51 54eph
51 54eph
51 54eph
51-54epH/rev.H
May 2006 Model 54e pH/ORP
2.0 INSTALLATION....................................................................................................... 5
2.1 Locating the Controller ............................................................................................ 5
2.2 Unpacking and Inspection ....................................................................................... 5
2.3 Mechanical Installation ............................................................................................ 5
i
MODEL 54e pH/ORP TABLE OF CONTENTS
LIST OF TABLES
ii
MODEL 54e pH/ORP TABLE OF CONTENTS
LIST OF FIGURES
iii
MODEL 54e pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 GENERAL DESCRIPTION The F1-F4 keys are multifunction. The active opera-
tion for that key is displayed as a label just above
The Model 54e pH/ORP analyzer/controller monitors
each function key as needed. For example, F1 is
and controls pH in chemical processes used in many
usually labeled Exit and F4 may be labeled Edit,
industries. This manual's sections and appendices
Save, or Enter. Pressing Enter 4 will access sub-
cover the system's configuration, calibration, and
menus, while pressing Edit allows changing values
maintenance, and provides a troubleshooting guide.
and Save stores the values in memory. Esc 3 can
All adjustments to the current outputs, alarm relays, be used to abort unwanted changes. Exit 1 returns
and calibration of the pH and temperature inputs can to the previous screen. Other labels may appear for
be made using the controller's membrane keypad. more specialized tasks.
The up t and down b keys are used to:
1. Move the cursor (shown in reverse video) up and
down on the menu screens.
2. Scroll through the list of options available for the
FIGURE 1-1. Main Display Screen The right and left keys are used to move the cursor
to the next digit of a number.
Green LEDs (labeled 1, 2, and 3) indicate when
alarm relays 1, 2, and 3 are energized. The fourth
1.2 DESCRIPTION OF CONTROLS relay indicates a fault condition. When a fault
Figure 1-1 shows a diagram of the main display occurs, the red LED (labeled FAIL) lights up, a
screen. Similar diagrams are used throughout this descriptive error message is displayed, and the
manual. The primary variable is continuously dis- action of the outputs and relays will be as described
played in large numerals. The process temperature in Section 5.6 and Section 5.7 under fault value
and primary current output value are always dis- (e.g. 22 mA).
played on the second line of the main display The red LED also indicates when the interval timer
screen. The third line can be configured to read routine is activated and when the time limit has been
several different items, as desired. In this case, it is reached on a feed limit timer. For more information
displaying setpoints for alarms 1 and 2. on these subjects, see Section 5.7.
1
MODEL 54e pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 SPECIFICATIONS
PHYSICAL SPECIFICATIONS - GENERAL
Enclosure: Epoxy-painted aluminum, NEMA 4X (IP65), 144 X 144 X 132 mm, DIN size (5.7 X 5.7 X 5.2 in.)
Front Panel: Membrane keyboard with tactile feedback and user selectable security. Light gray, blue and white
overlay. Light gray enclosure, dark gray bezel
Display: Back-lit dot matrix LCD (7.0 x 3.5 cm), blue on gray-green. The display contrast is compensated for
ambient temperature.
Process Variable Character Height: 16mm (0.6 in.)
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D.
T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
CSA-LR34186:
Max. relay contact rating: 28 Vdc; 110 Vac;
230 Vac; 6 amps resistive
FM: Max. relay contact rating: 28 Vdc resistive
150 mA - Groups A & B;
400 mA - Group C;
540 mA - Group D
Power:
Code -01: 100 - 127 VAC, 50/60 Hz ± 6%, 6.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 6.0 W
Code -02: 20 - 30 VDC, 6.0 W
Current Outputs:
Output 1: pH, ORP, temperature, glass impedance, or reference impedance.
Output 2: pH, ORP, temperature, glass impedance, or reference impedance.
Each output is galvanically isolated, 0-20 mA or 4-20 mA into 500 ohms maximum load at 115/230 Vac or
24 Vdc (Code -02) or 500 ohms maximum load at 100/200 Vac. Output 1 includes digital signal 4-20 mA
superimposed HART (Code -09 only).
RFI/EMI: EN-61326
The Model 54e when configured as a pH analyzer, The Model 54e when configured as an ORP
requires a dual (glass and reference) impedance pre- analyzer, measures over a range of -1400 mV to
amplifier (patent pending). This preamp converts the +1400 mV in either the American convention
high impedance pH glass electrode signal to a low (Oxidation Reduction Potential), or the European con-
impedance signal. The preamplifier may be located in vention (Reduction Oxidation-Redox). Although tem-
one of three areas: 1) in the pH sensor for best per- perature compensation is not used for ORP
formance, 2) in a remote junction box when process
measurements, the process temperature is measured
temperatures exceed 80°C (176°F) in submersion
applications, or 3) in the analyzer when the distance and displayed. Temperature measurement is made by
between the pH sensor and the analyzer is 4.5 meters an RTD located in the sensor assembly.
(15 feet) or less.
The Model 54e pH measures over the full range of 0-
14 pH. The current output may be calibrated to repre- ANALYZER SPECIFICATIONS @ 25°C
sent any 1 to 14 pH span. Measurement Range: –1400 to +1400 mV
A two-point calibration is made by immersing the sen-
Output Scale Expansion: Zero suppression: up to
sor in two different buffer solutions and entering the
pH values. When two buffers are used, the micro- ±1300 mV
processor automatically calculates the electrode slope Span: Any ORP range from 100 to 2800 mV
which is used for self-diagnostics. The electrode slope Accuracy: ± 1.0 mV
can be read on the display and manually adjusted. A
one-point process standardization is easily accom- Repeatability: ± 1.0 mV
plished by entering the pH value of a grab sample. Stability: ± 1.0 mV/month, non-cumulative
Temperature Coefficient: Input: ± 0.2 mV/°C
Output: ± 0.4 mV/°C
ANALYZER SPECIFICATIONS @ 25°C
Temperature Measurement: –15 to 120°C (5 to 248°F)
Measurement Range: 0 to 14 pH Pt 100 or Pt 1000 RTD
Output Scale Expansion: Zero suppression: up to 13
pH units
Span: Any pH from 1 to 14 RECOMMENDED SENSORS:
Accuracy: ± 0.01 pH Model 330 Flow Through ORP
Repeatability: ± 0.01 pH *Model 371 EuroSenz ORP
Stability: ± 0.01 pH/month, non-cumulative Model 381+ Insertion/Submersion/Flow Through ORP
*Model 389 Disposable ORP
Temperature Coefficient: Input: ± 0.003 pH/°C
Model 396P Disposable ORP
Output: ± 0.006 pH/°C Model 396R Retractable ORP
Temperature Compensation : Pt 100 or Pt 1000 RTD, Model 398 Insertion/Submersion ORP
Automatic or Manual Model 398VP Insertion/Submersion with VP 6.0 connector
–15 to 120°C (5 to 248°F) Model 398R Retractable ORP
Model 398RVP Retractable ORP with VP 6.0 connector
RECOMMENDED SENSORS:
Model 320B Flow Through pH
Model 320HP High Purity pH The Model 54e when ordered for ISE capability,
*Model 328A Steam Sterilizable pH is suitable for use with a number of ion-selective elec-
*Model 370 and 371 EuroSenz pH trodes. Consult the factory for available measure-
Model 381+ Insertion/Submersion/Flow Through pH ments and ranges.
*Model 389 Disposable pH
*Model 396/VP Disposable pH
Model 396P/VP Disposable pH
Model 396R/VP Retractable pH
Model 397 Quik Disconnect pH
Model 398/VP Insertion/Submersion pH
*Model 398R/VP Retractable pH
*Model 399 Disposable pH
*Model Hx338 Steam Sterilizable pH
*Model Hx348 Steam Sterilizable pH 3
MODEL 54e pH/ORP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
MODEL
54e pH/ORP MICROPROCESSOR ANALYZER
CODE OPTIONS
01 115/230 VAC, 50/60 Hz Power
02 24 VDC Power
CODE OPTIONS
09 HART Communications Protocol
20 Controller Outputs - PID and TPC
54e pH/ORP -01 -20 EXAMPLE
ACCESSORIES
4
MODEL 54e pH/ORP SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
This section is for installation of the controller. 2.3.2 Wall or Surface Mounting:
WARNING 1. Mount the bracket to the controller using the sup-
All electrical installation must conform to the plied four screws as shown in Figure 2-2.
National Electrical Code, all state and local 2. Mount controller mounting bracket to wall using
codes, and all plant codes and standards for any appropriate fastener such as screws, bolts,
electrical equipment. All electrical installations etc (see Figure 2-1 below).
must be supervised by a qualified and respon-
sible plant electrician.
