Download as pdf or txt
Download as pdf or txt
You are on page 1of 60

Operating Instructions

ATMOS E 341 Battery


English

GA1GB.320102.0

2021-11 Index: 13
Table of contents
1.0 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Information on operating instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Explanation of pictures and symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Intended use and side effects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.5 Intended operators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.6 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.7 Transport and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.0 Hints for your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 General safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Danger for users, patients and third parties. . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Damage to the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.0 Setting up and starting up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1 Device overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.1 Front and rear view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.2 Control panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2 Preparing the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3 Charging the battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4 Connection and removal of canister system and hoses. . . . . . . . . . . . . . . . . . 18
3.4.1 DDS canister system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4.2 Serres® canister system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4.3 Medi-Vac® canister system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5 Support for canister system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.5.1 DDS canister system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.5.2 Serres® canister system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.5.3 Medi-Vac® canister system / Universal bracket. . . . . . . . . . . . . . . . . . . . . . . 28
3.6 Hose rewind. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.7 Device base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.0 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.1 Switching on the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2 Switching off the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.3 Vacuum adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.4 Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.0 Cleaning and disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.1 Preparing for cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3 After cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.4 Recommended disinfectants. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.4.1 Instrument disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.4.2 Surface disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.5 Hygienic plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.6 Oversuction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

2
6.0 Maintenance and service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.1 Periodic tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2 Function check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2.1 Manual function check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2.2 Automatic function check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3 Sending in the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.4 Handling of batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5 Battery exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.6 Exchanging the release button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.0 Eliminating errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.0 Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.1 Wall and device support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.2 Retrofit kit canister system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.0 Spare parts and consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.0 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
11.0 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
12.0 Notes on EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Further information, accessories, consumables


and spare parts are available from:

ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Germany
Tel. + 49 7653 689-0
+ 49 7653 689-222 (Service Centre)
Fax: + 49 7653 689-190
+ 49 7653 689-292 (Service Centre)
[email protected]
www.atmosmed.de

3
1.0 Introduction
1.1 Information on operating instructions

These operating instructions contain important notes on how to


operate the ATMOS E 341 Battery safely, correctly and effectively.
The instructions are intended for the training and teaching of
operating personnel and are intended as a reference. Reproduction,
even partial, is only permitted with written permission from ATMOS.
These operating instructions must always be kept available near
the device.

Care, periodic tests, regular cleaning and proper application are indis-
pensable. They guarantee the operational safety and usability of the
ATMOS E 341 Battery.
Maintenance, repairs and periodic tests may only be carried out
by persons who have the appropriate technical knowledge and are
familiar with the product. To carry out these measures, the person
must have the necessary test devices and original spare parts.

Prior to first start-up, please peruse chapter “2.0 Hints for your safety”

§ on page 11. This helps you avoid potentially dangerous situations.


The product ATMOS E 341 Battery bears CE marking CE 0124 in
accordance with EC Council Directive 93/42/EEC concerning medical
devices and meets the basic requirements of Annex I to this directive.
The product ATMOS E 341 Battery complies with all applicable require-
ments of the Directive 2011/65/EU restricting the use of certain
hazardous substances in electrical and electronic equipment (“RoHS”).
The Declaration of Conformity and our general standard terms and
conditions can be obtained on our website at www.atmosmed.com.
The quality management system at ATMOS has been certified
according to international standard EN ISO 13485.

These operating instructions are valid for the following devices:


ATMOS E 341 Battery REF 319.0000.0
ATMOS E 341 Battery / DDS REF 319.1000.0
ATMOS E 341 Battery / Serres® REF 319.1100.0
ATMOS E 341 Battery / Medi-Vac ®
REF 319.1200.0
ATMOS E 341 Battery / Universal bracket REF 319.1300.0
Some figures show the ATMOS C 341 Battery. However, the devices do not differ in their
functionality described.

4 Introduction
1.2 Explanation of pictures and symbols

In the operating instructions

DANGER
Warning of a danger which causes immediate death or serious injury. Observe the
necessary measures.

WARNING
Beware of a danger which can cause death or serious injury. Observe the necessary
measures.

CAUTION
Beware of a danger which can easily hurt you. Observe the necessary measures.

NOTICE
Indication of a danger where the product or other items can be damaged. Observe
the necessary measures.

Warning of a danger which can cause death or serious injury.


Information regarding possible material damage which can be caused.

 Useful information on the handling of the device.


1. Action. Go step by step.
• Numeration.
» Result of an action.
Move, plug in this direction.

click Engage, check correct fit.

On device and type plate

Follow operating instructions


Consult operating instructions
(blue)

Manufacturer Date of manufacture

SN Serial number IP34D Degree of protection

REF Reference number KB Short-term operation

Professional disposal Type BF applied part

Connection suction hose/patient


Protection class II PATIENT
(Serres® canister system)
This device complies with the
For single use only (symbol is on
relevant requirements of EU
the consumables)
directives

Introduction 5
On the recharging accessories
This device complies with the
Consult operating instruc-
relevant requirements of EU regula-
tions
tions

Date of manufacture Professional disposal

Manufacturer REF Reference number

SN Serial number IP40 Degree of protection


Output voltage
Type CF applied part
(13.8 V / 3.5 A)
Input voltage
(100 – 240 V / 50 – 60 Hz / For indoor use only
1.1 A)

Alternating current Direct current

Protection class II device

On the battery
This device complies with the
Consult operating instruc-
relevant requirements of EU regula-
tions
tions

Date of manufacture Professional disposal

Manufacturer REF Reference number

SN Serial number EAN European Article Number


Installation position: on
Warning, pay special attention
top

Do not throw into fire European Recycling Platform

1.3 Intended use and side effects

Intended use
Name: ATMOS E 341 Battery
Main functions: Temporary and spontaneous suction of secretion, blood and body
fluids and also liquid, viscous and solid pieces of food in the medical field.
To evacuate vacuum mattresses and inflatable splints.
Medical indications / application: Suction of the upper and lower respiratory tract.
Specification of the main function: Drainage and temporary collection of body
fluids. By means of an electrical suction pump, a negative pressure will be created. The
integrated secretion canister allows temporary collection of the derived body fluids.
Application organ: Upper respiratory tract (nose, nasal cavity, throat), lower respiratory
tract (larynx, trachea, bronchial system)

6 Introduction
Application time: Temporary use on the patient (< 60 min)
Application site: The application site is the hospital, the practice, the accident &
emergency department, the nursing and home care field, as well as for outdoor appli-
cation and during transport. The device may only be applied by medically trained and
instructed staff.
Contraindications: Not suitable for
• The continuous operation by drainages in the low vacuum range (e.g. thorax
drainages or wound drainages).
• Permanent endoscopic use.
• Suction in medical rooms where a potential equalisation is necessary (e.g. heart
surgery).
• Use outside the medical field.
• Suction of flammable, corrosive or explosive substances.
• Suction in explosion-hazardous areas.
The product is: active
Sterility: Non-sterile product
Single-use product / reprocessing: The device and part of the accessories are
reusable. For information on reprocessing, cleaning and disinfection please see the
operating instructions.

Possible side effects during suction


• Bleeding in the nasal pharyngeal area
• Injury to the vocal cords
• Tracheal injury
• Hypoxaemia
• Cardiovascular instability
• Bradycardia, arrhythmia and asystole (caused by vagus stimulation)
• Tachycardia (caused by stress)
• Choking, nausea, vomiting and coughing
• Nosocomial infection of the respiratory tract
• Seizures by patients who tend to develop cramps
Attention must be paid to these operating instructions in order to keep the side effects
as minimal as possible.

Introduction 7
1.4 Function
The ATMOS E 341 Battery is a mobile, portable, mains-operated medical device for the
temporary application on adults, children and babies. The device is operated with an
electronically controlled, maintenance-free diaphragm pump.
The pump can be optionally operated by rechargeable battery or via an external DC
voltage source (12 V).
During operation, the pump generates a vacuum within the hose system and secretion
canister, thus suctioning off secretion, blood and body fluids as well as liquid, viscous
and solid pieces of food. The fluid is gathered in the secretion canister.
The predefined vacuum values enable a quick and precise adjustment of the vacuum in
different situations. It can be selected between four different vacuum values (−0.1 bar;
−0.2 bar; −0.5 bar and −0.8 bar). The control panel is illuminated so that you can read
the operating status even in the dark.
An overtemperature stop prevents overheating of the batteries.
DDS secretion canister:
The DDS secretion canister is affixed laterally to the device and is plugged via Direct
Docking onto the suction connection of the support for the DDS canister system.
Therefore, there is no intermediate hose. Now the user can/must only connect the
suction hose. A hydrophobic DDS bacterial and viral filter located in the canister lid
prevents bacteria, viruses and liquids from entering the device.
A mechanical oversuction stop (float ball) is integrated in the canister lid. This prevents
an accidental absorption of secretion into the pump head. The float ball rises to the top
of the secretion until it blocks the outlet.
Disposable secretion canister:
The disposable secretion canister is comprised of an external canister, disposable
suction bag, vacuum hose and the disposable suction hose.
The disposable secretion canister is affixed laterally to the device. The vacuum hose of
the canister is connected to the suction connection of the device. The secretion is trans-
ported to the disposable suction bag via the suction hose. The disposable suction bag
is a single-use product. As soon as the disposable suction bag is full, it is removed from
the external canister and disposed of. The disposable suction bag and the disposable
suction hose must not be reused.
A bacterial filter is integrated in the disposable suction bag. This prevents secretion,
liquid and bacteria from seeping into the device.

1.5 Intended operators


The ATMOS E 341 Battery may only be used by persons who were medically trained,
and were trained in medical suction. Prior to application, the user must be familiar with
the device. Please note the country-specific requirements and regulations.
ATMOS recommends: Instruction on the operation of the device must be performed by
an authorised person.

8 Introduction
1.6 Scope of delivery
1. Please compare the contents on completeness immediately upon receipt (see
delivery note).

