to eliminate the
interferences, to take one or more of the following actions:
2.2 CE Conformity
This equipment was built in compliance and is compatible with EC recommendations concerning
electrical safety and electromagnetic emissions. It complies with 89/336/EWG, 93/98/EWG, standards
EN 50 081-1, in 50 081-2, EN 50 082 - 1 and EN 50 082 - 2.
Note
Any modifications on the instrument which have not been approved in writing by the manufacturer will
automatically cancel the manufacturer warranty. If such modifications are nevertheless undertaken,
they are under the user responsibility; the manufacturer will under no circumstances be responsible for
any damages direct or indirect which they would cause
2
High Purity Nitrogen Generator
company who sold this product. If you want to get rid of this instrument, identify it as such and direct it
to a certified recycling centre.
3
High Purity Nitrogen Generator
4
High Purity Nitrogen Generator
The front face is divided into two parts, the upper part has the screen and tactile key board ; the
lower part fitted with hinges allow to accede to the pressure setting and inlet filter.
5
High Purity Nitrogen Generator
column, when the regeneration process is complete. By option, pure nitrogen is next carried to a
stainless steel catalytic oven where all hydrocarbons (HTC) and CH4 as well as the CO are oxidized.
The oven is fitted with a high performance catalyst based on a Platinum-Palladium mixture. Heated to
a high temperature, the catalyst allow to reduce the HTC and CO to less than 0.1ppm (100ppb).
The pressurized nitrogen is send to a tank, the capacity of which depends on the N2 capacity. The
N2 is pressure regulated and go through a fine filter which holds 99.99 % of particles larger than
0.5µm.
At this point the Nitrogen is available at the outlet fitting.
The compressor is driven by the tank pressure; it will stop itself when the upper set pressure value is
reached and start again when the lower set pressure is reached; the lower tank pressure is above the
outlet set pressure.
6
High Purity Nitrogen Generator
Filter
(external)
Evaporator
Catalysis Oven
Low noise fan
Ventilation
DRYING and
PURIFICATION
MODULE
Tank
(Volume
depends on flow Manual
Pressure
rate capacity) Pressure
Pressure sensor
sensor (tank) regulator (Outlet)
7
High Purity Nitrogen Generator
Generator responds by closing or opening a relay in the event of an alarm. The remote control can be
NC (normally closed) or NO (normally open).
8
High Purity Nitrogen Generator
Frame
Electronic
control module
Tank
N2 drying and
purification
module
Air
compression
module
9
High Purity Nitrogen Generator
AIR Compressor
Command
connector
Cooling coil
10
High Purity Nitrogen Generator
AIR Compressor
Cooling coil
11
High Purity Nitrogen Generator
Catalytic oven
with cooling coil
(option)
Connection fittings
with tank(s)
Gous3-HPN2 with
power supply
PS3550-50
N2 Outlet module
Coalescing filter
Coalescing filter Bloc of solenoid valves
(purge V4) Purge V5
12
High Purity Nitrogen Generator
The High Purity N2 Generator is set on an individual wooden mini-palette 40x80cm, which is the
basis of its transport packaging. Its upper part is protected by shaped foams and a thick cardboard.
The thick cardboard is screwed on the palette. The High Purity N2 Generator extraction from its
packaging start by unscrewing the cardboard, then cutting the straps. At this stage it is possible to
verify the visual integrity of the instrument.
Any damages must be immediately identified and photographed; It should be reported to the carrier
as well as to your local Distributor or YL Instrumentas.
For major damages, the High Purity N2 Generator shall be returned to the manufacturer after
synchronization with the after-sales service, which can be reached by e-mail at the
[email protected] .
13
High Purity Nitrogen Generator
To ensure proper ventilation, a clear space of at least 20cm is necessary at the back of
the instrument. The Air cooling admission is located on the lower front panel and
should under no circumstances be obstructed.
Make the connection to the consumer after the purge cycle described thereafter
The instrument pressure outlet is adjusted by the pressure regulator placed under the lower panel and
can be visualized permanently on the screen.
CAUTION : your High Purity N2 Generator was tested for several hours at the factory and all its tubes
have been cleaned of all ambient air contaminants. After a break of several days to several weeks of
operation, Air ambient has slightly dirty the circuit of the instrument. The walls of the tubes need again
to be cleaned. A purge cycle must be performed as described in the next chapter.
