t205 Data Sheet 2-7-11
t205 Data Sheet 2-7-11
resulting in extended battery life. The pump’s low power, small size, and
light weight play a critical role in portable gas detection and medical
applications. The T2-05 pump HE and LI pump models enable intrinsic
safety capabilities for sampling of hazardous gases, typical of industrial
and mining operations. The T2-05 IC pump is designed for compact and
wearable medical devices that require less than 1500 hours of pump
life. The T2-05 VBIC model is ideal for vacuum only medical applications
that benefit from improved sound quality, such as wound therapy.
Typical Markets
Features
• Safety
• Patient Therapy • The valve design has been optimized to provide the highest flow rates
• Patient Monitoring available with the lowest current draw, allowing for longer battery life
and smaller instrument size.
Typical Applications • The T2-05 model pump life ranges up to 10,000 rated hours
• Portable Gas Detection depending on motor (HE, LI and IC) options
• Gas Sampling
• The pump fits into the extremely tight spaces demanded of today’s
• Medical Instruments
handheld instruments, such as portable gas detectors and portable
• Trace Detection
negative pressure wound therapy devices for patient mobility. The
• Sidestream CO2
lightweight design minimizes instrument weight.
• Negative Pressure Wound Therapy
• RoHS compliant.
Product Specifications*
Physical Properties Electrical Pneumatic
Operating Environment : 1
Motor Type (DC): Head Configuration: Single
-4 to 122°F (-20 to 50°C) High Efficiency Coreless Brush (HE) Maximum Flow:
Storage Environment: Low Inductance Coreless Brush (LI) HE, LI: 800 smlpm, IC: 700 smlpm
-4 to 122°F (-20 to 50°C) PMDC Iron Core Brush (IC)
Maximum Intermittent Pressure7:
Media: Nominal Motor Voltages (DC)4 :
6.2 psi (430 mbar)
Air, Argon, Helium, Nitrogen, Oxygen, 3.3 VDC
Maximum Continuous Pressure:
and other non-reacting gases Max Power at Nominal Voltage:
2.0 psi (138 mbar)
Humidity: 0.36 Watts
Maximum Intermittent Vacuum7:
Most non-condensing gases Electrical Termination:
5-95% Relative Humidity 10.8 in Hg (274 mm Hg)
HE: Wire Leads
Noise Level2: Maximum Continuous Vacuum:
LI: Wire Leads
As low as 45dB IC: Solder Tabs or Wire Leads 4.1 in Hg (104 mm Hg)
Pump Assembly Rated Life : 3 Current Range : 5 Filtration:
Coreless Motor-Pump (HE): 34 - 105 mA 40 micron recommended
Up to 10,000 hours Inductance6: Efficiency at Free Flow8:
Coreless Motor-Pump (LI):
Up to 6,000 hours HE: 0.28 mH maximum @ 1kHz/50mV LPM/Watt: 4.66 @ 1.9 VDC
PMDC Iron Core-Pump (IC): LI: 0.05 mH maximum @ 1kHz/50mV (P/N T5-1HE-03-1EEB)
Up to 1,500 hours IC: 4.07 mH maximum @ 1kHz/50 mV LPM/Watt: 4.08 @ 1.9 VDC
(P/N T5-1LI-03-1EEB)
Weight: LPM/Watt: 3.12 @1.9 VDC
0.5 oz (14 g) HE and LI (P/N T5-1IC-03-1EEP)
0.4 oz (11 g) IC
Wetted Materials
Diaphragm: EPDM Valves: EPDM Pump Head: ABS (HE, LI), PBT (IC)
* See Appendix A for details.
3
Miniature Pumps
254 203 152 102 51 0 0.07 0.14 0.21 0.28 0.35
900 0.50
800 0.45
intermittent duty 0.40
700
intermittent duty 0.35
0.30
500 continuous duty
0.25
400
0.20
300
0.15
200 0.10
100 0.05
0 0.00
10 8 6 4 2 0 1 2 3 4 5
Vacuum (in Hg) ‐ Pressure (psig)
Max Flow‐4.5V Min Flow‐1.9V
T2-05 PMDC
T2‐05 Pump Iron Iron
‐ PMDC CoreCore
Brush Motor
Motor
Vacuum (mm Hg) ‐ Pressure (bar)
400 300 200 100 0 0.13 0.27 0.40 0.35
800 0.60
700
0.50
600
0.40
500
FLOW (ml/m)
POWER (W)
intermittent duty intermittent duty
400 0.30
300
0.20
200
0.10
100 continuous
duty
0 0.00
‐500 16
‐400 12
‐300 8
‐200 4
‐100 00 2
100 4
200 6
300 8
400 500
Vacuum (In Hg) ‐ Prssure (psig)
The above graphs represent examples of performance for the pumps series handling air at 800 feet (244M) above sea level at 75° F (24° C). Performance
will vary depending on barometric pressure and media temperature. Curves are representative of standard pump configurations. Pump configurations
could be customized for higher or lower flows, depending on specific customer requirements.
Please contact Parker Precision Fluidics Applications Engineering for other considerations
4
HE LI IC
Inductance6 Better Best N/A
Efficiency at Best Best Better
Free Flow8
Life3 Best Better Good
Cost Good Better Best
1.26 [32.0]
Coreless Brush/HE Version .12 [3.0]
.08 [2.0]
.62 [15.7]
Ø.47 [Ø12.0]
5
Miniature Pumps
.53 [13.5]
.08 [2.0]
.62 [15.7]
Ø.47 [Ø12.0]
Units
IN. (mm.)
