Basic Course Manual - WORKBOOK
Basic Course Manual - WORKBOOK
Basic Course Manual - WORKBOOK
INTRODUCTION 1
History
Definition
SONOGRAPHIC TERM INOLOGY 2
BASIC FORMULAS 2
J
CHARACTER]STICS OF SOU ND 4
GENERATIONS OF SOUND 5
Piezo-electric Effect
DigitalScan Converter
Grey Scale lnnaging
Beamformer
Scannlng Modes
Attenuation
Gain
Absorption
Reflection
Refraction
Scattering
Acoustic lmpedance
Tissue Echogenicity
ffi
ln collaboration with k5
I NSTITUTE OF DIAG NOSTIC M EDICAL SONOG RAPHY I NCORPORA"TED q1
IMAGE RESOLUTION 1L
COLOR DOPPLER L3
THE TRANSDUCER 13
Focus
Bandwidth
Frequency
Types of Transducers
. Frequency Selection
Scan Angle
Gain adjustment
Focus position
Dynamic Range
Frame Rate
Compound lmaging
IMAGE PROCESSING ?3
Pre-processing
Post-processing
TRANSDUCER MOVEMENT 25
BASIC SCANNING TECHNIQUES 26
IMAGE ORIENTATION 28
SAFETY OF ULTRASOUND 30
ROUTINE EQUIPM ENT MAINTENANCE 32
LI M ITATIONS OF ULTRASOUND 34
ffi
ln collaboration with
INSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPHY INCORPORATED
rH E t-t tsroEyllf uEBAsou N I
The use of ultrasound as a diagnostic modality is a relatively new practice. The idea
came after Langevin made use of a pulse-echo technique called SONAR {Sound Navigation and
Rangingi in 1916.
Later, in the late thirties, ultrasound was applied in industry by Firestone to detect metal
flaws present in equipment and machinery.
An Australian researcher named Dussik was the first to apply ultrasound to medicine.
He used the metal flaw detector to evaluate the cerebral ventricles for midline displacements
or defects.
ln 1949, Ludwig and Struthers irnproved the pulse-echo technique to detect foreign
bodies in soft tissue. Soon this practice was applied to the gallbladder, in the search for
gallstones.
- Howry and Bliss produced the flnst cross-sectiorral ultrascund innage in 1950. The idea
that this would be useful for imaging of the gravid uterus followed shortly thereafter.
lan Donald, in Scotland, was the first ta obtain sonographic images of evarian cysts and
BPDs in 1958"
ln the 40 years since Dussik's application sf ultrasound to rnedicine, the quallty of
ultrasonic evaluation has improved drarnatically to such a point that it is now considered an
integral part of diagnostic imaging. The relatively low cost, ease of examination, and absence of
significant bio-effects make ultrasound a fuvorable route to take among the many irnaging
rnodalities. As you will see, the ease with which one can master sonography makes this
modality integral to modern rnedical practice.
The human ear can hear sound in the frequency range of 20-20,000 Hertz {cycles per
second). Ultrasound is that sound which is at a frequency greater than 2Q&)0 Hertz. Diagnostic
Ultrasound operates in the range cf 1to 18 Megahertz.
Ultrasound travels in the forrn of a longitudinal wave, in which particle motion is along
the same direction as the rffave is traveling. These waves are generated by the ultrasound
transducer. Longitudinal waves transfer energy through the motion of regions of cornpression
and rarefaction within the wave. Because it is in $Jave form, ultrasound physics is governed by
many of the same principles as basic \n ave physics {see attached formulae).
1
PHILIPPIN[ SOCIEIYOT ULTRASOUND IN SURGERY
In collaboration with
I NSTITUTE OF DIACNOSTIC MEDICAL SONO6 RAPF.IY I NCORPORATED
,t
As the ultrasound bearn encounters tissues of different acoustic impedance, velocities
are altered such that neturning echoes are received by the transducer at different tirnes and
have different intensities. The differences in time and intensity are adjusted for by the user and
the computer within the systern" This inforrnation, along uvith the values for sound !\rave
velocities in tissues is used by the ultrasound device to generate the diagnostic image on the
rnonitor.
2.
BA$|C FOR[rUti$
3
PHILIPPISIE SOCIETY OF ULTRASOUND IN SURGERY
ln collaboratisn with
I NSTITUTE OF DIAGNOSTIC M EDICAL SONOGRAPHY I I{CORPORATED
Characteristics of Sou nd
Sound is a fsrm of mechanical energy that travels in a straight line through a conducting
mediunr {e.9., body tissue} as a iongitudinal wave pr*ducing alternating cornpression {high
pressr.*re) and rarefaction {low pressr.rre}. Sound propagaticr^r can be represented in a sinusoidal
waveforrn wlth a characteristic amplitude {A}, wavelength {U, frequency (f}, period {T} and
velocity lspeed {c} + directionJ.
LOW FREQUENCY
A
f = 1Hz
f = cycleslsec.
T
,1 A HIGH FREQUEI{CY
L
f=3Hz
fuurd is a. PrE*E rrE tYave
C
T
R L
tR C R C
t
R C
T
x.
