P0149-801 ACG+ User Manual
P0149-801 ACG+ User Manual
User Manual
Commercial in Confidence
Table of Contents
1 Contents checklist............................................................ 8
1.1 Standard contents ................................................................................... 8
1.2 Optional accessories ................................................................................ 8
2 Safety information ........................................................... 9
2.1 Capabilities to test gases against the EN 12021:2014 standard ....................... 9
2.2 Electrochemical sensors (oxygen and carbon monoxide) ................................ 9
2.3 PID lamp cleaning kit ............................................................................. 10
2.4 Inlet pressure & flow .............................................................................. 10
3 Introduction .................................................................. 11
3.1 ACG+ overview ..................................................................................... 11
4 Installation details ......................................................... 13
4.1 Mechanical overview .............................................................................. 13
4.2 Gas sample connection (Both fixed and portable versions)............................ 14
4.3 Installation (Portable) ............................................................................ 15
4.4 Installation (Fixed) ................................................................................ 17
4.5 Safety interlock relay (ACG+ Fixed).......................................................... 20
4.6 Safety interlock override key-switch (ACG+ Fixed) ...................................... 21
4.7 Switched 24V DC outputs (ACG+ Fixed) .................................................... 22
5 Operating instructions ................................................... 23
5.1 Start-up .............................................................................................. 23
5.2 Controls............................................................................................... 24
5.3 Inlet/purge valve................................................................................... 25
5.4 Display ................................................................................................ 26
5.5 Menu................................................................................................... 27
5.6 Adjusting the backlight ........................................................................... 28
5.7 Power sub-menu ................................................................................... 29
5.8 System flow ......................................................................................... 29
5.9 Sensor readings .................................................................................... 30
5.10 Alarms ................................................................................................ 32
5.11 Clearing alarms and hysteresis ................................................................ 33
5.12 Calibration adjustment ........................................................................... 37
5.13 Viewing device information ..................................................................... 44
5.14 Viewing sensor information ..................................................................... 44
5.15 Data-logging ........................................................................................ 45
5.16 Sensor replacement reminder .................................................................. 48
5.17 Timed sample draw ............................................................................... 49
5.18 Setting the date and time ....................................................................... 53
6 Configuration ................................................................. 54
6.1 Software installation .............................................................................. 54
6.2 Making a connection .............................................................................. 57
6.3 Changing options .................................................................................. 59
6.4 Downloading data-logs ........................................................................... 60
6.5 Setting the time .................................................................................... 61
7 Networking .................................................................... 62
7.1 Overview ............................................................................................. 62
7.2 Connection ........................................................................................... 62
7.3 Transmission method ............................................................................. 62
7.4 Setting the ACG Static IP Address ............................................................ 63
7.5 Retrieving Data using TCP/IP Requests ..................................................... 63
7.6 Retrieving Data using UDP ...................................................................... 64
7.7 Datagram Structure ............................................................................... 64
8 Troubleshooting ............................................................. 70
8.1 System faults ....................................................................................... 70
9 Recommended spares and accessories .......................... 75
Document ref: P0149-801-21 May 2022 Page 3 of 102
Commercial in Confidence
ACG+
User Manual Int. Approved
9.1 Service requirements ............................................................................. 77
10 Maintenance requirements .......................................... 78
10.1 Sensor & filter replacement ..................................................................... 78
10.2 Filter replacement ................................................................................. 91
10.3 Service requirements ............................................................................. 92
11 Specifications .............................................................. 99
11.1 Instrument specification ......................................................................... 99
11.2 Sensor specification .............................................................................. 100
12 UK Declaration of Conformity .................................... 101
13 Declaration of conformity .......................................... 102
Commercial in Confidence
ACG+
User Manual Int. Approved
List of figures
Figure 1 ACG+ Overview ............................................................................... 11
Figure 2 ACG+ Mechanical Overview ............................................................. 13
Figure 3 Typical ACG+ Portable Installation .................................................. 15
Figure 4 Typical ACG+ Fixed installation........................................................ 17
Figure 5 Connector Locations ........................................................................ 18
Figure 6 ACG+ Power.................................................................................... 19
Figure 7 Typical Interlock Relay Wiring ......................................................... 20
Figure 8 Typical Use of Switched 24V Outputs ............................................... 22
Figure 9 Start-Up Splash Screen.................................................................... 23
Figure 10 ACG+ Front Panel ........................................................................ 24
Figure 11 Main Display Overview ................................................................. 26
Figure 12 System Status Indicator ............................................................... 27
Figure 13 The Main Menu............................................................................. 27
Figure 14 The Backlight Sub-Menu............................................................... 28
Figure 15 Power Sub-Menu......................................................................... 29
Figure 16 Example of a Sensor that is in a Warm-Up State............................ 31
Figure 17 Sensor Bar Graph Indicator .......................................................... 31
Figure 18 Degraded State Sensor Bar Graph Indicator .................................. 31
Figure 19 Out Of Display Range Bar Graph Indicators ................................... 31
Figure 20 Sensor in Fault Bar Graph Indicator .............................................. 32
Figure 21 A Low-Going Alarm That Has Been Triggered ................................ 32
Figure 22 Mute Button................................................................................. 32
Figure 23 Hysteresis Band on a High Going Oxygen Alarm ............................ 33
Figure 24 Selecting a CO2 Alarm to Adjust.................................................... 35
Figure 25 Example of Alarm Set-Point Adjustment ....................................... 36
Figure 26 Flow Regulator Fitted to Calibration Bottle ................................... 39
Figure 27 Connecting Calibration Fitting to ACG+ Calibration Port ................ 39
Figure 28 Calibration Adjustment Screen Showing High Calibration Bottle .... 40
Figure 29 Calibration Adjustment Results - CO2 Sensor Failed to Adjust ........ 41
Figure 30 Excluding the VOC Sensor from the Calibration Adjustment ........... 42
Figure 31 Adjusting the CO2 Calibration Value .............................................. 43
Figure 32 Calibration Adjustment Reminder ................................................. 43
Figure 33 Device Information Screen ........................................................... 44
Figure 34 Sensor Information Screen for O2 ................................................. 44
Figure 35 Data-Log Download Screen .......................................................... 46
Figure 36 Data-Log Download in Progress.................................................... 46
Figure 37 Sensor Change Reminder ............................................................. 48
Figure 38 Sensor Replacement Reset Confirmation Screen ........................... 49
Figure 39 A Dräger Impactor Attached to the Outlet Port for Oil Mist Testing 51
Figure 40 Timed Sample Setup .................................................................... 52
Figure 41 Timed Sample Progress ................................................................ 52
Figure 42 Date And Time Adjustment Screen................................................ 53
Figure 43 ACG+ Software without Device Connected .................................... 57
Figure 44 Connecting a USB Cable to an ACG+ ............................................. 57
Figure 45 ACG Software with Device Connected Showing General Options .... 59
Figure 46 The Data Logs Available on a Connected ACG+.............................. 60
Figure 47 Downloading Data Logs from an ACG+.......................................... 61
Figure 48 Setting the Date and Time on an ACG+ ......................................... 61
Figure 49 System Showing a Fault Condition ................................................ 70
Commercial in Confidence
ACG+
User Manual Int. Approved
List of tables
Table 1 Electrochemical sensor first aid procedures ....................................... 9
Table 2 Connector J17 Pin Assignments ....................................................... 18
Table 3 Connector J3 Pin Assignments ......................................................... 19
Table 4 Relay and 24V Output Operation under Fault and Alarm Conditions .. 21
Table 5 Sensor Status Icons ........................................................................ 30
Table 6 Scaled Hysteresis Values ................................................................. 33
Table 7 Recommended Calibration Gas Concentrations ................................. 37
Table 8 Data-Log Formatting Example ......................................................... 45
Table 9 List of System Faults and Possible Solutions .................................... 71
Table 10 List of Sensor Faults and Possible Solutions ..................................... 72
Table 11 List of Spares .................................................................................. 75
Table 12 List of Accessories........................................................................... 76
Table 13 6 monthly service for ACG+ ............................................................ 77
Table 14 12 monthly service for ACG+ .......................................................... 77
Table 15 24 monthly service for ACG+ .......................................................... 77
Table 16 Maintenance Schedule of ACG+ Components .................................... 78
Table 17 Service Schedule of ACG+ Components ............................................ 92
Commercial in Confidence
ACG+
User Manual Int. Approved
Warnings, Cautions and Notes
Warnings and Cautions are used in this Manual to highlight potential hazards and safety
risks. Notes are used to provide supplementary information that is not hazard-related.
