9031 Sulfide
9031 Sulfide
EXTRACTABLE SULFIDES
2.1 If the sample contains solids that will interfere with agitation
and homogenization of the sample mixture, or so much oil or grease as to
interfere with the formation of a homogeneous emulsion in the distillation
procedure, the sample may be filtered and the solids extracted with water at pH
> 9 and < 11. The extract is then combined with the filtrate and analyzed by the
distillation procedure. Separation of sulfide from the sample matrix is
accomplished by the addition of sulfuric acid to the sample. The sample is
heated to 70EC and the hydrogen sulfide (H2S) which is formed is distilled under
acidic conditions and carried by a nitrogen stream into zinc acetate gas
scrubbing bottles where it is precipitated as zinc sulfide.
3.0 INTERFERENCES
3.4 Interferences have been observed when analyzing samples with high
metal content such as electroplating waste and chromium containing tannery waste.
4.3.3 Purge gas inlet tube - 24/40 joint with course frit.
4.3.4 Purge gas outlet - 24/40 joint reduced to 1/4 inch tube.
4.3.5 Gas scrubbing bottles - 125-mL, with 1/4 in. o.d. inlet and
outlet tubes. Impinger tube must not be fritted.
4.7 Flowmeter.
5.0 REAGENTS
5.9 Zinc acetate for the scrubber (approximately 0.5M). Dissolve 110
g zinc acetate dihydrate in 200 mL of water. Add 1 mL concentrated hydrochloric
acid, HCl, to prevent precipitation of zinc hydroxide. Dilute to 1 liter.
5.14 Titrant.
6.1 All samples must have been collected using a sampling plan that
addresses the considerations discussed in Chapter Nine of this manual.
6.2 All samples must be preserved with zinc acetate and sodium
hydroxide. Use four drops of 2N zinc acetate solution per 100 mL of aqueous or
multiphasic sample. Adjust the pH to greater than 9.0 with 50% NaOH. Fill the
sample bottle completely and stopper with a minimum of aeration. For solid
samples, fill the surface of solid with 2N zinc acetate until moistened. Samples
must be cooled to 4EC during storage.
7.0 PROCEDURE
7.1 Assemble the Buchner funnel apparatus. Unroll the glass wool and
fold the fiber over itself several times to make a pad about 1 cm thick when
lightly compressed. Cut the pad to fit the Buchner funnel. Dry and weigh the
pad, then place it in the funnel. Turn on the aspirator and wet the pad with a
known amount of water.
7.3 Transfer the solid and the glass fiber pad to a dried tared
weighing dish. Since most greases and oils will not pass through the fiber pad,
solids, oils, and greases will be extracted together. If the filtrate includes
an oil phase, transfer the filtrate to a separatory funnel. Collect and measure
the volume of the aqueous phase. Transfer the oil phase to the weighing dish
with the solid and glass fiber pad.
500 mL water
5 mL 50% w/v NaOH
1 g SnCl2 C 2H2O
50 mL n-hexane (if an oil or grease is present).
Cap the bottle with a Teflon or polyethylene lined cap and shake vigorously to
saturate the solution with stannous chloride. Direct a stream of nitrogen gas
at about 10 mL/min into the bottle for about 1 minute to purge the headspace of
oxygen. If the weight of the solids (Step 7.4) is greater than 25 g, weigh out
a representative aliquot of 25 g and add it to the bottle while still purging
with nitrogen. Otherwise, add all of the solids. Cap the bottle; avoid the
influx of air.
7.7 Place the bottle in the tumbler, making sure there is enough foam
insulation to cushion the bottle. Turn the tumbler on and allow the extraction
to run for about 18 hours.
7.10 If the extract contains an oil phase, separate the aqueous phase
using a separatory funnel. Neither the separation nor the filtration are
critical, but are necessary to be able to measure the volume of the aqueous
extract analyzed. Small amounts of suspended solids and oil emulsions will not
interfere with the extraction.
b
From Step 7.4. Weight of solids and oil phase with the dry weight of filter and
tared dish subtracted.
c
Includes volume of all rinses added to the filtrate (Steps 7.1 and 7.2).
d
500 mL water plus total volume of NaOH solution. Does not include hexane, which
is subsequently removed (Step 7.10).
CAUTION: Toxic hydrogen sulfide may be generated from the acidified sample.
This operation must be performed in the hood and the sample left
in the hood until the sample has been made alkaline or the sulfide
has been destroyed.
From the amount of sulfuric acid required to acidify the sample and the
mass or volume of the sample acidified, calculate the amount of acid
required to acidify the sample to be placed in the distillation flask.
7.12.4 Place the dropping funnel onto the flask making sure its
stopcock is closed. Add the volume of sulfuric acid calculated in Step
7.1.1 plus an additional 50 mL into the dropping funnel. The bottom
stopcock must be closed.
7.12.5 Attach the nitrogen inlet to the top of the dropping funnel
gas shut-off valve. Turn on the nitrogen purge gas and adjust the flow
through the sample flask to 25 mL/min. The nitrogen in the gas scrubbing
bottles should bubble at a rate of about five bubbles per second.
