E+H PromagH200 Manual 2017 EN
E+H PromagH200 Manual 2017 EN
Operating Instructions
Proline Promag H 200
HART
Electromagnetic flowmeter
Proline Promag H 200 HART
• Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
• To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to
these Instructions.
2 Endress+Hauser
Proline Promag H 200 HART Table of contents
Table of contents
Endress+Hauser 3
Table of contents Proline Promag H 200 HART
4 Endress+Hauser
Proline Promag H 200 HART Table of contents
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Endress+Hauser 5
Document information Proline Promag H 200 HART
1 Document information
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Symbol Meaning
Allen key
Open-ended wrench
6 Endress+Hauser
Proline Promag H 200 HART Document information
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
, , … Series of steps
Visual inspection
1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The CD-ROM provided for the device (depending on the device version, the CD-ROM
might not be part of the delivery!)
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
For a detailed list of the individual documents along with the documentation code
Endress+Hauser 7
Document information Proline Promag H 200 HART
Brief Operating Instructions Guide that takes you quickly to the 1st measured value
The Brief Operating Instructions contain all the essential information
from incoming acceptance to initial commissioning.
8 Endress+Hauser
Proline Promag H 200 HART Basic safety instructions
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage
caused by improper or non-designated use.
LWARNING
Danger of breakage of the sensor due to corrosive or abrasive fluids or from
environmental conditions!
‣ Verify the compatibility of the process fluid with the sensor material.
‣ Ensure the resistance of all fluid-wetted materials in the process.
‣ Keep within the specified pressure and temperature range.
Endress+Hauser 9
Basic safety instructions Proline Promag H 200 HART
Residual risks
The external surface temperature of the housing can increase by max. 10 K due to the
power consumption of the electronic components. Hot process fluids passing through the
measuring device will further increase the surface temperature of the housing. The surface
of the sensor, in particular, can reach temperatures which are close to the fluid
temperature.
Possible burn hazard due to fluid temperatures!
‣ For elevated fluid temperature, ensure protection against contact to prevent burns.
10 Endress+Hauser
Proline Promag H 200 HART Basic safety instructions
2.6 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
Endress+Hauser 11
Product description Proline Promag H 200 HART
3 Product description
The device consists of a transmitter and a sensor.
One device version is available: compact version - transmitter and sensor form a
mechanical unit.
1 2 3 4 5 6 7 8
–
+
E
9
A0014056
12 Endress+Hauser
Proline Promag H 200 HART Incoming acceptance and product identification
• If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center.
• Depending on the device version, the CD-ROM might not be part of the delivery!
The Technical Documentation is available via the Internet or via the Endress+Hauser
Operations App, see the "Product identification" section → 14.
Endress+Hauser 13
Incoming acceptance and product identification Proline Promag H 200 HART
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The chapters "Additional standard documentation on the device" → 8 and
"Supplementary device-dependent documentation" → 8
• The W@M Device Viewer: Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or
scan the 2-D matrix code (QR code) on the nameplate.
345
1
2
Order code: 14
Ser. no.:
Ext. ord. cd.:
6
15
7
8
9 i 16
i
Patents
17
Date: 322540-0001
10 11 12 13
A0013906
14 Endress+Hauser
Proline Promag H 200 HART Incoming acceptance and product identification
1 2
Order Code: 3
Ser.No.: 4
Ext. ord. cd.: 5
6 7
Tm: 8
Material: 9
10
11
i Patents i
13 12
Date:
15 14
A0017186
Order code
The measuring device is reordered using the order code.
Extended order code
• The device type (product root) and basic specifications (mandatory features) are
always listed.
• Of the optional specifications (optional features), only the safety and approval-
related specifications are listed (e.g. LA). If other optional specifications are also
ordered, these are indicated collectively using the # placeholder symbol (e.g. #LA#).
• If the ordered optional specifications do not include any safety and approval-related
specifications, they are indicated by the + placeholder symbol (e.g. XXXXXX-ABCDE
+).
Endress+Hauser 15
Incoming acceptance and product identification Proline Promag H 200 HART
Symbol Meaning
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious
or fatal injury.
Reference to documentation
Refers to the corresponding device documentation.
16 Endress+Hauser
Proline Promag H 200 HART Storage and transport
A0015604
A0015606
Endress+Hauser 17
Storage and transport Proline Promag H 200 HART
LCAUTION
Risk of damaging the magnetic coil
‣ If transporting by forklift, do not lift the sensor by the metal casing.
‣ This would buckle the casing and damage the internal magnetic coils.
A0023726
18 Endress+Hauser
Proline Promag H 200 HART Installation
6 Installation
Mounting location
h A0023343
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the
next pipe elbow: h ≥ 2 × DN
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
1
2
h
A0017064
Endress+Hauser 19
Installation Proline Promag H 200 HART
N
×D
³2
N
×D
³5
A0017063
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
An optimum orientation position helps avoid gas and air accumulations and deposits in
the measuring tube.
The measuring device also offers the empty pipe detection function to detect partially filled
measuring pipes in the event of outgassing fluids or variable process pressures.
Vertical
A0015591
Optimum for self-emptying pipe systems and for use in conjunction with empty pipe
detection.
Horizontal
2 2
A0019602
• The measuring electrode plane must be horizontal. This prevents brief insulation of
the two measuring electrodes by entrained air bubbles.
• Empty pipe detection only works if the transmitter housing is pointing upwards as
otherwise there is no guarantee that the empty pipe detection function will actually
respond to a partially filled or empty measuring tube.
20 Endress+Hauser
Proline Promag H 200 HART Installation
≥ 5 × DN ≥ 2 × DN
A0016275
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
Local display –20 to +60 °C (–4 to +140 °F), the readability of the display may be
impaired at temperatures outside the temperature range.
Liner Do not exceed or fall below the permitted temperature range of the liner .
If operating outdoors:
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
System pressure
A0015594
Never install the sensor on the pump suction side in order to avoid the risk of low pressure,
and thus damage to the liner.
Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps
are used.
• For information on the liner's resistance to partial vacuum → 127
• For information on the shock resistance of the measuring system → 126
• For information on the vibration resistance of the measuring system → 126
Endress+Hauser 21
Installation Proline Promag H 200 HART
Vibrations
In the event of very strong vibrations, the pipe and sensor must be supported and fixed.
• For information on the shock resistance of the measuring system → 126
• For information on the vibration resistance of the measuring system → 126
L
A0016266
Adapters
Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor
in larger-diameter pipes. The resultant increase in the rate of flow improves measuring
accuracy with very slow-moving fluids. The nomogram shown here can be used to
calculate the pressure loss caused by reducers and expanders.
The nomogram only applies to liquids with a viscosity similar to that of water.
[mbar] 100
8 m/s
7 m/s
6 m/s
10
5 m/s
max. 8° 4 m/s
3 m/s
d D
2 m/s
1
1 m/s
Display protection
‣ To ensure that the optional display protection can be easily opened, maintain the
following minimum head clearance: 350 mm (13.8 in)
22 Endress+Hauser
Proline Promag H 200 HART Installation
For transmitter
• For turning the transmitter housing: Open-ended wrench8 mm
• For opening the securing clamps: Allen key3 mm
For sensor
For flanges and other process connections:
• Screws, nuts, seals etc. are not included in the scope of supply and must be provided by
the customer.
• Appropriate mounting tools
1. Ensure that the direction of the arrow on the sensor matches the flow direction of
the medium.
2. To ensure compliance with device specifications, install the measuring device
between the pipe flanges in a way that it is centered in the measurement section.
3. Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0013964
The sensor is supplied to order, with or without pre-installed process connections. Pre-
installed process connections are firmly secured to the sensor by 4 or 6 hexagonal-headed
bolts.
The sensor may need to be supported or additionally secured depending on the
application and pipe length. In particular, it is absolutely essential to secure the sensor
additionally if plastic process connections are used. An appropriate wall mounting kit
can be ordered separately as an accessory from Endress+Hauser → 134.
Endress+Hauser 23
Installation Proline Promag H 200 HART
DN 2…25
(1/12…1")
A0018782
• If thin-walled pipes carrying food are welded correctly, the seal is not damaged by
the heat even when mounted. However, it is recommended to disassemble the
sensor and seal.
• It must be possible to open the pipe by approx. 8 mm (0.31 in) in total to permit
disassembly.
24 Endress+Hauser
Proline Promag H 200 HART Installation
In the case of plastic process connections (e.g. flange connections or adhesive fittings),
additional ground rings must be used to ensure potential matching between the sensor
and the fluid. If grounding rings are not installed, this can affect the measuring accuracy or
cause the destruction of the sensor as a result of the electrochemical decomposition of the
electrodes.
• Depending on the option ordered, plastic disks are used instead of grounding rings
on some process connections. These plastic disks only act as "spacers" and do not
have any potential matching function. Furthermore, they also perform a significant
sealing function at the sensor/process connection interface. Therefore, in the case
of process connections without metal grounding rings, these plastic disks/seals
should never be removed and should always be installed!
• Grounding rings can be ordered separately as an accessory from Endress+Hauser
→ 134. When ordering make sure that the grounding rings are compatible with
the material used for the electrodes, as otherwise there is the danger that the
electrodes could be destroyed by electrochemical corrosion!
Material specifications → 129.
• Grounding rings, including seals, are mounted inside the process connections.
Therefore the installation length is not affected.
Endress+Hauser 25
Installation Proline Promag H 200 HART
4
2 3 2
A0002651
1. Release the 4 or 6 hexagonal-headed bolts (1) and remove the process connection
from the sensor (4).
2. Remove the plastic disk (3), along with the two O-ring seals (2), from the process
connection.
3. Place the first O-ring seal (2) back into the groove of the process connection.
4. Fit the metal grounding ring (3) in the process connection as illustrated.
5. Place the second O-ring seal (2) into the groove of the grounding ring.
6. Mount the process connection back on the sensor. In doing so, make sure to observe
the maximum screw tightening torques for lubricated threads: 7 Nm (5.2 lbf ft)
max. 350°
8 mm 8 mm
A0013713
26 Endress+Hauser
Proline Promag H 200 HART Installation
–
+
E
3 mm
A0013905
1. Loosen the securing clamp of the electronics compartment cover using an Allen key.
2. Unscrew cover of the electronics compartment from the transmitter housing.
3. Optional: pull out the display module with a gentle rotational movement.
4. Rotate the display module into the desired position: Max. 8 × 45° in each direction.
5. Without display module pulled out:
Allow display module to engage at desired position.
