7D HIFU

USER MANUAL
Contents
1. Instrument overview ............................................................................... 2

2. Principle ..................................................................................................2

3. Technology advantages ........................................................................... 3

4. Technical parameters ...............................................................................4

5. Part of the instrument ..............................................................................5

6. Installation .............................................................................................. 8

7. Interface .................................................................................................. 9

8. Steps of operation ................................................................................. 13

9. Precautions for use and maintenance .................................................... 15

10. User contraindications ........................................................................ 16

11. Maintenance ........................................................................................ 17

1
1. Instrument overview
The company's latest product, HF7 Ultrasonic Scalpel Beauty Apparatus, can achieve the three perfect effects of

firming, removing wrinkles and shaping. The advanced point of the ultrasonic knife is the use of the most

advanced ultrasonic energy, deep penetration, can directly reach the SMAS layer of the skin, make collagen

produce immediate contraction and stimulate the collagen fiber network, thereby enhancing skin elasticity from

the bottom of the skin (self-repair mechanism is safer) ). This is currently the safest and most effective way to treat

aging and sagging facial skin in the beauty industry. It is a cross-age skin-beautifying device with no wounds, no

recovery period, no bleeding, and no need for multiple treatments. The magical wrinkle-removing and lifting

effect of the HF7 Ultrasonic Scalpel Beauty Apparatus has been recognized by authoritative organizations in the

beauty industry and is known as eternal The miracle light of youth!

2. Principle

Ultrasound focusing uses its unique high-energy focused ultrasound to directly reach the SMAS layer,

enhance the SMAS fascia suspension, and comprehensively solve the problem of facial sagging and

relaxation. It accurately locates the energy of the ultrasound on the 4.5mm fascia layer under the skin

and plays a role in the muscles. The membrane muscle grows and stretches to achieve the best effect of

shaping, lifting and tightening. It acts on the 3mm collagen layer under the skin to reorganize and

regenerate the collagen to achieve skin aging problems such as restoring elasticity, whitening, wrinkle

removal, and pore reduction. Acting on the subcutaneous fat layer, high temperature causes fat

degeneration and necrosis after being metabolized by the human body to make the fat layer thin. It is

mainly used to solve the problem of excessive fat layer and to shape the body.

At the same time, because the energy passes through the epidermis, there is no need to worry about

epidermal injury, and the skin can be lifted quickly, the contours are tightened, and wrinkles can be

2
smoothed quickly!

3. Technology advantages

Using the most advanced high-tech technology, 7 working probes for targeted use: 1.5mm probe is
suitable for epidermis, superficial fine lines, skin texture, skin tone, and fine pores; 2.0mm probe is
suitable for eyes and forehead lines, Neck lines, crow's feet and other areas; 3.0mm probe can
directly reach the dermis layer to stimulate the continuous regeneration of collagen; 4.5mm probe
can make heat reach the SMAS fascia layer, fundamentally improve skin aging problems. The
6.0m9.0mm, 13.0mm probe can effectively contour the entire body such as the chest, arms,
abdomen, thighs, underarms, calves, ankles and back. 100% without any damage. At the same time,
the depth of the skin of the treatment head is consistent with the set value to ensure that the
customer is painless and comfortable.
2. It has a thermal effect on dermal collagen and collagen fibers and thermal stimulation on the fat
layer and fascia layer (SMAS) at the same time. The therapeutic effect is far better than that of
Remaggi.
3. The HF7 operation can generally be completed in about 40 minutes, and the skin will feel
obvious tightness after completion.
This feeling may disappear after 2-3 days. This is a normal phenomenon of collagen breakdown and
reorganization.
4. The effect of tightening and shaping can be seen immediately after treatment. One treatment can
last for at least 18-24 months, and negative skin growth can be achieved once a year.
5. The operation is simple and convenient. No consumables are needed, greatly saving treatment
costs.
6. Put on makeup immediately after treatment, which will not affect normal life and work.

3
4. Technical parameters

Operating Voltage 110V-240V 50/60Hz

Maximum power 180W

Energy type High-energy focused ultrasound

Output energy 0.1J-3.0J

Output length 5.0-25mm

Probe distance 1-2mm

product weight 16kg

Product Size 380*380*190mm

Product Probe 1.5mm、2.0mm、3.0mm、4.5mm、6.0mm、9.0mm、13.0mm

4
5. Part of the instrument

1
2

3
4
4

1. Handle
2. LCD screen
3. Emergency stop switch
4. Key switch

5
Power outlet
Handle connected

Status
Indicator

Treatment
head

Button switch

6
Treatment head

Specification Photo The depth and efficacy of skin treatment

Suitable for epidermis, superficial fine lines, skin


1.5mm texture, complexion, fine pores

Suitable for eyes, forehead lines, neck lines, crow's


2.0mm feet and other areas

It can directly reach the dermis and stimulate the


3.0mm
continuous regeneration of collagen

Can make the heat reach the SMAS fascia layer,


4.5mm
fundamentally improve the skin aging problem

6mm

It can carry out effective contour shaping treatment for


9mm the entire body such as the chest, arms, abdomen,
thighs, underarms, calves, ankles and back

13mm

7
6. Installation

For installation of the treatment head, please refer to the installation video file of the
treatment head;

For adding purified water to the treatment head, please refer to the video ;

8
USER MANUAL OF HIFU

7. Interface

Start up

Check whether the connection of the instrument is normal, turn on the power of the
instrument, press the power switch of the host, and the machine will start up and enter the
professional work standby interface automatically (see Figure 1).

