RADPRO
RADPRO
RADPRO
a. 5 ft
b. 7 ft
c. 10 ft
d. 14 ft
2. If the exposure rate at 3 ft from fluoroscopic table is 40 mR/hr, what will be the exposure for
at a distance of 5 ft from the table?
a. 7 mR
b. 12 mR
c. 14 mR
d. 24 mR
3. What is the approximate dose reduction to the radiographer if the distance from the patient
during fluoroscopy is doubled?
a. ½
b. 1/4
c. 1/8
d. 1/9
4. If a patient received 2,000 mrad during a 10-min fluoroscopic examination, what was the dose
rate?
a. 0.2 rad/min
b. 2.0 rad/min
c. 5 rad/min
d. 200 rad/min
a. Silver bromide
b. Sodium sulfite
c. Lithium fluoride
d. Aluminum oxide
6. What is the most commonly employed detecting device for scattered radiation?
a. scintillation counter
b. pocket dosimeter
c. Thermoluminescent dosimeter
d. Ionization chamber
7. The automatic exposure device that is located immediately under the x-ray table is;
Question; A Radiologist Technologist is doing a work with a radiation source that delivers a
dose rate of 25 mSv/hr at point one meter from the source
9. 9. How long can the technologist work at point A without exceeding the dose limit of 0.20
mSv?
a. 28.8 sec
b. 18.6
c. 35.3 sec
d 14.4 sec
10. What is the dose rate at a point 2 meters from the source?
a. 12.5 mSv/hr
b. 6.25 mSv/hr
c. 35.5 mSv/hr
d. 5.25 mSv/hr
11. How long can he work at 2 meters from the source without exceeding the dose limit of 0.20
mSv?
a. 9.12 min
b. 4.35 min
c. 3.45 min
d. 1.92 min
12. What is the dose rate of the Technologist 3 meters from point A?
a. 20.8 mSv/hr
b. 27.8 mSv/hr
c. 2.78 mSv/hr
d. 0.28 mSv/hr
13. How much dose will be received if he stays for two hours at 2 meters from the source?
a. 12.5 mSv
b. 75 mSv
c. 50 mSv
d. 37.5 mSv
14. The dose rate for an x-ray 3 meters from the source is 150 mR/hr. What is the dose rate at a
distance of 4 meters?
a. 8.437 mR/hr
b. 8.437 mR/hr
c. 84.375 mR/hr
d 84 mR/hr
15. How many HVLS are required in order to reduce the intensity of a beam of monoenergetic
photons to less than 20 percent of its original value?
a. 2
b. 3
c. 4
d. 5
16. If the exposure rate to a body standing 5 ft from a radiation source is 10 mR/min, what will
be the dose rate to that body at a distance of 8 ft from the source?
a 25.6 mR/min
b. 16 mR/min
c. 6.25 mR/min
d. 3.9 mR/min
17. In which type of monitoring device do photons release electrons by their interaction with
air?
a. film badge
b. TLD
c. pocket dosimeter
d. OSL
18. All of the following statements regarding TLDs are true except,
19. The dose rate from a point 2 meters from the radiation source is 500 mR/hr. What is the
dose rate at a distance of 3.5 meters?
a. 160.34 mR/hr
b. 159 mR/hr
c. 161 mR/hr
d. 163.26 mR/hr
20. Which of the following personnel radiation monitors will provide an immediate reading?
a. TLD
b. Film badge
c. OSL
d. pocket dosimeter
22. The most efficient type of male gonadal shielding for use during fluoroscopy.
a. flat contact
b. shaped contact
C. OSL
d. cylindrical
23. Which term describes the amount of material required to reduce the intensity of the beam by
50% of its original value
a. LD 50/30
b. half-life
c. Half value layer
d. 50% rule
24. How many half-value layers of filtration must be added to the primary beam to reduce its
intensity to less than 1% of its original value?
a. 9
b. 7
c. 5
6. 6
25. If an equivalent dose of 240 mSv is received uniformly over a period of three hours, what is
the dose rate?
