Install Printer on Windows 10

TIP: This procedure is for Windows 10 only.


TIP: Check first if the printer has not already been automatically installed by the Windows
system.
TIP: Use only PCL5e or PCL 6 and higher compatible USB printer.
TIP: Older printer types might not be supported by Windows 10 anymore and can not be
installed.
******************************************************************************
*****************************************
Installation procedure with Control Unit HP rp5810
TIP: The installation of the printer can be done as user “cobasinstall” or “cobasoperator”
Installation procedure with user “cobasoperator”
1.
Make sure the user software is started and user “ADMIN” is logged in.
2.
Connect the printer to an UBS port on the control unit and switch ON the printer.
3.
The pinter will be automatically installed.
4.
Go to <Configuration>
5.
Choose <Printers>
6.
Check if the installed printer is shown.
Installation procedure with user “cobasinstall”
1.
Log on as user “cobasinstall”
2.
Connect the printer to an UBS port on the control unit and switch ON the printer.
3.
The pinter will be automatically installed.
4.
Choose the “Windows” key to display the task bar.
5.
Choose <Settings>
6.
Choose <Devices>
7.
Choose <Printers & Scanners
8.
Check if the installed printer is shown.
******************************************************************************
*****************************************
Installation procedure with HP printers on a Control Unit HP Flex Pro
TIP: The installation of HP printers will be done automatically.
TIP: Older HP printer models might not be supported by the Windows 10 anymore!

Installation procedure with user “cobasoperator” with HP printers


TIP: The installation of the printer as user “cobasoperator” can only be done with HP
printers!
1.
Make sure the user software is started and user “ADMIN” is logged in.
2.
Connect the printer to an UBS port on the control unit and switch ON the printer.
3.
The pinter will be automatically installed.
4.
Go to <Configuration>

5.
Choose <Printers>
6.
Check if the installed printer is shown.
Configuration
7.
Set the desired printer as “Set as Default Printer” with right-mouse click.
Set_default
8.
Print a test page to confirm the correct configuration of the printer.
Installation procedure with user “cobasinstall”
1.
Log on as user “cobasinstall”
2.
Connect the printer to an UBS port on the control unit and switch ON the printer.
3.
The pinter will be automatically installed.
4.
Press the “Windows” key to display the task bar.
5.
Choose <Settings>
6.
Choose <Devices>
7.
Choose <Printers & Scanners
8.
Check if the installed printer is shown.
TIP: The setup might me shown as incomplete. The printer still works and do NOT
connect to the Internet.
HP_Printer_cobasinstall
9.
Select the printer and chose <Manage>.
HP_Printer_Manage
10.
Set the printer as default and print a test page.

HP_Printer_Default
11.
Restart the control unit.
******************************************************************************
*****************************************
Installation procedure with NON HP printers on a Control Unit HP Flex Pro
TIP: For the installation of a NON HP printer the driver of the printer is required and needs to
be installed manually as user “cobasinstall”.
TIP: Older printer types might be not supported by the operating system and can not be
installed anymore.
1.
Log on as user “cobasinstall”.
2.
Connect the printer to an UBS port on the control unit and switch ON the printer.
3.
The printer will be recognized by the system but will not be installed automatically.
4.
Press the “Windows” key to display the task bar.
5.
Choose <Settings>

6.
Choose <Devices>
7.
Choose <Printers & scanners>.
8.
Choose <Add a printer or scanner>.
Add_Printer
9.
The printer will not be found and needs to be installed manually. Choose <The printer
that I want wasn’t listed>.
Printer_not_listed
10.
Choose <Add a local printer or network printer with manual settings> and click <Next>.

Local_Printer
11.
Choose the “Virtual printer port for USB” from the list and click <Next>.
USB_Port
12.
Insert the media with the corresponding printer driver and choose <Have Disk...>
Install_driver
13.
Browse to the printer driver and choose <Open>.
TIP: Only printer drivers with the ending .inf can be installed.
14.
Follow the instructions on the screen.
15.
After the installation of the printer, choose <Do not share this printer> and click
<Next>.
Share_Printer
16.
Print a test page and click <Finish>.
17.
Set the printer as default by choosing the desired printer and choose <Manage>.
Manage
18.
Choose <Set as default> and close the window.
Set_default_Lexmark
19.
Restart the control unit.

