BA01620F/00/EN/02.

18 Products Solutions Services


71394731
2018-04-12
01.00.zz (Device firmware)

Operating Instructions
Micropilot FMR67
HART
Free space radar
Micropilot FMR67 HART

Order code: XXXXX-XXXXXX


1. Ser. no.: XXXXXXXXXXXX
Ext. ord. cd.: XXX.XXXX.XX

Serial number

2. www.endress.com/deviceviewer Endress+Hauser
Operations App

3.

A0023555

2 Endress+Hauser
Micropilot FMR67 HART Table of contents

Table of contents

1 Important document information . . . . 5 6.2.3 Information concerning nozzles . . . . . 24


6.2.4 Information concerning threaded
1.1 Document function . . . . . . . . . . . . . . . . . . . . . 5 connections . . . . . . . . . . . . . . . . . . . . 25
1.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6.3 Installation: FMR67 - Flush-mounted
1.2.1 Safety symbols . . . . . . . . . . . . . . . . . . 5 antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
1.2.2 Electrical symbols . . . . . . . . . . . . . . . . 5 6.3.1 Aligning the antenna axis . . . . . . . . . 25
1.2.3 Tool symbols . . . . . . . . . . . . . . . . . . . . 5 6.3.2 Radial alignment of the antenna . . . . 26
1.2.4 Symbols for 6.3.3 Information concerning nozzles . . . . . 26
certain types of information . . . . . . . . . 6 6.4 FMR67 - Purge air connection . . . . . . . . . . . . 27
1.2.5 Symbols in graphics . . . . . . . . . . . . . . . 6 6.4.1 Purge air adapter for drip-off
1.2.6 Symbols at the device . . . . . . . . . . . . . 7 antennas . . . . . . . . . . . . . . . . . . . . . 27
1.3 Documentation . . . . . . . . . . . . . . . . . . . . . . . . 7 6.4.2 Integrated purge air connection for
1.4 Terms and abbreviations . . . . . . . . . . . . . . . . . 8 flush-mounted antennas . . . . . . . . . . 27
1.5 Registered trademarks . . . . . . . . . . . . . . . . . . . 9 6.4.3 Use . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.5 Container with heat insulation . . . . . . . . . . . . 28
2 Basic safety instructions . . . . . . . . . . . 10 6.6 Turning the transmitter housing . . . . . . . . . . 29
2.1 Requirements for the personnel . . . . . . . . . . . 10 6.7 Turning the display . . . . . . . . . . . . . . . . . . . . 29
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . 10 6.7.1 Opening cover . . . . . . . . . . . . . . . . . . 29
2.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . 11 6.7.2 Turning the display module . . . . . . . . 30
2.4 Operational safety . . . . . . . . . . . . . . . . . . . . . 11 6.7.3 Closing electronics compartment
2.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . 11 cover . . . . . . . . . . . . . . . . . . . . . . . . 30
2.5.1 CE mark . . . . . . . . . . . . . . . . . . . . . . 11 6.8 Post-installation check . . . . . . . . . . . . . . . . . . 30
2.5.2 EAC conformity . . . . . . . . . . . . . . . . . 11
2.6 Safety Instructions (XA) . . . . . . . . . . . . . . . . . 12 7 Electrical connection . . . . . . . . . . . . . . 31
7.1 Connection conditions . . . . . . . . . . . . . . . . . . 31
3 Product description . . . . . . . . . . . . . . . . 14 7.1.1 Terminal assignment . . . . . . . . . . . . 31
3.1 Product design . . . . . . . . . . . . . . . . . . . . . . . . 14 7.1.2 Cable specification . . . . . . . . . . . . . . . 34
3.1.1 Micropilot FMR67 . . . . . . . . . . . . . . . 14 7.1.3 Device plug connectors . . . . . . . . . . . 35
3.1.2 Electronics housing . . . . . . . . . . . . . . 15 7.1.4 Supply voltage . . . . . . . . . . . . . . . . . . 36
7.1.5 Overvoltage protection . . . . . . . . . . . 37
7.1.6 Connecting the measuring device . . . . 38
4 Incoming acceptance and product 7.1.7 Post-connection check . . . . . . . . . . . . 40
identification . . . . . . . . . . . . . . . . . . . . . 16
4.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . 16 8 Operation options . . . . . . . . . . . . . . . . . 41
4.2 Product identification . . . . . . . . . . . . . . . . . . . 16 8.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.2.1 Nameplate . . . . . . . . . . . . . . . . . . . . 17 8.1.1 Local operation . . . . . . . . . . . . . . . . . 41
8.1.2 Operation with remote display and
5 Storage, Transport . . . . . . . . . . . . . . . . 18 operating module FHX50 . . . . . . . . . . 42
5.1 Storage conditions . . . . . . . . . . . . . . . . . . . . . 18 8.1.3 Operation via Bluetooth® wireless
5.2 Transporting the product to the measuring technology . . . . . . . . . . . . . . . . . . . . 43
point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 8.1.4 Remote operation . . . . . . . . . . . . . . . 44
8.2 Structure and function of the operating
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6 Installation . . . . . . . . . . . . . . . . . . . . . . . 19 8.2.1 Structure of the operating menu . . . . 45
6.1 Installation conditions . . . . . . . . . . . . . . . . . . 19 8.2.2 User roles and related access
6.1.1 Orientation - Solid media . . . . . . . . . . 19 authorization . . . . . . . . . . . . . . . . . . 47
6.1.2 Optimization options . . . . . . . . . . . . . 21 8.2.3 Data access - Security . . . . . . . . . . . . 47
6.1.3 Beam angle . . . . . . . . . . . . . . . . . . . . 22 8.3 Display and operating module . . . . . . . . . . . . 52
6.1.4 External measurement through 8.3.1 Display appearance . . . . . . . . . . . . . . 52
plastic cover or dielectric windows . . . 23 8.3.2 Operating elements . . . . . . . . . . . . . . 55
6.2 Installation: Drip-off antenna, PTFE 50 mm / 8.3.3 Entering numbers and text . . . . . . . . 56
2" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8.3.4 Opening the context menu . . . . . . . . . 58
6.2.1 FMR67 - Aligning the antenna axis . . 24 8.3.5 Envelope curve on the display and
6.2.2 Radial alignment of the antenna . . . . 24 operating module . . . . . . . . . . . . . . . 59

Endress+Hauser 3
Table of contents Micropilot FMR67 HART

9 System integration via HART 15 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . 86


protocol . . . . . . . . . . . . . . . . . . . . . . . . . . 60 15.1 General information on repairs . . . . . . . . . . . . 86
15.1.1 Repair concept . . . . . . . . . . . . . . . . . 86
9.1 Overview of the Device Description files
15.1.2 Repairs to Ex-approved devices . . . . . 86
(DD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
15.1.3 Replacement of an electronics
9.2 Measured values via HART protocol . . . . . . . . 60
module . . . . . . . . . . . . . . . . . . . . . . . 86
15.1.4 Replacement of a device . . . . . . . . . . 86
10 Commissioning via SmartBlue 15.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 87
(app) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 15.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
15.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1 Requirements . . . . . . . . . . . . . . . . . . . . . . . . 61
10.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . 61
16 Accessories . . . . . . . . . . . . . . . . . . . . . . . 88
11 Commissioning via wizard . . . . . . . . . 65 16.1 Device-specific accessories . . . . . . . . . . . . . . . 88
16.1.1 Weather protection cover . . . . . . . . . 88
16.1.2 Adjustable flange seal . . . . . . . . . . . . 89
12 Commissioning via operating 16.1.3 Mounting bracket adjustable . . . . . . . 91
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 16.1.4 Remote display FHX50 . . . . . . . . . . . 92
16.1.5 Overvoltage protection . . . . . . . . . . . 93
12.1 Installation and function check . . . . . . . . . . . . 66
16.1.6 Bluetooth module for HART devices . . 94
12.2 Setting the operating language . . . . . . . . . . . . 66
16.2 Communication-specific accessories . . . . . . . . 95
12.3 Configuration of a level measurement . . . . . . 67
16.3 Service-specific accessories . . . . . . . . . . . . . . . 96
12.4 Recording the reference curve . . . . . . . . . . . . 69
16.4 System components . . . . . . . . . . . . . . . . . . . . 97
12.5 Configuration of the on-site display . . . . . . . . 70
12.5.1 Factory settings of the on-site
display . . . . . . . . . . . . . . . . . . . . . . . 70 17 Operating menu . . . . . . . . . . . . . . . . . . . 98
12.5.2 Adjustment of the on-site display . . . 70 17.1 Ovwerview of the operating menu
12.6 Configuration of the current outputs . . . . . . . . 70 (SmartBlue) . . . . . . . . . . . . . . . . . . . . . . . . . . 98
12.6.1 Factory setting of the current 17.2 Overview of the operating menu (display
outputs . . . . . . . . . . . . . . . . . . . . . . . 70 module) . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
12.6.2 Adjustment of the current outputs . . . 70 17.3 Overview of the operating menu (operating
12.7 Configuration management . . . . . . . . . . . . . . 71 tool) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
12.8 Protection of the settings against 17.4 "Setup" menu . . . . . . . . . . . . . . . . . . . . . . . . 115
unauthorized changes . . . . . . . . . . . . . . . . . . 72 17.4.1 "Mapping" wizard . . . . . . . . . . . . . . . 122
17.4.2 "Advanced setup" submenu . . . . . . . . 124
13 Diagnostics and troubleshooting . . . 73 17.5 "Diagnostics" menu . . . . . . . . . . . . . . . . . . . . 168
17.5.1 "Diagnostic list" submenu . . . . . . . . . 170
13.1 General trouble shooting . . . . . . . . . . . . . . . . 73
17.5.2 "Event logbook" submenu . . . . . . . . . 171
13.1.1 General errors . . . . . . . . . . . . . . . . . . 73
17.5.3 "Device information" submenu . . . . . 172
13.1.2 Error - SmartBlue operation . . . . . . . . 74
17.5.4 "Measured values" submenu . . . . . . . 175
13.1.3 Parametrization errors . . . . . . . . . . . 74
17.5.5 "Data logging" submenu . . . . . . . . . . 177
13.2 Diagnostic information on local display . . . . . . 76
17.5.6 "Simulation" submenu . . . . . . . . . . . 180
13.2.1 Diagnostic message . . . . . . . . . . . . . . 76
13.2.2 Calling up remedial measures . . . . . . 78 17.5.7 "Device check" submenu . . . . . . . . . . 185
17.5.8 "Heartbeat" submenu . . . . . . . . . . . . 186
13.3 Diagnostic event in the operating tool . . . . . . . 79
13.4 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . 80
13.5 Overview of diagnostic events . . . . . . . . . . . . . 81 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
13.6 Event logbook . . . . . . . . . . . . . . . . . . . . . . . . 82
13.6.1 Event history . . . . . . . . . . . . . . . . . . . 82
13.6.2 Filtering the event logbook . . . . . . . . 83
13.6.3 Overview of information events . . . . . 83
13.7 Firmware history . . . . . . . . . . . . . . . . . . . . . . 84

14 Maintenance . . . . . . . . . . . . . . . . . . . . . . 85
14.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . 85
14.2 Replacing seals . . . . . . . . . . . . . . . . . . . . . . . 85

4 Endress+Hauser
Micropilot FMR67 HART Important document information

1 Important document information

1.1 Document function


These Operating Instructions contain all the information that is required in various phases
of the life cycle of the device: from product identification, incoming acceptance and
storage, to mounting, connection, operation and commissioning through to
troubleshooting, maintenance and disposal.

1.2 Symbols

1.2.1 Safety symbols

Symbol Meaning

DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.

WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.

CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.

NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.

1.2.2 Electrical symbols

Symbol Meaning

Direct current

Alternating current

Direct current and alternating current

Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.

Protective Earth (PE)


A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.

1.2.3 Tool symbols

Symbol Meaning

Torx screwdriver

A0013442

Flat blade screwdriver

A0011220

Endress+Hauser 5
Important document information Micropilot FMR67 HART

Symbol Meaning

Cross-head screwdriver

A0011219

Allen key

A0011221

Hexagon wrench
A0011222

1.2.4 Symbols for certain types of information

Symbol Meaning

Permitted
Procedures, processes or actions that are permitted.

Preferred
Procedures, processes or actions that are preferred.

Forbidden
Procedures, processes or actions that are forbidden.

Tip
Indicates additional information.

Reference to documentation.

A Reference to page.

Reference to graphic.

Notice or individual step to be observed.

1. , 2. , 3. … Series of steps.

Result of a step.

Help in the event of a problem.

Visual inspection.

1.2.5 Symbols in graphics

Symbol Meaning

1, 2, 3 ... Item numbers

1. , 2. , 3. … Series of steps

A, B, C, ... Views

A-A, B-B, C-C, ... Sections

Hazardous area
- Indicates a hazardous area.

Safe area (non-hazardous area)


. Indicates the non-hazardous area.

6 Endress+Hauser
Micropilot FMR67 HART Important document information

1.2.6 Symbols at the device

Symbol Meaning

Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.

Temperature resistance of the connection cables


Specifies the minimum value of the temperature resistance of the connection cables.

1.3 Documentation
Document Purpose and content of the document

Technical Information Planning aid for your device


TI01304F The document contains all the technical data on the device and provides
an overview of the accessories and other products that can be ordered for
the device.

Brief Operating Instructions Guide that takes you quickly to the 1st measured value
KA01253F The Brief Operating Instructions contain all the essential information
from incoming acceptance to initial commissioning.

Description of Device Parameters Reference for your parameters


GP01101F The document provides a detailed explanation of each individual
parameter in the operating menu. The description is aimed at those who
work with the device over the entire life cycle and perform specific
configurations.

Special Documentation Functional Safety Manual


SD01087F The document is part of the Operating Instructions and serves as a
reference for application-specific parameters and notes.

Special Documentation Manual for Heartbeat Verification and Heartbeat Monitoring


SD01870F The document contains a description of the additional parameters and
technical data that are available with the Heartbeat Verification and
Heartbeat Monitoring application packages.

For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The W@M Device Viewer: enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.

Endress+Hauser 7
Important document information Micropilot FMR67 HART

1.4 Terms and abbreviations


Term/abbreviation Explanation

BA Document type "Operating Instructions"

KA Document type "Brief Operating Instructions"

TI Document type "Technical Information"

SD Document type "Special Documentation"

XA Document type "Safety Instructions"

PN Nominal pressure

MWP Maximum Working Pressure


The MWP can also be found on the nameplate.

ToF Time of Flight

FieldCare Scalable software tool for device configuration and integrated plant asset management
solutions

DeviceCare Universal configuration software for Endress+Hauser HART, PROFIBUS,


FOUNDATION Fieldbus and Ethernet field devices

DTM Device Type Manager

DD Device Description for HART communication protocol

εr (DC value) Relative dielectric constant

Operating tool The term "operating tool" is used in place of the following operating software:
• FieldCare / DeviceCare, for operation via HART communication and PC
• SmartBlue (app), for operation using an Android or iOS smartphone or tablet.

BD Blocking Distance; no signals are analyzed within the BD.

PLC Programmable Logic Controller

CDI Common Data Interface

PFS Pulse Frequence Status (Switching output)

8 Endress+Hauser
Micropilot FMR67 HART Important document information

1.5 Registered trademarks


HART®
Registered trademark of the HART Communication Foundation, Austin, USA
Bluetooth®
The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth
SIG, Inc. and any use of such marks by Endress+Hauser is under license. Other trademarks
and trade names are those of their respective owners.
Apple®
Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in
the U.S. and other countries. App Store is a service mark of Apple Inc.
Android®
Android, Google Play and the Google Play logo are trademarks of Google Inc.
KALREZ®, VITON®
Registered trademark of DuPont Performance Elastomers L.L.C., Wilmington, USA
TEFLON®
Registered trademark of E.I. DuPont de Nemours & Co., Wilmington, USA

Endress+Hauser 9
Basic safety instructions Micropilot FMR67 HART

2 Basic safety instructions

2.1 Requirements for the personnel


The personnel for installation, commissioning, diagnostics and maintenance must fulfill
the following requirements:
‣ Trained, qualified specialists must have a relevant qualification for this specific function
and task.
‣ Are authorized by the plant owner/operator.
‣ Are familiar with federal/national regulations.
‣ Before starting work, read and understand the instructions in the manual and
supplementary documentation as well as the certificates (depending on the
application).
‣ Follow instructions and comply with basic conditions.
The operating personnel must fulfill the following requirements:
‣ Are instructed and authorized according to the requirements of the task by the facility's
owner-operator.
‣ Follow the instructions in this manual.

2.2 Designated use


Application and media
The measuring device described in these Operating Instructions is intended for continuous,
non-contact level measurement primarily in bulk solids. Because of its operating frequency
of approx. 80 GHz, a maximum radiated peak power of 6.3 mW and an average power
output of 63 µW, unrestricted use outside of closed, metallic vessels is also permitted (for
example over heaps). Operation does not pose any danger whatsoever to humans and
animals.
If the limit values specified in the "Technical data" and the conditions listed in the
instructions and additional documentation are observed, the measuring device may be
used for the following measurements only:
‣ Measured process variables: level, distance, signal strength
‣ Calculable process variables: volume or mass in any shape of vessel
To ensure that the measuring device remains in proper condition for the operation time:
‣ Use the measuring device only for media against which the process-wetted materials
are adequately resistant.
‣ Observe the limit values in "Technical data".
Incorrect use
The manufacturer is not liable for damage caused by improper or non-designated use.
Verification for borderline cases:
‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance
in verifying the corrosion resistance of fluid-wetted materials, but does not accept any
warranty or liability.

Residual risks
Due to heat transfer from the process as well as power loss in the electronics, the
temperature of the electronics housing and the assemblies it contains (e.g. display module,
main electronics module and I/O electronics module) may rise to 80 °C (176 °F). When in
operation, the sensor may reach a temperature close to the medium temperature.
Danger of burns from contact with surfaces!
‣ For elevated fluid temperature, ensure protection against contact to prevent burns.

10 Endress+Hauser
Micropilot FMR67 HART Basic safety instructions

2.3 Workplace safety


For work on and with the device:
‣ Wear the required personal protective equipment according to federal/national
regulations.

2.4 Operational safety


Risk of injury.
‣ Operate the device in proper technical condition and fail-safe condition only.
‣ The operator is responsible for interference-free operation of the device.
Conversions to the device
Unauthorized modifications to the device are not permitted and can lead to unforeseeable
dangers.
‣ If, despite this, modifications are required, consult with the manufacturer.
Repair
To ensure continued operational safety and reliability,
‣ Carry out repairs on the device only if they are expressly permitted.
‣ Observe federal/national regulations pertaining to repair of an electrical device.
‣ Use original spare parts and accessories from the manufacturer only.
Hazardous area
To eliminate a danger for persons or for the facility when the device is used in the
hazardous area (e.g. explosion protection, pressure vessel safety):
‣ Based on the nameplate, check whether the ordered device is permitted for the
intended use in the hazardous area.
‣ Observe the specifications in the separate supplementary documentation that is an
integral part of these Instructions.

2.5 Product safety


This measuring device is designed in accordance with good engineering practice to meet
state-of-the-art safety requirements, has been tested, and left the factory in a condition in
which it is safe to operate. It meets general safety standards and legal requirements.

NOTICE
Loss of degree of protection by opening of the device in humid environments
‣ If the device is opened in a humid environment, the degree of protection indicated on
the nameplate is no longer valid. This may also impair the safe operation of the device.

2.5.1 CE mark
The measuring system meets the legal requirements of the applicable EC guidelines. These
are listed in the corresponding EC Declaration of Conformity together with the standards
applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

2.5.2 EAC conformity


The measuring system meets the legal requirements of the applicable EAC guidelines.
These are listed in the corresponding EAC Declaration of Conformity together with the
standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the EAC mark.

Endress+Hauser 11
Basic safety instructions Micropilot FMR67 HART

2.6 Safety Instructions (XA)


Depending on the approval, the following Safety Instructions (XA) are supplied with the
device. They are an integral part of the Operating Instructions.
The nameplate indicates the Safety Instructions (XA) that are relevant to the device.

Feature 010 Approval Feature 020 "Power Supply; Output"


1)
A B 2) C 3)

BA ATEX II 1G Ex ia IIC T6 Ga XA01549F XA01549F XA01549F

BB ATEX II 1/2G Ex ia IIC T6 Ga/Gb XA01549F XA01549F XA01549F

BC ATEX II 1/2G Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01552F XA01552F XA01552F

BD ATEX II 1/2/3G Ex ia/ic [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F
4) 4)
BE ATEX II 1D Ex ta IIIC Da * * * 4)

BF ATEX II 1/2D Ex ta/tb IIIC T85°C Da/Db XA01554F XA01554F XA01554F

BG ATEX II 3G Ex ec IIC T6 Gc XA01551F XA01551F XA01551F

BH ATEX II 3G Ex ic IIC T6 Gc XA01551F XA01551F XA01551F

BL ATEX II 1/2/3G Ex ia/ec [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F

B2 ATEX II 1/2G Ex ia IIC T6 Ga/Gb, 1/2D Ex ia IIIC T85°C Da/Db XA01555F XA01555F XA01555F

B3 ATEX II 1/2G Ex ia/db [ia Ga] IIC T6, Ga/Gb 1/2D Ex ta/tb IIIC T85°C Da/Db XA01556F XA01556F XA01556F

CB CSA IS Cl.I Div.1 Gr.A-D XA01612F XA01612F XA01612F

CD CSA DIP Cl.II,III Div.1 Gr.E-G [Ex ia] XA01613F XA01613F XA01613F

C2 CSA IS Cl.I,II,III Div.1 Gr.A-G, Ex ia, NI Cl.1 Div.2 [Ex ia] XA01612F XA01612F XA01612F

C3 CSA XP Cl.I,II,III Div.1 Gr.A-G, Zn0/1, NI Cl.I Div.2 [Ex ia] XA01613F XA01613F XA01613F

FA FM IS Cl.I Div.1 Gr.A-D XA01615F XA01615F XA01615F

FB FM IS Cl.I,II,III Div.1 Gr.A-G, AEx ia, NI Cl.1 Div.2 XA01615F XA01615F XA01615F

FC FM XP-IS Cl.I Div.1 Gr.A-D, AIS Cl.I Div.1 Gr.A-D XA01616F XA01616F XA01616F

FD FM XP-IS Cl.I Div.1 Gr.A-D, Zn0/1, DIP-IS Cl.II,III Div.1 Gr.E-G, NI Cl.I Div.2 XA01616F XA01616F XA01616F

FE FM DIP Cl.II,III Div.1 Gr.E-G XA01616F XA01616F XA01616F

GA EAC 0Ex ia IIC T6...T3 Ga X XA01617F XA01617F XA01617F

GB EAC Ga/Gb Ex ia IIC T6...T3 X XA01617F XA01617F XA01617F

GC EAC Ga/Gb Ex ia/db [ia Ga] IIC T6...T3 X XA01618F XA01618F XA01618F
4) 4)
GE EAC Ex ta IIIC Da * * * 4)

GF EAC Ex ta/tb IIIC T85°C Da/Db X XA01619F XA01619F XA01619F

IA IEC Ex ia IIC T6 Ga XA01549F XA01549F XA01549F

IB IEC Ex ia IIC T6 Ga/Gb XA01549F XA01549F XA01549F

IC IEC Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01552F XA01552F XA01552F

ID IEC Ex ia/ic [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F


4) 4)
IE IEC Ex ta IIIC Da * * * 4)

IF IEC Ex ta/tb IIIC T85°oC Da/Db XA01554F XA01554F XA01554F

IG IEC Ex ec IIC T6 Gc XA01551F XA01551F XA01551F

IH IEC Ex ic IIC T6 Gc XA01551F XA01551F XA01551F

IL IEC Ex ia/ec [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F

I2 IEC Ex ia IIC T6 Ga/Gb, Ex ia IIIC T85°C Da/Db XA01555F XA01555F XA01555F

I3 IEC Ex ia/db [ia Ga] IIC T6 Ga/Gb, Ex ta/tb IIIC T85°C Da/Db XA01556F XA01556F XA01556F
4) 4)
JA JPN Ex ia IIC T6 Ga XA01631F XA01631F XA01631F 4)

12 Endress+Hauser
Micropilot FMR67 HART Basic safety instructions

Feature 010 Approval Feature 020 "Power Supply; Output"


1)
A B 2) C 3)

JB JPN Ex ia IIC T6 Ga/Gb XA01631F 4) XA01631F 4) XA01631F 4)

JC JPN Ex d [ia] IIC T6 Ga/Gb XA01632F 4) XA01632F 4) XA01632F 4)

JG JPN Ex nA IIC T6 Gc XA01725F 4) XA01725F 4) XA01725F 4)

JH JPN Ex ic IIC T6 Gc XA01725F 4) XA01725F 4) XA01725F 4)

J2 JPN Ex ia IIC T6 Ga/Gb, JPN Ex ia IIIC T85°C Da/Db XA01728F 4) XA01728F 4) XA01728F 4)

J3 JPN Ex d [ia] IIC T6 Ga/Gb, JPN Ex ta/tb IIIC T85°C Da/Db XA01729F 4) XA01729F 4) XA01729F 4)

KA KC Ex ia IIC T6 Ga XA01623F XA01623F XA01623F

KB KC Ex ia IIC T6 Ga/Gb XA01623F XA01623F XA01623F

KC KC Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01624F XA01624F XA01624F

MA INMETRO Ex ia IIC T6 Ga XA01620F XA01620F XA01620F

MB INMETRO Ex ia IIC T6 Ga/Gb XA01620F XA01620F XA01620F


4) 4)
ME INMETRO Ex ta IIIC Da * * * 4)

MG INMETRO Ex ec IIC T6 Gc XA01621F XA01621F XA01621F

MH INMETRO Ex ic IIC T6 Gc XA01621F XA01621F XA01621F

NA NEPSI Ex ia IIC T6 Ga XA01625F XA01625F XA01625F

NB NEPSI Ex ia IIC T6 Ga/Gb XA01625F XA01625F XA01625F

NC NEPSI Ex ia/d [ia Ga] IIC T6 Ga/Gb XA01627F XA01627F XA01627F

NF NEPSI Ex tD A20/A21 IP6X T85°C XA01628F XA01628F XA01628F

NG NEPSI Ex nA IIC T6 Gc XA01626F XA01626F XA01626F

NH NEPSI Ex ic IIC T6 Gc XA01626F XA01626F XA01626F

N2 NEPSI Ex ia IIC T6 Ga/Gb, NEPSI Ex iaD 20/21 T85 XA01629F XA01629F XA01629F

N3 NEPSI Ex ia/d [ia Ga] IIC T6 Ga/Gb, NEPSI Ex tD A20/A21 IP6X T85°C XA01630F XA01630F XA01630F

8A FM/CSA IS+XP-IS Cl.I,II,III Div.1 Gr.A-G, AIS Cl.I,II,III Div.1 Gr.A-G XA01612F XA01612F XA01612F
XA01615F XA01615F XA01615F
XA01616F XA01616F XA01616F

* 4)

1) 2-wire; 4-20mA HART


2) 2-wire; 4-20mA HART, switch output
3) 2-wire; 4-20mA HART, 4-20mA
4) in preparation

Endress+Hauser 13
Product description Micropilot FMR67 HART

3 Product description

3.1 Product design

3.1.1 Micropilot FMR67

3 3 3

2.1 2.2
2.2

1 1 1
A0032714

1 Design of the Micropilot FMR67


1 Drip-off antenna PTFE
2.1 Process connection (Thread)
2.2 Process connection (Flange)
3 Electronics housing
4 Purge air connection

3 3 3

4 4 4
2.1
2.2
2.3

1 1
1
A0032782

2 Design of the Micropilot FMR67


1 PTFE antenna
2.1 Process connection (Flange)
2.2 Process connection (UNI-Flange)
2.3 Process connection (Flange with alignment device)
3 Electronics housing
4 Purge air connection

14 Endress+Hauser
Micropilot FMR67 HART Product description

3.1.2 Electronics housing

4
3
2
9
1
6

7
8

A0012422

3 Design of the electronics housing


1 Electronics compartment cover
2 Display module
3 Main electronics module
4 Cable glands (1 or 2, depending on instrument version)
5 Nameplate
6 I/O electronics module
7 Terminals (pluggable spring terminals)
8 Connection compartment cover
9 Grounding terminal

Endress+Hauser 15
Incoming acceptance and product identification Micropilot FMR67 HART

4 Incoming acceptance and product


identification

4.1 Incoming acceptance


Check the following during incoming acceptance:
• Are the order codes on the delivery note and the product sticker identical?
• Are the goods undamaged?
• Do the nameplate data match the ordering information on the delivery note?
• If required (see nameplate): Are the safety instructions (XA) present?
If one of these conditions does not apply, please contact your Endress+Hauser sales
office.