5
MODEL 54e pH/ORP SECTION 2.0
INSTALLATION
6
MODEL 54e pH/ORP SECTION 3.0
WIRING
SECTION 3.0
WIRING
NOTE
Wire only the analog and alarm outputs
required for your application. Be sure to read 3.4 ALARM RELAY OUTPUT WIRING
the warning at the beginning of Section 2.0. The controller has 3 "dry" alarm relay contacts which are
The Model 54e pH/ORP has five access holes in the bot- normally open. Alarm 1 is across terminals 4 and 5 on
tom of the instrument housing which accept ½-in. strain TB3. This alarm is typically used to control the pump in a
relief connectors or conduit fittings. Be sure to seal any chemical feed system. Alarm 2 across terminals 6 and 7
unused access holes. As you face the front of the unit, on TB3 is usually used to operate a light or horn as a
the rear openings are for input power, and alarm relay means of alerting the chemical process operator when
signals. The opening on the front left is for sensor wiring pH/ORP is outside the control range. Alarm 3 is across
only (DC). The front right is for analog output wiring. terminals 8 and 9 on TB3. All 3 of these alarms may be
activated on pH/ORP or temperature. They can also be
NOTE used to control other pumps or valves provided they are
For best EMI/RFI protection, the output cable programmed to do so. Refer to Section 5.0 to set up these
should be shielded and enclosed in an earth functions.
grounded, rigid, metal conduit. Connect the All three alarm contacts on the Model 54e pH/ORP are
output cable's outer shield to the earth ground rated for a maximum of 3 A (1.5A, 230 VAC, inductive
connection on TB2 (Figure 3-1) load). If your associated pump or valve exceeds this,
use a separate contact or relay rated for the external
device.
To use a contact output to control a pump, valve, or
3.2 POWER INPUT WIRING light, the contact must be wired into a circuit together
Figure 3-1 depicts the wiring detail for the Model 54e with a source of power for the device to be controlled.
pH/ORP. Code -01: connect AC power to TB3, termi- The power can be jumpered from the main power into
nals 1 and 2 for 115 VAC (terminals 2 and 3 for 230 the controller and the circuit can be wired as shown on
VAC). Code -02: connect DC power to TB3 terminals 1 the wiring diagrams, Figure 3-1.
and 2. Connect earth ground to the nearby ground lug.
A good earth ground is essential for proper operation of PREAMPLIFIER SELECTION
the controller. Be sure to provide a means of discon-
The pH sensor signal requires a preamplifier at some
necting the main power to the controller.
point in the measuring circuit. The preamp can be in-
CAUTION side the sensor, in the junction box, or in the controller.
Do not apply power to the controller until all To allow for these options, the Model 54e pH/ORP has
electrical connections are made. a jumper selectable preamp mounted on the CPU cir-
cuit board (Figure 3-3). The jumper is placed in the
WARNING "analyzer" position when there is no preamp in the sen-
Electrical connections to this equipment sor (or junction box). Generally, this jumper is in the
must be made in accordance with the cur- "sensor" position.
rent National and Local Electrical Codes in
effect for the installation location.
7
MODEL 54e pH/ORP SECTION 3.0
WIRING
FIGURE 3-1. Power Input and Relay Output Wiring for Model 54e pH/ORP
8
MODEL 54e pH/ORP SECTION 3.0
WIRING
9
MODEL 54e pH/ORP SECTION 3.0
WIRING
10
MODEL 54e pH/ORP SECTION 3.0
WIRING
FIGURE 3-5. Wiring for Sensors that have Solution Grounds (Models 381+, 385+, 396P, 396R)
11
MODEL 54e pH/ORP SECTION 3.0
WIRING
FIGURE 3-6. Wiring for Sensors Without Solution Grounds Directly to Analyzer (Models 389-02-54,
396-54, 397-54, 399-09). For wiring to a junction box, see Figure 3-7.
FIGURE 3-7. Wiring for Sensors Without Solution Grounds to a Junction Box (Models 389-02-54, 396-54, 399-09)
NOTES:
1. Interconnecting wire maximum length is 500 ft. Use PN 9200273 (no substitutes). Shields must be insulated from each other at all
cable ends. Connect to junction box as shown.
2. If distance to controller is short, the junction box is not required. Connect sensor leads directly to controller.
12
MODEL 54e pH/ORP SECTION 3.0
WIRING
13
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
SECTION 4.0
CALIBRATION
The following procedures are described in this section: This compensates for temperature-related changes in
• Temperature Calibration (Section 4.1) the response of the glass pH electrode which would
affect the pH measurement. The pH temperature com-
• Auto Buffer (Two-Point) Calibration (Section 4.2) pensation does not account for changes in chemical
• Manual Two-Point Calibration (Section 4.3) activity which affect the actual pH value of the solution
being measured. See Section 7.0 for information on
• Single-Point pH Calibration (Section 4.4) solution temperature compensation.
• Temperature Compensation Options (Section 4.5) To ensure the controller's accuracy, it is important to
• pH Slope Adjustment (Section 4.6) perform all the calibration procedures provided in this
section if you are:
• Hold Mode (Section 4.7)
• installing this unit for the first time
• changing or replacing electrodes or sensor ele-
INTRODUCTION
ments
Calibration is the process of adjusting or standardizing
• troubleshooting
the controller to a lab test (such as free acid titration) or
a calibrated laboratory instrument, or standardizing to
some known reference (such as a commercial pH IMPORTANT
buffer). Calibration ensures that the controller shows
an accurate, and therefore, repeatable reading of pH or Before attempting to calibrate, inspect the
temperature. pH sensor assembly. It must be clean,
undamaged and free from cracks or other
Since pH measurements are affected by temperature, signs of leakage or wear.
the Model 54e pH/ORP reads the temperature at the
sensor and compensates for the changing temperature
by referencing all pH measurements to 25°C (77 °F).
WARNINGS
Before performing any of these procedures, be sure to disable or disconnect the chemical feed pumps or
other external devices (see placing controller in hold, Section 4.7)
Perform the calibration procedures in this section only in the order they are given.
Do not attempt to calibrate the controller if the fault LED is lit or the display is showing fault messages.
14
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
25.1 °C 3. Press Edit (F4) with this display shown to adjust the temperature. The
screen below will then appear. Using the arrow keys, input the cor-
rect temperature value and press Save (F4). The controller will enter
Adjust temp: 25.1 °C the value into memory. To abort the change, press Esc (F3).
Afterwards, to continue with buffer calibration, go to Section 4.2 or
Exit Edit 4.3, otherwise press Exit (F1) three times for the main display.
NOTE
25.1 °C If hold mode was turned ON, be certain to install the sensor back
in the process and change the setting to OFF to resume normal
operation before leaving the controller. The screen on the top left
Adjust temp: +025.1 °C will appear again before the main display is shown. Follow the
same routine as in the Note for step 2 to turn the Hold Mode Off
Esc Save
and then press Exit (F1).
15
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
4.02 pH 2. With the sensor in the first buffer, press Cont (F3). To cancel the cal-
ibration, press Abort (F1).)
The word "Wait" flashes until the sensor is stabilized.
Autocal buffer 1
The controller is waiting for the pH reading to stabilize within the
Abort Cont parameters established in Section 5.9. If the controller appears to be
locked at this stage, the reading is not stable enough. Increase the
stabilize pH and/or decrease the stabilize time and retry the calibra-
4.02 pH tion.
Press Abort (F1) to cancel the calibration if it appears to be taking too
long.
Buffer 1 -- “Wait”
Abort Cont
16
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
3. A screen like the one on the left will appear after stabilization.
NOTE
The calibration calculation does not occur until buffer 2 is com-
pleted. Aborting the calibration at this point will not change the
reading of the controller.
5. A screen like the one on the left will appear after stabilization. Use the
10.02 pH arrow keys to select the correct buffer, and press Cont (F3). If
the correct buffer does not appear, refer to Section 5.9 and select a
buffer group that includes the desired buffers. As usual, to cancel the
Buf2 done: 10.01 @25°C calibration, press Abort (F1).
Abort Cont This completes the calibration. The controller has now recalculated
the slope and the zero offset of the sensor.
CALIBRATION NOTES
1. A two-point calibration should always be followed by a sin-
gle-point calibration (see Section 4.4).
Auto buffer cal done 3. If the sensor is not at the same temperature as the buffer the cal-
ibration will be in error.
Abort Cont
NOTE
If hold mode was turned ON, be certain to install the sensor back
in the process and change the setting to OFF to resume normal
operation before leaving the controller. The hold mode screen
will appear again before the main display is shown. Follow the
same routine as in the Note for step 1 to turn the Hold Mode Off
and then press Exit (F1).
17
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
Standardize pH 1. From the main display, press any key to obtain the main menu. With
the cursor on "Calibrate", press Enter (F4). With the cursor on "2-pt
Adjust temperature Calibration", press Enter (F4) again.
NOTE
Exit Enter
The Hold Mode screen may appear if the feature was enabled in
Section 5.6. Changing the Hold Mode to ON holds the outputs in
a fixed state, and avoids process upsets during calibration.
4.01 pH the first buffer solution. The large displayed pH shows the actual pH.
3. Allow the pH and temperature to stabilize. This may take several min-
utes, as the pH sensor may need to cool down to the buffer temper-
Calibrate point 1 ature. The display will flash “wait’ while the reading is stabilizing.
