Professional Documents
Culture Documents
OHTA 501 Manual Final Ixvybj
OHTA 501 Manual Final Ixvybj
STUDENT MANUAL
2023
ohtatraining.org
COPYRIGHT
Copyright © 2023 Occupational Hygiene Training Association Ltd (OHTA). All rights reserved.
You may view, download, and print this material for personal, non-commercial purposes, including
self-study and reference.
OHTA is a not-for-profit organisation with minimal income used for maintaining our website and
for administrative support. Please consider supporting OHTA through a donation at
https://www.ohtatraining.org/donate. OHTA is both a UK registered charity and a non-profit
organization (501c3) in the United States.
All materials are authored and edited by subject matter expert volunteers. If you use these
materials without providing support, it may not be possible for us to continue the development
and updating of our training materials. For that reason, we also ask you not to reproduce,
distribute or translate this material without seeking permission from [email protected]
Help power OHTA quality training materials for the years to come with a donation. All amounts
are appreciated.
DISCLAIMER
The information provided in this course is for educational purposes and is not intended as
professional or expert advice. The copyright holder has made efforts to ensure the accuracy of the
material but does not guarantee its completeness or suitability for any specific purpose.
Abbreviations ........................................................................................................................................ 9
References ........................................................................................................................................... 13
Index of Figures.................................................................................................................................... 16
Index of Tables..................................................................................................................................... 18
Chapter 5 Dusts, Fumes and Mists: Health Effects and Sampling Methods ............................................. 74
5.1 Introduction to Dusts, Fumes and Fibres .............................................................................................. 74
5.2 Particulate Deposition ........................................................................................................................... 74
5.3 Particulate Air Sampling ........................................................................................................................ 76
5.3.1 General ........................................................................................................................................... 76
5.3.2 Sample Filters ................................................................................................................................. 77
5.3.3 Basic Sample Collection Procedure ................................................................................................. 80
5.4 Inhalable Dust ........................................................................................................................................ 81
5.4.1 IOM Sampling Head:....................................................................................................................... 81
5.4.2 Conical Inhalable Sampler (CIS) ...................................................................................................... 81
5.4.3 SKC Button Aerosol Sampler ........................................................................................................... 81
5.4.4 Pre-Loaded Cassettes ..................................................................................................................... 81
5.5 Respirable Dust ...................................................................................................................................... 82
The Occupational Hygiene Training Association would like to express their appreciation to the
following individuals and organisations for their support and contributions during the
development and revision of this manual.
ACGIH (2007): Threshold limit values for chemical substances and physical agents and biological exposure
indices. ACGIH, 2007
American Industrial Hygiene Association (AIHA) publication “A strategy for assessing and managing
occupational exposures”
AS/NZ4360: Risk management. Australian and New Zealand Standard 4360, 2004
AS2290: Electrical equipment for coal mines – Maintenance and overhaul. Part 3: Maintenance of gas
detecting and monitoring equipment. Australian Standard 2290.3, 1990
AS2985: Workplace atmospheres – Method for sampling and gravimetric determination of respirable dust.
Australian Standard 2985, 2004
AS2986: Workplace atmospheres – Sampling and analysis of volatile organic compounds by solvent
desorption gas chromatography. Part 1: Pumped sampling method, Part 2: Diffusive sampling
method. Australian Standard 2986, 2003
AS3640: Workplace atmospheres – Method for sampling and gravimetric determination of inhalable dust.
Australian Standard 3640, 2004
AS3853: Health and safety in welding and allied processes – Sampling of airborne particles and gases in the
operator’s breathing zone. Australian Standard 3853.1, 2006
Australian Institute of Occupational Hygienist (AIOH) publication “Simplified monitoring strategies”
BOHS (1993): Sampling strategies for airborne contaminants in the workplace. BOHS Technical Guide
No.11, 1993
BOHS/NVvA (2011) “Testing Compliance with OELs for Airborne Substances”
BSEN689 (1996): Workplace atmospheres – Guidance for the assessment of exposure by inhalation to
chemical agents for comparison with limit values and measurement strategy. British and European
standard 689, 1996
COSHH Regulations (2002): The control of substances hazardous to health regulations 2002 (as amended).
Approved Code of Practice and Guidance L5 (5th Edn), HSE Books, 2005
Dost, A.A. (1996): Monitoring surface and airborne inorganic contamination in the workplace by a field
portable x-ray fluorescence spectrometer. Ann. Occup. Hyg. J. 5, 589-610
European Standard EN 689:2018 “Workplace exposure. Measurement of exposure by inhalation to
chemical agents. Strategy for testing compliance with occupational exposure limit values”
Grantham, D. (2001): Simplified Monitoring Strategies. AIOH, November 2001
1.1 Introduction
This Course has been delivered by OHTA approved training providers for over 10 years and
reviewed by international experts on two occasions to ensure the information is accurate and
current. The latest update was 2023.
The mention of any method, product, manufacturer or vendor as examples in this manual should not in any
way be construed as either endorsement or recommendation of such by either OHTA, authors,
contributors, or editors of this manual.
• Present the results in a form useful for health risk assessment purposes to enable
management to comply with relevant legislation.
In the final outcome, the aim of this manual is to transmit the principles of hazardous substances
measurement to attendees and provide guidance as to how those principles should be applied.
This section introduces the concept of risk assessment as it applies to occupational hygiene. Key
points of this chapter are as follows
✓ Hazard vs risk vs exposure
Risk assessment is a central part of occupational hygiene, the art and science of anticipation,
recognition, evaluation and control (AREC) of work place hazards. Risk assessment allows
occupational hygienists to evaluate whether controls are needed, or if they already exist whether
they need improvement.
The objective of this chapter is to discuss risk assessment as it applies to occupational hygiene
practice. Risk assessment encompasses all areas of human activities since all entail some degree of
risk. Much has been written about risk assessment, and the language used can be inconsistent in
definition and or application. The international standards organisation (ISO) approach (ISO 31000
Risk Management Practices: 2018) provides a useful frame work by which to examine and evaluate
occupational hazards as discussed below, and provides insight into what is involved in the risk
assessment process.
One way to express the relation between risk, hazard and exposure is as follows:
Where risk is can be seen as an uncertainty of adverse outcome (worker illness for
example). Ideally, we might want zero risk, but in reality, this is impractical, so the
issue becomes what is acceptable risk, or how safe is safe? The issue of what is an
acceptable risk has been much discussed and is a variable quantity depending on
The AIHA (American Industrial Hygiene Association) offers the following definition for risk
assessment
From a risk assessment perspective, it can be deduced from the above equation that making
either term in the above equation zero, then the risk would also be zero. So, in occupational
hygiene practice, one might eliminate the hazard by using a substance that has a lesser hazard. For
example, solvent based paints can be replaced with water-based paints.
However, hazard elimination may not always be feasible so occupational hygiene practice would
be to turn to find ways of reducing exposures. For example, use enclosed processes rather than
open vessels, or use adequate ventilation, or even require respiratory protection. Exposure
measurements can then provide a tool to evaluate how effective controls might be in reducing
them or containing them within acceptable limits.
ISO 31000 definitions provide a framework by which to consider occupational health risk
assessment.
Risk Identification, which is to find, recognize and describe risks that will impact a
particular organisation. This is where anticipation and recognition of potential hazards
come in.
Risk Analysis the goal of which is to better understand the nature and characteristics
of identified risks, and where necessary, estimate the level of risk. This is where hazard
evaluation and control start.
Risk Evaluation which supports decisions made in the risk analysis process. How well
does the risk analysis compare with established criteria? This is where we refine hazard
evaluation, and adjust controls as needed
The concept of risk assessment is central to the occupational health and safety legislative
framework in many countries.
An important consideration in identifying chemical hazards are potential routes of exposures. Can
the substance of interest be breathed in? Is skin contact an issue? Ingestion is often another
possibility, although this is usually managed through good personal hygiene practices.
Physical properties are also important. Is a substance volatile (high vapours pressure)? Is in gas or
solid form. If solid, will it generate respirable dust?
Two good sources for identifying risks related to chemical substances at work are Safety Data
Sheets (SDS) as well as labels and signs on products and or containers and vessels.
The UN’s Globally Harmonized System of Classification and Labelling of Chemicals (GHS) is a
worldwide accepted system for classifying and communicating the hazards of industrial and
consumer chemicals. The GHS provides for universal pictograms to provide an indication of a
hazard that a substance might pose. These pictograms are supported by standard hazard
statement language within a particular SDS. In some instances, more than one pictogram may be
relevant Figure 2-1 shows examples of GHS pictograms.
Acutely Toxic Burns Skin, Damages Eyes, Corrosive to Gases Under Pressure
(severe) Metals
Carcinogen, Respiratory Sensitiser, Toxic to aquatic environment Acutely toxic (harmful), Irritant to skin, eyes
Reproductive Toxicity, Target Organ or respiratory tract, Skin sensitiser,
Toxicity, Mutagenicity Aspiration Toxicity Hazardous to the Ozone layer.
Other ways to identify potential chemical risks would be through employee interviews, company
records (e.g., illness and injury statistics), government and industry standards and associations, as
well as the scientific literature. The important thing is to be systematic to reduce the risk of
missing something. Table 2-1 Sources of Information about Potential Chemical Risks lists typical
sources of information.
Tasks Processes
Interviews of workers, managers and
Work practices Exposure controls
engineers
Health issues Maintenance
Records:
Historic conditions Performance of engineering
Process standards controls
Standard operating procedures Chemical inventories
Production Past occupational hygiene
Usage amounts
Personnel monitoring reports
Medical Tasks
Past biological monitoring
Engineering Work histories results
Process flow diagrams
Epidemiological studies
Literature Emerging issues
Toxicological studies
• Who might be exposed directly and indirectly. Are there vulnerable workers (e.g.,
women of reproductive age)?
• Has this process been studied before? Is there previous data that can be used?
• Is there personal protective equipment (PPE) in use? What was the basis for selection?
• Are there standard operating procedures? Do they address the hazards of concern?
In some cases, it may not be possible to obtain complete information in which case one has to rely
on good occupational hygiene judgement.
We have seen that occupational health related risks will be determined by:
• How the worker is exposed to the substance (inhalation, skin contact, i.e., route of entry
to the body)
• How severe are the adverse health effects under the conditions of exposure (health
hazards)
• The duration and frequency of exposure (a single short exposure or continuous long-
term exposure)
Although occupational hygiene risk assessments rely on estimates of exposures, the usual
outcome is qualitative or semi qualitative. In practice, qualitative assessment is usually used first
to screen risks and to highlight activities with higher risk. Subsequently, it may be necessary to
Generally speaking the outcome of an occupational risk assessment can be summarized as follows:
Unacceptable Significant risk e.g., exposure exceeds OELs. Exposure risk is not adequately
controlled. Further risk reduction is required.
Acceptable Exposures consistently below OEL. May need to consider routine monitoring for
change. Adequate controls such that its adverse health effects on worked are not
expected.
Risk assessment can also be quantitative rather than descriptive. It may include factors such as
exposure monitoring data, modeling data, extrapolation from relevant studies to calculate
estimate risk values. Quantitative risk assessment is often used to determine OELs, or sometimes
to set exposure limits associated with a particular level of risk. Quantitative risk assessment
methods are beyond the scope of this course.
Keep in mind that there may be specific requirements in different geographic areas, so best to find
out ahead of time to avoid wasted effort.
The control banding approach focuses resources on exposure controls and describes how strictly a
risk needs to be managed. This qualitative risk assessment and management tool is intended to
It should also be recognized that all such systems provide general guidance based on the most
likely scenario and do not take account of individual process variations. While such systems are a
useful tool for small businesses, assessment of a workplace by an experienced occupational
hygienist may be (and in many cases is) required.
It is important to realize that non-sampling approaches such as COSHH Essentials and the ILO
Chemical Control Toolkit have some limitations, and may not be appropriate for situations, such as
some “hot” processes, open spray applications, gases, etc. However, COSHH Essentials scheme is
being progressively extended by the addition of industry and task-specific guidance for many
situations (see http://www.hse.gov.uk/pubns/guidance/index.htm. Control advice sheets are now
available for welding, metalworking fluids, silica exposures and low-level asbestos work. Particular
industries such as printing have also developed customized sheets for their own specialized
processes.
For some activities, processes, tasks or jobs, experts can specify that respiratory protective
equipment (RPE) (in combination with other control approaches) is always necessary. The most
developed model for control banding has been established by the Health & Safety Executive (HSE)
of the United Kingdom (UK).
The principle of control banding was first applied to dangerous chemicals, chemical mixtures, and
fumes. The control banding process emphasizes the controls needed to prevent hazardous
substances from causing harm to people at work. The greater the potential for harm, the greater
the degree of control needed to manage the situation and make the risk “acceptable”.
The basis of these bands for exposures to chemicals by inhalation is detailed in Table 2-3 Control
Bands for Exposures to Chemicals by Inhalation.
This approach has been developed into web-based applications specifically to assist small and
medium-sized enterprises to do risk assessments for chemicals and mixtures of chemicals.
• The type of task (e.g., mixing liquids, sack filling, manually cleaning and disinfecting
surfaces)
• The hazard classification from the material safety data sheet (MSDS) obtained from the
chemical manufacturer or supplier
• The amount used in the task (small quantities = grams or millilitres (ml); medium
quantities = kilograms or litres (L); large quantities = tons or cubic metres (m³))
• Produces advice on controlling risk from the chemical used in the specified task, and
In British law, the duty to control risk remains with the employer. Both the web and paper versions
of the COSHH Essentials tools are designed to assist the small or medium-sized employer meet
regulatory requirements. COSHH Essentials is a free service and was developed by the HSE in
collaboration with British industry and trade unions.
The ILO Toolkit has five (5) stages which need to be followed. These are:
Stage 1: Find the hazard classification and match it to a hazard group. For common
solvents this has already been done and the information provided on the ILO website.
For other substances there is a need to establish the risk phrases for the substance and
then find the hazard group from the ILO website.
Stage 2: Establish the amount of substance to be used and use this to determine the
scale of use from the table supplied by the ILO.
Stage 3: Establish how much of the substance will escape to the atmosphere. This is
done via looking at the physical state of solids (e.g., pellets – low, crystalline – medium,
fine powders – high) or via comparison of the boiling point of liquids to a table
provided by the ILO.
Stage 4: Find the control approach by using a selection guide that has been prepared
by the ILO.
Stage 5: Find the task-specific control guidance sheet(s) from a table which links the
task description and the control approach.
Once the appropriate control approach has been determined it needs to be implemented and
maintained.
Control banding approaches have been developed in Belgium (REGETOX project), The Netherlands
(Stoffenmanager), and Norway (KjemiRisk). The World Health Organisation is working with its
Collaborating Centres to pilot control banding programmes in more than a dozen countries.
It is important to realize that non sampling approaches such as COSHH Essentials and the ILO
Chemical Control Toolkit are not appropriate for many situations. Such situations could include
some “hot” processes, open spray applications, gases, etc. However, the COSHH Essentials scheme
is being progressively extended by the addition of industry and task-specific guidance on many
situations; see http://www.hse.gov.uk/pubns/guidance/index.htm. Sheets are now available for
welding, metalworking fluids, silica exposures and low-level asbestos work. Particular industries
such as printing have developed customized sheets for their own specialized processes.
It should also be recognized that all such systems provide general guidance based on the most
likely scenario and do not take account of individual process variations. While such systems are a
useful tool for small businesses, assessment of a workplace by an experienced occupational
hygienist may be (and in many cases is) required.
COSHH Essentials and the ILO’s Organisation’s Chemical Control Toolkit provide a simple method
to identify appropriate controls using basic toxicological information from labels and information
on volatility or dustiness and usage rates. The output from these models can be compared with
the control measures in use to assist in the evaluation of the suitability of the controls. More
sophisticated models, such as Stoffenmanager and the Advanced Reach Tool (ART) can provide an
estimate of exposure.
2.6 Documentation
Chemical exposure risk assessments should be documented. The information documented should
be proportional to anticipated risks. Besides regulatory requirements, reasons for documenting
the risk assessments include:
• To provide decision makers with a risk management plan for approval and subsequent
implementation;
Many statutory authorities require assessment records are kept for a number of years including
risk assessments, monitoring records, health surveillance records, maintenance examination, test
or training records (e.g., 5 to 50 years).
It is also important is the need to communicate the findings of risk assessments as well as
whatever requirements a risk management plan may have. This information needs to be shared
with both the management and the employees if there is benefit to be derived from risk
assessment.
Also critical is to implement a continuous cycle of improvement (e.g., plan, do, check, adjust), to
keep the risk management plan up to date and relevant.
• Can less hazardous substances be used? Can they be used in a less hazardous form (solid
rather than dust, or aqueous rather than solvent based)?
While engineering controls may appear to solve problems, they may not always be feasible due to
budgets and or other considerations. Some activities may be too infrequent or temporary to justify
a ventilation system. Moreover, controls such as ventilation require maintenance to ensure they
are operating as designed, and it’s important they are properly used. Nevertheless, engineering
controls are seen as superior because they are intended to engineer out exposures, and hence
• Reducing number of workers exposed (i.e., do it where/when there are fewer workers)
• Reducing the duration of exposure (e.g., use job rotation to reduce the exposure
duration for any individual)
• Are there safer ways to do the job (e.g., cover open containers when not in use)?
• Using codes of safe operating practices, and making sure affected people are trained on
them
This approach requires following specified procedures, but it can be challenging to modify human
behavior. Administrative controls in essence are about managing people to manage the risk.
Note that in practice a combination of these measures may be required to control exposures,
including short term and longer-term actions. For example, respiratory protection may be used
until a ventilation system can be brought online.
All control measures should be reviewed at regular intervals to ensure that adequate control is
maintained. Routine checks, regular maintenance and appropriate supervising procedures are also
necessary.
