Brief Operating Instructions


Deltapilot M FMB50, FMB51,
FMB52, FMB53
Hydrostatic level measurement

These Instructions are Brief Operating Instructions; they are not


a substitute for the Operating Instructions pertaining to the
device.
Detailed information about the device can be found in the Ope-
rating Instructions and the other documentation:
Available for all device versions via:
– Internet: www.endress.com/deviceviewer
– Smart phone/tablet: Endress+Hauser Operations App
Deltapilot M PROFIBUS PA

Order code: XXXXX-XXXXXX


1. Ser. no.: XXXXXXXXXXXX
Ext. ord. cd.: XXX.XXXX.XX

Serial number

2. www.endress.com/deviceviewer Endress+Hauser Operations App

3.

A0023555

2 Endress+Hauser
Deltapilot M PROFIBUS PA Table of contents

Table of contents
1 Document information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 Basic safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7


2.1 Requirements concerning the staff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Hazardous area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.6 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3 Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 CE mark, Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.3 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.4 General installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.5 Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.6 Mounting of the profile seal for universal process mounting adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.7 Closing the housing cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.8 Post-installation check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

5 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1 Connecting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.2 Connecting the measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.4 Overvoltage protection (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

6 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.1 Operating options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.2 Operation without an operating menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3 Operation with an operating menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.4 PROFIBUS PA communication protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

7 Commissioning without an operating menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35


7.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.2 Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

8 Commissioning with an operating menu (onsite display/FieldCare) . . . . . . . . . . . . . . . . . . 37


8.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.3 Position zero adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
8.4 Level measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8.5 Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.6 Pressure measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

Endress+Hauser 3
Document information Deltapilot M PROFIBUS PA

1 Document information
1.1 Document function
These Operating Instructions contain all the information that is required in various phases of
the life cycle of the device: from product identification, incoming acceptance and storage, to
mounting, connection, operation and commissioning through to troubleshooting, maintenance
and disposal.

1.2 Symbols used


1.2.1 Safety symbols

Symbol Meaning

DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
seriousor fatal injury.
A0011189-DE

WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
seriousor fatal injury.
A0011190-DE

CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minoror medium injury.
A0011191-DE

NOTICE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personalinjury.
A0011192-DE

1.2.2 Electrical symbols

Symbol Meaning Symbol Meaning

Direct current Alternating current

Direct current and alternating current Ground connection


A grounded terminal which, as far as
) the operator is concerned, is grounded
via a grounding system.

Protective ground connection Equipotential connection


A terminal which must be connected to A connection that has to be connected
ground prior to establishing any other to the plant grounding system: This may
connections. be a potential equalization line or a star
grounding system depending on
national or company codes of practice.

4 Endress+Hauser
Deltapilot M PROFIBUS PA Document information

1.2.3 Tool symbols

Symbol Meaning

Allen key

A0011221

Hexagon wrench

A0011222

1.2.4 Symbols for certain types of information

Symbol Meaning

Permitted
Indicates procedures, processes or actions that are permitted.
A0011182

Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011184

Tip
Indicates additional information.
A0011193

Reference to documentation

A0015482

Reference to page

A0015484

Reference to graphic

A0015487

1. , 2. , ... Series of steps

Result of a sequence of actions

A0018343

Visual inspection

A0015502

Indicates how to navigate to the parameter using the display and operating module

A0015502

Endress+Hauser 5
Document information Deltapilot M PROFIBUS PA

Indicates how to navigate to the parameter using operating tools (e.g. FieldCare)

A0015502

1.2.5 Symbols in graphics

Symbol Meaning

1, 2, 3, 4, ... Item numbers

1. , 2. , ... Series of steps

A, B, C, D, ... Views

1.2.6 Symbols at the device

Symbol Meaning

Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
A0019159

Connecting cable immunity to temperature change


t>85°C Indicates that the connecting cables have to withstand a temperature of 85°C at least.

1.2.7 Registered trademarks


KALREZ®, VITON®, TEFLON®
Registered label of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP®
Registered label of Ladish & Co., Inc., Kenosha, USA
PROFIBUS PA®
Trademark of the PROFIBUS User Organization, Karlsruhe, Germany
GORE-TEX®
Registered label of W.L. Gore & Associates, Inc., USA

6 Endress+Hauser
Deltapilot M PROFIBUS PA Basic safety instructions

2 Basic safety instructions


2.1 Requirements concerning the staff
The personnel for installation, commissioning, diagnostics and maintenance must fulfill the
following requirements:
• Trained, qualified specialists: must have a relevant qualification for this specific function and
task
• Are authorized by the plant owner/operator
• Are familiar with federal/national regulations
• Before beginning work, the specialist staff must have read and understood the instructions in
the Operating Instructions and supplementary documentation as well as in the certificates
(depending on the application)
• Following instructions and basic conditions
The operating personnel must fulfill the following requirements:
• Being instructed and authorized according to the requirements of the task by the facility's
owner-operator
• Following the instructions in these Operating Instructions

2.2 Designated use


The Deltapilot M is a hydrostatic pressure sensor for measuring level and pressure.

2.2.1 Incorrect use


The manufacturer is not liable for damage caused by improper or non-designated use.
Verification for borderline cases:
For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in
verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or
liability.

2.3 Workplace safety


For work on and with the device:
• Wear the required personal protective equipment according to federal/national regulations.
• Switch off the supply voltage before connecting the device.

2.4 Operational safety


Risk of injury!
‣ Operate the device in proper technical condition and fail-safe condition only.
‣ The operator is responsible for interference-free operation of the device.
‣ Only disassemble the device in pressurless condition!

Endress+Hauser 7
Basic safety instructions Deltapilot M PROFIBUS PA

Conversions to the device


Unauthorized modifications to the device are not permitted and can lead to unforeseeable
dangers:
‣ If, despite this, modifications are required, consult with Endress+Hauser.
Repair
To ensure continued operational safety and reliability,
‣ Carry out repairs on the device only if they are expressly permitted.
‣ Observe federal/national regulations pertaining to repair of an electrical device.
‣ Use original spare parts and accessories from Endress+Hauser only.

2.5 Hazardous area


To eliminate a danger for persons or for the facility when the device is used in the hazardous
area (e.g. explosion protection, pressure vessel safety):
• Based on the nameplate, check whether the ordered device is permitted for the intended use
in the hazardous area.
• Observe the specifications in the separate supplementary documentation that is an integral
part of these Instructions.

2.6 Product safety


This measuring device is designed in accordance with good engineering practice to meet
state-of-the- art safety requirements, has been tested, and left the factory in a condition in
which they are safe to operate. It fulfills general safety requirements and legal requirements. It
also conforms to the EC directives listed in the device-specific EC declaration of conformity.
Endress+Hauser confirms this fact by applying the CE mark.

8 Endress+Hauser
Deltapilot M PROFIBUS PA Identification

3 Identification
3.1 Product identification
The following options are available for identification of the measuring device:
• Nameplate specifications
• Order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer
(www.endress.com/deviceviewer): All information about the measuring device is displayed.
For an overview of the technical documentation provided, enter the serial number from the
nameplates in the W@M Device Viewer (www.endress.com/deviceviewer).

