Revco ULT2586
Revco ULT2586
Table of Contents
4 Pre-Installation .................................................................... 2
4.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
5 Installation ........................................................................ 2
5.1 Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
5.2 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
5.3 Leveling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
6.2 Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6.2.1 Turning the Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6.2.2 Setting the Cabinet Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6.2.3 Setting the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6.2.4 Alarm Reset & Status Lights (Revco Ultima II and Elite Freezers) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
6.2.5 Alarm Test (Revco Ultima II and Elite Freezers) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
i
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
7.2 Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
7.2.1 Turning the Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
7.2.2 Setting the Cabinet Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
7.2.3 Optional Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
10 Optional Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
10.1Backup Protection Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
10.1.1 Liquid Nitrogen (LN2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
10.1.2 Liquid CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
10.1.3 Voltage Safeguard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
11 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
ii
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
In this manual and on labels attached to this product, the words Avoid opening the door for extended time periods since chamber
WARNING and CAUTION mean the following: temperature air will escape rapidly. Room air, which is higher in
humidity, replacing chamber air may cause frost to develop in the
• WARNING: a potentially hazardous situation which, if not
avoided, could result in serious injury or death.
chamber more rapidly.
• CAUTION: a potentially hazardous situation which, if not 3.3 Initial Loading
avoided, may result in minor or moderate injury or damage to
the equipment. After reading and completing the Safety Considerations,
Pre-Installation, Installation, and Operation sections of this
Before installing, using or maintaining this product, please be sure
to read this manual and product warning labels carefully. Failure to
manual, turn the Key Switch to the POWER ON position and
follow these instructions may cause this product to malfunction, adjust the temperature setpoint if necessary.
which could result in injury or damage. The setpoint should be no warmer than -50ºC for two-stage
Below are important safety precautions that apply to this product: freezers or -10ºC for single-stage freezers. Allow the freezer to
• Use this product only in the way described in the product
operate at the desired temperature for a minimum of 12
literature and in this manual. Before using it, verify that this hours before loading.
product is suitable for its intended use. If your freezer is an upright model, load the freezer one shelf at a
• Do not modify system components, especially the controller. time, beginning with the top shelf. After loading each shelf,
Use OEM exact replacement equipment or parts. Before use, allow the freezer to recover to the desired setpoint before loading
confirm that the product has not been altered in any way. the next shelf. Repeat this process until the freezer is fully
• Your unit must be properly grounded in conformity with loaded.
national and local electrical codes. Never connect the unit to
overloaded power sources.
CAUTION! Failure to follow these procedures or
• Disconnect the unit from all power sources before cleaning, overloading the unit may cause undue stress on the
troubleshooting, or performing other maintenance on the compressors or jeopardize user product safety.
product or its controls.
1
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
2
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
3
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
BACKUP SYSTEM
SUPPLY SUPPLY BATTERY
check ON EMPTY LOW
fuse
Ultima cabinet
life-guard alert extreme ambient alert
temperature
off power on control
power failure set point clean filter
17
1 23 7 8 3 4 5 9 10 14 15 19 18
Figure 2. Revco Ultima II Series Control Panel with Optional Backup System
Before the initial start up, take some time to become familiar 9. Cold Alarm Setpoint. Press this pad to set the cold alarm
with the controls on your freezer. Figure 2 illustrates the Revco (refer to Section 6.2.3 on page 5).
Ultima II Chest and Upright Freezer control panel. 10. Warm Alarm Setpoint. Press this pad to set the warm alarm
(refer to Section 6.2.3 on page 5).
