D Ifiblp: Mskomno
D Ifiblp: Mskomno
d^ifiblp
pÉêîáÅÉ=j~åì~ä
båÖäáëÜ
GALILEOS Service Manual
Sirona Dental Systems GmbH
Service Manual GALILEOS
Table of contents
1 About this Service Manual........................................................................................ 12
båÖäáëÜ
2.1 Modifications to the unit ................................................................................ 15
2.2 Fixed connection ........................................................................................... 15
2.3 Electromagnetic compatibility........................................................................ 15
2.4 Electrostatic discharge .................................................................................. 15
2.5 Switching the unit on ..................................................................................... 15
2.6 Condensation ................................................................................................ 16
2.7 Laser light localizer ....................................................................................... 16
2.8 Ventilation slots ............................................................................................. 16
2.9 Qualifications of service personnel ............................................................... 16
2.10 Radiation protection ...................................................................................... 16
2.11 Safety checks................................................................................................ 16
3 Unit description......................................................................................................... 17
61 25 665 D3437
D3437.076.01.15.02 06.2012 3
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
4 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH
Service Manual GALILEOS
5 Messages ................................................................................................................. 82
båÖäáëÜ
5.5.2 Location 07: Easypad/DX7................................................................. 94
5.5.3 Location 10: System hardware........................................................... 98
5.5.4 Location 11: Power PC/Board DX11.................................................. 101
5.5.5 Location 12: CAN bus ........................................................................ 103
5.5.6 Location 14: Digital extension, SIDEXIS XG...................................... 104
5.5.7 Location 15: Configuration, update .................................................... 106
5.5.8 Location 41: Media interface card ...................................................... 107
5.5.9 Location 42: Remote control .............................................................. 111
5.5.10 Location 71: Multipad, board DX71.................................................... 115
5.5.11 Location 89: X-ray detector ................................................................ 119
6 Troubleshooting........................................................................................................ 125
61 25 665 D3437
D3437.076.01.15.02 06.2012 5
Sirona Dental Systems GmbH
Service Manual GALILEOS
7.1 General information about unit adjustment and calibration .......................... 157
7.1.1 Displays and help messages during adjustment/calibration ............. 158
7.1.2 "Service Functions" menu ................................................................. 159
7.1.2.1 Calling the "Adjustment/Calibration" menu...................................... 161
7.1.3 Enabling exposure readiness ............................................................ 162
7.1.4 Taking an exposure........................................................................... 162
7.1.5 Save values....................................................................................... 162
7.1.6 Test phantoms for adjustment and calibration .................................. 163
7.1.6.1 Distortion phantom .......................................................................... 163
7.1.6.2 Geometry phantom.......................................................................... 164
7.1.6.3 Constancy test phantom.................................................................. 166
7.2 Checking the mechanical system adjustment .............................................. 166
7.3 Adjustment and calibration via the "Service Functions" menu ..................... 167
7.3.1 Diaphragm image .............................................................................. 167
7.3.1.1 "Type 1/Type 2" diaphragm............................................................. 167
7.3.1.2 "Type 3" diaphragm......................................................................... 171
7.3.2 Checking the radiation field............................................................... 178
7.3.3 Dosimetry .......................................................................................... 179
7.3.4 Sensor adjustment ............................................................................ 181
7.3.5 Iris adjustment................................................................................... 182
7.3.6 Shading calibration............................................................................ 183
7.3.7 Distortion calibration.......................................................................... 185
7.3.8 Geometry calibration ......................................................................... 187
7.3.9 Service .............................................................................................. 189
7.4 Performing a white balance for FaceScan ................................................... 190
7.5 Checking and adjusting the touchscreen ..................................................... 191
61 25 665 D3437
6 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH
Service Manual GALILEOS
båÖäáëÜ
8.2 Service menu and service routines ............................................................... 206
8.2.1 Displays and symbols in the service menu ........................................ 206
8.2.1.1 Easypad ........................................................................................... 206
8.2.1.2 Multipad............................................................................................ 208
8.3 Basic operating procedures in the service menu .......................................... 210
8.3.1 Activating the service menu ............................................................... 210
8.3.1.1 Easypad ........................................................................................... 210
8.3.1.2 Multipad............................................................................................ 212
8.3.2 Selecting service routines and test steps........................................... 213
8.3.2.1 Selecting a service routine ............................................................... 213
8.3.2.2 Selecting a test step......................................................................... 214
8.3.2.3 Service routines with security access .............................................. 215
8.3.3 Select parameters .............................................................................. 217
8.3.4 Saving parameters ............................................................................. 219
8.3.5 Exiting the test step and service routine ............................................ 220
8.4 S002: Radiation without rotary movement, selectable kV/mA level and 221
maximum radiation time
8.4.1 S002: Test step 5 ............................................................................... 221
8.5 S005: General X-ray tube assembly service ................................................. 223
8.5.1 S005: Test step 1 ............................................................................... 223
8.5.2 S005: Test step 4 ............................................................................... 225
8.5.3 S005: Test step 5 ............................................................................... 226
8.5.4 S005: Test step 8 ............................................................................... 227
61 25 665 D3437
D3437.076.01.15.02 06.2012 7
Sirona Dental Systems GmbH
Service Manual GALILEOS
9 Repair....................................................................................................................... 288
båÖäáëÜ
9.3.1 Replacing the ring motor .................................................................... 301
9.3.2 Replacing the pinion at the ring motor ............................................... 303
9.3.3 Laying of cables when replacing the ring motor................................. 305
9.3.4 What has to be done after replacing the ring motor (M1_3)/pinion? .. 305
9.4 Rotary knob on the swivel arm ...................................................................... 306
9.4.1 Replacing the rotary knob .................................................................. 306
9.5 Control panel................................................................................................. 307
9.5.1 Replacing the Easypad user interface (GALILEOS Comfort) or 307
Multipad (GALILEOS Compact)
9.5.1.1 What has to be done after replacing the user interface? ................. 308
9.5.2 Laying of cables when replacing the user interface ........................... 309
9.6 X-ray tube unit............................................................................................... 310
9.6.1 Replacing the X-ray tube assembly ................................................... 310
9.6.2 Cables and connectors for replacement of the X-ray tube assembly. 316
9.6.3 What has to be done after replacing the X-ray tube assembly? ........ 317
9.7 Fan (X-ray tube assembly) ............................................................................ 318
9.7.1 Replacing the fan ............................................................................... 318
9.7.2 What has to be done after replacing the fan? .................................... 318
9.8 X-ray detector................................................................................................ 319
9.8.1 Replace X-ray detector ...................................................................... 319
9.8.2 What has to be done after replacing the X-ray detector?................... 324
61 25 665 D3437
D3437.076.01.15.02 06.2012 9
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
10 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH
Service Manual GALILEOS
båÖäáëÜ
11 Dismantling and disposal......................................................................................... 384
61 25 665 D3437
D3437.076.01.15.02 06.2012 11
1 About this Service Manual Sirona Dental Systems GmbH
1.1 Scope Service Manual GALILEOS
1.1 Scope
Scope
Scope of the Service Manual: GALILEOS
Wiring diagrams
● GALILEOS Wiring References: Order no. 61 25 640
GALILEOS installation
Installation Instructions
● GALILEOS Comfort: Order no. 61 25 574
● GALILEOS / ORTHOPHOS XG 3D Software installation: Order no.
61 42 389
● GALAXIS Operator's Manual: Order no. 61 23 488
● SIDEXIS XG Digital Radiography Installation Instructions:
Order no. 59 67 356
DVD text
You can order the technical documentation in paper form free of change
from our Customer Service Center by specifying the above mentioned
order numbers (REF).
A large portion of the technical documentation is also available our the
product DVD. To call up the instructions, insert the DVD in the DVD drive
of the PC. The DVD then starts automatically and a start screen opens.
The latest documentation can always be downloaded from the Sirona
homepage (www.sirona.de HOME ⇨ Service ⇨ Technical
Documentation).
61 25 665 D3437
12 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 1 About this Service Manual
Service Manual GALILEOS 1.3 Tools and auxiliary materials
båÖäáëÜ
● Mult-O-Meter 512L
● Soldering tool for repairing cables
● Cable ties
● Teflon tape
● Loctite
61 25 665 D3437
D3437.076.01.15.02 06.2012 13
1 About this Service Manual Sirona Dental Systems GmbH
1.4 Structure of the document Service Manual GALILEOS
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury
or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product
or an object in its environment.
IMPORTANT
Application instructions and other important information.
The formats and symbols used in this document have the following
meaning:
61 25 665 D3437
14 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 2 Safety instructions
Service Manual GALILEOS 2.1 Modifications to the unit
2 Safety instructions
Safety instructions
Modifications to this unit which might affect the safety of the system
owner, patients or other persons are prohibited by law!
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user is responsible for any damage resulting from the use
of non-approved accessories.
DANGER
Potentially lethal shock hazard!
båÖäáëÜ
Fixed connection!
Installing a mains plug instead of the specified fixed connection infringes
international medical regulatory actions and is prohibited. In case of
error, this puts patients, users, and other parties seriously at risk.
61 25 665 D3437
D3437.076.01.15.02 06.2012 15
2 Safety instructions Sirona Dental Systems GmbH
2.6 Condensation Service Manual GALILEOS
2.6 Condensation
Condensation
Safety information for condensation: Service engineer
Never cover the ventilation slots on the unit under any circumstances,
since this may obstruct air circulation. This can cause the unit to overheat.
61 25 665 D3437
16 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.1 Unit classes and versions
3 Unit description
Unit description
båÖäáëÜ
B
A GALILEOS unit*
B Control panel
C Facescan (optional)
*The unit is prepared for optional operation with Facescan from the
following unit serial numbers:
61 25 665 D3437
D3437.076.01.15.02 06.2012 17
3 Unit description Sirona Dental Systems GmbH
3.1 Unit classes and versions Service Manual GALILEOS
control panel (B). While the GALILEOS Comfort has a control panel with
GALILEOS Comfort
a color touchscreen (Easypad), the Compact has a simpler control panel
with a single-line display (Multipad). There are slight variations in how the
two unit classes operate because of the difference in the control panels.
GALILEOS Compact
Prog. S kV mA
T R
61 25 665 D3437
18 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
3.2 Hardware
Hardware
båÖäáëÜ
touched or interconnected without ESD protective measures. See also
"Electrostatic discharge" and "Electromagnetic compatibility" [ → 15].
61 25 665 D3437
D3437.076.01.15.02 06.2012 19
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
A B
61 25 665 D3437
20 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
båÖäáëÜ
D A
A Slide
B Stand
C Remote control [ → 24]
D FACESCAN [ → 24] (optional)
61 25 665 D3437
D3437.076.01.15.02 06.2012 21
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
3.2.3.1 Slide
Slide
MU
LS
DX11
DX1
DX6*
DX89
LS*
DX7*
GALILEOS Comfort
DX71*
GALILEOS Compact
61 25 665 D3437
22 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
3.2.3.2 Stand
Stand
Stand: GALILEOS
MHV
DX32*
båÖäáëÜ
DX41**
*) From unit serial number 3101, new units are supplied with a new
version of the board DX32 (see chapter "Board photos [ → 31]").
**) Starting with unit serial number 3201, new units will be delivered
without board DX41.
61 25 665 D3437
D3437.076.01.15.02 06.2012 23
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
DX42
3.2.3.4 Facescan
Facescan
61 25 665 D3437
24 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
L1
L5 DX11
DX1
L2 DX89
DX6
H1
L3
L3
båÖäáëÜ
A L4
L4
DX32
L1
L2
L5 B
A Power switch
B Line filter
C Wago terminal
61 25 665 D3437
D3437.076.01.15.02 06.2012 25
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
L6
DX1 / J306 – J307
5922526
DX6 / J2 – J3
L6_XG
J307
DX1/
526
5922
L6
J306
L6 526 DX1/
L15
5922
L6 DX11
DX1 DX89
L9
DX6
L9
H1
L10 L10
L15 DX7/DX71
A Power switch
61 25 665 D3437
26 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
L13
DX11
L19 DX1 DX89
L16 L29
LS
DX6
L16
HV
H1
L19
båÖäáëÜ
L27
DX7/DX71
L28*
L29
A Power switch
61 25 665 D3437
D3437.076.01.15.02 06.2012 27
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
L7
L8 DX1
L17
L25/L26
DX42
L7 L8
L17
DX41
E
L25
E Media converter
61 25 665 D3437
28 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
L7
L108 DX1
L117
båÖäáëÜ
L25/L26
DX42
L108 L7
L117
E
L25
E Media converter
61 25 665 D3437
D3437.076.01.15.02 06.2012 29
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
Facescan cabling
L77 L78
DX11
DX1 DX89
L77
A FACESCAN
L77
L78
PoE
A Power switch
61 25 665 D3437
30 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
Boards DX1/DX11 Installed up to unit serial number 8499 for "GALILEOS Comfort" and
48499 for "GALILEOS Compact".
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 31
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
Boards DX1/DX11 Installed from units with a serial number of 8500 and above for
"GALILEOS Comfort", and
48500 and above for "GALILEOS Compact"
IMPORTANT
The DX1/DX11V2 board can only be operated with unit software version
V04.04.00 or higher.
61 25 665 D3437
32 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
Board DX6 This board is not available as a spare part or a repair part. X-ray tube
assemblies can only be ordered as complete units.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 33
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
Board DX7
DX7: GALILEOS restriction
DX71: GALILEOS
61 25 665 D3437
34 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
Board DX71 This board is only used in the "GALILEOS Compact" (not in "GALILEOS
Comfort").
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 35
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
DX89: GALILEOS
Board DX89
61 25 665 D3437
36 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
båÖäáëÜ
Board DX32
DX32: GALILEOS: from serial no. 3201
A Line filter
DX41: GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 37
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
Board DX41 This board is not available as a repair part or spare part.
Board DX41 is omitted in units with a serial number of 3201 and above.
61 25 665 D3437
38 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
Board DX42 This board is not available as a spare part or a repair part.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 39
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
FACESCAN modular board This modular board is not available as a spare part or a repair part.
Facescan can only be ordered as a complete unit.
61 25 665 D3437
40 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 41
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
3.2.6 Covers
Covers
61 25 665 D3437
42 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
Covers: GALILEOS
GALILEOS
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 43
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
61 25 665 D3437
44 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
båÖäáëÜ
X-ray tube assembly: Focal spot size acc. to IEC 60336,
measured in the central X-ray beam: 0.5
kV: 85 kV
mA: 5 mA/7 mA
Pulsed mode: 10 ms – 30 ms
Total filtration of X-ray tube assembly > 2.5 Al / 90 IEC 60522
Cone-beam angle: collimated to approx. 24°
High voltage generation frequency: 80 kHz – 100 kHz
61 25 665 D3437
D3437.076.01.15.02 06.2012 45
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
Reconstruction:
Marking of focal spot:
61 25 665 D3437
46 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
båÖäáëÜ
Network: Network: 100 MB Ethernet, 1 Gbit Ethernet
recommended
Communication interface: RJ45 for LAN cable
61 25 665 D3437
D3437.076.01.15.02 06.2012 47
3 Unit description Sirona Dental Systems GmbH
3.2 Hardware Service Manual GALILEOS
3.2.7.1 Diagrams
Diagrams
61 25 665 D3437
48 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.2 Hardware
10°
Anode angle
båÖäáëÜ
16°
61 25 665 D3437
D3437.076.01.15.02 06.2012 49
3 Unit description Sirona Dental Systems GmbH
3.3 Software/compatibility Service Manual GALILEOS
3.3 Software/compatibility
Software/compatibility
Any software combinations other than those listed here are not allowed.
If a module software version does not match the main software version,
the main software version is identified with an asterisk on the info screen
(e.g. 04.03.01*).
Main software V03.03.02
GALILEOS Remote control
Board Software Board Software
DX6 02.88.00 DX42 02.45.06
DX7 02.57.00
DX7-L0 02.18.00
DX7-L1 02.18.00
DX7-L2 02.18.00
DX7-L3 02.18.00
DX7-L4 –
DX7-L5 –
DX71 –
DX11 02.61.01
DX41 02.30.00
DX89 01.10.06
DX89 FPGA 01.13.01
61 25 665 D3437
50 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.3 Software/compatibility
båÖäáëÜ
DX89 FPGA 01.13.01
61 25 665 D3437
D3437.076.01.15.02 06.2012 51
3 Unit description Sirona Dental Systems GmbH
3.3 Software/compatibility Service Manual GALILEOS
61 25 665 D3437
52 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.3 Software/compatibility
båÖäáëÜ
DX89 FPGA 01.15.00
61 25 665 D3437
D3437.076.01.15.02 06.2012 53
3 Unit description Sirona Dental Systems GmbH
3.3 Software/compatibility Service Manual GALILEOS
61 25 665 D3437
54 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.3 Software/compatibility
båÖäáëÜ
DX89 01.56.00
DX89 FPGA 01.54.00
61 25 665 D3437
D3437.076.01.15.02 06.2012 55
3 Unit description Sirona Dental Systems GmbH
3.3 Software/compatibility Service Manual GALILEOS
GALILEOS CD Remarks
Software index
V1.2 004 Requires unit main software V 03.03.01 and
SIDEXIS 2.0.