Basic device

ATMOS E 341 Hose rewind Power supply 2-pin mains


Battery with (mounted) and recharging connection ca-
device base unit 318.0035.0 ble 008.0920.0

DDS canister system (mounted)

Support for Reusable Secretion 10 x hydro- 10 x Fingertip


DDS canister suction hose, Ø canister 1 l with phobic DDS
system 10 mm, L 1.3 m canister lid, bacterial and
filter holder, viral filter
sealing ring

Serres® canister system (mounted)

Support for Vacuum hose Serres® external 10 x disposable


Serres® canister with angled canister 1 l suction hose
system connection with fingertip,
Ø 6 mm

Medi-Vac® canister system (not mounted)

Support for Vacuum hose Medi-Vac® 10 x disposable


Medi-Vac® with angled external suction hose
canister system connection canister 1 l with fingertip,
Ø 6 mm

Introduction 9
Universal bracket (not mounted)

Support for Vacuum hose


canister system with angled
connection
& A hydrophobic DDS bacterial and viral filter is not included in the scope of delivery
and must be ordered separately for use with a canister system without an integrat-
ed bacterial filter.

Not included in the scope of delivery:


• Suction catheter
• Adapter for vacuum mattress
• Serres® suction bag 1 l
• Medi-Vac® suction bag 1 l
• Wall and device support

1.7 Transport and storage


Only transport the device in a shipping container which is padded and offers sufficient
protection.
If damage occurs during transport:
1. Document and report the transport damage.
2. Fill in form QD 434 “Customer complaint/return shipment”. This form is enclosed
with each delivery and can be found at www.atmosmed.com.
3. Send in the device to ATMOS (chapter “6.3 Sending in the device” on page 44).

Environmental conditions for transport and storage:


• Temperature: −40...+70 °C
• Relative humidity: 5...95% without condensation
• Air pressure: 540...1100 hPa

10 Introduction
2.0 Hints for your safety
The safety of the ATMOS E 341 Battery complies with all the recognised rules of
technology and the Directives of the Medical Devices Act.
Read and follow the safety instructions carefully before using the product.

2.1 General safety information


Make yourself familiar with the device at an early stage, so you can use it even in hectic
situations.
Never operate the unit if it shows any obvious safety defects. Check the unit at regular
intervals for safety and function.

2.2 Danger for users, patients and third parties


Take care that the device is always functional and ready for use.
Your patient may suffocate.
• Ensure that the device is always ready for use in an emergency.
• Position the unit in an easily accessible location and keep access free.
• Make sure that the charging accessories are functional. Replace defective charging
accessories immediately.
• Recharge the battery at the latest after 6 months, even if you do not use the device.
• Perform a function check after each use. Perform a function check every 4 weeks in
case you do not use the device for a longer period.
• ATMOS recommends always having another suction device ready to hand in case of
any device failure. That way, you can suction even in the event of equipment failure.
• Please observe the notes on electromagnetic compatibility (EMC) of the device.
Avoid misapplication.
Your patient may be seriously injured.
• The ATMOS E 341 Battery may only be used by persons who were medically trained,
and were trained in medical suction.
• Please select the vacuum according to the patient and the application.
• Observe the valid guidelines.
Reduce the risk of infection for you and your patients!
Deadly diseases can be transmitted.
• Always wear disposable gloves if you could come into contact with secretion.
• Never use components marked with 2 more than once. These components are
intended for single use only.
• Only use sterile packaged parts when the packaging is undamaged.
• Never operate the device without a bacterial and viral filter.

Hints for your safety 11


Protect yourself against an electric shock.
Burns, cardiac arrhythmias and even death are possible.
• Do not operate the device if it has been dropped. In this case please clean the
device and send it in to ATMOS for repair.
• Disconnect the device from the mains power supply prior to cleaning or disin-
fection.
• Prior to each use, please check whether the device or the recharging accessories are
damaged. Never operate the device if you detect any failure. In this case please
clean the device and send it in to ATMOS for repair.
• Take care that no liquid penetrates the device. In case that liquid has penetrated the
device it may no longer be operated. In this case please clean the device and send it
in to ATMOS for repair.
• The ATMOS E 341 Battery cannot be sterilised.
• Use the recharging accessories in dry surroundings. The surroundings must be
non-conductive.
• Only use the recharging accessories according to the operating instructions.
• Only use original accessories and original spare parts from ATMOS. This specifically
applies to the recharging accessories and the battery.
• Please pay attention to the periodic tests in chapter “6.0 Maintenance and service”
on page 42.
• Assembly, new settings, alterations, extensions and repairs may only be carried out
by authorised persons.
• Do not modify the device without permission of the manufacturer.
Explosion and fire hazard!
Burns and injuries are possible.
• Never suction any explosive, flammable or corrosive gases or liquids. Please
observe the intended use in chapter “1.3 Intended use and side effects” on page 6.
• Never operate the device in explosion-hazardous areas or areas which are
oxygenated.
• Only use original accessories and original spare parts from ATMOS. This specifically
applies to the recharging accessories and the battery.
Danger of suffocation for children through accessories!
Children can strangle themselves or choke on small parts.
• Keep children away from hoses and connection cables.
• Keep children away from swallowable small parts. Small parts are, e.g. fingertip and
sealing ring.
Tripping hazard due to cables.
Injuries and fractures are possible.
• Lay connecting cables properly.
Only a fully functional product will meet the safety requirements of users, patients and
third parties. Please therefore read the following instructions carefully.

12 Hints for your safety


2.3 Damage to the device
Please observe the ambient conditions regarding transport, storage, operation and
recharging of the battery.
Take care that no liquid penetrates the device. In case liquid has penetrated the device,
it may no longer be operated. In this case, please clean the device and send it in to
ATMOS for repair.
Always place the device on a firm, level surface. The device must always be in a vertical
position when you use it. Otherwise, secretions may enter the unit.
Only use proper power connections and extension cords.
If possible, avoid transport at temperatures below −5 °C. After transport at tempera-
tures below −5 °C: The device must be acclimatised for up to 6 hours at room temper-
ature before you continue with the next steps.
The device may only be connected to the mains power supply when mains voltage and
frequency of device and mains power supply correspond.

Hints for your safety 13


3.0 Setting up and starting up
Please observe that insufficient battery charge can result in damage to the battery.
1. The battery must be fully charged prior to first use.

3.1 Device overview

3.1.1 Front and rear view

With DDS canister system


1
1 Control panel
2 Battery compartment cover
3 Device base

4 Release button wall and device


support
5 Hose rewind with suction hose
4

5
6 Connection charging acces-
sories
7 Guide for the wall and device
support

8 9 10 11 12 8 Filter holder
9 Sealing ring
10 Hydrophobic DDS bacterial
13 and viral filter
14 11 Inner canister lid
15
12 External canister lid
16 13 Canister lid lug
17
14 Float ball
15 Secretion canister with scale
16 Connection suction hose
17 Support for DDS canister
system

14 Setting up and starting up


With Serres® canister system

8
8 Angle (connection disposable
suction hose)
9 Serres® suction bag
10 Serres® external canister
9 11 Support for Serres® canister
system
13
12 Grey angle on the Serres®
12
external canister (connection
10 vacuum hose)
11 13 Vacuum hose with angled
connection

With Medi-Vac® canister system


8 Angle (connection disposable
8 suction hose)
9 Red hose
9 10 Medi-Vac® suction bag
10
14 11 Connection vacuum hose
12 Medi-Vac® external canister
11
13 Support for Medi-Vac® canister
system
12
14 Vacuum hose with angled
connection
13

With universal bracket


13 Support for canister system
14
14 Vacuum hose with angled
connection

The universal bracket is suitable


for a secretion canister with a
13 diameter of 11.5 – 12.5 cm.
Do not operate the device without
a bacterial filter.

Setting up and starting up 15


3.1.2 Control panel
1

5 4

1 LEDs for the display of the current vacuum


2 Display of battery status
3 Button battery status
4 Button to select the desired vacuum
5 On/off button

Display of the battery status


The following display values are not valid during battery charging.

Green
and red
Flashes LED flash

> 85% 60 – 85 % 35 – 60 % < 15 – < 10 – 15 % < 10 %


35 %
Prior to the battery going dead, a signal tone sounds every 5 seconds.
An error is present if all the green LEDs flash simultaneously or all LEDs are flashing.
Please observe chapter “7.0 Eliminating errors” on page 47.

3.2 Preparing the device


Prior to first operation, peruse the safety notes in chapter “2.0 Hints for your safety” on
page 11.
Damaged pump diaphragms due to cold temperatures during transport.
1. After transport at temperatures below −5 °C: The device must be acclimatised for up
to 6 hours before you continue with the next steps.
2. Check the device for any damage in transport.
3. If the device is damaged: Document and report the transport damage. Send in the
device to ATMOS (chapter “6.3 Sending in the device” on page 44).
4. If the device is not damaged: Place it on a safe and even surface.
5. Check the charging accessories for any damage.
6. Replace defective charging accessories immediately.
7. The battery must be fully charged: Chapter “3.3 Charging the battery” on page 17.
8. Remove the canister system from the support.

16 Setting up and starting up


9. With the DDS canister system: Prior to first use, clean the canister system and insert
a DDS bacterial and viral filter (chapter “5.0 Cleaning and disinfection” on page 35
as well as chapter “3.4 Connection and removal of canister system and hoses” on
page 18).
10. Connect the suction hose.

11. Place the canister system upright from above into the support: Chapter “3.4
Connection and removal of canister system and hoses” on page 18
12. Wrap the suction hose onto the hose rewind.
13. If you wish to use the device for vacuum mattresses: Check whether a suitable
adapter is available for the vacuum mattresses.