Note: Figure 2 shows all possible functions. These functions 11. Control Setpoint. Press this pad to set the cabinet temperature
vary depending on the model and options. You may not (refer to Section 6.2.2 on page 5).
have all of the functions on your equipment. Revco Elite 12. Cabinet Temperature Indicator. This indicator lights when the
models do not include the Check Fuse indicator light (6) temperature display window is showing the cabinet
or the Voltage Boost light (19) as these models do not temperature (refer to Section 6.2.1 on page 5).
come standard with these options. Likewise, only Revco 13. Digital Temperature Display Window. This window displays
Ultima II models will include the Set Point Security chamber temperature, alarm values, etc. depending on the
feature (23), Life-Guard Alert indicator light (5), and the operating status of the freezer and the procedure being
Extreme Ambient Alert indicator light (16). The CO2 or performed.
LN2 Backup System lights (20, 21, 22) represent status 14. Decrement Pad ( ). Use this pad to decrease temperature
indicators for a built-in CO2 or LN2 backup system, values.
which is optional on Revco Ultima II and Elite models. 15. Increment Pad ( ). Use this pad to increase temperature
values.
1. Three position keyed Power On Switch. 16. Extreme Ambient Alert (Revco Ultima II Models only). This
2. Power On Indicator. This indicator lights when power is indicator lights when the ambient temperature has exceeded
connected to the freezer. the upper limit of the recommended operating range. If the
3. Temperature Failure Indicator. This indicator lights when the Extreme Ambient Alert is activated, the ambient
freezer temperature deviates either above or below the alarm environment needs to be improved for proper function of the
temperature settings. freezer.
4. Power Failure Indicator. This indicator lights when there is a 17. Clean Filter Indicator. This indicator lights when the air
power failure to the freezer (refer to Section 8 on page 10). cooled condenser filter is dirty (refer to Section 8 on
5. Life-Guard Alert Indicator (Revco Ultima II Models only.) page 10).
This indicator lights when the first stage compressor has gone 18. Alarm Battery Low Indicator. This indicator lights when the
into a cautionary state of operation to help prevent failure. charge on the alarm battery is low. Press this button to display
6. Check Fuse Light (Revco Ultima II and Ultima Models the percent of full battery charge in the digital display
only). This indicator lights when the surge protector on the window.
main electronic panel is blown by a power surge (refer to 19. Voltage Boost Indicator. This indicator lights when incoming
Section 8 on page 10). voltage is low and the Voltage Safeguard has been activated
7. Alarm Test Pad. Press this pad to start an alarm test (refer to (refer to Section 8 on page 10). Press this pad to display the
Section 5.2.5 on page 6). current line voltage in the digital display window.
8. Alarm Reset Pad. Press this pad to reset slowly blinking 23. Alarm On/Set Point Security Indicator. This indicator lights
indicators. Slowly blinking indicators denote that the when the alarm is activated.Also indicates that Setpoint
condition has occurred but is now within the given operating Security is activated (Revco Ultima II Models only).
parameters. This pad is also depressed to temporarily silence
the audible alarm (refer to Section 6.2.4 on page 5).
4
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
The following indicators are available only when the freezer is 3. Press and hold or to adjust the Warm Alarm Setpoint.
equipped with the optional backup system (refer to Section 6.3.4 4. Release both pads when the digital temperature display
on page 7). window shows the correct setpoint value.
20. Backup System Supply On Indicator. When the cabinet temperature drops below the Warm Alarm
21. Backup System Empty Indicator setting, turn the key switch to ALARM ON. The freezer is ready
22. Backup System Battery Low Indicator. to operate.
6.2 Start Up Note: If a power failure lasting over 30 seconds occurs, the
POWER ON light goes out, the POWER FAILURE light
Refer to Section 6.1 and Figure 2 on page 4 as you complete the and digital temperature display flash simultaneously,
following procedures. and a buzzer sounds.
6.2.1 Turning the Power On WARNING! You must turn the three-position key
switch to the ALARM ON/SETPOINT SECURITY
To start up the Revco Ultima II, Ultima, and Elite Series position to activate the alarm and place it in security
Freezers, complete the following steps: operation.
