Spectrafuge 6C (CLSLN-AN-1003DOC REV1)
Spectrafuge 6C (CLSLN-AN-1003DOC REV1)
Instruction Manual
Catalog Numbers:
C0060
C0060-230V
3.0 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
5.0 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
7.0 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
The electrical warning indicates the presence of a potential hazard which could result in
electrical shock.
CAUTION This symbol refers you to important operating and maintenance (servicing)
instructions within the product Instruction Manual. Failure to heed this information may
present a risk of damage or injury to persons or equipment.
Instruction Manual | 1
3.0 Specifications
Maximum Speed 6,500 rpm
Maximum RCF 4,000 x g
Maximum Volume 6 x 15 mL
Admiss. Density 1.2 kg/dm3
Electrical / Fuse Rating 120V, 50-60Hz, 1.0A/1.25AT
230V, 50-60Hz, 0.6A/630mAT
Dimensions (W x D x H) 8.25 x 9.5 x 7.0 in. (21 x 24 x 18 cm)
Operating Conditions 5°C to 40°C / ≤80%RH
5.0 Installation
The centrifuge should be installed on a rigid, even surface such as a stable laboratory bench, cabinet,
etc. To guarantee sufficient ventilation, ensure that the centrifuge has at least 6 inches (15 cm) of free
space on all sides, including the rear.
The centrifuge should not be located in areas subject to excessive heat such as in direct sunlight or
near radiators or the exhaust of a compressor, as a buildup of heat may occur within the chamber.
Before operating the centrifuge, check that the power source corresponds to that on the
manufacturer’s rating label, then connect the power cord to the centrifuge and the power source.
2 | LABNET SPECTRAFUGE 6C
6.0 Installation of Rotors and Rotor Maintenance
The following accessories are available for the Spectrafuge™ 6C:
Radius (cm) 8.2 8.5 7.7 7.7 8.5 8.5 8.5 7.7 7.7 7.7 7.7 7.7
Max. RFC 3,873 4,015 3,637 3,637 4,015 4,015 4,015 3,637 3,637 3,637 3,637 3,637
Tubes per Rotor 6 6 6 6 6 6 6 6 6 6 6 6
Adapter (C0200-17A) – – Yes Yes – – – Yes Yes Yes Yes Yes
Instruction Manual | 3
A. Motor
B. Motor shaft adapter
C. Motor Shaft
4 | LABNET SPECTRAFUGE 6C
6.3 Overloading the Rotor
The maximum load of the rotor and the maximum speed has been established by the manufacturer.
Do not attempt to exceed these values.
The maximum speed of the rotor has been measured for liquids having a homogeneous density of
1.2 g/mL or less. In order to centrifuge liquids with a higher density it is necessary to reduce the speed.
NOTE: Failure to reduce the speed may result in damage to the rotor and centrifuge.
The revised maximum speed can be calculated with the following formula:
1.2
Reduced speed (nred) = × max speed (nmax)
higher density value
1.2
Reduced speed (nred) = × max speed (nmax)
1.2
Where higher density value
Example:
(nred) = × 6,500 = 5,461 rpm liquid is 1.7, the new maximum speed would be calculated as follows:
the density of the
1.7
1.2
(nred) = × 6,500 = 5,461 rpm
1.7
If in doubt concerning maximum speeds, please contact the manufacturer for assistance.
7.0 Operation
The following accessories are available for the Spectrafuge™ 6C:
CAUTION: Never attempt to operate the centrifuge with rotors or adapters that show signs
of corrosion or mechanical damage. Never centrifuge strongly corrosive materials that may
damage the rotors or accessories.
CAUTION: Do not attempt to open the lid of any centrifuge until the rotor has come to
a complete stop.
In the event of a power failure or malfunction, it may be necessary to open the lid manually.
1. Disconnect the power cord from the wall socket.
2. Remove the plastic plug, located on the left side of the unit, below the quick button.
3. Pull the cord (attached to the plug) to open the lid lock manually.
Instruction Manual | 5
7.5 Selection of Operating Time and Momentary Operation
Time can be selected in half minute intervals from 0.5 to 10 minutes, and in one minute intervals from
11 to 30 minutes. Time can also be set to continuous/hold by turning the timer knob past the thirty
minute position. This will display the continuous setting “On”.