!,
I
T I I I I
TiEE
fr
tri
4
PHILIPP}NE ST}C}ETY OF ULTRASOUND IN SUBGERY
ln collaboration with
I NSTITUTE OF DIAG NOSTIC MEDICAL SONOGRAPHY I NCORPORATED
The frequency cf an ultrasound wave is above 20,000 l'lz {or 20 KHz} and medical ultrasound
commanly is in the 2.5-X"8 MHz range. Hurnan hearing is in the 20-20,000 Hz range. The speed
of sound varies for different biological media but the average value is assumed to be 1,540
rn/sec (constant) for most human soft tissues. The speed of sound {ci can be calculated by
multiplying wavelength {}r} x frequency {f}. Thus sou*d with a high frequency has a short
wavelength and vice versa. For exarnple, the uravelength of a 2 MHz ultrascund wave = A.77
mm and that of a 15 MHz wave = 0.L0 mm.
"r., ,,i. i' The pulse length {PL} is the distance traveled per
-1. i
i . r I .
pulse. Waves of short pulse lengths improve axial
nasolution for ultrasound imaging. The PL cannot
be reduced to less than 2 or 3 sound cycles by the
damping materials within the transducer.
5
PHII-IPPINE SOCIETY OF ULTRASOUND IN SURGERY
ln callaboration with
I N STITUTE OF DIAGNO5TIC MEDICAI- 5ONOGRAPHY I NCORPORATED
PRF per time unit {S}
'li Lrf
PL PRP
Pulse Repetition Period {PRP} is the time from the start of one pulse to the start of the next
pulse.
Pul$e Bepetition Frequency {PRf} is the rate of pulses emitted by the transducer {number of
pulses per unit time). Ultrasound pulses must be spaced with enough time between pulses to
permit the sound to reach the target of interest and return to the transducer before the next
pulse is generated. The PRF for rnedical imaging ranges from t-10 kHz. For example, if the PRF =
5 kHz and the time between pulses is 0.2 msec., it willtake 0.1 msec to reach the target and 0.L
rnsec to return to the transducer. This means the pulse will travel 15.4 cm before the next pulse
is emitted {1,540 m/sec x 0.1 rnsec = 0.154 nr in S.1 msec = 15.4 cm}.
An ultrasound image is generated wiren the pulse wave emitted frorn the transducer is
transmitted into the body, reflected off the tissue interface and returned to the transducer. The
human body is composed of tissues of varying densities. Tissues of varying densities reflect
echoes of varying intensities. Dense or hand tissues like bones reflect high amplitude {strong)
echoes, softer tissr..les like the liver or kidney reflect lower arnplitude {weaker} echoes, while
fluids do not reflect any echoes at all.
The transducer transforms the echo {rnechanical energy} into an electrical signal. The stronger
the echo, the higher is the amplitude of the electrical signal. Weaker echoes are converted to
lower amplitude electrical signals {see Tissue Echogenicity}. The digital scan converter then
maps these {nange of} signals and assigns a nurnber frem 0 to 255 t255 erey levels}, where 0 is
mapped lplaced in the correct pasition an the pixel board - mapping) as black and 255 is
rnapped as white and all the nunrbers in between represent different shades of grey. This
process is called grey scale irnaging. These shades ef grey are then cornbined and displayed as
an image on the screen. {see also Dynarnic fronge o* page 22}
The position of the dots {shades of gray} on the monitor represents the depth from which the
returning echo was received. When the system hearnformer generates the electrical pulses, it
6
PHILIPPINT SOCIETY OF ULTRASOUhID IN SURGERY
ln coilabsration with
I NSTITUTE OF DIAGNOSTIC ME*ICAI- SONOGRAPHY I NCORPORATED
also starts the system clock and signals the start of transmission. The system then calculates
how long it takes for the echo to return to the transducer.
By using the following formula: Distanee = Velocity x Time,
Distance
where the average ve{ocity of sownd in tissue is assumed constant ot 754A mekrs / second,the
system can now determine the distance between the transducer surface and tissue interface.
Among the 4 mcdes, the B mode is rnost comrnonly used. There are 5 basic components of an
ultrasound scanner that are required for generation, display and storage of an ultrasound
image.
As the ultrasound beam travels thror:gh tissue layers, the amplitude of the original signal
becomes attenuated as the depth af penetration increases. Attenuation {energy loss) is due to:
7
PH I LIPPINI E SOCI ETY OF U LTRASOU T*D I I{ ST"I ftGERY
ln callaboration with
I NSTITUTE OF DIAGNOSTIC MTDICAL SONOGRAPI{Y I TICORPORATED
l") Abs*rption {conversion of acoustic energy to heat}
2) Retlectian
3) Scattering at interfaces
ln soft tissue, 8A% of the attenuatien of the sound wave is caused by absorption resulting in
heat production. Attenuation is rneasured in decibels per centimeter af tissue and is
represented by the attenuation coefficient af the specific tissr,re type {see table below}. A
reduction of 3 dB corresponds to dirninution of the original intensity by half. The higher the
attenuation coefficient, the mcre attenuated the ultrasound wave is by the specified tissue. For
example, bone with a very high attenuation c*efficient severely lirnits beam transmission.
The degree of attenuation also varles directly with the frequency of the ultrasound wave {see
figure below) and the" distance traveled. Generally speaking, a high frequenq wave is
associated with ttigh attenu{rficr: tfius ffmifing fiSsue penetratiar, whereas a low frequency
wave is asscciafed witlt lcw #ssue attenuatian and deep trbsre penetration.
x5
2 10
e .*"*MUSCLE
F
s
5
z
t- -LlvER
F
0
-BLOOD
Z 4 6810
FREQUENCY
To compensate for attenuation, it is possible to amplify the signal intensity of the returning
echo. The degree of receiver amplification is called the Gain. lncreasing the gain will amplifu
only the returning signal {echo} and not the transmit signal" An increase in the averall gain will
I
PHILIPPINE SOCIETY OT ULTRASOUND IN SURGERY
ln collaboration rr*ith
I NSTITUTE OF DIAGNCISTIC M EDICAL SONOGRAPI-IY I NCORPORATED
a
increase hrightness of the entire image, including the background noise. Preferably, the time
gain cempensatien tfCCI is adjusted to setectiveiy arnplify tlre weaker signals returning from
deeper structures.
