Professional Documents
Culture Documents
Acid-Soluble Chloride in Mortar and Concrete: Standard Test Method For
Acid-Soluble Chloride in Mortar and Concrete: Standard Test Method For
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C1152/C1152M − 20
Copyright by ASTM Int'l (all rights reserved); Tue Jul 13 17:20:53 EDT 2021
2
Downloaded/printed by
La Universidad de los Andes (La Universidad de los Andes) pursuant to License Agreement. No further reproductions authorized.
C1152/C1152M − 20
7.1.3 Powdered concrete obtained by use of a rotary impact NOTE 7—It is not necessary to clean all the slurry residue from the sides
drill is frequently used in determining chloride concentration of the beaker, nor is it necessary that the filter remove all the fine material.
The titration may take place in a solution that contains a small amount of
with depth in bridge decks, pavements, etc. Such samples may solid matter.
be unrepresentative, especially when the nominal maximum
coarse aggregate size is 25 mm (1 in.) or more. Thus, several 9.3 For instruments equipped with a dial readout it is
such samples should be combined, or the data used with care. necessary to establish an approximate “equivalence point” by
Procedures for this method of sampling are as follows: immersing the electrodes in a beaker of water and adjusting the
7.1.3.1 Using the rotary impact drill, drill perpendicular to instrument to read about 20 mV lower than mid-scale. Record
the concrete surface or parallel to the axis of a cored specimen the approximate millivoltmeter reading. Remove the beaker
to a specified depth or a depth sufficient to obtain a represen- and wipe the electrodes with absorbent paper.
tative sample of the concrete mixture of at least 20 g of 9.4 Pipet 2.00 mL of standard 0.05 N NaCl solution into the
powdered material. To prevent sample contamination, avoid cooled sample beaker (see Note 8). Place the beaker on a
contact of sample with hands and other sources of perspiration. magnetic stirrer and add a TFE-fluorocarbon coated stir bar.
Clean all sampling tools prior to each sampling operation (see Immerse the electrodes into the solution, taking care that the
Note 4). No lubricants shall be used during drilling. stir bar does not strike the electrodes. Begin stirring gently.
NOTE 4—Sampling tools may be cleaned with a brush, cloth, ethyl Place the delivery tip of the 10 mL buret filled to the zero mark
alcohol rinse, water rinse, or other method that will not contaminate the with standardized 0.05 N AgNO3 in (preferably) or above the
sample. sample solution (see Note 9).
7.1.3.2 Transfer powdered sample into sample container NOTE 8—It is advisable to maintain constant temperature during
using a spoon or other suitable means. measurement, as the solubility of silver chloride varies markedly with
temperature at low concentrations.
8. Sample Preparation NOTE 9—If the tip of the buret is out of the solution, any adhering
droplet should be rinsed into the beaker with a few millilitres of water
8.1 Pulverize the sample so that all the material will pass a following each titration increment.
850-µm (No. 20) sieve. Thoroughly blend the material by
9.5 Gradually titrate, record the amount of standard 0.05 N
coning as described in Practice C702/C702M from one glazed
silver nitrate solution required to bring the millivoltmeter
paper to another at least 10 times.
reading to –60.0 mV of the equivalence point determined in the
water.
9. Procedure
9.6 Continue the titration with 0.20 mL increments. Record
9.1 Determine the mass of approximately 10 g of sample to
the buret reading and the corresponding millivoltmeter reading.
the nearest 0.01 g and transfer to a 250 mL beaker. Disperse the
Allow sufficient time between each addition for the electrodes
sample with 75 mL of water. Slowly add 25 mL of dilute (1+1)
to reach equilibrium with the sample solution (see Note 10).
nitric acid and stir with a glass rod, breaking up any lumps of
sample. If the smell of hydrogen sulfide is strongly present, add NOTE 10—Experience has shown that acceptable readings are obtained
3 mL of hydrogen peroxide (see Note 5). Add 3 drops of when the minimum scale reading does not change within a 5-s period.
methyl orange indicator to the beaker and stir. Cover the beaker 9.7 As the equivalence point is approached, the equal
with a watch glass and allow to stand for 1 to 2 minutes. If the additions of silver nitrate solution will cause larger and larger
solution above the settled solids is not pink colored, add more changes in the millivoltmeter readings. Past the equivalence
nitric acid dropwise while stirring until a pink or reddish color point, the change per increment will decrease. Continue to
persists, then add 10 additional drops of nitric acid and stir. titrate until three readings past the approximate equivalence
Heat the covered beaker rapidly to boiling and remove from point have been recorded.
hot plate. Do not allow sample to boil more than a few seconds
(see Note 6). Make a blank determination using 75 mL of water 9.8 Calculate the precise equivalence point of the titration
in place of sample. (see Note 11) and subtract the results of the water blank (see
Note 12)
NOTE 5—Slags and slag cements contain sulfide sulfur in concentra-
tions that can interfere with chloride determination unless oxidized with NOTE 11—An example of recording solution, millivolt readings, and
hydrogen peroxide. calculation of the titration equivalence point can be found in Appendix X1
NOTE 6—To prevent loss of chloride by volatilization, it is important to of Test Methods C114.
keep the beaker covered during heating and digestion. Ten seconds of NOTE 12—For nonreferee analysis, the blank may be omitted.
boiling is sufficient. Excessive amounts of acid can erode the silver
chloride coating on the silver billet electrode prematurely. A slurry that is 10. Calculation
only slightly acidic is sufficient.