Nitrogen pressure should be limited to approximately 10 psi to prevent
excess stress on the glass system and fittings. Verify that there are no
leaks in the system. Open the nitrogen shut-off valve leading to the
dropping funnel. Observe that the gas flow into the sample vessel will
stop for a short period while the pressure throughout the system
equalizes. If the gas flow through the sample flask does not return
within a minute, check for leaks around the dropping funnel. Once flow
has stabilized, turn on the magnetic stirrer. Purge the system for 15
minutes with nitrogen to remove oxygen.
7.13.3 Pipet both of the gas scrubbing bottle solutions into the
flask, keeping the end of the pipet below the surface of the iodine
solution. If at any point in transferring the zinc acetate solution or
rinsing the bottles, the amber color of the iodine disappears or fades to
yellow, more 0.025N iodine must be added. This additional amount must be
added to the amount from Step 7.13.1 for calculations. Record the total
volume of standardized 0.025N iodine solution used.
7.13.5 Rinse any remaining traces of iodine from the gas scrubbing
bottles with water, and transfer the rinses to the flask.
8.1 All quality control data must be maintained and available for
reference or inspection for a period of three years. This method is restricted
to use by or under supervision of experienced analysts. Refer to the appropriate
section of Chapter One for additional quality control requirements.
8.3 Check standards are prepared from water and a known amount of
sodium sulfide. A check standard should be run with each analytical batch of
samples or once every twenty samples. Acceptable recovery will depend on the
level and matrix.
8.4 A matrix spiked sample should be run for each analytical batch or
twenty samples, whichever is more frequent, to determine matrix effects. If
recovery is low, acid-insoluble sulfides are indicated. A matrix spiked sample
is a sample brought through the whole sample preparation and analytical process.
Accuracy for the entire method for four synthetic waste samples 70-104%
recovery
10.0 REFERENCES
1. Test Methods for Evaluating Solid Waste, Physical/Chemical Methods, 3rd ed.;
U.S. Environmental Protection Agency. Office of Solid Waste and Emergency
Response. U.S. Government Printing Office: Washington, DC,1987; SW-846; 955-001-
00000-1.
3. CRC Handbook of Chemistry and Physics, 66th ed.; Weast, R., Ed.; CRC: Boca
Raton, FL, 1985.
4. Standard Methods for the Examination of Water and Wastewater, 16th ed.;
Greenberg, A.E.; Trussell, R.R.; Clesceri, L.S., Eds.; American Water Works
Association, Water Pollution Control Federation, American Public Health
Association: Washington, DC, 1985; Methods 427, 427A, 427B, and 427D.
7. Bateson, S.W.; Moody, G.J.; Thomas, J.P.R. Analyst 1986, 111, 3-9.
10. Franklin, G.O.; Fitchett, A.W. Pulp & Paper Canada 1982, 83(10), 40-44.
11. Fuller, W. Cyanide in the Environment; Van Zyl, D., Ed.; Proceedings of
Symposium; December 1984.
12. Gottfried, G.J. "Precision, Accuracy, and MDL Statements for EPA Methods
9010, 9030, 9060, 7520, 7521, 7550, 7551, 7910, and 7911"; final report to the
U.S. Environmental Protection Agency (EMSL-CI); Biopheric.
15. Landers, D.H.; David, M.B.; Mitchell, M.J. Int. J. Anal. Chem. 1983, 14,
245-256.
19. Snedecor, G.W.; Ghran, W.G. Statistical Methods; Iowa State University
Press: Ames, IA, 1980.
20. Umaña, M.; Beach, J.; Sheldon, L. "Revisions to Method 9010"; final report
to the Environmental Protection Agency on Contract No. 68-01-7266; Research
Triangle Institute: Research Triangle Park, NC, 1986; Work Assignment No. 1.
21. Umaña, M.; Sheldon, L. "Interim Report: Literature Review"; interim report
to the U.S. Environmental Protection Agency in Contract No. 68-01-7266; Research
Triangle Institute: Research Triangle Park, NC, 1986; Work Assignment No. 3.
24. Application Note 156; Princeton Applied Research Corp.: Princeton, NJ.
25. Guidelines for Assessing and Reporting Data Quality for Environmental
Measurements; U.S. Environmental Protection Agency Office of Research and
Development: Washington, DC, 1983.
27. The Analytical Chemistry of Sulfur and Its Compounds, Part I; Karchmer,
J.H., Ed.; Wiley-Interscience: New York, 1970.
28. Methods for the Examination of Water and Associated Materials; Department
of the Environment: England, 1983.
29. "Development and Evaluation of a Test Procedure for Reactivity Criteria for
Hazardous Waste"; final report to the U.S. Environmental Protection Agency on
Contract 68-03-2961; EAL: Richmond, CA.
30. 1985 Annual Book of ASTM Standards, Vol. 11.01; "Standard Specification for
Reagent Water"; ATSM: Philadelphia, PA, 1985; D1193-77.