3.3 Charging the battery


& The battery status can be checked by briefly pressing the button for the battery
status.
The battery must be fully charged prior to first use.
Damage to the battery due to deep discharge.
1. Charge the battery at the latest when the bottom green LED of the battery status
display flashes.
2. Only use the enclosed power supply and recharging unit 318.0035.0. Other charging
accessories must not be used.
3. Please observe the notes in chapter “6.4 Handling of batteries” on page 44.
During battery recharging, full suction capacity of the device is still available.
If the battery is fully discharged or defective, the device may be operated via the
charging accessories.
& If the ambient conditions are not adhered to, the charging time for the battery is
significantly increased. The charging process will be terminated if the temperature is
too high. Therefore, please protect the device from direct solar radiation and keep it
away from radiators.

Ambient conditions during charging


• Temperature: +0...+40 °C
• Relative humidity: 5...95 % without condensation
• Air pressure: 540...1100 hPa

Setting up and starting up 17


Charging with power supply and recharging unit
1. Connect the device plug from the power supply and recharging unit to the back of
the device 1 .

2. Connect the power cable to the power supply and recharging unit.
3. Plug the power plug of the power supply and recharging unit into the socket.
» The LEDs of the battery status display flash successively.
» One LED is continuously illuminated. This indicates the current battery status.
» The battery is fully recharged when the top red LED is continuously illuminated.

Recharging via the wall and device support


If you have attached the charging accessories to a wall and device support, then the
device will be charged automatically: Chapter “8.1 Wall and device support” on page 50.
1. Attach the device to the wall and device support.
» The LEDs of the battery status display flash successively.
» One LED is continuously illuminated. This indicates the current battery status.
» The battery is fully recharged when the top red LED is continuously illuminated.

3.4 Connection and removal of canister system and


hoses

3.4.1 DDS canister system


WARNING
Risk of infection from contaminated DDS bacterial and viral filter and canister lid.
Deadly diseases can be transmitted.
• Never operate the device without a DDS bacterial and viral filter. Always keep at
least one spare DDS bacterial and viral filter on hand.
• Wear disposable gloves when changing the DDS bacterial and viral filter.
• Prior to each use, check whether the DDS bacterial and viral filter is dry and clean.
Replace the DDS bacterial and viral filter with a new DDS bacterial and viral filter if it
is discoloured, contaminated or oversucked. The DDS bacterial and viral filter must
not be dried and reused.
• Replace the DDS bacterial and viral filter when changing the patient. ATMOS recom-
mends: Replace the DDS bacterial and viral filter after 14 days even if there is no
patient change.

18 Setting up and starting up


Removal
1. Remove the suction hose from the hose rewind and then from the hose guide.
2. Gently unlock the clip 1 of the canister lid from the support of the DDS canister
system and lift it upwards:

3. Lift the canister system upwards from the support.


4. Place the canister system on a safe and even surface.
5. Remove the suction hose from the secretion canister.
6. Turn the filter holder anti-clockwise by 90°.
& The filter holder is difficult to turn because it has to seal the canister lid tightly.

7. Remove the filter holder with DDS bacterial and viral filter from the canister lid.
8. If required: Remove the DDS bacterial and viral filter 7 and the sealing ring 6 from
the filter holder 5 .
5

Risk of infection due to overflowing secretion. Deadly diseases can be transmitted.


9. Hold the secretion canister with one hand and pull the canister lid upwards with
force.
» The canister system is open.
10. If required: Slide the inner canister lid out of the external canister lid.

11. If required: Remove the float ball 3 from the float ball compartment 2 of the inner
canister lid 1 .

1 2
3

Setting up and starting up 19


Connection
& When you pour 50–100 ml water or disinfectant into the secretion canister, then it is
easier to clean.
1 Inner canister lid

2 Float ball compartment


3 Float ball
1
4 External canister lid
2 3 4

1. Press the external canister lid 4 on the inner canister lid 1 until it clicks into place.
2. Open the float ball compartment 2 gently and insert the float ball 3 .
3. Gently press the float ball compartment together.
4. Check whether the float ball moves easily and does not fall out of the float ball
compartment.
5. Place the secretion canister on a firm surface.
6. Press the canister lid onto the secretion canister. The canister lid cannot be placed
in a wrong position.
7. Press the canister lid tightly with both hands as far as it will go onto the secretion
canister.
8. Place the sealing ring 6 onto the filter holder 5 .

6
5 6

9. Put a new DDS bacterial and viral filter 7 onto the filter holder 5 .
5

10. Insert the filter holder into the canister lid and turn it clockwise until it clicks into
place.

11. Connect the suction hose to the canister system.

20 Setting up and starting up


12. Place the canister system upright into the support and at the same time position
the suction hose in the hose guide.

click

13. Check that the clip 1 of the canister lid is correctly attached to the support of the
canister system.

» The vacuum connection from the pump to the canister system is established.
14. Check whether the hose has a kink. If applicable, remove the kink.
15. Insert the suction hose into the hose guide of the device base.

16. Wrap the suction hose onto the hose rewind.


17. If required: Connect a fingertip to the suction hose.
18. Perform a manual function check: Chapter “6.2.1 Manual function check” on page
42.

Setting up and starting up 21


3.4.2 Serres® canister system
WARNING
Risk of infection due to contaminated canister system and hoses.
Deadly diseases can be transmitted.
• Only use Serres® suction bags with integrated bacterial filter.
• Only use sterile packaged parts when the packaging is undamaged.
No vacuum or vacuum is too low because of incorrect connection.
Patient can suffocate.
• Please observe the operating instructions from the manufacturer of the Serres®
canister system.

Removal
1. Remove the disposable suction hose from the hose guide.
2. Remove the disposable suction hose and the angle 2 from the Serres® suction bag.

3. Close the connection “patient” on the Serres® suction bag with the green cap 3 .
4. Remove the vacuum hose from the Serres external canister (grey angle
® 1 ).
5. Remove the Serres canister system from the support.
®

6. If required: Remove the vacuum hose from the device.

Connection
1. Connect the vacuum hose to the device.

2. Place the Serres® external canister upright into the support. The scale must be
visible.
3. Insert the Serres® suction bag into the Serres® external canister.

22 Setting up and starting up


4. Connect the vacuum hose to the Serres® external canister (grey angle 1 ).

5. Check whether the foil of the Serres® suction bag is fully inserted into the Serres®
external canister and the lid tightly fits to the Serres® external canister.
6. Connect the disposable suction hose with the angle 2 to the Serres® suction bag.
7. Close the secondary air opening of the fingertip and close the front opening with
your thumb.
8. Switch on the device so that the pump builds up vacuum.
» The Serres® suction bag unfolds.
9. Insert the suction hose into the hose guide.

10. Wrap the suction hose onto the hose rewind.


11. Perform a manual function check: Chapter “6.2.1 Manual function check” on page
42.

Setting up and starting up 23


3.4.3 Medi-Vac® canister system
WARNING
Risk of infection due to contaminated canister system and hoses.
Deadly diseases can be transmitted.
• Only use Medi-Vac® suction bags with integrated bacterial filter.
• Only use sterile packaged parts when the packaging is undamaged.
No vacuum or vacuum is too low because of incorrect connection.
Patient can suffocate.
• Please observe the operating instructions from the manufacturer of the Medi-Vac®
canister system.

Removal
1. Remove the disposable suction hose from the hose guide.
2. Remove the disposable suction hose and the angle 2 from the Medi-Vac® suction
bag.

3
4

3. Close the connection “patient” on the Medi-Vac® suction bag with the blue cap 3 .
4. Remove the red hose 1 from the Medi-Vac suction bag.
®

5. Close the connection “vacuum” on the Medi-Vac® suction bag with the blue cap 4 .
6. Remove the vacuum hose from the red connection 5 of the Medi-Vac external
®

canister.
7. Remove the Medi-Vac® canister system from the support.
8. If required: Remove the vacuum hose from the device.

Connection
1. Connect the vacuum hose to the device.

2. Insert the Medi-Vac® suction bag into the Medi-Vac® external canister.
3. Connect the red hose 1 to the Medi-Vac® suction bag.
4. Place the Medi-Vac® external canister upright into the support.

24 Setting up and starting up


5. Connect the vacuum hose to the red connection 3 of the Medi-Vac® external
canister.
1

6. Check whether the lid tightly fits to the Medi-Vac® external canister.
7. Connect the disposable suction hose 2 to the Medi-Vac® suction bag.
8. Close the secondary air opening of the fingertip and close the front opening with
your thumb.
9. Switch on the device so that the pump builds up vacuum.
» The Medi-Vac® suction bag unfolds.
10. Insert the suction hose into the hose guide.

11. Wrap the suction hose onto the hose rewind.


12. Perform a function check: Chapter “6.2 Function check” on page 42.

Setting up and starting up 25


3.5 Support for canister system

3.5.1 DDS canister system

Removal
1. Remove the canister release.

2. Push the support for the DDS canister system backwards up to the middle and take
it out of the guides.

Mounting
1. Attach the support for the DDS canister system in the middle on the right side of
the device. The bars on the support must be fitted to the two guides on the device.

2. Slide the support for the DDS canister system forward until it is flush with the
device. The inlet to the pump must be visible.

26 Setting up and starting up


3. Attach the canister release.

& The canister release is at the same time the connection angle through which the
canister system is connected to the pump.

3.5.2 Serres® canister system

Removal
1. Remove the connection angle.
2. Push the support for the Serres® canister system backwards up to the middle and
take it out of the guides.

Mounting
1. Attach the support for the Serres® canister system in the middle on the right side of
the device. The bars on the support must be fitted to the two guides on the device.

2. Slide the support for the Serres® canister system forward until it is flush with the
device. The inlet to the pump must be visible.

Setting up and starting up 27


3. Connect the vacuum hose to the connection angle.

3.5.3 Medi-Vac® canister system / Universal bracket

Removal
1. Remove the connection angle.
2. Push the support of the canister system backwards up to the middle and take it out
of the guides.

Mounting
1. Attach the support of the canister system in the middle on the right side of the
device. The bars on the support must be fitted to the two guides on the device.

