MAINTENANCE MANUAL

VISERA ELITE XENON LIGHT SOURCE

CLV-S190

ISSUE 3
2015/11
148P-1264
CLV-S190

CONTENTS
Chapter 1: Introduction .......................................................................... 1-1
1 Introduction ......................................................................................................................... 1-1
2 Applicable unit .................................................................................................................... 1-1
3 Copyright ............................................................................................................................. 1-1
4 Trademarks.......................................................................................................................... 1-1

Chapter 2: Product Outline .................................................................... 2-1


1 Outline ................................................................................................................................. 2-1
2 Features ............................................................................................................................... 2-1
3 Limitations........................................................................................................................... 2-2
4 Product Specifications ....................................................................................................... 2-2
5 Nomenclature and Functions ............................................................................................ 2-7
6 Connector .......................................................................................................................... 2-13

Chapter 3: Specifications ...................................................................... 3-1


Chapter 4: Installation and Connection ................................................ 4-1
Chapter 5: Care, Storage, and Disposal ............................................... 5-1
Chapter 6: Inspection ............................................................................. 6-1
1 Inspection check Sheet ...................................................................................................... 6-1

Chapter 7: Safety Check ........................................................................ 7-1


1 Electrical safety tests ......................................................................................................... 7-1
2 Safety check card ............................................................................................................... 7-4

Chapter 8: Block Diagram ..................................................................... 8-1


Chapter 9: Troubleshooting .................................................................. 9-1

Chapter 10: Parts replacement ........................................................... 10-1

ISSUE3 1 CONTENTS
CLV-S190

Chapter 1: Introduction

1. Introduction
Repair and maintenance of this product requires highly specialized knowledge and techniques.
We recommend that you contact an Olympus service center in your area if a problem develops with the
product.

If repairs or modifications are made by personnel not authorized by Olympus, the warranty is void, and
Olympus shall not be liable for damage that occurs to or as a result of use of the modified product.

2. Applicable unit
VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190

3. Copyright
©2011 Olympus Corporation. All rights reserved.
Unauthorized reproduction or distribution in part or in whole is prohibited.

4. Trademarks
OLYMPUS is a registered trademark of Olympus Corporation.
The company names, product names, and proprietary technical terms in this document are the trade-
marks or registered trademarks of their respective owners.

ISSUE3 1-1 Introduction


CLV-S190

Chapter 2: Product Outline


1 Outline
(1) Intended use
When used in conjunction with an OLYMPUS endoscope, video system center, video
monitor, accessories and other related equipment, this product supports examination and
endoscopic treatment by enabling observation through a fiberscope or through a video
monitor.
(2) Compatibility
This product may be used in conjunction with V series videoscopes, VISERA series
videoscopes, rigid light guides, fiberscopes, as well as OTV-S190, OTV-S7Pro.

2 Features
(1) This is a light source unit for exclusive use with the OTV-S190 and the OTV-S7Pro. this
product is to be used for surgery only.
(2) A mechanical detection unit is installed in the endoscope socket, enabling automatic
switching to the maximum amount of light appropriate for any endoscope connected. It is
capable of distinguishing between LG connectors for use with the SP high-intensity function,
and LG connectors for use with the non-SP high-intensity function.
(3) When videoscopes or OES fiberscopes are not compatible with CLV-S40Pro are attached,
the unit is capable of generating same as CLV-180 GI mode (1.3 times the output light
intensity of the CLV-U40 EVIS 100 mode).
(4) When connected to a surgical fiberscope not compatible with the high-intensity RF
connector, it produces an output light intensity of the maximum permissible illumination
(white light) for the LG equivalent to the CLV-S40Pro OES mode.
(5) When used with an endoscope and light guide compatible with the high-intensity RF
connector, the high-intensity mode is available, creating a brighter illumination. Compared
to the illumination of the CLV-U40 EVIS 100 mode or OES mode, approximately 2 times the
maximum illumination (white light) is produced. In high-intensity mode, it is also possible to
select equal brightness or double brightness of the CLV-U40 through the brightness mode
switch.
(6) The optical filter used in NBI (Narrow-Band Imaging) observation has been loaded. NBI
observations are enabled when an NBI-compatible endoscope is connected to the unit.
(7) Two other optical filters for specialized observations may be loaded if required.
(8) The examination lamp will turn on when the power is switched on (autoignition function).
(9) The examination lamp can be manually turned off by continously pressing the LAMP switch
for approximately 1 second.
(10) Heat exhaust is through the rear (rear heat exhaust).
(11) The system automatically switches to the emergency lamp if the examination lamp does
not come on after pressing the lamp switch. As an emergency response, this will provide
sufficient brightness required for endoscope removal in situations when the examination
lamp does not come on or if it turns off. This will be indicated on the panel.
(12) Brightness can be adjusted using the 17 steps of automatic and manual brightness
available.
(13) Disconnection of the emergency lamp will be automatically indicated on the panel.
(14) Panel settings are retained, and the light source unit will display at the previous settings the
next time the unit is turned on. Filter switch settings are not backed up, however, and will
revert to the normal filter.
(15) The exterior and the operating panel may be disinfected with ethanol for disinfection (70%
ethyl or isopropyl alcohol) or with a mild detergent.
(16) This product has passed the OLYMPUS Eco-Product standard (Environmentally-friendly
design standard).