A

J
B

C I

Figure 1
D E F G
H

[A: Left-hand handle]: The handle is selected by default when the machine is turned on. The

color of the icon is colored. After the right handle is clicked and selected, this icon will turn

into a gray style as shown in Figure- (K: right handle); Can only be used for treatment after

being selected;

Page 9
USER MANUAL OF HIFU

[B: Handle energy setting 1]: After clicking this icon, the handle energy will be set to the

default value of 0.7J;

[C: Handle energy setting 2]: After clicking this icon, the handle energy will be set to the

default value of 1.0J;

[D: Select handle information]: After clicking this icon, the interface will switch to Figure 2,

the handle information interface;

[E: Handle energy display]: Display the current handle energy, click the down button on the

left to decrease the output energy; click the up button on the right to increase the output

energy;

[F: Equipment operating status]: Display the current operating status of the equipment.

After clicking, the equipment will switch between "STANDBY" and "START". The default

setting is "STANDBY" when it is turned on;

[G: Counting of the number of times of use]: Display the number of treatments of the

current handle;

[H: Number of repetitive treatments]: Display the number of repetitive treatments of the

device in the repetitive treatment mode, range: 1-999; click the down button on the left to

decrease the number of treatments; click the up button on the right to increase the number of

treatments;

[I: Background setting button]: Click this icon to switch to Figure 3, which is used to adjust

the system prompt sound volume;

[J: Handle mode selection button]: Display the current operating mode of the device. After

clicking, the device will switch between "Repeat" and "Single", "Repeat" is the repeat

treatment mode, and "Single" is the single treatment mode;

Page 10
USER MANUAL OF HIFU

[K: Right handle icon]: The handle is not selected by default when the machine is turned on.

The color of the icon is gray. After the right handle is clicked and selected, this icon will

change to a color style as shown in Figure- (A: left handle). The handle can be used for

treatment only after it is selected;

Figure 2
[L:Return]: Click this icon to return to the treatment interface in Figure 1;
[M: Treatment head information]: Display the current treatment head information;
[N: Clear treatment head information]: Click this icon to reset the treatment head count to
0;

Page 11
USER MANUAL OF HIFU

Figure 3
[O: Return icon]: Click this icon to return to the treatment interface in Figure 1;
[P: Treatment head information]: Display the current device's prompt sound volume, click the icon to set
the device's prompt sound volume.

Page 12
USER MANUAL OF HIFU

8. Steps of operation

1. Before using the instrument, disinfect and wipe the probes used with alcohol to dry the
water droplets on the face.

2. Apply the product (photosensitive gel) evenly to the area to be treated.

3. Before the operation of the instrument, replace the required treatment head, and the
energy range from low to high to the range that the experiencer can bear is better.

4. Keep the treatment head flat against the skin during the operation, and the operation
spacing should be closely arranged.

5. During the operation, the experiencer should be constantly asked about the comfort
level and the adjustment of the energy level to ensure that the experiencer accepts the
experience in a pleasant state

6.Preparation before starting;


First: Connect the product to the power cord. Note: Before plugging in, make sure that
the rated voltage of the product is consistent with the input voltage (110V-220V), otherwise
the internal components of the product will be damaged and the whole machine will not
work.
Secondly, turn on the ② key switch and check whether the emergency stop switch is in
the off state; Note: The machine can only work when the key switch is on and the emergency
stop switch is off. The open and closed state of each knob is shown in the figure below.

Warning: When connecting the power cord, make sure that the voltage of the connected
power supply is consistent with the rated voltage of the host to avoid damage and danger.

General description
The handle is the working component of the instrument treatment. According to the treatment
parameters set by the host, ultrasound is generated for treatment. The red button is an

Page 13
USER MANUAL OF HIFU

emergency switch that is an emergency stop switch. In an emergency, press this button to
disconnect the control system and power supply, and then rotate to the right and eject the
button to restore normal.

Starting program

1. Check to make sure that the handle is completely fixed to the device socket. (Note:
The handle must be connected to the designated position, otherwise the device will
operate incorrectly).

2. After completing step 1 above, insert the three-core power plug (power cord) into a
socket that can withstand the corresponding current.