2. Slide the support of the canister system forward until it is flush with the device. The
inlet to the pump must be visible.

28 Setting up and starting up


3. Connect the vacuum hose with the connection angle.

& The universal bracket is suitable for secretion canisters with a diameter of
11.5 – 12.5 cm.
& Do not operate the device without a bacterial filter.

3.6 Hose rewind

Removal
Prerequisite: The hose is unwound.
1. Pull the wings outwards so that the hose rewind can be released.
2. Pull the hose rewind from the device.

Mounting
Prerequisite: Device base and battery compartment cover are attached.
1. Turn the hose rewind so that the opening points upwards.
2. Attach the hose rewind with force to the support on the left side of the device until
it clicks into place.

click

Setting up and starting up 29


3.7 Device base

Removal
Prerequisite:
The following parts are removed:
• Canister system
• Support for canister system
• Hose rewind
• Battery compartment cover
1. Put the device carefully on the front.
2. Remove the device base in the following order A - B - C - D:

A
D

B
C

Mounting
NOTICE
Incorrectly mounted device base.
Device moves during the operation.
• Attach the device base with particular care, according to the operating instructions.

Prerequisite:
The following parts are removed:
• Canister system
• Support for canister system
• Hose rewind
• Battery compartment cover
1. Put the device carefully on the front.
2. Take care that the indentations at the device base are fitted to the protruding edges
on the device.
3. Attach the device base in the following order D - C - B – A. The arrows show the
points at which the device and the device base have to engage with each other.

A
D

B
C

4. Firmly press on all sides again.

30 Setting up and starting up


5. Afterwards, the following parts can be mounted:
- Battery compartment cover (chapter “6.5 Battery exchange” on page 45)
- Hose rewind (chapter “3.6 Hose rewind” on page 29)
- Support for canister system (chapter “3.5 Support for canister system” on page
26)
- Canister system (chapter “3.4 Connection and removal of canister system and
hoses” on page 18).

Setting up and starting up 31


4.0 Operation
WARNING
Risk of infection due to lack of hygiene or damaged components.
Deadly diseases can be transmitted.
• Please use new consumables and new disposable canister systems or reprocessed
DDS canister systems for each patient.
• Prior to each use, please check whether hoses or canister systems are damaged.
Replace any damaged parts.
Electric shock from damaged equipment.
Cardiac arrhythmias may be caused.
• Prior to each use, please check whether the device and the recharging accessories
are damaged.
• Replace any damaged parts immediately.
• Do not use the device if it is damaged.

Ambient conditions during operation


• Temperature: −5...+50 °C
• Relative air humidity: 5...95% without condensation
• Air pressure: 540...1100 hPa

4.1 Switching on the device


& The device should only be left on as long as you need it. This way you can increase
the battery life.
1. Push the on/off button to switch on the device.
» The pump starts. The vacuum is set which was selected prior to switch off.
» All LEDs on the control panel light up for about 1 second.
» The on/off button is illuminated as long as the device is switched on.

4.2 Switching off the device


1. Switch off the device by pressing the on/off button for at least 1 second.

4.3 Vacuum adjustment


WARNING
Vacuum is too high.
Patient may be seriously injured.
• Observe the valid guidelines.
• Please select the vacuum according to the patient and the application.

1. Push the on/off button to switch on the device.


» The pump starts. The vacuum is set which was selected prior to switch off.
2. Push the button for the required vacuum.
» The green LED above the selected button flashes.

32 Operation
4.4 Suction
WARNING
Device failure if the period of continuous operation is too long.
Patient can suffocate.
• Make sure not to use the device in continuous operation for more than 60 minutes.
Otherwise, the pump shuts off automatically. In this case, let the device cool down
for about 2 hours.
• Check the status of the battery regularly while you operate the device.
Risk of infection.
Deadly diseases can be transmitted.
• Always wear disposable gloves during suction.
CAUTION
Risk of injury due to inappropriate material or untrained users.
Injuries in the oral cavity and pharynx of the patient.
• Suction may only be carried out by persons who were medically trained and were
trained in medical suction.
• Use a suction catheter for tracheal or nasopharyngeal suction.
• If you suction viscous food ingredients in the oral cavity, use the suction hose
without the suction catheter.

Connecting the suction catheter


1. Remove the suction hose from the hose rewind.
If you suction viscous food ingredients in the oral cavity:
2. Use the suction hose without suction catheter.
For tracheal or nasopharyngeal suction:
1. Please choose a suction catheter in the appropriate size.
2. Connect the suction hose  and the suction catheter  using a fingertip :
   Suction hose
 Secondary air opening
 Suction catheter
 
 Fingertip

Connecting the special suction instruments


1. Please observe the operating instructions for the suction instruments.

Operation 33
Suction
CAUTION
Adherence due to careless suction.
Injuries in the oral cavity and pharynx of the patient.
• Briefly open the secondary air opening  if the suction catheter adheres to the
tissue.
• Suction particularly carefully in the tracheal area.

1. Push the on/off button to switch on the device.


» The pump starts. The vacuum is set which was selected prior to switch off.
Patient may be seriously injured if the set vacuum is too high.
2. Please select the vacuum according to the patient and the application. Push the
button for the required vacuum.
» The green LED above the selected button flashes.
& As long as the secondary air opening is open, the device does not suction.
1. Open the secondary air opening before inserting the suction catheter.
2. Apply the suction catheter in such a way as you were taught.
3. Close the secondary air opening so that the device suctions.
Suffocation is possible due to full canister system.
4. Pay attention to the filling level of the canister system.
5. Empty the secretion canister or change the suction bag once it is half full. As soon
as the canister system is too full, the float ball seals the intake area. You can then no
longer suction with the device.
Make sure that the hose is not kinked during suction. Otherwise, the suction capacity
applied on the patient is too low.
& If you want to interrupt the suction briefly, you can clamp the suction hose into the
opening of the hose rewind.

If secretion has penetrated into the device, please observe chapter “5.6 Oversuction” on
page 41.

After use
1. Switch off the device by pressing the on/off button for at least 1 second.
2. Clean the device after each use: Chapter “5.0 Cleaning and disinfection” on page
35.
3. Perform a function check after each cleaning: Chapter “6.2 Function check” on page
42.

34 Operation
5.0 Cleaning and disinfection
We recommend that you document any maintenance work and also any exchange of
parts.
WARNING
Risk of infection due to secretion on the device, accessories and consumables.
Deadly diseases can be transmitted.
• Always wear disposable gloves during any cleaning.
• Clean the device after every use:
• Clean and disinfect the device according to the operating instructions.
• The device must be reprocessed professionally if secretion has penetrated into the
device. Please observe chapter “5.6 Oversuction” on page 41.

5.1 Preparing for cleaning


1. Switch off the device.
2. Remove the recharging accessories from the device.
3. Remove the canister system from the device: Chapter “3.4 Connection and removal
of canister system and hoses” on page 18.
Risk of infection due to overflowing secretion. Deadly diseases can be transmitted.
4. Carefully remove the canister lid / the suction bag.
5. Dispose of the secretion / the suction bag. Please observe the notes in chapter “10.0
Disposal” on page 53.
6. Dispose of all disposables (e.g. suction Catheter, fingertip, disposable suction hose).
If you are using the DDS canister system: Dispose of the DDS bacterial and viral
filter.
7. Remove the hose rewind.
8. Remove the support for the canister system.

5.2 Cleaning
Please observe the operating instructions of the disinfectant manufacturers. Pay
particular attention to the information regarding the concentration of the disinfectants
and the material compatibility.
Some disinfectants may stain the parts of the canister lid and silicone hoses. Parts can
also stain due to autoclaving. However, this has no influence on the properties of the
materials.
Please only use disinfectants which are recommended by ATMOS (chapter “5.4 Recom-
mended disinfectants” on page 38). The use of other disinfectants may damage the
device or the canister system.

DDS canister system


Number of reprocessing cycles: max. 50.
1. Disassemble canister lid and filter holder into their individual parts. Chapter “3.4
Connection and removal of canister system and hoses” on page 18.

Cleaning and disinfection 35


1 2 3
5
4

1 Filter holder 3 Inner canister lid


2 Sealing ring 4 Float ball
5 External canister lid
2. Rinse the following parts of the DDS canister system with clear water:
- Secretion canister
- Inner canister lid
- External canister lid
- Float ball
- Filter holder
- Sealing ring
- Suction hose
- Support for canister system
3. Clean the mentioned parts with a brush or a cloth.
4. Disinfect the mentioned parts with a disinfectant which is recommended by ATMOS.
5. Let the individual parts of the canister lid and the filter holder dry.
As soon as the individual parts are dry:
6. Insert a new DDS bacterial and viral filter.
7. Reassemble the individual parts of the canister lid and the filter holder.

Serres®, Medi-Vac® canister system, other canister systems


• Please observe the instructions in the operating instructions for the canister system.
• Do not operate the device without a bacterial filter. The hydrophobic bacterial and
viral filter (REF 443.0738.0) must be used with a secretion canister system which has
no integrated bacterial filter.

Vacuum hose
After every suction process:
1. Rinse the vacuum hose with clear water for at least 10 seconds.
It must be exchanged after each patient or at least once a day:
2. Disinfect the vacuum hose with an instrument disinfectant recommended by
ATMOS.

36 Cleaning and disinfection


Device surface
WARNING
Electric shock due to liquid in the device.
• Disconnect the device from the mains power supply prior to cleaning.
• Do not rinse the device under running water and do not immerse it in any liquids.
• Make sure that the cleaning cloth is only damp and not wet.
• Do not autoclave the device.
• Do not immerse the device in disinfectant solution.

1. Clean the entire surface of the device and the hose rewind with a damp cloth.
2. Disinfect the entire surface of the device and the hose rewind with a surface disin-
fectant.