6. With display module pulled out:
Feed the cable into the gap between the housing and main electronics module and
plug the display module into the electronics compartment until it engages.
7. Reverse the removal procedure to reassemble the transmitter.
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping ?
Are the measuring point identification and labeling correct (visual inspection)?
Have the fixing screws been tightened with the correct tightening torque?
Endress+Hauser 27
Electrical connection Proline Promag H 200 HART
7 Electrical connection
The measuring device does not have an internal circuit breaker. For this reason,
assign the measuring device a switch or power-circuit breaker so that the power
supply line can be easily disconnected from the mains.
Electrical safety
In accordance with applicable federal/national regulations.
Signal cable
Current output
For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the
plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Plug-in spring terminals for device version without integrated overvoltage protection:
wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Screw terminals for device version with integrated overvoltage protection: wire cross-
sections 0.2 to 2.5 mm2 (24 to 14 AWG)
28 Endress+Hauser
Proline Promag H 200 HART Electrical connection
Transmitter
2 1 3 2 1 3
+ – + –
3 4 1 2 + – +–
3 4 1 2
A0013570 A0018161
Maximum number of terminals, without integrated Maximum number of terminals, with integrated
overvoltage protection overvoltage protection
Output 1 Output 2
Pulse/frequency/switch output
Option B 1) 4-20 mA HART (passive)
(passive)
Supply voltage
Transmitter
An external power supply is required for each output. The following supply voltage values
apply for the 4-20 mA HART current output:
Option A 1) 2)
: 4-20 mA HART For 4 mA: ≥ DC 18 V DC 35 V
For 20 mA: ≥ DC 14 V
Load
Load for current output: 0 to 500 Ω, depending on the external supply voltage of the power
supply unit
Endress+Hauser 29
Electrical connection Proline Promag H 200 HART
500
400
300
200
100
0
14 16 18 20 22 24 26 28 30 32 34 36 Us [V]
A0013563
1 Operating range
1.1 For order code for "Output", option A "4-20 mA HART"/option B "4-20 mA HART, pulse/frequency/switch
output" with Ex i
1.2 For order code for "Output", option A "4-20 mA HART"/option B "4-20 mA HART, pulse/frequency/switch
output" with non-Ex and Ex d
Sample calculation
Supply voltage of the power supply unit: US = 19 V
Maximum load: RB ≤ (19 V - 13.5 V): 0.022 A = 250 Ω
30 Endress+Hauser
Proline Promag H 200 HART Electrical connection
3 mm 20 mm
10 (0.4)
mm (in)
A0013836
Removing a cable
3 (0.12)
3 4 1 2
mm (in)
A0013835
‣ To remove a cable from the terminal, use a flat-blade screwdriver to push the slot
between the two terminal holes while simultaneously pulling the cable end out of the
terminal.
Endress+Hauser 31
Electrical connection Proline Promag H 200 HART
Requirements
LCAUTION
Electrode damage can result in the complete failure of the device!
‣ Same electrical potential for the fluid and sensor
‣ Company-internal grounding concepts
‣ Pipe material and grounding
For devices intended for use in hazardous locations, please observe the guidelines in
the Ex documentation (XA).
32 Endress+Hauser
Proline Promag H 200 HART Electrical connection
4
2 3 2
A0002651
2 1
3
4
A0017293
Endress+Hauser 33
Electrical connection Proline Promag H 200 HART
1 2 3 4
4...20 mA
+ +
5
- -
A0015511
Pulse/frequency output
1 2
– +
_ +
– 3
+
12345
A0016801
34 Endress+Hauser
Proline Promag H 200 HART Electrical connection
Switch output
1 2
_ +
_
+
_ 3
+
A0016802
HART input
1 2 3 4 5
4...20 mA
+ +
+
– – 7
+ -
–
–
+
+ +
+
– – 6
-
3 4
A0016029
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following
steps after the electrical connection:
1. Check that the housing seals are clean and fitted correctly. Dry, clean or replace the
seals if necessary.
2. Tighten all housing screws and screw covers.
3. Firmly tighten the cable glands.
Endress+Hauser 35
Electrical connection Proline Promag H 200 HART
4. To ensure that moisture does not enter the cable entry, route the cable so that it
loops down before the cable entry ("water trap").
A0013960
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
→ 35 ?
Depending on the device version: are all the device plugs firmly tightened ?
Does the supply voltage match the specifications on the transmitter nameplate ?
36 Endress+Hauser
Proline Promag H 200 HART Operation options
8 Operation options
SC
1 2 3 4 5
A0015607
Endress+Hauser 37
Operation options Proline Promag H 200 HART
Parameter 1
Parameter n
Submenu 1
Submenu n
Task-oriented
Wizard 1 / Parameter 1
Maintenance
Wizard n / Parameter n
Parameter 1
Parameter n
Submenu 1
Submenu n
Diagnostics Parameter 1
Parameter n
Submenu 1
Submenu n
Parameter n
System
Sensor
Expert
Intput
Output
Communication
Application
Diagnostics
A0018237-EN
38 Endress+Hauser
Proline Promag H 200 HART Operation options
Diagnostics "Maintenance" role Contains all parameters for error detection and analyzing
Fault elimination: process and device errors:
• Diagnostics and elimination of • "Diagnostic list" submenu
process and device errors Contains up to 5 currently pending diagnostic messages.
• Measured value simulation • "Event logbook" submenu
Contains up to 20 or 100 (order option " Extended
HistoROM") event messages that have occurred.
• "Device information" submenu
Contains information for identifying the device.
• "Measured values" submenu
Contains all current measured values.
• "Data logging" submenu (order option "Extended
HistoROM")
Storage and visualization of up to 1000 measured values
• "Heartbeat Technology" submenu
The functionality of the device is checked on demand and the
verification results are documented.
• "Simulation" submenu
Is used to simulate measured values or output values.
Expert function-oriented Tasks that require detailed knowledge Contains all the parameters of the device and makes it possible
of the function of the device: to access these parameters directly using an access code. The
• Commissioning measurements under structure of this menu is based on the function blocks of the
difficult conditions device:
• Optimal adaptation of the • "System" submenu
measurement to difficult conditions Contains all higher-order device parameters that do not
• Detailed configuration of the pertain either to measurement or the measured value
communication interface communication.
• Error diagnostics in difficult cases • "Sensor" submenu
Configuration of the measurement.
• "Input" submenu
Configuration of the input.
• "Output" submenu
Configuration of the outputs.
• "Communication" submenu
Configuration of the digital communication interface.
• "Application" submenu
Configuration of the functions that go beyond the actual
measurement (e.g. totalizer).
• "Diagnostics" submenu
Error detection and analysis of process and device errors and
for device simulation and Heartbeat Technology.
Endress+Hauser 39
Operation options Proline Promag H 200 HART
2 XXXXXXXXX F 3
4 1120.50
l/h
A0016502
1 Operational display
2 Device tag
3 Status area
4 Display area for measured values (4-line)
5 Operating elements → 45
Status area
The following symbols appear in the status area of the operational display at the top right:
• Status signals→ 98
– F: Failure
– C: Function check
– S: Out of specification
– M: Maintenance required
• Diagnostic behavior→ 99
– : Alarm
– : Warning
• : Locking (the device is locked via the hardware )
• : Communication (communication via remote operation is active)
Display area
In the display area, each measured value is prefaced by certain symbol types for further
description:
↓ ↓ ↓
Example
Measured variables
Symbol Meaning
Volume flow
Mass flow
40 Endress+Hauser
Proline Promag H 200 HART Operation options
Totalizer
displayed.
The measurement channel number indicates which of the three totalizers is
Output
displayed.
The measurement channel number indicates which of the two current outputs is
Measurement channel 1 to 4
The measurement channel number is displayed only if more than one channel is present for the same measured
variable type (e.g. totalizer 1-3).
Diagnostic behavior
The diagnostic behavior pertains to a diagnostic event that is relevant to the displayed measured variable.
For information on the symbols → 99
The number and display format of the measured values can be configured via the
"Format display" parameter → 72. "Operation" menu → Display → Format
display
1 1
5 5
A0013993-EN A0016327-EN
1 Navigation view
2 Navigation path to current position
3 Status area
4 Display area for navigation
5 Operating elements → 45
Navigation path
The navigation path - displayed at the top left in the navigation view - consists of the
following elements:
↓ ↓ ↓
/ ../ Display
Endress+Hauser 41
Operation options Proline Promag H 200 HART
For more information about the menu icons, refer to the "Display area" section
→ 42
Status area
The following appears in the status area of the navigation view in the top right corner:
• Of the submenu
– The direct access code for the parameter you are navigating to (e.g. 0022-1)
– If a diagnostic event is present, the diagnostic behavior and status signal
• In the wizard
If a diagnostic event is present, the diagnostic behavior and status signal
• For information on the diagnostic behavior and status signal → 98
• For information on the function and entry of the direct access code → 47
Display area
Menus
Symbol Meaning
Operation
Appears:
• In the menu next to the "Operation" selection
• At the left in the navigation path in the "Operation" menu
Setup
Appears:
• In the menu next to the "Setup" selection
• At the left in the navigation path in the "Setup" menu
Diagnostics
Appears:
• In the menu next to the "Diagnostics" selection
• At the left in the navigation path in the "Diagnostics" menu
Expert
Appears:
• In the menu next to the "Expert" selection
• At the left in the navigation path in the "Expert" menu
Submenu
Wizard
Parameter locked
When displayed in front of a parameter name, indicates that the parameter is locked.
• By a user-specific access code
• By the hardware write protection switch
42 Endress+Hauser
Proline Promag H 200 HART Operation options
Wizard operation
Symbol Meaning
2 20
0 1 2 3 4
5 6 7 8 9
3
A0013941 A0013999
1 Editing view
2 Display area of the entered values
3 Input mask
4 Operating elements → 45
Input mask
The following input symbols are available in the input mask of the numeric and text editor:
Numeric editor
Symbol Meaning
Endress+Hauser 43
Operation options Proline Promag H 200 HART
Text editor
Symbol Meaning
Toggle
Aa1 • Between upper-case and lower-case letters
• For entering numbers
• For entering special characters
44 Endress+Hauser
Proline Promag H 200 HART Operation options
Key Meaning
Minus key
In a menu, submenu
Moves the selection bar upwards in a choose list.