26. The dose rate 2 meters from a radiation source is 200 mR/hr. What will be the dose rate if a
person moves 1 meter nearer to the source?
a 1000 mR/hr
b. 600 mR/hr
c. 800 mR/hr
d. 400 mR/hr
27. How many half value layers are needed to reduce a dose rate of 4,000 mSv/hr to 125
mSv/hr?
a. 5 b. 4 c. 3 d. 2
28. Which type of shielding is suspended from the radiographic collimator to absorb the primary
beam photons before they reach the reproductive organs?
a. contact
b. shadow
c. contour
d. flat
a. 5 ft
b. 6 ft
c. 7 ft
d. 8 ft
30. In which type of monitoring device do photons release electrons by their interaction with
air?
a. film badge
b. TLD
c. ionization chamber
d. pair production
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
a. Lithium fluoride
b. Silver halide
c. Sodium iodide
d. Cesium iodide
33. The Philippines regulatory body in charge of safeguarding personnel and public from the
substances that make use of ionizing radiation.
a. PNRI
b. DOH
C. DOE
d. BHDT
34. A device for radiation detection and measurement that measures the amount of current flow
resulting from an interaction between ionizing radiation and suitable gas
a. Thermoluminescence type
b. Photographic emulsion
c. Scintillation counter type
d. lonization chamber
35. Which of the following factors will affect both the quality and the quantity of the primary
beam?
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
36. The generation of personnel radiation monitor devices depends on which of the following?
1. ionization
2. thermoluminescence
3. resonance
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
37. Which records exposure by discharging a wire due to ionization of the air within a chamber?
a. pocket dosimeter
b. film badge
c. cutie pie
d. TLD
38. A laboratory technician is working with a radiactive source tha gives off a dose of 75 µSv/hr
at his work station Calculate how long the technician can work in the said source so that he will
not exceed the daily exposure limit of 0.20 mSv?
a. 2.67 hr
b. 0.367 hr
c. 3.67 hr
d. 0.267 hr
39. The principle of operation of this instrument is the production of visible light which is
detected and measured.
a. dose
b. distance
c. exposure
d. height
41. According to ICRP, the radiation dose limit for persons working with radiation is;
a. 50 mSv/yr
b. 5 mSv/yr
c. 20 mSv/yr
d. 2 mSv/yr
42. A radiation measuring device using crystal which has to be heated in order to emit light
proportional to radiation absorbed dose.
a. Film badge
b. GM counter
c. Scintillation detector
d. TLD
43. A laboratory technician is working with a radioactive source that gives off a dose rate of 60
µSv/hr at his station. Calculate how long the technician can work from the said source so that he
will not exceed the daily exposure limit of 0.2 mSv.
a. 4.23 hr
b. 4 hr
c. 3.33 hr
d. 2 hr
44. The annual dose limit for medical imaging personnel includes radiation from:
a. 1 only
b. 2 only
c. 2 and 3
d. 1, 2 and 3
a. Geiger-Muller counter
b. Scintillation detector
c. Photostimulable phosphor
d. Photomultiplier tube
46. Which of the following can be used as a monitoring device that records exposure to
radiation by recording a silver-halide-based latent image?
a. TLD
b. photostimulable phosphor
c. Radiographic emulsion
d. Geiger-Muller counter
48. Which of the following is a device used for measuring the quantity of ionizing radiation?
a. densitometer
b. dosimeter
c. sensitometer
d. galvanometer
49. A system of dose limitation which state that no practice involving radiation shall be adopted
unless its introduction produces a positive net benefit.
a. ALARA
b. Justification
c. Optimization
d. Dose Limit
50. How does a film badge distinguish between different types of radiation such as heat, light, x-
ray, neutrons, etc?
1. flat contact
2. shaped contact (contour)
3. shadow
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
52. What is the most commonly employed detecting device for scattered radiation?
a. Scintillation counter
b. Pocket dosimeter
c. TLD
d. lonization chamber
53. What term describes the fraction of the time that a radiation beam is directed at a specific
barrrier?
a. workload factor
b. attenuation factor
c. Use factor
d. Occupancy factor
54. What is the primary purpose of placing filter in the primary beam of a diagnostic radiographic
unit?
55. Which of the following sets of technical factors would most likely produce the least patient
exposure?
56. Which of the following intensifying screen relative speeds would produce the minimum
patient exposures?
a. 40
b. 100
c. 250
d 400
57. Which of the following projections would afford the greatest radiation protection to the lens
of the patient's eye?