Attenuation also results from reflection and scattering of the ultrasound urave. The extent of
reflection is determined by the differance in acoustic impedances of the two tissues at the
interface {i"e., the degree of innpedance misrnatch}. Acoustic impedance is the resistance of a
tissue to the pas$age of ultrasound. The higher the degree of impedance rnismatch, the greater
the amount of reflection.
6
Body Tissu* Acoustic lmpedance {10 Rayls}
Air 0.0004
Lung 0.18
Fat 1.34
Liver 1.6s
Sloed 1.65
Kidney 1.63
Muscle t.7L
Bone 7.8
The degree of reflection is high for air because air has extremely low acoustic impedance
(0.0004) relative to other body tissues. Bone also produces a strong reflection because its
acoustic impedance is extremely high {7.8} relative to other body tissues. For this reason, it is
clinically irnportant to apply sufffcient conducting gel (an acoustic couptring nnediurn] on the
transducer surfuce to elirninate any air pcckets between the transducer and skin surface.
i-{{
Otherwise much of tlre ultrasound waves will be reflected limiting tissue penetration.
/@ e .:i,
Specular reflection [figures {a} & {b}] occurs at flat, srnooth interfaces where the transmitted
wave is reflected in a single direction depending on the angle of incidence. Examples of
specular reflectors are fasciatr sheaths, the diaphragm and walls of major vessels. Block needles
I
PHIUPPINE SOCIETY OT ULTRASCIUND IN SURGERY
ln collaboration with
I NSTITUTE OF DIAG NOSTIC MEDICAL SONOGRAPHY I I{CORPORATED
are al$o strsng specutrar reflectors. For specular reflection to occur, the wavelength of the
ultrasound wave must be smaller than the reflective strueture.
The angle of the inridance is a major deternrinant of reflection. An ultrasound wave hitting a
smooth mirror like interface {sp*cular reflector} at a 90 degree angle will result in a
perpendicular reflection. An incident wave h;tting the interface at an angtre less than 90 d*grees
will result in the wave being deflected away frann the transducer at an angle equalto the angle
of incidence but in the opposite direction {angle of reflection}. When this happens, the signal of
the returning echo is weakened and a darker innage is displayed. This explains why it is difficult
to visualize a needle inserted at a steep angle {> 45 degrees to the skin surface}.
Reflection in biological tissues is not always specular. Scattering [figures {c} & {d}] or diffuse
reflection occurs when the incident ura\re encounters an interface that is not perfectly smooth
{e.gi, surface of visceral organs}. Echoes frorn diffuse reflectsrs are generally weaker than those
returning from specular reflectors" Scattering [flgure te]] atso oecurs when the wavelength of
the ultrasound wave is larger tlran the dirrrensions of the reflective structure {e.9., red blood
cells). The reflected echo scatters in many different directions resulting in echoes of similar
weak amplitudes. Ultrasonic scaf,tering gives rise to much of the diagnostic information we
observe in rnedical ultrasound imaging.
Tissue Eehogenicity
When an echo returrrs to the transducer, its arnplitude is represented by the degree of
brightness {i.e. echoge*icity} of a dot on the display. Connbination of all the dots fsrms the final
image. Strong specular reflectisns give rise to bright dots {hyperechoic} e.g., diaphragm,
10
PHILIPPINE SOCIETYOF ULTRASOUND IN SURGERY
ln collaboration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPHY I NCSRPORATED
gallston€, bone, pericardium. Weaker diffuse reflections produce grey dots {hypoechoic} e.g.,
sclid organs. No reflection produces dark dots {a*echoic} e.g., fluid and blaod filled structures
hecause the beam passes easily through these structures without significant reflectlon. Also,
deep structures often appear hypoechoic because attenuation limits beam transmission to
reach the structures, resulting in a weak returning echo"
lmage Resolution
Resolution refers to the ability to discrimlnate between two objects (reflectors or scatterers)
that are close together in space (spatial resoluticn! as separate or two events that occur close
together in tirne {terrporal resolution}.
Axial or Longitudinal resclutian refers to the ability to distinguish two structures that lie along
the axis {i.e. paralfel] of the ultrtsound beam as separate and distinct" Axial resolution is
determined by the pulse length. A high frequency wave with a short pulse length will yield
better axial resolutian than e low frequency wave.
ln the figure A below, a 5 Ml'lz transducer generates ultrasound waves that travel 0.3 mm per
cycle {wavelength = 0.3 mm = speed of sound ,/ frequenry = 1,540 mfsec divided by 5 x 106
cycleslsec). The pulse length is the distance traveled by one echo {3 cycles in this case}. As seen
in figure & a 5 MHz transdueer {wavelength = 0.3 mm and pulse length = 0.9 mm; 3 cycles), the
axial resolution is sufficient to distinguish the 2 target objects as separate beeause the incident
wave hits target # 1 {brown} before hitting target # 2 {gre€n}.