3.2 Scope of delivery


The scope of delivery comprises:
• Device
• Optional accessories
Documentation supplied:
• The Operating Instructions BA00383P is available on the Internet.
See: www.endress.com Download
• Brief Operating Instructions: KA01034P Deltapilot M
• Final inspection report
• Additional Safety Instructions for ATEX, IECEx and NEPSI devices
• Optional: factory calibration form, test certificates

3.3 CE mark, Declaration of Conformity


The devices are designed to meet state-of-the-art safety requirements, have been tested and left
the factory in a condition in which they are safe to operate. The devices comply with the
applicable standards and regulations as listed in the EC Declaration of Conformity and thus
comply with the statutory requirements of the EC Directives. Endress+Hauser confirms the
conformity of the device by affixing to it the CE mark.

4 Installation
4.1 Incoming acceptance
• Check the packaging and the contents for damage.
• Check the shipment, make sure nothing is missing and that the scope of supply matches your
order.

Endress+Hauser 9
Installation Deltapilot M PROFIBUS PA

4.2 Storage and transport


4.2.1 Storage
The device must be stored in a dry, clean area and protected against damage from impact (EN
837-2).
Storage temperature range:
See Technical Information for Deltapilot M TI00437P.

4.2.2 Transport

! WARNING
Incorrect transportation
Housing, diaphragm and capillaries may become damaged, and there is a risk of injury!
‣ Transport the measuring device to the measuring point in its original packaging or by the
process connection.
‣ Follow the safety instructions and transport conditions for devices weighing more than 18
kg (39.6 lbs).
‣ Do not use capillaries as a carrying aid for the diaphragm seals.

4.3 Installation conditions


4.3.1 Dimensions
For dimensions, please refer to the Technical Information for Deltapilot M TI00437P,
"Mechanical construction" section.

4.4 General installation instructions


• Devices with a G 1 1/2 thread:
When screwing the device into the tank, the flat seal has to be positioned on the sealing
surface
of the process connection. To avoid additional strain on the process isolating diaphragm, the
thread should never be sealed with hemp or similar materials.
• Devices with NPT threads:
– Wrap Teflon tape around the thread to seal it.
– Tighten the device at the hexagonal bolt only. Do not turn at the housing.
– Do not overtighten the thread when screwing. Max. torque: 20 to 30 Nm (14.75 to 22.13
lbf ft)

4.4.1 Mounting sensor modules with PVDF thread

! WARNING
Risk of damage to process connection!
Risk of injury!
‣ Sensor modules with PVDF process connections with threaded connection must be installed
with the mounting bracket provided!

10 Endress+Hauser
Deltapilot M PROFIBUS PA Installation

! WARNING
Material fatigue from pressure and temperature!
Risk of injury if parts burst! The thread can become loose if exposed to high pressure and
temperatures.
‣ The integrity of the thread must be checked regularly and the thread may need to be
re-tightened with the maximum tightening torque of 7 Nm (5.16 lbf ft). Teflon tape is
recommended for sealing the ½" NPT thread.

4.5 Installing
• Due to the orientation of the Deltapilot M, there may be a shift in the zero point, i.e. when the
container is empty or partially full, the measured value does not display zero. You can correct
this zero point shift  ä 26, Section "Function of the operating elements" or  ä 39,
Section 8.3 "Position zero adjustment".
• The local display can be rotated in 90° stages.
• Endress+Hauser offers a mounting bracket for installing on pipes or walls.
 ä 15, Section 4.5.6 "Wall and pipe mounting (optional)".

4.5.1 General installation instructions


• Do not clean or touch process isolating diaphragms with hard or pointed objects.
• The process isolating diaphragm in the rod and cable version is protected against mechanical
damage by a plastic cap.
• If a heated Deltapilot M is cooled during the cleaning process (e.g. by cold water), a vacuum
develops for a short time, whereby moisture can penetrate the sensor through the pressure
compensation (1). If this is the case, mount the Deltapilot M with the pressure compensation
(1) pointing downwards.
1

• Keep the pressure compensation and GORE-TEX® filter (1) free from contamination.

Endress+Hauser 11
Installation Deltapilot M PROFIBUS PA

• The device must be installed as follows in order to comply with the cleanability requirements
of the ASME-BPE (Part SD Cleanibility).:

4.5.2 FMB50

Level measurement
• Always install the device below the lowest measuring point.
• Do not install the device at the following positions:
– in the filling curtain
– in the tank outflow
– in the suction area of a pump
– or at a point in the tank that can be affected by pressure pulses from the agitator
• The calibration and functional test can be carried out more easily if you mount the device
downstream of a shutoff device.
• Deltapilot M must be included in the insulation for media that can harden when cold.

Pressure measurement in gases


• Mount Deltapilot M with shutoff device above the tapping point so that any condensate can
flow into the process.

Pressure measurement in steams


• Mount Deltapilot M with siphon above the tapping point.
• Fill the siphon with liquid before commissioning.
The siphon reduces the temperature to almost the ambient temperature.

Pressure measurement in liquids


• Mount Deltapilot M with the shutoff device below or at the same level as the tapping point.

12 Endress+Hauser
Deltapilot M PROFIBUS PA Installation

4.5.3 FMB51/FMB52/FMB53
• When mounting rod and cable versions, make sure that the probe head is located at a point
as free as possible from flow. To protect the probe from impact resulting from lateral
movement, mount the probe in a guide tube (preferably made of plastic) or secure it with a
clamping fixture.
• In the case of devices for hazardous areas, comply strictly with the safety instructions when
the housing cover is open.
• The length of the extension cable or the probe rod is based on the planned level zero point.
The height of the protective cap must be taken into consideration when designing the layout
of the measuring point. The level zero point (E) corresponds to the position of the process
isolating diaphragm.
Level zero point = E; top of the probe = L.
17 (0.67)

E
L

Endress+Hauser 13
Installation Deltapilot M PROFIBUS PA

4.5.4 Mounting the FMB53 with a suspension clamp

A0018793
Fig. 1: Mounting with a suspension clamp
1 Extension cable
2 Suspension clamp
3 Clamping jaws

Mounting the suspension clamp:


1. Mount the suspension clamp (item 2). When selecting the place to fix the unit, take the
weight of the extension cable (item 1) and the device into account.
2. Raise the clamping jaws (item 3). Position the extension cable (item 1) between the
clamping jaws as illustrated in Figure.
3. Hold the extension cable in position (item 1) and push the clamping jaws (item 3) back
down.
Tap the clamping jaws gently from above to fix them in place.

14 Endress+Hauser
Deltapilot M PROFIBUS PA Installation

4.5.5 Seal for flange mounting

NOTICE
Distorted measurement results.
The seal is not allowed to press on the process isolating diaphragm as this could affect the
measurement result.
‣ Ensure that the seal is not touching the process isolating diaphragm.

1 2

A0017743
Fig. 2:
1 Process isolating diaphragm
2 Seal

4.5.6 Wall and pipe mounting (optional)

Mounting bracket
See operating instructions.

Endress+Hauser 15
Installation Deltapilot M PROFIBUS PA

4.5.7 Assembling and mounting the "separate housing" version

5
4
3
2

1
7
r ³ 120 (4.72)

A0028494
Fig. 3: "Separate housing" version
1 In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready
mounted.
2 Cable with connection jack
3 Pressure compensation
4 Connector
5 Locking screw
6 Housing mounted with housing adapter, included
7 Mounting bracket provided, suitable for pipe and wall mounting (for pipes from 1 1/4" up to 2" diameter)

Maßeinheit mm (in)

Assembly and mounting


1. Insert the connector (item 4) into the corresponding connection jack of the cable (item 2).
2. Plug the cable into the housing adapter (item 6).
3. Tighten the locking screw (item 5).
4. Mount the housing on a wall or pipe using the mounting bracket (item 7).
When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at
least 5 Nm (3.69 lbf ft).
Mount the cable with a bending radius (r) 120 mm (4.72 in).