ISSUE3 2-1 Product Outline


CLV-S190

3 Limitations
(1) Applicable video system centers:
OTV-S190, OTV-S7Pro.
(2) Use in a medical facility under the supervision of a medical doctor.
(3) Do not apply this light source directly to the heart. However, it may be used with the heart
when used in combination with a TYPE CF applied part.
(4) The outer casing of the unit must be grounded for safety.
(5) Do not use the light source in a combustible atmosphere.
(6) When the CLV-S190 is used with an endoscope, combined use with the following electrical
devices is prohibited.
a) Of devices designed to provide electrical treatment to patients, those that have not
been tested and found to be safe for use in conjunction with the CLV-S190.
b) Devices other than those intended to provide electrical treatment that have not been
tested and found to be safe from the standpoint of leakage current and the like.
(7) Power supply
a) Voltage: AC100-240V
b) Power: 500VA
c) Frequency: 50/60 Hz
d) Voltage fluctuation: ±10% or less
e) Frequency fluctuation: ±1 Hz or less

4 Product Specifications
4-1 Compatible endoscopes

(1) High-intensity compatible videoscopes


LTF-V, LTF-V2, LTF-V3, LTF-VP, LTF-S190-5, LTF-VH/VP-S, A4804A, A4805A, A50000A,
A50001A, A50010A, A50011A, A50020A, A50021A
(2) High-intensity uncompatible videoscopes
ENF-V, ENF-V2, ENF-VT/VT2/VQ, ENF-V3, ENF-VH, CYF-V/VA, CYF-V2/V2R/VA2,
CYF-VH/VHA/VHR, HYF-V, PEF-V, LF-V, VEF-V, URF-V, CHF-V
(3) Fiberscopes
CYF-3, CYF-4, CYF-4A, HYF-P, HYF-1T, HYF-XP, VEF, VEF-2, URF-P2, ENF-P3, ENF-P4,
ENF-XP, ENF-L3, ENF-T3, ENF-GP, PF-14, PF-22, PF-28, LTF, LF-1, LF-2, LF-P, LF-GP,
LF-TP, LF-DP, LF-T
(4) Light guide for use with rigid endoscopes
WA03200A, WA03202A, WA03212A, WA03210A, WA93200A
(5) Rigid endoscopes
Rigid endoscopes is available with light guide
(6) High-intensity compatible fiberscopes
CYF-5, CYF-5A, URF-P3, URF-P5, ENF-GP, CHF-CB30L/S, URF-P3, LF-TP, LF-DP,
LF-GP

Product Outline 2-2 ISSUE3


CLV-S190

4-2 ILLumination function


4-2-1 Optics
(1) The emergency lamp is automatically placed in the light path when the examination lamp
malfunctions.

4-2-2 Examination lamp


(1) Type: Xenon short arc lamp with an parabolic mirror
(2) Mode: MD-631
(3) Life: 500 hours (when lit continuously)
(4) The lamp will illuminate when the power is switched on(autoignition function). The lamp can
be turned off by pressing the LAMP switch continuously for one second.

4-2-3 Emergency lamp


(1) Type: Halogen 35 W (with reflector)
(2) Life: Average of at least 500 hours

4-2-4 Brightness Adjustment


(1) Manual adjusment only for fiberscopes and rigid-type endoscopes (excluding when used
with a OTV-S190/S7Pro and camera head).
(2) Automatic and manual adjustment available for videoscopes.
(3) Automatic adjustment is available when the OTV-S190/S7Pro and camera head is used
with rigid endoscopes and light-source cables.