1. Plug the freezer into the power outlet (refer to Section 4.2 on
page 2). 6.2.4 Alarm Reset & Status Lights (Revco Ultima II and
2. Turn the key switch to the POWER ON position. The Power Elite Freezers)
On LED and the Cabinet Temperature LED illuminate. The alarm reset feature ensures user acknowledgment of the
Note: The alarm function is not active at this time. Refer to
occurrence of certain alarm conditions. This provides greater
Section 5.2.3 for information about setting the alarms. In monitoring power by alerting the user if an alarm condition has
addition, for Revco Ultima II units only, the Setpoint occurred during periods when the freezer must be left
Security will be deactivated in this key position. unattended. The alarm conditions incorporating this feature are:
5
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
6.2.5 Alarm Test (Revco Ultima II and Elite Freezers) Note: Revco Value Series units use free-standing, field installed
Backup Systems.
To test the cold and warm alarms, complete the following steps:
1. Press and hold the Alarm Test pad. The digital temperature 6.3.1 CO2 and LN2 Precautions
display indicates rising temperature. When the temperature
reaches the warm alarm value, the alarm sounds. The following are precautions for using liquid CO2 and LN2
backup systems.
Note: The temperature of the refrigerated space does not
change during this procedure. Only the sensor is heated. WARNING! If a CO2 or LN2 cylinder falls and a
valve is knocked off, the cylinder becomes a deadly
2. Release the Alarm Test pad, the sensor and display return to and completely unguided missile. Transport the
cabinet temperature in a few minutes. cylinders in a handtruck or cart with secure chain ties
for the cylinder. After cylinders are connected to the
You can press the Audio Silence pad to silence the alarm but the equipment, securely attach them with chains to a
alarm will ring back in five to seven minutes. You can silence the solid, stationary object such as a building column.
alarm in this manner as many times as necessary until the sensor
cools below the warm alarm temperature. The alarm quits WARNING! CO2 and LN2 liquids are non-poisonous
automatically when the sensor cools below the warm alarm but are very cold and will burn unprotected skin.
value. Always wear protective eyewear and clothing when
changing cylinders or working on the piping systems
Note: There is a limit on the alarm test feature. The alarm test attached to an active source of liquid refrigerant.
will not commence if the warm alarm setpoint is more
than 15 degrees warmer than the operating temperature
WARNING! The gases produced by evaporation of
setpoint. CO2 or LN2 are non-poisonous but displace the
oxygen in a confined space and can cause
6.3 Backup System (Optional) asphyxiation. Do not store the cylinders in
When you purchase a built-in CO2 or LN2 optional backup subsurface or enclosed areas.
system for the freezer, the backup system control panel is located
behind a hinged panel adjacent to the grill (refer to Figure 3). CAUTION! When closing the cylinder valve, make
sure that the injection solenoid is energized to allow all
Note: Always purchase the cylinders which are equipped with the liquid to bleed off instead of being trapped in the
siphon tubes for withdrawing liquid from the bottom of supply tubing. Failure to do this results in activation of
the cylinder. CO2 cylinders must be kept at room the pressure relief device, which could damage the
temperature to function properly. LN2 bottles are freezer and requires replacing if it is activated.
functional at any reasonable temperature.
CAUTION! For chest and upright models ordered
Use this panel to set the backup initiation temperature, usually with factory installed built-in backup systems, the flow
10 to 15ºC warmer than the operating setpoint. of liquid CO2 or LN2 will be discontinued if the door or
lid is opened during operation of the backup system.
For units operated with free-standing, field installed
Digital Temperature Calibration Pot. type backup system, the flow of liquid CO2 or LN2 will
Display be discontinued upon door or lid opening only if the
switch provided with the free-standing package is
installed on the freezer.
TEMPERATURE °C
POWER PUSH
OFF TO SET
SETPOINT
ON ADJUST
6
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
6.3.2 Installation
Field installed systems are supplied with complete installation CO2 Inlet Conn.
and operating instructions. If your system is factory installed, the Copper Tubing
Factory Supplied
freezer is shipped with a coiled length of tubing to connect the
0.25" O.D.
freezer to the bottles:
• 3/8 in. OD copper tubing for connection to the CO2
supply.
