BSL-2 Facility in a mobile container required


Clean room of BSL equired specifications:

 Dimensions of container Length 20 20-25 ft Width 8-10 ft Height 8-9 ft


 Flooring: Vinyl or Epoxy coated 2 2-4mm
4mm thickness for easy cleaning
 Wall & Ceiling: Standard wall partitions with two skins of PPGI/PPGI (0.6-1mm
thickness) over GI frame work with a sealed and insulated interior or Fabricated
with 2-3 mm CRCA SheetSheet. Should be painted with anti-microbial
anti Epoxy paint
artitions and false ceiling should create an impervious shell, which form a fully
Partitions
sealed structure.

 Covings for entire wall to floor, wall to wall & wall to ceiling joints to be provided.
 Airtight clean room doors: Flush Door finishes to be 50 mm thick 1.2-1.4 1.2 mm
thick PCGI sheet suitable to fix on 50 mm thick wall panel with provisions for
double glazing glass for all door. Dimensions (900-950) 950) x (2100-2200).
(2100 Numbers
3-4 (Single leaf)
 Room pressure maintaining system: Change room pressure should be
maintained at 10 10-15 pascal while BSL-2 Lab should be maintained at 25-30
pascal.
 HVAC system or AHU with HEPA filter (Retention: 0.3 Micron Efficiency:
99.997% Pressure drop: 25 mm of W Water column) with booster fan, AC (Nos 3),
condenser, blower.
 A Fumigator needs to be provided
 Electricals: All the required cabling, switches
switches,, arrangements etc. for the purpose
of energizing the BSL BSL-2+ Laboratory facility needs to be provided. All the
electrical fittings and fixtures in the laboratory areas should
sh be suitable for clean
room application and shall be sealedsealed. Internal sockets & lights
li wiring with 3
Phase Raw power / Genset input supply supply.
 Loft water tank 01 and effluent water tank 01 to be provided
 The effluent water tank should be treatable so that when the effluent goes to
ground water it is free from contaminants.
 Stainless steel el sink (2’×2’) with bottom enclosure and steel eye wash assembly
to be provided
 Fire alarm system to be provided
 10KVa UPS with battery should be provided.
 Miscellaneous items to be provided: The change room should have shoe racks
(2), Shoe cover dispen ser unit (1), stainless steel gowning cupboard (600-
dispenser (
650mmmm W X300mm D X 1500 1500-2000 mm H with UV Lamp), stainless steel paddle
dustbin, used linen bin and an autoclave (50L).
The BSL-2 2 main Laboratory should have

Stainless steel work table (5


(5-7’ × 3-4’) with storage facility Nos 2, stainless steel
paddle dustbin Nos 2, stainless steel stool Nos 4. Stainless steel wall racks.

 The company should do the installations, validation and provide 5 years


comprehensive warranty on the whole assembly and associated components.