4-2-5 NBI special observation method


(1) The filter mode switch enables selection of the NBI special observation mode when a light
adjustment cable and theOTV-S190/S7Pro are connected to an NBI-compatible
endoscope.
(2) When the NBI special observation mode is selected, the NBI special filter is placed in the
illumination light path.
Note) The OTV-S190/S7Pro power switch must be ON.

4-2-6 High-intensity mode


(1) When a high-intensity compatible endoscope is inserted, the high-intensity mode may be
selected by pressing the high-intensity switch. High-intensity mode may not be selected
with endoscopes that are not high-intensity compatible.
(2) When a high-intensity compatible endoscope is inserted, the high-intensity mode switch
illuminates. This enables the selection of the high-intensity mode (LED above the switch
illuminates) or the normal mode (LED above the switch goes out).
(3) The current settings for the high-intensity mode switch are retained when the power is
turned OFF. When a high-intensity compatible endoscope is inserted, illumination will either
be the high-intensity mode or normal mode, depending on the settings retained by the
system. Initial factory setting: normal mode.
(4) When the high-intensity mode is terminated (normal mode), illumination is equivalent to the
CLV-U40 EVIS 100 and OES mode.
(5) High-intensity mode provides approximately twice the amount of illumination as the SP
normal observation mode.

4-2-7 Manual adjustment


(1) Mechanical diaphram (set on the panel)
(2) 17 steps

ISSUE3 2-3 Product Outline


CLV-S190

4-2-8 Automatic adjustment


(1) Mechanical diaphram with constant illumination control.
(2) 17 steps

4-2-9 Front panel display method in special observation mode


Observation mode Front panel display method
NBI special observation mode The NBI LED lights up (white)
PDD normal observation mode Filter button, “STBY” of filter mode
selection indicators and PDD LED of
filter mode selection indicator light up.
*Only Europe specification
PDD special observation mode Filter button, “ON” of filter mode
selection indicators and PDD LED of
filter mode selection indicator light up.
*Only Europe specification

4-2-10 Cooling
(1) Forced air cooling with a fan (rear exhaust)

4-3 Connection
4-3-1 Endoscope and light guide
(1) “One-touch” connection

4-3-2 Light-supply cable (MAJ-1411)


(1) Connector on the rear-panel (Light-supply cable for OTV-S7Pro connection)

4-3-3 Light-supply cable (MAJ-1959)


(1) Connector on the rear-panel (Light-supply cable for OTV-S190 connection)

4-3-4 Foot switch (MAJ-1391)


(1) Foot switch connector on the rear-panel

4-4 Emergency indication


4-4-1 Emergency lamp indicator
(1) Indicates if the emergency lamp is disconnected or if the emergency lamp is in use (when
the main lamp will not illuminate)
Lit up : Emergency lamp in use (the main lamp will not illuminate)
Blinking : Emergency lamp is disconnected or has been removed.

4-5 Panel
4-5-1 Setting retention
(1) Settings prior to powering off the unit are retained.
Manual / Automatic brightness settings
Light intensity level
High-intensity mode setting
Special observation function display
Lamp life indicator(with the power ON, the indicator may be reset when the examination
lamp is extinguished by continuously pressing the counter reset switch.)

Product Outline 2-4 ISSUE3


CLV-S190

4-6 Safety
4-6-1 Dimming
(1) When the endoscope is removed, the light emanating from the scope connector is dimmed
by the dimmer plate located in the endoscope socket.

4-6-2 Disinfection
(1) Disinfect with ethanol for disinfection (70% ethyl or isopropyl alcohol) or with a mild
detergent.

4-6-3 Warning
(1) To ensure safety, when the internal temperature rises above the specified value, the
temperature switch turns off and electrical current supply to the lamp is terminated. The
unit detects that the temperature switch has shut off, and sounds an alarm.

4-6-4 Electrical shock


(1) Class 1 unit (3P power supply).

4-7 Others
4-7-1 Usage restrictions
Ambient temperature: 10 - 40 degree C
Relative humidity: 30 - 85%
Air pressure: 700 to 1060hPa
Usage environment: Do not use in a combustible atmosphere
Type of protection against electrical shocks: Class 1 Type BF applied part
Voltage: 100 - 120 V (North America market)
100 V (Japan market)
100 - 240 V(The other market)
Frequency: 50/60 Hz
Power fluctuation: Within +/- 10%
Frequency fluctuation: Within +/- 1 Hz
Power consumption: 500VA

4-7-2 Estimated Replacement Components


User servicing: Examination lamp
OLYMPUS servicing: Emergency lamp replacement, special filter attachment