• Control measures have deteriorated or failed such that serious health effect could result:
e.g., carcinogens and allergens
• Required by regulation
• There are validated indicators of early signs of disease or adverse effect; and, or
• Exposures are liable to exceed limits prescribed in substance specific regulations that
require health surveillance.
Health surveillance should be conducted by or under the supervision of a health professional. Note
also that there may be requirements for preservation of patient confidentiality as well as for
retention of health surveillance data. US OSHA regulations for example may require employers to
retain medical and exposure records for 30 years following employee termination in some
instances, and European regulations require such records to be kept for 40 years from the last
entry.
This chapter describes and discusses the concept of Occupational Exposure Limits (OELs):
✓ What are they?
OELs are commonly used as a reference value to compare the findings of exposure monitoring to
assess potential risks and or to prioritize control actions. The idea of exposure assessment is to
obtain a reliable estimate of worker exposure and to compare it to well-defined criteria, as part of
the risk assessment process.
• Short Term Exposure Limit (STEL), usually averaged over 15 minutes (Mins)
OELs do not necessarily correspond between countries or sources for which reason it is important
to consider available OELs when evaluating air sampling data even though technically or legally
they may not be applicable to the work place under evaluation. Section 3.11 contains further
details on various OELs.
• Mass per unit volume (e.g., milligrams per cubic meter or mg/m³), typically used for
dusts, fumes and mists.
• Proportional by volume (e.g., parts per million or ppm), typically used for gases and
vapours. One percent is equivalent to 10,000 ppm, and one ppm is equivalent to
0.0001%.
It is possible to convert concentrations from mg/m³ to ppm, but for gases and vapours only, and
not dusts, fumes, or mists because the latter involves solids suspended in air rather than
gases/vapours mixed with air. The relevant conversion equation is
24.45 is the molar volume of gas at NTP (25 degrees C and 760 mm Hg)
Example:
The 8-hr OEL for Toluene is 50 ppm. What is the OEL in mg/m3?
24.45
𝐶 𝑝𝑝𝑚 = 𝐶 𝑚𝑔/𝑚³ ∗ 𝑀𝑊
, which re-arranged is same as
𝑀𝑊
𝐶 𝑚𝑔/𝑚3 = 𝐶 𝑝𝑝𝑚 ∗
24.45
Substituting in desired values… (MW for toluene is 92.1)
92.1
𝐶 𝑚𝑔/𝑚3 = 50 ∗
24.45
So OEL of 50 ppm toluene is equivalent to 188 mg/m3
This equation can be re-arranged to calculate OELs (as well as exposure data) in ppm from mg/m³
data.
Time is also a key component of OELs, and is used to average exposures over a specified period of
time to obtain a time weighted average (TWA). Figure 3-1 Time Weighted Average (TWA)
Summary provides a summary of the relationship between time and OEL.
In general, 8h-TWA OELs apply to substances that are chronically toxic (e.g., asbestos, heavy
metals), whereas 15-minute STELs commonly apply to substances that have acute (i.e., short term)
adverse effects (e.g., solvents, acid mists). Some substances have acute and chronic adverse
effects and so may have 15 mins STEL and 8h-TWA OELs. Irritant substances may also have a
ceiling OEL, which is an airborne concentration that should never be exceeded.
When using TWA based OELs, it is important to adjust your results accordingly. In other words, it is
necessary to calculate 8h-TWAs to compare the results against an 8h-TWA OEL.
𝐶1 ∗ 𝑇1 + 𝐶2 ∗ 𝑇2 + . . . . . . . 𝐶𝑛 ∗ 𝑇𝑛 …Equation 3-2
8ℎ 𝑇𝑊𝐴 =
8
Care is advised in making sure that time values are in the same units. If time T is
measured in minutes rather than hours, the denominator becomes 8*60 i.e., 480
minutes.
Example
You sampled an all-day painting project for xylene. The project involved three
samples: paint preparation for 30 minutes (result of 150 ppm), paint application
for 3 hours (i.e.) 180 minutes (result of 40 ppm), and then cleanup for 90 minutes
(result of 70 ppm). What is the 8h-TWA result? Assume zero exposure outside
the periods sampled.
Example
A worker is exposed to toluene for a 12-hour shift. The 8-hr OEL for toluene is
50 ppm. Calculate the daily adjusted OEL using the OSHA Model
8
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = ( ) ∗ 50 𝑝𝑝𝑚
12
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = 0.667 ∗ 50 𝑝𝑝𝑚
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = 33 𝑝𝑝𝑚
Note that OSHA has prescribed a specific method to derive long shift OELs for lead (see
29CFR1910.1025, and 29CFR1926.62), whereby the adjusted OEL for shifts longer than 8 hours is
derived as follows:
The OEL can also be adjusted for week long exposures as follows:
A worker is exposed to toluene for a 12-hour shift. The 8-hr OEL for Toluene is 50 ppm.
Calculate the daily adjusted OEL using the Brief and Scala Model
8 24 − 12
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = ( ) ∗ ( ) ∗ 50 𝑝𝑝𝑚
12 16
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = 0.5 ∗ 50 𝑝𝑝𝑚
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 𝑂𝐸𝐿 = 25 𝑝𝑝𝑚
NOTE: The adjusted exposure limit should be calculated daily and weekly reduction factors. The
most conservative value (i.e., lowest should be used).
As shown in the two worked examples, use of the Brief and Scala Model yields a more
conservative OEL value (25 ppm) whereas use of the OSHA model resulted in a higher value for the
adjusted OEL (33 ppm).
3.4.3 UK Approach
The UK’s HSE’s approach is to convert air sampling results collected over long shifts into 8h-TWAs
(for direct comparison to 8h-TWA WELs) using a proportional approach as follows:
An operator works a 12-hour shift each day for 5 days, and then has seven days’ rest.
WELs are based on an 8-hour reference period each 24 hours in which an exposure
occurs; the seven days’ rest makes no difference. While at work, the operator is
exposed to 4 mg/m3. The WEL for the substance of concern is 5 mg/m3 (8h-TWA). Was
the WEL exceeded?
𝑠ℎ𝑖𝑓𝑡 𝑑𝑢𝑟𝑎𝑡𝑖𝑜𝑛 𝑖𝑛 ℎ𝑜𝑢𝑟𝑠
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 8ℎ − 𝑇𝑊𝐴 𝐸𝑥𝑝𝑜𝑠𝑢𝑟𝑒 = ( ) ∗ 4 𝑚𝑔/𝑚3
8 ℎ𝑜𝑢𝑟𝑠
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 8ℎ − 𝑇𝑊𝐴 𝐸𝑥𝑝𝑜𝑠𝑢𝑟𝑒 = 1.5 ∗ 4 𝑚𝑔/𝑚3
𝐴𝑑𝑗𝑢𝑠𝑡𝑒𝑑 8ℎ − 𝑇𝑊𝐴 𝐸𝑥𝑝𝑜𝑠𝑢𝑟𝑒 = 6 𝑚𝑔/𝑚3
So, it can be concluded that the operator’s exposure would have exceeded the WEL.
Corrective action would be required.
STEL OELs are generally averaged over 15-minute intervals. Worker exposures should not exceed
the 15 min STEL for longer than 15 minutes, or for more than 4 such periods per work day.
Moreover, a minimum of 60 minutes should be allowed between successive exposures at the
STEL.
If real time data (i.e., direct measurement) is not feasible, it’s important to review the sampling
method to ensure that the sample volume allows the lab to report a limit of detection that is less
than the ceiling limit.
In the UK, an excursion limit of 3 times the workplace exposure limit (WEL) OEL is considered good
practice.
Relevant STELs or Ceiling limits take precedence over the peak exposure limits (formerly known as
excursion limits).
3.9 Notations
A notation is a designation in the OEL listing documentation that provides additional information
regarding a particular substance. These notations may indicate biological monitoring limits,
carcinogenicity, sensitisation and whether absorption through the skin should be considered. Each
of these types of notations are discussed below:
Correct application of biological monitoring limits requires consideration of all routes of exposure,
including non-occupational sources (e.g., diet). Employee resistance may be encountered with
biological monitoring because of the use of invasive collection techniques e.g., urine and blood
sampling.
Biological monitoring can be valuable in evaluating accidental exposures associated with incidents,
such as chemical spills.
3.9.2 Carcinogenicity
A carcinogen is an agent capable of causing cancer. Evidence of carcinogenicity comes from
epidemiology, toxicology, and mechanistic studies. Note that not all carcinogens may have
designated OELs.
The International Agency for Research on Cancer (IARC) has the following carcinogen classification
scheme based on the strength of published evidence for carcinogenicity.
The results of IARC’s evaluation on the carcinogenicity of more than 900 substances have been
published in a series of monographs published since 1972. (See https://shrtm.nu/Hrs2.
Different jurisdictions use different schemes, so care needs to be taken regarding local notations.
ACGIH, OSHA and National Institute of Occupational Safety & Health (NIOSH) in the US have their
own classifications. In the UK, carcinogenicity is indicated by ‘Carc’ meaning the chemical can
cause cancer and or heritable genetic damage.
3.9.3 Sensitisation
The notation SEN or Sen refers to the potential for the chemical to produce sensitisation in
exposed workers. Sensitisation may relate to respiratory, dermal or conjunctival exposures. A
sensitised, subsequent exposure(s), even at very low levels, usually results in adverse allergic
reactions.
In the UK, ‘Sen’ is used as the notation for sensitisers whereas the ACGIH TLV© uses ‘SEN’. The
AIHA uses DSEN for dermal sensitisers and RSEN for respiratory sensitisers.
• Occupational exposure limits are not meant to be protective of those who are sensitised.
3.9.4 Skin
The ‘Skin’ or ‘Sk’ notation indicates that the cutaneous route of exposure can contribute
significantly to the overall exposure. Cutaneous exposures include mucous membranes and the
eyes, either by contact with vapours or, of probable greater significance, by direct skin contact
with the substance. Typically, skin exposure results from handling substances without the use of
gloves or protective clothing, splashes or handling contaminated clothing and equipment.
Skin notations are not assigned on the basis of any harmful effects on the skin such as irritation or
allergic contact dermatitis but that it absorbs through the intact skin.
The use of a ‘Skin’/’Sk’ notation indicates that biological monitoring may be required to
supplement air sampling to quantify worker exposure. Skin notations also inform the selection and
use of control measures, for example the use of personal protective equipment to prevent skin
absorption.
Two commonly used toxicological concepts, used in the derivation of OELs are:
Appendix A provides a review of these concepts, which are also discussed in W 507 Health Effects
of Hazardous Substances Course.
OELs for chemicals are established based on a number of factors including toxicity, physiological
response (biologic action). Examples of such factors include:
Irritants Ability to cause inflammation of mucous membrane with which they come in
contact e.g., hydrochloric acid fumes, ammonia, ozone, acrolein.
Narcotic Depressant action upon the central nervous system, particularly the brain e.g.,
ether, chloroform.
Carcinogens Cancer causing substances e.g., asbestos, arsenic, vinyl chloride monomer
Overall, there is very limited human exposure data that can be used to derive OELs. In many cases,
OELs are derived from either animal studies, studies of similar but not identical industries or on
the notion that similar compounds (e.g., chlorinated hydrocarbons, or heavy metals) have similar
toxicological properties. In other cases, there are limited data, or the available data addresses
other routes of exposures. To address these issues, uncertainty factors (also known as safety
factors) are applied. These may range from 1 (as in the case of certain irritants for which there is
human data) to several thousand for carcinogens.
Before comparing results with relevant OELs, it is important to consider the following general
rules:
• Make sure that the units of measurements (i.e., results) are in the same units as the
relevant OEL
• Calculate relevant time weighted averages (TWA) of exposure, where relevant, to allow
direct comparison to TWA OELs. For extended shifts, see Section 3.4.
• If short term data was collected, it should be compared to STELs or Ceiling Values. If
there are no relevant listed value, then the data can be compared with the TWA value
(without calculating any time weighted average value).
This section provides an overview of various OELs commonly used by occupational hygienists in
various jurisdictions but is not intended to be an all-inclusive list. It’s wise to see what OELs are
relevant to any particular location an occupational hygienist may be working in. These are usually
available online. There are also specific publications and applications that are a compendium of
3. TLV-Ceiling (TLV-C)
3.13.1 TLV-TWA
The time weighted average (TWA) Threshold Limit Value (TLV) is defined as:
“The TWA concentration for a conventional 8-hour workday and a 40-hour work week,
to which it is believed that nearly all workers may be repeatedly exposed, day after
day, for a working lifetime without adverse effect.”
However, during this eight-hour averaging period, excursions above the TLV- TWA are permitted
providing these excursions are compensated for by equivalent excursions below the standard
during the working day. Because some substances can give rise to acute health effects even after
brief exposures to high concentrations, it is prudent that excursions above the TWA concentration
should be restricted, moreover, the magnitude of excursions is an indication of the true degree of
effective control over the release of contaminants from a process.
3.13.2 TLV-STEL
The ACGIH has recommended short-term exposure limits (STELs) for many substances that exhibit
acute (i.e., short-term) adverse health effects. STELs are defined as:
“A 15-minute TWA exposure that should not be exceeded at any time during a
workday, even if the TWA is within TLV-TWA. The TLV-STEL is the concentration to
which it is believed that workers can be exposed continuously for a short period
without suffering from:
• Irritation
The TLV-STEL is not a separate, independent exposure guideline, but it supplements the TLV-TWA
where the recognized acute effects from a substance whose toxic effects are primarily of a chronic
nature.
Exposures above the TLV-TWA up to the TWA-STEL should be less than 15 minutes, should occur
fewer than four times a day, with at least 60 minutes between successive exposures.
3.13.3 TLV-C
The ACGIH has also recommended ceiling limits for certain substances. These ceiling limits are
defined as:
“The concentration that should not be exceeded during any part of the working
exposure. If instantaneous measurements are not available, sampling should be
conducted for the minimum period of time sufficient to detect exposures at or above
the ceiling value.”
The ACGIH believes that TLVs® based on physical irritation should be considered no less binding
than those based on physical impairment. There is increasing evidence that physical irritation may
initiate, promote, or accelerate adverse health effects through interaction with other chemical or
biological agents or through other mechanisms.
Ceiling values are instantaneous guidelines that should not be exceeded at during any part of
working exposures. They are best evaluated using direct reading instruments. If none exist, then
sampling duration should be the minimum necessary to detect exposures (often 5-10 minutes)
Peak worker exposure levels may exceed 3 times the value of TLV-TWA for no more than 15
minutes during the workday, on no more than 4 occasions spaced one hour apart during a work
day. Under no circumstances should peak exposures exceed 5 times the TLV-TWA. Additionally,
the relevant 8h-TWA should not be exceeded for an 8-hour work period.
A process is not considered to be under reasonable control if these levels occur (3 times the
workplace exposure limit (WEL) in the UK), where the toxicological data exists to establish a TLV-
STEL or TLV-C these values take precedence over the excursion limits.
The ACGIH approach does not calculate a combined OEL for a mixture of substances. Instead,
measured exposures would be considered as exceeding TLVs when:
Where C1 is the airborne concentration and TLV1 is the corresponding threshold limit
value.
The additive formula applies to simultaneous exposures for hazardous agents with TWA, STEL and
Ceiling values.
A notation is a designation that appears as a component of the adopted TLV® value to provide
additional information with respect to the particular chemical:
Most BEIs® are based on a direct correlation with the TLV® (i.e., the concentration of the
determinant that can be expected when the airborne concentration is at the TLV) with an
assumption that there is no exposure by skin absorption or ingestion. Further information can be
found, in the TLV book (published annually), or in the documentation for the TLVs® and BEI® for
these substances.
3.13.6.2 Carcinogenicity
“A carcinogen is an agent capable of inducing benign or malignant neoplasms. Evidence of
carcinogenicity comes from epidemiology, toxicology, and mechanistic studies”.
• A2 Suspected Human Carcinogen: Human data are accepted as adequate in quality but
are conflicting or insufficient to classify the agent as a confirmed human carcinogen; OR,
the agent is carcinogenic in experimental animals at dose(s), by route(s) of exposure, at
site(s), of histologic type(s) or by mechanism(s) considered relevant to worker exposure.
A2 is used primarily when there is limited evidence of carcinogenicity in humans and
sufficient evidence of carcinogenicity in experimental animals with relevance to humans.
• A4 Not Classifiable as a Human Carcinogen: Agents which cause concern that they could
be carcinogenic for humans but which cannot be assessed conclusively because of a lack
of data. In vitro or animal studies do not produce indications of carcinogenicity which are
sufficient to classify the agent into one of the other categories.
3.13.6.3 Sensitisation
The notation SEN refers to the potential for the chemical to produce sensitisation which may
relate to respiratory, dermal or conjunctival exposures. Once a person has become sensitised,
subsequent exposure to the agent, even at very low levels, usually results in an adverse allergic
reaction.
Example: Toluene diisocyanate (TDI) often found in 2-pack paints is a respiratory sensitiser and
subsequent exposure can result in severe asthmatic reactions to those sensitised.
• Occupational exposure limits are not meant to be protective of those who are sensitised.
• Some bodies (e.g., AIHA) use different notation to indicate specific sensitisation, e.g.,
DSEN for dermal sensitisers, RSEN for respiratory sensitisers.
3.13.6.4 Skin
The Skin notation refers to the potential significant contributions to the overall exposure by
cutaneous route, including mucous membranes and the eyes, either by contact with vapours or, of
probable greater significance, by direct skin contact with the substance. Typically, skin exposure
occurs from splashes, wearing of contaminated clothing, or handling materials without adequate
protective clothing and or gloves (e.g., organophosphate pesticides, glycol ethers).
It is important to note that skin notations are not assigned on the basis of any harmful effects on
the skin such as irritation or allergic contact dermatitis. Substances with a skin notation are not
necessarily harmful to the skin.