ln figure B below, with a 2.5 MHz transducer {wave}ength = 0.6 mrn and pulse length 1.8 mm;
=
3 cycles), the axial resolution is no longer adequate. Because bcth target ff t(brown) and target
# 2 (green) are hit by the same wave, both target objects are seen as one.
1L
PHILIPPI$E SOCIETY OF ULTRASOUND IN SURGERY
ln collahoration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPHY I NCORPORATED
0 A.2 0.4 0.6 0.8 1.0 L.2 t.4 1.5 mm
0.6 mm wavelength
B
1.8 rnm pulse length
Lateral resolutisn refers to resolution of objects lying side by side {i.e., perpendicular to the
beam axis). Lateral resolution is directly related to the transducer beam width, which in turn is
inversely related to the ultrasound frequency. A hish frequency transducer emits a wave with a
short wavelength and a small beam width. Lateral resoluticn Is pcor when the 2 structures lying
side by side are located within the sarne beam width. Because the returning echoes overlap
with each other side by side, the 2 structures {t and 2 in tigure} will appear as one on the
display. lt is therefore clinically important to chnose the highest frequency transducer possible
tc keep the beam width as narrow as possible in order to prcvide the best possible lateral
resolution. Lateral resslution is also affected by gain settings - the higher the gain, the worse
the lateral resolution. Lateral resolution is typically around 1 mrn.
However, attenuaticn also increases with frequency thus one must strike a balance between
resolution and attenuation. The beam width can be further reduced by adjusting the focal zone
(FZ). Lateral resolution is the best at the FZ, where the beam is narrowest. lt is therefore
clinically usefu[ to facus the target structure within the focal zone to yield the best possible
lateral resolution. The beam is known to diverge {increased beam width} as it propagates deep
into the far field"
Temporatr resolution, or frame rate, is impcrtant in trying to distinguish events that occur close
together in time. Fname rate depends upon the time taken to eollect allthe data required to
create one irnage, which in turn depends upon the sector width and depth. M-mode irnaging
offers very high sampling rates, typically L800 times per second, because of the very na rrow
field of view {see above}. 2-D echo has a much slower frame rate, typically 20*30 frannes per
12
PHILIPPINE SOCIETY OF ULTNASOUND IN SURGERY
ln collaboratlon with
I N STITUTE O F D IAG NGSTIC M EDICAL SONOG RAPT+Y I }ICORPO RATE D
second, because cf the much greater amount of ultrasound dat* that must be collected to
create a single frarrre.
Color Doppler
Color Doppler is an instrurnent to characterize blacd flow" The Doppler Effect occurs when
there is a rnoving source {blood flcw of red blood cells, RBC} and a stationary listener
{ultrasound transducer}. There is an apparent change in the returning echoes due to the
relative motion between the sound source and the receiver. lf the source {RBC} is moving
towards the receiver {transdr.lcer}, the percelved frequency is I-IIGFIER {display in RED} and
when the source {RBC} is moving a},vay fronr the receiver, the perceived fnequency is LOWER
than the actual {display in BLUE}. lt is important to note that Color Dcppler detection of flow
and flow direction is worst when the transducer is perpendicular {90 degreesi to the vesseland
best when the transducer is parallel {0 degrees} to the blood flow.
Power Doppler is useful for differentiating vascular frcm non vascular structures. Power
Doppler is more sensitive than Color Doppler in flow detection but does nct indicate flow
direction.
The Transducer
Each piezoelectric crystal prcduces an ultraseund wave- The sumrnation of allwaves generated
by the piezoelectric crystals forrns the ultrasound beam. The transmitted sound beam has a
shape, a finite width and lenght- The sound is focused eitl"rer by the shape of the transducer, a
lens in front af the transducer, or a complex set of control pulses frorn the ultrasound scanner
machine (Seamformind. This focusing produces an arc-shaped sound wave from the face of
the transducer. The wave travels into the body and comes into focus at a desired depth"
Transducer
i
i tl
\ i!lt Near Zone {Fresnel} - the area between the transducer and the focus
il
;r
i
Focal Zone - the area where the diameter of the sound beam is at a minimum
il
,,lt
lii Far Zone {Fraunhofer} - the area extending beyond the focus
13
PHILIPPINE SOCIETY OT ULTRASOUND IN SUBGERY
ln cellaboratlon with
I NSTITUTE CIF DIAGNOSTIC MEDICAL SONOGRAPI*Y I NCORPORATED
Transducer characteristics, such as frequency and shape, deterrnine ultrasound image quality
{see lmage Resolution}. The transducer frequencies used f*r peripheral nerve blocks range from
3-L5 MHz Linear and curuilinear {or curved} transducers are most useful for nerve irnaging to
provide high resolution images. Sector phased array transducers are used for 2D-echo or
cardiac sonography.
Modern transducers are broad bandwidth transducers that are designed to generate more
than one frequency. Fcr example, a Linear 5-12 MHz transducer can generate waves ranging in
frequency frorn 5.L2 MHz. With broad bandwidth transducers, the operator can select the
examination frequency to match the target requirernent. The resonance freqt"rency is the one
frequency at which the piezoelectric transducer is most efficient in converting electrical energy
to acoustic energy and vice versa. The resonance frequency is deterrnined by the thickness of
the piezoelectric element.
o The diarneters of the strings are different. The smaller strings vibrate faster than the
larger strings, and create higher frequency ssund.
The same principal applies to the elements in a transducer.
I
The figurre below illustrates how the thickness of the crystal irnpacts the frequency of
the sound produced.
For superficial structr.rres {e.9. vascular, thyroid, breast, extremities}, it is ideal to use high
frequency transducers greater than or equal to 7 MHz. Transducers in the range of 10-15 MHz
are preferred but depth of penetration is often limited to 2-3 cm below the skin surface. For
visualization of deeper struct$res {e.9. in the abdornen; liver, spleen, gall bladder, urinary
hladder), it may be necessary to use a lower frequency transducer {less than or equal to 5 MHz)
because it offers ultrasound penetration of 10 cm or more belornr the skin surface.