4.2 Product identification


The following options are available for identification of the measuring device:
• Nameplate specifications
• Extended order code with breakdown of the device features on the delivery note
• Enter the serial number on the nameplate into W@M Device Viewer
(www.endress.com/deviceviewer): all the information about the measuring device is
displayed.
• Enter the serial number on the nameplate into the Endress+Hauser Operations App or
scan the 2-D matrix code (QR code) on the nameplate with the Endress+Hauser
Operations App: all the information about the measuring device is displayed.
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The W@M Device Viewer: enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or
scan the 2-D matrix code (QR code) on the nameplate.

16 Endress+Hauser
Micropilot FMR67 HART Incoming acceptance and product identification

4.2.1 Nameplate

92 (3.62)

2
1
Order code: 3 10
Ser. no.: 4
Ext. ord. cd.: 8
5
26

47 (1.85)
25
Tp max = 24 MWP: 6 9
23 LN = 7
Ta: 22 15
Mat.: 21
20
FW: 18 Dev.Rev.: 16 11
if modification
19 17 13 X = see sep. label
12
Date: 14
mm (in)
A0019444

4 Nameplate of the Micropilot


1 Device name
2 Manufacturer's address
3 Order code
4 Serial number (ser. no.)
5 Extended order code (Ext. ord. cd.)
6 Process pressure
7 Antenna length reference length
8 Certificate symbol
9 Certificate and approval relevant data
10 Degree of protection: e.g. IP, NEMA
11 Document number of the Safety Instructions: e.g. XA, ZD, ZE
12 2-D matrix code (QR code)
13 Modification mark
14 Manufacturing date: year-month
15 Temperature resistance of cable
16 Device revision (Dev.Rev.)
17 Additional information about the device version (certificates, approvals, communication protocol)
18 Firmware version (FW)
19 CE mark, C-Tick
20 Profibus PA: Profile Version; FOUNDATION Fieldbus: Device ID
21 Materials in contact with process
22 Permitted ambient temperature (Ta)
23 Size of the cable gland thread
24 Maximum process temperature
25 Signal outputs
26 Supply voltage

Up to 33 characters of the extended order code are indicated on the nameplate. If the
extended order code contains additional characters, these cannot be displayed.
However, the complete extended order code can also be displayed via the device
operating menu: Extended order code 1 to 3 parameter

Endress+Hauser 17
Storage, Transport Micropilot FMR67 HART

5 Storage, Transport

5.1 Storage conditions


• Permitted storage temperature: –40 to +80 °C (–40 to +176 °F)
• Use original packaging.

5.2 Transporting the product to the measuring point


NOTICE
Housing or sensor may become damaged or pull off.
Risk of injury!
‣ Transport the measuring device to the measuring point in its original packaging or by
the process connection.
‣ Always secure lifting equipment (slings, eyes, etc.) at the process connection and never
lift the device by the electronic housing or sensor. Pay attention to the center of gravity
of the device so that it does not tilt or slip unintentionally.
‣ Follow the safety instructions and transport conditions for devices over 18 kg (39.6
lbs), (IEC61010).

A0032300

18 Endress+Hauser
Micropilot FMR67 HART Installation

6 Installation

6.1 Installation conditions

6.1.1 Orientation - Solid media

1 2 3

A0016883

• Recommended distance A wall - nozzle outer edge: ~ 1/6 of the container diameter.
However, the device must not under any circumstances be mounted closer than
20 cm (7.87 in) to the container wall.
If the container wall is not smooth (corrugated iron, welding seams, joints, etc.) it is
recommended to maintain the largest possible distance from the wall. Where necessary
use an alignment unit to avoid interference reflections from the container wall
→  25.
• Not in the center (2) as interference can cause signal loss.
• Not above the filling curtain (3).
• The use of a weather protection cover (1) is recommended to protect the transmitter
from direct sunlight or rain.
• In applications with strong dust emissions, the integrated purge air connection can
prevent the antenna from becoming clogged .

Endress+Hauser 19
Installation Micropilot FMR67 HART

Internal container fittings

A0031814

Avoid the location of internal fittings (limit switches, temperature sensors, struts etc.)
inside the signal beam. Take into account the beam angle .

20 Endress+Hauser
Micropilot FMR67 HART Installation

Avoiding interference echoes

A0031817

Metal deflector plates, installed at an angle to scatter the radar signals, help prevent
interference echoes.

6.1.2 Optimization options


• Antenna size
The larger the antenna the smaller the beam angle α, resulting in fewer interference
echoes →  22.
• Mapping
Measurement can be optimized by electronically suppressing interference echoes.
See also Confirm distance parameter.
• Adjustable flange seal for FMR67
Adjustable flange seals sized from DN80 to DN150 (3" to 6") are available for the FMR67
with drip-off antenna 1). They can be used to align the device to the product surface.
Maximum angle of alignment: 8 °.
How to order:
– Order with the device 2)
– Order as an accessory: →  89
• Alignment unit for FMR67
Flanges from 4" / DN100 are optionally available with an alignment unit 3) They allow
the sensor to be optimally aligned to suit conditions in the container in order to prevent
interference reflections. The maximum angle is ±15 °.
The purpose of sensor alignment is primarily to:
– Prevent interference reflections
– Increase the maximum possible measuring range in conical outlets

1) Feature 070 in product structure, "Antenna", option GA


2) Feature 100 in product structure "Process connection", options PL, PM, PN, PO, PQ, PR
3) See feature 100 in product structure, "Process connection".

Endress+Hauser 21
Installation Micropilot FMR67 HART

6.1.3 Beam angle

α
W= 2 . D . tan_
2

D
α

W
A0031824

5 Relationship between beam angle α, distance D and beamwidth diameter W

The beam angle is defined as the angle α where the energy density of the radar waves
reaches half the value of the maximum energy density (3dB width). Microwaves are also
emitted outside the signal beam and can be reflected off interfering installations.

22 Endress+Hauser
Micropilot FMR67 HART Installation

FMR67

A0032084

A0032083

Antenna 1) Drip-off, PTFE 50 mm / 2" PTFE flush-mounted 80 mm /


3"

Beam angle α 6° 4°

Distance (D) Beamwidth diameter W

5 m (16 ft) 0.52 m (1.70 ft) 0.35 m (1.15 ft)

10 m (33 ft) 1.05 m (3.44 ft) 0.70 m (2.30 ft)

15 m (49 ft) 1.57 m (5.15 ft) 1.05 m (3.44 ft)

20 m (66 ft) 2.10 m (6.89 ft) 1.40 m (4.59 ft)

25 m (82 ft) 2.62 m (8.60 ft) 1.75 m (5.74 ft)

30 m (98 ft) 3.14 m (10.30 ft) 2.10 m (6.89 ft)

35 m (115 ft) 3.67 m (12.04 ft) 2.44 m (8.00 ft)

40 m (131 ft) 4.19 m (13.75 ft) 2.79 m (9.15 ft)

45 m (148 ft) 4.72 m (15.49 ft) 3.14 m (10.30 ft)

50 m (164 ft) 5.24 m (17.19 ft) 3.49 m (11.45 ft)

60 m (197 ft) - 4.19 m (13.75 ft)

70 m (230 ft) - 4.89 m (16.04 ft)

80 m (262 ft) - 5.59 m (18.34 ft)

90 m (295 ft) - 6.29 m (20.64 ft)

100 m (328 ft) - 6.98 m (22.90 ft)

110 m (361 ft) - 7.68 m (25.20 ft)

120 m (394 ft) - 8.38 m (27.49 ft)

125 m (410 ft) - 8.73 m (25.64 ft)

1) Feature 070 in product structure

6.1.4 External measurement through plastic cover or dielectric


windows
• Dielectric constant of medium: εr ≥ 10
• The distance from the tip of the antenna to the tank should be approx. 100 mm (4 in).
• If possible, avoid installation positions in which condensate or buildup can form between
the antenna and the vessel.
• In the case of outdoor installations, ensure that the area between the antenna and the
tank is protected from the weather.
• Do not install any fittings or attachments between the antenna and the tank that could
reflect the signal.

Endress+Hauser 23
Installation Micropilot FMR67 HART

Suitable thickness of tank roof or window


Material PE PTFE PP Perspex

εr 2.3 2.1 2.3 3.1


(Dielectric constant of
medium)

Optimum thickness 1.25 mm 1.3 mm 1.25 mm 1.07 mm


(0.049 in) 1) (0.051) 1) (0.049 in) 1) (0.042 in) 1)

1) or an integer that is a multiple of this value; it should be noted here that the microwave transparency
decreases significantly with increasing window thickness.

6.2 Installation: Drip-off antenna, PTFE 50 mm / 2"

6.2.1 FMR67 - Aligning the antenna axis


Align the antenna vertically to the product surface.
If necessary, the antenna can be aligned by means of a adjustable flange seal (available as
an accessory).
Attention:
The maximum reach of the antenna can be reduced if it is not installed perpendicular
to the product.

6.2.2 Radial alignment of the antenna


Based on the directional characteristic, radial alignment of the antenna is not necessary.

6.2.3 Information concerning nozzles


The maximum nozzle length Hmax depends on the nozzle diameter D:
Hmax

øD
A0032209

Nozzle diameter (ØD) Maximum nozzle length (Hmax) 1)

50 to 80 mm (2 to 3.2 in) 750 mm (30 in)

80 to 100 mm (3.2 to 4 in) 1 150 mm (46 in)

24 Endress+Hauser
Micropilot FMR67 HART Installation

Nozzle diameter (ØD) Maximum nozzle length (Hmax) 1)

100 to 150 mm (4 to 6 in) 1 450 mm (58 in)

≥150 mm (6 in) 2 200 mm (88 in)

1) In the case of longer nozzles, a reduced measuring performance must be anticipated.

Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle
should be rounded if possible.
• Mapping must be performed.
• Please contact Endress+Hauser for applications with nozzles that are higher than
indicated in the table.

6.2.4 Information concerning threaded connections


• When screwing in, turn by the hex bolt only.
• Tool: open-ended wrench 55 mm
• Maximum permissible torque: 50 Nm (36 lbf ft)

6.3 Installation: FMR67 - Flush-mounted antenna

6.3.1 Aligning the antenna axis


UNI flanges with an integrated alignment unit are available for FMR67 devices with a
flush mount antenna. An angle of inclination of up to 15° in all directions can be set for
the antenna axis using the alignment unit. The alignment unit is used to optimally align
the radar beam to the bulk solid.

Process connection with UNI flange Material Pressure rating Suitable for
Alignment unit 1)

XCA UNI 4" / DN100 / 100A Aluminum max. 14.5lbs / PN1 / 1K • 4" 150lbs
• DN100 PN16
• 10K 100A

XDA UNI 6" / DN150 / 150A Aluminum max. 14.5lbs / PN1 / 1K • 6" 150lbs
• DN150 PN16
• 10K 150A

XEA UNI 8" / DN200 / 200A Aluminum max. 14.5lbs / PN1 / 1K • 8" 150lbs
• DN200 PN16
• 10K 200A

XFA UNI 10" / DN250 / 250A Aluminum max. 14.5lbs / PN1 / 1K • 10" 150lbs
• DN250 PN16
• 10K 250A

1) Feature 100 in product structure

Endress+Hauser 25
Installation Micropilot FMR67 HART

A0032097

6 Micropilot FMR67 with alignment unit

Aligning the antenna axis


1. Release the screws
2. Align the antenna axis (up to max. ±15° possible in all directions)
3. Tighten the screws with 10 Nm (7.4 lbf ft)

6.3.2 Radial alignment of the antenna


Based on the directional characteristic, radial alignment of the antenna is not necessary.

6.3.3 Information concerning nozzles


Hmax

øD
A0032206

26 Endress+Hauser
Micropilot FMR67 HART Installation

Internal diameter of nozzle D Maximum nozzle height Hmax

min. 80 to 100 mm (3 to 4 in) 1 450 mm (57 in)

100 to 150 mm (4 to 6 in) 1 800 mm (71 in)

≥150 mm (6 in) 2 700 mm (106 in)

Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle
should be rounded if possible.
• Mapping must be performed.
• Please contact Endress+Hauser for applications with nozzles that are higher than
indicated in the table.

6.4 FMR67 - Purge air connection

6.4.1 Purge air adapter for drip-off antennas

Purge air connection 1) Meaning

A None

3 Purge air adapter G 1/4"

4 Purge air adapter NPT 1/4"

1) Feature 110 in product structure

A0032098

1 Purge air connection NPT 1/4" or G 1/4"

6.4.2 Integrated purge air connection for flush-mounted antennas

Purge air connection 1) Meaning

1 Purge air connection G 1/4"

2 Purge air connection NPT 1/4"

1) Feature 110 in product structure

Endress+Hauser 27
Installation Micropilot FMR67 HART

1 1

A0032099

1 Purge air connection NPT 1/4" or G 1/4"

6.4.3 Use
In applications with strong dust emissions, the integrated purge air connection can prevent
the antenna from becoming clogged. Pulse operation is recommended.
Purge air pressure range
• Pulse operation :
Max. 6 bar (87 psi)
• Continuous operation:
200 to 500 mbar (3 to 7.25 psi)
Purge air connection
• Tool:
– Open-ended wrench 13 mm (G 1/4")
– Open-ended wrench 14 mm (NPT)
– Open-ended wrench 17 mm (NPT "adapter")
• remains stable for a minimum of torque: 6 Nm (4.4 lbf ft)
• Max. torque: 7 Nm
Always use dry purge air.

In general, purging should only be performed to the extent necessary as excess


purging can cause mechanical damage (abrasion).

6.5 Container with heat insulation

A0032207

28 Endress+Hauser
Micropilot FMR67 HART Installation

If process temperatures are high, the device should be included in the usual container
insulation system (2) to prevent the electronics from heating as a result of thermal
radiation or convection. The insulation should not be higher than the neck of the device
(1).

6.6 Turning the transmitter housing


To provide easier access to the connection compartment or display module, the transmitter
housing can be turned:

2.
max. 350°

1. 3.

8 mm 8 mm

A0032242

1. Unscrew the securing screw using an open-ended wrench.


2. Rotate the housing in the desired direction.
3. Tighten the securing screw (1.5 Nm for plastic housing; 2.5 Nm for aluminum or
stainless steel housing).

6.7 Turning the display

6.7.1 Opening cover

2.

1.

3 mm
A0021430

1. Loosen the screw of the securing clamp of the electronics compartment cover using
an Allen key (3 mm) and turn the clamp 90 ° counterclockwise.
2. Unscrew cover and check lid gasket, replace if necessary.

Endress+Hauser 29
Installation Micropilot FMR67 HART

6.7.2 Turning the display module

2.

1.

A0036401

1. Pull out the display module with a gentle rotational movement.


2. Rotate the display module to the desired position: max. 8 × 45 ° in each direction.
3. Feed the coiled cable into the gap between the housing and main electronics module
and plug the display module into the electronics compartment until it engages.

6.7.3 Closing electronics compartment cover

1.

3 mm
2. 2,5 Nm

A0021451

1. Screw back firmly electronics compartment cover.


2. Turning securing clamp 90 ° clockwise and tighten the clamp with 2.5 Nm using the
Allen key (3 mm).

6.8 Post-installation check


 Is the device undamaged (visual inspection)?

Does the device conform to the measuring point specifications?


For example:
• Process temperature
 • Process pressure (refer to the chapter on "Material load curves" of the "Technical Information"
document)
• Ambient temperature range
• Measuring range

 Are the measuring point identification and labeling correct (visual inspection)?

 Is the device adequately protected from precipitation and direct sunlight?

 Are the securing screw and securing clamp tightened securely?

30 Endress+Hauser
Micropilot FMR67 HART Electrical connection

7 Electrical connection

7.1 Connection conditions

7.1.1 Terminal assignment

Terminal assignment 2-wire: 4-20 mA HART

A B

+
1 1 -
2 2
+
-

1 3
A0036498

7 Terminal assignment 2-wire: 4-20 mA HART


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage protection
2 Connection 4-20 mA HART passive: terminals 1 and 2, with integrated overvoltage protection
3 Terminal for cable screen

Block diagram 2-wire: 4-20 mA HART

3 4
2 5
1 6
+ + 1
I
- - 2

A0036499

8 Block diagram 2-wire: 4-20 mA HART


1 Active barrier with power supply (e.g. RN221N); observe terminal voltage
2 HART communication resistor (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device

Endress+Hauser 31
Electrical connection Micropilot FMR67 HART

Terminal assignment 2-wire: 4-20 mA HART, switch output

3
4
A B

+
3 3 -
4 4
+ +
- 1 -
2 1
+ 2
-

2
1 5
A0036500

9 Terminal assignment 2-wire: 4-20 mA HART, switch output


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage protection
2 Connection switch output (Open Collector): terminals 3 and 4, without integrated overvoltage protection
3 Connection switch output (Open Collector): terminals 3 and 4, with integrated overvoltage protection
4 Connection 4-20 mA HART passive: terminals 1 and 2, with integrated overvoltage protection
5 Terminal for cable screen

Block diagram 2-wire: 4-20 mA HART, switch output

3 4
2 5
1 6
+ + 1
I
- - 2

+ 3 Y
- 4

7
A0036501

 10 Block diagram 2-wire: 4-20 mA HART, switch output


1 Active barrier with power supply (e.g. RN221N); observe terminal voltage
2 HART communication resistor (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Switch output (Open Collector)

32 Endress+Hauser
Micropilot FMR67 HART Electrical connection

Terminal assignment 2-wire: 4-20 mA HART, 4-20 mA

3
4
A B

+
3 3 -
4 4
+ +
- 1 -
2 1
+ 2
-

2
1 5
A0036500

 11 Terminal assignment 2-wire: 4-20 mA HART, 4-20 mA


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection current output 1, 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage
protection
2 Connection current output 2, 4-20 mA: terminals 3 and 4, without integrated overvoltage protection
3 Connection current output 2, 4-20 mA: terminals 3 and 4, with integrated overvoltage protection
4 Connection current output 1, 4-20 mA HART passive: terminals 1 and 2, with integrated overvoltage
protection
5 Terminal for cable screen

Block diagram 2-wire: 4-20 mA HART, 4-20 mA

3 4
2 5
1 6
+ + 1
I
- - 2

+ + 3 Y
- - 4

7
8
A0036502

 12 Block diagram 2-wire: 4-20 mA HART, 4-20 mA


1 Active barrier with power supply (e.g. RN221N); observe terminal voltage
2 HART communication resistor (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Analog display device; observe maximum load
8 Active barrier with power supply (e.g. RN221N), current output 2; observe terminal voltage

Endress+Hauser 33
Electrical connection Micropilot FMR67 HART

Connection examples for the switch output

+
3+ 1
2
Ri + 3+
-
Ri
4-
A0015909
4-
 13 Connection of a relay A0015910

 14 Connection of a digital input


Suitable relays (examples):
• Solid-state relay: Phoenix Contact OV-24DC/480AC/5 with mounting 1 Pull-up resistor
rail connector UMK-1 OM-R/AMS 2 Digital input
• Electromechanical relay: Phoenix Contact PLC-RSC-12DC/21

For optimum interference immunity we recommend to connect an external resistor


(internal resistance of the relay or Pull-up resistor) of < 1 000 Ω.

7.1.2 Cable specification


• Devices without integrated overvoltage protection
Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
• For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.

HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept
of the plant.

34 Endress+Hauser
Micropilot FMR67 HART Electrical connection

7.1.3 Device plug connectors


For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be
connected without opening the housing.

Pin assignment of the M12 plug connector


Pin Meaning
4 3 1 Signal +

2 not connected
1 2 3 Signal -

A0011175
4 Ground

Pin assignment of the 7/8" plug connector


Pin Meaning
1 3 1 Signal -

2 Signal +
2 4 3 Not connected

A0011176
4 Screen

Endress+Hauser 35
Electrical connection Micropilot FMR67 HART

7.1.4 Supply voltage

2-wire, 4-20 mA HART, passive

"Power "Approval" 2) Terminal Maximum load R,


supply, voltage U at depending on the supply voltage
output" 1) device U0 of the power supply unit

A: • Non- 14 to 35 V 3) R [W]
2-wire; hazardous
4-20 mA • Ex nA 500
HART • Ex ic
• CSA GP

Ex ia / IS 14 to 30 V 3)

• Ex d(ia) / 14 to 35 V 3) 4)
XP
• Ex ic(ia) 0
• Ex nA(ia) 10 20 30 35 U0 [V]
• Ex ta / DIP 14 25
A0031745

Ex ia + Ex 14 to 30 V 3)
d(ia) / IS + XP

1) Feature 020 in the product structure


2) Feature 010 in the product structure
3) If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.
4) At ambient temperatures TTa ≤ –20 °C, a terminal voltage U ≥ 16 V is required to start the device with the
minimum failure current (3.6 mA).

"Power "Approval" 2) Terminal Maximum load R,


supply, voltage U at depending on the supply voltage
output" 1) device U0 of the power supply unit

B: • Non- 16 to 35 V 3) R [W]
2-wire; hazardous
4-20 mA • Ex nA 500
HART, • Ex nA(ia)
switch • Ex ic
outpu • Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
0
• Ex ia / IS 16 to 30 V 3)
• Ex ia + Ex 10 20 30 35 U0 [V]
16 27
d(ia) / IS + A0031746
XP

1) Feature 020 in the product structure


2) Feature 010 in the product structure
3) If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.

36 Endress+Hauser
Micropilot FMR67 HART Electrical connection

"Power "Approval" 2) Terminal Maximum load R,


supply, voltage U at depending on the supply voltage
output" 1) device U0 of the power supply unit

C: All 16 to 30 V 3) R [W]
2-wire; 4-20
mA HART, 500
4-20 mA

0
10 20 30 35 U0 [V]
16 27
A0031746

1) Feature 020 in the product structure


2) Feature 010 in the product structure
3) If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.

Integrated polarity reversal protection Yes

Permitted residual ripple with f = 0 to 100 Hz USS < 1 V

Permitted residual ripple with f = 100 to 10 000 Hz USS < 10 mV

7.1.5 Overvoltage protection


If the measuring device is used for level measurement in flammable liquids which requires
the use of overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), an overvoltage protection module has to
be installed.

Integrated overvoltage protection module


An integrated overvoltage protection module is available for 2-wire HART devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".

Technical data

Resistance per channel 2 × 0.5 Ω max.

Threshold DC voltage 400 to 700 V

Threshold impulse voltage < 800 V

Capacitance at 1 MHz < 1.5 pF

Nominal arrest impulse voltage (8/20 μs) 10 kA

External overvoltage protection module


HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.

Endress+Hauser 37
Electrical connection Micropilot FMR67 HART

7.1.6 Connecting the measuring device


LWARNING
Risk of explosion!
‣ Observe applicable national standards.
‣ Comply with the specifications in the Safety Instructions (XA).
‣ Use specified cable glands only.
‣ Check to ensure that the power supply matches the information on the nameplate.
‣ Switch off the power supply before connecting the device.
‣ Connect the potential matching line to the outer ground terminal before applying the
power supply.

Required tools/accessories:
• For devices with a cover lock: Allen key AF3
• Wire stripper
• When using stranded cables: One ferrule for every wire to be connected.

Opening connection compartment cover

2.