• 5/8 in. OD copper tubing covered with Armaflex®
insulating tubing for connection to the LN2 supply. 2.375"
Straighten the coiled tubing and connect one end to the labeled
connection on the freezer and the other end to the supply bottle or 24.75"
building supply fitting.
6.3.3 Operation
To activate the optional backup system, complete the following
steps:
1. Push the Power Switch to the ON position. The digital MACHINE COMPARTMENT
temperature display window shows the current cabinet REAR VIEW
temperature.
2. Press and hold the Push to Set button to display the setpoint Figure 4. Location of CO2 Connection Site,
temperature in the digital temperature display window. Chest Models Only
3. Use a small screwdriver to rotate the Setpoint Adjust screw
until the correct backup operating temperature shows in the
display window. The recommended setpoint is 10 to 15°C Table 1. Backup System Flow Rates – Cryogenic Freezers
above normal cabinet temperature.
Lbs/hr @ 75×F
Note: The temperature shown in the display window is Ambient Temperature
accurate to ±1°C. (Flow Rate)
Center Air
The backup system can run for a minimum of eight hours on Temperature
-150×C -140×C -130×C -120×C
battery power.
Liquid LN2
WARNING! You must push the Power Switch to the Chest Freezers
ON position in order to activate the backup system. Empty 11.0 9.5 8.4 7.3
5 ft3 1/2 Full 9.3 8.0 7.0 6.1
6.3.4 System Status Lights
Full 6.5 5.9 5.1 4.5
In addition to the backup control system, the built-in backup Empty 11.3 10.0 9.0 8.0
system includes main control panel system status lights. These
7 ft3 1/2 Full 9.7 8.6 7.6 6.8
lights are only available when the backup system is installed.
Full 7.8 6.8 6.0 5.2
• The Supply On Indicator lights when the backup system is
Empty 15.2 13.5 11.9 10.3
operating on battery power.
• The Backup System Battery Low Indicator lights when the 10 ft3 1/2 Full 12.0 10.4 9.4 8.3
backup system battery charge is low. The built-in battery Full 8.9 7.8 7.0 6.1
charger recharges the battery to full charge when power is
restored to the system.
Note: Since rechargeable batteries degrade over time, the
battery should be replaced after approximately three
years.
• The Backup System Empty Indicator lights when the standby
CO2 or LN2 bottle is empty.
7
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
8
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
7 Operation (Revco Value Series Freezers) 7.2.2 Setting the Cabinet Temperature
To set the cabinet temperature, complete the following steps:
7.1 Control Panel Features
1. Insert a small screwdriver into the slotted screw labeled
Before the initial start up, take some time to become familiar Control Setpoint and simultaneously press and hold the
with the controls on your freezer. Figure 5 illustrates the Revco Setpoint Display button. The temperature display changes to
Value Series Chest and Upright Freezer control panel. read the existing setpoint value.
2. Turn the setpoint screw (clockwise for a colder setting and
4 2 5 counterclockwise for a warmer setting) until the desired
setpoint shows in the digital temperature display.
3. Release the Setpoint Display button. The digital temperature
°C
display returns to the cabinet temperature.
Power
OFF
POWER ON 7.2.3 Optional Alarm
CONTROL
ALARM ON SETPOINT The optional Revco Value Series Alarm is operated by the freezer
electronic control system. The alarm is preset at the factory with
SETPOINT a warm alarm setpoint:
DISPLAY
• For units designed to operate at -75ºC and -85ºC, the alarm
1 3 setpoint is approximately 90% of the freezer temperature
setpoint. For example, if the freezer setpoint is -80ºC, the
Figure 5. Revco Value Series Control Panel alarm setpoint is -72ºC.
