Spec Tender HPLC Flexar

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 3

HPLC System

HPLC manual system with manual injection and UV-Vis Detector


Have Integrated LC tube management for streamlined chromatography
Have Hidden plug-and-play drainage automatically interconnects with components added to
system
Have solvent manager with integrated degasser and software communication link
Pump Specification
Flow-rate range : 0.01 to 10.0 mL/min

Flow-rate increments : 0.01 from 0 to 0.99 mL/min

0.1 from 1.0 to 10 mL/min

Flow precision : 0.3% RSD (typical 0.1%) at

1 mL/min water @ 1000 psi

Flow accuracy : ±1% of setting at 1 mL/min

and 1000 psi with water

Retention-time reproducibility : <0.3% RSD (typically <0.1%)

Pressure range : 0 to 6100 psi for the entire flow-rate range

Compositional range : 0 to 100%, solvent A to B

Compositional accuracy : Typically, 0.5% from 3 to 97% up to 5 mL/min

Compositional precision : Typically, <0.2% variation

Solvent blending increments : Settable to 0.1% increments

Solvent proportioning valves :2

Gradient forms Linear : exponential 1 to 9.9(positive and negative) or step (189 profile
selections)

Linearity : Within 1% from 10 to 90%

Step time : 0 to 999 minutes

Step-time increments : 0.1 min from 0 to 9.9 min

1.0 min from 10 to 999 min

Timed events : Two built-in independent momentary (2-second)contact closures

Events / method : Max of 9

Event increments : 0.1 min from 0 to 9.9 min;

1.0 min from 10 to 999 min


Dimensions (HxWxD) : 16 x 34 x 53 cm

Weight : 27 kg

Service Manager : software Pump setup, calibration and troubleshooting using


dedicated user interface

UV-Vis Detector Specification

Wavelength range : 190-700 nm


Wavelength accuracy : ±1 nm
Wavelength precision : ± 1 nm
Bandwidth : 5 nm
Linearity : ≥ 2.5 AU (with 5% deviation)
Optics : Dual beam
Maximum sampling rate : 50 pt/sec
Sensitivity range : 0.0005 to 3.000 AUFS in 0.0001 increments from
0.0005 to 0.1, and 0.01 increments above 0.1 AUFS
Noise : <7.5 x 10-6 AU, 210-280 nm, 2 sec response time,
standard test cell
Drift : <1 x 10-4 AU/hr, after warmup
Light sources : Deuterium (190-360 nm) or Tungsten
(360-700 nm), pre-focused, no adjustment
required on replacement
Flow cell : Flow cell sold separately; compatible
with PerkinElmer flow cells: 6 mm x 2.4 Μl
8 mm x 12 μL, 10 mm x 15 μL and 2 mm x 1.7 μL (prep)
Materials contacting solvent : Kel-F, PEEK, quartz
Leak detection : Heated thermistor sensor in glass envelope,
located in flow cell drip pan
Dimensions : (HxWxD)16 x 34 x 53 cm (6.5 x 13.5 x 21 in)
Weight : 11.4 kg (25.2 lbs)