When the pH reading is stable, press Cont (F3) to update the read-
Abort Cont ing. Press Edit (F4) and use the arrow keys to change the indicated
buffer value to the correct value for the buffer at the stabilized tem-
perature. The pH values at various temperatures are written on the
label of most buffer bottles.
4. Press Pt 2 (F3) to enter Buffer 1 into memory and proceed to buffer
Pt 2 Edit
10.01 pH 6. Place the pH sensor into a beaker containing the second buffer solu-
tion.
7. Allow the pH and temperature to stabilize (as before) and press Cont
Calibrate point 2 (F3). The display will flash “wait’ while the reading is stabilizing.
8. Enter the correct value for buffer 2 as described in step 3 and press
Abort Cont
Save (F4) to complete the calibration. Reinstall the sensor.
CALIBRATION NOTES
1. A two-point calibration should always be followed by a sin-
gle-point calibration (see Section 4.4)
2. After a buffer calibration, the electrode slope value can be
10.01 pH checked to see if the electrode is aging.
3. If the sensor is not at the same temperature as the buffer the cal-
ibration will be in error.
2-pt calibration done 4. If the hold mode is ON, be certain to reinstall the sensor in the
Esc Save process and change the setting to OFF to return to normal oper-
ation.
18
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
Standardize: 7.01 pH Use the arrow keys to change the second line standardize value to
the correct pH and press Save (F4) to complete the procedure. Esc
Exit Enter (F3) will cancel.
The pH reading in the large display will change to the new value and
the zero offset number will be recalculated.
4.01 pH NOTE
Before exiting the calibration mode, remember to change the
hold mode setting to OFF (if it was turned on in step 3).
Standardize: +07.00 pH
Esc Save
19
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
20
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
7.00 pH tenance or repair of the sensor, the Controller can be placed in hold (refer
to Section 5.6 to enable this feature) to prevent process upsets while the
reading is off-line. This will place the current outputs into the selected
Hold Mode: Off default states (see Section 5.6). The relays will act as selected in relay
default, see Section 5.7.
Exit Cont Edit Before removing the sensor from the process, press any key and then
Enter (F4). When the hold mode has been enabled, the hold mode
screen (on the left) will appear prior to calibration. To continue without
putting the controller in hold, simply press Cont (F3). To put the controller
in hold, press Edit (F4), use the arrow key to change the "Off" to "On" and
7.00 pH press Save (F4).
NOTE
26.2°C 12.0mA When the Hold Mode is activated ("On"), "Hold Mode Activated"
will always appear on the bottom line of the display.
Hold Mode Activated
Always calibrate after cleaning or repair of the pH sensor. After installing
the sensor back into the process, always change the Hold Mode setting
to OFF.
21
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
SECTION 5.0
SOFTWARE CONFIGURATION
This section contains the following:
• An introduction to using the configuration process
• A List of Settings for the controller
• Step-by-step instructions and explanations for each parameter on the List
INTRODUCTION TO CONFIGURATION
The controller arrives from the factory configured and ready to operate as a pH controller. Refer to Appendix A for ORP meas-
urements.
Figure 5-1 is an outline of the menu structure. Before attempting any changes refer to the parameter setup list shown in Table
5-1. This table presents a brief description and the possible options.
The factory setting is listed with a space for the user setting. It is recommended that the list be carefully reviewed before any
changes are made.
On initial configuration, it is recommended that the parameters be entered in the order shown on the worksheet. This will reduce
the chance of accidentally omitting a needed parameter.
By changing the standard output configuration, you can set up the Model 54e pH/ORP to perform a wide variety of control and monitoring tasks.
The configuration procedures allow you to program the controller to meet the specific control and monitoring requirements of your particular
plant. This is done by recording the desired configuration parameters on the List of Settings Form and then actually configuring them by using
the keys on the controller front panel.
24
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
Accessing Calibrate, Program and Configure Menus. Level 2 Program. To access the program level from the
Operating configuration changes are made at the levels main menu, place the cursor over "Program" with the
shown in Figure 5-1. Pressing any key from the main dis- down arrow key. Then press Enter (F4). From the program
play will access the main menu (top left). Refer to level menu, changes can be made to the alarm setpoints
Appendix A for ORP measurements. and the output setpoints.
Level 1 Calibrate. To access calibration selections from Level 3 Configure. To access the configure level from the
the main menu, with the cursor on "Calibrate" press Enter main menu place cursor over "Program" and Enter (F4),
(F4). All buffer calibration, pH standardization and tem- then place cursor over "Configure" and Enter (F4). This
perature adjustments are made at this level (refer to level contains advanced selections, such as alarms, diag-
Section 4.0 for these procedures). nostics, autocal and others.
Blank
Out 2
R1
GI
In
AL2
AL3
This section describes how the three alarm setpoints can be changed.
Alarm setpoints Move the cursor down by pressing the arrow key.
Output setpoints
1. From the main menu, move the cursor down to "Program" and press
Simulate tests Enter (F4). On the next display, with the cursor on "Alarm setpoints",
Exit Enter
press Enter (F4).
2. Select the desired alarm by moving the cursor down to highlight it.
Alarm 1 setpoint When the correct alarm is highlighted, press Enter (F4) to get to the
adjustment screen.
Alarm 2 setpoint In this example we have pressed the arrow key down once to access
Alarm 3 setpoint the alarm 2 setpoint.
3a. (normal alarm). The setpoint now being used for this alarm and the
Alarm High: 14.00pH kind of alarm (high or low) are displayed. If the alarm has been
turned off, then "off" will be displayed instead of "High". The "Enter"
key has now changed to the "Edit" key and will allow changing the
setpoint once the F4 key has been pressed. If the setpoint is ok,
then press Exit (F1).
Exit Enter After the Edit (F4) key is pressed, use the arrow keys to change the
display to the desired setpoint and press Save (F4) to enter into
memory. The plus (+) sign can be changed to a minus sign by press-
ing the down arrow key when the (+) is highlighted. To abort the
change, press Esc (F3) to return to the previous menu.
3b. (TPC alarm only). When the alarm has been configured as TPC, the
Setpoint: 07.00 pH setpoint is used for the TPC calculation of how long the alarm should
stay on. The "Enter" key has now changed to the "Edit" key and will
allow changing the setpoint once the F4 key has been pressed. If
the setpoint is ok, then press Exit (F1).
Exit Edit
After the F4 key is pressed, use the arrow keys to change the dis-
play to the desired setpoint and press Save (F4) to enter into mem-
ory. The plus + sign can be changed to a minus sign by pressing the
down arrow key. To abort the change, press Esc (F3) to return to the
previous menu.
26
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
This section describes how the two output setpoints can be changed.
Alarm setpoints This selection is only active if the current output control mode has been
set to "PID" (see Section 5.6). If the control mode is set to "normal",
Output setpoints then "Not applicable" will be displayed. For reranging outputs, go to
Section 5.3.
Simulate tests
1. From the main display, press any key to obtain the main menu. With
Exit Enter the down arrow key, move the cursor to "Program" and press Enter
(F4). With the cursor on "Output setpoints" (as on the left), press
Enter (F4).
Output 1 setpoints 2. Highlight the desired Output setpoints and press Enter (F4).
Output 2 setpoints
Exit Enter
3. The setpoint now being used is displayed. Press Edit (F4) and use
Setpoint: 7.00 pH the arrow keys to change the display to the new value.
4mA: 0.00 pH 4 mA is the deviation from setpoint that will result in a 4 mA out. 20
mA is the deviation from setpoint that will result in a 20 mA setpoint.
20mA: 2.00 pH Highlight the desired item and press Edit 4 and the arrow keys to
change the display to the new value.
Exit Enter
Example: A setpoint of 6 pH with a URV of +2 and a LRV of 0.0 pH.
When the pH is 7, the output will be (7 - 6)/(2 - 0) = 50% of range (12
mA). If the setpoint is changed to 6.5 pH, the output will be 7 - 6.5)/(2
- 0) = 25% of range (8 mA).
4. Press Save (F4) to enter into memory or Esc (F3) to abort the
Setpoint: + 06.90 pH change.
The Control setpoint is typically the condition where the current output is
at a minimum. The P and I control calculations use the setpoint to adjust
the current output to the desired level based on the parameters estab-
Exit Enter
lished in Section 5.6.
27
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
2. Select the desired output by moving the cursor down to highlight it.
Output 1 When the correct output is highlighted, press Enter (F4) to get to the
Output 2 adjustment screen.
Exit Enter
This screen allows changing the setpoints for output 1. A similar screen
4 mA: 0.00 pH is available for output 2. The live current output now being transmitted
by the controller is shown on the third line.
20 mA: 14.00 pH 4. Press Edit (F4) to make changes in the setpoints. The Edit key
Output 1: 12.00 mA changes to a Save key and the F3 key becomes active as an Esc
key. Use the arrow keys to make the display read the desired val-
Exit Enter ues for the high and low current output limits. When done, press
Save (F4) to enter the changes into memory. Press Esc (F3) to can-
cel changes.