With a Wizard
Confirms the parameter value and goes to the previous parameter.
With a text and numeric editor
In the input mask, moves the selection bar to the left (backwards).
Plus key
In a menu, submenu
Moves the selection bar downwards in a choose list.
With a Wizard
Confirms the parameter value and goes to the next parameter.
With a text and numeric editor
Moves the selection bar to the right (forwards) in an input screen.
Enter key
For operational display
• Pressing the key briefly opens the operating menu.
• Pressing the key for 2 s opens the context menu.
In a menu, submenu
• Pressing the key briefly:
– Opens the selected menu, submenu or parameter.
– Starts the wizard.
– If help text is open, closes the help text of the parameter.
• Pressing the key for 2 s for parameter:
If present, opens the help text for the function of the parameter.
With a Wizard
Opens the editing view of the parameter.
With a text and numeric editor
• Pressing the key briefly:
– Opens the selected group.
– Carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.
Plus/Enter key combination (press and hold down the keys simultaneously)
+
Increases the contrast (darker setting).
Endress+Hauser 45
Operation options Proline Promag H 200 HART
• Setup
• Conf. backup disp.
• Simulation
A0016326-EN
2. Press + simultaneously.
The context menu is closed and the operational display appears.
46 Endress+Hauser
Proline Promag H 200 HART Operation options
XXXXXXXXX
20.50
XX XX
Main menu 0104-1
Language
English
Operation
Setup
Main menu
Language
Operation
Setup
/ ../Operation 0091-1
Access stat.disp
Operator
Locking status
Display
/ ../Operation
Locking status
Display
/ ../Display 0098-1
Format display
1 value, max.
Contrast display
Display intervall
/ ../Format display 0098-1
1 value, max.
Bargr. + 1 value
2 values
Val. large+2val.
/ ../Format display 0098-1
1 value, max.
Bargr. + 1 value
2 values
Val. large+2val.
XXXXXXXXX
10.50
2s XX
19.00
XX
A0014010-EN
Endress+Hauser 47
Operation options Proline Promag H 200 HART
The direct access code consists of a 4-digit number and the channel number, which
identifies the channel of a process variable: e.g. 0914-1. In the navigation view, this
appears on the right-hand side in the header of the selected parameter.
0914-2 1
A0017223
A0014002-EN
2. Press + simultaneously.
The help text is closed.
48 Endress+Hauser
Proline Promag H 200 HART Operation options
001-FT-101 001-FT-101
ABC DEFG HIJK ABC DEFG HIJK
LMNO PQRS TUVW
3x XYZ Aa1@
001-FT-10
ABC DEFG HIJK
001-FT-10 001-FT-10
ABC DEFG HIJK ABC DEFG HIJK
LMNO PQRS TUVW
1x XYZ Aa1@
001-FT-10 001-FT-10
ABC DEFG HIJK ABC DEFG HIJK
LMNO PQRS TUVW
A a 1 @
1x XYZ Aa1@
001-FT-10 001-FT-10
ABC DEFG HIJK 012 3456 789
=+-* /[] ()
A a 1 @
2x <>{} Aa1@
001-FT-102 001-FT-10
012 3456 789 012 3456 789
=+-* /[] ()
0 1 2 0 1 2
4x <>{} Aa1@
001-FT-10 001-FT-102
012 3456 789 012 3456 789
=+-* /[] ()
0 1 2 0 1 2
2x <>{} Aa1@
A0014020-EN
A message is displayed if the value entered is outside the permitted value range.
Endress+Hauser 49
Operation options Proline Promag H 200 HART
Without access code With access code Without access code With access code
(from the factory) (from the factory)
Operator -- 1)
Maintenance
1) Despite the defined access code, certain parameters can always be modified and thus are excepted from
the write protection, as they do not affect the measurement. Refer to the "Write protection via access code"
section
If an incorrect access code is entered, the user obtains the access rights of the "Operator"
role.
The user role with which the user is currently logged on is indicated by the Access
status display parameter. Navigation path: Operation → Access status display
The locking of the write access via local operation can be disabled by entering the
customer-defined access code via the respective access option.
1. After you press , the input prompt for the access code appears.
2. Enter the access code.
The -symbol in front of the parameters disappears; all previously write-
protected parameters are now re-enabled.
50 Endress+Hauser
Proline Promag H 200 HART Operation options
If the user attempts to access the operating menu while the keypad lock is active, the
message Keylock on appears.
The keypad lock is switched on and off via the context menu.
If the user attempts to access the operating menu while the keypad lock is active, the
message Keylock on appears.
Endress+Hauser 51
Operation options Proline Promag H 200 HART
6 8
3
2
1 4 5 7
A0013764
3 –
+
E
2 1
A0014019
1 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device
2 Commubox FXA291
3 Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
52 Endress+Hauser
Proline Promag H 200 HART Operation options
Function scope
Field Xpert SFX350 and Field Xpert SFX370 are mobile computers for commissioning and
maintenance. They enable efficient device configuration and diagnostics for HART and
FOUNDATION fieldbus devices in the non-Ex area (SFX350, SFX370) and the Ex area
(SFX370).
For details, see Operating Instructions BA01202S
8.4.3 FieldCare
Function scope
FDT-based plant asset management tool from Endress+Hauser. It can configure all smart
field devices in a system and helps you manage them. By using the status information, it is
also a simple but effective way of checking their status and condition.
Access takes place via:
• HART protocol → 52
• Service interface CDI → 52
Typical functions:
• Configuring parameters of transmitters
• Loading and saving device data (upload/download)
• Documentation of the measuring point
• Visualization of the measured value memory (line recorder) and event logbook
For details, see Operating Instructions BA00027S and BA00059S
Endress+Hauser 53
Operation options Proline Promag H 200 HART
User interface
2 3 4 5 6 7
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx
1 Volume flow: 12.34 m³/h
Status: Good
10 11
A0021051-EN
1 Header
2 Picture of device
3 Device name
4 Device tag
5 Status area with status signal
6 Display area for current measured values
7 Event list with additional functions such as save/load, events list and document creation
8 Navigation area with operating menu structure
9 Operating range
10 Range of action
11 Status area
Function scope
Program from Emerson Process Management for operating and configuring measuring
devices via HART protocol.
Function scope
SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for
the operation, configuration, maintenance and diagnosis of intelligent field devices via
HART protocol.
54 Endress+Hauser
Proline Promag H 200 HART Operation options
Function scope
Industrial handheld terminal from Emerson Process Management for remote
configuration and measured value display via HART protocol.
Endress+Hauser 55
System integration Proline Promag H 200 HART
9 System integration
56 Endress+Hauser
Proline Promag H 200 HART System integration
The assignment of the measured variables to the dynamic variables can be modified and
assigned as desired via local operation and the operating tool using the following
parameters:
• Expert → Communication → HART output → Output → Assign PV
• Expert → Communication → HART output → Output → Assign SV
• Expert → Communication → HART output → Output → Assign TV
• Expert → Communication → HART output → Output → Assign QV
‣ Burst configuration
‣ Burst configuration 1 to 3
Burst mode 1 to 3
Burst command 1 to 3
Burst variable 0
Burst variable 1
Burst variable 2
Burst variable 3
Burst variable 4
Burst variable 5
Endress+Hauser 57
System integration Proline Promag H 200 HART
Burst variable 6
Burst variable 7
Burst mode 1 to 3 Activate the HART burst mode for burst • Off Off
message X. • On
Burst command 1 to 3 Select the HART command that is sent to the • Command 1 Command 2
HART master. • Command 2
• Command 3
• Command 9
• Command 33
• Command 48
Burst variable 0 For HART command 9 and 33, assign a • Volume flow Volume flow
HART device variable or process variable to • Mass flow
burst variable. • Totalizer 1
• Totalizer 2
• Totalizer 3
• Percent Of Range
• Measured current
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
• Not used
Burst variable 1 For HART command 9 and 33, assign a See Burst variable 0 Not used
HART device variable or process variable to parameter.
burst variable.
Burst variable 2 For HART command 9 and 33, assign a See Burst variable 0 Not used
HART device variable or process variable to parameter.
burst variable.
Burst variable 3 For HART command 9 and 33, assign a See Burst variable 0 Not used
HART device variable or process variable to parameter.
burst variable.
Burst variable 4 For HART command 33, assign a HART See Burst variable 0 Not used
device variable or process variable to burst parameter.
variable.
Burst variable 5 For HART command 33, assign a HART See Burst variable 0 Not used
device variable or process variable to burst parameter.
variable.
Burst variable 6 For HART command 33, assign a HART See Burst variable 0 Not used
device variable or process variable to burst parameter.
variable.
Burst variable 7 For HART command 33, assign a HART See Burst variable 0 Not used
device variable or process variable to burst parameter.
variable.
58 Endress+Hauser
Proline Promag H 200 HART System integration
Burst trigger mode Select the event that triggers burst message • Continuous Continuous
X. • Window
• Rising
• Falling
• On change
Burst trigger level Enter the burst trigger value. Positive floating-point number 2.0E-38
Together with the option selected in the
Burst trigger mode parameter the burst
trigger value determines the time of burst
message X.
Min. update period Enter the minimum time span between two Positive integer 1 000 ms
burst responses of one burst message.
Max. update period Enter the maximum time span between two Positive integer 2 000 ms
burst responses of one burst message.
Endress+Hauser 59
Commissioning Proline Promag H 200 HART
10 Commissioning
XXXXXXXXX
20.50
XX XX
Main menu 0104-1
Language
English
Operation
Setup
Language 0104-1
à English
Deutsch
Español
Français
Language 0104-1
à English
Deutsch
Español
Français
Hauptmenü 0104-1
Sprache
Deutsch
Betrieb
Setup
A0013996
60 Endress+Hauser
Proline Promag H 200 HART Commissioning
XXXXXXXXX
20.50
mA
Main menu 0104-1
Language
English
Display/operat.
Setup
Main menu
Display/operat.
Setup
Diagnostic
/ ../Setup
Curr.output 1
XXXXXXXXX
XXXXXXXXX
A0018774-EN
Navigation
"Setup" menu
Setup
Device tag → 62
‣ System units → 62
‣ Current output 1 → 64
‣ Pulse/frequency/switch output → 66
‣ Display → 72
‣ Output conditioning → 74
‣ Advanced setup → 78
Endress+Hauser 61
Commissioning Proline Promag H 200 HART
1 XXXXXXXXX
A0013375
Navigation
"Setup" menu → Device tag
Device tag Enter the name for the measuring point. Max. 32 characters, such as Promag
letters, numbers or special
characters (e.g. @, %, /).