However, the image resolution is often inferior to that cbtained with a higher frequency
transducer. Linear transducers less than or equal to 5 crn wide are available for high frequency
L4
PHIilPPINE SOCIETY OF I.'LTRASOUruD IN SI.'R6ERY
ln collaboration rpith
I NSTITUTE OF DIAG NOSTIC MIDICAI SONOGRAPHY NCORPORATID
I
transducers. SrnalNer transducers, i.e., transducer: with srnaller footprints are useful for
detailed scanning where the patient's anatorny prohibits the use cf h'ulkier transducers {e.g.,
the supraelavicular region where there is limited access). Curved transdueers are best suited for
scanning whenever a wide field of view ls required.
Frequency and lmage Resalution - lt is best to select the highest frequency transducer possible
for the required depth of penetration.
I rlns
A. The use of a Higher Frequency Transducer - A
\ 1 na:l ica 1o r higher frequency transducer (10-12 MHz)
Footprint
\ f; cnn ect rt provides the best image resolution for superficial
structures.
Types of Transdueers
I
I
I i
I
r
t
,a*
16
PHILIPPINE SOCIETY CIF ULTRASOUhID IN SURGERY
ln collaboration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPilY I NCORPORATED
Another type of etrectronic array transducer is the phased-array transducer. This transducer is
cor*monly used for scanning the heart {2* echo Sonographyi. ln order to generate 2-D
tomographic (sliee) images of the hea*, the ultrasound beam has to be scanned across a
section of the heart" The limited access to the heart afforded by spaces between the ribs and
lungs dictates that cardiac scanners are of the sectar scan type. The transducer is positioned on
the chest manually and held steady, but the ultrasound beam it generates sweeps rapidly to
and fro across a sector of an arc, creating a fan-shaped scan in the sarne way that a lighthouse
beam sweeps across the sea, illuminating abjects in its path. Tlris scanning rnotion is also similar
to that of a rnechanical transducer.
ln order to avoid blurring of the image by the heart's motion, at least 25 images per second are
required. The maximunr attainable irnage franre rate is primarily tha result of a trade-off
between the required image depth, which lirnits the nurnber of pulses transrnitted per second,
and the sector angle and image line density, but other factors such as the display rnode and
imaging processing power of the machine are now involved. lmage frarne rate shown on the
display screen may be as high as 150 s-L or as low as 6 s-1. lt can be improved by reducing
image depth and/or narrowing the scan angle.
Almost all commercial echo machines use "phased array'' technology to scan the ultrasound
beam. Using sophisticated equiprnent derived from that used to cut silicon for manufacturing
electronic "chips," a single piezoelectrie crystal is striced into as many as 255 very thin strips,
each connected individually to the electrle pulse generator, which activates them in a very rapid
and very precisely cantrolled sequence, as shown in the Fig. 1.
Figure 1
L7
PHILIPPINE SOCIETYOF ULTRASOUND IN SURGERY
in collabaration with
I NSTITUTE OF DIAG NOSTIC MEDICAL SONOG RAPFIY I NCSRPORATED
Steering the Phased-Array Transducer
All elements in a phased array transducer are fired for each scan line" The timing of the
excitation af the crystal elements deterrnines the direction of the bearn. A) When elements on
the right side of the transducer are fired first the bearn is directed to the left. B) When elements
on the left side of the transducer are fired first ttre beam is dlrected to the right" C) When fired
simultaneously the beam is directed straight ahead. D) Minor variations in the timing of crystal
excitatisn focus the bearn.
!!.!
tA) lrt
{B} il
tc} {D}
-{ ++,+4.+ ili;l ii
I i
I-J.
LIU i.jtj
illttt:
J- ,l
iiii
iiji ll
$$F++t lll:l
il
llrii
ii
il
i1
I1:ll .tt
itt
itli
irlll :i
ffiit** rllll
illtl
fril*ilil
tl
A+, i1
\"Jr&r\J\.J \.rrrj
\*.H^*H_H_KJ
ff #
k &
The user must choose a fransd*mr which has the highest frequency allowable
for the penetration required,
Scan all objects of interest in two planes 90 degrees to each ather {i.e., scan
each structure in its long and short axisi.
High freguency transducers have greater abitityto resolye minute structure$, butthe userrs limited
by decreased depth af penetration.
Application:
'Lower frequency probes are used fcr abdominal scanning, third trirnester
pregnancy, and obese patients.
'Higher frequency proks are for superficial exams such as breast, thyroid, and
testicles.
19
PHILIPPINT SCICISTYOF ULTRASOUND ISI SURGERY
ln collaboratian with
I N STITUTE OF DIAGNOSTIC MEDICAL SONOGRAPHY I NCORPORATED
-3.5 or 5.0 MHz prabes are generally adequate for abdorninal, 1st and 2nd
trimester 0B studies
lVofe; lt is best to select the highest frequency transducer possible for the required
depth of penetration.
The DEPTH setting determines how far the ultrasound beam 'looks' into the patient and
is an important deterrninant of frame rate. The greater the depth setting, the Ionger the
transducer will have to wait for the ultrasound pulse to make lts round trip before
repeating the pulse, and so the lower the fi"anne rate. The depth setting should be
shosen so that the whole area of interest can be seen, but not so deeply that it includes
inelevant structures beyond the region of interest.
lLe.tI
at lA.
a2 a*4.