UNITS
1/8" HOSE BARB SUCTION IN. [mm.]
C
L .16 [ 4.0 mm]
Units
IN. [mm]
Units
Ø.47 [12.00 mm] IN. (mm.)
NEGATIVE TERMINAL (-) .39 [10.00 mm]
ACROSS
FLATS
Lead Wire
.24 [6.10 mm]
.55 [13.89 mm]
C
L
.12 [3.0 mm] .08 [1.91 mm]
Units
IN. (mm.)
Units
IN. [mm]
.71 [18.1]
.73 [18.5]
.65 [16.5] WHITE MARK ON MOTOR
CL INDICATES POSITIVE TERMINAL (+)
.12 [3.1]
.11 [2.7] 1.52 [38.7] .05 [1.35] MAX
CL
Ø.47 [Ø12.0]
.37 [9.4]
Ø.16 [Ø4.0]
Miniature Pumps
T2-03 PUMP DIAGRAM
FILTER
IN FROM
ATMOSPHERE
T2-05
PUMP
CO2 SENSOR
X-VALVE
IN FROM
PATIENT
FILTER
APPLICATION T2-05
SPECIFIC PUMP
RESTRICTOR
8
Path Materials
1. EXCELLENT
Chemical EPDM ABS PBT Minimal or no effect
Air 1 1 1 2. GOOD
Ozone (1000 ppm) 4 2 1 Possible swelling and/or
loss of physical properties
Oxygen 1 1 1
Ethylene (Ethene) 4 1 1 3. DOUBTFUL
Moderate or severe swell-
Acetylene 1 2 2
ing and loss of physical
Propane 4 2 2 properties
Methane 4 4 2 4. NOT RECOMMENDED
Nitrogen 1 1 1 Severe effect and should
Carbon Dioxide 2 2 1 not be considered
Halothane (Up to 5%) 4 1 1 Note: Consult factory for
other gases.
*The above is an Abbreviated Chemical Compatibility Chart. Please
consult factory for details.
Ordering Information
T2-05 Micro Pumps
Vacuum: Free Pressure:
Configuration Max PCD* Wetted Materials
LPM @ Load Flow LPM @ Load
12 8 4 2 4 6
in Hg in Hg in Hg psig psig psig Vac Press Motor
0 VDC mA Diaphragm, Valves, Gasket
in Hg psig Type
305 203 102 134 276 414
mm Hg mm Hg mm Hg mbar mbar mbar
Coreless
T5-1HE-03-1EEB 0.2 0.5 0.8 0.6 0.3 10.8 6.2 4.5 438 EPDM
Brush
Coreless
T5-1LI-03-1EEB-1 0.2 0.5 0.8 0.6 0.3 10.8 6.2 4.5 438 EPDM
Brush
T5-1IC-03-1EEP 0.2 0.5 0.7 0.5 0.3 10.0 6.2 PMDC 4.5 240 EPDM
T5-1ICW-03-1EEP 0.2 0.5 0.7 0.5 0.3 10.0 6.2 PMDC 4.5 240 EPDM
T2-05-LI and HE model Sound Improvement options are also available, contact applications for more details.
Applications Engineering can assist with sound quality evaluations given the complex relationship between system
components and operational requirements of the customer’s specific application.
9
Miniature Pumps
T2-05 Micro Diaphragm Pump.
Serviceable – PPF products are designed for use through the rated life and Parker does not sell replacement parts,
nor is it recommended to service these in the field
Note: In addition to Parker’s innovative and flexible pump designs, we offer applications engineering expertise
to our customers in order to configure and recommend the optimal pump for the application. Contact Parker
Applications Engineering to discuss and configure alternate pump configurations to meet your specific
application requirements. Providing information on the following requirements will assist us in developing an
optimal solution for your application:
• Noise • Size ORDER
• Operating Pressure / Vacuum • Motor Control ON-LINE
• Power Consumption • Media
• Life Requirement • Voltage
Appendix A
All performance data is typical based on standard conditions: 70°F and 14.7 psia (21°C and 1 bar).
1. Duty Dependent. For operation above 122°F (50°C) consult factory
2. Noise is dependent on the configuration and operation of the pump in the application. Parker has the ability to tailor
the pump configuration when noise is a critical criterion in the effort to meet the performance requirements of the
application. Noise level is tested to Parker protocol P-105.
3. Life rating can vary depending on application and operating conditions.
4. Custom motor options available. Custom motors may require a significant application potential. The standard
motors can be configured with a special winding to meet a particular operation point at a specified voltage
5. Current range is dependent on motor type, voltage, pressure/vacuum and flow requirement. Lower levels possible
depending on application.
6. Inductance is an indicator of induced voltage with change in current and it is a key parameter to enable customers’
low energy intrinsic safety systems
7. Maximum intermittent pressure/vacuum data is a pump capability guideline for applications that go beyond the
maximum continuous levels for short periods of time. Please consult customer specific requirements with the factory
or Applications Engineering.
8. Pump efficiency is a measure of the flow rate generated per unit of power consumed. Efficiency may change
dependent on application and operating condition at free flow.