Commercial in Confidence
ACG+
User Manual Int. Approved
1 Contents checklist
1.1 Standard contents
Below is a list of standard contents which will be supplied to the customer
a) Li-Ion rechargeable battery pack kit (Includes battery mounting case & charger)
b) NiMH rechargeable battery pack kit (Includes battery mounting case & charger)
c) External PSU (If not using rechargeable battery pack option, connection via front
panel)
d) External PSU (For fixed installation using cable gland direct to interface PCB)
e) External beacon (For fixed units only)
f) Pressure regulator (2 stage)
g) Humidifier (For when the ACG+ is in constant use)
h) Mounting kit to install on either a solid wall or instrument panel (For fixed units
only)
i) Gland and Override Keyswitch kit (For fixed units only)
The humidifier (g) should be installed if the ACG+ is going to be continuously connected
for over 12 hours to the compressed gas source with no exposure to ambient air. The
humidifier is required to pull moisture from the ambient air to re-hydrate the
electrochemical cells within the ACG+. If dry gas is constantly being passed over the
sensors they will dry out and stop performing correctly.
Commercial in Confidence
ACG+
User Manual Int. Approved
2 Safety information
WARNING: READ THE SAFETY INFORMATION FULLY BEFORE
USING THE ANALOX ACG+.
The ACG+ can test the following gas types from the standard:
• Breathing air
• Oxygen compatible air
• Nitrogen depleted air
When the life of the sensor has expired or it is leaking or otherwise damaged it must be
disposed of safely in accordance with local regulations.
The carbon monoxide cell contains an acidic electrolyte (sulphuric acid) whilst the oxygen
cell contains caustic electrolyte (potassium hydroxide). In the event of an accident, use the
following first aid procedures.
Commercial in Confidence
ACG+
User Manual Int. Approved
Body
Effect First Aid Procedures
Part
Eye Persons with pre-existing eye Irrigate thoroughly with water for at
problems may be more susceptible least 15 minutes.
to the effects of the substance.
Obtain medical advice immediately.
Corrosive. May cause redness, pain,
blurred vision, and eye burns.
Handling:
• Do not breathe in the powder, avoid contact with skin, eyes and clothing
• Wear suitable protective clothing
• Follow industrial hygiene practices, wash face and hands thoroughly with soap and
water after use and before eating, drinking, smoking or applying cosmetics
• The powder carries a TVL (TWA) limit of 10mg/m3
Commercial in Confidence
ACG+
User Manual Int. Approved
3 Introduction
The ACG+ is a multi-sensor gas analyser, specifically designed for the analysis of
contaminants in compressed breathing air. The ACG+ can be used as a portable
instrument for use between compressors on site or can be permanently connected to a
compressor outlet, enabling you to continually verify the quality of your breathing air. This
delivers a much greater degree of system safety than a 3 or 6 month spot check. The unit
will provide continuous ‘on line’ monitoring of O2, CO2, CO, volatile organic compounds
(VOC) and water vapour in compressed breathing air lines.
1) Display
2) USB port for data-download, with waterproof cap
3) Power connection (Via either optional rechargeable battery pack or external mains
adaptor)
4) Gas inlet/purge valve
5) Gas inlet/calibration port
6) Sample support port
Document ref: P0149-801-21 May 2022 Page 11 of 102
Commercial in Confidence
ACG+
User Manual Int. Approved
7) Exhaust
8) Exhaust
9) Breather
10) Humidifier (Optional)
11) M16 cable gland (Optional)
12) Override keyswitch (Optional)
13) Exhaust
14) Blow out blanking plug
Commercial in Confidence
ACG+
User Manual Int. Approved
4 Installation details
4.1 Mechanical overview
Commercial in Confidence
ACG+
User Manual Int. Approved
4.2 Gas sample connection (Both fixed and portable
versions)
• The sample gas inlet to the ACG+ must be at a regulated pressure in the range of
5 to 10 bar gauge (72.5 to 145psig), taken directly from a sample line from a
compressor, after all filtering and regulation, which is supplying the bottles, air
banks or breathing masks.
• The ACG+ is supplied with a 2m length of 6mmOD x 3mmID PTFE tubing which the
inlet/calibration adaptor should be connected to. Longer pipes can be used but
purge times and response times will be longer.
• It is recommended that a 2-stage regulator is used to maintain the sample gas
inlet pressure within operating limits.
• The ACG+ provides a Linktech panel mount coupling for the connection of the
sample gas line or calibration gas and also a Linktech inline socket (Which accepts
up to ¼” OD tubing) for connection to the ACG+ panel mount coupling.
• Before fitting the pipe-work the PTFE tubing should be fitted to the rear of the
supplied Linktech inline socket.
• The ID at one end of the PTFE tubing will require stretching slightly to fit over the
rear of the Linktech inline socket.
• Remove the nut from the rear of the Linktech inline socket and place over the
supplied length of PTFE tubing.
• Fit the 6mmOD x 3mmID tubing over the end of the Linktech inline socket by
slightly stretching the inner of the tube so that it fits over the rear of the adaptor
and secure in place with the nut.
• At the free end of the PTFE tubing, a compression nut should be fitted to connect
to the regulated pressure source.
Commercial in Confidence
ACG+
User Manual Int. Approved
4.3 Installation (Portable)
The ACG+ Portable can be powered by either the optional rechargeable battery pack (Li-
Ion or NiMH variants) or an external PSU. Both options connect to the front panel power
connector located next to the USB port.
Commercial in Confidence
ACG+
User Manual Int. Approved
Connect the charger connector to the battery connection lead coming from
the battery case (This will only fit one way) and secure using the bayonet
nut on the battery cable.
Plug the charger into a mains supply, the charger will do a battery check
prior to charging (The charger the LED will be yellow).
If any faults are found the charger LED will flash orange/green.
If all is okay the charger will then start to charge the battery (The charger
LED will be orange).
Once the charge is complete, the charger will go into trickle mode until
disconnected (The charger the LED will go green).
Once complete, disconnect the charger, the battery is then ready for
connection to the ACG+.
Commercial in Confidence
ACG+
User Manual Int. Approved
4.4 Installation (Fixed)
Mounting details (ACG+ Fixed)
The ACG+ can be mounted as a fixed unit onto a wall or vertical flat surface using the
enclosure mounting holes. The minimum area required and the mounting centres for
installing the ACG+ are shown in Figure 2.
An optional mounting kit to install on either a solid wall or instrument panel can be
purchased.
Commercial in Confidence
ACG+
User Manual Int. Approved
Fit an M6 full nut to the screw thread at the back of the mounting panel
and fully tighten (alternatively, the screws could be fitted into M6 tapped
holes with the M6 nuts not required).
Electrical connection
The user terminals are located at J3 and J17 on the chassis PCB of the ACG+ as shown in
Figure 5.
Commercial in Confidence
ACG+
User Manual Int. Approved
Table 3 Connector J3 Pin Assignments
Connector J3 pin assignments
Terminal Function Details
No.
1 Solenoid drive (+ve)
Not for user connection
2 Solenoid drive (-ve)
3 Reserved
Not for user connection
4 Reserved
5 +24Vdc
6 Output 1 – switched to 0V Max total load of 500mA is available
7 Output 2 – switched to 0V between the 3 switched outputs
8 Output 3 – switched to 0V
Commercial in Confidence
ACG+
User Manual Int. Approved
4.5 Safety interlock relay (ACG+ Fixed)
The ACG+ has a double pole change-over relay that can be used to control external
equipment in the event of an alarm or fault condition.
Two sets of volt-free relay contacts are provided; a normally-open pair which are held
closed under ideal conditions, failing safe to their open state when a non-ideal state is
identified and a normally-closed pair which are held open under ideal conditions. Figure 6
shows the operational state of the relay under various system conditions.
A typical use of the interlock relay might be to use the normally-open pair to cut power to
a compressor or to close a filling valve on an alarm, fault condition, or in the case of power
loss. Alternatively, the normally-closed pair can be used to activate external, self-powered
alarm equipment such as a beacon.