3. Turn the key switch, then turn right to turn on the emergency stop switch to make it in
the released state. At this time, the HF7 instrument starts up and the LCD screen
starts to display the LOGO animation.

4. The device enters the standby interface (Figure 1). The first step is to select the
treatment handle, the second step is to set the probe point spacing, output distance,
and output energy. The third step is to set the treatment mode of the machine. If you

select the "Repeat" mode, You also need to set ;The fourth step is

to click the standby button on the display interface, and the device will switch to the
"START" state.

5. Press the control button on the handle to work, the indicator light flashes, and the
lamp is always on after the treatment. If you pause work during the treatment, press
the control button, the device will stop working, and press the control button again
to continue working.

6. The treatment head can be replaced in the standby state, and it is strictly forbidden to
pull the treatment head directly while working.

Page 14
USER MANUAL OF HIFU

9. Precautions for use and maintenance

1. After every use of the instrument, please wipe it with clean water and disinfect it with

alcohol and keep it in a safe place (to avoid scratches on the black film of the probe).

2. Before using the instrument, you must use a plug with a grounding pin, and make sure

that the power socket of the instrument is truly grounded.

3. Make sure that the voltage of the instrument is compatible. If the local power supply

voltage is unstable, we recommend that users add a regulated power supply with matching

power.

4. For the treatment effect of the instrument and to ensure the normal service life, please use

the designated accessories provided or recommended by the original manufacturer.

5. The instrument should not be placed in a humid place or near a water source, and the

instrument should not be directly exposed to sunlight.

6. Do not put the instrument close to a strong heat source, because this may affect the life

and normal use of the instrument.

8. Avoid bad contact when using it in winter or when the weather is dry. Please apply the gel

evenly on the face (the probe sticks flat).

9. Water, oil and other media shall not penetrate into the instrument, and shall not be

impacted, beaten or dropped to avoid damage to the instrument.

10. After using the instrument, please disinfect the probe with alcohol in time and

disconnect the power supply.

Page 15
USER MANUAL OF HIFU

10. User contraindications

1. Clean the skin of the physical treatment site before using the instrument to prevent dirt
from entering the skin with ultrasound or preventing the penetration of ultrasound. Please
remove all metal objects on your body before treatment, so as to avoid unpredictable
conditions that may affect the curative effect.
2. It is best to use a certain viscosity of the medium, which is conducive to the better
integration of ultrasound with the skin, and prevents reflections caused by voids and better
conduction of sound energy.
3. Each treatment area should not be repeated too much, at most three times, to avoid skin
redness and swelling.
4. The heat of the probe does not represent the output of the ultrasonic power; the liquid or
medicine with low concentration should not be directly penetrated, otherwise it will easily
cause dry skin.
5. When using the instrument, the probe should not be close to the eyes, can not pass the
eyeball, pregnant women and severe heart
patient.
6. Patients who are ill should also use this instrument with caution, unless the doctor's
permission is obtained.
7. Patients with malignant tumors, hemophilia, or severe bleeding.
8. Use with caution for patients with major diseases, skin diseases, and infectious diseases.
9. Heart, brain, neck nerve center, spine, eyeball, bleeding, injection.
10. Patients with hyperthyroidism, asthma, and hypertension.
11. Ultrasonic knife treatment can be performed only 20-30 days after injection of hyaluronic
acid.
12. Do not eat spicy food after ultrasonic knife treatment, and do not wash your face with hot
or cold water within a week. Do not use hot yoga, sauna, facial steaming, or facial massage
(avoid face squeezing) every day for a week to moisturize. Pay attention to sun protection. Do
not use radio frequency or light equipment for treatment within one month.

Page 16
USER MANUAL OF HIFU

11. Maintenance

1. After using the instrument, wipe it clean with alcohol and keep it properly.

2. Before use, you must use a plug with a grounding pin, and ensure that the power socket of

the instrument is truly grounded.

3. The voltage of the instrument is adapted. If the local power supply voltage is unstable, we

recommend that the user add a regulated power supply with matching power.

4. Please use the designated accessories provided or recommended by the original

manufacturer for the treatment effect and the guarantee of the normal service life of the

instrument.

5. Do not place the instrument in a humid place or near a water source, and do not expose

the instrument to a place directly exposed to sunlight.

6. The instrument is close to a strong heat source body, because this may affect the life and

normal use of the instrument.

7. Please remove all metal objects on your body first, so as to avoid unpredictable

conditions that may affect the curative effect.

8. Avoid poor contact, please pay attention to the place where the skin is in contact with a

certain degree of moisture, water-based products

Wipe with product or water can achieve the desired effect.

9. When the instrument is not in use, please turn off the power switch of the instrument, and

make sure to turn off the main power supply of the instrument before use to ensure the

safety of electrical products.

10. Strictly follow the instructions in the manual to use the instrument or train the operator

of the instrument.