FLOW-I Service Manual
FLOW-I Service Manual
Table of contents 1
1 Important 13
1.1 General 13
1.1.1 System version and system software version 14
1.1.2 Serial number 14
1.2 Symbols used in this manual 14
1.3 Hazard notices 15
1.4 Installation 16
1.5 System checkout 16
1.6 Service 16
1.7 To the responsible service personnel 17
1.8 Conditions for displayed measured values 17
1.8.1 BTPS, Body Temperature and Pressure Saturated 17
1.8.2 ATPD, Ambient Temperature and Pressure Dry 17
1.9 Environmental declaration 17
1.9.1 Components with special environmental concern 17
1.9.2 Printed circuit boards 17
1.9.3 Other electronics 18
1.10 Construction materials 19
1.10.1 Metal – total 70% 19
1.10.2 Polymeric material – total 10% 19
1.10.3 Electronics – total 20% 19
1.10.4 Others – very small amounts 20
1.11 Articles of consumption 20
1.12 Power consumption 21
1.13 Packing materials 22
1.14 Disassembly and disposal instructions 22
2 Introduction 23
2.1 General 23
2.2 Functional units 23
2.2.1 Front side of the system 24
2.2.2 Rear side of the system 26
2.3 Optional equipment 27
2.4 Electronic structure 27
2.4.1 General 27
2.4.2 CAN bus 28
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 1 (of 369)
Table of contents
Page 2 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Table of contents
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 3 (of 369)
Table of contents
Page 4 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Table of contents
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 5 (of 369)
Table of contents
Page 6 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Table of contents
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 7 (of 369)
Table of contents
Page 8 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Table of contents
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 9 (of 369)
Table of contents
Page 10 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Table of contents
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 11 (of 369)
1 Important
1.1 General
The information in this Service Manual covers Flow-i anesthesia system up to system version 4.7. Unless
otherwise stated, the Service Manual is also valid for system versions below 4.7.
The information in this manual refers to different models or versions of the Flow-i anesthesia system. These
models are mapped to their serial numbers (see section 1.1.2 on page 14). In order to recognize the manual
information referring to a certain model in an easy way, this manual is provided with labels indicating the
model described in the following paragraph. Where applicable, there is a table of content in the beginning of
each section, giving an overview where in the section to find the information referring to a certain model.
For some sections the information instead is mapped to the software version, the system version or to a
specific version of a component.
Unless otherwise stated all figures show Flow-i system with S/N > 20 000.
Note: Not all Products (Device types), System versions, System software versions and Software options
described in this document are available in all countries. For details, contact your Getinge representative.
Service documentation for the Flow-i anesthesia system consists of:
User’s Manual including Cleaning and Maintenance manual. The User’s Manual is an indispensable
complement to the Service Manual for proper servicing.
Service Manual
Description of error codes
Installation Instructions
Spare parts information
Documentation for all optional equipment included in the system
Product End-of-Life Disassembly Instructions (Disassembly and Disposal Instructions)
Contact MCC HSC or your local Getinge representative for more information.
Maquet Critical Care AB, part of Getinge, is the legal manufacturer of the Flow anesthesia systems. Maquet
Critical Care AB is referred to as “manufacturer” in this document.
The Flow-i anesthesia system is referred to as “the system” in this document.
The Flow-i anesthesia system User’s Manual is referred to as “the User’s Manual” in this document.
Maquet Headquarter Support Center is referred to as “MCC HSC” in this document. Note that MCC HSC
is available only for the Getinge service organization and authorized distributors. For other service
organizations, such as hospital biomedical departments, technical support is provided by the local
Getinge representative.
In addition to important information given here and in the related documents, always pay attention to
applicable local and national regulations.
Responsibility for the safe functioning of the equipment reverts to the owner or user in all cases in which
service or repair has been done by a non-professional or by persons who are not employed by or authorized
by Getinge, and when the equipment is used for other than its intended purpose.
The system complies with standards and requirements as stated in the User’s Manual.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 13 (of 369)
Chapter 1 – Important
References to IEC 60601-1 in the Service Manual refer to edition IEC 60601-1:2005 + A1:2012.
References to EN 80601-2-13 in the Service Manual refer to edition ISO EN 80601-2-13:2012 +
A1:2015.
References to IEC 60601-1-2 in the Service Manual refer to edition EN 60601-1-2:2015.
The system serial number is printed on a label that is affixed on the lower right part of the system, close to
entry of the mains power cable.
To display the serial number on the control panel, push the Menu button and go to System Info.
CAUTION: Indicates instructions that must be followed in order to ensure the proper operation
of the equipment.
Page 14 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 1 – Important
ESD sensitive components. When handling ESD sensitive devices, established procedures
must be observed to prevent damage.
Special waste. This product contains electronic and electrical components. Discard disposable,
replaced and left-over parts in accordance with appropriate industrial and environmental
standards.
Trained and authorized personnel. Only personnel trained and authorized by Getinge shall be
permitted to perform installation, service or maintenance of the system.
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit. All personnel must exercise extreme caution if fault tracing or
adjustments are performed with power supply connected and with covers removed.
CAUTION: With gas supply connected to the system, there are pressurized components inside
the unit. All personnel must be very cautious if fault tracing or adjustments are performed with
gas supply connected and with covers removed.
Some items, e.g., gas modules, are still under pressure when gas supply is disconnected.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 15 (of 369)
Chapter 1 – Important
1.4 Installation
Only personnel trained and authorized by Getinge shall be permitted to install the system. The
installation and handing-over procedures are described in the Installation Instructions.
1.6 Service
The system must not be serviced or maintained while in use with a patient, with the exception to fill a
vaporizer.
ESD sensitive components. When working with ESD sensitive components, always use a
grounded wrist band and a grounded work surface. Adequate service tools must always be
used.
ESD sensitive components. PC boards (spare parts) must always be kept in a package for
sensitive electronic devices. Getinge will not otherwise assume responsibility for the materials
used, the work performed or any possible consequences of same.
Special waste. This product contains electronic and electrical components. Discard disposable,
replaced and left-over parts in accordance with appropriate industrial and environmental
standards.
Trained and authorized personnel. The system must be serviced at regular intervals by
personnel trained and authorized by Getinge. It is recommended that service and maintenance
is done as a part of a service contract. Any maintenance or service must be noted in a log book.
Note: All repair or service requests resulting from a device failure to perform their intended function shall be
submitted as complaints. Please report a potential complaint to your local Getinge representative as soon as
possible for further processing.
Maintenance of the system must be performed by personnel trained and authorized by Getinge with the
following intervals:
Preventive maintenance (PM): At least once a year, or every 5000 hours of operation, whichever comes
first.
Preventive maintenance + (PM+): At least every second year.
The power backup battery shall be replaced every three or four years, depending on model.
The memory backup batteries on PC 1922 panel board, PC 2060/PC 1920 control board, and
PC 2061/PC 1921 monitoring board shall be replaced every four years.
Page 16 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 1 – Important
Only original spare parts from Getinge must be used in the system.
The contents of this document are not binding. If any significant difference is found between the system and
this document, please contact Getinge for further information.
We reserve the right to modify the system without amending this document or advising the user.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 17 (of 369)
Chapter 1 – Important
Page 18 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 1 – Important
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 19 (of 369)
Chapter 1 – Important
3 Vacuum- and serial tubing used with the 1 After each patient.
suction module. Applicable only if used
during treatment.
4 Gas sampling line (single use). 1 After each patient if no filter is used at the
Y-piece.
5 Exp. connection bacterial filter. 1 After each patient if no filter is used at the
Y-piece.
1
Must also be replaced on any extra patient cassettes.
Page 20 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 1 – Important
Operating mode and system Power consumption Total power Total power
configuration added consumption for a consumption for a
230 V / 50 Hz 110 V / 60 Hz
system system
2
Must also be replaced on any extra patient cassettes.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 21 (of 369)
Chapter 1 – Important
Page 22 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 2 – Introduction
2 Introduction
2.1 General
The System is intended for use in administering anesthesia while controlling the entire ventilation of patients
with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
The system is intended for use by healthcare professionals, trained in the administration of anesthesia.
The system is intended for use on neonate to adult patient populations.
The system is intended for use in hospital environments, except MRI and radiation environments.
When not in operation, the system is designed for in-hospital transport.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 23 (of 369)
Chapter 2 – Introduction
Figure 2-1: Front side of the system. Figure shows Flow-i system with S/N > 20 000.
Table 2-1: Description of callouts in Figure 2-1.
1 Control panel
See section 3.22 on page 107.
3 EVAC
See section 3.21 on page 105.
4 MSync (option)
See section 3.37 on page 143.
Page 24 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 2 – Introduction
5 Volume reflector
See section 3.16 on page 89.
8 Vaporizer
See section 3.10 on page 63.
9 AFGO (option)
See section 3.32 on page 129.
11 CO2 absorber
See section 3.15 on page 87.
13 Patient cassette
See section 3.12 on page 72.
15 Safety valve
See section 3.11 on page 70.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 25 (of 369)
Chapter 2 – Introduction
Page 26 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 2 – Introduction
26 Power section
See section 3.23 on page 110.
29 Power connection
See section 3.24 on page 115.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 27 (of 369)
Chapter 2 – Introduction
The communication link between the computer nodes above is CAN bus.
The system software is installed with a Getinge approved USB memory stick. Separation of the software
subsystems is handled by the installation program, refer to section 5.5.2, System software installation
procedure on page 249.
Page 28 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 2 – Introduction
2.4.3.2 Control
The main responsibility for the Control subsystem is patient treatment, how to regulate the inspiratory and
expiratory gas flow.
Breathing: Responsible for all modes of ventilation and delivery of anesthetic agents and gas mixtures.
Ventilator settings: Responsible for all ventilator settings.
Persistent storage: Maintain a persistent data storage area used to save parameter settings.
Metrics: Contains several metrics for various types of information.
System settings: Responsible for some of the system settings, e.g., storing default patient category and
default mode of ventilation. The system settings will be stored in persistent memory.
2.4.3.3 Monitoring
The main responsibilities for the Monitoring subsystem are:
alarm
monitoring functions.
Alarms: Responsible for detecting and handling alarm situations.
Metrics: Responsible for monitoring the system behavior.
Persistent storage: Maintain a persistent data storage area to save Alarm settings.
Internal network supervision: CAN bus-master responsible for connection of all nodes in the system and
supervision of the CAN network.
System settings: Responsible for some of the system settings, e.g., default alarm limits.
Supervision of anesthesia delivery: Responsible for supervision of the anesthetic agent concentration.
Optional functions: Responsible for optional functionality in the system. An option is marked in the
backplane EEPROM and can enable functions such as a ventilation mode or other features.
2.4.3.4 Panel
The main responsibility of the Panel subsystem is to enable the user to interact with the system.
Presentation and modification of ventilator settings: Responsible for implementing mechanisms to let
the users modify the ventilator and anesthesia settings, as well as displaying the ventilator and anesthesia
settings to the user.
Presentation of real-time metrics: Responsible for presentation of real-time metrics generated from
measurements in the Monitoring subsystem.
Presentation of alarms: Responsible for presentation of alarms generated by the Monitoring subsystem.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 29 (of 369)
Chapter 2 – Introduction
Drawing primitives and fonts: Will use its own graphics library that includes functions such as draw line,
draw text and draw circle.
Persistent storage: Maintain a persistent data storage area to save Log data.
Event logging: Responsible to maintain an Event log. Some of the information in the log will be generated
by the other subsystems. Date and time will always be stored together with the events.
Typical events saved in the log are:
Alarm activations
Changes in ventilator settings
Alarm limit changes
Activation of System checkout
Apnea periods
Activation of immediate functions, e.g., O2 flush
System diagnostics logging: Responsible for maintaining a system Service log. The Service log is divided
into five different logs and generated from all subsystems. Date and time will always be stored together with
the events as well as some identifier of the subsystem generating the event.
Typical events saved in the logs are:
1. Test results:
o System checkout results
o Individual test results
o Calibration results
2. Technical alarms:
o Details about technical alarms.
3. Configuration:
o Details about the systems startup configuration.
4. Installation:
o Installation version history.
5. Service report:
o Service notes created by the service technician.
o Calibration results.
FCI protocol: Responsible for the external communication protocol FCI (Flow Communication Interface).
Page 30 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
3 Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 31 (of 369)
Chapter 3 – Description of functions
OFF STANDBY
Page 32 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
PV7 Additional fresh gas outlet valve Patient cassette Patient cassette
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 33 (of 369)
Chapter 3 – Description of functions
Version B
See section 3.5.2 on page 37.
Version A
3.5.1 Version C
Page 34 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 35 (of 369)
Chapter 3 – Description of functions
The O-rings on the gas module docking must be replaced during Preventive maintenance + (PM+).
Page 36 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Figure 3-5: Schematic view of the Gas distribution section, Version A (left) and Version B (right).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 37 (of 369)
Chapter 3 – Description of functions
4 EVAC connector
5 N2O regulator
The gas delivered to the Reflector gas module can be set to O2 or Air. This is controlled by the Reflector gas
selector (ref.no 3 in Figure 3-6 on page 37). The default setting is O2 (selector in its upper position) and O2
must be used as reflector gas.
Note: The Reflector gas selector setting must not be changed.
Page 38 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Version B
See section 3.6.2 on page 41
Version A
3.6.1 Version C
Figure 3-7: Schematic view of the Version C drive gas supply section.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 39 (of 369)
Chapter 3 – Description of functions
Figure 3-8: Three-dimensional view of the Version C Drive gas supply section.
Table 3-4: Description of call-outs in Figure 3-8.
5 Drive gas to O2 flush connector 10 OV7 (O2 drive gas one-way valve)
Page 40 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Figure 3-9: Schematic view of the Version A drive gas supply section.
Figure 3-10: Schematic view of the Version B drive gas supply section.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 41 (of 369)
Chapter 3 – Description of functions
Figure 3-11: Three-dimensional view of the Version A (left) and Version B (right) Drive gas supply sections.
Table 3-5: Description of call-outs in Figure 3-11.
2 REG2 (pilot valve pressure regulator) 7 T2/T3 (pressure test points number 2
and 3)
3 SV3 (drive gas safety valve) 8 OV7 (drive gas one-way valve O2)
5 REG4 (instant O2 supply pressure 10 OV8 (drive gas one-way valve Air)
regulator)
Page 42 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
REG2 and REG3 are connected to the same output channel. With this design, the gas with the highest
output pressure will automatically be used as Pilot valve drive gas and drive gas will still be supplied if one of
the gases is lost. If the output pressure from REG2 and REG3 is the same, a mix of Air and O2 may be used
as Pilot valve drive gas.
The output from REG2/REG3 is also used to supply REG1.
The output flow from REG2/REG3 at PV1 is regulated to max. 10 l/min by a flow restrictor mounted at the
PV1 connector. This is a safety feature in case of a major O2 leakage e.g. disconnected or broken tubing.
The preset output pressure is 200 kPa ±10 kPa. Use Pressure test point 2/3 (T2/T3) to check or adjust the
pressure.
REG2/REG3 must be checked during Preventive maintenance + (PM+).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 43 (of 369)
Chapter 3 – Description of functions
Page 44 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 45 (of 369)
Chapter 3 – Description of functions
Page 46 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 47 (of 369)
Chapter 3 – Description of functions
MAN
The Manual ventilation valve opens and the manual ventilation bag is enabled. During the MAN mode the
APL potentiometer is connected and enables APL regulation by the APL/PEEP valve.
AUTO
The Manual ventilation valve closes and the manual ventilation bag is disabled. The APL potentiometer is
disabled and the APL/PEEP valve is controlled by the PEEP setting on the control panel.
Page 48 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
1 EMV 2
2 EMV 3
3 EMV 4
4 EMV 5
5 EMV 6
6 EMV 7
7 EMV 8
3
See section 3.7.5.1, Pressure supply pilot valves, on page 52 for detailed description.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 49 (of 369)
Chapter 3 – Description of functions
Figure 3-17: The MAN/AUTO ventilation switch. Version A (left) and Version B (right)
Page 50 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Note: The Vaporizer 2 lock pilot valve is deactivated for systems that are upgraded to software version 4.5.1
or higher.
EMV2
Manual ventilation pilot valve EMV2 is connected to the Manual ventilation valve actuator PP3.
When the Manual ventilation switch is set to AUTO, pilot valve EMV2 opens and supplies pressure to PP3.
This will close the Manual ventilation valve inside the Patient cassette and disable the Manual ventilation
bag.
When the Manual ventilation switch is set to MAN, pilot valve EMV2 closes. The valve actuator PP3 is
depressurized and the Manual ventilation valve opens to enable the Manual ventilation bag.
EMV3
System checkout pilot valve EMV3 is connected to the System checkout valve PV6 through gas channel
PP4.
At System checkout, the Y-piece is connected to the System checkout valve PV6. During most of the
subtests the valve is closed, i.e., the Y-piece is plugged. During parts of the Gas analyzer and Vaporizer
tests the Gas analyzer is used to measure concentrations of either O 2 or anesthetic agent in the fresh gas
flow. In order to enable quicker changes of gas concentrations, thus shortening the System checkout time,
the pilot valve EMV3 opens, supplies pressure and opens PV6. This will open a bypass channel from the
Inspiratory outlet to EVAC.
EMV4
Vaporizer bypass pilot valve EMV4 is connected to the Vaporizer bypass valve PV5. If no vaporizer is
connected or a connected vaporizer is switched off, pilot valve EMV4 is open, supplies pressure and opens
PV5, allowing fresh gas to pass through the bypass channel inside the vaporizer valve section. No
anesthetic gas is added to the fresh gas. If a vaporizer is connected and activated/switched on, EMV4 is
closed and PV5 disables the bypass channel. The fresh gas flows to the vaporizer, adding anesthetic gas
before the gas mix enters the Patient cassette.
EMV5
Vaporizer 2 selection pilot valve EMV5 is connected to Vaporizer inlet valve PV3 and Vaporizer outlet valve
PV4. When anesthetic agent is present in the vaporizer, EMV5 opens and supplies pressure to PV3 and PV4
to open, and fresh gas passes through the vaporizer.
EMV6
Vaporizer 1 selection pilot valve EMV6 is connected to Vaporizer inlet valve PV1 and Vaporizer outlet valve
PV2. When anesthetic agent is present in the vaporizer, EMV6 opens and supplies pressure to PV1 and PV2
to open, and fresh gas passes through the vaporizer.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 51 (of 369)
Chapter 3 – Description of functions
EMV18
AFGO pilot valve EMV18 controls the AFGO valve.
Pneumatic valves
Vaporizer 1 inlet and outlet valves (PV1 and PV2) distribute fresh gas via Vaporizer 1.
Vaporizer 2 inlet and outlet valves (PV3 and PV4) distribute fresh gas via Vaporizer 2.
Vaporizer bypass valve (PV5), distributes fresh gas when the vaporizers are disabled.
The AFGO valve PV7 controls the supply of fresh gas from the vaporizer, either to the patient cassette or to
the AFGO connector. With AFGO enabled, fresh gas will be supplied to the AFGO connector.
Page 52 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
the ambient pressure. The output signal is proportional to the measured pressure giving a linear
measurement in the range:
–40 cmH2O to +140 cmH2O (inaccuracy ±5% or ±1 cmH2O).
Information from the ID/Status PROM is read by the system.
A calibration of the APL potentiometer is required e.g., after replacement. Refer to chapters Disassembling
and assembling and also Service Procedures.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 53 (of 369)
Chapter 3 – Description of functions
Page 54 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 55 (of 369)
Chapter 3 – Description of functions
Page 56 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
LED
1 Gas inlet
3 Temperature sensor
6 Nozzle unit with valve diaphragm. Replaced during Preventive maintenance + (PM+).
7 Solenoid
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 57 (of 369)
Chapter 3 – Description of functions
are provided with matching guide pins. The pins prevent incorrect mounting of a filter cover on a fresh gas
module.
3.8.1.2 Filter
The filter protects the system from particles in the gas delivered to the fresh gas modules. The filter must be
replaced during Preventive maintenance + (PM+).
3.8.1.7 Solenoid
The solenoid regulates the gas flow through the gas module. The current supplied to the solenoid is
regulated so that the gas module delivers a gas flow according to settings on the control panel.
Page 58 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 59 (of 369)
Chapter 3 – Description of functions
Page 60 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 61 (of 369)
Chapter 3 – Description of functions
As long as the vaporizers are switched Off, EMV4 will be open and supply pressure to PV5. This will keep
PV5 open and allow fresh gas to pass through the bypass channel inside the Double channel plate. When an
anesthetic agent is set and accepted on a vaporizer, EMV4 will close and thus close PV5.