Figure 7 shows a typical example of the relay used in a system
Commercial in Confidence
ACG+
User Manual Int. Approved
4.6 Safety interlock override key-switch (ACG+ Fixed)
If an ACG+’s safety interlock relay (If fitted) is attached to external equipment, there may
be circumstances where it is not desirable for the ACG+ to disable the equipment attached
under alarm or fault conditions. The ACG+ provides an optional override key-switch,
mounted on the left side of the unit. With the key-switch set to ‘Normal’, the external
equipment will be switched in the event of an alarm or fault conditions. If the key-switch is
set to the ‘Override’ position then the external equipment will not be affected by alarm and
fault conditions. Table 4 shows how the key-switch affects the various system outputs.
Table 4 Relay and 24V Output Operation under Fault and Alarm Conditions
Interlock
Interlock
Key- relay Switched Switched Switched
System relay
switch (normally- 24V 24V 24V
condition (normally-
state closed output 1 output 2 output 3
open pair)
pair)
Open
Normal
ACG+ circuit Closed
Off Off Off
power off Closed circuit
Override
circuit
Ideal
Normal
conditions Closed
or Open circuit Off Off Off
(no alarm circuit
override
or fault)
Inlet alarm
only (flow Normal
Closed
or inlet or Open circuit On On On
circuit
pressure override
alarm)
Open
Sensor Normal
circuit Closed
alarm On On On
Closed circuit
active Override
circuit
Open
System or Normal
circuit Closed
sensor On On On
Override Closed circuit
fault active
circuit
Commercial in Confidence
ACG+
User Manual Int. Approved
4.7 Switched 24V DC outputs (ACG+ Fixed)
The ACG+ has 3 switched 24Vdc outputs that provide power to external equipment is the
event of an alarm condition or system fault. These can be used to power directly
equipment such as, for example, a beacon, an audible alarm, or to energise a relay. They
can also be used as an input signal to an external control system. Table 4 shows how the
switched 24V outputs operate under various system conditions.
Figure 8 shows a typical example of the switched 24V DC outputs used in a system.
Commercial in Confidence
ACG+
User Manual Int. Approved
5 Operating instructions
5.1 Start-up
An ACG+ will automatically start up when power is supplied to it. Initially, the ACG+ will
display a splash screen as shown in Figure 9 which will be displayed for a few seconds
after which the display will go blank for up to 15 seconds. The ACG+ will then show the
main display as described in section 5.4.1
Upon start-up, attached sensors may require a short time to warm up as described in
section 5.9.3
Also, to achieve the accuracy specified, a warm up period of 15 minutes is advised prior to
recording the ACG+ sensors displayed values.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.2 Controls
Figure 10 shows the elements of the ACG+ front panel.
1) Display
2) Sample/Calibration gas inlet
3) Gas outlet for timed sample support (section 5.17)
4) Gas inlet/purge valve
5) External power connection (Either by rechargeable battery pack or external PSU)
6) USB port
7) Up button
8) OK button
9) Down button
10) Mute button
11) Context sensitive buttons (function depends on current context)
Commercial in Confidence
ACG+
User Manual Int. Approved
5.3 Inlet/purge valve
Analox recommend that the purge system is used whenever a new pipe connection is
made to the ACG+, if the inlet line has been open to atmosphere or it has been exposed to
damp gas. The following instructions apply when using the supplied 2m PTFE pipe. If
shorter or longer lengths of pipework are used, the purge period should be adjusted
accordingly, depending on the total pipe length. Only stainless steel or PTFE/FEP pipework
should be used to connect to the ACG+.
Starting with the valve in the “off” position, apply pressure to the inlet (Between 5 and 10
bar gauge) using the supplied Linktech inlet/calibration connector.
Once pressure is applied, the valve should be turned to “Purge” and left for 2 minutes
(Using the supplied 2m PTFE pipe) so that the pipework prior to the ACG+ can be purged
of any stale, damp gas, and any moisture on the pipe walls can be dried out.
Once the purge has completed, the valve can then be turned to the “On” position to allow
the sensors to start analysing the gas sample. This process will ensure the ACG+ sensors
can stabilise as quickly as possible to the gas you wish to measure.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.4 Display
Main display screen
Under normal operation the ACG+ will display its main display screen. Figure 11 explains
the elements of the main screen.
Commercial in Confidence
ACG+
User Manual Int. Approved
System status indicator
The system status indicator shows the current status of the ACG+ system. If any alarm
condition has been identified then the system status indicator will turn red and display the
word ‘Alarm’. If any system fault is identified then the status indicator will turn yellow and
display the word ‘Fault’. If no current alarm or fault conditions have been identified then
the system status indicator will turn green. Figure 12 shows examples of system status
indicator states.
The system indicator also shows system diagnostic information regarding system
parameters of flow and input pressure. If an operational parameter is out of specification,
it shall be indicated in the system parameters information box. See troubleshooting
(section 8) for explanation of system fault codes.
5.5 Menu
The device main menu can be accessed from the main screen by pressing the left context
button (Figure 13).
Commercial in Confidence
ACG+
User Manual Int. Approved
The highlighted menu option can be changed by using the up and down buttons. An arrow
at the top or bottom of a menu indicates that there are more menu options above or below
those currently shown.
To select a menu item, highlight it and press the ‘OK’ button. Depending on the menu item
selected, the selected action will be performed or a sub-menu will be displayed. Figure 14
shows the backlight sub-menu.
Selecting the back option from a sub-menu will return to the previous menu level. The exit
context button can be pressed at any time whilst using the menu to return to the main
screen.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.7 Power sub-menu
You able to reset & put the system into sleep mode via the devices ‘Power’ menu.
When the reset option is selected the device performs a soft reset reloading all
configurations.
When the device is in sleep mode this disables the following:
• Networking capabilities
• Sounder
• Data logging
• Alarms
• Screen
To exit sleep mode you simply press ‘OK’ once, exiting sleep mode shall re-enable all of
the above sleep conditions.
In addition, when an unsuitable flow rate is measured, all sensor values will be marked as
uncertain (grey icon) along with a flashing indication that the flow rate is currently invalid.
The measured flow rate can be viewed on the system information screen as described in
section 5.13.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.9 Sensor readings
Sensor reading update
A live reading from each sensor will be displayed in a sensor window on the main display.
This reading will be updated on the display once per second. If no valid sensor reading is
currently available then three dashes (‘---‘) will be shown instead. This may occur if the
sensor has identified a fault in its operation or on occasions where the sensor reading is
unavailable due to routine processing such as after a calibration adjustment has been
performed.
Sensor status
The sensor status icon is shown to the left of each sensor reading. This icon indicates the
current status of the sensor. The meaning of each sensor icon is explained in Table 5
below.
Sensor warm-up
Under certain circumstances, when a digital sensor attached to the ACG+ cannot display a
valid reading, the display will show that the sensor is in a warm-up state. Warm-up occurs
when the ACG+ is initially powered up and may last 10 to 60 seconds depending on the
sensor type attached. Under certain circumstances the sensor may enter a warm-up state
whilst it processes information such as after a calibration adjustment has been attempted
or when a sensor is re-connected after being disconnected. Whilst in a warm-up state, the
display reading for the sensor is shown as three dashes (‘---‘) and a warm-up processing
animation is displayed (Figure 16).
Commercial in Confidence
ACG+
User Manual Int. Approved
For sensors where only a single high-going alarm set-point is defined, such as carbon
dioxide, the bar will be scaled from zero to an amount above the high-going set-point. This
allows any rising carbon dioxide levels to be seen with respect to the alarm set-point.
The bar will change colour depending on the current sensor status. Where a sensor is in a
non-alarm and non-fault state, the bar graph will be shown green as shown in Figure 17.a
below. Should the sensor value be in an alarm state, the bar will be shown in red, as per
Figure 17.b and Figure 17.c which show a low and high alarm respectively.
In cases where the reading falls outside of the displayed range of the bar graph, the bar
graph indicator will show a directional arrow to indicate the reading is outside of the
displayed region as shown in Figure 19 below.
In cases where a sensor fault has been identified and a reading can still be obtained from
the sensor, the bar graph will be show coloured yellow (Figure 20).