14
High Purity Nitrogen Generator
4.6.1 Start-up
Turn the instrument on with the switch placed on the back of the instrument. The identification image
appears during approximately 30 seconds the following information will be shown, the maker, model,
serial number as well as the software revision.
LNI SCHMIDLIN SA
Ultra Pure Nitrogen Station
Note : In order to facilitate the first start-up in your laboratory, the High Purity N2 Generator has been
set in the factory for an automatic start-up and produce a pressure of 5 bar.
II RUNNING Menu
It is possible to check the filling up progression of tank by pressing on MENU then select Diagnostic
and Internal Pressure.
The image present the tank pressure, the outlet pressure, the maximum pressure (alarm), and the
functioning state of the compressor.
Leave the instrument continuously working between 4 to 6 hours, limiting the outflow to about 100
NMl/min. The outflow could be limited with a restrictor or needle valve connected on the NITROGEN
output.
From this moment, it is recommended to connect the outlet tube to the consumer and
proceed to the pressure setting.
15
High Purity Nitrogen Generator
Note : The setting procedure of the outlet pressure automatically execute a rough setting (+/- 0.5bar)
of the alarms pressure setting. A different or finer setting can be done through the function N2
PRESSURE ADJUSTMENT.
At this stage, the instrument is working with its factory settings. In order to improve your skills on
performance, software, maintenance period, alarms and trouble shooting, we recommend that you
read the further chapters and paragraphs of this document.
This is the main menu allowing to set the instrument in mode : Generation (running) or stand-by.
Before generating some NITROGEN, it is useful to adjust some settings of the instrument. By pressing
« MENU » it is possible to access to the following menus :
Main Menu
=========================================
↑
16
High Purity Nitrogen Generator
N2 Pressure Setting
↓
ESC 1/6
The First menu « N2 pressure setting » (setting of the NITROGEN output pressure) allows as its
name indicate to adjust the N2 output pressure.
↑
Diagnostics
↓
ESC 2/6
Other menus are :
Tools :
Providing access to various tests including the compressor test
Main Menu
=========================================
↑
Tools
↓
ESC 3/6
Auxilary functions :
17
High Purity Nitrogen Generator
Allows managing the screen and clock as well as setting the working units.
Main Menu
=========================================
↑
Auxiliary Functions
↓
ESC 4/6
Special functions :
Which allow to set the restart mode in case of power cut as well as the remote control and the mode
alarms.
Main Menu
=========================================
↑
Special Functions
↓
ESC 5/6
↑
Alarm + Events
↓
ESC 6/6
18
High Purity Nitrogen Generator
Oven Temperature : 20 °C
Once the pressure has been selected (see next chapter), just press on « I> » to start the generation.
Then the following window appears:
II Running Menu
19
High Purity Nitrogen Generator
Alarm
=========================================
Alarm 1
Alarm 2
…
…
…
ESC
Shortcut :
The center of the screen below is an active area to directly reach the pressure adjustment window.
20
High Purity Nitrogen Generator
Main Menu
=========================================
↑
Air Pressure Setting
↓
ESC 1/6
N2 Pressure Setting
=========================================
4.8 bar
IIIIIIIIIIIIIIIIIII
Touch to memorize
ESC 1/1
ESC
Main Menu
=========================================
21
High Purity Nitrogen Generator
↑
Diagnostics
↓
ESC 2/6
Diagnostics Diagnostics
========================================= =========================================
↑ ↑
Air Pressure Case Temperature
↓ ↓
ESC 1/5 ESC 4/5
Diagnostics Diagnostics
========================================= =========================================
↑ ↑
Oven Temperature (option) Device Features
↓ ↓
ESC 2/5 ESC 5/5
Diagnostics
=========================================
↑
Functioning Time
↓
ESC 3/5
This menu allows only to observe the various functioning parameters of the instrument.
Moving within the tree is done by using the arrows « ↑ » and « ↓ ». Once the requested setting is
found, press on « » to enter into the sub-menu and « ESC » to come out.
This pressure is the outlet generator pressure, which will be applied on to the installation
22
High Purity Nitrogen Generator
This value indicate the functioning time of the instrument as well as the next maintenance.