16 Endress+Hauser
Deltapilot M PROFIBUS PA Installation

Routing the cable (e.g. through a pipe)


You require the cable shortening kit.
Order number: 71093286
For details on mounting, see SD00553P/00/A6.

4.5.8 Supplementary installation instructions


Sealing the probe housing
• Moisture must not penetrate the housing when mounting the device, establishing the
electrical connection and during operation.
• Always firmly tighten the housing cover and the cable entries.

4.6 Mounting of the profile seal for universal process mounting


adapter
For details on mounting, see KA00096F/00/A3.

4.7 Closing the housing cover


NOTICE
Devices with EPDM cover seal - transmitter leakiness!
Mineral-based, animal-based or vegetable-based lubricants cause the EPDM cover seal to swell
and the transmitter to become leaky.
‣ The thread is coated at the factory and therefore does not require any lubrication.
NOTICE
The housing cover can no longer be closed.
Damaged thread!
‣ When closing the housing cover, please ensure that the thread of the cover and housing are
free from dirt, e.g. sand.If you feel any resistance when closing the cover, check the thread
on both again to ensure that they are free from dirt.

Endress+Hauser 17
Installation Deltapilot M PROFIBUS PA

4.7.1 Closing the cover on the stainless steel housing

A0028497
Fig. 4: Closing the cover

The cover for the electronics compartment is tightened by hand at the housing until the stop.
The screw serves as DustEx protection (only available for devices with DustEx approval).

4.8 Post-installation check

O Is the device undamaged (visual inspection)?

O Does the device comply with the measuring point specifications?


For example:
• Process temperature
• Process pressure
• Ambient temperature range
• Measuring range

O Are the measuring point identification and labeling correct (visual inspection)?

O Is the device adequately protected against precipitation and direct sunlight?

O Are the securing screw and securing clamp tightened securely?

18 Endress+Hauser
Deltapilot M PROFIBUS PA Electrical connection

5 Electrical connection
5.1 Connecting the device
! WARNING
Supply voltage might be connected!
Risk of electric shock and/or explosion!
‣ Ensure that no uncontrolles processes are activated in the system.
‣ Switch off the supply voltage before connecting the device.
‣ When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or Installation
or Control Drawings.
‣ A suitable circuit breaker must be provided for the device in accordance with
IEC/EN61010.
‣ Devices with integrated overvoltage protection must be grounded.
‣ Protective circuits against reverse polarity, HF influences and overvoltage peaks are
integrated.
Connect the device in the following order:
1. Check that the supply voltage corresponds to the supply voltage indicated on the nameplate.
2. Switch off the supply voltage before connecting the device.
3. Remove housing cover.
4. Guide the cable through the gland. Preferably use a twisted, shielded two-wire cable.
5. Connect the device in accordance with the following diagram.
6. Screw down the housing cover.
7. Switch on the supply voltage.

Endress+Hauser 19
Electrical connection Deltapilot M PROFIBUS PA

1
4
- +
2

A0029967
PROFIBUS PA electrical connection
1 External ground terminal
2 Grounding terminal
3 Supply voltage: 9 to 32 VDC (Segment coupler)
4 Terminals for supply voltage and signal

5.1.1 Devices with M12 connector

PIN assignment for M12 connector PIN Meaning

1 Signal +

2 Not assigned
4 3 3 Signal –

4 Earth
1 2

A0011175

20 Endress+Hauser
Deltapilot M PROFIBUS PA Electrical connection

5.2 Connecting the measuring unit

For further information on the network structure and grounding and for further bus system
components such as bus cables, see the relevant documentation, e.g. Operating Instructions
BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning" and the PNO
Guideline.

5.2.1 Supply voltage

Electronic version

PROFIBUS PA, 9 to 32 V DC
version for non-hazardous areas

5.2.2 Current consumption


11 mA ±1 mA, switch-on current corresponds to IEC 61158-2, Clause 21.

5.2.3 Terminals
• Supply voltage and internal ground terminal: 0.5 to 2.5 mm2 (20 to 14 AWG)
• External ground terminal: 0.5 to 4 mm2 (20 to 12 AWG)

Endress+Hauser 21
Electrical connection Deltapilot M PROFIBUS PA

5.2.4 Cable specification


• Use a twisted, shielded two-wire cable, preferably cable type A.
• Cable outer diameter: 5 to 9 mm (0.2 to 0.35 in)

For further information on the cable specifications, see Operating Instructions BA00034S
"PROFIBUS DP/PA: Guidelines for planning and commissioning", the PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).

5.2.5 Shielding/potential equalization


• You achieve optimum shielding against disturbances if the shielding is connected on both
sides (in the cabinet and on the device). If potential equalization currents are expected in the
plant, only ground shielding on one side, preferably at the transmitter.
• When using in hazardous areas, you must observe the applicable regulations.
Separate Ex documentation with additional technical data and instructions is included with
all Ex systems as standard.

5.3 Potential equalization


Hazardous area applications: Connect all devices to the local potential equalization.
Observe the applicable regulations.

5.4 Overvoltage protection (optional)


See operating instructions.

5.5 Post-connection check


Perform the following checks after completing electrical installation of the device:
• Does the supply voltage match the specifications on the nameplate?
• Is the device properly connected?
• Are all screws firmly tightened?
• Are the housing covers screwed down tight?
As soon as voltage is applied to the device, the green LED on the electronic insert lights up briefly
or the connected onsite display lights up.

22 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

6 Operation
6.1 Operating options
6.1.1 Operation without an operating menu

Operating options Explanation Graphic illustration Description

Local operation without The device is operated  ä 24


device display using the operating key
and DIP switches on the
electronic insert. Display
Zero

1
damp 2 on
not used 3
delta p on:Ö
on: off:SW 4
only on:P2=High off:SW 5 off

on SW

off HW
Address

6.1.2 Operation with an operating menu


Operation with an operating menu is based on an operation concept with "user roles"  ä 27.

Operating options Explanation Graphic illustration Description

Local operation The device is operated using  ä 29


with device display the operating keys on the
device display.

– + E

Remote operation via The device is operated using  ä 33


FieldCare the FieldCare operating tool.

Endress+Hauser 23
Operation Deltapilot M PROFIBUS PA

6.1.3 Operation via PA communication protocol

Operating options Explanation Graphic illustration Description

Remote operation via The device is operated using  ä 34


FieldCare the FieldCare operating tool.

Remote operation via The device is operated using  ä 34


PDM the PDM tool.

6.2 Operation without an operating menu


6.2.1 Position of operating elements
The operating key and DIP switches are located on the electronic insert in the device.