NOTE
Outputs that have been configured as 0-20 mA in Section 5.6,
4 mA: +00.00 pH will show 0 mA instead of 4 mA on the top line. Outputs that are
based on temperature or impedance values will show matching
20 mA: 14.00 pH units such as °C, MΩ, or kΩ. See Section 5.6 for output config-
uration.
Output 1: 12.00 mA
Esc Save
28
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
This section describes how the current outputs and alarm relays can be
Alarm setpoints manually set for the purposes of checking devices such as valves,
pumps, or recorders.
Output setpoints 1. From the main menu, move the cursor down to "Program" and press
Simulated tests Enter 4. On this display, move the cursor to "Simulated tests" and
press Enter 4.
Exit Enter
Test output 1 2. At this point there are six separate screens for testing each of the
current outputs and each of the alarm relays. Highlight the desired
Test output 2 item by pressing the arrow b t keys as needed.
Test alarm 1
When the desired item is highlighted, press Enter 4 to continue.
Exit Enter Go to step 3a for outputs and 3b for alarms.
NOTE
Test alarm 2 A cautionary message will appear to warn that the out-
put or alarm that was selected will be changed by the fol-
Test alarm 3 lowing action. Be sure to alert plant personnel that
these changes are simulated and do not represent a
Test alarm 4 change in the actual process. Press Enter 4 to contin-
ue or Abort 1 to cancel the simulation.
Exit Enter
Test output 1: 10.00 mA 3a. The output is now being simulated. In the example to the left, out-
put 1 has been set to 10.00 mA. The output will remain at 10.00 mA
until either Exit 1 (or Edit 4 see below) is pressed or the test is
Simulating output1 concluded by timeout. The default value for the timeout is 10 min-
utes, so after 10 minutes, the output would go back to normal oper-
Exit Edit ation. To configure the timeout option, see Section 5.5.
If the displayed current is not the desired value, press the Edit 4
key and the next screen will allow changing the value. Use the arrow
Test output 1: 10.00 mA keys to change the display as needed, and press Test 4 to use that
value. Press Esc 3 to cancel the change in the value and contin-
ue simulating the previous current.
Esc Test
29
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
3b. The alarm relay is now being simulated. In the example to the left,
Test alarm 1: Open alarm 1 has been set to Open. This means that the relay is not ener-
gized (i.e. off). The alarm will remain open until either Exit 1 or Edit
4 is pressed or the test is concluded by timeout. The default value
Simulating alarm1 for the timeout is 10 minutes, so after 10 minutes, the alarm would
go back to normal operation and the display will return to the main
Exit Edit menu. To configure the timeout option, see Section 5.5.
If the displayed alarm action is not as desired, press the Edit 4 key
and the next screen will allow changing it. Use the arrow keys to
change the display as needed, and press Test 4 to enter the
Test alarm 1: Open change. Press Esc 3 to cancel the change in the value and con-
tinue simulating the previous action.
NOTE
Exit More Enter
Alarm relays may be simulated in the energized
(Closed) position or the de-energized (Open) position.
30
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
This section describes the options available for the changing of engineering
Display units and variables on the main display.
Outputs 1. From the main menu, move the cursor down to "Program" and press
Enter 4. From the program menu, move the cursor down using the
Alarms arrow key to highlight "Configure" and press Enter 4.
The first configuration menu is displayed. With the cursor on "Display"
Exit Enter
press Enter 4.
Exit Enter The values now being used by the controller are displayed. To change
any of these items, use the arrow key to highlight the desired item and
press Edit(F4). Use the arrow keys to make the change and press Save
(F4) to enter the change into memory.
WARNING: This WARNING
Press the arrow key three times to access this screen. The current out-
puts can be displayed as milliamps or as percent of full scale. The default
Display left: AL1 is mA. To change any of these items, use the arrow key to highlight the
Display right: AL2 desired item and press Edit (F4). Use the arrow keys to make the change
and press Save (F4) to enter the change into memory.
Display contrast: 5 Further menu items are available by pressing the arrow key repeatedly.
Exit Edit When the display is highlighting the item on the third line, the end of the
menu has been reached. To back up within the menu, use the up arrow
key.
Measure: pH
Resolution: 0.01
Temp Units: °C
Exit Enter
31
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
4. This screen allows you to choose the items displayed on the third line
Display left: AL1 left and right of the main display screen. The process temperature
and output 1 value (in mA or %) are always shown on line 2 of the
Display right: AL2 main display. This screen allows you to make the following choices:
Display contrast: 5
Lower Left of Main Display Lower Right of Main Display
Exit Edit
• AL 1 (alarm 1 setpoint - no units shown) • AL 2 (alarm 2 setpoint - no units shown)
• AL 3 (alarm 3 setpoint - no units shown) • AL 3 (alarm 3 setpoint - no units shown)
• In (sensor mV input) • In (sensor mV input)
• GI (glass impedance in Megohms) • RI (reference impedance in Kohms)
• RO (reference offset in mV) • Out 2 (Output 2 value in mA or %)
• Slp (slope of glass in mV/pH - • Blank (nothing displayed in lower right)
no units shown)
• Blank (nothing displayed in lower left)
5. The timeout feature works on both the display and simulated tests
using the current outputs and alarm relays.
Display contrast: 5 The display timeout will return the display to the main display screen
(from any other screen) if no key is pressed before the timeout value.
Timeout: On This is useful because the main display screen is usually the most
Timeout Value: 10 min important screen to the operator.
The timeout feature also allows simulating the current output and
Exit Edit alarm actions with an automatic return to normal operation. When the
feature is turned on (the default), simulated tests (see Section 5.4 for
details) will be completed automatically when the timeout value is
reached.
As before, to change these settings, use the arrow key to highlight the
desired item and press Edit 4. Use the arrow keys to make the
change and press Save 4 to enter the change into memory. Press
Esc 3 to abort.
SECURITY CAUTION
The Timeout Value is also used by the controller to acti-
vate security (Section 7.1). After unlocking the controller
by entering a security code, security is not re-activated
unless a display timeout occurs. If Timeout has been
turned off here, security will never re-activate.
32
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
This section describes the options available for configuration of the out-
Display puts. Each output can be configured to represent pH (or ORP/Redox, see
Appendix A), temperature, glass impedance, or reference impedance. In
Outputs addition, each output can be linear (normal) or nonlinear (PID). Several
options for both modes of outputs can be configured. This section
Alarms describes the features and shows how to make modifications in the set-
Exit Enter tings.
1. Beginning from the main menu, move the cursor down to "Program"
and press Enter 4. From the program menu, move the cursor down
using the arrow key b to highlight "Configure" and press Enter 4.
Use the arrow key again to highlight "Outputs" (as shown on the left)
and press Enter 4.
Output 1 control 2. There are 5 menu headers that relate to outputs. Each output has a
control header and a setup header. In addition, the hold feature is
Output 1 setup turned on here. The hold feature is mostly used to prevent upsets dur-
Output 2 control ing buffer calibration of the pH sensor.
To access each header, highlight the desired item and press the Enter
Exit Enter 4 key. To select another header, use the arrow keys. The bottom
menu header will only be highlighted if the end of the menu has been
reached.
Each output can be configured with the options above. The default
Exit Enter
options are that output 1 is Process (pH or ORP/Redox), output 2 is
Temperature and that both outputs are Normal (not PID). This is the
most common configuration and may not require changes. If no
changes are desired, skip to step 4.
Output: Process
To make changes in these parameters, highlight the desired menu
header and press Enter 4. The value now being used is displayed
and the F4 key can be pressed to Edit the item. Once Edit has been
pressed, change the item as needed and then press Save 4 to
Exit Edit store the value. Repeat for the other output and/or items as needed.
33
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
Hold: Last Value Note: 0-20 mA output range is disabled on output 1 with HART-
enabled version (-09 option).
Exit Edit
These parameters can be adjusted by highlighting the desired item
and pressing the Edit 4 key. Once Edit has been pressed, change
the item as needed and then press Save 4 to store the value.
"Range" determines whether the 4-20mA or 0-20 mA convention is
Hold: Last Value used for the current output. If the range is changed, be sure to
rerange the outputs as described in Section 5.3.
Fixed Hold: 21.00 mA
"Dampening" is used to time-average the current output, smoothing
Fault: 22.00 mA out the effect of a noisy reading. Higher values provide more smooth-
ing.
Exit Edit
Enabling the "hold" feature will give the user the option of placing the
output in hold during the calibration sequence.
A "fixed value" places the held output at a fixed value between 0
and 22 mA.
Exit Edit
The four parameters above are only available for outputs that have
been configured as PID outputs in step 3. These parameters can be
adjusted using the same technique as in step 4a, by highlighting the
desired item and pressing the Edit 4 key. Once Edit has been
Derivative: 0.0% pressed, change the item as needed and then press Save 4 to
store the value.
Range: 4-20 mA Use caution in changing the values of these parameters.