(Shown with Optional Alarm) • For units designed to operate at -50ºC and warmer, the alarm
setpoint is 80% of the freezer temperature setpoint. For
1. Keyed Power On/Alarm On switch. example, if the freezer setpoint is -45ºC, the alarm setpoint is
2. LED digital temperature display. -36ºC.
3. Recessed temperature control setpoint adjustment.
4. Temperature setpoint adjust button. To activate the alarm, wait until the freezer reaches operating
5. Calibration potentiometer. temperature and turn the key switch to the ALARM ON position.
9
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
8.2 Gasket Maintenance 4. Connect the battery terminals and replace the front grill.
Periodically check the gaskets around the door or lid for 8.5 Troubleshooting Procedures
punctures or tears. Leaks are indicated by a streak of frost which
forms at the point of gasket failure. Make sure that the cabinet is Table 3. Troubleshooting Procedures
level (refer to Section 5.3 on page 2 for leveling information).
Problem Solution
Keep the lid and door gaskets clean and frost free by wiping Check Fuse Surge protector is blown. Call an Electronics
gently with a soft cloth. indicator is on. Technician to check the equipment.
Clean Filter
Condenser filter is dirty. Refer to Section 8.1.2.
indicator is on.
Alarm battery is low on charge. Have a
Alarm Battery Low
technician check the alarm battery. Refer to
indicator is on.
Section 8.4.
Low incoming line voltage to the freezer. Call an
Voltage Boost
Electronics Technician to check the power
indicator is on.
supply.
10
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
CHANGE CHART
1 2
11
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
10.1 Backup Protection Devices This fully automatic device monitors the supply voltage and
increases voltage if necessary so the freezer always operates
Liquid Nitrogen and Liquid CO2 systems are available to protect within the factory recommended limits. Revco Ultima II models
your product in case of a prolonged power failure. These systems are factory equipped with this device.
are powered by a rechargeable battery with a built-in automatic
charger. Refer to Section 6.3.4 on page 7 for additional CAUTION! The Voltage Safeguard unit does not
information. allow you to operate a 115 volt product on 230 volts or
vice versa. It is designed to provide 5 to 15% voltage
10.1.1 Liquid Nitrogen (LN2) correction, depending on the freezer model.
This system uses portable and locally refillable bottles. These 10.2 Remote Alarm
containers keep the liquid at low temperature and pressure by
The remote alarm terminals on the back of the freezer are
continually venting a small amount of gaseous N2. You must
connected to freezer internal dry contacts. You can connect a
check the liquid level indicator daily to assure an adequate
remote alarm package with a separate 24 volt power supply.
emergency supply.
11 Accessories
Note: For accessories and options not described herein, or for special modifications, contact Customer Sales.
Inventory Rack Capacity Chart for Revco Ultima, Elite, and Value Series
Rack Catalog No. Cubic Foot Capacity
(Rack Quantity) Box Size in. 1 3.1 4.5 6.8 10.3 13.7 17.1 20.5 13.4 17.2 20.2 24.4
For ULT Chest Freezers
5972, Full Rack, (7) 2 in. boxes — 9 12 — — — — — — — — —
5973, Full Rack, (5) 3 in. boxes — 9 12 — — — — — — — — —
5978, Full Rack, (11) 2 in. boxes — — — 12 18 24 30 36 — — — —
6096, Full Rack, (8) 3 in. boxes — — — 12 18 24 30 36 — — — —
6230, Half Rack, (6) 2 in. boxes — — — 24 36 48 60 72 — — — —
6231, Half Rack, (4) 3 in. boxes — — — 24 36 48 60 72 — — — —
For ULT Upright Freezers
6112-1, (12) 2 in. or (9) 3 in. boxes — — — — — — — — 20 — — —
6113-1, (16) 2 in. or (12) 3 in. boxes — — — — — — — — — 20 25 30
12
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
Single full rack, holds (5) 3 in. boxes, for 3.1 and 4.5 cu.ft Red cell press. 6225
chest models only. 5973
Single full rack, holds (11) 2 in. boxes, for all chest models Blood Bank Freezer Inventory Capacity
except 3.1 and 4.5 cu.ft. 5978 Freezers Cu.ft/Total Capacity
Single full rack, holds (8) 3 in. boxes, for all chest models Rack Catalog No. and Chest Upright
except 3.1 and 4.5 cu.ft. 6096 Descriptiona
6.8 13.7 20.5 13.4 20.2
Single full rack, holds (8) plasma boxes, for 3 cu.ft chest Cu.ft Cu.ft Cu.ft Cu.ft Cu.ft
model. 6097
6225, Red Cell Press. 92 204 334 200 —
Single full rack, holds (13) plasma boxes, for all chest
models except 3.1 and 4.5 cu.ft. 6098 6098, Plasma Rack, 13
12 24 36 — —
Box Capacity.