24 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

2
1 Zero
Display

3
1
damp 2 on
not used 3
delta p on:Ö off:SW 4
only on:P2= High off:SW 5 off

on SW

off HW
Address

on
5 4
1 2 3 4 5 off
SW / P2=High
damping

SW /

10 9 8 7 6
A0023126
Fig. 5: PROFIBUS PA electronic insert
1 Green LED to indicate successful operation
2 Operating key for position zero adjustment (Zero) or reset
3 Slot for optional local display
4 DIP-switch for bus address SW / HW
5 DIP-switch for hardware address
6+7 DIP switch only for Deltabar M
8 Not used
9 DIP switch for switching damping on/off
10 DIP switch for locking/unlocking parameters relevant to the measured value

Function of the DIP switches

Switches Symbol/ Switch position


labeling
"off" "on"

1 The device is unlocked. The device is locked.


Parameters relevant to the measured Parameters relevant to the measured
value can be modified. value cannot be modified.

2 damping  Damping is switched off. Damping is switched on.


The output signal follows measured value The output signal follows measured value
changes without any delay. changes with the delay time .1)

Endress+Hauser 25
Operation Deltapilot M PROFIBUS PA

Switches Symbol/ Switch position


labeling
"off" "on"

4 SW/
(Deltabar)

5 SW/P2= High
(Deltabar)

6 Address Set the device address using switches 1-7

7 SW / HW Hardware addressing Software addressing

1) The value for the delay time can be configured via the operating menu ("Setup" -> "Damping").
Factory setting: = 2 s or as per order specifications.

Function of the operating elements

Operating key(s) Meaning

"Zero" Position adjustment (zero point correction)


pressed for at least Press key for at least 3 seconds. The LED on the electronic insert lights up briefly if the
3 seconds pressure applied has been accepted for position adjustment.
See also the following Section "Performing position adjustment on site.''

"Zero" Reset
pressed for at least All parameters are reset to the order configuration.
12 seconds

Performing position adjustment on site


• Operation must be unlocked.  ä 33, Section 6.3.5 "Locking/unlocking operation".
• The device is configured at the factory for the Level measuring mode.
• The pressure applied must be within the nominal pressure limits of the sensor. See
information on the nameplate.
Perform position adjustment:
1. Pressure is present at device.
2. Press key for at least 3 seconds.
3. If the LED on the electronic insert lights up briefly, the pressure applied has been accepted
for position adjustment.
If the LED does not light up, the pressure applied was not accepted. Observe the input
limits. For error messages, see operating instructions.

26 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

6.2.2 Locking/unlocking operation


Once you have entered all the parameters, you can lock your entries against unauthorized and
undesired access.

If operation is locked by means of the DIP switch, you can only unlock operation again by means
of the DIP switch. If operation is locked by means of the operating menu, you can only unlock
operation again using the operating menu.

Locking/unlocking via DIP switches


DIP switch 1 on the electronic insert is used to lock/unlock operation.
 ä 25, "Function of the DIP switches".

6.3 Operation with an operating menu


6.3.1 Operation concept
The operation concept makes a distinction between the following user roles:

User role Meaning

Operator Operators are responsible for the devices during normal "operation". This is usually limited to
reading process values either directly at the device or in a control room. If the work with the
devices extends beyond value read-off tasks, the tasks involve simple, application-specific
functions that are used in operation. Should an error occur, these users simple forward the
information on the errors but do not intervene themselves.

Service Service engineers usually work with the devices in the phases following device commissioning.
engineer/technician They are primarily involved in maintenance and troubleshooting activities for which simple
settings have to be made at the device.
Technicians work with the devices over the entire life cycle of the product.
Thus, commissioning and advanced settings and configurations are some of the tasks they have
to carry out.

Expert Experts work with the devices over the entire product life cycle, but their device requirements
are often extremely high. Individual parameters/functions from the overall functionality of the
devices are required for this purpose time and again.
In addition to technical, process-oriented tasks, experts can also perform administrative tasks
(e.g. user administration).
"Experts" can avail of the entire parameter set.

6.3.2 Structure of the operating menu

User role Submenu Meaning/use

Operator Language Only consists of the "Language" parameter (000) where the operating language
for the device is specified.
The language can always be changed even if the device is locked.

Endress+Hauser 27
Operation Deltapilot M PROFIBUS PA

User role Submenu Meaning/use

Operator Display/operat. Contains parameters that are needed to configure the measured value display
(selecting the values displayed, display format, etc.).
With this submenu, users can change the measured value display without
affecting the actual measurement.

Service Setup Contains all the parameters that are needed to commission measuring
engineer/technic operations. This submenu has the following structure:
ian • Standard setup parameters
A wide range of parameters, which can be used to configure a typical
application, is available at the start. The measuring mode selected determines
which parameters are available.
After making settings for all these parameters, the measuring operation
should be completely configured in the majority of cases.
• "Extended setup" submenu
The "Extended setup" submenu contains additional parameters for more
in-depth configuration of the measurement operation to convert the
measured value and to scale the output signal.
This menu is split into additional submenus depending on the measuring
mode selected.

Service Diagnosis Contains all the parameters that are needed to detect and analyze operating
engineer/technic errors. This submenu has the following structure:
ian • Diagnostic list
Contains up to 10 error messages currently pending.
• Event logbook
Contains the last 10 error messages (no longer pending).
• Instrument info
Contains information on the device identification.
• Measured values
Contains all the current measured values
• Simulation
Is used to simulate pressure, level and alarm/warning.
• Reset

Expert Expert Contains all the parameters of the device (including those in one of the
submenus). The "Expert" submenu is structured by the function blocks of the
device. It thus contains the following submenus:
• System
Contains general device parameters that neither affect measurement nor
integration into a distributed control system.
• Measurement
Contains all the parameters for configuring the measurement.
• Communication
Contains the parameters of the PROFIBUS PA interface.
• Application
Contains all the parameters for configuring the functions that go beyond the
actual measurement (e.g. totalizer).
• Diagnosis
Contains all the parameters that are needed to detect and analyze operating
errors.

For an overview of the entire operating menu: see operating instructions.

28 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

Direct access to parameters


The parameters can only be accessed directly via the "Expert" user role.

Parameter name Description

Direct access (119) Use this function to enter a parameter code for direct access.
Entry
User input:
• Enter the desired parameter code.

Menu path: Factory setting:
Expert  Direct access 0

6.3.3 Operation with a device display (optional)


A 4-line liquid crystal display (LCD) is used for display and operation. The onsite display shows
measured values, dialog texts, fault messages and notice messages.
For easy operation the display can be taken out of the housing (see figure steps 1 to 3). It is
connected to the device through a 90 mm (3.54 in) cable.
The display of the device can be turned in 90° stages (see figure steps 4 to 6).
Depending on the orientation of the device, this makes it easy to operate the device and read the
measured values.

1. 2. 3.

4. 5. 6.

A0028500

Endress+Hauser 29
Operation Deltapilot M PROFIBUS PA

Functions:
• 8-digit measured value display including sign and decimal point.
• Bar graph as graphic display of the standardized value of the Analog Input Block (see also
operating instructions)
• Three keys for operation
• Simple and complete menu guidance as parameters are split into several levels and groups
• Each parameter is given a 3-digit parameter code for easy navigation
• Possibility of configuring the display to suit individual requirements and preferences, such as
language, alternating display, display of other measured values such as sensor temperature,
contrast setting.
• Comprehensive diagnostic functions (fault and warning message etc.)

F 3
1
4
– + E 6

7 5

A0030013
Fig. 6: Display
1 Main line
2 Value
3 Symbol
4 Unit
5 Bar graph
6 Information line
7 Operating keys

The following table illustrates the symbols that can appear on the onsite display. Four symbols
can occur at one time.