The use of a skin notation is to alert the user that air sampling alone is not sufficient to quantify
worker exposure, biological monitoring may also be required in addition to changes to work
practices including the use of personal protective equipment to prevent cutaneous absorption.
The second edition was published in October 1991, and a third edition in May 1995 by the
National Occupational Health and Safety Commission. Australian Exposure Standards for more
than 700 substances are now published on the Safe Work Australia website:
https://tinyurl.com/y878tse3. These standards also have notations in regards to carcinogenicity,
sensitisation.
Australia’s Work Health Safety Act requires that employers maintain exposures not only below
exposure standards but as low as reasonably practical.
HSE’s publication EH40 Workplace Exposure Limits (see https://tinyurl.com/yajjckvs, includes the
list of substances assigned WELs and provides more detailed guidance on their use. WELs are
Note that WELs also address substances for which no short-term limit is specified. HSE
recommends that a figure of three times the long-term limit be used as a guideline for controlling
short-term peaks in exposure. Some workplace activities give rise to frequent short (less than 15
minutes) periods of high exposure which, if averaged over time, do not exceed either an 8-hour
TWA or a 15-minute TWA.
Indicative Occupational Exposure Limit Values (IOELVs) have been established when an
assessment of the available scientific data indicates that a threshold can clearly be
identified below which exposure to the substance should not have an adverse impact on
human health.
Binding Occupational Exposure Limit Values (BIOELVs) reflect socio- economic and
technical feasibility factors, as well as the same factors used to derive IOELVs
European Agency for Safety and Health at Work OEL values can be 8-hour TWA, short term, and/or
biological limit values and can be supplemented by further information such as notations and
routes of absorption.
A DNEL is required for each population anticipated to handle a particular substance (e.g., workers,
end users, humans via the environment, etc.). Certain vulnerable sub-populations (e.g., children
and pregnant women) may also need to be considered. Recall that OELs are generally focused on
healthy adult workers. Thus, any given substance may have several different types of DNELs as
follows:
ECHA has developed guidance documents for the development of DNELs (see
https://shrtm.nu/KH1c). The German Social Accident Insurance Scheme maintains a current list of
DNELs through the GESTIS program. They are available from
https://shrtm.nu/yMW9
OSHA PELs are contained in Title 29 of the US Code of Federal Regulations, which has three tables
of interest as follows:
Note also that more than half of US states have their own state OSHA programs that are required
to be “at least as effective as” Federal regulations. However, it is possible that state-set OELs
regulations may be more restrictive (i.e., lower) than Federal PELs.
3.18 NIOSH
National Institute of Occupational Safety and Health (NIOSH) in the US has established
Recommended Exposure Limits (RELs). They are available on CD-ROM or
http://www.cdc.gov/niosh.
It should be noted that NIOSH is directed to recommend limits that will ensure protection of “all”
workers rather than “nearly all” workers, as with ACGIH TLVs. Consequently, many RELs are lower
than existing OELs (e.g., PELs or TLVs).
3.19 AIHA
From 1980, the American Industrial Hygiene Association (AIHA) produced Workplace
Environmental Exposure Levels (WEELs) which, were updated annually until 2013, when the
development of WEELs was transferred to the Occupational Alliance for Risk Sciences (OARS),
managed by Toxicology for Risk Assessment. Updated WEEL values may be found on
https://www.tera.org/OARS/WEEL.html.
WEELs are intended to provide guidance on exposure levels for substances for which there is (was)
neither legal nor authoritative limits (e.g., benzyl alcohol, butylene oxide). WEELs include
recommendations for 8-hour TWA, Ceiling limit and a Short-Term TWA limit plus Skin, Dermal
sensitiser and Respiratory sensitiser notations.
The AIHA also publishes Emergency Response Planning Guidelines (ERPGTM). These should be
used for risk assessments when considering exposures of either the workforce or for the public for
accidental releases. There are three levels of ERPGs guidelines:
• ERPG–1: The maximum airborne concentration below which nearly all individuals could
be exposed for up to 1 hour without experiencing more than mild, transient adverse
health effects or without perceiving a clearly defined objectionable odor.
• ERPG–2: The maximum airborne concentration below which nearly all individuals could
be exposed for up to 1 hour without experiencing or developing irreversible or other
serious health effects or symptoms that could impair an individual’s ability to take
protective action.
• ERPG–3: The maximum airborne concentration below which nearly all individuals could
be exposed for up to 1 hour without experiencing or developing life-threatening health
effects.
• Safe vs Unsafe: OELs do not divide between safe and unsafe exposures but are guides to
assist in making conclusions about exposure risk. It is considered good practice to reduce
exposure to at least less than 50% of the OEL, or as far below it as is reasonably
practicable.
• Enforceability: Certain OELs in certain countries are legally enforceable, and may carry
civil and criminal penalties. Exceeding these values places a regulatory responsibility on
employers to control these exposures. However, there may be lower recommended
OELs that may represent more up to date information regarding the risks associated with
a particular substance although they may not necessarily be enforceable.
• Healthy Workforce: OELs are generally developed for working populations assumed to
be healthy. They do not consider the general population that may include those who are
very young, old, infirm or disabled. OELs are generally not relevant to community
exposures.
• Adjusting OELs for unusual and extended work schedules, needs to consider effects such
as synergism (where the combination of exposures can result in an effect greater than
the sum of effects of each individual substance) and potentiation (where one agent can
increase the potency of another).
• For some chemicals, there are likely to be several OELs in existence which have been set
by different countries and scientific authorities. When deciding on which OEL to use,
consider the OEL most likely to provide the best worker protection i.e., a health/science
based OEL such as the ACGIH TLVs as well as the applicable country OEL that must be
complied with which may be different.
• OELs apply to personal exposure data rather than to either area or static monitoring
data which may not correspond with worker exposure.
Expert help can be provided by toxicologists, exposure scientists and occupational hygienists with
experience in the setting of OELs as well as the application of OELs in the workplace, if needed.
4.1 Introduction
Occupational hygienists often use air sampling to evaluate worker exposures for various reasons,
including risk assessment and/or evaluating engineering controls. A sampling strategy is essentially
a plan that defines the goals of the air sampling and how those goals will be achieved.
The plan will depend on what its objectives are. In other words, what is (are) the question(s) that
need to be answered.
Thus, when developing any monitoring strategy, it is important to ask the fundamental question:
“How will the data generated from this exercise be used?” Without a reasonable answer to this
question the survey merely becomes the collection of data “for the sake of it”, which can turn out
to be a wasteful and meaningless exercise in retrospect.
The British Occupational Hygiene Society (BOHS 1993) also suggests other factors should be
considered before developing any monitoring programme. These include:
• The requirement for a qualitative risk assessment and appraisal of the workplace prior to
doing any measurements.
• Any requirements for biological monitoring and the integration of these into the overall
survey strategy.
• Any environmental or personal characteristics of the workers which may affect the
measurement.
Once these factors have been assessed it is appropriate to develop a workplace exposure sampling
strategy. In doing so it is appropriate to consider the following:
• What periods during the work day should the employee’s exposure be determined?
• What (if any) compounds are present which may interfere with the sampling (or
analytical) procedure?
• What analytical methods are to be used and what (if any) constraints will these places on
sampling techniques?
When developing a sampling strategy, it is important to understand that the variability of the
workplace environment is such that no universal approach is possible to cover all possible
situations.
The inconsistency of the workplace, in terms of density and intensity of activity, variability of
activity, variability of exposure cloud and the influence of uncontrolled factors such as wind
direction, employee practices, etc. results in the fact that data can only be related to the situation
being studied at the time it was studied.
Any exposure assessment based on a single worker for a single day will have errors of space
(location) and time and we will have little to link this outcome to the real situation.
Individual measurements will not necessarily represent the group, but by accounting for as many
influencing factors as is practicable, we can ensure that some assessments are substantially better
than others.
All the above factors need to be considered when considering a sampling strategy. It is important
to appreciate that monitoring the workplace does not in itself protect anyone, it merely provides
information; however, in some circumstances the mere act of monitoring does raise awareness of
the workforce and management which often results in initiatives to reduce exposure, regardless of
the actual results of the measurements.
The sampling system should be appropriate to the situation being studied and part of an overall
occupational hygiene monitoring strategy.
Guidance on the assessment of exposure can also be obtained from other sources such as BSEN
689 (1996) “Workplace Atmospheres – Guidance for the Assessment of Exposure by Inhalation to
Chemical Agents for Comparison to Limit Values and Measurement Strategy”.
In such situations it is not unusual for a common approach to be adopted with the following
components:
• Initial appraisal
• Basic survey
• Detailed survey
• Routine survey
While the names given to these components may be different in various countries and some
components may be combined (e.g., initial appraisal and basic survey), the concept remains the
same.
• Physical properties. For example, boiling point, vapours pressure, relative evapoursation
rate, dustiness, particle size distribution, ability to sublime, etc.
• Hazardous nature of the substance. This could include any known toxic effects in man
(both acute and chronic); other indications of toxicity (e.g., animal studies, in vitro tests,
structural factors, etc.); any special toxic potential (carcinogenicity, respiratory
sensitisation, reprotoxicity, etc.); and any indication of increased hazard from exposure
to mixtures of substances.
• Any effects on skin (e.g., corrosion, dermatitis) or mucous membranes (e.g., drying,
irritation).
During this initial information collection stage, the use of direct reading instruments or detector
tubes may be helpful in identifying emission sources or employees with potentially significant
exposures. Talking to the employees about the work that they do can also provide useful
information during a walk-through survey.
This information will be very limited and should only be used to support observations. At the
conclusion of the information collection exercise, it may be possible to make a reasonable
assessment of potential risk. It should at least provide sufficient information to decide if a more
detailed study is required or if a non-sampling approach would be effective.
• The initial appraisal suggests that unacceptable exposures may be present in the
workplace.
• An occupational exposure limit has been set where one did not previously exist.
A basic survey will have limited objectives but these should include obtaining sufficient
information to answer the following questions:
In some cases, an initial appraisal may be followed by a detailed survey without the intermediate
step of a basic survey. This depends on what was found during the initial assessment and the skill
and experience of the hygienist performing the evaluation.
In addition, all aspects of the survey need to be reviewed to ensure errors which may affect results
are minimized. In many cases statistical based sampling techniques are adopted and detailed
statistical analysis of the data undertaken.
No matter what the circumstances, the essential questions of: “Who?, When?, Where and How?”
remain central to the development of an effective monitoring strategy.
Irrespective of the above, there are a few simple guidelines which can be used to help in the
decision process regarding the frequency of routine surveys.
• How close are exposures to the relevant exposure standard – as exposures approach the
exposure standard more frequent monitoring will be required (as distinct from being
either well below or excessively above the exposure standard).
• The process cycle – monitoring frequency will need to match the process cycle. This is
especially important in situations where periodic events occur (e.g., maintenance
shutdowns) or irregular process cycles.
Such changes can affect the survey results from year to year and some understanding of these
issues is necessary if data from varying years is to be compared.
In the last 25 years there has been a gradual move to statistically based monitoring programmes
where the workforce is divided into groups of similar exposures called “Homogeneous or Similar
Exposure Groups” (HEGs or SEGs) and a statistically based subset of each group is monitored on a
random basis for an extended period of time. In essence, employees are placed into groups (SEGs)
based on past monitoring data or via using the knowledge of persons working in a plant as to
possible exposures.
Once sufficient data has been collected a statistical analysis of the exposures can be undertaken to
establish the level of compliance to the relevant exposure standard and to provide an indication in
the variability of the data.
While statistically based sampling and evaluation of workplace exposures is very useful in giving a
more accurate picture of employee exposures, it should not be considered as being the absolute
test. There are many assumptions (and thus potential errors) in such programmes but by
controlling as many influencing factors as is practicable a better estimate of exposure will be
guaranteed.
In many basic surveys the practice is to target “worse case” situations, however there is merit in
including some workers who are expected to have lower exposures. This provides a level of quality
control in respect to the initial appraisal and the choice of “worse case” individuals sampled.
4.5 When
The choice of when to monitor is directly related to what process or tasks give rise to significant
exposures. The other major factor that must be considered is the toxicology of the substance
under consideration.
For example, it is important to undertake short term sampling for acutely toxic substances
because they are fast acting, whereas longer sampling would be more appropriate for substances
that are chronically toxic.
The other point to consider when considering when to monitor is the type of exposure standard
appropriate to the substance of concern (e.g., TWA, STEL, Ceiling or Peak). These are generally
related to the toxicological properties of the substance.
As a general rule it is reasonable to state that if the objective of the survey is to evaluate the
exposure of a worker during a specific task, then the monitoring duration should equal the whole,
or a representative part, of the task.
Note that while fixed location (i.e., static) sampling is of use to identify either where workers may
be exposed, or whether controls are effective, or where there may be leaks, or other purposes, its
limitation is that its results represent a specific location rather than actual worker locations (See
Section 4.13). In many cases, sampling for compliance with OELs requires personal air monitoring.
4.7 How?
The selection of sampling equipment and analytical methods will in general result from the
properties of the contaminant under investigation. Other factors that will come into the equation
include:
• Legislative requirements
• Portability of equipment
In all cases it is prudent to use sampling methods from recognized authorities (e.g., National
Standards, NIOSH, OSHA, HSE).
Both the sampling method and the analytical method are subject to error and thus what may be
the most desirable choice from one standpoint may not be from the other.
Ultimately the choice will be a compromise, often dependent on the experience of the
occupational hygienist and the working relationship between the hygienist and the laboratory that
will perform the analysis.
The BOHS (1993) suggests the following considerations when selecting the sampling method.
• Is the sampling device (and collection medium) suitable for collecting the contaminant of
interest and is the medium compatible with the subsequent analytical method?
• Is sufficient known about the dynamics of the collection process so that any variables
can be accounted for in the design of the sampling programme?
• For aerosols, what is the most appropriate device to collect the size range of particles of
interest? Are wall losses (material which sticks to the sampling head and does not lodge
on the filter), either within the sampling head or train, of an order such that account
needs to be taken of them?
• For mists, especially, does possible vapours loss need to be taken into account?
• For gases and vapours sampled from a mixed atmosphere does preferential sorption of
one or more contaminants take place in the collection medium? Does the presence of
high water-vapours levels affect sorption characteristics of the sampling medium or the
presence of particulate material adversely affect the collection characteristics?
• With all contaminants, is the total capacity of the collecting medium sufficient to cope
with the likely loading of the contaminant given the intended sampling rate over the
proposed sampling period?
Other issues (such as the number of samples) need to be addressed but these will be discussed in
Section 4.8.
• Epidemiology – Such exercises invariably involve collecting as much data as possible and
is usually limited by time, budgets and resources.
Some general “rules of thumb” have been proposed (e.g., 1 in 10 workers should be sampled or a
minimum of 3 samples with a spread of less than 25%), however such approaches should be used
with care as they could significantly affect the quality of the data.
Where
𝑡_(𝑣𝑎𝑙𝑢𝑒 ∗) is the t-statistic for the number of degrees of freedom (n-1, with n
being the number of originals samples).
This is another method that requires prior data. It essentially determines the number of samples
needed to test the mean exposure of a lognormal distribution of exposures against an OEL. Table
4-1 summarizes the process.
Secondly, one needs to have an estimate of the Geometric Standard Deviation. You should be able
to estimate this using the preliminary data. An easy way to do this would be to use the AIHAs
IHSTAT, an Excel based approach (See https://shrtm.nu/S9VM. Or, it can be done in Excel.
Before using Table 4-1 Rappaport and Selvin Sample Number Model (α= 0.05, β= 0.10), you need
to understand what α and β mean:
α= 5% chance that it is claimed that the workplace complies with the exposure standard
when in fact it does not.
β= 10% chance that it is not claimed that the workplace complies with the Exposure
Standard when in fact it did.
Step 1 would be to look up the t-value for 4 degrees of freedom in a reference table.
That value corresponds to 2.776.
𝑆𝑡𝑎𝑛𝑑𝑎𝑟𝑑 𝐷𝑒𝑣𝑖𝑎𝑡𝑖𝑜𝑛
𝐶𝑉 =
𝑀𝑒𝑎𝑛
15
𝐶𝑉 = 60 or 0.25
Step 3 would be to insert values into the equation. So
0.25 2
𝑁𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑠𝑎𝑚𝑝𝑙𝑒𝑠 = (2.776 ∗ ) 𝑜𝑟 21.4 (22 𝑟𝑜𝑢𝑛𝑑𝑖𝑛𝑔 𝑢𝑝)
0.15
So, using the preliminary data from the boxed example above (60 ppm average), assuming an OEL
of 120 ppm, if the preliminary data had a Geometric Standard Deviation (GSD) of 2.0, one would
need to collect 21 samples. Note, however, that if the preliminary data represents a greater
percentage of the OEL than the 50 percent value in this example, the number or recommended
samples increases. In other words, as the mean of the exposures approaches the exposure
standard, more samples are necessary to make an accurate judgement as to whether the exposure
standard is exceeded. The same is the case where the data is more scattered (higher GSD), one
would need substantially more samples. However, if the data exceeds the OEL (calculated F > 1),
assuming no changes in the processed monitoring, additional sampling is less useful.
Table 4-1 Rappaport and Selvin Sample Number Model (α= 0.05, β= 0.10)
0.10 2 6 13 21 30
0.25 3 10 19 30 43
0.50 7 21 41 67 96
0.75 25 82 164 266 384
1.25 25 82 164 266 384
1.50 7 21 41 67 96
2.00 2 6 11 17 24
3.00 1 2 3 5 6
Sample size n for top 10% (τ= 0.1) and 95% confidence (α= 0.05)
Required No. of
Measured 11 12 13 14 15 16 17 18 19 20 21 29
Employees (n)
If N <= 11 then n = N
Such an approach should ensure that at least one result should be within the top 10% of
exposures with 95% confidence. However, following NIOSH’s recommendations may result in
collecting more samples than needed to obtain a reasonable estimate of exposures.
Source: AIHA 1998 – Used with permission of the American Industrial Hygiene Association (2007)
4.8.4 AIHA
The AIHA (1998 and 2006)
indicates that there is a point of
diminishing returns in respect to
the number of samples required
to adequately define an
exposure profile. Fewer than six
(6) measurements leave a great
deal of uncertainty about the
exposure profile, while more
than ten (10) provides additional
refinement in exposure
estimates but the marginal
improvement is rarely cost
effective, as indicated in Figure
4-1.
Sample numbers are based on statistical considerations. During the first phase, 3 samples from
workers selected randomly, are taken from a SEG and if all of these are below 10% of the OEL then
it can be concluded that the limit is unlikely to be exceeded. If one or more of the results is above
the limit, then this indicates that it can be exceeded. In either case no further sampling is required.
Otherwise at least 2 more samples are required from each of the same 3 workers. Including the
original 3 samples this will give a total of at least 9 results which can then be analysed using the
recommended statistical tests. Figure 4-2 provides a summary of the BOHS/NVvA sampling
scheme.
Figure 4-2 BOHS/NVvA Sampling Scheme
During a basic
survey, it may only
be possible to take a
few samples due to
practical
considerations or
because only a few
workers carry out
the work. It may still
be possible to arrive
at valid conclusions
about exposure in
such circumstances
by taking account of
observations and
other types of
evidence providing
care is taken when
interpreting the
results, bearing in
mind the typical
exposure profiles
discussed above.
Source: BOHS/NVvA
However, it is also important to review the analytical method of detection for the proposed
method of analysis. Although this information may be listed in the documentation for the
proposed method, it is important to discuss this with the laboratory BEFORE undertaking
sampling. Knowledge of the limit of detection (LOD) dictates the minimum sampling volume and
therefore the length of sampling time required. In some cases, it may not be possible to measure
15-minute exposures because the sample volume is not large enough to contain detectable
quantities of the contaminant of interest.
Example
You have been asked to collect air samples for a metal that has an OEL of 0.1 mg/m 3
(i.e., 100 µg/ m3). Exposure controls appear adequate so you estimate that exposures
might be about 10% of the OEL. The lab where you plan to have the samples analysed
tells you the analytical limit (LOD) of detection is 10 µg per sample. Using a flow rate (FR)
of 2 liters per minute (i.e., 0.002 m3/min), how long should you collect the sample for to
get a detectable amount of metal on the sample?
i.e.:
10
𝑡= 𝑚𝑖𝑛𝑢𝑡𝑒𝑠
10 ∗ 0.002
Which works out to be 500 minutes. So, you’d have to sample for at least a full shift (8 hours).
• Grab samples
• Grab sampling
• Short period sampling (less than the task duration and sometimes taken consecutively)
Irrespective of the nomenclature used the fundamental concept is similar. These different
approaches are shown Figure 4-3 Sampling Patterns
graphically in Figure 4-3.
Some international standards indicate that in situations where exposures are likely to be constant
as little as 50% of the full period need be sampled. In all cases professional judgement plays a
significant role in choosing the best approach.
• Full Period Consecutive Periods – these cover the full period of the relevant standard
(e.g., 8 hours for an 8-hour TWA exposure standard or 15 minutes for a STEL). This
approach is very useful in those situations where the process is intermittent, thus giving
• Full Period Single Samples – are normally carried out to establish the average exposure
of workers during their normal work day. Such samples enable the results to be
compared directly to an OEL based on an 8-hours TWA.
In some instances, it may be appropriate to sample for both the full shift and over short periods as
a substance may have both TWA and STEL exposure standards (e.g., trichloroethylene).
The first of these is cost effectiveness. Large statistically-based monitoring programmes are very
difficult to undertake in terms of the equipment required, the resources necessary to undertake
the exercise and the ongoing disruption to the process. Consequently, it is rare for such
programmes to be implemented outside of multi-national corporations and thus the question
arises “what can reasonably be done?”
For example, a single person operating without any assistance will find it difficult to calibrate,
distribute, monitor and recalibrate more than five sample collection devices at one time. Given
this, it is important that the quality of the monitoring be excellent, the persons and situations
determined for monitoring be appropriate and the collection of data be such that any
abnormalities in results can be explained.
Obviously, professional judgement and experience are major factors in this situation but provided
the basics are clearly understood and correctly applied, a good assessment of worker exposure
can be made.
The relationship between observations (work practices, control measures, dustiness of process,
etc.) and measurements cannot be over-stated; it is better to have fewer samples that can be
clearly interpreted than a large number of samples with limited data which can’t. The balance
between what is reasonably possible to achieve and what is necessary to obtain a picture of
exposure needs to be assessed for each and every exercise. If one person cannot achieve what is
necessary to obtain an exposure profile, then extra resources will be required.
The final limitation on sampling programmes, in many cases, is the process itself. In some
situations, the processes (e.g., batch process which occurs infrequently), do not lend themselves
well to statistically-based random sampling monitoring exercises. An evaluation of each process is
required before considering what can be reasonably achieved.
“A hemisphere of 300 mm radius extending in front of the face and measured from the midpoint
of a line joining the ears.”
Samples collected in the breathing zone of a worker are termed “personal samples” and are
directly linked to workplace exposure standards.
Research in wind tunnels has demonstrated that the location of the sampling head can result in
significant concentration differences over short distances. To avoid such variations, it is common
practice to attach sampling heads in the area of the worker’s lapel but still within the breathing
zone.
The other variable in the sampling head location equation is worker practices, which may have a
significant influence on exposure. One such case occurs when a worker inserts his or her head into
a reaction vessel to monitor the process.
Such actions may give rise to incredibly high exposures of short duration. The sampling device
needs to be positioned in such a manner within the breathing zone to collect the contaminant of
concern.
One approach to overcome (or at least minimize) some of the difficulties if factors are significantly
influencing the exposure cloud, is the use of dual lapel sampling – that is, collecting duplicate
samples on the worker with one sample collected on the left lapel and the other sample collected
on the right lapel. This at least gives some estimate over the variation in the exposure profile over
relatively short distances.
Such factors must be considered when designing a sampling strategy so as to ensure their
influence on the exposure levels is taken into account.
• As a surrogate for personal exposures, when a clear correlation between the results
from static samples and personal samples has been established.
• Are sometimes the only realistic means of measurement when certain types of
continuous monitoring are required.
• As the only realistic method of sampling high volumes of air (e.g., asbestos clearance
monitoring, or where there is a very low exposure limit and the sampling method LOD is
not low enough).
It should be understood that workplace exposure standards are linked to personal sampling and
the use of static or area samples for health assessment is not generally accepted.
This section describes and discusses airborne dusts, fumes and fibres and methods available to
evaluate airborne concentrations. This section addresses:
✓ What are dusts, fumes and fibres?
This section provides and overview of some generally accepted practices and procedures used to
evaluate airborne exposures to dusts, fumes and fibers, for which there are many accepted
methods, including those developed by HSE (UK), and NIOSH (US).
Fume can be defined as the condensation product of materials vapoursized during hot processes
(e.g., smelting, or welding). Fume particles are typically smaller than 0.05 µm and tend to
agglomerate.
Fibres are generally defined as particles with an aspect ratio (length to width) of ≥ 3:1.
Particulates is a collective noun generally used to refer to aerosols such as dust, fumes, mists, and,
smoke.
The adverse effects of dusts, fumes and fibres depend on particle size and chemical composition.
Particle size is important because it influences where in the body the particles may be deposited.
Chemical composition is important to understand the intrinsic toxic (i.e., hazard) properties of the
material in question.
Inhalable fraction: The mass fraction of total airborne particles inhaled through nose and mouth.
In general terms the inhalable fraction includes all particles <100 µm, though it may include larger
particles, although there is no data to support this.
Thoracic fraction: The mass fraction of inhaled particles that penetrate the respiratory system
beyond the larynx.
In general terms the thoracic fraction includes all particles <50 µm and having a 50% cut (of total
airborne particles) of about 10 µm.
Respirable fraction: The mass fraction of inhaled particles that reach unciliated airways (alveoli)
where gas exchange takes place. In general terms the respirable fraction includes all particles
<16 µm (majority <10 µm) and having a 50% cut at about 4 µm.
The importance of the above deposition curves cannot be overstated as this links the potential
health effect with the sampling device necessary to assess the potential health risk.
• Lead dust: Lead is a systemic poison which has been associated with kidney dysfunction,
increased blood pressure and sperm abnormalities. Historically the major toxic effect of
lead has been on the blood system, resulting in anemia. Thus, inhalable fraction is of
greatest interest.
5.3.1 General
Exposures to airborne particulates can be evaluated through traditional air sampling methods that
involve drawing air at a known flow rate through a pre-weighed membrane filter for a specified
time, and later submitting the filter to a laboratory for gravimetric analysis (i.e., post sample
weight to determine mass of particulate collected over a specified time). These filters may be
mounted in different housings (or sampling heads) as detailed in Sections 5.4 to 5.9.
In earlier times, it was common practice for industrial hygienists to weigh sample filters before and
after. Nowadays, sample filters are weighed before and after in a laboratory.
Airborne particulates are typically sampled using a sampling train that consists of a calibrated air
sampling pump connected to the sample filter via a short length of tubing as shown in Figure 5-2
which shows a sampling train for respirable dust.
Sampling trains can also be assembled for other types of particulates, as described in Sections 5.4
to 5.9.
Once assembled, and calibrated (see Section 6.3), the sampling train is installed on a worker of
interest. The pump is usually attached to the worker's belt. It's a good idea to have spare web
belts or a suitable harness to mount the air sampling pump if the worker is not wearing a belt.
Some practitioners have also used vests to hold the equipment and reduce worker inconvenience.
Exposures to airborne particulates can also be achieved through direct reading real-time
instruments as discussed in Section 5.3.1
The choice of collection media will normally be dictated by the choice of sampling method, and by
analytical considerations. In general, there are three types of mechanisms which capture particles
during filtration. These are:
• Interception (impingement) – This occurs when the particle is smaller than the pore of
the filter.
• Inertial Impaction – This occurs with a change in direction of airflow and requires high
velocities and dense fibre packing of filters.
• Diffusion – This occurs with very fine particles and occurs at low flow rates and is
assisted by electrostatic forces.
• Low cost
Not all these properties are achievable in one filter so the selection of a particular filter media for
a particular measurement becomes one of compromise.
The filter selection guide below shows which filters can be used for particular contaminants, but
local or statutory requirements may necessitate using an alternative.
Notwithstanding the information provided above, many Occupational Hygienists choose not to use
mixed cellulose ester filters for metal fume – metal dust analysis due to the poor electrostatic
properties which make them difficult to weigh. Alternatives commonly used include glass fibre or
polyvinyl chloride.
One aspect of filter selection that is sometimes confusing concerns pore size. When sampling for
respirable dust (50% cut at 4 µm), it is not uncommon to use a filter (PVC) of nominal pore size 5
µm. This seems illogical but since most membrane filters allow air to follow a tortuous path,
aerosols smaller than 1 µm are commonly captured. However, this does not apply to
polycarbonate filters which allow air to pass straight through because of how they are built.
Two other features of filters are critical and can cause significant errors in gravimetric analysis if
not considered. These are moisture and electrostatic charge.
In the case of some filters (especially membrane filters), moisture pick-up or loss can be
significant. This can be corrected for by the process of “equilibration”. This process requires that
sample filters and a suitable number of blanks be placed in clean containers with the lids slightly
ajar, in the balance room where they are to be weighed. They are then left for a suitable time to
come to equilibrium with the balance room atmosphere (overnight, but this may depend on the
filter type) before weighing. At the end of the sampling exercise the process is repeated and a
correction made for any gain or loss of mass in the blank filters (this should be minimal if the
balance room atmosphere is well controlled).
The other critical issue is electrostatic charge. This can be overcome by the use of a static
eliminator (usually an Americium 241 or Polonium 210 source). A high voltage static eliminator
may be used but it should be checked to ensure that it does not punch holes through the filter.
One final aspect needs to be considered and that is the transportation of dust-laden filters after
collection. Experience has shown that the layer of dust on the filter is fragile and any shocks or
vibration may cause loss of material unless precautions are taken. The safest way is to deliver
samples to the lab is by hand. If not possible, then filters should be carefully packed to avoid
dislodging collected dust from the filter.
• Calibrate pump flow rate with sampling head attached (see Section 5.4.1)
• Place pump on worker of interest. Record relevant information (e.g., Name, location, job
classification, work activities, date, temperature, engineering controls, personal
protective equipment used, anticipated breaks, etc.) The use of a standard form to
collect information consistently is recommended (e.g., AIHA’s).
• Turn pump on, and record time. Allow worker monitored to go about their business.
• Periodically verify that pump is still in operation by inspection of pump mounted flow
meter. Inspect and review processes of interest to verify activities monitored are as
expected. Keep detailed notes. You will find them useful later when reviewing the
results.
• At the end of the sampling period, note time, and record flow rate at end of sampling
period. Post sampling flow rates should be within ± 5% of pre-sample flow rates. If the
difference is greater, the sample should be considered invalid.
• Use the difference in the start and end times to calculate Figure 5-4 IOM Sampling
the period sampled in minutes. Head
There are several different approaches to measure the thoracic fraction of airborne dust. One
device, the “Respicon” (Figure 5-9 Respicon Sampler) is a multistage impactor that traps the
various size fractions on to individual collection filters of 37 mm diameter (Figure 5-10 Schematic
of Respicon Stage Impaction). A sampling pump operating at 3.1 L/mins is required as is a 4 µm
stage 1 cut module.
A second approach to measuring the thoracic fraction is the use of polyurethane foam filters
specifically designed to separate the individual fractions. These foam filters can be inserted into
5.7 Fibres
Samples for airborne asbestos or synthetic mineral fibres (SMF) are usually collected on to an
open-faced membrane filter in a three-stage cassette fitted with an electrically conductive cowl.
This method uses optical microscopy to count fibres in a section of the membrane filter that is
dissolved with acetone in the lab instead of gravimetric analysis used for particulates.
Airborne fibres are collected on to mixed cellulose ester (MCE) membrane filters instead of
commonly used pre-weighed PVC filters. Membrane pore size is typically 0.8µm although 1.2µm is
used in some countries.
Figure 5-12 Metal Cowl and Sampling Head for Fiber Sampling
Source: Gully Howard Technical - Reproduced with
Permission
Figure 5-12 Metal Cowl and Sampling Head for Fiber Sampling shows original metal design for fibre
filter collection.
5.10 Air Sampling Pumps Figure 5-15 Dual Sampling Head Set
Source: SKC
There are three types of air sampling pumps:
diaphragm, rotary vane and piston. Table 5-3
Comparison of Different Type of Air Sampling
Pumps compares these types of pumps.
The most commonly used type of pump in industrial hygiene work is the diaphragm pump. Its
principle of operation is illustrated in Figure 5-15.
Most particulate air sampling pumps have an operating range of 0.5 to 5 L/M, although most
particulate air sampling methods call for 1 to 2.5 L/M.
There are a variety of available air sampling pumps from different manufacturers. The following
are useful features for particulate air sampling pumps:
• Ability to set flowrates over a reasonable flow range: Necessary as capture devices vary
in flow rate requirements.
• Good battery capacity: This allows continuous operation for the full duration of a work
shift.
• Intrinsically safe: This is a mandatory requirement for those pumps that are used in
workplaces where the risk of an explosion may be high (e.g., coal mines, oil refineries).
Figure 5-17 Soap Film Meter
• Battery charge: Some types of pump battery (e.g.,
Nickel-Cadmium) have an unusual characteristic in
that they can develop a "memory effect" if
Source: SKC - Reproduced with Permission
• Internal flowmeters: Flow meters built into sampling pumps are generally not
considered accurate, and so should not be relied upon for calibration purposes (see
Section 5.11). However, they provide a quick visual indicator that the pump is running
(or not).
5.11.1 Basics
The accurate analysis of atmospheric dust concentrations is dependent on the determination of
the mass of dust, fume or fibre on the collection media (either gravimetrically, chemical analysis or
microscopy) and the total volume of air sampled (i.e., total number of m³ of air sampled).
The purpose of pump flow calibration is to set the sample flow rate at the specified rate, and to
verify sample flow rate at the end of the sampling period. The sample volume can be calculated
based on average flow rate (provided it remains within a range of accepted variability, usually ±
5%) and sample time (see Section 5.12).
Figure 5-18 Rotameter Figure 5-19 Electronic Flow Meter
Source: SKC- Reproduced with Permission
Good occupational hygiene practice requires that a path of traceability is established and
maintained. This is via use of either a primary or secondary standard. Primary standards are
directly traceable to a national standard and are not significantly affected by variables such as
temperature and pressure. However, primary standards are often impractical for field use, so it is
common practice to use secondary standards. Examples of primary standards are:
• Soap film meters (also known as bubble burette) - See Figure 5-17 Soap Film Meter
In certain countries, some electronic calibration units (e.g., BIOS Frictionless Piston) are considered
primary standards. However, third party accreditation bodies do not agree.
• Rotameters
• Always calibrate a sampling pump with a sample head identical to what will be used in
the field. Figure 5-20 Calibration with Bubble Flow Meter shows a calibration set up for
vapours sampling.
• Allow the sample pump to stabilize for at least 5 minutes after it has been switched on.
Adjust the flow to the required flow rate.
• Measure the pump flow rate until three consecutive results are within ±2% of the mean.
This accuracy is easily achievable using an electronic calibrator, which is the preferred
technique. It is also achievable with a soap film meter, but it may not be
• possible using a rotameter. Use the mean value of the three consecutive results for
airflow calculation (Section 5.12).
Rotameters Monthly for 3months. If results are Calibrated against a primary flowmeter
within ±3%, interval can be extended over range of use
(one-year small bore and two years large
bore)
Electronic Meters Monthly for three months then if Calibrated against a primary flowmeter
measurements are within ±3% of over range of use
expected results, the interval can be
extended to six months
Now, let's say you sampled for 7.75 hours at that average rate. That is equivalent to 7.75*
60 minutes.
Then, the volume, V litres, would be given by the following:
V = FR (L/min) ∗ Time (min)
V = 1.95 ∗ 7.75 ∗ 60
i. e. V = 906.75 litres or 0.907 m3
It's important to indicate air sample volume units (litres or m3) when submitting samples to
the lab.
𝑀𝑎𝑠𝑠 = 𝑃𝑜𝑠𝑡 𝑆𝑎𝑚𝑝𝑙𝑒 𝐹𝑖𝑙𝑡𝑒𝑟 𝑊𝑒𝑖𝑔ℎ𝑡 − 𝑃𝑟𝑒 𝑆𝑎𝑚𝑝𝑙𝑒 𝐹𝑖𝑙𝑡𝑒𝑟 𝑊𝑒𝑖𝑔ℎ𝑡 − 𝐵𝑙𝑎𝑛𝑘 𝐶𝑜𝑟𝑟𝑒𝑐𝑡𝑖𝑜𝑛 …Equation 5-1
𝑚 = 2.09 𝑚𝑔
So, say that your dust sample contained 2.09 milligrams of dust, and that you sampled 0.907 m³ of
air. That is equivalent to a dust air concentration of:
Care is advised in making sure that time values are in the same units. If time T is measured in
minutes rather than hours, the denominator becomes 8*60 i.e., 480 minutes.
What is the worker's 8h-TWA dust exposure based on the above results?
Note that there is no exposure data between 10:30 to 10:45, 12:45 to 13:30 and 15:30 to
15:45, which would correspond to breaks and lunch. Assume zero exposure for these
periods.
Using the equation above, the 8h-TWA exposure would be given by the following:
8ℎ 𝑇𝑊𝐴 = ((0.32 ∗ 2.5) + (0 ∗ 0.25) + (0.07 ∗ 2) + (0 ∗ 0.75) + (0.2 ∗ 2) + (0.25 ∗ 0) + (0.1 ∗ 1.25))/8
The TSI Dust Trak (Figure 5-22) uses laser photometry to detect light scattered by airborne dust
particles. It can be very useful for evaluating dust control procedures and for pinpointing emission
sources.
The instrument response depends on the size, shape and reflectivity of the airborne particles
rather than on particle mass. Some instruments can give a mass readout, but this is only accurate
if calibrated for the specific dust in question. Please refer to the manufacturer's instructions for
further details.
Note that environments with elevated airborne moisture (e.g., sprays, water mist, etc.), will cause
this instrument to over-respond (i.e., read higher
than true). This issue, common to most optical Figure 5-22 Dust Trak
detection-based instruments, needs to be taken
into consideration.
Source: TSI Inc. Reproduced with Permission
Collection Flowrate
Contaminant Sampling Head Comments
Medium (L/min)
0.8 µm pore size
Asbestos and MCE
Open face with 1–4 Requires conductive cowl (3-piece
synthetic mineral membrane
conductive cowl (8 – 16 in UK) cassette).
fibres filter
See Section 0
5.0 µm pore size
Respirable dust Miniature
PVC 1.7 – 3.0 Depends on type of cyclone used
(including silica) cyclone
See Section 0
5.0 µm pore size
PVC or
IOM (or Filter can be subsequently
Inhalable dust glass fibre 2
equivalent) analysed for metals, etc.)
filter
See Section 5.4
Welding + other IOM (or
PVC 2 0.8 µm pore size
metal fumes equivalent)
5.0 µm pore size
MCE
Rosin solder flux Millipore, Flow rate depends on fume load
membrane 1-2
fume Swinnex in atmosphere.
filter
See Section 5.9
PVC= Polyvinyl Chloride MCE= Mixed Cellulose Ester
This section describes and discusses general methods and procedures to evaluate airborne
concentrations of gases and vapours. This section addresses:
✓ What are gases and vapours?
✓ Sorbent Tubes
✓ Example Calculations
Review of relevant air sampling documents should be part of sample planning to get key
information such as sample flow rates, sample collection media, limits of detection. Documents to
consider would be NIOSH's Manual of Analytical Methods (NMAM) and the HSE's Methods for the
Determination of Hazardous Substances (MDHS). In some jurisdictions, relevant regulations
provide specific sampling details (e.g., OSHA certain substance-specific rules such as for asbestos,
benzene, cadmium, lead.). Agencies such as UK's Health and Safety Lab and commercial labs are
also useful resources as are some occupational hygiene equipment suppliers such as Casella, MSA,
SKC etc.
Vapours are the gaseous phase of a liquid at room temperature. They result from evapoursation of
molecules from the surface of a liquid into the gaseous phase. Examples of occupational hygiene
interest are solvents such as acetone, benzene, gasoline, as well as mercury in its elemental form.
You may hear the term fume in connection to solvents, paints etc. Technically, fume applies to
finely divided particulates associated with hot processes (e.g., welding), so it is incorrect to talk
about solvent fumes when you mean solvent vapours.
Some vapours are associated with certain solids. Here, molecules are going from solid to vapours
phase in a process known as sublimation. One common example is dry ice (solid carbon dioxide).
Other examples are arsenic, iodine, naphthalene (moth balls).
• Active Sampling
• Passive Sampling
• Grab Sampling
The first two approaches involve lab analysis. In the case of grab sampling, a volume of air is
collected into a container (e.g., an evacuated flask, or a Tedlar bag) which is then sent to a
laboratory for analysis for specified substances (see Figure 6-1 Sampling Train for Sorbent Tube. In
the case of active sampling, air is drawn through a sorbent tube that absorbs the gas or vapours of
interest (see Sections 6.3 and 6.4). The sorbent tube is submitted to the lab for analysis after the
sampling period. Active air sampling is the most commonly used approach in occupational hygiene
for solvent vapours.
The third method, passive sampling, is a method whereby vapours of interest are allowed to
diffuse into a sorbent material usually in a badge-type assembly, usually worn on the lapel. After a
specified period of exposure, the badge is sealed, packaged and shipped to a lab for analysis.
Further details on this method may be found in Section 6.9.
The fourth method, which involves the use direct reading instruments, is further discussed in
Section 6.12.
In organic vapours sampling, total volume of air collected is most important rather than low
pulsation flow; hence, some low flow pumps do not have as sophisticated flow control systems as
dust sampling pumps.
As with particulate air sampling, the sampling flow rate needs to be set to the rate specified in the
sampling method. In other words, the air sampling pump needs to be calibrated (with the sorbent
tube in place), just as with dust air sampling before use, and verified after use. If the final (i.e., post
sampling) flowrate differs from the initial flowrate by greater than ±5% (UK and US), the sample
should be discarded and sampling repeated if possible. Australian Standards (AS) allow a greater
variability (±10%) in flow rate before and after sampling. However, this is considered too high by
many occupational hygienists. A variability of ±5%
Figure 6-2 Air Sampling Train for
represents best practice.
Sorbent Tube
Keep in mind that adsorbent tubes need to be
sealed after use, and that there are limits to how
6.4.1 General
The adsorbent material is usually packed in a sealed glass tube as shown in Figure 6-3 Sorbent
Tube. Before sampling, both ends of the glass tube need to be carefully broken off and the tube
connected into the sampling train, usually via a sample tube holder. The printed arrow on the
sampling tube shows the direction of the airflow. Insert the tube so arrow points towards the
pump. If there is no arrow on the tube, insert the tube with the smallest sorbent section (referred
to as the back-up section) into the tube holder so air flows through the main (largest) bed first.
Sample analysis involves separate desorption of each section of sorbent tube (i.e., front and back
sections). This is commonly done using either solvents (e.g., carbon disulphide) or thermal
methods in the lab.
Figure 6-3 Sorbent Tube
Foam separator
To address breakthrough, and avoid underestimating, sorbent tubes contain two layers, front and
back. Each layer is analysed separately. NIOSH air sampling methods define breakthrough as
having occurred when the back layer contains 20% or more of the quantity of the contaminant(s)
collected in the front layer. UK Guidelines (MDHS para 69) define breakthrough as occurring when
the back layer contains 10% of the front layer. Where breakthrough is noted, the process or
activity should be re-evaluated after reducing sample volumes (e.g., reduced sample flow rates, or
multiple samples over shorter periods).
Charcoal Best for non-polar organic Can't use for polar Coconut shell, crushed and
vapours (e.g., substances. conditioned at high
hydrocarbons, ketones, temperature and low O2.
Poor recovery for reactive
esters, ethers).
compounds such as Carbon disulphide (CS2)
Also chlorinated solvents. aldehydes, amines, commonly used to desorb.
phenols, low molecular
Thermal desorption with
weight alcohols.
specialty tubes.
There are also specialty sorbent tubes such Figure 6-5 Multi-Tube Sample Tube Holder
as:
Basically, the problem is that it is not possible to extract 100% of a given contaminant from a
sorbent tube. If not accounted for, it will lead to errors in the calculation of an exposure. To
overcome this, the lab establishes a “desorption efficiency” for each batch of samples. The general
approach is to load (i.e., spike) sample tubes from a batch with varying amounts of the
contaminant of interest and to then analyse them as normal. In other words, the idea is to
compare mass detected via analysis against the mass added. The percentage recovered (e.g., 80%
or 0.8) is deemed the desorption efficiency for that particular batch of tubes, and for that
particular contaminant.
It is important that the laboratory understand the reasons for this process and be familiar with the
appropriate methods to establish such values. Some tube manufacturers publish a list of typical
desorption efficiencies for common contaminants as a guide for the laboratory.
• Humidity – charcoal's great affinity for water vapours reduces its collection of other
contaminants.
• Sampling flow rate - if sampling pump flow rates are too high, contaminant residence
times may be too short to be adsorbed resulting in collection losses.
• Channeling – Improper packing of sorbent can create channels or gaps in the bed
through which the gases can flow more easily and thus not be adsorbed on to the
sorbent.
• Overloading of sorbent tubes can occur if concentrations/sampling times are too long or
by the presence of other contaminants including water vapours.
• Potential low desorption efficiencies, meaning not all the contaminant is extracted.
Typical desorption efficiencies are in 80% range, less under high humidity conditions or
when sampling for substances with marked polar characteristics.
• About 0.1% of the solvent extract is actually analysed, thus resulting in a dilution factor
of 1,000, raising limit of detection.
• Work with CS2 in the lab requires appropriate measures to minimize its hazards.
Thermal desorption (i.e., the use of heat to drive off contaminants adsorbed onto the solid sample
media) addresses these limitations and offers the following advantages:
• Detection limits that are 103 to 104 times better than solvent extraction, making it
possible to detect contaminants in parts per billion (ppb) or parts per trillion (ppt) range.
For these reasons, some vapours and gas sample methods that require solvent extraction have
been superseded by thermal desorption methods in Europe.
However, thermal desorption sample collection requires specialty sampling tubes. The “industry
standard” is ¼ inch (6.4 mm) OD x 3½ inch (88.9 mm) long stainless-steel sorbent tube pre-packed
with the sorbent of choice, most commonly Tenax. In addition, a ¼ inch brass SwageLok type
storage cap (fitted with a PTFE ferrule) for the non-sampling end of the tube, and a diffusion cap at
the end of the tube is normal practice as shown in Figure 6-8.
In contrast with the glass sorbent tubes described in Section 6.4.3, the stainless-steel thermal
desorption tubes can be reused many times. It is therefore essential that the stainless-steel
thermal desorption tubes are always pre-conditioned before they are used for sample collection
to ensure that there is no contaminant on them, by the laboratory putting the tubes through a
pre-set heating sequence. Once sampling or analysis is completed, tubes should be recapped with
the brass storage caps as soon as possible and returned to a clean environment for storage.
An additional point to note is that each thermal desorption tube can only be analysed once. In
contrast, it may be feasible to re-analyse a glass sorbent tube sample if some of the liquid
desorption solution has been retained by the lab.
Specific details including the general handling of thermal desorption tubes, sorbent selection, tube
conditioning, post sampling short- and long-term storage should be obtained from the
manufacturer before use.
6.5 Filters
Some mists and vapours require samples to be collected on to specialty filters. For example,
certain pesticide and isocyanate air sampling methods require the use of glass fiber filters.
For some contaminants it may be necessary to use a filter impregnated with a stabilizing agent or
a backing pad treated with a collection media where the contaminant may be present in the
gaseous form or in both the
Figure 6-6 Thermal Desorption Tubes particulate and gaseous form.
The following examples illustrate approaches issues for mixed phase sampling:
The “traditional method” for sampling and measurement for coke oven emissions was to collect
samples onto membrane filters that were extracted with benzene to report results as the
“Benzene Soluble Fraction of the Total Particulate Matter”. However, polyaromatic hydrocarbons
emitted from coke ovens are present in both particulate and vapours phases and hence sampling
for just the particulate phase underestimates coke oven emission measurements. Modern
sampling trains for coke oven emissions include a sorbent layer behind the particulate membrane
filter to collect the vapours phase that passes through the membrane filter.
Example 2: Impingers
Practical difficulties associated with the use of impingers (see Section 9.5) have led to the
development of impregnated filters for contaminants such as isocyanates, formaldehyde and
glutaraldehyde.
In the case of the spraying of “two pack” isocyanate-based paints, isocyanates may be present
both as mist and as a vapour. In order to ensure that both phases of the isocyanate exposure are
collected a sampling train comprising of an impinger followed by an impregnated filter can be
used – for details see MDHS 25/4 published by the UK Health and Safety Executive
http://www.hse.gov.uk/pubns/mdhs/pdfs/mdhs25-4.pdf.
Example 3: Fluorides
Fluorides are commonly found contaminant in aluminum smelters. They may exist as particulates,
as a hydrofluoric acid (HF) mist or as hydrofluoric acid gas. They need to be sampled as described
in HSE MDHS 35/2. This method entails a Teflon filter mounted on a sodium carbonate
impregnated paper pad mounted in an inhalable sampler. The Teflon filter removes the particulate
fluorides, whilst the sodium carbonate impregnated pad collects the hydrogen fluoride.
Hydrofluoric acid mist is not retained on the Teflon filter and is collected on the sodium carbonate
impregnated pad.
The use liquid sample collection media requires careful handling to avoid spills, or introducing
liquid into the pump, and loss of sample volume through evapoursation. The need to keep sample
collection devices upright to avoid spills and the fragility of equipment limits practical use of this
methodology for personal sampling. There are alternative methods that do not involve liquid
sample collection media for many substances (e.g., isocyanates). These alternate methods
typically involve specially treated or impregnated filters, so the sampling is essentially a variation
of particulate air sampling.
D diffusion coefficient for the adsorbate in air in cm²/s– available from manufacturer
of the sampler for a given chemical
From Equation 6-1 above, if Co is zero (i.e., the collection medium is effective), then mass transfer
or collection rate is proportional to the ambient concentration C.
Sample rates using passive samplers depend on the diffusion coefficient of the contaminant and
the geometry of the monitor. Some monitors such as 3M, SKC monitors and Dräger ORSA monitors
have a diffusion path axial to the sorbent whereas others such as the Radiello badge have a
diffusion path to that is radial to the sorbent surface.
Sample rates remain constant as long as the sorbent media does not reach its capacity (i.e., does
not become saturated) and as long as adequate airflow is maintained across the face of the
monitor. These rates are available from monitor manufacturers.
Organic vapours diffusion monitors typically contain Figure 6-9 3M Diffusion Monitor
activated charcoal so that volatile contaminants
Source: 3M Australia – Reproduced with permission
Diffusion monitors meet or exceed an accuracy of ± 25% at 95% confidence for many workplace
contaminants. Diffusion monitors can also be used for area monitoring provided there is sufficient
airflow, defined as at least 25 ft/min or 0.13 m/sec in any orientation. Care needs to be taken not
to place diffusion area monitors away from corners and or other dead air spaces.
Although passive badges offer a number of advantages (e.g., ease of use and relatively
inexpensive), some of their disadvantages are as follows:
• Charcoal based diffusion monitors have the same moisture and recovery issues
associated with the use of active sampling tubes.
• With some diffusive samplers inaccuracies can occur at wind speeds >2.5 m/s,
depending on their design.
• “Sampling rates” are supplied by the manufacturer and differ for each compound.
• It can be difficult to know if breakthrough (see Section 6.4.2) has occurred, especially for
the more volatile compounds such as methylene chloride, as some diffusion monitors do
not have a back-up section.
Grab sampling is often used for sampling unknowns, for evaluating contaminant sources, where
the air contaminant concentration is known to be constant, or where peak concentrations are of
interest.
Sampling periods are short and generally last from a few seconds to a few minutes. However, in
the case of an evacuated container, it is possible to install specific flow controllers to regulate air
flow into the container over known periods of time.
Note that grab sampling is generally not a suitable method for personal air sampling.
SUMMA canisters have internal surfaces specially treated using a process (also known as summa
process) that combines electro polishing with chemical deactivation to produce nearly chemically
inert interior surfaces to maximize recoveries of air contaminants from the container in the lab.
Canisters are generally furnished from the lab that will analyse the sample(s) where they are
cleaned before use. It's important to review desired limits of detection with the lab to ensure that
canisters that have been suitably cleaned are provided.
Canisters are provided under vacuum, so sample collection involves simply opening the valve to fill
the container. This takes from a few seconds to a few minutes (depending on container size),
unless flow controllers (usually supplied by the lab) are used to sample over a longer specified
time (e.g., 1- 8 hours).
Grab sampling using canisters can yield limits of detections that are in the ppb (or µg/m³) range.
When used to sample for unknowns, the contents are analysed by gas chromatography/mass
spectrometry (GC/MS). While it can yield useful information, SUMMA canister sampling can be
expensive.
Grab sampling bags are relatively inexpensive, can be carried to site in a brief case, filled in
seconds and shipped easily to the laboratory for analysis. See Figure 6-10, although they come in
different sizes (up to 250 L), bags used for occupational hygiene sampling purposes bags are
typically between 5 and 15 litres. Pump flow rates of 1 L/min are typically used to fill these bags.
Keep in mind that overfilling gas bags in low temperatures can create leakage and or damage
when moving bags to warmer temperatures, and that insufficient filling at warmer temperatures
can reduce sample volumes if analysed in lower temperatures.
Start End
Note that post sampling flow rate was within 5% of initial flow rate, and so is acceptable.
Now, say that the lab reports a total 6.3 mg of toluene in the sample represented by the above air
sampling data. The lab report also shows that there were 5.6 mg in the front section and 0.7 mg in
the back section, and that desorption efficiency was 90%.
Since only 90% of toluene was recovered from the charcoal tube, the corrected amount collected
in the tube would be 6.3/(90%) or 7 mg.
This is equivalent to 17.0/24.45 ∗ 92.14 or 1.45 parts per million (ppm) by volume where 24.45
represents the molar volume of gas at NTP, and 92.14 is toluene's molecular weight.
Note that labs will usually report air concentrations in either in mg/m³ or ppm, provided sample
volumes were submitted with samples.
Air temperatures influence diffusion monitor sampling rates, as exemplified in Table 6-3 3M
Diffusion Monitor Temperature Correction Factors. Atmospheric pressure variations do not
require correction factors.
Source: 3M – Reproduced
Celsius (oC) -8 -3 2 7 13 19 25 31 37 44
with permission
Fahrenheit (oF) 18 27 36 45 55 66 77 88 99 111
Correction Factor 1.06 1.05 1.04 1.03 1.02 1.01 1 0.99 0.98 0.97
Note that for temperatures between 13°C and 37 °C (i.e., 55°F and 99 °F), correction factors are
quite small (2%).
The TWA concentration of contaminant in ppm can be calculated from the following expression:
Consider the following data for benzene sampled with a diffusion monitor:
Temperature 75 °F
The corresponding air concentration C (in mg/m3) would be derived using the first equation above,
using A:
Similarly, the concentration of benzene in ppm can be derived from application of Equation 3-1
(Chapter 3).
Note that no temperature correction was required for the sample since it was collected at 75 °F.
Note also that labs will usually provide results in terms of either in mg/m³ or ppm, so it's not
common to have to calculate air sample concentrations from lab data.
The biggest advantage of direct reading instruments is obtaining real time data that can be
compared to relevant exposure criteria. The advent of data logging technology now allows analysis
of instantaneous (seconds), short term 15-minute STEL concentrations and 8-hour TWA
concentrations for certain contaminants.
Although many direct reading devices are intended to measure specific contaminants (e.g., carbon
monoxide or hydrogen sulphide), some can monitor simultaneously multiple gases and vapours
(e.g., confined space entry monitors). Others (e.g., flame ionization detectors [FIDs] and photo
ionization detectors [PIDs]) are not able to differentiate between the different gases and vapours
that may be present and instead report a "composite" single result that may require careful
interpretation.
• To obtain real time data (employee concerns, leaks, emergency response, confined
space entry).
• To obtain information about peaks in exposure levels that traditional air sampling is
unable to measure.
• For stationary installations to record area exposure levels and as well as to sound an
alarm should concentrations of concern be detected.
Table 6-5 lists some commonly used direct reading instruments. Some will be discussed during the
practical session.
• Often costly to purchase though they may be available for rent in some countries.
• Sensors are generally substance specific, have a finite life and may have limited range.
May not always be user replaceable.
• Need for intrinsically safe instruments in situations where it is possible for a potentially
flammable atmosphere to be present.
Consider a carbon monoxide electrochemical sensor, often found as a feature of confined space
entry monitors. These sensors are responsive to gases other than carbon monoxide, and can give
either false positive results or false negative results that manufacturers address by using a sensor
filter. The following data from one particular supplier illustrates results obtained when applying
100 ppm of the listed gas to the CO sensor:
Colorimetric tubes are based on a specified colour change of a specific reactant when it comes
into contact with the contaminant of interest. Most tubes contain a solid reactant through which a
known volume of air is drawn through the tube using a manual pump. It's important to operate
the pump correctly and to track the number of strokes to collect the required number as specified
in the tube instructions. The tubes usually have an arrow that should point towards the pump to
ensure the tube is in the correct orientation. Air concentration of the particular contaminant, if
present, are read off directly from gradations on the colorimetric tube (i.e., length of stain, see
Figure 6-12 New (Left) and Used ). In some cases, the result is based on counting the
(Right) Colorimetric Tubes number of strokes required to match colour change in
Source: Dräger Safety – Reproduced with permission
There are also direct reading long term colorimetric tubes that use low flow battery operated
pumps or diffusion type badges for long term measurements of 1 to 4 hours.
• Immediate results
• They cannot be used to measure personal exposures, but they may be useful to provide
an on-the-spot indication of the order of magnitude of exposures, as a preliminary to a
decision on which exposures would warrant the collection of personal exposures.
It is very important to read the particular manufactures’ instructions for any colorimetric tube
before use to select the correct tube and to use it correctly, noting the effect of any potential
interferent, some of which can give false positive or false negative results.
Group II Equipment that used in severe conditions e.g., mounted on operating equipment,
where it could be exposed to vibration, high levels of dust and humidity.
Group III Equipment installed at a fixed location for appreciable periods of time with a local read
out of concentration.
The mining industry approach also sets out the requirements for a Certificate of Compliance,
recordkeeping, accuracy requirements and the minimum competencies for persons and accredited
authorities engaged in the examination, maintenance and testing of equipment covered.
The mining industry approach also provides advice for techniques and equipment to carry out
span and zero tests on gas detecting equipment. Span test is the test of response to certified test
gas(es). Zero test is test of response to zero gas conditions.
Test equipment for single point span checks consists of a cylinder containing the certified test gas
fitted with either a calibrated flow meter with a precision regulator or a flow restrictor and
pressure gauge.
For equipment in which the external atmosphere reaches the sensor or detector by diffusion, the
test procedure usually involves dispensing the certified test gas to the sensor via tubing and
suitable calibration cup. Calibration cups should conform to equipment manufacturer
requirements.
For sample-draw equipment containing an integral pump or hand-held aspirator, the sample inlet
is connected via tubing to grab sample bag that has been pre-flushed and filled with the certified
test gas.
The modern automation of equipment has led to its increased use in Explosive or Ex areas. Such
equipment is termed “Ex equipment” and is found in areas such as:
• Underground coalmines
• Woodworking areas
• Sugar refineries
• Oxygen
• An ignition source
Group I Equipment that is used in underground mines where it could be subjected to methane
and coal dust
Group II Equipment used in industries other than mining with the following subgroups
IIA - Where least readily ignited gases such as propane and benzene may be present
IIB - Where more readily ignited gases such as ethylene and diethyl ether may be
present
IIC - Where most readily ignited gases such as hydrogen and acetylene may be
present
Group I temperature designation requires that the temperature of the components and surfaces
exposed to dust and methane to be less than 150°C. Where components and surfaces are
protected from the ingress of dust, the maximum temperature of such must be less than 450°C.
For Group II designation, the maximum surface temperature must not exceed values shown Table
6-10 Maximum Surface Temperature / Ignition Temperature which corresponds to the
temperature class of the equipment. For convenience, a temperature class may be assigned to a
gas or vapours based on its ignition temperature.
T1 450
T2 300
T3 200
T4 135
permission
T5 100
T6 85
• “ia” – means that the type of protection ‘intrinsic safety” (no release of spark energy or
thermal energy that can cause ignition) is maintained with up to two faults.
Source: TestSafe –
“ia” Zones 0, 20
Reproduced with
permission
“ib” Zones 1, 21
“ic” Zones 2, 22
Further and much more detailed information for the use of gas detection equipment in potentially
explosive atmospheres including the Classification of Zones, Explosion Groups, Temperature
Classes, the Types of Protection provided by equipment, the requirements for Certification and
Marking is available from the different National Standards and Certification bodies.
• Use adequately designed equipment that reduces the probability of causing an explosion
(i.e., Ex equipment)
7.1 Introduction
Analysis of occupational hygiene samples may be done on the job using some form of direct
reading device or instrument. Alternatively, a sample is often collected at the workplace and sent
to a laboratory for analysis. This analysis could vary from a relatively simple weighing of the
contaminant on a filter to the determination of a metal using an inductively coupled plasma (ICP)
spectrometer or the use of a gas chromatograph linked to a mass spectrometer (MS) for the
determination of an organic solvent.
In most cases the hygienist does not perform the laboratory analysis, but an understanding of
some of the basics is required to:
The types of analysis can typically be divided into the following main types
• Spectroscopy
• Atomic
• Molecular
• Chromatography
7.2.1 Spectroscopy
The basic underlying principle of spectroscopy is that all elements or chemical compounds absorb
or emit electromagnetic radiation at specific frequencies. If a sample is radiated at a specific
frequency for a particular element, if that element is present the amount of radiation absorbed or
emitted is proportional to the concentration of that element in the sample.
Source: BP International
7.2.1.1.1 Flame Atomic Absorption Spectrometry (AAS)
The sample in solution is atomized by flame and the absorption of a specific wavelength of light
from the hollow cathode lamp in the flame is measured to quantify the element. This technique
typically used for the analysis of approximately 60 metals.
Figure 7-2 Atomic Absorption
7.2.1.1.2 Hydride Generation
Spectrometer
Arsenic and selenium have poor sensitivity using
conventional Flame AAS because their spectral lines
are in the far UV. Hydride generation overcomes this
issue. As and Se are converted to their respective
hydrides AsH3 and H2Se. When these hydrides are
swept through the flame or a through a heated quartz
cell a larger proportion of the element reaches the
light path resulting in increased sensitivity.
Source: University of Wollongong
The principle of the method is based on the absorption of ultraviolet and visible radiation by the
excitation of bonding electrons in molecules.
Source: BP International
Most chemicals species absorb UV or Visible radiation and thus can be quantified, e.g., oil. For
non-absorbing compounds a reaction with a colour producing reagent (a chromophore) may allow
its quantification.
E.g., the reaction of hexavalent chromium with s-diphenyl carbazide to produce a red complex
with an absorption peak at 540 nm.
7.2.1.5 IR Spectrometry
Infra-red (IR) spectrometry provides a way of identifying pure species as each molecular species
has its own unique absorption spectrum, i.e., fingerprint.
atoms and the number of bonds it contains. IR is particularly applicable to organics and covalently
bonded metal complexes.
. Note the distinctive quartz “doublet” at 798 and 779 cm-1 wavenumbers.
7.2.2 Chromatography
Chromatography is a separating method that
relies on differences in partitioning behavior
between a flowing mobile phase and a
stationary phase to separate the
components in a mixture.
A column or other support holds the stationary phase and the mobile phase carries the sample
through it. Sample components that partition strongly into the stationary phase spend a greater
amount of time in the column and are separated from components that stay predominantly in the
mobile phase and pass through the column faster.
As the components elute from the column they can be quantified by a detector and or collected
for further analysis. An analytical instrument can be coupled with a separation method for on line
analysis and includes gas and liquid chromatography with mass spectrometry.
Conventional analysis only showing the amounts of silica and oxygen is not helpful in this
situation; we need to know the form that the silica and oxygen is in. XRD is able to both identify
and quantify the different crystalline phases that have quite different potential health effects.
For example, if a welder is being sampled for “welding fumes” the gravimetric determination, i.e.,
the filter weighing, will be adversely affected if “grinding dusts” have also been sampled during
the fume collection period.
An example of chemical interference is calcium in the presence of phosphate as it forms the stable
compound calcium phosphate, which can reduce the absorbance of calcium.
• NIOSH Manual of Analytical Methods (NMAM) – a collection of over 1,700 methods for
sampling and analysis of contaminants in workplace air, and in the blood. Available on
line at: www.cdc.gov/niosh.nmam.
• UK HSE Methods for the determination of hazardous substances (MDHS Series) more
than 100 methods available on line at: www.hse.gov/uk/pubns/mdhsindex.htm.
• National Standard – A number of standards including the sampling for respirable and
inspirable dust, welding fumes and organic vapours are available through the National
Standards organisations of a number of countries.
• SKC Inc Comprehensive Catalog and Sampling Guide – annual publication and also on
their website www.skcinc.com provides references to the method, sampling parameter,
analysis and equipment for over 2,500 specific compounds.
• Although mainly aimed at measuring contaminant concentrations that are emitted from
a workplace rather contaminant concentrations within a workplace, methods published
by the Environmental Protection Agency (EPA) in the USA may also be worth
considering.
The analyst is often weighing sub milligram quantities of material and greater care has to be taken
during both filter/sample head preparation and filter reweighing after sampling.
Insufficient sampling time may mean not enough material is collected and cannot be detected
unless an appropriate laboratory balance is used.
Calibration of the microbalance is a key aspect and the following extract from AS3640 can be used
as a guide to what is required.
“The accuracy of the microbalance used in the gravimetric measurements shall be checked in the
following manner:
ii) Check the linearity of the balance inside or near to the working range.
4. After every weighing session: Check the calibration of the balance with a reference
weight at or near to full electrical capacity.
If a series of filters is being weighed the microbalance accuracy shall be checked at appropriate
intervals during the procedure.”
7.4 Microscopy
Polarized light microscopy (PLM) together with dispersion staining is the technique that is used for
the identification of types of asbestos fibre, and phase contrast microscopy (PCM) is used for the
counting of fibres.
Fibres are particles that have a needle-like or thread-like appearance with a specific length to
width ratio. Some examples of fibres include asbestos, fiberglass, rockwool and ceramic fibres.
Monitoring for asbestos fibres is carried out following the appropriate Standards methods such as:
Microscopy should only be performed by a trained and certified person. Typically, such persons
routinely participate in an inter laboratory system to maintain their skills and validate their
consistency with international standards.
Figure 7-9 Phase Contrast Microscopy – Amosite After sampling the filters are
Fibres & Synthetic Mineral Fibres mounted on a microscope slide by
collapsing the membrane using
Its accuracy may be tested by analysing known concentrations of the analyte. For example, by
adding known amounts of solvent to charcoal tubes, desorbing it and analysing it by gas
chromatography; or by spiking blood or urine samples with lead for example and analysing by
atomic absorption. The recovery of the analyte is the percentage of added analyte recovered, i.e.,
measured in the analysis.
Precision is determined by analysing enough replicate samples to enable the calculation of the
standard deviation or coefficient of variation. Several different concentrations over the range
should be selected.
The measurement range is a guide as to the usual operating range of the method. At the lower
end this involves an estimate of the limit of detection (LOD) and the limit of quantitation (LOQ).
• Interfering substances
• Capacity of the collection media (e.g., breakthrough volume for sorbent tubes)
• Stability of samples
There are well established and validated methods for many common chemicals.
7.5.1.2 Standards
Standard reagents: are chemicals of known purity and composition. These materials are often
available from external agencies e.g., Standard Reference Materials from the US National Bureau
of Standards.
Calibration standards: these are reference standards against which all test and control samples are
compared.
Where standard calibration curves are prepared at least 5 points should be used and appropriate
regression analysis should be undertaken to ensure the viability of the calibration curve.
7.5.1.3 Blanks
Field sampling blanks should be submitted with field samples to determine if there has been
contamination during sample handling and storage. The blank is treated in the same manner as
the field sample but with no air being drawn through it.
Reagent blanks are used in the laboratory to correct for any contribution made by the laboratory
reagents used in the analysis.
7.5.1.5 Recoveries
Recoveries should be assessed both as part of the method validation process, but also on an
ongoing basis as part of the quality control process.
7.5.1.6 Duplicates
Duplicate samples, i.e., from the field are more useful in assessing the reproducibility of the
sampling or analysis than are duplicate analysis, i.e., two chromatograph injection from the one air
sample.
• NIOSH - Proficiency Analytical Testing (PAT) – solvents on charcoal, asbestos, silica and
metals on filters
They involve the distribution of control samples to laboratories by an outside agency. The material
is analysed and the results returned to the coordinating body for statistical analysis.
This chapter describes and discusses other types of sampling occupational hygienists may be
called upon to do. These include
✓ Bulk Samples
✓ Surface Contamination
✓ Dermal Exposure
The same principles can be applied to other unknown substances found in workplaces. Before
developing a monitoring programme, bulk samples of an unknown material can be sent to a
laboratory for analysis to check on the specific contaminants present and to check for any
contaminants which may interfere with some sampling methods.
The results will guide what type of monitoring strategy is required and thus it is very useful in the
overall process.
There are various methods used for evaluating surface contamination, such as micro vacuuming,
disposable paper towels and manual wipe methods. The manual wipe method (also called smear
and wipe) is the most commonly used and involves a filter paper being drawn over a known area
of contaminated surface and then being analysed to produce an assessment of the level and
nature of the deposit.
Another method which has shown good results in laboratory trials (Wheeler & Stancliffe 1998) is
the use of adhesive tape, more specifically forensic tape. Such tapes are constructed of a clear
plastic top coat, a sticky middle layer and a base layer. By removing the clear plastic top coat, the
sticky layer can be pressed into a surface thus collecting what contaminants are present. In
general samples (both wipe and adhesive tape) are treated with acid to dissolve any contaminants
Other approaches to assessing contaminated surfaces involve the use of pH sticks or colorimetric
pads (acids and alkalis) or instrumentation such as mercury sniffers (the high vapours pressure of
mercury makes this a particularly effective technique).
The question as to why you would undertake surface contamination invariably arises. Such
sampling (especially during evaluation of contaminated waste sites) improves the characterization
of what hazards may be present and allows for better decision-making.
Surface contamination samples can indicate sources of leakage and help to track the spread of
contamination. They can give an indication of how and where skin contact might occur. However,
they are not a direct measure of exposure and cannot readily be compared with any exposure
limits.
In recent years small hand-held XRF analysers have been developed which are extremely useful for
measurements of samples within the field. One such application is their use to measure elements
in contaminated soils and unknown bulk materials. This is particularly useful for metal analysis.
It should be noted that particle size and surface preparation can influence results. Improved
analysis can be achieved if the sample is dried, sieved, ground or pressed.
Dost (1996) evaluated a field XRF unit in relation to the measurement of dusts from surfaces in
workplaces and commented on the ease with which the elemental nature and level of
contamination in the workplace could be determined. Dost also concluded that the XRF technique
had a distinct advantage over the traditional wipe method where the contaminant material was
on a rough and porous surface (e.g., concrete). Conversely, it was not suitable on surfaces such as
steel as it picked up the elements of this surface as well as the overlaying contaminant material.
A common use for XRF instruments is in the evaluation of coatings for the likely presence of
significant amounts of lead.
8.4.1 Direct
Direct means assessing what is deposited on to the skin; indirect means estimating dermal dose
either as attributable to some biologic indicator that is actually measured or that which could
potentially result from a contaminant measured on an accessible surface.
The most common direct method is the use of dermal dosimeters in the form of patches. Other
direct evaluation methods include skin washes and wipes, and the video detection of fluorescent
tracers.
8.4.2 Indirect
Indirect methods refer primarily to measuring a biologic response such as cholinesterase activity in
blood or urinary excretion, but also include measuring surface contamination.
In comparison to air sampling and even biological monitoring, dermal dosimetry is not a simple or
routine procedure.
An individual applying dermal dosimeters should be thoroughly trained regarding the placement
and retrieval of the dosimeters and recording of observations and other information about the
activity.
In addition to objective parameters, observed work practices can also have statistically significant
important influences on dermal exposure.
Each patch dosimeter is a sandwich holding a passive matrix (like a cotton gauze sponge) flat and
to protect it from skin perspiration. Either one or two sets of patch dermal dosimeters can be
used. The most important is the set placed against the skin under the clothing. It is believed that
errors will result from using patch dosimeters attached to the inside of clothing that is free to
move relative to the skin; such dosimeters will neither collect contaminants reaching the skin via
penetration through openings (such as the neck, sleeves, or cuffs) nor be affected by the air
motion carrying contaminant through the weave of the fabric. A second set of dosimeters may be
placed outside of any clothing; it is also important that no inner dosimeter is placed beneath an
outer dosimeter.
After dosimeters have been in place throughout an activity involving exposure, they are carefully
removed, prepared for extraction (the quantitative removal of the chemical from the collection
matrix), and the extract is analysed for the mass of chemical.
Whole body dosimeters are typically a set of long cotton underwear that minimizes the effect of
non-uniform depositions within a body part, but suffers from the lack of a barrier between the
skin and dosimeter and may add heat stress to the wearer. After use, the whole-body dosimeter
may still be dissected into portions covering individual body parts.
In other cases, such as chromium VI, Styrene or (Tetraethyl lead) where skin absorption can be a
significant route of exposure, a combination of environmental monitoring and biological
monitoring may give the most accurate picture of employee exposure.
The tool kit was constructed by analysing the major determinants of dermal hazard and control
exposure. The results were combined in the form of a decision tree. The tool kit does not show all
the details behind the assessment, but asks the user a series of questions that are translated by
the system into hazard and exposure categories that lead to an estimate of health risk from
dermal exposure together with suggested control strategies.
8.4.2.2 Hazard
The user is asked to enter the identification of the chemical and the risk phrases and any
additional information such as pH and the physical state of the chemical.
The information is translated into two hazard categories – one concerning local effects, the other
systemic effects after uptake through the skin. The hazards are rated – negligible, low, moderate
high, very high or extreme.
8.4.2.3 Exposure
User asked to enter information to identify the workplace or process that is assessed and which
one of the Dermal Exposure Operational units’ best fits with the sub category of exposure to solid
or liquid:
From the information the tool kit will apply default exposure rates, take into account duration and
the exposed body areas and the actual exposure score from local effects and the internal exposure
score from systemic effects are then calculated separately and ranked as health risk scores with
suggested controls ranging from no action required up to substitute in either case and stop
working.
The toolkit is an attempt to adapt elements of exact science to a situation where the necessary
input data are of limited quality and are only estimates. The purpose is to enable the user to
estimate the order of magnitude of hazard, exposure and risk and to encourage the user to deal
with issues of dermal hazard, exposure and control.
The RISKOFDERM project has been the subject of significant controversy and more detail can be
found in an overview by Oppl et al (2003).
Please be sure to understand the definition of a confined space in your jurisdiction, as definitions
and air monitoring requirements may vary.
A confined space is determined in part by the hazards associated with entry into such a space and
not just work performed in a physically restrictive location.
The presence of chemical agents (alone or in combination) may present a risk to personnel in a
confined space that would not otherwise occur in the general atmosphere.
Some of the hazards that may be associated with work in confined spaces are:
• Hazardous Substances: This includes the use of chemicals, previously stored substances
or their by-products (e.g., H2S from decomposing plant material), fumes from welding,
painting, etc.
• Flammable Atmospheres: This includes gases, vapours and dusts which are present in
the explosive range.
• Physical and Other Factors: This includes manual handling, ignition hazards, electrical
hazards, mechanical hazards, noise, radiation, biological hazards and heat stress.
As permit to enter procedures for confined spaces invariably involve a risk assessment, this
process should ensure that appropriate arrangements are put in place to test the atmosphere
within the confined space.
The common means of sampling the air to assess the risk of adverse health effects is to test for
specific materials with a suitable portable analyser. There are many different kinds of analysers
available but the results are only as good as the operator’s skill and the state of analyser
Instruments used for testing the atmosphere in a confined space should be selected for their
ability to measure hazardous concentrations and should be calibrated in accordance with the
manufacturer’s guidelines or manuals.
If atmospheres that are to be sampled are potentially explosive, intrinsically safe monitoring
equipment will be necessary. Initial monitoring should be performed from outside the confined
space by inserting a sample probe at appropriately selected openings. Telescopic extension probes
or probes attached to a line can be used to reach remote regions.
Some gases or vapours are heavier than air (for example, hydrogen sulphide) and in unventilated
areas will settle to the bottom of a confined space. Also, some gases are lighter than air (for
example, methane) and will be found around the top of the confined space. As it is possible for
contaminants to settle at different levels, the top, middle and bottom of a space should be
sampled. Horizontal spaces should also be sampled at representative intervals along their length.
Sampling should be such as to reflect accurately the conditions within the confined space.
When considering the appropriate time to monitor the atmosphere, it should be understood that
unless monitoring is undertaken immediately prior to entry, the results may not be relevant and
an unsafe condition may potentially exist.
While pre-entry testing indicates whether the atmosphere in the confined space is acceptable for
entry, atmospheric conditions in the confined space can change, therefore the atmosphere should
be re-tested during the work day.
Testing the atmosphere within the confined space while work is in progress will indicate whether
or not the ventilation system is adequate or if the work processes are making the atmosphere
unsafe.
• Because of the potential for later release or disturbance of hazardous material. Such
material includes sludge, scale or other deposits, brickwork and liquid traps. The
hazardous material may be released if disturbed or if heat is applied. Where harmful
• Because of the work undertaken in the space. For example, heat or fumes from
processes such as welding can build up rapidly in a confined space.
No matter what type of instrumentation is used to assess a confined space (or any other
workplace), it is important that the operator clearly understands the limitations of that
equipment. For example, an explosimeter exhibits different sensitivities towards different
flammable gases or vapours and thus to give accurate results it should be calibrated with known
concentrations of the gas or vapours likely to be present in the atmosphere being assessed.
Moreover, most chemical sensors used for the measurement of contaminant gases are fitted with
filters to minimize cross sensitivity from other contaminants. These filters need to be replaced
according to the manufacturer’s instructions and the potential problems of cross sensitivity well
understood by the instrument operator.
It should also be noted that monitoring is never a substitute for the systematic and verified
isolation of the confined space from any outside source of hazardous material.
In most commercial systems filters are installed to control moisture, oil mist and carbon
monoxide, but these have a finite life and need to be changed when expended.
There are varying approaches to monitoring these contaminants in the air but the advent of direct
reading devices has made the inline analysis of carbon monoxide on site relatively easy.
In modern systems continuous monitoring instrumentation for carbon monoxide and built-in
filtration are common. For older systems it may be necessary to sample the breathing air using
external procedures. In such cases air is drawn into a gas sampling bag from which it is extracted
and presented to the instrument (carbon monoxide monitor or indicator tube) for measurement.
Oil mist is usually sampled by passing a known volume of air through a small pore size filter. The
collect oil is either analysed gravimetrically or more accurately by infra-red or gas
chromatographic means.
This chapter addresses the basics of how to analyse and present the results of occupational
exposure measurements
9.1 Background
Accurate reporting of data is as important as sample collection as they provide a record of what
was done, where, when, why and by whom. It is therefore important to be clear on the objectives
of exposure monitoring effort, and to keep careful and detailed field notes in a consistent fashion.
Besides reporting findings made, reports should address the meaning of the data and should
include recommendations (including whether further sampling is necessary) as appropriate. Note
that in some jurisdictions such as the US, an exposure evaluation report is considered an exposure
record for which there are specific requirement regarding how long an employer is required to
retain the report.
Preparing an exposure monitoring report requires identifying the intended audience and affected
stakeholders such as:
The BOHS also published a guidance document for clear and concise report writing (see
https://www.bohs.org/information-guidance/bohs-resources/ ), which builds on previous
documents prepared by AIOH 2006. This guidance is the basis for BOHS assessment of reports as
part of personal learning portfolios (PLPs) for BOHS certificate and or diploma candidates.
There are also two national standards that address occupational exposure assessment report
structure and content: British Standard BSEN689 and Australian Standard AS 2985 (see text
boxes).
• Introduction - What the objectives of the survey were. Who did the work and where.
What were the limitations?
• Results and Discussion - What was the process investigated? What did the lab report?
Calculate 8h-TWAs as needed. What do the results mean? What was observed on site
that could explain the results?
It’s always a good idea to have a colleague review your report before sending it out. The checklist
provided in the current BOHS guidance is very useful.
“Reports shall be written of the occupational exposure assessment and of any periodic
measurement. Each report should give reasons for the procedures adopted in the particular
workplace.
• The report has to contain:
• The name of the person(s) or institutions undertaking the assessment and the
measurements;
• The description of the workplace factors including the working conditions during the
measurements;
• Date on which the test was carried out and sampling duration.
• If uncertainties are not formally derived, for sampling periods greater than 60
minutes the concentration should be reported to two decimal places and three
significant figures for six place microbalances, and to one decimal place and two
significant figures for five-place microbalances.
• The identity of any reference material used to assist in the validation of the test
results.
• Any observation, in relation to either the test sample or the performance of the test,
which may assist in the correct interpretation of the test results.
• Area/plant/process surveyed, e.g.,” a survey of the area known as cold press or CP was
conducted”.
• Conditions at the time (i.e., personnel, process conditions, risk controls in place) e.g.,
“usual operator unavailable”, “shutdown”, “worst case situation, with no controls”, “as
normal, believed to be a representative working day”, “only Blender No.2 was
operating”, “protective equipment worn other than overalls”.
• Identify any items examined, e.g., “Toolmaster serial number 123”, “machine called the
hot block curer”.
• Number of employees, duration of work shift(s) and task frequency and duration, e.g.,
“9 employees work an 8-hour day, 5-day week with 2 hours overtime worked
infrequently”, “it takes about 30 minutes for 5 bags to be opened and poured daily”.
• Results of personal sampling should be compared with the relevant exposure standard.
If there is no relevant exposure standard, it will be necessary to either modify or adapt
an existing guideline or develop a guideline. The rationale justifying the guideline used
should be provided.
• Peak/ceiling limits.
• Results should be compared with any previous surveys at the premises and data from
similar premises if available, e.g., “The process produced results that are similar to other
coating operations”. An explanation of general trends and unusual high or low trends
should be included.
• The level of risk should be determined (preferably quantitatively) to allow for the
adequacy of controls to be assessed and the prioritization of control options.
Conclusions should also be drawn about adequacy of control and any further practical actions to
eliminate or reduce the assessed risk so far as is practicable, e.g., “existing controls adequate if
maintained”,” existing controls not adequate and need to be upgraded”.
What is fundamental in all cases is that the information collected and evaluated is communicated
to all the stakeholders involved in the exercise in a manner and format that meets their needs or
expectations. In almost all cases this will be different for each of the stakeholders.
This primer is intended to provide a brief overview of statistics and statistical tools in the context
of occupational hygiene data. OHTA 501 Students are expected to be familiar with the general
concepts but will not be required to apply them as part of this course. However, competence and
board certifications in occupational hygiene will require a deeper understanding and application of
these and other concepts.
The following is suggested for further reading: Milz S, and J Mulhausen (2015) “Appendix IV:
Descriptive Statistics, Inferential Statistics, and Goodness of Fit” In A Strategy for Assessing and
Managing Occupational Exposures, 4th edition, AIHA, Fairfax, VA.
Note that statistical analysis assumes random sample collection, which may not always be the
case in occupational hygiene work. Often, processes vary, or anomalies may occur on the day of
sampling. Moreover, Nonetheless, statistical analysis provides useful insight.
Table 9-1. The equations shown below can be used to calculate those values, although most
spreadsheet programs calculate parameters such as mean and standard deviations easily, as does
the AIHA’s IHSTAT spreadsheet (see https://www.aiha.org/public-resources/consumer-
resources/topics-of-interest/ih-apps-tools).
1
𝑥̅ = ∑𝑥
𝑛 …Equation 9-1
Where x = result
n = number of sampler
𝑥̅ = arithmetic mean
1
ln(𝐺𝑀) = ∑ ln 𝑥
𝑛
1
𝐺𝑀 = 𝑒 𝑛 ∑ 𝑙𝑛(𝑥) …Equation 9-3
1
ln 𝐺𝑆𝐷 = ∑(ln(𝑥) − ln(𝑥̅ ))2
𝑛−1
1 2 …Equation 9-4
𝐺𝑆𝐷 = 𝑒 𝑛−1 ∑(𝑙𝑛(𝑥)−ln(𝑥̅ ))
Care needs to be taken when analysing data that contains some results reported as less than a
specified quantity (i.e., results are reported as non-detect, ND, or < LOD). Non detect results
represent neither zero nor a value just less that the limit of detection so should necessarily be
discarded when evaluating a collection of data. The University of Montreal has created an online
tool to address this issue (https://espum.umontreal.ca/lespum/departement-de-sante-
environnementale-et-sante-au-travail/production-scientifique/utilitaires/ndexpo/).
The Arithmetic means (AM and GM) scan be compared to OELs directly but are not informative as
to how likely it is that the criterion might be exceeded. Examination of Figure 9-1 indicates that it
is hypothetically possible to exceed any given value, so it is of value to get a sense of how likely
that might be. Standard deviations (SDev) and geometric standard deviations (GSD) provide some
insight into this.
Standard deviations essentially indicate how widely spread the underlying data is. Geometric
standard deviation values also can provide an insight into the nature of the underlying data as
shown on Table 9-2.
Table 9-2 Rules of Thumb GSD Values
1.0 No variability
Nowadays, these means and standard deviations (normal and log normal) are easily calculated in
most spread sheet applications. Online tools such as AIHA’s IHSTAT (https://www.aiha.org/public-
resources/consumer-resources/topics-of-interest/ih-apps-tools) simplify statistical calculations
and encourage the application of statistical concepts for evaluating data.
Knowledge of AM and SDev, as well as GM, and GSD, allow inferential analysis of the underlying
data by estimating various parameters such as confidence levels, and MVUE.
In some instances, estimated 95% UCL values are compared to exposure criteria to determine
whether a process is adequately controlled, particularly where the process involves substances
that exhibit acute toxicity.
It is easier to use spreadsheets or online tools such as AIHA’s IHSTA to estimate confidence limits.
However, they can also be estimated using log probability plots of data
Association 2007)
MVUE values are difficult to calculate because of the iterative nature of its calculation.
The plot can be done manually by ranking the results, 1 being used for the lowest value,
determining their plotting position as follow:
𝑅𝑎𝑛𝑘
𝑃𝑙𝑜𝑡𝑡𝑖𝑛𝑔 𝑃𝑜𝑠𝑖𝑡𝑖𝑜𝑛 (%) = 𝑥 100%
𝑛+1
…Equation 9-5
Where Rank is rank of result with lowest = 1
The resulting data can then be plotted on log normal paper, with % position on the y-axis, and
corresponding result on the x-axis. A linear plot indicates the data is log normally distributed. See
The GM can be found by reading off the concentration at the 50th percentile, so in the above
case, it would be 0.06 ppm (assuming the plot data is in ppm).
The 95th percentile (used as a measure of compliance by some organisations) can also be read off
the graph (around 0.40 ppm in the example above)
Log probability plots are easily done on software available on line such as AIHA’s IH Stat
(https://bit.ly/3SVKEfQ). They can also be done on linear plots on Excel, though you may need to
work with log converted concentrations, and will need to convert back when trying to derive
actual values (e.g., GM, 95%-ile) off the chart.
In contrast, workplace air monitoring and comparison of the results with exposure standards
provides information about the probable exposure of workers to inhalation hazards. It does not
provide information about the other exposure routes of skin absorption, ingestion, nor
information about any non-work-related exposures.