Hold: Last Value "Setpoint" is usually the desired value at which the process is being
controlled, typically the output will be 4(or 0) mA when the parameter
Exit Edit
is near the setpoint. This setting can also be changed using the pro-
cedure in Section 5.1.
34
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
Hold Feature Setup 5. The Hold feature is used to prevent problems that may occur during
calibration if the current outputs are used for control. The feature is
turned on (enabled) here and is specifically configured in step 4. The
Output 2 control controller starts out with the hold feature turned off (disabled).
Output 2 setup To enable the Hold feature, obtain the screen to the left with the hold
feature setup highlighted (see steps 1 and 2 for exact instructions).
Hold feature setup Press Enter 4 and the screen below will appear.
Exit Enter
7.00 pH When the hold feature has been enabled, this Hold Mode Screen will
appear when the Calibrate routine is entered. Possible actions are Exit
1 which cancels the calibration, Enter 4 which enters the calibrate
Hold mode: Off menu without putting the controller in hold, and Edit 4 which allows
turning Hold Mode On. Note that when hold has been enabled, this
Exit Enter screen requires pushing Cont 3 to enter and leave the calibrate menu.
35
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
Exit Edit
37
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
These parameters are available for alarms that have been config-
ured as TPC alarms in step 3. Parameters can be adjusted using the
Derivative: 0.0% same technique as in step 4a, by highlighting the desired item and
pressing the Edit 4 key. Once Edit has been pressed, change the
Time period: 30 sec item as needed and then press Save 4 to store the value.
URV: 2.00 pH "Setpoint" is usually the desired value at which the process is being
controlled, typically the alarm will not be on very much when the
Exit Enter process is at this value. This setpoint is also accessible in the
Program Menu under "Alarm Setpoints" (see Section 5.2)
"Proportional" is short for Proportional Band and indicates the
range over which control is being used. It is the opposite of the
URV: 2.00 pH process gain. Smaller values provide tighter control.
"Integral" is the number of seconds over which deviations from the
LRV: 0.00 pH setpoint are integrated to remove continuing offsets. Smaller values
Relay default: None provide higher response.
"Derivative" is a form of control that resists all changes in readings.
Exit Enter
Higher readings increase the derivative function. Use caution in set-
ting the derivative value to prevent process oscillation.
"Time period" is the cycle time for the TPC control. One cycle con-
sists of an energized (relay on) time and an deenergized (relay off)
time. The relative amounts of on time and off time depends on the
reading and the other settings listed here.
"URV" is the deviation from the setpoint that results in the alarm
being on all the time.
"LRV" is the deviation from the setpoint that results in the alarm being
off all the time. This should be set to 0.00.
EXAMPLE 1: A setpoint of 6 pH with URV of +2.0 pH and LRV of 0.0
pH, a time period of 30 seconds. When the pH is 7.0, the relay will
be on (7-6)/(2-0) = 50% of the time, 15 seconds each time. If the set-
point is changed to 6.5 pH, the relay will be on 7-6.5/(2-0) = 25% of
the time. The relay would then be on for 7 seconds and off for 23 sec-
onds.
"Relay Default" determines how the relay will act if there is a fault or
hold condition. Each alarm can be forced on (Close), off (Open) or
can remain unchanged (None). The original configuration is "None".
38
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
CAUTION
Understanding where to set the TPC parameters is not
trivial and is likely to require substantial trial and error to
yield acceptable results. pH and ORP/Redox are nonlin-
ear measurements and applying PID algorithms can
result in unintended effects.
Alarm 4 Setup 5. Alarm 4 is dedicated as a fault alarm. The only option for this alarm
is to enable it or to disable it. To disable the alarm, press Edit 4 and
use the arrow key to change "Fault" to "Off" .
Alarm: Fault When a fault condition exists, the relay will energize and the red LED
on the front display will turn on.
Exit Enter
7.00 pH pressed, at which time the controller returns to normal operation and
the feed limit's clock starts again. See Table 6-1, Controller Mode
Priority Chart, for controller action in the event of several modes
occurring at the same time.
26.2°C 12.0mA
NOTE
Feed limit alarm 1 Pressing the 2 (Ack) key will acknowledge all condi-
tions that turn the red LED on. If another event occurs
after the key is pressed, then the key must be pressed
again to acknowledge the new event. This is the only
way to clear a Feed Limit Timeout.
39
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
40
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
“On” “On”
Time Time
(sec) (sec)
RELAY RELAY
ACTIVATION ACTIVATION
Sequence A: One repeat per cycle Sequence B: Two repeats per cycle
41
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
This section describes the options available for configuration of the on-
Alarms line sensor diagnostics. The controller checks the integrity of the glass
electrode (for pH only) and the reference electrode by continuously meas-
Diagnostics uring the impedance of each. The original controller configuration is with
diagnostic messages turned off. To enable this feature, see below.
Auto Calibration
1. Beginning from the main menu, move the cursor down to "Program"
Exit Enter and press Enter (F4). From the program menu, move the cursor down
using the arrow key to highlight "Configure" and press Enter (F4).
Use the arrow key again to highlight "Diagnostics" (as shown on the
left) and press Enter (F4).
44
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
SECTION 6.0
THEORY OF OPERATION
6.1 THE pH SENSOR ASSEMBLY designed to avoid blocking of the internal liquid junc-
tion. It is essentially a complete electrode within an
The pH measurement is accomplished by means of a electrode outer body, using two liquid junctions.
measuring electrode (usually made of glass) which
develops a potential directly related to the hydrogen ion A non-reactive electrolyte (potassium chloride gel) is
concentration (pH) of the solution in which the elec- the filling solution in the outer body, and only this solu-
trode is immersed. A second electrode, called the ref- tion is in contact with the process. Clogging of the outer
erence electrode, is necessary to complete the electri- junction is minimized, since neither potassium ions nor
cal circuit and to serve as a constant reference poten- chloride ions form insoluble compounds with the
tial against which the potential of the glass electrode majority of materials found in process streams.
can be compared. Together, the two electrodes com- Contamination of the inner junction is not likely with this
prise the pH sensor. configuration.
45
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
46
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
• Time period value approaches the setpoint, the pump will be feed-
• URV point (or 100% on) ing for shorter and shorter intervals, and the chemicals
will be allowed to mix for longer and longer intervals of
• LRV point (or 0% on)
the period. This relationship is illustrated in Figure 6-1.
• Proportional
The exact amount of on time and off time per period is
• Integral determined by the settings for proportional, integral,
• Derivative and derivative bands. The proportional band (P) in %
is a separate adjustment that narrows (or widens) the
The setpoint is the desired value that you want to con- range of the TPC 0-100% action. Smaller values are
trol at. Time period is programmed in seconds and de- used for more control response. For a setpoint of 7 pH,
fines the interval during which the controller compares a URV of 2 pH, and P=100%, a pH reading of 8 would
the pH input from the sensor with the Setpoint. In the result in a relay on (8-7)/((2-0)*(100%)) or 50% of the
TPC mode the controller divides the period up into time. If P was changed to 50%, the same relay would
pump on-time (feed time) and pump off-time (blend be on (8-7)/((2-0)*50%) or 100% of the time.
time).
The integral band is set in seconds and acts to
The URV setting determines how far the pH must devi- increase the controller output as more time is spent
ate from the setpoint to get the pump to be on for the away from the setpoint. A smaller value in seconds will
entire period. The LRV setting determines how close result in faster integration response. Too low a value
the pH must be to the setpoint for the pump to be off for will result in excess oscillation.
the entire period. The LRV setting should always be set
at zero. When the error (the pH minus the setpoint) is The derivative band is set in % and acts to prevent
between the URV and LRV values, the relay will be en- changes in the reading. This setting should generally
ergized for some portion of the time period. As the pH be set to zero for pH and ORP applications.
47
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
TPC offers precise control by forcing the pump to feed 6.7 ANALOG OUTPUTS
chemical for shorter periods of time as you approach
the desired setpoint. If the process faces a large upset, The Model 54e pH/ORP controller includes a second
TPC mode forces the pump to feed chemical for longer analog output. An analog output produces an electri-
periods of time as the process deviates further from the cal current signal which varies in linear proportion to a
setpoint. This action continues until the pump is feed- value measured by the controller. You can configure
ing all the time, providing a speedy recovery from large the controller to produce a 4-20 mA or 0-20 mA cur-rent
up-sets. output proportional to pH or temperature. See Section
5.6 for programming details.