Power supply and recharging unit


WARNING
Electric shock due to liquid in the power supply.
• Disconnect the power supply and recharging unit from the mains power supply
prior to cleaning.
• Do not rinse the power supply and recharging unit under running water and do not
immerse it in any liquids.
• Make sure that the cleaning cloth is only damp and not wet.
• Do not autoclave the power supply and recharging unit, do not sterilise it and do
not thermally disinfect it.
• Do not immerse the power supply and recharging unit in disinfectant solution.

1. Clean the power supply and recharging unit with a damp cloth. You can use a mild
detergent.
2. Disinfect the power supply and recharging unit with a surface disinfectant. Recom-
mended: Terralin® Protect.

Wall and device support


1. Clean the wall and device support with a damp cloth.
2. Disinfect the wall and device support with a surface disinfectant.

5.3 After cleaning


Risk of injury to user and patient due to a damaged device.
1. Check after each cleaning whether the device is obviously damaged. In case the
device is damaged, please send it in to ATMOS.
2. Perform a function check: Chapter “6.2 Function check” on page 42.
3. Prepare the device for the next use.

Cleaning and disinfection 37


5.4 Recommended disinfectants
If you are using aldehyde and amine-containing disinfectants on the same object, this
may cause discolouration.

5.4.1 Instrument disinfection


Disinfectant Ingredients in 100 g Manufacturer
Mucocit®-T didecyldimethylammonium chloride 3.9 g Merz Dental, Lütjenburg
alkyl propylene diamine-1,5-bisguanidinium 4.5 g
acetate
bis(aminopropyl)laurylamine 2g
laurylpropylene diamine 2.8 g
non-ionic surfactants
Gigasept® FF (new) succindialdehyde 11.9 g Schülke & Mayr, Norder-
(Application concen- dimethoxytetrahydrofuran 3.2 g stedt
trate) anionic and non-ionic surfactants, perfumes,
methylisothiazolinones
Sekusept® PLUS glucoprotamin 25 g Ecolab, Düsseldorf
(Application concen-
trate)

5.4.2 Surface disinfection

Coated surfaces
Disinfectant Ingredients (in 100 g) Manufacturer
Green & Clean SK Di alkyl dimethyl ammonium chloride <1g Metasys, Rum
Alkyl dimethyl ethyl benzyl ammonium chloride <1g (Austria)
Alkyl dimethyl benzyl ammonium chloride <1g
Dismozon® pur magnesium monoperoxyphthalate hexahydrate 80 g Bode Chemie,
(Granulate) Hamburg
End of product 12/2014
Dismozon® plus magnesium monoperoxyphthalate hexahydrate 95.8 g Bode Chemie,
(Granulate) Hamburg
Kohrsolin® FF glutaral 5 g Bode Chemie,
(Application concentrate) benzyl-C12-C18-alkyldimethyl-ammonium 3 g Hamburg
chlorides 3g
didecyldimethylammonium chloride
Kohrsolin® extra (ethylenedioxy)dimethanol 14.1 g Bode Chemie,
(Application concentrate) glutaral 5 g Hamburg
didecyldimethylammonium chloride 8g
Perform® Potassium-bis(peroxymonosulfate)-bis(sulfate) 45 g Schülke & Mayr,
Norderstedt
Terralin® Protect benzyl-C12-16 alkyldimethyl-, chloride 22 g Schülke & Mayr,
(Application concentrate) 2-phenoxyethanol 17 g Norderstedt
aminoalkylglycine 0.9 g
non-ionic surfactants, perfumes

Other surfaces
Disinfectant Ingredients (in 100 g) (in 100 g) Manufacturer
Dismozon® pur magnesium monoperoxyphthalate hexahydrate 80 g Bode Chemie,
(Granulate) Hamburg
End of product 12/2014
Dismozon® plus magnesium monoperoxyphthalate hexahydrate 95.8 g Bode Chemie,
(Granulate) Hamburg
Kohrsolin® FF glutaral 5 g Bode Chemie,
(Application concentrate) benzyl-C12-18-alkyldimethyl-ammonium chlorides 3 g Hamburg
didecyldimethylammonium chloride 3g
Kohrsolin® extra (ethylenedioxy)dimethanol 14.1 g Bode Chemie,
(Application concentrate) glutaral 5 g Hamburg
didecyldimethylammonium chloride 8g

38 Cleaning and disinfection


Mikrobac® forte benzyl-C12-18-alkyldimethyl-ammonium chlorides 19.9 g Bode Chemie,
(Application concentrate) N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine 5 g Hamburg

Perform® Potassium-bis(peroxymonosulfate)-bis(sulfate) 45 g Schülke & Mayr,


Norderstedt
Terralin® Protect benzyl-C12-16 alkyldimethyl-, chloride 22 g Schülke & Mayr,
(Application concentrate) 2-phenoxyethanol 17 g Norderstedt
Suitable for power aminoalkylglycine 0.9 g
supply and recharging non-ionic surfactants, perfumes
unit.
SaniCloth® Active didecyldimethylammonium chloride < 1 g Ecolab,
Düsseldorf
Incidin® Active peracetic acid < 1 g Ecolab,
Düsseldorf
Mikrozid® Sensitive benzyl-C12-16 alkyldimethyl-, chloride; 0.26 g Schülke & Mayr,
Wipes didecyldimethylammonium chloride 0.26 g Norderstedt
benzyl-C12-14-alkyl [(ethylphenyl)methyl] 0.26 g
dimethyl-, chlorides
Gigasept® pearls sodium percarbonate 43 g Schülke & Mayr,
Suitable for DDS tetraacetylethylenediamine 22 g Norderstedt
secretion canister

Cleaning and disinfection 39


Cleaning and disinfection plan
ATMOS C / E 341 Battery
5.5 Hygienic plan
What How When Who

C D S

After each patient


Every 14 days*
Notices

Disinfection

Sterilization
Qualified and trained staff who

/ after each
Parts to be reprocessed
are familiar with reprocessing
Cleaning

suction
Daily*
Surfaces
Housing X X Wipe cleaning and disinfection X X

Device base X X1 Wipe cleaning and disinfection X X

Hose reel X X1 Wipe cleaning and disinfection X X

Wall and device support X X Wipe cleaning and disinfection X


Power supply and recharging Wipe cleaning and disinfection with a damp
X X X
unit cloth. Do not immerse into any liquid!

Secretion canister system


Cleaning with a brush; cleaning and
DDS secretion canister X X1 X2 disinfection (automatic or manual), X X
sterilization is possible
Cleaning with a brush; cleaning and
Outer canister lid (DDS) X X1 X2 disinfection (automatic or manual), X X
sterilization is possible
Cleaning with a brush; cleaning and
Inner canister lid (DDS) X X1 X2 disinfection (automatic or manual), X X
sterilization is possible
Cleaning with a brush; cleaning and
Float ball (DDS) X X1 X2 disinfection (automatic or manual), X X
sterilization is possible
Cleaning with a brush; cleaning and
Filter holder X X1 X2 disinfection (automatic or manual), X X
sterilization is possible
Sealing ring X X1 Cleaning and Disinfection X X
Hydrophobic DDS bacterial Exchange. If the filter is blocked it must also
X X
and viral filter be exchanged.
Fingertip Exchange. X X

Support for canister system X X1 Cleaning and Disinfection X X

X X1 X2 When using without suction catheter X X


Suction hose (DDS)
X X 1
X 2
When using with suction catheter X X

Vacuum hose X X Cleaning and Disinfection X X

Recommended disinfectants Important information The above stated hygiene requirements are based on
the regulations according to the Medical Devices Act, the
Surface disinfection for coated surfaces: Manual disinfection of instruments: Wipe cleaning and disinfection: Medical Devices Operator Ordinance, §18 IfSG and the
• Green & Clean SK (ATMOS) • Mucocit®-T (Merz Dental) All surfaces have to be wiped with recommendations of the Robert Koch Institute.
• Dismozon® pur (Bode Chemie) • Gigasept® FF neu (Schülke & Mayr) a clean (disposable) wipe which is Definition of the required reprocessing steps result
from the recommendations of the Robert Koch Institute:
• Dismozon® plus (Bode Chemie) • Sekusept® PLUS (Ecolab) damped with disinfectant solution. „Requirements for the reprocessing of medical products“,
• Kohrsolin® FF (Bode Chemie) The entire surface has to be wiped from Robert Koch Institute. The medical products were
• Kohrsolin® extra (Bode Chemie) thoroughly and may not be dried categorised in the risk groups uncritical, semi-critical and
• Perform® (Schülke & Mayr) For concentrations, contact time, tem- afterwards. critical. The reprocessing measures mentioned in this
cleaning and disinfection plan are a recommendation
• Terralin® Protect (Schülke & Mayr) perature, material compatibility, please of ATMOS MedizinTechnik. Any additional reprocessing
see the relevant information from the 1)
Preferred: mechanical cleaning and measures are at the operator‘s discretion.
Other surfaces: manufacturer. All the recommended disinfectants which are stated
• Dismozon® pur (Bode Chemie) disinfection in the washer disinfector herein are listed disinfectants (VAH/RKI) and have
• Dismozon® plus (Bode Chemie) with a device according to the ISO been tested on their suitability of use on the ATMOS C
• Kohrsolin® FF (Bode Chemie) 15883-1 (Program: Rinsing 1 min with / E 341 Battery. ATMOS MedizinTechnik cannot be hold
liable for any damage caused by wrong concentration
• Kohrsolin® extra (Bode Chemie) Wrong concentration of cold water, cleaning 5 min at 55°C,
of the disinfectants or by the application of any other
• Mikrobac® forte (Bode Chemie) disinfectants may lead to neutralisation 1 min with 1/3 cold and disinfectants.
• Perform® (Schülke & Mayr) material damage! 2/3 warm water, rinsing 1 min with Patients with suspicion of a clinical disease or who
purified water, thermal disinfection 5 developed a transmissible spongiform encephalopathy
• Terralin® Protect (Schülke & Mayr) (CJK, vCJK, etc.) have to be treated at facilities which are
• SaniCloth® Active (Ecolab) min at 93°C with purified water) able to provide for the necessary preventive measures
• Incidin® Active (Ecolab)
2)
If required hot-steam sterilization at against infection. The reprocessing of the reusable
134°C, 3 x fractionated pre-vacuum instruments and material may only be performed
• Mikrozid® Sensitive Wipes (Schülke at facilities which have an externally certified QM
& Mayr) method, sterilization time 5 min with a Management acc. to DIN EN ISO 13485.
• Gigasept® pearls (Schülke & Mayr) device according to EN285 The Medical Devices Act, IfSG, the RKI directives, BGR 250
* Homecare, unless there is no change and TRBA 250 always have to be considered.