,E tr{
#
*8" .4.&
t.
alr0B
rl{t
gtc{
u f
ltle
.
t*vae rEEl
tlrt
rlt*i llr:.
ilaa3r rrl.t aro:
lsaar rtffi
*tmr
ltt,, ri**
Ftd I
X?rt lrt
fk
BL.
tr tr
\". {xt
S;
Too shallow Sptimal Too deep
Sector width or scan angle determines the tield of view across which the ultrasound
beam sweeps. As with depth, sector width is an irnportant determinant of frame rate
and should be optimized for each view to include the region of interest. With the scan
angle constant, increasing the depth of the image reduces the frame rate. With the
depth setting constant, increasing the scan angle will also reduce the frame rate.
20
PHITIPHNE SOCIETY OF ULTRASOUND IN SLIRGERY
ln csllaboration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SSNOGRAPHY I NCORPORATED
The GAIN function compens&tes for attenuaticn (a reduction in sound amplitude) as
sound travels deep into the body. The intensity of the returning signatrs can ba amplified
by the receiver upon arrival so that the displayed image is brighter and more visible on
the screen.
Gain can be adjusted for the whole image {overall gain} or for part of the image (see
tirne-gain con'lper'rsation {TGG} belovr}. While a high gain setting can he useful for
detectlng weaker signals that rnight otheruvise not be visible" it r*duces lateral resofrtion
and also increases noise. Excessive increase in GAIN will add'*r'loise" to the irnage.
TGC or Time Gain compensation is also known as depth gain compensation, and
corrects for the attenuation of the ultrasound signal that accurs with increasing distance
from the transducer. TGC boosts the gain of the signals returning from the far field to
ensure an even 'echo brightness' acrCIss the whole deptFr of the image, The TGC
conlrols can be fine-tuned by the sonographer using slider-bars.
Applicatia*:
Bane appea{s fo Se white an the ultrasound image Semuse if is fiyper reftective.
Blood or fluids appear to be black on the image because tfiey are anechaic.
Soff fiss$e appears as grey {satt and Wpper} an the image &ecairse ff is of medium
echagenicity.
General Rule:
Use plenty of gel ta remave the air interface befween the tran#ucer and the
skin. The physical properties af air do nof allaw passage of the ultrasaund heam even
thouglt the acoustic impedance has a low value.
ffi
\ I it'
Focus can be fine*tuned with phased-array transducers and should be
adjusted for each view so that the beam is focused on the region of
\iii interest.
lmage quality {lateral resolution} and beam focus is best at the focal
i
zone. Most moder"n electronically steered transducers provide electronic
i
focusing adjustable for depth. lt is important to place the Focus at or
slightly below the level of the target structure of interest.
Application:
Since mosf madern equipmenf rs focus variable, cfiange the posifion of the
focus as you scan different anatomicaf regfons in order ta maximize resalution.
21.
PHILIPPINE SOCIETY OF TJLTRASCIUNS IN SURGERY
ln collaboration with
I NSTITUTE OF DIAG hIOSTIC MTDICAL SONOGRAPHY I NICORPORATED
BYNAMIC RANGE
Grey scale compression (dynamic range) adjusts the number of shades of grey that
are displayed in the image. This allows the sonographer to choose the degree of
contrast in the image. More shades of gray resu[t in befier contrast resclution, or the
ability to visualize very subtle differences in tissue signatures.
lncreasing the dynarnic range increases the amount of gray scale displayed so that the
weaker signals are inciuded and the image is softened.
Frame Rate
One complete sweep CIr scan wi{l produce one ultrasound image CIr one frame. Frame
Rafe is the number of frarnes dlsplayed on the monitor per second.
The function of the ultrasound machine frame rate is similar to that of a video camera.
To capture a clear picture of maving subject, you'll need a higher frame rate. When
scanning moving organs,; i.e. the heart, you'll need a higher frame rate.
22
PHILIPPINT SOCITTY OF ULTRASOUND IN SURGERY
ln collabaration with
I N STITUTE OF DIAG NOSTIC MEDICAL SONOG RAPHY I NCORPORATED
lqaqe P"nocJrsqinq
Digital manipulation of the ultrasound image can irnprove both image quality and
the diagnostic capability.
The user can alter the appearance of the image to enhance sonographic
findings.
{a} Pre-processing
(b) Fost-processing
Once the echo information is stored within the computer, it may be altered to suit the
user's needs through manlpulation of the image processing controls.
Pre- and Post -processing, which were described earlier, affect the assignment of binary
numbers and grey shades to tha retuming echoes.
For general sonograms, to maintain a smooth image, have a medium contrast level,
with pre and post-processing eurves which allow even distribution of numbers and grey
shades among all tissues displayed.
1a
LJ
(b) Persistence
td) Measurement
Joy sticks or roller balls are used to move the calipers about a structure which is
to be measured.
The caliper cross hairs should be placed exactly on the edge of the structure for
greatest accuracy.
Calipers can also be u*ed to measure the distance from the transducer to a
target during amniocentesis ol'PUBS. This ensures acsur&te needle placement.
(el Magnilication
Modern ultrasound devices possess the abitity ts "blow up" the entire image, or a
pre-selected porticn of it for cl*ser examinatlon.
24
PHiLIPPINE SOCIETY OF ULTRASOUND IN SURGERY
ln eollaboration with
I NSTITUTE OF DIAG NOSTIC MEDICAL SS$O6RAPI{Y' NCORPORATED
h{agnitication may allaw yor.r to visualize a small structure, but it generally
affects resslution negatively. A magnification s:ntrol witl blow up the entire field
of view.