4-7-3 Others
Weight Approximately 14.9 kg
Dimensions 383 (W) × 536 (D) × 162 (D) mm (maximum)
370 (W) × 475 (D) × 150 (D) mm
Power cord Cord set with a 3-prong hospital grade
With plug (100V CLASS market)
Plug-less (200V CLASS market)

ISSUE3 2-5 Product Outline


CLV-S190

4-7-4 Applicable regulations


Region Regulations
EU/EFTA MDD
Class: Class I (rule:12)
CE marking
Japan Pharmaceutical Law
FDC laws

USA
Canada Canadian Medical Device and Equipment Regulation
Class: I
Singapore Singaporean medical device regulations

4-7-5 Applicable external regulations


Region External Standard
Common to IEC 60601-1 (Safety of Medical Electronic Equipment); 1988+A1 , A2
all countries IEC 60601-1-1 (Medical Electronic Equipment); 2000
IEC 60601-1-2 (EMC); 2007
IEC 60601-2-18 (Safety of Endoscopic Apparatus); 1996+A1
ISO 14971 (Risk Management); 2007
ISO 7000(); 2004
IEC 60417-1: 2002, -2: 1998 (Graphical symbols) +A1, A2
IEC 60601-1-6(Usabillity); 2006
IEC 62366(Usabillity); 2007
IEC 60601-1-4(Programmable electrical medical systems);1999+A1
IEC 62304(Medical device software);2006
IEC 60601-1-9(Requirements for enviromentally conscious design);2007
EU/EFTA EN 980 (Graphical symbols for use on equipment): 2008
EN 1041(Information supplied by the manufacturer with medical devices);
1998
EN 60601-1(Safety of Medical Electronic Equipment); 1990+A1 , A2 , A3
EN 60601-1-1(Medical Electronic Equipment); 2001
EN 60601-1-2(EMC); 2007
EN 60601-2-18(Safety of Endoscopic Apparatus); 1996+A1
EN 60601-1-6(Usability) ;2007
EN 62366(Usability) ;2008
Japan JIS T0601-1 (Safety of Medical Electronic Equipment): 1999
JIS T0601-1-1(Medical Electronic Equipment): 2005
JIS T0601-1-2(EMC): 2002
JIS T0601-2-18(Safety of Endoscopic Apparatus); 2005
USA UL 60601-1: 2003
Canada CAN/CSA-C22.2 No. 601.1-M90: 1990

Product Outline 2-6 ISSUE3


CLV-S190

5-1 Front panel


Only Europe specification

Other specification

(1) Power switch


Press to turn the light source ON or OFF.
(2) Power indicator
Lights up when the light source is ON.
(3) Output socket
The endoscope or the light guide cable is connected to this socket.
This socket provides light to the endoscope.
(4) Control panel
Refer to the next page.

ISSUE3 2-7 Product Outline


CLV-S190

5-2 Control panel(buttons)

Only Europe specification

(5) Filter button


Press to switch the observation alternately between normal light observation and
optical-digital observation.
(6) Filter indicator
The indicator lights up in the optical-digital observation
(7) Filter mode selection indicators.
One of these indicators lights up to indicate the selected observation mode. “NBI” and
“PDD” indicate the NBI and PDD observations respectively.
(8) Intensity mode button
Press to switch between the high intensity mode and normal intensity mode.
(9) High intensity mode indicator
This indicator lights up when high intensity mode is selected.
(10) Lamp button
Press to turn ON the xenon lamp and long press while the lamp is on to turn it OFF.
(11) Lamp ON/OFF indicator
These indicators display whether the examination lamp (xenon lamp) lights up (“ON”) or not
(“OFF”).
(12) Emergency lamp indicator
This indicator lights up when the emergency lamp (halogen) is in use and blinks when the
emergency lamp (halogen) is broken, disconnected or not mounted.
(13) Brightness level buttons
These buttons are pressed to adjust the brightness level.

Product Outline 2-8 ISSUE3


CLV-S190

(14) Brightness level indicator


These indicators display the current brightness level.
(15) Counter reset button
After replacing the xenon lamp, the lamp hour indicator is reset by pressing this button for
at least 3seconds.
(16) Lamp hour indicator
This indicator displays the total working hours of the examination (xenon) lamp.
(17) Brightness mode button
This button is pressed to select automatic or manual brightness control.
(18) Brightness mode indicator
These indicators display the brightness adjustment setting (“AUTO” or “MANU”).

