Siwarex Calibration Procedure
Siwarex Calibration Procedure
Rev. 1.0
Ownership of this document is related to Extreme Automation S.r.l. All rights are reserved, included sharing, publishing
and copying
Rev’s
2
Index
SIWAREX Load cell calibration procedure ...............................................................................................................1
Rev’s 2
Index 3
1 Start set up ..................................................................................................................................................4
1.1 From main screen, push on “libra” symbol. ................................................................................................4
1.2 Push on adjustment button .......................................................................................................................5
1.3 Select “1.0 Setup” function .......................................................................................................................5
1.4 Select “1.2 Quick start” function ................................................................................................................6
1.5 Switch on service mode ............................................................................................................................7
1.6 Choosing calibration settings .....................................................................................................................8
1.7 Calibration parametrization ......................................................................................................................9
1.8 Calibration method ................................................................................................................................10
2a Calibration with samples weight .................................................................................................................10
2a.1 Set sample weight ..................................................................................................................................11
2a.2 Calibrations .............................................................................................................................................12
2b Automatic calibration .................................................................................................................................13
2b.1 Cells parameters ....................................................................................................................................13
3
1 Start set up
Once pushed, system will ask you access credentials. Log in as:
User: Electronico
Password: Electronic4
4
1.2 Push on adjustment button
5
1.4 Select “1.2 Quick start” function
6
1.5 Switch on service mode
First, push on red button << Switch on “Service mode” >> (1)
After this action, service mode status flag will become yellow (1).
7
1.6 Choosing calibration settings
In case of normal calibration, push on (a) button, to keep load cell settings already fixed.
Note: (b) option it will be selected ONLY in case of load cell substitution. In a calibration
procedure NEVER select (b) option
8
1.7 Calibration parametrization
9
1.8 Calibration method
At this point, you can choose if you would like to calibrate automatically (a) or with a sample
weight (b)
Both of procedure will be described.
10
2a.1 Set sample weight
11
2a.2 Calibrations
Arrived at this point, we must perform first “Set Calibration weight 0” with measurement point
completely empty, then wait some seconds, load known sample weight on weighing station and
perform “Set calibration weight 1”. Now load cell are parametrized.
12
2b Automatic calibration
By select << Calibrate without weights >> [option (b), as described on 1.8] you will have to leave
weighing station completely empty and with any kind of mechanical interference.
Leave parameters as they currently are! To be changed only in case of new type of load cells
installation. Push on “Next” button.
13
2b.2 Start automatic calibration
14