1 Vaporizer
4 PP1/PP2
The Vaporizer locking device will lock the activated vaporizer in its position.
When an anesthetic agent is set and accepted on:
Vaporizer 1: EMV7 will open and supply pressure to Vaporizer 1 locking device PP1. This will push out the
pneumatic piston in PP1 and lock the Vaporizer 1 release handle.
Page 62 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Vaporizer 2: EMV8 will open and supply pressure to Vaporizer 2 locking device PP2. This will push out the
pneumatic piston in PP2 and lock the Vaporizer 2 release handle.
Note: PP1 and PP2 are disabled for software version 4.5.1 or higher.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 63 (of 369)
Chapter 3 – Description of functions
Sintered metal filters protect the system from particles entering into the vaporizer. An optical Liquid level
sensor measures the liquid level in the Liquid indicator glass tube. The agent is filled into the container
through a key filling valve. The mechanism is protected by a lid with a Lid sensor.
Handle the vaporizer with care.
Never turn a vaporizer containing anesthetic agent upside down or lay it sideways, as the agent can leak
out.
The vaporizer is factory calibrated and leakage tested.
The vaporizer must not be disassembled.
In case of malfunction, repair or service must be carried out by the manufacturer only.
Perform an RMA procedure when returning a vaporizer to the manufacturer.
If the vaporizer is dropped, it must be serviced by the manufacturer to ensure proper functionality.
The vaporizer must be emptied before it is returned to the manufacturer. Refer to emptying instructions
in the User’s Manual.
Note that there may be small amounts of anesthetic agent in the vaporizer even when the indicator
shows that the vaporizer is empty.
The same type of vaporizer is used in all systems.
Page 64 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 65 (of 369)
Chapter 3 – Description of functions
5 Vaporizer injector
7 Vaporizing chamber
Page 66 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 67 (of 369)
Chapter 3 – Description of functions
Page 68 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 69 (of 369)
Chapter 3 – Description of functions
Six sending IR diodes used to monitor the anesthetic agent level in the level meter tube. The diodes
position corresponds to 5%, 10%, 25%, 50%, 75% and 100% contents in the container according to the
level meter tube. Receiving transistors located on PC 1913 vaporizer liquid level detector board. The
agent level can be shown on the Control Panel.
Page 70 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
4 Connector to EVAC
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 71 (of 369)
Chapter 3 – Description of functions
condition will also open the APL/PEEP valve and is an extra safety function and the situation will
normally not occur. Applicable in AUTO and MAN mode.
Page 72 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
2 Silicone gasket towards sample filters for Insp. gas sampling (S1). Currently not used.
3 Silicone gasket towards sample filters for Insp. gas sampling return (R1). Currently not used.
4 Silicone gasket towards sample filters for Insp. pressure sampling (Pi).
5 Silicone gasket towards sample filters for Exp. pressure sampling (Pe).
6 Silicone gasket towards sample filter for Exp. gas sampling return (R3)
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 73 (of 369)
Chapter 3 – Description of functions
Page 74 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
defective, it must be replaced. Remaining capacity (in %) for the membrane can be shown in the Service &
Setting’s Status window. The remaining capacity meter must be reset after replacement of the membrane.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 75 (of 369)
Chapter 3 – Description of functions
Figure 3-33: Three-dimensional view of the patient cassette lid. Version A (left) and Version B (right).
The patient cassette lid is mounted with hinges on the system. The lid must be opened to remove the patient
cassette. When the lid is opened the gas sampling and electrical connection to the patient cassette are
disconnected.
Page 76 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
For System versions 3.0 (or above) the Control gas analyzer is disabled and the Inspiratory sample filter
holder is used for inspiratory pressure only.
Systems that are field upgraded to System version 3.0 can use either the Version A filter holder with
three filters or the Version A filter holder with S1 and R1 plugged.
Systems that are delivered from factory with System version 3.0 (or above) will not include the Control
gas analyzer and must use the Version A filter holder with S1 and R1 plugged.
Systems using Patient cassette lid Version B must use the Version B filter holder.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 77 (of 369)
Chapter 3 – Description of functions
Versions Comments
Patient gas analyzer AION ADS with external patient O2
PGA Version A sensor.
Patient gas analyzer AION Platinum with integrated patient O2
PGA Version B sensor. Requires SW system version 4.04.01 or above. Was
introduced during Q4 2017.
Figure 3-34: Schematic view of the gas analyzer section with the PGA AION. Version B (left) and Version A (right).
Page 78 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Figure 3-35: Functional view of the Gas analyzer section with PGA AION Platinum (Version B).
Note: The Patient gas analyzer AION is a complete unit and must not be disassembled.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 79 (of 369)
Chapter 3 – Description of functions
Figure 3-37: Functional view of the Gas analyzer section with PGA AION ADS (Version A).
Page 80 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Figure 3-38: Exploded view of the PGA AION ADS (Version A).
Note: The Patient gas analyzer AION is a complete unit and must not be disassembled.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 81 (of 369)
Chapter 3 – Description of functions
Zero calibration valve (ZCV): To establish a gas zero concentration reference, ambient Air is routed
through the Gas measurement bench via the Zero calibration valve. The ambient gas intake is located
behind the Vaporizer locking handle. A particle filter is connected to the tubing.
Purge flow valve (AION ADS only): During normal sampling condition, when a water trap is used, there is
always a small purge of approximately 10% of the total flow through the AION. This flow is created by the
purge flow restrictor.
The automatic zero calibration of the AION will be performed:
During warm-up after cold start (power on).
At a temperature change >1 °C (in the GMB) during ISO/Full accuracy mode.
Every 4 hours during Full accuracy mode.
At gas span calibration.
When a new agent is identified.
If there is an obstruction in the sampling line, this will be detected by the flow sensor in the pump as a
decreased flow.
If the total flow from the two water trap outlets drops below 40 ml/min for more than one second, the forced
purge flow-function is activated during 12 s. The forced purge flow forces water and secretions in the
sampling line into the water trap container.
If the total flow remains below the limit value, the forced purge flow-function will be repeated up to four times
until further forced purge cycles are inhibited and an occlusion alarm is activated. The purge function is
automatic and is not possible to start manually.
Figure 3-39: Functional and three-dimensional view of the gas measurement bench.
The GMB consists of an IR light source, a rotating filter wheel with optical filters connected to a DC motor,
the gas measurement chamber and an IR light detector.
The measurement of CO2, N2O and the anesthetic agents in the sampling gas is based on the fact that the
different gas components absorb infrared light at specific wavelengths. The filter wheel has eight optical
filters to allow an accurate analysis of any mixture of these gases.
As O2 does not absorb infrared light to the same extent as other breathing gases, O 2 is measured by the
separate O2 sensor utilizing the paramagnetic properties of the O2 molecule. However, the presence of O2 in
the gas mixture causes some interference to other gas measurements. Therefore, information from the O2
sensor is used to compensate for that interference.
Page 82 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The GMB also includes a PC board with an EEPROM and components for pressure and temperature
measurements. Factory calibration data are stored in the EEPROM and the gas calibration data is updated
in the EEPROM during the gas span calibration.
Figure 3-40: Three-dimensional view of the O2 sensor. Servomex PM111E external O2 sensor used in PGA AION ADS
(left) and Hummingbird Paracube Sprint integrated O2 sensor used in PGA AION Platinum (right).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 83 (of 369)
Chapter 3 – Description of functions
A Suspension balanced
1 Current measurement
2 Amplifier
3 Photo diodes
4 Light source
5 Mirror
O2 is distinguished from most other common gases by its paramagnetic properties. This fact is used by the
paramagnetic O2 sensor.
Page 84 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The sensors incorporate two nitrogen-filled glass spheres mounted on a strong, noble metal taut-band
“suspension assembly”. This assembly is suspended in a symmetrical non-uniform magnetic field. When the
surrounding gas contains paramagnetic oxygen, the glass spheres are pushed further away from the
strongest part of the magnetic field. The strength of the torque acting on the suspension is proportional to the
oxygen content of the surrounding gases.
The measuring system is “null-balanced”. The zero position of the suspension assembly, as measured in
nitrogen, is sensed by a differential photo-sensor assembly that receives light reflected from a mirror
attached to the suspension assembly. The output from the photo-sensor is processed and then fed back to a
coil wound around the suspension assembly to achieve a null-balanced position. When oxygen is introduced
to the sensor, the torque acting upon the suspension assembly is balanced by a restoring torque due to the
feedback current in the coil.
The feedback current is directly proportional to the volume magnetic susceptibility of the sample gas and
hence, after calibration, to the partial pressure of oxygen in the sample. A voltage output is derived which is
proportional to the feedback current. In addition, the electromagnetic feedback stabilizes the suspension
(heavily damping oscillations).
The paramagnetic technology is non-depleting, which means that there are no consumable parts. This
ensures consistent performance over time. The selectivity of the paramagnetic measurement for oxygen
means that there is no interference from other respiratory gases. The small volume chamber allows a rapid
gas exchange, giving the capability for fast response oxygen measurement.
1 Purge inlet
2 Sample inlet
3 Upper switch
4 Lower switch
5 Nafion tube
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 85 (of 369)
Chapter 3 – Description of functions
6 Purge tube
8 PC board
The Water trap is connected to the Water trap receptacle. There are two gas connections on the receptacle;
one for the sampling gas flow and one for the purge gas flow.
The gas sample is routed via a bacteria filter and a Nafion™ tube. The Nafion tube equalizes the moisture
content and the temperature in the sampled gas to that of the ambient Air. This is to reduce the influence on
the measurements from humidity and to prevent from water condensation inside the sampling system.
Two electrical switches on the water trap receptacle detect if a water trap is present. If no water trap is
connected or if a water trap without container is connected, both switches are inactive. The pump, and thus
the gas flow through the Patient gas analyzer, is automatically stopped to prevent contamination.
The control signal cable from the water trap receptacle PC board is connected to the Main backplane.
The Nafion tube must be replaced during Preventive maintenance + (PM+).
The DRYLINE™ water trap protects the gas analyzer system from condensed water, secretions, bacterial
contamination and dust. It consists of two parts, a filter housing with hydrophobic anti-bacterial filter and a
container for separated waste.
The filter housing has one gas sample inlet (W1) to be connected to the sampling line.
Page 86 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 87 (of 369)
Chapter 3 – Description of functions
Figure 3-45: Three-dimensional view of the CO2 absorber. Version A (left) and Version B (right).
1 CO2 absorber
WARNING! Absorbent material may be very corrosive and must be handled with care. Avoid
coming into contact with dust or particles.
Page 88 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The volume reflector stores the expired gas from the patient allowing for partial re-administration of the gas
to the patient through the circle system.
CAUTION: As the Volume reflector and the Reflector socket are exposed to the patients expired
gas it must be included in the routine cleaning procedure, refer to the User’s Manual.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 89 (of 369)
Chapter 3 – Description of functions
The purpose of the Volume reflector is to store the patient’s exhaled gas so that the gas can be purified in
the CO2 absorber for partial re-use during next inspiration phase.
The inside of the reflector is designed as a long gas channel with a volume of 1.2 liter (1.3 liter including
Reflector adaptor).
During the expiration phase, exhaled gas from the patient will be routed into the reflector gas channel.
During the inspiration phase, reflector gas from the Reflector gas module will be routed into the other end of
the reflector gas channel. The reflector gas will, like a piston, push the expiratory gas out from the reflector
and through the CO2 absorber. Fresh gas according the Control Panel setting will then be mixed with the
purified gas to form the inspiratory gas to the patient.
Page 90 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 91 (of 369)
Chapter 3 – Description of functions
Page 92 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The gas delivered to the Reflector gas module can be set to O2 or Air. This is controlled by the Reflector gas
selector. The default setting is O2 (selector in its upper position) and O2 must be used as reflector gas.
Note 1: The Reflector gas selector setting must not be changed.
Note 2: The Reflector gas selector is removed on systems with S/N 4001 and above.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 93 (of 369)
Chapter 3 – Description of functions
1 Gas inlet
3 Temperature sensor
6 Nozzle unit with valve diaphragm. Replaced during Preventive maintenance + (PM+).
7 Solenoid
Page 94 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
3.17.3 Filter
The filter protects the system from particles in the gas delivered to the fresh gas modules. The filter must be
replaced during Preventive maintenance + (PM+).
3.17.8 Solenoid
The solenoid regulates the gas flow through the fresh gas module. The current supplied to the solenoid is
regulated so that the gas module delivers a gas flow according to settings on the control panel.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 95 (of 369)
Chapter 3 – Description of functions
Page 96 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Version C Same as Version B, but the tubing from OV13 goes directly to
the Gas control section. See section 3.19.1 on page 98.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 97 (of 369)
Chapter 3 – Description of functions
Figure 3-54: Schematic view of the emergency ventilation section, Version C (left) and Version B (right)
3 Mechanical APL
4
Same as Version B, but the tubing from OV13 goes directly to the Gas control section.
Page 98 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 99 (of 369)
Chapter 3 – Description of functions
Emergency APL actuator. Thus, turning the Mechanical APL knob, the rotation will be transmitted from the
knob to the cylinder on the Emergency APL actuator.
The cylinder on the Emergency APL actuator is connected to the movable APL/PEEP valve coil axis. The
axis will be pushed up towards the APL/PEEP valve diaphragm when rotating the cylinder (the rotation
created when turning the Mechanical APL knob clockwise) to set the APL value.
The cylinder is spring-loaded to assure that it returns towards neutral position when turning the knob
counterclockwise.
Page 100 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
3.19.2 Version A
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 101 (of 369)
Chapter 3 – Description of functions
2 Oxygen flow
Page 102 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 103 (of 369)
Chapter 3 – Description of functions
3 EVAC connector
The System checkout valve PV6 is a spring-loaded pneumatic valve. The valve can be either:
Closed, controlled by the internal spring.
Open, controlled by the System checkout pilot valve EMV3. The 240 kPa (S/N < 20 000: 200 kPa) drive
gas from EMV3 is connected to the drive gas connector and the System checkout actuator PP4 will open
the System checkout valve PV6.
During System checkout (SCO), the patient tubing’s Y-piece must be connected to the System checkout
valve.
When the valve is closed, the connector will act as a plug. With a plugged Y-piece, the SCO leakage
checks can be performed.
Page 104 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
When the valve is open, a fresh gas flow through the system will be allowed. The fresh gas from the Y-
piece will pass through the System checkout valve and be routed to the EVAC reservoir. This is used,
e.g., during gas analyzer checks.
The valve is only used during System checkout. During normal operation, the valve is kept closed by its
spring.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 105 (of 369)
Chapter 3 – Description of functions
1 EVAC reservoir
Page 106 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
CAUTION: Ensure that the EVAC system is correctly connected after any service intervention.
Running the system with a poor connection may result in anesthetic agent being emitted into the
operation environment. Leakages in the EVAC system will not be detected by the System
checkout.
Version B. S/N 300576 – 320000. Control Panel including Display with LED.
Version A. S/N < 300576. Control Panel including Display with lamps.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 107 (of 369)
Chapter 3 – Description of functions
Note: The Display/Touch screen for versions A and B is obsolete and is not available as spare part. If
replacement is required, then the LED Display kit (corresponding to Version C) must be installed. This kit
contains all parts required to mount an LED Display in an older version of the Control Panel.
1 Display/Touch screen
2 Membrane buttons
3 Rotary encoder
Page 108 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Version C
Version C requires PC 1924 of revision 08 (or above) and System software version 3.00.04 (or above).
Version C
The LEDs inside the Display/Touch screen are driven from circuits integrated in PC 1924. It is estimated that
the Version C Control Panel will maintain acceptable brightness during the unit’s lifetime. Use Service &
Settings to show a backlight time meter. This time meter is intended primarily for the Version A, but can be
used also on Version C. The time meter must be reset after replacement of the LEDs Display/Touch screen.
Version B
The LEDs inside the Display/Touch screen are driven from PC 1963 Backlight inverter. It is estimated that
the Version B Control Panel will maintain acceptable brightness during the unit’s lifetime. Use Service &
Settings to show a backlight time meter. This time meter is intended primarily for the Version A, but can be
used also on Version B. The time meter must be reset after replacement of the LEDs Display/Touch screen.
Version A
The Backlight lamps inside the Display/Touch screen are driven from the Backlight inverter. Estimated
lifetime (with acceptable brightness level) for the lamps is 30.000 hours. Use Service & Settings to show a
backlight time meter. This time meter must be reset after replacement of the lamps.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 109 (of 369)
Chapter 3 – Description of functions
Version C
The backlight inverter function is integrated in PC 1924 display connection board. A separate PC board is
not required.
Version B
PC board with driving stage for backlight (LEDs) mounted inside the Display/Touch screen. The supply
voltage delivered by PC 1963 Backlight inverter is 12 V.
Version A
PC board with driving stage for backlight (lamps) mounted inside the Display/Touch screen. The supply
voltage delivered by the Backlight inverter is 660 V.
WARNING! The PC board generates 660 V. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with power supply connected and with covers removed.
3.22.1.6 Loudspeaker
The Loudspeaker generates sound, e.g. alarm. The loudspeaker generates different tones with individual
sound volumes. At start-up and during System checkout the function of the loudspeaker is monitored by the
microphone on PC 1924 display connection board. During operation, it is continuously monitored through
current sensing.
Page 110 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 111 (of 369)
Chapter 3 – Description of functions
1 Battery connector
6 Lamp connector 12 V
8 AC/DC Converter
Page 112 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Mains power is supplied to the AC/DC converter via the AC/DC power inlet, which is equipped with two 6.3 A
fuses.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 113 (of 369)
Chapter 3 – Description of functions
o On PC 1903 revision 04 or lower, the fuses can be replaced. Always replace both fuses at the same
time. It is however recommended to replace the complete PC 1903.
o On PC 1903 revision 05 or higher, the fuses are fixed and cannot be replaced. In case of a blown
fuse, PC 1903 must be replaced.
Note: Before replacing PC 1903 external connectors board (or fuses), the root cause to the blown fuse
must be remedied.
Battery sense. Battery sense signal cables are used for measuring the battery voltage and the
temperature of the battery/battery compartment. The Battery sense cable connected to the negative
battery pole includes a thermistor.
Cylinder pressure. Connects the pressure transducers in the Backup gas rack in order to display the
cylinder gas pressure on the Control Panel.
LED lamp. Supplies power to the LED lamp mounted inside the Control Panel swivel arm.
o PC 1903 revision 06 (or below) is equipped with one connector (P40) to supply 12 V to the old
version of the LED lamp and its LED driver, see section 3.23.7 on page 115.
o PC 1903 revision 07 (or above) is equipped with two connectors to supply 12 V (P40) to the old
version of the LED lamp or 24 V (P43) to the new version of the LED lamp.
The LED driver/dimmer functions are now integrated on PC 1903 and an external LED driver is not
required.
Includes an ID PROM. The ID information can be read by the system.
3.23.5 Fan
The Fan is located behind the fan filter below the CO2 absorber. The fan forces cooling Air into the unit.
A tacho sensor integrated into the fan provides the fan rotation signal. The system can be switched On only
if fan rotation is detected. Refer to section 3.23.2, PC 2063/PC 1923 on page 113 for further information.
The fan speed is regulated by temperature sensors inside the unit:
Low fan speed. The fan runs with low speed if the temperature at the NTC resistor on PC 2057/PC
1907 valve drivers board is below 40 °C and the temperature at the Power control CPU is below 50 °C.
High fan speed. The fan runs with high speed if the temperature at the NTC resistor on PC 2057/PC
1907 valve drivers board is above 40 °C or the temperature at the Power control CPU is above 50 °C.
The fan also runs with high speed during the first 20 s after mains voltage is detected, or Power On is
made when no mains power supply cable is connected.
The fan filter must be replaced during Preventive maintenance + (PM+).
Page 114 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
5 The lifetime of the battery may be reduced if it is used frequently to supply power or if the battery
temperature is above 40˚C.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 115 (of 369)
Chapter 3 – Description of functions
Figure 3-68: Schematic view (left) and three-dimensional view (right) of Power connection with PC 1938
Table 3-35: Description of callouts in Figure 3-68.
Page 116 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 117 (of 369)
Chapter 3 – Description of functions
Figure 3-70: Three-dimensional view of the Power On/Off and Vaporizer PC board section
Table 3-36: Description of callouts in Figure 3-70.
