Test Kit Instructions: Romer Labs Agrastrip Total Aflatoxin Quantitative Watex
Test Kit Instructions: Romer Labs Agrastrip Total Aflatoxin Quantitative Watex
ROMER LABS
FORWARD
The instructions presented in this document cover only the procedure for performing the
analytical test for official inspections. For questions regarding this procedure, contact Dr. Ajit
Ghosh of the Technology and Science Division by phone at 816-891-0417 or email at
[email protected].
Refer to the Mycotoxin Handbook for information on use of this test kit in official inspections
including sampling, general sample preparation, grinding, reporting and certification of test
results, laboratory safety, and hazardous waste management. For questions regarding these
policies and/or instructions, contact Patrick McCluskey of PPMB by phone at 816-659-8403 or
email at [email protected].
The U.S. Department of Agriculture (USDA) prohibits discrimination in its programs on the basis of race,
color, national origin, sex, religion, age, disability, political beliefs, and marital or familial status. (Not all
prohibited bases apply to all programs.) Persons with disabilities who require alternate means for
communication of program information (Braille, large print, audiotape, etc.) should contact USDA’s
TARGET Center at (202) 720-2600 (voice and TDD).
To file a complaint, write to the USDA, Office of Civil Rights, Room 326-W, 1400 Independence Avenue,
SW, Washington, DC 20250-9410, or call (202) 720-5964 (voice and TDD). USDA is an equal employment
opportunity employer.
Page 2
1. GENERAL INFORMATION
The AgraStrip Total Aflatoxin Quantitative WATEX test is a one-step lateral flow
immunochromatographic assay for the quantitative screening of Aflatoxins in
samples. The test is based on a competition immunoassay format. Antibody-
colloidal gold complex (conjugate) lyophilized in a microwell is mixed with sample
extract. An AgraStrip Aflatoxin WATEX strip is placed into the microwell. The
mixed content is then wicked onto the membrane of the strip, which contains a
test zone and a control zone. The test zone captures free antibody-particle
complex (conjugate), allowing color particles to concentrate and form a visible
line. The color intensity of the line is inversely proportional to the concentration of
Aflatoxins in the sample. The line is always visible in the control zone regardless
of the presence of Aflatoxins. The strips are analyzed using an AgraVision
Reader and the results are determined.
Page 3
2. PREPARATION OF TESTING MATERIALS AND EQUIPMENT
All reagents and kit components must be at room temperature 20-24°C (68-75°F)
before use.
a. AgraStrip Incubator
b. AgraVision Reader
(1) Turn on the AgraVision reader using the power button located on
the back.
(2) Use the arrow keys on the keypad to highlight “TEST”. Select it
using the checkmark key.
(3) Use the arrow keys on the keypad to highlight “Mycotoxin”. Select it
using the checkmark key.
(4) The barcode scanner will turn on. Scan the barcode included in the
test kit on the certificate of analysis.
Note: Separate bar codes for the separate quantitation ranges are
provided in the test kit.
(5) Create a sample ID by using the alphanumeric keys on the keypad.
Use the checkmark key to enter.
(6) Since only one strip is being read; use the pound key to bypass to
the next screen.
(7) Enter the operator ID. Press the checkmark key to enter, and press
it a second time to move to the next screen.
(8) The reader is ready to read and will display “start measurement”.
Note: Strips should not be read more than 1 minute after
completion of the run.
Page 4
Allow all samples, water used for extraction, and test kit materials to
equilibrate to room temperature before use.
3. EXTRACTION PROCEDURE
a. Extraction procedure for Corn (including Dent or Field Corn, Corn Meal,
Corn Flour, Cracked Corn, Corn Grits or Polenta, and Corn screenings)
(1) Weigh out 50.0 ± 0.2 grams ground sample into one side of a
filtering Whirl-Pak bag.
(2) Place one soluble buffer bag on top of the ground sample in the
Whirl-Pak bag.
(4) Shake closed Whirl-Pak bag vigorously by hand for 2 min or shake
with similar shaking action (as hand shaking) for 2 minutes using a
mechanical shaker.
(5) Allow sample to settle for 2 minutes to obtain the filtered extract
(take extract from the opposite side of the filter from the sample).
This extract can only be used for the next 5 minutes.