ISSUE3 2-9 Product Outline


CLV-S190

Other specification

(4) Filter button


Press to switch the observation alternately between normal light observation and
optical-digital observation.
(5) Filter indicator
The indicator lights up in the optical-digital observation
(6) Filter mode selection indicators.
One of these indicators lights up to indicate which observation mode is selected. “NBI”and
“Op.” indicate the NBI and for expansion observations respectively.
(7) Intensity mode button
Press to switch between the high intensity mode and normal intensity mode.
(8) High intensity mode indicator
This indicator lights up when high intensity mode is selected.
(9) Lamp button
Press to turn ON the xenon lamp and long press while the lamp is on to turn it OFF.
(10) Lamp ON/OFF indicator
These indicators display whether the examination lamp (xenon lamp) lights up (“ON”) or not
(“OFF”).
(11) Emergency lamp indicator
This indicator lights up when the emergency lamp (halogen) is in use and blinks when the
emergency lamp (halogen) is broken, disconnected or not mounted.
(12) Brightness level buttons
These buttons are pressed to adjust the brightness level.
(13) Brightness level indicator
These indicators display the current brightness level.
(14) Counter reset button
After replacing the xenon lamp, the lamp hour indicator is reset by pressing this button for
at least 3seconds.

Product Outline 2-10 ISSUE3


CLV-S190

(15) Lamp hour indicator


This indicator displays the total working hours of the examination (xenon) lamp.
(16) Brightness mode button
This button is pressed to select automatic or manual brightness control.
(17) Brightness mode indicator
These indicators display the brightness adjustment setting (“AUTO” or “MANU”).

ISSUE3 2-11 Product Outline


CLV-S190

5-3 Rear and side panels

Rear-panel

Side-panel

(1) System terminal


It is not used.
(2) OTV terminal
This terminal is the receptacle for the light source cable ( MAJ-1411/MAJ-1959) for use the
videosystem center.
(3) Foot switch terminal
This terminal is the receptacle for the foot switch (MAJ-1391).
(4) Potential equalization terminal
This terminal is connected to a potential equalization terminal of the other equipment
connected to this instrument. The electrical potential of the equipment are made equal.
(5) Ventilation grills
The air in the light source is exhausted from the grills for cooling.
(6) AC power inlet
Connect the provided power cord to supply the AC power via this inlet.

Product Outline 2-12 ISSUE3


CLV-S190

(7) Lamp cover


This cover has to be removed to replace the examination lamp.
(8) Lamp chamber
The lamp is inserted to this chamber.

6 Connector
6-1 System terminal
View from rear panel
13 · · · · · · · · · · · · · · · · 1

25 · · · · · · · · · · · · · · 14

1 GND 10 N.C. 19 N.C.


2 TXD_EN 11 N.C. 20 N.C.
3 RXD_EN 12 N.C. 21 N.C.
4 RTS_EN 13 N.C. 22 N.C.
5 CTS_EN 14 N.C. 23 N.C.
6 DSR_EN 15 N.C. 24 N.C.
7 GND 16 N.C. 25 N.C.
8 DTR_EN 17 N.C.
9 N.C. 18 N.C.

6-2 OTV terminal

View from rear panel


25 · · · · · · · · · · · · · · · · · · · · · · · · · · · 1

50 · · · · · · · · · · · · · · · · · · · · · · · · · · · 26

1 GND 18 GND 35 DSR_RE


2 N.C. 19 N.C. 36 N.C.
3 N.C. 20 GND 37 TCK
4 GND 21 EE_DAT 38 GND
5 N.C. 22 EE_CLK 39 TDO
6 N.C. 23 EE_STB 40 V CE(+3.3V)
7 N.C. 24 GND 41 TMS
8 N.C. 25 CLVSEL_G 42 TDI
9 N.C. 26 DSR(CV_ON) 43 GND
10 GND 27 N.C. 44 N.C.
11 TXD 28 GND 45 DTR
12 RXD 29 TXD_RE 46 POT_DAT
13 RTS 30 RTS_RE 47 POT_CLK
14 CTS 31 GND 48 POT_STB
15 N.C. 32 RXD_RE 49 N.C.
16 +5V 33 CTS_RE 50 N.C.
17 N.C. 34 DTR_RE

ISSUE3 2-13 Product Outline


CLV-S190

6-3 Foot switch terminal

View from rear panel

A NC
B FOOT_GND
C FOOT_SW0
D FOOT_SW1
E FOOT_SEN
F NC

Product Outline 2-14 ISSUE3


CLV-S190

Chapter 3: Specifications
Refer to the “Specifications” of Instruction Manual.