V1.4 005 Requires unit main software V 03.04.00 and
SIDEXIS 2.2.
V1.4.3 SW 007 Requires unit main software V 03.04.01 and
SIDEXIS 2.2.
V1.5 009 Requires unit main software V 03.05.00 and
SIDEXIS 2.3.
V1.6 011 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.01 and SIDEXIS 2.3 or
above.
V1.6.1 012 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
V1.7 013 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
V1.7.1 014 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
V1.7.1.1 015 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
61 25 665 D3437
56 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 3 Unit description
Service Manual GALILEOS 3.3 Software/compatibility
GALILEOS CD Remarks
Software index
V1.7.2 016 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
V1.7.4 002 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.3 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.3 or
above.
V1.8 001 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.5.1 or
båÖäáëÜ
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.5.1
or above.
V1.9 003 For "GALILEOS Comfort": Requires unit main
software V 03.05.00 and SIDEXIS 2.5.1 or
above.
For "GALILEOS Compact": Requires unit
main software V 03.06.02 and SIDEXIS 2.5.1
or above.
V1.10 004 For "GALILEOS Comfort": Requires unit main
software V 04.09.00 and SIDEXIS 2.5.3 or
higher.
For "GALILEOS Compact": Requires unit
main software V 04.09.00 and SIDEXIS 2.5.3
or higher.
61 25 665 D3437
D3437.076.01.15.02 06.2012 57
4 General operating procedures Sirona Dental Systems GmbH
4.1 Switching the unit on Service Manual GALILEOS
WARNING
X-rays
Be sure to observe the radiation protection regulations applicable in
your country.
➢ No person may be positioned in the unit when it is switched on.
NOTICE
Damage to the unit
Check the room height before you raise the unit.
➢ If the room height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2")
for installation with the floor stand, you must limit the maximum
travel height [ → 272].
NOTICE
Fluctuations in temperature can cause condensation to form in the unit.
Electrical components are destroyed by short circuits.
➢ Do not switch the unit on until the temperature of the unit has
adapted to the ambient temperature and the condensation has
evaporated.
Waiting time when switching on and off
NOTICE
The unit must not be switched on/off constantly.
Constant switching on and off reduces the service life of individual unit
components and results in increased power consumption.
➢ After switching the unit off, wait for approx. 60 seconds before
switching it on again.
61 25 665 D3437
58 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.1 Switching the unit on
NOTICE
The surface of the touchscreen is sensitive.
The touchscreen can be damaged or its surface scratched.
➢ Never use pointed objects such as ballpoint pens, pencils, etc. to
operate the touchscreen.
➢ Only use your fingertips to operate the touchscreen.
IMPORTANT
After the unit is switched on, the touchscreen has only limited readability
for several minutes until the background lighting has completed its
warm-up phase.
After the unit is switched off with the main switch, the touchscreen
remains lit for approx. another 3 to 5 seconds.
båÖäáëÜ
1. Turn the main switch (A) to position I.
2. Wait for approx. 1 minute.
A B ª The X-ray radiation indicator (B) lights up for approx. 1 second as
a functional check.
ª After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
C ª The start screen is displayed on the touchscreen for several
seconds.
ª The program selection is then displayed on the touchscreen.
3. Check whether the patient symbols on the touchscreen can be
selected in exactly the right position.
If problems occur during selection, adjust the touchscreen [ → 191].
4. Press the R key.
ª The unit moves to its starting position.
5. Switch on the PC.
6. Start SIDEXIS XG.
ª As long as no connection has been made to SIDEXIS XG, the
message "Switch SIDEXIS to ready for exposure state" is
displayed in the comment line of the control panel.
61 25 665 D3437
D3437.076.01.15.02 06.2012 59
4 General operating procedures Sirona Dental Systems GmbH
4.1 Switching the unit on Service Manual GALILEOS
61 25 665 D3437
60 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.2 Updating the firmware
Read the information provided on the software CD supplied with the unit
and on the SIRONA dealer page on the Internet very carefully. These
sources always contain the latest information on software updates.
1. Start the "SIDEXIS Manager" under "Start"/"Programs"/"SIDEXIS"/
"SIDEXIS XG".
2. Click on "Configuration of the X-ray components".
ª The "Configuration of the X-ray components" menu opens.
båÖäáëÜ
3. Select the "Attributes" tab.
4. Click on the "Software update" button.
ª The dialog box for entering the service password opens.
5. Enter the service password.
Enter the first 4 digits of the current system date in reverse order as
the service password (e.g. on 05/24/1995, 5042 must be entered as
the service password.
If an incorrect service password or no password at all is entered, the
limited update menu for users will be started. This only supports an
automatic update option.
ª The dialog box for selecting the installation source opens.
61 25 665 D3437
D3437.076.01.15.02 06.2012 61
4 General operating procedures Sirona Dental Systems GmbH
4.2 Updating the firmware Service Manual GALILEOS
7. Select the path or the desired update file from the list and click "Open"
to confirm the selection.
The update file is located on the unit software CD. It is delivered with
each DX11 replacement board and also included in the country set.
The contents of the CD can be downloaded from the Dealer domain
of the SIRONA Internet home page (under Product Info/X-ray
Systems): www.sirona.com
8. Click on the "OK" button.
ª The software manager opens.
ª The left-hand window of the software manager displays the
modules and their current software versions.
ª In the left-hand window of the software manager, you can now
select the update modes "Automatic" and "Main version" using
the tabs.
9. Select the desired mode for the software update (see chapter entitled
"Update mode").
61 25 665 D3437
62 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.2 Updating the firmware
NOTICE
Unit inoperability!
Before starting the software update, make sure that no unit movements
are active. Otherwise the system may become inoperable in rare cases.
The X-ray detector must be installed as part of the update. Exposure
readiness must be deselected in SIDEXIS XG and the unit must not
already be in service mode.
61 25 665 D3437
D3437.076.01.15.02 06.2012 63
4 General operating procedures Sirona Dental Systems GmbH
4.2 Updating the firmware Service Manual GALILEOS
NOTICE
Effectiveness of the software update
The unit must be restarted after every software update. The new DX11
version will not run until the unit has been rebooted (see also chapter
"Measures following replacement of boards [ → 347]").
Any errors with the consecutive numbers 01, 03, 04, 06 or 07 displayed
immediately following the software update may be ignored. If these
messages appear again after the unit is rebooted, perform
troubleshooting as described in the section entitled "Error
messages [ → 85]".
If anything conspicuous occurs in connection with unit handling on
completion of the software update and restart of the unit, please repeat
the software update as the first measure.
12. Click on the "Show logfile" button and use the log files to check
whether the update was successfully performed.
If it features entries such as "Update of DXxx failed!", please perform
the update again. Repeat this procedure as often as necessary until
the "failed messages" no longer appear.
61 25 665 D3437
64 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.2 Updating the firmware
You can select two different update modes via the tabs "Automatic" or
"Main version":
● "Automatic"
The software of all components is automatically updated to the latest
software version.
The right window displays a list of the modules, their installed
software version and the latest software version offered by the
update.
båÖäáëÜ
● "Main version"
The software can be upgraded or downgraded to the desired version.
This update mode is required, for example, if a replacement module
arrives from the warehouse and features a newer release than the
existing main release of the unit. In this case, a main version update
to the overall system status (displayed on the info screen) must be
performed for the corresponding component with the appropriate
update file (*.SUI). The module is then reprogrammed.
The colored bars in front of the software releases indicate their validity
(see chapter entitled "Using the update manager to check the program
releases").
61 25 665 D3437
D3437.076.01.15.02 06.2012 65
4 General operating procedures Sirona Dental Systems GmbH
4.2 Updating the firmware Service Manual GALILEOS
which modules are connected to the unit and what their latest program
release is.
NOTE: From unit serial number 8500 (for "GALILEOS Comfort") or 48500
(for "GALILEOS Compact") and above, a DX11_FPGA version is also
displayed in addition to this list.
61 25 665 D3437
66 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.2 Updating the firmware
Opening Facescan
1. Unscrew the cover (A) from the Facescan.
båÖäáëÜ
2. Pull gray cable L78.4 from slot X2 of the FACESCAN modular board.
3. Unscrew the protective plate (T) from the Facescan unit.
ª The FACESCAN modular board is visible.
4. Plug cable L78.4 again into slot X2 on the FACESCAN modular
board.
61 25 665 D3437
D3437.076.01.15.02 06.2012 67
4 General operating procedures Sirona Dental Systems GmbH
4.2 Updating the firmware Service Manual GALILEOS
1. Insert the USB stick into the USB port (U) of the FACESCAN modular
board.
2. Switch the unit on again.
ª The Facescan will be updated.
Both LEDs in the status display light up during the update.
The process takes around 5 minutes (around 1 minute in the
event of an error).
ª The "USB_STICK_MODE" entry in the facescan_settings.cfg file
is reset to "CONFIG".
3. Wait until the green status display LED goes out.
The blue LED should then light up.
ª The update is completed.
4. Switch GALILEOS off.
5. Remove the Facescan USB stick safely from the USB port.
61 25 665 D3437
68 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.2 Updating the firmware
Closing Facescan
1. Pull gray cable L78.4 from slot X2 of the FACESCAN modular board.
2. Screw the protective plate (T) onto the Facescan unit.
båÖäáëÜ
3. Plug cable L78.4 again into slot X2 on the FACESCAN modular
board.
61 25 665 D3437
D3437.076.01.15.02 06.2012 69
4 General operating procedures Sirona Dental Systems GmbH
4.2 Updating the firmware Service Manual GALILEOS
61 25 665 D3437
70 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.3 Configuring the unit
The X-ray component must be set up and enabled using the SIDEXIS
Manager. For more information, refer to SIDEXIS XG Installation
Instructions.
båÖäáëÜ
4.3.2.1 Type 1: Configuration using the Facescan USB stick
Type 1: Configuration using the Facescan USB stick
61 25 665 D3437
D3437.076.01.15.02 06.2012 71
4 General operating procedures Sirona Dental Systems GmbH
4.3 Configuring the unit Service Manual GALILEOS
NOTICE
Always switch the device on before inserting the USB stick!
Before the Facescan USB stick can be inserted into the USB socket of
the FACESCAN modular board, GALILEOS must be switched on.
Otherwise, the Facescan configuration data will be reset to the factory
settings.
7. Insert the Facescan USB stick into the USB port of the FACESCAN
board.
ª The Facescan will be configured.
ª Both LEDs of the status display light up.
8. Wait until the light on both the LEDs of the status display goes out
(process lasts some 10 secs).
NOTICE
Faults during configuration
If there has been an error during configuration, only the blue LED of the
status display goes out.
Starting the device again is not necessary here.
➢ In the event of an error, check the log file "facescan_settings.log" on
the USB stick.
➢ Make sure to read section Syntax of the configuration file
"facescan_settings.cfg" [ → 74]“.
➢ Repeat the configuration process.
61 25 665 D3437
72 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.3 Configuring the unit
Closing Facescan
1. Pull gray cable L78.4 from slot X2 of the FACESCAN modular board.
2. Screw the protective plate (T) onto the Facescan unit.
båÖäáëÜ
3. Plug cable L78.4 again into slot X2 on the FACESCAN modular
board.
61 25 665 D3437
D3437.076.01.15.02 06.2012 73
4 General operating procedures Sirona Dental Systems GmbH
4.3 Configuring the unit Service Manual GALILEOS
IMPORTANT
Pay attention to the syntax!
Text entries should never have a space before and after "=".
Examples:
Correct: USB_STICK_MODE=CONFIG
Incorrect: USB_STICK_MODE= CONFIG
IMPORTANT
Changes to network addresses.
➢ Adjust the entries "IP" (IP address) and "Netmask" (subnet mask) as
required.
61 25 665 D3437
74 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.3 Configuring the unit
båÖäáëÜ
(http://<IP-Adresse>).
ª A password dialog box opens.
4. In the field "User" enter "service".
5. In the field "Password" enter "sirona".
ª The "Facescan Device Service " web dialog opens.
Selecting the configuration dialog
61 25 665 D3437
D3437.076.01.15.02 06.2012 75
4 General operating procedures Sirona Dental Systems GmbH
4.3 Configuring the unit Service Manual GALILEOS
Reset
1. Press the "Settings Reset" button.
ª A reset dialog opens.
NOTICE
Complete loss of user data
When the unit is reset to factory settings, all user data is overwritten,
including white balance and calibration data.
The network settings are, however, retained.
61 25 665 D3437
76 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.4 Reading unit data
"SIDEXIS XG".
2. Click on "Configuration of the X-ray components".
ª The "Configuration of the X-ray components" menu opens.
båÖäáëÜ
automatically.
61 25 665 D3437
D3437.076.01.15.02 06.2012 77
4 General operating procedures Sirona Dental Systems GmbH
4.4 Reading unit data Service Manual GALILEOS
61 25 665 D3437
78 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.5 Using demo mode – operation without radiation release
For demo use, the "X-ray detector dummy for GALILEOS" (Order No. 61
19 007) should be used instead of the actual X-ray detector. For further
information, please refer to the instructions included with the dummy.
If the volume tomograph is to be presented as a demo unit at trade fairs
or exhibitions, it must be ensured that radiation release is blocked.
When operated in demo mode, the unit must not release any radiation.
For this reason, you must take the following safety measures:
1. Switch off the unit.
DANGER
båÖäáëÜ
Potentially lethal shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.
3. Loosen the screws (A) and remove the cover plate (B).
A
61 25 665 D3437
D3437.076.01.15.02 06.2012 79
4 General operating procedures Sirona Dental Systems GmbH
4.5 Using demo mode – operation without radiation release Service Manual GALILEOS
6. Switch on the unit and check the mode on the info screen.
Demo mode: ON means that: Demo mode is switched on (radiation
release is not possible)
Demo mode: OFF means that: Demo mode is switched off
(radiography, X-ray radiation are possible!)
7. Switch the unit off again and reattach the cover plate (B) and the tube
assembly covers by following the dismantling procedure in reverse
order.
61 25 665 D3437
80 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 4 General operating procedures
Service Manual GALILEOS 4.5 Using demo mode – operation without radiation release
DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.
båÖäáëÜ
release is not possible)
Demo mode: OFF means that: Demo mode is switched off
(radiography, X-ray radiation are possible!)
7. Switch the unit off again and reattach the cover plate (B) and the tube
assembly covers by following the dismantling procedure in reverse
order.
61 25 665 D3437
D3437.076.01.15.02 06.2012 81
5 Messages Sirona Dental Systems GmbH
5.1 Help messages Service Manual GALILEOS
5 Messages
Messages
The help messages are displayed as help codes (Hx xx) on the Easypad
touchscreen (GALILEOS Comfort) or on the Multipad display (GALILEOS
Compact) as well as on the display of the remote control (if present). The
codes tell you how to operate the system if radiation release is not
possible due to a previous operator error.
The following list provides you with an overview of all help codes, their
meaning and the action required to eliminate the corresponding
problems.
61 25 665 D3437
82 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.1 Help messages
IMPORTANT: The measures listed only clear help messages that result
from operator errors. If it is not possible to clear a message by taking the
measures listed, another type of error is the cause. In such cases, you
should run an error diagnosis [ → 88].
båÖäáëÜ
H4 03 "Switch SIDEXIS to ready for exposure state" ● Make SIDEXIS XG ready for exposure.
H4 07 "SIDEXIS 3D Vorauswahl korrigieren" Correct SIDEXIS XG 3D preselection.
H4 08 "SIDEXIS 3D Aufnahme wählen" Correct SIDEXIS XG 3D preselection.
H4 20 "Get existing exposure" IMPORTANT: Do not switch the system off until
the help message has disappeared.
● Get exposure with "Sirona Control Admin"
(see SIDEXIS XG"Operator’s Manual" (REF
59 62 134).
61 25 665 D3437
D3437.076.01.15.02 06.2012 83
5 Messages Sirona Dental Systems GmbH
5.2 System messages Service Manual GALILEOS
On the Facescan
Facescan operating conditions
61 25 665 D3437
84 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.4 Error messages
The error messages are displayed as error codes (Ex yy zz) on the
Easypad touchscreen (GALILEOS Comfort) or on the Multipad display
(GALILEOS Compact) as well as on the remote control display (if there is
one).
The codes provide you with error type, error location and troubleshooting
information.
båÖäáëÜ
5.4.2 Ex - Error type
Ex - Error type
61 25 665 D3437
D3437.076.01.15.02 06.2012 85
5 Messages Sirona Dental Systems GmbH
5.4 Error messages Service Manual GALILEOS
61 25 665 D3437
86 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.4 Error messages
5.4.3 yy - Location
yy - Location
Identifier yy defines the location or logical function unit where the error
has occurred.
båÖäáëÜ
41 Media interface card DX41
42 Remote control DX42
89 X-ray detector DX89
61 25 665 D3437
D3437.076.01.15.02 06.2012 87
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
In the following table, the error codes are sorted by the location or function
unit where the error has occurred. For enhanced clarity, the
corresponding ID in the error code is printed in bold type.