9.2 Filter the sample through a 9-cm coarse-textured filter 10.1 Calculate percent chloride by mass of mortar or con-
paper in a 250 mL or 500 mL Buchner funnel and filtration crete to the nearest 0.001 % as follows:
flask using suction. Rinse the beaker and the filter paper twice C1, % 5 3.545@ ~ V 1 2 V 2 ! N # ⁄W
with small portions of water. Transfer the filtrate from the flask where:
to a 250 mL beaker and rinse the flask once with water. The
V1 = millilitres of 0.05 N AgNO3 solution used for sample
original beaker may be used (see Note 7). Cool the filtrate to
titration (equivalence point),
room temperature. The volume should not exceed 175 mL.
Copyright by ASTM Int'l (all rights reserved); Tue Jul 13 17:20:53 EDT 2021
3
Downloaded/printed by
La Universidad de los Andes (La Universidad de los Andes) pursuant to License Agreement. No further reproductions authorized.
C1152/C1152M − 20
V2 = millilitres of 0.05 N AgNO3 solution used for blank 11. Precision and Bias
titration (equivalence point), 11.1 Precision:
N = exact normality of 0.05 N AgNO3 solution, and
W = mass of sample, g. 11.1.1 Single-Operator Precision—The single-operator
standard deviation has been found to be 0.0015 % chloride.
10.2 Other useful measures of chloride concentration can be
Therefore, results of two properly conducted tests by the same
obtained as follows:
10.2.1 For calculating kilograms of chloride per cubic metre operator on the same material are not expected to differ by
[pounds of chloride per cubic yard] of concrete (see Note 13), more than 0.0042 %.6
multiply percent chloride by D1/100 or D2/100 to the nearest 11.1.2 Multilaboratory Precision—The multilaboratory
0.1 kg/m3 [lb/yd3], standard deviation has been found to be 0.0021 %. Therefore,
results of two properly conducted tests by two different
where:
laboratories on specimens of the same material are not ex-
D1 = oven dry density as determined in the section on pected to differ by more than 0.0059 %.6
Concrete Density in Test Method C1084, kg/m3 [lb/
yd3], and NOTE 14—These precision statements apply to tests of samples pre-
D2 = saturated-surface-dry density as determined in the sec- pared and ground by a single laboratory. For this reason, test results will
tion on Concrete Density in Test Method C1084, kg/m3 vary depending upon the method of obtaining the sample and the size of
[lb/yd3]. the sample before it is crushed and reduced to pass the 850-µm (No. 20)
sieve, or pulverized. The effect may be small for large concrete cores or
Report which density was used in the calculation. from mortars sectioned vertically to appreciable for large aggregate
NOTE 13—In the case of drilled samples or where the actual density can concretes sampled by rotary impact drilling with small diameter drills. A
not be determined: U = density of mortar or concrete, kg/m3 [lb/yd3]. The copy of the research report used to develop the precision statements is
air dry density of normal weight concrete for most purposes can be available by request.7
assumed to be 2263 kg/m3 [3815 lb/yd3].
11.2 Bias—Since there is no accepted reference material
10.2.2 For calculating percent chloride by mass of cement, suitable for determining the bias for the procedure in this test
multiply percent chloride by 100/P, method, no statement on bias is being made.
where:
P = percent cement by mass in the mortar or concrete, as
known or determined in accordance with Test Method
C1084. 6
These numbers represent the difference limits (d2s) as described in Practice
C670.
10.2.3 For calculating equivalent flake calcium chloride 7
Supporting data have been filed at ASTM International Headquarters and may
(calcium chloride dihydrate) concentrations, multiply percent be obtained by requesting Research Report RR:C09-1003. Contact ASTM Customer
chloride by 2.07. Service at [email protected].
SUMMARY OF CHANGES
Committee C09 has identified the location of selected changes to this standard since the last issue
(C1152/C1152M – 04 (2012)ε1) that may impact the use of this standard. (Approved June 1, 2020.)
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
Copyright by ASTM Int'l (all rights reserved); Tue Jul 13 17:20:53 EDT 2021
4
Downloaded/printed by
La Universidad de los Andes (La Universidad de los Andes) pursuant to License Agreement. No further reproductions authorized.