Navigation
"Setup" menu → Advanced setup → System units
‣ System units
Volume unit
Mass unit
Density unit
Temperature unit
62 Endress+Hauser
Proline Promag H 200 HART Commissioning
Volume flow unit Select volume flow unit. Unit choose list Country-specific:
• l/h
Result
• gal/min (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Mass flow unit Select mass flow unit. Unit choose list Country-specific:
• kg/h
Result
• lb/min
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Endress+Hauser 63
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Current output 1
Assign curr.
Density unit
Fixed density
Current span
4 mA value
20 mA value
Failure mode
Failure current
End of wizard
A0018498-EN
64 Endress+Hauser
Proline Promag H 200 HART Commissioning
Assign current output – Select process variable for • Off Volume flow
current output. • Volume flow
• Mass flow
Mass flow unit – Select mass flow unit. Unit choose list Country-specific:
• kg/h
Result
• lb/min
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Volume flow unit – Select volume flow unit. Unit choose list Country-specific:
• l/h
Result
• gal/min (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
4 mA value One of the following options is Enter 4 mA value. Signed floating-point Country-specific:
selected in the Current span number • 0 m3/h
parameter (→ 65): • 0 ft3/min
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
20 mA value One of the following options is Enter 20 mA value. Signed floating-point Depends on country
selected in the Current span number and nominal
parameter (→ 65): diameter
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
Failure mode One of the following options is Define output behavior in • Min. Max.
selected in the Assign current alarm condition. • Max.
output parameter (→ 65): • Last valid value
• Volume flow • Actual value
• Mass flow • Defined value
One of the following options is
selected in the Current span
parameter (→ 65):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
Failure current The Defined value option is Enter current output value in 3.59 to 22.5 mA 22.5 mA
selected in the Failure mode alarm condition.
parameter.
Endress+Hauser 65
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Pulse/frequency/switch output
Operating mode
Pulse
Assign pulse
Density unit
Fixed density
Pulse width
Failure mode
Invert outp.sig.
End of wizard
A0018551-EN
19 "Pulse/frequency/switch output" wizard in the "Setup" menu: "Operating mode" parameter"Pulse" option
Assign pulse output Select process variable for pulse output. • Off Off
• Mass flow
• Volume flow
66 Endress+Hauser
Proline Promag H 200 HART Commissioning
Value per pulse Enter measured value at which a pulse is Signed floating-point number 0
output.
Pulse width Define time width of the output pulse. 5 to 2 000 ms 100 ms
Failure mode Define output behavior in alarm condition. • Actual value No pulses
• No pulses
Endress+Hauser 67
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Pulse/frequency/switch output
Operating mode
Frequency
Assign freq.
Density unit
Fixed density
Val. at min.freq
Val. at max.freq
Failure mode
Failure freq.
Invert outp.sig.
End of wizard
A0018557-EN
20 "Pulse/frequency/switch output" wizard in the "Setup" menu: "Operating mode" parameter"Frequency" option
68 Endress+Hauser
Proline Promag H 200 HART Commissioning
Assign frequency output Select process variable for frequency output. • Off Off
• Volume flow
• Mass flow
Mass flow unit Select mass flow unit. Unit choose list Country-specific:
• kg/h
Result
• lb/min
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Volume flow unit Select volume flow unit. Unit choose list Country-specific:
• l/h
Result
• gal/min (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Measuring value at minimum Enter measured value for minmum Signed floating-point number 0
frequency frequency.
Measuring value at maximum Enter measured value for maximum Signed floating-point number 0
frequency frequency.
Failure frequency Enter frequency output value in alarm 0.0 to 1 250.0 Hz 0.0 Hz
condition.
Endress+Hauser 69
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Pulse/frequency/switch output
Operating
mode
Switch
Switch
out funct
Diag.
Off On
behavior
Assign
diag. beh
Alarm or
Alarm Warning
warning
Invert outp.sig.
End of wizard
A0018575-EN
21 "Pulse/frequency/switch output" wizard in the "Setup" menu: "Operating mode" parameter"Switch" option
(part 1)
70 Endress+Hauser
Proline Promag H 200 HART Commissioning
Operating
mode
Switch
Switch
out funct
Fl. direct.
Limit Status
check
Volume Mass
Unit
flow unit flow unit
Switch-on value
Switch-off
value
Switch-on
delay
Switch-off
delay
Failure mode
Invert
outp.sig.
End of wizard
A0018576-EN
22 "Pulse/frequency/switch output" wizard in the "Setup" menu: "Operating mode" parameter"Switch" option
(part 2)
Switch output function Select function for switch output. • Off Off
• On
• Diagnostic behavior
• Limit
• Flow direction check
• Status
Assign diagnostic behavior Select diagnostic behavior for switch output. • Alarm Alarm
• Alarm or warning
• Warning
Assign limit Select process variable for limit function. • Volume flow Volume flow
• Mass flow
• Totalizer 1
• Totalizer 2
• Totalizer 3
Endress+Hauser 71
Commissioning Proline Promag H 200 HART
Assign flow direction check Select process variable for flow direction • Volume flow Volume flow
monitoring. • Mass flow
Assign status Select device status for switch output. • Empty pipe detection Empty pipe detection
• Low flow cut off
Volume flow unit Select volume flow unit. Unit choose list Country-specific:
• l/h
Result
• gal/min (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Mass flow unit Select mass flow unit. Unit choose list Country-specific:
• kg/h
Result
• lb/min
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit totalizer Select process variable totalizer unit. Unit choose list l
Switch-on value Enter measured value for the switch-on Signed floating-point number 0 l/h
point.
Switch-off value Enter measured value for the switch-off Signed floating-point number 0 l/h
point.
Switch-on delay Define delay for the switch-on of status 0.0 to 100.0 s 0.0 s
output.
Switch-off delay Define delay for the switch-off of status 0.0 to 100.0 s 0.0 s
output.
Failure mode Define output behavior in alarm condition. • Actual status Open
• Open
• Closed
72 Endress+Hauser
Proline Promag H 200 HART Commissioning
Navigation
"Setup" menu → Display
Format display
Value 1 display
0% bargraph
value 1
100% bargraph
value 1
Value 2 display
Value 3 display
0% bargraph
value 3
100% bargraph
value 3
Value 4 display
End of wizard
A0013797-EN
Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size
are shown on the display. • 1 bargraph + 1
value
• 2 values
• 1 value large + 2
values
• 4 values
Value 1 display A local display is provided. Select the measured value that • Volume flow Volume flow
is shown on the local display. • Mass flow
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Current output 1
Endress+Hauser 73
Commissioning Proline Promag H 200 HART
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point 0 l/h
display. number
100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country
graph display. number and nominal
diameter
Value 2 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
Value 3 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
0% bargraph value 3 An option was selected in the Enter 0% value for bar graph Signed floating-point 0
Value 3 display parameter. display. number
100% bargraph value 3 An option was selected in the Enter 100% value for bar Signed floating-point Depends on country
Value 3 display parameter. graph display. number and nominal
diameter
Value 4 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
Navigation
"Setup" menu → Output conditioning
Display damping
Damping out. 1
Damping out. 2
Mode output 1
Mode output 2
End of wizard
A0018583-EN
74 Endress+Hauser
Proline Promag H 200 HART Commissioning
Damping output 2 The measuring device has a Set the reaction time of the 0 to 999.9 s 1s
second current output. output signal of the second
current output to fluctuations
in the measured value.
Damping output 2 The measuring device has a Set the reaction time of the 0 to 999.9 s 1s
pulse/frequency/switch output signal of the frequency
output. output to fluctuations in the
measured value.
Measuring mode output 1 – Select measuring mode for • Forward flow Forward flow
output. • Forward/Reverse
flow
• Reverse flow
compensation
Measuring mode output 2 – Select measuring mode for • Forward flow Forward flow
output. • Forward/Reverse
flow
• Reverse flow
compensation
Measuring mode output 2 – Select measuring mode for • Forward flow Forward flow
output. • Forward/Reverse
flow
• Reverse flow
• Reverse flow
compensation
Measuring mode output 2 – Select measuring mode for • Forward flow Forward flow
output. • Forward/Reverse
flow
• Reverse flow
• Reverse flow
compensation
Endress+Hauser 75
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Low flow cut off
Assign variable
On value
Off value
End of wizard
A0017209-EN
Assign process variable Select process variable for low flow cut off. • Off Volume flow
• Volume flow
• Mass flow
On value low flow cutoff Enter on value for low flow cut off. Signed floating-point number 0 l/h
Off value low flow cutoff Enter off value for low flow cut off. 0 to 100.0 % 50 %
Pressure shock suppression Enter time frame for signal suppression (= 0 to 100 s 0s
active pressure shock suppression).
76 Endress+Hauser
Proline Promag H 200 HART Commissioning
Navigation
"Setup" menu → Empty pipe detection
Off On
New adjustment
Progress
Ok Busy Not ok
Response time
End of wizard
A0017210-EN
Response time empty pipe detection – Enter the time before 0 to 100 s 1s
diagnostic message S862 ' Pipe
empty'' is displayed for empty
pipe detection.
Endress+Hauser 77
Commissioning Proline Promag H 200 HART
XXXXXXXXX
20.50
mA
Main menu 0104-1
Language
English
Display/operat.
Setup
Main menu
Display/operat.