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
59. What is the minimum distance for mobile radiographic exposure switch cord?
a. 1 ft
b. 5 ft
c. 6 ft
d. 20 ft
60. What term describes filtration that is part of the x-ray tube, housing and collimator?
61. What is the maximum time permitted on a fluoroscopic timer before it can be reset?
a. 2 minutes
b. 5 minutes
c. 10 minutes
d. 15 minutes
62. Which of the following structures would receive the greatest radiation exposure from a
abdomen?
a. liver
b. stomach
c. Kidneys
d. Urinary bladder
64. What is the minimum source-to skin distance for mobile fluoroscopic equipment?
a 12 cm (5")
b. 30 cm (12")
c. 38 cm (15")
d. 50 cm (20")
65. Acceptable method (s) of minimizing motion unsharpness is/are:
1. suspended respiration
2. short exposure time
3. patient instruction
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1,2 and 3
66. What minimum total amount of filtration (inherent plus added) is required in x-ray equipment
a. 2.5 mm Al equivalent
b. 3.5 mm Al equivalent
c. 2.5 mm Cu equivalent
d. 3.5 mm Cu equivalent
67. What is the minimum source-to skin distance for stationary fluoroscopic equipment?
a. 12 cm (5 in)
b. 30 cm (12 in)
c. 38 cm (15 in)
d. 50 cm (20 in)
68. Guidelines for the use of protective shielding state that gonadal shielding should be used
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 2 and 3 only
1. background radiation
2. leakage radiation
3. scattered radiation
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
72. How does the rare earth intensifying screen contribute to lowering the patient dose?
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 2 and 3 only
74. What is the required protective shielding of a conventional fluoroscopic barrier for operation
at less than 125 kVp?
a. 1.0 mm Pb/Eq
b. 2.0 mm Pb/Eq
c. 3.0 mm Pb/Eq
d. 4.0 mm Pb/Eq
75. Which of the following would be the best choice for an assistant when a pediatric patient
must be held during exposure?
a. student radiographer
b. radiology department aide
c. Nurse
d. Parent
76. Which of the following groups of exposure factors would deliver the lowest patient dose?
77. The collimator light and the actual irradiated area must be accurate to within;
a. 2% of the SID b. 5% of the SID c. 10% of the SID d. 15% of the SID
78. What is the minimum thickness for fluoroscopic protective aprons?
a. 0.15 mm Pb/Eqc
b. 0.25 mm Pb/Eq
c. 2.5 mm Al/Eq
d. 2.5 mm Pb/Eq
79. Inherent and added filtration in the x-ray tube function to;
80. Which of the following function (s) to reduce the amount of scattered radiation reaching the
film?
1. Grid devices
2 Restricted focal spots
3. Beam restrictors
a. 1 only
c. 1 and 3 only
b. 1 and 2 only
d. 1, 2 and 3
a. 0.15 mm Pb/Eq
b. 0,25 mm Pb/Eq
c. 2.5 Al/Eq
d. 2.5 Pb/Eq
82. Which of the following may be used to control the production of scatter radiation?
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
83. How can the radiographer reduce the amount of scatter radiation generated during a
radiographic examination?
84. Which of the following result (s) from the restriction of the x-ray beam?
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
86. The most efficient type of male gonadal shielding for use during fluoroscopy is,
a. flat contact
b. shaped contact
c. Shadow
d. Cynical
a. beam restriction
b. shielding
c. high-kVp, low mAs factors
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
88. Factors that determine the amount of scattered radiation produced include
1. radiation quality
2. field size
3. pathology
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
89. Which of the following technical factors would result in the lowest patient exposure dose?
1. accurate positioning
2. high-kVp, low mAs factors
3. rare earth screens
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
1. film speed
2. intensifying screen speed
3. focal spot size
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3 only
1. control booth
2. lead aprons
3. x-ray tube housing
a. 2 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
93. Any wall that the useful x-ray beam can be directed toward is called a;
a. secondary barrier
b. primary barrier
c. Leakage barrier
d. Scattered barrier
94. Which type of shielding is suspended from the radiographic collimator to absorb the primary
beam photons before they reach the reproductive organs?
a. shaped contact
b. contour
c. Flat
d. Shadow
95. How does the use of an appropriate back-up reductions in patient exposure dose?
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
96. If all factors remain the same, which of the following would result in the lowest patient
exposure dose?
97. For exposure to 1 rad of each of the following ionizing radiations, which would result the
greatest dose to the individual?
98. What is the relationship between film/screen combination and patient exposure dose?
99. Which of the following sets of technical factors would most likely produce the least patient
exposure?
a. 40
b. 100
c. 250
d. 400
101. Which of the following projections would afford the greatest radiation protection to the lens
of the patient's eye?