3 mm
1.
A0021490

1. Loosen the screw of the securing clamp of the connection compartment cover using
an Allen key (3 mm) and turn the clamp 90 ° clockwise.
2. Afterwards unscrew connection compartment cover and check lid gasket, replace if
necessary.

38 Endress+Hauser
Micropilot FMR67 HART Electrical connection

Connecting

1.

3.
10 (0.4)
2.
3.
10 (0.4)

A0036418

 15 Dimensions: mm (in)

1. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
2. Remove the cable sheath.
3. Strip the cable ends over a length of 10 mm (0.4 in). In the case of stranded cables,
also fit ferrules.
4. Firmly tighten the cable glands.
5. Connect the cable in accordance with the terminal assignment.

34
1
2
3
4
1
2

A0034682

6. If using shielded cables: Connect the cable shield to the ground terminal.

Plug-in spring-force terminals


In the case of devices without integrated overvoltage protection, electrical connection is via
plug-in spring-force terminals. Rigid conductors or flexible conductors with ferrules can be
inserted directly into the terminal without using the lever, and create a contact
automatically.

Endress+Hauser 39
Electrical connection Micropilot FMR67 HART

1. 2.

≤ 3 (0.12)

A0013661

 16 Dimensions: mm (in)

To remove cables from the terminal:


1. Using a flat-blade screwdriver ≤ 3 mm, press down on the slot between the two
terminal holes
2. while simultaneously pulling the cable end out of the terminal.

Closing connection compartment cover

1.

3 mm
2. 2,5 Nm

A0021491

1. Screw back firmly connection compartment cover.


2. Turning securing clamp 90 ° counterclockwise and tighten the clamp with
2.5 Nm (1.84 lbf ft) again using the Allen key (3 mm).

7.1.7 Post-connection check

 Is the device or cable undamaged (visual check)?

 Do the cables comply with the requirements ?

 Do the cables have adequate strain relief?

 Are all cable glands installed, securely tightened and leak-tight?

 Does the supply voltage match the specifications on the nameplate?

 Is the terminal assignment correct?

 If required: Has protective ground connection been established ?

If supply voltage is present, is the device ready for operation and do values appear on the display

module?

 Are all housing covers installed and securely tightened?

 Is the securing clamp tightened correctly?

40 Endress+Hauser
Micropilot FMR67 HART Operation options

8 Operation options

8.1 Overview

8.1.1 Local operation

Operation Pushbuttons Touch Control


with

Order code for Option C "SD02" Option E "SD03"


"Display;
Operation"

A0036312 A0036313

Display 4-line display 4-line display


elements white background lighting; switches to red in
event of device error

Format for displaying measured variables and status variables can be individually configured

Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.

Operating local operation with 3 push buttons (, , ) external operation via touch control; 3
elements optical keys: , , 

Operating elements also accessible in various hazardous areas

Additional Data backup function


functionality The device configuration can be saved in the display module.

Data comparison function


The device configuration saved in the display module can be compared to the current device
configuration.

Data transfer function


The transmitter configuration can be transmitted to another device using the display module.

Endress+Hauser 41
Operation options Micropilot FMR67 HART

8.1.2 Operation with remote display and operating module FHX50

1 2

A0036314

 17 FHX50 operating options


1 Display and operating module SD03, optical keys; can be operated through the glass of the cover
2 Display and operating module SD02, push buttons; cover must be removed

42 Endress+Hauser
Micropilot FMR67 HART Operation options

8.1.3 Operation via Bluetooth® wireless technology

Requirements

2
A0036790

 18 Device with Bluetooth module


1 Electronics housing of the device
2 Bluetooth module

This operation option is only available for devices with Bluetooth module. There are the
following options:
• The device has been ordered with a Bluetooth module:
Feature 610 "Accessory Mounted", option NF "Bluetooth"
• The Bluetooth module has been ordered as an accessory (ordering number: 71377355)
and has been mounted. See Special Documentation SD02252F.

Operation via SmartBlue (app)

1 2 3
A0034939

 19 Operation via SmartBlue (app)


1 Transmitter power supply unit
2 Smartphone / tablet with SmartBlue (app)
3 Transmitter with Bluetooth module

Endress+Hauser 43
Operation options Micropilot FMR67 HART

8.1.4 Remote operation

Via HART protocol

2 3 6 8

1 4 5 7
A0036169

 20 Options for remote operation via HART protocol


1 PLC (Programmable Logic Controller)
2 Transmitter power supply unit, e.g. RN221N (with communication resistor)
3 Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475
4 Field Communicator 475
5 Computer with operating tool (e.g. DeviceCare/FieldCare , AMS Device Manager, SIMATIC PDM)
6 Commubox FXA191 (RS232) or FXA195 (USB)
7 Field Xpert SFX350/SFX370
8 VIATOR Bluetooth modem with connecting cable
9 Transmitter

DeviceCare/FieldCare via service interface (CDI)

1
A0032466

 21 DeviceCare/FieldCare via service interface (CDI)


1 Service interface (CDI) of the instrument (= Endress+Hauser Common Data Interface)
2 Commubox FXA291
3 Computer with DeviceCare/FieldCare operating tool

44 Endress+Hauser
Micropilot FMR67 HART Operation options

8.2 Structure and function of the operating menu

8.2.1 Structure of the operating menu

Menu Submenu / Meaning


parameter

Language 1) Defines the operating language of the on-


site display

Commissioning 2) Launches the interactive wizard for guided


commissioning.
Additional settings generally do not need
to be made in the other menus when the
wizard is finished.

Setup Parameter 1 Once values have been set for these


... parameters, the measurement should
Parameter N generally be completely configured.

Advanced setup Contains additional submenus and


parameters:
• to adapt the device to special measuring
conditions.
• to process the measured value (scaling,
linearization).
• to configure the signal output.

Diagnostics Diagnostic list Contains up to 5 currently active error


messages.

Event logbook 3) Contains the last 20 messages (which are


no longer active).

Device information Contains information for identifying the


device.

Measured values Contains all current measured values.

Data logging Contains the history of the individual


measuring values.

Simulation Is used to simulate measured values or


output values.

Device check Contains all parameters needed to check


the measurement capability of the device.

Heartbeat 4) Contains all the wizards for the Heartbeat


Verification and Heartbeat Monitoring
application packages.

Expert 5) System Contains all higher-order device


Contains all parameters of the device parameters that do not concern the
(including those that are already in one of measurement or measured value
the other menus). This menu is organized communication.
according to the function blocks of the
device. Sensor Contains all parameters needed to
configure the measurement.
The parameters of the Expert menu are
described in: Output • Contains all parameters needed to
GP01101F (HART) configure the current output.
• Contains all parameters needed to
configure the switch output (PFS).

Endress+Hauser 45
Operation options Micropilot FMR67 HART

Menu Submenu / Meaning


parameter

Communication Contains all parameters needed to


configure the digital communication
interface.

Diagnostics Contains all parameters needed to detect


and analyze operational errors.

1) If operating via operating tools (e.g. FieldCare), the "Language" parameter is located under "Setup →
Advanced setup → Display"
2) Only if operating via an FDT/DTM system
3) only available with local operation
4) only available if operating via DeviceCare or FieldCare
5) On entering the "Expert" menu, an access code is always requested. If a customer specific access code has
not been defined, "0000" has to be entered.

46 Endress+Hauser
Micropilot FMR67 HART Operation options

8.2.2 User roles and related access authorization


The two user roles Operator and Maintenance have different write access to the
parameters if a device-specific access code has been defined. This protects the device
configuration via the local display from unauthorized access →  47.

Access authorization to parameters


User role Read access Write access

Without access With access code Without access With access code
code code
(from the factory) (from the factory)

Operator    --

Maintenance    

If an incorrect access code is entered, the user obtains the access rights of the Operator
role.
The user role with which the user is currently logged on is indicated by the Access
status display parameter (for display operation) or Access status tooling parameter
(for tool operation).

8.2.3 Data access - Security

Write protection via access code


Using the device-specific access code, the parameters for the measuring device
configuration are write-protected and their values can no longer be changed via local
operation.

Define access code via local display


1. Navigate to: Setup → Advanced setup → Administration → Define access code
→ Define access code
2. Define a max. 4-digit numeric code as an access code.
3. Repeat the same code in Confirm access code parameter.
 The -symbol appears in front of all write-protected parameters.

Define access code via operating tool (e.g. FieldCare)


1. Navigate to: Setup → Advanced setup → Administration → Define access code
2. Define a max. 4-digit numeric code as an access code.
 Write protection is active.

Parameters that can always be changed


The write protection does not include certain parameters that do not affect the
measurement. Despite the defined access code, they can always be modified, even if the
other parameters are locked.
The device automatically locks the write-protected parameters again if a key is not pressed
for 10 minutes in the navigation and editing view. The device locks the write-protected
parameters automatically after 60 s if the user skips back to the measured value display
mode from the navigation and editing view.
• If write access is activated via access code, it can be also be deactivated only via the
access code →  48.
• In the "Description of Device Parameters" documents, each write-protected
parameter is identified with the -symbol.

Endress+Hauser 47
Operation options Micropilot FMR67 HART

Disabling write protection via access code


If the -symbol appears on the local display in front of a parameter, the parameter is
write-protected by a device-specific access code and its value cannot be changed at the
moment using the local display →  47.

The locking of the write access via local operation can be disabled by entering the device-
specific access code.
1. After you press , the input prompt for the access code appears.
2. Enter the access code.
 The -symbol in front of the parameters disappears; all previously write-
protected parameters are now re-enabled.

Deactivation of the write protection via access code

Via local display


1. Navigate to: Setup → Advanced setup → Administration → Define access code
→ Define access code
2. Enter 0000.
3. Repeat 0000 in Confirm access code parameter.
 The write protection is deactivated. Parameters can be changed without entering
an access code.

Via an operating tool (e.g. FieldCare)


1. Navigate to: Setup → Advanced setup → Administration → Define access code
2. Enter 0000.
 The write protection is deactivated. Parameters can be changed without entering
an access code.

Write protection via write protection switch


Unlike parameter write protection via a user-specific access code, this allows write access
to the entire operating menu - except for the "Contrast display" parameter - to be locked.
The parameter values are now read only and cannot be edited any more (exception
"Contrast display" parameter):
• Via local display
• Via service interface (CDI)
• Via HART protocol

48 Endress+Hauser
Micropilot FMR67 HART Operation options

ON

CDI WP OFF
XXXXXXXXXXXXX 4. WP

2. 3.

3 mm
1.
A0026157

1. Loosen the securing clamp.


2. Unscrew the electronics compartment cover.
3. Pull out the display module with a gentle rotational movement. To make it easier to
access the lock switch, attach the display module to the edge of the electronics
compartment.

A0036086

Endress+Hauser 49
Operation options Micropilot FMR67 HART

4. Setting the write protection switch (WP) on the main electronics module to the ON
position enables hardware write protection. Setting the write protection switch (WP)
on the main electronics module to the OFF position (factory setting) disables
hardware write protection.
 If the hardware write protection is enabled: The Hardware locked option is
displayed in the Locking status parameter. In addition, on the local display the
-symbol appears in front of the parameters in the header of the operational
display and in the navigation view.

XXXXXXXXX

20.50
XX XX

A0015870

If the hardware write protection is disabled: No option is displayed in the Locking


status parameter. On the local display, the -symbol disappears from in front of
the parameters in the header of the operational display and in the navigation
view.
5. Feed the cable into the gap between the housing and main electronics module and
plug the display module into the electronics compartment in the desired direction
until it engages.
6. Reverse the removal procedure to reassemble the transmitter.

50 Endress+Hauser
Micropilot FMR67 HART Operation options

Enabling and disabling the keypad lock


The keypad lock makes it possible to block access to the entire operating menu via local
operation. As a result, it is no longer possible to navigate through the operating menu or
change the values of individual parameters. Users can only read the measured values on
the operational display.
The keypad lock is switched on and off via the context menu.

Switching on the keypad lock


For the SD03 display only
The keypad lock is switched on automatically:
• If the device has not been operated via the display for > 1 minute.
• Each time the device is restarted.

To activate the keylock manually:


1. The device is in the measured value display.
Press  for at least 2 seconds.
 A context menu appears.
2. In the context menu select the Keylock on option.
 The keypad lock is switched on.

If the user attempts to access the operating menu while the keypad lock is active, the
message Keylock on appears.

Switching off the keypad lock


1. The keypad lock is switched on.
Press  for at least 2 seconds.
 A context menu appears.
2. In the context menu select the Keylock off option.
 The keypad lock is switched off.

Bluetooth® wireless technology


Signal transmission via Bluetooth® wireless technology uses a cryptographic
technique tested by the Fraunhofer Institute
• The device is not visible via Bluetooth® wireless technology without the SmartBlue app
• Only one point-to-point connection between one sensor and one smartphone or tablet is
established

Endress+Hauser 51
Operation options Micropilot FMR67 HART

8.3 Display and operating module

8.3.1 Display appearance

2
1 2.1
2.3
2.2
1.1 2.6 2.4
1.3 2.5

OPEN

OPEN
1.2 3
ESC
1.4
3.1
E
3.2

5 4
User 20
ABC_
0 1 2 3 4
DEFG HIJK
LMNO PQRS TUVW 5 6 7 8 9
XYZ Aa1

A0012635

 22 Appearance of the display and operation module for on-site operation


1 Measured value display (1 value max. size)
1.1 Header containing tag and error symbol (if an error is active)
1.2 Measured value symbols
1.3 Measured value
1.4 Unit
2 Measured value display (1 bargraph + 1 value)
2.1 Bargraph for measured value 1
2.2 Measured value 1 (including unit)
2.3 Measured value symbols for measured value 1
2.4 Measured value 2
2.5 Unit for measured value 2
2.6 Measured value symbols for measured value 2
3 Representation of a parameter (here: a parameter with selection list)
3.1 Header containing parameter name and error symbol (if an error is active)
3.2 Selection list;  marks the current parameter value.
4 Input matrix for numbers
5 Input matrix for alphanumeric and special characters

52 Endress+Hauser
Micropilot FMR67 HART Operation options

Display symbols for the submenus

Symbol Meaning

Display/operation
Is displayed:
• in the main menu next to the selection "Display/operation"
A0018367
• in the header, if you are in the "Display/operation" menu

Setup
Is displayed:
• in the main menu next to the selection "Setup"
A0018364
• in the header, if you are in the "Setup" menu

Expert
Is displayed:
• in the main menu next to the selection "Expert"
A0018365
• in the header, if you are in the "Expert" menu

Diagnostics
Is displayed:
• in the main menu next to the selection "Diagnostics"
A0018366
• in the header, if you are in the "Diagnostics" menu

Status signals

"Failure"
A0032902
A device error is present. The measured value is no longer valid.

"Function check"
A0032903
The device is in service mode (e.g. during a simulation).

"Out of specification"
The device is operated:
S • Outside of its technical specifications (e.g. during startup or a cleaning)
A0032904
• Outside of the configuration carried out by the user (e.g. level outside configured
span)

"Maintenance required"
A0032905
Maintenance is required. The measured value is still valid.

Display symbols for the locking state

Symbol Meaning

Display parameter
Marks display-only parameters which can not be edited.
A0013148

Device locked
• In front of a parameter name: The device is locked via software and/or hardware.
A0013150
• In the header of the measured value screen: The device is locked via hardware.

Endress+Hauser 53
Operation options Micropilot FMR67 HART

Measured value symbols

Symbol Meaning

Measured values

Level

A0032892

Distance
A0032893

Current output
A0032908

Measured current

A0032894

Terminal voltage

A0032895

Temperature of the electronics or the sensor

A0032896

Measuring channels

Measuring channel 1
1
A0032897

Measuring channel 2
2
A0032898

Status of the measured value

Status "Alarm"
The measurment is interrupted. The output assumes the defined alarm value. A
A0018361 diagnostic message is generated.

Status "Warning"
The device continues measuring. A diagnostic message is generated.
A0018360

54 Endress+Hauser
Micropilot FMR67 HART Operation options

8.3.2 Operating elements

Key Meaning

Minus key
For menu, submenu
Moves the selection bar upwards in a picklist.
A0018330
For text and numeric editor
In the input mask, moves the selection bar to the left (backwards).

Plus key
For menu, submenu
Moves the selection bar downwards in a picklist.
A0018329
For text and numeric editor
In the input mask, moves the selection bar to the right (forwards).

Enter key
For measured value display
• Pressing the key briefly opens the operating menu.
• Pressing the key for 2 s opens the context menu.
For menu, submenu
• Pressing the key briefly
Opens the selected menu, submenu or parameter.
A0018328
• Pressing the key for 2 s for parameter:
If present, opens the help text for the function of the parameter.
For text and numeric editor
• Pressing the key briefly
– Opens the selected group.
– Carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.

Escape key combination (press keys simultaneously)


For menu, submenu
• Pressing the key briefly
+ – Exits the current menu level and takes you to the next higher level.
– If help text is open, closes the help text of the parameter.
A0032909
• Pressing the key for 2 s returns you to the measured value display ("home position").
For text and numeric editor
Closes the text or numeric editor without applying changes.

Minus/Enter key combination (press and hold down the keys simultaneously)
+
A0032910
Reduces the contrast (brighter setting).

Plus/Enter key combination (press and hold down the keys simultaneously)
+
A0032911
Increases the contrast (darker setting).

Endress+Hauser 55
Operation options Micropilot FMR67 HART

8.3.3 Entering numbers and text

Numeric editor Text editor

1 1

2 20 2 User
0 1 2 3 4 ABC_ DEFG HIJK
5 6 7 8 9 LMNO PQRS TUVW
3 3
XYZ Aa1

4 4

A0013941 A0013999

1 Editing view
2 Display area of the entered values
3 Input mask
4 Operating elements

Input mask
The following input symbols are available in the input mask of the numeric and text editor:

Numeric editor symbols


Symbol Meaning

Selection of numbers from 0 to 9.


0

9
A0013998

. Inserts decimal separator at the input position.

A0016619

Inserts minus sign at the input position.



A0016620

Confirms selection.

A0013985

Moves the input position one position to the left.

A0016621

Exits the input without applying the changes.

A0013986

Clears all entered characters.

A0014040

Text editor symbols


Symbol Meaning

Selection of letters from A to Z


ABC _

XYZ
A0013997

Toggle
Aa1 • Between upper-case and lower-case letters
• For entering numbers
A0013981
• For entering special characters

56 Endress+Hauser
Micropilot FMR67 HART Operation options

Confirms selection.

A0013985

Switches to the selection of the correction tools.

A0013987

Exits the input without applying the changes.

A0013986

Clears all entered characters.

A0014040

Correction symbols under


Symbol Meaning

Clears all entered characters.

A0032907

Moves the input position one position to the right.

A0018324

Moves the input position one position to the left.

A0018326

Deletes one character immediately to the left of the input position.

A0032906

Endress+Hauser 57
Operation options Micropilot FMR67 HART

8.3.4 Opening the context menu


Using the context menu, the user can call up the following menus quickly and directly from
the operational display:
• Setup
• Conf. backup disp.
• Env.curve
• Keylock on

Opening and closing the context menu


The user is in the operational display.
1. Press  for 2 s.
 The context menu opens.

XXXXXXXXXX
Setup
Conf.backup disp
Env. curve 50
Keylock on kg/h
A0033110-EN

2. Press  +  simultaneously.
 The context menu is closed and the operational display appears.

Calling up the menu via the context menu


1. Open the context menu.
2. Press  to navigate to the desired menu.
3. Press  to confirm the selection.
 The selected menu opens.

58 Endress+Hauser
Micropilot FMR67 HART Operation options

8.3.5 Envelope curve on the display and operating module


In order to assess the measuring signal, the envelope curve and - if a mapping has been
recorded - the mapping curve can be displayed:

E (2s)

2x +

E -

+ + -
+
+

E
+ + -

A0014277

Endress+Hauser 59
System integration via HART protocol Micropilot FMR67 HART

9 System integration via HART protocol

9.1 Overview of the Device Description files (DD)


Manufacturer ID 17 (0x11)

Device type 0x112B

HART specification 7.0

DD files For information and files see:


• www.endress.com
• www.hartcomm.org

9.2 Measured values via HART protocol


On delivery the following measruing values are assigned to the HART device varaibles:

Device variable Measuring value

Primary variable (PV) Level linearized

Secondary variable (SV) Distance

Tertiary variable (TV) Absolute echo amplitude

Quaternary variable (QV) Relative echo amplitude

The allocation of the device variables can be changed in the operating menu:
Expert → Communication → Output
In a HART multidrop loog only one device may use the output current for signal
transmission. For all other devices the follwing must be set:
• "Current span" parameter = "Fixed current" option
• "Fixed current" parameter = 4 mA

60 Endress+Hauser
Micropilot FMR67 HART Commissioning via SmartBlue (app)

10 Commissioning via SmartBlue (app)

10.1 Requirements
Device requirements
Commissioning via SmartBlue is only possible if the device has a Bluetooth module.
System requirements SmartBlue
SmartBlue is available as download for Android devices from the Google Play Store and for
iOS devices from the iTunes Store.
• iOS devices:
iPhone 4S or higher from iOS9.0; iPad2 or higher from iOS9.0; iPod Touch 5th
generation or higher from iOS9.0
• Devices with Android:
from Android 4.4 KitKat and Bluetooth® 4.0
Initial password
The ID of the Bluetooth module serves as the initial password used to establish the first
connection to the device. It can be found:
• on the information sheet which is supplied with the device. This serial number specific
sheet is also stored in W@M.
• on the nameplate of the Bluetooth module.

2
A0036790

 23 Device with Bluetooth module


1 Electronics housing of the device
2 Nameplate of the Bluetooth module; the ID on this nameplate serves as initial password.

All login data (including the password changed by the user) are not stored in the
device but in the Bluetooth module. This must be taken into account when the module
is removed from one device and inserted into a different device.

10.2 Commissioning
Download and install SmartBlue
1. To download the app, scan the QR code or enter "SmartBlue" in the search field

A0033202

 24 Download link

Endress+Hauser 61
Commissioning via SmartBlue (app) Micropilot FMR67 HART

2. Start SmartBlue

A0029747

 25 SmartBlue pictogram

3. Select device from displayed livelist (available devices only)


A0029502

 26 Livelist

Only one point-to-point connection can be established between one sensor and one
smartphone or tablet.

4. Perform login

admin

A0029503

 27 Login

5. Enter user name -> admin


6. Enter initial password -> ID of the Bluetooth module
7. Change the password after logging in for the first time

62 Endress+Hauser
Micropilot FMR67 HART Commissioning via SmartBlue (app)

8. By wiping from the side, additional information (e.g. main menu) can be dragged
into the image

A0029504

 28 Main menu

Envelope curves can be displayed and recorded


Additionally to the envelope curve, the following values are displayed:
• D = Distance
• L = Level
• A = Absolute amplitude
• In the case of screenshots, the displayed section (zoom function) is saved
• In video sequences, always the whole area without zoom function is saved
It is also possible to send envelope curves (video sequences) using the relevant
smartphone or tablet functions.

2
1 3

5
A0029486

 29 Envelope curve display (example) in SmartBlue; Android view


1 Record video
2 Create screenshot
3 Navigation to mapping menu
4 Start / stop video recording
5 Move time on time axis

Endress+Hauser 63
Commissioning via SmartBlue (app) Micropilot FMR67 HART

2
1 3

5
A0029487

 30 Envelope curve display (example) in SmartBlue; IoS view


1 Record video
2 Create screenshot
3 Navigation to mapping menu
4 Start / stop video recording
5 Move time on time axis

64 Endress+Hauser
Micropilot FMR67 HART Commissioning via wizard

11 Commissioning via wizard


A wizard guiding the user through the initial setup is available in FieldCare and
DeviceCare 4).

1. Connect the device to FieldCare or DeviceCare →  44.


2. Open the device in FieldCare or DeviceCare.
 The dashboard (home page) of the device appears:

A0027720

1 "Commissioning" button calls up the wizard.

3. Click on "Commissioning" to call up the wizard.


4. Enter or select the appropriate value for each parameter. These values are
immediately written to the device.
5. Click "Next" to switch to the next page.
6. After finishing the last page, click "End of sequence" to close the wizard.

If the wizard is cancelled before all necessary parameters have been set, the device
may be in an undefined state. A reset to the default settings is recommended in this
case.

4) DeviceCare is available for download at www.software-products.endress.com. The download requires a registration in the Endress+Hauser
software portal.

Endress+Hauser 65
Commissioning via operating menu Micropilot FMR67 HART

12 Commissioning via operating menu

12.1 Installation and function check


Make sure that all final checks have been completed before you start up your measuring
point:
• Checklist "Post-installation check" →  30
• Checklist "Post-connection check" →  40

12.2 Setting the operating language


Factory setting: English or ordered local language

XXXXXXXXX

20.50
XX XX
Main menu 0104-1
1. Display language
English
Operation
Setup
Display language 0104-1
2. Ã English
Deutsch
Español
Français
Display language 0104-1
3. Ã English
Deutsch
Español
Français
Hauptmenü 0104-1
4. Sprache
Deutsch
Betrieb
Setup

A0029420

 31 Using the example of the local display

66 Endress+Hauser
Micropilot FMR67 HART Commissioning via operating menu

12.3 Configuration of a level measurement

R
BD
100%

D
E F

L
0%

A0016934

 32 Configuration parameters for level measurements in bulk solids


R Reference point of the measurement
D Distance
L Level
E Empty calibration (= zero)
F Full calibration (= span)

1. Setup → Device tag


 Enter device tag.
2. Setup → Distance unit
 Select distance unit.
3. Setup → Bin type
 Select bin type.
4. Setup → Max. filling speed solid
 Enter maximum expected filling speed.
5. Setup → Max. draining speed solid
 Enter maximum expected draining speed.
6. Setup → Empty calibration
 Enter empty distance E (Distance from reference point R to the 0% level)
7. If the measuring range covers only an upper part of the tank or silo (E is much less-
than the tank/silo height), it is mandatory to enter the acutal tank or silo height into
the parameter. If there is an outlet cone, the tank or silo height should not be
adjusted as usually E is not much less-than the tank/silo height in these applications.
Setup → Advanced setup → Level → Tank/silo height
8. Setup → Full calibration
 Enter full distance F (Distance from the 0% to the 100% level).
9. Setup → Level
 Indicates the measrued level L.
10. Setup → Distance
 Indicates the measured distance from the reference point R to the level L.