Commercial in Confidence
ACG+
User Manual Int. Approved
5.10 Alarms
Every second, as the sensor reading is updated, each alarm associated with the sensor is
checked against the current reading. For high-going direction alarms, an alarm condition
will be triggered if the sensor reading is greater than the alarm set-point. Conversely, for
low-going direction alarms, an alarm condition will be triggered if the sensor reading is less
than the alarm set-point.
When an alarm condition is identified the horn will immediately sound and the system
status indicator will turn red, showing the word ‘alarm’. In addition the sensor reading will
be displayed with a flashing red status indicator and the reading bar graph will turn red,
showing the reading outside of the alarm threshold. Figure 21 shows an example of a
low-going alarm that has been triggered by a low oxygen reading.
Muting alarms
When an alarm condition is triggered and the horn sounds, the horn can be muted by
pressing the mute button. This will mute the horn on all occasions where the horn is
activated. In addition, when an alarm is muted, the sensor status indicator on the display
will stop flashing and will instead be shown in solid red to indicate that the alarm has been
acknowledged.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.11 Clearing alarms and hysteresis
To avoid nuisance re-triggering of alarms when a sensor reading is fluctuating around an
alarm set-point, each alarm has a hysteresis band applied to it. This hysteresis applies to
the clearing of an already triggered alarm. An alarm will always trigger immediately when
the sensor reading passes the set-point in the correct direction. However, when the sensor
returns in the opposite direction through the same set-point, it must go an amount beyond
the set-point before the alarm will clear. This extra amount required to clear an alarm is
the alarm’s hysteresis. Figure 23 shows an example of a hysteresis band on a high going
alarm.
The hysteresis applied to an alarm is adjusted automatically based on the value of the
alarm set-point. Hysteresis for each sensor range is defined by a percentage of set-point
and a maximum hysteresis value. The hysteresis applied to the alarm is the smaller of the
two values for a particular set-point. Table 6 shows the hysteresis applied to each sensor
type.
Commercial in Confidence
ACG+
User Manual Int. Approved
Latching alarms
Alarms can operate in either latching or non-latching mode.
With latching alarms disabled, if an alarm is triggered and the sensor reading
subsequently returns to a level that would not trigger the alarm, the visual and audio
indicators will be cancelled as soon as the reading passes the hysteresis point (see section
5.4.2)
With latching alarms enabled, if an alarm is triggered and the sensor reading
subsequently returns to a level that would not trigger the alarm, the visual and audio
indicators will continue until the mute button is pressed to acknowledge the alarm. Once
acknowledged, the visual and audio indicators will be cancelled. Whilst latching is
enabled, if an alarm is acknowledged by pressing the mute button prior to the sensor
reading passing back through the alarm set-point then the visual and audio indicators will
be cancelled as soon as the reading passes the hysteresis point (see section 5.4.2).
The alarm latching option can be turned on and off through the device menu. This can be
found in the main menu under ‘System options>Alarm latching’. The latching alarm option
can also be changed using the PC tool (see section 6.3).
Commercial in Confidence
ACG+
User Manual Int. Approved
Adjusting alarm set-points
Each of the ACG+’s alarm set-points can be individually adjusted to suit the monitoring
application. To adjust a set-point, first select the alarm using the ‘Sensor options’ sub-
menu from the main menu (Figure 24). Select the sensor of interest and select ‘Adjust
alarms’. A list of the available alarms will be shown in the next menu. Select which alarm
to adjust and the alarm adjustment screen will be shown.
Commercial in Confidence
ACG+
User Manual Int. Approved
Using the up and down adjustment keys, adjust the alarm set-point to the required value
(Figure 25). Once the desired new set-point is shown, press ’Accept’ to store the new
value. Press ‘Cancel’ to leave the alarm set-point unchanged.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.12 Calibration adjustment
Whilst in use, an ACG+’s sensors should be periodically calibrated at intervals deemed
necessary for the monitoring application, by exposing the sensor to gas of a known
concentration. Due to effects such as depletion of electro-chemical cells over time, sensor
readings shown when they are exposed to a gas of known concentration may differ from
the expected value. To compensate for these effects, calibration adjustments of the
sensors can be made where necessary as follows.
Each sensor’s output is defined by two calibration points, one low and one high. To
perform a successful calibration adjustment, both low and high calibration points should be
adjusted.
For calibration of standard ACG+ sensor parts, the following gas mixes are recommended.
Factory configurations for each ACG+ contain the calibration information for these gas
mixes which can be used for ease of repeat calibration. These values can be modified if
alternative mixes are used.
Commercial in Confidence
ACG+
User Manual Int. Approved
Performing an adjustment
Commercial in Confidence
ACG+
User Manual Int. Approved
To perform a calibration adjustment, first expose the sensors to the appropriate calibration
gas of known concentration via the sample/calibration gas inlet as follows:
Fit the 0.5 LPM gas flow regulator accessory (see section 9) or suitable
regulator to the calibration gas bottle (Figure 26).
Commercial in Confidence
ACG+
User Manual Int. Approved
Open the valve on the flow regulator attached to the bottle and then open
the inlet valve on the front panel. The ACG+ should register that there is
flow to the sensors.
• The values on the left show the current sensor readings. (Note that it is possible
for sensors to show negative readings when exposed to zero gas in this situation.
This is likely when attempting to measure values close to zero).
• The calibration bottle details are shown on the right. (If the values shown are not
correct for the bottle attached, the values can be changed. See section 5.12.3).
• The calibration details also show the type of calibration (low or high) which will be
performed for the selected bottle.
• Calibration details will not be shown for all sensors. Certain sensors that cannot be
user adjusted are excluded such as the H 2O sensor.
• If not all sensors shown are to be adjusted, sensors can be excluded from the
adjustment (see section 5.12.2).
Commercial in Confidence
ACG+
User Manual Int. Approved
• For each sensor marked for adjustment, a rotating progress symbol will be shown
whilst each sensor works out whether the adjustment is valid.
• After a short delay the result of the adjustment will be shown (Figure 29). If the
adjustment was successful, a green icon will be shown. If the adjustment was not
successful, a red icon will be shown. For troubleshooting of failed adjustments see
section 8.1.3
• After any calibration adjustment request is processed, the sensors may enter a
warm-up state for a few seconds during which time the sensor reading will not be
available.
Allow the sensors 1 minute to stabilise after calibration and check that the
readings shown are reasonable for the calibration gas applied.
To finish, close the valve on the bottle flow regulator and close the inlet
valve on the front panel then disconnect the calibration connector from
the ACG+ by pressing the release catch and gently removing.
Repeat for a high calibration as necessary. (Note that both low and high
calibration adjustment should be performed in order to reset calibration
due notifications).
Commercial in Confidence
ACG+
User Manual Int. Approved
Excluding sensors from calibration adjustment
When performing an adjustment, individual sensors may be excluded so that only certain
sensors are adjusted.
Commercial in Confidence
ACG+
User Manual Int. Approved
A calibration adjustment reminder will only be cleared once both a high and low calibration
adjustment have been performed. Both adjustments should be performed with as little
time between as possible. If not performed together, the calibration adjustment reminder
will be shown based on the time of the earlier of the two adjustments.
The sensor calibration due date can be viewed on the sensor information screen (section
5.14)
Commercial in Confidence
ACG+
User Manual Int. Approved
5.13 Viewing device information
For ease of access, the device information screen can be accessed from the main sensor
readings screen by using the right hand context button. Alternatively, to view information
about the system, select ‘System information’ from the main menu. This will display the
device information screen (Figure 33 Device Information Screen
The system information page shows the serial number of the device, information about the
software version and the current date and time. Readings are also shown for the current
(absolute) inlet pressure, the absolute atmospheric pressure and the flow rate of sample
gas to the sensors.
The sensor reading screen can be accessed from the device information screen by pressing
the right hand context button. This allows for quickly toggling between the two screens
when taking measurements.
Commercial in Confidence
ACG+
User Manual Int. Approved
From the sensor information screen, information about the next sensor in the list can be
easily displayed by pressing the ‘Next sensor’ context button.