When the preselected time expires, the centrifuge will stop automatically. To stop the centrifuge prior
to the expiration of set time, press the “Start/Stop” knob.
The centrifuge may be operated manually by pressing and holding the “Start/Stop” knob. The
centrifuge will continue to run as long as the knob is depressed.
6 | LABNET SPECTRAFUGE 6C
9.0 Troubleshooting Guide
Problem Possible Cause Solution
Centrifuge will not start No power supply Check that power is being supplied to
the outlet
Check that the power cord is plugged
into both the wall outlet and the back
of the centrifuge.
Check that power cord is not damaged
Blown fuse Check fuse and replace if necessary
Lid lock will not release Defective lid lock Open manually and have unit serviced
No power from PC board Call for service
Lid lock is jammed Call for service
Centrifuge is not receiving See “Centrifuge will not start”
power
Centrifuge cannot be Lid not closed correctly Close lid correctly
started, although power No speed or time has been Set speed and/or time
is on selected
Centrifuge displays error Lid opened prior to signal Close lid and open lid
“03” “00”
Should you have a question about the operation of the Labnet Spectrafuge™ 6C or if service is required,
contact Customer Service. Do not send in a unit for service without first calling to obtain a repair
authorization number. Should the unit require return for service, it should be properly packed to avoid
damage. Any damage resulting from improper packaging shall be the responsibility of the user.
Instruction Manual | 7
10.0 Determination of G-values
The centrifuging radius of the 15 mL rotor is 8.5 cm for round bottom tubes and 8.2 cm for conical
bottom tubes. See Section 6. for the correct radius when using adapters and smaller tubes. The chart on
the next page can be used to determine g-values.
8 | LABNET SPECTRAFUGE 6C
11.0 Limited Warranty
Corning Incorporated (Corning) warrants that this product will be free from defects in material
and workmanship for a period of two (2) years from date of purchase. CORNING DISCLAIMS ALL
OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE. Corning’s sole obligation shall be
to repair or replace, at its option, any product or part thereof that proves defective in material or
workmanship within the warranty period, provided the purchaser notifies Corning of any such defect.
Corning is not liable for any incidental or consequential damages, commercial loss or any other damages
from the use of this product.
This warranty is valid only if the product is used for its intended purpose and within the guidelines
specified in the supplied instruction manual. This warranty does not cover damage caused by accident,
neglect, misuse, improper service, natural forces or other causes not arising from defects in original
material or workmanship. This warranty does not cover motor brushes, fuses, light bulbs, batteries or
damage to paint or finish. Claims for transit damage should be filed with the transportation carrier.
In the event this product fails within the specified period of time because of a defect in material or
workmanship, contact Corning Customer Service at: USA/Canada 1.800.492.1110, outside the U.S.
+1.978.442.2200, visit www.corning.com/lifesciences, or contact your local support office.
Corning’s Customer Service team will help arrange local service where available or coordinate a return
authorization number and shipping instructions. Products received without proper authorization will
be returned. All items returned for service should be sent postage prepaid in the original packaging or
other suitable carton, padded to avoid damage. Corning will not be responsible for damage incurred by
improper packaging. Corning may elect for onsite service for larger equipment.
Some states do not allow limitation on the length of implied warranties or the exclusion or limitation
of incidental or consequential damages. This warranty gives you specific legal rights. You may have other
rights which vary from state to state.
No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the
period of this warranty.
For your reference, make a note of the serial and model number, date of purchase, and supplier here.
Instruction Manual | 9
To request certificates, please contact us at www.labnetlink.com.
Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended
for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the
performance of these products for clinical or diagnostic applications.
For a listing of trademarks, visit www.corning.com/clstrademarks. All other trademarks are the property of their respective owners.
© 2021 Corning Incorporated. All rights reserved. 6/21 CLSLN-AN-1003DOC REV1 www.labnetlink.com