Setup
Diagnostic
/ ../Setup
Curr.output 1
XXXXXXXXX
XXXXXXXXX
/ ../Setup
XXXXXXXXX
XXXXXXXXX
Advanced setup
A0018745-EN
Navigation
"Setup" menu → Advanced setup
‣ Advanced setup
‣ Sensor adjustment → 79
‣ Totalizer 1 to 3 → 79
‣ SIL confirmation
78 Endress+Hauser
Proline Promag H 200 HART Commissioning
‣ Deactivate SIL
‣ Display → 81
‣ Heartbeat setup
‣ Administration → 83
Navigation
"Setup" menu → Advanced setup → Sensor adjustment
‣ Sensor adjustment
Installation direction
Installation direction Set sign of flow direction to match the • Flow in arrow direction Flow in arrow direction
direction of the arrow on the sensor. • Flow against arrow direction
Navigation
"Setup" menu → Advanced setup → Totalizer 1 to 3
‣ Totalizer 1 to 3
Unit totalizer
Failure mode
Endress+Hauser 79
Commissioning Proline Promag H 200 HART
Assign process variable Select process variable for totalizer. • Off Volume flow
• Volume flow
• Mass flow
Unit totalizer Select process variable totalizer unit. Unit choose list l
Totalizer operation mode Select totalizer calculation mode. • Net flow total Net flow total
• Forward flow total
• Reverse flow total
80 Endress+Hauser
Proline Promag H 200 HART Commissioning
Navigation
"Setup" menu → Advanced setup → Display
‣ Display
Format display
Value 1 display
0% bargraph value 1
Decimal places 1
Value 2 display
Decimal places 2
Value 3 display
0% bargraph value 3
Decimal places 3
Value 4 display
Decimal places 4
Language
Display interval
Display damping
Header
Header text
Separator
Backlight
Endress+Hauser 81
Commissioning Proline Promag H 200 HART
Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size
are shown on the display. • 1 bargraph + 1
value
• 2 values
• 1 value large + 2
values
• 4 values
Value 1 display A local display is provided. Select the measured value that • Volume flow Volume flow
is shown on the local display. • Mass flow
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Current output 1
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point 0 l/h
display. number
100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country
graph display. number and nominal
diameter
Decimal places 1 A measured value is specified Select the number of decimal • x x.xx
in the Value 1 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
Value 2 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
Decimal places 2 A measured value is specified Select the number of decimal • x x.xx
in the Value 2 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
Value 3 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
0% bargraph value 3 An option was selected in the Enter 0% value for bar graph Signed floating-point 0
Value 3 display parameter. display. number
100% bargraph value 3 An option was selected in the Enter 100% value for bar Signed floating-point Depends on country
Value 3 display parameter. graph display. number and nominal
diameter
Decimal places 3 A measured value is specified Select the number of decimal • x x.xx
in the Value 3 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
Value 4 display A local display is provided. Select the measured value that Picklist (see Value 1 None
is shown on the local display. display)
Decimal places 4 A measured value is specified Select the number of decimal • x x.xx
in the Value 4 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
82 Endress+Hauser
Proline Promag H 200 HART Commissioning
Display interval A local display is provided. Set time measured values are 1 to 10 s 5s
shown on display if display
alternates between values.
Display damping A local display is provided. Set display reaction time to 0.0 to 999.9 s 0.0 s
fluctuations in the measured
value.
Header A local display is provided. Select header contents on local • Device tag Device tag
display. • Free text
Header text The Free text option is Enter display header text. Max. 12 characters ------------
selected in the Header such as letters,
parameter. numbers or special
characters (e.g. @,
%, /)
Navigation
"Setup" menu → Advanced setup → Administration
‣ Administration
Device reset
Endress+Hauser 83
Commissioning Proline Promag H 200 HART
Navigation
"Setup" menu → Advanced setup → Configuration backup display
Operating time
Last backup
Configuration management
Comparison result
Operating time – Indicates how long the device Days (d), hours (h), –
has been in operation. minutes (m), seconds
(s)
Last backup – Indicates when the last data Days (d), hours (h), –
backup was saved to the minutes (m), seconds
display module. (s)
84 Endress+Hauser
Proline Promag H 200 HART Commissioning
Configuration management A local display is provided. Select action for managing the • Cancel Cancel
device data in the display • Execute backup
module. • Restore
• Duplicate
• Compare
• Clear backup data
Comparison result – Comparison between present • Settings identical Check not done
device data and display • Settings not
backup. identical
• No backup
available
• Backup settings
corrupt
• Check not done
• Dataset
incompatible
Options Description
Execute backup The current device configuration is backed up from the integrated HistoROM to the
device's display module. The backup copy includes the transmitter data of the
device.
Restore The last backup copy of the device configuration is restored from the display
module to the device's integrated HistoROM. The backup copy includes the
transmitter data of the device.
Duplicate The transmitter configuration from another device is duplicated to the device using
the display module.
Compare The device configuration saved in the display module is compared with the current
device configuration of the integrated HistoROM.
Clear backup data The backup copy of the device configuration is deleted from the display module of
the device.
Integrated HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.7 Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Navigation
"Diagnostics" menu → Simulation
‣ Simulation
Endress+Hauser 85
Commissioning Proline Promag H 200 HART
Frequency value → 86
Pulse value → 86
Switch status → 87
Assign simulation process variable – Select a process variable for • Off Off
the simulation process that is • Volume flow
activated. • Mass flow
Process variable value A process variable is selected Enter the simulation value for Signed floating-point 0
in the Assign simulation the selected process variable. number
process variable parameter.
Value current output 1 The On option is selected in Enter the current value for 3.59 to 22.5 mA –
the Simulation current simulation.
output 1 parameter
(→ 86).
Frequency value The On option is selected in Enter the frequency value for 0.0 to 1 250.0 Hz 0.0 Hz
the Frequency output simulation.
simulation parameter.
Pulse output simulation The Down-count. val. option Switch simulation of the pulse • Off Off
is selected in the Simulation output on and off. • Fixed value
pulse output parameter. • Down-counting
Ifis the Fixed value option
selected, the Pulse value
width parameter defines
the pulse width of the
pulses output.
Pulse value The Down-count. val. option Enter the number of pulses for 0 to 65 535 0
is selected in the Simulation simulation.
pulse output parameter.
86 Endress+Hauser
Proline Promag H 200 HART Commissioning
Switch status The On option is selected in Select the status of the status • Open Open
the Switch output simulation output for the simulation. • Closed
parameter.
Simulation device alarm – Switch the device alarm on and • Off Off
off. • On
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
The device automatically locks the write-protected parameters again if a key is not pressed
for 10 minutes in the navigation and editing view. The device locks the write-protected
Endress+Hauser 87
Commissioning Proline Promag H 200 HART
parameters automatically after 60 s if the user skips back to the operational display mode
from the navigation and editing view.
• If write access is activated via access code, it can be also be deactivated only via the
access code → 50.
• The user role with which the user is currently logged on via the local display
→ 50 is indicated by the Access status display parameter. Navigation path:
"Operation" menu → Access status display
↓ ↓
OFF
CDI WP SIM
ON
XXXXXXXXXXXXX
–
+
E
3 mm
A0013768
88 Endress+Hauser
Proline Promag H 200 HART Commissioning
3. Pull out the display module with a gentle rotational movement. To make it easier to
access the lock switch, attach the display module to the edge of the electronics
compartment.
Display module is attached to the edge of the electronics compartment.
A0013909
4. Setting the write protection switch (WP) on the main electronics module to the ON
position enables hardware write protection. Setting the write protection switch (WP)
on the main electronics module to the OFF position (factory setting) disables
hardware write protection.
If hardware write protection is enabled, the Locking status parameter displays
the Hardware locked option . In addition, on the local display the -symbol
appears in front of the parameters in the header of the operational display and in
the navigation view.
XXXXXXXXX
20.50
XX XX
A0015870
Endress+Hauser 89
Operation Proline Promag H 200 HART
11 Operation
Navigation
"Operation" menu → Locking status
None The access status displayed in "Access status display" parameter applies → 50.
Only appears on local display.
Hardware locked The DIP switch for hardware locking is activated on the main electronics module.
This locks write access to the parameters .
Temporarily locked Write access to the parameters is temporarily lock due to device-internal
processing (e.g. data upload/download, reset). Once the internal processing has
been completed, the parameters can be changed once again.
Navigation
"Diagnostics" menu → Measured values → Process variables
‣ Process variables
Volume flow
Mass flow
90 Endress+Hauser
Proline Promag H 200 HART Operation
Volume flow Displays the volume flow currently measured. Signed floating-point number
Mass flow Displays the mass flow currently calculated. Signed floating-point number
11.4.2 Totalizer
The Totalizer submenu contains all the parameters needed to display the current
measured values for every totalizer.
Navigation
"Diagnostics" menu → Measured values → Totalizer
‣ Totalizer
Totalizer value 1 to 3
Totalizer overflow 1 to 3
Totalizer value 1 to 3 In the Assign process variable Displays the current totalizer Signed floating-point 0l
parameter in the Totalizer 1 counter value. number
to 3 submenu, one of the
following options is selected:
• Volume flow
• Mass flow
Totalizer overflow 1 to 3 In the Assign process variable Displays the current totalizer Integer with sign 0
parameter in the Totalizer 1 overflow.
to 3 submenu, one of the
following options is selected:
• Volume flow
• Mass flow
Navigation
"Diagnostics" menu → Measured values → Output values
‣ Output values
Output current 1
Measured current 1
Terminal voltage 1
Endress+Hauser 91
Operation Proline Promag H 200 HART
Output current 2
Pulse output
Output frequency
Switch status
Output current 1 Displays the current value currently 3.59 to 22.5 mA 3.59 mA
calculated for the current output.
Terminal voltage 1 Displays the current terminal voltage that is 0.0 to 50.0 V 0V
applied at the current output.
Output current 2 Displays the current value currently 3.59 to 22.5 mA 3.59 mA
calculated for the current output.
Pulse output Displays the value currently measured for Positive floating-point number 0 Hz
the pulse output.
Output frequency Displays the value currently measured for 0.0 to 1 250.0 Hz 0.0 Hz
the frequency output.
Switch status Displays the current switch output status. • Open Open
• Closed
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.
Preset + totalize The totalizer is set to the defined start value in Preset value parameterand the
totaling process is restarted.
92 Endress+Hauser
Proline Promag H 200 HART Operation
Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
Navigation
"Operation" menu → Operation
‣ Totalizer handling
Control Totalizer 1 to 3
Preset value 1 to 3
Preset value 1 to 3 Specify start value for totalizer. Signed floating-point number 0l
Reset all totalizers Reset all totalizers to 0 and start. • Cancel Cancel
• Reset + totalize
A0016222
Endress+Hauser 93
Operation Proline Promag H 200 HART
• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.
If the length of the logging interval or the assignment of the process variables to the
channels is changed, the content of the data logging is deleted.