Single split rack, holds (6) 2 in. boxes, for all chest models
except 3.1 and 4.5 cu.ft. 6230 6110-1, Plasma Rack,
— — — 20 —
12 Box Capacity.
Single split rack, holds (4) 3 in. boxes, for all chest models
except 3.1 and 4.5 cu.ft. 6231 6111-1, Plasma Rack,
— — — — 25
16 Box Capacity.
Basket/Rail Kit, for 6.8 cu.ft chest freezers. 6217
a
Example: A 13.7 cu.ft plasma chest freezer holds 24 total
Basket/Rail Kit, for 10.3 cu.ft chest freezers. 6218
Catalog No. 6098 Plasma Racks.
Basket/Rail Kit, for 13.7 cu.ft chest freezers. 6190
Basket/Rail Kit, for 17.1 cu.ft chest freezers. 6191
Basket/Rail Kit, for 20.5 cu.ft chest freezers. 6192
13
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
Catalog Free-standing, for Revco Elite and Value models only. Field
Description installed by qualified electrician or factory authorized
No.
service center. Specify freezer model number and
CO2 Backup, built-in, panel mounted, factory installed. voltage when ordering. 6402
Revco Ultima and Elite only. Low battery indictor lights
when battery needs replacing. Empty cylinder indicator
lights when cylinder is exhausted. 6594
CO2 Backup, free-standing, field installed. Specify freezer
model number and voltage when ordering. 6593
14
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
Remote Alarm/Monitoring
Catalog
Catalog Description
Description No.
No.
Extended 12-Month Comprehensive Coverage, for all
Standard Remote Alarm. Provides audible and visual models, provides additional 12-month warranty
signal in the event of temperature rise or power failure. coverage, parts and labor. Contact Sales for details. 6613
Adaptable to telephone switchboard. Can be located
up to 1/3 mile (1760 ft/536 m) from freezer. Must be
used with factory installed electronic alarm system and Interior Options
connected to normally-open or normally-closed
Catalog
remote alarm contacts. Wire not included. Specify Description
No.
freezer model number and voltage when ordering. 5612
Chest Freezer Interiors
Deluxe Electronic Remote Alarm System. User
programmable to sound alarm in the event of Stainless steel interior, 3.1 cu.ft chest. 6121
temperature rise or power failure. Can dial up to four
telephone numbers to advise of alarm condition across Stainless steel interior, 4.5 cu.ft chest. 6646
any telephone system which accepts pulse dialing. Stainless steel interior, 6.8 cu.ft chest. 6122
One system can monitor up to four individual freezers
or up to three groups of freezers. Contact Sales for Stainless steel interior, 10.3 cu.ft chest. 6123
detailed specifications. 6224
Stainless steel interior, 13.7 cu.ft chest. 6124
Stainless steel interior, 17.1 cu.ft chest. 6125
RS-232 Data Port and Software
Stainless steel interior, 20.5 cu.ft chest. 6126
Catalog
Description
No. Upright Freezer Interiors
For use with Revco Elite models only. Monitors freezer Stainless steel interior, 13.4 cu.ft upright. 6179
setpoint, high and low limit setpoints and power
supply. Advises of temperature deviation, voltage Stainless steel interior, 17.2 cu.ft upright. 6180
boost on, surge intercept, low alarm battery, and clean Stainless steel interior, 20.2 cu.ft upright. 6181
filter functions. Auto status report adjustable in
increments of 5 seconds to 2 hours. 6405 Stainless steel interior, 24.4 cu.ft upright. 6645
15
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
WEEE Compliance
Great Britain
WEEE Compliance. This product is required to comply with the European
Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96EC.
