Upgrading Instruction

Please copy the original RT1904C+ System Software and save them for future use
before starting the upgrading.

You need a SD card, a SD card reader and a USB mouse to complete the upgrade.
Please kindly follow the instruction step by step to upgrade.

STEP:
1. Save the system software and the software module 20060122.exe in your SD card as
figure 1. (Don’t change the software name, or the instrument can’t use the software.)

Figure 1
2. Put the SD card into the instrument. Then turn on the instrument.

3. Turn on the instrument, you will find the WinCE menu:


4. Double click My Computer then open the Storage Card folder.

Note : If you can not find "residentflash" folder, please cancel this procedure and
follow "Format the resident flash.doc". After solve this issue, please follow
this instruction to upgrade software again.
5. Put the mouse cursor on the “RT1904”, then click right button of the mouse and copy the
“RT1904” file.

6. Double click “ResidentFlash” folder.


7. Select and delete all the files in the folder. Click Yes and then click Yes to All. Later Paste
the software in the folder (paste the system file “RT9000” into “ResidentFlash” folder).
Note: Sometimes the instrument report flash memory is not enough, you need to follow
"Fail to find flash disk.doc" to upgrade WinCE again and then upgrade operation
software.
8. Press Start button then select Suspend and turn off the switch directly.

9. Software Upgrading is finished.


10. Turn on the instrument and press aspiration key to get into the test menu.

11. When you get into the test interface, please click Main button then get into the main menu
and then get into the System Setting interface.

12. Click the position (just right under “ly” of “instantly) within red ring several times to get into
the back door of software.

13. Input ID (e.g.40050000), Hospital (hospital name), T value, A value, and Pump speed;
Pump, Front board, Gain calibration, and Pump calibration in the back door of software.

Note: If you do not have the original parameters, please send the SN of this instrument
to us and we will give the parameters to you.