Symbol Meaning

Lock symbol
The operation of the device is locked. To unlock the device,  ä 33, Locking/unlocking
operation.

Communication symbol
Data transfer via communication

30 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

Symbol Meaning

Error message "Out of specification"


The device is being operated outside its technical specifications (e.g. during warmup or
cleaning processes).

Error message "Service mode"


The device is in the service mode (during a simulation, for example).

Error message "Maintenance required"


Maintenance is required. The measured value remains valid.

Error message "Failure detected"


An operating error has occurred. The measured value is no longer valid.

Operating keys on the display and operating module

Operating key(s) Meaning

O – Navigate downwards in the picklist


– Edit the numerical values and characters within a function

S – Navigate upwards in the picklist


– Edit the numerical values and characters within a function

– Confirm entry
F – Jump to the next item
– Selection of a menu item and activation of the editing mode

Contrast setting of onsite display: darker


O F and

Contrast setting of onsite display: brighter


S F and

ESC functions:

O S and
– Exit the edit mode for a parameter without saving the changed value.
– You are in a menu at a selection level. Each time you press the keys simultaneously,
you go up a level in the menu.

Operating example: Parameters with a picklist


Example: selecting "Deutsch" as the language of the menu.

Endress+Hauser 31
Operation Deltapilot M PROFIBUS PA

Language 000 Operation

1  English "English" is set as the menu language (default value). A  in front


of the menu text indicates the active option.

Deutsch

2 Deutsch Select "Deutsch" with  or .

 English

3  Deutsch 1. Confirm your choice with . A  in front of the menu text


indicates the active option ("Deutsch" is now selected as the
menu language).
English 2. Exit the edit mode for the parameter with .

Operating example: User-definable parameters


Example: setting "Set URV" parameter from 100 mbar (1.5 psi) to 50 mbar (0.75 psi).

Set URV 014 Operation

The local display shows the parameter to be changed. The value


highlighted in black can be changed. The "mbar" unit is specified
1 1 0 0 . 0 0 0 mbar in another parameter and cannot be modified here.

1. Press  or  to get to the editing mode.


2. The first digit is highlighted in black.
2 1 0 0 . 0 0 0 mbar

1. Use  to change "1" to "5".


2. Confirm "5" with . The cursor jumps to the next position
3 5 0 0 . 0 0 0 mbar (highlighted in black).
3. Confirm "0" with  (second position).

The third position is highlighted in black and can now be edited.

4 5 0 0 . 0 0 0 mbar

1. Switch to the "" symbol with the  key.


2. Use  to save the new value and exit the editing mode. 
5 5 0  . 0 0 0 mbar See next graphic.

32 Endress+Hauser
Deltapilot M PROFIBUS PA Operation

Set URV 014 Operation

The new value for the upper range value is 50.0 mbar (0.75 psi).

6 5  . 0 0 0 mbar – You exit the edit mode for the parameter with .
– You can get back to the editing mode with  or .

Operating example: Accepting the pressure present


Example: setting position adjustment

Pos. zero adjust 007 Operation

1  Abort The pressure for position adjustment is present at the device.

Confirm

2 Confirm Use  or  to switch to the "Confirm" option. The active option is


highlighted in black.

 Abort

3 Compensation Accept the pressure present as position adjustment with the 


accepted! key. The device confirms the adjustment and goes back to the
"Pos. zero adjust" parameter.

4  Abort Exit the edit mode for the parameter with .

Confirm

6.3.4 Operation via FieldCare


See operating instructions.

6.3.5 Locking/unlocking operation


See operating instructions.

6.3.6 Resetting to factory settings (reset)


See operating instructions.

Endress+Hauser 33
Operation Deltapilot M PROFIBUS PA

6.4 PROFIBUS PA communication protocol


6.4.1 Device identification and addressing
Please note the following:
• An address must be assigned to each PROFIBUS PA device. The control system/master can
only recognize the device if the address is set correctly.
• Each address can only be assigned once in any PROFIBUS PA network.
• Device addresses in the range from 0 to 125 are valid.
• The address "126" configured at the factory can be used for functional device testing and to
connect to a PROFIBUS PA network already in operation. This address must be changed
subsequently to add additional devices.
• On leaving the factory, all devices are delivered with the default address 126 and software
addressing.
• The FieldCare operating program is delivered with the default address 1.
There are two ways to assign the device address to a Deltapilot:
• Via an operating program of the DP Class 2 master, such as FieldCare or
• Onsite via DIP switches.

Zero
Display 2 + 8 = 10
1
damp 2 on
not used 3
delta p on:Ö off:SW 4
only on:P2= High off:SW 5 off

on SW
on
off HW SW
Address

– + E off
HW
1 2 3 4 5 6 7 8
Address
1 2

A0030184
Fig. 7: Setting the device address via DIP switches
1 If necessary, remove onsite display (optional)
2 Set the hardware address via the DIP switches

Hardware addressing
A hardware address is set as follows:
1. Set the DIP switch 8 (SW/HW) to "Off".
2. Set the address with DIP switches 1 to 7.
3. The change of address takes effect after 10 seconds. The device is restarted.

34 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning without an operating menu

DIP switch 1 2 3 4 5 6 7

Value when set to "On" 1 2 4 8 16 32 64

Value when set to "Off" 0 0 0 0 0 0 0

Software addressing
A software address is set as follows:
1. Set the DIP switch 8 (SW/HW) to "On" (factory setting).
2. The device is restarted.
3. The device reports its current address. Factory setting: 126
4. Set the address via the configuration program.
See the next section for information on how to enter a new address via FieldCare.
For other operating programs, see the corresponding operating manual.

7 Commissioning without an operating menu


The device is configured at the factory for the Level measuring mode. The measuring range and
the unit in which the measured value is transmitted correspond to the specifications on the
nameplate.

! WARNING
Exceeding the maximum allowable working pressure!
Risk of injury due to bursting of parts! Warning messages are generated if pressure is too high.
‣ If a pressure smaller than the minimum permitted pressure or greater than the maximum
permitted pressure is present at the device, the following messages are output in succession
(depending on the setting in the "Alarm behavior" (050) parameter):
"S140 Working range P" or "F140 Working range P"
"S841 Sensor range" or "F841 Sensor range"
"S971 Adjustment"
Use the device only within the sensor range limits.

NOTICE
Shortfall of the allowable working pressure!
Output of messages if pressure is too low.
‣ If a pressure smaller than the minimum permitted pressure or greater than the maximum
permitted pressure is present at the device, the following messages are output in succession
(depending on the setting in the "Alarm behavior" (050) parameter):
"S140 Working range P" or "F140 Working range P"
"S841 Sensor range" or "F841 Sensor range"

Endress+Hauser 35
Commissioning without an operating menu Deltapilot M PROFIBUS PA

"S971 Adjustment"
Use the device only within the sensor range limits.

7.1 Function check


Carry out a post-installation and a post-connection check as per the checklist before
commissioning the device.
• "Post-installation check" checklist  ä 18
• "Post-connection check" checklist  ä 22

7.2 Position adjustment


The following functions are possible by means of the key on the electronic insert:
• Position adjustment (zero point correction)
• Device reset  ä 26 (total reset)

• Operation must be unlocked.  ä 33, "Locking/unlocking operation"


• The device is configured for the "Pressure" measuring mode as standard.
• The pressure applied must be within the nominal pressure limits of the sensor. See
information on the nameplate.