rp5810 against Flex Pro


General Information
A replacement of a failed Control Unit is necessary if one of the following parts failed: For
example CPU Board, Power Supply of the Control Unit etc.
TIP: Describes a replacement of a defective control unit HP rp5810 against a HP Flex Pro.
An automatic migration of the database from the original drive D: can not be
performed. To keep the most recent database, a manual copy of a BOD_Export from drive D:
must be done, which will be imported afterwards. Be aware, depending on the workflow
performed after the BOD Export, data might get lost and the counters of the reagent cassettes
are not up to date anymore afterwards.
If development channels are active on the system, the dc.lic file needs to be
manually copied from the defective control unit if possible. If an access to drive C: is still
possible, browse to C:WINDOWS and copy the dc.lic file to an external media.
If the dc.lic file can not be copied and is not available anymore, contact GCS to get the dc.file
again.
Control Unit HP rp5810 Front Plate HP rp5810
Control Unit HP Flex Pro Front Plate HP Flex Pro
Parts required
CONTROL UNIT I400 PLUS WIN10 FLEX PRO (09241370001)
SW SYST 3.6.1 I400 plus Win10 CD (08791457001)
SW SYST 3.6.2 I400 plus Win10 CD (09239626001)
Procedure
1.
Make sure the instrument and the control unit are switched OFF.
2.
Disconnect all cables from the defective control unit.
3.
Remove the cover of the defective control unit by pulling the lever.
4.
Lift the power supply to get access to drive D:
5.
Unplug the data and power cable from drive D:
Hard disk D: is below the power supply: Light blue connector on the motherboard
(SATA1).
6.
Remove hard disk D: by pressing green plastic lever and place it aside for later use.
7.
Replace the defective control unit with the new control unit.
8.
Connect all cables to the new control unit.
Keyboard/Barcode Pen or Handheld Barcode Scanner
Mouse
LCD Display Cable (Monitor)
HIF serial cable COM 1 (Host, if installed)
Printer cable USB (if installed)
Second LAN connection (Remote service, if installed)
Cable twisted-pair RJ 45 (Instrument LAN)
Cable Serial DB9 M/F COM 2 (Instrument)

Flex_Pro_Back
9.
Switch ON the instrument and control unit.
10.
TIP: Only operating system Windows 10 is pre-installed on the new control unit.

Install System Software Version 3.6


11.
Check if the ISE power supply switch is ON.
ISE switch instruments serial-no below 420000/520000 f568
ISE switch instruments serial-no higher 420000/520000 ISE Switch Ext
12.
Check that no racks are on board.
13.
Insert the USB flash drive or CD with system software version 3.6 and wait until the
LED of the USB flash drive or DVD drive stops flashing.
14.
Log on as “cobasinstall” for user and enter the corresponding password.
15.
Press the “Windows” and “e” key together to open the Windows file explorer. Browse
to drive I400 Analyzer SW (E:) or "USB flash drive" (G:) to see the content of USB flash
drive or DVD drive. Double click on the file <Analyzer Software>.
TIP: Start this file only once. It takes a few moments until the software installation
starts.
16.
In the "Welcome" window click <Next>.
17.
Accept the Software License Agreement with <Yes>.
18.
Select the Language Group for your area and press afterwards next
- Western Languages (English, French, German, Italian, Portuguese and Spanish)
- Eastern European language (Polish language only)
19.
Start the installation by choosing <Install>.The installation starts and takes about 12
minutes.
20.
Complete the installation by choosing <Finish>.
21.
The "Bootmanager Process Entry" window appears (it might be hidden behind the
Explorer window).
Enter customer information in the text fields (examples: hospital name, instrument
name, serial number etc.) and choose <OK>.
TIP: It is necessary to write at least one character into one field.
TIP: Steps 22 and 23 are only valid if the ID Module was empty before.
22.
When prompted enter “CN” into the SMP field and the instrument serial number.
23.
The next window asks for components that are installed. Check the appropriate boxes
for FP and ISE.
24.
The control unit restarts and the instrument will initialize.
TIP: A warning will appear (Message ID 17.4453) because of missing calibration data.
Close it!
25.
Remove the USB flash drive or CD.
26.
Wait until the instrument has reached the ‘Standby’ mode. It may happen that
immediately BOD service actions are started. Stop these BOD actions. Log on as
‘ADMIN’ for user and enter the corresponding password. Select the ‘Service’ icon.
Within the ‘Service’ screen select the tab ‘All’. Select the Service Action ‘Backup
Database’ and click ‘Perform’. The window ‘Confirmation for database backup’ pops
up. During this initial "Backup database" a message window will appear:
"This database backup is part of the procedure to:
a) "Replace a failed C: disk" or
b) "Setup a new system"?
Select b)"Setup a new system" and press <Continue>.
27.
Shut down the control unit and remove the power cable.
28.
Remove the cover of the control unit Flex Pro by pulling the lever.
29.
Unplug the power and data cable from the DVD drive.
CU_overview
30.
Remove the short connection cable from the data cable.
Cable_DVD
31.
Connect the hard disk D: from the defective control unit to the connectors of the DVD
drive.
32.
Connect the power cable to the control unit and switch ON the control unit.
33.
When the Windows screen appear press and hold the “Shift” key.
34.
When the log on window appears log on as “cobasservice” for user and enter the
corresponding password.
35.
Press the “Windows” and “E” key to open the Windows file explorer and browse to the
drive UserDisk (G:).
36.
Copy the five folders DB_Export_Backup to an external media.
37.
Shut down the control unit and remove the power cable of the control unit.
38.
Unplug the additional connected drive D:, reconnect the short connection cable and
the DVD drive.
39.
Put back the cover of the control unit and switch it ON.
40.
When the Windows screen appear press and hold the “Shift” key.
41.
When the log on window appears log on as “cobasservice” for user and enter the
corresponding password.
42.
Perform a education import of the most recent DB_Export_Backup from the external
media.
Refer to Education Import (Database Import)
43.
An initial backup is performed. This may take up to 5 minutes.
44.
Check if the newest TAS version is installed on the system.
TIP: With software version 3.6 the default TAS file set needs to be executed
manually.
45.
TIP: Specific Windows settings are not stored and are set now the standard settings
from Windows.
Refer to Installation -> Configuration Setting
Refer to Install Printer on Windows 10
46.
If the dc.lic file for development channels could be copied from the defective control
unit, copy the file to folder C:\Roche\Instrument\DC-License