HPLC manual system with manual injection and UV-Vis Detector
Have Integrated LC tube management for streamlined chromatography Have Hidden plug-and-play drainage automatically interconnects with components added to system Have solvent manager with integrated degasser and software communication link Pump Specification Flow-rate range : 0.01 to 10.0 mL/min
Solvent blending increments : Settable to 0.1% increments
Solvent proportioning valves :2
Gradient forms Linear : exponential 1 to 9.9(positive and negative) or step (189 profile selections)
Linearity : Within 1% from 10 to 90%
Step time : 0 to 999 minutes
Step-time increments : 0.1 min from 0 to 9.9 min
1.0 min from 10 to 999 min
Timed events : Two built-in independent momentary (2-second)contact closures
Events / method : Max of 9
Event increments : 0.1 min from 0 to 9.9 min;
1.0 min from 10 to 999 min
Dimensions (HxWxD) : 16 x 34 x 53 cm
Weight : 27 kg
Service Manager : software Pump setup, calibration and troubleshooting using
dedicated user interface
UV-Vis Detector Specification
Wavelength range : 190-700 nm
Wavelength accuracy : ±1 nm Wavelength precision : ± 1 nm Bandwidth : 5 nm Linearity : ≥ 2.5 AU (with 5% deviation) Optics : Dual beam Maximum sampling rate : 50 pt/sec Sensitivity range : 0.0005 to 3.000 AUFS in 0.0001 increments from 0.0005 to 0.1, and 0.01 increments above 0.1 AUFS Noise : <7.5 x 10-6 AU, 210-280 nm, 2 sec response time, standard test cell Drift : <1 x 10-4 AU/hr, after warmup Light sources : Deuterium (190-360 nm) or Tungsten (360-700 nm), pre-focused, no adjustment required on replacement Flow cell : Flow cell sold separately; compatible with PerkinElmer flow cells: 6 mm x 2.4 Μl 8 mm x 12 μL, 10 mm x 15 μL and 2 mm x 1.7 μL (prep) Materials contacting solvent : Kel-F, PEEK, quartz Leak detection : Heated thermistor sensor in glass envelope, located in flow cell drip pan Dimensions : (HxWxD)16 x 34 x 53 cm (6.5 x 13.5 x 21 in) Weight : 11.4 kg (25.2 lbs)