Carrying out position adjustment

Pressure is present at device.

Press the "Zero" key for at least 3 s.

Does the LED on the electronic insert light up briefly?

Yes No

 
1)
Applied pressure for position adjustment has been Applied pressure for position adjustment has not been
accepted. accepted. Observe the input limits.

1) Observe warning on commissioning ( ä 35)

36 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

8 Commissioning with an operating menu


(onsite display/FieldCare)
The device is configured at the factory for the Level measuring mode. The measuring range and
the unit in which the measured value is transmitted correspond to the specifications on the
nameplate.

! WARNING
Exceeding the maximum allowable working pressure!
Risk of injury due to bursting of parts! Warning messages are generated if pressure is too high.
‣ If a pressure smaller than the minimum permitted pressure or greater than the maximum
permitted pressure is present at the device, the following messages are output in succession
(depending on the setting in the "Alarm behavior" (050) parameter):
"S140 Working range P" or "F140 Working range P"
"S841 Sensor range" or "F841 Sensor range"
"S971 Adjustment"
Use the device only within the sensor range limits.

NOTICE
Shortfall of the allowable working pressure!
Output of messages if pressure is too low.
‣ If a pressure smaller than the minimum permitted pressure or greater than the maximum
permitted pressure is present at the device, the following messages are output in succession
(depending on the setting in the "Alarm behavior" (050) parameter):
"S140 Working range P" or "F140 Working range P"
"S841 Sensor range" or "F841 Sensor range"
"S971 Adjustment"
Use the device only within the sensor range limits.

8.1 Function check


Carry out a post-installation and a post-connection check as per the checklist before
commissioning the device.
• "Post-installation check" checklist  ä 18
• "Post-connection check" checklist  ä 22

Endress+Hauser 37
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

8.2 Commissioning
Commissioning comprises the following steps:
1. Function check  ä 37
2. Selecting the language, measuring mode and pressure unit  ä 38
3. Position adjustment  ä 39
4. Configuring measurement:
– Pressure measurement  ä 51 ff
– Level measurement  ä 40 ff

8.2.1 Selecting the language, measuring mode and pressure unit

Language selection

Parameter name Description

Language (000) Select the language for the onsite display.


Options
Options:
• English
Menu path:
• Possibly another language (as selected when ordering the device)
Main menu  Language
• One further language (language of the manufacturing plant)
Factory setting:
English

Measuring mode selection

Parameter name Description

Measuring mode (005) Select the measuring mode.


Options The operating menu is structured differently depending on the measuring mode
selected.
Menu path:
Setup  Measuring mode
! WARNING
(005) Changing the measuring mode affects the span (URV)!
This situation can result in product overflow.
‣ If the measuring mode is changed, the span setting (URV) must be verified and, if
necessary, reconfigured!
Options:
• Pressure
• Level
Factory setting:
Pressure

38 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

Pressure unit selection

Parameter name Description

Press. eng. unit (125) Select the pressure unit.


Options If a new pressure unit is selected, all pressure-specific parameters are converted and
displayed with the new unit.
Menu path:
Options:
Setup  Press. eng. unit (125)
• mbar, bar
• mmH2O, mH2O,
• inH2O, ftH2O
• Pa, kPa, MPa
• psi
• mmHg, inHg
• kgf/cm2
Factory setting:
mbar or bar depending on the sensor nominal measuring range, or as per order
specifications

8.3 Position zero adjustment


The pressure resulting from the orientation of the device can be corrected here.

Parameter name Description

Corrected press. (172) Displays the measured pressure after sensor trim and position adjustment.
Display

Menu path:
Setup  Corrected press. (172) If this value is not equal to "0", it can be corrected to "0" by the position adjustment.

Pos. zero adjust (007) Position adjustment – the pressure difference between zero (set point) and the
(Gauge pressure sensor) measured pressure need not be known.
Options
Example:
– Measured value = 2.2 mbar (0.032 psi)
Menu path:
– Correct the measured value via the "Pos. zero adjust (007) (Gauge pressure sensor)"
Setup Pos. zero adjust (007)
parameter with the "Confirm" option. This means that you are assigning the value
(Gauge pressure sensor)
0.0 to the pressure present.
– Measured value (after pos. zero adjust) = 0.0 mbar
Options
• Confirm
• Abort
Factory setting:
Abort

Endress+Hauser 39
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

Parameter name Description

Calib. offset (192) (008) Position adjustment – the pressure difference between the set point and the measured
(absolute pressure sensor) pressure must be known.
Entry
Example:
– Measured value = 982.2 mbar (14.25 psi)
Menu path:
– You correct the measured value with the value entered (e.g. 2.2 mbar (0.032 psi))
Setup Calib. offset (192)
via the "Calib. offset (192)" parameter. This means that you are assigning the value
980.0 (14.21 psi) to the pressure present.
– Measured value (after calib. offset) = 980.0 mbar (14.21 psi)
Factory setting:
0.0

8.4 Level measurement


8.4.1 Information on level measurement
• The limit values are not checked, i.e. the values entered must be appropriate for the sensor
and the measuring task for the device to be able to measure correctly.
• Customer-specific units are not possible.
• There is no unit conversion.
• The values entered for "Empty calib. (028)/Full calib. (031)", "Empty pressure (029)/Full
pressure (032)", "Empty height (030)/Full height (033)" must be at least 1 % apart. The value
will be rejected, and a warning message displayed, if the values are too close together.
You have a choice of two methods for calculating the level: "In pressure" and "In height". The table
in the "Overview of level measurement" section that follows provides you with an overview of
these two measuring tasks.

8.4.2 Overview of level measurement

Measuring task Level Measured Description Measured value display


selection variable options

Calibration is "In pressure" Via the "Unit before – Calibration with reference The measured value
performed by entering lin (025)" pressure (wet calibration), display and the
two pressure-level parameter: %, level, see  ä 41 "Level before lin. (019)"
value pairs. volume or mass – Calibration without parameter display the
units. reference pressure (dry measured value.
calibration), see  ä 42

Calibration is "In height" – Calibration with reference


performed by entering pressure (wet calibration),
the density and two see  ä 45
height-level value – Calibration without
pairs. reference pressure (dry
calibration), see  ä 47

40 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

8.4.3 "In pressure" level selection


Calibration with reference pressure (wet calibration)

Example:
In this example, the level in a tank should be measured in "m". The maximum level is 3 m (9.8
ft). The pressure range is due to the filling height and the density.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled and emptied.

The values entered for "Empty calib. (028)/Full calib. (031)" and the pressures present at the
device must be at least 1% apart. The value will be rejected, and a warning message displayed,
if the values are too close together. Further limit values are not checked; i.e. the values entered
must be appropriate for the sensor and the measuring task so that the measuring device can
measure correctly.

Description

1 Perform "position adjustment"  ä 39.


B
2 Select the "Level" measuring mode via the 300 mbar
"Measuring mode (005)" parameter.
3m
Menu path: Setup Measuring mode (005)

3 Select a pressure unit via the "Press. eng. unit (125)" A


parameter, here "mbar" for example.
0 mbar
Menu path: Setup  Press. eng. unit (125) 0m
4 Select the "in pressure" level mode by means of the
"Level selection (024)" parameter.
A0030028
Menu path: Setup Extended Setup Level  Fig. 8: Calibration with reference pressure –
Level selection (024) wet calibration
5 Select a level unit by means of the "Unit before lin A See Table, Step 7.
(025)" parameter, here "m" for example. B See Table, Step 8.