5.15 Data-logging
If the data-logging feature is enabled on an ACG+ then it will maintain a log of system and
sensor data whenever the device is powered up. Data is logged every 10 seconds and
includes the system status, average, minimum and maximum readings for the 10 second
period for each sensor. An ACG+ operates a 90 day rolling log. Once 90 days’ worth of
data is collected, the oldest day’s data will be deleted so only the most recent 90 days are
recorded.
Data can be downloaded into daily log files in .CSV (comma-separated values) format for
easy import into popular spreadsheet applications. An example of a typical data log file is
shown below. (Note that only the first sensor data is shown here. Average, minimum and
maximum readings and units for each sensor will be logged).
Atm O2 CO2 CO
Flow pressure Temperature Average Average Average H2O Average
Date Time (l/m) (mbarA) (C) (%) (ppm) (ppm) (mg/m3)
15/09/2015 11:25:12 0 983 25.1 19.1 733 0 12120
15/09/2015 11:25:22 0 983 25.1 19.1 740 0 12119
15/09/2015 11:25:32 0 983 25.2 19.1 741 0 12118
15/09/2015 11:25:42 0 983 25.2 19.1 735 0 12114
15/09/2015 11:25:52 0 983 25.2 19.1 735 0 12102
15/09/2015 11:26:02 0 983 25.2 19.1 735 0 12063
Retrieving data-logs
There are two methods of downloading data-logs. To view all of the log files which are
available on the system, download and delete specific logs, the ACG+ PC Software can be
used as described in section 6.4.
Alternatively, for convenience, the full set of data-log files can be downloaded from the
device, directly onto a USB memory stick via the USB port on the front panel. To download
the data logs, prepare a USB memory stick, ensuring that there is enough space free to
download logs. Each daily log file is typically 1.5MB in size, with a full 90 days of logs
taking up approximately 135MB.
Once the USB memory stick is ready for downloading logs, unscrew the
protective cover from the USB socket on the front panel of the ACG+ and
insert the USB memory stick into the USB socket.
Commercial in Confidence
ACG+
User Manual Int. Approved
From the main menu, enter the ‘Data-logging’ sub-menu and select
‘Download data to USB’. The data-log download screen will be shown as
per Figure 35.
This will indicate how many logs are available to be downloaded. If there is insufficient
space on the USB memory stick to download all log files then an error will be shown.
To begin the download, press the ‘Start download’ context button. The
progress of the download will be shown on screen as per Figure 36. The
download may take several minutes depending on the number of files
currently stored.
The data-log download screen will indicate that the download is complete once all data-
logs have been downloaded.
The download can be cancelled at any time by pressing the ‘Cancel’ context button. It may
take a few seconds to cancel the download once requested. The USB device should remain
connected until the display shows that the download has been successfully cancelled.
When a download is cancelled, all full data-log files that have been downloaded up until the
point of cancellation will remain on the USB memory stick.
Each time a log download is performed in this way, all files will be downloaded. To avoid
having to download all data-logs each time, an option to delete all current data-logs exists
so that after a download, the logs may be cleared. To delete all log files, select ‘Data-
Commercial in Confidence
ACG+
User Manual Int. Approved
logging’ from the main menu and then select ‘Delete all data-logs’. Confirmation is
required before the logs will be deleted. Note that this operation is permanent and logs
that are deleted cannot be retrieved at a later date.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.16 Sensor replacement reminder
Certain ACG+ sensors have a requirement to be periodically replaced or have elements of
the sensor replaced, for example, to replace a depleted electrochemical cell or to return for
factory calibration. To aid with the timely replacement and/or maintenance of sensors the
ACG+ provides a sensor replacement reminder feature. This feature is only effective for
sensors which require periodic replacement and/or maintenance.
If the sensor is replaceable, the ACG+ will show an onscreen warning when the sensor/cell
is due to be replaced as shown in Figure 37.
The sensor/cell replacement reminder is shown once the number of days that have elapsed
since the cell was last replaced exceeds the sensor/cell replacement reminder period. The
sensor replacement due date can be viewed on the sensor information screen (5.14)
Commercial in Confidence
ACG+
User Manual Int. Approved
The ACG+ comes fitted with a fixed restriction prior to the sample outlet port suitable for
oil mist detection using a Drager Impactor. Another restrictor kit is supplied which can be
quickly fitted to provide suitable flows, for example for an Oilmist colorimetric tube (0.1l/m
sample for 3 minutes).
Remove the Impactor restrictor tube from the outlet of the solenoid valve
and the rear of the sample support bulkhead.
Commercial in Confidence
ACG+
User Manual Int. Approved
Remove the 8mm to 6mm push reducer from the Impactor restrictor tube
and fit it to the 0.1l/m restrictor tube.
Refit the 0.1l/m restrictor tube to the outlet of the solenoid valve and the
rear of the sample support bulkhead as shown.
A flow meter kit (1 to 10l/m) is also available as an accessory (P0149-611) which can be
fitted to the sample gas outlet port to provide a suitable flow for other sampling
techniques.
Commercial in Confidence
ACG+
User Manual Int. Approved
Attach the target gas to the sample inlet and the measurement device or
sample pipe to the gas outlet port on the ACG+ front panel (see Figure
39).
The protection label on the Dräger Impactor should not be removed until
after the timed sample has been completed.
Figure 39 A Dräger Impactor Attached to the Outlet Port for Oil Mist Testing
Commercial in Confidence
ACG+
User Manual Int. Approved
From the main menu select ‘Timed sample’ then select ‘Start new sample’.
The timed sample setup screen will be shown (Figure 40)
Throughout the sample period, the time remaining on the sample timer will be shown on
the screen (Figure 41).
When the sample is complete, the ACG+ will give three beeps and the sample valve will
close. The main screen shall show that the sample is complete (Figure 41).
A timed sample can be cancelled at any time whilst it is in progress. To cancel a sample,
select ‘Timed sample’ from the main menu and select ‘Cancel sample’.
For certain sampling applications, it may be necessary to purge the sample line prior to
beginning a sample. The simplest way to achieve this is to use the purge valve as
described in section 0.
Commercial in Confidence
ACG+
User Manual Int. Approved
5.18 Setting the date and time
The current date and time are stored on the device. This allows the ACG+ to log data with
a date and time stamp and also allows calibration and sensor replacement reminders to be
given. The date and time can be set or changed through the main menu under ‘System
options>Set date and time’.
Shows the date and time adjustment screen. To adjust the highlighted field, press the up
and down buttons. To highlight the next field, press the set button. Once the correct date
and time is shown, press the ‘Accept’ context button to save the date and time.
Alternatively, the date and time can be set using the PC tool (see section 6.5).
Commercial in Confidence
ACG+
User Manual Int. Approved
6 Configuration
In order to set options on an ACG+, the ACG+ software is provided to allow a device to be
configured using a USB connection. The ACG software is designed to run on Microsoft
Windows®.
Software requirements
The ACG+ software is designed to run on Microsoft Windows® and has the following
prerequisites:
Commercial in Confidence
ACG+
User Manual Int. Approved
Windows 10 Installation
®
Unplug all unnecessary USB devices from the computer keeping the ACG+
plugged in.
From the ACG+ software disk, run the installer file called ‘Setup’.
Keep the installation folder as default, then select Everyone and click
Next.
Commercial in Confidence
ACG+
User Manual Int. Approved
Click Next again to install the software (A warning may show, click Yes to
accept).
Commercial in Confidence
ACG+
User Manual Int. Approved
Running the software
To run the ACG+ software, select the program from the Windows® start menu, Analox
folder. When the software runs without a device connected, the window shown in Figure
43.
Always connect only one ACG+ to the computer at a time. To connect an ACG+ to the
computer, first open the hinged front door of the ACG+ device by loosening the four corner
screws. Locate the ACG+ display module which is mounted to the inside of the ACG+
hinged front door. Connect one end of the USB cable provided to the right hand USB port
(labelled ‘PC’) on the rear panel and connect the other end of the cable to a free USB port
on the computer. Power the ACG+ device and allow it to start up.
Commercial in Confidence
ACG+
User Manual Int. Approved
Once connected and powered, Windows should detect the device and display an
®
information bubble confirming the connection. On the first time connecting an ACG+ to the
computer, message bubbles may be shown explaining that device drivers are being
installed. This should take a few moments.
When the device connects, on Windows® 10 systems, the Windows Mobile Device Center
window will be shown. This window should be closed.
