Systec D-Series: Bench-top autoclaves.

Systec D-Series: Horizontal front loading bench-top autoclaves.


These autoclaves don't just seem to be different; they are different. Specially developed for sterilization procedures in the laboratory,
processes become easier, safer, more precise, more reproducible and validatable.

Systec D-23 Systec D-45 Systec D-65 Systec D-90

Dimensions and performance

Systec D-23 D-45 D-65 D-90 D-100 D-150 D-200

Chamber dimensions
Ø x depth in mm 260 x 420 344 x 500 400 x 500 400 x 700 500 x 500 500 x 750 500 x 1000
Chamber volume
in liters total / nominal 25 / 23 50 / 45 70 / 65 95 / 90 110 / 100 160 / 150 210 / 200
External dimensions in mm
W 555 618 750 750 850 850 850
H 500 550 930 630 730 730 730
D 650 740 770 970 810 1050 1300
Heating power kW
Systec DX 2.80 2.80 9.00 9.00 9.00 9.00 9.00
Systec DE and DB 2.30 3.00 4.50 4.50 4.50 7.25 7.25
Electrical connections for Systec D-23 and D-45: 220 – 240 V AC, 50/60 Hz, 16 A
Electrical connections for Systec D-65 – D-200: 380 – 400 V, 50/60 Hz, 3-phase to neutral, 16 A
Other voltages and single-phase operation available on request

Space advantage!
A compact instrument with ideal chamber
size in seven sizes. Each size has an optimal
loading capacity for most standard media
bottles and Erlenmeyer flasks.

Page_2 autoclaves_Systec_D-Series
A new dimension in technology and design.

Systec D-100 Systec D-150 Systec D-200

Loading capacity*

Erlenmeyer flasks
Systec D-23 D-45 D-65 D-90 D-100 D-150 D-200
250 ml 11 24 23 31 2 x 30 2 x 42 2 x 59
500 ml 8 12 15 21 2 x 15 2 x 24 2 x 40
1000 ml 3 8 9 13 12 18 23
2000 ml - 3 6 8 7 9 14
3000 ml - – 3 4 6 8 11
5000 ml - – – – 3 5 7

Loading capacity*

Schott–Duran media flasks


Systec D-23 D-45 D-65 D-90 D-100 D-150 D-200
250 ml 18 24 31 40 2 x 36 2 x 54 2 x 83
500 ml 10 18 23 31 2 x 26 2 x 40 2 x 59
1000 ml 4 10 15 18 18 26 40
2000 ml - 5 8 10 12 14 23
5000 ml - – 3 4 6 8 11
10000 ml - – – – 2 3 4
*At max. loading, partially without baskets

autoclaves_Systec_D-Series Page_3
Double-door (pass-through) autoclaves: Systec D-Series 2D.

Equipped with the same state-of-the-art technology as our horizontal front loading bench-top autoclave Systec D-Series,
DX the 2D series has the same modern design but now with two doors for fitting between separating walls.

Systec Systec DX-150 2D / DX-200 2D


DX-90 2D

View from the non-sterile side

Systec DX-90 2D Systec DX-150 2D Systec DX-200 2D

Chamber dimensions
Ø x depth in mm 400 x 750 500 x 750 500 x 1000
Chamber volume in liters
total / nominal 98 / 95 155 / 150 205 / 200

All types: heating power 9 kW, voltage 380 – 400 V, 50/60 Hz, 16 A.
Modifications to other voltage as well as single phase operation available on request.

The loading capacities are the same as for the Systec D-90, Systec D-150 and
Systec D-200 (see page 2).
View from the sterile side

Dimensions in mm
J Systec Systec Systec
Service side Clean room side Wall opening DX-90 2D DX-150 2D DX-200 2D
Wall frame
A 750 850 850
Wall opening

B 1030 1130 1130


I

C 1430 1530 1530


D 1570 1670 1670
Wall
D
C

E 240 260 260


H F max. 200 max. 200 max. 200
OKFF
K

G 900 900 1150


H 920 1020 1020
A
max.
I 780 880 880
B E F J 820 920 920
G
K 800 800 800
L 670 770 770

Page_6 autoclaves_Systec_D-Series
A new dimension in technology and design.

DX Triple safety aspects!


– One door only can be opened at any one View from the sterile side
time. If one door is open, the other is
automatically locked.

– If the autoclave is switched off or if no


current is available (e.g. power failure),
both doors remain locked.

– If the door at the non-sterile side is


opened, a sterilization program has to
be performed before the door at the
sterile side can be opened.

The locking system can be adapted to


customers’ wishes.

Doors and control panel are made of heat-


insulating plastic, the housing completely
of stainless steel and the edges specially
processed for smooth junctions with
walls.

Operation, however, can be carried out


from both sides, the position (open or
closed) of the opposite door being indicated
on the display.

For use under the most Pass-through autoclave in cross-section


build into dividing wall.
stringent clean room and
safety conditions.
– For biological safety laboratories.
Fitting as a sterilization and double-
door (pass-through) lock for protecting
the external environment.

– For clean rooms in laboratories and


production facilities as a sterilization
and double-door (pass-through) lock
separating sterile and non-sterile areas.