Once the « remaining time » reaches zero, an indication « Alarm » appears on the screen main
menu :
23
High Purity Nitrogen Generator
When the maintenance is done, go back to the menu « Diagnostic »\ « Total functioning time » and
press on « Maintenance done »
At this time « maintenance done » disappear, the messages « Next maintenance » is incremented by
4000H and the message « Remaining time » is up dated.
Moreover, the message « Alarm » on the main menu disappear.
This sub menu allows to view the instrument model, the serial number as well as the version of the
software.
24
High Purity Nitrogen Generator
Main Menu
=========================================
↑
Tools
ESC 3/6
Tools Tools
========================================= =========================================
↑ ↑
Valve Test Purge Test
↓ ↓
Tools Tools
========================================= =========================================
↑ ↑
Oven Test (option) Pump Test
↓ ↓
25
High Purity Nitrogen Generator
↑
Valve Test Are you sure to enter ?
↓
By pressing on « » It is possible to go into the menu and adjust the setting value
ESC ESC
This test allows direct access to the catalysis oven setting which can be modified as requested.
WARNING: in order to allow for a very different setting to factory default setting, this feature is little protected
against excessive adjustment values. A constant temperature setting above 750 ° C temperature can damage the
instrument.
26
High Purity Nitrogen Generator
↑
Oven Test Are you sure to enter ?
↓
Oven Test
=========================================
++ +
Regulate oven to 700°C
--
This test allows opening and closing the purge valve of the coalescent filter.
↑
Purge Test Are you sure to enter?
↓
By pressing on key « » it is possible to enter into the menu and opening or closing the purge.
27
High Purity Nitrogen Generator
ESC ESC
↑
Pump Test Are you sure to enter?
↓
By pressing on key « » It’s possible to enter into the menu and control the start of the compressor as
well as checking the Air pressure variation inside the tank.
Pump Test
=========================================
Tank pressure value : 3.5 bar
Touch to Stop Pump
IIIIIIIIIIIIIIIIIIIII 72%
ESC
Main Menu
=========================================
↑
Auxiliary Functions
↓
ESC 5/6
28
High Purity Nitrogen Generator
↑ ↑
Display Settings Touch Pad Calibration
↓ ↓
↑ ↑
Unit Change Sounds Menu
↓ ↓
This sub menu is used to set the time and date of the instrument.
Auxiliary Functions
=========================================
↑
Clock Settings
↓
ESC 1/6
++ + ++ +
2009 10
- - - - - -
29
High Purity Nitrogen Generator
+ ++ +
12 27
- - - -
++ + 17/12/2009
17 10 :27
- - -
ESC Day
Once entered in the menu « Clock setting », the software requires in turn the year (Year), the month
(Month),the day (Day), the time (hour) and the minute (Min). Once in the menu, the « + » and the
« ++ » allows to increase the value, the « - » and the « - - » allows to decrease the value. The « ++ »
and « - - » allow a rapid scroll of the values, the « + » and « - » allows a slow scroll of values.
The entered value is validated by pressing « ». The software automatically advances to the next
value.
Once all the values are entered the last "page" displays the date and time set.
↑ ++ +
Display Settings IIIIIIIIIIIIIIIIIIIIIII
↓ - - -
Display Settings
=========================================
++ +
IIIIIIIIIIIIIIIIIIIIIII
- - -
ESC Backlight
This sub menu allows to set the contrast (Contrast) as well as the reverse light screen (light back).
Proceed adjustments with buttons « + », « ++ » and « - » and « - - »
The value entered is validated be pressing on « ». The software automatically advances to the next
value.
4.7.5.3 Sub- Menu « Unit Change »
This menu allows choosing the working units
Auxiliary Functions
=========================================
↑
Unit Change
↓
ESC 3/6
31
High Purity Nitrogen Generator
Once in the « UNIT » menu it is possible with the keys « ↑ » and « ↓ » to change the following units :
Pressure and Temperature.
Once in the requested window (Pressure or Temperature) press on « » and choose with the arrows
« ↑ » and « ↓ » then validate with « » the requested unit.
For the pressure, the units are: Bar, Kilo Pascal and PSI
For the temperature the units are: Degree Celsius and Fahrenheit
↑ ↑
Alarm functions Buzzer ON
↓ ↓
Value recorded
This sub menu allows enabling or disabling the sound signal in case of alarm (Buzzer).
Once in the menu, choose between “ON” or “OFF” with the arrows then valid with
« ».