If not already open, run the ACG+ software. The ACG+ software will recognise the device
and read the current settings form the device. Once fully connected the window shown
Figure 45 will be displayed.
Connection troubleshooting
Disconnecting
To disconnect an ACG+ from the computer, first wait until all settings have been saved
correctly and that any file downloads have completed then remove the USB cable from the
rear of the ACG+. If the USB cable was removed from the ‘Tech’ USB port, ensure that it is
re-attached before closing and securing the ACG+ hinged front door.
Commercial in Confidence
ACG+
User Manual Int. Approved
6.3 Changing options
On connection to an ACG+, a copy of all of the device’s options is created in the ACG+
software. The options are presented over a set of tabs. To view the option tabs, click the
‘Options’ button. Using these tabs the options can be viewed and modified as required. No
changes are made to the device’s options until the changes are applied to the device. To
apply any changes made click the ‘Apply Changes’ button.). This will upload the new
options to the device. The ACG+ will confirm that the options have been saved by playing
a confirmation sound.
Any changes made to options can be discarded as long as they have not yet been applied
to the device. To re-download the current device settings from an ACG+, click the ‘Read
Options’ button.
General options
The general device options are displayed by clicking on the ‘General’ tab. As shown in
Figure 45.
Device information
This shows the serial number and the software version of the connected device.
Latching alarms
The latching alarms checkbox enables or disables latching alarms as described in section
5.11.1.
Commercial in Confidence
ACG+
User Manual Int. Approved
6.4 Downloading data-logs
The data stored on an ACG+ can be downloaded using the ACG software. To view the
data-log management controls, click the ‘Logs’ button (Figure 45). A list of the data-log
files available on the device will be shown. The data-log files are stored by date. Each
date’s file contains all of the data logged whilst the device was running during a particular
day. This list can be refreshed by pressing the ‘Refresh’ button.
Data-log files can be downloaded individually or in batches. To select the files that are to
be downloaded, click on the checkbox next to each of the files of interest. The ‘All’ and
‘None’ buttons can be used to select or deselect all of the files respectively. Once a
selection has been made, the log file download can be started by clicking the ‘Download’
button.
Commercial in Confidence
ACG+
User Manual Int. Approved
Although the internal data-logging operates a rolling log with automatic clean-up of older
data, the data-logs on the device can be managed manually using the ACG software. To
delete unwanted files from a device, select the files for deletion and click the ‘Delete’
button.
Commercial in Confidence
ACG+
User Manual Int. Approved
7 Networking
7.1 Overview
This section describes the Ethernet Networking Communications and the Sensor Datagram
of the ACG+. The ACG+ software communicates using two network protocols; TCP/IP and
UDP/IP. The same Datagram is used for both protocols.
7.2 Connection
For connection to the display module, a suitable networking cable should be fitted firstly
through the M16 x 1.5 EMC gland in the side of the instrument, then the cable should be
fitted with a suitable RJ45 connector ready for connection to the display module.
Connection should be made to the display module by plugging in the made up network
cable to the ‘NETWORK’ port on the rear of the display module as shown.
If your device supports Auto MDI-X capability, wherein a patch cable can be used in place
of a crossover cable, or vice versa, and the receive and transmit signals are reconfigured
automatically to yield the expected result.
If your device does not support Auto MDI-X capability, a cross over cable will be required
to connect the ACG+ to your system.
Although the XML document is listed as “UTF-16” the actual data is encoded into standard
“UTF-8” (ASCII) during physical transmission.
Commercial in Confidence
ACG+
User Manual Int. Approved
7.4 Setting the ACG Static IP Address
The ACG can be configured to run a TCP/IP server on a specified IP Address and port,
listening for client connections. The IP Address can be set in the ACG menu:
Menu -> System Options -> Server IP Settings -> Assign Static IP.
The ACG TCP/IP server uses Port 5021 for TCP/IP communications.
The client must use an IP address on the same subnet as the ACG.
For example, if the ACG is set to 10.0.0.2, the client can use 10.0.0.1 or (10.0.0.3 to
10.0.0.254).
Set the client’s subnet mask to 255.255.255.0.
The TCP/IP request “RemClient”. (case sensitive) is not required on the ACG since the
connection is closed after each datagram request.
Commercial in Confidence
ACG+
User Manual Int. Approved
7.6 Retrieving Data using UDP
The ACG also transmits the CAGASensorDatagram every 1 second using the UDP protocol.
The datagram format is the same as for TCP/IP.
The ACG transmits the UDP datagram to the broadcast IP Address (255.255.255.255)
using source and destination ports both set to Port 5020.
To receive the datagram, a client must set up a UDP client listening to port 5020. The UDP
datagram is received by the client as one block, so there is no need to assemble message
blocks and detect end of message.
<CAGASensorDatagram>
<Sensors> (CAGASensorData)
Holds the list of sensor data and overall fault and alarm flags
<SensorData>
The sensor data, status, faults and alarms for an individual sensor
Commercial in Confidence
ACG+
User Manual Int. Approved
Name Data Type Description Example
<DisplayDecimalPlaces> String Number of “One”, “Two”
decimal places
the ACG will
display
<MaxRange> Float The highest valid 5.0
reading of the
sensor before
setting a range
fault
<MinRange> Float The lowest valid 1.0
reading of the
sensor before
setting a range
fault
<WorkingReading> Float Current reading 123.456
from the sensor
updated at least
once a second in
the units specified
in
<WorkingUnits>
<Status> SensorStatus. Status flags for
See description the sensor
below.
(End of base sensor data)
<SensorMeasurand> String Item being “Helium”,
measured by the “Oxygen”
sensor
<WorkingUnits> String The units used for “mBar”,”ppm”
the working
reading.
<IsReplaceableable> Bool Set to true if the “True”, ”False”
sensor can be
replaced.
<IsCalibratable> Bool Set to true if the “True”, ”False”
sensor can be
calibrated.
<StatusFlags> Hex Hex 0
representation of
the status flags
for the attached
sensor.
<PressureStatusFlags> Hex Hex 0
representation of
the status flags
for the attached
pressure sensor.
<CalibrationIntervalSeconds> Int The number of 7776000
seconds which
can elapse since
the last sensor
calibration until
the calibration
due warning can
be displayed.
Commercial in Confidence
ACG+
User Manual Int. Approved
Name Data Type Description Example
<CellReplacementIntervalDays> Int The number of 365
days which can
elapse since the
last cell
replacement
before a cell
replacement
warning will be
displayed.
<SerialNumber> String Serial Number of “MEC-123456”
the sensor
<PartNumber> String The Analox part “MECHEMBAP”
number of the
sensor
<Alarms> (AlarmManager)
Holds the alarm list and a status flag
<Status> (SensorStatus)
Status data and flags for a sensor
Commercial in Confidence
ACG+
User Manual Int. Approved
Name Data Type Description Example
<GeneralFaultState> <State> Set in
combination with
other fault flags
as a general fault
flag.
<FatalErrorState> <State> Set if the sensor
has triggered a
fatal error.
<SensorRangeFaultState> <State> Set if the sensors
reading is outside
the acceptable
range.
<MonitorUnderRangeFaultState> <State> Set if the sensors
reading is outside
the minimum and
maximum range
set on the ACG.
<PowerFaultState> <State> Set if the sensors
power is outside
the acceptable
range.
<LampFaultState> <State> Set if a problem
has occurred with
the lamp on the
sensor.
<PidFaultState> <State> Set if a PID error
has occurred on
the sensor.
<CrcFaultState> <State> Set if the CRC of
the program
memory on the
sensor has
become corrupt.
<TemperatureFaultState> <State> Set if the
temperature on
the sensor is
outside of
acceptable limits.
<CalibrationFaultState> <State> Set if the unit has
failed to calibrate.
<IsInvalidFlow> Bool Set to true if the “True”,
flow is outside ”False”
acceptable limits.
<IsInvalidTempAndPressure> Bool True if “True”,
temperature and ”False”
pressure are
outside
acceptable limits.
<IsWarmup> Bool True if the sensor “True”,
is currently ”False”
warming up.
<IsBusy> Bool True if the sensor “True”,
is busy ”False”
<IsCalibrationDue> Bool True if the sensor “True”,
is due to be ”False”
calibrated
<IsChangeDue> Bool True if the sensor “True”,
is due to be ”False”
changed.
<IsReadingValid> Bool True if the “True”,
reading can be ”False”
considered valid.
Commercial in Confidence
ACG+
User Manual Int. Approved
Name Data Type Description Example
<CalDueDateTime> Date/Time The date and time 2014-05-
for the next 30T23:59:59
calibration
<ChangeDueDate> Date/Time The date and time 2014-05-
that the cell is due 30T23:59:59
for replacement
<Flow>
<Pressure>
<InterfaceBoard>
Commercial in Confidence
ACG+
User Manual Int. Approved
Name Data Type Description Example
<VersionString> String Software version “1.0”
number of the
Interface Board
<State>
Commercial in Confidence
ACG+
User Manual Int. Approved
8 Troubleshooting
8.1 System faults
An instrument constantly monitors itself for fault conditions. Any fault conditions that are
identified will trigger a fault alarm. When a fault alarm is raised, the horn will sound
immediately. Faults will be indicated on the display with the status bar coloured yellow as
shown in Figure 49. Fault alarms can be muted in the same way as standard alarms by
pressing the mute button.
All fault alarms are latching so if an intermittent fault is identified and subsequently
cleared, the alarm will continue to be displayed and sounded until the mute button is
pressed to acknowledge the fault.
Information about the fault condition identified is displayed alternating with the sensor
reading if valid as shown in Figure 49 in which a sensor communication fault is shown. In
cases where a fault prevents a valid reading from being obtained from a sensor then the
reading will not be shown on the display and will instead be replaced by ‘---‘ interleaved
with the fault code.
Commercial in Confidence
ACG+
User Manual Int. Approved
System fault conditions
The following table explains the possible system fault conditions that may be displayed in
the status bar:
Commercial in Confidence
ACG+
User Manual Int. Approved
Sensor fault conditions
The following table explains the possible sensor specific fault conditions that may be
displayed for each sensor:
An inappropriate concentration of gas has been applied for the calibration point. For
example, a full scale gas has been applied when attempting to perform a low calibration
adjustment.
The concentration of gas specified by the user does not match the concentration of the
calibration gas.
The amount by which the calibration value differs from the current sensor reading is too
great.
The electrochemical cell (oxygen or carbon monoxide) has depleted over time. Its output
may have dropped too far to perform an adequate adjustment. This can be an indicator
that the cell should be replaced, even if the cell replacement warning is not yet shown.
The PID detector (VOC sensor) may have become overly dirty due to exposure to oil or
other contaminants.
Check that the gas value being entered for the adjustment corresponds to the
concentration given on the calibration bottle.
Commercial in Confidence
ACG+
User Manual Int. Approved
Check that an appropriate gas concentration is being applied for the type of calibration
(low calibration close to minimum sensor range and high calibration close to maximum
sensor range).
Check that there is no flow warning currently being shown. See Table 9 for information
about flow faults.
Check that there are no other sensor faults currently shown.
For O2 or CO sensor, replace the sensor’s electrochemical cell (section 10.1).
For VOC sensor, clean or replace the PID lamp component (section 10.1).
If the problem persists, please contact your supplier for assistance.
Diagnostic logging
Connection troubleshooting
ActiveSync Drivers
The PC tool requires Windows Mobile Device Center to connect to the ACG unit. Once
installed, the following guide can be used to resolve connection issues for Windows 10.
• Open the start menu and type in Services followed by 'Enter' to launch the
Services Console.
• On the Log On tab, enable the Local System Account option and enable the Allow
service to interact with desktop option.
• On the Recovery Tab, set First failure, Second failure, and Subsequent failures to
Restart the Service.
• Click on Apply and you will be notified this setting won't take effect until the
service is restarted.
Commercial in Confidence
ACG+
User Manual Int. Approved
• Do not restart the service yet as you will also need to update the below service
first. Proceed to step 2 below.
• Double click the Windows Mobile-based device connectivity service to open the
settings for it.
• On the Log On tab, enable the Local System Account option and enable the Allow
service to interact with desktop option.
• On the Recovery Tab, set First failure, Second failure, and Subsequent failures to
Restart the Service.
• Click on Apply and you will be notified this setting won't take effect until the
service is restarted.
• Right-click this service and choose Restart. This will restart the service with the
new changes.
• Two Reg keys need to be added. To do so open a Command prompt as admin and
run the following commands:
Commercial in Confidence
ACG+
User Manual Int. Approved
9 Recommended spares and accessories
Commercial in Confidence
ACG+
User Manual Int. Approved
Table 12 List of Accessories
Item Description Part no.
1 External PSU, 24VDC, 1.5A, 36W for connection to ACGPEXTPSU
front panel via panel mount connector (For use
with Portable & Fixed variants)
2 External PSU, 24VDC, 2.7A, 65W for connection ACGEXTPSU
through gland direct onto interface PCB (For use
with fixed variants)
3 Li-Ion rechargeable battery pack kit, universal P0149-605LU
(Includes charger and battery casing – for
portable variants)
4 Spare Li-Ion rechargeable battery pack (Includes P0149-605LS
battery casing lid – for portable variants)
5 NiMH rechargeable battery pack kit, universal P0149-605NU
(Includes charger and battery casing – for
portable variants)
6 Spare NiMH rechargeable battery pack (Includes P0149-605NS
battery casing lid – for portable variants)
7 Inlet/Calibration adaptor 6000-0201
8 0.5LPM 110L bottle regulator Contact Analox or
source locally
9 Zero calibration gas (not including CO Sensor) – Contact Analox or
8% oxygen, balance nitrogen source locally
10 Zero calibration gas (For CO Sensor only) – Contact Analox or
20.9% oxygen balance nitrogen. source locally
Oxygen only span calibration gas (For part
number ACGPABXXXXX)
11 Nitrogen background span calibration gas, (18ppm Contact Analox or
carbon monoxide, 18ppm isobutylene, 20.9% source locally
oxygen, 900ppm carbon dioxide, balance
nitrogen)
12 Carbon dioxide only span calibration gas (For part Contact Analox or
number ACGPABXXXXX) – 9000ppm carbon source locally
dioxide, balance nitrogen
13 Pressure regulator, compact piston sensed, 300bar 6000-0164
max inlet, 0 to 7bar adjustable outlet (Suitable
1/8“ NPT fittings required for inlet and outlet),
complete with pressure relief valve set at 7.5bar
14 Combined beacon/sounder, 24V DC, IP65 (For 2562-0008
fixed variant)
15 30mm open ended spanner 9300-1020
16 Gastec Airtec Tube (Oil Mist) 109A, Pk10 9100-3500
17 Rubber Tube Holder for colorimetric tubes GAS357
18 8mm to 4mm Push fit reducer (Required for item 6000-0165
17)
19 Drager Oil Impactor, Pk10 9100-3501
20 Override Keyswitch & Gland Kit P0149-610
21 Humidifier Kit P0149-660
22 External flow meter sample support kit P0149-611
23 VOC electrode stack removal tool 9300-1021
24 VOC lamp cleaning kit 9300-1022
25 ACG+ Fixing Kit P0075-615
Commercial in Confidence
ACG+
User Manual Int. Approved
9.1 Service requirements
The following tables specify the description of the service parts and their associated part
numbers for the 6 monthly, 12 monthly and 24 monthly service requirements.
Commercial in Confidence
ACG+
User Manual Int. Approved
10 Maintenance requirements
Certain elements of the ACG+ system require regular, preventative maintenance to ensure
optimum operation. Table 13 shows the maintenance that is required and the scheduled
interval for performing the maintenance.
Commercial in Confidence
ACG+
User Manual Int. Approved
Oxygen cell replacement
The part number for your sensor’s replacement cell is: 9100-9212-9HM
When a replacement cell is received check that the cell has not leaked. The cells are
themselves sealed and do not under normal circumstances present a health hazard
however if leakage of the electrolyte has occurred use rubber gloves and wear chemical
splash goggles to handle and clean up. Rinse contaminated surfaces with water. If contact
is made with the electrolyte, please refer to Section 1.
Loosen the corner screws of the enclosure and remove the lid.
Commercial in Confidence
ACG+
User Manual Int. Approved
Using a terminal screwdriver push down the clamp release button to
release the clamp and gently pull the cell wire from the clamp. Repeat this
for both the red and black wires.