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
103. What is the minimum distance for mobile radiographic exposure switch cord?
a. 1 ft b. 5 ft
c. 6 ft d. 20 ft
104. What is the maximum time permitted on a fluoroscopic timer before it can be
reset?
a. 2 minutes b. 5 minutes
c. 10 minutes d. 15 minutes
105. Which of the following structures would receive the greatest radiation exposure
from a PA abdomen?
a. liver b. stomach
c. Kidneys d. Urinary bladder
a. teletherapy b. brachytherapy
c. crossfire treatment d. Linear accelerator
110. The determination of the location and extent of the tumor as related to the
surrounding anatomic landmarks is known as;
a. simulation b. Localization
c. field selection d. verification
1. interarticular
2. interstitial
3. intracavitary
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 ony
d. 1, 2 and 3
1. suspended respiration
2. short exposure time
3. patient instruction
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1,2 and 3
113. What is the minimum source-to skin distance for stationary fluoroscopic
equipment?
a. 12 cm (5 in)
b. 30 cm (12 in)
c. 38 cm (15 in)
d. 50 cm (20 in)
a. inherent filtration
b. added filtration
c. SID
d. Focal spot size
115. How does the rare earth intensifying screen contribute to lowering the patient
dose?
a. 1 only
b. 1 and 2 only.
c. 1 and 3 only
d. 2 and 3 only
116. Which of the following would be the best choice for an assistant when a pediatric
patient must be held during exposure?
a. student radiographer
b. radiology department aide
c. Nurse
d. Parent
117. Which of the following groups of exposure factors would deliver the lowest patient
dose?
120. Which of the following function (s) to reduce the amount of scattered radiation
reaching the film?
1. Grid devices
2. Restricted focal spots
3. Beam restrictors
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
121. Which of the following may be used to control the production of scatter radiation?
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
122. How can the radiographer reduce the amount of scatter radiation generated during
a radiographic examination?
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
123. Which of the following result (s) from the restriction of the x-ray beam?
a. 1 only
b. 1 and 2 only
c. 2 and 3 only
d. 1, 2 and 3
124. Which of the following procedures would most likely contribute higher gonadal
dose to a female than a male?
a. AP abdomen
b. AP knee
c. Lateral hip
d. PA chest
125. Which of the following sets of technical factors would most likely produce the
greatest patient exposure?
126. Which of the following would most likely result in greatest skin dose?
a. short SID
b. high-kVp
c. Increased filtration
d. Increased mA
127. Which of the following projections would afford the greatest radiation protection to
the breast?
128. The use of all of the following will function to reduce patient dose except,
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
131. The risk of internal contamination among technologists is relatively higher in;
132. A Co-60 machine was found to exhibit high levels of leakage radiation during a
routine survey procedure. What will you do first?
133. Which of the following primary beam field sizes would produce the greatest
radiation exposure to a patient?
a. 4 x4"
b. 6 x14"
c. 8 x8"
d. 8 x10"
134. Which of the following technical factors would result in the lowest patient exposure
dose?
1. accurate positioning
2. high-kVp, low mAs factors
3. rare earth screens
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2 and 3
1. film speed
2. intensifying screen speed
3. focal spot size
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
137. If all factors remain the same, which of the following would result in the lowest
patient exposure dose?
138. What is the relationship between film/screen combination and patient exposure
dose?
140 Which of the following is likely to improve image quality and decrease patient dose?
1. beam restriction
2. low kVp, high mAs
3. grids
a. 1 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
141. Which of the following most effectively minimizes radiation exposure to the patient?
143. Which of the following technical factors would result in lowest patient exposure
dose?
1. fixed kVp
2. variable kVp
3. automatic exposure controls
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2 and 3
145. Which of the following technical factors would result in the greatest patient
exposure dose?
146. Which of the following technical factors would produce the greatest patient
exposure dose?
147. The international symbol for ionizing radiation is black on a yellow background.
a. triangle
b. trefoil
c. Propeller
d. Symbol
149. In radiation protection, the product of absorbed dose and the correct modifying
factor is used to determine,
a. cosmic rays
b. atmospheric radioisotopes
c. radiation therapy treatments
d. radioactive minerals
a. Gray
b. Joules/ kilogram
c. Sievert
d. rad
a. Roentgen
b. Becquerel
c. Gray
d. Sievert
In radiation protection, the product of absorbed dose and the correct modifying factor is used to
determine
a. Roentgen (C/kg)
b. rem (Sv)
c. rad (Gy)
d. radiation quality
a. Roentgen (C/kg)
b. rad (Gy)
c. rem (Sv)
d. RBE