32
High Purity Nitrogen Generator
ESC 5/6
Calibration
Calibration
Calibration
Once in the sub menu « touch pad setting » the software ask to press on the screen (Touch the pad).
After having press on the screen, the software ask to press on the square placed on the top left hand
corner and to keep the finger on it (touch black square). At that time a counter starts up. Once
arrived at the value 10, the next page ask to remove the finger (remove finger) appears. Then another
calibration point is requested (bottom right hand corner). When done, the message « calibration
done » (Pad Calibration done) appears.
↑ ↑
Sounds Menu IIIIIIIIIIIIIIIIIIIIIII
↓ 50 % ↓
33
High Purity Nitrogen Generator
This sub menu lets you adjust the duration of the sound emitted when the screen is touched. A setting to 0 %
totally removes the sound.
↑
Special Functions
↓
ESC 5/6
↑ ↑
Power Up Control Pressure Level Alarm
↓ ↓
↑ ↑
Remote Control Service Alarm Settings
↓ ↓
↑ ↑
Log Book download Pass Word
↓ ↓
34
High Purity Nitrogen Generator
This menu is used to determine, in the event of power failure, in which state the instrument must
restart. There are two possibilities:
Restart in the same Status it was before the power failure (On Last Running Status)
Restart and go in « stand-By » Status (On Standby Status)
↑ ↑
Power Up Control On Standby Status
↓ ↓
Power Up Control
=========================================
↑
On Last Running Status
↓
ESC 2/2
Special Functions
=========================================
↑
Remote Control
↓
ESC 2/6
35
High Purity Nitrogen Generator
This menu allows you to download the log book on a PC in order to send it to the after sale service for
a help in the diagnosis.
↑
Log Book Download Please Connect to a PC
↓
36
High Purity Nitrogen Generator
37
High Purity Nitrogen Generator
Please Connect to a PC
ESC
OK
ESC 4/6
38
High Purity Nitrogen Generator
++ +
IIIIIIIIIIIIIIIIIIIIIII
- - 4.2 bar -
↑ ↑
Service Alarm Settings 000000
↓ ↓
↑ ↑
Pass Word 000000
↓ ↓
This sub menu allows access to the function « USINE » is not accessible to persons who have not
received adequate training.
39
High Purity Nitrogen Generator
40
High Purity Nitrogen Generator
Chapter 5.Maintenance
Open the lower part of front face, in order to access to the inlet filter.
Unscrew the filter and replace it by a new one
Pressure regulator’s
adjustment button
Inlet filter
To replace the coalescing filter and silencer, we recommend to remove the Air purifier module:
Unscrew the lateral protection plates on both sides
Disconnect the cable and tubes on the rear face
Unscrew the 4 screws which fix the drawer
41
High Purity Nitrogen Generator
Coalescing filter
Coalescing filter (Purge V5)
The following table lists the available YL Instruments items for light maintenance on the NITROGEN
generator (High Purity N2), as well as the recommended accessories::
42
High Purity Nitrogen Generator
IMPORTANT!
The manufacturer reserves the right to change or modify its products without prior notice.
In the event your instrument must be returned to the factory for a repair which cannot be carried out on
the spot, please contact your agent by specifying in detail the problem as well as the instrument
operating environment, the type and the serial number.
This done, your agent will provide all the necessary instructions to return the instrument in the best
conditions.
In case where the warranty period would have expired, or in case the default would be due to misuse
of the instrument, all transport and repair costs incurred would be the sole responsibility of the user,
unless special arrangement between the manufacturer and the client had been undertaken.
Make sure that the instrument will never be exposed during transport or storage, to a
temperature below - 5 ° C.
43
High Purity Nitrogen Generator
6.1 Troubleshooting
The following table describes major cases of malfunctions which the user will be able to fix without the
help of trained technicians on this product. In case of more complex failures, the modular concept of
the LINEA10 line, allow to do a standard exchange of the faulty module without having to send back
the full instrument to the retailer or the factory. Refer to the service manual.
44
High Purity Nitrogen Generator
Oven temperature over 1) The heating cartridge is probably melted. The temperature sensor
750°C (thermocouple K-type) inside is out of order. Verify voltage by
“Diagnostics” Menu and eventually wiring. If heating cartridge is
melted, the complete oven must be replaced.
45