Page 118 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
O2 flush. Detects if the O2 flush valve is open. For redundancy, a 2-pole switch is integrated in the O2
flush button. Both poles must be closed to detect the O 2 flush valve as open.
Vaporizer status LED lamps. Show vaporizer status with the Vaporizer LEDs, visible just above the
vaporizer slots. The LEDs are mounted on this board. There are two sets of LEDs, one for each
vaporizer slot. Each set of LED contains three LEDs; green, yellow and red.
The LEDs indicate:
o Green: The selected vaporizer has passed the System checkout or Vaporizer check, and is ready to
deliver agent.
o Yellow:
- The vaporizer slot is selected but no vaporizer is present.
- The selected vaporizer slot holds an unchecked vaporizer.
- The selected slot's vaporizer is being filled.
- The selected slot's vaporizer has triggered an alarm, i.e. vaporizer nearly empty.
o Red:
- The vaporizer is disabled.
- The power to the vaporizer is shut off (the condition is applicable to both slots, i.e. not only the
selected slot).
o Not lit/Off: The vaporizer slot is not selected.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 119 (of 369)
Chapter 3 – Description of functions
cassette handling. Expiratory flow measurement is performed by the Expiratory flow transducer in the
cassette.
The PC 1984/PC 1919 Expiratory channel board is connected to the Patient cassette via PC 1918 Expiratory
cannel connector.
The PC 1984/PC 1919 Expiratory channel board contains electronics including a microprocessor for
handling of:
Expiratory flow measurement performed by the Expiratory flow transducer inside the cassette. The
output signal Exp. Flow is an analogue signal used in sub-systems Control and Monitoring.
All electronic connections to and from the Patient cassette.
Measurement of Fresh gas pressure and Inspiratory pressure via PC 1781 (F) and PC 1781 (I). PC
1984/PC 1919 contains holders and electrical connectors for these pressure transducer boards.
Control of the APL/PEEP valve and the Fresh gas safety valve functions.
A DC/DC converter on PC 1984/PC 1919 generates 200 VDC to the Expiratory flow transducer drivers. A
discharge circuitry makes sure that the capacitor that carries 200 V is discharged when the cassette is
disconnected.
WARNING! The PC board generates 200 V. All personnel must exercise extreme caution if
fault tracing or adjustments are performed with power supply connected and with covers
removed.
The CAN bus is responsible for all communication excluding the analog flow signal.
Page 120 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
A thermistor on PC 1984/PC 1919 expiratory channel board monitors the temperature. An alarm is activated
if the temperature is 77 ±5 °C or higher or if the temperature sensor is not connected (e.g., due to poor
connection).
Includes an ID PROM. The ID information can be read by the system.
Green and red LEDs indicate activities that are related to the PC 1984/PC 1919 expiratory channel board.
The LEDs are described in section 6.4, PC board LED indicators on page 321.
Note: The System software must be reinstalled if the PC 1984/PC 1919 expiratory channel board is
replaced.
Uses PC 2060
Uses PC 1920
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 121 (of 369)
Chapter 3 – Description of functions
The PC 2060/PC 1920 control board comprises electronics including microprocessor for handling of:
Inspiratory control and regulation (fresh gas modules, fresh gas safety valve, fresh gas pressure
transducer, insp. pressure transducer).
Anesthesia control and regulation.
Vaporizer control (bypass valve, chamber valves and pressure valves)
Expiratory control and regulation (expiratory flow transducer, APL/PEEP valve, water trap and expiratory
pressure transducer)
Volume reflector
Reflector gas control (reflector gas module and reflector pressure transducer).
Page 122 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Uses PC 2061
Uses PC 1921
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 123 (of 369)
Chapter 3 – Description of functions
The PC 2061/PC 1921 monitoring board comprises electronics including microprocessor for handling of:
Pressure failure alarm PFA
Barometric pressure
Backup buzzer
Real-Time clock RTC
Temperature monitoring
System voltage monitoring
Includes an ID PROM. The ID information can be read by the system.
Green and red LEDs indicate activities related to PC 2061/PC 1921 monitoring board. The LEDs are
described in section 6.4, PC board LED indicators on page 321.
Information required by the system is stored on PC 2061/PC 1921 monitoring board. Refer to section 4.4.5,
Information stored on the PC boards on page 149.
A memory backup battery on PC 2061/PC 1921 monitoring board supplies power to the internal memory on
the PC board. The memory backup battery must be replaced after four years. If the battery is disconnected
or if the battery voltage is too low, information stored in the internal memory will be lost.
Note: If the PC 2061/PC 1921 monitoring board is replaced, then:
The System software must be reinstalled.
The barometer must be calibrated.
Page 124 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
4 VGA connector
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 125 (of 369)
Chapter 3 – Description of functions
5 USB connector.
The intended use of the USB connector is only for USB memory sticks. Restricted items include,
but are not limited to, external hard drives, USB hubs and any equipment drawing power from
the USB connector. Getinge recommends using a Getinge USB memory stick for this purpose.
7 Remote Services Adapter with RJ45 connector. Required for systems with PC 1922 of revision
10 or older that are provided with RJ12 connector.
PC 1922 panel board comprises electronics including microprocessor for handling of:
The graphical user interface on the Control panel.
Metrics and curves on the Control panel.
Settings and display of parameters on the Control panel.
Show trends of important metrics on the Control panel. Trends can be displayed either numerically or
graphically.
User interface for System checkout.
Authorization for service engineers using an USB Service key and present a user interface for Service &
Settings.
Alarm limit configurations and display alarm notifications.
Presentation of various event logs to the user.
User inputs from the touch screen, the knob and the membrane buttons.
Data exporting to a USB memory stick. Data that can be exported includes trends, logs, screen images
and settings. This data, except for screen images, can also be exported via the Ethernet connector using
Remote Services.
Includes an ID PROM. The ID information can be read by the system.
LED indicators on PC 1922 panel board are not visible (covered by the connector panel) and cannot be used
during troubleshooting.
The CompactFlash® card on PC 1922 panel board contains the language that is installed during System
software installation. The Vaporizer software is stored also on the CompactFlash card. This allows updating
the Vaporizer without use of USB memory stick.
The CompactFlash® should not be disconnected and moved to another PC 1922 panel board.
Information required by the system is stored on PC 1922 panel board. Refer to section 4.4.5, Information
stored on the PC boards on page 149.
A memory backup battery on PC 1922 panel board power supplies the internal memory on the PC board.
The memory backup battery must be replaced after four years. If the battery is disconnected or if the battery
voltage is too low, information stored in the memory will be lost.
Note: The System software must be reinstalled if PC 1922 panel board is replaced.
The external output/input ports are mounted on PC 1922 panel board, refer to the illustration below.
Cables listed below have passed EMC requirements specified in IEC 60601-1-2:
RS-232 cables (shielded, length 2.5 m)
Page 126 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Pin Description
1 –
2 RXD_ISO
3 TXD_ISO
4 DTR_ISO
5 GND_ISO
6 –
7 RTS_ISO
8 –
9 –
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 127 (of 369)
Chapter 3 – Description of functions
Page 128 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Version B
See section 3.32.1, Version B, on page 130.
Version A
See section 3.32.2, Version A on page 131.
Exception within Control gas analyzer’s sampling gas return point (R2), see
Version A section 3.32.2.1 on page 132.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 129 (of 369)
Chapter 3 – Description of functions
3.32.1 Version B
2 AFGO valve
Page 130 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
3.32.2 Version A
3 To AFGO connector
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 131 (of 369)
Chapter 3 – Description of functions
Control gas analyzer’s sampling gas return point (R2). Not used in System version 3.0 and above.
Plugged on units with S/N 2001 or above.
Page 132 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
Figure 3-86: Schematic (left) and three-dimensional (right) view of Extra mains power outlets, IEC
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 133 (of 369)
Chapter 3 – Description of functions
2 Power outlets
Figure 3-87: Schematic (left) and three-dimensional (right) view of Extra mains power outlets, US 5-15R.
Page 134 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The US 5-15R power outlets version (100–120 VAC systems) contains the following fuses:
Power outlets fuse (circuit breaker) rated 2 A for outlet 1.
Power outlets fuses (circuit breakers) rated 1 A for outlet 2 and 3.
Patient monitor fuse (circuit breaker) rated 2 A.
Isolation transformer fuses S1 and S2 rated 8 AF. Included in section 20. Power Connection.
Figure 3-88: Schematic view (left) and three-dimensional view (right) of Extra mains power outlets, CEE 7/4 (Type F)
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 135 (of 369)
Chapter 3 – Description of functions
1 Isolation transformer
2 Power IN
3 Power OUT
The isolation transformer is included in the (factory-mounted) option Extra mains power outlets. The isolation
transformer enables mains power connection for auxiliary equipment while maintaining the total system earth
leakage current below the limits given in IEC 60601-1. The transformer provides galvanic isolation from the
mains outlet in the wall.
The transformer supplies power to the three extra mains power and to the integrated patient monitor power
cable. The transformer is located below the power backup battery compartment.
The Isolation transformer is a complete unit that contains a transformer switching relay and a ring core
transformer. There are two different Isolation Transformers available, one for 100–120 V AC and one for
220–240 V AC.
The transformer switching relay is a current limitation that connects power to the transformer in a controlled
and safe way. This will prevent fuses to trip due to high start current in the transformer. Power supply to the
transformer is protected by fuses. There are different types of fuses used depending on the version of Extra
mains power outlet that is connected. Refer to section 3.33, Extra mains power outlets (Option) (29) on
page 133 for information regarding the fuses. The transformer is protected by a thermal fuse that will trip at
approximately 110 °C.
Page 136 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
On/Off switch and the suction unit regulating valve. The vacuum gauge shows the current suction
pressure.
Refer to the User’s Manual for information regarding:
Testing
Cautions
Cleaning
Technical specifications.
1 Air inlet
2 O2 inlet
4 Flowmeter tube
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 137 (of 369)
Chapter 3 – Description of functions
WARNING! Do not use lubricants that are not recommended by Getinge in high O2 exposed
areas. Risk of fire or explosion.
Page 138 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The pressure transducers are connected to PC 1916 Cylinder pressure connection inside the rack. A cable
connects PC 1916 with PC 1903 External connectors to supply cylinder pressure information to the system.
The backup gas supply should only be turned ON (valves open) when the backup gas supply is in use, or
during System Checkout.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 139 (of 369)
Chapter 3 – Description of functions
Figure 3-93: Three-dimensional view of the Backup gas rack (left) and the Backup gas trolley (right)
Page 140 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 141 (of 369)
Chapter 3 – Description of functions
Note: Gas cylinders must be secured, by means of a chain or wall mount, when not mounted on the system
or when kept in storage.
Page 142 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
The cylinder gas pressure is reduced to a preset pressure by the pressure regulators (REG23):
340 kPa on holders produced Q1 2016 or earlier.
310 kPa on holders produced from Q2 2016.
The gas is connected to the Gas distribution section via tubing.
The pressure regulator is preset in factory and cannot be adjusted during field service.
Trained and authorized personnel. MSync is a highly technically advanced product that
requires adequate training before order and installation.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 143 (of 369)
Chapter 3 – Description of functions
1 LAN connector
2 RS-232 connector
3 USB connector
4 Power supply
Page 144 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 3 – Description of functions
After connecting any external electric device to the system, perform an electrical safety test of the complete
system. For electrical safety test of the system, refer to 7.6.6, Safety inspection on page 353.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 145 (of 369)
4 Disassembling and assembling
4.1 General
In this chapter disassembling of the system is described. If not stated otherwise, the assembling procedure is
reverse of the disassembling procedure.
When disassembling and assembling, refer to the Spare parts list.
After any installation, maintenance or service intervention in the system, perform a System checkout and a
manual check of Emergency ventilation system according to instructions in the User’s Manual.
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit. All personnel must exercise extreme caution if fault tracing or
adjustments are performed with power supply connected and with covers removed.
CAUTION: With gas supply connected to the system, there are pressurized components inside
the unit. All personnel must be very cautious if fault tracing or adjustments are performed with
gas supply connected and with covers removed.
Some items, e.g., gas modules, are still under pressure when gas supply is disconnected.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 147 (of 369)
Chapter 4 – Disassembling and assembling
The PC boards contain components that are highly sensitive to static electricity. When
working with ESD sensitive components, always use a grounded wrist band and a grounded
work surface. Adequate service tools must be used.
Those who come into contact with circuit boards containing sensitive components must take certain
precautions to avoid damaging the components (ESD protection).
PC boards (spare parts) must always be kept in protective packaging for sensitive electronic device.
PC boards must not be inserted or removed while the mains power or battery power is applied to the PC
boards.
Remove and insert the PC boards carefully to avoid damage to the connectors.
Page 148 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Stored information PC 1900 PC 1917 PC 2060/ PC 2061/ PC 1922 PC 2063/ PC 1925 AION
PC 1920 PC 1921 PC 1923 PCB
System info: X
Serial number
System version
Installed software
options
Operating time (total and
running hours)
Clinical trends X
Barometer calibration X
Performed preventive X X1
maintenance (counter)
Replaced expiratory X X1
membrane (counter)
Replaced backlight X1 X
(counter)
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 149 (of 369)
Chapter 4 – Disassembling and assembling
Replaced memory X1
backup batteries
Startup configurations X X X
made in Service & (ventilation (alarm (units, language,
Settings waveforms,
settings) limits)
displayed values,
biomed access
code, etc.)
Anesthetic agent X
consumption, clinical use
Anesthetic agent X
consumption, service
use
M3 0.95 Nm ±15%
M4 3.1 Nm ±15%
M5 5.0 Nm ±15%
M6 7.0 Nm ±15%
Page 150 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
WARNING! Do not use lubricants that are not recommended by Getinge in high O2 exposed
areas. Risk of fire or explosion.
Lubricants should normally not be used when servicing the unit. If lubricants must be used, use only small
amounts of grease with P/N 9573700. This special grease is O 2 compatible.
CAUTION: Tubing included in the gas sampling system must not be cut before reconnected.
The tubing length is calculated and adapted to the sampling functionality. Proper operation of
the sampling system cannot be ensured if the tubing length is incorrect. Cut tubing must be
replaced.
CAUTION: Ensure that the EVAC system, including tubing from scavenging valves, is correctly
connected. Running the system with a poor connection may result in anesthetic agent being
emitted into the operation environment. Leakages in the EVAC system will not be detected by
the System checkout.
When disassembling, disconnecting internal tubing may be required. Disconnect carefully to prevent
damaging the tubing or connectors.
When assembling the system, it is important that all tubing is properly reconnected. Faulty or poor
connections lead to system malfunction.
Necessary tools:
7 mm open-end wrench
8 mm open-end wrench
10 mm open-end wrench
14 mm open-end wrench
Follow this procedure:
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 151 (of 369)
Chapter 4 – Disassembling and assembling
Necessary tools:
Flat-head screwdriver
Follow this procedure:
1. Use a flat-head screwdriver and carefully lift the
locking sleeve.
2. Place the blade under the tubing end and
carefully use the screwdriver end to gently
loosen the tube.
3. Pull off the tube from the nipple.
Version A. S/N < 300576. Control Panel including Display with lamps.
Page 152 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit, e. g. the backlight lamps in Control panel Version A that are
supplied with 660 V by the Backlight Inverter. All personnel must exercise extreme caution if
fault tracing or adjustments are performed with power supply connected and with the user
interface rear cover removed.
Necessary tools:
5 mm Hex key
Phillips PH2 screwdriver
Follow this procedure:
1. Remove the plastic cable holder (ref. no. 1).
2. Disconnect the control panel cable (ref. no. 2).
3. While holding the control panel securely,
remove the screws (ref. no. 3).
4. Put the control panel with the display side down
on a soft surface.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 153 (of 369)
Chapter 4 – Disassembling and assembling
Page 154 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.7.1, Control panel – removal, page 153.
Necessary tools:
T20 Torx® key
14 mm open-end wrench
Follow this procedure:
1. Place the control panel on a flat surface as
shown in the image.
2. Unscrew the eight screws (ref. no. 1).
3. Lift off the rear cover (ref. no. 3), a few
centimeters only.
4. Slide the rear cover slightly and carefully to the
right to get access to the rotary encoder (ref.
no. 2). The cables between the rear cover and
PC 1925 display CPU board do not need to be
disconnected.
5. Pull off the knob (ref. no. 5).
6. Unscrew the nut (ref. no. 4).
7. Disconnect and remove the rotary encoder (ref.
no. 2).
8. Mount the (new) encoder into its position.
9. Fasten the encoder using the plastic nut that
comes with the encoder. Do not use the metal
nut and washer from the old encoder.
Tightening torque: 0.95 Nm ±15%.
10. Push the knob (ref. no. 5) into position.
11. Lift and slide the rear cover (ref. no. 3) slightly
and carefully to the left.
12. Fasten the eight screws (ref. no. 1).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 155 (of 369)
Chapter 4 – Disassembling and assembling
Page 156 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Necessary tools:
Flat-head screwdriver
Follow this procedure:
1. Remove the Evac filter cover.
2. Remove the filter.
3. The Evac filter holder is mounted with adhesive
tape. Use a flat-head screwdriver to carefully
unbind one corner of the holder.
4. If needed, carefully rub off any adhesive tape
from the surface of the system.
5. Clean the surface with a suitable solvent.
6. Mount the new holder.
7. Install filter and cover.
Necessary tools:
2.5 mm hex key
Follow this procedure:
1. Open the patient cassette lid.
2. Unscrew the two screws on the Inspiratory
sample filter holder.
3. Pull off the filter holder.
Version B
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 157 (of 369)
Chapter 4 – Disassembling and assembling
Version A
Necessary tools:
2.5 mm hex key
Follow this procedure:
1. Open the patient cassette lid.
2. Unscrew the screw on the Expiratory sample
filter holder.
3. Pull off the filter holder.
Version B
Version A
Page 158 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 159 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.11.1, Patient cassette – removal,
page 159
Follow this procedure:
1. Turn the absorber valve counterclockwise.
2. Pull out the absorber valve.
Prerequisite tasks:
Section 4.11.1, Patient cassette – removal,
page 159
Follow this procedure:
1. Push the absorber valve into position.
2. Turn the valve clockwise.
Page 160 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.11.1, Patient cassette – removal,
page 159
Necessary tools:
Retaining ring pliers
Follow this procedure:
Use a pair of retaining ring pliers to carefully lift off
the membrane and retaining ring.
Prerequisite tasks:
1. Section 4.11.1, Patient cassette – removal,
page 159
2. Section 4.11.4, APL/PEEP valve membrane –
removal, page 161
Necessary tools:
Retaining ring pliers
Necessary consumables:
Distilled water
Lint-free cloth
Follow this procedure:
Note: In the following step, make sure that the
membrane is not deformed from the installation.
1. Place the retaining ring (ref. no. 1) into the
membrane. Make sure that the retaining ring
(ref. no. 1) and washer (ref. no. 2) are correctly
mounted.
2. Use distilled water and a lint-free cloth to clean
the valve membrane and the valve membrane
seat.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 161 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.11.1, Patient cassette – removal,
page 159
Necessary tools:
Retaining ring pliers
Follow this procedure:
1. Use a pair of retaining ring pliers to carefully lift
off the membrane and retaining ring (ref. no. 1).
2. Lift off the expiratory outlet one-way valve
OV12 (ref. no. 2).
Page 162 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
1. Section 4.11.1, Patient cassette – removal,
page 159
2. Section 4.11.4, APL/PEEP valve membrane –
removal, page 161
3. Section 4.11.6, Expiratory outlet one-way valve
OV12 – removal, page 162
Necessary tools:
Needle-nose pliers
Necessary consumables:
Distilled water
Follow this procedure:
1. Lubricate the expiratory outlet one-way valve
with a small amount of distilled water
2. Push the valve into place.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 163 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.11.1, Patient cassette – removal,
page 159
Necessary tools:
Flat-head screwdriver
Follow this procedure:
1. Turn the inspiratory and expiratory valve caps
counterclockwise
2. Pull up the valve caps.
3. Lift up the valve cages.
4. Pull up the valve seats. If needed, use a small
flat-head screwdriver on the outside of the
seats to pry them loose.
Prerequisite tasks:
1. Section 4.11.1, Patient cassette – removal,
page 159
2. Section 4.11.8, Inspiratory and expiratory valve
– removal and disassembly, page 164
Necessary consumables:
Distilled water
Follow this procedure:
Note: Handle the valve seats very carefully. Make
sure the sharp edge that seals towards the ceramic
valve disc is not damaged by using any tool.
1. Lubricate the O-rings on the valves seats with a
small amount of distilled water.