Zoom controls will allow you tc blow up a specific area of the image. You will be
able to select the area you wish to enlarge, and rnay scan in this enlarged forrnat.
Transducer Movement
It is best to gently rest the arm holding the transducer on the patienfs body or bed to
maintain hand steadiness and avoid fatigue during scanning.
1) Sliding; with the transducer in the upright pasition, slide the transducer longitudinally
2) Rotation; rotating the transducer {clockwise I counter-clockwise}
3) Tilting; tilting or angling the transducer
Alignment refers to the task of sliding the transducer longitudinally to follow the course
of the target e.9., a nerve or a needle. This first transducer movernent is also most
useful in locating the block needle.
When the transducer {ultrasound beam} and the needle are perpendicular to each other
(out of plane approach), it is irnpcrtant to slide the transducer along the shaft of the
needle (alignmenti to identify the needle tip. Both the needle tip and shaft in cross
section appear as a hyper-echsic white dot on the screen.
ln-plane Cut-af-piane
:
:::?j1,. :)ii;4.:::;:
:*
Transducer and Needle Alignment - Rctating the transducer is most useful for aligning
the ultrasound beam with the needle {in plane approach} so that the enilre length of the
needle (both shaft and tip) can be clearly seen.
25
PHILIPPINE SOCIETY OF ULTRASOUfiID IN SURGERY
ln collaboration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAFHY I NCORPOfiATED
Transdulcer, Needle and Nerve Alignrnent - Rotating the transducer is also required to
accurately align the transducer {u}trasound beami. the needle and the nerve al} in one
plane. This allaws visualization of hoth the ner*r{e and needle in the long axis.
Tilting the transducer can improve irnage quality by aligning the ultrasound beam
perpendicular to the target (nerve or needle). The angle of the incidence is also a
majcr determinant of reflection. An ultrasound wave hitting a smooth mirror like interface
at a 90 degree angle will result in a perpendicular reflection. An incident wave hitting the
interlace at an angle less thar: 90 degrees will result in the wave being deflected away
from the transducer at an angle equal to the angle of incidence but in the appesite
direction iangle of reflection). Vdhen this happens. the signal of the returning eche is
weakened and a darl<er image is displayed" This explains why it is difficult to visualize a
needle inserted at a steep arrgle {t 45 degrees to the skin surface}.
mffi
26
PHILIPPINE SOCIETY OF ULTRASOI"'ND IN SURGERY
ln callaboration with
I NSTITUTE OF DIAGNOSTIC M EDICAN- SONOGRAPHY INCORPORATTD
Freparing the ultrasound scanner
r The ultrasound canner should be turned on and allowed to warm up for at least 5
minutes before scanning commenses.
. Any image recording devices which will be used should also be ttirned on at this
time.
. Transducers should be cleansed and placed within easy reach before starting the
exam.
. Power settings and techniques should be checked to insure adequate imaging.
Transducer selection
Depending upon the patient's body size, weight and habitus, the proper transducer must
be Selected. Different applications reqr*ire different types of transducers so select the
proper transducer for the required examination.
For example, a 3.5 MHz or 5.S MHz sonvex transducer with a variable focus is
adequate in most abdominal scans while a higher frequency linear transducer is needed
for vascular, breast and other superficial scans.
After selecting a transducer, perform a quick initial scan to evaluate image orientation,
depth and penetration capabilities, and processing needs.
Image polarity
The irnage should be presented as a series of white dots placed on a black background.
Polarity can be reversed {black image on white background), but the standard is white
on black.
Black areas on the screen are hypaeehoic regions such as ftuid;
White regrbns are hyperechoic structures such as bone.
Scanning Planes
27
PHILIPPINE SOCITTY OF ULTRASOUND IN SURCERY
ln collaboration with
1 NSTITUTE OF DIAC NOSTIC MESICAL SONOGRAPHY I NCORPORATED
ta) Longitudinal plane
The longitudinal or sagittal scanning plane is that which runs frorn the head to the
foot and would divide the patient into right and ieft halves
The coronal plane divides the patient i*to anterior and posterior halves. This
plane is en'lplayed during evaluation of the adult kidneys sr other laterally placed
structures within the body.
The abave-mentianed terms are importafifu understand, but wfien yau are actually
scanning ff is mosf irnpofta*t that the scanning pla*e yau chaa$e, or variati*n cf it,
a//ows the beam ta reach fhe fargef at a perpendicular incidence" I*rs fl?eans that yaur
emplayed plane will continually change thraughout the exarn, depeMing upan the
posifion af the target argarr. flhfs is further complicated by the fact that yaur target rnay
bea
lmage Orientation
The sonographic image must be properly oriented in order to obtain correct anatomic
information. You must determine if you are correctly oriented before starting to scan.
iti ilr,1i.l'
I
I
I
. i.l:l! li
"riF* "--"?
posterior
I
ffi
?8
PHILIPPINE SOCIE TY OF UTTRASOIjND IN SURGERY
ln collaboration with
I NSTITUTE OF DIAGNOSTIC M EDICAL SONOGRAPHY I NCORPORATED
When scanning in a lonqiludlnal plane:
' iili ii
{* t' ,
i --{.
ieft
I
posteriQf
lf you are scanning in the transverse plane, and you angle the transducer toward the
patient's right hand side, information pertaining to the right should appear on the left
side of the screen.