Unscrew the green locking ring from the chimney of the cell housing and
remove from the enclosure lid.
Fit the new cell in to the enclosure lid and secure in place using the
locking ring.
Commercial in Confidence
ACG+
User Manual Int. Approved
Using a terminal screwdriver, push down the clamp release button to
release the clamp. Push the ferrule of the cell wire in to the clamp and
remove the screwdriver from the clamp release button. Gently pull on the
wire to ensure that the wire is held by the clamp. Repeat this for both the
red and black wires, taking note of the correct polarity as shown below.
Red Black
Refit the enclosure lid in place and secure in place using the corner
screws.
When a replacement cell is received check that the cell has not leaked. The cells are
themselves sealed and do not under normal circumstances present a health hazard
however if leakage of the electrolyte has occurred use rubber gloves and wear chemical
splash goggles to handle and clean up. Rinse contaminated surfaces with water. If contact
is made with the electrolyte, please refer to Section 1.
Commercial in Confidence
ACG+
User Manual Int. Approved
Use a flat blade screwdriver to pop out the cell holder from the sensor
bulkhead fitting.
Rotate the cell PCB connector clockwise to release from the cell holder.
Commercial in Confidence
ACG+
User Manual Int. Approved
Connect the new cell to the PCB connector (Note that it will only connect
in one orientation).
Fit the cell in to the housing and rotate anti-clockwise to lock in place.
Refit the cell holder into the sensor bulkhead.
Refit the flow adaptor into the sensor inlet.
Commercial in Confidence
ACG+
User Manual Int. Approved
Use a terminal screwdriver to pop out the sensor holder from the sensor
bulkhead fitting.
Commercial in Confidence
ACG+
User Manual Int. Approved
Remove the lid from the CO sensor housing and feed the cell PCB through
the bulkhead.
Disconnect the damaged cell PCB wiring loom from the MEC PCB
connector.
Connect the sensor wires of the new cell PCB to the corresponding ‘Red’,
‘Black’ and yellow terminals of connector J3.
Commercial in Confidence
ACG+
User Manual Int. Approved
Feed the cell PCB through the bulkhead and refit the lid.
Connect the CO cell to the new PCB connector. Note that it will only
connect in one orientation.
Fit the cell in to the housing and rotate anti-clockwise to lock in place.
Refit the cell holder into the sensor bulkhead.
Refit the flow adaptor into the sensor inlet.
Commercial in Confidence
ACG+
User Manual Int. Approved
VOC maintenance
Maintenance of the VOC sensor can be done by the following methods, clean or replace the
lamp, or replace the electrode stack. Cleaning the lamp should be performed as required if
the sensitivity of the sensor reduces over time. Replacement of the PID lamp should be
performed after 5000 hours of use.
Replacement of the electrode stack should be performed only if the sensor has been
damaged by contamination.
The lamp and electrode stack are both part of the PID component of the sensor. The
cleaning an replacement processes are documented below.
Remove the sensor housing from the MEC and then remove the sensor
from the housing by twisting the PCB on the rear of the sensor, remove
the PCB from the sensor
Commercial in Confidence
ACG+
User Manual Int. Approved
Using the stack removal tool (9300-1021) insert the ends of the tool into
the slots on the sides of the sensor as shown, keeping a finger on top of
the stack so as it does not spring away from the sensor and so that the
spring and lamp cannot pop out upon removal.
Carefully remove the lamp from the sensor, making sure that the window
of the lamp is not touched.
Using the lamp cleaning kit (9300-1022) and suitable protective clothing,
open the container of cleaning compound, with a clean cotton bud, apply a
small amount to one end.
Using a circular action, clean the lamp by applying light pressure to the
lamp window (do not touch the lamp window with fingers), continue
polishing until a light squeaking can be heard.
Commercial in Confidence
ACG+
User Manual Int. Approved
Remove any residual cleaning compound from the lamp using the clean
end of the cotton bud.
Ensure that the lamp is completely free from any signs of contamination
prior to refitting.
Carefully place the lamp into the rear recess of the electrode stack, this
will be held by the O-ring seal in the electrode pellet.
Take the electrode stack and the mated lamp and insert back into the
sensor housing, push home until the electrode stack snaps into place, the
top of the electrode stack should be flush with the top of the sensor
housing.
Commercial in Confidence
ACG+
User Manual Int. Approved
Disconnect the cable from the end of the water vapour sensor.
Using a 30mm spanner (part number 9300-1020), loosen the sensor from
its housing and unscrew the sensor fully from its mount. Take care when
performing this step as the sensor head is delicate.
Unscrew the protective cap from the new sensor and place it over the
sensing head of the removed sensor.
Dowty
washer
Ensure that the dowty washer is fitted to the new sensor as shown and
insert the new sensor into the sensor housing, taking care not to damage
the sensor head. Screw the sensor in to its mount and fully tighten using a
30mm spanner
Commercial in Confidence
ACG+
User Manual Int. Approved
Re-connect the cable to the new sensor, into the port labelled 'II',
checking for correct orientation. Push the connector in full to ensure it is
secure.
Return the old sensor to your supplier for maintenance, ensuring that the
sensor head is securely covered by the protective cap.
Remove filter from clip and remove the elbows from the filter.
Commercial in Confidence
ACG+
User Manual Int. Approved
Unscrew the filter from both of the male and female luers attached to the
6mm pipework from the inlet manifold.
Before attempting to replace any sensors ensure that the ACG+ is switched off and the gas
supply has been isolated. Using a Pozi-drive screwdriver loosen the 4 enclosure lid screws
and open the lid.
Commercial in Confidence
ACG+
User Manual Int. Approved
To replace the PID component, remove the flow adaptor from the sensor
inlet.
Commercial in Confidence
ACG+
User Manual Int. Approved
Use a flat blade screwdriver to pop out the PID holder from the sensor
bulkhead fitting.
Rotate the PID PCB connector clockwise to release from the PID holder.
Connect the new PID to the PCB connector. Note that it will only connect
in one orientation.
Commercial in Confidence
ACG+
User Manual Int. Approved
Fit the PID in to the housing and rotate anti-clockwise to lock in place.
Refit the PID holder in to the sensor bulkhead.
Refit the flow adaptor in to the sensor inlet.
Use a terminal screwdriver to pop out the PID holder from the sensor
bulkhead fitting.
Commercial in Confidence
ACG+
User Manual Int. Approved
Rotate the PID PCB connector clockwise to release from the PID holder.
Remove the lid from the MEC sensor housing and feed the cell PCB
through the bulkhead.
Commercial in Confidence
ACG+
User Manual Int. Approved
Disconnect the damaged cell PCB wiring loom from the MEC PCB
connector.
Connect the sensor wires of the new cell PCB to the corresponding red,
blue and yellow terminals of connector J3.
Commercial in Confidence
ACG+
User Manual Int. Approved
Feed the PCB back through the bulkhead and refit the lid.
Connect the PID to the PCB connector PCB. Note that it will only connect
in one orientation.
Fit the PID into the housing and rotate anti-clockwise to lock in place.
Refit the PID holder into the sensor bulkhead.
Refit the flow adaptor into the sensor inlet.
Commercial in Confidence
ACG+
User Manual Int. Approved
11 Specifications
Analox reserve the right to upgrade, develop or change specifications without prior notice.
Commercial in Confidence
ACG+
User Manual Int. Approved
11.2 Sensor specification
Oxygen (O2)
Range 0 to 50%
Accuracy ±(0.035% O2 + 1% of reading + Temp coefficient)
(over ±10°C
range) Temp coefficient =
0.4% of reading/°C or 0.22% of reading/°F
Detection mode Electrochemical
Expected life 2 years for the electrochemical cell in atmospheric air
Range 0 to 10000ppm
Accuracy ±(50ppm CO2 + 1% of reading + Temp coefficient)
(over ±10°C
range) Temp coefficient =
5ppm CO2/°C or 2.8ppm CO2/°F
Detection mode Infra-red
Expected life 6 years in operation
Commercial in Confidence
ACG+
User Manual Int. Approved
12 UK Declaration of Conformity
Commercial in Confidence
ACG+
User Manual Int. Approved
13 Declaration of conformity
Commercial in Confidence