2. Place the new valve seats with O-rings into the
holders.
Page 164 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 165 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Remove the drawer(s).
Follow this procedure:
Page 166 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Front door version B with metal vaporizer slot housing integrated with the front door.
Front door version A with a separate plastic vaporizer slot housing.
Prerequisite tasks:
1. Remove any accessories mounted on the rails.
2. Remove the vaporizers.
Necessary tools:
3 or 4 mm Hex key
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 167 (of 369)
Chapter 4 – Disassembling and assembling
Note: The document “Vaporizer alignment tool, Flow-i – Adjustment Instructions” gives further information
on how to close the front door correctly. The instruction also describes adjustment of PC 1909 Vaporizer
adapter, using the Vaporizer alignment tool, to assure proper connection between the vaporizers and
PC 1909.
PC 1909 with new design (available as spare part) has improved connection to the vaporizers.
Page 168 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 169 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
1. Section 4.13.1, Front door – opening, page
167.
2. Section 4.13.3, Rear cover – removal, page
169.
Necessary tools:
T6 Torx® key
Adhesive tape
Page 170 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 171 (of 369)
Chapter 4 – Disassembling and assembling
Page 172 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
4 mm hex key
Follow this procedure:
1. Slightly loosen (a half to one turn) the screws
that hold the gas module to the docking. There
are two screws for each gas module.
2. Carefully move the gas module several times
from one side to the other. This is to release the
gas pressure in a controlled manner.
3. Unscrew the two screws.
4. Pull out the gas module.
Note: For systems with Gas distribution section of Version A (S/N 4000 and below) and revision 01–07
(e.g. xxxxxxx/05), see label in detail in figure above:
Do not remove all four gas modules at the same time. These systems do not have screws securing the
position for the internal gasket. If all four gas modules are removed, the gasket may come out of position
and cause leakage. From revision 08, the gasket position is secured with screws.
Prerequisite tasks:
1. Section 4.13.3, Rear cover – removal, page
169.
2. Section 4.15.1, Gas module – removal, page
173.
Necessary tools:
4 mm hex key
Follow this procedure:
Unscrew the four screws to open the gas module.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 173 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
1. Section 4.15.1, Gas module – removal, page
173.
2. Section 4.15.2, Gas module – disassembly,
page 173.
Necessary consumables:
Lint-free cloth
Follow this procedure:
1. Pull out the filter with its rubber seal from the
gas module.
2. Clean the rubber seal with a lint-free cloth. Dirt
particles on the rubber seal may cause
leakage.
3. Put the rubber seal on the new filter.
4. Check if there is moisture in the gas module
filter housing.
Note: Moisture can be an indication of too high
moisture content in the gas supply. Inform the
customer that gas supply must fulfill
requirements for medical grade gases
according to applicable standards. Refer to
technical specifications in the User’s Manual.
Page 174 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.15.1, Gas module – removal, page 173
Follow this procedure:
1. Unlock the lock at the top of the gas module.
2. Pull out the nozzle unit.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 175 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.15.1, Gas module – removal, page 173,
at least the two modules to the left (see figure).
Follow this procedure:
1. Replace the four O-rings on the gas module
docking.
2. Mount the gas module.
Prerequisite tasks:
Section 4.15.1, Gas module – removal, page 173,
only the two modules to the left (as shown in
figure).
Necessary tools:
2.5 mm hex bit
Page 176 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
Phillips PH0 screwdriver
2.5 mm Hex key
Follow this procedure:
1. Disconnect the pilot valve cable from the
PC 2057/PC 1907 valve drivers board.
2. Unscrew the two screws on the pilot valve.
3. Pull out the pilot valve.
A. Pilot valve
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 177 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
2.5 mm hex key
Follow this procedure:
1. Disconnect all cables and the tube to PC 1781
pressure transducer board.
2. Unscrew the five screws on PC 2057 valve
drivers’ board.
3. Cut the cable tie on PC 1781 pressure
transducer board.
4. Remove the PC 1781 pressure transducer
board and keep for use together with the
replacement PC 2057 board.
5. Install the new PC 2057 valve drivers’ board in
the reverse order.
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
2.5 mm hex key
Follow this procedure:
1. Remove the Fresh gas filter (ref. no. 3).
2. Remove the Fresh gas pipes (ref. nos. 1 and 2)
together with their connecting joints.
Page 178 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
3. Disconnect all cables and the tube to PC 1781 pressure transducer board.
4. Unscrew the five screws on PC 1907 valve drivers’ board.
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
2 mm hex key
Follow this procedure:
1. Unscrew the two screws (ref. no.1).
2. Carefully pull out the regulator. Pressurized gas
may be released.
3. Install the new regulator(s) in reverse order.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 179 (of 369)
Chapter 4 – Disassembling and assembling
Page 180 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.16.5.1, Double channel plate – removal,
page 180.
Necessary tools:
5 mm Hex key
Follow this procedure:
1. Place the plate onto the screws (1) and press it
firmly to the right as shown in illustration.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 181 (of 369)
Chapter 4 – Disassembling and assembling
4.17 PC boards
4.17.1 PC 1984/PC 1919, PC 2060/PC 1920, PC 2061/PC 1921, PC 1922, PC 2063/PC 1923 – removal
Refer to section 4.4, PC boards and electrical components on page 148 for detailed information.
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
3 mm hex key
Phillips PH2 screwdriver
Follow this procedure:
1. Unscrew the two screws (ref. no. 1) on the PC
board cover and remove the cover.
The order of the PC boards (from top to bottom) is
the following:
PC 1984/PC 1919 expiratory channel board
Page 182 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
3 mm hex key
Follow this procedure:
1. Remove the aluminum cover (ref. no. 1).
2. Mark and remove all cables (ref. no. 2). The
AC/DC inlet cable is attached with a cable
clamp.
3. Remove the two screws on the right side (ref.
no. 4).
4. Remove the two screws (ref. no. 3).
5. Pull out the assembly.
Necessary tools:
3 mm hex key
Follow this procedure:
Referring to the removal instructions, install the
components in reverse order.
Note: When mounting the assembly, make sure
that the lower edge of the assembly fits in the board
rail and that the upper edge fits in the recessed
parts (see marked areas in figure).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 183 (of 369)
Chapter 4 – Disassembling and assembling
4.17.3.1 General
Lithium batteries are used on PC 2060/PC 1920 control board, PC 2061/PC 1921 monitoring board and PC
1922 panel board. These batteries must be replaced every four years. A Technical error message will
appear on the screen if the battery voltage level is too low:
Battery on PC 2060/PC 1920 control board: TE 501 Memory backup battery depleted.
Battery on PC 2061/PC 1921 monitoring board: TE 20 Memory backup battery depleted.
Battery on PC 1922 panel board: TE 614 Memory backup battery depleted.
Note: Always replace all three batteries at the same time, to keep the same replacement date for the
batteries.
ESD sensitive components. The lithium batteries are mounted on ESD sensitive PC
boards. Refer to chapter Disassembling and assembling, section 4.4.2, Handling PC
boards, on page 148 for further information regarding ESD sensitive components.
Prerequisite tasks:
Switch off the system using the Power button.
Disconnect the mains power cable.
Disconnect the gas supplies (wall and/or
cylinder).
Section 4.13.3, Rear cover – removal, page
169.
Necessary tools:
Side-cutting pliers
3 mm hex key
Phillips PH2 screwdriver (for removal of PC
1922 only)
Follow this procedure:
Remove the PC board cover.
Pull out the PC board(s) that accommodate(s)
the battery(ies) to replace.
Page 184 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 185 (of 369)
Chapter 4 – Disassembling and assembling
Necessary tools:
T6 Torx® key
Follow this procedure:
1. Unscrew the two screws.
Note: In the following step, the movement is limited
by attached sampling lines and cables.
2. Carefully pull out the water trap receptacle.
Page 186 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
2.5 mm hex key
Side-cutting pliers
Follow this procedure:
1. Mark the tubes to ensure correct connection
when installing the new AION Platinum unit:
o Return line from AION gas analyzer
(ref.no.1)
o Zero calibration gas tube (ref. no.2)
o Sampling line inlet (ref.no.3)
o Purge line (ref.no.4)
2. Disconnect the marked tubes.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 187 (of 369)
Chapter 4 – Disassembling and assembling
Page 188 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.18.4.1, Gas analyzer AION ADS –
removal, page 188.
Necessary tools:
3 mm Hex key
Follow this procedure:
1. Put the AION Platinum assembly in its position
and tighten the three screws (ref. no. 6).
2. The peg (ref. no. 7) replaces the screw in the
upper left corner. The peg is used to hold
tubing in place and prevent the tubing from
being pinched by the door.
3. Carefully connect the tubing:
o Zero calibration gas (ref. no. 1). Connect at
the filter end.
o Purge line (ref. no. 2).
o Sampling line (ref. no. 3).
o Return line (ref. no. 4).
4. Carefully connect the AION Platinum cable
connector (ref. no. 5) to PC 1900 Main back-
plane.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 189 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
1. Remove the four screws (or three screws and a
peg in the upper left corner) holding the AION
ADS assembly.
2. Lift off the AION ADS assembly.
3. Remove the two screws holding the O2 sensor
(ref. no. 2) to the PGA mounting plate (ref. no.
1).
Necessary tools:
3 mm Hex key
7 mm open-end wrench
Follow this procedure:
1. Mount the O2 sensor bracket (ref. no. 3) to the
PGA mounting plate. Use the screws (ref. no.
4) and nuts (ref. no. 5) included in the kit.
2. Mount the O2 sensor on the O2 sensor bracket.
Use the screws (ref. no. 6) previously holding
the O2 sensor to the PGA mounting plate.
3. Reinstall the complete AION ADS assembly.
4. Connect the new O2 sensor cable (ref. no. 7)
included in the kit. Place the cable in the space
between the O2 sensor and the AION.
Page 190 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
2.5 mm Hex key
3 mm Hex key
8 mm open-end wrench
Follow this procedure:
1. Disconnect the fresh gas sampling connector
(ref. no. 1) from the fresh gas filter.
2. Disconnect the fresh gas pipe (ref. no. 2) from
the fresh gas sampling connector (ref. no. 1).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 191 (of 369)
Chapter 4 – Disassembling and assembling
Page 192 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Replacing the version B switch implies replacing the whole MAN/AUTO – APL control assembly, see
section 4.18.7.1, Version B – removal, page 193.
For version A, only the MAN/AUTO switch needs to be replaced, see section 4.18.7.3, Version A –
removal, page 194.
Prerequisite tasks:
1. Section 4.13.1, Front door – opening, page
167.
2. Section 4.13.3, Rear cover – removal, page
169.
Necessary tools:
2.5 mm hex key
Follow this procedure:
1. Disconnect the cables from the MAN/AUTO
switch (ref. no. 2) and the APL potentiometer
from PC 2057/PC 1907 valve drivers board.
2. Put a sheet of paper on top of the EVAC unit to
prevent screws from falling into the machine.
3. Unscrew the three screws (ref. no. 1) from
below.
4. Lift up the MAN/AUTO switch.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 193 (of 369)
Chapter 4 – Disassembling and assembling
Page 194 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 195 (of 369)
Chapter 4 – Disassembling and assembling
Version B, APL potentiometer units with knob scale in zinc metal, see section Version B – removal, page
196, and section Version B – installation, page 197.
Version A, APL potentiometer units with knob scale in stainless steel, see section Version A – removal,
page 198, and section Version A – installation, page 200.
Version B – removal
Prerequisite tasks:
Section 4.13.1, Front door – opening, page
167.
Section 4.13.3, Rear cover – removal, page
169.
Necessary tools:
2.5 mm hex key
6 mm socket wrench
13 mm socket wrench
Follow this procedure:
1. Disconnect the APL potentiometer cables from
PC 1907 valve drivers board.
Pull out the cables to the front side.
2. Pull off the cap from the APL knob (ref. no. 1).
3. The potentiometer shaft is attached to the APL
knob with a collet (clamping device) (ref. no. 2).
Loosen (do not remove) the collet and remove
the coupling device (ref. no. 3) together with the
collet.
4. Remove the locking nut (ref. no. 4) and lift out
the potentiometer from the MAN/AUTO – APL
control assembly.
Page 196 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Version B – installation
Prerequisite tasks:
Section 4.18.7.1, Version B – removal, page 193.
Necessary tools:
2.5 mm hex key
6 mm socket wrench
13 mm socket wrench
Follow this procedure:
1. Mount the APL potentiometer in the
MAN/AUTO – APL control assembly.
2. Mount and tighten the locking nut (ref. no. 4).
Tightening torque is 2 Nm.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 197 (of 369)
Chapter 4 – Disassembling and assembling
Version A – removal
Prerequisite tasks:
Section 4.13.1, Front door – opening, page
167.
Section 4.13.3, Rear cover – removal, page
169.
Necessary tools:
2.5 mm hex key
6–8 mm flat tip screwdriver or 6 mm socket
wrench
13 mm socket wrench
Follow this procedure:
1. Disconnect the MAN/AUTO switch and APL
potentiometer cables from PC 1907 valve
drivers board.
2. Put a sheet of paper on top of the EVAC unit to
prevent screws from falling into the machine.
3. Unscrew the three screws (ref. no. 1) from
below and remove the MAN/AUTO – APL
control assembly from the system.
Page 198 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 199 (of 369)
Chapter 4 – Disassembling and assembling
Version A – installation
Prerequisite tasks:
Section APL potentiometer – re, page 196.
Necessary tools:
2.5 mm hex key
6–8 mm flat tip screwdriver or 6 mm socket
wrench
13 mm socket wrench
Follow this procedure:
1. Mount the new potentiometer (ref. no. 8) on the
mounting plate (ref. no.7).
Tightening torque for the nut is 2 Nm.
2. Mount the potentiometer mounting plate. Do not
tighten the collet; the potentiometer shaft must
remain free from the APL knob.
3. Mount the MAN/AUTO – APL control assembly
to the system. Make sure that the rubber seal is
in correct position.
4. Connect the MAN/AUTO switch and the APL
potentiometer cables to PC 1907.
5. Start the system.
6. With the system in standby mode, the APL
value is shown on the screen.
Page 200 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
3.5 mm flat-head screwdriver
Follow this procedure:
1. Unlock the left and right connector (ref. no. 2)
by turning the center screw (ref. no. 1) 90°
counterclockwise.
2. Pull off the connectors (ref. no. 2) from the
button and the P152 connector on the board.
3. Remove the cable (ref. no. 3) from the P153
connector on the board.
4. Tap the locking nut (ref. no. 4) gently
counterclockwise to loosen it.
5. Unthread the locking nut fully.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 201 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
2.5 mm hex key
3 mm hex key (for the calibration step)
7 mm open-end wrench
T10 Torx® key
Follow this procedure:
1. Turn the APL knob counterclockwise until it
stops.
Page 202 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 203 (of 369)
Chapter 4 – Disassembling and assembling
Page 204 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
3 mm hex key
Follow this procedure:
1. Remove the two screws (ref. no. 1).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 205 (of 369)
Chapter 4 – Disassembling and assembling
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
2 mm hex key
Follow this procedure:
1. Open front cover right.
Note: The following step exposes two loose parts
under spring load in the valve housing.
2. Remove the pneumatic valve (ref. no. 1 or ref.
no. 2) by loosening the two screws (ref. no. 3).
3. Turn the collar (ref. no. 4) counterclockwise to
release the switch.
6
There are two existing versions of the switch. However, the removal procedures are the same for the both versions.
Page 206 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 207 (of 369)
Chapter 4 – Disassembling and assembling
Version B – Replacing
Necessary tools:
9 mm socket wrench
side-cutting pliers
pliers
Follow this procedure:
1. Cut the three cables close to the DIM knob (ref.
no. 1).
2. Remove the cap on the knob and unscrew the
screw holding the knob.
3. Remove the knob.
Page 208 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
CAUTION: To avoid short circuit – always disconnect the minus (black) pole first when removing
an old battery and connect the positive (red) pole first when installing a new battery.
4.19.3 Batteries
There are two types of power backup batteries:
CAUTION: When replacing the power backup battery, make sure that the replacement battery
is of the same type and capacity (18 Ah;black respective 38 Ah;orange) as the old battery.
Installing wrong battery can cause damage to the battery.
7
The lifetime of the battery may be reduced if it is used frequently to supply power or if the battery temperature is above 40˚C.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 209 (of 369)
Chapter 4 – Disassembling and assembling
4.19.5.1 Removal
Necessary tools:
Version A
2.5 mm hex key
5 mm hex key
10 mm open-end wrench
Version B
4 mm Hex key
10 mm open-end wrench
Follow this procedure:
1. Remove the drawers to access the battery
compartment.
2. Release the screw (ref. no. 1) and lift off the
battery compartment cover (ref. no. 2).
Page 210 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
Version A
Version B
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 211 (of 369)
Chapter 4 – Disassembling and assembling
4.19.5.2 Installation
Prerequisite tasks:
Section 4.19.5.1, Removal, page 210.
Necessary tools:
Version A
2.5 mm hex key
5 mm hex key
10 mm socket wrench
Version B
4 mm Hex key
10 mm socket wrench
Follow this procedure:
1. Connect the battery cables to the new battery. Version A
Tighten the nuts firmly.
2. Place the new battery in correct position in the
compartment and secure the battery with the
battery clamp(s) (ref. no. 4).
3. Version B only: If required, tighten the screws
(ref. no. 5) to secure the battery clamp.
4. Fasten and tighten the screws (ref. no. 3).
5. Mount the battery compartment cover.
Version B
4.19.6.1 Removal
Necessary tools:
2.5 mm hex key
5 mm hex key
10 mm socket wrench
Follow this procedure:
1. Remove the drawers to access the battery
compartment.
2. Remove the screws (ref. no. 1) and lift off the
battery compartment cover (ref. no. 2).
Page 212 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
4.19.6.2 Installation
Prerequisite tasks:
Section 4.19.6, Power backup battery –
replacement, Version C, page 212.
Necessary tools:
2.5 mm hex key
5 mm hex key
10 mm socket wrench
Follow this procedure:
1. Connect the battery cables to the new battery.
Tighten the nuts firmly.
2. Place the new battery in correct position in the
compartment and secure the battery with the
battery clamp.
3. Mount the battery compartment cover.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 213 (of 369)
Chapter 4 – Disassembling and assembling
4.19.7.1 Removal
Prerequisite tasks – special for C40:
Disconnect the system from the ceiling pendant
and move the system to a suitable place for the
battery replacement:
o Remove the vaporizers and all loose
assessories
o Make sure that the swivel arm is secured
before tipping the system.
o With the pendant console as a support,
carefully tip system so that it is lying on the
lifting console.
Necessary tools:
2.5 mm hex key
5 mm hex key
10 mm socket wrench
Follow this procedure:
1. Remove the screws (ref. no. 1) holding the
battery compartment cover.
2. Remove the battery compartment cover from
underneath the system.
Page 214 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
4.19.7.2 Installation
Prerequisite tasks:
Section 4.19.7, Power backup battery –
replacement, Version D, page 214.
Necessary tools:
2.5 mm hex key
5 mm hex key
10 mm socket wrench
Follow this procedure:
1. Starting with the red plus pole (ref. no. 5)
connect the battery cables to the new battery.
2. Mount the screws (ref. no. 3) holding the clamp
(ref. no. 4) that secure the battery to the
compartment cover.
3. Mount the battery compartment cover.
4. Carefully lift up the system to standing position.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 215 (of 369)
Chapter 4 – Disassembling and assembling
4.19.8.1 Removal
Necessary tools:
4 mm hex key
5 mm hex key
8 mm socket wrench
Follow this procedure:
1. Remove the drawers to access the battery
compartment.
2. Remove the screws (ref. no. 1) and lift off the
battery compartment cover (ref. no. 2).
Page 216 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
4.19.8.2 Installation
Prerequisite tasks:
Section 4.19.8, Power backup battery –
replacement, Version E, page 216.
Necessary tools:
1. 4 mm hex key
2. 5 mm hex key
3. 8 mm socket wrench
Follow this procedure:
1. Measure the voltage over the battery poles.
A voltage of 12.5V ± 0.1V is required.
2. Put the new battery in its position and install the
mounting plate. Tighten the screws.
3. Starting with the plus (red) pole, connect the
battery cables (ref. nos. 3 and 4) to the new
battery. Place the temperature sensor cable
(ref. no. 7) on top.