Navigation
"Diagnostics" menu → Data logging
‣ Data logging
Assign channel 1
Assign channel 2
Assign channel 3
Assign channel 4
Logging interval
‣ Display channel 1
‣ Display channel 2
‣ Display channel 3
‣ Display channel 4
94 Endress+Hauser
Proline Promag H 200 HART Operation
Assign channel 1 to 4 The Extended HistoROM Assign process variable to • Off Off
application package is logging channel. • Volume flow
available. • Mass flow
• Electronic
The software options
currently enabled are temperature
• Current output 1
displayed in the
• Current difference
Software option
potential
overview parameter.
Logging interval The Extended HistoROM Define the logging interval for 1.0 to 3 600.0 s 10.0 s
application package is data logging. This value
available. defines the time interval
between the individual data
The software options
currently enabled are points in the memory.
displayed in the
Software option
overview parameter.
Clear logging data The Extended HistoROM Clear the entire logging data. • Cancel Cancel
application package is • Clear data
available.
Endress+Hauser 95
Diagnostics and troubleshooting Proline Promag H 200 HART
Local display dark and no output Supply voltage does not match that Apply the correct supply voltage .
signals specified on the nameplate.
Local display dark and no output Supply voltage has incorrect Reverse polarity of supply voltage.
signals polarity.
Local display dark and no output No contact between connecting Check the connection of the cables
signals cables and terminals. and correct if necessary.
Local display dark and no output Terminals are not plugged into the Check terminals.
signals I/O electronics module correctly.
Local display dark and no output I/O electronics module is defective. Order spare part → 113.
signals
Local display is dark, but signal Display is set too bright or too dark. • Set the display brighter by
output is within the valid range simultaneously pressing + .
• Set the display darker by
simultaneously pressing + .
Local display is dark, but signal The cable of the display module is Insert the plug correctly into the
output is within the valid range not plugged in correctly. main electronics module and
display module.
Local display is dark, but signal Display module is defective. Order spare part → 113.
output is within the valid range
Backlighting of local display is red Diagnostic event with "Alarm" Take remedial measures → 103
diagnostic behavior has occurred.
Text on local display appears in a Incorrect operating language is 1. Press + for 2 s ("home
foreign language and cannot be configured. position").
understood. 2. Press .
3. Set the desired language in the
Language parameter.
Message on local display: Communication between the • Check the cable and the
"Communication Error" display module and the electronics connector between the main
"Check Electronics" is interrupted. electronics module and display
module.
• Order spare part → 113.
Signal output outside the valid Main electronics module is Order spare part → 113.
range defective.
Signal output outside the valid I/O electronics module is defective. Order spare part → 113.
current range
(< 3.6 mA or > 22 mA)
Device shows correct value on local Configuration error Check and correct parameter
display, but signal output is configuration.
incorrect, though in the valid range.
Device measures incorrectly. Configuration error or device is 1. Check and correct parameter
operated outside the application. configuration.
2. Observe limit values specified in
the "Technical Data".
96 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
For access
Problem Possible causes Remedial action
No write access to parameters Hardware write protection enabled Set the write protection switch on
the main electronics module to the
OFF position .
No write access to parameters Current user role has limited access 1. Check user role → 50.
authorization 2. Enter correct customer-specific
access code → 50.
No connection via HART protocol Missing or incorrectly installed Install the communication resistor
communication resistor. (250 Ω) correctly. Observe the
maximum load → 29.
No connection via HART protocol Commubox Observe the documentation for the
• Connected incorrectly Commubox.
• Configured incorrectly
• Drivers not installed correctly FXA195 HART: Document
"Technical Information"
• USB interface on computer TI00404F
configured incorrectly
No connection via service interface Incorrect configuration of USB Observe the documentation for the
interface on PC or driver not Commubox.
installed correctly.
FXA291: Document
"Technical Information"
TI00405C
Endress+Hauser 97
Diagnostics and troubleshooting Proline Promag H 200 HART
2 1
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
3
4
X XX i Menu
5
A0013939-EN
1 Status signal
2 Diagnostic behavior
3 Diagnostic behavior with diagnostic code
4 Short text
5 Operating elements
If two or more diagnostic events are pending simultaneously, only the message of the
diagnostic event with the highest priority is shown.
Other diagnostic events that have occurred can be called up in the Diagnostics menu:
• Via parameters → 105
• Via submenus → 106
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
The status signals are categorized according to VDI/VDE 2650 and NAMUR
Recommendation NE 107: F = Failure, C = Function Check, S = Out of Specification, M
= Maintenance Required
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
A0013956
Function check
The device is in service mode (e.g. during a simulation).
A0013959
Out of specification
The device is operated:
• Outside its technical specification limits (e.g. outside the process temperature range)
A0013958 • Outside of the configuration carried out by the user (e.g. maximum flow in parameter
20 mA value)
Maintenance required
Maintenance is required. The measured value remains valid.
A0013957
98 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Diagnostic behavior
Symbol Meaning
Alarm
• Measurement is interrupted.
• Signal outputs and totalizers assume the defined alarm condition.
A0013961 • A diagnostic message is generated.
• For local display with touch control: the background lighting changes to red.
Warning
Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic
A0013962 message is generated.
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault. In addition, the corresponding symbol for the
diagnostic behavior is displayed in front of the diagnostic information on the local display.
Diagnostic information
Diagnostic code
Diagnostic Diagnostic
Status signal Short text
behavior number
↓ ↓ ↓
Operating elements
Key Meaning
Plus key
In a menu, submenu
A0013970
Opens the message about the remedial measures.
Enter key
In a menu, submenu
A0013952
Opens the operating menu.
Endress+Hauser 99
Diagnostics and troubleshooting Proline Promag H 200 HART
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
X XX i Menu
Diagnostic list S
Diagnostics 1
1 S801 Supply voltage
Diagnostics 2
Diagnostics 3
A0013940-EN
The user is in the Diagnostics menu at an entry for a diagnostics event, e.g. in the
Diagnostic list submenu or the Previous diagnostics parameter.
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message for the remedial measures closes.
100 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx Volume flow: 12.34 m³/h
Status signal: Function check (C)
Xxxxxx
P Diagnostics 1: C485 Simu...
P Remedy information: Deactivate...
P Access status tooling: Mainenance Failure (F)
A0021799-EN
Furthermore, diagnostic events that have occurred can be viewed in the Diagnostics
menu:
• Via parameters → 105
• Via submenu → 106
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault. In addition, the corresponding symbol for the
diagnostic behavior is displayed in front of the diagnostic information on the local display.
Diagnostic information
Diagnostic code
Diagnostic Diagnostic
Status signal Short text
behavior number
↓ ↓ ↓
Endress+Hauser 101
Diagnostics and troubleshooting Proline Promag H 200 HART
A0014048-EN
You can assign the following options to the diagnostic number as the diagnostic behavior:
Options Description
Alarm Measurement is interrupted. Signal outputs and totalizers assume the defined alarm
condition. A diagnostic message is generated.
For local display with touch control: the background lighting changes to red.
Warning Measurement is resumed. The signal outputs and totalizers are not affected. A
diagnostics message is generated.
Logbook entry only The device continues to measure. The diagnostic message is entered in the Event
logbook (events list) submenu only and is not displayed in alternation with the measured
value display.
Off The diagnostic event is ignored, and no diagnostic message is generated or entered.
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
A0013956
Function check
The device is in service mode (e.g. during a simulation).
A0013959
102 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Symbol Meaning
Out of specification
The device is operated:
• Outside its technical specification limits (e.g. outside the process temperature range)
A0013958 • Outside of the configuration carried out by the user (e.g. maximum flow in parameter
20 mA value)
Maintenance required
Maintenance is required. The measured value remains valid.
A0013957
A0023076
Diagnostic of sensor
Diagnostic of electronic
Endress+Hauser 103
Diagnostics and troubleshooting Proline Promag H 200 HART
Diagnostic of configuration
Diagnostic of process
104 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Navigation
"Diagnostics" menu
Previous diagnostics
Operating time
Endress+Hauser 105
Diagnostics and troubleshooting Proline Promag H 200 HART
Actual diagnostics 1 diagnostic event has Displays the current diagnostic Symbol for diagnostic –
occurred. event along with the behavior, diagnostic
diagnostic information. code and short
message.
Ifoccur
two or more messages
simultaneously,
the message with the
highest priority is shown
on the display.
Previous diagnostics 2 diagnostic events have Displays the diagnostic event Symbol for diagnostic –
already occurred. that occurred prior to the behavior, diagnostic
current diagnostic event along code and short
with the diagnostic message.
information.
Operating time from restart – Shows the time the device has Days (d), hours (h),
been in operation since the last minutes (m), seconds
device restart. (s)
Operating time – Indicates how long the device Days (d), hours (h), –
has been in operation. minutes (m), seconds
(s)
/ ../Diagnose list
Diagnostics
F273 Main electronic
Diagnostics 2
Diagnostics 3
A0014006-EN
106 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
/ ../Eventlist F
I1091 Config. change
I1157 Mem.err. ev.list
0d01h19m10s
F311 Electr. failure
A0014008-EN
I1089 Power on
Endress+Hauser 107
Diagnostics and troubleshooting Proline Promag H 200 HART
108 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Navigation
"Setup" menu → Advanced setup → Administration → Device reset
‣ Administration
Device reset
Options Description
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Navigation
"Diagnostics" menu → Device information
‣ Device information
Device tag
Serial number
Firmware version
Endress+Hauser 109
Diagnostics and troubleshooting Proline Promag H 200 HART
Device name
Order code
ENP version
Device revision
Device ID
Device type
Manufacturer ID
Device tag Enter the name for the measuring point. Max. 32 characters such as Promag
letters, numbers or special
characters (e.g. @, %, /)
Serial number Shows the serial number of the measuring Max. 11-digit character string 79AFFF16000
device. comprising letters and
numbers.
Firmware version Shows the device firmware version installed. Character string with the 01.01.zz
following format:
xx.yy.zz
Device name Shows the name of the transmitter. Character string composed of Promag
letters, numbers and certain
The name can be found on the
nameplate of the transmitter. punctuation marks.
Order code Shows the device order code. Character string composed of –
letters, numbers and certain
The order code can be found on the
nameplate of the sensor and punctuation marks
transmitter in the "Order code" field.
Extended order code 1 Shows the 1st part of the extended order Character string –
code.
Extended order code 2 Shows the 2nd part of the extended order Character string –
code.
110 Endress+Hauser
Proline Promag H 200 HART Diagnostics and troubleshooting
Extended order code 3 Shows the 3rd part of the extended order Character string –
code.