E6 06 02 GALILEOS
E6 06 03
E6 06 04
E6 06 05
61 25 665 D3437
88 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 06 06
E6 06 07
båÖäáëÜ
addressed by the master: (see wiring diagrams)
E6 06 08
61 25 665 D3437
D3437.076.01.15.02 06.2012 89
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 06 10
E7 06 12
E6 06 13
E2 06 20
E6 06 21
E6 06 22
61 25 665 D3437
90 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E3 06 23
E5 06 30
båÖäáëÜ
● Check the CAN bus. S. [ → 128]
● Please contact the Sirona Customer S. [ → 61]
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
E5 06 31
E5 06 32
E1 06 40 GALILEOS
E1 06 41 GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 91
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E1 06 42 GALILEOS
E1 06 43 GALILEOS
E1 06 44 GALILEOS
E1 06 45 GALILEOS
E6 06 51 GALILEOS
E6 06 52 GALILEOS
E6 06 53 GALILEOS
E6 06 54
E6 06 55
E6 06 56
61 25 665 D3437
92 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 06 60
E6 06 65
E6 06 66
båÖäáëÜ
if necessary. S. [ → 310]
E6 06 67
E6 06 68 GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 93
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 07 02
E6 07 03
E6 07 04
E6 07 05
61 25 665 D3437
94 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 07 06
E6 07 07
båÖäáëÜ
addressed by the master: (see wiring diagrams)
E6 07 08
61 25 665 D3437
D3437.076.01.15.02 06.2012 95
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 07 10
E7 07 12
E6 07 20
E6 07 21
61 25 665 D3437
96 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E3 07 30/33/34/35/36 GALILEOS
E6 07 40
båÖäáëÜ
necessary.
Check whether selected language set is S. [ → 61]
already installed, perform software update
if necessary.
Please contact the Sirona Customer S. [ → 61]
Service Center (CSC) to find out whether a
bug fix by means of a software update is
possible and perform such an update if
necessary.
61 25 665 D3437
D3437.076.01.15.02 06.2012 97
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 10 02
E1 10 03
E1 10 04
E1 10 05: GALILEOS
E7 10 06 GALILEOS
61 25 665 D3437
98 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E1 10 07 GALILEOS
båÖäáëÜ
replace if necessary.
● Replace board DX89.
● Replace board DX1.
● Replace the X-ray detector.
● Replace board DX11.
61 25 665 D3437
D3437.076.01.15.02 06.2012 99
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E1 10 23/24/25/26 GALILEOS
61 25 665 D3437
100 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 11 02: GALILEOS
båÖäáëÜ
If the error occurs repeatedly ... S. [ → 350]
● Replace board DX11.
E6 11 03: GALILEOS
E7 11 04
E6 11 05: GALILEOS
E7 11 07
61 25 665 D3437
D3437.076.01.15.02 06.2012 101
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E5 11 09
E7 11 11
E7 11 12 GALILEOS
E7 11 14
E7 11 15 GALILEOS
61 25 665 D3437
102 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E1 11 19 GALILEOS
E1 11 20 GALILEOS
båÖäáëÜ
● This error is a sequential fault. watch for
additional causal error messages and take
the respective action.
E2 11 22 GALILEOS
E1 11 23 GALILEOS
E1 11 88
E6 12 02
61 25 665 D3437
D3437.076.01.15.02 06.2012 103
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 14 02
E5 14 04
E6 14 05
E6 14 06
61 25 665 D3437
104 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 14 10/12 GALILEOS
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 105
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 15 03
E6 15 04
*) OI = Operating instructions
E6 15 05
E6 15 10
61 25 665 D3437
106 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 41 02
båÖäáëÜ
● Replace board DX41.
E6 41 03
E6 41 04
E6 41 05
61 25 665 D3437
D3437.076.01.15.02 06.2012 107
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 41 06
E6 41 07
E6 41 08
E6 41 09
61 25 665 D3437
108 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E7 41 10
E7 41 12
båÖäáëÜ
● Check unit configuration (with or without
DX41) by running service routine S017.9,
correct the configuration if necessary.
If the error occurs repeatedly ...
● Repeat procedure and observe causal
error messages.
E3 41 20
E6 41 21
61 25 665 D3437
D3437.076.01.15.02 06.2012 109
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 41 23
E6 41 24
E6 41 25
61 25 665 D3437
110 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 42 02
båÖäáëÜ
a bug fix by means of a software update is
possible and perform such an update if
necessary.
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ... S. [ → 343]
● Replace board DX42.
E6 42 03
E6 42 04
61 25 665 D3437
D3437.076.01.15.02 06.2012 111
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 42 05
E6 42 06
E6 42 07
E6 42 08
61 25 665 D3437
112 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E7 42 10
E7 42 12 GALILEOS
båÖäáëÜ
booting.
4. Repeat procedure and check function.
● Check unit configuration (with or without
DX41) by running service routine S017.9,
correct the configuration if necessary.
If the error occurs repeatedly ... S. [ → 343]
● Repeat procedure and observe causal
error messages.
● Check the signal path from board DX1 to
board DX42, replace module if necessary
E6 42 20 GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 113
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E7 42 21 GALILEOS
E3 42 30
E3 42 31
61 25 665 D3437
114 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 71 02
båÖäáëÜ
necessary.
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ... S. [ → 343],
● Replace board DX71 or Multipad. S. [ → 307]
E6 71 03
E6 71 04
E6 71 05
61 25 665 D3437
D3437.076.01.15.02 06.2012 115
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 71 06
E6 71 07
E6 71 08
E7 71 10
61 25 665 D3437
116 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 71 12
E6 71 20
båÖäáëÜ
S. [ → 141],
● Check the CAN bus.
S. [ → 362]
● Check cable L9, replace if necessary.
E6 71 21
61 25 665 D3437
D3437.076.01.15.02 06.2012 117
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E3 71 30/33/34/35/36/37/38/39/40/41
61 25 665 D3437
118 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
båÖäáëÜ
E6 89 02
61 25 665 D3437
D3437.076.01.15.02 06.2012 119
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 89 04
E6 89 05
E6 89 06
E6 89 07
61 25 665 D3437
120 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E6 89 08
E7 89 10
båÖäáëÜ
LED states).
If the board remains in the bootloader stage ... S. [ → 61],
● Run software update. S. [ → 134],
● Check board DX89, replace if necessary. S. [ → 343]
E7 89 12
E5 89 13
E6 89 20
61 25 665 D3437
D3437.076.01.15.02 06.2012 121
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 89 21
E5 89 22
E5 89 23
E7 89 26
E7 89 27
61 25 665 D3437
122 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 5 Messages
Service Manual GALILEOS 5.5 List of error messages
E7 89 28
E7 89 29
båÖäáëÜ
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and check function.
If the error occurs repeatedly ... S. [ → 343]
● Check that the memory modules on board
DX89 are firmly fixed, replace board DX89
if necessary.
E7 89 30
E6 89 32
61 25 665 D3437
D3437.076.01.15.02 06.2012 123
5 Messages Sirona Dental Systems GmbH
5.5 List of error messages Service Manual GALILEOS
E6 89 33
E1 89 34
E2 89 35
E7 89 37
E2 89 38
E1 89 39
61 25 665 D3437
124 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.1 Error logging memory
6 Troubleshooting
Troubleshooting
DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and wait at least 1 minute before
removing a cover.
Switch OFF the X-ray unit before connecting a measuring instrument.
Perform continuity tests only on units which are switched off.
NOTICE
Risk of damage to unit
Select the correct current/voltage type and adjust the measuring range
to match the expected readings.
båÖäáëÜ
Keep to the prescribed cool-off periods if several exposures have to be
taken to check a measured value.
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
CAN bus cable: When unplugging CAN bus cables, it is essential to
unplug the power supply as well.
61 25 665 D3437
D3437.076.01.15.02 06.2012 125
6 Troubleshooting Sirona Dental Systems GmbH
6.1 Error logging memory Service Manual GALILEOS
61 25 665 D3437
126 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.1 Error logging memory
båÖäáëÜ
[PC Date / Time]
61 25 665 D3437
D3437.076.01.15.02 06.2012 127
6 Troubleshooting Sirona Dental Systems GmbH
6.2 Checking the CAN bus Service Manual GALILEOS
NOTICE
Risk of damage to unit
The power supply MUST be plugged in and switched on when cables
are attached and plugged in. For example, if no power cable is
connected to the DX71, the module has no ground connection to the unit
and there is no potential equalization. If the CAN cable is plugged in, the
CAN transceiver (IC on the DX71) can be destroyed by the voltage
difference. In other words, when the unit is switched on, CAN cables
may only be plugged in on modules that are connected to the power
source and ground.
For troubleshooting, you can disconnect the CAN bus cable and/or plug
it back in and observe the (unit's) behavior.
61 25 665 D3437
128 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.2 Checking the CAN bus
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 129
6 Troubleshooting Sirona Dental Systems GmbH
6.2 Checking the CAN bus Service Manual GALILEOS
61 25 665 D3437
130 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.2 Checking the CAN bus
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 131
6 Troubleshooting Sirona Dental Systems GmbH
6.2 Checking the CAN bus Service Manual GALILEOS
The jumpers are located on board DX1 at sockets X302, X303, X306,
X307, X309, X500, and X503 (see also wiring diagrams). If a cable is
connected to the socket, the corresponding jumpers must be set to the
outer position. If no cable is plugged in, the jumpers must be set to the
inner position. If a jumper is set to the inner position without a cable
plugged in, the CAN bus is interrupted at this location. Modules located
behind this location can no longer be connected to the CAN bus and,
therefore, do not function.
61 25 665 D3437
132 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.2 Checking the CAN bus
Jumper outside
If the jumpers are set to the outer position, the module is connected
(i.e. the connector is plugged in).
XXXX
båÖäáëÜ
Jumper inside
If the jumpers are set to the inner position, the module is not connected
(i.e. the connector is not plugged in).
XXXX
61 25 665 D3437
D3437.076.01.15.02 06.2012 133
6 Troubleshooting Sirona Dental Systems GmbH
6.3 Checking the boards Service Manual GALILEOS
61 25 665 D3437
134 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.3 Checking the boards
GALILEOS LEDs
båÖäáëÜ
V101 lit not lit
V102 lit not lit
DX71 V101 lit not lit
V103 lit not lit
V107 flashing at 1 Hz not lit flashing at 2 Hz
DX32 V132 lit not lit
V133 lit not lit
DX41* V103 flashing at 1 Hz not lit flashing at 2 Hz
V202 lit not lit
V204 lit not lit
DX42 V101 lit not lit
V103 lit not lit
DX89 V201 lit not lit
V202 lit not lit
V203 lit not lit
V204 flashing at 1 Hz not lit flashing at 2 Hz
V205 lit not lit
V207 lit not lit
*) Starting with unit serial number 3201, new units will be delivered
without board DX41.
61 25 665 D3437
D3437.076.01.15.02 06.2012 135
6 Troubleshooting Sirona Dental Systems GmbH
6.3 Checking the boards Service Manual GALILEOS
61 25 665 D3437
136 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.3 Checking the boards
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 137
6 Troubleshooting Sirona Dental Systems GmbH
6.4 Checking the motors Service Manual GALILEOS
61 25 665 D3437
138 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.5 Checking the light barriers
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 139
6 Troubleshooting Sirona Dental Systems GmbH
6.6 Device leakage current too high Service Manual GALILEOS
61 25 665 D3437
140 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.7 Checking the cables
NOTICE
You can use a standard Cat5 cable as a test cable for L8 (up to unit
serial number 3201), L10, L12, L40 and L37. This cable must not be
permanently installed.
IMPORTANT: Most cables have the same plug at both ends and are
connected 1:1.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 141
6 Troubleshooting Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path Service Manual GALILEOS
61 25 665 D3437
142 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.8 Error analysis of X-RAY control signal path
båÖäáëÜ
E3 41 20 occurs once after the unit is
switched on:
61 25 665 D3437
D3437.076.01.15.02 06.2012 143
6 Troubleshooting Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path Service Manual GALILEOS
61 25 665 D3437
144 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.8 Error analysis of X-RAY control signal path
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 145
6 Troubleshooting Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path Service Manual GALILEOS
61 25 665 D3437
146 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.8 Error analysis of X-RAY control signal path
båÖäáëÜ
E3 42 31 occurs once after the unit is
switched on:
61 25 665 D3437
D3437.076.01.15.02 06.2012 147
6 Troubleshooting Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path Service Manual GALILEOS
61 25 665 D3437
148 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.8 Error analysis of X-RAY control signal path
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 149
6 Troubleshooting Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path Service Manual GALILEOS
61 25 665 D3437
150 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.9 Fault diagnosis of the X-ray detector and on board DX89
NOTICE
Do not damage the image tube!
The image tube of the X-ray detector is sensitive to mechanical stress,
and therefore must be handled with extreme care. Avoid bumps and
jolts. Please consider this point during transport and installation.
båÖäáëÜ
4. CAUTION! Risk of injury! The cover plate has sharp edges.
Carefully pull the cover plate upwards to remove it from the X-ray
detector (see also Replace X-ray detector [ → 319]).
5. WARNING! Potentially lethal shock hazard! Do not touch any live
parts while observing board DX89.
Remove the cover plate of board DX89. Switch the unit on again and
observe board DX89.
The LEDs on the board can provide information about the possible
cause of the error (LEDs on board DX89 [ → 152]).
61 25 665 D3437
D3437.076.01.15.02 06.2012 151
6 Troubleshooting Sirona Dental Systems GmbH
6.9 Fault diagnosis of the X-ray detector and on board DX89 Service Manual GALILEOS
61 25 665 D3437
152 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 6 Troubleshooting
Service Manual GALILEOS 6.9 Fault diagnosis of the X-ray detector and on board DX89
båÖäáëÜ
If all LEDs light up after the power-up phase, this leads to conclusions
concerning a defect on board DX89. See the procedure outlined above
for troubleshooting.
PLL_CCD
NOTICE
The unit must be switched off before disconnecting any plugs or cables.
If all LEDs light up after the power-up phase, this indicates that there is a
defective FPGA on board DX89. For troubleshooting, see the action
under "LED OFF for PLL_FPGA".
Gettering
LED on: Gettering is o.k.
LED flashing (after a waiting Gettering is not o.k.
period of 12 minutes):
Action: ● Replace the X-ray detector [ → 319].
The free ions are pumped out of the vacuum of the X-ray detector by the
getters (hence the name "getter pump"). The getter current is measured
during operation. If this does not drop below a certain value within 12
minutes, the gettering is not o.k. In this case, the error is probably caused
by a defective x-ray detector.
61 25 665 D3437
D3437.076.01.15.02 06.2012 153
6 Troubleshooting Sirona Dental Systems GmbH
6.9 Fault diagnosis of the X-ray detector and on board DX89 Service Manual GALILEOS
NOTICE
The unit must be switched off before disconnecting any plugs or cables.
Operating voltages (28V, 24V, 5V, 3.3V) The four LEDs are powered directly by the four operating voltages and all
must light up after the system start.
If this is not the case, check connector X201 for firm seating. If the
connector is OK and the LEDs still do not light up, then replace the X-ray
detector [ → 319].
Operating voltages on DX89 and X-ray These two LEDs PLL_FPGA and PLL_CCD must light up following the
detector system start. If this is not the case, replace the x-ray detector [ → 319].
61 25 665 D3437
154 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS
DANGER
X-rays
When performing the following tests, be sure to observe the radiation
protection regulations applicable in your country (see Operating
Instructions).
DANGER
X-rays
"Radiation" is signaled by the message "X-RAY active!", a beep, and an
X-RAY LED.
GALILEOS service set
båÖäáëÜ
● Geometry phantom [ → 164]
● Distortion phantom [ → 163]
● Steel tape measure, 300mm
IMPORTANT
If you encounter problems with unit calibration, check whether the
required EMC conditions have been met. No heavy-duty electric
equipment (e.g. air conditioners, fan motors, etc.) should be present in
the vicinity of the unit.
Tip: Move the unit to a typical working height (bite block height (A) =
approx. 1,520 mm (60")) with the Up/Down keys on the control panel
before commencing calibration.
61 25 665 D3437
D3437.076.01.15.02 06.2012 155
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
Service Manual GALILEOS
As of unit software version V 03.04.00, you can even adjust the height of
the unit during the calibration procedure. If the unit is ready for an
exposure (after the "Image acquisition" button has been pressed in the
SIDEXIS XG service menu), the corresponding service routine (S002.6/
S010.10-14/S011.8, and S030.5) is displayed on the control panel. All of
these service routines allow the height adjustment menu to be opened by
pressing the Test key. The current unit height is displayed in selection
field 1 in this menu. You can then set the unit to the desired height using
the UP/DOWN keys on the control panel.
Press the Service key or the double arrow key (Easypad) or the arrow key
above selection field 3 (Multipad) to exit the height adjustment menu.
61 25 665 D3437
156 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.1 General information about unit adjustment and calibration
båÖäáëÜ
● Distortion calibration [ → 185]
● Geometry calibration [ → 187])
Tip: It may be helpful to use the SIDEXIS XG coloring function to evaluate
the image.
61 25 665 D3437
D3437.076.01.15.02 06.2012 157
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.1 General information about unit adjustment and calibration Service Manual GALILEOS
The most frequent help and status messages during calibration are listed
below.
Help messages H3 01: Move unit to starting position, press the R key.
H3 07: Remove the occlusal bite block.
H3 21: Close the door.
H323/3244 GALILEOS
Easypad Multipad
"Ready for exposure" no special display;
kV level and mAs are displayed
"Exposure not possible" S110
"Please wait" Progress bar
"Ready for exposure in XXs" XXs
"X RAY Active!" LED lights up on control panel
61 25 665 D3437
158 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.1 General information about unit adjustment and calibration
båÖäáëÜ
A Message window
B Tabs
C Preview image
D Status column
E Tools pictograph
Message window The message window displays text messages about the calibration
process.
Tabs / index cards (submenus) The "Service exposure " menu guides you through the calibration process
and contains the following tabs via which you can open and change back
and forth between eight index cards (submenus):
● "Diaphragm"
● "Radiation field"
● "Dose"
● "Sensor "
● "Iris"
● "Shading"
● "Distortion"
● "Geometry"
● "Service"
61 25 665 D3437
D3437.076.01.15.02 06.2012 159
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.1 General information about unit adjustment and calibration Service Manual GALILEOS
Preview image The "Diaphragm" and "Shading" submenus each contain a preview
image that symbolizes the exposure to be taken during the calibration
step. Due to the varying geometry of the "GALILEOS Comfort" and the
"GALILEOS Compact" diaphragms, the preview images displayed in
these submenus differ slightly.
Tools pictograph The tools pictograph shows which (if any) test phantom must be used for
this particular calibration step.
61 25 665 D3437
160 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.1 General information about unit adjustment and calibration
You can call the "Service functions" menu via SIDEXIS XG:
"Utilities"/"Constancy test..."/"3D"/"Select X-ray device"/"Service
exposure "/Password prompt/"Select X-ray component"/"Service
functions"
The "Select X-ray device" and "Select X-ray component" prompts are
only displayed if more than one unit has been set up in SIDEXIS XG.
Password protection
GALILEOS password
Service mode
GALILEOS service
When you open the "Service functions" menu, the unit switches from user
båÖäáëÜ
mode to the PC service mode logged by the PC. In PC service mode, the
control options that are available on the control panel are determined by
SIDEXIS XG and the service routine currently selected. General control
of the unit by means of the control panel (as in the user mode) is not
possible in this mode.
Easypad overview (GALILEOS)
Easypad
Service mode is displayed on the Easypad via the PC service image.
GALILEOS service mode
Multipad
"SERVICE" is displayed on the Multipad to indicate that the service mode
is active.
61 25 665 D3437
D3437.076.01.15.02 06.2012 161
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.1 General information about unit adjustment and calibration Service Manual GALILEOS
To take an exposure in SIDEXIS XG, the system must first be made ready
for exposure.
61 25 665 D3437
162 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.1 General information about unit adjustment and calibration
båÖäáëÜ
You must clip the distortion phantom onto the X-ray detector cover for the
radiation field check [ → 178] and for the distortion calibration [ → 185].
This phantom must be removed again for all other calibration steps.
61 25 665 D3437
D3437.076.01.15.02 06.2012 163
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.1 General information about unit adjustment and calibration Service Manual GALILEOS
61 25 665 D3437
164 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.1 General information about unit adjustment and calibration
båÖäáëÜ
IMPORTANT
Once it has been inserted into the bite block holder, the geometry
phantom must be aligned vertically and horizontally with the spirit level,
so that calibration can be performed correctly.
➢ Insert the geometry phantom (A) into the pan bite block holder (B) on
the unit and secure it with the screw (C).
61 25 665 D3437
D3437.076.01.15.02 06.2012 165
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.2 Checking the mechanical system adjustment Service Manual GALILEOS
The constancy test phantom is inserted in the bite block holder of the unit
for for the constancy and acceptance tests as well as for the check of the
mechanical unit adjustment [ → 166].
1. Insert the constancy test phantom in the bite block holder of the unit.
P1 P2 M
P3 M
B2 B2
M
B1
61 25 665 D3437
166 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
✔ The "Service functions" menu is selected.
1. Click on the "Image acquisition" button.
ª SIDEXIS XG makes the unit ready for exposure [ → 162].
ª Service routine S030.5 is displayed on the control panel.
2. Take an exposure (85 kV/21 mAs) [ → 162].
3. For "GALILEOS Comfort": Evaluate the image.
61 25 665 D3437
D3437.076.01.15.02 06.2012 167
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
A+B Adjustment OK
C Adjustment not OK
E Permissible tolerance: 30 pixels ± 5 pixels
61 25 665 D3437
168 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
A+B Adjustment OK
C Adjustment not OK, diaphragm not centered
D Adjustment not OK, diaphragm too small
E Permissible tolerance: 65 pixels ± 5 pixels
61 25 665 D3437
D3437.076.01.15.02 06.2012 169
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
170 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
Adjusting the "diaphragm open" ✔ The "Service functions" menu is called [ → 161].
diaphragm setting
✔ The "Diaphragm" tab is selected.
✔ The "Collimator open" is selectable in the "Image acquisition" menu
area.
1. Set the rotary knob on the tube assembly to the "open diaphragm"
position.
61 25 665 D3437
D3437.076.01.15.02 06.2012 171
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
A+B Adjustment OK
C Adjustment not OK
E Permissible tolerance: 30 pixels ± 5 pixels
61 25 665 D3437
172 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
A+B Adjustment OK
C Adjustment not OK, diaphragm not centered
D Adjustment not OK, diaphragm too small
E Permissible tolerance: 65 pixels ± 5 pixels
61 25 665 D3437
D3437.076.01.15.02 06.2012 173
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
174 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
A Adjustment OK
B Adjustment not OK
61 25 665 D3437
D3437.076.01.15.02 06.2012 175
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
Adjustment of the "Lower jaw" ✔ The "Service functions" menu is called [ → 161].
diaphragm setting
✔ The "Diaphragm" tab is selected.
✔ The "Lower jaw" is selectable in the "Image acquisition" menu area.
1. Set the rotary knob on the tube assembly to the "lower jaw" position.
61 25 665 D3437
176 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
A Adjustment OK
B Adjustment not OK
61 25 665 D3437
D3437.076.01.15.02 06.2012 177
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
6. To confirm that the lighting strips on the distortion phantom are not lit,
click the check box underneath the message window.
ª The box will appear checked.
ª The radiation field check is now complete.
7. Continue the calibration procedure with the dosimetry [ → 179].
Tip: Leave the distortion phantom on the unit for the next calibration
step.
61 25 665 D3437
178 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
7.3.3 Dosimetry
Dosimetry
båÖäáëÜ
✔ The "Service functions" menu is called [ → 161].
✔ The distortion phantom is clipped onto the cover of the X-ray detector
for protection against scratching.
1. Click the "Dose" tab.
ª The corresponding tab card is selected.
2. Attach the Mult-O-Meter sensor approximately in the middle of the
distortion phantom mounted on the X-ray detector.
61 25 665 D3437
D3437.076.01.15.02 06.2012 179
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
180 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
✔ The "Service functions" menu is called [ → 161].
1. Click the "Sensor " tab.
ª The corresponding tab card is selected.
61 25 665 D3437
D3437.076.01.15.02 06.2012 181
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
182 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
✔ The "Service exposure " menu is called [ → 161].
1. Click the "Shading" tab.
ª The corresponding tab card is selected.
61 25 665 D3437
D3437.076.01.15.02 06.2012 183
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
184 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
✔ The "Service functions" menu is called [ → 161].
1. Click the "Distortion" tab.
ª The corresponding tab card is selected.
2. Read off the serial number of the distortion phantom from the ID label
of the phantom and enter it in the "Seriennr. Kalbrierphantom" text
box of the "Distortion" tab card.
3. Clip the distortion phantom onto the X-ray detector cover [ → 163].
4. Click on the "Image acquisition" button.
ª SIDEXIS XG makes the unit ready for exposure [ → 162].
ª Service routine S010.12 is displayed on the control panel.
5. Press the R key to move the unit back to the starting position.
6. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the distortion calibration is
displayed in the message window.
If the information in the message window indicates that
calibration was not successful, check the calibration phantom to
make sure that it is not damaged.
If the phantom checks out OK (all balls are present and correctly
positioned), repeat the procedure starting with point d) as often
as required until the calibration is OK.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
61 25 665 D3437
D3437.076.01.15.02 06.2012 185
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
61 25 665 D3437
186 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
IMPORTANT
For an installed and configured Facescan:
The face scanner is automatically calibrated during the geometry
calibration of GALILEOS. In this case, make sure that the normal room
lighting is switched on during the calibration process. The room does not
have to be darkened during calibration.
båÖäáëÜ
✔ The "Service functions" menu is called [ → 161].
1. Click the "Geometry" tab.
ª The corresponding tab card is selected.
2. Read off the serial number of the geometric phantom from the ID
label of the phantom and enter it in the "Seriennr. Kalbrierphantom"
text box of the "Geometry" tab card.
3. Insert the geometric phantom in the bite block holder of the
unit [ → 164].
4. For Facescan: Check to make sure that the room lighting is switched
on. Switch it on if necessary.
5. Click on the "Image acquisition" button.
ª SIDEXIS XG makes the unit ready for exposure [ → 162].
ª Service routine S010.13 is displayed on the control panel.
6. Press the R key to move the unit back to the starting position.
61 25 665 D3437
D3437.076.01.15.02 06.2012 187
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu Service Manual GALILEOS
7. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the calibration is displayed
in the message window.
If the information in the message window indicates that
calibration was not successful, check the calibration phantom to
make sure that it is not damaged.
If you have repeated the procedure three times and still have not
attained a positive result, check the mechanical geometry of the
unit [ → 166]. Adjust the unit if necessary and then repeat the
calibration.
If this still does not lead to a positive result, please contact the
SIRONA Customer Service Center (CSC).
8. If the calibration is OK or possible, click the "Save values" button.
ª The calibration is saved.
ª Calibration of the geometry is now complete.
9. Remove the geometric phantom from the bite block holder of the unit.
Calibration of the unit is now complete. IA test image can be
generated [ → 189] via the "Service" tab card if necessary.
61 25 665 D3437
188 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.3 Adjustment and calibration via the "Service Functions" menu
7.3.9 Service
Service
You can create a test image with the "Service" tab card.
It is not necessary to execute this menu for the calibration of the unit!
båÖäáëÜ
✔ The "Service functions" menu is called [ → 161].
1. Click the "Service" tab.
ª The corresponding tab card is selected.
2. Click on the "Image acquisition" button.
ª SIDEXIS XG makes the unit ready for exposure [ → 162].
ª Service routine S032.41 is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
ª The CCD test image is displayed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 189
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.4 Performing a white balance for Facescan Service Manual GALILEOS
IMPORTANT
Constant lighting conditions
If the lighting conditions during the white balance process are different
from those during the subsequent Facescan process, this will lead to
inaccurate colors.
➢ Ensure that the lighting conditions during white balancing
correspond with those of the later Facescan operation.
61 25 665 D3437
190 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.5 Checking and adjusting the touchscreen
ª Once the unit software has started up, the start screen is
displayed on the touchscreen.
2. As soon as the start screen appears, press the 3 keys light localizer,
T, and R simultaneously (see (A)).
IMPORTANT: These keys must be pressed while the unit is booting
and the start screen is still displayed.
ª The first adjustment screen appears on the display.
A ª You can abort the adjustment procedure at any time by pressing
the "Abort" button.
3. Touch the top left corner of the screen.
ª The second adjustment screen appears on the display.
båÖäáëÜ
4. Touch the bottom right corner of the screen.
ª The third adjustment screen appears on the display.
5. Touch the center of the green square in the top right corner of the
screen.
IMPORTANT: Touch the green squares as close to the center as
possible so that the black dots that then appear will be positioned
exactly in the center of the squares.
ª The fourth adjustment screen appears on the display.
61 25 665 D3437
D3437.076.01.15.02 06.2012 191
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.5 Checking and adjusting the touchscreen Service Manual GALILEOS
6. Touch the center of the green square in the bottom left corner of the
screen.
ª The fifth adjustment screen appears on the display.
61 25 665 D3437
192 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.6 Mechanical adjustments
NOTICE
Perform this adjustment only if the measured values are out of
tolerance [ → 166].
båÖäáëÜ
A A
61 25 665 D3437
D3437.076.01.15.02 06.2012 193
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.6 Mechanical adjustments Service Manual GALILEOS
C C
61 25 665 D3437
194 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.6 Mechanical adjustments
NOTICE
Perform this adjustment only if the measured values are out of
tolerance [ → 166].
båÖäáëÜ
2. Loosen screw (A) slightly.
3. NOTICE! Do not forget to tighten screw (A) again. Otherwise, the
B clearance and play of the swivel arm is not ensured!
Adjust the swivel arm with the eccentric screw (B). Hold the eccentric
screw securely in place and tighten screw (A) again.
IMPORTANT: The swivel arm is shown here without the control panel
for purposes of clarity (*).
61 25 665 D3437
D3437.076.01.15.02 06.2012 195
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.6 Mechanical adjustments Service Manual GALILEOS
61 25 665 D3437
196 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.6 Mechanical adjustments
Preparations 1. With the "Type 3" diaphragm:Pull off the adjusting knob with the
silicone ring.
2. Remove the "Front tube assembly" and "Rear tube assembly" covers.
7.6.3.1 Adjusting the "type 1" diaphragm (up to serial no. 2200)
Adjusting the "type 1" diaphragm (up to serial no. 2200)
C
D
båÖäáëÜ
E B
A A
61 25 665 D3437
D3437.076.01.15.02 06.2012 197
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.6 Mechanical adjustments Service Manual GALILEOS
7.6.3.2 Adjusting the "Type 2" diaphragm (serial no. 2201 and higher)
Adjusting the "Type 2" diaphragm (serial no. 2201 and higher)
61 25 665 D3437
198 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.6 Mechanical adjustments
Adjusting the diaphragm size In order to adjust the size of the diaphragm opening, adjust the diaphragm
distance.
båÖäáëÜ
1. Loosen screw (D) slightly (approx. 1 turn).
2. Adjust the diaphragm distance with knurled nut (E).
Turn the knurled nut toward the rear:
The image on the X-ray detector becomes larger
Turn the knurled nut toward the front
The image on the X-ray detector becomes smaller
3. Retighten screw (D) firmly.
4. Repeat the diaphragm exposure [ → 167].
61 25 665 D3437
D3437.076.01.15.02 06.2012 199
7 Adjusting/calibrating the unit Sirona Dental Systems GmbH
7.6 Mechanical adjustments Service Manual GALILEOS
Adjusting the complete diaphragm unit If the exposure taken in the "diaphragm open" diaphragm setting is not
OK, the complete diaphragm unit must be adjusted.
1. Set the rotary knob on the tube assembly to the "open diaphragm"
position.
A
A
61 25 665 D3437
200 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 7 Adjusting/calibrating the unit
Service Manual GALILEOS 7.6 Mechanical adjustments
Adjusting the lower lead diaphragm If the exposure taken in the "upper jaw" diaphragm setting is not OK, the
lower lead diaphragm must be adjusted.
1. Set the rotary knob on the tube assembly to the "maxillary exposure"
position.
båÖäáëÜ
B
61 25 665 D3437
D3437.076.01.15.02 06.2012 201
8 Service Sirona Dental Systems GmbH
8.1 Overview of service routines Service Manual GALILEOS
8 Service
Service
61 25 665 D3437
202 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.1 Overview of service routines
båÖäáëÜ
(fixed address)
S037.4 Manual input of static network settings (IP address, default S. [ → 283]
gateway address, and subnet mask)
61 25 665 D3437
D3437.076.01.15.02 06.2012 203
8 Service Sirona Dental Systems GmbH
8.1 Overview of service routines Service Manual GALILEOS
61 25 665 D3437
204 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.1 Overview of service routines
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 205
8 Service Sirona Dental Systems GmbH
8.2 Service menu and service routines Service Manual GALILEOS
You can use the service routines to check the function of certain unit
components and modules, as well as to set important unit parameters.
Description of Service Routines: Service Manual
This chapter describes all service routines that can be selected and
started via the service menu on the control panel. Service routines S010
and S030 cannot be selected manually and, therefore, are not described
here. They are used solely for unit adjustment and calibration.
8.2.1.1 Easypad
Easypad
There are many different control symbols and display fields on the
touchscreen; these are activated on a context-sensitive basis depending
on the procedure step.
61 25 665 D3437
206 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.2 Service menu and service routines
båÖäáëÜ
service routine. Most, however,
confirm a selection or the activation of
the next test step.
K T(est rotation) key Start a test.
L R(eturn) key Move the unit to the starting position
or confirm a save operation.
M Double arrow key Return to the main menu.
N Wrench symbol Displayed if level 4 (service menu) is
activated.
61 25 665 D3437
D3437.076.01.15.02 06.2012 207
8 Service Sirona Dental Systems GmbH
8.2 Service menu and service routines Service Manual GALILEOS
8.2.1.2 Multipad
Multipad
K L H J I
B C D
61 25 665 D3437
208 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.2 Service menu and service routines
båÖäáëÜ
confirm a selection or the activation of
the next test step.
K T(est rotation) key Start a test.
L R(eturn) key Move the unit to the starting position
or confirm a save operation.
61 25 665 D3437
D3437.076.01.15.02 06.2012 209
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
8.3.1.1 Easypad
Easypad
61 25 665 D3437
210 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.3 Basic operating procedures in the service menu
båÖäáëÜ
ª After you have entered the key combination correctly, the service
menu is displayed. You can return to the next higher level with the
double arrow key (I) at any time.
61 25 665 D3437
D3437.076.01.15.02 06.2012 211
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
8.3.1.2 Multipad
1. Press and hold down Service key (D) until the patient symbol keys
Multipad
E F G H
ª After you have entered the key combination correctly, the service
menu is displayed. You can return to the next higher level with the
arrow key above selection field 3 (I) at any time.
61 25 665 D3437
212 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.3 Basic operating procedures in the service menu
- 1 +
båÖäáëÜ
B
S005_______1____
61 25 665 D3437
D3437.076.01.15.02 06.2012 213
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
61 25 665 D3437
214 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.3 Basic operating procedures in the service menu
båÖäáëÜ
follows:
1. Select the service routine or the test step, and confirm your selection
with the Service key [ → 213].
ª After you have confirmed your selection, a "0" appears in
selection field 2.
61 25 665 D3437
D3437.076.01.15.02 06.2012 215
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
61 25 665 D3437
216 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.3 Basic operating procedures in the service menu
båÖäáëÜ
8.3.3 Select parameters
Select parameters
Easypad touchscreen
If arrow keys are displayed in the selection fields once the required
service routine has been selected, you can use these arrow keys to
choose between different parameters.
Example You want to run service routine S017.6 to activate the remote control.
✔ Once you have selected service routine S017.6, the code "00" is pre-
selected for the "Remote control disabled" option.
61 25 665 D3437
D3437.076.01.15.02 06.2012 217
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
➢ Touch the + or - arrow keys (A) to select the code 01 (B) for the
"Remote control enabled" option.
ª Once the selected parameter has been changed (in this case the
code for the activation of the remote control), the Memory key (C)
lights up.
Selecting parameters in GALILEOS
Multipad
Example You want to run service routine S017.6 to activate the remote control.
✔ Once you have selected service routine S017.6, the code "00" is pre-
selected for the "Remote control disabled" option.
➢ Press the UP or DOWN arrow key (A) to select the code 01 (B) for the
"Remote control enabled" option.
ª Once the selected parameter has been changed (in this case the
code for the activation of the remote control), the LED above
Memory key (C) lights up.
61 25 665 D3437
218 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.3 Basic operating procedures in the service menu
Easypad touchscreen
Example You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔ The Memory key (C) lights up.
1. Touch the Memory key (C).
ª The R key (D) lights up.
2. Touch the R key (D).
ª The selected setting is saved to non-volatile memory.
Saving parameters in GALILEOS
båÖäáëÜ
Multipad
Example You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔ The LED above Memory key (C) lights up.
1. Press Memory key (C).
ª The LED above the R key lights up.
2. Press R key (D).
ª The selected setting is saved to non-volatile memory.
61 25 665 D3437
D3437.076.01.15.02 06.2012 219
8 Service Sirona Dental Systems GmbH
8.3 Basic operating procedures in the service menu Service Manual GALILEOS
Easypad
Touch the Service key (A) or the double arrow key (B) to go back to the
menu for selecting service routines.
Touch the double arrow key (B) in the service menu to go back to the
main menu.
Multipad
Press the Service key (A) or the up arrow key above selection field 3 (B)
to go back to the menu for selecting service routines.
In the service menu, touch the up arrow key above selection field 3 (B) to
go back to the main menu.
61 25 665 D3437
220 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.4 S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
WARNING
båÖäáëÜ
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
* Factory setting
61 25 665 D3437
D3437.076.01.15.02 06.2012 221
8 Service Sirona Dental Systems GmbH
8.4 S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time Service Manual GALILEOS
61 25 665 D3437
222 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.5 S005: General X-ray tube assembly service
båÖäáëÜ
S005: Test step 1
IMPORTANT
Only CB tube D151 R (indicator number 03) is permissible for use with
the GALILEOS volume tomography unit.
IMPORTANT
The X-ray tube assembly automatically queries the tube type
information. If no defined value is saved, the security access will be
shown instead of the tube type.
S005.1 GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 223
8 Service Sirona Dental Systems GmbH
8.5 S005: General X-ray tube assembly service Service Manual GALILEOS
If the tube type is invalid ... 1. Confirm the security access [ → 215].
ª Selection field 1 shows the indicator number of the (invalid) tube
type detected by the X-ray tube assembly.
2. Use the arrow keys in selection field 1 to select the code for the tube
type (see table).
ª The Memory key (B) (Easypad) or the LED above the Memory
key (B) (Multipad) lights up.
3. Save the selected parameter [ → 219].
4. Exit the service routine [ → 220].
After calling the service routine S005.1 again, the security access is no
longer requested and the stored tube type is displayed.
61 25 665 D3437
224 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.5 S005: General X-ray tube assembly service
Fan test
* Factory setting
1. Call service routine S005.4 [ → 213].
2. Use the arrow keys (A) in selection field 1 to select the code "01" (see
table).
3. Confirm your selection by pressing the R key (B).
ª The fan starts up.
4. Check the fan for running noise.
båÖäáëÜ
5. Exit the service routine [ → 220].
Upon exiting the service routine, the fan is automatically switched off
again.
61 25 665 D3437
D3437.076.01.15.02 06.2012 225
8 Service Sirona Dental Systems GmbH
8.5 S005: General X-ray tube assembly service Service Manual GALILEOS
61 25 665 D3437
226 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.5 S005: General X-ray tube assembly service
båÖäáëÜ
2. Start the automatic adjustment by pressing and holding the release
button.
IMPORTANT: Keep pressing the release button until adjustment is
completed and the new offset value for preheating is displayed. If you
interrupt the adjustment procedure prematurely by letting go of the
release button, the message "EEEE" appears in selection field 2. This
message must be acknowledged by pressing the R key.
ª When pressing the release button, radiation is released for 2 s to
warm up the tube assembly to operating temperature. This is
followed by the automatic tuning routine.
61 25 665 D3437
D3437.076.01.15.02 06.2012 227
8 Service Sirona Dental Systems GmbH
8.5 S005: General X-ray tube assembly service Service Manual GALILEOS
61 25 665 D3437
228 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.6 S007: Error logging memory
båÖäáëÜ
Title only for GALILEOS
Easypad touchscreen
* Factory setting
1. Call service routine S007.1 [ → 213].
2. Use the patient symbol keys (B, C, D) to select the step width for
scrolling between the error events (see table).
ª The selected patient symbol key lights up.
61 25 665 D3437
D3437.076.01.15.02 06.2012 229
8 Service Sirona Dental Systems GmbH
8.6 S007: Error logging memory Service Manual GALILEOS
3. Use the arrow keys (A) in selection field 1 to select the required error
event (66 in the example).
ª The corresponding error message is displayed in selection field
2 [ → 82].
ª Selection field 3 displays the date and time of the error event.
4. Exit the service routine [ → 220].
S007_1 GALILEOS
Multipad
1. Call service routine S007.1 [ → 213].
2. Use the arrow keys (A) in selection field 1 to select the required error
event (66 in the example).
Use the patient symbol keys to set the increment for scrolling
between the error numbers:
Patient symbol key 1 (left) = increment 1 (factory setting)
Patient symbol key 2 = increment 10
Patient symbol key 3 = increment 100
The LED above the selected patient symbol key is lit up.
61 25 665 D3437
230 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.6 S007: Error logging memory
3. Use the arrow keys (B) to scroll and display the corresponding error
code [ → 82], the time, and the date of the error event in selection field
2.
4. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 231
8 Service Sirona Dental Systems GmbH
8.6 S007: Error logging memory Service Manual GALILEOS
61 25 665 D3437
232 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.6 S007: Error logging memory
2. To clear the memory, press the Memory key (A) (R key (B) (Easypad)
or LED above the R key (B) (Multipad) lights up) followed by the R key
(B).
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 233
8 Service Sirona Dental Systems GmbH
8.6 S007: Error logging memory Service Manual GALILEOS
61 25 665 D3437
234 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.6 S007: Error logging memory
NOTICE
This service routine may only be called up subject to the approval of and
with the support of the Sirona Customer Service Center (CSC).
* Factory setting
1. Call service routine S007.5 [ → 213].
S007.5 GALILEOS
båÖäáëÜ
2. Use the arrow keys (A) to select the required setting (see table).
ª Once the required setting has been selected, the T key (B)
(Easypad) or the LED above the T key (B) (Multipad) lights up.
3. Touch the T key (B) to enable the selected setting.
ª All CAN bus events occurring from now on during operation of the
unit will be logged and can be displayed with a web browser (e.g.
Internet Explorer). This log will help you when consulting the
Sirona Customer Service Center (CSC) for error diagnosis.
4. Exit the service routine [ → 220].
61 25 665 D3437
D3437.076.01.15.02 06.2012 235
8 Service Sirona Dental Systems GmbH
8.6 S007: Error logging memory Service Manual GALILEOS
61 25 665 D3437
236 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.6 S007: Error logging memory
båÖäáëÜ
ª The CAN bus protocol is displayed in the browser and can be
saved as an HTML page, printed out, or sent to the Sirona
Customer Service Center (CSC).
61 25 665 D3437
D3437.076.01.15.02 06.2012 237
8 Service Sirona Dental Systems GmbH
8.7 S008: Update service Service Manual GALILEOS
Easypad touchscreen
1. Call service routine S008.2 [ → 213].
ª The software versions currently installed on the modules are
displayed on an info screen on the touchscreen display.
2. Exit the service routine [ → 220].
S008_2 GALILEOS
61 25 665 D3437
238 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.7 S008: Update service
Multipad
1. Call service routine S008.2 [ → 213].
ª "SYSTEMSOFTWARE" is displayed on the display line of the
Multipad.
2. Select the required module in selection field 1 with the arrow keys (A)
and confirm your selection by pressing the Memory key (B).
båÖäáëÜ
ª The software version of the selected module is displayed in
selection field 1.
3. Exit the service routine [ → 220].
61 25 665 D3437
D3437.076.01.15.02 06.2012 239
8 Service Sirona Dental Systems GmbH
8.7 S008: Update service Service Manual GALILEOS
IMPORTANT
If the backup copy of the old unit serial number does not match the new
one after replacing a module, the entry of the serial number is activated.
If an incorrect serial number is entered, the message "FFFF" appears
on the display. In this case, the service routine can be run again.
61 25 665 D3437
240 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.8 S009: Flash file system
IMPORTANT
The unit has to be completely recalibrated after formatting the flash file
system [ → 155]. When the flash file system is formatted, the content of
the error logging memory is lost.
S009 GALILEOS
båÖäáëÜ
S009: Test step 4
2. To initialize the flash file system, press the Memory key (A) (R key
(Easypad) or LED above R key (Multipad) lights up) followed by the
R key (B).
61 25 665 D3437
D3437.076.01.15.02 06.2012 241
8 Service Sirona Dental Systems GmbH
8.8 S009: Flash file system Service Manual GALILEOS
61 25 665 D3437
242 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.8 S009: Flash file system
2. To test the flash file system, press the Memory key (A) (R key
(Easypad) or LED above R key (Multipad) lights up) followed by the
R key (B).
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 243
8 Service Sirona Dental Systems GmbH
8.8 S009: Flash file system Service Manual GALILEOS
ª Once the system has passed the test without errors, "OK"
appears in selection field 1.
ª If the test fails, "ERROR" is displayed. In this case, the flash file
system must be formatted with service routine S009.4 [ → 241].
3. Exit the service routine [ → 220].
61 25 665 D3437
244 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.8 S009: Flash file system
båÖäáëÜ
Only one practical direction of data transfer is offered at any one
time. If both locations contain valid data,
"---" is displayed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 245
8 Service Sirona Dental Systems GmbH
8.8 S009: Flash file system Service Manual GALILEOS
2. To trigger the memory process, press the Memory key (A) (R key (B)
(Easypad) or LED above the R key (B) (Multipad) lights up) followed
by the R key (B).
61 25 665 D3437
246 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.9 S011: Dosimetry (without ring movement)
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
S011: GALILEOS
båÖäáëÜ
* SR=service routine, ** SHZ=security access
61 25 665 D3437
D3437.076.01.15.02 06.2012 247
8 Service Sirona Dental Systems GmbH
8.9 S011: Dosimetry (without ring movement) Service Manual GALILEOS
61 25 665 D3437
248 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.10 S012: CAN bus service
IMPORTANT
The CAN bus service is not yet implemented for the module DX11!
båÖäáëÜ
values
1 Subassembly DX1 - DX88
2 ● Counter value of
CAN bus events
● Presence code behind
the counter value:
P = module present
L = module lost
61 25 665 D3437
D3437.076.01.15.02 06.2012 249
8 Service Sirona Dental Systems GmbH
8.10 S012: CAN bus service Service Manual GALILEOS
S012.1: GALILEOS
61 25 665 D3437
250 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.10 S012: CAN bus service
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 251
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
* Factory setting
61 25 665 D3437
252 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
båÖäáëÜ
2. Use the arrow keys (A) to select the code for the required hardware
version in selection field 1 (see table).
ª Once the hardware version has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
D3437.076.01.15.02 06.2012 253
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
* Factory setting
1. Call service routine S017.3 [ → 213].
S017_3 GALILEOS
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
61 25 665 D3437
254 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
2. Use arrow keys (A) to select the required country group code in
selection field 1 (see table).
ª Once the country group code has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 255
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
Select a language
61 25 665 D3437
256 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
2. Use the arrow keys (A) to select the code for the required language
in selection field 1 (see table).
ª Once the language has been selected, the Memory key (B) lights
up.
3. Save the setting [ → 219].
IMPORTANT: If the selected language is not in the installed language
set (S017: Test step 5 [ → 258]), "English" is set by default.
4. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 257
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
IMPORTANT
A software update must be performed [ → 61] every time the language
set changes, in order to install the corresponding languages in the
system.
2. Use the arrow keys (A) to select the code for the required language
set in selection field 1 (see table).
ª Once the language set has been selected, the Memory key (B)
lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
258 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
* Factory setting
1. Call service routine S017.6 [ → 213].
S017_6 GALILEOS
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 259
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
2. Use arrow keys (A) to select the code for the required setting in
selection field 1 (see table).
ª Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
260 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
* Factory setting
1. Call service routine S017.7 [ → 213].
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 261
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
2. Use arrow keys (A) to select the required country group code in
selection field 1 (see table).
ª Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
262 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
* Factory setting
1. Call service routine S017.9 [ → 213].
ª Once the service routine has been selected, the code for the
båÖäáëÜ
current setting is displayed in selection field 1.
61 25 665 D3437
D3437.076.01.15.02 06.2012 263
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
2. Use arrow keys (A) to select the required code in selection field 1 (see
table).
ª Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
264 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
* Factory setting
1. Call service routine S017.13 [ → 213].
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
båÖäáëÜ
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
ª Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
D3437.076.01.15.02 06.2012 265
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
* Factory setting
1. Call service routine S017.14 [ → 213].
ª Once the service routine has been selected, the code for the line
currently selected is displayed in selection field 1.
2. Use the arrow keys (A) to select the required line in selection field 1
(see table).
ª The activation status code is displayed in selection field 2.
61 25 665 D3437
266 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
3. Use the arrow keys (B) to select the code for the required state of the
line selected in selection field 1 in selection field 2 (see table).
ª Once the required setting has been selected, the Memory key (C)
lights up.
4. Save the setting [ → 219].
5. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 267
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
* Factory setting
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
61 25 665 D3437
268 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
ª Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 269
8 Service Sirona Dental Systems GmbH
8.11 S017: Configuration service Service Manual GALILEOS
* Factory setting
1. Call service routine S017.25. [ → 213]
ª Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
61 25 665 D3437
270 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.11 S017: Configuration service
2. Use arrow keys (A) to select the required code in selection field 1 (see
table).
ª Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 271
8 Service Sirona Dental Systems GmbH
8.12 S018: Service for height adjustment Service Manual GALILEOS
61 25 665 D3437
272 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.12 S018: Service for height adjustment
3. To save the maximum travel height, press Memory key (B) (R key (C)
lights up) followed by R key (C).
båÖäáëÜ
4. Set the mechanical limit stop at the unit:
Loosen nut (D) and move mechanical limit stop (E) for the limit switch
until it engages. Tighten nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the
limit switch.
5. Exit the service routine [ → 220].
61 25 665 D3437
D3437.076.01.15.02 06.2012 273
8 Service Sirona Dental Systems GmbH
8.12 S018: Service for height adjustment Service Manual GALILEOS
ª Once the service routine has been selected, the current height
position is displayed in selection field 1.
ª Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
2. To undo the maximum travel height setting, press Memory key (B) (R
key (C (Easypad) or LED above R key (C) lights up) followed by R key
(C).
3. Exit the service routine [ → 220].
61 25 665 D3437
274 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.12 S018: Service for height adjustment
båÖäáëÜ
Patient symbol key 3 lit Upper limit switch activated
not lit Upper limit switch not
activated
ª Once the service routine has been selected, the current height
position is displayed in selection field 1.
ª Patient symbol keys 1 to 3 (F) show the switching state of the limit
switches (see table).
If the patient symbol key (Easypad) or the LED above the patient
symbol key (Multipad) is lit, the corresponding switch is activated,
i.e. the unit is at a position value greater than 1500.
61 25 665 D3437
D3437.076.01.15.02 06.2012 275
8 Service Sirona Dental Systems GmbH
8.12 S018: Service for height adjustment Service Manual GALILEOS
2. Use UP/DOWN keys (A) on the control panel to move the unit up and
down and use patient symbol keys (F) to check the switching states.
3. Exit the service routine [ → 220].
61 25 665 D3437
276 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
båÖäáëÜ
S037_1 GALILEOS
If all network data is set to default, the system is in UDP boot mode.
* Factory setting
* Factory setting
61 25 665 D3437
D3437.076.01.15.02 06.2012 277
8 Service Sirona Dental Systems GmbH
8.13 S037: Network service Service Manual GALILEOS
61 25 665 D3437
278 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
IMPORTANT
The network address can only be restored to the factory setting (default
value) in fixed address boot mode (STATIC or no DHCP).
båÖäáëÜ
selection field 1
61 25 665 D3437
D3437.076.01.15.02 06.2012 279
8 Service Sirona Dental Systems GmbH
8.13 S037: Network service Service Manual GALILEOS
2. Before restoring the factory settings, check the network data that is
still in the system:
A = Show IP address
B = Show default gateway
C = Show subnet mask
ª The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
3. To reset the network data, press Memory key (E) (R key (Easypad)
or LED above R key (Multipad) lights up) followed by R key (F).
ª The default network data (factory default setting) is displayed. To
switch between the display of the different network data, proceed
as in test step 1.
4. Restart the unit.
5. Exit the service routine [ → 220].
61 25 665 D3437
280 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
* Factory setting
1. Call service routine S037.3 [ → 213].
ª Once the service routine has been selected, the current boot
mode of the unit is displayed in selection field 1.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 281
8 Service Sirona Dental Systems GmbH
8.13 S037: Network service Service Manual GALILEOS
2. Use arrow keys (A) to select the required boot mode "automatic
address assignment" (DYNAMIC) or "fixed address" (STATIC) in
selection field 1 (see table).
ª Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
3. Save the setting [ → 219].
4. Exit the service routine [ → 220].
61 25 665 D3437
282 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
båÖäáëÜ
number pad B2 is selected
Patient symbol key 3 (C) lit The subnet mask is displayed in
selection field 1
or - after pressing the T key -
number pad B3 is selected
Patient symbol key 4 (D) or - after pressing the T key -
number pad B4 is selected
* Factory setting
* Factory setting
61 25 665 D3437
D3437.076.01.15.02 06.2012 283
8 Service Sirona Dental Systems GmbH
8.13 S037: Network service Service Manual GALILEOS
61 25 665 D3437
284 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
4. Now use the patient symbol keys to select the required number pad
1-4 (A-D) (see also table):
A = Number pad B1
B = Number pad B2
C = Number pad B3
D = Number pad B4
ª The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 285
8 Service Sirona Dental Systems GmbH
8.13 S037: Network service Service Manual GALILEOS
5. Use arrow keys (E) to select the digit to be changed in selection field
1 ("Digit No. 12" in the example).
ª The corresponding patient symbol key (D) or the LED above the
corresponding patient symbol key (D) lights up.
ª Selection field 2 displays the value of the currently selected digit.
61 25 665 D3437
286 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 8 Service
Service Manual GALILEOS 8.13 S037: Network service
6. To change the value for the digit, use arrow keys (F) in selection field
2.
ª Memory key (G) (Easypad) or the LED above Memory key (G)
(Multipad) lights up.
7. Save the setting [ → 219].
8. Exit the service routine [ → 220].
9. Restart the unit.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 287
9 Repair Sirona Dental Systems GmbH
Service Manual GALILEOS
9 Repair
Repair
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least 1 minute, or 4
minutes if disconnecting the tube assembly (cable L3), before starting
the repair or taking off a cover panel!
When replacing parts in the vicinity of the power connection, power
switch, board DX32 or X-ray tube assembly, the unit must be
disconnected from the junction box of the main building!
CAUTION
Make sure to reattach all ground cables to ensure correct grounding of
all modules.
CAUTION
Product safety
Modifications to this unit which might affect the safety of the system
owner, patients or other persons are prohibited by law! For reasons of
product safety, this product may be operated only with original Sirona
accessories or third-party accessories expressly approved by Sirona.
The user is responsible for any damage resulting from the use of non-
approved accessories.
NOTICE
Do not damage the cables
Be careful not to kink the cables when removing or installing them. Take
particular care with fiber-optic cables L5, L6, L7, and L15. Tighten cable
ties only as far as the contact and do not apply force.
NOTICE
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
IMPORTANT
After replacing boards or modules containing boards, check to make
sure that the software version of the module corresponds to the current
software status of the system. The software versions of the modules can
be queried by running service routine S008.2 or using the extended
detail query in SiXABCon. You can also check the info screen in
advance to determine whether the current software constellation is
permissible. If this is not the case, the version number of the entire
software is labeled with an asterisk (e.g. V03.03.01*).
61 25 665 D3437
288 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.1 Safety checks
IMPORTANT
When replacing modules, be sure to note which ones contain boards
and follow the instructions in the section titled Measures following
replacement of boards. Also check whether the current software CD or
the SIRONA dealer page contains any additional more up-to-date
information about module replacement.
Be sure to follow the instructions about how to proceed following module
replacement. You will find this information at the end of each set of
repair instructions.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 289
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
CAUTION
Risk of injury due to uncontrolled movement of the slide
If the slide can no longer be moved electrically, it must be moved
mechanically.
It must be ensured that no one is located underneath the ring arm during
the repair.
61 25 665 D3437
290 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.2 Height adjustment motor (M1_4)/spindle
9.2.1.1.1 Moving the slide with the "height adjustment" service kit, REF. 62
57 518
Moving the slide with the "height adjustment" service kit, REF. 62 57 518
2. Loosen the two screws (C) and remove the cover (D).
båÖäáëÜ
C
C
61 25 665 D3437
D3437.076.01.15.02 06.2012 291
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
Locking the slide ➢ Now move the clamp (A) directly underneath the slide.
ª The slide is now locked in this position for further repair work.
9.2.1.1.2 Moving the slide without the "height adjustment" service kit, REF.
62 57 518
Moving the slide without the "height adjustment" service kit, REF. 62 57 518
2. Loosen the 1st of the two screws (E) on the spindle holder (F).
F
E
3. Attach the socket wrench (SW19) to the spindle. Hold it firmly in place
while you unscrew the 2nd of the two screws (E).
CAUTION! If the socket wrench has to be reset, secure the spindle
holder against turning, e.g. by using a screw.
4. Rotate the spindle holder using a socket wrench (SW19) to move the
slide to the required height.
Clockwise rotation of spindle = slide moves up
Counterclockwise rotation of spindle = slide moves down
5. After reaching the desired target position, secure the position again
using the two screws (E).
CAUTION! Before replacing the height adjustment motor, the slide
must be secured in this position.
Locking the slide: GALILEOS
61 25 665 D3437
292 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.2 Height adjustment motor (M1_4)/spindle
Locking the slide 1. Make a mark at the position of the upper limit stop.
2. Loosen nut (G) on the upper profile clamp (H) and remove the upper
limit stop (J) from the stand.
3. Install limit stop (J) above the lower limit stop so that there is a
distance of 31 cm between the upper edge of the upper screw on
board DX41 and the lower edge of the limit stop.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 293
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
5. Loosen all the left-hand screws (B) of the protective plates (E).
61 25 665 D3437
294 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.2 Height adjustment motor (M1_4)/spindle
L3
F
G
H
DX32
L3
I
X1
E X2
båÖäáëÜ
L2
X100
61 25 665 D3437
D3437.076.01.15.02 06.2012 295
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
Removing the spindle 1. Loosen the two screws (E) on the spindle holder (F) (if you have not
already done so) [ → 290].
2. Turn spindle holder (D) (with an 18 mm A/F socket wrench)
counterclockwise until the motor comes to rest on the limit stop and
spindle (L) has been turned all the way out of the motor.
3. Remove the straight pin (K).
4. Remove the spindle (L).
Tip: First, pull spindle (L) downward along the motor, and then
diagonally upward and out of the unit.
Removing the defective motor 1. Pull the pulse generator cable connector X402 off board DX1.
2. Detach the motor cable from the cable harness and carefully pull it
out of the stand.
3. Pull the motor connecting cable off of the filter.
4. Loosen the three screws (M).
5. Remove the motor while carefully pulling the motor cable out of the
stand.
61 25 665 D3437
296 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.2 Height adjustment motor (M1_4)/spindle
Installing the new motor Install the height adjustment motor in the reverse order of removal.
Please observe the following:
Nuts: When fastening the motor, make sure that all three screws are
tightened uniformly and protrude approx. 3 mm out of the nut.
Acorn nuts: If acorn nuts have been installed in the unit, turn the acorn
nuts to the end stop.
CAUTION! Do not forget to reattach all connectors or cables, route them
in their original position and reattach all cable ties and cable clamps.
Make sure that none of the cables are crushed by the cover plates of the
DX32 connection box.
Final work With the "height adjustment" service kit
1. After reinstalling the spindle, screw the first of the two screws (E)
back into the spindle holder.
2. Then remove the locking pin and screw in the second of the two
screws (E).
båÖäáëÜ
3. Attach the cover (D).
4. Remove the clamp (A).
5. Only then should you check the travel function of the slide.
Without the "height adjustment" service kit
1. Attach the spindle holder (F) with the two screws (E).
2. Attach the cover (D).
3. Reattach the upper limit stop to the previously marked position.
4. Only then should you check the travel function of the slide.
61 25 665 D3437
D3437.076.01.15.02 06.2012 297
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
2. Connect the protective ground wire (P) and lay it as shown in the
photo.
61 25 665 D3437
298 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.2 Height adjustment motor (M1_4)/spindle
3. Attach cable L2 first to the lower strain relief (photo on the left) and
then to the upper strain relief (photo on the right) (Q) of board DX32.
4. Connect cable L2 to board DX32 (R) and attach the protective ground
wire (S).
båÖäáëÜ
5. Run cable L3 (T) and the motor cable (U) around the height
adjustment motor.
61 25 665 D3437
D3437.076.01.15.02 06.2012 299
9 Repair Sirona Dental Systems GmbH
9.2 Height adjustment motor (M1_4)/spindle Service Manual GALILEOS
6. Lay the motor cable in the cable harness (V) on the rear of the unit
and secure in position with the cable clamps.
7. Route the cable into the arm.
IMPORTANT: The green mark must lie in the recess (X).
8. Plug connector X402 (W) into board DX1.
1. After inserting the new spindle above and below the height
adjustment motor, grease it thoroughly with Chesterton 622.
2. Use the Up/Down keys on the control panel to check the function of
the height adjustment motor.
3. Reset the travel height.
61 25 665 D3437
300 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.3 Ring motor (M1_3)
A D X813
båÖäáëÜ
E
X804
1. Detach the motor cable from the cable harness and pull it off of
connector X813 on board DX1.
2. Loosen the four screws (A) on the ring motor and remove the motor
including the screws and the serrated washers (B).
61 25 665 D3437
D3437.076.01.15.02 06.2012 301
9 Repair Sirona Dental Systems GmbH
9.3 Ring motor (M1_3) Service Manual GALILEOS
Installing the new motor 1. Insert the new motor including coupling and absorber in the ring.
Tip: While inserting the motor, turn it back and forth slightly until the
pinion engages in the ring gear.
2. Use the screws (A) and serrated washers (B) to screw the new motor
onto the motor support ring.
3. Run the ring motor cable along its original path and plug it back into
connector X813 on board DX1.
IMPORTANT: Don't forget to reattach all cable ties and clamps.
Attaching the covers ➢ Reattach the covers.
61 25 665 D3437
302 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.3 Ring motor (M1_3)
båÖäáëÜ
C
A B
61 25 665 D3437
D3437.076.01.15.02 06.2012 303
9 Repair Sirona Dental Systems GmbH
9.3 Ring motor (M1_3) Service Manual GALILEOS
Replacing the pinion 1. Loosen the set screws (A) and pull off the defective pinion (B).
2. IMPORTANT: Ensure that the pinion is seated in the coupling so that
the set screws (A) are sitting on the flattened surface (C) of the pinion
during subsequent tightening to prevent the pinion from turning.
Insert the new pinion.
3. IMPORTANT: Apply Loctite 242 to the set screws (B) before
tightening.
Retighten the set screws (A).
Installing the motor ➢ Reinsert the motor in the ring, route the cable and connect the motor
as described in the chapter Replacing the ring motor [ → 301] .
Attaching the covers ➢ Reattach the covers.
61 25 665 D3437
304 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.3 Ring motor (M1_3)
båÖäáëÜ
1. Lay the cable (G) parallel to cable L3 and secure it with the clamps.
2. Plug connectors X804 (I) and X813 (H) into board DX1.
61 25 665 D3437
D3437.076.01.15.02 06.2012 305
9 Repair Sirona Dental Systems GmbH
9.4 Rotary knob on the swivel arm Service Manual GALILEOS
61 25 665 D3437
306 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.5 Control panel
båÖäáëÜ
GALILEOS Compact
GALILEOS Comfort
1. Press into slit (A) of the housing cover with a screwdriver (do not pry!)
and remove the defective user interface from the control panel.
2. Pull cables L9 and L10 off of connectors X102 (L9) and X103 (L10)
on board DX7 (Easypad, "GALILEOS Comfort") or DX71 (Multipad,
"GALILEOS Compact") of the defective user interface.
3. Plug the cables into connectors X102 (L9) and X103 (L10) of board
DX7 (Easypad, "GALILEOS Comfort") or DX71 (Multipad,
"GALILEOS Compact") on the new user interface.
4. Clip the new user interface onto the control panel.
61 25 665 D3437
D3437.076.01.15.02 06.2012 307
9 Repair Sirona Dental Systems GmbH
9.5 Control panel Service Manual GALILEOS
IMPORTANT: So that the board is also replaced with the user interface,
you MUST also follow the instructions in the chapter titled "Measures
following replacement of boards [ → 347]“.
1. Check that the user interface and the display elements are
functioning correctly: When the unit is switched on, all of the display
elements must light up briefly.
2. Perform a software update to the latest version [ → 61].
GALILEOS with Easypad
Easypad only Following replacement of the user interface, the language set on board
DX7 is set to the factory setting by default (00 = German, English, French,
Italian). If the configured unit language set (which can be queried by
running service routine S017.5 or using the “extended detail query“ in
SiXABCon) has a configuration other than 00, this configuration will be
copied to board DX7 by the update function.
61 25 665 D3437
308 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.5 Control panel
Easypad
båÖäáëÜ
1. Plug the green cable L10 (A) into connector X103 on board DX7.
2. Plug the gray cable L9 (B) into connector X102 on board DX7.
Laying of cables: Multipad
Multipad
1. Plug the green cable L10 (D) into connector X103 on board DX71.
2. Plug the gray cable L9 (C) into connector X102 on board DX71.
61 25 665 D3437
D3437.076.01.15.02 06.2012 309
9 Repair Sirona Dental Systems GmbH
9.6 X-ray tube unit Service Manual GALILEOS
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least another 4 minutes
before starting the repair or taking off a cover panel!
NOTICE
Oil leakage!
The screws (E) from the ring (X-ray window) must not be loosened, as
otherwise oil may leak out.