The controller can be programmed to be direct or re-
verse acting, depending on the pH (or temperature) The analog output must be "scaled" so that 4 (or 0) mA
value selected for URV. For example, if the controller is corresponds to the low end of the scale and 20 mA cor-
direct acting based on pH, such as in caustic chemical responds to the high end. The operator can scale the
addition control, the pH will rise as chemical is added, output as in the following example:
so the URV value will be below the LRV (i.e. below The Model 54e pH/ORP is connected to a strip chart
zero). As the pH rises toward the control point value, recorder with a 0 to 100% scale. The average value of
the pump will be on for gradually less time. Conversely, the bath is 10.0 pH, plus or minus 0.5 pH. The opera-
if the controller is reverse-acting based on pH, such as tor wants to match this value with the 50% mark on the
in acid addition for control, the pH will drop as acid is recorder. To do so, the operator selects 9.5 pH as the
added, and the URV value will be positive. The pH will 4 mA value and 10.5 pH as the 20 mA value by enter-
fall toward the control point value, and the pump will be ing them as in Section 5.3. The chart on the recorder
on for gradually less time. will display 0% when the pH is 9.5 or below and 100%
Complete TPC configuration is explained and typical when the pH equals 10.5 or greater. (this is for a 4-20
settings for these parameters are listed in Section 5.0. mA recorder).
After startup, the operator needs to adjust only the 0% The Model 54e pH/ORP is also capable of PID control
On to maintain the desired chemical concentration. where the analog output will be proportional to the dif-
ference between the setpoint and the measured vari-
able, either pH or temperature. This control mode is
6.6 NORMAL MODE used to modulate a pump or valve, rather than to turn
Normal mode is on-and-off control based on an alarm a device off and on.
setpoint. To prevent nuisance alarms, a hysteresis
(deadband) setting, and/or a time delay can be pro-
grammed during configuration. You can configure each
alarm to trigger above the setpoint as a high alarm or
below the setpoint as a low alarm. The operator need
only raise or lower the alarm setpoint as necessary.
48
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
Your Model 54e pH/ORP can function in different modes Priority is in the following order (from lowest to highest):
depending on both how it is configured, what process normal, fault, timer, hold, feed limit, test. Each output or
conditions exist, and actions an operator may have relay acts as if it is only in the state of highest priority.
made. To reconcile these possible modes, there is a set
NOTE
priority that determines exactly what will happen to the 2
current outputs and the 4 alarm relays in the event of mul- Some of these features may not be in use for your con-
tiple modes occurring at the same time. See Table 6-1 troller.
below.
1 Indicates the state IF that item has been configured or selected (i.e. if it is an interval timer or a feed limit
timer or it is the one being tested). Unconfigured or unselected items are not affected by that mode.
Condition Definitions:
1. Normal refers to conditions when no other mode is present.
2. Fault is when the instrument has diagnosed a fault condition.
A fault message is displayed and the red LED will be on.
3. Interval Timer is only while the timer sequence is occurring.
4. Hold Mode is while hold is activated by the operator (i.e. during calibration).
5. Feed Limit occurs when a feed limit timer has reached it's limit and is turned off after being on for too long.
6. Simulate tests are described in Section 5.4.
Action Definitions:
1. Normal is determined by process conditions or how the item has been configured (Sections 5.5, 5.6)
2. Open is a deenergized alarm relay. (alarm off)
3. Default is the setting configured for each item if there is a fault. (Sections 5.5, 5.6)
4. Closed is an energized alarm relay. (alarm on)
5. Hold is the setting for the current output configured in Section 5.5 (this could be a fixed mA value or the last
normal value)
6. Prior is the state the alarm had before that mode occurred.
7. Test is the value input by the operator (mA for current, on or off for a relay).
49
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
50
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
Proportional Mode (Gain) The reset mode repeats the proportional action as long
The simplest control is proportional. Proportional may as an offset from the set point exists. Reset action is
also be referred to as sensitivity or gain. Although these cumulative. The longer the offset exists, the more the
terms may refer to a different version of proportional, output signal is increased.
the control function is still fundamentally the same - the The controller configured with reset continues to
error from set point is multiplied by this factor to pro- change until there is no offset. If the offset persists, the
duce the output. reset action eventually drives the controller output to its
The Model 54e pH/ORP's proportional mode is 100% limit - a condition known as "reset windup". To
referred to as proportional "band" which is config- prevent reset windup, a controller with reset mode
urable from 0 to 299%. For good control of a specific should never be used to control a measured variable
process, the proportional band must be properly influenced by uncorrectable conditions. Once the con-
adjusted. The proportional band is the percent of the troller is "wound up", the deviation must be eliminated
analog output span (the difference between the 4 (or 0) or redirected before the controller can unwind and
mA and 20 mA settings) through which the measured resume control of the measured variable. The integral
variable must move to change the output from mini- time can be cleared and the "windup" condition quickly
mum to maximum. The larger the proportional band, eliminated by manually overriding the Model 54e
the less the controller reacts to changes in the pH/ORP's analog output using the simulate tests
measured variable. As the proportional band is feature (detailed in Section 5.4).
made smaller, the reaction of the controller
increases. At 0 proportional band, the proportional- Control Loop Adjustment and Tuning
only controller behaves like an on/off controller (an
alarm set at 20 mA). There are several methods for tuning PID loops includ-
ing: Ziegler-Nichols frequency response, open loop
Most processes require that the measured variable be step response, closed loop step response, and trial and
held at the set point. The proportional mode alone will error. Described in this section is a form of the open
not automatically do this. Proportional alone will only loop response method called the process reaction
stabilize the measured variable at some offset to the curve method. The reaction times and control charac-
actual control point. To control at an exact setpoint, teristics of installed equipment and real processes are
proportional plus integral mode is used. difficult to predict. The Process Reaction Curve Method
of tuning works well because it is based on the
Proportional (Gain) Plus Integral (Reset) response of the installed system. This procedure, out-
lined in the following paragraphs, can be used as a
For the automatic elimination of deviation, I (Integral starting point for the P and I settings. Experience has
mode), also referred to as Reset, is used. The propor- shown that PID controllers will do a fair job of control-
tional function is modified by the addition of automatic ling most processes with many combinations of rea-
reset. With the reset mode, the controller continues to sonable control mode settings.
change its output until the deviation between measure-
ment and set point is eliminated.
The action of the reset mode depends on the propor-
tional band. The rate at which it changes the controller
output is based on the proportional band size and the
reset adjustment. The reset time is the time required
for the reset mode to repeat the proportional action
once. It is expressed as seconds per repeat,
adjustable from 0-2999 seconds.
51
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
Process Reaction Curve Method This can be done by observing the reading on the
A PID loop can be tuned using the Process Reaction Model 54e pH/ORP and noting values at intervals
Curve Method. This method involves making a step timed with a stop watch. A strip chart recorder can be
change in the chemical feedrate (usually about 50% of used for slower reacting processes. To collect the data,
the pump or valve range) and graphing the response of perform the following steps:
the Model 54e pH/ORP reading versus time. 1. Let the system come to a steady state where the
The process reaction curve graphically shows the reac- measured variable (pH, conductivity or tempera-
tion of the process to step change in the input signal. ture) is relatively stable.
Figure 6-2 shows an example of a tuning process for a 2. Observe the output current on the main display of
pH controller. Similar results can be obtained for the the controller.
conductivity controller. 3. Using the simulate test, manually set the controller
To use this procedure with a Model 54e pH/ORP and a output signal at the value which represented the
control valve or metering pump, follow the steps out- stable process measurement observed in step 1,
lined below. then observe the process reading to ensure steady
The Model 54e pH/ORP should be wired to the control state conditions (a stable process measurement).
valve or metering pump. You will introduce a step 4. Using the simulate test, cause a step change in the
change to the process by using the simulate test func- output signal. This change should be large enough
tion to make the step change in the output signal. to produce a significant change in the measured
The change in the measured variable (conductivity, pH, variable in a reasonable amount of time, but not
or millivolts) will be graphed as shown in Figure 6-2. too large to drive the process out of desired limits.
52
MODEL 54e pH/ORP SECTION 6.0
THEORY OF OPERATION
5. The reaction of the system, when graphed, will In the example, the process Delay Time (D) was four
resemble Figure 6-2, showing a change in the seconds and the Response Time period (L) was 12
measured variable over the change in time. After a seconds, so:
period of time (the process delay time), the meas-
ured variable will start to increase (or decrease) L 12 seconds
rapidly. At some further time the process will begin R= =3
D 4 seconds
to change less rapidly as the process begins to sta-
bilize from the imposed step change. It is important The last parameter used in the equations is a plant gain
to collect data for a long enough period of time to (C). The plant gain is defined as a percent change in the
see the process begin to level off to establish a tan- controlled variable divided by the percent change in
gent to the process reaction curve. manipulated variable; in other words, the change in the
6. When sufficient data has been collected, return the measured variable (pH, conductivity, temperature) divid-
output signal to its original value using the simulate ed by the percent change in the analog output signal.
test function. Maintain the controller in this manual The percent change in the controlled variable is
mode until you are ready to initiate automatic PID defined as the change in the measured variable (pH,
control, after you have calculated the tuning con- conductivity, temperature) compared to the measure-
stants. ment range, the difference between the 20 mA (Hi) and
4 (or 0) mA (Lo) setpoints, which you determined when
Once these steps are completed, the resulting process
configuring the analog output.
reaction curve is used to obtain information about the
overall dynamics of the system. It will be used to cal- In the example shown in Figure 6-2:
culate the needed tuning parameters of the Model 54e The percent change in pH was:
pH/ORP controller.