4. Tighten the terminal screws.
5. Mount the battery compartment cover.
4.19.9.1 Removal
Prerequisite tasks – special for C40:
Disconnect the system from the ceiling pendant
and move the system to a suitable place for the
battery replacement:
o Remove the vaporizers and all loose
assessories.
o Make sure that the swivel arm is secured
before tipping the system.
o With the pendant console as a support,
carefully tip system so that it is lying on the
lifting console.
Necessary tools:
4 mm hex key
10 mm socket wrench
Follow this procedure:
1. Remove the four screws (ref. no. 1).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 217 (of 369)
Chapter 4 – Disassembling and assembling
4.19.9.2 Installation
Prerequisite tasks:
Section 4.19.9, Power backup battery –
replacement, Version F, page 217.
Necessary tools:
1. 4 mm hex key
2. 10 mm socket wrench
Follow this procedure:
1. Measure the voltage over the battery poles.
A voltage of 12.5V ± 0.1V is required.
2. Put the new battery in its position and install the
mounting plate. Tighten the nut (ref. no. 4).
Page 218 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 4 – Disassembling and assembling
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 219 (of 369)
Chapter 4 – Disassembling and assembling
Page 220 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5 Service procedures
5.2 Tools
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 221 (of 369)
Chapter 5 – Service procedures
3. Tap Tools.
Two service tools appear on the screen:
Gas analyzer tool
Vaporizer tool
AGC tools (System version 4.0 and above)
Page 222 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 223 (of 369)
Chapter 5 – Service procedures
4. Tap Technician.
The Start remote services button is displayed on
the Service & Settings user category screen.
Remote services network parameters are set in
Settings > Network.
Page 224 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.3.2 Status
Prerequisites:
Log in to Service & Settings, see section 5.3.1, Logging in to Service & Settings, page 223.
Follow this procedure:
Tap Status.
To view the content in the Status tab, see section 10, Diagrams, on page 369.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 225 (of 369)
Chapter 5 – Service procedures
5.3.3 Logs
A number of logs are accessible in Service & Settings. The logs can be viewed and used for troubleshooting
of the unit.
Prerequisites:
Log in to Service & Settings, see section 5.3.1, Logging in to Service & Settings, page 223.
Necessary tools:
USB memory stick for downloading logs. Only Getinge approved USB memory sticks must be used.
Page 226 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
1. Tap Logs.
3. Tap OK.
All log files have now been saved in a Logs and
Data folder on the USB memory stick.
5.3.4 Tests
Prerequisite tasks:
Log in to Service & Settings, see section 5.3.1, Logging in to Service & Settings, page 223.
Necessary tools:
A 14-18-20 mm rubber plug included in the Leakage detector tool kit (required for the AFGO leakage
check only).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 227 (of 369)
Chapter 5 – Service procedures
Page 228 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 229 (of 369)
Chapter 5 – Service procedures
Internal leakage
AFGO leakage
Perform the AFGO leakage check test during:
Installation of AFGO option. The AFGO option is available as a factory-mounted option, or as an on-site
upgrade.
Preventive maintenance
Troubleshooting
The AFGO leakage check is available in all systems, not only systems with AFGO option active. The AFGO
leakage check can only be performed on systems equipped with an AFGO valve.
Page 230 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
4. Tap Tests.
5. Tap Leakage check.
6. Select AFGO leakage.
7. Tap Start. The test now pressurizes the system
and measures any leakage between the fresh
gas module and the plugged AFGO connector.
If the leakage is above 30 ml/min at a pressure
of 50 cmH2O, the check fails.
If the leakage check fails the Test result log
includes Check Failure (CHK) logs/data. Each
CHK has a unique Check Failure Identifier
(CFI).
8. Remove the rubber plug from the AFGO
connector when the test is completed.
5.3.5 Calibration
Prerequisites:
Log in to Service & Settings, see section 5.3.1, Logging in to Service & Settings, page 223.
Necessary tools:
Equipment listed in each calibration procedure.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 231 (of 369)
Chapter 5 – Service procedures
1. Tap Calibration.
2. Tap APL knob.
3. Follow the on-screen instructions.
5.3.5.2 Barometer
The barometer must be calibrated:
If the barometer test in system checkout fails.
If a technical error code indicates barometer error.
If the Barometric pressure check performed during installation or Preventive maintenance shows that
calibration is required, meaning that the Current barometer pressure differs more than ±5% from the
actual barometric pressure.
After replacing PC 2061/PC 1921 monitoring board or in case of malfunction in the memory backup
battery on PC 2061/PC 1921 monitoring board.
After any installation, maintenance or service intervention in the system, perform a system checkout and a
manual check of the Emergency ventilation system according to instructions in the User’s Manual.
Necessary equipment:
Barometer or information about the barometric pressure at the installation site.
Note: The actual barometric pressure is affected by altitude and weather. Make sure that the absolute
barometric pressure at your current location and elevation is used. Do not use barometric pressure readings
from a weather service as these typically report barometric pressure at sea level.
Page 232 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Important
Before starting any check or calibration of AION or O2 sensor, read the complete instructions below to
be familiar with the procedures.
Leakage in the system will result in a faulty calibration. Always perform a System checkout or at least a
leakage check before starting the calibration. To perform a leakage check, select either Leakage check
on the Standby menu or AUTO ventilation leakage and MAN ventilation leakage in Service & Settings.
After any installation, maintenance or service intervention in the system, perform a System checkout and
a Manual check of Emergency ventilation system according to instructions in the User’s Manual.
CAUTION: The calibration gas contains substances that may be detrimental to your health.
During use of calibration gas, assure that the system is connected to an effective gas
evacuation system, e.g., the hospital’s EVAC system.
When to calibrate
The gas analyzer must be calibrated:
If the Patient gas analyzer AION check or the O2 sensor check indicates that calibration must be
performed. It is only required to calibrate the unit (AION or O2 sensor) that failed during the test. If both
units failed, always start with O2 sensor calibration.
If either AION ADS or O2 sensor Servomex PM1111E has been replaced, the O2 sensor must be
calibrated to save the O2 sensor calibration values in AION ADS.
In case of Technical errors and/or System Checkout failure.
Required equipment
Calibration gas and equipment (see further information below):
o P/N 7265908 Collection bag
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 233 (of 369)
Chapter 5 – Service procedures
Page 234 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
1. Tap Calibration.
2. Tap Gas analyzer.
3. Check that the Prerequisites are fulfilled.
4. Tap Start.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 235 (of 369)
Chapter 5 – Service procedures
Page 236 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
A completed calibration sequence will display the last page in the calibration procedure. A dialog on this
page will display the result of the calibration:
o If the calibration check resulted in measured calibration values already within valid intervals, the text
Calibration not necessary will appear.
o If calibration was performed and accepted, the text Calibration accepted by user will appear.
o A cancelled calibration will also end at the last page if the cancellation was submitted while showing
a page displaying calibration values. In this case, the text Calibration cancelled by user. Ensure that
the calibration gas bottle valve is closed will appear.
The result of the calibration procedure is saved in the Service report log in Service & Settings. See Logs
> Service report.
The result of the calibration procedure (e.g., Last calibration and Last calibration check) is also saved in
Gas analyzer status in Service & Settings. See Status > Gas analyzer.
Note: This setup shows equipment delivered by Getinge. If other calibration gas and equipment is used, see
information in section, Required equipment on page 233.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 237 (of 369)
Chapter 5 – Service procedures
Note: This check, as described above, is performed with the gas and equipment delivered by Getinge. If
other calibration gas and equipment is used, the procedure and the acceptance values must be adapted.
The gas measurement accuracy for different O2 concentrations is stated in the User’s Manual, chapter
Technical specifications, section Gas analyzer.
Follow this procedure:
1. Connect the USB service key.
2. Set the system to Manual ventilation (MAN).
3. Push the Start case button.
4. Push the Menu button.
5. Tap Tools > Gas analyzer tool.
6. Connect the calibration gas kit.
7. Open the calibration gas bottle valve and carefully fill the collector bag with calibration gas. The bag will
not withstand high gas pressure. The 3-way-valve must be set to “room Air”.
8. Set the 3-way-valve to ”calibration gas”. If required, adjust the calibration gas bottle valve to keep a
sufficient amount of calibration gas in the collector bag during the calibration procedure.
9. Wait 30 seconds for stabilization.
10. Check the displayed value for Patient O2 in the Gas analyzer tool.
11. If the displayed value is within valid range 47.7–56.3% when using the Getinge calibration gas,
calibration is not required. If other calibration gas is used, see Gas measurement accuracy in the User’s
Manual (Technical specifications > Gas analyzers).
1. Close the calibration gas bottle valve.
2. Allow the system to purge the gas in the calibration setup, including the collection bag, into the gas
evacuation system.
3. Set the 3-way-valve to “room Air”.
4. Remove the calibration gas kit.
5. Select End case.
12. If the displayed value is outside valid range, calibration is required:
1. Close the calibration gas bottle valve.
2. Allow the system to purge the gas in the calibration setup, including the collection bag, into the gas
evacuation system.
3. Set the 3-way-valve to “room Air”.
4. Select End case.
5. Continue with section 5.3.6.3, Patient O2 sensor calibration on page 238.
Page 238 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 239 (of 369)
Chapter 5 – Service procedures
Page 240 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
On tab Step 3:
5. Connect the calibration gas kit. The 3-way-
valve must be set to ”room Air”.
6. Follow the instruction on the Step 3 tab.
8. Click Continue.
9. Click OK.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 241 (of 369)
Chapter 5 – Service procedures
Note:
With the O2 sensor Servomex 1111E mounted on
the O2 sensor bracket (ref. no.1), the potentio-
meters used during calibration of the O2 sensor are
easier to access.
For installation instructions, see section 4.18.5, O2
sensor bracket – installation (AION ADS only),
page 189.
Page 242 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 243 (of 369)
Chapter 5 – Service procedures
Page 244 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 245 (of 369)
Chapter 5 – Service procedures
5.3.10 Settings
The Settings menu is used for setting date and time, for calculating total anesthetic agent consumption, and
for configuring the remote services network connection.
Prerequisites:
Log in to Service & Settings, see section 5.3.1, Logging in to Service & Settings, page 223.
Tap Settings. To view the content in the Settings tab, see section 10, Diagrams, on page 369.
Page 246 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Checks the internal sampling components, from the water trap through the PGA to R3 in the Patient cassette
docking. This test can also be used to check repair or maintenance performed on these components, for
example replacement of the Nafion tubing.
Part 2, steps 5–7 (refer to section 5.4.3, Sampling line pressure drop test procedure on page 247).
Sampling line connected.
Recommended upper pressure drop limit is approximately 60 hPa for System software version 4.04.00
and below.
Recommended upper pressure drop limit is approximately 75 hPa for System software version 4.04.01
and above.
This part of the test will check the complete sampling system with the sampling line connected between the
water trap and the Y-piece. The Getinge sampling line is designed (as specified in the User’s Manual) for the
System and should be recommended.
Sampling lines with characteristics outside these specifications are improper to use with the system. Such
sampling lines may cause a pressure drop above 70/85 hPa and disable AGC.
The 60/75 hPa limit in this test is chosen to ensure that the pressure drop in the system is, with margin, lower
than the System checkout limit.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 247 (of 369)
Chapter 5 – Service procedures
Figure 5-1: Test results log. The pressure drop in this example is 54.4 hPa.
Only personnel trained and authorized by Getinge shall be permitted to install system and
vaporizer software.
For functional enhancement the latest released system software version is recommended. This also applies
to the vaporizer. Make sure to include the vaporizer in every system software and vaporizer software
installation.
Vaporizer software installation is included in the system software installation. From System version 3.0,
Vaporizer software can be installed without performing the complete System software installation. Both
vaporizer slots can be used during the installation procedure. If more than two vaporizers are used, then the
installation procedure must be repeated.
A system software installation may change system functionality and a new version of the User’s Manual may
be required.
The system software package includes:
Installer application
System software
If the included installer application is of a different version than the version installed in the system, the
installer application is installed during the first phase of the installation. A restart of the installation is required
to run the new installer application.
CAUTION: The system must be connected to the mains power during the installation process.
Trying to install the software when connected to battery will either fail or cause malfunction to
the system.
CAUTION: Do not switch off the system during the installer application installation process. An
interrupted process may cause damage to the PC boards and the PC boards must be replaced.
Note that the installer application sometimes makes pauses in the installation process.
Page 248 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
The system software is distributed to a number of PC boards, refer to section 4.17, PC boards, on page 182.
If any of these PC boards is replaced, then a system software installation must be performed to make sure
that the correct software version is installed on all PC boards in the system. A compatibility check of the PC
boards is performed at every system start-up.
Note: After any installation, maintenance or service intervention in the system, perform a system checkout
and a manual check of the Emergency ventilation system according to instructions in the User’s Manual.
1. Tap Status to check the installed system software version. It is not recommended to install a system
software of a lower version number than the system software installed in the system.
2. Check the system software version stored on the USB memory stick. Remove old software versions on
the USB memory stick before copying new files to the USB memory stick.
3. Before any software installation, make sure that the system is fully compatible to the software.
Prerequisites and other important information for the different system versions can be found in section
9.1, System revision history on page 363.
Necessary tools:
USB memory stick with the new system software version. Only Getinge approved USB memory sticks
must be used in the system8.
Tool for pressing the Boot enable switch, e.g., a 1.5 mm hex key.
Service key.
8
From System version 3.0 the Vaporizer software can be installed without performing the complete System software installation, and
USB memory stick with System software is not required.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 249 (of 369)
Chapter 5 – Service procedures
Page 250 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 251 (of 369)
Chapter 5 – Service procedures
Figure 5-2: Installation of the subsystems on different PC boards is shown on this illustration.
Installation procedure does not start automatically. Installer application on PC 1922 panel board may
The installer screen does not appear. be corrupt.
PC board must be replaced.
Installation procedure starts, but the Installer status Installer application on PC 2060/PC 1920 control
bar for Monitoring and/or Control is not green. board and/or PC 2061/PC 1921 monitoring board
may be corrupt.
PC boards must be replaced.
Installer update failed. System software file on the USB memory stick
corrupt, transfer to flash memory failed, checksum
error, etc.
Do not set the power button to off.
Remove the USB memory stick and insert again to
restart the installation.
Page 252 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
System software installation failed. System software file on the USB memory stick
corrupt, transfer to flash memory failed, checksum
error, etc.
Set the power button to off and restart the system
software installation.
9
Valid only for systems with System software version 3.0 or above.
10
From System version 3.0 the Vaporizer software is also copied to, and stored on, the Compact Flash memory on PC 1922 Panel
board.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 253 (of 369)
Chapter 5 – Service procedures
Page 254 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
1. Copy the Startup configuration to a USB memory stick prior to a Software option installation and reinstall
the configuration afterwards. Some parts of the configuration may be changed during Software option
installation.
2. Make sure that the system is fully compatible to the software before starting any software installation.
Prerequisites and other important information for the different system versions can be found in section
9.1, System revision history on page 363.
Necessary equipment:
USB memory stick with the software option. Only Getinge approved USB memory sticks must be used in
the system.
Service key.
A 14-18-20 rubber plug included in the Leakage detector tool kit, for test of the AFGO option.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 255 (of 369)
Chapter 5 – Service procedures
Page 256 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Note: After any installation, maintenance or service intervention in the system, perform a system checkout
and a manual check of the Emergency ventilation system according to instructions in the User’s Manual.
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit. All personnel must exercise extreme caution if fault tracing or
adjustments are performed with power supply connected and with covers removed.
CAUTION: With gas supply connected to the system, there are pressurized components inside
the unit. All personnel must be very cautious if fault tracing or adjustments are performed with
gas supply connected and with covers removed.
Note that in the leakage detection illustrations below in this section, a calibration manometer and a blood
pressure bulb with valve are used. However, the procedure when using a cuff filling/measuring device with
pressure gauge is the same.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 257 (of 369)
Chapter 5 – Service procedures
Page 258 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 259 (of 369)
Chapter 5 – Service procedures
Page 260 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 261 (of 369)
Chapter 5 – Service procedures
Page 262 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 263 (of 369)
Chapter 5 – Service procedures
5.7.11 Test case 8: APL/PEEP valve coil, Manual ventilation valve actuator and Fresh gas safety
valve
Page 264 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 265 (of 369)
Chapter 5 – Service procedures
Page 266 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 267 (of 369)
Chapter 5 – Service procedures
General
When performing the fresh gas line tests, it is recommended to cross-check the system overview, which can
be found in section 10, Diagrams on page 369.
5.7.15.1 Version B
Preparations
These tests should be performed if a leakage is detected in section 5.7.8, Test case 5: Fresh gas line on
page 262.
Page 268 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Use the Valve tests function in Service & Settings to facilitate troubleshooting.
1. Connect the gas supplies.
2. Open Service & Settings > Service functions > Valve tests.
3. Close the Vaporizer bypass valve. This will divide the fresh gas line in two sections:
Section A: Section A includes the following parts:
o Subtask 1: Isolating faults to the safety valve housing in a removed state.
o Subtask 2: Isolating faults to the O2 flush button.
o Subtask 3: Isolating faults to the PV7 AFGO valve or the OV13 valve.
o Subtask 4: Isolating faults to the PV7 AFGO valve.
o Subtask 5: Isolating faults to the fresh gas pressure transducer.
Section B: Section B includes the following parts:
o Subtask 6: Isolating faults to the fresh gas modules or docking seals.
Test order
Select the applicable tests and in which order to perform the tests, for example if the leakage is detected in
Section A or Section B.
Test procedure
Follow this procedure:
1. Undock the vaporizers. The vaporizers should not be connected during the fresh gas line leakage tests
(as it was tested in section 5.7.7, Test case 4: Vaporizer on page 261).
2. With the Vaporizer bypass valve closed, pressurize the fresh gas line as described in Test case 5: Fresh
gas line:
If the leakage remains when the Vaporizer bypass valve is closed, the Fresh gas line leakage is located
in Section A. Check for leakage in main parts included in section 5.7.15.2, Section A, starting on page
270.
If no leakage is detected when the Vaporizer bypass valve is closed, the Fresh gas line leakage is
located in Section B. Check for leakage in main parts included in section 5.7.15.3, Section B, starting on
page 273273.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 269 (of 369)
Chapter 5 – Service procedures
5.7.15.2 Section A
Necessary tools:
4 mm hex key
Necessary consumables:
Distilled water
Follow this procedure:
1. Loosen the screw 8 mm (ref. no. 1).
2. Remove the safety valve housing.
3. Remove the valve membrane (ref. no. 3) from
the housing.
4. Make sure that the valve membrane (ref. no. 3)
and valve seat (ref. no. 2) are clean.
5. Mount the valve membrane. Use distilled water
as lubricant and make sure that the membrane
seal mates with the groove.
6. Plug the fresh gas inlet on the safety valve.
7. Connect the leakage detector to the fresh gas
outlet on the safety valve.
8. Push the membrane to close it.
9. Pressurize to 80–100 cmH2O.
If there is a leakage, the leakage is isolated to
the safety valve housing.
Page 270 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Subtask 3: Isolating faults to the PV7 AFGO valve or the OV13 valve
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 271 (of 369)
Chapter 5 – Service procedures
Page 272 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.7.15.3 Section B
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 273 (of 369)
Chapter 5 – Service procedures
5.7.15.4 Version A
These tests should be performed if a leakage is detected in section 5.7.8, Test case 5: Fresh gas line on
page 262.
Use the Valve tests function in Service & Settings to facilitate troubleshooting.
1. Connect the gas supplies.
2. Open Service & Settings > Service functions > Valve tests.
3. Close the Vaporizer bypass valve. This will divide the fresh gas line in two sections:
Section A: Section A includes the following parts, from the fresh gas outlet in the safety valve to the
double channel plate outlets.:
o Subtask 1: Isolating faults to the fresh gas safety valve housing in a removed state.
o Subtask 2: Isolating faults to the O2 flush button.
o Subtask 3: Isolating faults to the PV7 AFGO valve or the OV13 valve.
o Subtask 4: Isolating faults to the PV7 AFGO valve.
o Subtask 5: Isolating faults to the fresh gas filter outlet connector.
o Subtask 6: Isolating faults to the fresh gas filter.
o Subtask 7: Isolating faults to the vaporizer valve section outlet.
Section B: Section B includes the following parts, from the double channel plate inlets to the fresh gas
modules nozzle units:
o Subtask 8: Isolating faults to the vaporizer valve section inlet.
o Subtask 9: Isolating faults to the fresh gas supply.
o Subtask 10: Isolating faults to the fresh gas pressure transducer.
o Subtask 11: Isolating faults to the fresh gas modules or docking seals.