Endress+Hauser 67
Commissioning via operating menu Micropilot FMR67 HART

11. Setup → Signal quality


 Indicates the quality of the evaluated level echo.
12. Setup → Mapping → Confirm distance
 Compare distance indicated on the display to real distance in order to start the
recording of an interference echo map.
13. Setup → Advanced setup → Level → Level unit
 Select level unit: %, m, mm, ft, in (Factory setting: %)

It is strongly recommended to adjust the maximum filling and draining speed to the
actual process.

68 Endress+Hauser
Micropilot FMR67 HART Commissioning via operating menu

12.4 Recording the reference curve


After the configuration of the measurement it is recommended to record the current
envelope curve as a reference curve. The reference curve can be used later on in the
process for diagnostic purposes. To record the reference curve use the Save reference
curve parameter.
Navigation in the menu
Expert → Diagnostics → Envelope diagnostics → Save reference curve
Meaning of the options
• No
No action
• Yes
The current envelope curve is saved as reference curve.
The reference curve can only be displayed in the envelope curve diagram of FieldCare
after it has been loaded from the device into FieldCare. This is performed by the "Load
Reference Curve" function in FieldCare.

 33 The "Load Reference Curve" function

Endress+Hauser 69
Commissioning via operating menu Micropilot FMR67 HART

12.5 Configuration of the on-site display

12.5.1 Factory settings of the on-site display

Parameter Factory setting

Language English

Value 1 display Level linearized

Value 2 display None

Value 3 display None

Value 4 display None

12.5.2 Adjustment of the on-site display


The on-site display can be adjusted in the following submenu:
Setup → Advanced setup → Display

12.6 Configuration of the current outputs

12.6.1 Factory setting of the current outputs

Current output Allocated measuring 4 mA value 20 mA value


vlaue

1 Level linearized 0% or the corresponding 100% or the corresponding


linearized value linearized value

2 1) Distance 0 Empty calibration

1) for devices with 2 current outputs

12.6.2 Adjustment of the current outputs


The current outputs can be adjusted in the following menus:
Basic settings
Setup → Advanced setup → Current output 1 to 2
Advanced settings
Expert → Output → Current output 1
See "Description of Device Parameters", GP01101F

70 Endress+Hauser
Micropilot FMR67 HART Commissioning via operating menu

12.7 Configuration management


After commissioning, you can save the current device configuration, copy it to another
measuring point or restore the previous device configuration. You can do so using the
Configuration management parameter and its options.
Navigation path in the operating menu
Setup → Advanced setup → Configuration backup display → Configuration management
Meaning of the options
• Cancel
No action is executed and the user exits the parameter.
• Execute backup
A backup copy of the current device configuration in the HistoROM (built-in in the
device) is saved to the display module of the device. The backup copy comprises the
transmitter and sensor data of the device.
• Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device. The backup copy comprises the transmitter and sensor data of
the device.
• Duplicate
The transmitter configuration is duplicated to another device using the transmitter
display module. The following parameters, which characterize the individual measuring
point are not included in the transmitted configuration:
– HART date code
– HART short tag
– HART message
– HART descriptor
– HART address
– Device tag
– Medium type
• Compare
The device configuration saved in the display module is compared to the current device
configuration of the HistoROM. The result of this comparison is displayed in the
Comparison result parameter angezeigt.
• Clear backup data
The backup copy of the device configuration is deleted from the display module of the
device.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
If an existing backup is restored to a different device using the Restore option, it may
occur that some device functionalities are no longer available. In some cases even a
device reset will not restore the original status.
In order to transmit a configuration to a different device, the Duplicate option should
always be used.

Endress+Hauser 71
Commissioning via operating menu Micropilot FMR67 HART

12.8 Protection of the settings against unauthorized


changes
There are two ways to protect the settings against unauthorized changes:
• Via parameter settings (software locking) →  47
• Via locking switch (hardware locking) →  48

72 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

13 Diagnostics and troubleshooting

13.1 General trouble shooting

13.1.1 General errors

Error Possible cause Remedial action

Device does not respond. Supply voltage does not match Connect the correct voltage.
the value indicated on the
nameplate.

The polarity of the supply Correct the polarity.


voltage is wrong.

The cables do not contact the Ensure electrical contact between the cable
terminals properly. and the terminal.

Values on the display invisible Contrast setting is too weak or • Increase contrast by pressing  and 
too strong. simultaneously.
• Decrease contrast by pressing  and 
simultaneously.

The plug of the display cable is Connect the plug correctly.


not connected correctly.

Display is defective. Replace display.

"Communication error" is Electromagnetic interference Check grounding of the device.


indicated on the display when
starting the device or Broken display cable or display Replace display.
connecting the display plug.

Duplicating of the parameters Display with backup is not Connect display (with the backup) and
from one device to another via recognized if no data backup restart the device.
the display doesn't work. has been performed at the
Only the "Save" and "Abort" device before.
options are available.

Output current <3.6 mA Signal cable connection Check connection.


incorrect.

Electronics is defective. Replace electronics.

HART communication does not Communication resistor Install the communication resistor(250 Ω)
function. missing or incorrectly installed. correctly.

Commubox connected Connect Commubox correctly.


incorrectly.

Commubox not switched to Set the selection switch of the Commubox to


HART mode. the HART position.

CDI communication does not Wrong setting of the COM port Check the setting of the COM port on the
work. on the computer. computer and change it if necessary.

Device measures incorrectly. Parametrization error Check and adjust parameterization.

Device not accessible via No Bluetooth connection Enable Bluetooth function onsmartphone or
SmartBlue tablet.

Device already linked to Disconnect device from smartphone/tablet.


another smartphone / tablet

Bluetooth module not Connect Bluetooth module (see SD02252F).


connected.

Login via SmartBlue not Device is being put into Enter initial password (ID of the Bluetooth
possible operation for the first time module) and change.

Device cannot be operated via Incorrect password entered Enter correct password
SmartBlue
Password forgotten Contact Endress+Hauser Service
(www.addresses.endress.com)

Endress+Hauser 73
Diagnostics and troubleshooting Micropilot FMR67 HART

13.1.2 Error - SmartBlue operation

Error Possible cause Solution

Device is not visible in the No Bluetooth Enable Bluetooth® function on smartphone or tablet
live list connection
Bluetooth® function of sensor disabled, perform
recovery sequence

Device is not visible in the The device is already Only one point-to-point connection is established
live list connected with between a sensor and a smartphone or tablet
another smartphone/
tablet

Device is visible in the live list Android end device Is the location function enabled for the app, was it
but cannot be accessed via approved the first time?
SmartBlue
GPS or positioning function must be activated for
certain Android versions in conjunction with
Bluetooth®

Activate GPS - close the app fully and restart - enable


the positioning function for the app

Device is visible in the live list Apple end device Log in as standard
but cannot be accessed via Enter user name "admin"
SmartBlue Enter initial password (ID of the Bluetooth module)
paying attention to lower/upper case

Login via SmartBlue not Device is being put Enter initial password (ID of the Bluetooth module) and
possible into operation for the change; paying attention to lower/upper case
first time

Device cannot be operated via Incorrect password Enter correct password


SmartBlue entered

Device cannot be operated via Password forgotten Contact the Endress+Hauser Service department
SmartBlue (www.addresses.endress.com)

13.1.3 Parametrization errors

Error Possible cause Remedy

Measured value If measured distance (Setup • Check and adjust Empty calibration parameter
incorrect → Distance) matches the real if necessary.
distance: • heck and adjust Full calibration parameter if
Calibration error necessary.
• Check and adjust linearization if necessary
(Linearization submenu).

Level correction set incorrectly Enter correct value in Level correction parameter.

If measured distance (Setup Carry out tank mapping (Confirm distance


→ Distance) does not match parameter).
the real distance:
Interference echo

No change of measured Interference echo from • Carry out tank mapping (Confirm distance
value on filling / installations, nozzle or build- parameter).
emptying up on the antenna. • If possible, align antenna in the direction of the
solid surface in order to prevent interference
echoes.
• If necessary, clean antenna (purge air).
• If necessary, select a better mounting position
and/or larger antenna.

74 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

Error Possible cause Remedy

During filling/emptying Signal is weakened (e.g. by • Carry out tank mapping (Confirm distance
or measurement, the fluidisation of the surface, parameter).
measured value jumps extreme dust formation) - the • Increase integration time (Expert → Sensor
sporadically to a higher interference echoes are → Distance → Integration time)
level. sometimes stronger. Strong • Optimize alignment of antenna
build-up, filling stream in path • If necessary, select a better mounting position
of beam. and/or larger antenna.
• If necessary, clean antenna (purge air).

Error message F941 or Level echo is too weak. • Optimize alignment of antenna
S941 "Echo lost" • If necessary, select a better mounting position
Possible causes:
and/or larger antenna.
• Fluidisation of the surface
• Extreme dust formation
• Angle or repose

Measured values jumps • Build-up at container • Periodic cleaning


to higher levels and • Build-up at antenna • Carry out tank mapping (Confirm distance
remains there. • Strong condensate parameter).
formation at antenna • Increase integration time (Expert → Sensor
→ Distance → Integration time)
• Optimize alignment of antenna
• If necessary, select a better mounting position
and/or larger antenna.

Device displays a level Interference echo Carry out mapping over entire measuring range
when the tank is empty. when the silo is empty (Confirm distance
parameter).

Wrong slope of the level Bin property or process • Select the correct option in Bin type parameter.
throughout the complete propertis incorrect • Enter the actual values in "Max. filling speed
measuring range solid" parameter and "Max. draining speed
solid" parameter.

Endress+Hauser 75
Diagnostics and troubleshooting Micropilot FMR67 HART

13.2 Diagnostic information on local display

13.2.1 Diagnostic message


Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the measured value display.

Measured value display in alarm condition Diagnostic message

2 1

XXXXXXXXX S XXXXXXXXX S

20.50 S801
Supply voltage
3
4

X XX i Menu

5
A0029426-EN

1 Status signal
2 Status symbol (symbol for event level))
3 Status symbol with diagnostics event
4 Event text
5 Operating elements

Status signals

"Failure (F)" option


A0032902
A device error is present. The measured value is no longer valid.

"Function check (C)" option


A0032903
The device is in service mode (e.g. during a simulation).

"Out of specification (S)" option


S The device is operated:
A0032904
• Outside of its technical specifications (e.g. during startup or a cleaning)
• Outside of the configuration carried out by the user (e.g. level outside configured span)

"Maintenance required (M)" option


A0032905
Maintenance is required. The measured value is still valid.

Status symbol (symbol for event level)

"Alarm" status
 The measurement is interrupted. The signal outputs take on the defined alarm condition.
A diagnostic message is generated.

"Warning" status

The device continues to measure. A diagnostic message is generated.

76 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

Diagnostics event and event text


The fault can be identified using the diagnostics event. The event text helps you by
providing information about the fault. In addition, the corresponding symbol is displayed
before the diagnostics event.

Diagnostics event

Status symbol Status signal Event number Event text

↓ ↓ ↓

Example  S 441 Curr.output 1


A0032904

3-digit number

If two or more diagnostic messages are pending simultaneously, only the message with the
highest priority is shown. Additional pending diagnostic messages can be shown in the
Diagnostic list submenu.
Past diagnostic messages that are no longer pending are shown as follows:
• On the local display:
in the Event logbook submenu
• In FieldCare:
via the "Event List /HistoROM" function.

Operating elements

Operating functions in menu, submenu

Plus key

Opens the message about the remedial measures.

Enter key

Opens the operating menu.

Endress+Hauser 77
Diagnostics and troubleshooting Micropilot FMR67 HART

13.2.2 Calling up remedial measures

XXXXXXXXX S XXXXXXXXX S

20.50 S801
Supply voltage

X XX i Menu

1.

Diagnostic list S
Diagnostics 1
1 S801 Supply voltage
Diagnostics 2
Diagnostics 3

2.

2 Supply voltage (ID:203) 3


4 S801 0d00h02m25s 5
Increase supply voltage
6

3.

A0029431-EN

 34 Message for remedial measures


1 Diagnostic information
2 Short text
3 Service ID
4 Diagnostic behavior with diagnostic code
5 Operation time of occurrence
6 Remedial measures

The user is in the diagnostic message.


1. Press  (-Symbol).
 Diagnostic list submenu opens.
2. Select the desired diagnostic event with  or  and press .
 The message for the remedial measures for the selected diagnostic event opens.
3. Press  +  simultaneously.
 The message for the remedial measures closes.

The user is in the Diagnostics menu at an entry for a diagnostics event, e.g. in Diagnostic
list submenu or in Previous diagnostics.
1. Press .
 The message for the remedial measures for the selected diagnostic event opens.
2. Press  +  simultaneously.
 The message for the remedial measures closes.

78 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

13.3 Diagnostic event in the operating tool


If a diagnostic event is present in the device, the status signal appears in the top left status
in the operating tool along with the corresponding symbol for event level in accordance
with NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)

A: Via the operating menu


1. Navigate to the Diagnostics menu.
 In the Actual diagnostics parameter, the diagnostic event is shown with event
text.
2. On the right in the display range, hover the cursor over the Actual diagnostics
parameter.

A tool tip with remedial measures for the diagnostic event appears.

B: Via the "Create documentation" function


1.

Select the "Create documentation" function.


2.

Make sure "Data overview" is marked.

Endress+Hauser 79
Diagnostics and troubleshooting Micropilot FMR67 HART

3. Click "Save as ..." and save a PDF of the protocol.


 The protocol contains the diagnostic messages and remedy information.

C: Via the "Eventlist / Extended HistoROM" function


1.

Select the "Eventlist / Extended HistoROM" function.


2.

Select the "Load Eventlist" function.


 The list of events, including remedy information, is shown in the "Data overview"
window.

13.4 Diagnostic list


In the Diagnostic list submenu submenu, up to 5 currently pending diagnostic messages
can be displayed. If more than 5 messages are pending, the messages with the highest
priority are shown on the display.
Navigation path
Diagnostics → Diagnostic list

Calling up and closing the remedial measures


1. Press .
 The message for the remedial measures for the selected diagnostic event opens.
2. Press  +  simultaneously.
 The message about the remedial measures closes.

80 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

13.5 Overview of diagnostic events


Diagnostic Short text Remedy instructions Status Diagnostic
number signal behavior
[from the [from the
factory] factory]

Diagnostic of sensor

046 Build-up detected Clean sensor F Alarm 1)

102 Sensor incompatible 1. Restart device F Alarm


error 2. Contact service

151 Sensor electronic Replace sensor electronic module F Alarm


failure

Diagnostic of electronic

242 Software incompatible 1. Check software F Alarm


2. Flash or change main
electronics module

252 Modules incompatible 1. Check electronic modules F Alarm


2. Change I/O or main electronic
module

261 Electronic modules 1. Restart device F Alarm


2. Check electronic modules
3. Change I/O Modul or main
electronics

262 Module connection 1. Check module connections F Alarm


2. Change electronic modules

270 Main electronic failure Change main electronic module F Alarm

271 Main electronic failure 1. Restart device F Alarm


2. Change main electronic module

272 Main electronic failure 1. Restart device F Alarm


2. Contact service

273 Main electronic failure 1. Emergency operation via display F Alarm


2. Change main electronics

275 I/O module defective Change I/O module F Alarm

276 I/O module faulty 1. Restart device F Alarm


2. Change I/O module
276 I/O module failure F Alarm

282 Data storage 1. Restart device F Alarm


2. Contact service

283 Memory content 1. Transfer data or reset device F Alarm


2. Contact service

311 Electronic failure Maintenance required! M Warning


1. Do not perform reset
2. Contact service

Diagnostic of configuration

410 Data transfer 1. Check connection F Alarm


2. Retry data transfer

411 Up-/download active Up-/download active, please wait C Warning

412 Processing download Download active, please wait C Warning

431 Trim 1 to 2 Carry out trim C Warning

435 Linearization Check linearization table F Alarm

437 Configuration 1. Restart device F Alarm


incompatible 2. Contact service

Endress+Hauser 81
Diagnostics and troubleshooting Micropilot FMR67 HART

Diagnostic Short text Remedy instructions Status Diagnostic


number signal behavior
[from the [from the
factory] factory]

438 Dataset 1. Check data set file M Warning


2. Check device configuration
3. Up- and download new
configuration

441 Current output 1 to 2 1. Check process S Warning


2. Check current output settings

484 Failure mode Deactivate simulation C Alarm


simulation

485 Simulation measured Deactivate simulation C Warning


value

491 Current output 1 to 2 Deactivate simulation C Warning


simulation

494 Switch output Deactivate simulation switch C Warning


simulation output

495 Diagnostic event Deactivate simulation C Warning


simulation

585 Simulation distance Deactivate simulation C Warning

586 Record map Recording of mapping C Warning


please wait

Diagnostic of process

801 Energy too low Increase supply voltage S Warning

803 Current loop 1. Check wiring F Alarm


2. Change I/O module

825 Operating temperature 1. Check ambient temperature S Warning


2. Check process temperature
825 Operating temperature F Alarm

921 Change of reference 1. Check reference configuration S Warning


2. Check pressure
3. Check sensor

941 Echo lost Check parameter 'DC value' S Warning 1)

942 In safety distance 1. Check level S Alarm 1)


2. Check safety distance
3. Reset self holding

943 In blocking distance Reduced accuracy S Warning


Check level

950 Advanced diagnostic Maintain your diagnostic event M Warning 1)


1 to 4 occured

952 Foam detected Check process conditions F Alarm 1)

1) Diagnostic behavior can be changed.

13.6 Event logbook

13.6.1 Event history


A chronological overview of the event messages that have occurred is provided in the
Event list submenu 5).

5) This submenu is only available for operation via local display. In the case of operation via FieldCare, the event list can be displayed with the "Event
List / HistoROM" functionality of FieldCare.

82 Endress+Hauser
Micropilot FMR67 HART Diagnostics and troubleshooting

Navigation path
Diagnostics → Event logbook → Event list
A maximum of 100 event messages can be displayed in chronological order.
Die Ereignishistorie umfasst Einträge zu:
• Diagnostic events
• Information events
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
– : Event has occurred
– : Event has ended
• Information event
: Event has occurred

Calling up and closing the remedial measures


1. Press 
 The message for the remedial measures for the selected diagnostic event opens.
2. Press  +  simultaneously.
 The message about the remedial measures closes.

13.6.2 Filtering the event logbook


Using the Filter options parameter, you can define which category of event messages is
displayed in the Event list submenu.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information

13.6.3 Overview of information events

Info number Info name

I1000 --------(Device ok)

I1089 Power on

I1090 Configuration reset

I1091 Configuration changed

I1092 Embedded HistoROM deleted

I1110 Write protection switch changed

I1137 Electronic changed

I1151 History reset

I1154 Reset terminal voltage min/max

I1155 Reset electronic temperature

I1156 Memory error trend

I1157 Memory error event list

Endress+Hauser 83
Diagnostics and troubleshooting Micropilot FMR67 HART

Info number Info name

I1184 Display connected

I1185 Display backup done

I1186 Restore via display done

I1187 Settings downloaded with display

I1188 Display data cleared

I1189 Backup compared

I1256 Display: access status changed

I1264 Safety sequence aborted

I1335 Firmware changed

I1397 Fieldbus: access status changed

I1398 CDI: access status changed

I1512 Download started

I1513 Download finished

I1514 Upload started

I1515 Upload finished

I1554 Safety sequence started

I1555 Safety sequence confirmed

I1556 Safety mode off

13.7 Firmware history


Date Firmware Modifications Documentation (FMR67, HART)
version
Operating Instructions Description of Parameters
1)
01.2017 01.00.zz Original software BA01620F/00/EN/01.17 GP01101F/00/EN/01.17
BA01620F/00/EN/02.18 2)

1) Contains information on the Heartbeat wizards which are available in the latest DTM version for
DeviceCare and FieldCare.
2) Contains information on the Bluetooth interface.

The firmware version can explicitly be ordered via the product structure. In this way it
is possible to ensure compatibility of the firmware version with an existing or planned
system integration.

84 Endress+Hauser
Micropilot FMR67 HART Maintenance

14 Maintenance
The measuring device requires no special maintenance.

14.1 Exterior cleaning


When exterior-cleaning the device, always use cleaning agents that do not attack the
surface of the hosuing and the seals.

14.2 Replacing seals


The process seals of the sensors (at the process connection) must be replaced periodically,
particularly if molded seals (aseptic construction) are used. The period between changes
depends on the frequency of cleaning cycles and on the temperature of the measured
substance and the cleaning temperature.

Endress+Hauser 85
Repairs Micropilot FMR67 HART

15 Repairs

15.1 General information on repairs

15.1.1 Repair concept


The Endress+Hauser repair concept assumes that the devices have a modular design and
that repairs can be done by the Endress+Hauser service or specially trained customers.
Spare parts are contained in suitable kits. They contain the related replacement
instructions.
For more information on service and spare parts, contact the Service Department at
Endress+Hauser.

15.1.2 Repairs to Ex-approved devices


When carrying out repairs to Ex-approved devices, please note the following:
• Repairs to Ex-approved devices may only be carried out by trained personnel or by the
Endress+Hauser Service.
• Comply with the prevailing standards, national Ex-area regulations, safety instructions
(XA) and certificates.
• Only use original spare parts from Endress+Hauser.
• When ordering a spare part, please note the device designation on the nameplate. Only
replace parts with identical parts.
• Carry out repairs according to the instructions. On completion of repairs, carry out the
specified routine test on the device.
• Only Endress+Hauser Service may convert a certified device into a different certified
variant.
• Document all repair work and conversions.

15.1.3 Replacement of an electronics module


If an electronics module has been replaced, it is not necessary to perform a new basic setup
as the calibration parameters are stored in the HistoROM which is located in the housing.
However, after exchanging the main electronics module it may be necessary to record a
new mapping (interference echo suppression).

15.1.4 Replacement of a device


After a complete device or electronic module has been replaced, the parameters can be
downloaded into the instrument again in one of the following ways:
• Via the display module
Condition: The configuration of the old device has been saved in the display module
→  162.
• Via FieldCare
Condition: The configuration of the old device has been saved to the computer via
FieldCare.
You can continue to measure without carrying out a new setup. Only a linearization and a
tank map (interference echo suppression) have to be recorded again.

86 Endress+Hauser
Micropilot FMR67 HART Repairs

15.2 Spare parts


• A few interchangeable measuring device components are identified by a spare part
nameplate. This contains information about the spare part.
• The connection compartment cover of the device contains a spare part nameplate that
includes the following information:
– A list of the most important spare parts for the measuring device, including their
ordering information.
– The URL for the W@M Device Viewer (www.endress.com/deviceviewer):
There, all spare parts for the measuring device are listed, including the order code, and
can be ordered. If available, the corresponding Installation Instructions can also be
downloaded there.

Spare parts for: Levelflex FMP5x


Ser.-no.: 8A21AC098AF4
Pos. no. Spare part Spare no./structure
1 Sensor XPF0002- / XPF0003-
2 Main electr. XPF0001-
3 IO module 71023451
4 Cover (front) 71023451
5 Cover (right) 71023451
6 Display 71023451

002847-A
6 3
4 1 5
Additional information: www.endress.com/deviceviewer

A0014979

 35 Example for spare part nameplate in connection compartment cover

Measuring device serial number:


• Is located on the device and spare part nameplate.
• Can be read out via the "Serial number" parameter in the "Device information"
submenu.

15.3 Return
The measuring device must be returned if it is need of repair or a factory calibration, or if
the wrong measuring device has been delivered or ordered. Legal specifications require
Endress+Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.
To ensure safe, swift and professional device returns, please refer to the procedure and
conditions for returning devices provided on the Endress+Hauser website at
http://www.endress.com/support/return-material

15.4 Disposal
Observe the following notes during disposal:
• Observe valid federal/national regulations.
• Ensure proper separation and reuse of the device components.

Endress+Hauser 87
Accessories Micropilot FMR67 HART

16 Accessories

16.1 Device-specific accessories

16.1.1 Weather protection cover

Accessory Description

Weather protection cover

222 (8.74)

A0015466

300 (11.8) 275 (10.8)


255.1 (10) 164 (6.46) 54
(2
.13
)
55
5)
(1.

4)
38

35°
65°
(6.1
156

A0015472

 36 Weather protection cover; Dimensions: mm (in)

 Enclosed",
The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory
option PB "Weather Protection Cover").
Alternatively, it can be separately ordered as an accessory; order code 71162242.

88 Endress+Hauser
Micropilot FMR67 HART Accessories

16.1.2 Adjustable flange seal

Accessories Description

Adjustable
flange seal

4° ød
øD
1

2
h

±8°
hmax

hmin
A0032292

1 UNI slip-on flange


2 Adjustable flange seal
3 Nozzle

 (temperature,
The material properties and process conditions of the adjustable flange seal must be compatible with the properties
pressure, resistance) of the process.

 enclosed", options PL, PM, PN, PO, PQ, PR).