Clearly laid out control panel with


membrane display. Door operated by
an ergometric handle.

autoclaves_Systec_D-Series Page_7
Innovative design – to the last detail!

Guaranteed sterilization of
liquids
During the entire sterilization process the
temperature is measured by a PT 100
flexible temperature sensor in a reference
vessel. In this way it is guaranteed that
the sterilization process begins only when
the sterilization temperature of the liquid
has been attained. The cooling temperature
is also continuously monitored. As required
by the operating standards and to prevent
delayed boiling, the door can only be
opened when the temperature of the liquid
has cooled to at least 80 °C.

Page_12 autoclaves_Systec_D-Series
High, all-round quality!
The pressure vessel is made of corrosion-
resistant electro-polished stainless steel
1.4571 (V4A) AISI 316 Ti and is thus easy
to clean. The excess pressure release safety
valve is of the approved component type.
Autoclave framework and housing are also
made of stainless steel. The highly efficient,
high-quality Hanno-Tect insulation mate-
rial releases no particles; thus, the Systec
D-Series can be used under clean room
conditions.

Compressed air Interface RS 232/485


connection (optional)

Demineralized water Flexible power connection


feed for steam with CEE plug
generation

Cooling water feed

Collection outlet
(drain)

autoclaves_Systec_D-Series Page_13
Intelligent process engineering for quicker and safer
processes!
The actual sterilization time of e. g. 15 – 20 minutes at 121 °C is only a fraction of the total time required for the entire
sterilization process. Especially when sterilizing liquids, the heating up and cooling down times are far longer (fig. 1).

The conventional process. fig. 1


°C/bar
In conventional systems, the set steriliza-
= 10 min
tion temperature may in fact be reached
within the autoclave; however, the liquids
to be sterilized may only have reached
approx. 100 °C. The temperature equilibrium 121°C
2,1 bar
time between chamber and liquid temper-
ature is usually much longer.

The cooling process for liquids is also very 100 °C


slow as the heat under 100 °C can only be 1 bar
dispersed by convection via the chamber
insulation if there is no rapid cooling facil-
ity.
80 °C

t
H S C

Temperature H = Heating up phase


Pressure S = Sterilization phase
C = Cooling down phase

* The times indicated in the diagram are


dependent on the size and number of
items to be sterilized.

Page_14 autoclaves_Systec_D-Series
System accessories
for ease of handling!

Support Tables Loading trolley Stainless steel quality!


There is a special table for each of the Large autoclaves in particular can be easily All parts are made of stainless steel and
Systec D-Series autoclaves. These are and securely filled using a special loading cleanly welded. Bench-top autoclaves are
custom-dimensioned for the instrument trolley. The items to be sterilized can either fitted with adjustable leveling screws to
in question. The tables are 72 cm high be placed on the sliding platform of the ensure stability; the filling trolleys have
(other heights available at no extra trolley or in a filling basket. The trolley large rollers, two of them fitted with
charge) and make for ease of handling. can now be moved to the autoclave and brakes, to ensure smooth running.
They have a practical shelf e. g. for fixed in position by a clamp. The clamp
storing baskets etc. can then be loosened to allow the plat-
form to slide into the autoclave on fixed
rails.
.

Loading shelves
To fully utilize the available space in the
chamber, especially when sterilizing small
items, every autoclave can be fitted with
a set of loading shelves. The entire shelv-
ing system or individual trays can be
removed.

Page_18 autoclaves_Systec_D-Series
Loading baskets and inserts

Wire mesh baskets out of stainless steel, stackable


Inner Dimensions D-90 D-150 D-200
Art. No. L x W x H (mm) D-23 D-45 D-65 DX-90 2D D-100 DX-150 2D DX-200 2D
Hold capacity per autoclave
1552 390 x 168 x 132 1
1553 490 x 265 x 180 1
4072 490 x 310 x 210 1
5074 688 x 310 x 210 1
6071 490 x 360 x 290 1 2
6072 490 x 360 x 140 2 4
7071 355 x 360 x 290 2
7072 735 x 360 x 290 1
7075 355 x 360 x 140 4
7076 735 x 360 x 140 2

Closed tubs out of stainless steel for waste sterilization


Inner Dimensions D-90 D-150 D-200
Art. No. L x W x H (mm) D-23 D-45 D-65 DX-90 2D D-100 DX-150 2D DX-200 2D
Hold capacity per autoclave
1554 395 x 180 x 135 1
1555 495 x 265 x 180 1
4073 495 x 318 x 219 1
5075 696 x 318 x 219 1
6070 495 x 368 x 300 1 2
7070 368 x 368 x 300 2
7073 747 x 368 x 300 1

Our Quality
Assurance
Each and every component used is subjected
to stringent controls and each autoclave is
thoroughly checked for function before
being delivered. An acceptance protocol is
issued.

Step-by-step documentation
optionally available!
Our Quality Management fulfills the high-
est requirements with regard to testing and
documentation. Within the scope of our
service, we also provide qualification and
validation work with GMP-compliant docu-
mentation, e. g.:
– DQ Design Qualification
– IQ Installation Qualification
– OQ Operation Qualification
– PQ Performance Qualification

autoclaves_Systec_D-Series Page_19