(1) Dilute the settled and filtered extract 1:21 with Dilution Buffer. Using
a 1000 µL pipette and a blue pipette tip, add 1000 µL of Dilution
Buffer to a dilution tube. Then, using a 100 µL pipette and a yellow
or white pipette tip, add 50 µL of filtered extract into the dilution
tube.
(2) Close tube, mix well by inverting a few times and the sample is
ready for assay. This diluted extract can only be used for next
one hour.
Note: Extract should be pipetted from the side of filter bag that is
opposite of the ground sample. If the sample has a large foam
head, tilt the bag for easier access to the supernatant.
Page 5
5. TEST PROCEDURES
(1) Place the cover of the heat block on the top of the heat block.
Break off one conjugate coated microwell for the sample to be run.
Remove the sealing cap from conjugate microwell, and place the
conjugate microwell inside the heat block. Ensure that the well is
fully seated in the heat block. Only one sample should be run at
a time.
(3) Put one test strip into the microwell. Place the cover back into the
heat block to cover the microwell and test strip.
(4) Start the timer and develop the test strip for 3 minutes. Remove the
test strip immediately once the 3 minutes are complete.
(5) Wipe the end of the test strip onto an absorbent paper and insert
the strip into the strip holder/tray for reading.
Note: After the test, the used microwell can be removed with
tweezers.
(1) Use the AgraVision Reader to read the strip and interpret the result.
Strips should not be read more than 1 minute after completion of
the run.
(2) Insert the strips into the tray, and insert the tray into the reader. The
strips should go in the tray with the white end facing into the reader.
Press the checkmark key to read.
(3) After completion of reading, press the checkmark key to save the
result in the AgraVision Reader’s memory or the pound key to print
the result.
Page 6
(4) If the reader gives a value greater than 100 ppb, the previously
diluted sample extract needs to be further diluted and run the test
following “Quantitation Range 101 to 300 ppb” below.
If the reader gives a concentration greater than 100 ppb, the diluted
extract needs to be further diluted and retested.
(1) Dilute the diluted extract in a ratio of 1:10, i.e. using an adjustable
1000 µL pipette, add 900 µL of Dilution Buffer to a dilution tube.
(3) Perform the test by repeating steps (1) – (5) of section 5 “Test
procedures” sections a. “Analysis Procedure for Quantitation
Ranges 5 to 100 ppb and 100 to 300 ppb” followed by section b.
“Reading the results.”
(4) To interpret the results for the second quantitation range, a different
bar code must be scanned using the AgraVision Reader prior to
repetition of the analysis.
A final result less than 53 ppb (using 101 – 300 ppb quantitation range) is
indicative of a problem, and troubleshooting is needed. Verify the
procedure is being followed properly. Perform the analysis using 5.0 – 100
ppb quantitation range and only perform the analysis using 101 – 300 ppb
quantitation range again if the value is greater than 100 ppb.
d. Interpretation of results
(1) A color line always appears in the upper section of the test strip to
indicate that the test strip is working properly. This line is the
Control Line (CL). A line in the lower section of the test strip
indicates the test result. This line is the Test Line (TL).
(2) Invalid results: If there is no control line in the control zone, the
test is invalid, and the sample should be re-tested by using a new
test strip. The AgraVision Reader will also indicate “invalid” if the
strip is invalid.
Page 7
(3) Valid results: 2 lines or only CL are visible. The intensity of the line
in the test zone is dependent on the Aflatoxin concentration in the
sample.
Refer to the Mycotoxin Handbook for reporting and certification of test results.
For questions regarding these instructions, contact Patrick McCluskey (816-659-
8403 or [email protected]).
a. Storage Conditions:
(1) Store test kits at 4-8°C (40-47°F) when not in use, and do not use
beyond the expiration date. Do not freeze. Do not leave it in direct
sunlight.
(2) Conjugate microwells and the Test strips must be kept inside their
original tubes.
Page 8
(2) Adhere to the FGIS issued instructions of test procedures.
(4) Consider all materials, containers and devices that are exposed to
the sample to be contaminated with toxin. Wear protective gloves
and safety glasses when using the kit.
(5) The components in this test kit have been quality control tested as
a standard batch unit. Do not mix components from different lot
numbers.
Page 9
(7) Romer #10002634: Single channel pipette capable of pipetting up
to 100 µL
9. REVISION HISTORY
Page 10
10. FLOW CHART
Page 11