Biological monitoring is one of the three tools used in the prevention of disease from hazardous
substances in the work environment, the other two being occupational hygiene or environmental
monitoring and health surveillance.
Biological monitoring means the assessment of exposure to chemicals (substances) that are
present in the workplace, through the measurement of appropriate determinants in biological
specimens from exposed workers. In most cases, the specimen used for biological monitoring is
urine, blood or exhaled air.
The risks associated with the obtaining and handling of bodily fluids, in terms of potential
exposure to possible pathogens, i.e., HIV, Hepatitis, viruses etc. have to be considered.
In many countries only a qualified doctor or nurse can obtain such samples. Local advice must be
sought before such work is to be carried out.
• Expired air – e.g., for carbon monoxide and organic solvents – e.g., benzene or analysis
of its metabolites
• Zinc protoporphyrin in blood – these levels increase with exposure to lead, because lead
inhibits the biosynthesis of heme.
Biological effect monitoring is not health surveillance through which individuals with early signs of
adverse health effects are identified.
The loss of chemical from the body or elimination depends on metabolism and excretion.
Chemicals may be eliminated by numerous routes including fecal, urinary, exhalation, perspiration
and lactation.
A chemical may be excreted from the body without metabolism, i.e., the particular chemical can
be measured directly. In other cases, the chemical may be metabolized through oxidation,
reduction, hydrolysis or combination of these followed by often very complex biochemical
reaction in the body. Hence the choice of the indicator of exposure and even the timing of when
to take a sample is critical.
An exposure that occurs through the ingestion route typically takes longer to enter the blood
stream than an exposure through the inhalation route. So, it is possible to collect the sample too
soon after the exposure occurs, although this is not usually an issue in relation to workplace
exposures.
End of the work week After 4 or 5 consecutive working days with exposure
The UK Health & Safety Executive (HSE) in the Guidance Note EH56 “Biological Monitoring for
Chemical Exposures in the Workplace” (HSE 1992) uses the following Table 10-2 Optimum Time
for Collecting Samples) to provide advice on the timing of sample collection.
• By correcting for the creatinine level in the urine, as creatinine excretion from the body
occurs naturally at a nearly constant rate.
The World Health Organisation has adopted the following guidelines for acceptable limits to assist
in overcoming the issues associated with highly diluted and highly concentrated urine samples:
Some Biological Exposure Indices BEIs® for determinants whose concentrations is dependent on
urine output are expressed as relative to creatinine concentration. For other determinants
correction for urine output is not appropriate.
Biological Exposure Indices (BEIs®) are guidance values for assessing biological monitoring results.
BEIs® represent the levels of determinants that are most likely observed in specimens collected
from healthy workers who have been exposed to chemicals to the same extent as workers with
inhalation exposure at the TLV®.
In a similar fashion to TLVs®, BEIs® are to be used as guidelines in the evaluation of occupational
hygiene health hazards. BEIs® do not indicate a sharp distinction between hazardous and
nonhazardous exposures. Due to the often-varied nature of concentration in biological specimens’
great care and caution must be exercised in the interpretation of the results from a single
specimen.
BEIs® apply to 8-hour exposures, 5 days per week. The BEI® Committee does NOT recommend
adjusting or applying a correction factor to the BEIs® for altered or extended shift patterns.
Use of BEIs® should only be done by experienced occupational health professionals in consultation
with the associated documentation for them. The BEI® is a guideline for the control of potential
health hazards for workers and the values are inappropriate for use for the general public and for
non-occupational exposures. In the application of BEIs® reference must be made to the current
edition of the Documentation of the Threshold Limit Values and Biological Indices from the
ACGIH®.
10.8.2 Notations
A notation is a designation that appears as a component of the adopted BEI® value to provide
additional information with respect to the particular chemical:
“B” = Background
The determinant may be present in biological specimens collected from subjects who have
not been occupationally exposed, at a concentration which could affect the interpretation
of the result.
“Nq” = Non-quantitative
Biological monitoring should be considered for this compound based on the review;
however, a specific BEI® could not be determined due to insufficient data.
“Ns” = Non-specific
“Sq” = Semi-quantitative
The biological determinant is an indicator of exposure to the chemical, but the quantitative
interpretation of the measurement is ambiguous.
These determinants should be used as a screening test if a quantitative test is not practical or as a
confirmatory test if the quantitative test is not specific and the origin of the determinant is in
question.
10.8.3 UK Limits
In the UK the HSE has established a system of non-statutory biological monitoring guidance values
as an aid in the interpretation of biological monitoring data.
Biological Monitoring Guidance Values (BMGVs) are set where they are likely to be of practical
value, suitable monitoring methods exist and there are sufficient data available. The type of data
that are available will vary between substances and therefore the route taken to deriving the
BMGV will vary between substances. BMGVs are either based on a relationship between biological
concentrations and health effects, between biological concentrations and exposure at the level of
the WEL or are based on data collected from a representative sample of workplaces correctly
applying the principles of good occupational hygiene practice. The technical basis for each BMGV
will be clearly described in supporting documentation such as an EH64 summary or other
guidance.
BMGVs are non-statutory and any biological monitoring undertaken in association with a guidance
value needs to be conducted on a voluntary basis (i.e., with the fully informed consent of all
concerned). BMGVs are intended to be used as tools in meeting the employer’s primary duty to
ensure adequate control under COSHH. Where a BMGV is exceeded, it does not necessarily mean
that any corresponding airborne standard has been exceeded nor that ill health will occur. It is
intended that where they are exceeded this will give an indication that investigation into current
control measures and work practices is necessary.
Of course, that is not necessarily to say that because biological monitoring results are below a
particular guidance value an employer need take no further action to reduce exposure; but it
should be noted that BMGVs are not an alternative or replacement for airborne occupational
exposure limits.
10.9 Confidentiality
There are several ethical and confidentiality issues that must be considered and implemented
before commencing a biological monitoring programme.
• The informed consent from the participants is needed. This consent must only be given
when the participant feels no fear of reprisals if their consent is not given.
• Results of the monitoring should be kept confidential and shared only with the
occupational health professional and the participant.
This section contains a basic review of physiology and toxicology. This subject is covered in
greater detail in student manual of the OHTA module W507 Health Effects of Hazardous
Substances.
Arteries bring the oxygenated blood, pumped from the heart, to the tissues and the veins bring
the deoxygenated blood back to the heart. Blood passes from arteries to veins through capillaries,
which are the thinnest and most numerous of the blood vessels.
Hormones act as messengers and are carried by the bloodstream to different cells in the body
which then interpret the message and act on them. Examples include the pituitary gland, the
thyroid gland, adrenal gland and the pancreas and gonads.
The skin is often known as the largest organ of the body and as the interface with the
surroundings it provides protection against the physical hazards such as heat, radiation and
abrasion, chemicals and bacteria. Its other important functions are insulation and temperature
regulation, sensation and Vitamin D and B synthesis.
• Absorption of fatty acids and subsequent transport of fat to the circulatory system.
In general, the function of muscle is to produce movement, maintain posture, stabilize joints and
to generate heat.
Muscles are attached to bone by tendons and other tissues. They exert force by converting
chemical energy into force. Nerves link the muscles to the central nervous system.
The peripheral nervous system is divided into the somatic nervous system and the autonomic
nervous system.
The somatic nervous system is responsible for coordinating the body's movements, and also for
receiving external stimuli. It is the system that regulates activities that are under conscious
control.
The autonomic nervous system is then split into the sympathetic division, parasympathetic
division, and enteric division. The sympathetic nervous system responds to impending danger or
stress, and is responsible for the increase of one's heartbeat and blood pressure, among other
physiological changes, along with the sense of excitement one feels due to the increase of
adrenaline in the system. The parasympathetic nervous system, on the other hand, is evident
when a person is resting and feels relaxed, and is responsible for such things as the constriction of
the pupil, the slowing of the heart, the dilation of the blood vessels, and the stimulation of the
digestive and genitourinary systems.
The role of the enteric nervous system is to manage every aspect of digestion, from the esophagus
to the stomach, small intestine and colon.
The female reproductive system consists of the internal organs including the ovaries, fallopian
tubes, uterus, cervix and vagina.
Air then travels along the bronchioles to its ending (the terminal bronchiole) which is covered in
tiny multi lobed sacs called alveoli where most of the gas exchange occurs.
The most obvious function of bone is to support the body. It is also the site of hematopoiesis, the
manufacture of blood cells that takes place in bone marrow and why bone marrow cancer is very
often a terminal disease. The skeleton is also necessary for the protection of vital organs. Human
The kidneys are one of the various organs (together with the lungs, intestine and skin) that
participates in the elimination of the wastes of the organism. The kidneys are bean-shaped organs
about the size of a bar of soap. They are near the middle of the spine, just below the ribcage. They
are situated behind the organs of digestion within the abdominal cavity. Situated on the superior
surface of each kidney is an adrenal gland.
A kidney consists of about 1 million filtering units termed nephrons, each consisting of a
glomerulus, ball-shaped network of capillaries, and a network of tubules. Blood plasma is filtered
by the glomerulus, and the resultant "prourine" passes through the tubular system where water,
and nutrients are reabsorbed under the supervision of hormone activity and the autonomic
nervous system.
Humans produce about 1.5 litres of urine over 24 hours, although this amount may vary according
to circumstances. Increased fluid intake generally increases urine production, while increased
perspiration and respiration may decrease the amount of fluid excreted through the kidneys. A
reduced intake of water will normally result in less urine production as well. Some medications
interfere directly or indirectly with urine production, such as diuretics.
The kidney plays a crucial role in regulating electrolytes in the human blood (e.g., sodium,
potassium, calcium). pH balance is regulated by the removal of excess hydrogen ions (H+) from
blood. In addition, they remove urea, a nitrogenous waste product from the metabolism of
proteins from amino acids. The metabolism process forms ammonia which is transported by blood
to the liver and detoxified to a less harmful byproduct called urea.
Inhalation: The requirements of a man in a normal day are approximately 3.4 kg food and water
(water is obtained in the food we eat and as direct ingestion). For light physical work an average
person breathes in between 1-1.2 m³ of air per hour. This rate would be much higher for heavy
physical exertion.
Therefore, it is easy to understand why inhalation is by far the most common route of entry due to
both the volume of air coming into contact with the large surface area of the lungs and the thin
cell layer in the lungs separating the air from the blood, with skin absorption next (especially
pesticides) and ingestion last. Inhalation is the major route of entry of dusts, fumes, mists, gases
and vapours into the body.
Eye: The eye is a relative minor route of entry into the body. It should also be noted that the eye is
also at risk from direct contact with chemicals.
Ingestion: Ingestion is usually a relatively minor route of absorption of chemicals in the workplace.
It is commonly as a result of an accidental ingestion, or from poor personal hygiene e.g., eating
with dirty/contaminated hands. In some workplaces inadvertent ingestion or skin absorption can
result from contact with contaminated surfaces.
It should be noted that insoluble aerosols can end up in the digestive tract from where they can be
absorbed into the body. Additionally, involuntary ingestion as a result of clearance mechanisms in
the upper respiratory tract can also be another route of entry, especially in the case of large
particles of toxic substances.
Target organs are defined as organs in which critical effects are observed as the result of exposure
to a harmful input. There are many identifiable instances of inputs which affect a number of
critical organs. Which they affect depends upon the circumstances of exposure, the interplay of
defense processes and the susceptibility of the individual, as well as the tissues of the target
organ. Thus, in discussing effects it is required that all possible target organs are considered.
The definition of 'target organs' must, necessarily, be wide, and must include, where appropriate,
systems and tissues as well as organs.
For example, the target organ of hydrogen sulphide, which attacks the nerve tissue and causes
respiratory paralysis, might be categorized as the central nervous system.
Crocidolite induces serious disease of the pleura and peritoneum (the tissue lining in the inner
surface of the chest wall, and the lungs or the inner surface of the abdominal cavity and the
abdominal organs). In this instance the pleura and peritoneum are the target organs.
A series of target organs and an outline of their principal functions are given in Table A-3.
Skin Protects against friction, water/fluid loss, entry of harmful inputs; thermal
insulation; self-greasing by means of sebaceous glands; thermoregulatory
by means of sweat glands; receives afferent information.
Respiratory tract Oxygen and carbon dioxide exchange; defense against aerosols; warming
and moistening of incoming air; excretion of gases, vapours.
Kidney, urinary tract Excretion: Water, salts and nitrogenous wastes (includes homeostasis
as well as bio- dumping).
Storage: a) Vitamins
b) Iron (for hemoglobin)
c) Glycogen-energy store substance
Brain and nervous system Information processing and control of bodily activities.
Bone Support framework for movement and protection (certain bones house
blood-forming organs; but those are functionally separate from bone).
Lymphoid system and Tissue drainage; filtration; site of defensive processes such as immune
lymphatics response and phagocytosis.
Ideally dose should be defined as the concentration of a substance at the site of effect, regard
being made for the time for which the substance concentration is maintained. For practical
This simplistic equation does not account for the following factors:
• Dose may be less than the amount inhaled if most is exhaled without any absorption
(e.g., many gases).
• Heavy physical workload results in higher breathing rates than light workloads and thus
have higher doses.
• Additional exposure may come from non-occupational sources (carbon monoxide from
smoking).
Effect can be any observable, biological change associated with the input concerned, and ideally it
should be quantifiable. It is implicit in dose-effect relations that effect is related to and caused by
the dose.
Effect does not necessarily denote an adverse biological change, but embraces any biological
change. Certain effects can be beneficial and only become adverse if the dose is excessive or
remains for a critical period of time.
Acute or immediate effects occur during or immediately after exposure and last for a short period
of time. Examples of acute effects include the immediate eye and respiratory tract response to
exposure to, and inhalation of, chlorine or burns to the skin caused by direct contact with strong
acids or alkalis.
Chronic effects are long lasting and may be, but not necessarily, permanent. Some examples of
chronic exposures are pneumoconiosis from long term exposure to coal dust, silicosis after
exposures to quartz dusts.
Local effects occur at the point of entry to the body of the toxin and systemic effects are
associated with distant target organs (e.g., with lead the main route of entry is by inhalation but
the toxic effect is upon the blood forming process, nervous system, kidneys and reproductive
functions).
Critical organ concentration seems, given the present state of knowledge, to be the parameter of
greatest utility in estimating dose. Whole body concentration provides a less useful criterion,
because the organs in which greatest accumulation occurs may not be critical organs.
At some time in the future, it will, no doubt, be possible to estimate dose in terms of critical cell
concentrations - or subcellular concentration - but at present this is impracticable.
There are complexities in the specification of effect, since certain effects, such as death, are of an
all-or-none character, while others are of a graded nature, such as occupational deafness.
Specification is further complicated by the fact that certain all-or-none effects (cancer, for
example) require only a trigger. Once triggered they continue by self-propagation or by the other
processes independent of the dose of the triggering input. On the other hand, many observable
and gradable effects are both trivial and reversible.
However, the complexities do not end here. The specification of dose needs to take account of all
possible modes of input, and the non-occupational as well as the occupational possibilities. For
example, in the case of metals like lead, in most if not in all countries, input by ingestion from the
normal diet is inevitable. Any occupational exposure, probably by inhalation, will be supplemented
by the non-occupational dose. Combination of the two may cause a critical organ concentration to
be reached in the bone marrow or in other organs.
In toxicology, the NOAEL is specifically the highest tested dose or concentration of a substance at
which no adverse effect is found in the exposed test species (usually animals or cells).
This level is commonly used in the process of establishing a dose response relationship, a
fundamental component in most risk assessment strategies.
Another important toxicological concept is “lowest observed adverse effect level” (LOAEL) or the
lowest dose or concentration that causes any observed adverse effect. Thus, by definition the
NOAEL is less than the LOAEL.
As these determinations of exposure and effect have generally been established in species other
than humans, various safety factors or uncertainties are applied before this data is used in the
establishment of workplace exposure standards.
The precise threshold for a given effect can, and usually does, vary within certain limits with
species, with individuals within a species, and perhaps even with time in the same individual.
For a given population, as illustrated by the dose response relationship (see Figure 10-3 Dose
Response Curve), it is clear that thresholds exist because it can be determined experimentally that
certain low levels of exposure will produce no detectable effect, and that as the dosage is
increased the effect appears.
Since the dose-response relationship is a continuum, somewhere between the experimental no-
effect and effect levels is the turning point known as the threshold.
Dose-response curves typical of those plotted from data obtained in chronic toxicity experiments
exist for a number of contaminants. It is very important to recognize that such a curve is drawn
from only several points, one for each exposure group in the experiment. The greater the number
of exposure groups, the greater the number of points, and hence, the greater the accuracy of the
curve that is drawn. But without an infinite number of points, the precise shape of the dose-
response curve cannot be known.
The curve is interpreted as follows: with chronic exposure of increasing doses up to the threshold,
no effect is detectable because some biochemical or physiologic mechanism, handles the chemical
in a manner that prevents an effect from occurring. At the threshold, the defense mechanism is
saturated, or in some manner
overwhelmed, for the more Figure 10-3 Dose Response Curve
Source: Tranter 1999 – Reproduced with
Another toxicologic question relates to the shape of dose-response curves for carcinogens as they
approach zero dose. The inability of toxicology to answer this question by experiment has given
In regard to carcinogens, it is important to note that it is rare to have any data except for high
doses, so the estimate of the shape of the dose response curve below the lowest actual data point
must typically cover many orders of magnitude. Where a threshold cannot be identified, limits are
generally risk based and dependent upon the dynamics of the particular substance.
What should be recognized is that in any group of test subjects there are some susceptible
individuals (hypersensitive) who are affected at low concentrations of the test contaminant and
there are also some highly resistant individuals (hyposensitive) who are not affected at high
concentrations but there are the vast majority of “average” individuals in the middle (Figure 10-4
Variability of Human Exposure to Dose).