ISSUE3 3-1 Specifications


CLV-S190

Chapter 4: Installation and Connection


Refer to the “Installation and Connection” of Instruction Manual.

ISSUE3 4-1 Installation and Connection


CLV-S190

Chapter 5: Care, Storage, and Disposal


Refer to the “Care, Storage, and Disposal” of Instruction Manual.

ISSUE3 5-1 Care, Storage, and Disposal


CLV-S190

Chapter 6: Inspection
Refer to the “Inspection” of Instruction Manual.

1. Inspection Check Sheet


Please copy and use check sheet on next page.

A check item may change with the destination and specifications.


There may be a check item which is on this list and is not in the instruction manual.
Please refer to the instruction manual (Chapter of Inspection or Inspection Before Use) for an
actual check item and the check procedure.
There may be a blank column in a Check Sheet.
Please utilize this blank column, if needed.

ISSUE3 6-1 Inspection


CLV-S190

Inspection 6-2 ISSUE3


CLV-S190

Chapter 7: Safety Check

1. Electrical safety tests


1-1 Standard
Perform test of leakage current on following items. Values shall not exceed the following:
At single failure condition
Leakage current At normal condition (NC)
(SFC)
Patient DC 10 50
leakage
AC 100 500
current
Exterior leakage
100 300/500*
current
Ground leakage
500 1000
current
300: UL region, 500:Other region. (μA)

1-2 Chassis leakage current test


(1) Use power code to connect CLV-S190 with leak current controller. Set Xenon Lamp to
automatic lighting mode.
(2) Connect alligator clip from leak current controller to the outer case screw farthest from inlet.
(3) Perform leak current test using automatic leak current meter.
Refer to manual of automatic leak current meter for measuring method.

Connect to leakage current


controller

CLV-S190

Connect alligator clip

ISSUE3 7-1 Inspection


CLV-S190

1-3 Patient leakage current test


(1) Use power code to connect CLV-S190 with leak current controller. Set Xenon Lamp to
automatic lighting mode.
(2) Attach GI Scope Jig and connect alligator clip from leak current controller.
GI Scope jig
(3) Perform leak current test using automatic leak current meter.
Refer to manual of automatic leak current meter for measuring method.

Connect to leakage current controller

CLV-S190

Connect the alligator clip

1-4 Ground leakage current test


(1) Measure leak current at ground terminal of inlet.
(2) Perform leak current test using automatic leak current meter.
Refer to manual of automatic leak current meter for measuring method.

Safety check 7-2 ISSUE3


CLV-S190

1-5 Procedures for ground resistance test


(1) Connect one side of clip to ground terminal of product inlet, and the other side to outer
case screw of cover (as far as possible from position of inlet).
(2) Measure ground resistance using automatic ground resistance meter. Value shall be 0.1
ohms or less at 25A.
Refer to manual of automatic ground resistance meter for measuring method.

Screw

Inlet of earth pin

ISSUE3 7-3 Safety check


CLV-S190

2. Safety check card


Date of inspection
Product CLV-S190 Inspected by
Serial number Checked by

Electrical safety tests

Inspection item Standard Measurement Value Result


Equipment No.
Exterior leakage current NC 100μA or less μA
test SFC 300μA or less (UL region) μA
500μA or less (Other regio
n)
Patient leakage NC 10μA or less μA
DC
current test SFC 50μA or less μA
NC 100μA or less μA
AC
SFC 500μA or less μA
Ground leakage current NC 500μA or less μA
test SFC 1000μA or less μA
Ground resistance test 25A, 0.1ohoms or less

Safety check 7-4 ISSUE3


CLV-S190

Chapter 8: Block Diagram

Halogen
Xenon lamp
lamp

Halogen lamp Power circuit for lamp

Power Wall mains outlet


Control panel System control circuit circuit

Brightness adjustment/
driving circuit

Endoscope / Peripheral equipment

ISSUE3 8-1 Block Diagram


CLV-S190

Chapter 9: Troubleshooting
Refer to the “Troubleshooting” of Instruction Manual.

ISSUE3 9-1 Troubleshooting


CLV-S190

Chapter 10: Parts Replacement


Refer to the “Lamp Replacement” of Instruction Manual

ISSUE3 10-1 Lamp Replacement


OLYMPUS CORPORATION
2951 Ishikawa-machi, Hachioji-shi, Tokyo Japan

Medical Service Engineering Department


Medical Repair Division

PRINTED IN JAPAN FIRST EDITION : 10.2011