Tattoo removal machine user manual - 看图王 BM20
Tattoo removal machine user manual - 看图王 BM20
Tattoo removal machine user manual - 看图王 BM20
1
User Notice
Safety measures
User training
Device electromechanical safety
Page number1-2 2
Product Introduction
Product description
Product structure and properties
3 Device structures
Product technical parameters
Device Installation
Product treatment range
Device installation process
Product parts
Device related components installation
Page number2-5
and water injection instructions
Water injection instructions
Water cycle debugging
Treatment handle installation
Page number5-8
4
Device Software Operating Instructions
Main interface
Page number9-9
Content
5
Other system settings
Device details procedure
Treatment course
Treatment parameters
Care and precautions
6
Equipment maintenance
Page number11-13
Clean the device
Therapy handle cleaning
and replacement
System trouble shooting
Page number13-15
7
Specifications
System specification chart
Page number
15-16
Chapter 1 User Notice
Before use, the operator must read the user manual carefully and check the
device functions are working properly or not to avoid unnecessary damage.
If you need to connect the device to peripheral devices, peripheral devices
must read detailed instructions.
Notice: Only after the distributor trained personnel have the right to repair
the facilities inside the instrument, not entitled to any treatment may cause
damage to the instrument and cause warranty invalid.
1
dangerous high voltages exist in the device,After the power is turned off,high
voltage devices is still possible with residual voltage,So please do not open the
protective cover to avoid high voltage shock hazard. Only authorized personnel
can repair the device, Otherwise it will cause warranty invalid.
Key switch
3
The laser head of the 1064nm wavelength laser treatment handle is circular, the
laser head of the 532nm wavelength laser treatment handle is square.
As the picture shows:
4
2.4 Product parts
PicoSure product accessories composition as shown in table:
5
3.2 Device related components installation and water injection instructions
When install the parts of the device, the first to be is water injection operation,
before run the device each time, Must check the water injection position‘s water
level mark whether is above the blue sign the water column or not. If do not reach
it must be distilled water refilling.
Vent
Water mark
Water
injection positio
Respectively connected the funnel and white connectors with the both ends
of the pipe, As the picture shows:
Connected the white connector port with water injection position, through a
funnel to inject water into the device, until the water reaches the blue area, as
the picture shows:
6
After the injection is completed, click stainless steel keys on white connector,
The water appliances can be removed from the water injection position.
Notice: Water drain and water Injection is in the same order,only need to
remove the funnel, then put the injection pipe is placed to the water container.
Notice:Every 1-2 months changing the water inside the device,open the
injection box when changing the water, as much as possible to discharge all
remaining water in the cooling system.
7
3.2.3 Treatment handle installation
When install the handle, need to lower than the machine height,in order to
discharge bubble from the handle.After installed,start the device,let the water
circulation run for at least one minute, then can do treatment.
As the picture shows: Align the handle and interfaces, first, connect the right
followed by the left, connect it carefully, when heard in a row "Kaka" sound,
plug connector is completed, finally, both sides of the plug head can be screwed .