Menu path: Setup  Extended Setup  Level 


Unit before lin (025)

6 Select the "Wet" option by means of the "Calibration


mode (027)" parameter.

Menu path: Setup  Extended Setup  Level 


Calibration mode (027)

Endress+Hauser 41
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

Description

7 The pressure for the lower calibration point is


present at the device, here 0 mbar for example. h
Select the "Empty calib. (028)" parameter. [m]
Menu path: Setup  Extended Setup  Level  B 3
Empty calib. (028)

Enter the level value, here 0 m for example. The


pressure value present is assigned to the lower level
value by confirming the value.

8 The pressure for the upper calibration point is


present at the device, here 300 mbar (4.35 psi) for
example.

Select the "Full calib. (031)" parameter. A 0


0 300 p
Menu path: Setup  Extended Setup  Level 
Full calib. (031)
[mbar]

Enter the level value, here 3 m (9.8 ft) for example. A0017658
The pressure value present is assigned to the upper Fig. 9: Calibration with reference pressure –
level value by confirming the value. wet calibration

9 If calibration is performed with a medium other A See Table, Step 7.


than the process medium, enter the density of the B See Table, Step 8.
calibration medium in "Adjust density (034)".

Menu path: Setup  Extended Setup  Level 


Adjust density (034)

10 If calibration was performed with a medium other


than the process medium, specify the density of the
process medium in the "Process density (035)"
parameter.

Menu path: Setup  Extended setup  Level 


Process density (035).

11 Result:
The measuring range is set for 0 to 3 m (9.8 ft).

The measured variables %, level, volume and mass are available for this level mode.
See operating instructions "Unit before lin (025)".

8.4.4 "In pressure" level selection


Calibration without reference pressure (dry calibration)
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters (264 gal) corresponds to a pressure of 450 mbar (6.53 psi). The minimum volume

42 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

of 0 liters corresponds to a pressure of 50 mbar (0.72 psi) since the device is mounted below the
start of the level measuring range.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the pressure and volume values for the lower and upper
calibration point must be known.

• The values entered for "Empty calib. (028)/Full calib. (031)", "Empty pressure (029)/Full
pressure (032)" must be at least 1% apart. The value will be rejected, and a warning message
displayed, if the values are too close together. Further limit values are not checked; i.e. the
values entered must be appropriate for the sensor and the measuring task so that the
measuring device can measure correctly.
• Due to the orientation of the device, there may be pressure shifts in the measured value, i.e.
when the container is empty or partly filled, the measured value is not zero. For information
on how to perform position adjustment, see  ä 39, "Position zero adjustment".

Description

1 Select the "Level" measuring mode via the


"Measuring mode (005)" parameter.

Menu path: Setup Measuring mode (005) B


g 1000 l
2 Select a pressure unit via the "Press. eng. unit (125)" r =1 450 mbar
parameter, here "mbar" for example. cm 3
Menu path: Setup  Press. eng. unit (125) A
3 Select the "in pressure" level mode by means of the 0l
"Level selection (024)" parameter. 50 mbar
Menu path: Setup  Extended Setup  Level 
Level selection (024)

4 Select a volume unit via the "Unit before lin (025)"


parameter, here "l" (liter) for example.
A0030030

Menu path: Setup  Extended Setup  Level  Fig. 10: Calibration without reference pressure –
Unit before lin (025) dry calibration
A See Table, Steps 7 + 8.
B See Table, Steps 9 + 10.

Endress+Hauser 43
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

Description

5 Select the "Dry" option by means of the "Calibration


mode (027)" parameter. V
Menu path: Setup  Extended Setup  Level  [l]
Calibration mode (027)
C 1000
6 "Adjust density (034)" contains the factory setting
1.0, but this value can be changed if required. The
entered value pairs must correspond to this density.

Menu path: Setup  Extended Setup  Level 


Adjust density (034)

7 Enter the volume value for the lower calibration


point via the "Empty calib. (028)" parameter, here 0
liters for example.
A 0
Menu path: Setup  Extended Setup  Level  50 450 p
Empty calib. (028) [mbar]
B D
8 Enter the pressure value for the lower calibration
A0031028
point via the "Empty pressure (029)" parameter, Fig. 11: Calibration with reference pressure –
here 50 mbar (0.72 psi) for example. wet calibration
Menu path: Setup  Extended Setup  Level  A See Table, Step 7.
Empty pressure (029) B See Table, Step 8.
C See Table, Step 9.
9 Enter the volume value for the upper calibration D See Table, Step 10.
point via the "Full calib. (031)" parameter, here
1000 liters (264 gal) for example.

Menu path: Setup  Extended Setup  Level 


Full calib. (031)

10 Enter the pressure value for the upper calibration


point via the "Full pressure (032)" parameter, here
450 mbar (6.53 psi) for example.

Menu path: Setup  Extended Setup  Level 


Full pressure (032)

11 If calibration was performed with a medium other


than the process medium, specify the density of the
process medium in the "Process density (035)"
parameter.
Menu path: Setup  Extended setup  Level 
Process density (035).

12 Result:
The measuring range is set for 0 to 1000 l (264
gal).

The measured variables %, level, volume and mass are available for this level mode.
See operating instructions "Unit before lin (025)".

44 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

8.4.5 "In height" level selection


Calibration with reference pressure (wet calibration)

Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters (264 gal) corresponds to a level of 4.5 m (14.8 ft). The minimum volume of 0 liters
corresponds to a level of 0.5 m (1.6 ft) since the device is mounted below the start of the level
measuring range.
The density of the medium is 1 g/cm3 (1 SGU).
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled and emptied.

The values entered for "Empty calib. (028)/Full calib. (031)" and the pressure values present at
the device must be at least 1% apart. The value will be rejected, and a warning message
displayed, if the values are too close together. Further limit values are not checked; i.e. the
values entered must be appropriate for the sensor and the measuring task so that the measuring
device can measure correctly.

Description

1 Perform position adjustment. See  ä 39.

2 Select the "Level" measuring mode via the


"Measuring mode (005)" parameter. C
g 1000 l
Menu path: Setup  Measuring mode (005) A r =1 4.5 m
cm 3
3 Select a pressure unit via the "Press. eng. unit (125)"
parameter, here "mbar" for example.
B
Menu path: Setup Press. eng. unit (125) 0l
0.5 m
4 Select the "in height" level mode via the "Level
selection (024)" parameter.

Menu path: Setup  Extended Setup  Level 


Level selection (024)

5 Select a volume unit via the "Unit before lin (025)" A0031027
Fig. 12: Calibration with reference pressure –
parameter, here "l" (liter) for example.
wet calibration
Menu path: Setup  Extended Setup  Level  A See Table, Step 8.
Unit before lin (025) B See Table, Step 9.
B See Table, Step 9.