in patient. ATMOS MedizinTechnik GmbH & Co. KG


Ludwig-Kegel-Str. 16 n 79853 Lenzkirch/Germany
Phone +49 7653 689-0 n Fax +49 7653 689-190
GA3GB.320100.0 2021-09 Index: 04 [email protected] n www.atmosmed.com

40 Cleaning and disinfection


5.6 Oversuction
If you use the ATMOS E 341 Battery according to instructions with the DDS bacterial
and viral filter and the float ball, the device cannot be oversucked during normal use.
Nevertheless, should secretion penetrate into the interior of the device, the device is
oversucked.
This can happen, for example, if no hydrophobic DDS bacterial and viral filter is used
and the device tips over.
Reduced suction capacity is an indication for an oversucked device. If you suspect that
your device might be oversucked, proceed as follows:
WARNING
Risk of infection due to secretion on and within the device.
Deadly diseases can be transmitted.
• Always wear disposable gloves when touching the oversucked device.
• Clean and disinfect the device.
• Send in the device to ATMOS or an authorised ATMOS service partner, chapter “6.3
Sending in the device” on page 44.

Cleaning and disinfection 41


6.0 Maintenance and service
Maintenance, repairs and periodic tests may only be carried out by persons who have
the appropriate technical knowledge and are familiar with the product. To carry out
these measures, the person must have the necessary test devices and original spare
parts.
ATMOS recommends: Work should be carried out by an authorised ATMOS service
partner. This ensures that repairs and testing are carried out professionally, original
spare parts are used and warranty claims remain unaffected.

6.1 Periodic tests


Please comply with the country-specific guidelines regarding regular testing especially
for the electrical safety.
ATMOS recommends a test every 24 months.

6.2 Function check


Perform a function check:
• prior to each use
• after each use or cleaning
• every 4 weeks in case the device is not used
• after every maintenance work, service or repair

6.2.1 Manual function check


1. Please check whether the following parts are damaged or torn:
- all hoses
- canister system
2. In case parts are visibly damaged: Please replace them.
3. Switch on the device.
4. Check whether all LEDs are illuminated.
5. Check the battery status.
6. Connect a fingertip to the suction hose and close the secondary air opening.
7. Close the front opening of the fingertip with your thumb.
8. Select the vacuum −0.5 bar.
9. Please check whether the device reaches the vacuum after approx. 20 seconds:
The pump switches off and the green LED above the −0.5 bar button is illuminated
continuously.
10. In case the device does not reach the vacuum within 20 seconds: Please check the
device for any possible sources of error and remedy them: chapter “7.0 Eliminating
errors” on page 47.
11. You may now use the device or switch it off.

42 Maintenance and service


6.2.2 Automatic function check

Termination
1. Press the button for the battery status.
» The device switches off.

Performing a function check


The automatic function control checks successively the following functions:
• Remaining operating hours
• Battery service life
• Tightness (duration: approx. 10 seconds)
The display of battery status shows the results in quick succession. Therefore, first read
this section completely.
Between the individual tests only the red LED is illuminated and a signal tone sounds. If
two signal tones are sounded in direct succession, then the previous test has failed.
1. Connect a fingertip to the suction hose.
2. Close the secondary air opening of the fingertip and close the front opening with
your thumb.
3. Press the button for the battery status for approx. 3 seconds.
» The function check starts as soon as all LEDs are illuminated. The first result is
shown immediately.
4. Check the results by means of the battery status indication.
5. The fingertip may only be opened when all LEDs are off. The function check is
completed.

Result
Remaining operating hours
1. 2. 3. 4. Short signal tone:
< 50 h
1. Please contact your
service.
> 750 h > 500 h > 250 h < 250 h
Remaining battery capacity
1. 2. 3. 4. Short signal tone:
< 10%
1. Exchange the battery.

100% 70% 40% 20%


Tightness
1. 2. 3. 4. Short signal tone:
System has a leakage.
1. See chapter “7.0 Elimi-
nating errors” on page
−0.5 bar −0.4 bar −0.3 bar −0.2 bar 47.

Maintenance and service 43


6.3 Sending in the device
1. Remove and properly dispose of consumables.
2. Clean and disinfect the products and accessories according to the operating instruc-
tions.
3. Place used accessories with the device.
4. Fill in the form QD 434 “Delivery complaint / return shipment” and the respective
Decontamination certificate.
& This form is enclosed with each delivery and can be found at www.atmosmed.com.
5. The device must be well padded and packed in suitable packaging.
6. Place the form QD 434 “Delivery complaint / return shipment” and the respective
decontamination certificate in an envelope.
7. Affix the envelope to the outside of the package.
8. Send the product to ATMOS or to your dealer.

6.4 Handling of batteries


Batteries are wearing parts and therefore excluded from the general warranty. There is
a function guarantee of 6 months.
Please observe the following notes in order to reach the maximum service life of your
battery:
• Only use the original lithium-ion battery 4IMR 19/66-2 BM18650Z3.
• Please observe the operating instructions of the battery manufacturer.
• Prior to first use, the battery must be fully charged.
• Battery-run devices should only be stored when they are charged.
• Please fully recharge the battery every 6 months, even if the device is not used.
• Protect the batteries from direct solar radiation and keep them away from radiators.
The perfect storage temperature is between 8 and 15 °C.
• Exchange the battery when the remaining battery service life noticeably decreases.
• Batteries are run-down after approx. 500 charging cycles.

44 Maintenance and service


6.5 Battery exchange
NOTICE
Damage to the electronics due to the use of a third-party battery.
• Only use the original lithium-ion battery 4IMR 19/66-2 BM18650Z3. This battery is
included in the scope of delivery and is available at ATMOS. The warranty claim shall
not be applicable if non-original spare parts are used.

Prerequisite: The support for the canister system is removed.


1. Switch off the device.
2. Disconnect the device from the power supply.
3. Remove the hose rewind: Chapter “3.6 Hose rewind” on page 29.
4. Place the device on its back with the control panel facing upwards.
5. Press the battery compartment cover 1 from the right a little towards the left.

6. Lift the battery compartment cover slightly and remove it from the upper guide.
The detent 2 could break off due to the incorrect removal of the cable.
7. Remove the cable from the device by pressing the detent 2 of the plug against the
plug and simultaneously pull the plug.

8. Remove the battery in an upward direction.


9. Insert the new battery. Please observe that the shape of the battery compartment
corresponds to the shape of the battery and the symbol points towards the
socket in the battery compartment.
10. Plug in the connector of the battery cable to the socket in the battery compartment.
11. Stow the cables in the battery compartment so that they cannot be damaged by the
battery compartment cover.
12. Insert the battery compartment cover in the upper rail.
13. Slide the battery compartment cover to the right until it stops.
14. Press down the battery compartment cover.
15. Slide the battery compartment cover completely to the right.
16. Attach the hose rewind: Chapter “3.6 Hose rewind” on page 29.
17. Affix the support for the canister system: Chapter “3.4 Connection and removal of
canister system and hoses” on page 18.
18. Perform a function check.

Maintenance and service 45


6.6 Exchanging the release button
& There are 2 springs under the release button. Pay attention that they are not
misplaced.
1. Place the screwdriver in the middle of the release button and lift the release button
up.

2. Replace with a new release button. Pay attention that the two springs are
positioned in the guide of the release button.

3. Press the release button downwards until it engages.

46 Maintenance and service


7.0 Eliminating errors
The ATMOS E 341 Battery was subjected to thorough quality control in the factory.
Nevertheless, if a problem does occur, you can possibly solve it yourself.

Recharging and battery status


Error indication Possible cause Remedy
Device cannot be The plug from the 1. Check the connection to the
recharged. charging accessories is mains supply.
poorly fitted.
Charging accessories 1. Exchange charging accessories.
are defective.
Battery is not 1. Check the plug connections in
connected properly. the battery compartment.
Battery temperature 1. After long-term use: Let the
too high or too low. device cool down.
2. Extreme ambient temperature:
Position the device where
appropriate to a cooler or
warmer place.
Deep discharge of the 1. Exchange the battery.
battery.
Defective electronics. 1. Send in the device for repair.
LED on the power supply Defective power 1. Exchange the power supply
and recharging unit is not supply and recharging and recharging unit.
illuminated. unit.
The power plug is 1. Check the connection to the
poorly fitted. mains supply.
When recharging the Battery service life 1. Exchange the battery.
battery, 100% cannot be is exhausted or the
achieved. The charging battery is defective.
time can take up to 3 Wrong charging acces- 1. Only use the provided charging
hours. sories. accessories or an original spare
part.
The bottom green and Battery is almost 1. Recharge the battery.
the red LED on the completely
display of battery status discharged.
flashes and a signal tone
sounds every 5 s.
During switch on: All Remaining battery 1. Perform a function check.
LEDs of the display of level is low. 2. Exchange the battery.
battery status flash for 5
s; a signal tone sounds.
All the green LEDs on the A non-ATMOS battery 1. Only use the provided battery
display of battery status is used. or an original spare part.
flash permanently.