Test order
Select the applicable tests and in which order to perform the tests, for example if the leakage is detected in
Section A or Section B.
Page 274 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
2. With the Vaporizer bypass valve closed, pressurize the fresh gas line as described in Test case 5: Fresh
gas line:
o If the leakage remains when the Vaporizer bypass valve is closed, the Fresh gas line leakage is
located in Section A. Check for leakage in main parts included in section 5.7.15.25.7.15.5, Section
A, starting on page 275.
o If no leakage is detected when the Vaporizer bypass valve is closed, the Fresh gas line leakage is
located in Section B. Check for leakage in main parts included in section 5.7.15.6, Section B,
starting on page 279.
5.7.15.5 Section A
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 275 (of 369)
Chapter 5 – Service procedures
Subtask 3: Isolating faults to the PV7 AFGO valve or the OV13 valve
Page 276 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 277 (of 369)
Chapter 5 – Service procedures
Page 278 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.7.15.6 Section B
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 279 (of 369)
Chapter 5 – Service procedures
Page 280 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Subtask 11: Isolating faults to the fresh gas modules or docking seals
Follow this procedure:
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 281 (of 369)
Chapter 5 – Service procedures
5.7.16.1 Version B
Page 282 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 283 (of 369)
Chapter 5 – Service procedures
Page 284 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.7.16.2 Version A
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 285 (of 369)
Chapter 5 – Service procedures
Page 286 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.8.1 General
The pressure regulators are located on the back of the gas inlets and outlets and are accessed by opening
the right front cover. There are two (S/N < 20 000: three) test points where a pressure gauge can be
connected to check and adjust the regulators.
The pressure regulators must be checked or adjusted:
During preventive maintenance
During troubleshooting
After any installation, maintenance or service intervention in the system, perform a System checkout and a
Manual check of Emergency ventilation system according to instructions in the User’s Manual.
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit. All personnel must exercise extreme caution if fault tracing or
adjustments are performed with power supply connected and with covers removed.
CAUTION: With gas supply connected to the system, there are pressurized components inside
the unit. All personnel must be very cautious if fault tracing or adjustments are performed with
gas supply connected and with covers removed.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 287 (of 369)
Chapter 5 – Service procedures
5.8.2 Version B
5.8.2.2 Procedure
Prerequisite tasks:
Section 4.13.1, Front door – opening, page 167.
Necessary tools:
4 mm hex key
Follow this procedure:
1. Connect the Connector for pressure gauge
(ref. no. 1) and the pressure gauge (ref. no. 2)
to the relevant test point.
Page 288 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.8.3 Version A
5.8.3.2 Procedure
Prerequisite tasks:
Section 4.13.1, Front door – opening, page
167.
Connect mains power and start the system.
Disconnect gas supply (AIR and O2).
Start case and let the system ventilate a few
breaths to depressurize the system.
Switch the system to Standby mode.
Necessary tools:
3 mm hex key
13 mm open-end wrench
Follow this procedure:
1. Connect the Connector for pressure gauge
(ref. no. 1) and the pressure gauge (ref. no. 2)
to the relevant test point (see below).
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 289 (of 369)
Chapter 5 – Service procedures
Page 290 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
5.9.1 General
The Mechanical APL must be calibrated:
If a check of the Mechanical APL shows that calibration is required.
After replacement of parts included in the Emergency APL device.
After any installation, maintenance or service intervention in the system, perform a System checkout and a
Manual check of Emergency ventilation system according to instructions in the User’s Manual.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 291 (of 369)
Chapter 5 – Service procedures
Prerequisite tasks:
Section 4.13.1, Front door – opening, page
167.
Necessary tools:
Screwdriver
Follow this procedure:
1. Connect O2 to the system.
2. Connect patient tubing and Y-piece (ref.
no. 1).
3. Connect the pressure gauge to the Y-piece
(ref. no. 2).
4. Plug the pressure gauge inlet (ref.no.3).
5. Plug the manual ventilation bag connector (ref.
no. 4).on the patient cassette.
Page 292 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 293 (of 369)
Chapter 5 – Service procedures
5.10.1 Procedure
Prerequisite tasks:
Section 4.13.3, Rear cover – removal, page 169.
Necessary tools:
8 mm open-end wrench
13 mm open-end wrench
19 mm open-end wrench
Follow this procedure:
1. Loosen the screw (ref. no. 1) slightly to release
the gas connection.
2. Loosen the lock nut (ref. no. 2).
3. Turn the actuator (ref. no. 3), clockwise or
counterclockwise, to the correct position A
indicated in the illustration, where A= 6.6 ± 0.5
mm.
The valve piston is magnetic. A hex key can be
attached horizontally to be used as a reference.
4. Secure this position with the lock nut (ref.
no. 2).
Page 294 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
WARNING! The swivel adjustment knob must be tightened, holding the arm firmly in its
position.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 295 (of 369)
Chapter 5 – Service procedures
WARNING! The tilt adjustment screw must be tightened, holding the monitor/control panel
firmly in its position.
WARNING! The swivel adjustment knob must be tightened, holding the arm firmly in its
position.
Necessary tools:
8 mm hex key
5 mm hex key
Follow this procedure:
Tighten or loosen the screws (ref. no. 1 or 2) for
swivel adjustment.
Page 296 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
WARNING! This device, when removed from its wall or ceiling mount, does not meet the
stability requirements of ISO 80601-2-13 and IEC 60601-1 respectively. Special caution has to
be taken.
CAUTION: Disconnect all electrical cables and gas connections before undocking the system
from the ceiling pendant.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 297 (of 369)
Chapter 5 – Service procedures
Front view
Page 298 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
Rear view
Verify that the ceiling pendant arm complies with IEC 60601-1.
Verify that the ceiling pendant arm complies with ISO 11197.
Verify that the ceiling pendant arm have a net load capacity of at least 200 kg.
The net weight of the C40 unit, including ceiling pendant interface but without accessories, is approx.
150 kg:
o Verify that the combined weight of the C40 unit, with additional accessories, do not exceed the net
load capacity of the ceiling pendant.
o Verify that the combined weight of C40 with additional accessories do not exceed 250 kg, which is
the net load capacity of the ceiling pendant interface.
Verify that the vertical movement of the ceiling pendant arm fulfills the following criteria:
o Movement is in the operator’s field of view.
o Movement is only possible by continuous activation of the control.
o An emergency stopping device is provided.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 299 (of 369)
Chapter 5 – Service procedures
Perform an electrical safety test of the complete system. For electrical safety test of the system, refer to
chapter Preventive Maintenance, section Safety inspection.
After any installation, maintenance or service intervention in the system, perform a System checkout and a
Manual check of Emergency ventilation system according to instructions in the User's Manual.
Drawer N/A 5 5 5
Top shelf15 1 20 20 20
11
To maintain the lift function, the combined load of the extra equipment and accessories mounted on the height adjustable part should
not exceed 30 kg.
12
Maximum configurations is dependent on the lift capacity (minimum requirement 200 kg) of ceiling pendants approved for C40 and the
maximum capacity of the C40-pendant interface (250 kg). This gives a range of approx. 50-100 kg for accessories and extra equipment,
depending on the ceiling pendant.
13
Applies when using a single arm per system, attached to 1 or 2 shafts and with a load at a maximum distance of 500 mm from the
system.
14
C30 can only be equipped with the additional arm or the backup gas holder.
15
The system can be equipped with either Top shelf or Patient monitor.
Page 300 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 5 – Service procedures
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 301 (of 369)
Chapter 5 – Service procedures
Page 302 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
6 Troubleshooting
6.1 General
Before starting troubleshooting, try to eliminate all possibilities of operational errors. If the malfunction
remains, use the troubleshooting guides below as well as the information in section 3, Description of
functions, starting on page 31, to locate the faulty part.
There should be a logical trend to your troubleshooting. First try to trace the faulty function. Then the most
suitable procedure is to track down the exact fault by systematically replacing individual spare parts, one at a
time. One purpose of section 3, Description of functions starting on page 31 and section 10, Diagrams,
starting on page 369 in this Service Manual is to make it easier to trace faults.
When the fault is corrected, carry out a complete System checkout as described in the User’s Manual.
The troubleshooting guides below are focused only on technical problems. Information about clinical related
problems can be found in the User’s Manual.
For functionality enhancement, the latest released System software version is always recommended.
The information in this chapter applies to System software version 4.7.
WARNING! With power supply connected to the system, there are energized electrical
components inside the unit. All personnel must exercise extreme caution if fault tracing or
adjustments are performed with power supply connected and with covers removed.
CAUTION: With gas supply connected to the system, there are pressurized components inside
the unit. All personnel must be very cautious if fault tracing or adjustments are performed with
gas supply connected.
Possible causes to malfunction not mentioned in the following troubleshooting guides are:
The system has not been correctly assembled after cleaning, maintenance or service.
Disconnection or bad connection in cable connectors, PC board connectors, and interconnection boards
(PC 1900 main backplane board, PC 1909 vaporizer adapter board, etc.).
Pinched cables. Defective cables may short-circuit the power supply, CAN bus signals, etc.
Disconnected or defective gas tubes, breathing gas as well as drive gas.
Note: Repeated disconnection/connection of tubes in quick couplings may cause a leakage.
These possible causes to malfunction must always be considered during troubleshooting.
Troubleshooting is divided into four sections:
1. System checkout: This section describes the System checkout to give knowledge about all tests
included. The Test results log, and associated log interpretation, is available only for personnel trained
and authorized by Getinge. Refer to section 6.2, System checkout on page 304.
2. Technical error codes and messages: The Technical error code interpretation is available only for
personnel trained and authorized by Getinge. Refer to section 6.3, Technical error codes and messages
on page 320.
3. PC board LED indicators: This section describes LED indicators on some of the PC boards used to
indicate activities and errors. Refer to section 6.4, PC board LED indicators on page 321.
4. Other errors and remedies: This section describes possible causes and recommended actions in case
of other technical errors in the system. Refer to section 6.5, Other errors and remedies on page 329.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 303 (of 369)
Chapter 6 – Troubleshooting
2 O2 flush test
4 Internal tests
Page 304 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
5 Barometer test
14 Battery test
15 Vaporizer 1 test
16 Vaporizer 2 test
INF Info
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 305 (of 369)
Chapter 6 – Troubleshooting
RST Reset
Page 306 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 307 (of 369)
Chapter 6 – Troubleshooting
This test checks that the Insp. and Exp. valve discs
are in place and are able to move.
1. Tap Start check to start the Insp. and Exp.
valves test.
2. Visually check that the ceramic Insp. and Exp.
valve discs in the patient cassette are moving
If the system detects that the patient cassette lid is open after the user tapped Yes, a dialog box will appear.
The user gets the choice to close the lid, or to tap Bypass in case the system has problems detecting the lid
position. Tapping Bypass will not affect the test result, but generate an event in the log.
At the start of this test, it is verified that the Air inlet supply pressure is within range. The APL/PEEP valve is
open. The test is started by opening the Fresh gas module Air. The Vaporizer bypass valve is closed in order
to build-up a pressure of approx. 110 cmH2O (measured by the Fresh gas pressure transducer). When the
target pressure is reached, the bypass valve is opened. This creates a pressure wave which will cause the
inspiratory and expiratory valve to move. Actual generated pressure will be shown in the test log. If the
pressure reaches 140 cmH2O or fails to reach 50 cmH2O, the test will be aborted.
The pressure wave is repeated for 90 s or until the test is manually deactivated.
Page 308 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 309 (of 369)
Chapter 6 – Troubleshooting
Page 310 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
Figure 6-8: 1 = Central gas supply pressure transducer; 2 = Gas module pressure transducer.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 311 (of 369)
Chapter 6 – Troubleshooting
The gain factors must be within ±5% from the factory calibration for the test to pass.
The pressures are measured one more time at 60 cmH2O, using the calculated gain factors. The Monitoring
and Control value for each pressure transducer is compared. The pressures are not allowed to differ more
than 1 cmH2O. All measured pressures must be within 55–65 cmH2O.
Leakage may cause these tests to fail. If so, a dialog will appear and give the user the possibility to check for
leakage and redo the test.
Page 312 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
Zeroing
The DC signals from the Gas modules and Patient cassette are digitalized and converted to pressure values.
PC 2060/PC 1920 control board must adjust the offset for these signals.
All Gas modules are set to deliver no flow.
PC 2060/PC 1920 control board will now measure the flow from the Gas modules and Patient cassette.
These values are set to correspond to 0 l/min in the Control software.
The offset must be within 0.000–0.040 V for each Gas module and 2.480–2.520 V for the Patient
cassette.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 313 (of 369)
Chapter 6 – Troubleshooting
The false flow correction and dynamic common mode correction is calculated for each Gas module and
is used for the control of the modules.
The false flow correction must be within 225 to 775 DAC steps and the dynamic common mode
correction must be within –1371 to 1371 DAC steps.
Zeroing
The DC signals from the Gas modules and Patient cassette are digitalized and converted to pressure values.
The PC 2061/PC 1921 monitoring board must adjust the offset for these signals.
All of the Gas modules are set to deliver no flow.
PC 2061/PC 1921 monitoring board will now measure the flow from the Gas modules and Patient
cassette. These values are then set to correspond to 0 l/min in the Monitoring software.
The offset must be within 0.000–0.040 V for each Gas module and 2.480–2.520 V for the Patient
cassette.
Flow test
If the N2O module is supplied with gas, the tests of the Fresh gas modules are tested two by two. The
Reflector gas module is tested individually.
Order: Air + N2O, O2 + N2O, Air + O2 and Reflector.
If the N2O module is not supplied with gas, the tests of the Fresh gas modules are first tested individually and
then two by two. The Reflector gas module is tested individually.
Order: Air, O2, Air + O2 and Reflector.
The modules to be tested are set to deliver 30 l/min.
The Control and Monitoring system receives the measured values from all Gas modules and the Patient
cassette. The measured value from the modules must be within the following limits in order for the test to
pass:
o Active modules: 29.4–30.6 l/min
o Inactive modules: -0.096–0.096 l/min
o Patient cassette: 90–110% of the sum of all Gas modules flow as measured by the same subsystem
(CON and MON).
Note: When the Fresh gas module O2 is active, the value from Reflector gas module will be ignored and
when the Reflector gas module is active, the value from Fresh gas module O2 will be ignored. This is due to
that the modules are using the same gas supply.
Noise measurement
The amplitude of the sound pulses generated by the ultrasonic transducers/receivers in the Expiratory flow
transducer is measured at zero gas flow.
The amplitude of the received sound pulse is measured in an earlier timeframe than during ordinary
operation. The measured amplitude gives an indication of how much of the transmitted sound pulse that
Page 314 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
travels through the solid cassette material instead of through the gas mixture in the Expiratory flow
transducer.
The values from both upstream and downstream directions are logged.
k value
First the sum of the flow values from all of Gas modules in the flow tests is calculated.
Secondly the sum of all the flow values from the Patient cassette is calculated, where the flow value from the
gas reflector tests is absolute.
The k value is produced by dividing the Gas module flow sum with the Patient cassette flow sum. The k
value must be within:
0.92 – 1.08 for System version 4.2 and below.
0.88 – 1.08 for System version 4.3 and above.
The k value is tested by multiplying each Patient cassette flow value measured by the Monitoring system
with the k value. If the sum of the flow values from the Gas modules differ more than 8% from the new
Patient cassette flow value in a subtest, the test will fail.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 315 (of 369)
Chapter 6 – Troubleshooting
Compressible volume compensation factor is calculated during this test and shall be within the limits of:
2.6 – 6.0 ml/cmH2O for System version 4.2 and below.
2.6 – 8.0 ml/cmH2O for System version 4.3 and above.
Page 316 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 317 (of 369)
Chapter 6 – Troubleshooting
4. The O2 concentration is checked by PGA and must be within 92–108%. The primary aim of the check is
to:
o Verify tightness, rather than the accuracy of measurement.
o Verify that O2 delivery is correct (no erroneous gas supply connection).
5. The pressure drop in the sampling line is continuously measured. The measured values must be within
25 hPa to 125 hPa. The measured pressure drop is saved persistently, for use to detect if sampling tube
is disconnected or not in run mode. The pressure drop value is shown in the Test results log, refer to
section 5.4, Sampling line pressure drop test on page 246.
6. The O2 fresh gas flow is stopped.
7. The SCO valve is closed and the Manual ventilation valve and the APL/PEEP valve are opened.
8. The pump flow is stopped.
Page 318 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 319 (of 369)
Chapter 6 – Troubleshooting
Page 320 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
Uses PC 2057
Uses PC 1907
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 321 (of 369)
Chapter 6 – Troubleshooting
The LEDs on PC 2057 valve drivers board, as described here, indicates that a control signal to the
concerned valve is enabled.
Page 322 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
D35 VAP 1 Lock pilot valve EMV7 D42 VAP bypass pilot valve EMV4
D36 VAP 1 selection pilot valve EMV6 D43 VAP 2 pressure valve EMV11
D37 VAP 1 scavenging valve EMV10 D44 SCO pilot valve EMV3
D38 VAP 1 pressure valve EMV9 D45 Manual vent. pilot valve EMV2
D39 CGA inlet valve EMV14 (not used) D46 AFGO valve EMV18
D40 VAP 2 selection pilot valve EMV5 D47 VAP 2 scavenging valve EMV12
D41 VAP 2 lock pilot valve EMV8 D48 CGA outlet valve EMV16 (not used)
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 323 (of 369)
Chapter 6 – Troubleshooting
Uses PC 1984
Uses PC 1919
LED#
D38 – Heating foil on Expiratory flow transducer in operation (power supply monitored).
D2 Processor reset.
D3 D3 Normal operation.
D4 D4 Startup process.
Page 324 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
Uses PC 2060
Uses PC 1920
D7 L_1G = Flashing in sequence (back and forth) with D8, D9 and D10 indicates checksum (SW or
HW) error.
D9 L_2G = Flashing in sequence (back and forth) with D7, D8 and D10 indicates checksum (SW or
HW) error.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 325 (of 369)
Chapter 6 – Troubleshooting
Uses PC 2061
Uses PC 1921
LED#
– D78 DISV = Disable valves. Will close gas modules and open Fresh gas safety valve and
APL/PEEP valve.
D1 D1 DISV = Disable valves. Will close gas modules and open Fresh gas safety valve and
APL/PEEP valve.
D8 D8 L_1G = Flashing in sequence (back and forth) with D8, D9 and D10 indicates
checksum (SW or HW) error.
D9 D9 L_1R = Flashing in sequence with D10 indicates memory (RAM) error. Flashing in
sequence (back and forth) with D7, D9 and D10 indicates checksum (SW or HW)
error.
D10 D10 L_2G = Flashing in sequence (back and forth) with D7, D8 and D10 indicates
checksum (SW or HW) error.
Page 326 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
LED#
D11 – L_2R = Flashing in sequence with D8 indicates memory (RAM) error. Flashing in
sequence (back and forth) with D7, D8 and D9 indicates checksum (SW or HW)
error.
Uses PC 2063
Uses PC 1923
LED #
LED indication
PC 2063 PC 1923
D2 D2 CPU EXEC = Solid green indicates normal CPU operation. Flashing green indicates
that PC board calibration is required (not possible during field service).
D97 – +5V_START-UP
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 327 (of 369)
Chapter 6 – Troubleshooting
LED #
LED indication
PC 2063 PC 1923
D11 D11 +24V_INSP_1_2 = +24 V power for Gas modules 1 and 2 (upper modules) available.
D12 D12 +24V_INSP_3_4 = +24 V power for Gas modules 3 and 4 (lower modules) available.
+12V_UNREG:
D14 D14 _EXP_S_VALVES = +12V_UNREG for Fresh gas safety valve and APL/PEEP valve
available.
D18 D18 _EXPANSION = +12V_UNREG for LED Lamp and Cylinder pressure (functions
controlled by PC 1903 external connectors board) available.
Page 328 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
The system will not start up. The The Emergency ventilation switch 1. Turn off the switch.
ON/OFF switch is not lit. is active. 2. Replace the switch in case of
malfunction.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 329 (of 369)
Chapter 6 – Troubleshooting
The system is ON, but not Malfunction in the Power button. Check/replace the Power button.
possible to switch OFF.
Poor cable connections. 1. Check Power button
connection to PC 1928 power
on/off board.
2. Check PC 1928 power on/off
board cable connector to P29
on PC 1900 main backplane
board.
The system will only run on power The AC/DC converter will not 1. Check cables and cable
backup battery. supply power to PC 2063/PC 1923 connectors between AC/DC
power control board. converter and PC 2063/
PC 1923 power control board.