Endress+Hauser 45
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

Description

6 Select a level unit by means of the "Height unit


(026)" parameter, here "m" for example. h p
h=
Menu path: Setup  Extended Setup  Level  [m] r ×g
Height unit (026)
4.5
7 Select the "Wet" option by means of the "Calibration
mode (027)" parameter.
Menu path: Setup  Extended Setup  Level  A
Calibration mode (027) g
r =1
8 If calibration is performed with a medium other cm 3
than the process medium, enter the density of the
calibration medium in the "Adjust density (034)"
parameter, here 1 g/cm3 (1 SGU) for example.
Menu path: Setup  Extended Setup  Level 
Adjust density (034) 0.5
50 450 p
9 The pressure for the lower calibration point is
present at the device, here 0.5 m covered / 49 mbar V [mbar]
(0.71 psi) for example. [l]
Enter the volume value for the lower calibration
point via the "Empty calib. (028)" parameter, here 0 C 1000
liters for example.

Menu path: Setup  Extended Setup  Level 


Empty calib. (028)

10 The pressure for the upper calibration point is


p
present at the device, here 4.5 m covered / 441 h=
mbar (6.4 psi) for example. r ×g
Enter the volume value for the upper calibration
point via the "Full calib. (031)" parameter, here
B 0
"1000 liters" (264 gal) for example. 0.5 4.5 h
[m]
Menu path: Setup  Extended Setup  Level 
Full calib. (031) A0031029
Fig. 13: Calibration with reference pressure –
11 If calibration was performed with a medium other wet calibration
than the process medium, specify the density of the A See Table, Step 8.
process medium in the "Process density (035)" B See Table, Step 9.
parameter. C See Table, Step 10.
Menu path: Setup  Extended Setup  Level 
Process density (035)

12 Result:
The measuring range is set for 0 to 1000 l (264
gal).

The measured variables %, level, volume and mass are available for this level mode operating
instructions "Unit before lin (025)".

46 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

8.4.6 "In height" level selection


Calibration without reference pressure (dry calibration)
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters (264 gal) corresponds to a level of 4.5 m (14.8 ft). The minimum volume of 0 liters
corresponds to a level of 0.5 m (1.6 ft) since the device is mounted below the start of the level
measuring range.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the height and volume values for the lower and upper
calibration point must be known.

• The values entered for "Empty calib. (028)/Full calib. (031)", "Empty height (030)/Full height
(033)" must be at least 1% apart. The value will be rejected, and a warning message displayed,
if the values are too close together. Further limit values are not checked; i.e. the values
entered must be appropriate for the sensor and the measuring task so that the measuring
device can measure correctly.
• Due to the orientation of the device, there may be pressure shifts in the measured value, i.e.
when the container is empty or partly filled, the measured value is not zero. For information
on how to perform position adjustment, see  ä 39, "Position zero adjustment".

Endress+Hauser 47
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

Description

1 Select the "Level" measuring mode via the


"Measuring mode (005)" parameter.

Menu path: Setup Measuring mode (005) C


g 1000 l
2 Select a pressure unit via the "Press. eng. unit (125)" A r =1 4.5 m
parameter, here "mbar" for example. cm 3
Menu path: Setup Press. eng. unit (125) B
3 Select the "in height" level mode via the "Level 0l
selection (024)" parameter. 0.5 m
Menu path: Setup  Extended Setup  Level 
Level selection (024)

4 Select a volume unit via the "Unit before lin (025)"


parameter, here "l" (liter) for example.
A0031027
Menu path: Setup  Extended Setup  Level 
Fig. 14: Calibration without reference pressure –
Unit before lin (025)
dry calibration
5 Select a level unit by means of the "Height unit A See Table, Step 7.
(026)" parameter, here "m" for example. B See Table, Step 8 and 10.
C See Table, Step 9 and 11.
Menu path: Setup  Extended Setup  Level 
Height unit (026)

6 Select the "Dry" option by means of the "Calibration


mode (027)" parameter.

Menu path: Setup  Extended Setup  Level 


Calibration mode (027)

7 Enter the density of the medium via the "Adjust


density (034)" parameter, here "1 g/cm3" (1 SGU)
for example.

Menu path: Setup  Extended Setup  Level 


Adjust density (034)

48 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

Description

8 Enter the volume value for the lower calibration


point via the "Empty calib. (028)" parameter, here 0 h p
liters for example. h=
[m] r ×g
Menu path: Setup  Extended Setup  Level 
Empty calib. (028) 4.5
9 Enter the height value for the lower calibration
point via the "Empty height (030)" parameter, here A
0.5 m (1.6 ft) for example. g
r =1
Menu path: Setup  Extended Setup  Level 
cm 3
Empty height (030)

10 Enter the volume value for the upper calibration


point via the "Full calib. (031)" parameter, here
1000 liters (264 gal) for example. 0.5
50 450 p
Menu path: Setup  Extended Setup  Level 
Full calib. (031)
V [mbar]
[l]
11 Enter the height value for the upper calibration
point via the "Full height (033)" parameter, here 4.5
m (14.8 ft) for example. D 1000

Menu path: Setup  Extended Setup  Level 


Full height (033)

12 If the process uses a medium other than that on


which the calibration was based, the new density p
must be specified in the "Process density (035)" h=
parameter. r ×g
Menu path: Setup  Extended setup  Level  B 0
Process density (035). h
0.5 4.5
13 Result: C E [m]
The measuring range is set for 0 to 1000 l (264
A0031066
gal).
Fig. 15: Calibration with reference pressure –
wet calibration
A See Table, Step 7.
B See Table, Step 8.
C See Table, Step 9.
D See Table, Step 10.
E See Table, Step 11.

The measured variables %, level, volume and mass are available for this level mode operating
instructions "Unit before lin (025)".

Endress+Hauser 49
Commissioning with an operating menu (onsite display/FieldCare) Deltapilot M PROFIBUS PA

8.4.7 Required parameters for Level measuring mode

Parameter name Description

Level selection (024) see operating instructions

Unit before lin (025) see operating instructions

Height unit (026) see operating instructions

Calibration mode (027) see operating instructions

Empty calib. (028) see operating instructions

Empty pressure (029) see operating instructions

Empty height (030) see operating instructions

Full calib. (031) see operating instructions

Full pressure (032) see operating instructions

Full height (033) see operating instructions

Density unit (127) see operating instructions

Adjust density (034) see operating instructions

Process density (035) see operating instructions

Level before lin. (019) see operating instructions

8.5 Linearization
See operating instructions.

50 Endress+Hauser
Deltapilot M PROFIBUS PA Commissioning with an operating menu (onsite display/FieldCare)

8.6 Pressure measurement


8.6.1 Calibration without reference pressure (dry calibration)
Example:
In this example, a device with a 400 mbar (6 psi) sensor is configured for the 0 to +300 mbar
(4.35 psi) measuring range, i.e. 0 mbar and 300 mbar (4.35 psi) are assigned.
Prerequisite:
This is a theoretical calibration, i.e. the pressure values for the lower and upper range are
known.

Due to the orientation of the device, there may be pressure shifts in the measured value, i.e. the
measured value is not zero in a pressureless condition. For information on how to perform
position adjustment, see  ä 39. Calibration is possible only using FieldCare.

Description

1 Select the "Pressure" measuring mode via the "Measuring mode (005)" parameter.

Menu path: Setup Measuring mode (005)

2 Select a pressure unit via the "Press. eng. unit (125)" parameter, here "mbar" for example.

Menu path: Setup Press. eng. unit (125)

3 Where necessary scale the "Output value (Out Value)" of the Analog Input Block, see operating instructions,
parameter descriptions for "Proc value scale" and "Output scale".

4 Result:
The measuring range is configured for 0 to +300 mbar (4.35 psi).

Endress+Hauser 51
71316892

www.addresses.endress.com