The adjustable flange seal can also be ordered directly with the device (product structure: feature 620 "Accessories

Technical data: version DN/JIS

Order number 71074263 71074264 71074265

Compatible with DN80 PN10/40 DN100 PN10/16 • DN150 PN10/16


• JIS 10K 150A

Recommended screw length 100 mm (3.9 in) 100 mm (3.9 in) 110 mm (4.3 in)

Recommended screw size M14 M14 M18

Material EPDM

Process pressure –0.1 to 0.1 bar (–1.45 to 1.45 psi)

Process temperature –40 to +80 °C (–40 to +176 °F)

D 142 mm (5.59 in) 162 mm (6.38 in) 218 mm (8.58 in)

d 89 mm (3.5 in) 115 mm (4.53 in) 169 mm (6.65 in)

h 22 mm (0.87 in) 23.5 mm (0.93 in) 26.5 mm (1.04 in)

hmin 14 mm (0.55 in) 14 mm (0.55 in) 14 mm (0.55 in)

hmax 30 mm (1.18 in) 33 mm (1.3 in) 39 mm (1.45 in)

Endress+Hauser 89
Accessories Micropilot FMR67 HART

Accessories Description

Technical data: version ASME/JIS

Order number 71249070 71249072 71249073

Compatible with • ASME 3" 150lbs ASME 4" 150lbs ASME 6" 150lbs
• JIS 80A 10K

Recommended screw length 100 mm (3.9 in) 100 mm (3.9 in) 110 mm (4.3 in)

Recommended screw size M14 M14 M18

Material EPDM

Process pressure –0.1 to 0.1 bar (–1.45 to 1.45 psi)

Process temperature –40 to +80 °C (–40 to +176 °F)

D 133 mm (5.2 in) 171 mm (6.7 in) 219 mm (8.6 in)

d 89 mm (3.5 in) 115 mm (4.53 in) 168 mm (6.6 in)

h 22 mm (0.87 in) 23.5 mm (0.93 in) 26.5 mm (1.04 in)

hmin 14 mm (0.55 in) 14 mm (0.55 in) 14 mm (0.55 in)

hmax 30 mm (1.18 in) 33 mm (1.3 in) 39 mm (1.45 in)

90 Endress+Hauser
Micropilot FMR67 HART Accessories

16.1.3 Mounting bracket adjustable

Accessories Description

Mounting bracket
adjustable

260 (10.2)
99.5 (3.9)

9(0.4)
ø9(0.4)

65 (2.6)
122 (4.8)

48(1.9)

12(0.5)
35(1.4) 100 (3.9) 16 (0.6)
A0032295

Material: 304 (1.4301)


Suitable for housing 1):
– A: GT19 dual compartment, plastic PBT
– C: GT20 dual compartment, Alu, coated
Suitable for antenna 2):
GA: Drip-off, PTFE DN50
Suitable for process connection 3):
– GGJ: thread ISO228 G1-1/2, 316L
– RGJ: thread ANSI MNPT1-1/2, 316L
Order number: 71336522

 •• Risk
There is no conductive connection between the mounting bracket and the transmitter housing.
of electrostatic charge.
• Integrate the mounting bracket in the local potential equalization system.
• Only to be fastened at stable materials (e.g. metal, brick, concrete) using suitable fastening material (to be
supplied by customer).

1) Feature 040 in the product structure


2) Feature 070 in the product structure
3) Feature 100 in the product structure

Endress+Hauser 91
Accessories Micropilot FMR67 HART

16.1.4 Remote display FHX50

Accessories Description

Remote display FHX50

2.
1.
4.
3.

A0019128

• Material:
– Plastic PBT
– 316L/1.4404
– Aluminum
• Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for display modules:
– SD02 (push buttons)
– SD03 (touch control)
• Connecting cable:
– Cable supplied with device up to 30 m (98 ft)
– Standard cable supplied by customer up to 60 m (196 ft)
• Ambient temperature range: –40 to 80 °C (–40 to 176 °F)
• Ambient temperature range (option): –50 to 80 °C (–58 to 176 °F) 1)

 • IfL, the remote display should be used, order the device version "Prepared for display FHX50" (feature 030, version
M or N). For the FHX50, you must select option A: "Prepared for display FHX50" under feature 050 "Measuring
device version".
• If the device version "Prepared for display FHX50" was not originally ordered and a FHX50 display is to be
retrofitted, you must select version B "Not prepared for display FHX50" under feature 050: "Measuring device
version" when ordering the FHX50. In this case, a retrofit kit for the device is supplied with the FHX50. The kit
can be used to prepare the device so that the FHX50 can be used.

 FHX50
Use of the FHX50 may be restricted for transmitters with an approval. A device can only be retrofitted with the
if the option L, M or N ("Prepared for FHX50") is listed under Basic specifications, item 4 "Display, operation"
in the Safety Instructions (XA) for the device.
Also pay attention to the Safety Instructions (XA) of the FHX50.

 Retrofitting is not possible on transmitters with:


• An approval for use in areas with flammable dust (dust ignition-proof approval)
• Ex nA type of protection

 For details, see document SD01007F.


1) This range is valid if option JN "Ambient temperature transmitter –50 °C (–58 °F)" has been selected in ordering feature 580 "Test, Certificate". If
the temperature is permanently below –40 °C (–40 °F), failure rates may be increased.

92 Endress+Hauser
Micropilot FMR67 HART Accessories

16.1.5 Overvoltage protection

Accessory Description

Overvoltage protection for


2-wire-devices
OVP10 (1 channel)
OVP20 (2 channel)

A0021734

Technical data
• Resistance per channel: 2 * 0.5 Ωmax
• Threshold DC voltage: 400 to 700 V
• Threshold impulse voltage: < 800 V
• Capacitance at 1 MHz: < 1.5 pF
• Nominal arrest impulse voltage (8/20 μs): 10 kA
• Suited for wire cross-sections: 0.2 to 2.5 mm2 (24 to 14 AWG)

 Ordering with device


The overvoltage protection module is preferably ordered with the device. See product structure, feature 610
"Accessory mounted", option NA "Overvoltage protection". Separate ordering of the module is only necessary if a
device is to retrofitted with the overvoltage protection.

 Order code for retrofitting


• For 1-channel devices (feature 020, option A)
OVP10: 71128617
• For 2-channel devices (feature 020, option B, C, E or G)
OVP20 : 71128619
Hosuing lid for retrofitting
In order to keep the necessary safety distances, the housing lid needs to be replaced if the device is retrofitted with
the overvoltage protection. Depending on the housing type, the order code of the suitable lid is as follows:
• GT18 housing: Lid 71185516
• GT19 housing: Lid 71185518
• GT20 housing: Lid 71185516

 Restrictions for retrofitting


Depending on the approval of the transmitter the usage of the OVP module may be restricted. A device may only be
retrofitted with an OVP module if the option NA (overvoltage protection) is quoted unter Optional Specifications in
the Safety Instructions (XA) pertaining to the device.

 For details refer to SD01090F.

Endress+Hauser 93
Accessories Micropilot FMR67 HART

16.1.6 Bluetooth module for HART devices

Accessory Description

Bluetooth module

A0036493

• Quick and easy commissioning via SmartBlue (app)


• No additional tools or adapters required
• Signal curve via SmartBlue (app)
• Encrypted single point-to-point data transmission (tested by Fraunhofer institue) and password protected
communication via Bluetooth® wireless technology
• Range under reference conditions:
> 10 m (33 ft)

 When using the Bluetooth module the minimum supply voltage increases by up to 3 V.
 Ordering with device
The Bluetooth module is preferably ordered with the device. See product structure, feature 610 "Accessory
Mounted", option NF "Bluetooth". A separate order is only necessary in case of retrofitting.

 Order code for retrofitting


Bluetooth module (BT10): 71377355

 Restrictions in case of retrofitting


Depending on the approval of the transmitter, application of the Bluetooth module may be restricted. A device may
only be retrofitted with a Bluetooth module if the option NF (Bluetooth) is listed in the associated Safety
Instructions (XA) under Optional specifications.

 For details refer to SD02252F.

94 Endress+Hauser
Micropilot FMR67 HART Accessories

16.2 Communication-specific accessories


Accessory Description

Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface.
HART
 For details refer to Technical Information TI00404F

Accessory Description

Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a computer.
Order code: 51516983

 For details refer to Technical Information TI00405C

Accessory Description

HART Loop Converter Evaluates the dynamic HART variables and converts them to analog current signals
HMX50 or limit values.
Order code: 71063562

 For details refer to Technical Information TI00429F and Operating


Instructions BA00371F

Accessory Description

WirelessHART Adapter Connects field devices to a WirelessHART network.


SWA70 The WirelessHART adapter can be mounted directly at a HART device and is easly
integrated into an existing HART network. It ensures safe data transmission and
can be operated in parallel with other wireless networks.

 For details refer to Operating Instructions BA00061S

Accessories Description

Connect Sensor Fully integrated, battery-powered gateway for simple applications with SupplyCare Hosting.
FXA30/FXA30B Up to 4 field devices with 4 to 20 mA communication (FXA30/FXA30B), serial Modbus
(FXA30B) or HART (FXA30B) can be connected. With its robust design and ability to run for
years on the battery, it is ideal for remote monitoring in isolated locations. Version with LTE
(USA, Canada and Mexico only) or 3G mobile transmission for worldwide communication.

 BA01710S.
For details, see "Technical Information" TI01356S and Operating Instructions

Accessories Description

Fieldgate Fieldgates enable communication between connected 4 to 20 mA, Modbus RS485 and
FXA42 Modbus TCP devices and SupplyCare Hosting or SupplyCare Enterprise. The signals are
transmitted either via Ethernet TCP/IP, WLAN or mobile communications (UMTS). Advanced
automation capabilities are available, such as an integrated Web-PLC, OpenVPN and other
functions.

 BA01778S.
For details, see "Technical Information" TI01297S and Operating Instructions

Endress+Hauser 95
Accessories Micropilot FMR67 HART

Accessories Description

SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Enterprise pressures, densities or other tank parameters. The parameters are recorded and transmitted
SCE30B by means of gateways of the type Fieldgate FXA42.
This Web-based software is installed on a local server and can also be visualized and
operated with mobile terminals such as a smartphone or tablet.

 BA00055S
For details, see "Technical Information" TI01228S and Operating Instructions

Accessories Description

SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Hosting SCH30 pressures, densities or other tank parameters. The parameters are recorded and transmitted
by means of gateways of the type Fieldgate FXA42, FXA30 and FXA30B.
SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the
Endress+Hauser portal, the user is provided with the data over the Internet.

 BA00050S.
For details, see "Technical Information" TI01229S and Operating Instructions

Accessory Description

Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area.

 For details, see Operating Instructions BA01202S

Accessory Description

Field Xpert SFX370 Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area and the Ex area.

 For details, see Operating Instructions BA01202S

16.3 Service-specific accessories


Accessory Description

DeviceCare SFE100 Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus devices

 Technical Information TI01134S


 • DeviceCare is available for download at
www.software-products.endress.com. The download requires a
registration in the Endress+Hauser software portal.
• Alternatively, a DeviceCare DVD can be ordered with the device. Product
structure: Feature 570 "Service", Option IV "Tooling DVD (DeviceCare
Setup)".

FieldCare SFE500 FDT-based Plant Asset Management tool.


Helps to configure and maintain all field devices of your plant. By supplying status
information it also supports the diagnosis of the devices.

 Technical Information TI00028S

96 Endress+Hauser
Micropilot FMR67 HART Accessories

16.4 System components


Accessory Description

Graphic Data Manager The graphic data manager Memograph M provides information on all the relevant
Memograph M process variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on an SD card or USB stick.

 For details refer to Technical Information TI00133R and Operating


Instructions BA00247R

RN221N Active barrier with power supply for safe separation of 4 to 20 mA current circuits.
Provides bi-directional HART transmission.

 For details refer to Technical Information TI00073R and Operating


Instructions BA00202R

RNS221 Transmitter supply for 2-wire sensors or transmitters exclusively for non-Ex areas.
Provides bi-directional communication using the HART communication sockets.

 For details refer to Technical Information TI00081R and Operating


Instructions KA00110R

Endress+Hauser 97
Operating menu Micropilot FMR67 HART

17 Operating menu

17.1 Ovwerview of the operating menu (SmartBlue)


Navigation  Operating menu

 Setup →  115

Device tag →  115

Distance unit →  115

Bin type →  115

Max. draining speed solid →  116

Max. filling speed solid →  116

Empty calibration →  117

Full calibration →  118

Level →  119

Distance →  119

Signal quality →  119

Confirm distance →  119

Present mapping →  121

Mapping end point →  121

Record map →  121

‣ Advanced setup →  124

Locking status →  124

Access status tooling →  124

Enter access code →  125

‣ Level →  126

Medium type →  126

98 Endress+Hauser
Micropilot FMR67 HART Operating menu

Medium property →  126

Max. filling speed liquid →  127

Max. draining speed liquid →  127

Level unit →  128

Blocking distance →  129

Level correction →  129

Tank/silo height →  129

‣ Linearization →  132

Linearization type →  134

Unit after linearization →  135

Free text →  136

Level linearized →  136

Maximum value →  137

Diameter →  137

Intermediate height →  137

Table mode →  138

Table number →  139

Level →  139

Level →  140

Customer value →  140

Activate table →  140

‣ Safety settings →  141

Output echo lost →  141

Value echo lost →  141

Endress+Hauser 99
Operating menu Micropilot FMR67 HART

Ramp at echo lost →  142

Blocking distance →  129

‣ Current output 1 to 2 →  146

Assign current output →  146

Current span →  147

Fixed current →  147

Damping output →  148

Failure mode →  148

Failure current →  149

Output current 1 to 2 →  149

‣ Switch output →  150

Switch output function →  150

Assign status →  151

Assign limit →  151

Assign diagnostic behavior →  151

Switch-on value →  152

Switch-on delay →  153

Switch-off value →  153

Switch-off delay →  154

Failure mode →  154

Switch status →  154

Invert output signal →  154

 Diagnostics →  168

Actual diagnostics →  168

Timestamp →  168

100 Endress+Hauser
Micropilot FMR67 HART Operating menu

Previous diagnostics →  168

Timestamp →  169

Operating time from restart →  169

Operating time →  162

‣ Diagnostic list →  170

Diagnostics 1 to 5 →  170

Timestamp →  170

‣ Device information →  172

Device tag →  172

Serial number →  172

Firmware version →  172

Device name →  172

Order code →  173

Extended order code 1 to 3 →  173

Device revision →  173

Device ID →  173

Device type →  174

Manufacturer ID →  174

‣ Measured values →  175

Distance →  175

Level linearized →  136

Output current 1 to 2 →  149

Measured current 1 →  176

Endress+Hauser 101
Operating menu Micropilot FMR67 HART

Terminal voltage 1 →  176

Sensor temperature →  176

‣ Simulation →  181

Assign measurement variable →  182

Process variable value →  182

Current output 1 to 2 simulation →  182

Value current output 1 to 2 →  183

Switch output simulation →  183

Switch status →  183

Device alarm simulation →  184

Diagnostic event simulation →  184

102 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.2 Overview of the operating menu (display module)


Navigation  Operating menu

Language →  156

 Setup →  115

Device tag →  115

Distance unit →  115

Bin type →  115

Max. draining speed solid →  116

Max. filling speed solid →  116

Empty calibration →  117

Full calibration →  118

Level →  119

Distance →  119

Signal quality →  119

‣ Mapping →  122

Confirm distance →  122

Mapping end point →  122

Present mapping →  121

Record map →  121

Distance →  119

Prepare recording map →  122

‣ Advanced setup →  124

Locking status →  124

Access status display →  125

Enter access code →  125

Endress+Hauser 103
Operating menu Micropilot FMR67 HART

‣ Level →  126

Medium type →  126

Medium property →  126

Level unit →  128

Blocking distance →  129

Level correction →  129

Tank/silo height →  129

‣ Linearization →  132

Linearization type →  134

Unit after linearization →  135

Free text →  136

Maximum value →  137

Diameter →  137

Intermediate height →  137

Table mode →  138

‣ Edit table

Level →  139

Customer value →  140

Activate table →  140

‣ Safety settings →  141

Output echo lost →  141

Value echo lost →  141

Ramp at echo lost →  142

Blocking distance →  129

‣ SIL/WHG confirmation →  144

104 Endress+Hauser
Micropilot FMR67 HART Operating menu

‣ Deactivate SIL/WHG →  145

Reset write protection →  145

Code incorrect →  145

‣ Current output 1 to 2 →  146

Assign current output →  146

Current span →  147

Fixed current →  147

Damping output →  148

Failure mode →  148

Failure current →  149

Output current 1 to 2 →  149

‣ Switch output →  150

Switch output function →  150

Assign status →  151

Assign limit →  151

Assign diagnostic behavior →  151

Switch-on value →  152

Switch-on delay →  153

Switch-off value →  153

Switch-off delay →  154

Failure mode →  154

Switch status →  154

Invert output signal →  154

‣ Display →  156

Language →  156

Endress+Hauser 105
Operating menu Micropilot FMR67 HART

Format display →  156

Value 1 to 4 display →  158

Decimal places 1 to 4 →  158

Display interval →  158

Display damping →  159

Header →  159

Header text →  160

Separator →  160

Number format →  160

Decimal places menu →  160

Backlight →  161

Contrast display →  161

‣ Configuration backup display →  162

Operating time →  162

Last backup →  162

Configuration management →  162

Backup state →  163

Comparison result →  163

‣ Administration →  165

‣ Define access code →  167

Define access code →  167

Confirm access code →  167

Device reset →  165

 Diagnostics →  168

Actual diagnostics →  168

106 Endress+Hauser
Micropilot FMR67 HART Operating menu

Previous diagnostics →  168

Operating time from restart →  169

Operating time →  162

‣ Diagnostic list →  170

Diagnostics 1 to 5 →  170

‣ Event logbook →  171

Filter options →  171

‣ Event list →  171

‣ Device information →  172

Device tag →  172

Serial number →  172

Firmware version →  172

Device name →  172

Order code →  173

Extended order code 1 to 3 →  173

Device revision →  173

Device ID →  173

Device type →  174

Manufacturer ID →  174

‣ Measured values →  175

Distance →  175

Level linearized →  136

Output current 1 to 2 →  149

Measured current 1 →  176

Endress+Hauser 107
Operating menu Micropilot FMR67 HART

Terminal voltage 1 →  176

Sensor temperature →  176

‣ Data logging →  177

Assign channel 1 to 4 →  177

Logging interval →  177

Clear logging data →  178

‣ Display channel 1 to 4 →  179

‣ Simulation →  181

Assign measurement variable →  182

Process variable value →  182

Current output 1 to 2 simulation →  182

Value current output 1 to 2 →  183

Switch output simulation →  183

Switch status →  183

Device alarm simulation →  184

Diagnostic event category →  184

Diagnostic event simulation →  184

‣ Device check →  185

Start device check →  185

Result device check →  185

Last check time →  185

108 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.3 Overview of the operating menu (operating tool)


Navigation  Operating menu

 Setup →  115

Device tag →  115

Distance unit →  115

Bin type →  115

Max. draining speed solid →  116

Max. filling speed solid →  116

Empty calibration →  117

Full calibration →  118

Level →  119

Distance →  119

Signal quality →  119

Confirm distance →  119

Present mapping →  121

Mapping end point →  121

Record map →  121

‣ Advanced setup →  124

Locking status →  124

Access status tooling →  124

Enter access code →  125

‣ Level →  126

Medium type →  126

Medium property →  126

Max. filling speed liquid →  127

Endress+Hauser 109
Operating menu Micropilot FMR67 HART

Max. draining speed liquid →  127

Level unit →  128

Blocking distance →  129

Level correction →  129

Tank/silo height →  129

‣ Linearization →  132

Linearization type →  134

Unit after linearization →  135

Free text →  136

Level linearized →  136

Maximum value →  137

Diameter →  137

Intermediate height →  137

Table mode →  138

Table number →  139

Level →  139

Level →  140

Customer value →  140

Activate table →  140

‣ Safety settings →  141

Output echo lost →  141

Value echo lost →  141

Ramp at echo lost →  142

Blocking distance →  129

‣ SIL/WHG confirmation →  144

110 Endress+Hauser
Micropilot FMR67 HART Operating menu

‣ Deactivate SIL/WHG →  145

Reset write protection →  145

Code incorrect →  145

‣ Current output 1 to 2 →  146

Assign current output →  146

Current span →  147

Fixed current →  147

Damping output →  148

Failure mode →  148

Failure current →  149

Output current 1 to 2 →  149

‣ Switch output →  150

Switch output function →  150

Assign status →  151

Assign limit →  151

Assign diagnostic behavior →  151

Switch-on value →  152

Switch-on delay →  153

Switch-off value →  153

Switch-off delay →  154

Failure mode →  154

Switch status →  154

Invert output signal →  154

‣ Display →  156

Language →  156

Endress+Hauser 111
Operating menu Micropilot FMR67 HART

Format display →  156

Value 1 to 4 display →  158

Decimal places 1 to 4 →  158

Display interval →  158

Display damping →  159

Header →  159

Header text →  160

Separator →  160

Number format →  160

Decimal places menu →  160

Backlight →  161

Contrast display →  161

‣ Configuration backup display →  162

Operating time →  162

Last backup →  162

Configuration management →  162

Backup state →  163

Comparison result →  163

‣ Administration →  165

Define access code →  165

Device reset →  165

 Diagnostics →  168

Actual diagnostics →  168

Timestamp →  168

Previous diagnostics →  168

112 Endress+Hauser
Micropilot FMR67 HART Operating menu

Timestamp →  169

Operating time from restart →  169

Operating time →  162

‣ Diagnostic list →  170

Diagnostics 1 to 5 →  170

Timestamp →  170

‣ Device information →  172

Device tag →  172

Serial number →  172

Firmware version →  172

Device name →  172

Order code →  173

Extended order code 1 to 3 →  173

Device revision →  173

Device ID →  173

Device type →  174

Manufacturer ID →  174

‣ Measured values →  175

Distance →  175

Level linearized →  136

Output current 1 to 2 →  149

Measured current 1 →  176

Terminal voltage 1 →  176

Sensor temperature →  176

Endress+Hauser 113
Operating menu Micropilot FMR67 HART

‣ Data logging →  177

Assign channel 1 to 4 →  177

Logging interval →  177

Clear logging data →  178

‣ Simulation →  181

Assign measurement variable →  182

Process variable value →  182

Current output 1 to 2 simulation →  182

Value current output 1 to 2 →  183

Switch output simulation →  183

Switch status →  183

Device alarm simulation →  184

Diagnostic event simulation →  184

‣ Device check →  185

Start device check →  185

Result device check →  185

Last check time →  185

‣ Heartbeat →  186

114 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.4 "Setup" menu


• : Marks the navigation path to the parameter via the display and operating
module.
• : Marks the navigation path to the parameter via an operating tool (e.g.
FieldCare).
• : Marks parameters which can be locked via the software locking.

Navigation  Setup

Device tag 

Navigation  Setup → Device tag

Description Enter a unique name for the measuring point to identify the device quickly within the
plant.

Factory setting FMR6x

Distance unit 

Navigation  Setup → Distance unit

Description Used for the basic calibration (Empty / Full).

Selection SI units US units


• mm • ft
•m • in

Factory setting m

Bin type 

Navigation  Setup → Bin type

Prerequisite Medium type (→  126) = Solid

Description Optimizes the signal filters for the respective bin type. Note: 'Workbench test' deactivates
all filters. This option should exclusively be used for tests.

Selection • Buffer silo (fast)


• Bin / Pile
• Crusher / belt
• Silo
• Workbench test

Endress+Hauser 115
Operating menu Micropilot FMR67 HART

Factory setting Silo

Additional information Workbench test deactivates all filters. This option is intended for tests only.

Max. draining speed solid 

Navigation  Setup → Max. draining speed solid

Prerequisite Medium type (→  126) = Solid

Description Select maximum expecte draining speed.

Selection • Very slow < 0.5m (1.6ft) /h


• Slow < 1m (3.3ft) /h
• Standard < 2m (6,5ft) /h
• Medium < 4m (13ft) /h
• Fast < 8m (26ft) /h
• Very fast > 8m (26ft) /h
• No filter / test

Factory setting No filter / test

Additional information By specifying the maximum filling and draining speed the signal evaluation is
automatically optimized for the process.

Max. draining speed liquid Step response time / s

Very slow < 0.5m (1.6ft) /h 850

Slow < 1m (3.3ft) /h 710

Standard < 2m (6,5ft) /h 300

Medium < 4m (13ft) /h 155

Fast < 8m (26ft) /h 70

Very fast > 8m (26ft) /h 6

No filter / test <1

The filling and draining speed can be set separately as filling and draining may may
be different processes.
When selecting the No filter / test option all filters of the signal evaluation are
deactivated. This option is intended for tests only.

Max. filling speed solid 

Navigation  Setup → Max. filling speed solid

Prerequisite Medium type (→  126) = Solid

Description Select expected maximum filling speed.

116 Endress+Hauser
Micropilot FMR67 HART Operating menu

Selection • Very slow < 0.5m (1.6ft) /h


• Slow < 1m (3.3ft) /h
• Standard < 2m (6,5ft) /h
• Medium < 4m (13ft) /h
• Fast < 8m (26ft) /h
• Very fast > 8m (26ft) /h
• No filter / test

Factory setting No filter / test

Additional information By specifying the maximum filling and draining speed the signal evaluation is
automatically optimized for the process.

Max. draining speed liquid Sprungantwortzeit / s

Very slow < 0.5m (1.6ft) /h 850

Slow < 1m (3.3ft) /h 710

Standard < 2m (6,5ft) /h 300

Medium < 4m (13ft) /h 155

Fast < 8m (26ft) /h 70

Very fast > 8m (26ft) /h 6

No filter / test <1

The filling and draining speed can be set separately as filling and draining may may
be different processes.
When selecting the No filter / test option all filters of the signal evaluation are
deactivated. This option is intended for tests only.

Empty calibration 

Navigation  Setup → Empty calibration

Description Distance between process connection and minimum level (0%).

User entry Depending on the antenna

Factory setting Depending on the antenna

Endress+Hauser 117
Operating menu Micropilot FMR67 HART

Additional information

0%

A0019488

 37 Empty calibration (E) for level measurements in bulk solids.

The measuring range starts at the point at which the radar beam hits the tank or silo
bottom. In the case of dished boiler ends or conical outlets levels below this point can
not be measured.

Full calibration 

Navigation  Setup → Full calibration

Description Distance between minimum level (0%) and maximum level (100%).

User entry Depending on the antenna

Factory setting Depending on the antenna

Additional information

100%

0%

A0019489

 38 Full calibration (F) for level measurements in bulk solids

118 Endress+Hauser
Micropilot FMR67 HART Operating menu

Level

Navigation  Setup → Level

Description Displays measured level L (before linearization).

Additional information

L
0%

A0019484

 39 Level in case of bulk solid measurements

The unit is defined in the Level unit parameter (→  128).

Distance

Navigation  Setup → Distance

Description Distance between lower edge of flange or thread and medium surface.

Signal quality

Navigation  Setup → Signal quality

Description Shows the quality of the evaluated level signal.

Confirm distance 

Navigation  Setup → Confirm distance

Description Specify, whether the measured distance matches the real distance. Depending on the
selection the device automatically sets the range of mapping.

Endress+Hauser 119
Operating menu Micropilot FMR67 HART

Selection • Manual map


• Distance ok
• Distance unknown
• Distance too small *
• Distance too big *
• Tank empty
• Factory map

Factory setting Distance unknown

Additional information Meaning of the options


• Manual map
To be selected if the range of mapping is to be defined manually in the Mapping end
point parameter. In this case it is not necessary to confirm the distance.
• Distance ok
To be selected if the measured distance matches the actual distance. The device
performs a mapping.
• Distance unknown
To be selected if the actual distance is unknown. A mapping can not be performed in this
case.
• Distance too small 6)
To be selected if the measured distance is smaller than the actual distance. The device
searches for the next echo and returns to the Confirm distance parameter . The distance
is recalculated and displayed. The comparison must be repeated until the displayed
distance matches the actual distance. After this, the recording of the map can be started
by selecting Distance ok.
• Distance too big
To be selected if the measured distance exceeds the actual distance. The device adjusts
the signal evaluation and returns to the Confirm distance parameter. The distance is
recalculated and displayed. The comparison must be repeated until the displayed
distance matches the actual distance. After this, the recording of the map can be started
by selecting Distance ok.
• Tank empty
To be selected if the tank is completely empty. The device records a mapping covering
the complete measuring range as defined by the Tank/silo height parameter. By
default, Tank/silo height = Empty calibration.
Take into account that in case of conical outlets, for example, a measurement is only
possible up to the point at which the radar hits the bottom of the tank or silo. If the
Tank empty option is used, Empty calibration (→  117) and Tank/silo height may
not reach below this point as otherwise the empty signal is suppressed.
• Factory map
The factory map permanently stored in the device is used.
No factory map is required for the FMR6x device generation. Therefore, a constant
line of –116 dB is stored as the factory map. On commissioning, a map can be
recorded which optimally fits the actual installation.
When operating via the display module, the measured distance is displayed together
with this parameter for reference purposes.
If the teaching procedure with the Distance too small option or Distance too big
option is quit before the distance has been confirmed, a map is not recorded and the
teaching procedure is reset after 60 s.

* Visibility depends on order options or device settings


6) Only available for "Expert → Sensor → Echo tracking → Evaluation mode parameter " ≠ " History off option"

120 Endress+Hauser
Micropilot FMR67 HART Operating menu

Present mapping

Navigation  Setup → Present mapping

Description Present end of mapping.

Mapping end point 

Navigation  Setup → Mapping end point

Prerequisite Confirm distance (→  119) = Manual map or Distance too small

Description New end point of mapping.

User entry 0.0001 to 999 999.9 m

Factory setting 0.1 m

Record map

Navigation  Setup → Record map

Prerequisite Confirm distance = Manual map or Distance too small

Selection • No
• Record map
• Overlay map
• Factory map
• Delete partial map

Factory setting No

Endress+Hauser 121
Operating menu Micropilot FMR67 HART

17.4.1 "Mapping" wizard


The Mapping wizard is only available when operating via the local display. When
operating via an operating tool, all parameters concerning the mapping are located
directly in the Setup menu (→  115)
In the Mapping wizard two parameters are displayed simultaneously on the display
module at any one time. The upper parameter can be edited, whereas the lower
parameter is displayed for reference purposes only.

Navigation  Setup → Mapping

Confirm distance 

Navigation  Setup → Mapping → Confirm distance

Description →  119

Mapping end point 

Navigation  Setup → Mapping → Mapping end point

Description →  121

Record map 

Navigation  Setup → Mapping → Record map

Description →  121

Distance

Navigation  Setup → Mapping → Distance

Description →  119

Prepare recording map

Navigation  Setup → Mapping → Prepare recording map

Description Zeigt Status der Aufnahme der Ausblendung.

122 Endress+Hauser
Micropilot FMR67 HART Operating menu

User interface • Init. recording


• In progress
• Finished

Endress+Hauser 123
Operating menu Micropilot FMR67 HART

17.4.2 "Advanced setup" submenu

Navigation  Setup → Advanced setup

Locking status

Navigation  Setup → Advanced setup → Locking status

Description Indicates the write protection with the highest priority that is currently active.

User interface • Hardware locked


• SIL locked
• WHG locked
• Temporarily locked

Additional information Meaning and priorities of the types of write protection


• Hardware locked (priority 1)
The DIP switch for hardware locking is activated on the main electronics module. This
locks write access to the parameters.
• SIL locked (priority 2)
The SIL mode is activated. Writing access to the relevant parameters is denied.
• WHG locked (priority 3)
The WHG mode is activated. Writing access to the relevant parameters is denied.
• Temporarily locked (priority 4)
Write access to the parameters is temporarily locked on account of internal processes in
progress in the device (e.g. data upload/download, reset etc.). The parameters can be
modified as soon as the processes are complete.
On the display module, the -symbol appears in front of parameters that cannot be
modified since they are write-protected.

Access status tooling

Navigation  Setup → Advanced setup → Access status tooling

Description Shows the access authorization to the parameters via the operating tool.

Additional information The access authorization can be changed via the Enter access code parameter
(→  125).
If additional write protection is active, this restricts the current access authorization
even further. The write protection status can be viewed via the Locking status
parameter (→  124).

124 Endress+Hauser
Micropilot FMR67 HART Operating menu

Access status display

Navigation  Setup → Advanced setup → Access status display

Prerequisite The device has a local display.

Description Indicates access authorization to parameters via local display.

Additional information The access authorization can be changed via the Enter access code parameter
(→  125).
If additional write protection is active, this restricts the current access authorization
even further. The write protection status can be viewed via the Locking status
parameter (→  124).

Enter access code

Navigation  Setup → Advanced setup → Enter access code

Description Enter access code to disable write protection of parameters.

User entry 0 to 9 999

Additional information • For local operation, the customer-specific access code, which has been defined in the
Define access code parameter (→  165), has to be entered.
• If an incorrect access code is entered, the user retains his current access authorization.
• The write protection affects all parameters marked with the -symbol in this document.
On the local display, the -symbol in front of a parameter indicates that the parameter
is write-protected.
• If no key is pressed for 10 min, or the user switches from the navigation and editing
mode back to the measured value display mode, the device automatically locks the write-
protected parameters after another 60 s.
Please contact your Endress+Hauser Sales Center if you lose your access code.

Endress+Hauser 125
Operating menu Micropilot FMR67 HART

"Level" submenu

Navigation  Setup → Advanced setup → Level

Medium type 

Navigation  Setup → Advanced setup → Level → Medium type

Description Specify medium type.

User interface • Liquid


• Solid

Factory setting Solid

Additional information This parameter determines the value of several other parameters and strongly
influences the complete signal evaluation. Therefore, it is strongly recommended not
to change the factory setting.

Medium property 

Navigation  Setup → Advanced setup → Level → Medium property

Description Specify relative dielectric constant εr of the medium.

Selection • Unknown
• DC 1.4 ... 1.6
• DC 1.6 ... 1.9
• DC 1.9 ... 2.5
• DC 2.5 ... 4
• DC 4 ... 7
• DC 7 ... 15
• DC > 15

Factory setting Depending on the Medium type (→  126) and Medium group parameters.

Additional information Dependence on "Medium type" and "Medium group"


Medium type (→  126) Medium group Medium property (→  126)

Solid Unknown

Liquid Water based (DC >= 4) DC 4 ... 7

Others Unknown

For dielectric constants (DC values) of many media commonly used in various
industries refer to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)

126 Endress+Hauser
Micropilot FMR67 HART Operating menu

Max. filling speed liquid 

Navigation  Setup → Advanced setup → Level → Max. filling speed liquid

Prerequisite Medium type (→  126) = Liquid

Description Select expected maximum filling speed.

Selection • Slow < 1cm (0.4in) /min


• Medium < 10cm (4in) /min
• Standard < 1m (40in) /min
• Fast < 2m (80in) /min
• Very fast > 2m (80in) /min
• No filter / test

Factory setting Depending on the Tank type parameter

Additional information By selecting the maximum expected filling and draining speed the signal evaluation is
automatically optimized for the process.

Max. draining speed liquid Step response time / s

Slow < 1cm (0.4in) /min 90

Medium < 10cm (4in) /min 50

Standard < 1m (40in) /min 19

Fast < 2m (80in) /min 8

Very fast > 2m (80in) /min 6

No filter / test <1

The filling and draining speeds can be set separately as the filling and draining
procedures may be different.
With the No filter / test option all signal evaluation filters are deactivated. This
option should exclusively be used for tests.
Max. filling speed liquid is preset by Tank type. It can, however, be adjusted to the
process in the vessel at any time. If Tank type is changed again, it may be necessary
to repeat the fine adjustment.

Max. draining speed liquid 

Navigation  Setup → Advanced setup → Level → Max. draining speed liquid

Prerequisite Tank type = Liquid

Description Select expected maximum draining speed.

Selection • Slow < 1cm (0.4in) /min


• Medium < 10cm (4in) /min
• Standard < 1m (40in) /min

Endress+Hauser 127
Operating menu Micropilot FMR67 HART

• Fast < 2m (80in) /min


• Very fast > 2m (80in) /min
• No filter / test

Factory setting Depending on the Tank type parameter

Additional information By selecting the maximum expected filling and draining speed the signal evaluation is
automatically optimized for the process.

Max. draining speed liquid Step response time / s

Slow < 1cm (0.4in) /min 90

Medium < 10cm (4in) /min 50

Standard < 1m (40in) /min 19

Fast < 2m (80in) /min 8

Very fast > 2m (80in) /min 6

No filter / test <1

The filling and draining speeds can be set separately as the filling and draining
procedures may be different.
With the No filter / test option all signal evaluation filters are deactivated. This
option should exclusively be used for tests.
Max. draining speed liquid is preset by Tank type. It can, however, be adjusted to
the process in the vessel at any time. If Tank type is changed again, it may be
necessary to repeat the fine adjustment.

Level unit 

Navigation  Setup → Advanced setup → Level → Level unit

Description Select level unit.

Selection SI units US units


•% • ft
•m • in
• mm

Factory setting %

Additional information The level unit may differ from the distance unit defined in the Distance unit parameter
(→  115):
• The unit defined in the Distance unit parameter is used for the basic calibration (Empty
calibration (→  117) and Full calibration (→  118)).
• The unit defined in the Level unit parameter is used to display the (unlinearized) level.

128 Endress+Hauser
Micropilot FMR67 HART Operating menu

Blocking distance 

Navigation  Setup → Advanced setup → Level → Blocking distance

Description Dead band in front of the process connection.

User entry 0 to 200 m

Factory setting • Empty calibration - Full calibration - 200 mm (8 in)


• Minimum value: 150 mm (6 in)

Additional information The blocking distance can be used to suppress interference echos in the vicinity of the
antenna.

BD

A0023041

 40 Blocking distance (BD) for bulk solid measurements

Level correction 

Navigation  Setup → Advanced setup → Level → Level correction

Description Specify level correction (if required).

User entry –200 000.0 to 200 000.0 %

Factory setting 0.0 %

Additional information The value specified in this parameter is added to the measured level (before linearization).

Tank/silo height 

Navigation  Setup → Advanced setup → Level → Tank/silo height

Description Total height of the tank or silo (measured from the process connection)

Endress+Hauser 129
Operating menu Micropilot FMR67 HART

User entry –999.9999 to 999.9999 m

Factory setting Empty calibration (→  117)

Additional information If the parametrized measuring range differs significantly from the tank or silo height, it is
recommended to enter the tank or silo height. Example: Continuous level monitoring in
the upper third of a tank or silo.

0%
H

A0019868

 41 '"Tank/silo height" parameter' for measurements in bulk solids


E Empty calibration (→  117)
H Tank/silo height (→  129)

For tanks with conical outlet, Tank/silo height should not be changed as in this type
of applications Empty calibration (→  117) is usually not much less-than the tank
or silo height.

130 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Linearization" submenu

2
A
L
100%
L 1 B L‘
0%
M L‘
L
100%

L2
C L1
0%
L‘1 L‘2 L‘
L
100%

h
D
0%
M L‘
L
100%

h
E
0%
M L‘
L
100%

h
0%
M L‘
L
100%

G d

0%
M L‘
L
100%

H d

0%
M L‘

A0019648

 42 Linearization: Transformation of the level and (if relevant) the interface height into a volume or weight;
the transformation is dependent on the shape of the vessel.
1 Selection of linearization type and unit
2 Configuration of the linearization
A Linearization type (→  134) = None
B Linearization type (→  134) = Linear
C Linearization type (→  134) = Table
D Linearization type (→  134) = Pyramid bottom
E Linearization type (→  134) = Conical bottom
F Linearization type (→  134) = Angled bottom
G Linearization type (→  134) = Horizontal cylinder
H Linearization type (→  134) = Sphere
L Level before linearization (measured in distance units)
L' Level linearized (→  136) (corresponds to volume or weight)
M Maximum value (→  137)
d Diameter (→  137)
h Intermediate height (→  137)

Endress+Hauser 131
Operating menu Micropilot FMR67 HART

Structure of the submenu on the display module

Navigation  Setup → Advanced setup → Linearization

‣ Linearization

Linearization type →  134

Unit after linearization →  135

Free text →  136

Maximum value →  137

Diameter →  137

Intermediate height →  137

Table mode →  138

‣ Edit table

Level →  139

Customer value →  140

Activate table →  140

132 Endress+Hauser
Micropilot FMR67 HART Operating menu

Structure of the submenu in an operating tool (e.g. FieldCare)

Navigation  Setup → Advanced setup → Linearization

‣ Linearization

Linearization type →  134

Unit after linearization →  135

Free text →  136

Level linearized →  136

Maximum value →  137

Diameter →  137

Intermediate height →  137

Table mode →  138

Table number →  139

Level →  139

Level →  140

Customer value →  140

Activate table →  140

Endress+Hauser 133
Operating menu Micropilot FMR67 HART

Description of parameters

Navigation  Setup → Advanced setup → Linearization

Linearization type 

Navigation  Setup → Advanced setup → Linearization → Linearization type

Description Select linearization type.

Selection • None
• Linear
• Table
• Pyramid bottom
• Conical bottom
• Angled bottom
• Horizontal cylinder
• Sphere

Factory setting None

Additional information
A B
100%
100%

0% 0%

C/D E
100% 100%

0% 0%

F/G

100%

0%
A0021476

 43 Linearization types
A None
B Table
C Pyramid bottom
D Conical bottom
E Angled bottom
F Sphere
G Horizontal cylinder

134 Endress+Hauser
Micropilot FMR67 HART Operating menu

Meaning of the options


• None
The level is transmitted in the level unit without linearization.
• Linear
The output value (volume/weight) is directly proportional to the level L. This is valid, for
example, for vertical cylinders. The following additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
• Table
The relationship between the measured level L and the output value (volume/weight) is
given by a linearization table consisting of up to 32 pairs of values "level - volume" or
"level - weight", respectively. The following additional parameters have to be specified:
– Unit after linearization (→  135)
– Table mode (→  138)
– For each table point: Level (→  139)
– For each table point: Customer value (→  140)
– Activate table (→  140)
• Pyramid bottom
The output value corresponds to the volume or weight in a silo with pyramid bottom.
The following additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
– Intermediate height (→  137): The height of the pyramid
• Conical bottom
The output value corresponds to the volume or weight in a tank with conical bottom. The
following additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
– Intermediate height (→  137): The height of the conical part of the tank
• Angled bottom
The output value corresponds to the volume or weight in a silo with an angled bottom.
The following additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
– Intermediate height (→  137): Height of the angled bottom
• Horizontal cylinder
The output value corresponds to the volume or weight in a horizontal cylinder. The
following additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
– Diameter (→  137)
• Sphere
The output value corresponds to the volume or weight in a spherical tank. The following
additional parameters have to be specified:
– Unit after linearization (→  135)
– Maximum value (→  137): Maximum volume or weight
– Diameter (→  137)

Unit after linearization 

Navigation  Setup → Advanced setup → Linearization → Unit after linearization

Prerequisite Linearization type (→  134) ≠ None

Description Select unit of the lineaized value.

Endress+Hauser 135
Operating menu Micropilot FMR67 HART

Selection SI units US units Imperial units


• STon • lb impGal
•t • UsGal
• kg • ft³
• cm³ • ft
• dm³ • in
• m³
• hl
•l
•%
• mm
•m

Custom-specific units
Free text

Factory setting %

Additional information The selected unit is only used to be indicated on the display. The measured value is not
transformed according to the selected unit.
It is also possible to configure a distance-to-distance linearization, i.e. a
transformation from the level unit to a different distance unit. To do so, select the
Linear linearization mode. In order to define the new level unit, select the Free text
option in the Unit after linearization parameter and enter the required unit into the
Free text parameter (→  136).

Free text 

Navigation  Setup → Advanced setup → Linearization → Free text

Prerequisite Unit after linearization (→  135) = Free text

Description Enter unit symbol.

User entry Up to 32 alphanumerical characters (letters, numbers, special characters)

Factory setting Free text

Level linearized

Navigation  Setup → Advanced setup → Linearization → Level linearized

Description Displays linearized level.

Additional information The unit is defined by the Unit after linearization parameter →  135.

136 Endress+Hauser
Micropilot FMR67 HART Operating menu

Maximum value 

Navigation  Setup → Advanced setup → Linearization → Maximum value

Prerequisite Linearization type (→  134) has one of the following values:


• Linear
• Pyramid bottom
• Conical bottom
• Angled bottom
• Horizontal cylinder
• Sphere

Description Linearized value corresponding to a level of 100%.

User entry –50 000.0 to 50 000.0 %

Factory setting 100.0 %

Diameter 

Navigation  Setup → Advanced setup → Linearization → Diameter

Prerequisite Linearization type (→  134) has one of the following values:


• Horizontal cylinder
• Sphere

Description Diameter of the cylindrical or spherical tank.

User entry 0 to 9 999.999 m

Factory setting 2m

Additional information The unit is defined in the Distance unit parameter (→  115).

Intermediate height 

Navigation  Setup → Advanced setup → Linearization → Intermediate height

Prerequisite Linearization type (→  134) has one of the following values:


• Pyramid bottom
• Conical bottom
• Angled bottom

Description Height of the pyramid, conical or angled bottom.

User entry 0 to 200 m

Factory setting 0m

Endress+Hauser 137
Operating menu Micropilot FMR67 HART

Additional information

H
0%

A0013264

H Intermediate height

The unit is defined in the Distance unit parameter (→  115).

Table mode 

Navigation  Setup → Advanced setup → Linearization → Table mode

Prerequisite Linearization type (→  134) = Table

Description Select editing mode of the linearization table.

Selection • Manual
• Semiautomatic
• Clear table
• Sort table

Factory setting Manual

Additional information Meaning of the options


• Manual
The level and the associated linearized value are entered manually for each linearization
point.
• Semiautomatic
The level is measured by the device for each linearization point. The associated
linearized value is entered manually.
• Clear table
Deletes the existing linearization table.
• Sort table
Rearranges the linerization points into an ascending order.

Conditions the linearization table must meet:


• The table may consist of up to 32 pairs of values "Level - Linearized Value".
• The table must be monotonic (monotonically increasing or decreasing).
• The first linearization point must refer to the minimum level.
• The last linearization point must refer to the maximum level.

Before entering a linearization table, the values for Empty calibration (→  117)
and Full calibration (→  118) must be set correctly.
If values of the table need to be changed after the full or empty calibration have been
changed, a correct evaluation is only ensured if the existing table is deleted and the
complete table is entered again. To do so delete the existing table (Table mode
(→  138) = Clear table). Then enter a new table.

138 Endress+Hauser
Micropilot FMR67 HART Operating menu

How to enter the table


• Via FieldCare
The table points can be entered via the Table number (→  139), Level (→  139)
and Customer value (→  140) parameters. As an alternative, the graphic table editor
may be used: Device Operation → Device Functions → Additional Functions →
Linearization (Online/Offline)
• Via local display
Select the Edit table submenu to call up the graphic table editor. The table is displayed
and can be edited line by line.

The factory setting for the level unit is "%". If you want to enter the linearization table
in physical units, you must select the appropriate unit in the Level unit parameter
(→  128) beforehand.
If a decreasing table is entered, the values for 20 mA and 4 mA of the current output
are interchanged. That means: 20 mA refers to the lowest level, whereas 4 mA refers
to the highest level.

Table number 

Navigation  Setup → Advanced setup → Linearization → Table number

Prerequisite Linearization type (→  134) = Table

Description Select table point you are going to enter or change.

User entry 1 to 32

Factory setting 1

Level (Manual) 

Navigation  Setup → Advanced setup → Linearization → Level

Prerequisite • Linearization type (→  134) = Table


• Table mode (→  138) = Manual

Description Enter level value of the table point (value before linearization).

User entry Signed floating-point number

Factory setting 0%

Endress+Hauser 139
Operating menu Micropilot FMR67 HART

Level (Semiautomatic)

Navigation  Setup → Advanced setup → Linearization → Level

Prerequisite • Linearization type (→  134) = Table


• Table mode (→  138) = Semiautomatic

Description Displays measured level (value before linearization). This value is transmitted to the table.

Customer value 

Navigation  Setup → Advanced setup → Linearization → Customer value

Prerequisite Linearization type (→  134) = Table

Description Enter linearized value for the table point.

User entry Signed floating-point number

Factory setting 0%

Activate table 

Navigation  Setup → Advanced setup → Linearization → Activate table

Prerequisite Linearization type (→  134) = Table

Description Activate (enable) or deactivate (disable) the linearization table.

Selection • Disable
• Enable

Factory setting Disable

Additional information Meaning of the options


• Disable
The measured level is not linearized.
If Linearization type (→  134) = Table at the same time, the device issues error
message F435.
• Enable
The measured level is linearized according to the table.
When editing the table, the Activate table parameter is automatically reset to
Disable and must be reset to Enable after the table has been entered.

140 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Safety settings" submenu

Navigation  Setup → Advanced setup → Safety settings

Output echo lost 

Navigation  Setup → Advanced setup → Safety settings → Output echo lost

Description Output signal in case of a lost echo.

Selection • Last valid value


• Ramp at echo lost
• Value echo lost
• Alarm

Factory setting Last valid value

Additional information Meaning of the options


• Last valid value
The last valid value is kept in the case of a lost echo.
• Ramp at echo lost 7)
In the case of a lost echo the output value is continously shifted towards 0% or 100%.
The slope of the ramp is defined in the Ramp at echo lost parameter (→  142).
• Value echo lost 7)
In the case of a lost echo the output assumes the value defined in the Value echo lost
parameter (→  141).
• Alarm
In the case of a lost echo the device generates an alarm; see the Failure mode parameter
(→  148)

Value echo lost 

Navigation  Setup → Advanced setup → Safety settings → Value echo lost

Prerequisite Output echo lost (→  141) = Value echo lost

Description Output value in case of a lost echo

User entry 0 to 200 000.0 %

Factory setting 0.0 %

Additional information Use the unit which has been defined for the measured value output:
• without linearization: Level unit (→  128)
• with linearization: Unit after linearization (→  135)

7) Only visible if '"Linearization type (→  134)" = "None"

Endress+Hauser 141
Operating menu Micropilot FMR67 HART

Ramp at echo lost 

Navigation  Setup → Advanced setup → Safety settings → Ramp at echo lost

Prerequisite Output echo lost (→  141) = Ramp at echo lost

Description Slope of the ramp in the case of a lost echo

User entry Signed floating-point number

Factory setting 0.0 %/min

Additional information
100%
B

A
C

0%
t

min

A0013269

A Delay time echo lost


B Ramp at echo lost (→  142) (positive value)
C Ramp at echo lost (→  142) (negative value)

• The unit for the slope of the ramp is "percentage of the measuring range per minute" (%/
min).
• For a negative slope of the ramp: The measured value is continuously decreased until it
reaches 0%.
• For a positive slope of the ramp: The measured value is continuosly increased until it
reaches 100%.

Blocking distance 

Navigation  Setup → Advanced setup → Safety settings → Blocking distance

Description Specify blocking distance BD.

User entry 0 to 200 m

Factory setting 0 mm (0 in)

Additional information Signals in the blocking distance are only evaluated if they have been outside the blocking
distance when the device was switched on and move into the blocking distance due to a

142 Endress+Hauser
Micropilot FMR67 HART Operating menu

level change during operation. Signals which are already in the blocking distance when the
device is switched on, are ignored.
If required, a different behavior for signals in the blocking distance can be defined by
the Endress+Hauser service.

BD

A0023041

 44 Blocking distance (BD) for bulk solid measurements

Endress+Hauser 143
Operating menu Micropilot FMR67 HART

"SIL/WHG confirmation" wizard


The SIL/WHG confirmation wizard is only available for devices with SIL or WHG
approval (Feature 590: "Additional Approval", option LA: "SIL" or LC: "WHG overfill
prevention" ) which are currently not in the SIL- or WHG-locked state.
The SIL/WHG confirmation wizard is required to lock the device according to SIL or
WHG. For details refer to the "Functional Safety Manual" of the respective device,
which describes the locking procedure and the parameters of the sequence.

Navigation  Setup → Advanced setup → SIL/WHG confirmation

144 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Deactivate SIL/WHG" wizard


The Deactivate SIL/WHG wizard (→  145) is only visible if the device is SIL-locked
or WHG-locked. For details refer to the "Functional Safety Manual" of the respective
device.

Navigation  Setup → Advanced setup → Deactivate SIL/WHG

Reset write protection 

Navigation  Setup → Advanced setup → Deactivate SIL/WHG → Reset write protection

Description Enter unlocking code.

User entry 0 to 65 535

Factory setting 0

Code incorrect 

Navigation  Setup → Advanced setup → Deactivate SIL/WHG → Code incorrect

Description Indicates that a wrong unlocking code has been entered. Select procedure.

Selection • Reenter code


• Abort sequence

Factory setting Reenter code

Endress+Hauser 145
Operating menu Micropilot FMR67 HART

"Current output 1 to 2" submenu


The Current output 2 submenu (→  146) is only available for devices with two
current outputs.

Navigation  Setup → Advanced setup → Current output 1 to 2

Assign current output 1 to 2 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Assign current output

Description Select process variable for current output.

Selection • Level linearized


• Distance
• Electronic temperature
• Relative echo amplitude
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
• Area of incoupling

Factory setting • Current output 1: Level linearized


• Current output 2 8): Level linearized

Additional information Definition of the current range for the process variables
Process variable 4 mA value 20 mA value
1)
Level linearized 0 % or the associated 100 % 2) or the associated linearized value
linearized value

Distance 0 (i.e. level is at the reference Empty calibration (→  117) (i.e. level is at
point) 0 %)

Electronic temperature –50 °C (–58 °F) 100 °C (212 °F)

Analog output adv. depending on the parametrization of the Advanced Diagnostics


diagnostics 1/2

1) the 0% level is defined by Empty calibration parameter (→  117)


2) The 100% level is defined by Full calibration parameter (→  118)

This can be done by the following parameters:


• Expert → Output → Current output 1 to 2 → Turn down
• Expert → Output → Current output 1 to 2 → 4 mA value
• Expert → Output → Current output 1 to 2 → 20 mA value

8) only for devices with two current outputs

146 Endress+Hauser
Micropilot FMR67 HART Operating menu

Current span 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Current span

Description Determines the current range used to transmit the measured value. ‘4…20mA’: Measured
variable: 4 …20 mA ‘4…20mA NAMUR’: Measured variable: 3.8 … 20.5 mA ‘4…20mA US’:
Measured variable: 3.9 … 20.8 mA ‘Fixed current’: Measured variable transmitted via
HART only Note: Currents below 3.6 mA or above 21.95 mA can be used to signal an
alarm.

Selection • 4...20 mA
• 4...20 mA NAMUR
• 4...20 mA US
• Fixed current

Factory setting 4...20 mA NAMUR

Additional information Meaning of the options


Option Current range for process Lower alarm signal level Upper alarm signal level
variable

4...20 mA 4 to 20.5 mA < 3.6 mA > 21.95 mA

4...20 mA NAMUR 3.8 to 20.5 mA < 3.6 mA > 21.95 mA

4...20 mA US 3.9 to 20.8 mA < 3.6 mA > 21.95 mA

Fixed current Constant current, defined in the Fixed current parameter (→  147).

• In the case of an error, the output current assumes the value defined in the Failure
mode parameter (→  148).
• If the meausred value is out of the measuring range, diagnostic message Current
output is issued.
In a HART multidrop loop only one device can use the analog current to transmit a
signal. For all other devices one must set:
• Current span = Fixed current
• Fixed current (→  147) = 4 mA

Fixed current 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Fixed current

Prerequisite Current span (→  147) = Fixed current

Description Define constant value of the output current.

User entry 4 to 22.5 mA

Factory setting 4 mA

Endress+Hauser 147
Operating menu Micropilot FMR67 HART

Damping output 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Damping output

Description Reaction time of the output signal on fluctuation in the measured value.

User entry 0.0 to 999.9 s

Factory setting 0.0 s

Additional information Fluctuations of the measured value affect the output current with an exponential delay,
the time constant τ of which is defined in this parameter. With a small time constant the
output reacts immediately to changes of the measrued value. With a big time constant the
reaction of the output is more delayed. For τ = 0 (factory setting) there is no damping.

Failure mode 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Failure mode

Prerequisite Current span (→  147) ≠ Fixed current

Description Defines which current the output assumes in the case of an error. ‘Min.’: < 3.6mA ‘Max.’: >
21.95mA ‘Last valid value’: Last valid value before occurrence of the error. ‘Actual value’:
Output current is equal to the measured value; error is ignored. ‘Defined value’: User
defined value.

Selection • Min.
• Max.
• Last valid value
• Actual value
• Defined value

Factory setting Max.

Additional information Meaning of the options


• Min.
The current output adopts the value of the lower alarm level according to the Current
span parameter (→  147).
• Max.
The current output adopts the value of the upper alarm level according to the Current
span parameter (→  147).
• Last valid value
The current remains constant at the last value it hat before the error occurred.
• Actual value
The output current follows the actual measured value; the error is ignored.
• Defined value
The output current assumes the value defined in the Failure current parameter
(→  149).
The error behavior of other output channels is not influenced by these settings but is
defined in separate parameters.

148 Endress+Hauser
Micropilot FMR67 HART Operating menu

Failure current 

Navigation  Setup → Advanced setup → Current output 1 to 2 → Failure current

Prerequisite Failure mode (→  148) = Defined value

Description Defines which current the output assumes in case of an error.

User entry 3.59 to 22.5 mA

Factory setting 22.5 mA

Output current 1 to 2

Navigation  Setup → Advanced setup → Current output 1 to 2 → Output current 1 to 2

Description Shows the actual calculated value of the output current.

Endress+Hauser 149
Operating menu Micropilot FMR67 HART

"Switch output" submenu


The Switch output submenu (→  150) is only visible for devices with switch
output. 9)

Navigation  Setup → Advanced setup → Switch output

Switch output function 

Navigation  Setup → Advanced setup → Switch output → Switch output function

Description Defines the function of the switch output. 'Off' The switch output is always open (non-
conductive) 'On' The switch output is always closed (conductive). 'Diagnostic behavior' The
switch output is normally closed and is only opened if a diagnostic event is present. 'Limit'
The switch output is normally closed and is only opened if a measured variable exceeds a
defined limit. 'Digital output' The switch output is controlled by one of the digital output
blocks of the device.

Selection • Off
• On
• Diagnostic behavior
• Limit
• Digital Output

Factory setting Off

Additional information Meaning of the options


• Off
The output is always open (non-conductive).
• On
The output is always closed (conductive).
• Diagnostic behavior
The output is normally closed and is only opened if a diagnostic event is present. The
Assign diagnostic behavior parameter (→  151) determines for which type of event
the output is opened.
• Limit
The output is normally closed and is only opened if a measured variable exceeds or falls
below a defined limit. The limit values are defined by the following parameters:
– Assign limit (→  151)
– Switch-on value (→  152)
– Switch-off value (→  153)
• Digital Output
The switching state of the output tracks the output value of a DI function block. The
function block is selected in the Assign status parameter (→  151).
The Off and On options can be used to simulate the switch output.

9) Ordering feature 020 "Power supply; Output", option B, E or G

150 Endress+Hauser
Micropilot FMR67 HART Operating menu

Assign status 

Navigation  Setup → Advanced setup → Switch output → Assign status

Prerequisite Switch output function (→  150) = Digital Output

Selection • Off
• Digital output AD 1
• Digital output AD 2
• Digital output AD 3
• Digital output AD 4

Factory setting Off

Additional information The Digital output AD 1/2/3/4 options refer to the Advanced Diagnostic Blocks. A switch
signal generated in these blocks can be transmitted via the switch output.

Assign limit 

Navigation  Setup → Advanced setup → Switch output → Assign limit

Prerequisite Switch output function (→  150) = Limit

Selection • Off
• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Relative echo amplitude
• Area of incoupling

Factory setting Off

Assign diagnostic behavior 

Navigation  Setup → Advanced setup → Switch output → Assign diagnostic behavior

Prerequisite Switch output function (→  150) = Diagnostic behavior

Description Defines to which behavior of diagnostic events the switch output reacts.

Selection • Alarm
• Alarm or warning
• Warning

Factory setting Alarm

Endress+Hauser 151
Operating menu Micropilot FMR67 HART

Switch-on value 

Navigation  Setup → Advanced setup → Switch output → Switch-on value

Prerequisite Switch output function (→  150) = Limit

Description Defines the switch-on point. The output is closed if the assigned process variable rises
above this point.

User entry Signed floating-point number

Factory setting 0

Additional information The switching behavior depends on the relative position of the Switch-on value and
Switch-off value parameters:
Switch-on value > Switch-off value
– The output is closed if the measured value is larger than Switch-on value.
– The output is opened if the measured value is smaller than Switch-off value.

B
t

D
t

A0015585

A Switch-on value
B Switch-off value
C Output closed (conductive)
D Output opened (non-conductive)

Switch-on value < Switch-off value


– The output is closed if the measured value is smaller than Switch-on value.
– The output is opened if the measured value is larger than Switch-off value.

152 Endress+Hauser
Micropilot FMR67 HART Operating menu

A
t

D
t

A0015586

A Switch-on value
B Switch-off value
C Output closed (conductive)
D Output opened (non-conductive)

Switch-on delay 

Navigation  Setup → Advanced setup → Switch output → Switch-on delay

Prerequisite • Switch output function (→  150) = Limit


• Assign limit (→  151) ≠ Off

Description Defines the delay applied before the output is switched on.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

Switch-off value 

Navigation  Setup → Advanced setup → Switch output → Switch-off value

Prerequisite Switch output function (→  150) = Limit

Description Defines the switch-off point. The output is opened if the assigned process variable falls
below this point.

User entry Signed floating-point number

Factory setting 0

Additional information The switching behavior depends on the relative position of the Switch-on value and
Switch-off value parameters; description: see the Switch-on value parameter
(→  152).

Endress+Hauser 153
Operating menu Micropilot FMR67 HART

Switch-off delay 

Navigation  Setup → Advanced setup → Switch output → Switch-off delay

Prerequisite • Switch output function (→  150) = Limit


• Assign limit (→  151) ≠ Off

Description Defines the delay applied before the output is switched off.

User entry 0.0 to 100.0 s

Factory setting 0.0 s

Failure mode 

Navigation  Setup → Advanced setup → Switch output → Failure mode

Prerequisite Switch output function (→  150) = Limit or Digital Output

Description Defines the state of the switch output in case of an error.

Selection • Actual status


• Open
• Closed

Factory setting Open

Additional information

Switch status

Navigation  Setup → Advanced setup → Switch output → Switch status

Description Current status of the switch output.

Invert output signal 

Navigation  Setup → Advanced setup → Switch output → Invert output signal

Description 'No' The switch output behaves as per its parameter setting. 'Yes' The switching behavior is
inverted as compared to its parameter setting.

Selection • No
• Yes

154 Endress+Hauser
Micropilot FMR67 HART Operating menu

Factory setting No

Additional information Meaning of the options


• No
The behavior of the switch output is as described above.
• Yes
The states Open and Closed are inverted as compared to the description above.

Endress+Hauser 155
Operating menu Micropilot FMR67 HART

"Display" submenu
The Display submenu is only visible if a display module is connected to the device.

Navigation  Setup → Advanced setup → Display

Language

Navigation  Setup → Advanced setup → Display → Language

Description Set display language.

Selection • English
• Deutsch *
• Français *
• Español *
• Italiano *
• Nederlands *
• Portuguesa *
• Polski *
• русский язык (Russian) *
• Svenska *
• Türkçe *
• 中文 (Chinese) *
• 日本語 (Japanese) *
• 한국어 (Korean) *
• Bahasa Indonesia *
• tiếng Việt (Vietnamese) *
• čeština (Czech) *

Factory setting The language selected in feature 500 of the product structure.
If no language has been selected: English

Additional information

Format display

Navigation  Setup → Advanced setup → Display → Format display

Description Select how measured values are shown on the display.

Selection • 1 value, max. size


• 1 bargraph + 1 value
• 2 values
• 1 value large + 2 values
• 4 values

Factory setting 1 value, max. size

* Visibility depends on order options or device settings

156 Endress+Hauser
Micropilot FMR67 HART Operating menu

Additional information

4841.000
1 Â mm
A0019963

 45 "Format display" = "1 value, max. size"

1 Â
93.5 %
1 Â 159.0
mm
A0019964

 46 "Format display" = "1 bargraph + 1 value"

1 Â 93.5
%
1 Â 159.0
mm
A0019965

 47 "Format display" = "2 values"

1 Â 159.0 mm
V 1 Â 23.5 V
1 Â 93.5
%
A0019966

 48 "Format display" = "1 value large + 2 values"

1 Â 93.5 %
1 Â 159.0 mm
V 1 Â 93.5 V
1 Â 26.3 °C

A0019968

 49 "Format display" = "4 values"

• The Value 1 to 4 display →  158 parameters specify which measured values are
shown on the display and in which order.
• If more measured values are specified than the current display mode permits, the
values alternate on the device display. The display time until the next change is
configured in the Display interval parameter (→  158).

Endress+Hauser 157
Operating menu Micropilot FMR67 HART

Value 1 to 4 display 

Navigation  Setup → Advanced setup → Display → Value 1 display

Description Select the measured value that is shown on the local display.

Selection • Level linearized


• Distance
• Absolute echo amplitude
• Relative echo amplitude
• Area of incoupling
• Current output 1
• Measured current
• Current output 2 *
• Terminal voltage
• Electronic temperature
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
• Analog output adv. diagnostics 3
• Analog output adv. diagnostics 4

Factory setting • Value 1 display: Level linearized


• Value 2 display: None
• Value 3 display: None
• Value 4 display: None

Decimal places 1 to 4 

Navigation  Setup → Advanced setup → Display → Decimal places 1

Description This selection does not affect the measurement and calculation accuracy of the device.

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xx

Additional information The setting does not affect the measuring or computational accuracy of the device.

Display interval

Navigation  Setup → Advanced setup → Display → Display interval

Description Set time measured values are shown on display if display alternates between values.

* Visibility depends on order options or device settings

158 Endress+Hauser
Micropilot FMR67 HART Operating menu

User entry 1 to 10 s

Factory setting 5s

Additional information This parameter is only relevant if the number of selected measuring values exceeds the
number of values the selected display format can display simultaneously.

Display damping 

Navigation  Setup → Advanced setup → Display → Display damping

Description Set display reaction time to fluctuations in the measured value.

User entry 0.0 to 999.9 s

Factory setting 0.0 s

Header 

Navigation  Setup → Advanced setup → Display → Header

Description Select header contents on local display.

Selection • Device tag


• Free text

Factory setting Device tag

Additional information
1 XXXXXXXXX

A0029422

1 Position of the header text on the display

Meaning of the options


• Device tag
Is defined in the Device tag parameter (→  115)
• Free text
Is defined in the Header text parameter (→  160)

Endress+Hauser 159
Operating menu Micropilot FMR67 HART

Header text 

Navigation  Setup → Advanced setup → Display → Header text

Prerequisite Header (→  159) = Free text

Description Enter display header text.

Factory setting ------------

Additional information The number of characters which can be displayed depends on the characters used.

Separator 

Navigation  Setup → Advanced setup → Display → Separator

Description Select decimal separator for displaying numerical values.

Selection •.
•,

Factory setting .

Number format 

Navigation  Setup → Advanced setup → Display → Number format

Description Choose number format for the display.

Selection • Decimal
• ft-in-1/16''

Factory setting Decimal

Additional information The ft-in-1/16'' option is only valid for distance units.

Decimal places menu 

Navigation  Setup → Advanced setup → Display → Decimal places menu

Description Select number of decimal places for the representation of numbers within the operating
menu.

160 Endress+Hauser
Micropilot FMR67 HART Operating menu

Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx

Factory setting x.xxx

Additional information • Is only valid for numbers in the operating menu (e.g. Empty calibration, Full
calibration), but not for the measured value display. The number of decimal places for
the measured value display is defined in the Decimal places 1 to 4 →  158
parameters.
• The setting does not affect the accuracy of the measurement or the calculations.

Backlight

Navigation  Setup → Advanced setup → Display → Backlight

Prerequisite The device has the SD03 local display (with optical keys).

Description Switch the local display backlight on and off.

Selection • Disable
• Enable

Factory setting Disable

Additional information Meaning of the options


• Disable
Switches the backlight off.
• Enable
Switches the backlight on.
Regardless of the setting in this parameter the backlight may be automatically
switched off by the device if the supply voltage is too low.

Contrast display

Navigation  Setup → Advanced setup → Display → Contrast display

Description Adjust local display contrast setting to ambient conditions (e.g. lighting or reading angle).

User entry 20 to 80 %

Factory setting Dependent on the display.

Additional information Setting the contrast via push-buttons:


• Darker: press the E
buttons simultaneously.
• Brighter: press the buttons simultaneously.
E

Endress+Hauser 161
Operating menu Micropilot FMR67 HART

"Configuration backup display" submenu


This submenu is only visible if a display module is connected to the device.

The configuration of the device can be saved to the display module at a certain point of
time (backup). The saved configurateion can be restored to the device if required, e.g. in
order to bring the device back into a defined state. The configuration can also be
transferred to a different device of the same type using the display module.

Navigation  Setup → Advanced setup → Configuration backup display

Operating time

Navigation  Setup → Advanced setup → Configuration backup display → Operating time

Description Indicates how long the device has been in operation.

Additional information Maximum time


9 999 d ( ≈ 27 years)

Last backup

Navigation  Setup → Advanced setup → Configuration backup display → Last backup

Description Indicates when the last data backup was saved to the display module.

Configuration management 

Navigation  Setup → Advanced setup → Configuration backup display → Configuration


management

Description Select action for managing the device data in the display module.

Selection • Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data

Factory setting Cancel

162 Endress+Hauser
Micropilot FMR67 HART Operating menu

Additional information Meaning of the options


• Cancel
No action is executed and the user exits the parameter.
• Execute backup
A backup copy of the current device configuration in the HistoROM (built-in in the
device) is saved to the display module of the device.
• Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device.
• Duplicate
The transmitter configuration is duplicated to another device using the transmitter
display module. The following parameters, which characterize the individual measuring
point are not included in the transmitted configuration:
– HART date code
– HART short tag
– HART message
– HART descriptor
– HART address
– Device tag
– Medium type
• Compare
The device configuration saved in the display module is compared to the current device
configuration of the HistoROM. The result of this comparison is displayed in the
Comparison result parameter (→  163).
• Clear backup data
The backup copy of the device configuration is deleted from the display module of the
device.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
If an existing backup is restored to a different device using the Restore option, it may
occur that some device functionalities are no longer available. In some cases even a
device reset will not restore the original status.
In order to transmit a configuration to a different device, the Duplicate option should
always be used.

Backup state

Navigation  Setup → Advanced setup → Configuration backup display → Backup state

Description Displays which backup action is currently in progress.

Comparison result

Navigation  Setup → Advanced setup → Configuration backup display → Comparison result

Description Comparison between present device data and display backup.

Endress+Hauser 163
Operating menu Micropilot FMR67 HART

Additional information Meaning of the display options


• Settings identical
The current device configuration of the HistoROM is identical to the backup copy in the
display module.
• Settings not identical
The current device configuration of the HistoROM is not identical to the backup copy in
the display module.
• No backup available
There is no backup copy of the device configuration of the HistoROM in the display
module.
• Backup settings corrupt
The current device configuration of the HistoROM is corrupt or not compatible with the
backup copy in the display module.
• Check not done
The device configuration of the HistoROM has not yet been compared to the backup copy
in the display module.
• Dataset incompatible
The data sets are incompatible and can not be compared.
To start the comparison, set Configuration management (→  162) = Compare.

If the transmitter configuration has been duplicated from a different device by


Configuration management (→  162) = Duplicate, the new device configuration
in the HistoROM is only partially identical to the configuration stored in the display
module: Sensor specific properties (e.g. the mapping curve) are not duplicated. Thus,
the result of the comparison will be Settings not identical.

164 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Administration" submenu

Navigation  Setup → Advanced setup → Administration

Define access code 

Navigation  Setup → Advanced setup → Administration → Define access code

Description Define release code for write access to parameters.

User entry 0 to 9 999

Factory setting 0

Additional information If the factory setting is not changed or 0 is defined as the access code, the parameters
are not write-protected and the configuration data of the device can then always be
modified. The user is logged on in the Maintenance role.
The write protection affects all parameters marked with the symbol in this
document. On the local display, the symbol in front of a parameter indicates that
the parameter is write-protected.
Once the access code has been defined, write-protected parameters can only be
modified if the access code is entered in the Enter access code parameter
(→  125).
Please contact your Endress+Hauser Sales Center if you lose your access code.

For display operation: The new access code is only valid after it has been confirmed in
the Confirm access code parameter (→  167).

Device reset 

Navigation  Setup → Advanced setup → Administration → Device reset

Description Reset the device configuration - either entirely or in part - to a defined state.

Selection • Cancel
• To factory defaults
• To delivery settings
• Of customer settings
• To transducer defaults
• Restart device

Factory setting Cancel

Endress+Hauser 165
Operating menu Micropilot FMR67 HART

Additional information Meaning of the options


• Cancel
No action
• To factory defaults
All parameters are reset to the order-code specific factory setting.
• To delivery settings
All parameters are reset to the delivery setting. The delivery setting may differ from the
factory default if customer specific settings have been ordered.
This option is only visible if customer specific settings have been ordered.
• Of customer settings
All customer parameters are reset to their factory setting. Service parameters, however,
remain unchanged.
• To transducer defaults
Every measurment-related parameter is reset to its factory setting. Service parameters
and communication-related parameters, however, remain unchanged.
• Restart device
The restart resets every parameter which is stored in the volatile memory (RAM) to the
factory setting (e.g. measured value data). The device configuration remains unchanged.

166 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Define access code" wizard

The Define access code wizard is only available when operating via the local display.
When operating via an operating tool, the Define access code parameter is located
directly in the Administration submenu. The Confirm access code parameter is not
available for operation via operating tool.

Navigation  Setup → Advanced setup → Administration → Define access


code

Define access code 

Navigation  Setup → Advanced setup → Administration → Define access code → Define access
code

Description →  165

Confirm access code 

Navigation  Setup → Advanced setup → Administration → Define access code → Confirm access
code

Description Confirm the entered access code.

User entry 0 to 9 999

Factory setting 0

Endress+Hauser 167
Operating menu Micropilot FMR67 HART

17.5 "Diagnostics" menu


Navigation  Diagnostics

Actual diagnostics

Navigation  Diagnostics → Actual diagnostics

Description Displays current diagnostic message.

Additional information The display consists of:


• Symbol for event behavior
• Code for diagnostic behavior
• Operating time of occurrence
• Event text
If several messages are active at the same time, the messages with the highest
priority is displayed.
Information on what is causing the message, and remedy measures, can be viewed via
the symbol on the display.

Timestamp

Navigation  Diagnostics → Timestamp

Description Displays the timestamp for the currently active diagnostic message.

Previous diagnostics

Navigation  Diagnostics → Previous diagnostics

Description Displays the last diagnostic message which has been active before the current message.

Additional information The display consists of:


• Symbol for event behavior
• Code for diagnostic behavior
• Operating time of occurrence
• Event text
The condition displayed may still apply. Information on what is causing the message,
and remedy measures, can be viewed via the symbol on the display.

168 Endress+Hauser
Micropilot FMR67 HART Operating menu

Timestamp

Navigation  Diagnostics → Timestamp

Description Shows the timestamp of the previous diagnostic message.

Operating time from restart

Navigation  Diagnostics → Operating time from restart

Description Displays the time the device has been in operation since the last device restart.

Operating time

Navigation  Diagnostics → Operating time

Description Indicates how long the device has been in operation.

Additional information Maximum time


9 999 d ( ≈ 27 years)

Endress+Hauser 169
Operating menu Micropilot FMR67 HART

17.5.1 "Diagnostic list" submenu

Navigation  Diagnostics → Diagnostic list

Diagnostics 1 to 5

Navigation  Diagnostics → Diagnostic list → Diagnostics 1

Description Display the current diagnostics messages with the highest to fifth-highest priority.

Additional information The display consists of:


• Symbol for event behavior
• Code for diagnostic behavior
• Operating time of occurrence
• Event text

Timestamp 1 to 5

Navigation  Diagnostics → Diagnostic list → Timestamp

Description Timestamp of the diagnostic message.

170 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.5.2 "Event logbook" submenu


The Event logbook submenu is only available when operating via the local display.
When operating via FieldCare, the event list can be displayed in the FieldCare
function "Event List / HistoROM".

Navigation  Diagnostics → Event logbook

Filter options 

Navigation  Diagnostics → Event logbook → Filter options

Description Define which category of event messages is shown in the Events list submenu.

Selection • All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)

Factory setting All

Additional information • This parameter is only used for operation via the local display.
• The status signals are categorized according to NAMUR NE 107.

"Event list" submenu


The Event list submenu displays the history of past events of the category selected in the
Filter options parameter (→  171). A maximum of 100 events are displayed in
chronological order.
The following symbols indicate whether an event has occurred or has ended:
• : Event has occurred
• : Event has ended
Information on what is causing the message, and remedy instructions, can be viewed
via the -button.
Display format
• For event messages in category I: information event, event text, "recording event" symbol
and time the event occurred
• For event messages in category F, M, C, S (status signal): diagnostics event, event text,
"recording event" symbol and time the event occurred

Navigation  Diagnostics → Event logbook → Event list

Endress+Hauser 171
Operating menu Micropilot FMR67 HART

17.5.3 "Device information" submenu

Navigation  Diagnostics → Device information

Device tag

Navigation  Diagnostics → Device information → Device tag

Description Enter the name for the measuring point.

Factory setting FMR6x

Serial number

Navigation  Diagnostics → Device information → Serial number

Description Shows the serial number of the measuring device.

Additional information Uses of the serial number


• To identify the device quickly, e.g. when contacting Endress+Hauser.
• To obtain specific information on the device using the Device Viewer:
www.endress.com/deviceviewer
The serial number is also indicated on the nameplate.

