AP18 Minimum Information For Reliance MD v4
AP18 Minimum Information For Reliance MD v4
AP18 Minimum Information For Reliance MD v4
One of the main objectives of SEARN is reliance. To address this need, the Assembly adopted on 27 July
2023 a Strategy to facilitate reliance. To address this need for medical devices, the Assembly of SEARN
adopted on 27 July 2023 as part of the 2023-2024 work plan an action point to ‘Facilitate reliance for
medical devices, including IVDs’.
This document was developed to support the implementation of this strategy. There would be 2 main
usages of this information:
Acknowledging that the implementation of these recommendations may require addressing practical
and technical challenges, and in some cases further discussions and agreement of other authorities, the
adoption by the Assembly of SEARN will be followed by an implementation period of two years.
Effective communication between reference authority, applicant and NRAs is also required to ensure
that the sameness of the product is maintained after approval. The sharing of some information may
require a Memorandum of Understanding and/or a Confidential Disclosure Agreement.
The information identified in the below table intends to present the minimum information required
for reliance. The below minimum information would generally be expected for higher risk medical
devices and would not usually be required for low-risk medical devices. For reliance, the ultimate
requirement is that the relying NRA should have sufficient trust in the reference authority to use the
output of their work in their own regulatory decision-making system. Abridged assessment may
require additional information.
The above information may also be accessed through requesting for the full
technical dossier, similar to the technical dossier that was submitted to the
reference regulatory authority.
Regulatory The following information should be accessible from the reference NRA
Inspections (publicly, or shared by the NRA, e.g. upon request):
References: