Professional Documents
Culture Documents
Amulet Innovality Operation Manual
Amulet Innovality Operation Manual
Operations
Troubleshooting
Appendices
Original Instructions
When the manual is provided in languages other This Operation Manual describes details on basic operations of the FUJIFILM
than English, it is considered to be a “Translation Digital Mammography System FDR MS-3500 and cautions to be observed when
of the original instructions”.
operating it. Please read the Operation Manual thoroughly before actually
In this case, it must bear the words “Translation of
the original instructions” in place of “Original operating the FDR MS-3500 system.
Instructions” above. After reading this manual, store it nearby the FDR MS-3500 system so that you
can see it whenever necessary.
Refer to the reference guide for more details of operation methods and functions
(including options).
897N102525X
Introduction
FUJIFILM Digital Mammography System FDR MS-3500 is a system designed for obtaining digital
mammographic images and providing the obtained images for diagnosis.
This Operation Manual provides necessary information for using the FUJIFILM Digital Mammography
System FDR MS-3500 system (referred to as the system hereafter), such as the overview of the
system, system operation procedures and precautions to observe, as well as daily system inspections
and maintenance.
The FDR MS-3500 consists of the FDR-3500DRLH and FDR-3000AWS workstation.
The FDR-3500DRLH is hereinafter referred to as the exposure unit.
The FDR-3000AWS is hereinafter referred to as AWS.
This Operation Manual applies to the following software.
• FDR-3500H Application Software V5.7 or later
• FDR-3000DRL Application Software V3.8 or later
• AWS Application Software V9.2 or later
Installation may only be conducted by authorized service personal.
CAUTIONS
1 No part or all of this manual may be reproduced in any form without prior permission.
2 The information contained in this manual may be subject to change without prior notice.
3 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by dealers other than
those specified by FUJIFILM Corporation.
4 FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM
Corporation products due to products of other manufacturers not supplied by FUJIFILM
Corporation.
5 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specified by
FUJIFILM Corporation.
6 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of the precautions and operating methods contained in this manual.
7 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range specified for this product, such as the
power supply, installation environment, etc., as described in this manual.
8 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
natural disasters, such as fires, earthquakes, floods, lightning, etc.
Part of the components contains harmful substances which may pollute the ambient
environment if disposed carelessly. In particular, lead is contained in concentrations > 0.1 wt%
in X-ray tube. For details on product disposal, contact our official dealer or FUJIFILM
Representative.
Regulation (EU) 2017/745 requires that any serious incident that has occurred in relation to the
device should be reported to the manufacturer or European Authorized Representative, and the
competent authority of your country. In case you become aware of any serious incident, please
report immediately to our official dealer and the competent authority of your country.
Trademarks
FDR is a trademark or a registered trademark of FUJIFILM Corporation.
Other holders’ trademarks
Windows is the registered trademark of US Microsoft Corporation in the U. S. A. and other countries.
All other company names and product names described in this manual are the trademarks or registered
trademarks of their respective holders.
Copyright © 2012-2020 FUJIFILM Corporation. All rights reserved.
FUJIFILM Digital Mammography System FDR MS-3500 Operation Manual 897N102525X iii
FUJIFILM Digital Mammography System
Operation Manuals
Manage and store all the Operation Manuals of the devices constituting the system together as a set.
FDR MS-3500 Option Tomosynthesis Exposure Tomosynthesis Exposure Software Operation Manual
Indicates hazardous situations that may lead to serious injuries or even death
WARNING if the precaution is not or cannot be followed.
Chapter 3 Operations
3.1 System Basic Operation Procedures......................................................................... 36
3.2 Changing the Compression Plate.............................................................................. 39
3.3 Emergency Stop........................................................................................................ 41
3.4 Sleep Mode................................................................................................................ 42
Chapter 4 Troubleshooting
4.1 Diagnosis System Troubleshooting and Solutions.................................................... 43
4.2 Example of Diagnosis System messages.................................................................. 43
The institution where the equipment is installed is responsible for its use and
maintenance.
In addition, this equipment should not be used by persons other than doctors or
suitably trained staff.
Process waste correctly, as stipulated by local law or any regulations that apply.
• You shall not modify the System and/or its components, and shall not disassemble,
decompile, or in any other way reverse engineer the System and/or its components.
FUJIFILM shall not be liable to you for any defects caused by such modification,
disassembly, decompilation or reverse enginerring, and shall not liable to you for
any damages caused by said defetcs.
• You shall not alter, remove, or obscure any patent markings, copyright notices,
trademark notices, or other proprietary or confidentiality notices that are (i) placed
or embedded in the System and/or its components, (ii) displayed when the System
and/or its components are run, or (iii) applied to the System and/or its components,
their packaging, and labels.
• You shall comply with the applicable U.S. and foreign export laws and regulations
and shall not export, re-export or otherwise transfer the System and/or its
components in violation of the applicable export laws and regulations.
• The System and its components may include parts and/or software supplied by
thrid parties. Any inquiries on the third-party parts and/or software in the System
shall be forwarded to FUJIFILM. Unless explicitly provided otherwise, the third-party
suppliers shall not be liable to end-users of the Systems for their parts and/or
software.
1. This equipment should be used only by people who have the proper skills.
1 2. Observe the following precautions before beginning to use the device.
2-1. Inspect the switch contacts, polarities, dial settings and meters and confirm that the
equipment is functioning properly.
For Safe Operation
2-2. Confirm that the ground wire has been completely connected.
2-3. Make sure that all cords have been connected properly and safely.
2-4. Be aware that correct diagnosis can be hindered and danger can result from using different
pieces of equipment together.
3. Heed the following precautions after using the equipment.
3-1. Using the established procedure, restore the operation switches, dials, etc., to what they
were prior to use and then turn the power off.
3-2. When unplugging cords, do not pull on the body of the cord itself or apply unnecessary
force.
3-3. After using the accessories, recollect them and put them back in order.
4. If there is trouble with the equipment, do not attempt to fix it randomly. Instead, do what is
indicated and entrust repairs to a professional.
5. Do not remodel the equipment.
6. Maintenance and Inspection
6-1. Inspect the equipment and parts periodically.
6-2. If the equipment has not been used for a long time, make sure that it operates normally and
safely prior to using it again.
7. Other Items
7-1. When subjecting patients (particularly infants and pregnant women) to radiation, make sure
not to exceed the necessary time and dose. Also, ensure that radiation is contained within
the region for exposure.
7-2. For cleaning and disinfecting the parts that patients come in contact with in a normal usage
environment, slightly moisten a soft cloth (non-woven cloth: Kimwipe®) in neutral detergent
or ethanol for disinfection and wipe away any dirt. If neutral detergent is used, please also
wipe with disinfectant ethanol. Do not use any other cleansers or disinfectants.
Also, do not scrub the equipment’s surface vigorously.
7-3. Follow the Operation Manual and operate the equipment correctly.
CAUTIONS
• Follow the specified procedure when turning off the equipment. Otherwise, the flat panel
sensor could be damaged by thermal shock.
• Do not store magnetic media near the exposure stand or control cabinet of the exposure unit.
Otherwise, magnetism generated by the equipment may cause the data to be lost.
Connection instructions
WARNING
Make sure that the devices to be connected to the equipment are authorized for connection.
WARNING
• Operate the device by following the instructions given in the Operation Manual and supplied
documentation. Wrong operations could expose a technologist or patient to unnecessary
X-ray exposure.
• Make sure that no one other than a technologist or patient is in the study room when making
X-ray exposures. If it is necessary for a care assistant or other person to remain in the room,
provide sufficient radiation protection when using the remote exposure switch or remote
exposure foot switch.
• Use the longest possible focal spot to skin distance to minimize the absorbed dose.
• Reduce the radiation field size to the minimum necessary size for the exposure. If X-rays are
unavoidably radiated to areas outside of the subject region, take other protective measures
such as readjusting the collimator or wearing a Pb apron.
5
mm
11..5
technologist can make an exposure safely, using the figure below Patient
as a guide.
45°
Protective shield
1
(optional)
WARNING
WARNING
Dangerous parts on the exposure stand
• A patient’s or technologist’s hand, foot or other body (Movable parts)
parts may hit or be caught by the potentially dangerous
parts. Eliminate all dangers completely. Swivel arm
This part rotates.
• When operating the equipment by the foot switch, swivel Face guard
arm control panel or control pad, make sure that no Space between the compression
hand, foot, or other part of the technologist’ s or patient’ plate and face guard or swivel arm
(when the face guard is not installed)
Exposure table
This part moves up/down.
where the protective shield (optional) covers and protects all of his/her body from exposure to
X-rays.
Face guard
The face guard is designed to prevent the patient’s head from entering the radiation field. It is not X-ray
proof.
WARNING
When making an X-ray exposure without using the face guard, make sure that the patient’s
head is physically positioned outside of the X-ray path.
Disinfection instructions
WARNING
• Confirm that the respiratory density of disinfectant including solvent is under legal regulation.
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
• Do not use any of the disinfectants described below as damage to the equipment may result.
Quality, performance and safety of the equipment cannot be assured.
• Chloric disinfectant which is strongly corrosive to metals and rubber parts.
• Disinfectant whose uses on metals, plastics, and coating are forbidden according to the
instructions supplied with the disinfectant.
• Formalin gas and disinfectant sprays that may get inside the equipment.
Disinfectant ethanol is recommended for disinfection.
Carefully read the instructions and cautions supplied with the disinfectant before use.
Disinfectant ethanol is ethanol of 76.9 to 81.4 vol% without other additives.
Cautions on Network
CAUTIONS
After connecting this system to the network with other systems, confirm that the other systems
are not affected. If they are affected, take countermeasures such as network separation.
Connection instructions
WARNING
Make sure that the devices to be connected to the equipment are authorized for connection.
Classification
1) According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
2) According to the degree of protection against electrical shock
Type B applied part
3) According to the degree of protection against harmful ingress of water
Main Unit : IPX0
Foot switch (optional) : IPX1
Remote exposure foot switch (optional) : IPX8
4) According to the degree of safety of application in the presence of a flammable anesthetics mixture
with air or with oxygen or nitrous oxide.
Equipment not suitable for use in the presence of a flammable anesthetics mixture with air or with
oxygen or nitrous oxide.
5) According to the mode of operation
Continuous operation with intermittent loading
If the problem cannot be solved with the above measures, stop using this equipment and consult
the manufacturer, our official dealer or FUJIFILM Representative for help.
WARNING
● Do not place devices generating electromagnetic wave near this equipment.
● If a device(s) other than those specified is connected, predetermined EMC performance
cannot be guaranteed.
● Do not use mobile RF communications equipment within 30cm (11.8 in.) of this equipment.
● The product may be interfered with by other equipment, even if that other equipment
complies with CISPR EMISSION requirements.
● Do not use this equipment near devices that generate strong electromagnetic waves, such
as thermotherapy devices or HF surgical equipment.
● If this equipment is used near commercially available electronic devices, such as mobile
phones, laptop computers or home appliances, which generate electromagnetic waves, this
equipment may malfunction due to electromagnetic interference.
The FDR MS-3500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FDR MS-3500 should assure that it is used in such an environment.
The FDR MS-3500 uses RF energy only for their internal function.
RF emissions
CISPR 11
Group 1 Therefore, their RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
1
RF emissions
Class A
CISPR 11
Table 2
Guidance and manufacturer’s declaration - electromagnetic immunity
The FDR MS-3500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FDR MS-3500 should assure that it is used in such an environment.
The FDR MS-3500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FDR MS-3500 should assure that it is used in such an environment.
IEC 60601-1-2
Immunity test Compliance level Electromagnetic environment - guidance
test level
1 Conducted RF
IEC61000-4-6
3V
150 kHz to 80 MHz
3V Portable and mobile RF communications equipment
should be used no closer to any part of the FDR
MS-3500, including cables, than the recommended
I S M F r e q e n c y b a n d * See Table 5. separation distance calculated from the equation
Amateur radio band applicable to the frequency of the transmitter.
For Safe Operation
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the FDR MS-3500 is used exceeds the applicable
RF compliance, the FDR MS-3500 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the FDR MS-3500.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
The FDR MS-3500 is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled.
The customer or the user of the FDR MS-3500 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the FDR MS-3500
as recommended below, according to the maximum output power of the communications equipment.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
1 26.957 6 40.66 6
Amateur radio band compliance level
MHz V MHz V
1.8 6 3.5 6
5.3 6 7 6
10.1 6 14 6
18.07 6 21 6
24.89 6 28 6
50 6
Proximity fields from RF compliance level
Preventing accidents
• Keep the exposure table, compression plate or other parts touched by patients or technologists clean. If
the parts are soiled with the blood, etc. of an injured patient, infection may result. Disinfect the soiled
parts immediately with a disinfectant or other cleaning solution.
• Before making an X-ray exposure, make sure that there is no sign of abnormality on the exposure table.
A flaked or scratched surface could injure the patient.
• Before using the compression plate, make sure that it is not worn out. If the plate has any crack, replace
it with the new one. Otherwise, the plate may break while it is used and the sharp edges could injure the
patient.
• Before adjusting the compression force on breasts, make sure that no part or all of the patient’s body is
at the risk of hitting the equipment or being caught by the equipment.
• Pay attention to the patient’s behavior to prevent the patient from hanging onto or leaning against the
swivel arm. Otherwise, the patient may be injured or the arm may be deformed.
• Before adjusting the angle or height of the swivel arm, make sure that a part or all of the patient’s body is
not at the risk of hitting the arm or being caught by the part.
• Before rotating the swivel arm, make sure that the patient is away from the exposure stand.
• Check the conditions of the patient (whether she has an implanted pacemaker/defibrillator or breast
implants, etc.) before executing a study.
• When compression of the breast is completed, the foot switch should be moved away from the patient
and technologist. If the foot switch is accidentally operated, it may endanger the patient or technologist.
• During an X-ray exposure, a technologist or doctor should be where the emergency stop button is within
their reach.
• Install the magnification table horizontally. If slanted, the stainless stay will fall down under its own weight
• When installing or removing the magnification table, tightly hold the bottom part with both hands. If not,
the table might fall and cause an injury.
• When cleaning the magnification table, make sure that it is correctly installed. Do not apply excessive
force to the table, otherwise it may break and cause an injury.
• Do not lean against or sit on the operation desk. The operation desk may break or topple over, possibly
causing an injury.
• Do not apply a strong impact to the protective shield on the operation desk.
As the protective shield is made of X-ray-proof glass, it may be cracked if a strong impact is applied. If
cracked, contact our official dealer to have a protective shield replaced with a new one. Though highly
safe laminated glass or shatterproof film is adopted, using a cracked shield is highly dangerous.
• When making an exposure of a patient in a wheelchair, the swivel arm or exposure table may hit the
patient’s foot and cause injury at the time of rotating or moving up/down the swivel arm. Be careful of
themoving parts.
• When installing the under-armrest (optional) and conducting an exposure of a patient in a wheelchair,
there is the risk that the under-armrest may hit the patient’s foot and cause injury when the swivel arm is
rotating or moving up or down. Take care of moving parts.
Preventing malfunctions
• Do not use inappropriate detergent or disinfectant. Otherwise, it may cause cracks on the surface or
inside of the covers. Such detergent or disinfectant penetrates into the equipment and could cause fire.
• Use a disinfectant that will cause no damage to the equipment or harm to the human body when
disinfecting the equipment.
Other limitations
• Minimum necessary interval between successive exposures
When making X-ray exposures one after another, allow a longer interval between exposures than that
indicated in the graph “Minimum exposure interval for energy input to X-ray tube”. Do not make
exposures more than 60 times in an hour. Otherwise, X-ray exposure may be disabled for a long time
due to heat generated from the X-ray tube.
See the graph “Minimum exposure interval for energy input to X-ray system (tube)” in “A.1
Specifications”.
When operating parts of the equipment successively, observe the following operation time limit and
operation interval for safety.
Part Operation time limit Operation interval
Swivel arm up/down movement 16 seconds 60 seconds or more
Swivel arm rotation 24 seconds 23 seconds or more
Compression plate *1 30 seconds 120 seconds or more
Grid mounting/dismounting *2 90 seconds or more
*1: Successive compression plate operation without making X-ray exposures
*2: The operation interval is the time period that should be left between a grid dismounting and the next mounting.
the system shutdown procedure, and wait until the equipment is automatically powered off.
• When the monitor arm is not in use, fix the monitor to prevent it from falling off the operation table.
Preventing malfunctions
• Do not install any software or hardware in addition to the preinstalled one. Do not uninstall the
preinstalled software or hardware.
The AWS is preinstalled with the appropriate software or hardware. If other software or hardware is
added, various operational errors may result.
• Before operating the AWS, make sure that the monitor has appropriate brightness (light intensity). If not,
contact our official dealer. Regularly adjust the monitor settings to keep optimum display of images read
by the AWS. Apart from regular adjustments, do not change the settings.
If a fluorescent tube for the monitor backlight is used for over a prolonged period, brightness (light
intensity) of the monitor decreases because the lifetime of the fluorescent tube is running out.
In this case, use of the monitor adjustment tool or service by our official dealer is recommended.
Managing data
• When disposing of the AWS, completely delete all personal information.
3 3
1
2
5. Type B applied
part label
Do not touch the biopsy positioner
connector. Touching it may cause a
system failure.
1
2. Foot Switch Label
(Right-hand Side)
For Safe Operation
XXXXXXX
or
n Generator
1. High voltage warning label
2
3 2. System Label 3. Rating Label
1 401Y120024
1 401Y120025
1
401Y100124
2
1
401Y100121
2
1
401Y100131
1 401Y120026
2
1
401Y120005
1
401Y120006
1 401Y100122
For Safe Operation
1 401Y100130
2
401Y120033
1
2
1
401Y120038
1 401Y120046
401Y200001
1
2
1
401Y100118
1 401Y100115
401Y100116
401Y100119
401Y120009
1
1
401Y120011
1 401Y100114
1 1
When changing the position of the monitor, be careful
so that your fingers are not caught between the
monitors or between the monitor and the monitor arm.
or
n Control pad
1 1. Warning Label 2. Stop Label
4
2 3. Caution Label (for Connection) 4. Successive Operation Condition Label
1.8倍拡大撮影台S 357Y100432
XXXXXXX
4. Type B applied part label
2. Magnification Table
1 23 4
Name Plate
: Type B applied part
XXXXXXXX
リモート撮影スイッチ 128Y100144
1 XXXXXXX
遮蔽板キット 898Y101477
1
XXXXXXX
1
For Safe Operation
Identification Label
Identification Label
• The month and year of actual manufacture are printed on the label.
• Store the CD case with care.
3 Conduct exposures
Change the compression plate as needed.
F
or details, see “3.2 Changing the Compression Plate”.
Adjust the height and angle of the swivel arm to suit the patient. Use Single-Touch if required.
Position the patient.
Apply compression if required. Check the exposure field.
Set the exposure mode with AWS.
F
or details, see “A.3 Automatic Exposure Control (AEC)”.
Press the exposure button to conduct an exposure. Release the compression plate.
Optimize the image with AWS.
Product Overview
the breast thickness and compression force that have been automatically measured by the system.
• The automatic exposure control (AEC) enables to accurately detect the mammary region and
automatically controls the mAs value and others.
• This equipment is provided with the two-step automatic decompression function of compression plate.
When this function is enabled, compression of the breast is automatically released at the end of exposure
to shorten patient’s pain.
• A compression plate automatic release function is included. When this function is enabled, compression
of the breast is automatically released at the end of exposure to shorten patient’s pain.
• This equipment is provided with the foot switch (optional) so that raising/lowering operations of the swivel
arm and compression/decompression operations of compression plate are possible even when both
hands are used to support a patient.
• This equipment is provided with the armrests as well as optional chest wall pad and axilla pads. These
items help relieve discomfort of the patient when she is positioned for exposure.
• The AWS includes the X-ray control panel, which allows a doctor or technologist to set and check the
exposure conditions in the same window.
• The automatic positioning function allows the swivel arm to automatically rotate to and stop at a desired
angle with a single touch of a button.
• The exposure stand is provided with the patient information display function, which displays patient ID,
patient name, and date of birth registered on the AWS. With this function, a doctor or technologist can
have accurate information on the patient before positioning.
• Using the optional tomosynthesis exposure software allows exposure images to be acquired from several
angles, and a 3D slice image to be re-organized from those images.
• When the optional tomosynthesis exposure software and the optional tomosynthesis biopsysoftware are
used together, targeting is possible for positioning on slice images. This enables the user to perform
highly accurate studies easily.
• With the use of the biopsy positioner, the three-dimensional coordinates of an object are calculated
based on the image obtained from a stereotactic exposure in order to support positioning of the needle.
• CEDM (Optional) can capture images consecutively under two different tube voltage conditions during
one compression, and then create and display a subtraction image of the two acquired images. A
subtraction image of the acquired images is created from a low tube-voltage (low energy) exposure taken
under conditions similar to standard mammography exposure and a high tube-voltage (high energy)
exposure that applies a Cu filter.
Control pad
• The FDR MS-3500 consists of the exposure unit and the AWS. Peripheral devices such as a server and an
image viewer are not included.
• The FDR-3500DRLH consists of the exposure stand, control cabinet, flat panel sensor, control pad and
generator.
• The AWS consists of a personal computer, monitor (optional), keyboard, mouse and hub.
ee “Workstation FDR-3000AWS Operation Manual” for optional items for the AWS.
S
See “Biopsy Connection FDR-2000BPY Operation Manual”, “FDR MS-3500 Option Tomosynthesis
Exposure Software Operation Manual” and “FDR MS-3500 Option Tomosynthesis BiopsySoftware
Operation Manual” for details on the biopsy positioner, tomosynthesis exposure software and
tomosynthesis biopsysoftware
WARNING
• The exposure unit can be used in an environment where a patient comes in contact with the equipment.
• When installing the AWS in the study room, the technologist should be where the protective shield
(optional) covers and protects all of his/her body from exposure to X-rays.
• When installing the AWS in a study room, use the protective shield (optional) to protect the technologist
from exposure to X-rays.
CAUTIONS
Handle and keep all the Operation Manuals for each component of FDR MS-3500 as a whole.
Standard attachment
n Exposure Unit 2
• Face guard • Axillary pad R
• 24×30 compression plate (High) (401Y120004) *2 • Fuse
Product Overview
• X-ray power radiation cable • Operation manual, supplied documentation
• Chest wall pad *1 • CD-ROM (Reference Guide)
• Axillary pad L • Stand anchor
*1 Replacement parts for a chest wall pad and axilla pads are optional.
*2 Accessories: Extra facilities to the FDR MS-3500 which easily can be mounted by the user.
Optional items
n Exposure Unit
• 24×30 compression plate (Low) (401Y120024)*2 • Remote exposure foot switch
• 24×30 compression plate (Shift) (401Y120025)*2 • Protective shield
• 24×30 compression plate (Shift Small) (401Y100124)*2 • Protective shield cover
• 24×30H compression plate (Flex) (401Y100122)*2 • Monitor arm
• 24×30H compression plate (FS) (401Y100130)*2 • Chest wall pad *1
• 18×24 compression plate (Low) (401Y120026)*2 • Axillary pad L *1
• 18×24 compression plate (High) (401Y120005)*2 • Axillary pad R *1
• 18×24 compression plate (Small) (401Y120006)*2 • Operation desk
• 18×24L compression plate (Flex) (401Y100121)*2 • Biopsy positioner (FDR-2000BPY) *2
• 18×24L compression plate (FS) (401Y100131)*2 • Lateral approach kit (898Y101490)*2
• 18×24 compression plate (FS Small) (401Y120033)*2 • Stereo up-grade kit
• 18×24 compression plate (Shift P) (401Y120038)*2 • Tomosynthesis exposure software
• 18×24 compression plate (Shift P Small) (401Y120046)*2 • Tomosynthesis exposure software Excellent
• 18×24 compression plate (FS Shift P) (401Y200001)*2 • Tomosynthesis biopsysoftware
• Rectangle spot compression plate (401Y100118)*2 • Shielding Plate
• Compression plate for axilla radiography (401Y100115)*2 • Under armlest
• Compression plate for magnifications (401Y100116)*2 • Face guard T (898Y101468)*2
• Rectangle spot magnification compression plate (401Y100119)*2 • Face guard T Comfort (898Y200541)*2
• Round spot compression plate (401Y120009)*2 • Foot switch array modification kit Frip Type
• Round spot magnification compression plate (401Y120011)*2 • Foot switch array modification kit Cross Type
• 2D biopsy compression plate (401Y100114)*2 • Compression plate storage holder
• Foot switch • 1 Shot Phantom M Plus 24x30
• Magnification table (898Y101487)*2 • CEDM Software
• Remote exposure hand switch • Comfort Comp
*1 For replacement.
*2 Accessories: Extra facilities to the FDR MS-3500 which easily can be mounted by the user.
n AWS
Refer to “FUJIFILM Mammography Workstation FDR-3000AWS Operation Manual”.
Separately-sold items
n Exposure Unit
• Biopsy positioner (FDR-2000BPY) • Anchor plate • Decoration seal
• Lateral approach kit
Contact a designated contractor to purchase optional items and separately-sold items.
Compression force
progress bar indicates
the image reading status
Column after the exposure. The
Compressed breast
The swivel arm moves up or time remaining until
thickness
down his column.
Product Overview
reading completion is
Intermediate compression force also displayed on the
progress bar.
Progress bar
CAUTIONS
• The 24 x 30H compression plate (Flex), 18 x 24L compression plate (Flex), 24 x 30H
compression plate (FS), 18 x 24L compression plate (FS), 18 x 24 compression plate (FS
Small) and 18x24 compression plate (FS Shift P) are designed to properly distribute the
compression force over the breast.
With this feature, the compression plate tilts according to the consistency and/or size of the
breast. The compressed breast thickness is properly adjusted by the compression force of
approximately 80N or more along the tilt of the compression plate.
The thickness of the thickest part on the chest wall side of the compressed breast is
displayed on the exposure stand display panel.
• Patient information is displayed on the panel according to the character size setting of the
exposure stand. For setting, contact our official dealer.
• T he compression pressure configurable maximum is 200N. Even after the maximum
compression pressure is set, additional compression of about 200 to 230N is possible with
the manual compression dial.
Left-hand side Right-hand side 1 Swivel arm up foot Step on this switch to raise the
switch swivel arm.
3 6 6 3
2 Swivel arm down Step on this switch to lower the
foot switch swivel arm.
Product Overview
Use this to fine-adjust
the compression force.
Compression force
Rotation angle display Compression plate
display Compressed breast This compresses breast
thickness display during exposure.
Bottom LCD
switching button
Manual Flat panel sensor
Light field on/off
compression Flat panel detector which
button Exposure table
dial converts X-ray images
Press this to turn on
or off the light field Set breast on this table directly into electric signals.
lamp and switch the during exposure.
exposure field size.
Swivel arm
counterclock- Swivel arm
wise rotation clockwise
button Swivel arm up/down rotation button
Press this to rotate the button Press this to
swivel arm Press this to raise/lower rotate the swivel
counterclockwise (-). the swivel arm. arm clockwise (+).
CAUTIONS
The swivel arm cannot be raised, lowered, or rotated when the compression force is more than
30N. In this case, unless the compression force is set to 0N, the swivel arm operations cannot
be performed.
Breaker ON
Breaker OFF
Circuit breaker
CAUTIONS
Do not turn off the power to the control cabinet.
Decompression button
2
Press this to release the compression plate
when it is in the compressing position.
n Operation desk
Product Overview
Operation desk
Place the AWS on this.
Monitor (Optional)
Mouse
Hub
Monitor Keyboard 2
See “Workstation FDR-3000AWS Operation Manual” for optional items for the AWS.
Product Overview
n X-ray Control Panel
This panel displays the status of the exposure stand connected to the AWS and allows the various
settings to be made. During a study, the actual radiation parameter information is displayed in real time
on the X-ray control panel in the “Study window”.
See “Workstation FDR-3000AWS Operation Manual” for optional items for the AWS.
(4) Exposure unit setting button
(5) Flat panel detector (6) Exposure stand status (7) Exposure (9) Time display
(FPD) status display display message display
(8) Exposure status indicator
After the exposure stand power is turned on, the buzzer sounds and the exposure button turns on
until system startup is complete, however there is no X-ray exposure in this state.
• When it passes 1:00 a.m. without shutting down the system, calibration starts automatically.
• If the system is restarted within 30 minutes after shutdown, an exposure can be made after
approximately 3 minutes, as automatic calibration is skipped.
• In an emergency, X-ray exposure can be made without performing daily calibration. Select “Abort”
from the calibration menu, while “Initializing” is displayed in the flat panel detector (FPD) status
display of the AWS. However, daily calibration must be performed every day to keep the flat panel
sensor in optimal conditions. If daily calibration is cancelled or aborted, artifacts may appear in the
image.
Operations
Conduct exposures
Press the exposure switch to conduct an exposure. Release the compression plate.
Optimize the image with AWS.
CAUTIONS
• D uring exposure, the buzzer sounds and the exposure switch lights yellow.
Exposure ends automatically.
• If the buzzer sound or the yellow illumination of the exposure switch continues for
more than 15 seconds, release your hand from the exposure switch.The buzzer
sound and the yellow illumination of the exposure switch stop.
• If the buzzer sound or the yellow illumination of the exposure switch does not stop
even after releasing the exposure switch, press the emergency stop button.
Exposure is forcibly stopped.
Select . The application software of the AWS ends, and the power of the personal computer
is automatically turned off.
Select to cancel the operation.
• Before shutting down the system, make sure that no compression is still being applied.
• When it passes 0:00 a.m. with the exposure stand turned on, calibration automatically starts at 1:00
a.m. Since the system cannot be operated during calibration, wait until it finishes before starting
operation.
• As the AWS is not supposed to operate continuously all day long, turn the personal computer off
once a day by performing shutting down operation.
Do not press the power switch when turning off the personal computer.
• To deactivate AWS, select .
3
Operations
3
Unlock lever (1)
Operations
When replacing the compression plate,
raise it to 100 mm (3.93 in.) or more above
the exposure table.
3 CAUTIONS
• Be careful not to pinch your fingers when attaching and removing the compression
plate.
Operations
• Make sure that the compression plate is securely attached. If it is not correctly
locked, it could drop off and cause injury to the patient or technologist.
• Hold the compression plate to prevent it from falling when removing it with the
swivel arm rotated.
• When using the compression plate for rectangle spot magnification, round spot
magnification or magnification, be sure to use the magnification table (optional).
Otherwise, radiation may not be delivered properly.
• When the compression plate is correctly attached, the collimator automatically adjusts itself to
produce the radiation field of the plate and the light field lamp lights on the exposure table
approximately 20 seconds.
• The compressed breast thickness display may change when the compression plate is attached;
however, the height between the exposure table and the compression table is accurate.
• If using the optional magnification table, attach it in this step.
See “O.1 Using the Magnification Table” for details.
(1)
Operations
3.3.2 Exposure Unit Emergency Stop Operation
1 Press one of the emergency stop buttons (1).
The exposure stand operation (swivel arm movement, compression plate operation or X-ray
radiation) stops.
CAUTIONS
• If the system is not restored properly, shut it down, and then restart it after more
than 30 seconds have elapsed.
S
ee “3.1.6 Shutting Down the Diagnosis System” for how to shut down the system.
See “3.1.2 Starting up the Diagnosis System” for how to start up the system.
• Before restoring the system, make sure that the patient is away from the exposure
stand.
3
Operations
Sleep Mode settings can be disabled in the service settings, however leaving it enabled is
recommended, as it reduces the load on the flat panel sensor.
If a message is displayed recommending that you contact a designated contractor for the solution,
or if you conduct the steps in the displayed solution and the problem is not resolved, contact a
designated contractor.
Contact a designated contractor if the Exposure Unit does not start up, or if there is a problem that
does not have a solution displayed in the AWS screen.
Troubleshooting
This section displays several messages that are often displayed by the system. Conducting the steps in
the displayed solutions resolves the majority of problems.
See “Reference Guide 4.2 When an Error Message Appears” for details of other messages.
Message/Cause Remedy
The door is open. Close the door before exposure.
The position of the tube is incorrect. Return the tube back to 0° before exposure.
The compression plate has not been set correctly. Reinstall the compression plate.
Images may not be obtained correctly due to a high temperature. Lower the temperature of the room.
A calibration error has occurred. Conduct the daily calibration again.
5 CAUTIONS
If plates with a prickly feel or cracks are used, they may
break during compression and cause serious injury to
patients. Make sure to replace them.
Daily Inspection and Maintenance
During the startup of the exposure unit, the exposure switch briefly
turns on yellow.
1 Put an ACR phantom or 40-mm acrylic board on the exposure table. Attach the
compression board to the exposure stand, and set the height of compression to 45 mm.
2 Select [Auto] (auto mode) by using the (Set Exposure Mode button).
i-Choose AEC (provisional) OFF.
3 Make an exposure.
The exposure switch lights for a while (approximately one second).
X-ray radiation finishes and the image is output.
4 Remove the phantom, and enclose the exposure table with the lead cover.
A Pb apron, etc. can be used.
7 Follow the message and select [OK] to start the restore processing. 5
When the restore processing finishes, the AEC normally operates.
Make sure to protect the exposure table by using any Pb protector to prevent the FPD from
deterioration due to a test exposure.
2 Set the exposure menu items. Enter any information for making a test exposure.
See “Reference Guide 3.2.4 Selecting Exposure Menu Items” for details.
Cleaning
Before studies, make sure to clean parts that touch the patient. Cleaning is necessary for all the parts
other than those that only require disinfection.
For cleaning, use a cloth slightly moistened with neutral detergent diluted with cold water or lukewarm
water and tightly wrung out. Use a lint-free non-woven or 100 % cotton cloth.
Disinfectant ethanol is ethanol of 70 to 75 wt% or 76.9 to 81.4 vol% without other additives.
Clean according to the following procedure.
1 Wipe all patient contact surfaces with a lint-free non-woven or 100 % cotton cloth
dampened with commercially available disinfectant ethanol (wet but not dripping) for a
minimum of 5 minutes and until visibly clean. Replace soiled wipes as needed.
2 Use a small sterile swab for hard to clean locations (i.e. needle guard lumen) for a
minimum of 30 seconds. Replace soiled swabs as needed.
CAUTIONS
• Make sure to use a recommended detergent to clean the equipment. Improper detergent may
damage the equipment.
• Make sure not to let detergent or disinfectant ethanol get inside the equipment during
cleaning.
• Clean the magnification plate under stable conditions. In addition, do not apply load to it while
cleaning. The magnification may be broken, causing an injury.
Disinfection
Disinfectant ethanol is recommended for disinfection.
Carefully read the instructions and cautions supplied with the disinfectant before use.
Disinfectant ethanol is ethanol of 70 to 75 wt% or 76.9 to 81.4 vol% without other additives.
Using lint-free non-wovens or 100 % cotton cloths dampened with commercially available disinfectant
ethanol (wet but not dripping), wipe all surfaces of the device paying close attention to seams and
recessed areas. Maintain contact of the disinfectant on all surfaces of the device per the manufacturer’s
recommended minimum contact time (i.e. a minimum of 10 minutes).
• Compression plate • Exposure table/Magnification table • Face guard
• Face Guard T • Face Guard T Comfort
WARNING
Confirm that the respiratory density of disinfectant is under legal regulation.
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
n Personal computer
• Checking for normal startup and shutdown of the AWS as well as the establishment of proper
communication with the connected devices
• Checking the fan for its operation
Check if strange noise or odor is not perceived.
• Checking the built-in clock for the time and adjustment
The AWS identifies and manages images based on time information. Make sure that the time of
the clock is always set correctly.
• Cleaning the color LCD monitor 5
• Cleaning the keyboard and mouse
n Hub
CAUTIONS
If the swivel arm still operates, turn off the breaker that supplies power to the
exposure stand and contact our official dealer.
CAUTIONS
Daily Inspection and Maintenance
If the compression force changed more than ± 20N, turn off the exposure stand and
contact our official dealer.
*1 : Equal to 60,000 times. However, this number is based on FUJIFILM’s evaluation conditions.
*2 : Equal to 900 days if exposures are made 200 times per day (four exposures (in two directions for each breast) for 50 patients).
*3 : Equal to 1,800 days if exposures are made 100 times per day (four exposures (in two directions for each breast) for 25 patients).
The cycles of periodical maintenance and inspection and of parts replacement differ depending on
the usage and the daily operation time.
For details, contact us directly or our official dealer.
“FDR Mammography QC Program” (optional) is available for you to easily perform quality control of
multiple items, especially those related to image quality.
Contact our official dealer for details.
Percentage of
Maximum (60A~63A)
Reference ambient temperature (40°C)
Maximum (40A~50A) Reference ambient temperature (45°C)
Ambient temperature (°C)
Operation time
Minimum
A Maximum
Main Specifications
Attenuation Equivalent
• Collimator : At least 0.4 mmAl (to the conditions of 49kV radiation)
• Exposure table : 0.1 mmAl (to the conditions of 30kV radiation)
• Magnification table : 0.21 mmAl (to the conditions of 30kV radiation)
• Biopsy protection cover : 0.18 mmAl (to the conditions of 30kV radiation)
Operating Mode
Continuous operation with intermittent loading
Continuous Operation
60 shots/hour (28kV, W/Rh, 71mA, 4 consecutive hours)
40 shots/hour (28kV, W/Rh, 71mA, 8 consecutive hours)
However, maintain the interval shown in the Minimum exposure interval for energy
input to X-ray tube graph at least.
SID (Focal Spot to Image Receptor Distance)
65 cm (25.59 in.)
Characteristics of X-ray Exposure
Tube voltage range 22kV to 49kV (at 1kV increments)
Tube voltage accuracy ±1kV
mAs setting range Large focal spot: 2mAs to 600mAs
Tomosynthesis:
(HR mode) 16mAs to 125mAs
(ST mode) 16mAs to 90mAs
Small focal spot: 2mAs to 300mAs
(Depending on load conditions such as tube voltage)
mAs setting accuracy ±10%+0.2mAs
Tube current Large focal spot: 20mA to 200mA
Small focal spot: 20mA to 61mA
Nominal maximum output power 7kW (35kV x 200mA) * Load time: 40ms
Nominal maximum output power (at 30kV, 40ms) 5kW (168mA)
Maximum output power for making exposures at 2.8kW (28kV x 100mA)
1-minute intervals Load time 1s
Total allowable mAs per hour 7,560mAs (35kV), 5,400mAs (49kV)
Anode material 10%Re/Tungsten face on molly alloy
Large focal spot: 0.3
X-ray tube focal spot size Small focal spot: 0.1
* Per IEC 60336, determined with the slit camera method.
Specific filter 0.63mm Be
Anode and filter combination W/50μmRh, W/700μmAl, W/(250μmCu+700μmAl) A
Main Specifications
■ : 2D imaging
● : Tomosynthesis imaging
Time [min]
Condition: 25°C, large focal spot
A.1.2 AWS
Refer to “FUJIFILM Mammography Workstation FDR-3000AWS Operation Manual”
(Flex)
18 cm (7.08 in.)
18 cm (7.08 in.)
18 cm (7.08 in.)
24 × 30H compression plate (FS)
(Hold the light exposure field ON/
OFF button to switch the radiation
(0.19 in.)
(0.19 in.)
(0.19 in.)
0.5 cm
0.5 cm
0.5 cm
Chest wall edge Chest wall edge
field sizes between 18 × 24 cm Chest wall edge
24 cm (9.44 in.)
(0.19 in.)
0.5 cm
24 × 30 compression plate (Shift) 36.5˚ to 146.5˚ ±36.4˚ or less/±146.6˚ or more -36.5˚ to -146.5˚
24 × 30 compression plate (Shift 24 cm (9.44 in.) 24 cm (9.44 in.) 24 cm (9.44 in.)
Small)
18 cm (7.08 in.)
18 cm (7.08 in.)
(0.19 in.)
(0.19 in.)
0.5 cm
0.5 cm
0.5 cm
18 × 24 compression plate (Shift Compression plate position: Left Compression plate position: Center Compression plate position:
P) Right
18 24 compression plate (Shift P
× 24 cm (9.44 in.) 24 cm (9.44 in.) 24 cm (9.44 in.)
Small)
18 cm (7.08 in.)
18 cm (7.08 in.)
18 cm (7.08 in.)
(0.19 in.)
(0.19 in.)
compression plate.)
0.5 cm
0.5 cm
0.5 cm
A
18 × 24 compression plate (High)
18 × 24 compression plate (Low) 24 cm (9.44 in.)
18 × 24 compression plate
(Small)
0.5 cm 18 cm (7.08 in.)
(Flex)
(0.19 In.)
18 cm (7.08 in.)
9 cm
field sizes between 9.2 × 9 cm
(3.54 in.)
and 18 × 24 cm.)
(0.19 in.)
0.5 cm
Chest wall edge Chest wall edge
(7.24 in.)
18.4 cm
(0.19 in.)
0.5 cm
Chest wall edge
18 cm (7.08 in.)
field sizes between 14.5 × 14.5
14.5 cm
(5.7in.)
cm and 18 × 24 cm.)
(0.19 in.)
(0.19 in.)
0.5 cm
0.5 cm
Chest wall edge Chest wall edge
Round spot compression plate 8.1 × 11.6 cm radiation field 18 × 24 cm radiation field
(Hold the light exposure field ON/ 24 cm (9.44 in.)
OFF button to switch the radiation
field sizes between 8.1 × 11.6 cm
(0.19 In.)
Chest wall edge Chest wall edge
(9.3 in.)
and 18 × 24 cm.)
(0.19 in.)
(0.19 In.)
0.5 cm
A
Chest wall edge
Main Specifications
A
Main Specifications
A.3.1 Coverage
■ Description
AEC operates in i-AEC mode or regular mode (i-AEC OFF).
In i-AEC mode, the x-ray exposure image acquired by preliminary exposure is analyzed to
implement optimum dose control.
i-AEC
[ON] : Used to take images of the breast. Dose control is performed through image analysis.
[OFF] : Used to take images of the phantom and other matters than the breast.
■ Subjects
[AUTO] : Automatically controls the target/filter, X-ray tube voltage and mAs.
[Semi] : Automatically controls the mAs.
Manually select the target/filter and X-ray tube voltage.
[Manu] : The automatic control is not executed.
Manually select the above parameters.
■ Controlling conditions
In [Auto] mode, exposure is automatically controlled as follows.
Tube voltage (kV)
Compressed breast
Target/Filter Preliminary
thickness (mm (in.)) L-mode N-mode H-mode
exposure
6 (0.23) ~ 10 (0.39) W/Rh 24 26 26 26
11 (0.43) ~ 15 (0.59) W/Rh 25 26 26 26
16 (0.62) ~ 20 (0.78) W/Rh 26 26 26 26
21 (0.82) ~ 25 (0.98) W/Rh 27 26 26 26
26 (1.02) ~ 30 (1.18) W/Rh 28 26 26 26
31 (1.22) ~ 35 (1.37) W/Rh 28 27 27 27
36 (1.41) ~ 40 (1.57) W/Rh 28 28 28 28
41 (1.61) ~ 45 (1.77) W/Rh 28 28 28 28
46 (1.81) ~ 50 (1.96) W/Rh 28 28 28 28
51 (2.00) ~ 55 (2.16) W/Rh 29 29 29 29
56 (2.20) ~ 60 (2.36) W/Rh 30 30 30 30
61 (2.40) ~ 65 (2.55) W/Rh 30 30 30 30
66 (2.59) ~ 70 (2.75) W/Rh 31 30 30 30
A 71 (2.79) ~ 75 (2.95)
76 (2.99) ~ 80 (3.15)
W/Rh
W/Rh
31
32
31
32
31
32
31
32
81 (3.18) ~ 85 (3.34) W/Rh 32 32 32 32
86 (3.38) ~ 90 (3.57) W/Rh 33 33 33 33
Main Specifications
A
Main Specifications
■ REFERENCE
The followings are the causes of variations in AGD value and the variation range.
- Different dosimeter and PMMA thickness were used. Variation range: ±20% approx
- The measurement was made without a compression plate. Variation range: ±10% approx
- The performance
■ INSTRUCTION
The AGD value for check is the total value of pre-exposure and real exposure.
- Add the value of [Pre exposure average glandular dose (mGy)] shown in the exposure result list
to the AGD value displayed on the AWS screen. (Because, the AGD value only for real exposure
is displayed on the AWS screen.)
(1) Install the 24x30 compression plate (High) to the compression unit.
(2) Place the 20mm-thick acrylic board (for 24x30 size) on the exposure stand.
(3) Set the compression plate to 21mm above the exposure plane.
(4) Set the exposure conditions on the AWS.
Exposure Menu: RCC
AEC: Auto
Dose Level: N-mode
D-Tap: 0 or hide (Display preference of D-Tap can be selected at “IIP ServiceUtility” > [Image
Modality] > [94 Enable D-Tap].)
iAEC: OFF
(5) Press the exposure button on the control pad to make an exposure.
(6) Calculate the AGD value, and check that the calculated AGD value is within the allowable
range.
(6-1) Take a note of the AGD value for the real exposure shown on the AWS screen.
(6-2) Take a note of the AGD value for the pre-exposure shown on the list of exposure
results.
(6-3) Check that the AGD value, obtained by adding the value for the real exposure in Step
(6-1) and the value for the pre-exposure in (6-2), is within the range below.
- Japan (Type A) : 0.99mGy ± 22%
- US (Type B) : 0.52mGy ± 22%
- EU (Type C) : 0.52mGy ± 22%
(7) Refer to the table of the AGD target values, and repeat Steps (2) to (6).
[JAPAN(Type A)]
Breast Thickness PMMA Thickness
L-mode (mGy) N-mode (mGy) H-mode (mGy)
(mm) (mm)
21 20 0.32 0.76 1.08
32 30 0.42 0.76 1.09
45 40 0.55 1.00 1.43
60 50 0.75 1.14 1.73
75 60 1.00 1.44 2.34
A
[US(Type B)/EU(Type C)]
Breast Thickness PMMA Thickness
L-mode (mGy) N-mode (mGy) H-mode (mGy)
(mm) (mm)
Main Specifications
■ REFERENCE
X-ray dose can be calculated with the following formula.
AGD = E • g • c • s
E: Entrance surface dose
g and s: Factor dependent on radiation quality and object’s thickness
c: Factor dependent upon radiation quality, object’s thickness and mammary gland density
For details on g, c, and s values, refer to “European guidelines for quality assurance in breast
cancer screening and diagnosis”.
Example of surface dose and average entrance dose calculated with the formula above
PMMA: 40mm (breast thickness of 45mm)
N-mode
Surface Dose AGD
(mGy) (mGy)
JAPAN (Type A) 3.2 1.00
US (Type B)
2.9 0.92
EU (Type C)
A
Main Specifications
n Mammographic stand
A
Main Specifications
B
Test Report
B
Test Report
H MODE
PMMA thickness Target Filter Tube Voltage [kV] mAs AGD CNR
20mm WRh 26kVp 47 0.7mGy 9.3
30mm WRh 27kVp 69 0.9mGy 8.2
40mm WRh 28kVp 98 1.2mGy 7.2
50mm WRh 30kVp 124 1.7mGy 6.2
60mm WRh 31kVp 160 2.1mGy 5.2
70mm WRh 33kVp 194 2.7mGy 4.4
Table 18-1: AEC Performance – H Mode Exposure Conditions
N MODE
PMMA thickness Target Filter Tube Voltage [kV] mAs AGD CNR
20mm WRh 26kVp 32 0.5mGy 7.9
30mm WRh 27kVp 47 0.6mGy 6.9
40mm WRh 28kVp 68 0.9mGy 6.1
50mm WRh 30kVp 85 1.2mGy 5.1
60mm WRh 31kVp 108 1.4mGy 4.3
70mm WRh 33kVp 133 1.9mGy 3.6
Table 18-2: AEC Performance – N Mode Exposure Conditions
L MODE
PMMA thickness Target Filter Tube Voltage [kV] mAs AGD CNR
20mm WRh 26kVp 18 0.3mGy 6.0
B 30mm
40mm
WRh
WRh
27kVp
28kVp
28
39
0.4mGy
0.5mGy
5.4
4.7
50mm WRh 30kVp 49 0.7mGy 3.9
60mm WRh 31kVp 63 0.8mGy 3.2
Test Report
0
Fibers Specks Masses
H-MODE N-MODE
ACR Phantom
42 mm L-MODE
0
Fibers Specks Masses
B
Figure 18-9: ACR Phantom (4.2cm) Scoring
Test Report
Table 18-4 provides the relevant exposure information for each ACR MAP image included on the
ACR MAP Phantom Images CD in the media sleeve at the beginning of this submission.
PMMA20mm Hmode Nmode Lmode
AGD 0.86 mGy 0.59 mGy 0.33 mGy
Target/Filter W/Rh W/Rh W/Rh
Tube Voltage 26 kV 26 kV 26 kV
25312 25315 25318
25313 25316 25319
File Name 25314 25317 25320
PMMA42mm Hmode Nmode Lmode
AGD 1.28 mGy 0.87 mGy 0.50 mGy
Target/Filter W/Rh W/Rh W/Rh
Tube Voltage 28 kV 28 kV 28 kV
25321 25334 25337
25322 25335 25338
File Name 25323 25336 25339
PMMA60mm Hmode Nmode Lmode
AGD 1.63 mGy 1.14 mGy 0.66 mGy
Target/Filter W/Rh W/Rh W/Rh
Tube Voltage 30 kV 30 kV 30 kV
File Name 25340 25343 25346
25341 25344 25347
25342 25345 25348
Table 18-4: ACR MAP Phantom Images - Exposure Conditions.
B
Test Report
H-MODE
Percent Simulated Breast Thickness (㎝)
Glandular/Adipose
Tissue 2.0 4.2 6.0
30/70 kVp 26 28 29
(BR3070) mAs 38 70.8 89
Anode material/x-ray filtration W/Rh W/Rh W/Rh
Entrance Surface Exposure level (mGy) 1.21 3.03 4.39
Average Glandular Dose (mGy) 0.58 0.95 1.16
50/50 kVp 26 28 30
(BR12) mAs
Anode material/x-ray filtration
41.7
W/Rh
81.3
W/Rh
95
W/Rh
B
Entrance Surface Exposure level (mGy) 1.32 3.48 5.09
Average Glandular Dose (mGy) 0.64 1.08 1.36
Test Report
70/30 kVp 26 28 30
(PMMA) mAs 48.2 103.8 125.7
Anode material/x-ray filtration W/Rh W/Rh W/Rh
Entrance Surface Exposure level (mGy) 1.53 4.43 6.74
Average Glandular Dose (mGy) 0.73 1.37 1.78
Table 18-5: Patient Radiation Dose – H Mode
L-MODE
Percent Simulated Breast Thickness (㎝)
Glandular/Adipose
Tissue 2.0 4.2 6.0
30/70 kVp 26 28 29
(BR3070) mAs 14.7 27.7 35.6
Anode material/x-ray filtration W/Rh W/Rh W/Rh
Entrance Surface Exposure level (mGy) 0.46 1.18 1.75
Average Glandular Dose (mGy) 0.23 0.37 0.50
50/50 kVp 26 28 30
(BR12) mAs 16 31.9 37.9
Anode material/x-ray filtration W/Rh W/Rh W/Rh
Entrance Surface Exposure level (mGy) 0.50 1.36 2.03
Average Glandular Dose (mGy) 0.25 0.45 0.58
70/30 kVp 26 28 30
(PMMA) mAs 18.6 41.1 50.0
Anode material/x-ray filtration W/Rh W/Rh W/Rh
Entrance Surface Exposure level (mGy) 0.59 1.75 2.68
Average Glandular Dose (mGy) 0.29 0.57 0.74
Table 18-7: Patient Radiation Dose – L Mode
B
Test Report
B
Test Report
B These radiologists met the radiologists’ qualifications as required by the guidance document.
An overall assessment was made by the two independent reviewers that image sets exposed by the
Fujifilm AMULET Innovality FFDM system collectively are of sufficiently acceptable quality for
Test Report
B
Test Report
(14.21 in.)
361 mm
650 mm (25.5 in.)
Subject
(1.18 in.)
30 mm
Magnification table
(1)
(2)
(1) (2)
Stay
Hinge lock
(3) (4)
(4)
(3)
Stay
Hinge lock
CAUTIONS
Be careful not to pinch your hand when using the hinge lock and stay.
Slip-proof sheet
Exposure table
Magnification table
(3)
CAUTIONS
• When the magnification table is in use, the face guard comes in contact with the
compression plate. Remove the face guard before use.
• Install the magnification table horizontally. If slanted, the stay will fall down under
its own weight and could cause injury to the patient or technologist.
• Make sure that the magnification table is securely installed. If it is not correctly
locked, it could drop off and cause injury to the patient or technologist.
• Be careful not to pinch your fingers when installing the magnification table.
• Be careful not to pinch your fingers when removing the magnification table.
CAUTIONS
Make sure that the magnification table is securely installed. If it is not correctly locked,
the patient may be subject to an unnecessary X-ray dose.
table.
(1)
CAUTIONS
Do not use tape other than the supplied double-sided tape.
O
Exposure table
(1)
The axilla pads are intended to be used symmetrically.
Affix them on both the right- and left-hand sides.
Use of Optional Items
CAUTIONS
Do not use tape other than the supplied double-sided tape.
1 Insert the modular plug of the remote exposure switch into the modular jack on the
rear of the control pad.
Control box Modular plug
Modular jack
Modular plug
2 Insert the cable of the remote exposure switch into the modular jack on the control pad.
3 After using the remote exposure switch, hang it on the hook attached to the operation table etc.
CAUTIONS
Read “Radiation safety instructions” and “Radiation protection area” in “1.3 Safety”,
before making an X-ray exposure in the study room with the remote exposure switch
and/or remote exposure foot switch.
Shielding Plate O
Use of Optional Items
CAUTIONS
Before conducting exposures in a case where shielding is required, make sure that
the shielding plate is correctly set.
FUJIFILM Digital Mammography System FDR MS-3500 Operation Manual 897N102525X 75
O.5.2 Detaching the Shielding Plate
1 Slide and detach the shielding plate.
Face guard T
O
Use of Optional Items
CAUTIONS
• The Face guard T Comfort cannot be used in combination with the biopsy positioner,
magnification table, compression plate for magnification, rectangle spot magnification
compression plate or round spot magnification compression plate (all of them are optional).
Before attaching the Face guard T Comfort, be sure to check that these optional parts are not
attached.
• The Face guard T Comfort cannot be used simultaneously with the face guard (standard
attachment) or Face Guard T (optional).
Detachable arms
O
Partially locked: The right or left button is Buttons
projected.
Use of Optional Items
Locked Unlocked
CAUTIONS
If the locks are released, the Face guard T Comfort may fall down and make contact
with the compression plate. When attaching or detaching the Face guard T Comfort,
hold the detachable arms with both hands.
O
Use of Optional Items
2 Slide the compression part in the desired direction while the lever is held up (2).
3 Make sure that the radiation field also shifts in the same direction.
4 Make sure that the lever returns to its original position and that the digits on the swivel arm
operation panel display are not blinking.
Lever
(1) (1)
(2) (2)
CAUTIONS
Be careful not to pinch your fingers when sliding the compression part.
O
Use of Optional Items