INSTRUMENT CONTROL PARAMETERS: GCMS

--------------------------------------

C:\MassHunter\GCMS\1\METHODS\TDUBetteraveGazSaver2022.M
Wed Feb 15 23:26:42 2023

Control Information
------- -----------

Sample Inlet : GC
Injection Source : External Device
Mass Spectrometer : Enabled

Injection Location: Rear

No Sample Prep method has been assigned to this method.

GC
GC Summary
Run Time 41.875 min
Post Run Time 0 min

Oven
Temperature
Setpoint On
(Initial) 35 °C
Hold Time 0 min
Post Run 50 °C
Program
#1 Rate 8 °C/min
#1 Value 290 °C
#1 Hold Time 10 min

Equilibration Time 0.05 min


Max Temperature 350 °C
Maximum Temperature Override Disabled
Slow Fan Disabled

Front SS Inlet He
Mode Pulsed Splitless
Heater Off
Pressure Off
Total Flow Off
Septum Purge Flow Off
Septum Purge Flow Mode Standard
Pre-Run Flow Test Off
Injection Pulse Pressure 25 psi Until 0.5 min
Purge Flow to Split Vent 100 mL/min at 1 min
Liner Agilent 5190-2293: 900 μL (Splitless,
single taper, wool, Ultra )

Back PTV Inlet He


Mode Solvent Vent
Pressure On 6.7776 psi
Total Flow On 54 mL/min
Septum Purge Flow On 3 mL/min
Septum Purge Flow Mode Standard
Pre-Run Flow Test Off
Gas Saver On 20 After 3 min mL/min
Purge Flow to Split Vent 50 mL/min at 1.2 min
Vent Flow 100 mL/min per min
Vent Pressure 7.5 psi Until 0.05 min
Liner Agilent 18740-80200: 140 μL (Direct,
straight (gas samples, HS, P&)

Column
Column #1
Flow
Setpoint On
(Initial) 1 mL/min
Post Run 1 mL/min

Column Information Agilent 19091S-433UI: 0684951H


HP-5MS UI
Temperature Range -60 °C—325 °C (350 °C)
Dimensions 30 m x 250 μm x 0.25 μm
Column lock Locked
In Back PTV Inlet He
Out MSD
(Initial) 35 °C
Pressure 6.7776 psi
Flow 1 mL/min
Average Velocity 36.169 cm/sec
Holdup Time 1.3824 min
Control Mode Constant Flow

Column Outlet Pressure 0 psi

Detector Evaluation
Perform Detector Evaluation Test On
Signal Selected No Signal Selected
Checkout Sample NONE

MSD Transfer Line


Temperature
Setpoint On
(Initial) 250 °C

Signals
Signal #1:
Description None

Signal #2:
Description None

Signal #3:
Description None

Signal #4:
Description None

Signal #5:
Description None

Signal #6:
Description None

Signal #7:
Description None

Signal #8:
Description None

GERSTEL MAESTRO

SYSTEM SETTINGS
Maestro Runtime : 41.88 min
GC Cool Down Time : 4.00 min
Cryo Timeout : 20.00 min

GERSTEL CIS

CIS : used

CRYO COOLING
Cryo Cooling : used

TEMPERATURE PROGRAM
Heater Mode : Standard

Liner Type : 013882-010-00 glass liner CIS4, baffled, non


deactivated
Liner Max Temperature : 400 'C

Initial Temperature : -40 'C


Equilibration Time : 0.05 min
Initial Time : 0.20 min
Ramp 1
Rate : 12.00 'C/s
End Temp : 280 'C
Hold Time : 5.00 min
Ramp 2
Rate : 0.0 'C/s

Low Split : not used

GERSTEL TD3.5+

TEMPERATURE PROGRAM
Initial Temperature : 30 'C
Delay Time : 0.50 min
Initial Time : 20.20 min
Ramp 1
Rate 1 : 60.0 'C/min
End Temp 1 : 280 'C
Hold Time 1 : 10.00 min
Ramp 2
Rate 2 : 0.0 'C/min

TD3.5+ SETTINGS
Transfer Temp : 300 'C
Transfer Temp Mode : Fixed
Desorption Mode : Solvent Venting
Vent Time : 20.00 min
Sample Mode : Retain Tube - Standby Cooling
Standby Temp : 50 'C

TD3.5+ LIQUID OPTION


Liquid Option : not used - use these parameters if it becomes 'used'
Sample Tray : Rack 1,VT54/2-CVM
Fixed Liquid Vial : used
Liquid Vial : 1
Inject into Micro vial : not used

GERSTEL DHS

DHS : not used - use these parameters if it becomes 'used'


DHS Module : DHS

INJECTION TUBE
Tube Tray : TDUTray1,VT40t/Tube
Sync. Vial/Tube Numbers : used

PURGE TUBE
Purge Tube : not used - use these parameters if it becomes 'used'
Purge Tube Tray :
Tube Number :

INCUBATION SETTINGS
Incubation Temperature : 30 'C
Incubation Time : 5.00 min

AGITATION SETTINGS
Agitator On Time : 10 s
Agitator Off Time : 1 s
Agitator Speed : 500 rpm

TRANSFER HEATER SETTINGS


TEMP : 50 'C

PURGE PHASE SETTINGS


Volume : 0.0 mL
TRAPPING PHASE SETTINGS
Volume : 200.0 mL
Flow : 20.0 mL/min
Trap Temperature : 25 'C
Incubation Temperature : 30 'C

DRYING PHASE SETTINGS


Volume : 50.0 mL
Flow : 10.0 mL/min
Trap Temperature : 25 'C

MS Information
-- -----------

General Information
------- -----------

Acquisition Mode : Scan


Solvent Delay (minutes) : 0
Tune file : C:\MassHunter\GCMS\1\5977\ATUNE.U
EM Setting mode Gain : 1.000000

Normal or Fast Scanning : Normal Scanning


Trace Ion Detection : Off
Run Time (if MS only) : 10 minutes

[Scan Parameters]
Start Time : 0
Low Mass : 30
High Mass : 400
Threshold : 150
A/D Samples: : 4

[MSZones]

MS Source : 230 C maximum 250 C


MS Quad : 150 C maximum 200 C

Timed Events
----- ------
Number Events= 0

END OF MS ACQUISITION PARAMETERS

TUNE PARAMETERS for SN: US2104R033


---------------------------------

Trace Ion Detection is OFF.

34.593 : EMISSION
70.007 : ENERGY
25.536 : REPELLER
79.473 : IONFOCUS
25.226 : ENTRANCE_LENS
991.027 : EMVOLTS
1026.4 : Actual EMV
1.00 : GAIN FACTOR
2382.000 : AMUGAIN
139.375 : AMUOFFSET
2.000 : FILAMENT
0.000 : DCPOLARITY
11.861 : ENTLENSOFFSET
0.000 : Ion_Body
0.000 : EXTLENS
-55.000 : MASSGAIN
-21.000 : MASSOFFSET

END OF TUNE PARAMETERS


----------------------

END OF INSTRUMENT CONTROL PARAMETERS


------------------------------------