2. Check/replace the AC/DC
converter.
The system will not run on power The power backup battery will not 1. Check power backup battery
backup battery. supply power to PC 2063/PC 1923 cables and cable connectors.
power control board. 2. Check power backup battery.
3. Check fuses F1 and F2 on
PC 1903 external connectors
board. If fuse(s) are blown, the
root cause must be found.
Probable causes are
Short circuit of the positive
battery pole to
chassis/ground. Check
battery position in the
battery compartment.
Malfunction on PC
2063/PC 1923 power
control board. Check PC
2063/PC 1923 power
control board for
component damages.
When the root cause is
remedied, replace PC 1903
external connectors board.
4. Replace PC 2063/PC 1923
power control board.
During a Leakage check the Drive gas supply is added into the Check/replace the O2 flush valve.
message: Leakage into the system patient circuit.
appears
Fresh gas or reflector gas is added Check all four gas modules. Check
to patient circuit. that nozzle units are properly
inserted and not damaged.
Page 330 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
The Mechanical APL knob does Poor adjustment of the Mechanical Adjust regulator according to
not respond with a distinct feeling APL pressure regulator REG5. instructions in section 5, Service
or gives an incorrect pressure procedures on page 221.
level.
Defective drive gas supply tube to Check tubing.
emergency ventilation section.
The set APL value will not When Emergency ventilation is Set Mechanical APL to minimum.
correspond to the value displayed switched Off, the Mechanical APL
on the Control panel. must be set to minimum. If the
Mechanical APL is set to a higher
pressure, the Emergency APL
actuator may not be depressurized
through the Excess pressure vent
in the regulator. This may affect
the normal APL function.
The set APL value will not APL knob is not calibrated. Perform an APL knob calibration
correspond to the value displayed Calibration lost e.g., during according to instructions in section
on the Control panel. System software installation. 5, Service procedures on
page 221.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 331 (of 369)
Chapter 6 – Troubleshooting
No or low vacuum Suction On/Off switch is closed Switch on the suction unit.
Pressure level set too low. Adjust the pressure level with the
regulating valve.
Septic fluid jar volume. Ensure that the septic fluid jar
used with the unit must have a
usable volume of at least 500 ml.
Page 332 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 6 – Troubleshooting
Difficult to switch on the suction The suction unit module is 1. Loosen the screws that fasten
unit. tightened too hard to the system. the suction unit.
2. Suction unit service
Too high pressure. Pressure level set too high. Adjust the pressure level with
regulating valve.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 333 (of 369)
Chapter 6 – Troubleshooting
Page 334 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
7 Preventive maintenance
7.1 General
A cleaning of the system shall be performed before carrying out Preventive maintenance. Only personnel
trained and authorized by Getinge are permitted to perform the cleaning. Instruct the system owner to
perform this cleaning and include extra patient cassettes, extra volume reflectors any other accessories and
auxiliary equipment used for the system. Refer to the User’s Manual – Cleaning and Maintenance.
In some parts of the Preventive maintenance, as described in this section, access to Service & Settings is
required.
Optional equipment not covered by this Service Manual, may also demand maintenance actions. Refer to
the documentation for the optional equipment.
The system must not be serviced or maintained while in use with a patient, with the exception to fill a
vaporizer.
Trained and authorized personnel. The system must be serviced at regular intervals by
personnel trained and authorized by Getinge. It is recommended that service and maintenance
is done as a part of a service contract. Any maintenance or service must be noted in a log book.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 335 (of 369)
Chapter 7 – Preventive maintenance
PM PM+
Preparation
Parts to be cleaned
EVAC system16 - X
Parts to be replaced
Parts to be checked
Cables – external X X
Expiratory outlet X X
Cables – internal - X
Fan X X
16
Only for units without EVAC filter.
Page 336 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Control panel X X
Barometric pressure X X
Vaporizer section17 X X
Carrier X X
Safety inspection
Completing actions
Table 7-2: Additional maintenance actions relating to power and memory backup batteries.
17
Only systems with Locking device, see section 3.9.7, Vaporizer 1 and 2 locking device, on page 63.
18
Depending on model, see section 4.19.3, Batteries, on page 209.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 337 (of 369)
Chapter 7 – Preventive maintenance
Pressure gauge, local purchase. Refer to section 5.8,Pressure regulators, on page 287.
Calibration equipment. Refer to section 5.3.5, Calibration on page 231.
O2 compatible grease, P/N 9573700. Lubricants should normally not be used when servicing the unit. If
lubricants must be used, use only small amounts of the O2 compatible grease.
Gas evacuation system.
Barometer, or information about the actual barometric pressure.
Access to Service & Settings (S&S).
If the silicone gaskets are difficult to mount, lubricate the gaskets with a small amount of distilled water.
Page 338 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Absorber valve
7.6.1 Preparation
Follow this procedure:
1. Make a general inspection/visual check for external defects or damages on the system.
2. Replace defective or damaged parts. The check shall also cover extra patient cassettes, extra volume
reflectors and other optional equipment used for the system.
3. Make sure that gas supply hoses, connectors and color codings are correct.
4. Check and note serial number, system software version and operating time. Refer to Service & Settings.
5. Check that a User’s Manual corresponding to the installed System version is available. Also check that
operating manuals for optional equipment connected to the system are available.
6. Check if there are any unexpected Technical alarms in the logs. A Service key with access to Service &
Settings is required.
7. Perform a complete System checkout according to instructions in the User’s Manual.
8. After completed System checkout, download all logs.
9. Perform a manual check of emergency ventilation system according to instructions in the User’s Manual.
10. Switch off the system.
11. Disconnect the mains power cable.
12. Disconnect the gas supplies (wall and/or cylinder).
13. Disconnect the following items:
patient tubing
gas sampling tube
water trap
manual breathing bag
bacteria filter (-s) from the patient cassette, if any.
WARNING! With mains power supply connected to the unit, there are energized electrical
components inside the equipment. All personnel must exercise extreme caution when in the
vicinity of this equipment if maintenance, fault tracing or adjustments are performed with mains
power supply connected and with the covers/doors opened or removed.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 339 (of 369)
Chapter 7 – Preventive maintenance
CAUTION: The anesthetic gas and the calibration gas contain substances that may be
detrimental to your health. During use of such gases, assure that the system is connected to an
effective gas evacuation system, e.g., the hospital’s AGSS system.
Ref.no Description
1 EVAC reservoir
2 EVAC indicator
3 Hytrel tubing
CAUTION: Ensure that the EVAC system is correctly connected after any service intervention.
Running the system with a poor connection may result in anesthetic agent being emitted into the
operation environment. Leakages in the EVAC system will not be detected by the System
checkout.
Page 340 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 341 (of 369)
Chapter 7 – Preventive maintenance
Page 342 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Nozzle unit
Replace the nozzle unit on all gas modules.
Refer to section 4.15.4, Nozzle unit – replacement,
on page 175, for instructions.
Optional, perform a leakage test.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 343 (of 369)
Chapter 7 – Preventive maintenance
Only for systems with Gas distribution section of Version A (S/N 4000 and below):
Check the position of the Reflector gas selector, see section 4.15.4, step 5, page 175.
Version B
Version A
Page 344 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Fan filter
Remove and replace the fan filter.
Refer to section 4.8, Fan filter – on page 156 for
instructions.
EVAC filter
Remove and replace the EVAC filter.
Refer to section 4.9.1, Filter – replacement, page
156 for instructions.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 345 (of 369)
Chapter 7 – Preventive maintenance
Page 346 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Absorber valves
Remove and replace the absorber valves (ref. no.
1).
Refer to section 4.11.2, Absorber valves – removal
on page 160 and section 4.11.3, Absorber valves –
installation on page 160 for instructions.
Optional, perform a leakage test.
Necessary tools:
4 mm hex key
Follow this procedure:
1. Pull off the gasket.
2. Insert the hex key to remove the filter holder.
3. Pull out the filter holder and filter.
4. Replace the filter.
5. Mount the filter holder.
6. Mount the new gasket.
Note: The illustration shows the PISS (pin index
safety system) cylinder connection standard.
Other standards include DIN, which also has a filter
that shall be replaced.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 347 (of 369)
Chapter 7 – Preventive maintenance
Page 348 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 349 (of 369)
Chapter 7 – Preventive maintenance
Page 350 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
Check the service log Test results for the System checkout (performed during Gas analyzer
check/calibration), see work step 9, Gas analyzer section. The value can be found in the Flow transducer
test.
A measured Fresh Gas Pressure at 60 l/min that exceeds 10 cm H2O indicates a clogged Fresh gas
filter. The filter should be replaced. In the log example below the measured value is 7.86 cm H2O.
2017-05-26 14:54:48 FIN Fresh gas filter test Result Passed
2017-05-26 14:54:48 MEA Measured O2 flow 0.99 l/s]
2017-05-26 14:54:48 MEA Fresh Gas Filter pressure drop at 60 l/min 1.44 cmH2O
2017-05-26 14:54:48 MEA Fresh Gas Pressure at 60 l/min 7.86 [N/A, 11.00] cmH2O
2017-05-26 14:54:46 MEA Measured O2 flow 1.49 l/s
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 351 (of 369)
Chapter 7 – Preventive maintenance
WARNING! Make sure that all mounted arms are properly secured, including friction joints for
adjustment of swivel, tilt and height. Where applicable all mounting screws must be tightened
with the correct torque, see section 5.11, Adjustment of friction joints, for details.
Page 352 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 7 – Preventive maintenance
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 353 (of 369)
Chapter 7 – Preventive maintenance
capacity meter must be reset for each cassette. Enter the cassette S/N in the text field for each
cassette in Service report > Replaced expiratory membrane.
o Replaced backlight. Resets the backlight operating time counter on the Status menu.
o Replaced internal battery. Updates the replaced date for power backup battery on the Status menu.
o Replaced memory backup batteries.
26. Save system log files.
27. Make a note in the customer’s log book that Preventive maintenance has been performed.
28. Inform the customer that Preventive maintenance has been performed.
The power backup battery shall be replaced every three or four years, depending on model.
Refer to section 4.19, Power backup battery – replacement, page 209.
Note: When removing or installing the battery, be careful not to short-circuit the battery.
Note: The power backup battery replacement date must be reset after replacement. The replacement
date is set in Service & Settings > Service report > Replaced internal battery).
Page 354 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
8 Index
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 355 (of 369)
Chapter 8 – Index
Page 356 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 8 – Index
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 357 (of 369)
Chapter 8 – Index
Page 358 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 8 – Index
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 359 (of 369)
Chapter 8 – Index
Page 360 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 8 – Index
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 361 (of 369)
Chapter 8 – Index
Page 362 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
9 Revision history
System version 1.0 Not valid Replaced by System version 1.1 in mandatory update
MCC/10/009/NU.
System version 1.1 Not valid Replaced by System version 1.2 in mandatory update
MCC/11/001/IU.
System version 1.2 Not valid Replaced by System version 2.1 in mandatory update
MCC/11/006/IU.
System version 2.0 Not valid Replaced by System version 2.1 in mandatory update
MCC/11/006/IU.
System version 2.1 Valid only in Replaced by System version 4.2 in the mandatory update
countries where MCC/16/001/NU in all countries where System version 4.2 is
System version 4.2 regulatory approved.
is not yet regulatory
System conditions:
approved
Control Gas Analyzer CGA and Patient Gas Analyzer
PGA installed and active.
Double sampling used for CGA.
System checkout valve deactivated. Not mounted on
units with S/N 1500–2000.
System version 2.2 Valid only in Replaced by System version 4.2 in the mandatory update
countries where MCC/16/001/NU in all countries where System version 4.2 is
System version 4.2 regulatory approved.
is not yet regulatory
Software improvements:
approved
New system software installer interface.
The APL calibration values are not lost during System
software installation.
More robust monitoring of the Power Backup Battery.
Increased MAN/AUTO switch reliability.
Improved EEPROM write algorithm.
Possibility to reset the EEPROM on PC 1900 in Service &
Settings.
Improved Service & Settings functions.
Flow-i Communication Interface (FCI) Protocol version
0004 introduced.
Other changes compared to System version 2.1:
Improved Fresh Gas Sampling IFGS (single sampling)
containing:
o Fresh Gas Filter.
o Improved gas sampling point S2.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17 Page 363 (of 369)
Chapter 9 – Revision history
System version 3.0 Valid only in Replaced by System version 4.2 in the mandatory update
countries where MCC/16/001/NU in all countries where System version 4.2 is
System version 4.2 regulatory approved.
is not yet regulatory
Software improvements:
approved
System software installer improvements.
Vaporizer software can be installed without using an USB
memory stick.
Remote Services is implemented.
Preventive Maintenance counter is now reset to 5000
hours when checking and saving with the “Performed
preventative maintenance”-button in the Service &
Settings menu.
The Units menu in Service & Settings is allowed for
changes from the access level Biomed.
Stabilized TE16 alarm. I2C messages are sent twice
before triggering the TE16 alarm.
Flow-i Communication Interface (FCI) Protocol version
0005 introduced.
Other changes compared to System version 2.2:
Control Gas Analyzer CGA not used:
o For Flow-i field upgraded to System version 3.0, CGA
will remain mounted but will be disabled. CGA
sampling and return points S1/S2 and R1/R2 will
remain connected but not used.
o For Flow -i delivered from factory with System
version 3.0, CGA will not be included. CGA sampling
and return points S1/S2 and R1/R2 will be plugged.
Note that the Fresh gas filter introduced with IFGS (in
System version 2.2) is required even though the
improved sampling point S2 is not used.
System checkout valve reintroduced. Units with S/N
1500–2000 were manufactured without the System
Checkout Valve PV6 and the System Checkout Pilot
Valve EMV3. These units must be retrofitted with a
System checkout valve kit prior to System version 3.0
upgrade.
System version 4.0 Valid only in Replaced by System version 4.2 in the mandatory update
countries where MCC/16/001/NU in all countries where System version 4.2 is
System version 4.2 regulatory approved.
Page 364 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 9 – Revision history
System version 4.1 Valid only in Replaced by System version 4.2 in the mandatory update
countries where MCC/16/001/NU in all countries where System version 4.2 is
System version 4.2 regulatory approved.
is not yet regulatory
Major software improvements related to service:
approved
Conditions for displayed measured flow and volume
values adapted to BTPS.
Improved vaporizer stability.
Vaporizer 1 and 2 test will not be performed if the Gas
analyzer test fails during System checkout.
Flow-i Communication Interface (FCI) Protocol version
0007 introduced.
Other changes of the system:
From S/N 4001:
o New design of Gas distribution section and Drive gas
supply section.
o New design of Emergency manual ventilation section.
From S/N 5001:
o New design of Patient Cassette Docking.
o New design of CO2 Absorber locking switch.
Other hardware changes:
New design of Display/Touch screen and PC 1924
Display Connection from Control Panel S/N 320000.
New design of PC 1903 External Connectors due to new
LED Lamp. Introduced in production Q2 2015.
Changed fuses on PC 1938 Power Connection.
Auxiliary power outlets, US version new option.
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 365 (of 369)
Chapter 9 – Revision history
Page 366 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 9 – Revision history
A B C Section Comment
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 367 (of 369)
Chapter 9 – Revision history
Note: Revision 17 is created only for editorial changes on the cover page.
Page 368 (of 369) © 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17.
Chapter 10 – Diagrams
10 Diagrams
© 2019, Maquet Critical Care AB. Flow-i Anesthesia System Service Manual. Publ. No. 6676980, rev. 17. Page 369 (of 369)
7
1 N7 N8
PC1781
A10
A10 A1 A1
MODUL1
B10 B1 B1
C10 C1 C1
C10
R1 R2
C3 C4
C8
R5
C7
A1 B1 C1
N6A
1
P11
C1
P17
P15
P16
1 1 1
P14
2
1
C1 C32
B1 B32
A1 A32
N1B N1A
C2
IC1
C1 C32
1 2 B1 B32
A1 A32
N2B N2A
P18
C1 C32
B1 B32
A1 A32
N3B N3A
N6B
A1-H1
A2-H2
C1 C32
B1 B32
A3-H3
A1 A32
N9 N10
N4
A4-H4
A10 A1
B10 B1
C1 C32
B1 B32
A1 A32
R3 R4
N5A
C5 C6
Z4 Z8 Z12 Z16 Z20 Z24 Z28 Z32
N5B Pb
PC1900E
200 205
---------
PC1900
G
F
1
2
E
P31
1
2
D
P32
GETINGE GROUP
C
A
8
F
2
Ben{mning/Denomination
Typnr/Type No
E426E
B
Skala/Scale
1:1
2 P29 16
PCB Assembly
Blad/Sheet
2(2)
PC1900
Artikelnr/Part No
A
Bottom side view
66 72 837
Rev./Ver
06
1
6
7
2
4
5
Critical Care AB
Critical Care AB
Critical Care AB
Critical Care AB
Service & Settings
System version 4.7
User category: Field service engineer (SSU/SSP)
Access code: Service key
Page 1 (2)
Overview
Service
Status Logs Tests Calibration Service report
functions
System Gas analyzer Hardware Software Options LOGS System checkout: APL knob barometer Touch screen Valve tests Gas analyzer SERVICE TASKS:
Gas analyzer
-Alarms -O2 flush -Performed preventive
Central gas pressure PATIENT System item System item Installed software -Ventilation settings -Insp. & Exp. valves -1. Set APL value to SP -Current barometer - Calibration process for Touch Screen calibration System Valves - Connect maintenance
-O2 Software version PC board name Version no. options. -Events -Internal tests -2. Set APL Value to 30 pressure the Patient gas analyzer - Interface for calibration -Manual to AION through COM2 -Replaced expiratory
-Air Article no. -Vaporizer 1 -Barometer -3. Set APL Value to 80 -New barometer AION Platinum of the touch screen Ventilation(EMV2) membrane
AION -Replaced backlight
-N2O Version no. -Vaporizer 2 SERVICE LOGS -Gas supply pressure pressure -Fresh gas Safety (SV1)
.Type -Replaced internal
Serial no. -Test results -Pressure transducer -AFGO (EMV18)
Backup gas pressure -Serial number
-Technical alarms -Safety valve -PEEP/APL battery
-O2 -Hardware revision
-Configuration -Vaporizer inlet/outlet -System checkout (EMV3) -Reset internal battery
-Air -Operating time
-Installation valve errors
-N2O -Gas measurement bench
-Service report -Flow transducer Vaporizer valves -Replaced memory
Vaporizer 1 -Serial number
-AUTO ventilation -Vaporizer 1 in/out backup batteries
-Agent -Hardware revision
leakage (EMV6)
-Liquid level -Operating time
-MAN ventilation leakage -Vaporizer 2 in/out Write comment and sign.