Firmware version

Navigation  Diagnostics → Device information → Firmware version

Description Shows the device firmware version installed.

User interface xx.yy.zz

Additional information For firmware versions differing only in the last two digits ("zz") there is no difference
concerning functionality or operation.

Device name

Navigation  Diagnostics → Device information → Device name

Description Shows the name of the transmitter.

172 Endress+Hauser
Micropilot FMR67 HART Operating menu

Order code 

Navigation  Diagnostics → Device information → Order code

Description Shows the device order code.

Additional information The order code is generated from the extended roder code, which defines all device
features of the product structure. In contrast, the device features can not be read directly
from the order code.

Extended order code 1 to 3 

Navigation  Diagnostics → Device information → Extended order code 1

Description Display the three parts of the extended order code.

Additional information The extended order code indicates the version of all the features of the product structure
and thus uniquely identifies the device.

Device revision

Navigation  Diagnostics → Device information → Device revision

Description Shows the device revision with which the device is registered with the HART
Communication Foundation.

Additional information The device revision is used to allocate the correct Device Description file (DD) to the device.

Device ID

Navigation  Diagnostics → Device information → Device ID

Description Shows the device ID for identifying the device in a HART network.

Additional information In addition to the Device type and Manufacturer ID, the Device ID is part of the unique
device identification (Unique ID) which characterizes each HART device unambiguously.

Endress+Hauser 173
Operating menu Micropilot FMR67 HART

Device type

Navigation  Diagnostics → Device information → Device type

Description Shows the device type with which the measuring device is registered with the HART
Communication Foundation.

Additional information The device type is needed to allocate the suitable Device Description (DD) to the device.

Manufacturer ID

Navigation  Diagnostics → Device information → Manufacturer ID

Description Use this function to view the manufacturer ID with which the measuring device is
registered with the HART Communication Foundation.

User interface 2-digit hexadecimal number

Factory setting 0x11 (for Endress+Hauser)

174 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.5.4 "Measured values" submenu

Navigation  Diagnostics → Measured values

Distance

Navigation  Diagnostics → Measured values → Distance

Description Distance between lower edge of flange or thread and medium surface.

Additional information

A0019485

 50 Distance for bulk solid measurements

Level linearized

Navigation  Diagnostics → Measured values → Level linearized

Description Displays linearized level.

Additional information The unit is defined by the Unit after linearization parameter →  135.

Output current 1 to 2

Navigation  Diagnostics → Measured values → Output current 1 to 2

Description Shows the actual calculated value of the output current.

Endress+Hauser 175
Operating menu Micropilot FMR67 HART

Measured current 1

Navigation  Diagnostics → Measured values → Measured current 1

Prerequisite Only available for current output 1

Description Shows the current value of the current output which is currently measured.

Terminal voltage 1

Navigation  Diagnostics → Measured values → Terminal voltage 1

Description Shows the current terminal voltage that is applied at the output.

Sensor temperature

Navigation  Diagnostics → Measured values → Sensor temperature

Description Indicates the current sensor temperature.

176 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.5.5 "Data logging" submenu

Navigation  Diagnostics → Data logging

Assign channel 1 to 4 

Navigation  Diagnostics → Data logging → Assign channel 1 to 4

Description Assign a process variable to logging channel.

Selection • Off
• Level linearized
• Distance
• Current output 1
• Measured current
• Current output 2 *
• Terminal voltage
• Electronic temperature
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
• Analog output adv. diagnostics 3
• Analog output adv. diagnostics 4

Factory setting Off

Additional information A total of 1000 measured values can be logged. This means:
• 1000 data points if 1 logging channel is used
• 500 data points if 2 logging channels are used
• 333 data points if 3 logging channels are used
• 250 data points if 4 logging channels are used
If the maximum number of data points is reached, the oldest data points in the data log
are cyclically overwritten in such a way that the last 1000, 500, 333 or 250 measured
values are always in the log (ring memory principle).
The logged data are deleted if a new option is selected in this parameter.

Logging interval 

Navigation  Diagnostics → Data logging → Logging interval


 Diagnostics → Data logging → Logging interval

Description Define the logging interval tlog for data logging. This value defines the time interval
between the individual data points in the memory.

User entry 1.0 to 3 600.0 s

Factory setting 30.0 s

* Visibility depends on order options or device settings

Endress+Hauser 177
Operating menu Micropilot FMR67 HART

Additional information This parameter defines the interval between the individual data points in the data log, and
thus the maximum loggable process time T log :
• If 1 logging channel is used: T log = 1000 ⋅ t log
• If 2 logging channels are used: T log = 500 ⋅ t log
• If 3 logging channels are used: T log = 333 ⋅ t log
• If 4 logging channels are used: T log = 250 ⋅ t log
Once this time elapses, the oldest data points in the data log are cyclically overwritten such
that a time of T log always remains in the memory (ring memory principle).
The logged data are deleted if this parameter is changed.

Example
When using 1 logging channel
• Tlog = 1000 ⋅ 1 s = 1 000 s ≈ 16.5 min
• Tlog = 1000 ⋅ 10 s = 1 000 s ≈ 2.75 h
• Tlog = 1000 ⋅ 80 s = 80 000 s ≈ 22 h
• Tlog = 1000 ⋅ 3 600 s = 3 600 000 s ≈ 41 d

Clear logging data 

Navigation  Diagnostics → Data logging → Clear logging data


 Diagnostics → Data logging → Clear logging data

Description Clear the entire logging data.

Selection • Cancel
• Clear data

Factory setting Cancel

178 Endress+Hauser
Micropilot FMR67 HART Operating menu

"Display channel 1 to 4" submenu


The Display channel 1 to 4 submenus are only available for operation via the local
display. When operating via FieldCare, the logging diagram can be displayed in the
FieldCare function "Event List / HistoROM" .
The Display channel 1 to 4 submenus invoke a diagram of the logging history of the
respective channel.

• x-axis: depending on the number of selected channels, 250 to 1000 measured values of
a process variable are displayed.
• y-axis: covers the approximate measured value span and constantly adapts this to the
measurement.
To return to the operating menu, press  and  simultaneaously.

Navigation  Diagnostics → Data logging → Display channel 1 to 4

Endress+Hauser 179
Operating menu Micropilot FMR67 HART

17.5.6 "Simulation" submenu


The Simulation submenu is used to simulate specific measuring values or other
conditions. This helps to check the correct configuration of the device and connected
control units.

Conditions which can be simulated


Condition to be simulated Associated parameters

Specific value of a process variable • Assign measurement variable (→  182)


• Process variable value (→  182)

Specific value of the output current • Current output simulation (→  182)


• Value current output (→  183)

Specific state of the switch output • Switch output simulation (→  183)


• Switch status (→  183)

Existence of an alarm Device alarm simulation (→  184)

180 Endress+Hauser
Micropilot FMR67 HART Operating menu

Structure of the submenu

Navigation  Diagnostics → Simulation

‣ Simulation

Assign measurement variable →  182

Process variable value →  182

Current output 1 to 2 simulation →  182

Value current output 1 to 2 →  183

Switch output simulation →  183

Switch status →  183

Device alarm simulation →  184

Endress+Hauser 181
Operating menu Micropilot FMR67 HART

Description of parameters

Navigation  Diagnostics → Simulation

Assign measurement variable 

Navigation  Diagnostics → Simulation → Assign measurement variable

Selection • Off
• Level
• Level linearized

Factory setting Off

Additional information • The value of the variable to be simulated is defined in the Process variable value
parameter (→  182).
• If Assign measurement variable ≠ Off, a simulation is active. This is indicated by a
diagnotic message of the Function check (C) category.

Process variable value 

Navigation  Diagnostics → Simulation → Process variable value

Prerequisite Assign measurement variable (→  182) ≠ Off

User entry Signed floating-point number

Factory setting 0

Additional information Downstream measured value processing and the signal output use this simulation value. In
this way, users can verify whether the measuring device has been configured correctly.

Current output 1 to 2 simulation 

Navigation  Diagnostics → Simulation → Current output 1 to 2 simulation

Description Switch the simulation of the current output on and off.

Selection • Off
• On

Factory setting Off

Additional information An active simulation is indicated by a diagnostic message of the Function check (C)
category.

182 Endress+Hauser
Micropilot FMR67 HART Operating menu

Value current output 1 to 2 

Navigation  Diagnostics → Simulation → Value current output 1 to 2

Prerequisite Current output simulation (→  182) = On

Description Defines the value of the simulated output current.

User entry 3.59 to 22.5 mA

Factory setting 3.59 mA

Additional information The current output assumes the value specified in this parameter. In this way, users can
verify the correct adjustment of the current output and the correct function of connected
control units.

Switch output simulation 

Navigation  Diagnostics → Simulation → Switch output simulation

Description Switch the simulation of the switch output on and off.

Selection • Off
• On

Factory setting Off

Switch status 

Navigation  Diagnostics → Simulation → Switch status

Prerequisite Switch output simulation (→  183) = On

Description Current status of the switch output.

Selection • Open
• Closed

Factory setting Open

Additional information The switch status assumes the value defined in this parameter. This helps to check correct
operation of connected control units.

Endress+Hauser 183
Operating menu Micropilot FMR67 HART

Device alarm simulation 

Navigation  Diagnostics → Simulation → Device alarm simulation

Description Switch the device alarm on and off.

Selection • Off
• On

Factory setting Off

Additional information When selecting the On option, the device generates an alarm. This helps to check the
correct output behavior of the device in the case of an alarm.
An active simulation is indicated by the diagnostic message C484 Failure mode
simulation.

Diagnostic event category 

Navigation  Diagnostics → Simulation → Diagnostic event category

Description Select event category for the simulation.

Selection • Sensor
• Electronics
• Configuration
• Process

Factory setting Process

Additional information Only events of the selected category are available in the selcetion list of the Diagnostic
event simulation parameter (→  184).
When operated via tool, all diagnostic messages are always available in Diagnostic
event simulation. Therefore, Diagnostic event category appears only on the local
display.

Diagnostic event simulation 

Navigation  Diagnostics → Simulation → Diagnostic event simulation

Description Select the diagnostic event to be simulated. Note: To terminate the simulation, select 'Off'.

Factory setting Off

Additional information When operated via the local display, the selection list can be filtered according to the event
categories (Diagnostic event category parameter (→  184)).

184 Endress+Hauser
Micropilot FMR67 HART Operating menu

17.5.7 "Device check" submenu

Navigation  Diagnostics → Device check

Start device check 

Navigation  Diagnostics → Device check → Start device check

Description Yes starts a device check.

Selection • No
• Yes

Factory setting No

Additional information • In case of an echo loss a device check cannot be performed.


• The minimum distance to the medium is 1.5 m (5 ft).

Result device check

Navigation  Diagnostics → Device check → Result device check

Last check time

Navigation  Diagnostics → Device check → Last check time

Description Indicates the operating time at which the last device check has been performed.

Endress+Hauser 185
Operating menu Micropilot FMR67 HART

17.5.8 "Heartbeat" submenu


The Heartbeat submenu is only available via FieldCare or DeviceCare. It contains the
wizards which are part of the Heartbeat Verification and Heartbeat Monitoring
application packages.
Detailed description
SD01870F

Navigation  Diagnostics → Heartbeat

186 Endress+Hauser
Micropilot FMR67 HART Index

Index
A Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 11
Access authorization to parameters Define access code . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Define access code (Parameter) . . . . . . . . . . . . 165, 167
Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Define access code (Wizard) . . . . . . . . . . . . . . . . . . 167
Access code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Incorrect input . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Device alarm simulation (Parameter) . . . . . . . . . . . . 184
Access status display (Parameter) . . . . . . . . . . . . . . 125 Device check (Submenu) . . . . . . . . . . . . . . . . . . . . . 185
Access status tooling (Parameter) . . . . . . . . . . . . . . 124 Device Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Accessories Device ID (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 173
Communication specific . . . . . . . . . . . . . . . . . . . . 95 Device information (Submenu) . . . . . . . . . . . . . . . . 172
Device-specific . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Device name (Parameter) . . . . . . . . . . . . . . . . . . . . 172
Service specific . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Device replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Activate table (Parameter) . . . . . . . . . . . . . . . . . . . . 140 Device reset (Parameter) . . . . . . . . . . . . . . . . . . . . . 165
Actual diagnostics (Parameter) . . . . . . . . . . . . . . . . 168 Device revision (Parameter) . . . . . . . . . . . . . . . . . . . 173
Administration (Submenu) . . . . . . . . . . . . . . . . . . . 165 Device tag (Parameter) . . . . . . . . . . . . . . . . . . 115, 172
Advanced setup (Submenu) . . . . . . . . . . . . . . . . . . . 124 Device type (Parameter) . . . . . . . . . . . . . . . . . . . . . 174
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Diagnostic event
Assign channel 1 to 4 (Parameter) . . . . . . . . . . . . . . 177 In the operating tool . . . . . . . . . . . . . . . . . . . . . . . 79
Assign current output (Parameter) . . . . . . . . . . . . . . 146 Diagnostic event category (Parameter) . . . . . . . . . . . 184
Assign diagnostic behavior (Parameter) . . . . . . . . . . 151 Diagnostic event simulation (Parameter) . . . . . . . . . 184
Assign limit (Parameter) . . . . . . . . . . . . . . . . . . . . . 151 Diagnostic events . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Assign measurement variable (Parameter) . . . . . . . . 182 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Assign status (Parameter) . . . . . . . . . . . . . . . . . . . . 151 Diagnostic list (Submenu) . . . . . . . . . . . . . . . . . . . . 170
Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . . 76
B Diagnostics
Backlight (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 161 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Backup state (Parameter) . . . . . . . . . . . . . . . . . . . . 163 Diagnostics (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . 168
Bin type (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 115 Diagnostics 1 (Parameter) . . . . . . . . . . . . . . . . . . . . 170
Blocking distance (Parameter) . . . . . . . . . . . . . 129, 142 Diagnostics event . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Bluetooth® wireless technology . . . . . . . . . . . . . . . . . 43 Diameter (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 137
DIP switch
C see Write protection switch
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Display (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . 156
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Display and operating module FHX50 . . . . . . . . . . . . 42
Clear logging data (Parameter) . . . . . . . . . . . . . . . . 178 Display channel 1 to 4 (Submenu) . . . . . . . . . . . . . . 179
Code incorrect (Parameter) . . . . . . . . . . . . . . . . . . . 145 Display damping (Parameter) . . . . . . . . . . . . . . . . . 159
Comparison result (Parameter) . . . . . . . . . . . . . . . . 163 Display interval (Parameter) . . . . . . . . . . . . . . . . . . 158
Configuration backup display (Submenu) . . . . . . . . . 162 Display module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Configuration management (Parameter) . . . . . . . . . 162 Display symbols for submenus . . . . . . . . . . . . . . . . . . 53
Configuration of a level measurement . . . . . . . . . . . . 67 Display symbols for the locking state . . . . . . . . . . . . . 53
Confirm access code (Parameter) . . . . . . . . . . . . . . . 167 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Confirm distance (Parameter) . . . . . . . . . . . . . 119, 122 Distance (Parameter) . . . . . . . . . . . . . . . . 119, 122, 175
Context menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Distance unit (Parameter) . . . . . . . . . . . . . . . . . . . . 115
Contrast display (Parameter) . . . . . . . . . . . . . . . . . . 161 Document
Current output 1 to 2 (Submenu) . . . . . . . . . . . . . . . 146 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Current output 1 to 2 simulation (Parameter) . . . . . . 182 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Current span (Parameter) . . . . . . . . . . . . . . . . . . . . 147
Customer value (Parameter) . . . . . . . . . . . . . . . . . . 140 E
Electronics housing
D Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Damping output (Parameter) . . . . . . . . . . . . . . . . . . 148 Empty calibration (Parameter) . . . . . . . . . . . . . . . . . 117
Data logging (Submenu) . . . . . . . . . . . . . . . . . . . . . 177 Enter access code (Parameter) . . . . . . . . . . . . . . . . . 125
DD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Envelope curve display . . . . . . . . . . . . . . . . . . . . . . . . 59
Deactivate SIL/WHG (Wizard) . . . . . . . . . . . . . . . . . 145 Event history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Decimal places 1 (Parameter) . . . . . . . . . . . . . . . . . 158 Event level
Decimal places menu (Parameter) . . . . . . . . . . . . . . 160 Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Endress+Hauser 187
Index Micropilot FMR67 HART

Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Locking status (Parameter) . . . . . . . . . . . . . . . . . . . 124


Event list (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 171 Logging interval (Parameter) . . . . . . . . . . . . . . . . . . 177
Event logbook (Submenu) . . . . . . . . . . . . . . . . . . . . 171
Event text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 M
Events list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Extended order code 1 (Parameter) . . . . . . . . . . . . . 173 Manage device configuration . . . . . . . . . . . . . . . . . . . 71
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Manufacturer ID (Parameter) . . . . . . . . . . . . . . . . . 174
Mapping (Wizard) . . . . . . . . . . . . . . . . . . . . . . . . . . 122
F Mapping end point (Parameter) . . . . . . . . . . . . 121, 122
Failure current (Parameter) . . . . . . . . . . . . . . . . . . . 149 Max. draining speed liquid (Parameter) . . . . . . . . . . 127
Failure mode (Parameter) . . . . . . . . . . . . . . . . 148, 154 Max. draining speed solid (Parameter) . . . . . . . . . . . 116
FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Max. filling speed liquid (Parameter) . . . . . . . . . . . . 127
Field of application Max. filling speed solid (Parameter) . . . . . . . . . . . . . 116
Residual risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Maximum value (Parameter) . . . . . . . . . . . . . . . . . . 137
Filter options (Parameter) . . . . . . . . . . . . . . . . . . . . 171 Measured current 1 (Parameter) . . . . . . . . . . . . . . . 176
Filtering the event logbook . . . . . . . . . . . . . . . . . . . . 83 Measured value symbols . . . . . . . . . . . . . . . . . . . . . . 54
Firmware version (Parameter) . . . . . . . . . . . . . . . . . 172 Measured values (Submenu) . . . . . . . . . . . . . . . . . . 175
Fixed current (Parameter) . . . . . . . . . . . . . . . . . . . . 147 Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Format display (Parameter) . . . . . . . . . . . . . . . . . . . 156 Medium property (Parameter) . . . . . . . . . . . . . . . . . 126
Free text (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 136 Medium type (Parameter) . . . . . . . . . . . . . . . . . . . . 126
Full calibration (Parameter) . . . . . . . . . . . . . . . . . . . 118 Menu
FV (HART device variable) . . . . . . . . . . . . . . . . . . . . . 60 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
H
Hardware write protection . . . . . . . . . . . . . . . . . . . . . 48 N
HART device variables . . . . . . . . . . . . . . . . . . . . . . . . 60 Number format (Parameter) . . . . . . . . . . . . . . . . . . 160
HART integration . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
HART protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 O
Header (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 159 Operating elements
Header text (Parameter) . . . . . . . . . . . . . . . . . . . . . 160 Diagnostics message . . . . . . . . . . . . . . . . . . . . . . . 77
Heartbeat (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 186 Operating module . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Housing Operating time (Parameter) . . . . . . . . . . . . . . . 162, 169
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Operating time from restart (Parameter) . . . . . . . . . 169
Turning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Order code (Parameter) . . . . . . . . . . . . . . . . . . . . . . 173
I Output current 1 to 2 (Parameter) . . . . . . . . . . 149, 175
Input mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Output echo lost (Parameter) . . . . . . . . . . . . . . . . . . 141
Intermediate height (Parameter) . . . . . . . . . . . . . . . 137 Overvoltage protection
Invert output signal (Parameter) . . . . . . . . . . . . . . . 154 General information . . . . . . . . . . . . . . . . . . . . . . . 37

K P
Keypad lock Prepare recording map (Parameter) . . . . . . . . . . . . . 122
Disabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Present mapping (Parameter) . . . . . . . . . . . . . . . . . 121
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Previous diagnostics (Parameter) . . . . . . . . . . . . . . . 168
Process variable value (Parameter) . . . . . . . . . . . . . . 182
L Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Language (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 156 PV (HART device variable) . . . . . . . . . . . . . . . . . . . . . 60
Last backup (Parameter) . . . . . . . . . . . . . . . . . . . . . 162
Last check time (Parameter) . . . . . . . . . . . . . . . . . . 185 R
Level (Parameter) . . . . . . . . . . . . . . . . . . 119, 139, 140 Ramp at echo lost (Parameter) . . . . . . . . . . . . . . . . . 142
Level (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Level correction (Parameter) . . . . . . . . . . . . . . . . . . 129 Record map (Parameter) . . . . . . . . . . . . . . . . . 121, 122
Level linearized (Parameter) . . . . . . . . . . . . . . 136, 175 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . . 9
Level measurement configuration . . . . . . . . . . . . . . . 67 Remedial measures
Level unit (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 128 Calling up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Linearization (Submenu) . . . . . . . . . . . . . 132, 133, 134 Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Linearization type (Parameter) . . . . . . . . . . . . . . . . 134 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Repair concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
see Diagnostics message Replacing a device . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
see In alarm condition Requirements for personnel . . . . . . . . . . . . . . . . . . . . 10

188 Endress+Hauser
Micropilot FMR67 HART Index

Reset write protection (Parameter) . . . . . . . . . . . . . 145 System components . . . . . . . . . . . . . . . . . . . . . . . . . . 97


Result device check (Parameter) . . . . . . . . . . . . . . . . 185
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 T
Table mode (Parameter) . . . . . . . . . . . . . . . . . . . . . 138
S Table number (Parameter) . . . . . . . . . . . . . . . . . . . . 139
Safety instructions Tank/silo height (Parameter) . . . . . . . . . . . . . . . . . 129
Basic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Terminal voltage 1 (Parameter) . . . . . . . . . . . . . . . . 176
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . 12 Timestamp (Parameter) . . . . . . . . . . . . . . 168, 169, 170
Safety settings (Submenu) . . . . . . . . . . . . . . . . . . . . 141 Transmitter
Sensor temperature (Parameter) . . . . . . . . . . . . . . . 176 Turning the display . . . . . . . . . . . . . . . . . . . . . . . 29
Separator (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 160 Turning the display module . . . . . . . . . . . . . . . . . 30
Serial number (Parameter) . . . . . . . . . . . . . . . . . . . 172 Transmitter housing
Service interface (CDI) . . . . . . . . . . . . . . . . . . . . . . . . 44 Turning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Setting the operating language . . . . . . . . . . . . . . . . . 66 Turning the display . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Settings Turning the display module . . . . . . . . . . . . . . . . . . . . 30
Manage device configuration . . . . . . . . . . . . . . . . 71 TV (HART device variable) . . . . . . . . . . . . . . . . . . . . . 60
Operating language . . . . . . . . . . . . . . . . . . . . . . . 66
Setup (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 U
Signal quality (Parameter) . . . . . . . . . . . . . . . . . . . . 119 Unit after linearization (Parameter) . . . . . . . . . . . . . 135
SIL/WHG confirmation (Wizard) . . . . . . . . . . . . . . . 144 Use of the measuring device
Simulation (Submenu) . . . . . . . . . . . . . . . . . . . 181, 182 see Designated use
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Use of the measuring devices
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Borderline cases . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Start device check (Parameter) . . . . . . . . . . . . . . . . . 185 Incorrect use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Status signals . . . . . . . . . . . . . . . . . . . . . . . . . . . 53, 76
Störungsbehebung . . . . . . . . . . . . . . . . . . . . . . . . . . 73 V
Submenu Value 1 display (Parameter) . . . . . . . . . . . . . . . . . . 158
Administration . . . . . . . . . . . . . . . . . . . . . . . . . . 165 Value current output 1 to 2 (Parameter) . . . . . . . . . . 183
Advanced setup . . . . . . . . . . . . . . . . . . . . . . . . . 124 Value echo lost (Parameter) . . . . . . . . . . . . . . . . . . . 141
Configuration backup display . . . . . . . . . . . . . . . 162
Current output 1 to 2 . . . . . . . . . . . . . . . . . . . . . 146 W
Data logging . . . . . . . . . . . . . . . . . . . . . . . . . . . 177 W@M Device Viewer . . . . . . . . . . . . . . . . . . . . . . . . . 87
Device check . . . . . . . . . . . . . . . . . . . . . . . . . . . 185 Wizard
Device information . . . . . . . . . . . . . . . . . . . . . . . 172 Deactivate SIL/WHG . . . . . . . . . . . . . . . . . . . . . . 145
Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . 170 Define access code . . . . . . . . . . . . . . . . . . . . . . . 167
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156 Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Display channel 1 to 4 . . . . . . . . . . . . . . . . . . . . 179 SIL/WHG confirmation . . . . . . . . . . . . . . . . . . . . 144
Event list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Event logbook . . . . . . . . . . . . . . . . . . . . . . . . . . 171 Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Events list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Write protection
Heartbeat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186 Via access code . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Via write protection switch . . . . . . . . . . . . . . . . . . 48
Linearization . . . . . . . . . . . . . . . . . . . 132, 133, 134 Write protection switch . . . . . . . . . . . . . . . . . . . . . . . 48
Measured values . . . . . . . . . . . . . . . . . . . . . . . . 175
Safety settings . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . 181, 182
Switch output . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
SV (HART device variable) . . . . . . . . . . . . . . . . . . . . . 60
Switch output (Submenu) . . . . . . . . . . . . . . . . . . . . 150
Switch output function (Parameter) . . . . . . . . . . . . . 150
Switch output simulation (Parameter) . . . . . . . . . . . 183
Switch status (Parameter) . . . . . . . . . . . . . . . . 154, 183
Switch-off delay (Parameter) . . . . . . . . . . . . . . . . . . 154
Switch-off value (Parameter) . . . . . . . . . . . . . . . . . . 153
Switch-on delay (Parameter) . . . . . . . . . . . . . . . . . . 153
Switch-on value (Parameter) . . . . . . . . . . . . . . . . . . 152
Symbols
For correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
In the text and numeric editor . . . . . . . . . . . . . . . . 56

Endress+Hauser 189
*71394731*
71394731

www.addresses.endress.com

You might also like