29
PHILIPPINE SOCIETY OF ULTRASOUND IN SURGERY
In collaboration with
I NSTITUTE OF DIAGNOSTIC MESICAL SONO6 RAPHY I NCORPORATED
rEsr rg A$sFs$ tMAcE QUALiTY
You want your examination to offer the mast diagnostic information possible. Therefore,
you should adjust and set ali image parameters before starting scan.
After choosing a transducer and any preset technique settings, and after your have
confirmed that you are scannlng in the right direction (carrect irnage orientation], yot-r
are ready to perform a quick test to asse$s the image quality.
take a quick image of something in your field of view. FREEZE this image.
Iook carefully at the dlsptrayed image. ,{re there:
Using TGC (time gain conrpensationi gain pods or dials, correct areas noted above
Once entire image, top to bottom, is uniform then assess the overafi image quality:
lJsing the OVERALL gain control (pod or dial), correct image gain
Safetv of ultrasound
Ultrasound involves the delivery cf externai energy to body tissues and so it is rmportant
to ccnsider the potentiai adver"se biclogical effects ihat this could entaii. The irrtensity of
exposure to uitrassund is expressed a* power per ilnit of area (wattsl*mZ) expressed
as the rnaximum intensity within the ultrasound beam (the spatial pear) averaged over
the duration of *xposLire {ternporal average), the spatlal peak temporal average
{SPTA). There are two main biologieal effects af exposure to ultrasound energy: thermal
{heating) and mechanical {e,g" cavitation).
Thermal effects are caused by conversion of the mechanical energy of the ultrasound
into heat energy as it passes though the tissues. The amount of heating Is hard to
predict but relates tc several factors including transducer frequency, transmit power,
focus and depth. Heat may be generated not just by the ultrasound but also directly by
the probe itself. lt is prudent to keep imaging time to a rninimum and to ensure that the
probe is repasitioned regularly, and to manitor the temperature of the prabe.
30
PHILIPPINE SOCIETY OF ULTRASOUND IN SURGTRY
ln collaboration with
I NSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPFTY INCORPORATED
Mechanica! affeets include cavitation, in which gas bubbles sre *reated as ultrasound
pas$es through the tissues. lt is net thought tc he a problem during standard trans-
thoracic studies, but is important when bubble contrast agents are used as it can cause
resonance and even disruption of the bubbles. Mechanical effects of riltrassund can
also be measured hy rnechanical index {Ml), which is the peak negative {rarefactional)
presslire divlded by the square root of the transducer frequency. An Ml of <1 is
considered safe.
At much higher po\Mer levels, heat formation and cavitation have been demonstrated
Prudent application of diagnostic ultrasound shot-ild follow the "ALARA" (As Low As
Reasonably Achievable) principle for pCIwer settings
ln the low rnegahertz frequency range there have been no independently confirmed
significant biological effects in marnmalian tissues exposed to intensities" below 100
MW/cm'. Furtherrnore, for ultrasonic exposure times* less than 50CI seconds and
greater than 1 second, such effects have not been dernonstrated even at higher
when the product of intensity* and exposure time*" is less than SCI
oul
Diagnostic Ultrasound has been in use for over 25 years. Given its known benefits and
recognized efficacy for medical diagnosis, including its use during human pregnancy,
the American lnstitute of Ultrasound in Medicine herein addresses the clinical safety of
such use.
Although the possibility exists that such biological effects may be identified in the future,
cunent data indicate that the benefits to patients of the prudent use of diagnostic
ultrasound outweigh the risks, if any, to the patient.
o [Jse power settings which atre "as low as reasonably achievable" {ALARA} yet still
provide diagnostic nfonnation.
i
a Clean transducers with damp cloth before and after each examination.
o Perform quick reference scan before beginning the examination to assure proper
image orientation and transducer selection.
Routine Maintenance
Froper care and maintenance af ultrasound equipment ensures safety and pr^olongs
equipment life and reduces mashine downtime. tselaw are general procedures in srder
tc keep scanning quality at a maxirnum. ALWAyS REFER TG TftE ltlACl{/rVE'S USEE
fvIANUAL for user maintenance procedures"
32
PHILIPPINE SOCIETY SF ULTRASOUND IN SURGERY
ln collaboration with
INSTITUTE OF DIAGNOSTIC MEDICAL SONOGRAPHY INCORPORATED
Ctrean and store ultrasound transducers when not in use.
Clean probe with suitable solution {e.g. cutasept} after each patient.
Clean the unit with a damp cloth at the end of the day.
.Do not hang or drape transducers around the device" Flace transducers i":pright in
specially designed transducer holders. USE THE Df$IGNATED HANGHRS FOR THE
TRAN$DUCER CABLES,
.Do not allow transducer to ssak in Cidex for more than one hour
{Cidex or bleach will
destroy the aeoustic rnernbrane on the transducer face). Always follow the length ot
submerging time specified in the n:anufacturer's instructions. The solution must satisfy
the-CDC {Center for Disease Controt) "SterilanUDisinfectant' requirernents.
'Never gas or autoclave a transducer, these processes will destroy the transducer
crystals.
'Use acoustic coupling gel specified by the manufacturer of the scanner. Do not use
mineral oil.
Ultrasound waves do not pass through air; therefore, an evaluation of the stomach,
small intestine and large intestine rnay be limlted. lntestinal gas may also prevent
visualizaticn of deeper structures such as the pancreas and aorta.
Patients suffering from obesity are more diflicult to image * this is because fat causes
considerable refractian and image distorticn.
-0-
34
PHIUPPINE SOCIETYOF ULTRASOUND IN SURGERV
ln collaboration rarith
I NSTITUTE OF DIAGNOSTIC M EDICAL SONOGRAPHY I NCORPORATE D