-Status -Pump module
-Gas analyzer (EMV5)
Vaporizer 2 -Serial number
-Battery -Vaporizer Bypass
-Agent -Hardware revision
-Vaporizer 1 (EMV4)
-Liquid level -Operating time
-Vaporizer 2 -Vaporizer 1 pressure
-Status -Calibration dates
-Technical alarms (EMV9)
Battery status -Factory
-Remaining time -Last calibration -Vaporizer 1 Scavenging
Control: (EMV10)
-Voltage -Last calibration check
-Control disable -Vaporizer 2 pressure
-Battery replaced O2 sensor vaporizers (EMV11)
Product -Serial number -AFGO valve -Vaporizer 2 Scavenging
Serial number -Last calibration -Control pressure (EMV12)
transducer
System version -APL/PEEP valve
System software version -Control safety valve
Operating time -Vaporizer inlet/outlet
valve
Next preventive -Control flow transducer
maintenance -AUTO ventilation
Patient cassette status leakage
- APL/PEEP valve -MAN ventilation leakage
membrane -Vaporizer 1
-Remaining capacity -Vaporizer 2
-First day of use
-Operating time Monitoring:
-Initial gas supply
Network status
-O2 flush
-Remote Services URL
Absorber switch
-IP address
-Water trap
-Netmask
-Mains connection
-Gateway
-Monitoring disable
-DNS server
vaporizer
-Network address
-Power failure
-Network interface status
-Barometer
Control panel status -Gas supply pressure
-Backlight operating time -Monitoring pressure
transducer
-Monitoring safety valve
-Monitoring flow
transducer
-Gas analyzer
-Battery
-Technical alarms
Panel:
-Panel audio
Leakage check:
-Multiple pressure
leakage
-Internal leakage
-AFGO leakage
-AUTO ventilation
leakage
-MAN ventilation leakage
Service & Settings
System version 4.7
User category: Field service engineer (SSU/SSP)
Access code: Service key
Page 2 (2)
Overview
General Units Screen layout Displayed Waveforms Alarm general Adult AUTO Infant AUTO Biomed Date & Agent Network
Ventilation and gas
measurements alarm limits alarm limits access code time consumption
-Date format - Airway pressure -Loops Ventilation Waveform 1: -Alarm sound 1(2) 1(2) -Current code Date -ISO -Remote Services
-Time format - Flow -Gas mix measurements -Pressure level Lower/ Upper Lower/ Upper -Enter new -Day -SEV URL
-Decimal -Dynamic rotameter and placements Waveform 2 -20-100% -Ppeak -Ppeak code -Month -DES -Address setting
separator compliance -Numerical gas -Flow -Apnéa mute -MVe -MVe Startup Startup 1(6) Startup 2(6) Startup 3(6) Startup 4(6) Startup 5 and -Confirm new -Year Reset - IP address
-Language - CO2 pressure Page 1 -Volume -Heart Lung -RR -RR -Start Patient range Ventilation Pause Gas delivery 6 (6) code Time - Netmask
-Medical gas - Gas supply -Graphical gas Page 2 -CO2 Machine/Cardio -PEEP -PEEP Infant age mode: function: settings Summary of -Hour - Gateway
colour code pressure pressure Enable MAC -O2 Bypass mode Apnea time Apnea time Infant, VT/ Infant pause settings -Minute - DNS server
-Remote services - Barometric -Short trends presentation -Agent -Abbreviation PBW Parameter period -Second
-Flashing Standby pressure -Volume reflector -Blank -HLM 2(2) 2(2) Adult age Widgets Pause - Network
Message indicator (VRI) Waveform 3 -CPB Lower/ Upper Lower/ Upper Adult, VT/ for chosen function: Interface status
-Prediceted body -Flow -FiCO2 -FiCO2 PBW ventilation Adult pause
weight (PBW) -Volume -EtCO2 -EtCO2 Volume mode period
-CO2 -FiO2 -FiO2 setting Vaporizer
-O2 -EtO2 -EtO2 Breath Cycle slot
-Agent -FiISO -FiISO setting Activate N2O
-Blank -EtISO -EtISO Activate Activate N2O
Waveform 4 -FiDES -FiDES insp./exp cylinder
-Flow -EtDES -EtDES hold check during
-Volume -FiSEV -FiSEV PBW-based System
-CO2 -EtSEV -EtSEV values used checkout
-O2 automaticall Activate air
-Agent Reminder to turn Reminder to turn y cylinder
-Blank agent on agent on check during
Waveform 5 System
-Flow checkout
-Volume Activate O2
-CO2 Guard
-O2
-Agent
-Blank
Waveform 6
-Flow
-Volume
-CO2
-O2
-Agent
-Blank
Sweep speed
-5 mm/s
-10 mm/s
-20 mm/s
A B C D E F G
30 Auxiliary O2 and
1931 P50
Vaporizer 1 Vaporizer 1 Safety fill
scavenging pressure suction module 18 Control panel
MOD1
Serial com. 2
Backup O2 supply To EVAC Vaporizer valve
REG21 O2 120 kPa T
heating foil
I2C Vap 2
O2 240 kPa
1
TFT / LED display
1
IN Profile
Vaporizer PC boards
inlet filter System From PV 1 Gas escape
AUTO checkout Vaporizing Touch screen
EMV5 EMV11 EMV12 EMV4 EMV6 EMV3 EMV10 EMV9 pipe filter CPU T Start
MAN
valve PV1 Vaporizer chamber Stop
Reset
actuator inlet valve
O2 cylinder O2 cylinder SV21 I2C-Onboard
"8051"
CPU
gas pressure gas pressure 650 kPa EMV6 Trends Vaporizer LEDs
transducer regulator O2 cylinder E2Prom ”000”
Vap 2 Vap 2 Vap 2 Vaporizer Vap 1 System Vap 1 Vap 1
PV2 Vaporizer
Save End
gas safety selection pressure scavenging bypass selection checkout scavenging pressure outlet valve Screen Case From O2 P217 PC 1927 O2 Flush P216
pilot valve valve valve pilot valve pilot valve pilot valve valve valve
OUT
valve T Vaporizer PC 1925 Flush Valve
P56 P60 P61 P57 P55 P58 P66 P65 To PV 2 liquid Display CPU
PC1910 Power
PC 1912
PC 1913
Pc 1916 Exp. pressure transducer pilot Vaporizer P10 Screen
pressure to E2Prom ”001” valve Config.
Menu Switch
Cylinder pressure connection Vap = Vaporizer injector Silencer
P149 P151 P154 P153 P152
P70
PC 1781 (F) I/O port “001”
E2Prom “111”
on PC 1984 EMV2
PC 1781
Internal Boot enable switch Mains LED
MOD1
N259
ADC ”00”
P50
Pe gas channel P
PC 1911
Liquid agent P1 P4
34 Patient Monitor Backup Air supply From I/O port “000” PP3
SV2 Vaporizer filter
Suction P2
P67
REG22 Patient P3 P8 PC 1928 Power On/Off
Manual Manual PC 1924 Display connection
E
To P42 on PC 1903 Vaporizer pressure
Cylinder gas 310 kPa cassette vent. ventilation Drive gas O2 (from O2 Out) AIR (from AIR Out) N1A
pilot safety valve transducer
P59
inlet filter External connectors valve valve 240 kPa Drain plug
ADC ”01” P68
actuator
N1B P1B P1A
Serial com. 2
Fresh gas Vaporizer valve trap pressure Pi
bypass valve T W2 W2
heating foil
I2C Vap 2
Vaporizer to vaporizer from Patient
W3 W3 cassette NTC
PV1 inlet valve PV3 IN
27 Lift
Purge line
32 Backup gas holder
To Vaporizer 1 O2 P111
Vaporizing Gas escape Water Fresh gas Pf
AFGO pipe filter RS-232 (2) pressure
Pf PV3 Vaporizer chamber trap
valve CPU from Gas
Backup N2O supply To inlet valve detector Patient
RS-232 (4) control
AFGO
"8051"
CPU
REG23 EMV18 EMV5 O2 sensor section
Cylinder gas 310 kPa W1 P32
PV7 Vaporizer F I
Fresh gas PV4 Vaporizer
inlet filter outlet
from vaporizer outlet valve
Patient gas analyzer PC 1781 PC 1781
valve OUT Agent gas T Vaporizer P15 AION Platinum E2Prom “011” E2Prom “010”
From Vaporizer 1 liquid
PC1910
PC 1912
PC 1913
SV23 Vaporizer pressure pressure
650 kPa P81 P82 injector transducer transducer
To P42 O2 flush
N 2O PC 1909
on PC 1903 valve P110
cylinder
9 Patient cassette
To P217 on Vaporizer adapter
22 PC 1984
External gas safety P
N2O/O2 PC 1927 O2 Flush P30
lid
connectors Liquid agent
PC 1911
valve
SV2 Vaporizer filter Expiratory channel
Power 20 N30
Vaporizer
safety valve
pressure
transducer
R3
N2A
Connection Drain plug Pe
Expiratory pressure to N2B P2B P2A
6A PC 1781 (E) on PC 2057
AUT. 12 A
Isolation
Transformer
1 Gas distribution section 2 Drive gas supply section Flush
Fresh gas from vaporizer to patient cassette
RS-232
Mux
OV1-OV3 = Gas inlet One-Way valves P103 P45 E2Prom “000”
Instant O2 supply Emerg
OV4-OV6 = Backup gas Inlet One-Way valves
3 P225 PC 1938
Mains OV7 Pressure test point 2 T2
Reflector gas
P104
33 MSync CAN 1 CPU
I2C-Onboard
3 3
PC 1906 I2C- PC 1918
Connection PLD
Gas block Monitor Expiratory I2C-
P223 Pilot valve drive gas PV1
channel Control 48 MB Flash
16 System checkout valve
ADC ”00”
REG3 240 kPa SV3 connector
Pilot valve 300 kPa Vaporizer I2C- I2C 128 MB SDRAM
E2Prom “101”
N253 P249 O2 Out pressure Drive gas drive gas P46 Vap 1 Mux
V1 PP4 System checkout Pi 1 MB SRAM
(Auxiliary O2 ) regulator safety valve Inspiratory pressure to
valve PV6 and PC 1781 (I) I2C-
N233 T1 Test plug actuator PP4 Battery
on PC 1984 Vap 2
7 Safety
P OV1 N3A
Fuse F5 Pressure
N239
Fuse F1 transducer O2 4 Fresh gas supply 28 AFGO valve N3B P3B P3A
N243 Air out Barometric
Fuse F2 (Suction module) 1 SV1 CAN 1 NTC pressure
T P Additional fresh gas From AFGO in
19 Power section
Fresh gas safety transducer
P247 P222 outlet connector Gas control section valve 117 cmH2O
Serial Ser.
comm. 2 Mux PS1
S1 PLD
8 Patient cassette
OV5 CPU
Air In R1 LED Lamp 24 V Serial
Central Fresh gas comm. 3 E2Prom “000”
Pi RS-232
Filter
OV6
13 Reflector gas supply Exp.
chan.
PC 1903 AC/DC
converter
I2C-
Monitor
2 MB SRAM RTC
4
External
4 To AC/DC Power Inlet
in External connectors
To connector
P146 on PC2063 2
Reflector pressure
to PC1781(R) OV12 Expiratory
APL/
PEEP N46 12 V connectors I2C-
Battery
4
Power Control T P on PC 1900
Pr
outlet one-way valve valve E Prom “000”
2 CO2 P40 AC/DC
Power
PC 2061
PC 1917
24 V
P43 P6A N6A 24 Monitoring
Absorber
Extra mains 29 N 2O
Central Reflector
P17
APL/PEEP
Expiratory sensor
Cylinder
pressure P42
power outlet
Filter
channel N4
gas supply P OV3 gas module A
REG 6 I2C- valve coil B Air /O2 / N2O
Pressure O2 cassette HB ADC “10” P4
400kPa Valve HA
transducer N2O E Pot “0110” ”1110”
2 T1 T2 I C-
2
CAN 1
N2O gas supply To P67 on I/O Port" 000"
P1
1A N8 Monitor
pressure regulator Reflector PC 2057 120
AUT. 2A
adaptor Emergency T Valve drivers E2Prom " 110" CAN 2
Patient Monitor
APL actuator T CAN 1 120
CPU
I2C-
IEC
power
Gas module docking Expiratory flow transducer
Pe
Battery
sense
P38
120 P41
Panel
I2C-Onboard
outlets P18 P19
RS-232 (FCI 1) N137
P39 Fuse F1 P6B N6B E2Prom ”000”
PP3
APL/PEEP
Transformer 17 EVAC AGS flow Mechanical
valve coil
5 5
Switching indicator 10 APL
Fresh gas
14 Manual
Relay N5A
Emergency
Filter
T
1 2 34 module I2C- ventilation
Air N 2O Valve switch ventilation P5B P5A
E2Pot “0010” ”1010”
0
section + N5B
P1
N10
Power backup battery I2C- ADC “01”
Power
Oxygen flow
O2 l/min. ADC “00”
P142
A B C E F G
For revision history,
Flow-i Anesthesia System S/N <20 000 For information regarding other versions of this functional unit, refer to chapter 3. Description of functions in the Service Manual. EVU-123791_16
see EVU-191414 System Overview
A B C D E F G H
XXX- = Internal serial
Xxxxx communication
26 PC 1900 Main Back-plane Reflector Pressure
From Reflector gas
Pr
To connector supply
Backup Gas Holder 32 P42 on PC 1903 Backup Gas Trolley/Rack Gas Control From AFGO Pilot Manual Ventilation switch APL Vaporizer (VAP1) Vaporizer (VAP 2) XXX Xnn = Connector on PC 1900 Main Back-plane
External Valve EMV18 MAN E2Prom “000” R
N2O/O2
Connectors O2 Cylinder Gas 31 Section 6 Vaporizer Drive Docking
PC Lid Gas Filter Valve
6 Vaporizer Drive Docking
PC Lid Gas Filter Valve
I2C- PC 1781
Pressure Regulator AUTO Safety Fill 1931 Sensor Safety Fill 1931 Sensor Monitor E2Prom “100”
Cylinder Gas Cylinder Gas
Pressure N2O/O2 Cylinder
Gas Pressure
Inlet Filter 3 P62 P70
Valve Valve
Control Panel 18 Reflector Pressure P50
PC1910 PC1910 Transducer
P64
Transducer Pin Index OV9 OV9 MOD1
Regulator Vaporizer Gas Vaporizer
P63
O2 Exp. Pressure Gas
Heating Foil Heating Foil
"8051"
REG23 Escape
1
REG21 Escape
"8051"
from Patient PC 1907
CPU
SV21 T T
Power On/Off and 21
CPU
310 kPa 310 kPa Valve Drivers Pipe Pipe
Cylinder Gas O2 In 650 kPa Cassette P60 P61 P65 P66 From Absorber Sensor
Filter Vaporizing Filter Vaporizing
Inlet Filter O2 Cylinder
Pin Index Gas Pressure
O2 Cylinder Gas
Safety Valve E2Prom ”001”
Pe P67 EMV2 PP3 Chamber IN Chamber IN
120
N5 Display/Touch Panel
Alarm
profile Vaporizer PC boards
N2O/O2 In SV23 N256 N257 N258 EMV3 PP4 Liquid PC 1914 PC 1926 PC 1914 PC 1926 Vaporizer LEDs
PC 1912
PC 1913
PC 1912
PC 1913
Liquid
650 kPa O2 O2 N2O AIR P57 I/O port “000” Container VI Container VI
I2C-
Onboard P8 Version B and C:
Trends
P217 PC 1927 O2 Flush P216
N2O/O2 Cylinder E
I2C Vap 1
PC 1916 P58 With LED
I2C Vap 2
P56 ADC ”01” Vaporizer
Gas Safety Valve Cylinder Pressure Connection PC 1781 Vaporizer E2Prom ”000” Save End
EMV4 PV5 Injector screen case
From O2 From
N2O connected in diagram P55 P69 E2Prom “111” Injector
AIR/N2O AIR/N2O Liquid Agent P SV2 Liquid Agent P SV2 PC 1925 Display CPU Flush Valve Emergency Power
button
PC 1911
PC 1911
P57 Ventilation
N2O/O2 Cylinder Gas Cylinder Gas
Pressure Pressure
P54 I2C-
Panel
I2C-
Monitor
P50
MOD1 PV3
Filter
Vaporizer Vaporizer Filter Vaporizer Vaporizer P9 P10 Version A: Switch Power
P53 EMV5 PV4 Drain Pressure Safety Valve Drain Pressure Safety Valve Backlight Inverter On LED
Transducer Regulator I2C- Plug Transducer Plug Transducer P29
P68 Version B:
Patient Monitor Control P14
P56 PC 1963 Screen
Menu
P259
Pressure Supply Pilot Valves: N30
650 kPa EMV2 = Manual Ventilation Pilot Valve
EMV7 PP1 Valve PC 1928 Power On/Off
AIR/N2O Cylinder EMV3 = System Checkout Pilot Valve
Gas Safety Valve EMV4 = Vaporizer Bypass Pilot Valve P54 EMV9 P81 P30 P82 N1A
Vaporizer 1
AIR connected in diagram EMV5 = Vaporizer 2 Selection Pilot Valve EMV8 PP2 Scavenging PC 1909
2
EMV6 = Vaporizer 1 Selection Pilot Valve P66 Vaporizer Adapter N1B P1B P1A
Valve
AIR/N2O EMV7 = Vaporizer 1 Lock Pilot Valve PV1 PV2 PV3 PV4
EMV8 = Vaporizer 2 Lock Pilot Valve P53
Gas Analyzer Section 10
PV3
PV4
PV1
PV2
CAN 1 P108
EMV10
Vaporizer 2 I2C- I2C-Onboard
To connector P42 on PC 1903 External Connectors PV2 E1 Pressure Water trap CPU E2Prom “000”
P60 Exp.
Valve receptacle Zero calibration chan.
PV5 W2 gas inlet
Instant O2 Supply Pressure T4 I2C-
V2 E2Pot “01”
Lift 27 Gas Distribution Section 1 REG4
200 kPa
Pressure Regulator TestPoint 4 EMV11 Vaporizer 2
Scavenging
PV5
W3
Patient Gas Analyzer AION
ADS
Monitor
Version B P61 Vaporizer Bypass Valve Water 0
Inspiratory
Valve trap
REG3 Pilot Valve Pressure W3 Pressure Pi
Lift O2 200 kPa detector from Patient
Regulators REG2/REG3
Sampling line
actuator E2 NTC
Pressure Test EMV12 Water Patient Cassette
Purge line
PC 1906 Drive Gas Drive Point 2/3 Fresh Gas Filter O2 sensor O2
P111
Gas Block Safety Valve Gas trap P15 Fresh Gas Pf
T2/T3 Pressure
OV7
AC/DC
I2C- ADC ”00”
REG2
Fresh Gas to Vaporizer Vaporizer Valve Fresh Gas from Vaporizer W2 from Fresh
SV3 240 kPa Gas Supply
Monitor E2Prom “101” 200 kPa Pilot Valve Section 5 R3
Return line
Converter PV1
Drive Gas
O2 Out
AIR W1 P31 F I
OV8 Vaporizer Pressure Vaporizer Version A P32 PC 1781 PC 1781
REG1 Regulator Drive Gas E2Prom “011” E2Prom “010”
Cylinder Gas 120 kPa
Supply OV4 V1 Exp. Pressure Insp. Pressure Zero calibration Patient Gas Analyzer AION P50 P50
Pressure Test To PC 1781 on gas inlet MOD2 MOD1
Point 1 To PC 1781 Platinum
T1 on PC 1907 PC 1919/PC 1984 Insp. Pressure
Fresh Gas Pressure
Power 20 O2 In
Gas Inlet Filters
and Backup Gas
Pressure
Transducer Version B
R3
P1 Transducer Transducer
Connection Inlet Filters O2 2 Drive Gas Supply Section P2
P110
PC 1919/PC 1984
0 Exp. Channel
3 22
6A Central Gas Exp.
Pi
AUT. 12 A Supply P OV1 Reflector Gas Insp.
W2
N2A
Sample
Isolation 1 T P Reflector Gas Supply 13
Pe
Filter Sample W3
O2 Flush
N7 on PC 1907 P104
PC 1918 Version B PLD
PV7
N253 P249 Pressure Valve Drivers
Exp. Channel 19 Power Section I2C-
Gas Inlets / Gas Outlets
4
P11 12 V CAN 1
N2O In Gas Module Docking P40 NTC pressure
Reflector Cylinder Serial transducer
Absorber Sensor
Filter
Auxiliary O2 P17
APL/PEEP 11 P = Pressure measuring point:
E2Prom ”000”
Input/Output ports
2A Silencer RS-232 (FCI 2) N136 ADC ”00”
Valve Coil PP3 Pe = Expiratory pressure
AUT. 4 A LVDS
Power Oxygen Pf = Fresh gas pressure N135 64 MB Flash
PGA Gas
OV13 Oxygen Flow
CO2 Pi = Inspiratory pressure Power backup battery
PP3
Outlets Connector analyzer 128 MB SDRAM
Block One-way Valve Pr = Reflector pressure FLOW-i VGA N134
RJ45
Suction
zero calibr.
gas intake Manual Absorber R = Gas return point Remote
Services USB N133
2 MB SRAM
Mechanical Ventilation S = Gas sampling point
O2 (from O2 Out) AIR (from AIR Out) Valve Adapter Battery
Power Out
10 APL S1-R1 = Not used N132
Power In Emergency To connector
Ventilation
Actuator
P67 on PC 1907
S2-R2 = Not used, not shown
S3-R3 = Patient Gas Analyzer
RJ12 (RJ45)
Ethernet
PC 1922 Panel 25
1 2 1 3 Switch Valve Drivers S3
0 Y-piece measurement
AGS Flow
EVAC 17 0 Emergency T = Temperature measuring point 33 N5A
Indicator APL Actuator W1 - W3 = Water trap connections
I
Oxygen Flow
MSync N5B P5B P5A
6 Switching
Relay
Transformer
EVAC Reservoir below VAP2)
Bleed-flow Device
Version B
Manual
Ventilation AFGO
PV7
PATIENT +12V RS-232 LAN
Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and for life
science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts, healthcare
professionals and medtech specialists, we are improving the everyday life for people - today and tomorrow.
www.getinge.com