NOTE pH2 - pH1 8.2 - 7.2 pH
x 100% = = 33.3%
The tuning procedure outlined below is pH “Hi” - pH “Lo” 9.0 - 6.0 pH
adapted from "Instrumentation and
Process Measurement and Control", The change in the output signal was:
by Norman A. Anderson, Chilton Co.,
Radnor, Pennsylvania, ©1980. 6 - 4 milliamps
x 100% = 12.5%
Information derived from the process reaction curve 20 - 4
will be used with the following empirical formulas to
predict the optimum settings for proportional and inte- So the Plant Gain is:
gral tuning parameters.
33.3
Four quantities are determined from the process reac- C= = 2.66
tion curve for use in the formulas: time delay (D), time 12.5
period (L), a ratio of these two (R), and plant gain (C).
Once R and C are calculated, the proportional and inte-
A line is drawn on the process reaction curve tangent gral bands can be determined as follows:
to the curve at point of maximum rise (slope) as shown C
in Figure 6-2. The Time Delay (D), or lag time, extends Proportional band (%) = P = 286
R
from "zero time" on the horizontal axis to the point Integral Time (seconds per repeat) = I = 3.33 D x C
where the tangent line intersects the time axis. The
Response Time period (L), extends from the end of So for the example:
delay period to the time at which the tangent line inter- 286 (2.66)
P= = 254%
sects the 100% reaction completion line representing 3
the process stabilization value. The ratio (R) of the
Response Time period to the Time Delay describes the I = 3.33 (4 sec.) 2.66 = 36 seconds
dynamic behavior of the system. To enter these parameters, use the procedure detailed
in Section 5.6.
53
MODEL 54e pH/ORP SECTION 7.0
SPECIAL PROCEDURES AND FEATURES
SECTION 7.0
SPECIAL PROCEDURES AND FEATURES
This section covers features of the Model 54e Level 1 - 3 Password Privileges
pH/ORP controller that are used less frequently.
Level 1 access is usually given to an operator who
Use of the features outlined in the appendix is
simply needs to calibrate during the course of normal
optional.
operation. A separate section of the manual contains
Special procedures and features outlined in this operating procedures normally used by this type of per-
appendix include the following: sonnel. Level 1 restricts the operator from changing the
• Password Protection major control mode configuration by preventing access
to the Program Menu.
• Temperature Slope Calculation
The Level 1 user can do the following:
• Temperature Sensor
1. Access Diagnostic Variables (Section 8.1).
• Reference Temperature
2. Zero the controller in air (Section 4.3).
• Controller Mode Priority
3. Enter the Temperature Slope (Section 4.4).
• PID Control
4. Change Temperature Compensation from Auto to
Before using this section, you should become familiar Manual and select a temperature (Section 4.7).
with the basic Theory of Operation of the controller as
outlined in Section 6.0, the keypad functions in Section 5. Calibrating pH and Temperature readings (Section
1.0, and the List of Settings Table and configuration 4.1 and Section 4.6).
procedures outlined in Section 5.0. A Level 2 user can do all of the above and:
As with all the settings in your Model 54e pH/ORP, the 1. Change control setpoints for PID current outputs
first step to configuration is obtaining a good under- (Section 5.1).
standing of how the feature works, before determining
2. Change alarm setpoints for normal and TPC
the values of the settings to achieve the desired con-
alarms (Section 5.2).
trol. This appendix provides more background for
deciding on the appropriate settings. Configuring the 3. Rerange both 4-20 (or 0-20) mA outputs
settings is done using the instructions in this appendix (Section 5.3).
and Section 5.0, Software Configuration. 4. Manually test both outputs and all 4 alarm relays
for operation.
7.1 PASSWORD PROTECTION A Level 3 user has total access to the Configure Menu
and can make any changes that are deemed neces-
Your Model 54e pH/ORP can be programmed so that a
sary.
3-digit password must be entered before any changes
in the configuration are allowed. This protects your These privileges should be given only to an indi-
controller from tampering by unauthorized users. There vidual who fully understands the controller, the
are three levels of password access, Level 1 (calibra- process and the potential effects of modifying the
tion only), Level 2 (lockout of Configure Menu), and setup.
Level 3 (total access). Password privileges for each An individual with no password access privilege can
level are described below. only view the main display, containing conductivity,
If password protection is not desirable, you can config- temperature, current output 1, and the lower line dis-
ure all security codes to be 000. This will leave the con- play items configured in Section 5.5.
troller unlocked so the configuration can be changed
NOTE
without entering a password. The controller is shipped
from the factory with the password set at 000. You must have level 3 access to
change any security code.
54
MODEL 54e pH/ORP SECTION 7.0
SPECIAL PROCEDURES AND FEATURES
1. Beginning from the main menu, move the cursor down to "Program"
and press Enter 4. From the program menu, move the cursor down
using the arrow key b to highlight "Configure" and press Enter 4.
Use the arrow key again to highlight "Security" (as shown on the left)
and press Enter 4.
NOTES ON SECURITY:
a. A code of 000 disables security for that level.
b. The security feature will not activate until the keypad has not
been pressed for a short period of time (the timeout value pro-
grammed in Section 5.5).
c. A hold condition will indefinitely prolong the time out period.
d. Security will activate immediately if power is removed from the
controller and then restored.
e. Forgotten Code: Press and hold F4 Key for 5 seconds when the
security screen appears and the code for that level will appear on
the display.
55
MODEL 54e pH/ORP SECTION 7.0
SPECIAL PROCEDURES AND FEATURES
Temp coeff: +.000 Sensor Temperature Compensation. The millivolt output of a pH sen-
sor changes with temperature, which in turn, changes the measured pH
Operate iso: +07.00 pH value. The Sensor Isopotential pH (Sensor ISO) is the pH where the pH
sensor's output does not change with temperature. The Controller uses
Sensor iso: 07.00 pH the Sensor Isopotential pH value to correct for the temperature effect on
the pH sensor. Except in rare instances (antimony or glass on metal elec-
Exit Edit
trode systems), the Sensor Isopotential pH is always 7.0 pH.
Solution Temperature Compensation. The actual pH of a solution can
sometimes noticeably change with temperature. This can occur in solu-
tions with a pH above 6.5 pH and in solutions containing weak acids,
bases, and their salts. This effect can be compensated for by entering a
Temperature Coefficient (Temperature Coeff) which is the change in solu-
tion pH per degree centigrade (°C). If there is no noticeable solution pH
change with temperature, or solution temperature compensation is not
desired, the Temperature Coefficient (Temperature Coeff) should be set
to 0.00.
Operating Isopotential pH. The Operating Isopotential pH (Operate
ISO) is the Isopotential pH, which the Controller is using for overall tem-
perature compensation. It represents a combination of both the sensor
and the solution temperature compensations.
If no solution temperature compensation is being used (Temperature
Coeff = 0), the Operating Isopotential pH (Operate ISO) is equal to the
Sensor Isopotential pH (Sensor ISO).
If there is a non-zero Temperature Coefficient (Temperature Coeff) being
used, the Controller automatically calculates the Operating Isopotential
pH (Operate ISO) from the following relation:
Operate ISO = Sensor Iso - (Temperature Coeff) X (298.16)
Conversely, if an Operating Isopotential pH is known for a particular solu-
tion, the Controller will automatically calculate the corresponding
Temperature Coefficient.
Example: High Purity water with Ammonia, being measured with a normal
pH sensor (Sensor Isopotential pH of 7.0 pH), has a Temperature
Coefficient of -.033pH per °C and an Operating Isopotential pH of 16.84.
56
MODEL 54e pH/ORP SECTION 7.0
SPECIAL PROCEDURES AND FEATURES
57
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
SECTION 8.0
TROUBLESHOOTING
The Model 54e pH/ORP automatically searches for Many control problems are unrelated to the pH meas-
fault conditions that would cause an error in the meas- urement system. When problems arise, first check
ured pH reading. If such a condition occurs, the current other systems that affect chemical concentration.
outputs and alarm relays will act as configured in Consider what may have changed in the system that
Section 5.6 and Section 5.7, the red "FAIL" LED on the can cause poor control. Some causes for poor control
controller panel will be lit and a diagnostic message will other than controller malfunction are:
be displayed. If more than one fault exists, the display 1. An empty chemical drum.
will sequence through the diagnostic messages. This
will continue until the cause of the fault has been cor- 2. Malfunction of a chemical feed pump, pump motor,
rected or until the Ack 2 key is pressed. or motor starter.
3. Water inlet or drain valves stuck or left open by
operators.
Troubleshooting is easy as 1, 2, 3…
4. Check flow interlocks (if used).
Step 1 Look for a diagnostic fault message on the dis-
play to help pinpoint the problem. Refer to 5. A temperature control malfunction.
Table 8-1 for an explanation of the message 6. Broken or blocked chemical feed lines.
and a list of the possible problems that trig-
gered it. 7. A conductivity probe that has been left out of the
bath.