Eliminating errors 47
Error indication Possible cause Remedy
All the LEDs on the Battery is not inserted. 1. Insert the battery.
display of battery status Battery is not 1. Check the plug connections in
flash permanently. connected properly. the battery compartment.
Defective battery. 1. Exchange the battery.
Defective electronics. 1. Send in the device for repair.
Battery compartment Battery is not fitted 1. Fit the battery correctly.
cover cannot be closed. correctly.
Battery compartment 1. Mount the battery
cover is installed incor- compartment cover correctly
rectly. according to operating instruc-
tions.

Switching on and off


Error indication Possible cause Remedy
Device cannot be Battery is discharged. 1. Recharge the battery.
switched on or off. Battery is not connected 1. Check the plug connections in
properly. the battery compartment.
The plug from the 1. Check the connection to the
charging accessories is mains supply.
poorly fitted.
Defective electronics. 1. Send in the device for repair.
During switch on: Device was stored 1. Switch on the device again.
Battery LEDs blink outside the operating
once, but the device temperature (battery is
does not start. in standby mode).
During switch on: The Device is not ready. 1. Perform a function check.
red LED of the display 2. Send in the device for repair.
of battery status
flashes for 5 s; a signal
tone sounds.
Pump does not start Vacuum is already built 1. Do not switch on the device if
up. up. the vacuum is already built up.
Device switches off Self-protection of the 1. Let the device cool down for
after 60 min. device. approx. 2 hours.
Device switches off Battery is discharged. 1. Recharge the battery.
after < 60 min. Battery temperature is 1. Let the device cool down or
too high. select a lower vacuum.

48 Eliminating errors
Vacuum and suction capacity
Error indication Possible cause Remedy
Vacuum is not built Battery is discharged or 1. Recharge or exchange the battery.
up or cannot be defective.
reached. Leakages in the hoses 1. Check canister lid and hoses for
or the canister system. tight fit.
2. DDS canister system: Firmly insert
the hydrophobic DDS bacterial
and viral filter and check the
sealing ring and filter holder.
Fingertip is not closed. 1. Close both openings of the
fingertip.
Liquid has penetrated 1. Send in the device for repair.
the device.
Pump is defective or 1. Send in the device for repair.
the device has a leak.
Low ambient pressure Not possible.
(e.g. high altitude).
Low suction capacity Hydrophobic DDS 1. Replace the hydrophobic DDS
although the vacuum bacterial and viral filter bacterial and viral filter.
has been reached. is blocked.
Hose is kinked. 1. Check the hoses.
DDS canister system: 1. Check and, if needed, clean
Float ball closes the the float ball and the float ball
suction area. compartment.

Eliminating errors 49
8.0 Accessories
Accessories REF
Serres® external canister 1 l 312.0465.0
Medi-Vac® external canister 1 l 312.0473.0
DDS canister system 1 l, complete 318.1000.0
Wall and device support for ATMOS Emergency Suction Devices 318.1250.0
Retrofit kit DDS canister system 318.1350.0
Retrofit kit Serres canister system
®
318.1450.0
Retrofit kit Medi-Vac® canister system 318.1650.0

8.1 Wall and device support

Affixing the power supply and recharging unit


Prerequisite: The Velcro fastener 1 is attached to the wall and device support.
Damaged cable due to incorrect mounting.
1. Please ensure that the power supply and recharging unit are positioned with the
writing towards the wall or the standard rail. Otherwise, the cable could be clamped.
2. Affix the parts 2 and 3 and fix the cable 4 with the Velcro fastener 1 .
2

1 4

Wall mounting
& Only affix the power supply and recharging unit after you have marked the holes to
be drilled.
& Screws are not included in delivery.
1. Only use screws (max. 4 mm) which are suitable for the material of the wall.
2. Position the wall and device support at an easily accessible place.
3. Check whether the wall is smooth and vertical at the mounting position.
4. Hold the wall and device support to the mounting location and align it with a spirit
level.
5. Mark the holes to be drilled on the wall.
6. Drill the holes with a drill bit appropriate for the wall material and the chosen
screws.
7. Mount the power supply and recharging unit to the wall and device support. The
power supply and recharging unit must not be connected to the mains supply.
8. Screw the wall and device support to the wall with suitable screws.
9. Connect the power supply and recharging unit to the mains supply.

50 Accessories
10. Check whether the charging accessory is correctly attached by installing the device.
» The battery will be charged.

Attaching to / removing from a standard rail


Prerequisite: The power supply and recharging unit is mounted.
Mounting Removal

Attaching the device


1. Slide the device from above onto the wall and device support until it clicks into
place.
» If the charging accessory is attached, the battery is charged automatically.

Removing the device


1. Press the release button 1 and pull the device simultaneously vertically upwards.

8.2 Retrofit kit canister system


You may change the canister system. The retrofit kits comprise the canister system as
well as the required support for the canister system. The retrofit kits for the disposable
canister systems also comprise the vacuum hose.

Retrofitting
1. Remove the existing canister system, chapter “3.4 Connection and removal of
canister system and hoses” on page 18.
2. Remove the existing support for the canister system.
3. Attach the new support for the canister system.
4. Insert the new canister system.

Accessories 51
9.0 Spare parts and consumables
Consumables REF
Reusable suction hose, Ø 10 mm 318.1012.0
Fingertip for reusable suction hose, Ø 10 mm, 10 pcs. 318.1100.0
Hydrophobic DDS bacterial and viral filter for ATMOS DDS secretion
340.0054.0
canister, pack of 10 pcs.
Suction hose, disposable, Ø 6 mm, L = 1.30 m, 10 pcs. 006.0057.0
Suction hose, disposable, Ø 6 mm, L = 1.30 m, 50 pcs. 006.0059.0
Serres suction bag 1 l, not autoclavable, 36 pcs.
®
312.0466.0
Medi-Vac® suction bag 1 l, not autoclavable, 50 pcs. 312.0474.0
Vacuum hose for disposable canister system 318.1211.0
Suction catheter Unomedical , size: CH 12, L = 50 cm, 100 pcs.
®
000.0294.0
Suction catheter Unomedical®, size: CH 14, L = 50 cm, 100 pcs. 000.0295.0
Suction catheter Unomedical®, size: CH 16, L = 50 cm, 100 pcs. 000.0296.0
Hydrophobic bacterial and viral filter, Ø 8 mm 443.0738.0

Spare parts REF


DDS canister system
DDS secretion canister 1 l 318.1013.0
DDS external canister lid 318.1002.0
DDS inner canister lid 318.1004.0
Float ball 000.0839.0
Filter holder 318.1003.0
Sealing ring 055.0112.0
Device
Battery for ATMOS E 341 Battery 319.0015.0
Battery compartment cover 318.0012.0
Hose rewind 319.0004.0
Device base 319.0003.0
Support for DDS canister system 318.1010.0
Support for Serres® canister system 318.1210.0
Support for Medi-Vac canister system
®
318.1500.0
Release button 318.0013.0
Spring for release button 000.1029.0
Power supply and recharging unit 318.0035.0
2-pin power cable 008.0920.0

52 Spare parts and consumables


10.0 Disposal
Packaging
1. Please recycle the packaging.

Secretion and blood


1. Please dispose of secretion, blood and contaminated parts in line with country-spe-
cific regulations.
In the Federal Republic of Germany the “Requirements on the implementation aid for
disposal of waste from healthcare institutions” are valid, a statement of the Federal /
State Working Group on Waste.

Canister system
Disposable products may not be reprocessed and may not be reused! Please dispose of
disposable products professionally.
The following notes are only applicable for reusable products.
1. Clean and disinfect the reusable products of the canister system.
2. Recycle the disinfected reusable products.

ATMOS E 341 Battery


Do not dispose of the device or the battery in domestic waste.
There is a lithium-ion battery included in the ATMOS E 341 Battery which
must be disposed of in accordance with applicable guidelines.
1. Clean and disinfect the device.
2. In Germany: Send in the device to ATMOS or your specialised dealer. They will
dispose of the device professionally.
3. In other countries: Dispose of the device professionally and according to
country-specific laws and regulations.
In Germany the device is excluded from the Electrical and Electronic Equipment Act
(ElektroG) according to the National Register for waste electric equipment because it
may be contaminated. Do not dispose of the device with electronic waste.
The housing is fully recyclable. Refer to the country-specific laws and regulations.

Disposal 53
11.0 Technical data
Device
Dimensions (W x H x D):
with DDS canister system 370 x 277 x 146 mm
with Serres® canister system 370 x 277 x 146 mm
with Medi-Vac® canister system 370 x 277 x 136 mm
with universal bracket 370 x 277 x 136 mm
Weight:
Device with battery / without canister 3.65 kg
system and support
DDS canister system with support for 1.00 kg
DDS canister system
Serres® canister system with support for 0.65 kg
Serres® canister system
Medi-Vac® canister system with support 0.295 kg
for Medi-Vac® canister system
Universal bracket 0.2 kg
Operation:
Temperature range −5 °C up to +50 °C
Relative air humidity 5% to 95% without condensation
Air pressure 540 hPa to 1100 hPa
Transport / storage:
Temperature range −40 °C up to 70 °C
Relative air humidity 5% to 95% without condensation
Air pressure 540 hPa to 1100 hPa
Charging:
Temperature range 0 °C up to 40 °C
Relative air humidity 5% to 95% without condensation
Air pressure 540 hPa to 1100 hPa
Maximum operating altitude 5000 m (NN)
Contamination level Class 1 (fully-sealed housing)
Overvoltage category II
Maximum power consumption 45 W
Maximum current consumption 3.7 A
Mains voltage 12 V DC nominal (at least 10 V, max. 15 V)
at the charging interface or via the power
supply and recharging unit.
Pump Vacuum pump (diaphragm pump), 1 head
Suction capacity at the device inlet 34 l/min ± 4 l/min
(without canister system) at −0.8 bar,
fully recharged battery and 21 °C / 1013
hPa (determined by buffer canister 1 l)
Suction capacity at the inlet of the 30 l/min ± 3 l/min
DDS canister system at −0.8 bar, fully
recharged battery and 21 °C / 1013 hPa
Maximum achievable vacuum 0.8 bar* ± 5% resp. 80% of the air pressure
Vacuum adjustment Via predefined steps: −0,1 bar, −0,2 bar,
−0,5 bar and −0,8 bar, electronically
controlled