It is marked with the following symbol. Thermo Fisher Scientific has contracted
with one or more recycling/disposal companies in each EU Member State, and this
product should be disposed of or recycled through them. Further information on
Thermo Fisher Scientific compliance with these Directives, the recyclers in your
country, and information on Thermo Scientific products which may assist the
detection of substances subject to the RoHS Directive are available at
www.thermo.com/
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Deutschland
Equipment (WEEE) Richtlinie 2002/96EC erfüllen. Das Produkt ist durch
folgendes Symbol gekennzeichnet. Thermo Fisher Scientific hat Vereinbarungen
getroffen mit Verwertungs/Entsorgungsanlagen in allen EU-Mitgliederstaaten und
dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden.
Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo
Fisher Scientific, die Verwerter und Hinweise die Ihnen nützlich sein können, die
Thermo Scientific Produkte zu identifizieren, die unter diese RoHS.
Anweisungfallen, finden Sie unter www.thermo.com/
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione
Europea 2002/96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici Italia
(WEEE). È marcato col seguente simbolo.Thermo Fisher Scientific ha stipulato
contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli Stati
Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste
medesime. Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con
queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni
sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze
soggette alla Direttiva RoHS sono disponibili sul sito www.thermo.com/
16
Important
For your future reference and when contacting the factory, please have the
following information readily available:
Model Number:
Serial Number:
Date Purchased:
SALES DIVISION
Phone: 828/658-2711
800/252-7100
FAX: 828/645-3368
TECHNICAL SUPPORT
Phone: 800/438-4851
Thermo Scientific Revco Chest and Upright Freezers Installation and Operation
WEEE Compliance
Great Britain
WEEE Compliance. This product is required to comply with the European
Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96EC.
It is marked with the following symbol. Thermo Fisher Scientific has contracted
with one or more recycling/disposal companies in each EU Member State, and this
product should be disposed of or recycled through them. Further information on
Thermo Fisher Scientific compliance with these Directives, the recyclers in your
country, and information on Thermo Scientific products which may assist the
detection of substances subject to the RoHS Directive are available at
www.thermo.com/
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Deutschland
Equipment (WEEE) Richtlinie 2002/96EC erfüllen. Das Produkt ist durch
folgendes Symbol gekennzeichnet. Thermo Fisher Scientific hat Vereinbarungen
getroffen mit Verwertungs/Entsorgungsanlagen in allen EU-Mitgliederstaaten und
dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden.
Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo
Fisher Scientific, die Verwerter und Hinweise die Ihnen nützlich sein können, die
Thermo Scientific Produkte zu identifizieren, die unter diese RoHS.
Anweisungfallen, finden Sie unter www.thermo.com/
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione
Europea 2002/96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici Italia
(WEEE). È marcato col seguente simbolo.Thermo Fisher Scientific ha stipulato
contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli Stati
Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste
medesime. Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con
queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni
sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze
soggette alla Direttiva RoHS sono disponibili sul sito www.thermo.com/
16
Important
For your future reference and when contacting the factory, please have the
following information readily available:
Model Number:
Serial Number:
Date Purchased:
SALES DIVISION
Phone: 1-866-984-3766 (866-9-THERMO)
TECHNICAL SUPPORT
Phone: 1-800-438-4851
Thermo Fisher Scientific Inc.
275 Aiken Road
Asheville, NC 28804
United States
www.thermofisher.com
39125H01 Rev. M