61 25 665 D3437
310 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.6 X-ray tube unit
NOTICE
Oil leakage!
The screws (E) from the ring (X-ray window) must not be loosened, as
otherwise oil may leak out.
båÖäáëÜ
1. Loosen the two lower screws (A) (approx. 2 to 3 turns).
2. Push the diaphragm (B) upward and then toward the front.
3. Unscrew the two sleeves (C) with washers.
4. Remove the aluminum filter (D).
5. Keep diaphragm (B) safe and store aluminum filter (D) in a safe
place.
IMPORTANT: These components will be reused when attaching the
diaphragm unit to the new tube assembly.
61 25 665 D3437
D3437.076.01.15.02 06.2012 311
9 Repair Sirona Dental Systems GmbH
9.6 X-ray tube unit Service Manual GALILEOS
1. Loosen the four screws (A) and remove cover plate (B) incl. the cable
shielding (L3). CAUTION! Also pull cable L3 off connector X3 and the
ground cable off connector X304 on board DX6.
Tip: The ferrite core and cable shielding can remain on the cover
plate.
2. Detach cables L5, L6 and L15 from the rubber grommets and pull the
cables off of sockets J6 (L5), J2-J3 (L6) and J5 (L15) on board DX6.
61 25 665 D3437
312 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.6 X-ray tube unit
båÖäáëÜ
3. Loosen the two rear screws (C) on the tube assembly.
4. CAUTION! The tube assembly is heavy!
Hold the tube assembly firmly in place, loosen the two front screws
(D) (3-4 turns) and remove the tube assembly toward the front.
Preparing the new X-ray tube assembly
NOTICE
for installation
Oil leakage!
The screws (E) from the ring (X-ray window) must not be loosened, as
otherwise oil may leak out.
IMPORTANT
Note the order for attaching the filters when reattaching the diaphragm
unit to the new tube assembly. If the filters are installed in the wrong
order, this will impair the image quality. Insert the aluminum filter (D)
first, followed by the copper filter (M).
61 25 665 D3437
D3437.076.01.15.02 06.2012 313
9 Repair Sirona Dental Systems GmbH
9.6 X-ray tube unit Service Manual GALILEOS
F F
D
M
F L
C
61 25 665 D3437
314 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.6 X-ray tube unit
Installing the diaphragm unit ➢ NOTICE! Do not forget to re-install the filters or in the case of "Type
1" diaphragm unit, the primary diaphragm.
The diaphragm is installed in the reverse order to removal.
With the "Type 3" diaphragm:
IMPORTANT
Note the order for installing the filters!
The image quality will be impaired if the filters are installed in the wrong
sequence. Insert the aluminum filter (D) first, followed by the copper filter
(M).
båÖäáëÜ
Set the adjusting knob in place with the silicone ring.
Updating the ID label
61 25 665 D3437
D3437.076.01.15.02 06.2012 315
9 Repair Sirona Dental Systems GmbH
9.6 X-ray tube unit Service Manual GALILEOS
61 25 665 D3437
316 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.6 X-ray tube unit
båÖäáëÜ
A Cable L5 → Socket J6 on board DX6
B Cable L6 → Socket J2/J3 on board DX6
C Grounding cable → Connector X304 on board DX6
D Cable L3 → Connector X3 on board DX6
E Laying cables correctly on the cover plate
F Cable L15 → Socket J5 on board DX6
G Cable routed on left side of tube assembly:
2x L21 and L20
H Cable L12 routed on right side of tube assembly.
IMPORTANT
Since board DX6 is also replaced with the tube assembly, you MUST
also follow the instructions in the chapter entitled Measures following
replacement of boards [ → 347].
61 25 665 D3437
D3437.076.01.15.02 06.2012 317
9 Repair Sirona Dental Systems GmbH
9.7 Fan (X-ray tube assembly) Service Manual GALILEOS
2. NOTICE! Cable! Loosen the three screws (A) and carefully remove
the cover plate including the fan.
3. Pull the fan cable off of connector X2 on board DX6.
4. Install the new fan in the reverse order of removal.
➢ Check the function of the fan using service routine S005.4 [ → 225].
61 25 665 D3437
318 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.8 X-ray detector
IMPORTANT
For Facescan units:
The FaceScan must be removed [ → 325] from units with
FacescanFacescan fitted before the X-ray detector can be replaced.
båÖäáëÜ
1. CAUTION! Risk of injury! The cover plate may have sharp edges.
Carefully pull cover plate (A) upwards to remove it from the X-ray
detector.
2. Loosen the screws (B) and (C), as well as clamps (D) and (E).
3. Remove cable L13 from connector X201 (F) on board DX89.
61 25 665 D3437
D3437.076.01.15.02 06.2012 319
9 Repair Sirona Dental Systems GmbH
9.8 X-ray detector Service Manual GALILEOS
4. NOTICE! Make sure that the grounding cable does not slip into the
ring. Secure it with a cable tie or piece of adhesive tape if necessary.
Loosen screw (G) and disconnect the grounding cable (H).
Removing the X-ray detector
➢ CAUTION! The x-ray detector is heavy!
Loosen the screw (J), swing the X-ray detector slightly upwards and
lift it out of the holder on the ring.
IMPORTANT: Depending on the unit hardware version involved,
there may be a second screw located on the side opposite screw (J).
If so, this screw(*) must be loosened in order to remove the X-ray
detector. This second screw does not have to be used during
reassembly.
61 25 665 D3437
320 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.8 X-ray detector
1. Hook the new x-ray detector into holder (K) from above, using the ring
on the unit. The dead weight of the X-ray detector will cause it to tilt
into the correct position.
båÖäáëÜ
2. Secure it in place using the screw (J).
3. CAUTION! Risk of injury! The cover plate may have sharp edges.
Carefully pull the cover plate (K) upwards to remove it from the X-ray
detector.
4. Plug cable L13 (from the ring) onto connector X201 on PCB DX89
and use the two screws to secure it.
61 25 665 D3437
D3437.076.01.15.02 06.2012 321
9 Repair Sirona Dental Systems GmbH
9.8 X-ray detector Service Manual GALILEOS
5. Connect the grounding cable from the ring with screw (G) as well as
with washer (L), serrated washer (M) and contact washer (N).
61 25 665 D3437
322 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.8 X-ray detector
6. NOTICE! Ensure that cable L13 is correctly laid in the nut (O) of the
X-ray detector.
Route the cable L13 as illustrated in the diagram, and attach the
cover shielding on the X-ray detector housing using the 2 clamps (D)
and screws (B).
Depending on the unit hardware version, the brackets on your unit
may differ slightly from those shown.
7. Secure the cable with clamp (E) and screw (C).
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 323
9 Repair Sirona Dental Systems GmbH
9.8 X-ray detector Service Manual GALILEOS
IMPORTANT
For Facescan units:
The FaceScan must be refitted [ → 327] on units with Facescan fitted
after the X-ray detector has been replaced.
61 25 665 D3437
324 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
9.9 Facescan
Facescan
båÖäáëÜ
2. Disconnect the cable (H) from the display board (I) on the inside of
the cover (F).
3. Undo the six screws (G) and remove the cover (F).
61 25 665 D3437
D3437.076.01.15.02 06.2012 325
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
6. Pull gray cable L78.3 from socket X1 off the FACESCAN modular
board.
7. Refit the panel (T) with the two screws (S).
61 25 665 D3437
326 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
båÖäáëÜ
2. Unscrew the six screws (D) and detach the scan unit from the X-ray
detector brackets to the rear.
NOTICE
Risk of damage
Mechanical stress can cause damage to the scanning unit.
➢ For transporting and aligning the scan unit hold on to the transport
lock (B) only.
NOTICE
➢ Do not remove the wing nuts.
IMPORTANT
Risk of damage
When hanging the scan unit, the housing of the X-ray detector can get
scratched.
➢ Hang the scan unit carefully on the X-ray detector.
61 25 665 D3437
D3437.076.01.15.02 06.2012 327
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
1. Hang the scan unit from behind in the inlets of the X-ray detector.
2. Attach the scan unit loosely with 6 screws (D) - do not screw tightly!
3. Mount the installation aid (C) from below on both sides onto the first
cooling vent (H) of the face scan unit.
4. Insert the fastening element (F) into the installation aid (C).
5. Turn the holding plates (G) over the chassis plate of the face scan
unit.
6. Push the installation aid (C) with the scan unit on to the sensor
surface (H) of the X-ray detector.
7. Align the installation aid (C) to the center of the sensor surface (H).
IMPORTANT
Possible faulty alignment
Displacement of the installation aid (C) can be caused through further
fastening of the scan unit with the safety screws (E).
➢ When screwing in the safety screws (E), please ensure that the
installation aid (C) always remains aligned.
61 25 665 D3437
328 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
8. Move the safety screw (E) on both sides from the parked position (L)
(delivered condition) to the function position (M).
9. Screw the scan unit down from below with the safety screws (E). In
so doing, the screws should pierce through the housing of the X-ray
detector.
ª The distance (K) between the plate of the scan unit and the cover
of the X-ray detector should be 6 mm.
båÖäáëÜ
10. Screw down the 6 screws (D).
11. Remove the installation aid (C) (stays with the customer).
NOTICE
Keep the transport lock (B) and the wing screws (A) in case the
Facescan unit has to be returned for repairs (these stay with the
customer).
61 25 665 D3437
D3437.076.01.15.02 06.2012 329
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
61 25 665 D3437
330 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
IMPORTANT
Permitted cleaning agents
- A dry, lint-free cloth
- A cleaning agent approved by Sirona
If you have no access to the Internet, please contact your dental depot,
to request the list.
REF 59 70 905
båÖäáëÜ
2. Clean the surface of the mirror and the vision panel inside the
Facescan cover.
3. Screw down the covering bonnet (F) from below with 6 screws (G).
4. Connect the cable (H) with the display board (I) to the covering
bonnet (F).
61 25 665 D3437
D3437.076.01.15.02 06.2012 331
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
5. Read the serial number on the identification plate (S) and record this
on the device certificate.
6. Screw the right end cap (J) with two screws (K) down on the scan unit.
7. If network configuration is to be completed later using the Facescan
USB stick:
Do not screw the cover cap (A) tight.
or
➢ If network configuration is to be completed later using a network
cable:
Screw the cover cap (A) down with two screws.
61 25 665 D3437
332 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
båÖäáëÜ
3. Detach cable L77.WH (white) and L77.BN (brown) from terminals
K1.2 and K1.1 (orange).
4. Pull cables L71 and L7 from sockets SC:SC and BU1.
61 25 665 D3437
D3437.076.01.15.02 06.2012 333
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
7. Unscrew the two screws (K) and the nuts (L) and remove the PoE
module.
61 25 665 D3437
334 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
(J) (the serrated washer must sit behind the fitting plate of the PoE
module).
2. Screw down the PoE module (P) to the place provided in the stand
with two screws (K) and a nut (L).
båÖäáëÜ
IMPORTANT
The locking catch of the RJ45 plug of Ethernet cable L77 is secured with
adhesive tape (N).
➢ Remove the adhesive tape (N) from the RJ45 plug of Ethernet cable
L77.
3. Plug Ethernet cable L77 into the POE socket of the PoE module.
61 25 665 D3437
D3437.076.01.15.02 06.2012 335
9 Repair Sirona Dental Systems GmbH
9.9 Facescan Service Manual GALILEOS
IMPORTANT
Cable L76 must be routed behind the mains filter plate (J).
9. Route cable L76 behind the mains filter plate to the terminal K1.
61 25 665 D3437
336 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.9 Facescan
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 337
9 Repair Sirona Dental Systems GmbH
9.10 Head fixation device Service Manual GALILEOS
Remove defective receptacle element. 1. Using the Up/Down buttons on the control panel, switch the device on
and move it to a comfortable working height to remove the acquisition
unit.
2. Remove the head fixation device (see operating instructions).
3. If the bore hole (C) on the defective acquisition unit is not available:
Press the locking button (A) and move the flange (B) forward to
expose the bore hole (C).
61 25 665 D3437
338 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.10 Head fixation device
båÖäáëÜ
2. IMPORTANT: Do not tighten the screw yet. It should not be possible
to rotate the acquisition unit.
Screw the new receptacle element on to the unit with screw (D) so
that the laser localizer (E) is facing forward.
3. Push the head fixation device (F) into the acquisition unit (see
operating instructions).
4. Press the locking button (A) and push the flange (B) including the
head fixation device backwards so that the light localizer (E) is
exposed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 339
9 Repair Sirona Dental Systems GmbH
9.10 Head fixation device Service Manual GALILEOS
61 25 665 D3437
340 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.11 Light barriers
båÖäáëÜ
The following light barriers can be replaced:
● Light barrier at ring motor, starting position of rotation: V1_3
● Light barrier at HA motor, height adjustment: V1_4
61 25 665 D3437
D3437.076.01.15.02 06.2012 341
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
9.12 Boards
Boards
NOTICE
Touching the boards can damage them.
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
Replacing the boards DX6 (X-ray tube Never replace these boards at the same time. After replacing one of these
assembly) and DX11 or DX89 and DX11 boards, you must first perform the measures specified in the chapter and
then restart the unit. Only then may you begin replacement of the other
module.
Replacing DX11: GALILEOS
61 25 665 D3437
342 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
CAUTION
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD).
Touch a ground point to discharge static electricity before touching any
boards.
Replacing PC boards: GALILEOS
båÖäáëÜ
*) Board DX41 is omitted in units with serial numbers of 3201 and higher.
61 25 665 D3437
D3437.076.01.15.02 06.2012 343
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
IMPORTANT
The software version of the "DX1/DX11 board" must be compatible with
the main software version of the unit.
After replacing the board DX1, you must observe the notes provided in
the chapter .
IMPORTANT
The software version of the "DX1/DX11 board" must be compatible with
the main software version of the unit.
IMPORTANT
The cover plates of the DX1 (REF 59 24 142 and REF 62 82 052) board
versions are not compatible with each other.
9.12.2.2.1 In the case of boards with REF 59 24 142 (DX1) and REF 59 25
214 (DX11)
In the case of boards with REF 59 24 142 (DX1) and REF 59 25 214 (DX11)
61 25 665 D3437
344 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
9.12.2.2.2 In the case of boards with REF 62 82 052 (DX1) and REF 63 17
056 (DX11)
In the case of boards with REF 62 82 052 (DX1) and REF 63 17 056 (DX11)
båÖäáëÜ
Replacing DX11: GALILEOS
After replacing the board DX11, you must observe the notes provided in
the chapter .
DX1
After replacing the board DX32, you must observe the notes provided in
the chapter .
61 25 665 D3437
D3437.076.01.15.02 06.2012 345
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
3. Remove the four screws (B) and remove the defective DX89 board
from the X-ray detector.
4. Pull the connectors of cables L13 (X201), L27 (X203) and L28 (X400)
off of the defective board DX89.
5. Install the new board DX89 by following the steps for removal in
reverse order.
IMPORTANT: After replacing the board DX89, always observe the notes
provided in the chapter .
61 25 665 D3437
346 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
båÖäáëÜ
required for the corresponding situations following board replacement.
61 25 665 D3437
D3437.076.01.15.02 06.2012 347
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
61 25 665 D3437
348 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
båÖäáëÜ
● Inserting a new DX89 ● Switch the unit on. S. [ → 61]
detector
● GALILEOS Comfort ● Perform a software update of the unit
to a version V03.03.01 or higher.
● System software version
V03.03.01 or higher ● Restore the configuration data of S. [ → 245]
board DX89 by using service routine
S009.7.
● Inserting a new DX89 ● Switch the unit on. S. [ → 61]
● GALILEOS Compact ● Perform a software update of the
device to a version V03.06.01 or
● System software version
higher.
V03.06.01 or higher
● Restore the configuration data of S. [ → 245]
board DX89 by using service routine
S009.7.
*) Board DX41 is omitted in unit serial number 3201 and higher. Board
DX41 is available as a spare part for units up to unit serial number 3199.
61 25 665 D3437
D3437.076.01.15.02 06.2012 349
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
NOTICE
After a new DX11 is inserted, the IP address is initially reset to the
factory setting. Before you set the unit to a new IP address, make sure
that the IP address you're assigning has not been assigned to any other
unit.
61 25 665 D3437
350 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
15. Wait for approx. 1 minute. Then switch the unit back on.
The error message E6 15 05 (undefined system serial number) is
displayed.
16. Acknowledge the error message with the R key .
17. Error message E6 15 04 (undefined activation data) is displayed.
18. Acknowledge the error message with the R key .
19. Call the service menu [ → 210].
20. Call the service routine S008.3, check the serial number, and confirm
this if necessary [ → 240].
The unit serial number is located on the nameplate of the unit.
NOTICE! If the serial number is incorrect, exit the update process and
contact the Sirona Customer Service Center.
21. Switch off the unit.
22. Wait for approx. 1 minute. Then switch the unit back on.
The message "No Key" should no longer appear.