Step 2 Refer to the Quick Troubleshooting Guide,
Table 8-2, for common pH hardware problems 8. The level of bath is below the probe and the probe
and the recommended actions to resolve them. is dry.
Step 3 Follow the step by step troubleshooting 9. The probe needs to be cleaned.
approach offered in Table 8-3 to diagnose less 10. The condition of the incoming metal has changed,
common or more complex problems. i.e., temperature, cleanliness, speed.
11. The condition of the incoming water has changed,
CAUTION i.e., temperature, cleanliness, flow rate, hardness,
pH.
Do not attempt to troubleshoot unless
you have familiarized yourself with this 12. Unauthorized personnel have tampered with the
manual. Only trained, qualified techni- controller settings.
cians should perform these procedures. 13. Standardizing procedure is not accurate due to a
Do not attempt to troubleshoot, repair, or malfunctioning laboratory instrument or contami-
modify the printed circuit cards or elec- nated chemical standard solutions.
tronic components inside the controller.
WARNING
IMPORTANT
Always recalibrate the controller To prevent chemical feed into the
before returning it to service. process or injury to operating personnel,
disconnect or disable the chemical feed
pump and other external devices while
you are servicing and troubleshooting
the controller.
58
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
* Off line error messages will not initiate fault condition, and will display only once. The message will clear from
the screen when a key is pressed.
59
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
SYMPTOM ACTION
pH reading won't change in different buffers. Clean the electrode, check wiring. Replace electrode.
"Cracked glass failure". 1. Replace electrode if cracked.
2. Check wiring for short.
pH sensor has sluggish response. Clean the electrode; if still sluggish soak in 1% HCl for 1 hr.
Replace electrode if not rejuvenated by the HCl soak.
"Old glass warning". Check sensor in buffers; replace if calibration unsuccessful.
"Lo or Hi slope error". 1. Verify internal preamp switch is in the appropriate position.
2. Improper buffer calibration: check buffer accuracy; wait for reading
to stabilize;
3. Clean the electrode; if fault persists; replace electrode.
4. If new electrode doesn't resolve the fault; replace preamp.
"Calibration warning". Check sensor in buffers; replace if calibration unsuccessful.
"Wait" flashing continuously; won't stabilize 1. Readjust stabilization pH or time for Auto Cal (Section 5.9)
during Auto Buffer Calibration. 2. Clean the sensor and retry Auto Cal in buffers.
3. Try a Manual Buffer Calibration. (see Section 5.9 for configuration)
Calibrates in buffers but not in the process 1. Verify process reading to be correct
2. Possible ground loop. Make sure that the shield wires do NOT
touch grounded metal. (Process noise).
Incorrect temperature reading. 1. Standardize the temperature
Suspected temp. compensation problem. 2. Verify sensor's RTD resistance vs. temperature
"Temp. error high" 3. Verify temperature reading to be correct
"Temp. error low"
Display segments missing. Replace Display board.
Analyzer locks up; won't respond. Replace CPU board.
Erratic display and relays chattering. Check alarm set points, configuration (Sections 5.2, 5.7)
Analyzer not responding to key presses. Verify and clean ribbon cable connection on CPU board.
Key press gives wrong selection. Replace enclosure door/keyboard assembly.
Wrong or no current output. 1. Verify that output is not being overloaded (max load is 600 ohms)
2. Rerange outputs (Section 3.3)
3. Replace Power board
No display or indicators. Verify that the removable fuse module is securely seated.
Alarm relay closure problems. Check Power board.
Power board cut-off. Replace the Power board.
Sensor diagnostics faults keep appearing 1. Calibrate sensor in buffers; replace if unsuccessful
2. Check the diagnostic variables for inappropriate settings (Section 5.8)
3. Turn analyzer diagnostics "OFF" (Section 5.8)
4. Verify proper wiring and preamp switch position.
5. Perform Systematic Troubleshooting procedures.
60
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
Input: 120 mV 3. Diagnostic variables are displayed three at a time. More variables are
viewable until the cursor (showing highlighted text) is brought down
Glass imped: 300 MΩ to the bottom line.
Ref imped: 8 kΩ Use the down arrow key to view the items on the lower screen.
Exit The up arrow key can be used to return to a previously viewed item .
Press the Exit (F1) key to return to the main menu above.
NOTE
Many diagnostic variables can be read directly on the main dis-
Zero offset: 0 mV play in the lower left or lower right positions. For details, see
Section 5.5.
pH slope: 59.02 mV/pH
Version: pH1.01
Exit
61
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
62
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
63
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
64
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
65
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
Displayed pH value not the same Grab sample incorrect - Calibration, Section 4.0
as grab sample of process. Reevaluate sample technique and
equipment.
Unclear which is correct - Calibration, Section 4.0
Retest with new buffer solution.
Controller out of calibration - Calibration, Section 4.0
Recalibrate per Calibration,
(See Section 3.0)
Outputs do not change Controller is in Hold or simulate mode - Software Configuration, Section 5.6
"Hold Mode activated" To remove from Hold, press any key Calibration , Section 4.7
"simulating output 1 or 2" and then press Enter (F4). Press Edit (F4),
"simulating alarm 1,2,3, or 4" change "On" to Off", and press Save (F4).
To stop simulation, press Exit (F1).
When the unit has been put into "Hold",
all outputs go to their default states.
Red LED light is on Unit has gone into fault mode - Section 8.0
Various fault messages Read the message and take action as
shown on lower display. indicated in Table 8-2.
68
MODEL 54e pH/ORP SECTION 8.0
TROUBLESHOOTING
NOTE: Individual printed circuit boards cannot be ordered for Model 54e. Replacement boards for Model 54e
are assembled and calibrated as an integrated board stack.
69
MODEL 54e pH/ORP SECTION 9.0
RETURN OF MATERIAL
SECTION 9.0
RETURN OF MATERIAL
9.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is
important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)
number.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individ-
ual to be contacted should additional information be needed.
3. Do Steps 3 and 4 of Section 9.2.
NOTE
Consult the factory for additional information regarding service or repair.
70
MODEL 54e pH/ORP APPENDIX A
ORP CONFIGURATION
APPENDIX A
ORP CONFIGURATION
71
MODEL 54e pH/ORP APPENDIX A
ORP CONFIGURATION
72
MODEL 54e pH/ORP APPENDIX A
ORP CONFIGURATION
73
MODEL 54e pH/ORP APPENDIX A
ORP CONFIGURATION
Accessing Calibrate, Program and Configure Level 2 Program. To access the program level from
Menus. Operating configuration changes are made at the main menu, place the cursor over "Program" with
the levels shown in Figure A-1. Press any key from the down arrow key. Then press Enter (F4). From the
the main display to access the main menu (top left). program level menu, changes can be made to the
Refer to Section 5 for pH measurements. alarm setpoints and the output setpoints.
Level 1 Calibrate. To access calibration selections Level 3 Configure. To access the configure level
from the main menu, with the cursor on "Calibrate" from the main menu place cursor over "Program" and
press Enter (F4). ORP standardization and tempera- Enter (F4) then place cursor over "Configure" and
ture adjustments are made at this level (refer to Enter (F4). This level contains advanced selections,
Section A.1 and Section 3.1 for these procedures). such as alarms and diagnostics.
Calibrate
Diagnostic variables
Program PRESS TWO ARROWS b
and then ENTER
Exit Enter
Display
Alarm setpoints Outputs
PRESS ENTER Output setpoints Alarms
Simulated tests Exit More Enter
Exit Enter
Standardize PRESS
ARROWS
Adjust temperature PRESS
ARROWS b
Exit Enter b
CALIBRATE MENUSEC- Diagnostics
TION Security
Buffer Calibration 4.2, 4.3
Standardize pH 4.4 Simulated tests Exit More Enter
Adjust Temperature 4.1 Configure
Temp Compensation4.5 CONFIGURE MENUSEC-
pH slope 4.6 TION
Exit Enter
Display 5.5
PRESS
PROGRAM MENU SEC- Outputs 5.6
ENTER
TION Alarms 5.7
Alarm setpoints 5.1 Diagnostics 5.8
Output setpoints (PID)5.2 Security 7.1
Output setpoints 5.3
Simulated tests 5.4
74
MODEL 54e pH/ORP APPENDIX A
ORP CONFIGURATION
400 mV 5. The number on the large display is the live process ORP reading. The
next line displays the ORP reading when the display was accessed.
Press Edit (F4) to perform the standardize.
Standardize: 400 mV Use the arrow b key to change the reading to the desired value and then
press Save (F4) to complete the procedure. Esc (F3) will cancel.
Exit Enter
NOTE
The + sign can be changed to a - sign after the numerical portion
of the number has been entered by highlighting the + and press-
ing the down arrow key.
400 mV Before exiting the calibration mode, remember to change the hold mode
setting to OFF (if it was turned on in step 3).
Standardize: +0402 mV
Esc Save
75
RETURN OF MATERIALS REQUEST •IMPORTANT!
This form must be completed to ensure expedient factory service.
ADDRESS ___________________________________________________________________________________________________
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
right now.
http://www.raihome.com