When removing the handle, first removed the fixing screws which are in both
sides of the plug head with the "cross" screwdriver, Then hold down the two
buttons on the plug head, then you can remove the handle ,
as the picture shows:
8
Chapter 4 Device software
operating instructions
Counter
Status column
successful pre-combustion, The status bar indicates turned to ,at this time
step on the pedals to transmit pulses. In the working time transmit each a pulse,
When turn off the device, press the "Stop" button to stop working,then turn off
the power switch, when it is working you can also press the "Trigger" button to
9
Chapter 5 Other system settings
Allow work
instructions
Treatment course
Items First treatment Second treatment Remark
time/Day intervals/Day
Eyebrow tattoo 1~3 30~45 Protect eyes during treatment
Eyeliner tattoo 1~3 45~60 Protect eyes during treatment
Traumatic 1~3 30~45 Select 532nm
tattoo remove
10
Tattoo remove 1~4 30~45 ooze blood when treatment
effect is good
Age spots remove 1~4 30~45 ooze blood when treatment
effect is good
Birthmarks remove 3~7 30~45 ooze blood when treatment
effect is good
Treatment parameters
Skin Items Times Interval/ Wavelength/ Energy/ Frequency/
types Month nm mJ HZ
Black、Blue tattoo 1~2 1~2 1064 3~6
Red、Brown eyebrow 1~3 2~3 532 3~6
Black、Blue eyelines 1~2 1~2 1064 1~4
I Black、Blue tattoo 1~3 2~3 1064 4~6
II Red、 Brown eyebrow 1~4 2~3 532 600~700 4~6
Other color tattoo 3~5 2~3 1064/532 4~6
Melanin mole remove 1~3 1 1064 1~2
Nevus of ota remove 3~6 3 1~3
~3
Lentigines remove 2~3 1~1.5 1~2
~2
11
Black、Blue eyebrow 1~2 1~2 1064 3~6
Red、 Brown eyebrow 1~3 2~3 532 3~6
Black、Blue eyelines 1~2 1~2 1064 1~4
III Black、Blue tattoo 1~3 2~3 1064 4~6
IV Red、 Brown eyebrow 1~4 2~3 532 4~6
Other color tattoo 3~5 2~3 1064/532 4~6
Melanin mole remove 1~3 1 1064 1~2
Nevus of ota remove 3~6 3 1064 1~3
Lentigines remove 2~3 1~1.5 532 1~2
Black、Blue eyebrow 1~2 1~2 1064 3~6
Red、Brown eyebrow 1~3 2~3 532 3~6
Black、Blue eyelines 1~2 1~2 1064 1~4
Black、Blue tattoo 1~3 2~3 1064 4~6
V Red 、 Brown tattoo 1~4 2~3 532 500~700 4~6
Other color tattoo 3~5 2~3 1064/532 4~6
Melanin mole remove 1~3 1 1064 1~2
Nevus of ota remove 3~6 3 1064 1~3
Lentigines remove 2~3 1~1.5 532 1~2
12
After treatment, should take prevent bask, if necessary, taking acetylspiramycin
tablet or antibiotics under the doctor advised.
72 hours after the treatment, applying appropriate amount of erythromycin
ointment, meantime, erasing repair daub or eyebrow liquid on the treatment
parts.
After treatment within one week, avoid to eat much spicy or dark food, also
can not drink wine.
When treatment, the energy of the laser is weakest when the treating head
contacts with the skin, the laser energy is maximum when the treating head
and skin within 4-6cm.
After treatment, the operator should make the details record according to
the treatment data.
13
6.3 System trouble shooting
Check whether the laser mirrors dirt or not, please wipe clean up if has
The output of
the laser Internal handle damaged, replace treatment head
energy is weak
or empty gun - Remove the treatment head, check the front-end mirror damage or not
no laser
Treatment head is overheated, closed the device,cooling a half
hours before use next time
Check the handle and whether treatment head is leaking
Laser treatment Check whether the laser treatment head is damaged , if damaged
head naked to replace
Equipment unused for a long time, triggering difficulties
Treatment
The indoor temperature is too low, increasing the indoor temperature
process, click the
trigger button,
Check whether the circuit malfunction
appear a "creak"
sound
Indoor too wet , reduce indoor humidity
14
Cooling system
Close the system , let the system cooling for a period
temperature too
,then restart .If still not ok , please contact the dealer.
high
Check whether there are impurity in reflector surface,
wipe cleaning if have
Treatment handle Treatment handle overheating , shut off and cooling half
energy is too low an hour then used again
Check whether the front mirror is complete , if damaged ,
please contact the dealer
Treatment handle Without lifetime of treatment head, please contact the
low energy dealer
The tubes in the laser device is replaced , the treatment
Laser damage head is damaged , please contact the dealer
Chapter 7 Specifications
This chapter describes the most important technical parameters and system
categories of treatment system.
Parameter Date
System category
15
Climatic conditions(during operation)
Length 420mm
Width 330mm
Height 440mm
16