54 Technical data
Vacuum display By means of LEDs on the control panel
Display By means of LEDs on the control panel:
on/off, selected vacuum, actual vacuum,
display of battery status, warning (red
LED)
Power cycles (short-term operation) 60 min On, 120 min Off
Noise level: mean sound pressure level at < 60 dB(A)
1 m distance and at −0.8 bar
Classification according to EN 60601-1:
• Protection class against electric shock Protection class II (during mains and
• Degree of protection against electric battery operation)
shock Type BF applied part
Degree of protection against: IP34D
• Ingress of solid foreign objects
• Penetration of dust
• Ingress of water with harmful effects
Periodic tests Recommended: testing every 24 months
Suspension Compatible with ATMOS wall and device
support
Classification according to EN ISO High vacuum / high flow
10079-1
Product class according to Directive IIa
93/42/EEC
UMDNS code 15-016 Suction device, emergency
GMDN code 36616, Suction unit, transport and
emergency
*1 bar = 100 kPa

Battery
Type Lithium-ion; 4INR 19/66-2 US 18650VTC6
Dimensions (W x H x D) 43 x 73 x 75 mm
Weight 0.4 kg
Nominal capacity 6 Ah
Nominal voltage 14.4 V
Charging time Battery status 80%: 3 h 45 min at 20 °C
without operation; battery status 100%:
approx. 5 h 40 min
Automatic switch-over to trickle charging
Recharging interval during long-term Every 6 months
storage
Battery operating time during continuous 85 min at −5 °C
operation with fully recharged battery / 85 min at +21 °C
new battery (>20 l/min, setting −0.8 bar) 42 min at +50 °C
Life cycle Approx. 500 recharging cycles
Display Display of battery status during operation
and recharging
Typical battery operating life* −0.2 bar: 200 min
−0.5 bar: 140 min
−0.8 bar: 85 min
* Measured at +21 °C, continuous operation, without battery recharging and at free air flow.

Technical data 55
DDS canister system
Capacity 1000 ml
Connection reusable suction hose Ø 10 mm I.D.
Reusable suction hose:
Diameter Ø 10 mm I.D.
Length 1300 mm
Connection to the suction device Direct connection (without intermediate hose)
DDS bacterial and viral filter DDS bacterial and viral filter cartridge for use
in the secretion canister lid, single-use product
Bacterial filtration efficiency (BFE) 99.999778%*
Viral filtration efficiency (VFE) 99.73%*
Overall filtration efficiency >99.95%*
Filter class H13 (High-Efficiency Particulate Air/
Arrestance)*
*External test report (test laboratory)

Disposable canister system


Capacity 1000 ml
Connection disposable suction hose Ø 7 mm I.D.
Disposable suction hose:
Diameter Ø 6 mm I.D.
Length 1300 mm
Connection to the suction device By means of a vacuum hose (intermediate
hose)
Bacterial filter Integrated in the suction bag

Power supply and recharging unit


Dimensions (W x H x D) 130 x 36 x 60 mm
Weight 280 g
Operation:
Temperature range 0 °C up to +40 °C
Relative air humidity 10% to 90% without condensation
Air pressure 700 hPa to 1100 hPa
Transport / storage:
Temperature range −40 °C up to +70 °C
Relative air humidity 10% to 95% without condensation
Air pressure 700 hPa to 1100 hPa
Electrical connection 100 V AC to 240 V AC, 50 Hz to 60 Hz
Maximum current consumption 1.1 A
Output nominal 13.8 V DC, 3.5 A
Classification according to EN 60601-1:
• Protection class against electric shock Protection class II
• Degree of protection against electric Type CF applied part
shock
Degree of protection against: IP40
• Ingress of solid foreign objects
• Penetration of dust
• Ingress of water with harmful effects
Length of output line 1.8 m
Length of power supply cord approx. 2 m

56 Technical data
12.0 Notes on EMC
• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed
according to the following EMC notes.
• Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.

Guidelines and Manufacturer’s Declaration – Emissions


The ATMOS E 341 Battery is intended for use in the environment specified below. The customer or user of the
ATMOS E 341 Battery should ensure that it is used in such an environment.

Emissions Test Compliance Electromagnetic Environment – Guidance


RF Emissions acc. to CISPR 11 Group 1 The ATMOS E 341 Battery uses RF energy
only for its internal function. Therefore, its HF
emissions are very low and it is unlikely that
nearby electronic devices will be affected.
RF Emissions acc. to CISPR 11 Class B
The ATMOS E 341 Battery is suitable for use
Harmonic emissions according to Class A in all establishments, including domestic
IEC 61000-3-2 ones, and those directly connected to the
public low-voltage power supply network that
Voltage fluctuations/flicker Corresponds
supplies buildings used for domestic purposes.
according to IEC 61000-3-3

Guidelines and Manufacturer’s Declaration – Immunity for ATMOS E


341 Battery
The ATMOS E 341 Battery is intended for use in the electromagnetic environment specified below. The
customer or user of the ATMOS E 341 Battery should ensure that it is used in such an environment.

Electromagnetic Environment –
Immunity Test IEC 60601 Test Level Compliance Level
Guidance
ESD acc. to ± 6 kV contact ± 6 kV contact Floors should be made of wood or
IEC 61000-4-2 concrete or laid with ceramic tiles.
± 8 kV air ± 8 kV air
If floors are synthetic, the relative
humidity should be at least 30%.
Fast electrical ± 2 kV mains ± 2 kV mains Mains power quality should be that
transient/burst IEC of a typical commercial or hospital
± 1 kV I/Os
61000-4-4 environment.
Surges IEC 61000-4-5 ± 1 kV common ± 1 kV common mode Mains power quality should be that
mode of a typical commercial or hospital
± 2 kV differential
environment.
± 2 kV differential mode
mode
Voltage dips, short < 5% UT (> 95% dip of < 5% UT (> 95% dip of Mains power quality should be that
interruptions and the UT for 0.5 cycle) the UT for 0.5 cycle) of a typical commercial or hospital
voltage variations environment. If the user of the
40% UT (60% dip of 40% UT (60% dip of
according to IEC ATMOS E 341 Battery demands
the UT for 5 cycles) the UT for 5 cycles)
61000-4-11 continued operation even in case
70% UT (30% dip of 70% UT (30% dip of of interruptions of the energy
the UT for 25 cycles) the UT for 25 cycles) supply, it is recommended to supply
< 5% UT (> 95% dip of < 5% UT (> 95% dip of the ATMOS C 341 Battery from an
the UT for 5 s) the UT for 5 s) uninterruptible power supply or a
battery.
Magnetic field at 3 A/m Inapplicable Power frequency magnetic
power frequency fields should be that of a
50/60 Hz acc. to IEC typical commercial or hospital
61000-4-8 environment.
NOTE
UT is the AC mains voltage prior to application of the test level.

Notes on EMC 57
Guidelines and Manufacturer’s Declaration – Immunity
The ATMOS E 341 Battery is intended for use in the electromagnetic environment specified below. The
customer or user of the ATMOS E 341 Battery should ensure that it is used in such an environment.

IEC 60601 Test Compliance


Immunity Test Electromagnetic Environment – Guidance
Level Level
Conducted RF IEC 3 Veff 10 V Portable and mobile communications
61000-4-6 150 kHz to 80 MHz equipment should be separated from the
ATMOS E 341 Battery, including cables, by no
less than the distances as calculated/listed
Radiated RF IEC 3 V/m 10 V/m below.
61000-4-3 80 MHz to 2.5 GHz
Recommended distances:
d = 0.35 √ P
d = 0.35 √ P
80 MHz to 800 MHz
d = 0.70 √ P
800 MHz to 2.5 GHz
where P is the max. power in watts (W) and d
is the recommended separation distance in
metres (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compliance
level (b).
Interference may occur in the vicinity of
equipment bearing the following symbol:

NOTE 1
With 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by
absorption and reflection of buildings, objects and people.
a
The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain
radio equipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predetermined
exactly.
To determine the electromagnetic environment in regard to stationary transmitters, a study of the location
is to be considered. If the field strength measured at the location where the ATMOS E 341 Battery is used
exceeds the above compliance level, the ATMOS E 341 Battery is to be observed to verify the intended
function. If abnormal performance characteristics are noted, additional measures may be necessary, such
as reorienting or relocating the ATMOS E 341 Battery.
b
Within the frequency range of 150 kHz to 80 MHz, the field strength should be below 3 V/m.

The ATMOS E 341 Battery may not be used directly next to other devices or piled up with other devices. If
operation next to or piled with other devices is necessary, please watch the device to check its intended
operation in this arrangement.

Recommended separations between portable and mobile RF


communications equipment and the ATMOS E 341 Battery
The ATMOS E 341 Battery is intended for use in an electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the ATMOS E 341 Battery can help prevent electro-
magnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the ATMOS E 341 Battery as recommended below, according to the
maximum output power of the communications equipment.
Safety distance, depending on transmit-frequency m

58 Notes on EMC
Nominal output of the trans- 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
mitter W d = [0.35] √ P d = [0.35] √ P d = [0.35] √ P
0.01 0.035 0.035 0.07
0.1 0.11 0.11 0.22
1 0.35 0.35 0.7
10 1.1 1.1 2.2
100 3.5 3.5 7.0
For transmitters for which the maximum nominal output is not indicated in the above table, the recom-
mended safety distance d in meters (m) can be determined using the equation belonging to the respective
column whereas P is the maximum nominal output of the transmitter in watts (W) acc. to manufacturer’s
specification.
NOTE 1
With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by
absorption and reflection of buildings, objects and people.

Notes on EMC 59
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16
79853 Lenzkirch / Germany
Phone: +49 7653 689-0
[email protected]

www.atmosmed.com

You might also like