23. Call the service menu [ → 210].
båÖäáëÜ
24. Call the service routine S017 and perform the unit configuration (test
step 2-15) [ → 252].
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the welcome screen or acoustic exposure signal.
Activate these functions if they are required.
25. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
26. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
The error message E1 11 20 (invalid unit calibration) is displayed.
27. Acknowledge the error message with the R key .
28. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
29. Call up the "Extended Details" via SiXABCon.
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 351
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
NOTICE
After inserting the board, you must reconfigure the IP address to match
the IP address of the existing X-ray component. Before you set the unit
to a new IP address, make sure that the IP address you're assigning has
not been assigned to any other unit.
61 25 665 D3437
352 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
14. Call the service routine S017 and perform the unit configuration (test
step 2-15) [ → 252].
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the welcome screen or acoustic exposure signal.
Activate these functions if they are required.
15. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
16. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
17. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
18. Call up the "Extended Details" via SiXABCon.
båÖäáëÜ
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 353
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
NOTICE
After a new DX11 is inserted, the IP address is initially reset to the
factory setting. Before you set the unit to a new IP address, make sure
that the IP address you're assigning has not been assigned to any other
unit.
61 25 665 D3437
354 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
15. Wait for approx. 1 minute. Then switch the unit back on.
The error message E6 15 05 (undefined system serial number) is
displayed.
16. Acknowledge the error message with the R key .
17. Error message E6 15 04 (undefined activation data) is displayed.
18. Acknowledge the error message with the R key .
19. Call the service menu [ → 210].
20. Call the service routine S008.3, check the serial number, and confirm
this if necessary [ → 240].
The unit serial number is located on the nameplate of the unit.
NOTICE! If the serial number is incorrect, exit the update process and
contact the Sirona Customer Service Center.
21. Switch off the unit.
22. Wait for approx. 1 minute. Then switch the unit back on.
The message "No Key" should no longer appear.
23. Call the service menu [ → 210].
båÖäáëÜ
24. Call the service routine S017 and perform the unit configuration (test
step 2-15) [ → 252].
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the acoustic exposure signal. Activate these
functions if they are required.
25. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
26. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
The error message E1 11 20 (invalid unit calibration) is displayed.
27. Acknowledge the error message with the R key .
28. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
29. Call up the "Extended Details" via SiXABCon.
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 355
9 Repair Sirona Dental Systems GmbH
9.12 Boards Service Manual GALILEOS
NOTICE
After inserting the board, you must reconfigure the IP address to match
the IP address of the existing X-ray component. Before you set the unit
to a new IP address, make sure that the IP address you're assigning has
not been assigned to any other unit.
61 25 665 D3437
356 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.12 Boards
14. Call the service routine S017 and perform the unit configuration (test
step 2-15) [ → 252].
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the welcome screen or acoustic exposure signal.
Activate these functions if they are required.
15. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
16. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
17. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
18. Call up the "Extended Details" via SiXABCon.
båÖäáëÜ
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.
61 25 665 D3437
D3437.076.01.15.02 06.2012 357
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
9.13 Cable
Cable
IMPORTANT
Remove cable ties
For the following steps, all necessary cable ties should be removed with
wire cutters.
DANGER
Danger of fatal electrocution!
➢ Before you remove the energy chain, switch off the power supply.
61 25 665 D3437
358 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.13 Cable
4. If not present: Mark the position of the screw (B) on the stand.
5. Make a note of the position of the screw (B) for when you later install
the new energy chain.
6. Remove the screw (B) from the energy chain 1.
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 359
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
9. Remove the cable covers (I) and (J) of the right-hand cable duct.
10. Remove the energy chain along with the mains filter plate from the
stand.
1. Lead the energy chain up through the stand from its base.
2. Insert new energy chain 1 in the stand.
IMPORTANT
Assembly instructions
➢ Pay attention to the energy chain's rolling direction.
61 25 665 D3437
360 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.13 Cable
IMPORTANT
Possible assembly errors
The one end piece has been removed to make the new energy chain
easier to lay .
The missing end piece is enclosed.
➢ Attach this end piece to the energy chain only when the energy
chain has been laid in the stand.
3. Screw the new energy chain down in the position (marking) of the old
energy chain.
4. Screw down the mains filter plate.
5. Screw the power cable (A) to the terminal K1 and the strain relief.
6. Lead the external PE cable (B) over the mains filter plate through the
ferrite core twice (C) (a hose).
båÖäáëÜ
7. Screw the external PE cable (B) down on the ground bolts (D).
61 25 665 D3437
D3437.076.01.15.02 06.2012 361
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
CAUTION
Switch the unit off before you start replacing cables or removing
connectors.
NOTICE
Be careful not to twist the cables or kink the fiber-optic light guides when
installing them.
Cable replacement; GALILEOS
61 25 665 D3437
362 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.13 Cable
båÖäáëÜ
2. NOTICE! Do not kink or twist fiber-optic cables, the bending radius
may not be less than 20mm, otherwise it is at risk of breaking!
Attach the radius limiter (A) close to the connector, which is plugged
onto board DX1, onto the cable.
3. Plug the connector of the new fiber-optic cable to the same color
assignment on the board DX1.
4. Lay the fiber-optic cables up to point (B), and clip the radius limiter (A)
at point (B) (approx. 900 mm from the connectors on DX1) onto the
cable.
5. Guide the fiber-optic cable to board DX6 and plug the connector of
the new fiber-optic cable to the same color assignment on board DX6.
61 25 665 D3437
D3437.076.01.15.02 06.2012 363
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
9.13.2.2 Cable replacement (L3, L5, L6, and L15)/Laying the cable/
corrugated tube at the rotation unit
Cable replacement (L3, L5, L6, and L15)/Laying the cable/corrugated tube at the rotation unit
NOTICE
The connectors and cables must be protected by inserting them in the
fabric tube (A) supplied with the cables.
1. For L3: Remove the connector of the cable with tool W1.
2. Remove the corrugated tube and the spiral spring (B) from the cable
loom.
3. Remove the defective cable and run the new cable up to the rotary
ring in the original position.
4. Bunch the cables together again to form a loom.
5. Fasten the defective cable to the loom and use it as a pull wire to pull
the loom through the fabric tube (A).
6. Pull the fabric tube over the connector and as far over the cable loom
as possible.
7. Use the pull wire to pull the fabric tube into the spiral spring (B).
61 25 665 D3437
364 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.13 Cable
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 365
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
13. Unscrew the angle brackets on both sides of the cable track.
61 25 665 D3437
366 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 9 Repair
Service Manual GALILEOS 9.13 Cable
14. Remove the motor-side end piece from the cable track.
15. With defective cable L7/L117: Unscrew cable L7/L117 from the
interface board and remove the shield.
If cable L7/L117 should be intact and used again, this step is not
necessary. Unless it is not possible to lay down the cable track flat
near the stand (see next step).
16. Remove the cable ties from the cable track and lay the cable track
down on a flat surface stretched out.
båÖäáëÜ
17. CAUTION! You must observe the position of connector X303 of cable
L108 (see image).
Carefully pull both cables (together) out of the cable track and the
fabric tube.
18. With defective cable L108: Wrap the connector X303 of the new cable
L108 with adhesive tape to protect the detent against breaking off.
19. Lay the cable track down on a flat surface stretched out.
20. Fasten the two (new) cables together with adhesive tape above the
flag labels.
61 25 665 D3437
D3437.076.01.15.02 06.2012 367
9 Repair Sirona Dental Systems GmbH
9.13 Cable Service Manual GALILEOS
21. CAUTION! Push the green cable. The white cable is carried along. In
this way, you can prevent the sensitive fiber optic cable from being
damaged. Push both cables (together) into the cable track up to the
cable markings.
New cables do not have cable markings. Orientate yourself according
to the marking on the second (old) cable and make sure that both
cables protrude equally far out of the cable track once they have been
drawn in. Then make a mark on the new cable.
22. NOTICE! The cable ties should only fix the position of the cables.
They must not be tightened too much, otherwise fiber-optic cable L7
could be damaged.
Before installing the cable track in the stand, fix the cables at both
ends of the cable track with a cable tie.
23. Reinstall the cable track in the stand.
Installation of the cable track is performed in reverse order of the
removal.
The procedure for replacing cable L1 and the grounding strap is basically
analogous to the procedure described in chapter Replacing cable L7/
L117 or L108 in cable track 2 [ → 366].
61 25 665 D3437
368 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.1 Calibrating the unit
10 Maintenance
Maintenance
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least 1 minute, or 4
minutes if disconnecting the tube assembly (cable L3), before starting
the maintenance or taking off a cover panel!
CAUTION
Risk of electric shock!
Always switch the unit off before ...
...connecting a measuring instrument or
...carrying out continuity checks.
båÖäáëÜ
NOTICE
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
NOTICE
Risk of damage to tube assembly
Keep to the prescribed cool-off periods if several exposures have to be
taken to check a measured value.
61 25 665 D3437
D3437.076.01.15.02 06.2012 369
10 Maintenance Sirona Dental Systems GmbH
10.2 Checking the height adjustment Service Manual GALILEOS
Check the threaded rod and motor for ➢ Perform a visual inspection of height adjustment motor (A) and
abrasion spindle (B) for abrasion.
If significant abrasion is present:
ª Replace the height adjustment motor including spindle [ → 290].
Check whether the height adjustment ➢ Use the Up/Down keys on the control panel to move the unit up and
produces atypical running noises down through its entire adjustment range.
If the mechanics of the height adjustment is defective, a speed-
dependent hammering noise may occur which points to bearing
damage at the height adjustment motor.
If a hammering noise occurs:
ª Replace the height adjustment motor including spindle [ → 290].
Check whether precise, jolt-free height If the unit is not used for a longer period of time, a slight jolt may occur the
adjustment is possible first time it starts moving. However, the next time it starts moving, it must
execute a jolt-free soft start.
➢ Use the Up/Down keys on the control panel to move the unit and
observe the movement of the slide. The slide must start in gentle
starting and then change over to a faster movement.
If the height adjustment cannot be correctly positioned in detail using
the gentle start:
ª Lubricate the spindle with a light coat of Chesterton 622.
61 25 665 D3437
370 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.3 Checking the fan and temperature sensor
50
D
båÖäáëÜ
Check whether an audible signal can be ➢ Use the Up/Down keys on the control panel to move the unit up and
Checking height adjustment signal: GALILEOS
Check whether the fan is functioning ➢ Check the function of the fan using service routine S005.4 [ → 225].
ª If the fan is defective: Replace the fan [ → 318].
Check whether the temperature sensor ➢ Read the temperature in the single tank with service routine
is supplying plausible values S005.5 [ → 226].
ª If the displayed temperature reading is not plausible: Replace the
tube assembly.
Check whether the cables feeding the ➢ Perform a visual inspection of the power cable, protective ground
unit are OK wire, control cables and data cables.
If cables exhibit external damage:
ª Replace the respective cable [ → 362].
61 25 665 D3437
D3437.076.01.15.02 06.2012 371
10 Maintenance Sirona Dental Systems GmbH
10.5 Checking the idling rollers Service Manual GALILEOS
Check whether the idling rollers (A) are ➢ Manually turn the ring (B) and check it for smooth and easy
OK movement.
If the ring does not move smoothly and easily:
ª Remove the housing covers and check the idle rollers (A) for dirt
and foreign particles. Clean and remove foreign particles if
necessary.
61 25 665 D3437
372 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.6 Checking the grounding straps
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 373
10 Maintenance Sirona Dental Systems GmbH
10.6 Checking the grounding straps Service Manual GALILEOS
Check whether the grounding straps 1. Perform a visual and “hands-on” inspection of the grounding straps to
have complete and firm contact ensure that they have complete and firm contact at the positions
marked.
If the grounding straps do not have correct contact:
ª Fasten the grounding straps correctly.
2. Perform a visual inspection of the grounding straps for damage.
If the grounding straps are damaged:
ª Replace the grounding straps.
61 25 665 D3437
374 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.7 Checking the cable shields
båÖäáëÜ
Shield on the X-ray detector: GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012 375
10 Maintenance Sirona Dental Systems GmbH
10.7 Checking the cable shields Service Manual GALILEOS
61 25 665 D3437
376 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.8 Checking the protective ground wires
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least one more 1 minute
before beginning the check!
1. Switch the line voltage off at the main switch of the building
installation.
2. Disconnect the power cable and the second protective ground wire
from the building installation.
3. Remove the "bottom profile", "top tube assembly", and "bottom
tube assembly" covers.
Measuring setup - Ground wire test
båÖäáëÜ
쓓
A A
A B C
V D
B
C E
D
E
A Ammeter
B Power source
C Measuring point A, central protective ground wire
D Voltmeter
E Measuring points B - E
61 25 665 D3437
D3437.076.01.15.02 06.2012 377
10 Maintenance Sirona Dental Systems GmbH
10.8 Checking the protective ground wires Service Manual GALILEOS
61 25 665 D3437
378 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.8 Checking the protective ground wires
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 379
10 Maintenance Sirona Dental Systems GmbH
10.8 Checking the protective ground wires Service Manual GALILEOS
61 25 665 D3437
380 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.8 Checking the protective ground wires
båÖäáëÜ
D Board cage DX32
E Tube assembly housing
61 25 665 D3437
D3437.076.01.15.02 06.2012 381
10 Maintenance Sirona Dental Systems GmbH
10.9 Checking the device leakage current Service Manual GALILEOS
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least one more 1 minute
before beginning the check!
B
L
N
PE
A 쓓
A B
MD
1kOhm
0,015µF
10kOhm
V D
61 25 665 D3437
382 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 10 Maintenance
Service Manual GALILEOS 10.9 Checking the device leakage current
1. Switch the line voltage off at the main switch of the building
installation.
2. Disconnect the power cable and the second protective ground wire
from the building installation.
3. Remove the "profile" cover.
4. Check whether the unit power switch is turned on.
5. Connect a high-resistance measuring voltage source between the
short-circuited power cable (B) and the protective ground wire (A).
6. Measure the voltage drop via MD.
The measured value must not exceed 5 mA.
ª If the leakage current is not correct: Troubleshoot the problem
following the instructions in Device leakage current too
high [ → 140] .
7. Reconnect the unit to the building installation (see the installation
instructions for the unit).
båÖäáëÜ
61 25 665 D3437
D3437.076.01.15.02 06.2012 383
11 Dismantling and disposal Sirona Dental Systems GmbH
11.1 Dismantling and reinstallation Service Manual GALILEOS
11.2 Disposal
Your product is marked with the adjacent symbol. Within the European
Disposal
Disposal procedure
We advise that this product is subject to the stipulations in EC guideline
2002/96 governing waste electrical and electronic equipment and must
be disposed of in line with these special requirements within the
European Union (EU).
Prior to disassembly / disposal of the product, it must be fully prepared
(cleaned / disinfected / sterilized).
When disposing of equipment permanently, please proceed as follows:
61 25 665 D3437
384 D3437.076.01.15.02 06.2012
Sirona Dental Systems GmbH 11 Dismantling and disposal
Service Manual GALILEOS 11.2 Disposal
In Germany:
To initiate return of the electrical device, please send a disposal order to
"enretec GmbH".
1. You can find a form for placing a disposal order on the company's
homepage (www.enretec.de) under the menu item "Entsorgung
elektrischer und elektronischer Geräte" (Disposal of electric and
electronic devices). The form can either be downloaded or completed
online.
2. Fill out the form with the corresponding details and send it either as
an online order or fax it to enretec GmbH at +49(0)3304 3919 590.
You can also get in touch with the following contacts for disposal
orders and any questions relating to this you may have:
Phone: +49(0)3304 3919 500;
By e-mail: [email protected]
Mailing address: enretec GmbH, Geschäftsbereich eomRECYCLING
Kanalstrasse 17, 16727 Velten
båÖäáëÜ
ª Your equipment that is not permanently installed will be collected in
the practice, while your permanently installed equipment will be
collected curbside at your address at the agreed time and date.
All disassembly, transport and packaging costs are to be borne by the
owner/operator of the equipment. The disposal itself is free of charge.
The X-ray tube assembly for this product contains an X-ray tube with a
potential implosion hazard, a small amount of beryllium, a lead lining and
mineral oil.
The unit contains counterbalancing weights made of lead.
61 25 665 D3437
D3437.076.01.15.02 06.2012 385
12 Service Manual History Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
386 D3437.076.01.15.02 06.2012
tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK
páêçå~=aÉåí~ä=póëíÉãë=dãÄe
áå=íÜÉ=rp^W
c~Äêáâëíê~≈É=PN páêçå~=aÉåí~ä=póëíÉãë=ii` lêÇÉê=kç SN=OR=SSR=aPQPT
SQSOR=_ÉåëÜÉáã QUPR=páêçå~=aêáîÉI=pìáíÉ=NMM
dÉêã~åó `Ü~êäçííÉI=k`=OUOTP
ïïïKëáêçå~KÅçã rp^