ENP version Shows the version of the electronic Character string with the 2.02.00
nameplate (ENP). following format:
xx.yy.zz
Device revision Shows the device revision with which the 2-digit hexadecimal number 2
device is registered with the HART
Communication Foundation.
Device type Displays the device type with which the 0 to 255 0x48
measuring device is registered with the
HART Communication Foundation.
Flashing the firmware to the current version or to the previous version is possible via
the service interface (CDI) .
For the compatibility of the firmware version with the previous version, the installed
device description files and operating tools, observe the information about the device
in the "Manufacturer's information" document.
The manufacturer's information is available:
• In the Download Area of the Endress+Hauser Internet site: www.endress.com →
Download
• Specify the following details:
– Text search: Manufacturer's information
– Search range: documentation
Endress+Hauser 111
Maintenance Proline Promag H 200 HART
13 Maintenance
For a list of some of the measuring and test equipment, refer to the "Accessories"
chapter of the "Technical Information" document for the device.
112 Endress+Hauser
Proline Promag H 200 HART Repair
14 Repair
Endress+Hauser 113
Repair Proline Promag H 200 HART
www.endress.com/deviceviewer
1 Sensor
Additional information:
2 Main electronic
3 I/O module
4 Cover (front)
5 Cover (right)
6 Transmitter
7 Display
8
9
10
6
2
3
7 5
4
1
A0014017
14.4 Return
The measuring device must be returned if it is need of repair or a factory calibration, or if
the wrong measuring device has been delivered or ordered. Legal specifications require
Endress+Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.
To ensure safe, swift and professional device returns, please refer to the procedure and
conditions for returning devices provided on the Endress+Hauser website at
http://www.endress.com/support/return-material
14.5 Disposal
114 Endress+Hauser
Proline Promag H 200 HART Repair
Endress+Hauser 115
Accessories Proline Promag H 200 HART
15 Accessories
Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is
available from your local Endress+Hauser sales center or on the product page of the
Endress+Hauser website: www.endress.com.
Accessories Description
Accessories Description
Seal set For the regular replacement of seals for the sensor.
Welding jig Welded connection as process connection: welding jig for installation in pipe.
116 Endress+Hauser
Proline Promag H 200 HART Accessories
Grounding rings Are used to ground the fluid in lined measuring tubes to ensure proper
measurement.
Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface.
HART
For details, see "Technical Information" TI00404F
HART Loop Converter Is used to evaluate and convert dynamic HART process variables to analog current
HMX50 signals or limit values.
BA00371F
For details, see "Technical Information" TI00429F and Operating Instructions
Wireless HART adapter Is used for the wireless connection of field devices.
SWA70 The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
BA00053S
For details, see "Technical Information" TI00025S and Operating Instructions
Fieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART
measuring devices via a Web browser.
Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area.
Endress+Hauser 117
Accessories Proline Promag H 200 HART
Memograph M graphic The Memograph M graphic display recorder provides information on all relevant
display recorder measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
BA00247R
For details, see "Technical Information" TI00133R and Operating Instructions
RN221N Active barrier with power supply for safe separation of 4-20 mA standard signal
circuits. Offers bidirectional HART transmission.
BA00202R
For details, see "Technical Information" TI00073R and Operating Instructions
RNS221 Supply unit for powering two 2-wire measuring devices solely in the non-Ex area.
Bidirectional communication is possible via the HART communication jacks.
118 Endress+Hauser
Proline Promag H 200 HART Technical data
16 Technical data
16.1 Application
The measuring device is only suitable for flow measurement of liquids with a minimum
conductivity of 20 μS/cm.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
adequately resistant.
Measuring principle Electromagnetic flow measurement on the basis of Faraday's law of magnetic induction.
16.3 Input
Measuring range Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
25 1 9 to 300 75 0.5 1
Endress+Hauser 119
Technical data Proline Promag H 200 HART
½ 15 1 to 27 6 0,05 0,1
16.4 Output
Resolution < 1 µA
Pulse/frequency/switch output
Pulse output
Frequency output
120 Endress+Hauser
Proline Promag H 200 HART Technical data
Switch output
Current output
4-20 mA
Failure mode Selectable (as per NAMUR recommendation NE 43):
• Minimum value: 3.6 mA
• Maximum value: 22 mA
• Defined value: 3.59 to 22.5 mA
• Actual value
• Last valid value
HART
Device diagnostics Device condition can be read out via HART Command 48
Pulse/frequency/switch output
Pulse output
Failure mode Choose from:
• Actual value
• No pulses
Frequency output
Failure mode Choose from:
• Actual value
• 0 Hz
• Defined value: 0 to 1 250 Hz
Switch output
Failure mode Choose from:
• Current status
• Open
• Closed
Endress+Hauser 121
Technical data Proline Promag H 200 HART
Local display
Backlight Additionally for device version with SD03 local display: red lighting indicates a
device error.
Operating tool
• Via digital communication:
HART protocol
• Via service interface
Load → 29
Low flow cut off The switch points for low flow cut off are user-selectable.
Galvanic isolation All outputs are galvanically isolated from one another.
Terminal assignment → 29
Option A 1) 2)
: 4-20 mA HART For 4 mA: ≥ DC 18 V DC 35 V
For 20 mA: ≥ DC 14 V
122 Endress+Hauser
Proline Promag H 200 HART Technical data
Electrical connection → 30
Potential equalization → 32
Terminals • For device version without integrated overvoltage protection: plug-in spring terminals
for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• For device version with integrated overvoltage protection: screw terminals for wire cross-
sections 0.2 to 2.5 mm2 (24 to 14 AWG)
Cable entries • Cable gland: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
Cable specification → 28
Overvoltage protection The device can be ordered with integrated overvoltage protection for diverse approvals:
Order code for "Accessory mounted", option NA "Overvoltage protection"
1)
Input voltage range Values correspond to supply voltage specifications
Endress+Hauser 123
Technical data Proline Promag H 200 HART
[%]
2.5
2.0
0.5 %
1.5
1.0
0.5
0
0 1 2 4 6 8 10 [m/s]
v
0 5 10 15 20 25 30 32 [ft/s]
A0003200
Accuracy of outputs
o.r. = of reading
124 Endress+Hauser
Proline Promag H 200 HART Technical data
Current output
Accuracy ±10 µA
Pulse/frequency output
Pulse/frequency output
16.7 Installation
"Mounting requirements" → 19
16.8 Environment
Ambient temperature → 21
range
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature The storage temperature corresponds to the operating temperature range of the
measuring transmitter and the appropriate measuring sensors.
• Protect the measuring device against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• Select a storage location where moisture cannot collect in the measuring device as
fungus or bacteria infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed
before installing the measuring device.
Endress+Hauser 125
Technical data Proline Promag H 200 HART
Mechanical load • Protect the transmitter housing against mechanical effects, such as shock or impact.
• Never use the transmitter housing as a ladder or climbing aid.
16.9 Process
TA [°F] [°C]
140 60
100 40
20 1
0 PFA
-20
-40 -40
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
TF
-40 0 100 200 300 360 [°F]
A0019805
TA Ambient temperature
TF Medium temperature
126 Endress+Hauser
Proline Promag H 200 HART Technical data
Flow limit The diameter of the pipe and the flow rate determine the nominal diameter of the sensor.
The optimum velocity of flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the
velocity of flow (v) to the physical properties of the fluid:
• v < 2 m/s (6.56 ft/s): for abrasive fluids
• v > 2 m/s (6.56 ft/s): for fluids producing buildup
A necessary increase in the flow velocity can be achieved by reducing the sensor
nominal diameter.
For an overview of the measuring range full scale values, see the "Measuring range"
section → 119
Pressure loss • No pressure loss occurs as of nominal diameter DN 8 (3/8") if the sensor is installed in a
pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545
→ 22
System pressure → 21
Vibrations → 22
Design, dimensions For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section
25 1 4.0 8.8
Endress+Hauser 127
Technical data Proline Promag H 200 HART
Measuring tube Nominal diameter Pressure rating 1) Process connection internal diameter
specification
EN (DIN) PFA
3
A0020640
Adapter for cable entry with For non-Ex and Ex Nickel-plated brass
internal thread NPT ½" (except for CSA Ex d/XP)
Sensor housing
Coated aluminum AlSi10Mg
128 Endress+Hauser
Proline Promag H 200 HART Technical data
Measuring tubes
Stainless steel 1.4301 (304)
Liner
PFA (USP Class VI, FDA 21 CFR 177.1550, 3A)
Process connections
• Stainless steel 1.4404 (F316L)
• PVDF
• PVC adhesive sleeve
Electrodes
• Standard: 1.4435 (316L)
• Optional: Alloy C22, tantalum, platinum
Seals
• O-ring seal: EPDM, FKM, Kalrez
• Aseptic molded seal: EPDM 1), FKM, silicone 1)
Accessories
Weather protection cover
Stainless steel 1.4404 (316L)
Grounding rings
• Standard: 1.4435 (F316L)
• Optional: Alloy C22, tantalum
Fitted electrodes Measuring electrodes and empty pipe detection electrodes (only DN 25 (1")):
1.4435 (316L), Alloy C22, platinum, tantalum
Endress+Hauser 129
Technical data Proline Promag H 200 HART
Surface roughness Stainless steel electrodes, 1.4435 (F316L); Alloy C22, 2.4602 (UNS N06022); platinum;
tantalum:
≤ 0.3 to 0.5 µm (11.8 to 19.7 µin)
(All data relate to parts in contact with fluid)
Liner with PFA:
≤ 0.4 µm (15.7 µin)
(All data relate to parts in contact with fluid)
Stainless steel process connections:
≤ 0.8 µm (31 µin)
(All data relate to parts in contact with fluid)
16.11 Operability
Order code for "Display; Operation", option C "SD02" Order code for "Display; Operation", option E "SD03"
1 1
– –
+ +
E E
A0015544 A0015546
Display elements
• 4-line display
• With order code for "Display; operation", option E:
White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individually
configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
Operating elements
• With order code for "Display; operation", option C:
Local operation with 3 push buttons: , , E
130 Endress+Hauser
Proline Promag H 200 HART Technical data
1 2 3
A0013137
6 8
3
2
1 4 5 7
A0013764
Endress+Hauser 131
Technical data Proline Promag H 200 HART
3 –
+
E
2 1
A0014019
1 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device
2 Commubox FXA291
3 Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
CE mark The measuring system is in conformity with the statutory requirements of the applicable
EC Directives. These are listed in the corresponding EC Declaration of Conformity along
with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications
and Media Authority (ACMA)".
Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this
document on the nameplate.
Functional safety The measuring device can be used for flow monitoring systems (min., max., range) up to
SIL 2 (single-channel architecture) and SIL 3 (multichannel architecture with
homogeneous redundancy) and is independently evaluated and certified by the TÜV in
accordance with IEC 61508.
132 Endress+Hauser
Proline Promag H 200 HART Technical data
Endress+Hauser 133
Technical data Proline Promag H 200 HART
HistoROM extended Comprises extended functions concerning the event log and the activation of the
function measured value memory.
Event log:
Memory volume is extended from 20 message entries (basic version) to up to 100
entries.
Data logging (line recorder):
• Memory capacity for up to 1000 measured values is activated.
• 250 measured values can be output via each of the 4 memory channels. The
recording interval can be defined and configured by the user.
• Data logging is visualized via the local display or FieldCare.
16.14 Accessories
Overview of accessories available for order → 116
16.15 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The CD-ROM provided for the device (depending on the device version, the CD-ROM
might not be part of the delivery!)
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
Device Parameters
134 Endress+Hauser
Proline Promag H 200 HART Technical data
Technical Information
NEPSI Ex d XA01179D
NEPSI Ex i XA01178D
INMETRO Ex d XA01309D
INMETRO Ex i XA01310D
INMETRO Ex nA XA01311D
Special Documentation
Installation Instructions
Endress+Hauser 135
Index Proline Promag H 200 HART
Index
A Degree of protection . . . . . . . . . . . . . . . . . . . . . 35, 126
Access authorization to parameters Design
Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Measuring device . . . . . . . . . . . . . . . . . . . . . . . . . 12
Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Access code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Device components . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Incorrect input . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Device description files . . . . . . . . . . . . . . . . . . . . . . . 56
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Device documentation
Adapting the diagnostic behavior . . . . . . . . . . . . . . . 102 Supplementary documentation . . . . . . . . . . . . . . . . 8
Adapting the status signal . . . . . . . . . . . . . . . . . . . . 102 Device locking, status . . . . . . . . . . . . . . . . . . . . . . . . 90
Ambient temperature Device name
Influence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Ambient temperature range . . . . . . . . . . . . . . . . . . . 21 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
AMS Device Manager . . . . . . . . . . . . . . . . . . . . . . . . 54 Device repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Device revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 119 Device type ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Applicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Diagnostic behavior
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
B Diagnostic information
Burst mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Design, description . . . . . . . . . . . . . . . . . . . . 99, 101
FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
C Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Cable entries Remedial measures . . . . . . . . . . . . . . . . . . . . . . 103
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . 123 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Cable entry Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . 35 Diagnostics
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 132 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 DIP switch
Checklist see Write protection switch
Post-connection check . . . . . . . . . . . . . . . . . . . . . 36 Direct access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Post-installation check . . . . . . . . . . . . . . . . . . . . . 27 Direct access code . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cleaning Disabling write protection . . . . . . . . . . . . . . . . . . . . . 87
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . 112 Display
Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 112 Current diagnostic event . . . . . . . . . . . . . . . . . . . 105
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Previous diagnostic event . . . . . . . . . . . . . . . . . . 105
Advanced settings . . . . . . . . . . . . . . . . . . . . . . . . 78 see Local display
Configuring the measuring device . . . . . . . . . . . . . 60 Display area
Communication-specific data . . . . . . . . . . . . . . . . . . . 56 For operational display . . . . . . . . . . . . . . . . . . . . . 40
Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 In the navigation view . . . . . . . . . . . . . . . . . . . . . 42
Connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Display values
Connecting the measuring device . . . . . . . . . . . . . . . . 30 For locking status . . . . . . . . . . . . . . . . . . . . . . . . . 90
Connection Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
see Electrical connection Document
Connection examples, potential equalization . . . . . . . 32 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Connection preparations . . . . . . . . . . . . . . . . . . . . . . 30 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Connection tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Context menu Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Down pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Opening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 E
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . 123 Electrical connection
Commubox FXA195 . . . . . . . . . . . . . . . . . . . . . . 52
D Commubox FXA195 (USB) . . . . . . . . . . . . . . . . . 131
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 10 Commubox FXA291 . . . . . . . . . . . . . . . . . . . 52, 132
Define access code . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . 35
136 Endress+Hauser
Proline Promag H 200 HART Index
Endress+Hauser 137
Index Proline Promag H 200 HART
Navigation view . . . . . . . . . . . . . . . . . . . . . . . . . . 41 N
see Diagnostic message Nameplate
see In alarm condition Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
see Operational display Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 Navigation path (navigation view) . . . . . . . . . . . . . . . 41
Navigation view
M In the submenu . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Main electronics module . . . . . . . . . . . . . . . . . . . . . . 12 In the wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Maintenance tasks . . . . . . . . . . . . . . . . . . . . . . . . . 112 Numeric editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Replacing seals . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Managing the device configuration . . . . . . . . . . . . . . 84 O
Manufacturer ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . 120
Manufacturing date . . . . . . . . . . . . . . . . . . . . . . . 14, 15 Operating elements . . . . . . . . . . . . . . . . . . . . . . . 45, 99
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 Operating keys
Maximum measured error . . . . . . . . . . . . . . . . . . . . 124 see Operating elements
Measured variables Operating menu
Calculated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Menus, submenus . . . . . . . . . . . . . . . . . . . . . . . . 38
Measured . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
see Process variables Submenus and user roles . . . . . . . . . . . . . . . . . . . 39
Measuring and test equipment . . . . . . . . . . . . . . . . 112 Operating philosophy . . . . . . . . . . . . . . . . . . . . . . . . 39
Measuring device Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Operation options . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Operational display . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14, 15
Integrating via HART protocol . . . . . . . . . . . . . . . . 56 Orientation (vertical, horizontal) . . . . . . . . . . . . . . . . 20
Mounting the sensor . . . . . . . . . . . . . . . . . . . . . . 23 Outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Cleaning with pigs . . . . . . . . . . . . . . . . . . . . . . 24 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Mounting grounding rings . . . . . . . . . . . . . . . . 25 Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Mounting the seals . . . . . . . . . . . . . . . . . . . . . 24
Welding nipples . . . . . . . . . . . . . . . . . . . . . . . 24 P
Preparing for electrical connection . . . . . . . . . . . . 30 Packaging disposal . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preparing for mounting . . . . . . . . . . . . . . . . . . . . 23 Parameter
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 Changing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Enter a value . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Parameter settings
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 119 Administration (Submenu) . . . . . . . . . . . . . . 83, 108
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Burst configuration 1 to 3 (Submenu) . . . . . . . . . . 57
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Configuration backup display (Submenu) . . . . . . . 84
Measuring tube specification . . . . . . . . . . . . . . . . . . 128 Current output 1 (Wizard) . . . . . . . . . . . . . . . . . . 64
Mechanical load . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Data logging (Submenu) . . . . . . . . . . . . . . . . . . . . 93
Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Device information (Submenu) . . . . . . . . . . . . . . 109
Medium temperature range . . . . . . . . . . . . . . . . . . . 126 Diagnostics (Menu) . . . . . . . . . . . . . . . . . . . . . . 105
Menu Display (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 81
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Display (Wizard) . . . . . . . . . . . . . . . . . . . . . . . . . 72
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Empty pipe detection (Wizard) . . . . . . . . . . . . . . . 77
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60, 61 Low flow cut off (Wizard) . . . . . . . . . . . . . . . . . . . 75
Menus Operation (Submenu) . . . . . . . . . . . . . . . . . . . . . . 92
For measuring device configuration . . . . . . . . . . . . 60 Output conditioning (Wizard) . . . . . . . . . . . . . . . . 74
For specific settings . . . . . . . . . . . . . . . . . . . . . . . 78 Output values (Submenu) . . . . . . . . . . . . . . . . . . . 91
Mounting dimensions Process variables (Submenu) . . . . . . . . . . . . . . . . . 90
see Installation dimensions Pulse/frequency/switch output (Wizard) . . 66, 68, 70
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Sensor adjustment (Submenu) . . . . . . . . . . . . . . . 79
Mounting preparations . . . . . . . . . . . . . . . . . . . . . . . 23 Setup (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Mounting requirements Simulation (Submenu) . . . . . . . . . . . . . . . . . . . . . 85
Installation dimensions . . . . . . . . . . . . . . . . . . . . 21 System units (Submenu) . . . . . . . . . . . . . . . . . . . . 62
Mounting tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Totalizer (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 91
Totalizer 1 to 3 (Submenu) . . . . . . . . . . . . . . . . . . 79
Partially filled pipe . . . . . . . . . . . . . . . . . . . . . . . . . . 19
138 Endress+Hauser
Proline Promag H 200 HART Index
Endress+Hauser 139
Index Proline Promag H 200 HART
T
Technical data, overview . . . . . . . . . . . . . . . . . . . . . 119
Temperature range
Ambient temperature range for display . . . . . . . . 130
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . 17
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . 29, 30
Terminal voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Text editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Tool tip
see Help text
Tools
Electrical connection . . . . . . . . . . . . . . . . . . . . . . 28
For mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Transmitter
Connecting the signal cables . . . . . . . . . . . . . . . . . 30
Turning the display module . . . . . . . . . . . . . . . . . 27
Turning the housing . . . . . . . . . . . . . . . . . . . . . . . 26
Transporting the measuring device . . . . . . . . . . . . . . 17
Troubleshooting
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Turning the display module . . . . . . . . . . . . . . . . . . . . 27
Turning the electronics housing
see Turning the transmitter housing
Turning the transmitter housing . . . . . . . . . . . . . . . . 26
U
Use of the measuring device
Borderline cases . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Incorrect use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
see Designated use
User roles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
V
Version data for the device . . . . . . . . . . . . . . . . . . . . 56
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . 126
Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
W
W@M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112, 113
W@M Device Viewer . . . . . . . . . . . . . . . . . . . . 13, 113
Weight
Transport (notes) . . . . . . . . . . . . . . . . . . . . . . . . . 17
Wizard
Current output 1 . . . . . . . . . . . . . . . . . . . . . . . . . 64
140 Endress+Hauser
www.addresses.endress.com